Fig. 5
PANA SPRAYPANA SPRAY
E-type Spray Nozzle
Handpiece
Cotton Swab
Glass Rod Optic End
Handpiece
Cautions on Lubrication
Cautions on Sterilization
Fig. 6
Fig. 4
Caution
4. Irrigation Nozzle (Ti95EX)
When irrigation is required please connect
the irrigation tube to the irrigation nozzle.
5. Lubrication
Supply PANA SPRAY after each use and/or before autoclaving.
q Push E-type spray nozzle attachment over the PANA SPRAY nozzle until it
firmly seats.
w Shake the can 3-4 times to well mix lubricant and propellant.
e Insert the E-type spray nozzle in the rear of the handpiece and spray for
approximately 2-3 seconds until the oil comes out of the handpiece head.
6. Cleaning and Sterilization
NSK recommends heat sterilization by autoclave.
Sterilization required after each patient as noted below.
Autoclave Procedure
q
Scrub dirt and debris from the handpiece, and wipe clean with alcohol-
immersed
cotton swab or cloth. Do not use a wire brush.
w Lubricate with PANA SPRAY. Refer to 5. Lubrication.
e Insert into an autoclave pouch. Seal the pouch.
r Autoclavable up to a max. 135˚C.
ex.) Autoclave for 20 min. at 121℃, or 15 min. at 132℃.
t Keep the handpiece in the autoclave pouch to keep it clean until you use it.
✻ Sterilization at 121˚C for more than 15 minutes is recommended by EN13060
or EN554.
✻ Performing sterilization according to our instruction has minimal effect on the
instruments. Life span is generally determined by wear and damage due to
use.
7. Cleaning Glass Rod Optic End (Ti85L, Ti85LA, Ti95L, Ti95LA)
In case the light had become dim, clean the glass rod optic end with an alcoholsoaked cotton swab. (Fig.6)
・Be sure to hold the handpiece firmly to prevent the handpiece from
slipping out by the spray pressure when lubricating.
・Supply lubricant until it comes out of the handpiece head (for approx.
2 seconds).
・Keep the PANA SPRAY can upright.
・Use of lubricants other than NSK’s PANA SPRAY can cause excessive
heat generation and possible handpiece damage.
・In case the sterilizer chamber temperature could exceed 135℃ during
the dry cycle. Skip the dry cycle.
・Always place the handpiece in the center or upper shelf of the
chamber, as the local temperature at the chamber bottom could rise
beyond the set value.
・Do not autoclave the handpiece, even when it is bagged, together with
other instruments, from which chemicals are not sufficiently cleaned.
This is to avoid discoloration and / or damage to the plating of the
handpiece.
・Do not heat or cool the handpiece quickly. Rapid change in tempera-
ture could break the glass rod give abnormal strain to other metals.
8. Repair Service
For genuine NSK service and parts always return the handpiece to your NSK dealer.
Irrigation Nozzle
Irrigation Tube
Do not use a pointed tool or sharp
edges to clean the glass rod optic
end. They may damage the mirror
finish and reduce the light transmission efficiency. If the glass rod
optic end has become damaged,
send the handpiece to the NSK
dealer for repair.
Powerful Partners
®
’07.06.02 M
※ Specifications may be changed without notice.
www.nsk-inc.com
NAKANISHI INC.
700 Shimohinata, Kanuma, Tochigi
322-8666, Japan
www.nsk-europe.de
NSK EUROPE GmbH
Westerbachstraße 58 60489 Frankfurt,
Germany
This handpiece can be washed via Thermo Disinfector.