NSK NAC-Y, BB-Y, NM-Y, NBBW-Y Operation Manual

Contra Angle Handpiece Latch Bur
NAC/BB/NM/NBBW
OM-C0363E
This operation manual for :
NAC - EC / ER4 / ER10 / ER16 / ER64 / ECM / ER4M / ER10M / ER16M / ER64M BB - EC / ER4 / ER10 / ER16 / ER64 / ECM / ER4M / ER10M / ER16M / ER64M NM - EC / ER4 / ER10 / ER16 / ER64 / ECM / ER4M / ER10M / ER16M / ER64M NBBW - EC / ER4 / ER10 / ER16 / ER64 / ECM / ER4M / ER10M / ER16M / ER64M
Indications for Use
The handpiece is designed only for clinical dental use by qualified personnel.
Caution
When operating the handpiece always consider the safety of the patient. Check for vibration, noise and overheating outside the patient's oral
cavity before use. If any abnormalities are found, stop using the Product immediately and contact your authorized dealer.
Do not disassemble or alter the handpiece except as recommended by
NSK in this Operation Manual.
Do not allow any impact on to the Product. Do not drop the handpiece.
Do not exceed the allowable max speed recommended on the Specification.
Be sure to connect or disconnect the handpiece only after the motor
completely stops rotation.
Prior to clinical use, inspect the handpiece. Check for vibration, noise
and overheating. If any abnormalities are found, stop using the handpiece immediately and contact your Authorized NSK Dealer. (Refer to "5. Checking the Handpiece Before Each Use")
Operators and all others in the area must wear eye protection and a
mask when operating this handpiece.
Perform periodical maintenance checks.
(Refer to "9. Periodical Maintenance Checks")
If the handpiece has not been used for a long period, check for noise,
vibration, and overheating before use.
003
1. Specifications
Model
Allowable Max. Speed
(motor)
Bur Type
Chuck Length
Max. Bur Length
Max. Working Part Diameter
Sheath
Gear Ratio
2.
Connecting and Disconnecting the Handpiece from the Motor
(1) Connecting
Insert the handpiece direct into the motor.
(2) Disconnecting
Hold the motor and the handpiece, then pull apart.
NAC-Y BB-Y NM-Y NBBW-Y
30,000 min-140,000 min-130,000 min-140,000 min
ISO 1797-1 (EN ISO 1797-1
13.2mm
22.5mm
4.0mm
φ
EC/ECM
1
ER4/ER4M
1
13.7mm
22.5mm
4.0mm
φ
ER10/ER10M
4 1
) Type1 Ø2.35mm CA Bur
10.6mm
22.5mm
4.0mm
φ
ER16/ER16M
10 1
16 64
1
12.7mm
22.5mm
φ 4.0mm
ER64/ER64M
1
Caution
・ Do not connect or disconnect the handpiece until the motor has
completely stopped.
・ Connect ONLY to E type motors (ISO 3964 (EN ISO 3964)).
Do not exceed the Max. Rotation Speed (Motor) shown in "1. Specifications".
3. Inserting and Removing the Bur
-1
Caution
・ Do not insert or remove the bur until the rotation hascompletely
stopped.
(1) Inserting
1) Swing the Latch Plate to open the chuck.
2) Insert the bur into the chuck until the bur "notch" mechanism engages.
3) Swing the Latch Plate back to closed position.
(2) Removing
Swing the Latch Plate to open position and pull the bur out.
Caution for bur
・ Do not use burs with problems listed below as the bur may break,
seize up or disengage from the chuck.
- Bent, deformed, worn, rusted, broken, deficient bur.
- Bur which is cracked on the edge or axis.
- Non-ISO(EN ISO) standard, or tampered bur.
・ Make sure that the bur is safely inserted before use. Incorrect
installation may cause the bur to come out during rotation and is very dangerous.
・
Do not exceed the bur speed recommended by the bur manufacturer.
4. Connecting the Irrigation Tube
NBBW Cut the Irrigation Tube (Option) to the desired length. Connect the Irrigation Tube to the irrigation nozzle. (Fig.1)
* Tube (RSM) : Order Code Z166101
Irrigation Tube
Irrigation Nozzle
Fig.1
NAC / BB / NM Attach the Spray Nozzle (Option) to the handpiece head.Cut the Irrigation Tube to the desired length then connect to the Spray Tube Nozzle on the motor. (Fig. 2)
* Spray Nozzle NAC,BB : Order Code C032701 NM : Order Code C039701
Spray Nozzle
Irrigation Tube
Fig.2
5. Checking the Handpiece Before Each Use
Follow the check procedure below before use. If any abnormalities are found, stop using the handpiece immediately and contact your Authorized NSK Dealer.
1) Check the union nut (Fig.5) on the head is firmly tightened. If loosened,
tighten the union nut.
2)
Check the end cap (Fig.6) is firmly tightened. If loosened, take the following actions.
NAC / BB / NBBWTighten the end cap with the provided cap wrench.
NMContact your dealer. (The cap wrench is not provided due to not being
able to change the cartridge.)
3) Insert the bur.
4) Rotate the handpiece for about one
minute at the Max. rotation speed stated
in "1. Specifications". During rotation,
check for abnormalities such as
abnormal rotation, vibration, noise.
5) After the handpiece rotation has completely stopped, touch the
handpiece head to confirm the head is NOT heating abnormally (Fig. 3).
Fig.3
Caution
・ Do not use the handpiece when the end cap is loose.
Using the handpiece under such condition will cause looseness of the internal parts in the handpiece head, which may cause the end cap to come off and patient’s accidental ingestion.
・ To avoid injury, keep your hands away from the bur during rotation.
6. Lubrication
PANA SPRAY Plus / PANA SPRAY
Supply “PANA SPRAY Plus /PANA SPRAY” after each use and/or before
autoclaving.
1) Push E-type Spray Nozzle attachment over the lubricant spray nozzle until it firmly seats.
2) Remove the bur.
3) Insert the E-Type Spray Nozzle in the rear of the handpiece and spray for approximately 2 – 3 seconds until the oil comes out of the handpiece head.
E-type Spray Nozzle
Handpiece
PANA SPRAY Plus/PANA SPRAYPANA SPRAY Plus/PANA SPRAY
Caution
Be sure to hold the handpiece firmly to prevent the handpiece from
slipping out of hand by the spray pressure when lubricating.
Supply lubricant until it comes out of the handpiece head
(for approx. 2 – 3 seconds).
Hold the Lubricate spray can upright.
7. Sterilization
The products listed below are washable with Thermo Disinfector. ECM, ER4M, ER10M, ER16M, ER64M
Steam autoclave is recommended. Sterilization is required first time you use and after each patient.
Autoclave Procedure :
1) Scrub dirt and debris from the handpiece, and wipe clean with alcohol-immersed cotton swab or cloth. Do not use a wire brush.
2) Lubricate the handpiece. Refer to 6. Lubrication.
3) Insert into an autoclave pouch. Seal the pouch.
4) Autoclavable up to max. 135°C. ex.) Autoclave for 20 min. at 121°C or 15 min. at 132°C.
5) Keep the handpiece in the autoclave pouch to keep it clean until you
use it.
* NSK recommends Class B sterilizers as stated in EN 13060.
Fig.4
Caution
・ Do not wipe nor immerse the handpiece in acidic water or acidic
solutions.
Keep the product in suitable atmospheric pressure,temperature,
humidity, ventilation, and sunlight. The air should be free from dust, salt and sulphur.
Do not touch the product immediately after autoclaving as it will be
very hot and must remain in a sterile condition.
8. Cartridge Replacement (NAC,BB,NBBW)
1) Loosen the Union Nut and remove the head from the sheath.
Pull out the Middle Gear from the head. (Fig. 5)
Union Nut
2) Place Cap Wrench on the End Cap and turn
counter-clockwise. Unscrew and remove the cap. (Fig. 6)
3) Swing to open the latch plate. Insert a bur
blank or round rod into a hole from the head end, and push out the cartridge. At this time, if the inside of the head is dirty, clean the inside by applying PANA SPRAY Plus/PANA SPRAY.
4) Insert the new cartridge, with its pin aligned
in the slot of the head. (Fig. 7)
5) Mount each part in a reverse procedure to
the removal procedure, and firmly tighten the end cap with the cap wrench.
Middle Gear
End Cap
Cap Wrench
Slit
Fig.5
Fig.6
Pin
Fig.7
* Cartridge (NAC-03) NAC-Y : Order Code C030011
(BB-03) BB-Y : Order Code C037015 (NBB-03) NBBW-Y : Order Code C210015
* Cap Wrench(Black) : Order Code C071751 * Middle Gear (NAC-02) NAC,BB,NBBW : Order Code C032012
9. Periodical Maintenance Checks
Every 3 months perform periodical maintenance checks, referring to the check sheet below. If any abnormalities are found, contact your Authorized NSK Dealer.
Points to check Details
Union nut is loose
End cap is loose
Rotation
Check the union nut on the head is firmly tightened.
Check the end cap is firmly tightened.
Rotate the handpiece and check for abnormalities such as
abnormal rotation, vibration, noise, and overheating.
2017.04.28 03