Digital Differential Refractometer
Cat.No.3930
DD-7
Application examples:
♦
Te a
♦
Unsweetened or diet beverages
♦
Coffee or herbal extracts
♦
Cleaning and sanitizing solutions
♦
Surfactants
The DD-7 measures solutions of up to 2% Brix concentration at the
The DD-7 measures solutions of up to 2% Brix concentration at the
extremely high-accuracy level of
sales@novatech-usa.com
www.novatech-usa.com
Tel: (866) 433-6682 Fax: (866) 433-6684
Tel: (281) 359-8538 Fax: (281) 359-0084
extremely high-accuracy level of
(Note that the range of measurement gets narrower at concentrations higher than 10% Brix.) When a reference solution of 8.000% Brix is used, for
example, the DD-7 can measure concentrations in the range of 8% to 10% Brix at resolution of 0.001% Brix and precision of measurement of
±0.005% Brix. Note: High viscosity samples may not be suitable for measurement.
±0.005% Brix.
• Measurement is very simple. Inject a reference solution and a sample solution to respective injection ports,
and press the Start Switch.
• Digital readings eliminate reading errors resulting from user subjectivity.
Measurement data can be exported to a printer or computer via RS-232C connection.
•
Measure low-concentration food and
beverages, such as coffee, tea, diet sodas,
and herbal extracts.
Liquid Crystal Display (LCD)
Operation Panel
Reference Port
Measure concentrations of sanitizers and
disinfectants, such as hydrogen peroxide
solutions, at a precision level of ±0.012% or
higher.
Specifications
• Method of measurement : Differential optical refraction
method
• Measurement range : 0.000 to 2.000% in concentration
It is possible to measure samples with refractive index
up to 1.50 (nD) by reference solutuion.
(
Samples must be low in viscosity.
• Measurement accuracy : ±0.005% (In the case of
measurement of sucrose
solution (0 to 2%) at 30˚C
• Unit of minimum display : 0.001%
• Display area : Two 24-characer lines (LCD
• Light source : LED
• Light sensor : Photo-diode
• Output RS-232C Printer : Digital printer DP-63,
DP-AD(optional)
• Ambient temperature : 10 to 30˚C
• Power supply : AC100 to 240V, 50/60Hz
• Power consumption : 50VA
• Dimensions : 36(W) × 35(D) × 14(H)cm
Sample Port
• Weight : 5.8kg
Measure surfactants, anti-rust agents, metalworking fluids, and other industrial solutions
at ±0.005% Brix accuracy.
(
for sucrose solution
)
(
dual type
)
)
)
)
All ATAGO products are designed and manufactured in Japan.
Headquarters: 32-10, Honcho, Itabashi-ku, Tokyo 173-0001 Japan
TEL : 81-3-3964-6156 FAX : 81-3-3964-6137
overseas@atago.net http : //www.atago.net/
TEL : 1-425-637-2107 customerservice@atago-usa.com
TEL : 91-22-2833-8038 / 8076 customerservice@atago-india.com
TEL : 662-982-8718-9 customerservice@atago-thailand.com
TEL : 55 16 3916-6000 customerservice@atago-brasil.com
TEL : 39 2 36557267 customerservice@atago-italia.com
TEL : 86 20-38106065 info@atago-china.com
ISO9001
REGISTERED
ATAGO products comply with HACCP,
GMP, and GLP system standards.
*Specifications and appearance are subject to change without notice.