Digital Differential Refractometer
Cat.No.3930
DD-7
Application examples:
♦
Te a
♦
Unsweetened or diet beverages
♦
Coffee or herbal extracts
♦
Cleaning and sanitizing solutions
♦
Surfactants

The DD-7 measures solutions of up to 2% Brix concentration at the
sales@novatech-usa.com
www.novatech-usa.com
Tel: (866) 433-6682 Fax: (866) 433-6684
Tel: (281) 359-8538 Fax: (281) 359-0084
extremely high-accuracy level of
(Note that the range of measurement gets narrower at concentrations higher than 10% Brix.) When a reference solution of 8.000% Brix is used, for
example, the DD-7 can measure concentrations in the range of 8% to 10% Brix at resolution of 0.001% Brix and precision of measurement of
±0.005% Brix. Note: High viscosity samples may not be suitable for measurement.
±0.005% Brix.
• Measurement is very simple. Inject a reference solution and a sample solution to respective injection ports,
and press the Start Switch.
• Digital readings eliminate reading errors resulting from user subjectivity.
Measurement data can be exported to a printer or computer via RS-232C connection.
•
Measure low-concentration food and
beverages, such as coffee, tea, diet sodas,
and herbal extracts.
Measure concentrations of sanitizers and
disinfectants, such as hydrogen peroxide
solutions, at a precision level of ±0.012% or
higher.
Measure surfactants, anti-rust agents, metalworking fluids, and other industrial solutions
at ±0.005% Brix accuracy.
Liquid Crystal Display (LCD)
Operation Panel
Reference Port
Sample Port
All ATAGO products are designed and manufactured in Japan.
Headquarters: 32-10, Honcho, Itabashi-ku, Tokyo 173-0001 Japan
TEL : 81-3-3964-6156 FAX : 81-3-3964-6137
[email protected] http : //www.atago.net/
Specifications
• Method of measurement : Differential optical refraction
method
• Measurement range : 0.000 to 2.000% in concentration
It is possible to measure samples with refractive index
up to 1.50 (nD) by reference solutuion.
(
Samples must be low in viscosity.
• Measurement accuracy : ±0.005% (In the case of
measurement of sucrose
solution (0 to 2%) at 30˚C
• Unit of minimum display : 0.001%
• Display area : Two 24-characer lines (LCD
• Light source : LED
• Light sensor : Photo-diode
• Output RS-232C Printer : Digital printer DP-63,
DP-AD(optional)
• Ambient temperature : 10 to 30˚C
• Power supply : AC100 to 240V, 50/60Hz
• Power consumption : 50VA
• Dimensions : 36(W) × 35(D) × 14(H)cm
• Weight : 5.8kg
ISO9001
REGISTERED
*Specifications and appearance are subject to change without notice.
(
for sucrose solution
)
(
dual type
ATAGO products comply with HACCP,
GMP, and GLP system standards.
)
)
)
)
11122000PP Printed in Japan