Nova-Tech CON 700 User Manual

Instruction Manual
gyM
sy...
sales@novatech-usa.com www.novatech-usa.com Tel: (866) 433-6682 Fax: (866) 433-6684 Tel: (281) 359-8538 Fax: (281) 359-0084
CON 700 Conductivity/ºC/ºF Bench Meter
Part of Thermo Fisher Scientific 68X541702 Rev. 1 Jan 2010
Tech nol o
adeEa
TABLE OF CONTENTS
1. INTRODUCTION.............................................................. 1
2. GETTING STARTED........................................................ 2
2.1 Keypad Functions ...................................................................2
2.2 LCD Annunciators ................................................................... 3
2.3 Meter Connections .................................................................. 3
3. ELECTRODE INFORMATION......................................... 4
4. CALIBRATION................................................................. 4
4.1 Automatic or Manual Calibration ............................................. 4
4.2 Single or Multi-Point Calibration.............................................. 5
4.3 General Calibration Tips .........................................................6
4.4 Automatic Conductivity Calibration Procedure ........................ 6
4.5 Manual Conductivity & TDS Calibration Procedure................. 7
4.6 Temperature Calibration .........................................................7
5. MEASUREMENT.............................................................. 7
5.1 Taking Measurements............................................................. 7
5.2 Automatic and Manual Ranging .............................................. 8
5.3 HOLD Function .......................................................................8
5.4 Storing and Recalling Data .....................................................8
6. SETUP FUNCTIONS........................................................ 9
6.1 P1.0 CAL (Calibration) ............................................................9
6.2 P2.0 ELE (Electrode Information)............................................ 9
6.3 P3.0 ConF (Configuration) ......................................................9
6.4 P3.1 rdY (Ready / Stability Indicator) ...................................... 9
6.5 P3.2 ºC ºF (Celcius Or Fahrenheit) ......................................... 10
6.6 P3.3 AtC (Automatic Temperature Compensation) ................. 10
6.7 P3.4 tdS (Total Dissolved Solids factor).................................. 10
6.8 P3.5 t.CO (Temperature Coefficient)....................................... 10
6.9 P3.6 t.nr (Normalization Temperature in ºC) ...........................11
6.10 ACAL (Automatic Conductivity Calibration)............................. 11
6.11 SPC (Single Point Calibration) ................................................ 11
6.12 CELL (Nominal Cell Constant) ................................................ 11
6.13 P4.0 rSt (Reset) .....................................................................12
6.14 P5.0 CLr (Clear Memory) ....................................................... 12
7. CALCULATING TDS CONVERSION FACTOR.............. 13
8. CALCULATING TEMPERATURE COEFFICIENTS........ 14
9. REPLACEMENTS AND ACCESSORIES........................ 15
10. TROUBLESHOOTING GUIDE......................................... 16
11. SPECIFICATIONS............................................................ 17
12. WARRANTY..................................................................... 18
13. RETURN OF ITEMS......................................................... 18

1. INTRODUCTION

Thank you for purchasing our Con 700 series benchtop meter. This microprocessor-based meter is economical and simple to use. The design incorporates a large LCD for clear viewing, yet offers a small footprint to conserve space.
The CON 700 measures conductivity or TDS, and temperature (ºC or ºF) simultaneously.
Each meter includes a convenient slide-out card for quick reference.
Some configurations include an electrode arm and metal bracket which can be easily attached to the left or right side of the meter according to your preference.
The CON 700 series benchtop meter replaces our popular CON 510 series meter that was introduced in 2002.
We take great pride in every instrument we manufacture and hope this one serves you well.
1

2. GETTING STARTED

2.1 Keypad Functions

Powers the meter on and off. Upon power on, the meter automatically begins in the mode that was last used. Calibration and memory values are retained even if meter is unplugged.
Toggle between available measurement modes— Conductivity with Temperature or TDS with Temperature. Also used to switch to Temperature calibration during Conductivity or TDS calibration mode.
Press and hold for 5 seconds to enter SETUP mode.
Toggles between measurement and calibration modes. In SETUP mode, returns user to the measurement mode.
Confirms calibration values in CAL mode. Confirms selections in SETUP mode. Changes range and resolution in MEAS mode.
View recalled values in memory mode.
MI (Memory Insert) stores values into memory.
Increase value. Scroll up in SETUP & CAL modes.
MR (Memory Recall) recalls values from memory
Decrease value. Scroll down in SETUP & CAL modes.
Freezes measured reading. Press again to resume live reading.
2

2.2 LCD Annunciators

2.3 Meter Connections

CON/
Temp
DC
8-pin DIN connection for 2-cell Con/TDS/Temp electrode
Power supply
3

3. ELECTRODE INFORMATION

The CON700 includes an electrode with a nominal cell constant of k = 1.0, built-in temperature sensor, and 1 meter cable. The Ultem body housing has good chemical resistant properties. The electrode design offers fast temperature response and reduces air entrapment, ensuring accurate, repeatable, and stable readings.
he wetted materials of the probe include:
T
Polyetherimide (Ultem) – protective probe guard
Polybutylterphalate (Valox) – sensor housing
Stainless Steel (SS 304) – 2 bands
The protective probe guard can be removed temporarily for cleaning however it must be re-attached during measurement and
calibration. Erroneous results will occur while the probe guard
is removed.
Always immerse the probe beyond upper steel band for best results. Use the fill line on the outside of the probe guard for
reference.

4. CALIBRATION

4.1 Automatic or Manual Calibration

The CON 700 is capable of automatic or manual calibration for conductivity, and manual calibration for TDS.
In the automatic calibration mode, the meter will automatically select one of (4) conductivity calibration standard values depending o
range and normalization temperature being used (TABLE 1). If you only use calibration standards that are listed in TABLE 1,
automatic calibration is recommended. If you intend to calibrate with one or more standards that are not listed in must be set for manual calibration instead.
TABLE 1, the CON 700
n the
4
The factory default is automatic conductivity calibration. See Section
6.10 to change this setting. There is no automatic calibration value
available for conductivity range 1.
TABLE 1
Automatic Calibration Values
Range # Conductivity
Range
r 1 0.00 – 20.00 μS None None
r 2 20.1 – 200.0 μS 84 μS 76 μS
r 3 200.1 – 2000 μS 1413 μS 1278 μS
r 4 2.01 – 20.00 mS 12.88 mS 11.67 mS
r 5 20.1 – 200.0 mS 111.8 mS 102.1 mS
Normalization Temperature
25 ºC 20 ºC
Range
#
r 1 0 – 10.00 ppm none
r 2 10.1 – 100.0 ppm
r 3 101 – 1000 ppm
r 4 1.01 – 10.00 ppt
r 5 10.1 – 100 ppt

4.2 Single or Multi-Point Calibration

Use Single-Point Calibration to apply a single calibration value across all ranges.
Use Multi-Point Calibration for individual calibration in each range. This will restrict an individual calibration so that it is applied to one range only. When using multi-point calibration, perform a calibration in each range that you expect to use for best results.
TDS Range
(using 0.5 TDS factor)
Automatic Calibration
Values
none none none none
5
The factory default is Single-Point Calibration. See Section 6.11 to
change this setting.

4.3 General Calibration Tips

or best results, periodic calibration with known accurate standards is
F recommended. A maximum of one calibration point per range can be performed. If multiple calibration points are used in the same range, the most recent one will replace the previous one.
When the electrode is replaced, it is best to clear the calibration to the
factory default values (see Section 6.13).
Rinse or immerse the probe before calibration and between samples with clean water (deionized water is ideal).
The CON 700 has non-volatile memory which will retain all calibration values, as well as meter settings and memory values upon meter shut down or unexpected power loss.
To protect from erroneous calibrations, the allowable tolerance is ±40% of the factory default value.
Low conductivity standard solutions (less than 20 µS) are unstable and are very temperature dependent. As a result, reproducible calibration results are challenging in lowest measurement range #1 (0.00 to 20.0 µS).

4.4 Automatic Conductivity Calibration Procedure

1) Press
2) Dip the electrode into the conductivity standard and press . Provide stirring for best results.
3) The primary display will show the factory default value, while the secondary display will lock on the appropriate automatic standard
value from TABLE 1.
4) When the READY indicator appears, press
primary reading will flash briefly before returning to measurement mode upon successful calibration.
as needed to select conductivity (μS or mS).
to accept. The
6

4.5 Manual Conductivity & TDS Calibration Procedure

1) Press as needed to select conductivity (μS or mS) or TDS (ppm or ppt) calibration.
2) Dip the electrode into the calibration standard and press
.
Provide stirring for best results.
3) The primary display will show the current reading, while the secondary display will be the factory default value. Adjust the
temperature reading using
or . Press to accept. The
primary reading will flash briefly before returning to measurement mode upon successful calibration.

4.6 Temperature Calibration

T
he thermistor sensor used for automatic temperature compensation and measurement is accurate and stable, so frequent calibration isn’t required. Temperature calibration is recommended upon probe replacement, whenever the temperature reading is suspect, or if matching against a certified thermometer is desired.
1) Place the probe into a solution with a known accurate temperature
such as a constant temperature bath.
2) Press
followed by . The primary display shows the measured temperature while the secondary display shows the factory default temperature.
4) Adjust the temperature using
to cancel. The meter allows an adjustable maximum value of ± 5 (or ± 0.9
º
F) from the factory default temperature.
or . Press to accept or
º

5. MEASUREMENT

C

5.1 Taking Measurements

1) Rinse the electrode with de-ionized or distilled water before use to remove any impurities. Gently shake excess water droplets.
7
2) Dip the probe into the sample beyond the upper steel band (utilize the fill line on the outside of the probe guard for reference).
3) Allow time for the reading to stabilize. Note the reading on the display. The clear yellow protective probe guard must be attached during measurement. Erroneous results will occur while the probe guard is removed.

5.2 Automatic and Manual Ranging

The CON 700 automatically selects the optimum range in which your
readings appear. Refer to TABLE 1 for a list of the available ranges.
To turn-off automatic ranging, press
. MEAS will flash, indicating
that manual ranging is active. To manually select the next range,
press again. After range 5, automatic ranging is resumed.
For example, a reading of 465 μS will automatically settle in range 3. Using manual range advancement this will read as 0.47 mS in range 4, and 0.5 mS in range 5. For best resolution, use auto-ranging.

5.3 HOLD Function

F
or prolonged observation of a reading, press
during measurement mode to freeze the display. The “HOLD” indicator will display when the reading is held. To release the held value and
resume measurement, press
memory by pressing
.
again or insert the held value into

5.4 Storing and Recalling Data

The CON 700 can retain up to 100 data points into memory for later retrieval.
1) In the measurement mode, press
to insert the measured value into memory. The stored memory location value (StO) is briefly displayed.
8
2) To recall data from memory, press . The location of the most
recent stored data is displayed first. Press or to select the
location of the desired data, then press
3) Press to measurement mode. To erase stored data, see Section 6.14.
to return to the stored data location. Press to return
to accept.

6. SETUP FUNCTIONS

Use the set up mode to customize your instrument operation. During
measurement, press and hold mode.
or to change programs or change options. Press
Press to select the program or confirm selection.
Press

6.1 P1.0 CAL (Calibration)

6.2 P2.0 ELE (Electrode Information)

6.3 P3.0 ConF (Configuration)

6.4 P3.1 rdY (Ready / Stability Indicator)

to go back one level or return to measurement mode.
Press
Press
Press
to view stored calibration points in each range.
to view cell efficiency in each range.
to access set-up programs 3.1 thru 3.9.
for 5 seconds to enter SETUP
Press
Press or Auto HOLd.
.
to choose READY “On”, READY “OFF”, or
9
Press to confirm.
6.5 P3. (C hrenheit) 2 ºC ºF elcius Or Fa
Press
Press or to select ºC or ºF.
Press to
6.6 P3. perature Compensation)
6.7 P3. ot Solids factor)

6.8 P3.5 t.CO (Temperature Coefficient)

T degree temperature (% per ºC). The CON 70 temperature coefficient of 2.1 % per ºC. For m provide good results. The meter allows adjustment from 0.0 to 10.0.
TIP: Select 0.0% for uncompensated measurements. The temperature
will be measured by the electrode and displayed in measurement mode—without compensation.
See Section 8 – Calculating Temperature Coefficients.
3 AtC (Automatic Tem
Press
Press or to select “Yes” or “No”.
Press
4 tdS (T al Dissolved
ress
P
Press or to select the desired TDS factor (.40 to
1.00).
Press
he temperature coefficie ount of change in conductivity per
.
confirm.
.
to confirm.
.
to confirm.
nt is the am
0 is factory set to a
ost applications this will
10
Press .
Press or to select the desired value (0.00 to 10.0).
to confirm. Press

6.9 P3.6 t.nr (Normalization Temperature in ºC)

W
hen Au
tomatic Temperature Compensation is used, measurements are adjuste the temp coefficient to the normalization tempera s 25 ºC.

6.10 ACAL (Automatic Conductivity Calibration)

See Sec
6.11 SP le int Calibration)
See Sec
d by erature
re. The default value itu
Press .
Press or to select the desired value (15.0 to 30.0).
Press to confirm.
Press .
Press or to select “Yes” (Automatic) or “No” (Manual).
Press to confirm.
C (Sing Po
tio
n 4.2 for details on Single Point & Multi Point Calibration.
Press .
n. tion 4.1 for more details on Automatic Conductivity Calibratio
Press or to select “Yes” (Single) or “No” (Multi).
Press to confirm.
6.12 CELL (Nomin Cell Constant)
The CON 700 inclu 1.0. Use probes with k = 0.1 and 10 (sold separately) for improved
al
des a probe with a nominal cell constant (k) of
11
performa ce in extreme sample ranges. Use this setup function to change the cell constant if necessary. M included probe. k = 0.1 ideal for low measurements <20 µS (<10 ppm). k = 1.0 ideal for mid-range measurements k = 10 ideal for high measurements >20 mS (>10 ppt).
n
eter default is 1.0 to match the
Press
Press or to select 0.1, 1.0, or 10.0
Press to confirm.
.

6.13 P4.0 rSt (Reset)

Press .
or Press to select “Yes” (Reset) or “No” (Cancel).
If “Yes”, press
r “F (complete reset to factory default settings). only) o Ct”
Press

6.14 P5.0 CLr (Clear Memory)

Press
Press or to select “Yes” (Erase memory) or “No”.
Press
to confirm.
.
to confirm.
or to select “Cal” (calibration reset
12

7. CALCULATING TDS CONVERSION FACTOR

You can calibrate TDS using the value of the calibration standard solution at a standard temperature such as 25 ºC. To determine the conductivity­to-TDS conversion factor use the following formula:
Factor = Actual TDS ÷ Actual Conductivity @ 25 ºC
Actual TDS: Value from the solution bottle label or as a
standard made using high purity water and precisely weighed
salts.
ppm = milligram of salt(s) per liter of water = mg/L
ppt = gram of salt(s) per liter of water = g/L
Actual Conductivity: Value measured using a properly
calibrated Conductivity/Temperature meter.
Both the Actual TDS and the Actual Conductivity values must be in the same magnitude of units. For example, if the TDS value is in ppm the conductivity value must be in µS; if the TDS value is in ppt the conductivity value must be in mS.
Check your factor by multiplying the conductivity reading by the factor in the above formula. The result should be in TDS value.
When the TDS factor is set to 1.0, Conductivity = TDS.
13

8. CALCULATING TEMPERATURE COEFFICIENTS

To determine the temperature coefficient of your sample solution use this formula:
Where:
tc = Temperature coefficient 25 = 25 ºC
C
= Conductivity at Temp 1 CT2 = Conductivity at Temp 2
T1
T1 = Temp 1 T2 = Temp 2
NOTE: A controlled temperature water bath is ideal for this procedure.
1. Immerse the probe into a sample of your solution and adjust the temperature coefficient to 0% (that is, no compensation) by following
instructions as described in Section 6.8.
2. Wait for 5 minutes. Note T1 and CT1 (conductivity at T1).
3. Condition the sample solution and probe to a temperature (T2) that is about 5 ºC to 10 ºC different from T1, and note the conductivity reading C
.
T2
NOTE: Record your results for future reference. Ideally T1 and T2 should
bracket your measurement temperature, and should not differ by more than 5 ºC.
4. Calculate the temperature coefficient of your solution according to the formula shown above.
5. Enter the calculated temperature coefficient into the meter.
The calculated temperature coefficient will now be applied to all the meter readings.
14

9. REPLACEMENTS AND ACCESSORIES

Item Description
CON 700 Benchtop with electrode and integral stand,100/240 VAC
Replacement electrode, k = 1.0 CONSEN9501D 35608-74
Epoxy/platinum electrode, k = 0.1 - 35608-72
Glass/platinum electrode, k = 1.0 - 35608-76
Epoxy/platinum electrode, k = 10.0 - 35608-78
SMPS, 100/240 VAC, 9V, 6W power adapter
10 μS, (20) x 20mL Sachets ECCON10BS 35653-09
84 µS, 480 mL bottle* ECCON84BT 00653-16
100 μS, 480 mL bottle ECCON100BT -
447 μS, 480 mL bottle ECCON447BS 00653-47
500 μS, 480 mL bottle ECCON500BT -
1413 μS, 480 mL bottle* ECCON1413BT 00653-18
1413 μS, (20) x 20mL Sachets* ECCON1413BS 35653-11
2764 μS, 480 mL bottle ECCON2764BT 00653-20
2764 μS, (20) x 20mL Sachets ECCON2764BS 35653-12
5.0 mS, 480 mL bottle ECCON5000BT -
12.88 mS, 480 mL bottle* ECCON1288BT 00606-10
15 mS, (20) x 20mL Sachets ECCON15000BS 35653-13
111.8 mS, 480 mL bottle* ECCON1118BT -
Part number Ordering Code
Eutech Instruments Oakton Instruments
ECCON70043S 35411-00
60X030130 35615-07
*Automatic Calibration Standard Value at 25 ºC
15

10. TROUBLESHOOTING GUIDE

PROBLEM CAUSE SOLUTION
No display Main power not switched
“Ur” (Under range)
“Or” (Over range)
in primary or secondary display
Slow response
Meter not responding to key press
Secondary display continually scrolls
on. AC Adapter socket not inserted properly.
Measured value is out of range.
Electrodes not connected.
Electrode clogged, dirty or broken.
Meter not calibrated.
Wrong temp value.
Dirty electrode.
Temperature of sample is changing.
Manual HOLD or Auto HOLD is active.
Worn keypad.
Invalid key; Button is not functional in the current operation mode.
Conductivity calibration standard is not within 40% of expected value during automatic calibration.
a) Switch on the power supply.
b) Re-insert AC Adapter.
Check electrode is connected.
Clean or replace electrode.
Recalibrate the meter.
Confirm measurement condition.
Treat samples to bring within meter measuring range.
Reset meter.
Clean the electrode.
Allow temperature to stabilize.
Press HOLD. See Section 6.4 to disable Auto Hold feature if enabled.
Contact Technical Service.
Press alternate key. Select valid key depending on current mode.
Ensure electrode guard has not been removed.
Use fresh or different calibration standard.
to deactivate manual
16

11. SPECIFICATIONS

Conductivity Range
TDS Range (0.5 TDS factor)
Resolution 0.05 % Full Scale
Accuracy ±1% Full Scale
Temperature
Range (Meter)
Compensation Type & Range
Resolution Accuracy Calibration Coefficient Normalization
Cell Constant 0.1, 1.0, 10.0 (selectable)
TDS Factor 0.40 to 1.00 (adjustable)
Calibration points
Auto-ranging Yes
Hold Function Yes
Memory 100 data sets
Input 8-pin DIN
Power AC/DC 9V, 6W Adapter (100/240 VAC, 50-60Hz)
Dimensions (mm) meter only = 175 (L) x 155 (W) x 69 (H)
0 to 10.00, 100.0 ppt
0.0 to 80 ºC / 32.0 to 176.0 ºF with supplied cell
Offset in 0.1 º increments; Offset range: 5 ºC / 9 ºF
0 to 20.00, 200.0, 2000 µS
0 to 20.00, 200.0 mS
0 to 10.00, 100.0, 1000 ppm
(max 200.0 ppt @ 1.0 TDS factor)
0.0 to 100.0 ºC / 32.0 to 212.0 ºF
Automatic with supplied cell or Manual.
0.0 to 100 ºC / 32.0 to 212.0 ºF,
0.1 ºC / ºF
± 0.5ºC / ± 0.9ºF
0.0 to 10.0% per ºC
15.0 to 30.0 ºC (adjustable)
Automatic (4 points); Maximum 1 per range
Manual (5 points); Maximum 1 per range
17

12. WARRANTY

This meter is supplied with a warranty against significant deviations in material and workmanship for a period of THREE years from date of purchase whereas probe with a SIX month warranty.
If repair or adjustment is necessary and has not been the result of abuse or misuse within the designated period, please return – freight prepaid – and correction will be made without charge. Eutech Instruments/Oakton Instruments will determine if the product problem is due to deviations or customer misuse.
Out of warranty products will be repaired on a charged basis.
The warranty on your instrument shall not apply to defects resulting from:
Improper or inadequate maintenance by customer
Unauthorized modification or misuse
Operation outside of the environment specifications of
the products

13. RETURN OF ITEMS

Authorization must be obtained from our Customer Service Department or authorized distributor before returning items for any reason. A “Return Goods Authorization” (RGA) form is available through our authorized distributor. Please include data regarding the reason the items are to be returned. For your protection, items must be carefully packed to prevent damage in shipment and insured against possible damage or loss. Eutech Instruments will not be responsible for damage resulting from careless or insufficient packing. A restocking charge will be made on all unauthorized returns.
NOTE: Eutech Instruments Pte Ltd /Oakton Instruments reserve the right
to make improvements in design, construction, and appearance of products without notice.
18
sales@novatech-usa.com www.novatech-usa.com Tel: (866) 433-6682 Fax: (866) 433-6684 Tel: (281) 359-8538 Fax: (281) 359-0084
For more information on our products, please contact our channel partner or visit our websites listed below:
Oakton Instruments
625 E Bunker Court Vernon Hills, IL 60061 USA Tel: (1) 888-462-5866 Fax: (1) 847-247-2984

info@4oakton.com www.4oakton.com

Distributed by:
Part of Thermo Fisher Scientific
Loading...