Model C10
sales@novatech-usa.com
www.novatech-usa.com
Tel: (866) 433-6682 Fax: (866) 433-6684
Tel: (281) 359-8538 Fax: (281) 359-0084
z Benchtop Abbe Precision
z Liquid & Solids Testing
z Brix & Refractive Index Scales
The VEE GEE Model C10 Abbe Refractometer is an excellent instrument for the
Refractive Index (nD) analysis of liquids, glass, plastic, & film.
Easy to Use.
Operation consists of placing a sample on the prism; while looking
through the eyepiece, the control knob is turned until the shadowline is centered in
the crosshairs (see image below left); the reading is taken where the vertical line
crosses the scale. Calibrates with water.
Industrial Features. Oversized, easy-to-read scale; Refractive Index (nD) & Brix
scales; primary & secondar y prisms; color compensator dial; large measurement
dial; focusable eyepiece.
Scale View
Ordering Information
Prism Temperature Control. Featuring 2 circulator nozzles for the primar y prism
and 2 for the secondar
y prism, Model AR10 can be connected to a circulating bath to
accurately regulate prism temperature. The prism temperature is displayed on the
external glass thermometer. The effective temperature range is 0-70°.
Film Testing. Transparent or translucent film samples can be measured with the use
of a glass standard (aids in the illumination of the sample), contact liquid (adheres the
sample to the glass standard and to the primar y prism), and an external sodium light
source (sold separately).
Note: The refractive index (nD) value of the sample must be lower than that of the contact liquid and the glass standard. The
included bromonaphthalene contact liquid has a refractive index (nD) value of 1.6300. The included glass standard has
a refractive index value (nD) of 1.5163.
Glass & Plastic Testing. Transparent or translucent glass and plastic samples can be
measured with the use of a contact liquid (adheres the sample to the primar y prism), and
an external sodium light source (sold separately).
Note: The refractive index (nD) of the sample must be lower than that of the contact liquid. The included bromonaphthalene
contact liquid has a refractive index (nD) value of 1.6300.
Specifications
Measuring System: Optical Refraction Critical Angle Detection
Measuring Range: Refractive Index (nD): 1.3000-1.7000
Resolution: Refractive Index (nD): 0.0005
Accuracy: Refractive Index (nD): ±0.0003
Dispersion (nF-nC): Dial Readings w/ Conversion Charts
Ambient Temperature: 5-35°C
Measuring Temperature: 0.0-70.0°C
Calibration Liquid / Medium: Distilled Water
Prism: Optical Glass
Prism Housing: SUS316 Stainless Steel
Circulator Nozzles: 2 for Primary Prism, 2 for Secondary Prism
Dimensions: 140 (L) x 90 (W) x 240 (H) mm
Weight: 2.3kg
Supplied With: Glass Thermometer (1 ea.)
Brix: 0.0-95.0%
Brix: 0.25%
Brix: ±0.1 5%
Glass Standard, 1.5163 nD
(w/ Bromonaphthalene Contact Liquid)
Glass Standard, 1.5163 nD (1 ea.)
Contact Liquid, Bromonaphthalene, 4ml Bottle (1 ea.)
Scale Adjustment Tool (1 ea.)
Carrying Case (1 ea.)
Instruction Manual (1 ea.)
Cat. No.
C10
Printed in the USA No. PS032004RFC10DSL-A
Model
C10
Scale
Refractive Index (nD)
Brix
Range
1.3000-1.7000 nD
0.0-95.0%
Resolution
0.0005 nD
0.25%
Accuracy
±0.0003 nD
±0.15%
Dimensions / Weight
140 (L) x 90 (W) x 240 (H) mm / 2.3kg