Nova-Tech ATR User Manual

ATR-Series
Automatic Refractometers
ATR-Series Refractometers
Cat. No.
Measuring Range
Resolution
Optical Wavelength: Measuring Light Source:
RI: Brix: RI: Brix: RI: Brix:
Sensor: Prism: Prism Sealant: Prism Assembly Insert: Stage: Working Temperature: Ambient Temperature: Temperature Correction Range: Temperature Sensor Accuracy: Temperature Control: Temperature Control Stability: Flow Through Measurement: Sample Quantity (Standard): Sample Quantity (Flow Through): Zero Calibration Liquid: Custom Calibration Liquid: Calibration History: Display: Keypad: Password Protection: Languages: Printer Interface: Computer Port: USB Port: Programmable Scales: Power Requirements: Power Supply: Dimensions Control:
Weight Control:
Sensor:
Sensor:
Manufactured In:
ATR-ST
45001
1.3320 - 1.5230
0.0 - 95.0%
0.0001
0.05%
±0.0001
±0.05%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
®
Teflon
Stainless Steel
+5 to +50°C +10 to +40°C +10 to +40°C
±0.03°C
N/A N/A N/A
>0.3mL
N/A
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C Optional
10
90-240V AC, 50/60Hz
5V DC
290 x 220 x 120mm
3.0kg
Germany
BASIC: Single instrument design offers the core features of the ATR-Series line. Recommended for users who don’t require temperature control and therefore are conducting measurements on samples at or near room temperature. Models ATR-ST
plus
offer enhanced ranges, resolution, and accuracy.
ATR-ST
plus
45001P
1.30000 - 1.54000
0.00 - 100.00%
0.00002
0.02%
±0.00005
±0.03%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
®
Teflon
Stainless Steel
+5 to +50°C
+10 to +40°C
+5 to +50°C
±0.03°C
N/A N/A N/A
>0.3mL
N/A
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C
Optional
10
90-240V AC, 50/60Hz
5V DC
290 x 220 x 120mm
3.0kg
Germany
ATR-ST
S plus
45001SP
1.28000 - 1.54000
0.00 - 100.00%
0.00001
0.01%
±0.00003
±0.02%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
®
Teflon
Stainless Steel
+5 to +50°C
+10 to +40°C
+5 to +50°C
±0.03°C
N/A N/A N/A
>0.3mL
N/A
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C
Optional
10
90-240V AC, 50/60Hz
5V DC
290 x 220 x 120mm
3.0kg
Germany
plus
and ATR-ST S
ATR-SW
45004
1.33200 - 1.53200
0.00 - 95.00%
0.00002
0.02%
±0.00004
±0.04%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
N/A
Stainless Steel
+5 to +50°C +10 to +40°C +10 to +40°C
±0.03°C External Circulator External Circulator
Yes
>0.3mL
>35mL
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C
Optional
10
90-240V AC, 50/60Hz
5V DC 290 x 220 x 120mm 200 x 160 x 150mm
3.0kg
4.5kg
Germany
INTERMEDIATE: Separate sensor in stainless steel housing. Users with external water circulators can accurately control prism well temperature. Model ATR-SW ranges, resolution, and accuracy.
ATR-SW
plus
45004P
1.30000 - 1.54000
0.00 - 100.00%
0.00001
0.01%
±0.00002
±0.02%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
N/A
Stainless Steel
+5 to +50°C
+10 to +40°C
+5 to +50°C
±0.03°C External Circulator External Circulator
Yes
>0.3mL
>35mL
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C
Optional
10
90-240V AC, 50/60Hz
5V DC 290 x 220 x 120mm 200 x 160 x 150mm
3.0kg
4.5kg
Germany
plus
offers enhanced
ATR-W1
45006
1.33200 - 1.53200
0.00 - 95.00%
0.00002
0.02%
±0.00004
±0.04%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
N/A
Stainless Steel
+10 to +80°C +10 to +40°C +10 to +50°C
±0.03°C
Electronic Peltier
±0.02°C
N/A
>0.3mL
N/A
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C
Optional
10
90-240V AC, 50/60Hz
15V DC 290 x 220 x 120mm 200 x 160 x 140mm
3.0kg
4.5kg
Germany
ADVANCED: Separate sensor in stainless steel housing. Features internal electronic Peltier temperature control. Model ATR-W1
plus
accuracy.
offers enhanced ranges, resolution, and
ATR-W1
plus
45006P
1.33200 - 1.53200
0.00 - 100.00%
0.00001
0.01%
±0.00002
±0.02%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
N/A
Stainless Steel +10 to +80°C +10 to +40°C +10 to +50°C
±0.03°C
Electronic Peltier
±0.02°C
N/A
>0.3mL
N/A
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C
Optional
10
90-240V AC, 50/60Hz
15V DC 290 x 220 x 120mm 200 x 160 x 140mm
3.0kg
4.5kg
Germany
ATR-W2
45007
1.33200 - 1.70000
0.00 - 95.00%
0.00002
0.02%
±0.00004
±0.04%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
N/A
Stainless Steel
+10 to +80°C +10 to +40°C +10 to +50°C
±0.03°C
Electronic Peltier
±0.02°C
N/A
>0.3mL
N/A
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C
Optional
10
90-240V AC, 50/60Hz
15V DC 290 x 220 x 120mm 200 x 160 x 140mm
3.0kg
4.5kg
Germany
FULLY-LOADED: Separate sensor in stainless steel housing. Offers the full Abbe refractive index range. Features int ernal electronic Peltier temperature control. Model ATR-W2 offers enhanced ranges, resolution, and accuracy.
ATR-W2
plus
45007P
1.33200 - 1.70000
0.00 - 100.00%
0.00001
0.01%
±0.00002
±0.02%
589.3nm LED
Linear Hi-Res CCD
Synthetic Sapphire
®
Viton
N/A
Stainless Steel +10 to +80°C +10 to +40°C +10 to +50°C
±0.03°C
Electronic Peltier
±0.02°C
N/A
>0.3mL
N/A
Distilled Water
Liquid Standard
Yes
LCD, 16x16 Characters
20-Key Alphanumeric
Yes
English & German
Centronics (Parallel)
RS-232C Optional
10
90-240V AC, 50/60Hz
15V DC 290 x 220 x 120mm 200 x 160 x 140mm
3.0kg
4.5kg
Germany
plus
Printed in the USA
No. PS062008RFATRSROL-A