Nova-Tech 4470 User Manual

SpecificationS
Range: –76 to 932°F / –60 to 500°C Resolution: 0.1° (-9.9 to 199.9) 1° (otherwise)
opeRation
Press and continue to hold down the trigger to take temperature readings. The unit will take temperature readings while the trigger is pressed and held (SCAN appearing on the display indicates measurements are being taken).
After the trigger is released, the last temperature measured will be displayed for approximately 15 seconds and the display will turn off automatically. HOLD will appear on the display when the trigger has been released to indicate that the unit is no longer taking temperature measurements.
Selecting unitS of meaSuRe
Each press of the °C/°F button will toggle between °C and °F.
laSeR taRgeting
Laser targeting is always active when the trigger is being pressed. Each time the trigger is pressed and held, the laser beam will be emitted. CAUTION: Eye damage may result from direct exposure to laser light.
maXimum tempeRatuRe
While taking measurements, the unit will automatically record the maximum temperature reading that has been measured. This value will appear on the display to the right of MAX.
DiStance Spot/Ratio
There are other factors that may affect measurement accuracy. The target
must completely ll the spot diameter
seen by the infrared sensor; otherwise
readings will be inuenced by the
surface surrounding the target. The ratio of the distance to the size of the spot being measured is 11:1. For example, an object’s diameter of 100mm can be measured from 1100mm.
DiSplaY meSSageS
Er2” will appear on the display when the unit is exposed to rapid changes in ambient temperature.
Er3” will appear on the display when the ambient temperature is below 32° F (0° C) or above 122° F (50° C).
Er” will appear on the display when the unit needs to be reset. To reset the unit, remove the batteries, wait 2 minutes and then re-install the batteries.
all opeRational DifficultieS
If this thermometer does not function properly for any reason, replace the batteries with new high quality batteries (see Battery Replacement section). Low battery power can occasionally cause any number of “apparent” operational
difculties. Replacing the batteries
with new fresh batteries will solve most
difculties.
BatteRY Replacement
Erratic readings, faint readings, no display, or appearing on the display are all indications that the batteries must be replaced. Open the battery compartment by sliding the battery cover in the direction indicated by the arrow and then lift the battery cover. Remove the exhausted bat­teries and replace them with two (2) new AAA alkaline batteries. Make certain to install the new batteries in the directions in­dicated in the battery compartment. Close the battery cover.
WaRRantY, SeRVice, oR RecaliBRation
For warranty, service, or recalibration, contact:
CONTROL COMPANY
4455 Rex Road
Friendswood, Texas 77546 USA
Ph. 281 482-1714 • Fax 281 482-9448
E-mail sales@control3.com • www.control3.com
Control Company is ISO 9001 Quality-Certied by
DNV and ISO 17025 accredited as a Calibration
Laboratory by A2LA.
Traceable® is a registered trademark of Control Company.
©2007 Control Company. 92-4470-10 Rev. 1 050307
TRACEABLE
INFRARED
THERMOMETER
GUN
INSTRUCTIONS
®
Loading...