Novametrix VentCheck 101 User Manual

User’s Manual
VentCheck
Model 101
October 30, 1997
Catalog No. 6800-23-02
Copyright ©1997. Novametrix Medical Systems Inc.
5 Technology Drive, Wallingford, Connecticut, U.S.A.
06492. All rights reserved.
About this manual
This manual is written for clinical personnel using the VENT (VentCheck)
Handheld Respiratory Mechanics Monitor, Model 101, and the flow sensors and accessories intended for use with the monitor.
This document contains information which is proprietary and the property of Novametrix Medical Systems Inc., and may not be reproduced, stored in a retrieval system, translated, transcribed, or transmitted, in any form, or by any means, without the prior explicit written permission of Novametrix Medical Systems Inc.
Novametrix reserves the right to change specifications without notice. VentCheck
and the VENT✔ logo are trademarks of Novametrix Medical Systems Inc.
VENT
Per requirements of IEC 601-1, VENT is classified as class II equipment,
internally powered, with type BF applied part, and an enclosure protection rating of IPX1.
Transport/Storage: -10 to +55° C (14-131° F), 10-95% R.H. non-condensing Operating Conditions: +10 to +40° C (50-104° F), 10-90% R.H. non-condensing
The VENT (VentCheck) Handheld Respiratory Mechanics Monitor, Model 101,
contains no user serviceable parts. Refer servicing to qualified service personnel. (See Accessories on page 21 for listings of the Service Manual and Service Test kit for use by technical personnel.)
Technical Description
Declaration of Conformity with European Union Directive
The Authorized Representative for Novametrix equipment is:
D.R.M. Green European Compliance Services Limited, Oakdene House, Oak Road, Watchfiled Swindon, Wilts SN6 8TD United Kingdom
Manual Revision History
31-Jul-97 Release at Rev. 00. 17-Oct-97 Rev. 01. Including software release version 1.1. 30-Oct-97 Rev. 02. Including software release 1.2.
2 VENT✔ Users Manual Rev. 02
Table of Contents
General Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Indications for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Contra-indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Patient Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Preparation for Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Getting Acquainted . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Keypad Controls and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Power supply options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
External Battery Charger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Features Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
Monitor Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Power On/Off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Patient Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
Patient Disconnect . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Screen Display Flowchart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
DATA Display Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
GRAPH Display Screens . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
System Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
Parameter Alerts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Battery Alert . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Cleaning and Disinfecting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Parameter List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Rev. 02 VENT Users Manual 3
4 VENT✔ Users Manual Rev. 02
General Description
The VENT (VentCheck) Handheld Respiratory Mechanics Monitor, Model 101,
from Novametrix Medical Systems Inc., is an easy-to-operate stand-alone respiratory mechanics monitor. It is intended to be used in patient monitoring environments by technically skilled clinical personnel.
Indications for Use
Contra­indications
VENT✔ is indicated for use by technically skilled clinical personnel as a mechanics
monitor for mechanically ventilated or spontaneously breathing persons in monitoring environments including the intensive care unit (ICU), the operating room (OR) and in emergency/transport care. The device is intended for short­term use (nominally 1 hour) and is NOT intended for continuous monitoring.
VENT✔ compatible flow sensors are a single patient use devices having fittings
that comply with ISO 5356-1 and may be used in conjunction with endotracheal tubes, face masks, and breathing circuit devices that also comply with the specification.
Pediatric/adult flow sensors are indicated for use with endotracheal tubes with
internal diameters of 6 mm. Neonatal flow sensors are indicated for use with endotracheal tubes with internal diameters of ≤4 mm. Further, for endotracheal
tubes with internal diameters of 4.5 to 5.5 mm, in general, neonatal sensors are indicated if the patient weight <10 kg or tidal volume <100 ml, and pediatric/
adult sensors are indicated where patient weight 10 kg and tidal volume ≥100
ml.
VENT✔ is intended for short-term use (nominally 1 hour) and is NOT intended as
a continuous monitoring device.
VENT✔ does NOT compensate for the presence of high oxygen concentrations,
anesthetic gases or helium-oxygen mixtures. When such mixtures are present, errors in flow measurements may exist.
Principles of Operation
Rev. 02 VENT Users Manual 5
VENT measures the respiratory parameters of flow and pressure. Many other
parameters are calculated from the flow and pressure signals, including: volume
(tidal and minute), airway resistance, and lung compliance. VENT uses fixed-
orifice, differential pressure pneumotachs. The pneumotach is the airway flow sensor that converts volumetric flow rate into a differential pressure signal. Gas flowing through the flow sensor causes a small pressure drop across the two ports connected to the sensor. This pressure drop is transmitted through the tubing to the monitor where differential and absolute pressure transducers are located.
This pressure drop is correlated to flow according to the factory stored calibration. (The relationship between the volumetric flow-rate and the resulting differential pressure is governed by standard fluid mechanics equations.) User calibration is not required because of the ability of the plastic injection mold to produce precision flow sensors. Empirical calibration for the specific pneumotach type (pediatric/adult, neonatal, etc.) is performed at the factory and is stored in the monitor. The pressure transducer is automatically "zeroed" to correct for changes in ambient temperature, electronics and position (gravity effects).
Patient Safety
Warnings
 Keep the VENT and its accessories clean. Do not operate the VENT when
it is wet due to spills or condensation.
 Connect only Novametrix Series 3 Flow Sensors to the VENT✔. For
maximum performance; keep the pressure sensor ports oriented upward, and keep the sensor clear of moisture and secretions by proper breathing circuit maintenance.
 Connect the sensor first to the VENT and then to the patient breathing
circuit in order to limit circuit volume loss and to avoid excessive moisture build-up in the flow sensor tubing.
VENThas electrically isolated inputs. Patient leakage current flowing from
the instrument to ground is limited to less than 50 µA at 120 V, 60 Hz. Patient isolation is greater than 10 MΩ, 2500 V rms at 60 Hz.
 Where electromagnetic devices (i.e., electrocautery) are used, patient
monitoring may be interrupted due to electromagnetic interference. Electromagnetic fields up to 3 V/m will not adversely affect system performance.
VENT contains no user serviceable parts. Refer servicing to qualified
service personnel. (See Accessories on page 21 for listings of the Service Manual and Service Test kit for use by technical personnel.)
 This product and its accessories which have patient contact are latex free.
WARNING:
Indicates a potentially harmful co ndition that ca n lead to personal injury.
!
Explosion Hazard: Do NOT use the VENT in the presence of flammable
anesthetics. Use of this instrument in such an environment may present an explosion hazard.
Electrical Shock Hazard: Always turn the monitor off before cleaning it.
Do NOT use a damaged monitor or sensor. Refer servicing to qualified service personnel.
Fire Hazard: The VENT should not be exposed to elevated oxygen levels
at elevated pressures. Use in such an environment may present a fire hazard.
Failure of Operation: If the monitor fails to respond as described, do not
use it until the situation has been corrected by qualified personnel.
 Do not apply tension to the sensor tubing while connected to a patient
breathing circuit, as accidental extubation may result.
 Do not position the flow sensors tubing in any manner that may cause
entanglement or strangulation.
 Use the optional external battery charger in non-patient areas only.
6 VENT✔ Users Manual Rev. 02
Cautions
Notes
CAUTION:
Indicates a condition that may lead to equipment damage or malfunction.
!
 Federal (U.S.A.) law restricts this device to sale, distribution, or use by or
on the order of a licensed medical practitioner.
 Electrical Shock Hazard: Always turn the monitor off before cleaning. Do
NOT use a damaged monitor.
 Do NOT use a damaged flow sensor.
 Do NOT immerse the monitor or sensors in liquids.
 Do NOT sterilize the monitor or the sensors.
 No user serviceable parts inside. Refer servicing to qualified service
personnel.
 Operate at temperatures between +10° C to +40° C (50-104° F), < 90%
relative humidity (non-condensing).
 Avoid storing the monitor at temperatures less than -20° C or greater than
+55° C (<-4° F or >131° F).
NOTE
Indicates points of particular interest or emphasis and intended to provide for more efficient or convenient operation.
 The VENT operates with Novametrix Series 3 Flow Sensors only.  The VENT✔ performs an automatic zero (self calibration) periodically and as
needed. During this time, monitoring is interrupted for less than three seconds.
 The automatic zero can be manually initiated by simultaneously pressing
the DATA and GRAPH keys. After changing the sensor from Adult to
Neonatal (while the VENT is operational), wait 30 seconds then perform
an automatic zero.
 This product and its accessories which have patient contact are free of latex.
 The C
 To determine the VENT software version, turn the monitor on. During the
self test performed at power up, the software level is shown on the third line as main-101-xx, where xx is the software version.
 Some VENT monitors were produced with the statement Use only
Novametrix approved devices, 13 VDC, 1A located on the label at the Flow Sensor Input Connector. This erroneous statement does NOT apply to the
VENT✔ monitor and should be ignored.
/C Compliance Ratio (neonatal) parameter is not supported.
20
Rev. 02 VENT Users Manual 7
Preparation for Use
Getting Acquainted
The VENT Handheld Respiratory Mechanics Monitor, Model 101 is shown below.
Front View Top View
Keypad &
Display
Connect Series 3
Flow Sensor
Flow Sensors
t
n
e
i
t
a
P
o
T
r
o
t
a
l
i
t
n
e
V
o
T
Neonatal Pediatric/Adult
Sensor
#6718/6720
Sensor
#6717
Side/Bottom View
Features Connector
Quick Guide
Battery Access
Compartment
8 VENT Users Manual Rev. 02
Keypad Controls and Indicators
1 2 3
4
POWER GRAPH DATA
A
# Key Action Function
POWER
1 Press Turns VENT on/off.
GRAPH
2Press
DAT A
3Press
Press Turns display backlight on/off.
4
Press & Hold
Display Graph Screens.
Additional presses causes VENT✔ to sequence
through available Graph Screens.
Display Data Screens.
Additional presses causes VENT✔ to sequence
through available Data Screens.
Adjusts contrast/viewing angle of display (1 step/sec.)
NEO
B
Illuminated if powered from battery
ANone
Green; battery charged Yellow flashing slowly; capacity getting low Red flashing quickly; exhausted in 10-15 min.
NEO
BNone
Illuminates when a Neonatal sensor is connected.
Symbols
Symbol Description
Patient Isolation
Identifies patient isolation connection as type BF.
Attention
!
Rev. 02 VENT Users Manual 9
Consult manual for detailed information.
Power supply options
The VENT can be powered from seven AA disposable alkaline batteries, or a
rechargeable battery. Battery capacity is shown in the chart below; it may be reduced in colder temperatures or with power cycling, and slightly increased if
VENT✔ is operated with its backlight off.
Standard
AA alkaline batteries
(7 ea. - disposable)
Power Source
Capacity, continuous 8 hours 12 hours
Recharge Time n/a 1.5 hours
Catalog Number
AA Alkaline Batteries. To power VENT✔ from AA alkaline batteries, insert seven
disposable alkaline batteries (Panasonic AM3X or equivalent) into the Battery Case (Cat. No. 6862-00) following the polarity markings on the case.
400038 (battery)
6862-00 (case)
Rechargeable battery,
Optional
(NiMH 7.2 vdc)
400043
WARNING:
Do not recharge or incinerate alkaline batteries. Attempting to do so m ay
!
cause the batteries to leak or explode.
Battery Installation
Rechargeable Battery. VENT✔ can operate from the optional Duracell® DR30
(or equivalent) rechargeable battery. If a rechargeable battery has not been used for three months or more, recharge it before use.
WARNING:
Battery can explode, leak or catch on fi re if heated or exposed to fire or high
!
temperatures.
Additional Battery Information.
 Dispose of batteries in accordance with local laws.
 Do not mix battery types (e.g. disposable and rechargeable AA batteries).
VENT may not power up if the batteries are nearly depleted.
To install or remove the battery, grasp the finger grips on each end of the VENT
battery cover. Squeeze together and pull so that the hinged cover opens. The battery is keyed and can be installed in only one orientation (see illustration inside
10 VENT✔ Users Manual Rev. 02
battery compartment). The contacts should go in first and be located toward the top left of the monitor. Close the battery cover before operating the monitor.
Connect only Novametrix approved devices
Swing battery
Hinged Battery Compartment Cover
Finger Grips
cover open
External Battery Charger
Features Connector
An optional external charger for the Duracell® DR30 is available. The external
charger allows the battery to be recharged outside of the VENT✔ monitor.
 The external charger is for use with the rechargeable Duracell DR-30 battery
only.
 In a non-patient area, connect the external charger to an AC source. Remove
the battery from the VENT and insert it into the external charger. The
battery will be fully charged in approximately 1 1/2 hours.
 Refer to the instructions supplied with the charger for additional information.
WARNING:
The external battery charger should NOT be operated near or in close
!
proximity to patients and/or other medical equipment in operation.
The VENT features connector is for use only with future optional Novametrix
accessories. The connector meets the patient safety requirements set forth by the
safety agencies as labelled on the VENT monitor.
Rev. 02 VENT✔ Users Manual 11
Monitor Operation
Prior to operating the VENT monitor, please note the following;
VENT is intended for short-term use (nominally 1 hour) and is NOT
intended as a continuous monitoring device.
VENT does NOT compensate for the presence of high oxygen
concentrations, anesthetic gases or helium-oxygen mixtures. When such mixtures are present, errors in flow measurements may exist.
Power On/Off
Flow Sensor
 The
The instructions below assume the VENT has a charged battery installed and
that no flow sensor is connected. Use of equipment other than mentioned here may yield different results.
1. Press the POWER key to turn
2. Verify both the NEO and battery
3. Press and hold the LAMP key to
4. If desired, press the LAMP key to
5. Verify INSERT FLOW SENSOR
6. Select a Series 3 Flow Sensor based on the patient type.
VENT
performs an automatic zero (self calibration) periodically and as
needed. During this time, monitoring is interrupted for less than three seconds.
VENT on (or off).
indicators illuminate as part of the self-test performed at power up.
adjust the contrast/viewing angle of the display.
POWER GRAPH DATA
turn the display backlight on/off.
message appears.
 #6717, Pediatric/Adult sensor adds approximately 6.5 cc of deadspace.  #6718/6720, Neonatal sensors add approximately 0.8 cc of deadspace.
NEO
ET Tube Size Series 3 Flow Sensor
6 mm  Pediatric/Adult
 Pediatric/Adult if weight ≥10
4.5 to 5.5 mm
4 mm Neonatal
kg and V
 Neonatal if weight <10 kg or
V
<100 ml
T
100 ml
T
Neonatal
Sensor
#6718/6720
o
T
Pediatric/Adult
Sensor
#6717
t
n
e
i
t
a
P
o
T
r
o
t
a
l
i
t
n
e
V
12 VENT✔ Users Manual Rev. 02
7. Remove the sensor from the package. Verify the sensor is clean, dry and physically intact with no broken edges or disconnected tubing.
8. Connect the Flow sensor to the VENT✔.  Verify VENT✔ briefly displays the selected
sensor type.
 Verify the NEO indicator is illuminated only if a
neonatal sensor is in use.
Patient Connection
9. After connecting the flow sensor to the VENT, place the flow sensor into
the breathing circuit proximal to the patient.
NOTE
Connect the sensor first to the monitor and then to the patient br eathing circuit in order to limit circuit volume loss and to avoid excessive moisture build-up in the flow sensor tubing.
 The sensor is best placed between the endotracheal tube and the ventilator
circuit wye.
 Do NOT position the sensor between the ET tube and the elbow (pediatric/
adult circuit), as this may allow patient secretions to build up in the sensor.
 If other airway appliances (e.g., suction adapter, elbow, heat moisture
exchanger (HME), CO more distal positionaway from the patientcloser to the circuit wye.
 To prevent rain-out and moisture from draining into the sensor, do NOT
place the sensor in a gravity dependent position where moisture and secretions may accumulate.
 Periodically check the flow sensor and tubing for excessive moisture and/or
secretion build-up.
 After changing the sensor from Adult to Neonatal (while the VENT is
operational), wait 30 seconds then perform an automatic zero by simultaneously pressing the DATA and GRAPH keys.
adapter, etc.) are in use, place the flow sensor in the
2
To
VENT
To ventilator
Point tubing
upward
To
VENT
Pediatric/Adult
Flow Sensor
To patient
Rev. 02 VENT✔ Users Manual 13
To patient
Point tubing
upward
To ventilator
Neonatal
Flow Sensor
Patient Disconnect
VENT✔ does NOT provide audible alarms and will not alert the user to inadvertent
disconnection of the flow sensor from either the breathing circuit of the monitors sensor input connector.
NOTE
To limit circuit volume loss and avoid excessive moisture build-up in the flow sensor tubing, disconnect the flow sensor from the breathing circuit before disconnecting the flow sensor from the monitor.
Screen Display Flowchart
10. Press the GRAPH or DATA keys and the most recently viewed data or graph
screen appears in the display. Additional key presses cause VENT✔ to
sequence through the available data screens as shown below.
1
GRAPH
2
GRAPH
3
GRAPH
1
DATA
2
DATA
3
DATA
4
GRAPH
5
GRAPH
4
DATA
5
DATA
14 VENT Users Manual Rev. 02
DATA
DATA Display Screens
Each of the data screens is shown below.
1
DATA
2
DATA
3
DATA
VOLUME DATA
 Spon - spontaneous breath,
Mech - mechanical breath.
 Values from rolling 8-breath aver-
age.
 Activity bar is breath-by-breath
using Volume Graph scale.
PRESSURE DATA
 Breath-by-breath reporting.  If auto-PEEP is detected, the word
auto appears under PEEP and the display reflects total PEEP present.
 Activity bar is breath-by-breath
using Pressure Graph scale.
TIMING DATA
 Breath-by-breath reporting.  Activity bar is breath-by-breath
using Flow Graph scale.
FLOW/VOLUME DATA
4
DATA
 Breath-by-breath reporting.
MECHANICS DATA
5
DATA
 Breath-by-breath reporting.  RSBI - Pediatric/Adult sensor in use,
blank when Neonatal sensor in use.
 Activity bar is breath-by-breath
using Flow Graph scale.
Rev. 02 VENT✔ Users Manual 15
GRAPH
GRAPH Display Screens
Each of the graph screens is shown below.
1
GRAPH
2
GRAPH
3
GRAPH
VOLUME GRAPH
 Automatically scaled.  Breath-by-breath reporting.
PRESSURE GRAPH
 Automatically scaled.  Breath-by-breath reporting.
FLOW GRAPH
 Automatically scaled.  Breath-by-breath reporting.
System Errors
FLOW/VOLUME GRAPH
4
GRAPH
 Automatically
scaled.
 4 breaths overlayed
then cleared.
 Breath-by-breath.
PRESSURE/VOLUME GRAPH
5
GRAPH
 Automatically
scaled.
 4 breaths overlayed
then cleared.  Breath-by-breath.  RSBI - Pediatric/
Adult sensor in use, blank when
Neonatal sensor in use.
If VENT✔ detects that a system error has occurred,
patient monitoring is interrupted and an error message is displayed. Correcting the error condition will clear the error message from the display and monitoring will resume. The
SERVICE
messages cannot be cleared from the
CALL
display without power cycling the monitor.
F
V
V
P
16 VENT✔ Users Manual Rev. 02
The following errors are possible;
Error Message
CALL SERVICE Faulty monitor
Possible Cause
or sensor.
INVALID SENSOR TYPE Incorrect sensor
type connected.
Damaged sensor.
Damaged monitor.
FLOW OUT OF RANGE Flow rate
greater than 180 L/min.
PRESSURE > 120 CM H2O Pressures
>120 cm H
2
were detected.
O
Corrective Action
Remove monitor and flow sensor from use. Contact qualified service personnel and note number shown in triangle.
Replace sensor with a valid sensor type, pediatric/adult or neonatal.
Connector is damaged. Replace sensor.
Monitors sensor connector is damaged. Remove from use and contact qualified service personnel.
VENT✔ does not operate with flow
rates in excess of 180 L/min.
Do not expose monitor/sensor to pressures in excess of 120 cm H
O.
2
Parameter Alerts Battery Alert
Cleaning and Disinfecting
VENT does not provide audible or visible parameter alerts.
VENT monitors, and indicates to the
user, the remaining battery capacity. The battery indicator illuminates
GREEN
to indicate a charged battery, slowly flashes getting low, and quickly flashes
YELLOW to indicate capacity is
RED to
indicate the battery will become exhausted within 10-15 minutes. When the battery indicator flashes
BATTERY appears in the title stripe
LOW
RED, VERY
POWER GRAPH DATA
area of the display (example at right).
 Turn VENT off before cleaning. Do not immerse the monitor. Do not
attempt to sterilize the monitor.
 The monitor can be cleaned and disinfected with solutions such as isopropyl
alcohol 70% solution, 2% glutaraldehyde solution, or bleach 10% solution, then wiped with a clean water dampened cloth to rinse and dry before use.
 Treat Series 3 Flow Sensor in accordance with established protocol for
single-patient use items.
NEO
Rev. 02 VENT✔ Users Manual 17
P arameter List
Parameters reported by the VENT✔ are summarized below.
Abv. Units Parameter
Cdyn ml/cmH
I:E Inspiratory:Expiratory Ratio
MAP cmH
MV Liters Minute Volume
NIP cmH
PEEP cmH
PEF LPM (L/min) Peak Expiratory Flow
PIF LPM (L/min) Peak Inspiratory Flow
PIP cmH
Plat cmH
Raw cmH2O/L/sec Airway Resistance
RR br/min Respiratory Rate
RSBI br/min/L Rapid Shallow Breathing Index
Te sec Expiratory Time
Ti sec Inspiratory Time
O Dynamic Compliance
2
O Mean Airway Pressure
2
O Negative Inspiratory Pressure
2
O Positive End Expiratory Pressure
2
O Peak Inspiratory Pressure
2
O Plateau Pressure
2
VT ml Tidal Volume (Averaged)
VTi ml Tidal Volume Inspiratory
VTe ml Tidal Volume Expiratory
18 VENT✔ Users Manual Rev. 02
Specifications
Below are specifications for the Novametrix VENT Handheld Respiratory
Mechanics Monitor, Model 101. These specifications are listed for informational purposes only, and are subject to change without notice.
Principle of Operation
Flow Sensors
Fixed orifice differential pressure pneumotach. No calibration required.
Series 3 Flow Sensors, Single Patient Use
ET Tube Size Series 3 Flow Sensor Catalog No.
6 mm Pediatric/Adult 6717
4.5-5.5 mm
4 mm Neonatal 6718/6720
Flow:
 Range: Adult Sensor: ± 2 to 180 L/min.
Range: Neonatal Sensor: ± 0.25 to 35 L/min.
Accuracy:
Greater of ± 0.5 L/min or 5% of reading (pediatric/adult sensor). Greater of ± 0.125 L/min or 5% of reading (neonatal sensor).
Pressure:
 Range: 120 to 120 cmH
Accuracy: Greater of ± 0.5 cmH
Respiratory Rate
 Range: 0 to 120 br/min
Accuracy: ±1 br/min
Resolution: 1 br/min
Pediatric/Adult if weight ≥10 kg and V Neonatal if weight <10 kg or V
O
2
O or 2% of reading.
2
<100 ml
T
100 ml
T
6717
6718/6720
Sensor Type Identification: Automatic.
Automatic Barometric Pressure Compensation: 400-800 mmHg.
Physical
Display
Rev. 02 VENT Users Manual 19
VENT monitor
 Size (maximum height, width, depth): 20 × 9.5 × 4.8 cm (8 × 3.75 × 1.875 in.).
 Weight (typical):
without battery: 370 grams, (13 oz.), with alkaline battery pack: 595 grams (21 oz.), with rechargeable battery: 710 grams (25 oz.).
 Operating Environment:
+10 to +40° C (50-104° F), 10 to 90% relative humidity non-condensing.
 Transport/Storage Environment:
-10 to +55° C (14-131° F), 10 to 95% relative humidity non-condensing.
Liquid Crystal Display
 128 × 64 pixels, 2.5 × 1.25 inches.
 Adjustable contrast with backlight and plastic shield.
Indicators
LED indicators
 Remaining battery capacity (green, yellow, red).  Neonatal flow sensor connected (green).
Alerts
AA Alkaline batteries
Rechargeable Battery:
External Battery Charger:
Alerts for measured patient parameters: None. Very low battery alert: audible and visual.
Panasonic AM3X or equivalent, (7 required).
 Type: AA, Alkaline, 1.5 vdc, 1500 mAH.
 Continuous battery life: 8 hours minimum at 21 °C (70 °F).  Up to 85% capacity remaining after 4 years storage at 21 °C (70 °F).
 Recharge: Disposable, do not recharge.
Duracell® DR30 or equivalent
 Type: DR30, Nickel-Metal-Hydride (NiMH), 7.2 vdc, 2400 mAH, >250 cycles.
 Continuous battery life: 12 hours minimum at 21 °C (70 °F).  Up to 70% capacity remaining after 30 days storage at 21 °C (70 °F).  Charge Conditions: 0 to 45 °C (32 to 113 °F).  Discharge Conditions: -20 to 50 °C (-4 to 122 °F).
 Recharge time: 1.5 hr with external charger.
Charger compatible with Duracell® DR30 or equivalent battery
 Storage Conditions: -40 to +70 °C (-40 to 158 °F).  Operating Conditions: 0 to 50 °C (32 to 122 °F); optimum +25°C (77 °F).
 Recharge time: 1.5 hour typical
20 VENT✔ Users Manual Rev. 02
Accessories
Cat. No. Description
Monitor (English language)
6800-00 6800-23 6800-90
420037 420038 420039
Flow Sensors
6717-00 Series 3, Pediatric/Adult Flow Sensor. 10/Box
6718-00 Series 3, Neonatal Flow Sensor. 10/Box
6720-00 Series 3, Neonatal Combined CO
Power Options
6862-00 Battery Case for AA Batteries (batteries not included)
400038 Alkaline Battery, 1.5 vdc, AA Size
400043 Rechargeable Battery, NiMH, 7.2V, 2.4AH,
VENT✔ Handheld Respiratory Mechanics Monitor, Model 101
User's Manual Service Manual
VENT✔ Quick Guide Inservice Video
VHS video tape, NTSC format (U.S.A.) VHS video tape, PAL format VHS video tape, SECAM format
/Flow Sensor. 10/Box
2
(Panasonic AM3X or equivalent)
(Duracell
® DR30 or equivalent)
400049 External Battery Charger (DR30), Universal Voltage, NiMH
(power line cord not included)
600026 Power Line Cord, 120 vac, (U.S.A.) for External Battery Charger
Miscellaneous
140084 Pole/Shelf Mount Kit
315107 Carrying Case
Rev. 02 VENT✔ Users Manual 21
Warranty
Equipment manufactured or distributed by Novametrix Medical Systems Inc., is fully guaranteed, covering materials and workmanship, for a period of one year from the date of shipment, except for certain disposable products and products with stated guarantees other than one year. Novametrix reserves the right to perform guarantee service(s) at its factory, at an authorized repair station, or at the customers installation.
Novametrix obligations under this guarantee are limited to repairs, or at Novametrix option, replacement of any defective parts of our equipment, except fuses, batteries, and calibration gasses, without charge, if said defects occur during normal service.
Claims for damages during shipment must be filed promptly with the transportation company. All correspondence concerning the equipment must specify both the model name and number, and the serial number as it appears on the equipment.
Improper use, mishandling, tampering with, or operation of the equipment without following specific operating instructions will void this guarantee and release Novametrix from any further guarantee obligations.
Service Department
For factory repair service:
Call toll free: 1-800-243-3444
To Call Direct: (203) 265-7701
Facsimile (203) 284-0753
http://www.novametrix.com
techline@novametrix.com
Caution: Federal (U.S.A.) law restricts this device to sale, distribution, or use by or on the order of a licensed medical practitioner.
Copyright 1997. Novametrix Medical Systems Inc.,
5 Technology Drive, Wallingford, Connecticut, 06492.
All rights reserved.
22 VENT Users Manual Rev. 02
Loading...