WarrantyWe guarantee the meter described for 3 years from the date of purchase.
The use of advanced technology and the high quality standard of our meters
are the result of a continuous development. This may result in differences
between this operating manual and your meter. Also, we cannot guarantee
that there are absolutely no errors in this manual. Therefore, we are sure you
will understand that we cannot accept any legal claims resulting from the
data, figures or descriptions.
The meter warranty covers manufacturing faults that are discovered within
the warranty period.
The warranty does not cover components that are replaced during
maintenance work, e. g. batteries.
The warranty claim extends to restoring the meter to readiness for use but
not, however, to any further claim for damages. Improper handling or
unauthorized opening of the meter invalidates any warranty claim.
To ascertain the warranty liability, return the meter and proof of purchase
together with the date of purchase freight paid or prepaid.
The compact M 545P precision pH meter enables you to perform pH
measurements rapidly and reliably. The M 545P provides the
maximum degree of operating comfort, reliability and measuring
certainty for all applications.
The proven calibration procedures and special AutoRead function (AR)
support your work with the pH meter.
pH
6
9
9
2
2
Pinnacle series
4
M
8
5
4
°
5
C
TP
P
3
2
1Keypad
2Display
3Socket field
1
ba75564e0302/2007
7
OverviewM 545P
STO
1.1Keypad
Pinnacle series
_
_>:
CAL
RCL
Switch the measuring instrument on/off
Call up calibration data
Select measured parameter
Open setting menu for calibration and
measurement
Call up calibration procedure
Call up initialization
MODEAR
STO
In this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<OK
_>).
<On/Off>:
<On/
Off
_>:
<MODE>:
MODE
CAL
<MODE
>:
<CAL>:
<CAL
<AR>:
_>:
AR
STO
RCL
8
<AR
<STO>:
_>:
<STO
<RCL>:
<RCL
_>:
<▲>: Increment values, scroll
Activate / deactivate AutoRead
Set the interval for data transmission
Store measured value
Set the interval for automatic storing
processes
Display / transmit measured values and
calibration records
Erase stored measured values
ba75564e0302/2007
M 545POverview
<▼>: Decrement values, scroll
Function display
indicators
<OK>:
_>:
<OK
1.2Display
SET
1
Time
Day.Month
Year
AutoCal
LoBat
DIN
SETDuring calibration: number of the buffer set
BUFFERDuring calibration: number of the buffer
U pH
8
8
Baud
No.
Ident
88
Auto
BUFFER
8
8
Store
Confirm entries
Open setting menu for system settings
Status display
Sensor symbol
8
8
CalError
ARConCal
%
mV/pH
°F
°
C
RCL
TP
Measured value
display
Function and
temperature display
indicators
AutoCalCalibration with automatic buffer recognition
AutoStoreAutomatic storing is switched on
CalErrorAn error occurred during calibration
LoBatWith battery operation: batteries almost empty
ARAutoRead is active
RCLMemory display / memory download
TPTemperature measurement active
ba75564e0302/2007
9
OverviewM 545P
1.3Socket field
Connectors:
1pH electrode
2Reference electrode
3Temperature sensor
4Power pack
5RS 232 interface
Caution
Only connect sensors to the meter that cannot return any voltages
or currents that are not allowed (> SELV and > current circuit with
current limiting).
Almost all customary sensors fulfill these conditions.
10
ba75564e0302/2007
M 545PSafety
2Safety
This operating manual contains basic instructions that you must follow
during the commissioning, operation and maintenance of the meter.
Consequently, all responsible personnel must read this operating
manual before working with the meter.
The operating manual must always be available within the
vicinity of the meter.
Target groupThe meter was developed for work in the laboratory.
Thus, we assume that, as a result of their professional training and
experience, the operators will know the necessary safety precautions
to take when handling chemicals.
Safety instructionsSafety instructions in this operating manual are indicated by the
warning symbol (triangle) in the left column. The signal word (e.g.
"Caution") indicates the level of danger:
Further notes
Warning
indicates instructions that must be followed precisely in order to
avoid possibly great dangers to personnel.
Caution
indicates instructions that must be followed precisely in order to
avoid the possibility of slight injuries or damage to the meter or
the environment.
Note
indicates notes that draw your attention to special features.
Note
indicates cross-references to other documents, e.g. operating
manuals.
ba75564e0302/2007
11
SafetyM 545P
2.1Authorized use
This meter is authorized exclusively for pH and ORP measurements in
the laboratory.
The technical specifications as given in chapter 7 T
ECHNICALDATA must
be observed. Only the operation and running of the meter according to
the instructions given in this operating manual is authorized.
Any other use is considered unauthorized.
2.2General safety instructions
This meter is constructed and tested in compliance with the IEC 1010
safety regulations for electronic measuring instruments.
It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the meter can only be
guaranteed if the generally applicable safety measures and the specific
safety instructions in this operating manual are followed during
operation.
The smooth functioning and operational safety of the meter can only be
guaranteed under the environmental conditions that are specified in
chapter 7 T
ECHNICALDATA.
If the meter was transported from a cold environment to a warm
environment, the formation of condensate can lead to the faulty
functioning of the meter. In this event, wait until the temperature of the
meter reaches room temperature before putting the meter back into
operation.
Caution
The meter is only allowed to be opened by authorized personnel.
12
ba75564e0302/2007
M 545PSafety
Safe operationIf safe operation is no longer possible, the meter must be taken out of
service and secured against inadvertent operation!
Safe operation is no longer possible if the meter:
z has been damaged in transport
z has been stored under adverse conditions for a lengthy period of
time
z is visibly damaged
z no longer operates as described in this manual.
If you are in any doubt, please contact the supplier of the meter.
Obligations of the
purchaser
The purchaser of this meter must ensure that the following laws and
guidelines are observed when using dangerous substances:
z EEC directives for protective labor legislation
z National protective labor legislation
z Safety regulations
z Safety datasheets of the chemical manufacturers.
ba75564e0302/2007
13
SafetyM 545P
14
ba75564e0302/2007
M 545PCommissioning
3Commissioning
3.1Scope of delivery
z M 545P laboratory meter
z Power pack
z 4 batteries 1.5 V Mignon type AA
z Stand with stand base
z Operating manual
3.2Initial commissioning
Perform the following activities:
z Switch on the meter
Switching on the meter
Setting the date and
time
Connecting the power
pack
z Set the date and time
z Connect the power pack (for line power operation only).
1Switch on the meter with <On/Off>.
A display test is briefly displayed.
2See section 4.7.1
Caution
The line voltage at the operating site must lie within the input
voltage range of the original power pack (see chapter 7 T
DATA).
ECHNICAL
Caution
Use original power packs only (see chapter 7 TECHNICALDATA).
ba75564e0302/2007
3Insert the plug into the socket of the pH meter.
4Connect the original power pack to an easily accessible power
outlet.
Note
You can carry out measurements without the power pack.
15
CommissioningM 545P
16
ba75564e0302/2007
M 545POperation
4Operation
4.1Switching on the meter
1Place the meter on a flat surface and protect it from intense
light and heat.
2Press the <On/Off>
A display test is briefly displayed.
Subsequently, the meter switches to the measuring mode
(measured value display).
Note
The meter has an energy saving feature to avoid unnecessary battery
depletion during battery operation.
The energy saving feature switches off the meter if no key was pressed
during the specified interval (setting the switch-off interval see section
4.7.1).
The energy saving feature is not active:
z if the power is supplied by the power pack
z if the AutoStore function is active (see section 4.5.2)
z if a PC is connected (with communication cable to RS232 interface)
z if the printer cable is connected (for external printers).
key.
ba75564e0302/2007
17
OperationM 545P
4.2General operating principles
This section contains basic information on the operation of the M 545P.
4.2.1Operating modes
The meter has the following operating modes:
z Measurement
The display indicates the measurement data in the measured value
display
z Calibration
The display guides you thru a calibration procedure with calibration
information
z Storage in memory
The meter stores measuring data automatically or manually
z Data transmission
The meter transmits measuring data and calibration records to a
serial interface automatically or manually.
z Configuration
The system menu or a sensor menu with submenus, settings and
functions is displayed
4.2.2Operation
KeysThe meter is operated via keys. The keys can have different functions
with long or short keystrokes.
FunctionsGenerally, with a short keystroke a function is carried out. A long
keystroke opens a setting menu.
In a setting menu, settings are selected with the <▲><▼> keys.
A setting is confirmed with <OK>. With confirming, the setting is
finished and the next setting is displayed.
RepresentationIn this operating manual, keys are indicated by brackets <..> .
The key symbol (e.g. <OK>) generally indicates a short keystroke
(under 2 sec) in this operating manual. A long keystroke (approx.
2 sec) is indicated by the underscore behind the key symbol (e.g.
<OK
_>).
18
ba75564e0302/2007
M 545POperation
4.3Measuring
Preparatory activitiesPerform the following preparatory activities when you want to measure:
1Connect an electrode to the meter.
2Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.
3Calibrate or check the meter with the electrode.
4Select the measured parameter with <MODE>.
Note
Incorrect calibration of pH electrodes leads to incorrect measured
values. Calibrate regularly before measuring.
Caution
When connecting an earthed PC/printer, measurements cannot be
performed in earthed media as incorrect values would result.
The RS232 interface is not galvanically isolated.
Temperature sensorYou can measure with or without a temperature sensor. If a
temperature sensor is connected, it is indicated on the display by TP.
Note
The pH meter automatically recognizes the type of the temperature
sensor used. Therefore, you can connect electrodes with an NTC30 or
Pt1000.
The temperature measurement is absolutely essential for a
reproducible pH measurement. If the measurement is made without a
temperature sensor, proceed as follows:
1Measure the current temperature using a thermometer.
2Set the temperature value with <▲><▼>.
Note
When calibrating without temperature sensor, also set the current
temperature of each buffer solution manually with the <▲><▼>
keys.
ba75564e0302/2007
19
OperationM 545P
4.3.1Measuring the pH value
1Perform the preparatory activities according to section 4.3.
2Immerse the pH electrode in the test sample.
3Using <MODE>, scroll as necessary until the measured
parameter pH is displayed.
pH
AutoRead AR
(drift control)
9
4
2
8
TP
°
C
9
6
2
The AutoRead function (drift control) checks the stability of the
measurement signal. The stability has a considerable effect on the
reproducibility of the measured value.
For identical measurement conditions, the following criteria apply:
Drift within 15 sec < 0.02 pH units.
1If necessary, call up the measured variable pH with <MODE>.
2With <AR>, activate the AutoRead function.
The AR function display indicator appears.
The current measured value is frozen (hold function).
3Start measurement with AutoRead with <OK>.
The AR display indicator flashes until a stable value is
measured.
This measured value is downloaded to the interface.
20
4If necessary, start the next measurement with AutoRead with
<OK>.
5To terminate the AutoRead function: press the <MODE> or
<AR>
key.
Note
The current measurement with AutoRead can be terminated at any
time (accepting the current value) by pressing <OK>.
ba75564e0302/2007
M 545POperation
4.3.2Measuring the ORP
The meter can, in conjunction with an ORP electrode, measure the
ORP (mV) of a solution.
1Perform the preparatory activities according to section 4.3.
2Immerse the ORP electrode in the test sample.
3If necessary, call up the measured parameter U with <MODE>.
4Wait for a stable measured value.
U
°
mV
TP
C
4
8
2
Note
ORP electrodes are not calibrated. However, you can check ORP
electrodes using a test solution.
6
1
3
ba75564e0302/2007
21
OperationM 545P
4.4Calibration
Why calibrate?pH electrodes age. This changes the zero point (asymmetry) and slope
of the pH electrode. As a result, an inexact measured value is
displayed. Calibration determines the current values of the zero point
and slope of the electrode and stores them in the meter. Thus, you
should calibrate at regular intervals.
When do you have to
calibrate?
AutoCalis adapted to the permanently programmed buffer solutions as a fully
Buffer sets for
calibration
z After connecting another electrode
z When the sensor symbol flashes on the display:
– after the calibration interval has expired
– after a voltage interruption (empty batteries)
automatic single-point, two-point or three-point calibration. The buffer
solutions are automatically recognized by the meter.
You can use the buffer sets quoted in the table for an automatic
calibration. The pH values are valid for the specified temperature
values. The temperature dependence of the pH values is taken into
consideration during the calibration.
No.Buffer set *pH valuesat
1Buffers according to DIN 19266 and
NIST Retraceable Buffers
1.679
4.006
25 °C
6.865
9.180
12.454
22
2Technical buffers 2.000
4.010
7.000
10.011
3Merck1*4.000
7.000
9.000
4Merck2 *1.000
6.000
8.000
13.000
5Merck3 *4.660
6.880
9.220
ba75564e0302/2007
25 °C
20°C
20°C
20°C
M 545POperation
No.Buffer set *pH valuesat
6DIN 19267 11.090
4.650
6.790
9.230
7Mettler Toledo USA *1.679
4.003
7.002
10.013
8Mettler Toledo TEC *1.995
4.005
7.002
9.208
9Fisher *2.007
4.002
7.004
10.002
10Fluka BS *4.006
6.984
8.957
11Radiometer *1.678
4.005
7.000
9.180
25 °C
25 °C
25 °C
25 °C
25 °C
25 °C
12Baker *4.006
25 °C
6.991
10.008
13Metrohm *3.996
25 °C
7.003
8.999
14Beckman *4.005
25 °C
7.005
10.013
15Hamilton Duracal *4.005
25 °C
7.002
10.013
16Precisa *3.996
25 °C
7.003
8.999
* Brand names or trade names are trademarks of their respective owners
protected by law (see page 69).
ba75564e0302/2007
23
Loading...
+ 52 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.