Repair Instructions – Service No./Modi, Error list
1Service and Access Codes/Release Option
Enter the number under “service diagnosis” (Evita) or ventilation (Evita 2 dura) during
operation. For release Hardware Option wait 30 seconds and switch Evita off and on.
Release Software Option:
Refer to section of "Repair Instructions - Service No./ Modi, Error list" chapter 3.5.2,
Diagnosis page “Option Release” (release of software options), page 76.
1.1Service Numbers for Software Versions 1.00 to 1.04 and 1.08
3 0 3 2to change configuration of ventilation
1 3 1 1to clear user logbook
4 6 5 5to select Service Mode
4 6 4 4
+
4 6 5 5
to select Service Mode with info text in DS error list
1 3 2 2to clear DS error list
4 7 4 7to select calibration of PEEP valve
4 7 4 8 to clear PEEP valve calibration data
4 8 1 1to activate DC module
4 8 1 0to deactivate DC module
6 7 9 5to activate CO
8 3 9 6to activate SpO
module
2
module
2
9 4 9 5to deactivate all options
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 1
1.2Service Numbers for Evita 4 and Evita 2 dura
Evita 4 with Software 1.05, 1.06, 1.07, 1.09 and 2.n and higher; for Evita 2 dura as of
Software 3.00
3 0 3 2To change configuration of ventilation.
1 3 9 9 + 1 3 1 1 To clear user logbook.
3 9 9 9 + 3 9 4 8Deactivation of leakage compensation (re-activated by
switching the Evita off/on).
3 9 9 9 + 3 9 5 8Deactivation of BTPS conversion (re-activated after
switching the Evita off/on). Important: This is only
possible if P Ambient is larger than 960 mbar,
otherwise the Evita will perform a re-start (see Test
Certificate).
7 2 9 9 + 7 2 3 5 + 0 0 0 0Deactivation of the hose compliance correction (re-
activated after switching the Evita off/on).
4 6 5 5To select Service Mode. With Evita 2 dura units without
“Service Plus” option, connect a download cable
(service coding) to COM1.
4 6 4 4 + 4 6 5 5To select Service Mode with info text in DS error
list.(Evita 4 only). With Evita 2 dura units, software
version 3.00 or higher, without “Service Plus” option,
connect a download cable (service coding) to COM1.
1 3 9 9 + 1 3 2 2To clear DS error list
4 7 9 9 + 4 7 4 7To select calibration of PEEP valve
4 7 9 9 + 4 7 4 8 To clear PEEP valve calibration data
4 8 1 1To activate DC module
4 8 9 9 + 4 8 1 0To deactivate DC module
6 7 9 5To activate CO
6 7 9 9 + 6 7 9 4To deactivate CO
8 3 9 6To activate SpO
8 3 9 9 + 8 3 9 5To deactivate SpO
module
2
module
2
module
2
module
2
For internal use only. Copyright reserved.
R5664500T025664800.fm
9 4 9 9 + 9 4 9 5To deactivate all options
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 2
2List of Evita 2 dura/Evita 4 error numbers
Error list as at February 1998 (corresponding to SW 3.n of Evita 4/Evita 2 Dura)
Overview of error number assignment
02Error recognised by safety software which runs on CPU 68332 PCB,
Pneumatic Controller PCB and Graphics Controller PCB (Evita 4) or Front
Panel PCB (Evita 2 dura)
03Error in control console components not located on printed circuit card of
Graphics Controller PCB (Evita 4) or Front Panel PCB (Evita 2 dura), e.g.
Touchscreen
04Error in electronic unit components not located on printed circuit cards, e.g.
power pack
05Error in pneumatic system components
06Error in extension box components
07Software error in Rosy operating system which runs on CPU 68332 PCB,
Pneumatic Controller PCB, Graphics Controller PCB (Evita 4) or Front Panel
PCB (Evita 2 dura) and Communication PCB.
08Exception
09For future expansion
10Error on Pneumatic Controller PCB
11Error on HPSV Controller (Air) PCB or in Air mixer cartridge. On the Evita 4,
such errors are stored with software version 1.09 and as of version 2.21; on the
Evita 2 dura with versions greater than 3.00. With older software versions, such
errors are only displayed as "mixer fault". The error status can be read out in
diagnosis mode if the error is currently present.
12Error on HPSV Controller (O2) PCB or in O2 mixer cartridge. On the Evita 4,
such errors are stored with software version 1.09 and as of version 2.21; on the
Evita 2 dura with versions greater than 3.00. With older software versions, such
errors are only displayed as "mixer fault". The error status can be read out in
diagnosis mode if the error is currently present.
13Error on CPU 68332 PCB
14Error on CO2 Carrier PCB
15Error on Communication PCB
16Error on Paediatric Flow PCB
17Error on Graphics Controller PCB (Evita 4) or Front Panel PCB (Evita 2 dura)
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 3
2.1Error code 02 Safety Software
The service codes 02.00.. to 02.22.. are recognised on the CPU 68332 PCB; they
correspond to 02.40.. to 02.62.. as recognised on the Pneumatic Controller PCB.
FaultItemCause/Remedy
02.00.001si0_Err, New start, cause
incorrect, reset
02.00.002si0_Err, Fast data chksum
error, reset
02.00.003si0_Err, Safety software of
partner CPU not started, reset
02.00.004si0_Err, Input data, chksum
error, reset
02.00.005si0_Err, Vector for NMI
incorrect, reset
Monitoring of cold/warm start storage in
RAM on CPU 68332 PCB. Cause:
Software error on CPU 68332 PCB
Data packets between CPU 68332
PCB and Pneumatic Controller PCB
cannot be transferred in full. CPU
68332 PCB or Pneumatic Controller
PCB defective.
After starting, it was not possible within
10 seconds to establish synchronisation
between CPU 68332 PCB and
Pneumatic Controller PCB. Cause:
CPU 68332 PCB or Pneumatic
Controller PCB defective.
Data packets between CPU 68332
PCB and Pneumatic Controller PCB
cannot be transferred in full. CPU
68332 PCB or Pneumatic Controller
PCB defective.
RAM or SW error on CPU 68332 PCB
For internal use only. Copyright reserved.
R5664500T025664800.fm
02.00.020si0_Err, Si boot, case
incorrect, reset
5664.500/800 Evita 4/Evita 2 dura01.2002
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
Repair Instructions
– 4
FaultItemCause/Remedy
02.01.001si1_ Err, Adjuster has incorrect
pixel sum, X alarm, reset:
Screen content (with Evita 2 dura also 7segment displays) is read out and
compared to request. Cause: Hardware
or SW error on Graphics Controller
PCB (Evita 4)/Front Panel PCB
(Evita 2 dura) or CPU 68332 PCB or
Pneumatic Controller PCB. If display
plausible, then SW error; otherwise
high probability of hardware error on
Graphics Controller PCB (Evita 4) or
Front Panel PCB (Evita 2 dura). Table
for X:
1Paw high adjuster selected
2 Paw high arrow selected
3 MV min adjuster selected
4 MV min arrow selected
5 O2 adjuster selected
6 Last adjuster plus one
7 Paw high adjuster not selected
8Paw high arrow not selected
9MV min adjuster not selected
10 MV min arrow not selected
11 O2 adjuster not selected
12 Paw high line 30 top
13 Pressure scale values top 30
14 Paw high line 60 top
15 Pressure scale values top 60
16 Paw high line 100 top
17 Pressure scale values top 100
18 Paw high line 30 bottom
19 Pressure scale values bottom 30
20 Paw high line 60 bottom
21 Pressure scale values bottom 60
22 Paw high line 100 bottom
23 Pressure scale values bottom 100
24 Paw high line 30 bargraph
25 Pressure scale values bargraph 30
26 Paw high line 60 bargraph
27 Pressure scale values bargraph 60
28 Paw high line 100 bargraph
29 Pressure scale values bargraph
100
30 Paw high display 1
31 Paw high display 2
32 MV min display 1
33 MV min display 2
34 Cold start screen
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 5
FaultItemCause/Remedy
02.01.002si1_Err, Paw high, Si copy,
consistency X alarm, reset
02.01.003si1_Err, MV min, Si copy,
consistency X alarm, reset
02.01.004si1_Err, O2 adjust, Si copy,
consistency X alarm, reset
02 .01.005si1 _Er r, Paw high, invalid value,
reset
02.01.006si1_Err, MV min, invalid value,
reset
Alarm values relevant to safety are
checked for plausibility.
Cause: RAM or SW error on CPU
68332 PCB or Pneumatic Controller
PCB.
Alarm values relevant to safety are
checked for plausibility.
Cause: RAM or SW error on CPU
68332 PCB or Pneumatic Controller
PCB.
Alarm values relevant to safety are
checked for plausibility.
Cause: RAM or SW error on CPU
68332 PCB or Pneumatic Controller
PCB.
Alarm values relevant to safety are
checked for plausibility.
Cause: RAM or SW error on CPU
68332 PCB or Pneumatic Controller
PCB.
Alarm values relevant to safety are
checked for plausibility.
Cause: RAM or SW error on CPU
68332 PCB or Pneumatic Controller
PCB.
Screen content (with Evita 2 dura also 7segment displays) is read out and
compared to request. Cause: Hardware
or SW error on Graphics Controller
PCB (Evita 4) / Front Panel PCB (Evita
2 dura) or CPU 68332 PCB or
Pneumatic Controller PCB. If display
plausible, then SW error; otherwise
high probability of hardware error on
Graphics Controller PCB (Evita 4) or
Front Panel PCB (Evita 2 dura). Refer to
error 02.01.001 for X table.
Refer to error 02.02.001.
Refer to error 02.02.001.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
SW or transmission error from
Graphics Controller PCB (Evita 4) or
Front Panel PCB (Evita 2 dura) via CPU
68332 PCB to Pneumatic Controller
PCB. Cause: Probably error on
Graphics Controller PCB (Evita 4) or
Front Panel PCB (Evita 2 dura).
02.05.004si5_Err, Factor for NTPS to
BTPS incorrect, X alarm, reset
02.05.020si5_Err, Si boot, case
incorrect, reset
02.06.001si6_Err, MV deactivation,
confirmation X signal incorrect,
reset
HPSV actuation monitoring signals
error. Cause: HPSV cartridge (Air or
O2) or one of the two HPSV actuation
PCBs or Pneumatic Controller PCB
defective. Error status can be read out in
diagnosis mode “Valves“ if error is
present at the time.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
SW error or impermissible activation of
BTPS deactivation at ambient pressure
less than 961 mbar.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
Probably software error on Graphics
Controller PCB (Evita 4) or Front Panel
PCB (Evita 2 dura) or also CAN
communication problem with CPU
68332 PCB and Pneumatic Controller
PCB.
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 8
FaultItemCause/Remedy
02.06.002si6_Err, MV deactivation,
request Y signal incorrect,
reset
02.06.003si6_Err, Paed MV deactivation,
confirmation X signal incorrect,
reset
02.06.004si6_Err, Paed MV deactivation,
request Y signal incorrect,
reset
02.06.020si6_Err, Si boot, case
incorrect, reset
02.06.021si6_Err, Si Y signal, case
incorrect
02.07.020si7_Err, Si boot, case
incorrect, reset
02.08.001si8_Err, O2 measurement
deactivation, confirmation X
signal incorrect, reset
Refer to error 02.06.001.
Refer to error 02.06.001.
Refer to error 02.06.001.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
Refer to error 02.06.020, however
defective Graphics Controller PCB
(Evita 4) or Front Panel PCB (Evita 2
dura) also possible.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
Probably software error on Graphics
Controller PCB (Evita 4) or Front Panel
PCB (Evita 2 dura) or also CAN
communication problem with CPU
68332 PCB and Pneumatic Controller
PCB.
02.08.002si8_Err, O2 measurement
deactivation, request Y signal
incorrect, reset
02.08.020si8_Err ,Si boot, case
incorrect, reset
02.08.021si8_Err, Si Y signal, case
incorrect, reset
02.09.001si9_Err, Disconnection
manoeuvre, disconnection too
long, reset
02.09.002si9_Err, Disconnection
manoeuvre, post-oxygenisation
too long, reset
02.09.003si9_Err, Disconnection
manoeuvre, pre-oxygenisation
too long, reset
Refer to error 02.08.001.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
Refer to error 02.08.021, however
defective Graphics Controller PCB
(Evita 4) or Front Panel PCB (Evita 2
dura) also possible.
CPU 68332 PCB monitors timing of
Pneumatic Controller PCB. One of the
two systems is defective.
Refer to error 02.09.001.
Refer to error 02.09.001.
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 9
FaultItemCause/Remedy
02.09.004si9_Err, Disconnection
manoeuvre, no button, reset
02.09.020si9_Err, Si boot, case
incorrect, reset
02.09.021si9_Err, Si status change, case
incorrect, reset
SW error on Graphics Controller PCB
(Evita 4) / Front Panel PCB (Evita 2
dura) or CPU 68332 PCB or
Pneumatic Controller PCB.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
CPU 68332 PCB monitors timing of
Pneumatic Controller PCB. One of the
two systems is defective.
CPU 68332 PCB monitors timing of
Pneumatic Controller PCB. One of the
two systems is defective.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
Inspiration pressure too high for more
than 1 second and cannot be dissipated
via emergency vent Y1.3. Inspiration
pressure is measured on Pneumatic
Controller PCB and CO2 Carrier PCB.
Possible causes:
− Measurement of inspiration
pressure on Pneumatic Controller
PCB (1st channel) or CO2 Carrier
PCB (2nd channel) defective. Read
out values by way of service mode.
If error is on CO2 Carrier PCB,
contact problems may be the
cause. CO2 Carrier PCB must be
secured in position with both
screws.
− Inspiration pressure sensor
defective.
− Pneumatic fault.
− Too high a measured value from
the expiratory pressure sensor
(Note: If ground (GND) is missing,
the Paw sensors will measure
approx. 115 mbar.)
RAM or SW error probably on Graphics
Controller PCB (Evita 4) or Front Panel
PCB (Evita 2 dura), but may also be on
CPU 68332 PCB or Pneumatic
Controller PCB.
RAM or SW error probably on Graphics
Controller PCB (Evita 4) or Front Panel
PCB (Evita 2 dura), but may also be on
CPU 68332 PCB or Pneumatic
Controller PCB.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB, but may
also be on Graphics Controller PCB
(Evita 4) or Front Panel PCB (Evita 2
dura).
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB or
Graphics Controller PCB (Evita 4) /
Front Panel PCB (Evita 2 dura).
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 11
FaultItemCause/Remedy
02.18.002si18_Err, Silence too long,
reset
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB or
Graphics Controller PCB (Evita 4) /
Front Panel PCB (Evita 2 dura).
02.18.003si18_Err, Confirmation signal
for standby incorrect, reset
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB or
Graphics Controller PCB (Evita 4) /
Front Panel PCB (Evita 2 dura).
02.18.020si18_Err, Si boot, case
incorrect, reset
02.19.001si19_Err, Memory consistency,
reset
02.19.002si19_Err, Auxiliary alarm mute,
number 1 incorrect, reset
02.19.003si19_Err, Auxiliary alarm mute,
number 2 incorrect, reset
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
RAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
02.21.000sic_Err, Task recognises errorRAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
02.22.000sil_Err, Task recognises errorRAM or SW error on CPU 68332 PCB
or Pneumatic Controller PCB.
The service codes 02.40.. to 02.62.. are recognised on the Pneumatic Controller PCB; they
correspond to 02.00.. to 02.22.. as recognised on the CPU 68332 PCB.
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 12
Defective Hardware
FaultItemCause/Remedy
02.71.001No acoustic alarmActuation signal from CPU 68332 PCB to
loudspeaker (approx. 50 ohms) is tapped via
series resistor (5 ohms) on Graphics
Controller / Front Panel PCB and supplied to
the system as digital information. This signal is
used for desired/actual comparison of the
loudspeaker actuation. The loudspeaker is in
the control console/front section.
Possible cause of trouble:
-Software error
if error occurs between several aspiration
modes without intervening ventilation. This
error has been rectified on the Evita 4 as of
SW 2.21 (Japan as of SW 1.09) and on the
Evita 2 dura with versions greater than 3.00.
-Cable fault between CPU 68332 PCB and
loudspeaker in control console
-Loudspeaker defect
Trouble-shooting:
-Measure resistance of loudspeaker via
connecting cable of control console, incl.
cable between pins 9 and 10 of SUB-D cable
connector. Test value = 55 plus/minus 8
ohms. This value incl. shunt resistance must
be 5 ohms higher than loudspeaker value,
which can be measured after opening control
console.
-Using diagnosis mode with lowest volume
setting, Refer to section of "Repair
Instructions - Service No./ Modi, Error list"
chapter 3.2.1, Diagnosis page
“Microprocessor” of “Front”, page 57.
-Using external DS mode via PC with lowest
volume setting, Refer to section of "Repair
Instructions - Service No./ Modi, Error list"
chapter 9.4.7, Loudspeaker test, page 141. or
Refer to section of "Repair Instructions Service No./ Modi, Error list" chapter 9.5.7,
Testing of loudspeaker, page 151.
instructions.
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Note: See also error s 02.71.012 and
02.71.013
Repair Instructions
– 13
02.71.002Flow measurement on
Pneumatic Controller
PCB defective
Remedy:
As of SW 1.05: Renew Pneumatic Controller
PCB.
Up to SW 1.04: Fault is set if ambient
pressure is less than 961 mbar on switch-on,
but is not displayed on screen. Only renew
Pneumatic Controller PCB if error message is
displayed on screen during operation.
02.71.003Gold Cap capacitor,
capacitance too low
In the event of power failure, Gold Cap on
CO2 Carrier PCB supplies piezo alarm tone
generator with voltage and thus signals mains
failure. The duration of this mains failure alarm
is governed by the capacitance of the Gold
Cap. Both the function and the capacitance
are therefore tested once an hour, starting 5
minutes after switch-on of the unit. A unit fault
is signalled if one of the following criteria is
met:
a. Final charging voltage too high
(greater than 11 V), voltage regulator
defective
b. Final charging voltage too low (less
than 8 V), charging circuit defective
c. Reference voltage too high (greater
than 6.3 V), discharge circuit
defective
d. Reference voltage too low (less than
4.5 V), no Gold Cap
e. Capacitance less than 30mF, Gold
Cap capacitance too low
For internal use only. Copyright reserved.
R5664500T025664800.fm
02.71.004Gold Cap capacitor,
voltage too high
02.71.005Gold Cap capacitor,
voltage too low
02.71.006Gold Cap capacitor,
tester defective
5664.500/800 Evita 4/Evita 2 dura01.2002
Remedy:
refer to section of "Repair Instructions Electronic Components" Chapter 5.5,
Conversions, page 34 CO2 Carrier PCB.
see error 02.71.003
see error 02.71.003
see error 02.71.003
Repair Instructions
– 14
02.71.007Boot test defectiveCause of trouble:
CPU 68332 PCB (more likely) or Pneumatic
Controller PCB defective
Note:
Additionally displayed with other error
messages
In addition to signalling mains failure, the
piezo assumes the alarm function of the
loudspeaker if CPU 68332 is defective
(stopped). To ensure that this is the case, the
piezo on the CO2 Carrier PCB is suppressed
by the CPU 68332 PCB with a square-wave
signal. If this cyclical clocking fails, the
associated muting of the piezo is cancelled
and a continuous beep sounds. Muting is
deliberately cancelled to test the piezo. The
current through the piezo, which then
increases as a result of cancelling the muting,
can be measured as voltage drop across a
series resistor by the A/D converter on the
CO2 Carrier PCB.
Possible cause of trouble:
a. CO2 Carrier PCB defective
b. CPU 68332 PCB defective
Trouble-shooting:
-With diagnosis mode, Refer to section of
"Repair Instructions - Service No./ Modi, Error
list" chapter 3.3.2, Diagnosis page
“Microprocessor” of “Electronics”, page 63.
-With external DS mode via PC, Refer to
section of "Repair Instructions - Service No./
Modi, Error list" chapter 9.4.6, Horn test, page
140.
instructions
For internal use only. Copyright reserved.
R5664500T025664800.fm
02.71.009Piezo alarm generator,
current too high
02.71.010Piezo alarm generator,
current too low
02.71.011Monitoring of nebuliser
valve defective
5664.500/800 Evita 4/Evita 2 dura01.2002
refer to error 02.71.008
refer to error 02.71.008
The required nebuliser valve position is
constantly compared to the checkback signal
of the actuation electronics.
Remedy: Renew Pneumatic Controller PCB
Note: There is no recognition of whether or
not valve is connected.
Repair Instructions
– 15
02.71.012Loudspeaker too seldom
recognised
The actuation signal from the CPU 68332
PCB to the loudspeaker (approx. 50 ohms) is
tapped via a series resistor (5 ohms) on the
Graphics Controller PCB and supplied to the
system as digital information. This signal is
used for desired/actual comparison of the
loudspeaker actuation. A unit fault is signalled
if an error is established in the course of this
desired/actual comparison. The loudspeaker
is located in the control console/front section.
Possible cause:
a. Cable fault between CO2 Carrier
PCB and loudspeaker in control
console
b. Loudspeaker defective
c. Graphics Controller PCB defective
d. CPU 68332 PCB defective
Trouble-shooting:
- Measure resistance of loudspeaker
via connecting cable of control
console incl. cable between pins 9
and 10 of SUB-D cable connector.
Test value = 55 plus/minus 8 ohms.
This value incl. shunt resistance must
be 5 ohms higher than loudspeaker
value, which can be measured after
opening control console.
- Using diagnosis mode with lowest
volume setting, Refer to section of "Repair
Instructions - Service No./ Modi, Error list"
chapter 3.2.1, Diagnosis page
“Microprocessor” of “Front”, page 57.
141. or Refer to section of "Repair
Instructions - Service No./ Modi, Error list"
chapter 9.5.7, Testing of loudspeaker, page
151.
instructions.
Note: Refer also to error 02.71.001
For internal use only. Copyright reserved.
R5664500T025664800.fm
02.71.013Loudspeaker recognised
too frequently
5664.500/800 Evita 4/Evita 2 dura01.2002
refer to error 02.71.012
Repair Instructions
– 16
02.71.014+ 15 V too lowThe operating voltages accessible via an A/D
converter on the CO2 Carrier PCB are
cyclically compared to the defined tolerances.
An undershoot of + 14 V produces a unit fault
message.
Possible cause:
a. Power pack defective (source)
b. CO2 Carrier PCB defective
Check voltages from power pack, various
possibilities:
-Measure voltage with multimeter,
refer to section of "Repair Instructions -
Electronic Components" Chapter 1, Power
Supply Unit, page 1
-With diagnosis mode, Refer to section of
"Repair Instructions - Service No./ Modi, Error
list" chapter 3.2.1, Diagnosis page
“Microprocessor” of “Front”, page 57.
-With external service mode, Refer to
section of "Repair Instructions - Service No./
Modi, Error list" chapter 9.4.1, Testing
voltages, page 131.
If OK = Renew CO2 Carrier PCB
If not OK = Renew power pack
02.71.015+10V reference voltage
too low
02.71.016AD conversion for O2
measurement defective
This voltage is stepped down from the +15 V
on the CO2 Carrier PCB, where it is also
measured. An undershoot of + 9 V produces
this unit fault message.
Remedy:
Renew CO2 Carrier PCB
The pre-amplified O2 sensor voltage is read in
on the Pneumatic Controller PCB by two
independent AD converters. These values
differ.
Trouble-shooting:
-With diagnosis mode, Refer to section of
"Repair Instructions - Service No./ Modi, Error
list" chapter 3.4.2, Diagnosis page “Sensors”
of “Pneumatics”, page 70.
-With external service mode, Refer to
section of "Repair Instructions - Service No./
Modi, Error list" chapter 9.3.4, Sensors, page
The required O2/Air switching valve position
is constantly compared to the checkback
signal of the actuation electronics on the
Pneumatic Controller PCB.
Trouble-shooting:
-With diagnosis mode, Refer to section of
"Repair Instructions - Service No./ Modi, Error
list" chapter 3.4.1, Diagnosis page “Valves” of
“Pneumatics”, page 67.
-With external service mode, Refer to
section of "Repair Instructions - Service No./
Modi, Error list" chapter 9.3.1, Valve switching,
page 121.
Remedy:
Renew Pneumatic Controller PCB
Note:
There is no recognition of whether or not valve
is connected
02.71.018Incorrect nebuliser gasDuring nebulisation, the required air/O2
switching valve position is thus constantly
compared to the checkback signal of the
actuation electronics.
Trouble-shooting:
-With diagnosis mode, Refer to section of
"Repair Instructions - Service No./ Modi, Error
list" chapter 3.4.1, Diagnosis page “Valves” of
“Pneumatics”, page 67.
-With external service mode, Refer to
section of "Repair Instructions - Service No./
Modi, Error list" chapter 9.3.1, Valve switching,
page 121.
Remedy:
Renew Pneumatic Controller PCB
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 18
02.71.019Cold/warm start
recognition defective
Cold/warm start recognition is effected by
way of mains-switch auxiliary contacts in
power pack, which are read in via CO2
Carrier PCB.
Trouble-shooting:
- With diagnosis mode, Refer to section of
"Repair Instructions - Service No./ Modi, Error
list" chapter 3.3.2, Diagnosis page
“Microprocessor” of “Electronics”, page 63.
- With external service mode, Refer to section
of "Repair Instructions - Service No./ Modi,
Error list" chapter 9.4.2, Power pack status,
page 133.
Possible cause of trouble:
- Mains-switch auxiliary contacts in power
pack defective
The data for initialising the interrupt masks
have not been properly stored on the CPU
68332 PCB.
Cause:
Fault in one of the following printed circuit
cards (RAM or SW error):
a. CPU 68332 PCB
b. Pneumatic Controller PCB
c. Graphics Controller / Front Panel
PCB
02.71.021Quartz time discrepancyQuartz times of CPU 68332 PCB and
Pneumatic Contoller PCB are compared and
an error message generated if the two differ
Cause:
Defect in one of the following printed circuit
cards:
a. CPU 68332 PCB
b. Pneumatic Controller PCB
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 19
2.2Error Code 03 Control-Console Components
FaultItemCause/Remedy
03.01.001Touchscreen is not recognised
during BOOT
Following switch-on of Evita 4,
touchscreen is not recognised by
Graphics Controller PCB. The
touchscreen communicates with the
Graphics Controller PCB via RS232.
Trouble-shooting:
- With diagnosis mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 3.2.1,
Diagnosis page “Microprocessor” of
“Front”, page 57.
- With external service mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 9.5.5,
Testing of touchscreen, page 148.
Defect in one of the following printed
circuit cards:
a. Touchscreen
b. Graphics Controller PCB
03.01.002TOUCHSCREEN NO RESET ACK
similar to error 03.01.001, but occurs
during operation
03.01.003OUCHSCREEN NO REPORTsimilar to error 03.01.001, but occurs
during operation
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 20
03.07.001Reset line faultReason:
Interruption in reset line between CPU
68332 PCB and Graphics Controller /
Front Panel PCB.
Possible cause:
a. Control console / Front Panel PCB
not connected
b. Fault in cable to control console /
Front Panel PCB
c. CPU 68332 PCB defective
d. Graphics Controller / Front Panel
PCB defective
e. Connector between printed circuit
cards defective
Trouble-shooting:
1. Check and if necessary renew front-
section cable
2. Remove housing cover to reveal
service LEDs on CPU 68332 PCB. If
necessary, insert card in one of the
upper slots (all identical).
3. Connect up control console ready for
operation.
4. Switch on unit.
5. LED 6 must light briefly after
switching on:
If not = CPU 68332 PCB defective
If yes = CPU 68332 PCB probably
OK, renew Graphics Controller /
Front Panel PCB
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 21
2.3Error Code 04 Components of Electronics Unit
FaultItemCause/Remedy
04.01.001Power pack fault, 5V supply
voltage greater than 6 V
The operating voltages accessible via an
A/D converter on the CO2 Carrier PCB
are cyclically compared to the defined
tolerances of plus/minus 20 %.
Possible cause:
a. Power pack defective (source)
b. CO2 Carrier PCB defective
(monitoring)
Trouble-shooting:
Check voltages from power pack,
various possibilities:
-Measure voltage with multimeter,
refer to section of "Repair Instructions Electronic Components" Chapter 1,
Power Supply Unit, page 1
instructions
-With diagnosis mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 3.3.2,
Diagnosis page “Microprocessor” of
“Electronics”, page 63.
- With external service mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 9.4.1,
Testing voltages, page 131.
If OK = Renew CO2 Carrier PCB
If not OK = Renew power pack
Important:
001Power pack +5V too high
(for digital components)
002Power pack +5 V too low
003Power pack +12V too high
(for HPSVs, fan in power pack)
004Power pack +12V too low
005Power pack +15V too high
(for analog components)
006Power pack +15V too low
007Power pack -15V too high
(for analog components)
008Power pack -15V too low
009Power pack +24V too high
(for valves and control console)
010Power pack +24V too low
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 22
04.01.002Power pack fault, 5V supply
voltage less than 4 V
refer to error 04.01.001
04.01.003Power pack fault, 12V supply
voltage greater than 14.4 V
04.01.004Power pack fault, 12V supply
voltage less than 9.6 V
04.01.005Power pack fault, 15V supply
voltage greater than 18 V
04.01.006Power pack fault, 15V supply
voltage less than 12 V
04.01.007Power pack fault, -15V supply
voltage greater than -18 V
04.01.008Power pack fault, -15V supply
voltage less than -12 V
04.01.009Power pack fault, 24V supply
voltage greater than 28.8 V
04.01.010Power pack fault, 24V supply
voltage less than 19.2 V
refer to error 04.01.001
refer to error 04.01.001
refer to error 04.01.001
refer to error 04.01.001
refer to error 04.01.001
refer to error 04.01.001
refer to error 04.01.001
refer to error 04.01.001
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 23
2.4Error Code 05 Pneumatic System Components
FaultItemCause/Remedy
05.01.001Short circuit in fan in
pneumatic system
Fan for cooling pneumatic-system
printed circuit cards defective or not
connected
Recognition:
a. Actuation current of fan is measured
with an A/D converter via a shunt. In
normal operation, this fan current
must be subject to fluctuations
greater than 50 mV. The minimum
and maximum levels of this fan
voltage are measured for 60 s each.
If the max./min. difference is less
than 50 mV, this indicates a fault.
b. Measured value of fan voltage = 0 V
(open circuit)
c. Measured value of fan voltage
greater than or equal to 2.9 V (short
circuit)
Trouble-shooting:
- With diagnosis mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 3.4.3,
Diagnosis page “Microprocessor” of
“Pneumatics”, page 74.
- With external service mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 9.3.5,
Voltages, page 127.
- Fan connected?
-Fan operation?
No = Renew fan
Yes = Renew Pneumatic Controller
PCB
For internal use only. Copyright reserved.
R5664500T025664800.fm
05.01.002Fan in pneumatic system not
connected
05.01.003No operation of fan in
pneumatic system
5664.500/800 Evita 4/Evita 2 dura01.2002
refer to error 05.01.001
refer to error 05.01.001
Repair Instructions
– 24
05.02.001Reset-up line faultInterruption in reset line between CPU
68332 PCB and Pneumatic Controller
PCB.
Possible cause:
a. CPU 68332 PCB defective
b. Pneumatic Controller PCB defective
c. Connector between printed circuit
cards defective
Trouble-shooting:
1. Remove housing cover to reveal
service LEDs on CPU 68332 PCB.
If necessary, insert card in one of the
upper slots (all identical).
2. Switch on unit.
3. Following LEDs must light briefly
after switching on:
LED 4 (together with 3) for reset-up
LED 5 for reset-down
If not = CPU 68332 PCB defective
If yes = CPU 68332 PCB probably
OK, renew Pneumatic Controller
PCB
05.02.002Reset-down line faultrefer to error 05.02.001
05.02.003Interruption in disable lineInterruption in disable line between
CPU 68332 PCB and Pneumatic
Controller PCB.
Possible cause:
a. CPU 68332 PCB defective
b. Pneumatic Controller PCB defective
c. Connector between printed circuit
cards defective
Trouble-shooting:
1. Removing housing cover to reveal
service LEDs on CPU 68332 PCB.
If necessary, insert card in one of the
upper slots (all identical).
2. Switch on unit.
3. LED 7 must flash/light after switching
on until pneumatic system is activated
(approx. 10 s):
If not = CPU 68332 PCB defective
If yes = CPU 68332 PCB probably
OK, renew Pneumatic Controller
PCB
Note:
Only possible with software versions
1.00 to 1.05, 2.00 and 2.10
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 25
05.04.001O2/Air switching valve cannot
be switched on
The required O2/Air switching valve
position is constantly compared to the
checkback signal of the actuation
electronics
If required position and checkback
signal do not tally, a unit fault message
is generated.
Trouble-shooting:
- With diagnosis mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 3.4.1,
Diagnosis page “Valves” of
“Pneumatics”, page 67.
- With external service mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 9.3.1,
Valve switching, page 121.
Remedy:
Renew Pneumatic Controller PCB
Note:
There is no recognition of whether or
not valve is connected.
05.04.002O2/Air switching valve not
switched off
05.04.003O2/Air switching valve not
switched over
05.04.004Incorrect O2/Air switching
valve position
05.04.005O2/Air switching valve does
not switch to Air
refer to error 05.04.001
refer to error 05.04.001
refer to error 05.04.001
In unit check, a test pressure of 30 mbar
is only built up in tubing system with one
connected pressure supply (O2 or Air)
in each case. The O2/Air switching
valve Y1.1 then effects switching to the
other compressed gas. This causes the
tubing system to be vented to below
10 mbar.
Trouble-shooting:
-With diagnosis mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 3.4.1,
Diagnosis page “Valves” of
“Pneumatics”, page 67.
- With external service mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 9.3.1,
Valve switching, page 121.
Remedy:
Renew valve Y1.1 in gas connection.
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 26
05.04.006O2/Air switching valve does
not switch to O2
refer to error 05.04.005
05.05.001Safety valve cannot be
switched on
Up to SW 1.04 only
Cause: Safety valve or actuation on
Pneumatic Controller PCB defective
05.05.002Safety valve cannot be
switched off
Up to SW 1.04 only
Cause: Safety valve or actuation on
Pneumatic Controller PCB defective
05.05.003Safety valve cannot be
switched
Up to SW 1.04 only
Cause: Safety valve or actuation on
Pneumatic Controller PCB defective
05.05.004Incorrect safety valve positionUp to SW 1.04 only
Cause: Safety valve or actuation on
Pneumatic Controller PCB defective
05.05.005Pressure venting via safety
valve Y1.3 defective
Test pressure of 30 mbar is built up in
patient system during unit check. This is
followed by venting via safety valve Y1.3
to below 10 mbar.
Trouble-shooting:
- With diagnosis mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 3.4.1,
Diagnosis page “Valves” of
“Pneumatics”, page 67.
- With external service mode, Refer to
section of "Repair Instructions - Service
No./ Modi, Error list" chapter 9.3.1,
Valve switching, page 121.
Remedy:
Renew valve Y1.3 in gas connection.
For internal use only. Copyright reserved.
R5664500T025664800.fm
05.05.006Test pressure of 30 mbar
cannot be built up during unit
check
5664.500/800 Evita 4/Evita 2 dura01.2002
Pressure of 30 mbar should be built up
in tubing system during unit check with
test lung connected.
Cause:
-Leak in tubing system
-Test lung not connected
Note:
As of SW 1.04, this fault is only
indicated by an “ F “ in the unit check.
The following applies to all Rosy errors: Operating-system software errors. The operating
system runs on the CPU 68332 PCB, the Pneumatic Controller PCB, the Graphics
Controller PCB (Evita 4)/Front Panel PCB (Evita 2 dura) and the Communication PCB.
Please report any problems to Lübeck.
2.6Error Code 08 Exceptions (SW)
The following applies to all Exception errors: Software problem on CPU 68332 PCB,
Pneumatic Controller PCB, Graphics Controller PCB (Evita 4) / Front Panel PCB (Evita 2
dura) or Communication PCB. Please report any problems to Lübeck.
10.97.xxxROSY SHUTDOWNrefer to: Rosy error codes (07..)
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 28
SW:
FaultItemCause/Remedy
10.99.000PNT UNEXPECTED END
OF_TASK
10.99.020LM UNEXPECTED END OF
TASK
10.99.030DSPT UNEXPECTED END OF
TASK
Software error on Pneumatic Controller
PCB.
Software error on Pneumatic Controller
PCB.
Software error on Pneumatic Controller
PCB.
10.99.040VS INVALID MSG RECEIVEDSoftware error on Pneumatic Controller
PCB.
10.99.050UE INVALID MSG RECEIVEDSoftware error on Pneumatic Controller
PCB.
10.99.060BEAT ILLEGAL MSGSoftware error on Pneumatic Controller
PCB.
10.99.061BEAT ILL MSGSIZESoftware error on Pneumatic Controller
PCB.
10.99.062BEAT ILL BREATH STATESoftware error on Pneumatic Controller
PCB.
10.99.063BEAT ILL VENTILATION
MODE
Software error on Pneumatic Controller
PCB.
10.99.064BEAT ACTIVATE ILL BMODESoftware error on Pneumatic Controller
PCB.
10.99.065BEAT ILL DEACTIVATIONSoftware error on Pneumatic Controller
PCB.
10.99.070bau: BAU Task should never
reach this point
10.99.071BAU ILLEGAL SETTING,
incorrect Ti Te, calculation in
Software error on Pneumatic Controller
PCB.
Software error on Pneumatic Controller
PCB.
parameter adjustment, apnoea
ventilation
10.99.080eba: EBA Task should never
reach this point
Software error on Pneumatic Controller
PCB.
10.99.081EBA ILL MSG IDSoftware error on Pneumatic Controller
PCB.
10.99.082EBA LONG MSGSoftware error on Pneumatic Controller
PCB.
10.99.083EBKB TERMINATE
OPERATION
Software error on Pneumatic Controller
PCB.
For internal use only. Copyright reserved.
R5664500T025664800.fm
5664.500/800 Evita 4/Evita 2 dura01.2002
Repair Instructions
– 29
10.99.084EBKB O2 ILL STATESoftware error on Pneumatic Controller
PCB.
10.99.085EBKB PAED ILL STATESoftware error on Pneumatic Controller
PCB.
10.99.086EBKB ILL BRONCHSTATESoftware error on Pneumatic Controller
PCB.
10.99.087EBKB ILL FLOWSTATE1Software error on Pneumatic Controller
PCB.
10.99.088EBKB ILL FLOWSTATE2Software error on Pneumatic Controller
PCB.
10.99.101BB INVALID MSG RECEIVEDSoftware error on Pneumatic Controller
PCB.
10.99.110MSF INVALID MSG
RECEIVED
Software error on Pneumatic Controller
PCB.
10.99.111MSF, AD VALUES shiftedIncorrect AD conversion on Pneumatic
Controller PCB.
Cause:
Pneumatic Controller PCB defective or
supply voltages not stable. Further
cause: In the case of abnormal mixer
noise on switch-on, an HPSV is
impermissibly loading the minus 15 V.
10.99.112MSF, AD VALUES on ZERORefer to error 10.99.111.
10.99.801EEPROM WRITE QUEUE
FULL
10.99.901DL UNEXPECTED END OF
TASK
10.99.9111ST CD LENGTH CORRUPT,
false central data length for
Software error on Pneumatic Controller
PCB.
Software error on Pneumatic Controller
PCB.
Software error on Pneumatic Controller
PCB.
first object in frame arrived
10.99.9121ST LONG MSG CORRUPT,
start frame data of multi packet
Software error on Pneumatic Controller
PCB.
receive message RPC corrupt
For internal use only. Copyright reserved.
R5664500T025664800.fm
10.99.913CAN ERROR INT, CAN
controller marked ill interrupt
by sending error frames
10.99.914NET QUEUE OVERFLOW,
data overflow in applications
receive queue
5664.500/800 Evita 4/Evita 2 dura01.2002
CAN Controller error on CPU 68332
PCB or Pneumatic Controller PCB or
Graphics Controller PCB
(Evita 4)/Front Panel PCB
(Evita 2 dura). Most likely cause:
Pneumatic Controller PCB.
Software error on Pneumatic Controller
PCB.
Repair Instructions
– 30
Loading...
+ 123 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.