BREATHING SYSTEM
SENSOR INTERFACE PANEL
(VPO ASSEMBLY)
SV45024
1-0Rev. F
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4INTRODUCTION
1.0 Recommendations
Because of the sophisticated nature of Draeger Medical, Inc. anesthesia equipment and its critical
importance in the operating room setting, it is highly recommended that only appropriately
trained and experienced professionals be permitted to service and maintain this equipment.
Please contact DrägerService® at (800) 543-5047 for service of this equipment.
Draeger Medical, Inc. also recommends that its anesthesia equipment be serviced at three-month
intervals. Periodic Manufacturer’s Service Agreements are available for equipment manufactured
by Draeger Medical, Inc. For further information concerning these agreements, please contact us
at (800) 543-5047.
Draeger Medical, Inc. products/material in need of factory repair shall be sent to:
DrägerService
3124 Commerce Drive
Telford, PA 18969
(Include RMA Number)
HOW TO USE THIS MANUAL
The manual is divided into several sections. The DIAGNOSTICS section describes self-test and
service diagnostics for checking the system functions. An understanding of the on-board service
capabilities is necessary before any attempt is made to troubleshoot the unit. The
TROUBLESHOOTING section lists error codes and provides troubleshooting guides to assist the
TSR in locating the source of a problem. The REPLACEMENT PROCEDURES section contains
instructions for removal and replacement of the assemblies that are considered field-replaceable.
The ADJUSTMENT AND CALIBRATION PROCEDURES section contains the field procedures
needed to restore original system specifications. The Periodic Manufacturer’s Service (PMS)
PROCEDURE section outlines the steps required to verify the electrical, mechanical and
pneumatic safety of the unit and also identifies components requiring periodic replacement.
GENERAL TROUBLESHOOTING GUIDELINES
Troubleshooting the Narkomed 4 should always begin by communicating withthose who observed
or experienced a problem with the unit. This may eliminate unnecessary troubleshooting steps.
Once a general problem is identified, refer to the troubleshooting flow charts in Section 3 to
determine the proper corrective action to be taken.
After a component has been replaced, verify that the unit is operating properly by running the
appropriate diagnostic procedure. The PMS PROCEDURE in Section 6 must also be performed
after any component has been replaced.
The general arrangement of the Narkomed 4 Anesthesia System is shown on the opposite page.
WARNINGS are used in this manual before procedures which if not performed correctly could
result in personal injury.
CAUTIONS are used in this manual to alert service personnel to the possibility of damage to the
equipment if a procedure is not performed correctly.
CliniDAS, Datagrip, NAD Information Systems, NAD Logo, Narkomed, O.R. Data Manager,
O.R. Link, ORM, PC Prep/View, Quality Service For Life, Vigilance Audit, Vitalert, Vitalink
and Narkomed GS are registered trademarks of Draeger Medical, Inc. All other products or
name brands are trademarks of their respective owners.
Disclaimer
The content of this manual is furnished for informational use only and is subject to change
without notice. Draeger Medical, Inc. assumes no responsibility or liability for any errors or
inaccuracies that may appear in this manual.
1-2
Rev. M
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4DIAGNOSTICS
2.0 DIAGNOSTICS
The NARKOMED 4 contains a diagnostic system that monitors certain system functions and
records their operational status. A series of tests is performed when the system is powered
up and the results are displayed on the diagnostics screen shown in Figure 2-1. Further
diagnostic functions are available through a series of service screens that can be called up
by a Technical Service Representative at the display panel. The following paragraphs provide
a description of each of the service screens that can be accessed through touch keys on the
display panel. If no display is present upon system power-up, refer to Section 3 of this
manual for troubleshooting assistance.
A PREVENTIVE MAINTENANCE DUE message will appear on the screen if the current
date exceeds the Periodic Manufacturer’s Service due date stored in the machine.
DIAGNOSTICS
VIDEO TESTPASS
FIRMWARE TESTPASS
MEMORY TEST BANK1PASS
MEMORY TEST BANK2PASS
TIMER/INTERRUPT TESTPASS
ANALOG TESTPASS
AUDIO TEST -PRIMARYPASS
NARKOMED 4 SYSTEM SOFTWARE
(C) COPYRIGHT 1988-1995, NAD INC.
VERSION:1.3097 CF
MONITORCHECKOUT
Rev. C
Figure 2-1: POWER-UP DIAGNOSTICS SCREEN
2-1
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
2.1Main Service Screen
2.1.1View Mode
To access the Main Service Screen, press and hold the Selection DIAL, and
simultaneously press the MONITOR and SYSTEM CONFIG keys. The View
mode service screen shown in Figure 2-2 will appear, displaying the machine
serial number, last service date and machine hours accumulated. Version
numbers for each software set in the machine are also displayed.
If the MONITORS service screens are accessed in the View mode, current
values will be displayed but the operator will not be able to perform
calibrations. A VIEW entry will not be logged into the Service Log.
MACHINE SERIAL NUMBER: 00124001
LAST SERVICE DATE: 03-14-96
HOURS RUNNING SINCE LAST SERVICE: 97
TOTAL HOURS RUNNING: 5610
Touch the Service Code box on the screen, then scroll through the following list
of service codes with the Selection DIAL:
PMS: When performing scheduled Periodic Manufacturer’s Service.
INST: For initial installation of the NM4. This selection will establish
the start date for the warranty period.
SRVC: For unscheduled service visits requiring component
replacements or calibrations.
Press the Selection DIAL to enter the service event into the Service Log.
The Technical Service Representative ID boxes will appear on the screen as
shown in Figure 2-3. Touch the first box, scroll to the desired alpha-numeric
with the Selection DIAL, and press the dial to enter the first character of the
ID. Enter the remaining three characters in the same manner.After the ID has
been entered, the RESET key will be displayed.
NOTE: Following any service procedures, return to this screen and touch the
RESET key to enter the current date as the last service date, and to reset the
Hours Running Since Last Service to zero. Touching the RESET key also
places an entry into the service log.
MACHINE SERIAL NUMBER: 00124001
LAST SERVICE DATE: 03-14-96
HOURS RUNNING SINCE LAST SERVICE: 97
TOTAL HOURS RUNNING: 5610
Touch the SERVICE LOG key to display a list of stored events. Figure 2-4
shows a typical service log screen. The codes displayed are classified as Normal
Events, Failed Events, Error Events, Service Events, and Calibration Events.
These are listed in Tables 2-1 thru 2-6.
Up to 1000 events may be logged. Touching the PREV PAGE key will move the
display backward in time; touching the NEXT PAGE key will move the display
forward in time.
SERVICE LOG
DATETIMEPARAMETERCODE DESCRIPTION
07-14-9118:100000FB640000POWER ON
07-14-9118:13JOHNC009RESP FLOW ZERO
07-14-9118:19JOHNC008RESP FLOW SPAN
PREV
PAGE
NEXT
PAGE
PRINT
SERVICE LOG
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
PERIPHERALS
Figure 2-4: SERVICE LOG SCREEN
2-4
Rev. A
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4DIAGNOSTICS (continued)
Table 2-1: NM4 SERVICE LOG NORMAL EVENT CODES
CodeDescription
0000Power on
Table 2-2: NM4 SERVICE LOG FAILED EVENT CODES
CodeDescriptionRecommended Action
F011QPDM Address Register Error
F012QPDM Data Register Error
F013QPDM Vblank Int Timeout Err
F014QPDM VRAM Memory Error
F021PROM CRC Test Error
F031SRAM Memory Test Error
F051MFP Bus Error
F052MFP TimerA Error
F053MFP TimerB Error
F054MFP TimerC Error
F055MFP TimerD Error
F056System Clock Error
F057FIFO Empty Interrupt Error
F058ADC Interrupt Error
F059CNTOVF Interrupt Error
F061ADC Interrupt Error
F06212V to -12V Out of Tol Error
Replace Processor
F0635V to GND Out of Tol Error
F071Pri Spkr Static Test Error
F072Pri Spkr Dynamic Test Error
F073Pri Spkr FIFO Test Error
F074Pri Spkr No Vol Test Error
Rev. L
2-5
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
Table 2-2 (continued): NM4 SERVICE LOG FAILED EVENT CODES
CodeDescriptionRecommended Action
F081Backup Spkr Dynamic Test Err
F082Backup Spkr No Vol Test Err
F091SAO2 Spkr Static Test Error
F092SAO2 Spkr Dynamic Test Error
F093SAO2 Spkr No Vol Test Error
F111Real Time Clock Test Error
F121Bad File Warning
F122Bad Files Error
F123Event Memory Error
F130Alt Processor Fail
F200UART1 Internal Init Fail
F201UART1 Register Readback Fail
F202UART1 Transmit Interrupt Fail
F203UART1 Receive Interrupt Fail
F204UART1 Receive Err Int Fail
F205UART1 Baud Rate Test 1 Fail
F206UART1 Baud Rate Test 2 Fail
F207UART1 Baud Rate Test 3 Fail
F210UART1 Channel 0 Fail
F211UART1 Channel 1 Fail
F212UART1 Channel 2 Fail
F213UART1 Channel 3 Fail
Replace Processor
2-6
Rev. L
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4DIAGNOSTICS (continued)
Table 2-2 (continued): NM4 SERVICE LOG FAILED EVENT CODES
CodeDescriptionRecommended Action
F214UART1 Channel 4 Fail
F215UART1 Channel 5 Fail
F216UART1 Channel 6 Fail
F217UART1 Channel 7 Fail
F300UART2 Internal Init Fail
F301UART2 Register Readbk Fail
F302UART2 Transmit Interrupt Fail
F303UART2 Receive Interrupt Fail
F304UART2 Receive Err Int Fail
F305UART2 Baud Rate Test 1 Fail
F306UART2 Baud Rate Test 2 Fail
F307UART2 Baud Rate Test 3 Fail
F310UART2 Channel 0 Fail
F311UART2 Channel 1 Fail
F312UART2 Channel 2 Fail
F313UART2 Channel 3 Fail
F314UART2 Channel 4 Fail
F315UART2 Channel 5 Fail
F316UART2 Channel 6 Fail
F317UART2 Channel 7 Fail
Replace Processor
Rev. L
2-7
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
Table 2-3: NM4 SERVICE LOG ERROR EVENTS
Touch Panel Events
CodeDescriptionRecommended Action
T001Bad Device RecalibrationClear obstructions. If recurring,
replace touch panel
Head Assembly Events
CodeDescriptionRecommended Action
H001Head Temperature ErrorCall Tech. Service
A001Pattern AdjustedNone
Power Supply Events
CodeDescriptionRecommended Action
P001AC FailRestore AC power
P002AC GoodNone
P003Battery LowRecharge battery. If persistent,
replace battery
P004Battery GoodNone
P005Brick BadReplace Pwr Supply PCB
MSPEC Events
CodeDescriptionRecommended Action
M001Channel Err Active CPU
M002Channel Err Inactive CPU
M003Command Retry
M004Comm Error
M005System Fault Max
Replace MSPEC
M006Versiion ID Retry
M007Span Fail
M008Solenoid Stuck
M009Zero Timeout
M010Zero Fail Max
2-8
Rev. L
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4DIAGNOSTICS (continued)
Table 2-3 (continued): NM4 SERVICE LOG ERROR EVENTS
MSPEC Events (continued)
CodeDescriptionRecommended Action
M011RX Buffer OverflowReplace MSPEC
M012No Continuous DataReplace MSPEC
M013Line Block DiffClear Blockage
M014Line Block Less Than MinClear Blockage
M016Multispec ShutdownReplace MSPEC
M101CD Chan Err
M102Chan Err O2 Info P1
Call Tech. Service
M103Chan Err O2 Info P2
Datagrip Events
DG00Datagrip Internal Error
DG01Datagrip Internal Error
Replace Datagrip
DG02Datagrip Internal Error
Table 2-3 (continued): NM4 SERVICE LOG ERROR EVENTS
NIBP Events
CodeDescriptionRecommended Action
N001NIBP Mon ErrorReplace NIBP
N002NIBP Com ErrorReplace NIBP
N003NIBP Bad DataCall Tech. Service
N004Infl Pres Not GT 30Check for Leaks
N005Error Rcvd from NIBPReplace NIBP
N006NIBP Mode ConflictReplace NIBP
2-8ARev. L
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
Table 2-4: NM4 SERVICE LOG SERVICE EVENT CODES
CodeDescription
S001INST
S002PMS
S003SRVC
S004Secondary Serv Screen Entry
S005Reset Service Date
S006Service Log Clear
Table 2-5: NM4 SERVICE LOG CALIBRATION EVENT CODES
CodeDescriptionCodeDescription
C001O2 Store Zero AC009Respiratory Flow Store Span
C002O2 Store Zero BC010Respiratory Flow Store Zero
C003Line Block Store ReferenceC011Pressure Store Span
C004Gas Analyzer - StartC012Pressure Store Zero
C005Gas Analyzer - ContinueC013NIBP Infl Check
C006NIBP InflateC014NIBP Defl Ck
C007NIBP StopC015NIBP Leak Ck
C008NIBP Take Reading
Rev. L2-8B
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4DIAGNOSTICS (continued)
Table 2-6: NM4 O.R. DATA MANAGER EVENT CODES
CodeDescriptionRecommended Action
V001VTLK Com Error
Replace ORDM if recurring
V002VTLK Com Lost
V003VTLK Com RegainedNone
V010ORDM Startup Failure
Replace ORDM
V011ORDM Inact Startup Failure
Table 2-7: NM4 TEMPLATE EVENT CODES
CodeDescriptionRecommended Action
TEM0Template Defaults Loaded
TEM1Recreating Tmplt Database
TEM2Templates Reinitalized
None
TEM3Templates Reinitalized
TEM4Factory Defaults Installed
2-8CRev. L
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
Touch the PRINT LOG key to obtain a Service Log Printout from the internal
printer. Figure 2-5 shows the format of the printed service log.
Touch the PMS CRITERIA key to bring up the screen shown in Figure 2-6.
Touch the Month window, scroll to the next PMS due month with the Selection
DIAL, and press the dial to enter the month in which the PREVENTIVE
MAINTENANCE DUE message will appear on the power-up screen.
SELECT CRITERIA FOR "PREVENTIVE MAINTENANCE" MESSAGE
MONTH :1991
DEC
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
PERIPHERALS
Figure 2-6: PMS CRITERIA SCREEN
2-9
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
2.4Monitors
Touch the MONITORS key to bring up a second column of touch keys. These
keys select the calibration screens for the system sensors.
2.4.1Oxygen Monitor Screen
The Oxygen Monitor Screen displays current oxygen cell readings and
provides a calibration procedure forthe oxygen sensor circuitry. A touch
key allows entry of a new zero value. See Figure 2-7.
CURRENT CELL A: 110
CURRENT CELL B: 108
ZERO CALIBRATION PROCEDURE:
- REMOVE FUEL CELL FROM SENSOR HOUSING.
- LET CURRENT CELL VALUES STABILIZE.
- TOUCH "STORE ZERO" KEY TO ENTER
CURRENT VALUES AS CALIBRATED VALUES.
- DISCONNECT O2 SENSOR HOUSING CABLE
BEFORE RE-INSTALLING FUEL CELL.
- RE-INSTALL FUEL CELL.
- RE-CONNECT CABLE
STORED CELL A ZERO: 3
STORED CELL B ZERO: 6
STORE
ZERO
OXYGEN
PRESSURE
RESPIRATORY
FLOW
SpO2
NIBP
GAS
ANALYZER
LINE BLOCK
CALIBRATION
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
PERIPHERALS
Figure 2-7: OXYGEN MONITOR SCREEN
2-10
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4DIAGNOSTICS (continued)
2.4.2Breathing Pressure Monitor Screen
The Breathing Pressure Monitor Screen displays current airway
pressure and provides a calibration procedure for the pressure sensing
circuitry. Touch keys allow entry of new zero and span settings. See
Figure 2-8.
CURRENT PRESSURE VALUE : 205
ZERO CALIBRATION PROCEDURE:
- REMOVE PRESSURE SAMPLE LINE FROM
ABSORBER; EXPOSE TO AIR.
- LET CURRENT VALUE STABILIZE.
- TOUCH "STORE ZERO" KEY TO ENTER THE
CURRENT VALUE AS THE CALIBRATION ZERO.
SPAN CALIBRATION PROCEDURE:
- REMOVE PRESSURE SAMPLE LINE FROM
ABSORBER.
- APPLY 50 CM H2O CONSTANT PRESSURE AT
SAMPLE LINE, VERIFIED BY A KNOWN,
CALIBRATED METER.
- LET CURRENT VALUE STABILIZE.
- TOUCH "STORE SPAN" KEY TO ENTER THE
CURRENT VALUE AS THE CALIBRATION SPAN.
STORED ZERO: 244
STORED SPAN: 540
STORE
SPAN
STORE
ZERO
OXYGEN
PRESSURE
RESPIRATORY
FLOW
SpO2
NIBP
GAS
ANALYZER
LINE BLOCK
CALIBRATION
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
PERIPHERALS
Figure 2-8: BREATHING PRESSURE MONITOR SCREEN
2-11
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
2.4.3Respiratory Flow Monitor Screen
The Respiratory Flow Monitor Screen displays the current flow reading
and also shows a calibration procedure for the flow sensing circuitry.
Touch keys are provided for entry of new zero and span settings. See
Figure 2-9.
CURRENT FLOW VALUE : 76
ZERO CALIBRATION PROCEDURE:
- REMOVE SPIROMED SENSOR FROM ABSORBER.
- LET CURRENT VALUE STABILIZE.
- TOUCH "STORE ZERO" KEY TO ENTER THE
CURRENT VALUE AS THE CALIBRATION ZERO.
SPAN CALIBRATION PROCEDURE:
- REMOVE SPIROMED SENSOR FROM ABSORBER.
- ATTACH A GAS SUPPLY AT INPUT OF SENSOR,
AND A CALIBRATED FLOWMETER AT OUTPUT.
- PROVIDE A STEADY 10 L/MIN FLOW THROUGH
SPIROMED SENSOR.
- LET CURRENT VALUE STABILIZE.
- TOUCH "STORE SPAN" KEY TO ENTER THE
CURRENT VALUE AS THE CALIBRATION SPAN.
STORED ZERO: 78
STORED SPAN: 161
STORE
SPAN
STORE
ZERO
OXYGEN
PRESSURE
RESPIRATORY
FLOW
SpO2
NIBP
GAS
ANALYZER
LINE BLOCK
CALIBRATION
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
PERIPHERALS
Figure 2-9: RESPIRATORY FLOW MONITOR SCREEN
2-12
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4DIAGNOSTICS (continued)
2.4.4SpO2 Monitor Screen
The SpO2Screen displays current values for SpO2, Pulse, SpO2Pleth
and SpO
Bargraph. See Figure 2-10.
2
SpO2 VALUE :98
PULSE VALUE :75
PLETH VALUE :125
BARGRAPH VALUE : 12
OXYGEN
PRESSURE
RESPIRATORY
FLOW
SpO2
NIBP
GAS
ANALYZER
LINE BLOCK
CALIBRATION
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
PERIPHERALS
Figure 2-10: SpO2MONITOR SCREEN
2-13
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
2.4.5NIBP Monitor Screen
The NIBP Monitor Screen displays Real Time, Systolic, Diastolic and
Mean blood pressure and Pulse.
The TAKE READING touch key will initiate a single NIBP
measurement cycle.
A procedure for checking NIBP cuff inflation pressure against an
external gauge is also displayed, with touch keys to perform the test.
See Figure 2-11.
Refer to Section 6 of this manual for inflation, deflation, and leak
testing procedures.
REAL TIME PRESSURE : 0
SYSTOLIC:130
DIASTOLIC:78
MEAN:98
PULSE:75
NIBP PRESSURE CALIBRATION VERIFICATION:
- PLACE A "T" IN LINE WITH THE BP CUFF.
- ATTACH A PRESSURE GAUGE AT "T".
- TOUCH THE "INFLATE" KEY.
- NIBP MODULE WILL INFLATE CUFF TO
180 MM Hg.
- VERIFY WITH PRESSURE GAUGE.
- TOUCH THE "STOP" KEY TO END THE TEST.
NIBP INFLATION, DEFLATION, AND LEAK CHECKS:
- REFER TO SERVICE MANUAL FOR PROCEDURES.
INFLATION
CHECK
DEFLATION
CHECK
READING
LEAKAGE
INFLATESTOP
TAKE
CHECK
OXYGEN
PRESSURE
RESPIRATORY
FLOW
SpO2
NIBP
GAS
ANALYZER
LINE BLOCK
CALIBRATION
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
PERIPHERALS
Figure 2-11: NIBP MONITOR SCREEN
2-14
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
NM4DIAGNOSTICS (continued)
2.4.6Gas Analyzer Screen
Current values of CO2,N2O and anesthetic agent are displayed on the
Gas Analyzer Screen along with a span calibration procedure. See
Figure 2-12. (The scrubber bottle replacement instruction is not
displayed on machines having a Model 4610 gas analyzer.)
The MODE window on the gas analyzer screen displays the following
messages:INHIBIT during warm-up
READY after 30 minute warm-up
The STATUS window on the gas analyzer screen displays the following
messages:IDLE before starting calibration
ZERO IN PROGRESS during calibration
ZERO COMPLETE during calibration
SPAN IN PROGRESS during calibration
PASSED or FAILED after calibration
The complete calibration procedure is outlined in Section 5.
CURRENT CO2: *INV*MODE:
CURRENT N2O: 10STATUS:
CURRENT AGENT: *INV*
SPAN CALIBRATION PROCEDURE:
- WAIT FOR MODE = "READY".
- PERFORM GAS ANALYZER ACCURACY TEST.
- IF TEST FAILS, PERFORM SPAN CAL:
- WITH PATIENT CIRCUIT ENABLED:
- TOUCH "START" KEY,
- WAIT FOR STATUS = "ZERO COMPLETE"
- ATTACH SAMPLE LINE TO CALIBRATION BOTTLE.
- TURN FLOW ON BOTTLE TO 250 ML/MIN
- LET THE CURRENT CELL VALUES STABILIZE
- REMOVE THE SAMPLE LINE FROM THE BOTTLE
- TOUCH "CONTINUE" KEY.
- WAIT FOR STATUS = "PASSED" OR "FAILED"
- REPLACE SCRUBBER BOTTLE ANNUALLY.
OXYGEN
PRESSURE
RESPIRATORY
FLOW
SpO2
NIBP
GAS
ANALYZER
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
Figure 2-12: GAS ANALYZER SCREEN
Rev. C
2-15
STARTCONTINUE
LINE BLOCK
CALIBRATION
PERIPHERALS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
DIAGNOSTICS (continued)NM4
2.4.7Line Block Screen
The current sample pressure is displayed on the Line Block screen
along with a calibration procedure and a touch key to enter a setpoint
value. See Figure 2-13. The stored reference value is displayed at the
lower left corner of the screen.
CURRENT SAMPLE PRESSURE:200
LINE BLOCK CALIBRATION PROCEDURE:
- WITH PATIENT CIRCUIT ENABLED:
- ENSURE THAT SAMPLE LINE IS
CONNECTED AND UNOCCLUDED.
- PLACE A FLOWMETER AT THE SAMPLE
EXHAUST.
- ATTACH FLOW RESTRICTOR, FIX 0336, TO
PATIENT SAMPLE LINE.
- WAIT FOR VALID CURRENT SAMPLE PRESSURE.
- ADJUST RESTRICTION IN ORDER TO
OBTAIN A FLOW OF 100 ML/MIN.
- TOUCH "STORE REFERENCE" KEY.
- FULLY OCCLUDE SAMPLE LINE AND OBSERVE
THAT LINE BLOCK ALARM BECOMES
ACTIVE AFTER AT LEAST 15 SECONDS.
- REMOVE FLOW RESTRICTOR FROM SAMPLE LINE.
STORED REFERENCE :200
REFERENCE
STORE
OXYGEN
PRESSURE
RESPIRATORY
FLOW
SpO2
NIBP
GAS
ANALYZER
LINE BLOCK
CALIBRATION
MAIN
SERVICE
LOG
PMS
CRITERIA
MONITORS
PERIPHERALS
Figure 2-13: LINE BLOCK CALIBRATION SCREEN
2-16
Loading...
+ 428 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.