NORTH AMERICAN DRÄGER Drager Narkomed GS User manual

RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Setup and
Installation
Manual
Part Number: 4113852-006
Date: 25 February 2004
© 2004 Draeger Medical, Inc.
Rev: J
Narkomed GS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Contents
Section 1. Introduction
Introduction. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Warnings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Cautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Section 2. Moving the Narkomed GS
Moving the Narkomed GS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Section 3. Setup and Installation
Setup Checklist. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Installing the Absorber Swivel Arm Stop Pin . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Installing the Vaporizers (Fixed Mount) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Installing the Vaporizers (Removeable Mount). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Filling the Vaporizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Handling the Vaporizer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Installing the Absorber System. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Installing the Breathing Pressure Pilot Line . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Installing the Auto-Bag Sensor Cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Installing the Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Installing the Ultrasonic Flow Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Installing the Bain Circuit Adapter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Installing the Fresh Gas Oxygen Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Connecting the Open Reservoir Scavenger System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Connecting the Scavenger Interface for Passive Systems. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Installing the Patient Line Boom Arm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Installing the Manual Sphygmomanometer Gauge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
Installing the Second Shelf . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Installing the Shelf Extender. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
Installing the Outlet Strip. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-29
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Section 4. Gas Supply Connections
Gas Supply Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Connecting the Cylinders . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Connecting the Pipelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Section 5. Power-Up and System Configuration
Connecting the Narkomed GS to Electrical Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
Power-Up Diagnostics Test. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
Section 6. Testing and PMC
6.0 PMC Procedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
6.1 Self-Diagnostics. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6.2 Electrical Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
6.3 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6.4 Service Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
6.5 Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
6.6 Absorber Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14
6.7 High Pressure Leak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.8 Breathing System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-16
6.9 Oxygen Analyzer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-20
6.10 Flowmeters Concentrations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-21
6.11 High Pressure Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
6.12 Low O2 Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
6.13 Oxygen Supply Failure Protection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
6.14 Pressure Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
6.15 Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
6.16 Bellows Adult . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
6.17 Bellows Pediatric External . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-29
6.18 Bellows Pediatric Internal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-30
6.19 Open Reservoir Scavenger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-31
6.20 A/C Scavenger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-33
6.21 Bag Scavenger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
6.22 Suction Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
6.23 Manual Sphygmomanometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
6.24 Final Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
ii Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
1
Introduction
Section contents:
Introduction.......................................................................... 1-2
Warnings ............................................................................. 1-2
Cautions .............................................................................. 1-2
Notes ................................................................................... 1-2
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
1
Introduction
Introduction This manual provides instructions for moving, setting up, testing, and
maintaining a Narkomed GS anesthesia machine.
Moving the Machine—Section 2 shows how to move the
machine safely.
Setup and Installation—The following sections explain how
to set up and install the anesthesia machine. The setup and installation must be performed by, or under the supervision of, an authorized representative of DrägerService.
Section 3 outlines installation of the breathing circuit components,
sensors, and optional equipment such as vaporizers, scavenger systems, and second shelf.
Section 4 explains how to connect the gas supplies to the machine
and notes the precautions to be observed when making those connections.
Warnings, Cautions, and Notes
Section 5 provides instructions for connecting the anesthesia
machine to electrical power and verifying correct power-up.
Tes t P ro ced u re—Section 6 contains instructions for testing
the machine, using the Periodic Manufacturer's Certification (PMC) procedure. The machine must be tested following setup and installation. The test procedure must be performed by, or under the direct supervision of, an authorized representative of DrägerService.
NOTE: The PMC procedure must also be performed periodically to
ensure that the machine is ready for service.
Refer to the Operator's Manual for cleaning and routine maintenance procedures.
This manual contains warning, caution, and note statements. Each type of statement has a specific interpretation.
Warnings—Give important information that, if ignored, could
result in personal injury.
Cautions—Give important information that, if ignored could
lead directly to equipment damage and, indirectly, to a patient's injury.
Notes—Clarify information that may not be readily obvious,
such as a secondary cause and effect.
1-2 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
2
Moving the Narkomed GS
Section contents:
Getting the Machine Ready to Move ................................... 2-2
Moving the Machine ............................................................ 2-2
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual
2
Moving the Narkomed GS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Getting the Machine Ready to Move
Moving the Machine
Prepare the machine so that it can be moved safely:
unplug the power cord
retract the boom arm
disconnect all external hoses
remove all external monitors and equipment
remove the absorber system
Perform the following steps to move the anesthesia machine.
WARNING: The anesthesia machine should only be moved by people who
are physically capable of handling its weight. Draeger Medical recommends that two people move the anesthesia machine to aid in maneuverability. Exercise special care so that the machine does not tip when moving up or down inclines, around corners, and across thresholds (for example, in door frames and elevators). Do not attempt to pull the machine over any hoses, cords, or other obstacles on the floor.
1. Unlock the front casters by stepping on their locking levers. Do not attempt to move the machine while the casters are locked.
2. Using only the handle or push/pull bar shown in the illustration, move the machine.
Do not push or pull the anesthesia machine using the absorber system, vaporizers, ventilator bellows, cord wrap, or boom arm.
2-2 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Moving the Narkomed GS
2
HANDLE
CASTER LOCKING LEVERS
PUSH/PULL
BAR (OPTIONAL)
SU87026
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 2-3
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
3
Setup and Installation
Section contents:
Setup Checklist.................................................................... 3-2
Installing the Absorber Swivel Arm Stop Pin ....................... 3-3
Installing the Vaporizers (Fixed Mount) ............................... 3-4
Installing the Vaporizers (Removeable Mount) .................... 3-5
Filling the Vaporizer ............................................................. 3-7
Handling the Vaporizer ........................................................ 3-7
Installing the Absorber System ............................................ 3-7
Installing the Breathing Pressure Pilot Line ....................... 3-10
Installing the Auto-Bag Sensor Cord ................................. 3-12
Installing the Oxygen Sensor............................................. 3-13
Installing the Ultrasonic Flow Sensor................................. 3-14
Installing the Bain Circuit Adapter ..................................... 3-15
Installing the Fresh Gas Oxygen Sensor ........................... 3-18
Connecting the Open Reservoir Scavenger System ......... 3-19
Connecting the Scavenger Interface for Passive Systems 3-20
Installing the Patient Line Boom Arm ................................ 3-22
Installing the Manual Sphygmomanometer Gauge............ 3-23
Installing the Second Shelf................................................ 3-25
Installing the Shelf Extender .............................................. 3-27
Installing the Outlet Strip ................................................... 3-29
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual
3
Setup and Installation
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Setup Checklist
Following is a checklist for the initial setup of a Narkomed GS Anesthesia Machine. This setup must be performed by or under the direct supervision of an authorized representative of DrägerService. Items 1 through 16 are described in this section; items 17 through 19 are outlined in subsequent sections.
1. Install the absorber swivel arm stop pin
2. Install the vaporizers (optional)
3. Install the absorber system (optional)
4. Install the breathing pressure pilot line
5. Install the AUTO/BAG sensor cord
6. Install the oxygen sensor
7. Install the ultrasonic flow sensor
8. Install the Bain Circuit adapter (optional)
9. Install the fresh gas oxygen sensor adapter (optional)
10. Connect the open reservoir scavenger system (optional)
11. Connect the passive scavenger system (optional)
12. Install the patient line boom arm (optional)
13. Install the manual sphygmomanometer gauge (optional)
14. Install the second shelf (optional)
15. Install the shelf extender (optional)
16. Install the outlet strip (optional)
17. Connect the gas supplies
18. Connect the machine to electrical power
19. Perform a Periodic Manufacturer's Certification (PMC) procedure. This procedure must be performed by an authorized representative of DrägerService as part of the initial setup.
3-2 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the Absorber Swivel Arm Stop Pin
Setup and Installation
3
Install the stop pin as follows:
1. Locate the roll pin supplied in the drawer of the machine.
2. Rotate the swivel arm so that the hole near the bottom of the arm faces the back of the machine.
3. Hammer the pin into the hole until it is flush with the swivel arm.
SWIVEL ARM
Selecting the Vaporizer Order
(Fixed Vapor Mount System)
PIN
SU00606
You must install the vaporizers in the following order, as viewed from the front of the machine, unless otherwise specified by the customer:
Left position: Halothane
Center position: Enflurane
Right position: Isoflurane (or desflurane)
NOTE: To install a TEC 6 desflurane vaporizer, see Service Procedure
SP00091.
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 3-3
3
Setup and Installation
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the Vaporizers
(Fixed Vapor Mount System)
The following procedure applies to the fixed vapor mount system used on 2-vaporizer and 3-vaporizer machines, for Dräger 19.1 vaporizers.
If your machine is equipped with a removable vaporizer system, skip to the next procedure.
Install each vaporizer as follows:
1. Remove the back cover from the vaporizer exclusion system.
2. Remove the vaporizer from its shipping container.
3. Remove the tape from the back of the vaporizer. Verify that the vaporizer's O-rings are present at the inlet and outlet ports.
4. Unscrew the plastic plates that are mounted over the vapor ports. Save the screws that you remove.
5. Using the saved screws from the previous step, secure the vaporizer to the appropriate location on the manifold (see “Selecting the Vaporizer Order” above), and tighten the screws.
WARNING: Use only the screws from the plastic plates.
6. When all vaporizers are installed, test the exclusion system for proper operation. If adjustment is needed, refer to the Narkomed GS Service Manual. Reinstall the back cover.
COVER SCREWS
AND
WASHERS (4X)
BACK
COVER
VAPORIZER
MOUNTING SCREWS (2X)
MANIFOLD
O-RINGS
SU00109
VAPORIZER
3-4 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the Vaporizers
(Removable Vapor Mount System)
Setup and Installation
3
The removable vapor mount system will accommodate Dräger 19.3 and Dräger 2000 vaporizers.
If the customer does not specify the vaporizer positions, install the vaporizers in the following order, as viewed from the front of the machine:
Left position: Halothane
Right position: Isoflurane
Install the vaporizers as follows:
1. Examine the ports on the vaporizer manifold and verify that each port has an O-ring.
2. Move the selector lever to the right.
3. Place the vaporizer down over the left manifold ports, and turn its locking lever 90
° clockwise to secure the vaporizer to the manifold.
4. Move the selector lever to the left.
5. Place the second vaporizer down over the right manifold ports, and turn its locking lever 90
° clockwise to secure the vaporizer to the
manifold.
6. Test the interlock for proper operation. Verify that only one vaporizer at a time can be turned on. Return all vaporizers to their Zero position at completion of the test.
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 3-5
3
R
G
D
Setup and Installation
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
LOCKING
LEVER
O-RINGS
VAPORIZE
MOUNTIN
MANIFOL
REMOVABLE VAPORIZER
MOUNTING ARRANGEMENT
(TOP VIEW)
SU00114
SELECTOR LEVER
3-6 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Filling a Vaporizer
Handling the Vaporizer
Installing the Absorber System
Setup and Installation
3
See the appropriate separate manual, supplied with vaporizers available for use with the Narkomed 6000.
If a vaporizer is dropped during handling, or if its handwheel exhibits a lack of resistance (spins freely), or if a gas analyzer maintains a zero reading after the handwheel has been turned to a labeled concentration, do not use the vaporizer. Return the vaporizer to DrägerService.
Install the absorber system as described below. This installation procedure presumes the use of an appropriate scavenger system.
WARNING:The Narkomed GS's patient breathing system must not be
used in conjunction with any additional components that establish a flow direction.
WARNING:Hoses and bags attached to the 22 mm hose terminals of the
inspiratory valve, expiratory valve, ventilator hose connection, and breathing bag mount must comply with current ANSI standards.
1. Place the absorber mounting stud into the top of the absorber pole.
2. Tighten the set screw on the absorber pole to lock the absorber in place.
WARNING:Do not pinch or kink the fresh gas hose leading from the fresh
gas common outlet to the absorber.
3. Pull out the fresh gas locking bar, located on the front of the Narkomed GS, to its extended position. Insert the 15 mm male fitting on the fresh gas hose axially into the 15 mm female terminal. Release the spring­loaded locking bar over the fitting, allowing it to “lock” the fitting into place.
WARNING:To prevent leaks and misdirection of gas pathways, all hoses
should be correctly and tightly fitted, as shown in the illustrations. Take special care to attach all 19 mm hoses to the proper 19 mm connectors. Possible machine malfunction and harm to the patient could occur if the scavenger hoses are attached to any 22 mm connection.
4. Attach a 22 mm breathing hose between the ventilator bellows 22 mm terminal marked VENTILATOR HOSE, and the 22 mm terminal on the rear of the manual/automatic selector valve marked VENTILATOR HOSE.
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 3-7
3
Setup and Installation
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
5. Attach a 22 mm breathing hose betwen the 22 mm hose terminal on the inspiratory valve marked INSPIRATION, and one side of the Y-piece.
WARNING:To ensure proper gas flow direction during inspiration and
expiration, the disks in the inspiratory and expiratory valves must move freely, without sticking.
6. Attach a 22 mm breathing hose between the other side of the Y-piece and the expiratory hose terminal on the ultrasonic flow sensor.
7. Attach the breathing bag to the swivel bag mount 22 mm terminal marked BREATHING BAG.
ULTRASONIC FLOW SENSOR W/22MM BREATHING HOSE TERMINAL
INSPIRATORY VALVE W/22MM BREATHING HOSE TERMINAL
22MM BREATHING HOSES
(P/N 9995132)
ABSORBER
ABSORBER MOUNTING STUD
EXPIRATORY
VALVE
APL VALVE W/19MM
SCAVENGER HOSE
TERMINAL
19MM SCAVENGER
HOSE
MANUAL/AUTOMATIC
SELECTOR VALVE
W/22MM HOSE TERMINAL
LOCKING DEVICE
W/15MM FEMALE
FRESHGAS HOSE
(P/N 9995230)
TO SCAVENGER
SET SCREW
(P/N HW10002)
FRESHGAS
FITTING
W/15MM MALE
FITTING
Y-PIECE
SWIVEL BAG MOUNT W/22MM BREATHING BAG TERMINAL
SU14718B
3-8 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
BREATHING BAG
ABSORBER POLE
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Setup and Installation
WARNING:The breathing bag acts as a pressure limiting device during
manually assisted and spontaneous ventilation. Therefore, breathing bags used with the Absorber system must comply with the pressure/volume requirements of current ANSI standards. A bag that has been stretched may have drastically altered compliance characteristics, and altered conductivity in the case of conductive bags.
8. Connect the 19 mm scavenger hose between the 19 mm terminal marked SCAVENGER HOSE on the rear of the APL valve and the 19 mm terminal marked SCAVENGER HOSE on the scavenger.
VENTILATOR BELLOWS
W/22MM VENTILATOR
HOSE TERMINAL
SU98101B
APL
VALVE
PRESSURE
SENSING PORT
QUICK-CONNECT
FITTING
OXYGEN SENSOR
3
VENTILATOR 22MM BREATHING HOSE
19MM SCAVENGER HOSE
TO SCAVENGER
ULTRASONIC
FLOW SENSOR
ABSORBER POLE
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 3-9
3
Setup and Installation
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the Breathing Pressure Pilot Line
For Absorber Monitoring
Draeger Medical anesthesia systems are supplied with two breathing pressure pilot lines:
A short line for breathing pressure monitoring at the absorber
A long line for breathing pressure monitoring at the Y-piece
For breathing pressure monitoring at the absorber, install the short pilot line (which has quick-connect fittings on both ends) as follows:
1. Connect one end of the pilot line to the fitting mounted on the rear of the pipe extending from the absorber top assembly.
2. Connect the other end of the pilot line to the BREATHING PRESSURE interface underneath the rear panel of the ventilator box.
REAR PANEL OF
VENTILATOR BOX
VOLUME SENSOR
2
OXYGEN SENSOR
BREATHING PRESSURE
AUTO/BAG SELECTOR
QUICK-CONNECT
FITTINGS
SHORT PILOT LINE
(P/N 4109368)
SU00603A
ABSORBER
TOP ASSEMBLY
3-10 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
For Y-piece Monitoring
Setup and Installation
3
For breathing pressure monitoring at the patient Y-piece, install the long pilot line (which has a quick-connect fitting on one end and a Luer type fitting on the other end) as follows:
1. Connect the quick-connect fitting on the pilot line to the BREATHING PRESSURE interface underneath the rear panel of the ventilator box.
NOTE: The quick-connect fitting on the absorber is self-closing. You can
leave it unused when installing the long pilot line for Y-piece monitoring.
2. Connect the Luer fitting on the other end of the pilot line to a Luer fitting on either the patient Y-piece or a 15 mm adapter on the patient side of the Y-piece.
3. Using the four plastic hose clips attached to the pilot line, mount the pilot line on either of the breathing hoses leading to the Y-piece.
PILOT LINE
(P/N 4108528)
REAR PANEL OF VENTILATOR BOX
LONG
VOLUME SENSOR
MALE
LUER-LOCK
FITTING
2
OXYGEN
BREATHING
SENSOR
PRESSURE
PLASTIC
HOSE CLIPS
AUTO/BAG SELECTOR
QUICK-CONNECT
FITTING
SU00604A
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 3-11
3
Setup and Installation
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the AUTO/BAG Sensor Cord
1. Push the manual/automatic selector sensor into the underside of the manual/automatic selector valve housing until it snaps into place. Note that the sensor is keyed and only fits into the housing one way.
2. Align the key on the sensor cord connector with the keyway on the AUTO/BAG SELECTOR interface underneath the ventilator box. Then push the connector in until it snaps into place.
REAR PANEL OF
VENTILATOR BOX
SU00602
VOLUME SENSOR
SENSOR CORD CONNECTOR
2
OXYGEN SENSOR
BREATHING PRESSURE
AUTO/BAG SELECTOR
INTERFACE CABLE
MANUAL/
AUTOMATIC
SELECTOR
VALVE
MANUAL/
AUTOMATIC
SELECTOR
SENSOR
3-12 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
Y
Y
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the Oxygen Sensor
Setup and Installation
3
1. Remove the inspiratory valve dome plug from the inspiratory valve dome.
2. Insert the sensor cord connector into the OXYGEN SENSOR interface underneath the rear panel of the ventilator box.
3. Remove any protective covering from the sensor housing.
4. Perform an oxygen sensor calibration as described in Operation ­Oxygen Monitoring in the
Narkomed GS Operator's Manual.
5. Insert the sensor assembly into the inspiratory valve dome by pressing it into place.
SENSOR
REAR PANEL OF
VENTILATOR BOX
HOUSING
VOLUME SENSOR
2
OXYGEN SENSOR
BREATHING
AUTO/BAG
PRESSURE
SELECTOR
SENSOR
CORD
CONNECTOR
INSPIRATOR VALVE DOME
INSPIRATOR VALVE DOME
PLUG
(P/N 4106387)
SU00601
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 3-13
3
Setup and Installation
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the Ultrasonic Flow Sensor
1. Slide the flow sensor down onto the bracket attached to the expiratory valve mount.
2. Install the connector hose on the threaded port of the flow sensor, and join the other end of the hose to the expiratory valve. Ensure that the expiratory valve is oriented as shown in the illustration below.
3. Connect the sensor plug to the VOLUME SENSOR interface on the rear panel of the ventilator box.
2
SENSOR PLUG
ULTRASONIC
FLOW SENSOR
CONNECTOR
HOSE
EXPIRATORY HOSE TERMINAL
FLOW SENSOR BRACKET
EXPIRATORY VALVE
ABSORBER ASSEMBLY
SU00009A
3-14 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the Bain Circuit Adapter (Optional)
Absorber Mount
Setup and Installation
3
Two types of Bain circuit adapters are available. One mounts to the absorber; the other mounts to the pole.
To install the absorber-mounted Bain Circuit Adapter, slip the mounting bracket of the Bain Circuit Adapter down over the slide mount on the absorber assembly (on the block below the APL valve).
WARNING:To avoid confusion between hose terminals on the absorber
system and the Bain Circuit Adapter, always remove the Bain Circuit Adapter from the absorber mount when the Bain Circuit Adapter is not in use.
TO BREATHING PRESSURE MONITOR INTERFACE PANEL
APL VALVE
TO 19MM HOSE TERMINAL
ON SCAVENGER
OXYGEN SENSOR
22MM M/15MM F EXPIRATION TERMINAL
BAIN CIRCUIT
BAIN CIRCUIT HOSE BARB FOR FRESH GAS HOSE
ABSORBER
APL VALVE
BAIN CIRCUIT
ADAPTER
BREATHING BAG
SU14730A
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 3-15
FRESH GAS
OUTLET
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
3
Setup and Installation
Pole Mount The Bain Circuit Adapter that mounts on the absorber pole may be
supplied with an optional positive end-expiratory pressure (PEEP) valve.
To install a pole-mounted Bain Circuit Adapter, with or without a PEEP valve:
1. Loosen the set screw on the absorber pole and remove the absorber system (if the machine is so equipped).
2. Slip the Bain Circuit Adapter mounting stud into the absorber pole.
3. Tighten the set screw on the absorber pole to lock the Bain Circuit Adapter in place.
BREATHING PRESSURE PILOT LINE
BAIN CIRCUIT
BAIN CIRCUIT HOSE BARB FOR FRESH GAS HOSE
MOUNTING STUD
OXYGEN SENSOR
19 MM SCAVENGER HOSE
(P/N 9995230)
(CONNECTS TO 19MM
TERMINAL ON SCAVENGER)
22MM HOSE
(P/N 9995123)
(CONNECTS TO
VENTILATOR HOSE
TERMINAL ON
VENTILATOR
BELLOWS ASSEMBLY)
SET SCREW
FRESH GAS
COMMON
OUTLET
BREATHING
FRESH GAS HOSE
SU14719A
3-16 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.
BAG MOUNT
BREATHING BAG
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Bain Circuit Oxygen Sensor Installation
Setup and Installation
3
1. Insert the oxygen sensor into the port adjacent to the expiratory terminal. When the port is not in use, close it with the provided plug. Insert the sensor cable into the OXYGEN SENSOR interface as described earlier in this section.
WARNING:A functioning oxygen analyzer must always be used with the
Bain Circuit Adapter.
2. Connect the breathing pressure pilot line to the quick-connect fitting on the bain circuit adapter. Insert the other end of the pilot line into the BREATHING PRESSURE interface as described earlier in this section.
WARNING:A functioning breathing pressure monitor must always be
used with the Bain Circuit Adapter.
3. Bain Circuit ConnectionsInstall the breathing bag onto the terminal labeled BREATHING BAG.
WARNING:Breathing bags attached to the Bain Circuit Adapter's 22 mm
terminals must comply with current ANSI standards.
4. Connect the 19 mm scavenger hose between the 19 mm terminal at the rear of the APL valve marked SCAVENGER HOSE and the 19 mm terminal on scavenger.
5. If applicable, connect a 22 mm hose between the terminal on the rear of the manual/automatic selector valve, and the VENTILATOR HOSE terminal on the bottom of the ventilator bellows assembly.
WARNING:Hoses connected to the Bain Circuit Adapter terminals must
comply with current ANSI standards.
6. Connect the Bain Circuit to the 22 mm male/15 mm female terminal on the Bain Circuit Adapter marked EXPIRATION.
7. Connect the fresh gas hose between the fresh gas outlet on the Narkomed GS, and the hose barb fitting on the Bain Circuit (inner tube connection).
8. Set the APL valve fully open, set the PEEP valve (if supplied) at minimum, and set the Man/Auto selector to BAG.
WARNING:The fresh gas hose must not be pinched or kinked.
Draeger Medical, Inc. Narkomed GS Setup and Installation Manual 3-17
3
Setup and Installation
RETURN TO THIS MANUAL'S TABLE OF CONTENTS
RETURN TO CD-ROM TABLE OF CONTENTS
Installing the Fresh Gas Oxygen Sensor Adapter (Optional)
1. Hold the locking bar on the anesthesia machine's fresh gas outlet in the extended position.
2. With the oxygen sensor port facing up, slide the notched portion of the adapter stem into the retaining slot of the locking bar, so that the adapter's 15 mm male fitting faces the anesthesia machine's fresh gas outlet.
3. Release the locking bar, guiding the adapter's 15 mm male fitting into the fresh gas outlet of the anesthesia machine.
4. Insert the oxygen sensor into the port on top of the adapter.
WARNING:If the oxygen sensor is removed, the oxygen sensor port must
be closed with the oxygen sensor port plug.
5. Attach the non-rebreathing circuit fresh gas hose to the adapter's barbed fitting. (For non-rebreathing circuits equipped with a 15 mm fresh gas connector, cut off the connector and press the hose onto the barbed fitting of the adapter.)
FRESH GAS OUTLET
OXYGEN SENSOR
15MM MALE FITTING
ADAPTER STEM
OXYGEN SENSOR PLUG
SU20997
FRESH GAS
LOCKING BAR
(EXTENDED POSITION)
OXYGEN SENSOR PORT
FRESH GAS HOSE
BARBED FITTING FOR
NON-REBREATHING
GAS CIRCUIT
3-18 Narkomed GS Setup and Installation Manual Draeger Medical, Inc.