Nortec UL10 Operation Manual

Page 1
OPERATION MANUAL
TKZ: 315 011 002 002
Ausgabe: 01 / 2015
CHARGER / DISCHARGER
UL10
0 – 40V / 0 – 40A
TKZ: UL10.040000
Software-Version 20.108a
Dated 22.01.2013
COPYRIGHT © 2003-2015 by
NorTec Electronics GmbH & Co. KG
An der Strusbek 32 B
D – 22926 Ahrensburg
Tel.: +49 4102 42002 Fax: +49 4102 42840
Email: info@nortec-electronics.de
Web: www.nortec-electronics.de
Page 2
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 1
Page 3
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
2 © NorTec Electronics 2014
CHARGER / DISCHARGER UL10
Page 4
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 3
Contents
1 General ................................................................................................................................. 6
1.1 Application............................................................................................................................. 6
1.1.1 Range of functions ................................................................................................................ 6
1.1.2 Design ................................................................................................................................... 6
1.2 Sphere of application ............................................................................................................. 7
2 Organs of control and indication ............................................................................................ 8
2.1 Front instrumentation panel ................................................................................................... 8
2.2 Rear Instrumentation wall ...................................................................................................... 9
2.2.1 Mains cable ........................................................................................................................... 9
2.2.2 Source of acoustic signals ..................................................................................................... 9
3 Principle of operation ........................................................................................................... 10
3.1 Connection .......................................................................................................................... 10
3.2 Installation ........................................................................................................................... 10
3.3 Commissioning .................................................................................................................... 10
3.3.1 Mains Connection................................................................................................................ 10
3.3.2 Warning ............................................................................................................................... 11
4 Ratings ................................................................................................................................ 12
4.1 Input & Output Parameters .................................................................................................. 12
4.2 Working temperature range ................................................................................................. 12
5 Controls ............................................................................................................................... 13
5.1 General ............................................................................................................................... 13
5.1.1 Indikation ............................................................................................................................. 13
5.1.2 Keyboard ............................................................................................................................. 14
5.2 Self-Test .............................................................................................................................. 16
6 System adjustment .............................................................................................................. 17
6.1 Language Selection ............................................................................................................. 17
6.2 Date Input............................................................................................................................ 18
6.3 Time Input ........................................................................................................................... 18
6.4 Print out functions................................................................................................................ 19
6.4.1 General ............................................................................................................................... 19
6.4.2 Print Equipment ................................................................................................................... 19
6.4.3 Printing Programs................................................................................................................ 20
6.5 Program Selection ............................................................................................................... 22
Page 5
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
4 © NorTec Electronics 2014
6.5.1 Principles of Program Selektion ........................................................................................... 22
6.6 Types of batteries ................................................................................................................ 24
6.7 Programming Matrix ............................................................................................................ 24
6.8 Input of battery data ............................................................................................................ 25
6.8.1 General ............................................................................................................................... 25
6.8.2 Operation without battery Р:00 ............................................................................................ 25
6.8.3 Operation without Lead acid batteries ................................................................................. 25
6.8.4 Operation with nickel-cadmium batteries ............................................................................. 28
6.9 Readings on the LC display of program data 420 ................................................................ 30
6.9.1 Program 420 Stage 1 .......................................................................................................... 31
6.9.2 Program 420 stage 2 ........................................................................................................... 32
6.9.3 Program 420 stage 3 ........................................................................................................... 33
6.9.4 Program 420 stage 4 ........................................................................................................... 34
6.9.5 Program 420 stage 5 ........................................................................................................... 35
6.9.6 Program 420 stage 6 ........................................................................................................... 36
7 Tabulated program progress display.................................................................................... 37
7.1 Program 0. Discharge – any battery type ............................................................................ 37
7.2
Program 10. Discharge – air-tight aircraft lead acid battery ................................................. 38
7.3 Program 11. Discharge – Pb battery sealed commercial ..................................................... 39
7.4 Program 16. Discharge – Pb battery vented commercial ..................................................... 40
7.5 Program 20. Discharge – enclosed aircraft nickel-cadmium battery .................................... 41
7.6 Program 31. Discharge – enclosed aircraft nickel-cadmium battery .................................... 42
7.7 Program 110. Commissioning – air-tight aircraft lead acid battery ....................................... 43
7.8 Program 111. Commissioning – Pb battery sealed commercial ........................................... 44
7.9 Program 116. Commissioning – Pb battery vented commercial ........................................... 45
7.10 Program 120. Commissioning – enclosed aircraft nickel-cadmium battery .......................... 46
7.11 Program 210. Charging – air-tight aircraft lead acid battery ................................................. 47
7.12 Program 211. Charging – Pb battery sealed commercial ..................................................... 48
7.13 Program 216. Charging – Pb battery vented commercial..................................................... 49
7.14 Program 220. Charging – aircraft nickel-cadmium battery ................................................... 50
7.15 Program 221. Charging – aircraft nickel-cadmium battery VARTA ...................................... 51
Page 6
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 5
7.16 Program 230. Charging – aircraft nickel-cadmium battery type SAFT .................................. 52
7.17 Program 231. Charging – aircraft nickel-cadmium battery type НКБН ................................. 53
7.18 Program 410. Maintenance – air-tight aircraft lead acid battery ........................................... 54
7.19 Program 411. Maintenance – Pb battery sealed commercial ............................................... 55
7.20 Program 416. Maintenance – Pb battery vented commercial ............................................... 57
7.21 Program 420. Maintenance – enclosed aircraft nickel-cadmium battery .............................. 58
7.22 Program 421. Maintenance – enclosed aircraft nickel-cadmium battery type VARTA .......... 59
7.23 Program 430. Maintenance – enclosed aircraft nickel-cadmium battery type SAFT ............. 60
7.24 Program 431. Maintenance – enclosed aircraft nickel-cadmium battery type НКБН ............ 61
7.25 Program 510. Direct current charging – air-tight aircraft lead acid battery............................ 62
7.26 Program 511. Direct current charging – Pb battery sealed commercial ............................... 63
7.27 Program 516. Direct current charging – Pb battery vented commercial ............................... 64
7.28 Program 520. Direct current charging – enclosed aircraft nickel-cadmium battery ............... 65
7.29 Program 800. Constant power supply – without battery ...................................................... 66
7.30 Program 810. Constant power supply – air-tight aircraft lead acid battery .......................... 67
7.31 Program 811. Constant power supply – Pb battery sealed commercial .............................. 68
7.32 Program 816. Constant power supply – Pb battery vented commercial .............................. 69
7.33 Program 820. Constant power supply – enclosed aircraft nickel-cadmium battery .............. 70
8 Replacement of printing paper and ribbon ........................................................................... 71
8.1 Ribbon replacement ............................................................................................................ 71
8.2 Replacement of printing paper............................................................................................. 71
Page 7
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
6 © NorTec Electronics 2014
1 General
1.1 Application
1.1.1 Range of functions
The battery charger-discharger UL10 serves to test, charge and discharge all types of batteries based on their particular approved characteristics and using various programs.
A menu and a touch-input keyboard control all the programs. Also, the device can serve as a direct current source without connecting any battery.
The following software is available to measure and register voltages of individual batteries according to the current charging program or independently of operations underway:
- Measurement of voltage of an individual battery (Р)
Sphere of application
The device is contained in a body to be used indoors. Its rugged design enables to install it on vibra­tion-proof racks or to carry it in a handbag during field exercises. It is used by the navy as a tester.
Types of batteries:
The types of the batteries are the following:
00 no battery connected 10 air-tight aircraft lead acid electrolyte-filled batteries 11 Pb battery sealed commercial 16 Pb battery vented commercial 20 enclosed aircraft nickel-cadmium batteries 21 enclosed nickel-cadmium batteries type VARTA 30 enclosed nickel-cadmium batteries type SAFT 31 enclosed nickel-cadmium batteries type НКБН
Different programs serve to enter parameters of various batteries (nominal voltage, number of cells in a battery and nominal capacity) into the UL 10 enabling to ensure their long service life.
1.1.2 Design
The device consists of three main components:
1. A functional pull-out unit 19 inches. Up to 5 units can be installed in a pile. The performance of the device is tested during manufacture.
2. A read-only memory specified by a customer of energy independent reprogrammable sto­rage with a relevant instruction manual.
3. A body
4. A set of cables (specified by the customer)
Page 8
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 7
1.2 Sphere of application
The normal body of the device is designed for indoor operation without any special ventilation.
It can be installed in a laboratory, a charging room, etc.
The device is fully adapted for field service when packed into a polyethylene or corrugated card­board transport case.
The fully functional pull-out unit can be used in a separate vibration protected body or in groups of mobile or stationary framework racks.
Page 9
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
8 © NorTec Electronics 2014
2 Organs of control and indication
Organs of control, indication and connection of the UL 10 device for testing batteries are easily ac­cessible from the front panel. Connections of specifically requested cells for automatic registration of voltages of batteries, such as aircraft batteries and batteries for torpedoes or connections of several PCs to the mains, are arranged on the rear face.
2.1 Front instrumentation panel
1) The POWER ON / OFF switch with a LED indicator
а) triggers the device on and off
b) The red LED indicates that the device is connected to the mains
2) The automatic protection circuit 16 А protects 220 V AC power supply.
The indicator
The LC graphic display (2 lines, 40 characters) serves to transmit information to the user and to in­dicate the parameters of operation of programs in the progress of charging.
2) The touch-input keyboard
The keys BREAK, CLEAR, PRINT, ENTERserve to interrupt programs, to delete erroneously en­tered data, to print out results and to confirm input of requested data, respectively. The digital unit enables the user to enter digital data, such as parameters of batteries and the opera­tor number. The key * ensures access to the information about the system menu.
3) The printer
The built-in printer for printing out protocols can be enabled at any moment before and after opera­tions with a battery. It prints out all basic parameters of a battery and all errors, if necessary.
6) The battery tester
The battery tester is connected to the UL10 device with a plug-and-socket connector type VG 95 328 C10-6SN (6-contact). The battery tester enables to measure voltages of batteries between
-4.0 to + 4.0 V.
7) Mains connection to the battery
The battery is connected via special connecting cables to the UL 10 device with two 4-pin Plug and socket connectors type VG complying with requirements VG 95 234.
8) The manufacturer/ designation of the UL 10 device
Page 10
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 9
2.2 Rear Instrumentation wall
2.2.1 Mains cable
2.2.2 Source of acoustic signals
The source of acoustic signals informs the operator of the condition of the device; accordingly them the operator receives relevant instructions to execute. An error of input, a failure or operation finish can induce an acoustic signal.
Page 11
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
10 © NorTec Electronics 2014
3 Principle of operation
3.1 Connection
Connection is made following the instructions below:
- connect to the 220 V AC mains
- connect the battery
- put the mains switch POWER ON / OFF (2.1.1a) into position ON
- check if the red LED indicator is on.
After that the automatic self test follows.
NOTE!
The procedure is interrupted in case of an error message. After a successful self-test the parame­ters of the connected battery should be entered by selecting a corresponding item. This is done with
the keys 4; 6 and ENTER. The available types of batteries correspond to item 1.1.2.
3.2 Installation
The device should be installed so that uninterrupted ventilation of and air exhaust from the device are ensured (at least 5 cm free space from the back face of the device). The cooling air is delivered from below and exits through the rear face. When the device operates in the open air, e.g. under a tent, check that it is placed n a dust-free base otherwise the cooling process can be inhibited.
The device should be operated in the buildings designed for this purpose on corresponding bases because the pull-out unit is fitted with a fan without any extra case so that the device without a pro­tective body can be damaged or the servicing personnel can be traumatized.
3.3 Commissioning
3.3.1 Mains Connection
- put the switch POWER ON/OFF into the position OFF
- connection to the mains 220 V AC is made with a mains cable extending through the rear face and
fitted with a connector type CEBEC (at an angle 90
о)
- connect the cable from the battery to the device (+ and -). It is impossible to confuse the polarity
because the connector is provided with a special key
- connect the measuring cable when operating with nickel-cadmium batteries
- the battery is connected to the UL10 with the cables from the delivery set (the cables with clamps
(+ and -)). Check the polarity! When connecting the battery with the cable with tapered outlet clamps with built-in temperature sensors it is impossible to confuse the polarity because of the geometry of the poles of the battery.
Page 12
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 11
3.3.2 Warning
Be cautious when touching!
The rear face of the pull-out unit can heat, specifically when executing the discharge programs.
The device disconnects in case of an intolerable temperature rise.
Do not open the device when it is on!
Even when the device is disconnected from the 220 V AC mains there remains inside the device a level of voltage dangerous to touch. Exclusively the manufacturer can perform servicing of the de­vice otherwise the warranty becomes null.
Note:
The manufacturer bears no responsibility for losses due to improper installation, commissioning or operation. Only the materials approved by the manufacturer should be used.
- Only the manufacturer can repair and calibrate the device UL 10 when it is used in aircraft.
Page 13
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
12 © NorTec Electronics 2014
4 Ratings
4.1 Input & Output Parameters
input voltage 220 V + 10% (single-phase alternating)
input frequency 45-60 Hz
input power 2000 VА
external mains protection 16 А, mean inertia
fan power about 100 / 300 l/min
charge current range 0.5 40 А
discharge current range 0.5 – 40 А
charge voltage range 0.1 – 40 V
discharge voltage range 0.1 – 40 V
voltage error < 0.1 % of rated parameter plus ±5 discharge
current power error < 3 % of rated parameter plus ±5 discharge
temperature error < ±2
С
Insulation VDE 0160, class 1
Protection type IP 54
working temperature -20
С to +45 С
Weight about 50 kg (in 19” body, without battery cable)
Dimensions width 400 mm, height 504 mm, depth 242 mm
Free space for exhaust > 50 mm
4.2 Working temperature range
The device operation is guaranteed within the temperature range -20
С to + 45 С.
The accuracy of the readings of the indicator and printed out discharge and charge parameters correspond to the above indicated.
The accuracy of output values within the temperature range -20
оС to +45 ОС is guaranteed for
2 years since the time of delivery of the device.
Page 14
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 13
5 Controls
5.1 General
5.1.1 Indikation
The display of the device а UL 10 is divided into two parts (region 1 and region 2 below).
Region 1 Region 2
The following Figure shows when information each region displays:
Program status: Stage Status Current time Input, information or operation parameters
Program Status
This line indicates the general condition of the device. Among other messages there can be: PROGRAM SELECTION, PROGRAM START and PRO­GRAM COMPLETION
Stage Status
This line indicates the program current stage and description of the program stages in the process of functioning.
Current time
The time is indicted in the format hours – minutes – seconds
Input, information or operation parameters
All the current parameters are displayed in this region, such as:
- input parameters and data
- Messages about errors in open text
- Operation parameters (current, voltage, etc.) in the process of operation of the battery
- Information about menu execution, temperature indication
Page 15
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
14 © NorTec Electronics 2014
5.1.2 Keyboard
Functions of keys
Key C = CLEAR Key B = BREAK Key P = PRINT Key E = ENTER Key * = Special Program Key = Decimal Point Key 0-9 = Digits 0-9 Key 4 = Digit 4 corresponds to command "forward" Key 6 = Digit 6 corresponds to command "backward" Key 2 = Digit 2 corresponds to command "downward" Key 8 = Digit 8 corresponds to command "upward"
31
7 9
0
B
BREAK
P
PRINT
C
CLEAR
E
8
4
2
6
Page 16
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 15
to 9 and .
Input of digital data separated by a decimal point.
ENTER
Confirmation of data input and messages or exit from information windows.
BREAK
Interruption and /or completion of the current program at any moment of time.
PRINT
In the process of operation with the battery or upon completion of the program all current parameters are printed out. In case of a failure an error message can be printed out by hitting the key PRINT.
CLEAR
Cancel the last data input.
4
Return to the previous item in the menu in the mode of program input.
6
Proceed to the next item in the menu in the mode of program input.
Page 17
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
16 © NorTec Electronics 2014
5.2 Self-Test
Put the switch POWER ON/OFF (2.1.1а) into the position ON. The red LED indicator on the front panel (2.1.1b) should light; no voltage in the mains. The Display is lit. After voltage is switched on the display should show the following information for 2 seconds
- the rights of the manufacturer
- the device model
- the software version
- the software date
- the check status
Test OK
The test is complete Concurrently a short sound beep sounds indicating that the device is ready for operatation. If the self-test detects any failure in the device the following message is displayed:
The appendix lists possible failure messages. These messages are in English, as a rule. If the device operates without failures the information about the type of the connected battery is to be input (see item 6.6) Then a pre-installed program 0-8 to service the battery and the program P and * are selected. The key B enables to return to the item of selection of the battery type.
COPYRIGHT 2000-2003 BY NORTEC UL10 V20.108 REVISION 11.10.2003 SELFTEST
FAILURE: EQUIPMENT KEYBORD
Page 18
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 17
6 System adjustment
After self test and selection of the battery type the key * opens the menu to show options of Chang­ing various parameters of the device. The input data are saved in the energy-independent memory of the device and remain available when the device is switched on again or when voltage fails in the mains. When the key * is hit a menu of system adjustment is displayed in order to select a language of messages and to select the date and specific time indication.
6.1 Language Selection
By hitting the key 0 the current language selection is confirmed (English by default), use the key 1 to scan consecutively the available languages. After that the display shows messages in the selected language. So far the following languages are available in the UL-10 device:
- English
- German
- French
- Russian
(- Dutch: only four languages can be choose!)
Page 19
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
18 © NorTec Electronics 2014
6.2 Date Input
After the language is selected the time is set.
Date Time
1
3.07.01 09:20:07
The cursor appears under the first item of the date. When the key ENTERis hit the input data are saved. The cursor moves to the item of time. The digital keys (0 … 9) serve to enter the current time. The point is entered by default. Any error in the date input is rejected.
6.3 Time Input
The internal clock is adjusted with digital keys (0 … 9). The colon (separating hours from minutes) is entered by default. When the input is confirmed by the key ENTERor the last digit of indication of minutes is entered the display switches over to the main menu. The counting of seconds starts automatically.
Page 20
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 19
6.4 Print out functions
6.4.1 General
The device is provided with a printer with electronic controls. It enables to register results of any op­eration with a battery in the form of a printed out acknowledgement. The print function in the menu can be enabled irrespective of the program input.
6.4.2 Print Equipment
The printer includes the following components:
- a printer body
It is attached to the front panel. The rear face has a 21-pin port for the printing mechanism.
- a printing mechanism with a matrix EPSON printer
- a removable front panel
- normal printing paper 50 mm wide
- 2 knurled head screws to secure the printer
- a self-test start pushbutton
- a color ribbon
- a rear 13-pin connector
- a printer controller
The board BT20-9 controller on the rear face of the printer body inside the device controls printing functions.
Page 21
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
20 © NorTec Electronics 2014
6.4.3 Printing Programs
6.4.3.1
General
By hitting the key P the last printout appears in the main menu or in the submenu in the PP mode.
6.4.3.2 Program Р: printer and measurements of voltages of individual batteries
The Program Р is a special program serving to print out the protocol while another program is run­ning, in particular when measuring and registering voltages of individual batteries. The following op­tions are available:
- Automatic protocol print-out in the program
- Manual protocol print-out in the program
NOTE: Automatic protocol print-out is possible only when the cable is available to measure voltages of individual batteries (optional).
DATA OF ALL PRINT-OUT PARAMETERS
The capacity data in % relate to the battery nominal capacity. The line TEMPERATURE is printed out only when both temperature sensors are connected. The reading relates to a battery with a higher temperature. INSTRUCTIONS HOW TO MEASURE VOLTAGES OF INDIVIDUAL BATTERIES In the program х type of batteries 20. The voltages of individual batteries are possible to measure only in the manual protocol print-out mode is indicated by a short beep. At this moment it is possible to measure voltage with the measur­ing cable and to print out voltages of individual batteries. The maximum number of batteries is 99. If during voltage measurement a beep sounds the measurement should be repeated. The counting of cells is continued only providing the measurement is correct. The voltage of a battery is displayed and printed out simultaneously. If the sign * is printed before the relevant battery voltage it means that it is less than 1.55 V (the vol­tage at the end of charging of an individual nickel-cadmium battery). If the measurement is made at a proper moment of time (according to the requirement of the battery manufacturer) the 1.55 V may prove that the battery is faulty. If individual batteries are not measured the lines CELL 01 – CELL # are not printed. In case the pro­gram print-out does not respond to the data input the device receives a repeated inquiry after 30 seconds, etc., and resumes the normal program run mode.
Page 22
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 21
6.4.3.3 Determination of battery voltage
================================
COPYRIGHT
NORTEC
2000-2003 UL10 V20.108 SN: 110 350
-------------------------------- DATE: 03.04.12 BATTERY NO: 12345 OPERATOR NO: 0002 NOM. VOLTAGE: 24:00 V RATED CAP: 004.0 AH PROGRAM: 420
-------------------------------­PROGRAM START 08:42:24 VOLTAGE = 26.66 V
-------------------------------­P420:S1: DISCHARGE 1 08:42:24 VOLTAGE = 25.91 V CURRENT = 04.01 A CAPACITY = 000.0 AH = 000.0 % TIME = 000 MIN
-------------------------------­* CELL 01: 1.292 V * CELL 02: 1.291 V * CELL 03: 1.294 V
--------------------------------
Page 23
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
22 © NorTec Electronics 2014
6.5 Program Selection
6.5.1 Principles of Program Selektion
All working programs of the UL 10 device have digital designations:
- The first digit represents the operation type
- The second and third digits represent the battery type
6.5.1.1 Types of Operation
6.5.1.1.1 Discharge (=Operation type 0) Any battery with the nominal voltage within up to 40 V can be discharged with the UL 10 device;
with the discharge current up to 40 А. It is performed with discharge load circuits that enable to bring the battery’s voltage to 0 V. The energy the battery receives is dissipated into the environment by radiators and fans.
6.5.1.1.2 Commissioning (=Operation type 1) Depending on the process of manufacture of batteries the manufacturer delivers them charged or
uncharged. Deliveries are possible via intermediate storages. Battery manufacturers provide only the instructions how to use them before their first start, the in­structions may differ depending on the manufacturing processes. For some manufacturers the pa­rameters of commissioning of the batteries are saved in the UL 10 memory depending on the battery type.
6.5.1.1.3 Charging (=Operation type 2) To fulfill the «CHARGING» operation the memory of the UL 10 device contains programs of charg-
ing of each battery type recommended by manufacturers or specified by the customer. The pro­grams required for optimum charging of batteries run fully automatic. In the process of charging the validity of the input battery data is checked, primarily the following:
- Initial voltage
- Final voltage after charging
- Current at the charging start
- Current at the charging end
- Charging duration
- Temperature After the charging process is over the battery condition is rated. Faulty batteries are identified and marked as faulty according to the program used.
6.5.1.1.4 Maintenance (=Operation type 4) This program serves to analyze the condition of the battery that the user of the UL 10 is unaware of.
This program enables:
- To achieve the maximum possible power
- To determine fitness of a battery for further use
- To re-charge the battery for further use. The battery type is selected in 7 stages. The user is guaranteed failure-free operation of the battery after it is checked. The batteries need periodic servicing and maintenance. Such maintenance is performed using operations type 4.
Page 24
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 23
6.5.1.1.5 Charging with direct current (=operation type 5) The operator selects the charging with direct current and the charging time.
6.5.1.1.6 Charging with direct voltage (=operation type 8) In this type of operation the device UL10 can be used for both, with and without the connected bat-
tery as a direct voltage source.
Without the battery = (battery type 00) The device operates as a source of voltage of direct current with a power 40 V / 40 А. The voltage and current can be set in steps 0.01 V / 0.01 А.
With the battery = (battery type 10-31) If the battery type >0 is selected the device automatically switches over to the optimal direct voltage selected for the battery in order to ensure its prolonged charging without affecting the battery’s cha­racteristics.
Page 25
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
24 © NorTec Electronics 2014
6.6 Types of batteries
The data of the main types of batteries are pre-installed in the UL 10device. Other types of batteries can also be checked and they can be charged using the available types.
00: the battery is not connected ♦ 10: air-tight aircraft lead acid electrolyte-filled batteries ♦ 11: Pb battery sealed commercial ♦ 16: Pb battery vented commercial ♦ 20: enclosed aircraft nickel-cadmium batteries ♦ 21: enclosed nickel-cadmium batteries type VARTA ♦ 30: enclosed nickel-cadmium batteries type SAFT ♦ 31: enclosed nickel-cadmium batteries type NKBN
The following types of operations can be used:
0: Discharge 1: In service ♦ 2: Charge 4: Maintenance ♦ 5: I-Charge 8: U-Charge
6.7 Programming Matrix
The above types of operations (0-8) and the 6 types of batteries (00-31) enable up to 43 possible combinations of different charging programs.
BA Type
Operation
00 10 11 16 20 21 30 31
0 - 10 11 16 20 21 30 31 1 - 110 111 116 120 121 130 131 2 - 210 211 216 220 221 230 231 4 - 410 411 416 420 421 430 431 5 - 510 511 516 520 521 530 531 8 800 810 811 816 820 821 830 831
The program for the battery type SAFT and НКБН is executed upon customer’s request. The above programs have no emergency program disable function. The operations should be used extremely cautiously. See additional information about individual programs.
Page 26
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 25
6.8 Input of battery data
The charging program by the UL 10 device depends on the entered data of batteries and the se­lected operation type. The data is required about the type and the battery parameters for the entered data to enable proper execution of the program.
6.8.1 General
When inputting the battery type and parameters the operation type is selected and the program is assigned automatically that the UL10 device should use. For proper execution of programs the data is needed about the composition of the connected battery.
6.8.2 Operation without battery Р:00
When there is no battery (the battery is at 00) only one operation type 8 is possible, or the device is enabled just as a source of direct current voltage. There is an option «the battery is not connected + discharge» = program 000. In this case the final discharge voltage and final current are indicated. This program is disabled; it is enabled by an additional request from the customer because there is a risk of ruining the battery.
6.8.3 Operation without Lead acid batteries
The information below relates to aluminum types of batteries with the data input starting with 1х. Below is the full list of the types in question:
Type 10: air-tight aircraft lead acid electrolyte-filled batteries
Type 11: standard air-tight lead acid electrolyte-filled batteries
Type 16: standard enclosed lead acid electrolyte-filled batteries
After the type is selected with the keys 4 and 6 it is confirmed by hitting Enterand the operation type is to be entered. For lead acid batteries all methods of maintenance can be used. For example, when "maintenance of air-tight aircraft lead acid batteries” is selected hit 6 and proceed to the required operation type, confirm by hitting Enter. The LED message is:
The input of parameters Nominal voltage
The battery nominal voltage should be entered. It is shown on the next tag. It is 12 V usually. The UL 10 device can be used for connecting several batteries in series in exceptional cases.
Batteries with different nominal capacities should be not connected to the mains in series.
P 410: PARAMETER INPUT NOMINAL VOLTAGE 00 0V
Page 27
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
26 © NorTec Electronics 2014
Note:
The lead acid batteries consist, as a rule, of several batteries each with 2 V voltages. The following are standard:
the battery 2 V
the battery 6 V
the battery 12 V
the battery 24 V
The device approves the input of the data as multiples of 2.0 V, otherwise the data are rejected as false. The input voltage is confirmed by hitting ENTER. The LED reads:
Input of parameters nominal capacity
The nominal capacity in Ampere/hours (Аh) characterizes the physical value of an individual battery that determines the current magnitude to which it is possible to charge or discharge it. The label can also have an indication of power. Depending on the input of discharge current magnitude the batte­ries of a similar design can be discharged to different powers. The following rule applies: the higher the selected discharge current the lower is the output power of the battery.
The aircraft lead acid batteries are often designed for a 5-hour discharge.
The capacities of other batteries are calculated depending on their application and the effective standard of discharge current within a range 10 to 100 hours. This is provided for the types of batte­ries in the memory. After the capacity is input (with steps of 0.1 Аh) it is confirmed by hitting
ENTERand the start menu appears:
(0) the device returns to the main menu to select:
the battery type and
the operation type
P 420: PARAMETER INPUT RATED CAPACITY 000.0Ah
P 410: PARAMETER INPUT 1 = START P = WITH PRINT 0 = NEW
Page 28
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 27
(1) starts the checking process from the first stage of the program : discharge by 5-hour current. The
protocol is not printed out usually. Still it can be printed out at any moment of time by hitting the key (P).
During the first print-out the lacking data are requested:
the battery number
the operator identification number
and the last data input appears as a tentative setting. After the input is confirmed by hitting ENTERthe protocol is printed out together with its relevant heading.
The subsequent print-outs can be produced by hitting the key (Р) once again. The reference «Addi­tionally printed out» is printed in the acknowledgement and the current parameters are indicated.
Page 29
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
28 © NorTec Electronics 2014
6.8.4 Operation with nickel-cadmium batteries
The information below relates to all types of batteries with the data input beginning with 2х and 3х. Below is the list of the types in question:
Type 20: enclosed aircraft nickel-cadmium batteries
Type 21: enclosed nickel-cadmium batteries type VARTA
Type 30: enclosed nickel-cadmium batteries type SAFT
Type 31: enclosed nickel-cadmium batteries type NKBN
After the battery type is selected with the keys ← → and confirmed by hitting the key ENTER the type of operation should be selected. Any type of operation can be entered. For example let us select the type 4: a enclosed aircraft nickel-cadmium battery. The LED reads:
Input of parameters The number of cells
For the nickel-cadmium battery the number of batteries should be indicated and the battery voltage is calculated by the device.
Note:
The nominal voltage of a nickel-cadmium battery is 1.2 V. The voltage 1.55 V is selected to charge it fully. Unlike the lead acid batteries the nickel-cadmium batteries have sometimes a nonstandard number of batteries.
The standard number of batteries 5 cells 5 х 1.2 V = battery 6 V 10 cells 10 х 1.2 V = battery 12 V 20 cells 20 х 1.2 V = battery 24 V
An example of the non-standard number 2 cells in portable lamps – 2.4 V 4 cells in portable lamps – 4.8 V 6 cells in piles of batteries for electric drills – 7.2 V 8 cells in piles of batteries for electric drills – 9.6 V 11 cells in piles of batteries for video recorders – 13.2 V 19 cells – special aircraft batteries 26 cells aircraft / guided flying missile Cormoran
Note:
It means that any input of the data and the number of nickel-cadmium batteries requires utmost caution not to damage a battery.
Because of flexible data input the UL 10 is a perfect tool to charge and maintain batteries with uncommon configurations of cells that masks the risk of their improper use!
P 420: PARAMETER INPUT NUMBER OF CELLS 00 CELLS
Page 30
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 29
The maximum tolerable input of the number of cells is 26 (in the program 31:25) that correspond to the maximum:
nominal voltage 31.2 V and
final charge voltage 40.3 V
To avoid false calculation the user should input just the number of batteries and hit ENTER↵. To determine the maximum currents the battery capacity is indicated in Аh.
Input of parameters
nominal capacity
After the capacity is entered and confirmed by hitting ENTERthe start menu appears
P 420: PARAMETER INPUT
RATED CAPACITY 000.0Ah
P 420: PROGRAM START 1 = START P = WITH PRINT 0 = NEW
Page 31
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
30 © NorTec Electronics 2014
♦ (0) or (V) the device returns to the main menu. ♦ (1) Starts the process of checking with the first stage– discharge with the nominal cur-
rent during 6 minutes. The key Р is hit to print out the protocol.
The additional data
the battery number
the user identification number
are input or confirmed by hitting the key ENTER↵.
After the key Р is hit and the additional data are input the process of checking of each battery starts. The print-out reads:
the battery number
the operator identification number
An intermediate print-out is made by hitting Р once again. If the start key is pressed on the positive probe of the measuring cable the voltages of individual batteries are registered.
6.9 Readings on the LC display of program data 420
For enclosed nickel-cadmium aircraft batteries = type 20 according to requirements VG 95 238 Т 110 BS with the nominal voltage 24 V and nominal capacity 4 Аh the check cycle = the operation type should be selected by the key 4. Hence, after input of special data and confirmation by hitting the key Р the program # 420 is started.
Page 32
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 31
6.9.1 Program 420 Stage 1
1 2 3 4
P420:S1:DISCHARGE 1 19:31:43
24.06V 04.00A 000.5AH +008MIN --.-°C
5 6 7 8 9
1 Program number
Operation type 4 = maintenance Battery type 20 = enclosed aircraft nickel-cadmium battery
2 number stage а = Stage 1 3 stage name = discharge 1 4 19:31:43 = current time 5 24,06V = current battery voltage 6 04.00A = current discharge current
The discharge current of the enclosed nickel-cadmium aircraft battery С is equal to the nominal capacity (А), i.e. the current that discharges the battery capacity 4 Аh ful­ly within one 1 hour. This current is 4 А.
7 000.5Ah = the current discharge capacity at the moment of
print out
8 +008MIN Maintenance program # 4 for aircraft batteries stage
1 provided for discharge duration of 6 minutes during which the check current C should flow. It is 4 A in this case.
9 --,- °C = the temperature measurement is disabled. The re-
levant protocol print-out shows the same data in addition to the data of the opera­tor and also the current time. The protocol is drawn up automatically at the end of stage 1 when the remaining charging time is -00:00:00.
Page 33
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
32 © NorTec Electronics 2014
6.9.2 Program 420 stage 2
1 2 3 4
P420:S2:I-CHARGE 1 19:37:38
24.64V 04.00A 000.5AH +007MIN --.-°C
5 6 7 8 9
1 Program number operation type 4 = maintenance battery type 20 = enclosed aircraft nickel-cadmium battery 2 stage number = Stage 2 3 stage name = I-Charge 1 4 19:37:38 = current time 5 24,64V = current battery voltage 6 04.00A = current charging current
The charging current in this program of checking of the aircraft nickel-cadmium bat-
tery restores the nominal capacity during 1 h. 4 А is for the battery with 4Аh.
Page 34
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 33
6.9.3 Program 420 stage 3
1 2 3 4
P420:S3:TOPPING CHARGE 1 20:42:38
29.64V 4.00A 4.019AH +058MIN --.-°C
5 6 7 8 9
1 Program number
operation type 4 = maintenance
battery type 20 = enclosed aircraft nickel-cadmium battery 2 stage number = Stage 3 3 stage name = topping charge 1 4 20:42:38 = current time 5 29.64 V = current battery voltage. No charging voltage limits at the
31 V level.
6 04.00 А = current charging current
Extra charging with the 10-hour current 4 Аh/10 h = 0.4 А during 2 hours. This stage serves to equalize voltages of Individual batteries
7 4.019 Аh The capacity indication continues.
The capacity charged with direct current is 1 plus the capacity after extra charging1.
Page 35
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
34 © NorTec Electronics 2014
6.9.4 Program 420 stage 4
1 2 3 4
P420:S4:DISCHARGE 2 23:34:38
10.5V 04.00A 3.60 AH +054MIN --.-°C
5 6 7 8 9
1 Program number
operation type 4 = maintenance
battery type 20 = enclosed aircraft nickel-cadmium battery 2 stage number = Stage 4 3 stage name = discharge 2 4 23:34:38 = current time 5 10.5 V = current battery voltage 6 4.0 А = current discharge current
The Discharge current of the aircraft nickel-cadmium battery is equal to the nominal capacity 1 С (А), i.e. the current is selected that discharges the battery with the capacity 4 Аh fully within 1 hour. This current is 4 А.
7 3.6 Аh = at the moment of prints-out of the currently selected
capacity
8 0054 MIN When the discharge is 2 the battery is discharged with a on
hour current 4А until the voltage of the battery reaches 0.5 V/battery, i.e. up to 10 V.
9 --, -оС = the temperature measurement is disabled. The available
printed out protocol shows the same data, the identification data of the operator and the time when it is made. At the end of stage 4, when the cut-off voltage of 10 V is reached, the protocol is prepared automatically.
Page 36
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 35
6.9.5 Program 420 stage 5
1 2 3 4
P420:S5:I-CHARGE 2 00:34:38
30.98V 4.00A 4.00AH +060MIN --.-°C
5 6 7 8 9
1 Program number
operation type 4 = maintenance
battery type 20 = enclosed aircraft nickel-cadmium battery 2 stage number = Stage 5 3 stage name = I-Charge 2 4 00:34:38 = current time 5 30.98 V = current voltage battery 6 4.0 А = charging current
The charging current in this program of checking of the aircraft
nickel-cadmium battery restores the nominal capacity within 1h.
It is 4 А for the battery with 4Аh. 7 4.00 Аh The discharged capacity in Аh after 60 minutes 8 0060 MIN The indication of charging time with the positive sign (the
forward counting).
Page 37
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
36 © NorTec Electronics 2014
6.9.6 Program 420 stage 6
1 2 3 4
P420:S6:TOPPING CHARGE 2 00:36:38
29.60V 0.40A 4.019AH +060MIN --.-°C
5 6 7 8 9
1 Program number
operation type 4 = maintenance battery type 20 = enclosed aircraft nickel-cadmium battery.
2 stage number = Stage 6 3 stage name = topping charge 2 4 00:36:38 = current time 5 29.6 V = current battery voltage 6 0.40 А = current charging current
The charging voltage grows and can exceed 31 V. Because of this all the batteries reach the final charging voltage. The capacities of the cells equalize. Extra charging 2 with the current 0.4 А (1/10 of the nominal capacity) is supplied into the cells for 90 minutes.
7 4.019 Аh The capacity counting continues. The main charging plus extra
charging. The total is 4 Аh + 2 minutes of 0.4 А = 0.019 Аh; the total is 4.019 Аh.
8 0060 MIN Indication of the charging time.
After 75 min a beep sounds. а) the voltage of the cells should measured with the measuring cable. b) distilled water should be added into the batteries.
Page 38
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 37
7 Tabulated program progress display
7.1 Program 0. Discharge – any battery type
Stage name Maximum
duration
Program number 0 Operation number 0 Operation type Discharge time and final Discharge
voltage after discharge Battery type all Manufacturing process all Access to valves All types Battery type 0 Minimum battery capacity 0.4 Maximum battery capacity 200 Аh Minimum battery voltage 1.2 Maximum battery voltage 31.2 Maximum number of 26 batteries in a battery
Stage 1 (S1) Discharge Current = x (A) x=0,01 – 40
Stage 1 (S1) Discharge Discharge depth V/battery 1.5
Stage 1 (S1) Discharge Time t=1min – 999min
Page 39
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
38 © NorTec Electronics 2014
7.2 Program 10. Discharge – air-tight aircraft lead acid battery
Stage name Maximum duration
Program number 10 Operation number 0 Operation type Discharge Battery type Lead acid Manufacturing
process Gel / absorbed electrolyte
Access to valves unavailable Battery type 10 Minimum battery capacity 2 Аh Maximum battery capacity 200 Аh Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
Stage 1 (S1) Discharge Current = xС (A) x=0.2 max 6 h
Stage 1 (S1) Discharge Discharge depth V/battery 1.5
Stage 1 (S1) Discharge Time t=1min – 999min
Page 40
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 39
7.3 Program 11. Discharge – Pb battery sealed commercial
Stage name Maximum
duration
Program number 11 Operation number 0 Operation type Discharge Battery type Lead acid Manufacturing process Gel / absorbed
electrolyte
Access to valves unavailable Battery type 11 Minimum battery capacity 2 Аh Maximum battery capacity 200 Аh Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
Stage 1 (S1) Discharge Current = xС (A) x=0.2 Max 6 h
Stage 1 (S1) Discharge Discharge depth V/battery 1.5
Stage 1 (S1) Discharge Time t=1min– 999min
Note: the program is similar to program 10
Page 41
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
40 © NorTec Electronics 2014
7.4 Program 16. Discharge – Pb battery vented commercial
Stage name Maximum
duration
Program number 16 Operation number 0 Operation type Discharge Battery type Lead acid Manufacturing process Liquid electrolyte Access to valves Open Battery type 16 Minimum battery capacity 10 Maximum battery capacity 400 Аh Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries in a battery 15
Stage 1 (S1) Discharge Current = xС (A) x=0.1 max 10 h
Stage 1 (S1) Discharge Discharge depth V/battery 1.5
Stage 1 (S1) Discharge Time t=1min – 999min
Page 42
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 41
7.5 Program 20. Discharge – enclosed aircraft nickel-cadmium battery
Stage name Maximum duration
Program number 20 Operation number 0 Operation type Discharge Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves Open Battery type 20 Minimum battery capacity 4 Аh Maximum battery capacity 50 Аh Minimum battery voltage 1.2 Maximum battery voltage 31.2 Maximum number of batteries 26
Stage 1 (S1) Discharge Current = xС (A) x= 1 max 1.2 h
Stage 1 (S1) Discharge Discharge depth V/battery 1
Stage 1 (S1) Discharge Time t=1min – 999min
Programs 21, 30 are identical to program 20
Page 43
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
42 © NorTec Electronics 2014
7.6 Program 31. Discharge – enclosed aircraft nickel-cadmium battery
Stage name Maximum duration
Program number 31 Operation number 0 Operation type Discharge Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves Open Battery type 20 Minimum battery capacity 4 Аh Maximum battery capacity 50 Аh Minimum battery voltage 1.2 Maximum battery voltage 31.2 Maximum number of batteries 26
Stage 1 (S1) Discharge Current = xС (A) x= 0.4
Stage 1 (S1) Discharge Discharge depth´V/battery 1
Stage 1 (S1) Discharge Time t=1min – 999min
Page 44
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 43
7.7 Program 110. Commissioning – air-tight aircraft lead acid battery
Stage name Maximum duration
Program number 110 Operation number 1 Operation type Commissioning Battery type Lead acid Manufacturing process Gel / absorbed electrolyte Access to valves unavailable Battery type 10 Minimum battery capacity 2 Ah Maximum battery capacity 200 Ah Minimum battery voltage 2 V Maximum battery voltage 30 V Maximum number of batteries 15
Stage 1 (S1) Discharge Discharge current а =xС (A) x= 0.2
Stage 1 (S1) Discharge Discharge depth V/battery 1,5
Discharge voltage <1.5 V Error
Message Use program 410
Stage 2 (S2) I-Charge Current = xС (A) x= 0.4 8h
Stage 2 (S2) I-Charge Umax = V/battery 2.4 8h
Stage 3 (S3) U-Charge Umax = V/battery 2.4 8h
Stage 3 (S3) U-Charge Ia = xС (A) x= 0.04 8h
Note: when the maximum output current of the device is exceeded the maximum time para­meters change accordingly.
Page 45
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
44 © NorTec Electronics 2014
7.8 Program 111. Commissioning – Pb battery sealed commercial
Stage name Maximum duration
Program number 111 Operation number 1 Operation type Commissioning Battery type Lead acid Manufacturing process Gel / absorbed electrolyte Access to valves unavailable Battery type 11 Minimum battery capacity 2 Ah Maximum battery capacity 200 Ah Minimum battery voltage 2 V Maximum battery voltage 30 V Maximum number of batteries 15
Stage 1 (S1) Discharge Discharge current = xС (A) x= 0.2
Stage 1 (S1) Discharge Discharge depth V/battery 1.5
Discharge voltage <1,5 V Error
message. Use program 410!
Stage 2 (S2) I-Charge Current = xС (A) x= 0,36 8h
Stage 2 (S2) I-Charge Umax = V/battery 2.4 8h
Stage 3 (S3) U-Charge Umax = V/battery 2.4 8h
Stage 3 (S3) U-Charge Ia = xС (A) x= 0.04 8h
Note: when the maximum output current of the device is exceeded the maximum time para­meters change accordingly.
Page 46
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 45
7.9 Program 116. Commissioning – Pb battery vented commercial
Stage name Maximum duration
Program number 116 Operation number 1 Operation type Commissioning Battery type Lead acid Manufacturing process Liquid electrolyte Access to valves open Battery type 16 Minimum battery capacity 10 Ah Maximum battery capacity 400 Ah Minimum battery voltage 2 V Maximum battery voltage 30 V Maximum number of batteries 15
Stage 1 (S1) Discharge The electrolyte is filled 20 min or
Enter
Stage 2 (S2) Discharge Discharge current а =xС (A) 0.2
Stage 2 (S2) Discharge Discharge depth V/battery 1.5
Stage 2 (S2) Voltage < 1,5 V Error
message. Use program 416!
Stage 3 (S3) I-Charge Current = xС (A) Х= 0.2
Stage 3 (S3) I-Charge Umax = V/battery 2.4
Stage 4 (S4) U-Charge Umax = V/battery 2.4
Stage 4 (S4) U-Charge Ia = xС (A) x= 0.04
Note: when the maximum output current of the device is exceeded the maximum time para­meters change accordingly.
Page 47
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
46 © NorTec Electronics 2014
7.10 Program 120. Commissioning – enclosed aircraft nickel-cadmium battery
Stage name Maximum duration
Program number 120 Operation number 1 Operation type Commissioning Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves Open Battery type 20 Minimum battery capacity 4 Ah Maximum battery capacity 50 Ah Minimum battery voltage 1.2 V Maximum battery voltage 31.2 V Maximum number of batteries 26
Stage 1 (S1) Precharge Current = xС (A) х = 1
Stage 1 (S1) Precharge Umax = V/battery 1.55 max 1 h
Stage 2 (S2) Discharge 1 Discharge current а = xС (A) х = 1
Stage 2 (S2) Discharge 1 Discharge depth V/battery 1 about 30 min
Stage 3 (S3) I-Charge 1 Current = xС (A) х= 1
Stage 3 (S3) I-Charge 1 Umax = V/battery 1.55 about 60 min
Stage 4 (S4) Topping charge 1 Current = xС (A) х = 0.2 90 min
, Stage 5 (S5) Discharge 2 Discharge current а = xС (A) х = 1
Stage 5 (S5) Discharge 2 Discharge depth V/battery 1 about 60 min
Stage 6 (S6
) I-Charge 2 Current = xС (A) х= 1
Stage 6 (S6) I-Charge 2 Umax = V/battery 1.55 about 60 min
Stage 7 (S7) Topping charge 2 Current = xС (A) х = 0.2 90 min
Stage 7 (S7) Topping charge 2 Electrolyte level equalization Sound signal 75 min
Program 121, 130 and 131 are similar to program 120.
Page 48
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 47
7.11 Program 210. Charging – air-tight aircraft lead acid battery
Stage name Maximum
duration
Program number 210 Operation number 2 Operation type Charging Battery type Lead acid Manufacturing process Gel / absorbed
electrolyte
Access to valves Open Battery type 10 Minimum battery capacity 2 Ah Maximum battery capacity 200 Ah Minimum battery voltage 2 V Maximum battery voltage 30 V Maximum number of batteries 15
For voltages < 2.25 V/battery
Stage 1 (S1) Precharge Umax = V/battery 2.4 2 h
Stage 1 (S1) Precharge Iа = xС (A) 0.36 2 h
Stage 2 (S2) I-Charge I = xС (A) 0.36 8 h
Stage 2 (S2) I-Charge Umax = V/battery 2.4 8 h
Stage 3 (S3) U-Charge Umax = V/battery 2.4 8 h
Stage 3 (S3) U-Charge I а= xС (A) 0.02 8 h
Stage 4 (S4) U-Charge U = V/battery 2.4 2 h
Note: when the maximum output current of the device is exceeded the maximum time para­meters change accordingly.
Page 49
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
48 © NorTec Electronics 2014
7.12 Program 211. Charging – Pb battery sealed commercial
Stage name Maximum
duration
Program number 211 Operation number 2 Operation type Charging Battery type Lead acid Manufacturing process Gel / absorbed
electrolyte
Access to valves unavailable Battery type 11 Minimum battery capacity 2 Ah Maximum battery capacity 200 Ah Minimum battery voltage 2 V Maximum battery voltage 30 V Maximum number of batteries 15
For voltages < 2.25 V/battery
Stage 1 (S1) Precharge Umax = V/battery 2.4 2 h
Stage 1 (S1) Precharge Iа = xС (A) 0.36 2 h
Stage 2 (S2) I-Charge Current = xС (A) 0.36 8 h
Stage 2 (S2) I-Charge Umax = V/battery 2.4 8 h
Stage 3 (S3) U-Charge Umax = V/battery 2.4 8 h
Stage 3 (S3) U-Charge I а= xС (A) 0.02 8 h
Stage 4 (S4) U-Charge Voltage = V/battery 2.4 2 h
Note: when the maximum output current of the device is exceeded the maximum time para­meters change accordingly.
Page 50
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 49
7.13 Program 216. Charging – Pb battery vented commercial
Stage name Maximum
duration
Program number 216 Operation number 2 Operation type Charging Battery type Lead acid Manufacturing process Liquid electrolyte Access to valves open Battery type 16 Minimum battery capacity 10 Ah Maximum battery capacity 400 Ah Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
For voltages > 2.25 V/battery
Stage 1 (S1) Precharge Umax = V/battery 2,4 6 h
Stage 1 (S1) Precharge Iа = xС (A) 0,25 6 h
Stage 2 (S2) I-Charge Current = xС (A) 0,25 8 h
Stage 2 (S2) I-Charge Umax = V/battery 2,4 8 h
Stage 3 (S3) U-Charge Umax = V/battery 2,4 8 h
Stage 3 (S3) U-Charge I а= xС (A) 0,04 8 h
Stage 4 (S4) U-Charge U = V/battery 2,4 2 h
Note: when the maximum output current of the device is exceeded the maximum time para­meters change accordingly.
Page 51
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
50 © NorTec Electronics 2014
7.14 Program 220. Charging – aircraft nickel-cadmium battery
Stage name Maximum
duration
Program number 220 Operation number 2 Operation type Charging Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves Open Battery type 20 Minimum battery capacity 4 Ah Maximum battery capacity 50 Ah Minimum battery Voltage 1,2 V Maximum battery Voltage 31,2 V Maximum number of Batteries 26
Stage 1 (S1) I-Charge 1 Current = xС (A) х = 1 72 min
Stage 1 (S1) I-Charge 1 Umax = V/battery 1.55 72 min
Stage 2 (S2) Topping charge 1 I а= xС (A) х = 0.2 90 min
Stage 2 (S2) Topping charge 1 Electrolyte level Sound signal 75 min
equalization
Page 52
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 51
7.15 Program 221. Charging – aircraft nickel-cadmium battery VARTA
Stage name Maximum
duration
Program number 220 Operation number 2 Operation type Charging Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves Open Battery type 20 Minimum battery Capacity 4 Ah Maximum battery Capacity 50 Ah Minimum battery Voltage 1,2 V Maximum battery Voltage 31,2 V Maximum number of Batteries 26
Stage 1 (S1) I-Charge 1 Current = xС (A) х = 1 72 min
Stage 1 (S1) I-Charge 1 Umax = V/battery 1.55 72 min
Stage 2 (S2) Topping charge 1 I а= xС (A) х = 0.2 120 min
Stage 2 (S2) Topping charge 1 Electrolyte level Sound signal 105 min
equalization
Page 53
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
52 © NorTec Electronics 2014
7.16 Program 230. Charging – aircraft nickel-cadmium battery type SAFT
Stage name Maximum duration
Program number 230 Operation number 2 Operation type Charging Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves open, SAFT Co. Battery type 30 Minimum battery capacity 4 Ah Maximum battery capacity 50 Ah Minimum battery voltage 1,2 V Maximum battery voltage 31,2 V Maximum number of batteries 26
Stage 1 (S1) I-Charge 1 Current = xС (A) х = 1
Stage 1 (S1) I-Charge 1 Umax = V/battery 1.57
Stage 2 (S2) Topping charge 1 I а= xС (A) х = 0.1 240 min
Stage 2 (S2) Topping charge 1 Electrolyte level Sound signal 225 min
equalization
Page 54
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 53
7.17 Program 231. Charging – aircraft nickel-cadmium battery type НКБН
Stage name Maximum
duration
Program number 231 Operation number 2 Operation type Charging Battery type Nickel-cadmium Manufacturing process used plates Access to valves open, type НКБН Battery type 31 Minimum battery capacity 2 Ah Maximum battery capacity 50 Ah Minimum battery voltage 1.2 V Maximum battery voltage 31.2 V Maximum number of batteries 25
Stage 1 (S1) I-Charge 1 Current = xС (A) х = 0.4 180 min
Stage 1 (S1) I-Charge 1 Umax = V/battery unlimited
Stage 2 (S2) Topping charge 1 I а= xС (A) 0.2 120 min
Stage 2 (S2) Topping charge 1 Electrolyte level Sound signal 105 min
equalization
Note: this program can only be used with discharged batteries!
Page 55
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
54 © NorTec Electronics 2014
7.18 Program 410. Maintenance – air-tight aircraft lead acid battery
Stage name Maximum
Duration
Program number 410 Operation number 4 Operation type Maintenance Battery type Lead acid Manufacturing process Gel / absorbed
electrolyte
Access to valves Unavailable Battery type 10 Minimum battery capacity 2 Ah Maximum battery capacity 200 Ah Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
Stage 1 (S1) Discharge Discharge current 0.2 6 h
а = xС (A)
Stage 1 (S1) Discharge Discharge depth 1.5 6 h
V/battery
Stage 2 (S2) Precharge Voltage chargingа 2.4 2 h
Umax = V/battery
Stage 2 (S2) Precharge Current = xС (A) Х=0.4 2 h
Stage 3 (S3) I-Charge 1 Current = xС (A) 0.4 6 h
Stage 3 (S3) I-Charge 1 Umax = V/battery 2.4 6 h
Stage 4 (S4) U-Charge 1 Umax = V/battery 2.4 6 h
Stage 4 (S4) U-Charge 1 I а= xС (A) 0.02 6 h
Stage 4 (S4) U-Charge 1 Umax = V/battery 2.4 2 h
Stage 5 (S5) Discharge Discharge current 0.2 6 h
а = xС (A)
Stage 5 (S5) Discharge Discharge depth 1.5 6 h
V/battery
Stage 6 (S6) I-Charge 2 Current = xС (A) 0.4 6 h
Stage 6 (S6) I-Charge 2 Umax = V/battery 2.4 6 h
Stage 7 (S7) U-Charge 2 Umax = V/battery 2.4 6 h
Stage 7 (S7) U-Charge 2 I а= xС (A) 0.02 6 h
Stage 7 (S7) U-Charge 2 Umax = V/battery 2.4 2 h
Note: the program is identical to program 310 Note: when the maximum output current of the device is exceeded the maximum time para­meters change accordingly.
Page 56
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 55
7.19 Program 411. Maintenance – Pb battery sealed commercial
Stage name Maximum
duration
Program number 411 Operation number 4 Operation type Maintenance Battery type Lead acid Manufacturing process Gel / absorbed electrolyte Access to valves Unavailable Battery type 11 Minimum battery capacity 2 Ah Maximum battery capacity 200 Ah Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
Stage 1 (S1) Discharge Discharge current 0.2 6 h
а = xС (A)
Stage 1 (S1) Discharge Discharge depth 1.5 6 h
V/battery
Stage 2 (S2) Precharge Voltage charging 2.4 2 h
Umax = V/battery
Stage 2 (S2) Precharge Current = xС (A) Х= 0.36 2 h
Stage 3 (S3) I-Charge 1 Current = xС (A) 0.36 6 h
Stage 3 (S3) I-Charge 1 Umax = V/battery 2.4 6 h
Stage 4 (S4) U-Charge 1 Umax = V/battery 2.4 6 h
Stage 4 (S4) U-Charge 1 I а= xС (A) 0.02 6 h
Stage 4 (S4) U-Charge 1 Umax = V/battery 2.4 2 h
Stage 5 (S5) Discharge Discharge current 0.2 6 h
а = xС (A)
Stage 5 (S5) Discharge Discharge depth 1.5 6 h
V/battery
Stage 6 (S6) I-Charge 2 Current = xС (A) 0.36 6 h
Stage 6 (S6) I-Charge 2 Umax = V/battery 2.4 6 h
Stage 7 (S7) U-Charge 2 Umax = V/battery 2.4 6 h
Stage 7 (S7) U-Charge 2 I а= xС (A) 0.02 6 h
Stage 7 (S7) U-Charge 2 Umax = V/battery 2.4 2 h
Note: the program is identical to program 311
Page 57
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
56 © NorTec Electronics 2014
Note: when the maximum output current of the device is exceeded the maximum time para­meters change accordingly.
Page 58
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 57
7.20 Program 416. Maintenance – Pb battery vented commercial
Stage name Maximum
duration
Program number 416 Operation number 4 Operation type Maintenance Battery type Lead acid Manufacturing process Liquid electrolyte Access to valves open Battery type 16 Minimum battery capacity 10 Ah Maximum battery capacity 400 Ah Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
Stage 1 (S1) Discharge Electrolyte level Sound signal
equalization
Stage 1 (S1) Discharge Discharge current 0.05 24h
а = xС (A)
Stage 1 (S1) Discharge Discharge depth 1.5 24 h
V/battery
Stage 2 (S2) Precharge Voltage chargingа 2.4 2 h
Umax = V/battery
Stage 2 (S2) Precharge Ia = xС (A) Х=0.25 2 h
Stage 3 (S3) I-Charge 1 Current = xС (A) Х = 0.25 8 h
Stage 3 (S3) I-Charge 1 Umax = V/battery 2.4 8 h
Stage 4 (S4) U-Charge 1 Umax = V/battery 2.4 8 h
Stage 4 (S4) U-Charge 1 I а= xС (A) 0.04 8 h
Stage 4 (S4) U-Charge 1 Umax = V/battery 2.4 8 h
Stage 5 (S5) Discharge Discharge current 0.05 24 h
а = xС (A)
Stage 5 (S5) Discharge Discharge depth 1.5 24 h
V/battery
Stage 6 (S6) I-Charge 2 Current = xС (A) 0.25 8 h
Stage 6 (S6
) I-Charge 2 Umax = V/battery 2.4 8 h
Stage 7 (S7) U-Charge 2 Umax = V/battery 2.4 8 h
Stage 7 (S7) U-Charge 2 I а= xС (A) 0.04 8 h
Stage 8 (S8) U-Charge 3 Umax = V/battery 2.4 2 h
Note: when the maximum output current of the device is exceeded the maximum time parameters change accordingly.
Page 59
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
58 © NorTec Electronics 2014
7.21 Program 420. Maintenance – enclosed aircraft nickel-cadmium battery
Stage name Maximum
duration
Program number 420 Operation number 4 Operation type Maintenance Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves open Battery type 20 Minimum battery capacity 4 Ah Maximum battery capacity 50 Ah Minimum battery voltage 1,2 V Maximum battery voltage 31,2 V Maximum number of batteries 26
Stage 1 (S1) Discharge 1 Discharge current х = 1 max. 6 min.
а = xС (A)
Stage 1 (S1) Discharge 1 Discharge depth 1 max. 6 min.
V/battery
Stage 2 (S2) I-Charge 1 Current = xС (A) х = 1 max. 72 min.
Stage 2 (S2) I-Charge 1 Umax = V/battery 1.55 max. 72 min.
Stage 3 (S3) Topping charge 1 Current = xС (A) х = 0.1 120 min.
Stage 4 (S4) Discharge 2 Discharge current х = 1 1 h
а = xС (A)
Stage 4 (S4) Discharge 2 Discharge depth 0.5 1 h
V/battery
Stage 5 (S5) I-Charge 2 Current = xС (A) х = 1 1 h
Stage 5 (S5) I-Charge 2 Umax = V/battery 1.55 1 h
Stage 6 (S6) Topping charge 2 Current = xС (A) х = 0.2 90 min.
Stage 6 (S6) Topping charge 2 Measurement of Sound signal 75 min.
voltages of individual batteries
Stage 6 (S6) Topping charge 2 Electrolyte level Sound signal 75 min.
equalization
Page 60
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 59
7.22 Program 421. Maintenance – enclosed aircraft nickel-cadmium battery type VARTA
Stage name Maximum
duration
Program number 421 Operation number 4 Operation type Maintenance Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves open, type VARTA Battery type 21 Minimum battery capacity 4 Ah Maximum battery capacity 50 Ah Minimum battery voltage 1.2 V Maximum battery voltage 31.2 V Maximum number of batteries 26
Stage 1 (S1) I-Charge 1 Current = xС (A) х = 1 72 min.
Stage 1 (S1) I-Charge 1 Umax = V/battery 1.55 72 min.
Stage 2 (S2) Topping charge 1 Current = xС (A) х = 0.2 120 min.
Stage 2 (S2) Topping charge 1 Measurement of Sound signal 105 min.
voltages of individual batteries
Break 30 min.
Stage 4 (S4) Discharge Discharge current х = 1 54 min.
а = xС (A)
Stage 4 (S4) Discharge Discharge depth 1 54 min.
V/battery
Stage 4 (S4) Discharge Measurement of Sound signal 54 min.
voltages of individual batteries
Page 61
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
60 © NorTec Electronics 2014
7.23 Program 430. Maintenance – enclosed aircraft nickel-cadmium battery type SAFT
Stage name Maximum
duration
Program number 430 Operation number 4 Operation type Maintenance Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves open, type SAFT Battery type 30 Minimum battery capacity 4 Ah Maximum battery capacity 50 Ah Minimum battery voltage 1.2 V Maximum battery voltage 31.2 V Maximum number of batteries 26
Stage 1 (S1) I-Charge 1 Current = xС (A) х = 1 75 min.
Stage 1 (S1) I-Charge 1 Umax = V/battery 1.57 75 min.
Stage 2 (S2) Topping charge 1 Current = xС (A) х = 0.1 240 min.
Stage 2 (S2) Topping charge 1 Measurement of Sound signal 225 min. voltages of individual
batteries
Break 30 min
Stage 4 (S4) Discharge Discharge current х = 1 54 min.
а = xС (A)
Stage 4 (S4) Discharge Discharge depth 1 54 min.
V/battery
Stage 4 (S4) Discharge Measurement of Sound signal 54 min.
voltages of individual batteries
Page 62
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 61
7.24 Program 431. Maintenance – enclosed aircraft nickel-cadmium battery type НКБН
Stage name Maximum
duration
Program number 431 Operation number 4 Operation type Maintenance Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves open, type НКБН Battery type 31 Minimum battery capacity 4 Ah Maximum battery capacity 50 Ah Minimum battery voltage 1.2 V Maximum battery voltage 31.2 V Maximum number of batteries 25
Stage 1 (S1) Discharge 1 Discharge current х = 0.4 30 min.
а = xС (A)
Stage 1 (S1) Discharge 1 Discharge depth 1 30 min.
V/battery
Break 30 min.
Stage 2 (S2) I-Charge 1 Current = xС (A) х = 0.4 30 min.
Stage 2 (S2) I-Charge 1 Umax = V/battery unlimited 30 min.
Stage 3 (S3) Topping charge 1 Current = xС (A) х = 0.2 120 min.
Stage 3 (S3) Topping charge 1 Measurement of Sound signal 105 min
voltages of individual batteries
Note: this program can only be used with discharged batteries!
Page 63
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
62 © NorTec Electronics 2014
7.25 Program 510. Direct current charging – air-tight aircraft lead acid bat­tery
Stage name Maximum
duration
Program number 510 Operation number 5 Operation type Direct current charging Battery type Lead acid Manufacturing process Gel / absorbed electrolyte Access to valves Unavailable Battery type 10 Minimum battery capacity 2 Ah Maximum battery capacity 200 Ah Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
Stage 1 (S1) I-Charge I free selection
Stage 1 (S1) I-Charge Umax = V/battery 2.4
Stage 1 (S1) I-Charge time free selection
Page 64
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 63
7.26 Program 511. Direct current charging – Pb battery sealed commercial
Stage name Maximum
duration
Program number 511 Operation number 5 Operation type Direct current charging Battery type Lead acid Manufacturing process Gel / absorbed electrolyte Access to valves Unavailable Battery type 11 Minimum battery capacity 2 Ah Maximum battery capacity 200 Ah Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
Stage 1 (S1) I-Charge Current free selection
Stage 1 (S1) I-Charge Umax = V/battery 2.4
Stage 1 (S1) I-Charge time free selection
Note: the program is identical to program 510.
Page 65
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
64 © NorTec Electronics 2014
7.27 Program 516. Direct current charging – Pb battery vented commercial
Stage name Maximum
duration
Program number 516 Operation number 5 Operation type Direct current charging Battery type Lead acid Manufacturing process Liquid electrolyte Access to valves open Battery type 16 Minimum battery capacity 10 Ah Maximum battery capacity 400 Ah Minimum battery voltage 2 V Maximum battery voltage 30 V Maximum number of batteries 15
Stage 1 (S1) I-Charge Current free selection
Stage 1 (S1) I-Charge Umax = V/battery 2.4
Stage 1 (S1) I-Charge time free selection
Page 66
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 65
7.28 Program 520. Direct current charging – enclosed aircraft nickel­cadmium battery
Stage name Maximum
duration
Program number 520 Operation number 5 Operation type Direct current charging Battery type Nickel-cadmium Manufacturing process Fused electrodes Access to valves open Battery type 20 Minimum battery capacity 4 Ah Maximum battery capacity 50 Ah Minimum battery voltage 1.2 Maximum battery voltage 31.2 Maximum number of batteries 26
Stage 1 (S1) I-Charge Current free selection
Stage 1 (S1) I-Charge Umax = V/battery free selection
Stage 1 (S1) I-Charge time free selection
Program 521, 530 and 531 are identical.
Page 67
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
66 © NorTec Electronics 2014
7.29 Program 800. Constant power supply – without battery
Stage name Maximum
duration
Program number 800 Operation number 8 Operation type Source of voltage Battery type unavailable Manufacturing process ­Access to valves - Battery type - Minimum battery capacity ­Maximum battery capacity ­Minimum battery voltage ­Maximum battery voltage ­Maximum number of batteries -
Arbitrary
Stage 1 (S1) voltage 0.01, 40 V max. 40 А
Stage 1 (S1) current 0.01, 40 А max. 40 V
Page 68
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 67
7.30 Program 810. Constant power supply – air-tight aircraft lead acid bat­tery
Stage name Maximum
duration
Program number 810 Operation number 8 Operation type Source of voltage Battery type Lead acid Manufacturing process Gel / absorbed electrolyte Access to valves Unavailable Battery type 10 Minimum battery capacity 2 Ah Maximum battery capacity arbitrary Minimum battery voltage 2 Maximum battery voltage 30 Maximum number of batteries 15
Stage 1 (S1) Direct extra charging mode Ufloat = V/battery 2.25 arbitrary
Stage 1 (S1) Direct extra charging mode Imax=xC(A) х = 0.4
Page 69
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
68 © NorTec Electronics 2014
7.31 Program 811. Constant power supply – Pb battery sealed commercial
Stage name Maximum
duration
Operation number 811 Operation type 8 Battery type Source of voltage Manufacturing process Lead acid Access to valves Gel / absorbed electrolyte Battery type Unavailable Minimum battery capacity 11 Maximum battery capacity 2 Ah Minimum battery voltage arbitrary Maximum battery voltage 2 V Maximum number of batteries 30 V Operation number 15
Stage 1 (S1) Direct extra charging mode Ufloat = V/battery 2.25 arbitrary
Stage 1 (S1) Direct extra charging mode Imax=xC(A) х = 0.,4
Note: the program is identical to program 810
Page 70
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 69
7.32 Program 816. Constant power supply – Pb battery vented commercial
Stage name Maximum
duration
Operation number 816 Operation type 8 Battery type Source of voltage Manufacturing process Lead acid Access to valves Liquid electrolyte Battery type open Minimum battery capacity 16 Maximum battery capacity 10 Ah Minimum battery voltage arbitrary Maximum battery voltage 2 Maximum number of batteries 30 Operation number 15
Stage 1 (S1) Direct extra charging mode Ufloat = V/battery 2.25 arbitrary
Stage 1 (S1) Direct extra charging mode Imax=xC(A) х = 0.25
Page 71
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
70 © NorTec Electronics 2014
7.33 Program 820. Constant power supply – enclosed aircraft nickel­cadmium battery
Stage name Maximum
duration
Operation number 820 Operation type 8 Battery type Source of current Manufacturing process Nickel-cadmium Access to valves Fused electrodes Battery type Open Minimum battery capacity 20 Maximum battery capacity 4 Ah Minimum battery voltage 40 Ah Maximum battery voltage 1.2 Maximum number of batteries 31.2 Operation number 26
Stage 1 (S1) Direct current charging float Current=xC(A) Х=0.025 arbitrary
Stage 1 (S1) Direct current charging float Ufloat = V/battery 1.55
Page 72
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 71
8 Replacement of printing paper and ribbon
8.1 Ribbon replacement
Name: ribbon (a pack of 2 pcs.)
HTZ/TKZ: 101 033 111 000 The ribbon is replaced in the following procedure:
- disconnect the device 220 V from the mains
- remove 2 knurl-head screws from the head of the printer and extract it from the body
- lift the front panel of the printer and remove it
- extract the worn ribbon from the opened printer
- pull out several centimeters from the printer
- insert the paper into the ribbon and insert the ribbon by lightly pressing the printer. The position of adjustment of the printing ribbon: a free portion of the ribbon should be down, the wheel of the trans­fer mechanism should be leftward
- insert the front panel into the printer
- insert the printer into its body
- tighten 2 knurl-head screws
- connect the device to the 220 V mains
- perform the self-test of the printer: the printer prints a test print-out of 3 lines when the device is switched on by pressing the paper feed key
- switch off the device
8.2 Replacement of printing paper
Name: printing paper (a pack of 6 pcs.)
HTZ/TKZ: BT2000.40071 The paper is replaced in the following procedure:
- disconnect the device from the 220 V mains
- remove 2 knurl-head screws from the printer
- extract the printer from the body
- lift the front panel of the printer and remove it
- pull the ribbon out of the opened printer
- pull the used paper roll from the printer and tear the paper slightly
- pull the paper protruding from the printer in the direction of the paper feed
- insert the printing paper as it is shown on the body of the printer and move it as it is shown on the body of the printer
- insert the printer into its body
- switch on the device
- actuate the paper feed and use the free hand to insert the printing paper into the receiving slot of the printer until the paper is moved by the printer
- insert the printing paper into the ribbon and then insert the ribbon into the printer without pressure. The position of adjustment of the printing ribbon: a free portion of the ribbon should be down, the wheel of the transfer mechanism should be leftward
- insert the front panel into the printer
- insert the printer into its body
- tighten 2 knurl-head screws
- connect the device to the 220 V mains
- perform the self-test of the printer: the printer prints a test print-out of 3 lines when the device is switched on by pressing the paper feed key
Page 73
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
72 © NorTec Electronics 2014
- switch off the device
Page 74
CHARGER / DISCAHRGER UL10 OPERATION MAUNAL
© NorTec Electronics 2014 73
Protection according to DIN 34