are connected to the LLZ cabinet.
The Remote Control unit is intended for installation in the tower or a technical room to give
remote control and status indication. An optional Remote Slave panel can be used if control
and status indication is required i n ad ditional positions.
The RMM system comprises the built-in RMS system in the LLZ cabinet, and a data program
running on a standard PC. The PC can be connected directly to the cabinet , or by mo dems
through leased or switched telephone lines.
1.2Abbreviations
ACAlternating Current
ADCAnalog to Digital Converter
AGCAutomatic Gain Control
CLCourse Line
CLRClearance
COUCourse
CPUCentral Processing Unit
CSCourse Sector
DACDigital to Analog Converter
DCDi rect Current
DDMDifference in Depth of Modulation
DFDifference Frequency
DLDC Loop
DSDisplacement Sensitivity
DSPDigital Signal Processor
EEPROMElectrically Erasable Progammable Read Only Memory
EMCElectroMagnetic Compatibility
EMIElectroMagnetic Interference
EPROMErasable Programmable Read Only Memory
FFTFast Fourier Transfor m
FIFOFirst-In-First-Out
FPGAField Programmable Gate Array
GPAGlidepath Power amplifier Assembly
I/FInterFace
I²CInter Integrated Circuit
IICSame as I²C
ILSInstrument Landing System
LEDLight Emitting Diode
LFLow Frequency
LLZLocalizer
LPALocalizer Power amplifier Assembly
LRULine Replacable Unit
MCUMonitor Combiner Unit
NAVNAVigation signals
NFNear Field
PCPersonal Computer
RAMRandom Access Memory
RFRadio Frequency
RMMRemote Maintenance Monitor
RMSRemote Monitoring System
ROMRead Only Memory
RTCReal Time Clock
SCStation Control
SDMSum in Depth of Modulation
SPASame Parameter Alarm
SRAMStatic Random Access Memory
STBStandby
SWSoftWare
TRMTeRMinator
TXTransmitter
UARTUniversal Asynchronous Receiver Transmitter
Frequency range108-112 MHz
Frequency tolerance+
Output power (CSB)5-15 W adjustable
Harmonic radiation2.5 uW maximum
RF difference frequency (2-freq. only)10 kHz +
Spurious25 uW maximum
Output power stability+
CSB/SBO stability+
Modulator
Modulation depth 90/150 Hz20%
Adjustable range 10-25%
SDM stability+
DDM stability+
Frequency tolerance+
Total harmonic distortion (90/150 Hz)1% maximum
Phase locking (90 Hz to 150 Hz)5° maximum ref 150 Hz
SBO phaser adjustment range+
OPERATING MANUALNORMARC 7013
0.002%
2 kHz
0.2 dB
0.3 dB / +-5°
0.3% SDM
0.1% DDM
0.05 Hz
10°
Identity keyer
Modulation frequency1020 Hz +
10 Hz
Modulation depth5-15% adjustable
Distortion6% maximum
Speed of identification7 Words/min. approx.
Monitoring
Alarm functionsAdjustable range (*)
RF power reduction1-5 dB
Change of nominal CL+
Change of nominal DS from nominal value+
Change of nominal CLR (2-freq. only)+
Change of nominal NF+
Change of nominal SDM+
Difference frequency (2-f req. only)+
4-25 uA
10-50 uA
10-60 uA
4-25 uA
2-6% SDM
2-5 kHz
Total period of radiation out of tolerance1-10 sec
Additional NF delay0-20 sec
Line break, ILS-remote control (disable
optional).
Identification lost or continuous (optional)
Monitor input levels
RF power reduction40-75% of alarm limit
Change of nominal CL40-75% of alarm limit
Change of nominal DS40-75% of alarm limit
Change of nominal CLR40-75% of alarm limit
Change of nominal NF40-75% of alarm limit
Change of nominal SDM40-75% of alarm limit
Difference frequency40-75% of alarm limit
Mains failure
Remote control
Data transmission medium2-wire line, 600 ohm
Data modulationserial, FSK
Transmitter level-10dBm +
Receiver dynamic range-10dBm to -34dBm
5 dB
0,2 dB
1 uA
0,5% SDM
2 dB
(*): Asymmetrical limits are possible
2.2Environmental characte ristics
Operating temperature:-10 to +55 °
Storage temperature- 30 to +60 °
2.3EMC characteristics
EMR:EN 55022 class B
Spurious and harmonics:CISPR 22
2.4Mechanical character istics
Dimensions: (HxBxW)
ILS Rack:1020x600x500 mm
Remote control:129x71x170 mm
Slave panels:129x41x170 mm
Weight:85 kg - 95 kg depending on model
The ILS rack is wall mounted. The remote control an d slave panel s fit a st andard 3U (132mm)
high 19" subrack.
The power ON/OFF switches are located on front of the main cabinet. Adjacent to the
switches are GND sockets for connecting wrist st rap to ensure ESD-protected environment
when performing maintenance operations.
remote control), the ILS wil l turn of f al l active trans mitters and not allow them to
be turned on before the interlock condition is removed. This signal overrides
manual mode of operation.
3.2.2Pushbuttons
OPERATING MANUALNORMARC 7013
ON/OFF
Used to:
•Toggle the ILS on/off.
Valid when:
•The LOCAL/REMOTE switch must be in LOCAL position, and
•The interlock signal is not active if the ILS is configured for
interlock.
CHANGEOVER
Used to:
•Toggle the coax relay and transmitters between TX1/
TX2 as the active transmitter(s).
Valid when:
•The LOCAL/REMOTE switch is in LOCAL position, and
•The ILS is «on» when the MANUAL/AUTO switch is in
AUTO position.
MAIN SELECT
Used to:
•Toggle between TX1/TX2 as the main transmitter.
Valid when:
•The LOCAL/REMOTE switch is in LOCAL position.
STBY
ON/
OFF
HBK604-1
TX1
NORMAL
ON/
OFF
HBK603-1
CHANGE
OVER
MAIN
TX2
COURSE TX1/TX2
Used to:
•Toggle the COURSE TX1/TX2 on/off.
Valid when:
•LOCAL/REMOTE switch is in LOCAL position, and
•MANUAL/AUTO switch is in MANUAL position, and
•The interlock signal is active if the ILS is configured for interlock.
CleaRance TX1/TX2
Used to:
•Used to toggle the CLEARANCE TX1/TX2 on/off.
Valid when:
•CLEARANCE transmitters are present, and
•LOCAL/REMOTE switch is in LOCAL position, and
•MANUAL/AUTO switch is in MANUAL position, and
•The interlock signal is active if the ILS is configured for interlock.
•Select between LOCAL and REMOTE mode of operation. REMOTE mode of
operation will inhibit use of the local push-button (as described above) and
the use of the local seri al communication port for enter ing RMS access level
2 and 3. LOCAL mode of operation will inhi bit use o f the remote con trol pushbuttons (ON/OFF, CHANGEOVER) and the use of the remote serial communication ports for entering RMS access level 2 and 3.
Valid when:
•The LOCAL/REMOTE switch is in LOCAL position.
MANUAL / AUTO
Used to:
•Select between AUTOMATIC and MANUAL mode of operation.
Valid when:
•The LOCAL/REMOTE switch is in LOCAL position.
WRITE PROTECT
Used to:
•Prevent changing of alarm limits and other ILS parameter adjustments. Setting this switchl o ck in vertical/horizontal po sition will prevent/not prevent entering of access level 3
Valid when:
•Always
HBK611-1
LOCAL
REMOTE
HBK609-1
MANUAL
AUTO
HBK610-1
WRITE
PROTECT
INTERLOCK OVERRIDE
Used to:
•Override the interlock input to allow testing of transmitters. Using this
function will allow the local panel pushbuttons to be used (ON/OFF,
CHANGEOVER and the individual TX on/off keys).
Valid when:
•The LOCAL/REMOTE switch is in LOCAL position, and
•The MANUAL/AUTO switch is in MANUAL position.
INTERLOCK
OVERRIDE
HBK612-1
3.2.4System Status Indications
SERVICE
Used to:
•Indicate that the ILS is currently in SERVICE mode of operation.
This will also set the remote control to alarm state.
Activated by:
•EXTERNAL SERVICE line forced low (by LF1223A set in service
condition), or
PARAM
DISAGR
BATT
IDENT
MAINT
STBY
HBK608-1
SERVICE
ALARM
WARNING
NORMAL
•RMS in access level 2 or 3, or
•The LOCAL/REMOTE switch is in LOCAL position, or
•The MANUAL/AUTO switch is in MANUAL position, or
•MANUAL mode entered from RMS, or
•Mismatch between the WRITE PROTECT switch and configuration strap setting on U6
(strap position 6-15).
The front panel menu system includes a 20x4character LCD with 7 dedicated pushbutton
keys. The system is organised as a top-down menu tree-st ructure where the main menu is the
default entering point.
READ
HBK550-1
The main menu screen has three menu-options. The options are div ided into the th ree access
levels; level 1 with basic readout functions, level 2 with test and control func ti ons and level 3
with permanent system parameter setti ngs and configuration changes.
>Read syst para L1
Local settings L2
Syst settings L3
Figure 3-4 The top level main menu screen.
3.3.2Handling the different menuscreens.
The menu system consists of five basic screen types; menu-screens, readout-screens, toggle-screens, input/sto ring-screens, and the quick read screens.
3.3.2.1 The menu screen
The menu screen is controlled by the key-buttons: <PREV> <NEXT> <ENTER> and <ESC>.
The menu options are formatted as a l ong li st wher e eac h option hol ds one char act er li ne. If a
menu screen includes more than four options, a small arrow-indicator will be displayed in the
lower-right corner. This informs the user that more than four menu options are available. The
blinking arrow cursor left to the menu-t ext is controlled by the <NEXT> and <PREV> keys.
The <ENTER> key activates the menu options left to the arrow-cursor.
The <ESC> key brings the user one menu-level upwards or back in the menu-tree-structure.The menu scrolls down if the user pushes the <NEXT> key when the arrow-cursor is at
the lower line and the arrow indicator is displayed in the lower-right corner. The menu will in
the same way scroll upwards by pushing the <PREV>-key in a corresponding situation.
3.3.2.2 The readout screen
The readout screens display the various monitor and maintenance parameter readings. The
screens are entered by stepping through the menu options. The <ESC> key brings the user
back to the last menu-screen again. The parameter readings are continuously updated.
CL readout M1 M2
DDM.: 0.1% 0.1%
Al U +1.5% +1.5%
Wa U -1.5% -1.5%
Figure 3-5 A typical readout screen: CL DDM.
3.3.2.3 The toggle screens
The toggle screens offer the user to choose between two or more options. The options text is
enclosed by two blinking square-brackets. The plus «+» and minus «- » keys step through the
various options. The <ENTER> key activates the chosen toggle-option. The <ESC> keys
leaves the screen without activating any of the opti ons.
Figure 3-6 A typical toggle screen: TX1 90Hz on/off.
3.3.2.4 Input/Store screen
The user can change the various parameter alarm and warning settings in the input/storingscreens. A virtual cursor is controlled by the <PREV> and <NEXT> keys. An input field number will blink if the cursor is moved to the correct positi on. The blinking limit val ues can now be
changed by pushing the plus «+» and minus «-» keys. If the input field is digital, the valid
options will be «False» and «T rue» when the plus «+» and minus «-» keys are pushed.
When all the input numbers in the screen are changed to th e preferred new valu es, the cursor
must be moved to the (STORE)-field in t he upper left corner. When the cursor is placed at this
(STORE)-field, the brackets will start blin king. The <ENTER> key will now store the new values on the screen in the ILS. The <ESC> key leaves the screen unchanged.
CL <store> M1 M2
DDM.: 0.1% 0.1%
Al U +1.5% +1.5%
Wa U -1.5% -1.5%
Figure 3-7 A typical input/store screen: CL DDM
3.3.2.5 The quick read screens
The quick read screens are access by pushing the <QUICK READ> key. The key toggles
between the original menu tree and the quick read screens. The <QUICK READ> button can
be pushed at all times, not changing the original menu-screens. The quick read screens
include only the basic monitor readings. The <PREV> and <NEXT> keys step through the
quick read screens.
CL %DDM %SDM RFv
M1 0.1 40.0 3.0
M2 0.1 40.0 3.0
Figure 3-8 A typical quick read screen: CL.
3.3.3The access levels.
The level 3 menu options must be entered by typing a 4 character password. The level 2
menu can be entered as an ordinary menu-option.
3.3.3.1 Level 2
To enter the level 2 menu option, the front panel switches must be in the LOCAL and MANUAL positions. The remote panel acces s switch must also be in ACCESS position. The user is
denied access if the switches are not correctly set.
Local switch on
Front panel in
REMOTE position
Push <ESC>
Figure 3-9 An access denial screen: Switch in REMOTE position.
3.3.3.2 Level 3
To enter the level 3 menu option, the front panel switches must be as for level 2 access. In
addition the WRITE PROTECT switch must be in «ACCESS» position. The user must key a 4
character password to enter level 3 access. The access-screen is operated similarly to an
input/store-screen. The blink ing cursor is moved by the <PREV> and <NEXT> keys. A blinking input-field is changed by the plus «+» and minus «-» keys.
When the correct password is entered, the cursor must be moved to the <level 3 access>field. The brackets will then start blinking. The <ENTER> key will now bring the user to the
level3 options. The <ESC> key leaves the screen back to the main menu.
LEVEL ACCESS 3
m m
m m
<level3 access>
Figure 3-10 Level 3 access screen.
3.3.3.3 Leaving the access levels
If the user enters level 2 or 3, an extra menu line is added in the main menu screen. This is a
leave current access level option. The user must enter this menu option to quit the current
access level - typically when leaving th e ILS.
The access levels are also left when the front panel switches are changed to an illegal position. (e.g. switch in REMOTE position)
>Read syst para L1
Local settings L2
Syst settings L3
Leave access 3
Figure 3-11 The top level main menu screen: User i n level 3 access.
3.3.4Function listing
The menu system includes the following functionality.
3.3.4.1 Level 1
Readout of all the monitor values, warning and alarm limits.
Readout of all the maintenance values and warning limits.
Readout of all the delays.
Readout of all the TX1 and TX2 configurations.
Ident. configurations with morse si gnal sett ings to normal, cont inuous, TST or of f on TX1/TX2.
Local morse speaker on/off on the transmitters.
TX1 and TX2: test-signal 90 Hz or 150 Hz dominance on/off.
TX1 and TX2: 90/150 Hz on/off.
Communications.
3.3.4.3 Level 3
Settings of all the monitor warning and alarm li mits.
Settings of all the maintenance warning limits.
Settings of all the delays.
Settings of all the TX1 and TX2 configurations.
Communications
Bitrate local port
Bitrate rem1 port
Bitrate rem2 port
Init modem
Mon. parameters
Mon. 1&2 alarm par
Mon. 1&2 warn. par
Maint. val. & lim.
Monitor 1
Monitor 2
LF generator 1
LF generator 2
Mon 1 frontend
Mon 2 frontend
Oscillator 1
Oscillator 2
Course TX1 PA
Course TX2 PA
Clearance TX1 PA
Clearance TX2 PA
DC supplies
System delays
Mon1 delays
Mon2 delays
Maint delays
TX configuration
TX1 configuration
TX2 configuration
Test signals TX1
Test signals TX2
HBK595-2
INSTRUMENT LANDING SYSTEM
NORMARC 7013
Page 33
INSTRUMENT LANDING SYSTEM
4Tower Equipment
4.1Remote Control Operation
PARAM
OPERATING MANUALNORMARC 7013
SILENCE
DISAGR
BATT
IDENT
MAINT
STBY
ALARM
WARNING
NORMAL
ON /
OFF
RMM
ACCESSGRANT
CHANGE
OVER
DENY
HBK566-1
Figure 4-1 RF1242A front panel.
4.1.1Glossary
INTERLOCK When an ILS is in active interlock mode (the interlock signal is an input to the
remote control), the ILS wil l turn of f al l active trans mitters and not allow them to
be turned on before the interlock condition is removed. This signal overrides
manual mode of operation.
•The LOCAL/REMOTE switch must be in REMOTE position, and
•The interlock signal is not active if the ILS is configured for interlock.
CHANGE OVER
Used to:
•Toggle the coax relay and transmitters between TX1/TX2 as the active transmitte r(s).
Valid when:
•The LOCAL/REMOTE switch on the ILS is in REMOTE position, and
•The ILS is «on» when he MANUAL/AUTO switch is in AUTO position.
SILENCE
Used to:
•Turn off the audio alarm on the remote control (caused by remote status change from
NORMAL to ALARM). Also used for lamp test (all lamps are illuminated when this pushbutton is held down).
Valid when:
•Always.
4.1.3Switches/Switchlocks
GRANT/DENY
Used to:
•Give access grant (access level 2/3) for RMS control of the ILS when the switch is in
GRANT position. NOTE! This switch may be override by configuration straps on U6 on
TC1216A (strap position 7-14). When this switch is in the DENY position no RMS access
(at access level 2/3) is allowed.
Valid when:
•Always.
4.1.4System Status Indications
ALARM
Used to:
•Indicate that the ILS has detected an alarm condition.
Activated by:
•One or more alarms present on the ILS, or
•Communic a ti on failure with ILS.
WARNING
Used to:
•Indicate that the ILS has detected one or more warning conditions.
Activated by:
•Warni ng condition(s) detected by RMS.
NORMAL
Used to:
•Indicate that no alarm conditions are detected by the ILS.
•Indicate that access level 2/3 has been granted for RMS control.
Activated by:
•Grant/Deny switchlock is in GRANT position.
4.2Remote Slave Operation
SILENCE
NORMARC 7013
INSTRUMENT LANDING SYSTEM
ALARM
WARNING
NORMAL
ON/
OFF
HBK599-1
Figure 4-2 SF1344A front panel.
4.2.1Glossary
INTERLOCK When an ILS is in active interlock mode (the interlock signal is an input to the
remote control), the ILS wil l turn of f al l active trans mitters and not allow them to
be turned on before the interlock condition is removed. This signal overrides
manual mode of operation.
•The LOCAL/REMOTE switch must be in REMOTE position, and
•The interlock signal is not active if the ILS is configured for interlock.
SILENCE
Used to:
•Turn off the audio alarm on the slave panel (generated by remote control). Also used for
lamp test (all lamps are illuminated when this push- button is held down).
Valid when:
•Always.
4.2.3System Status Indications
ALARM
Used to:
•Indicate that the ILS has detected an alarm condition.
Activated by:
•One or more alarms present on the ILS, or
•Communic a ti on failure with ILS.
WARNING
Used to:
•Indicate that the ILS has detected one or more warning conditions.
Activated by:
•Warni ng condition(s) detected by RMS.
NORMAL
Used to:
•Indicate that no alarm conditions are detected by the ILS.
Activated by:
•No alarms present, and
•Communication with ILS is OK.
4.3Interlock Switch Operation
The interlock function is used when mutual exclusive operation on different ILS systems is
required. An active interlock for a station shall prevent transmission of ILS signals. The interlock signal is connected to the remote contr ol RC1241 and transmitted to the ILS through the
remote control connection.
The interlock function on the ILS is enabled/disabled from configuration settings on the
TC1216A. The rest of this chapter describes an ILS configured for interlock.
When an active ILS receives an active interlock signal, the ILS performs a shutdown. If the
interlock signal is deactivated, the ILS will automatically be turned on after a 20 second delay
period.
For maintenance purposes, an interlock override function has been implemented. This function makes it possible to operate the ILS locally in manual mode with an active interlock signal. To activate the interlock override function the foll owing conditions must be met:
•LOCAL/REMOTE switch must be in LOCAL position,
•AUTO/MANUAL switch must be in MANUAL position and
•INTERLOCK OVERRIDE switchlock must be set in horizontal position.
The remote control will always be set in ALARM state when the inter lock override function is
activated.
This chapter is the user manual for the Remote Maintenance and Moni toring (RMM) s oftware.
Version 3.9 of the RMM software is covered in this chapter. This version can be used to con-
nect to ILS cabinets with version 5 up to and including version 12 of the RMS software.
5.1Introduction
This paragraph gives a brief description of this user manual, the program and the system
requirements.
5.1.1Description
This manual describes how to use the NM70x x remote monitoring software. The remote monitoring software consists of software running in the ILS rack and on software running on IBM/
PC's and compatibles. The purpose of the system is to retrieve status and measurements
from the ILS rack and to change ILS operation and parameters from a remote site. The
retrieved data can be displayed and further processed by programs on the PC.
For example, the ILS data can be
•Displayed on the screen.
•Saved on disk for later use.
•Exported to other programs for graphi cal display of parameter variati ons.
5.1.2Available versions
The NM7000 RMM program is available in a DOS version and a 32 bit version for Windows
95/NT. Both versions are distributed together . Depending on the operating syst em installed on
the PC, the user must install one of thes e versions.
5.1.3System requirements
The PC running the DOS version of the software must be equipped with:
•386DX33 based computer
•DOS 3.0 or later
•a hard disk
•4 Mbyte RAM
•HIMEM installed
•a RS-232 serial port connected to a Hayes compatible 2400 baud modem
•a printer port & printer, if printouts are desired
For Windows 95 and NT the following are acceptable minimum configurations:
95: 486x33 and 8 MB RAM
NT: Minimum 486x66 and 16-20 MB RAM
A recommended configuration for running the DOS version in cludes the following:
•a 486 based computer or better
•a mathematical co-processor (486DX internal is satisfactory)
If a 386- or a slow 486-based PC is used, it is recommended to drop the bitrate to 2400 or
4800 bps.
If a mouse is available, it is supported by the RMM software.
5.1.4Running under Microsoft Windows 3.xx
Neither the DOS version nor the Win95 /NT v ersion of the RMM program runs under Windows
3.xx. The DOS version can, of course, be start ed on a PC with Windows 3.xx inst alled as long
as Windows is not also started.
If the RMM-software is run ning on a sl ow PC with Wind ows 95 or NT, avoid to st art up sever al
other applications. This to ensure sufficient processor-time is available for data processing in
the NM 7000 RMM.
5.1.5How to use this manual
The user interface of the RMM software is very similar to the one used in Microsoft Windows.
If you are familiar with using Windows programs, you will easi ly understand how to operate
this program, and you can skip Chapter 5.2. The next chapters describe each feature of the
program. You may want to read these chapters to learn how use the RMM software.
5.1.6Starting the RMM software
To install the RMM PC Software, create a directory named NM7000 on the hard disk. Then
copy the contents of the NMS152A (DOS) or NMS152B (Win95/NT) RMM PC SW floppy to
this directory.
To start the program, make the RMM system directory your current directory. Then start
NM7000.EXE (DOS) or NM7000.BAT (Win95/NT).
If you are not familiar wit h DOS commands, the abov e step s can be perf ormed with t he follo wing commands:
CD\NM7000↵
NM7000↵
5.2User interface components
There are three visible components t o the RMM software: the menu bar at t he top, the window
area in the middle and the status line at the bottom of the screen. The next few sect ions
describe these components, and the information that they display.
5.2.1The menu bar and menus
The menu bar is your primary access to all the menu commands. The menu bar is always visible. If a menu command is followed by an ellipsis (. ..) choosing the command displays a dialog box. If the command does not have an ellipsis, the action occurs when you choose the
command.
Here is how to choose menu commands using the keyboard:
1. Press F10. This makes the menu bar active; the next thing you type wil l relate to the items
in the menu bar.
2. Use the arrow keys to select the menu you want to display. Then press Enter.
As a shortcut for this step, you can just press the highlighted letter of the menu title. For
example, when the menu bar is active, press F to move to and display the File menu. At
any time, press Alt and the highlighted letter (such as Alt+F) to display the menu you want.
3. Use the arrow keys again to select a command from the menu you have opened. Then
press Enter.
At this point, the ILS Data Sampler either carries out the command, displays a dialog box,
or displays another menu.
There are two ways to choose commands with a mouse:
•Click the desired menu title to display the menu and click the desired command.
•Or, drag straight from the menu title down to the menu command. Release the mouse button on the command you want. (If you change your mind, just drag off the menu; no command will be chosen.)
Note that some menu commands are unavailable when it would make no sense to choose
them. However, you can always get on-line Help about currently unavailable commands.
5.3Shortcuts
The software offers a number of quick ways to choose menu commands. The click-drag
method for mouse users is an example. From the keyboard, you can use a number of keyboard shortcuts (or hot keys) to access the menu bar, choose commands, or work within dialog boxes. You need to hold down Alt while pressing the highlighted letter when moving from
an input box to a group of buttons or boxes. Here is a list of the shortcuts available:
Do thisTo accomplish this...
Press Alt plus th e highlighted letter of the command (just press
the highlighted letter in a dialog box).
Display the menu or carry
out the command.
Type the keystrokes next to a menu command.Carry out the command.
Table 5-1 Shortcut commands.
For example, to display a prev iously saved log , press Alt+F D (for File |Di splay log) or you can
just press F3, the shortcut displayed next to it.
5.3.1The desktop area
Most of what you see and do in the ILS Data Sampler environment happens in a window. A
window is a screen area that you can open, move, resize, zoom, til e, and overlap. Every window is positioned on the desktop, the area between the menu bar and the status line.
You can have many windows open at the same time, but only one window can be active at
any time. Any command you choose or text you type generally applies onl y to the active window.
You can spot the active window easily: It is the one wi th th e double-lined border ar ound it. T he
active window usually has a close box, a zoom box, and scroll bars. If your windows are
overlapping, the active window is always the one on top of all the others.
There are several typ es o f windows , and they may h ave all o r some of t he foll owing cha ract er-
•A window that controls a modem also has a flow display icon. This icon is normal ly chang-
ing, indicating that new data is received from the ILS station.
The close box of a window is the box in the upper left corner. Click this box to quickly close the
window. (Or choose Window | Close, with shortcut Alt+F3). The Help window is considered
temporary; you can also close it by pressing Esc.
The title bar, the topmost horizontal bar of a window, contains the name of the window. Dou-
ble-clicking the title bar zoom the window. You can also drag the title bar to move the window
around.
The zoom box of a window appears in the upper right corner. If the icon is an up arrow, you
may click the arrow to enlarge the window to the largest size possi ble. If the icon is a double
headed arrow, the window is already at its maximum size. I n that case, clicking it returns the
window to its previous size. To zoom a window from the keyboard, choose Window | Zoom,
with shortcut F5).
Shortcut: Double-click the title bar of a window to zoom or restore it.
5.3.1.1 Scrolling the contents of a window
Scroll bars are horizontal or vertical bars with an arrow in each end, and a box between these.
You use the scroll bars with a mouse to scroll the contents of an ILS parameter display window. Cl ick the arrow at either end to scr oll one line at a ti me. (Keep the mouse butt on pressed
to scroll continuously.) Y ou can cli ck the shaded area to eit her side of the scroll box to sc roll a
page at a time. Finally, you can drag the scroll box to any spot on the bar to quickly move to a
place in the window relative to the position of the scr oll box.
5.3.1.2 Resizing a window
You can resize a window by dragging the resize corner (the bottom, right corner) of the window. The resize cor ner is di splayed as singl e outline part of the window frame. To resize using
the keyboard, choose Size/Move from the Window menu. Pressing the arrow keys will then
move the window, and Shift+arrow keys resizes the window. Note that resizing only applies to
windows which has a resize corner.
5.3.1.3 Quick window management
The next table gives yo u a quick run-down of how to handle windows in the ILS Dat a Sampler.
Note that you don't need a mouse to perform these actions - a keyboard works just fi ne.
Every window that display s ILS parameter s has a small cont rol window. This control window is
placed at the upper left corner of the desktop when you choose to display a log fi le, or connect
to an ILS station.
Open and display a saved log fileChoose File | Display log.
Close a windowSelect the window, and press Alt+F3.
Hang up phone and close data window Select the corresponding cont rol window, and press
Alt+F3.
Move the active windowDrag its title bar. Or choose Window | Size/Move
and use the arrow keys to place the window where
you want it, then press Enter.
Resize the active windowDr ag the bottom right corner. Or choose Window |
Size/Move and press Shift while you use the arrow
keys to resize the window, then press Enter.
Zoom the active windowClick the zoom box in the upper right corner of the
window, or
Double-click the window's tit le bar, or
Choose Window | Zoom
Table 5-2 Quick window operations.
5.3.2The status line
The status line appears at the bottom of the screen. It
•reminds you of basic keystrokes and shortcuts (or hot keys) applicable at that mome nt in
the active window.
•lets you click the shortcuts to carry out the action instead of choosing the command from
the menu or pressing the shortcut keystroke.
•offers one-line hints on any selected menu command and dialog box items.
•tells you how much memory is currently available
The status line changes as you switch windows or activities. One of the most common status
lines is the one you see when you are browsing a log file, or logging data. When you've
selected a menu title or command, the status line changes to display a one-line summary of
the function of the selected item.
5.3.3Dialog boxes
A menu command with an ellipsis (...) after it leads to a dialog box. Dialog boxes off e r a con-
venient way to view and set multiple options. When you are making settings in dialog boxes,
you work with five basic types of on screen controls : Action buttons, radio buttons, check
boxes, input boxes, and list boxes. Here is a sample dialog box that illustrates some of these
items:
This dialog box has two standard buttons: OK and Cancel. If you choose OK, the choices in
the dialog box are accepted; if you choose Cancel, nothing changes and no action takes
place, and the dialog box is put away. You can press F1 to open a Hel p window for this dial og
box. Esc is always a keyboard shortcut for Cancel (even if no Cancel button appears).
If you are using a mouse, click the dialog box button you want. When you are using the keyboard, press Alt and the highli ghted letter of an item t o activate it . For example, ALT+O selects
OK because the O in OK is highlighted. Press Tab or Shift+Tab to move forward or back from
one item to another in a dialog box. Each element is highlighted when it becomes active.
In this dialog box, OK is the default button , which means you need only pres s Enter to choose
that button. (On monochrome systems, arrows indicat e the default; on color monitors, default
buttons are highlighted.) Be aware that tabbing to a button makes that button the default. You
can then press Space to choose that button.
5.3.3.2 Radio buttons and check boxes
Radio buttons are like car radio buttons. They come in groups, and only one radio button in
the group can be on at any one time. To choose a radio button, cli ck it or it s te xt. Fro m the keyboard, select Alt and press the highlighted letter (if any), or press Tab until the group is highlighted and then use the arrow keys to choose a particular radio button. Press Tab or
Shift+Tab again to leave the group with the new radio button chosen.
Check boxes differ from radio buttons in that you can have any n umbe r of check boxes
checked at any time. When you select a check box, an x appears in it to show you it's on. An
empty box indicates it's off. To change the status of a check box, click it or its text, press Tab
until the check box is highlighted and then press Spacebar, or select Alt and the highlighted
letter (if any).
If several check boxes apply to a topic, they appear as a group. In t hat case, tabbing moves to
the group. Once the group is selected, use the arrow keys to select the item you want, and
then press Spacebar to check or uncheck it. On monochrome monitors, the active check box
or group of check boxes will have a chevron symbol (>>) to the left and right. When you press
Tab, the chevrons move to the next group of check boxes or radio buttons.
Toggles let you select one setting out of a group defined settings. They are similar to Radio
buttons (chapter 5.3.3.2); except that only the selected setting is displayed. You select other
settings by clicking on the text showing the toggle setting, or by pressing the space bar.
5.3.3.4 Input and list boxes
Input boxes let you ty pe in t ext . Most bas ic te xt-edi ting k eys work in the text b ox ( for exa mple,
arrow keys, Home, End, and Ins). If you continue to type once you reach the end of the box,
the contents automatically scr oll. If ther e is more text than what shows in the box, arrowheads
appear at the end. You can click the arrowheads to scroll or drag the text.
If an input box has a down-arrow icon to its right, there is an associated history list. Click the
down arrow icon, or press the down arrow key to display the list. You will find text you typed
the last few times you used the input box. Press Enter to choos e an item from this li st. You can
also edit an item in the history list. Press Esc to exit from the history list without making a
selection.
A final component of many dialog boxes is a list box, which lets you scr oll through and select
from variable-length lists (often file names) without leaving a dialog box. If a blinking cursor
appears in the list box and you know what you are looking for, you can type the word (or the
first few letters of the word) and the ILS Dat a Sampler will search for it.
You make a list box active by clicking it or by choosing the highlighted letter of the list title (or
press Tab until it's highlighted). Once a list box is displayed, you can use the scroll box to
move through the list or press the arrow keys on the keyboard.
5.4Configuring the RMM system
This chapter describes how to configure the RMM system. The configurat ion settings cov ered
in this chapter does not affect the operation of remote IL S systems in any manner. It applies to
the “PC part” of the system only.
The configuration dialogs is accessed via the Options menu. The following sub-chapters covers each of the menu items on this menu.
5.4.1Communications setup
The software allows you to connect the modems to two of four communications port s; COM1 COM4, with configurable IRQ's for COM3 and COM4. The ILS st a tion may al so be connect ed
directly to the PC by the use of a null modem cable.
The communication with the ILS is done with the use of a Channel construct. In the stati on
setup for an ILS station one selects a Channel in which the communication with the remote
ILS shall flow . Each channel is then associated with a specific set of configurable communication settings.
Choosing Communications setup brings up the following dialog
Selecting one of the Channel setup buttons opens up a dialog like Figure 5-3 shows.
Figure 5-3 Channel setup dialog
Note: DOS version also has the “Modem hangup String” and “Use DTR to hangup” fields in
this window. In the Win95/NT version DTR will always be used to hangup the modem.
The items in this dialog has the following purposes.
ItemPurpose
PortSpecifies the serial communications adapter to which the
modem or ILS is connected.
Bit rateSpecifies the bit rate to use when communicating with the
modem or ILS.
Use DTR to hang-upSelect this if the software should use the DTR line instead of t he
"Modem hang-up string" to make the modem hang up. This is
faster and safer than using the Modem hang-up string.
Direct connectSelect this if the PC is connected to the ILS via a null modem
Leased lineSelect this if the PC is connected to the ILS via dedicated
leased line modems.
Normal dial-upSelect this if the PC shall connect the ILS via modem on a pub-
lic telephone network.
Modem Init Stri ngThis string will be sent to the modem before each dialing proc-
ess starts. The default is ATZ, which performs a software reset
of the modem.
Hang-up StringCommand to make the modem hang up the phone.
Table 5-3 Channel setup operations.
Modem configuration
5.4.1.1 Close Connecti on
Selecting the "Use DTR to hang-up" is recommended, as this method for closing the connection is both safer and faster than the "+++" escape sequence (Hang-up string). Note that the
modem must be configured with the AT&D2 command for the "Use DTR to hang-up" method
to work (some modems do not include this command in their default initiation).
5.4.1.2 Recommended init-strings
Refer to the Technical notes at the end of this chapter for information on how to set up the
modems on different types of connection.
5.4.2IRQ setup
Selecting the IRQ setup button in the Com port setup dialog opens the following dialog.
Figure 5-4 IRQ setup dialog
This dialog allows alternate IRQ selections for COM3 and COM4.
NOTE
The Win95/NT version does not have this dialog, as these settings are handled at the
The Printer setup dialog contains options for an optionally connected printer. The dialog is
shown below.
Figure 5-5 Print setup dialog
The purpose of each field in the dialog is described below.
ItemPurpose
PortDetermines the printer port to use for the printing.
If “File” is selected the output will be sen t to the file specified in
the string input field. The file will be created in the “Log directory” for the current stat ion (see 5.5).
Print framesAllows enabling/disabling the printing of special frame charac-
ters. Uncheck this checkbox i f your pri nter does n ot support the
PC-8 bit character set.
Send form-feedAllow for sending a form-feed (ASCII 12) character to the pr inter
following each print-out. The extra form-feed is required by
most laser printers.
Table 5-4 Print setup operations.
To print the report, select File|Print report.
5.4.4Preferences
The Preferences dialog allows the user to set general program preferences
The DDM setting applies to all user input and all screen or printer output of DDM measurements. Selecting % means that input/output is in percent, likewise, select ing uA results in
micro ampere input/output.
The date format field selects day-month-year or month-day-year display of ti me.
5.4.5Toggle video mode
This menu selection toggles the current video mode between 25 lines and 43/50 lines.
5.4.6Set date/time
To set the date and time of the PC, select Set date/time from the Options menu. A dialog as
show next is shown.
Figure 5-7 Set date/time dialog
In the Date field, enter the current date in day.month.year format if DD.MM.YY is selected in
the Preferences dialog ( Figure 5-6) . Els e enter t he date in MM.DD.YY for mat. In t he time fi eld,
enter the current time of day in standard 24 hour format.
5.5Setting up ILS stations
Before connecting to an ILS station the st ation must be added to the “station list”. Some information about each station must also be entered, e.g. the number to dial to connect to the station. This chapter describes how to set up ILS stat ions.
Selecting the ILS | Station list menu command brings up a dialog like the one shown below.
This dialog contains a list showing the name and number of the configured stations. Use the
scroll bar to the left of the list to scr oll or user the up/down arrow keys. If the selected Channel
for a station is a “Direct connect” channel, the number field will show DIRECT/CHx. Up to 50
stations may be configured.
The following table describes how use this dialog.
To accomplish thisDo this
Change the configuration for
a specific station.
Select the station from the list, and press the Edit button. You
can also press ENTER, when the desired stati on is hi ghlight ed.
Double-clicking the desired station also activates the Edit function.
Add a new station to the list Press the Add button, or edit an empty name/number line from
the list.
Delete a stationSelect the station you want to delete and press the Delete but-
ton. You may also press Del on your keyboard.
Connect to the station
Press the Connect button.
(see the Displaying measurements chapter)
Table 5-5 Station list operations.
When you have selected Add or Edit from the above dialog, another dialog appears. This dialog is shown in Figure 5-9.
The table below describes the meaning of each field in the ILS Station setup dialog.
Use fieldTo specify
Station nameA name for the ILS station. Holds up to 20 characters. This
name is saved in the log files you create.
NumberThe telephone number of the ILS station. This string is passed
directly to the modem, meaning that special chara cters like ',' or
other characters can be included if the modem understands
them.
Log dir. nameThe name of the directory where downloaded measurements
shall be stored. The directory is made if it do not exist.
Auto download MED
storage
Whether or not the medium time periodic storage should be
downloaded automatically each day.
Start timeThe time of day to start the automatic MED download.
I/O channelWhich I/O channel to use when connecting to this station.
(See the Communications setup chapter)
Table 5-6 ILS station setup operations.
5.6Connecting to an ILS station
To connect to an ILS station the station must first be added to the Station List, as descri bed in
chapter 5.5. When this has been done, just press the “Connect” button in the ILS St ation List
dialog. The software will try to make a connection to the station. Once a connection has been
established you will be prompted to enter your use rname and passwords. You enter this information in the dialog box shown in Figure 5-10.
Your assigned access level depend on the number of passwords and the username entered.
Different passwords must be separated with a space character. The following table outlines
the access level properties, and how to gain them. Increasing access levels adds more
access rights, and retai n ing the rights for the lower access levels.
NORMARC 7013
Access levelYou enterAccess rights
1your username and 1
password
2your username and 2
passwords
3your username and 3
passwords
MASTERusername MASTER
and 3 passwords
Table 5-7 Login access rights.
When you have entered correct login i nformation and pr essed the OK butto n, the pr ogram will
start to retrieve ILS configuration data. Windows that was open the last time you was connected to the ILS will automatically be opened when the ILS configu ration has been received.
If you enter invalid login information the ILS will not grant you access rights to the ILS. The
software will display a message telling you what went wrong, and let you perform another
login attempt. You have three attempts. If all attempts are used without success the ILS will
disconnect and not allow more login attempts for 3 minutes.
When the program has received the configura tion dat a, th e ILS configur ation will be display ed
in a window in the upper lef t corne r of the deskt op area. This wi ndow will be re ferr ed to as the
“Link” window. Figure 5-11 shows a typical Link window.
Read all measurements/status and show all settings,
and download.
Change front panel settings and change settings in
the “Ident and mod. switches” window.
To shut down an ILS connection, close the Link window for the connection you want to shut
down. For information on how to close a window, see Table 5-2.
Do not close the windows you want reopened the next time you connect to this ILS station.
5.6.2Baud rate recommendations
When connecting to the ILS at 9600 bps, Remote port 2 or Local port on the ILS should be
used. For performance reasons, remote port 1 should not be used at 4800 and 9600 bps. This
will affect the response times of the Local RMS front panel as well as message handling on
the other RS-232 ports.
5.7Monitor measurements and status
5.7.1General
The functions described in this and the following chap ters are accessed via the ILS|Measurement sub-menu.
The windows described here will automatically be update d with new measurements as long
as the windows remain open.
The monitors signal the state of each par ameter. For analog measurements, the stat es wi ll b e
displayed by a character code preceding the measurement and by the col or of the displayed
number.
StateCharacter codeColour code
Normal(None)(Black)
Raw warningwYellow
Delayed warningWYellow
Raw AlarmaOrange (light red)
Delayed AlarmARed
Table 5-8 Monitor codes.
For digital measurement only two states are defined; OK and ERROR. The ERROR string is
shown in red, while the OK string is shown in black.
To display the parameters shown in the window below select the ILS|Measurements|Monitor
Window 1 menu item. This window displays the standard monitor parameters (DDM, SDM
and RF-level) for each of the monitor channels (Course line, Displacement Sensitivity, Near
field and Clearance Transmitter (2-freq. systems)). In addition, Difference Frequency (2freq.systems), DC-loop and Ident in formation are di splayed. The res ult of the cont inuous monitor self test is also shown.
Figure 5-12 Monitor window
The DC loop status shown in this window will be OK if all the DC loop readings are within the
warning limits, else ERROR is displayed (DC loop r eadings for each loop is displayed in Monitor window II).
5.7.3Additional monitor parameters window
Selecting ILS|Measurements|Monitor 2 bri ngs up the fol lowing window. This window shows
secondary monitor parameters.
Due to the nature of the monitor self test signals, the values wi ll toggle between alarm and
normal. If one of these erroneous te st signals f ails to generate alar m, a self test erro r will be i n
Monitor Window 1. The CL/CLR MORSE parameters shows the modulation level of the ident
signal. No activity on this signal for the time peri od specified in the ident delay setting, will
result in an IDENT warning in Monitor Window 1.
Depending on the antenna system installed, up to 4 DC-loops can be active. The DC LOOP
parameter in Monitor Window 1 shows accumulated status of the 4 DC-loops.
Selecting ILS|Measurements|External brings up t he following window. This window shows the
measurements on the external channels. The NM 70xx can provide executive monitoring of
signals from external equipment, defined by the user (option).
For external channels that are not used, no measurement field will be shown (e.g. the standby
measurement field will be shown (e.g. the st andby measurements in Figure 5-14)
Figure 5-14 External parameter s window
5.7.5Maintenance monitor window
Selecting ILS|Measurements|Maintenance brings up the window shown below. The maintenance warning parameters are displayed in groups corresponding to the modules on which
they are measured. A separate group consists of the main power supplies and current measurements. In this way, the origin of a maintenance warning can easily be deduced.
If one maintenance parameter on a module is in warning status, the module to which it
belongs is given the same status. Therefore, it can be easily determined the maintenance
warning status of the system by looking at the top window (Figure 5-15) only.
The following table shows the defined status codes.
All parameters OKOK
Raw warning on at least one param.WARN (R)
Delayed warning on at least one param.WARN (D)
Internal I
2
C error on at least one param.
ERROR
No measurements applicableN/A
Table 5-9 Status codes.
Digital parameters are displayed as outlined in the above table. Analog parameters are displayed with the same character and color codes as the monitor measurements.
By selecting a board on the list and double-c licking it or pressin g Enter opens another window
which displays the status/measurements on that specific board.
Figure 5-15 Maintenance monitor window
The following figures show the available status/measurement windows.
Figure 5-22 User parameters window (sample configuration).
If a digital user paramet er is defined as a out put and you have at least acces s level 2, you can
change the output value by changing the state of the corresponding toggle in this window.
5.8Monitor settings
The functions described in this and the following sub-chapter are accessed via the ILS|Monitor settings sub-menu.
The current access level determin es which fun cti ons wil l be avai lable. If a functi on is not avai lable because of to low access level, the function will not be possible to activate.
All access levels allow the currently stored sett ing to be displayed; but in order to change a
setting an access level of at least 3 is required.
5.8.1Limits Monitor 1&2
When the ILS|Monitor settings|Limits Monitor 1&2 menu option is sel e cted a window showing
the currently configured monitor 1 and 2 alarm and warning limits is display ed. A window with
sample data is displayed below.
The alarm limit field will show “OFF” if the alarm limit is equal to the maximum or minimum
value.
The warning limit field for CL IDENT, CLR IDENT and external channels (i.e. FFM RF and
VOICE) will show “n/a” as the warning limits for these parameters is not def ined as a percentage of the alarm limits.
The “MON 1 Limits” and the “MON 2 Limits” buttons will only be displayed for access
level 3 and master user access level.
In order to change the alarm and/or warning limits for a specific parameter; first select the
parameter in the list box, then selec t the “MON 1 Limits” butt on or the “MON 2 Limits” button if
you want to change either monitor 1 or monitor 2 alarm/warning.
When one of the above mentioned buttons is selected a dialog similar to the one below is
opened. In this dialog one may enter new alarm limits and warni ng limit for the currently
selected parameter. The alarm limits is always relative to t he nominal value. The warn ing limit
is always entered in percent of the selected alarm limit.
If the Symmetric button is pressed the alarm limit is entered as a value which is symmetric
around the nominal value; hence only one alarm limit wi ll have t o be ente red in the sy mmetric
mode.
The arrowheads after each input field in this dialog can be clicked on to increase or decrease
the displayed limit value. Hold down Shift to increment/decrement by a larger value. The up
and down arrow keys may be used instead of the mouse.
Due to the digital represent a tion of t he alarm and warning limits, the entered value is rounded
to the nearest 12-bit digital value. The actual stored value, which is read back to the dialog
box, can therefore differ slightly from the entered value.
The text OFF may be entered in alarm limits field to inhibit low, high or any alarms on the current monitor parameter. If OFF is entered, one would normally enter 100 (%) in the warning
field, in order to also inhibit warning indications. For parameters where the warning limits is
not in % of the alarm limits , large negat ive and pos itive numbers must be entered to i nhibit t he
warning. See Appendix B for recommended values.
Figure 5-24 Monitor parameter input dialog
The NAME button is used to change the parameter name. This applies to the ext ernal channels only.
Refer to appendix B for the factory default values o f all monitor parameters.
5.8.2Limits Standby monitor
When the ILS|Monitor settings|Limits Stdby monitor menu option is selected a window showing the currently configured standby monitor alarm and warning limits is displayed. This window is used in the same way as the Monitor 1&2 limits window, described above.
5.8.3Delays
Selecting ILS|Monitor settings|Delays brings up the Monitor Delays window, shown below.
The NM70xx is configured with three separately adjustable delays: Normal, Near Fiel d, Ident
and Far Field Monitor. This window displays the currently configured Monitor 1&2 (equal
delays) and standby monitor delays. To change a delay; first select the desired parameter on
the list and then press the “Mon 1 and 2” button if the delay for Monitor 1 and 2 should be
altered, else press the “Stdby” button for a standby monitor delay change.
A standby column will be shown if the system is equipped with a standby monitor. The resolution of the delay parameters is 1/10 second.
Once the “Mon 1 and 2” or the “S t dby” bu tton has be pressed a n input dialog l ike the one show
below is opened. Enter the desired delay in the field, or use the up and down arrows to increment/decrement the current value. When the “Update” button is selected, the ILS will be
updated.
Figure 5-26 Delay input dialog
Valid input range in this window is 0.4 to 102.3 seconds. The tot al changeover-shutdown
delay will be two times this value. See Appendix B for recommended values.
5.8.4Limits Maintenance monit or
When the ILS|Monitor settings|Analog maintenance limits or Digital maintenance limits menu
item is selected a window showing the currently configu red anal og/digital maintenance monitor warning limits is displa yed. Windows with sampl e data is di splayed below. The board name
is shown idented to the left. The maintenance parameters is show after each board name.
For analog parameters the “Low” and “High” columns sh ows the low and high warning limits.
For digital parameters a single column show the “normal state” for the parameter. If the
parameter state is not equal this state a maintenance warning will be generated by the ILS
RMM software. If no warning is wanted for a parameter the columns will display “O FF”.
Figure 5-27 Analog maintenance monitor limits window
This list is scrollable, all defined parameters are listed in the Maintenance monitor chapter.
Figure 5-28 Digital maintenance monitor limits window
This list is scrollable, all defined parameters are listed in the Maintenance monitor chapter.
To change a warning limit; first selec t the desired p arameter on the list and p ress the “Change”
button. Depending on the type (digit al/analog) of the dialog, one of t he dialogs shown next wi ll
Figure 5-29 Input dialog for digital maintenance parameters
To change the setting for a digital parameter cl ick in the normal st ate field or use the space ba r
to toggle. The normal state field will toggle between Low, High and OFF. The settings for the
CHARGER 1 STATUS and CHARGER 2 STATUS parameters is On and Off.
Press the Update button when the desired state is shown.
Figure 5-30 Input dialog for analog maintenance parameters
To change the setting for a analog parameter, type in the wanted low and high warning limits
in the corresponding fields. Then press the Update butt on.
5.8.5Delays Maintenance monitor
To display or change the maintenance monitor delay parameters, select the ILS|Monitor settings|Delays maintenance monitor menu item. The window shown below will be displ a yed.
The operation of the window is similar to the operation of the Monitor delays window .
Figure 5-31 Maintenance delay window
The resolution of the maintenance delay parameters is 1 second. Valid input range is 1 to
1023 seconds.
5.8.6Analog user parameters
Selecting ILS|Monitor sett ings|Analog user para meters opens the window i n Figure 5- 32. This
window shows the current settings for the analog user parameters.
Figure 5-32 Analog user parameters definiti ons
The following definitions c an be made for t he analog user p arameters: Name, lower and upper
warning limits, and scale and offset constants for conversion to engineering units. There is
also a field for enabling/disabling gener ation of a maintenance warning.
To change the settings for a parameter, select the parameter on the lis t and press the
“Change” button. This brings up the dialog shown below.
Figure 5-33 Analog user parameter settings dialog
5.8.7Digital user parameters
Selecting ILS|Monitor settings|Digital user parameters opens the window shown below. This
window shows the current settings for the digital user parameters. There are six p arameters
which operate as digital input or output , and two parameters which count number of positive
transitions on the corresponding digital inputs. The counter will count from 0 to 99, and be
automatically reset to 0.
The following definitions can be made for the digital user parameters: Name, input or output
selection, whether the value shall be inverted or not, and warning state of the parameter
(TRUE / FALSE / NONE).
To change the setting for a I/O parameter, select the parameter on the list and press the
“Change” button. This brings up the dialog shown below.
Figure 5-35 Digital user parameters settings dialog
For the counter parameters; the “Change” button opens a dialog where the label text can be
modified.
5.9TX settings
5.9.1TX Adjustments
Selecting the ILS|TX settings|TX adjustments menu item opens the window shown below.
Figure 5-36 TX adjustments window
The following TX parameter s can be changed for Course and Clear ance (2-freq. systems) for
transmitter 1 and 2:
•90Hz & 150Hz modulation balance (DDM),
•90Hz & 150Hz modulation sum (SDM),
•RF power
•Ident modulation depth
•In addition, the runway Morse code identification (2-4 letters) is entered here.
To change TX adjustment values click the TX1 or TX2 buttons. A dialog similar to the one
below will be opened. To change the parameter value you have the following options:
•Click on the up and down arrows to increment/decrement the value. The ILS will automati-
cally be updated when you stop changing the value.
•Press the up and down arrows on your keyboard to increment/ decrement the value. Holding Shift down will increment/decrement by a larger value. The ILS will automatically be
updated when you stop changing the value.
•Enter a new value in the numeri c input f ield. The ILS will be updated when you p ress Enter
or click the Update button.
Figure 5-37 TX Adjustment input dialog
The TX-adjustment values are the positions of digital potentiometers used and stored in the
EEPROM on the LF generator boards. The values are unit-less integers in the range 0-4095.
See Appendix B for typical values used.
5.9.2Ident. and modulation switches
Selecting ILS|TX settings|Ident. and modulation swit ches brings up the window shown below.
Figure 5-38 Ident. and modulation switches window
In this window the LF generator 90 Hz and 150 Hz signals may be switched on or off and the
Morse code and VOICE settings may be changed (LLZ only). VOICE on/ off is only applicable
for VOICE configured Localizers. The following t able descr ibes the ef fec t of each Morse code
setting.
SettingTransmitted ident.
OFFNone (silent)
NORMALThe one selected in TX adjustments - Morse code
CONT.Continuous 1020 Hz tone.
TSTThe Morse code for TST
Table 5-10 Morse code settings.
5.9.3CLR mod. and DME settings
Selecting ILS|TX settings|CLR mod. and DME settings brings up the window shown below.
The fields in this window specifies the operati on of the DME interface (LLZ only) and quadrature modulation of the CLR LF signal (2-freq. systems only).
The following table describes the each selection.
Keying masterLLZLLZ is the keying master
DMEDME is the keying master
DME active signalOPENOpen circuit indicates active DME (enables colocation)
CLOSEDClosed circuit indicates active DME (enables colocation)
Quadrature mod.OFFDo not quadrature modulate the clearance LF signal.
ONEnable quadrature modulation of the clearance LF signal.
Table 5-11 CLR modulation and DME settings.
5.10Power scaling
NOTE
The scaling values in this window should only be changed on racks with the first ver-
sion of the PA module (LPA1230A/GPA1231A/GPA1232A version 1). On newer PA mod-
ules, the power reading is pre-adjusted, and the scaling value shoul d be set to 1.0.
Selecting ILS|TX Settings|Power scaling when connected to an ILS with RMS software version 10, or later, will bring up the window below.
Figure 5-40 Power scaling settings
The scaling values will normally be in the range 0.5-2.0.
The values in this window can be used to adjust the power measurements done on the trans-
mitters. The measurement values are displayed in the maintenance windows for the transmit-
ters.
These scaling values are stor ed in the RMS s ystem in t he ILS. Thi s implies that the values will
have to be changed if a PA module is replaced.
5.11Front panel operation
When ILS|Front panel is selected a window displaying the same information as the ILS front
panel is opened. See Figure 5-41.
Figure 5-41 Front panel window
Pressing buttons in this window has the sa me effect as pressing the corresponding switch on
the ILS front panel.
Note that some switches on the ILS front panel is not possible to change from the RMM system (e.g. write protect). The settings of these switches is displayed as “LED’s” in this window .
5.12Get historical data
5.12.1General
Selecting ILS|Get historical dat a opens the Storage download window, shown in Figure 5-42.
This window is used to download one of the data storages.
The different storages contain the following information. All stored items are given a corresponding time stamp:
•Alarm storage: Samples of all monitor and maintenance parameters for a period starting
from 30s before an alarm occurred and ending with the actual alarm sample. Up to 7 alarm
storages can be stored in the ILS. The RMM will download all storages into one alarm storage file.
•Warning storage: One sample of al l monitor and maintenance parameters at the ins tant
when a warning LED is activated. The 25 last warnings are stored.
•Medium time periodic storage: One sample of all monitor and maintenance parameters
stored periodically at user specifi ed intervals (see 5.22). The storage contai ns the last 96
samples.
•Long time periodic storage: For every day through a 180 days period, the mean value and
standard deviation of all monitor and analog maintenance parameters.
Figure 5-42 Storage download window
E.g. to download the alarm storage buffer; just select the Alarm storage radio button and
press the “Download“ button. The software will then retrieve the alarm storage and store the
data on the local hard disk.
The software automatically creates a name for the fi le in which the data is st ored. The name is
of the form YYMMDD.ext. Where YY is the last two digits of the cur rent year, MM is the month
and DD is the day of month number. The fi lename extension field is defined in the following
table.
Storage typeExtension
Medium time periodic.MED
Long time.LON
Warning.WRN
Alarm.ALA
Table 5-12 Filename extensions.
The file will be created in the directory specified in the Log dir. field of the current station
setup.
5.12.2Alarm storage specific
The software will check the .ALA file for the current day (if it exists) to see wheth er or not a
new alarm has occurred. If no new alarm is available in the ILS storage buffer no data will be
downloaded, and you will be notified about the situation.
5.13Continuous data logging
When ILS|Continuous logging is selecte d the dialog in Figure 5- 43 is opened. From this di alog
the user controls the writing of received data to a file. The data written to a file can be displayed on screen and exported to text format f or later use in a spreadsheet program. The user
may select which types of data to be stored on disk, as displayed in Figure 5-43.
The time interval between each time a sample is stored can also be configured in t his dialog.
The intervals must be entered in seconds and tenths of a second. The storage interval information entered in this dialog will be stored and associated with the current station.
The software automatically creates a name for the fi le in which the data is st ored. The name is
of the form YYMMDD.ext. Where YY is last two digit s of the current y ear , MM is the mont h and
DD is the day of month number. The filename extension field is CNT.
The file will be created in the directory specified in the Log dir. field of the current station
setup.
The "Stop cont logging" button in this window can be used to stop logging continuous data
(once it has been enabled). When continuous logging has been turned on the "Continuous
log" field in the Link window displays "ON".
5.14Events
Selecting ILS|Events brings up the window shown in Figure 5-44. This window displays the
last 50 events (user operations , log -in/out, alarms, war nings) i n a scroll able li st. The di splayed
fields are:
•The time when the event occurred
•The user caused the event (if any)
•The event type description
Event diagnostics is also invoked from this window, see chapter 5.24 for a description of this
feature.
The ILS|Test DDM function is for use when a DDM flight test is performed on the system. The
Test DDM window is shown in Figure 5-45. The window contains the following fields:
•The current CL and NF DDM nominal, measurement, and low and high alarm limit.
•Buttons for quick access to the monito r alarm/warning configuration dialog (see chapter
5.8.1).
•DDM adjustment “potentiometers” for adjusting dominance levels.
The Inc and Dec buttons increase/decrease the dominance level.
NOTE: Pressing Inc on a 150 Hz level will decreas e the potentiometer value, while Dec will
increase the value.
•Toggles for setting the type of dominance (toggles between Off, 90 Hz and 150 Hz)
The required access levels for the fields in this dialog is as follows:
•To change monitor limits (use of the “monitor CL/NF DDM” buttons) requires access
level 3.
The test DDM values are unit-less integers in the range 0-4095.
5.16Set ILS date/time
The functionality of the ILS|Set ILS date/time window depends on whether or not the user is
logged in as an ordinary user or as master user. When logged in as a master user this window
allows setting of the ILS RMS system clock. When logged in as an ordinary user, this window
displays the current ILS RMS system date/time.
Figure 5-46 ILS date/time window.
The available master user operations is outlined in the table below.
FieldDescription
New DateHolds the date to be sent to the ILS upon an Update but-
ton activation.
New Time Holds the time to be sent to the ILS upon an Update but-
ton activation.
Update (button)Sets the ILS RMS clock to the date/time specified in the
Sync with PC clock (button)Sets the ILS RMS clock equal to the local PC system
clock.
Close (button)Closes the ILS date/time window.
Table 5-13 Master user available operations.
5.17User administration
The functionality of the ILS|User list/adm. window depends on whether or not the user is
logged in as an ordinary user or as master user. When logged in as a master user this window
allows configuring of existing and new users which shall be granted access to the RMM system. When logged in as an ordinary user, this window only displays the last time each user
was logged on.
The following is a description of the available operations when logged in as a master user.
Figure 5-47 User administration window
The available master user operations is outlined in the table below.
Operation
(button)
Description
EDITEdit the username and/or passwords for the currently selected
user
ADDAdd a new user
DELETEDelete the currently selected user
MASTEREdit the passwords for the master user
DISPLAYEdit the password for the ILS front panel display pa nel.
Table 5-14 Master user operations.
When Edit, Add or Master is selected in the User administration window a dialog showing the
currently defined username and p asswords i s opened. In t his dia log y ou may enter or c hange
the username and password(s).
Note that passwords must be 6-8 characters in length. Both the username and passwords
should only consist of characters in the range A-Z, a-z or 0-9.
The username for the master user cannot be changed.
With access level 3 you have access to change ILS modem bit rate and initiat ion strings and
make the ILS initiate the modems. Selecting ILS|Modem configuration brings up the window
shown below.
Figure 5-49 ILS Modem settings window.
The window displays the currently configured bit rates and init strings. Pressing the Change
button brings up a dialog where you can change the bit rate and/or the init string for the
selected port. Note that i t is no i nit stri ng f or t he local por t, s ince a modem is not su pported f or
this port.
Pressing the Send init button st arts modem initiation at the sel e cted port. When the ILS is
done initiating a dialog showing the result of the modem initiation will appear.
If the init-string is l eft blank or t he word «default» i s written, th e ILS will use a defaul t init-string
that shall work on standard Hayes compat ible modems under normal conditions.
Please consult the modem manual for further in formation about modem commands and initialisation.
5.19Upload configuration to ILS
Several boards in the NORMARC 7000 cabinet store settings or operationa l parameters in
internal EEPROM. Whenever a new board is inserted in a cabinet, these parameters need to
configured. This can be done by “uploading” the parameter values to the cabinet with the use
of the “ILS | Configuration options | Upload configuration to ILS” menu item. When this menu
item is chosen, a window for selecting a previously stored “CFG” file (see section 5.20) is displayed.
Once the user has selected the file containing the de sired configuration parameters, the window shown in Figure 5-50 is opened. This window lists boards for which the configuration may
be uploaded. Configuration for one board or all boards (normally used only for new cabinets)
can be uploaded.
This is an access level 3 operation.
Figure 5-50 Configuration upload
When the OK button is clicked, the RMM software starts the upload and displays a window
telling the user that the operati on is in progress. The upload may take a few minutes, depending on the board type selected for upl oad. In order to sp eed up the upload proc ess, it is a good
idea to close all windows that display monitor or maintenance measurements (this frees up
communication bandwidth).
Note 1: Refer to chapter 7 for detailed info rmation on uploading configuration after e.g. a
board replacement.
Note 2: The following configuration data for the RMS board are not stored in the c onfigur ation
file and therefore require manual updating: “Lamp mappings ”, “ILS modem configu ration”, and
“User list/ adm.”.
5.20Write configuration to file
As described in section 5.19, different boards store configuratio n data in internal EEPROM.
The configuration data can be stored on disk with the use of the “ILS | Configuration options |
Write configuration to file” menu item. Sel ecting this menu it em brings up the window shown in
Figure 5-51 The user must enter a new filename, or select an existin g file for storing the configuration parameter s. If an existing file is selected, the user is asked whether or not the file
should be overwritten.
The configuration file is a plain ASCII text file. The file may be inspected by the user, but to
avoid introducing possible errors, it is not recommended to manually edit it
5.21Lamp mappings
At access level 3 the user may configure which of the small warning LED’s that are “mapped”
to the main (big) warning LED on the local an remote control panels.
Figure 5-52 shows the window that appears when the “ILS | Configuration options | Lamp
mappings” menu item is selected. An “X”-mar k to the lef t of the war ning type indi cates th at an
active warning of that type will cause the corresponding main warning lamp to be activated.
If the configuration is changed, remember to click the “Update” button to send the new settings to the cabinet.
Figure 5-52 Lamp mappings HBK1142
5.22Medium storage sample interval
The sample intervals for the medium time storage may be selected by choosing ILS| Configuration options| Medium storage sample interval. This brings up the window shown below.
Figure 5-53 Medium storage interval setting dialog
The following intervals way be selected
Storage intervalTotal medium storage time
15 Minutes24 Hours
30 Minutes48 Hours
60 Minutes4 Days
120 Minutes8 Days
Table 5-15 Medium time storage interval settings
NORMARC 7013
INSTRUMENT LANDING SYSTEM
5.23Disp laying stored measurements
5.23.1General
This chapter describes how t o open and di splay a st orage buf fer after i t has been downl oaded
and stored on the local hard disk, as described in chapter 5.12.
The desired file is opened by selecting File|Open log. A dialog l ike the one shown below will
be displayed. This dialog shows the files matching the wild card mask in the Name field. The
directory may be changed by double clicking (or by pressing Enter) a list entry with a back
slash (“\”) appended to the name.
Figure 5-54 Open log file dialog
When a selection has been made a new window will be opened. This window displays information about the data stored in the opened file. The different types of storage files have different information displayed in this window, as described in the following chapters.
When a this window is opened, you would display the stored status/measurements by open-
ing the appropriate windows by selecting menu items from the ILS menu. Note t hat some
menu items will not be accessible because some information types is not stored in the log file.
To move back and forth in the stored data, press F7 and F8.
5.23.2Medium time periodic (MED)
When a MED file has been opened a window as shown below will be opened.
This window displays the time stamps of the first, the currently displayed one, and the last
sample stored in the file.
Figure 5-55 MED file log window
5.23.3Alarm storage (ALA)
When an ALA file has been opened a window as shown below will be opened. This window
shows the time stamps of the currently displayed Monitor 1, Monitor 2, Standby (if installed)
and Maintenance monitor samples. This window also has a list where you select an alarm to
study.
The operation of the window is outlined in the table below.
To accomplish thisDo this
Study a specific alarmSelect the desired alarm on the alarm selection list and press
When a CNT file has been opened a window as shown below will be opened. This window
shows the same information as the Alarm storage window, except that this window does not
have the alarm specific functions.
Figure 5-57 CNT file log window.
5.24DIAGNOSTICS
5.24.1General description of the diagnostic functions
The NM7000 RMM system has two algorithms to isolate a failure in the ILS. The algorithms
are the Current data diagnostics and the Alarm Event diagnostics. See chapter 7 for a thor-
ough description of these alghorithms.
5.24.2Current data diagnostics
This diagnostic function are invoked by selecting ILS|Current data diagnost ics. Based upon
the current status of the equipment, the program will suggest the faulty module, if any.
5.24.3Alarm Event diagnostics
This diagnostic function are invoked from the Event l ist window. A CHANGEOVER or SHUTDOWN event must be selected in the event list, then may the diagnostic function be invoked
by pressing the Event diagnostics button below the event list. Based upon the nature of the
CHANGEOVER or SHUTDOWN event, combined with the contents of the corresponding
alarm storage, the program will suggest the faulty module which originated the alarm. Note
that the event storage diagnostics function will only be able to isolate faults to module level
when a single fault has occurred
5.25Exporting data
You can export stored measurements to a spreadsheet with the function described in thi s section. Exported data will be stored in an ASCII tex t file. This text file may then be imported into
a spreadsheet program or a graphical present ation program.
Data from all types of log files may be exported.
The Export data function i s located on the fil e menu. In order to use the function, y ou must first
open the log file from which you want to export data. This is done by using the File|Open log
function (described in chapter 5.22) . When a log file has been opened you may select the
Export data function. The dialog shown in Figur e 5-58 wil l appear. Select the type of data to
export and press OK. You will be prompted to enter a file name for the exported data. The
Decimal point field selects the type decimal point character used when stored data are
exported to a text file. Select the p roper setting for use by your spreadsh eet progra m or graphical presentation pro gram.
Each field in the export file are separated by a TAB character. Each line contains all measurements in a sample. The line begins with the time stamp of sample (resolution 1 second). The
next field is 1/10 seconds of t he time st amp. The r est of the l ine is the measur ement s/ p arameters of the sample.
The type of decimal point used for analog measurements is defined by Decimal point setting
in the Preferences dialog (see chapter 5.4.4).
Digital parameters (OK/ ERROR) are export ed as numbers, 0 and 1. 1 equals OK, and 0 indicates ERROR.
5.26Printing reports
When you are logged in on an ILS you can make printouts of several measuremen ts and configurations.
The Print Report function is located on the file menu. To use it you must first connect and log
in on a ILS rack (access level 1). Figure 5-59 shows the Print Report dialog which is opened
when you select the File|Print report menu item.
Figure 5-59 Print Report dialog
The system will print the data you select in the ‘P rint’ section of the dialog. The Print setup
button is a shortcut to the Options|Print setup di alog, described in chapter 5.4.3.
5.27Print screen
To print the screen contents in the Windows 95/NT version, select the File|Print screen menu
option. This will send the cur rent sc reen cont ent s t o t he pri nter por t select ed in t he Pri nt setu p
window.
To print the screen under DOS use the "Print Screen" key on the PC keyboard. This will print
the screen on LPT1 (independent of the port selected in the Print setup window).
5.28Technical note - Leased line SETUP (USRobotics)
USING NM7000 RMM SOFTWARE WITH LEASED LINE CONNECTION
Applies to USRobotics Courier Dual Standard V.34+ modems
5.28.1Definitions
Local modemThe modem attached to the PC running
NM7000 RMM software.
Remote modemThe modem attached to the NM7000
cabinet.
5.28.2Configuring the remote modem
1. Set DIP switches on the remote modem.
•Switch 1 must be set to the DTR ALWAYS ON position.
•Switch 4 must be set to the ON position - Do not echo offline commands
•Switch 5 must be set to the AUTO ANSWER ON RING position.
•The remaining switches should be left in their factory default positions.
2. Connecting the modem to the cabinet.
•The remote modem must be connected to Remote Port 2 on the connection interface
board in the ILS cabinet.
3. Configure remote modem init string.
•With a PC running NM7000 RMM software, log in with MASTER access on the Local Port.
•Select «Modem configuration» from the ILS menu.
•Set the Remote port 2 modem configuration as follows (all ‘0’ letters are zeroes):
Figure 5-60 ILS port settings for leased-l ine connections
4. Initializing the remote modem.
•Pressing the «Send init» butt on in the ILS modem se ttings windo w makes the ILS send the
initialization string on the port selected in the port list.
•The ILS will also send the these init strings when the I LS is powered on.
5.28.3Configuring the local modem
1. All modem DIP switches should be left in their factory default positions.
2. Configure the «Channel Setup» for the leased line connection as follows (all ‘0’ letters are
zeroes):
Figure 5-61 RMM channel setup for USR leased-line connections
The modem will be automatically set up for leased when a connection with this channel is
selected by the user.
Users of the DOS version should also select «Use DTR to hangup». The hangup string is
ignored.
5.29Technical note - Dial up connection SETUP (USRobotics)
USING NM7000 RMM SOFTWARE WITH DIAL UP CONNECTION
Applies to USRobotics Courier Dual Standard V.34+ modems
5.29.1Definitions
Local modemThe modem attached to the PC running
NM7000 RMM software.
Remote modemThe modem attached to the NM7000
cabinet.
5.29.2Configuring the remote modem
1. Set DIP switches on the remote modem.
•Switch 1 must be set to the DTR ALWAYS ON position.
•Switch 4 must be set to the ON position - Do not echo offline commands
•Switch 5 must be set to the AUTO ANSWER ON RING position.
•The remaining switches should be left in their factory default positions.
2. Connecting the modem to the cabinet.
•The remote modem must be connected to Remote Port 2 on the connection interface
board in the ILS cabinet.
3. Configure remote modem init string.
•With a PC running NM7000 RMM software, log in with MASTER access on the Local Port.
•Select «Modem configuration» from the ILS menu.
•Set the Remote port 2 modem configuration as foll ows (All ‘0’ letters are zeroes.):
Figure 5-62 ILS port settings for USRobotics dial up connection
4. Initializing the remote modem.
•Pressing the «Send init» butt on in the ILS modem se ttings windo w makes the ILS send the
initialization string on the port selected in the port list.
•The ILS will also send the these init strings each time the ILS is powered on.
5.29.3Configuring the local modem
1. All modem DIP switches should be left in their factory default positions.
2. Configure the «Channel Setup» for the dial up connection as follows:
Figure 5-63 Local modem configuration
The modem will automatically dial and connect to the remote ILS whe n a connection with this
channel is selected by the user.
Of course, the teleph one number to dial wh en con necting to this st at ion must be conf igu red in
the ILS Station Setup window.
Users of the DOS version should also select «Use DTR to hangup». The hangup string is
ignored.
NOTE
Some PBX dial tones are not recognized by the modem. In this case the modem will not
start dialing when the user has selected a station to connect to. The problem is cor-
rected by entering X3 in the init string field. E.g.: AT&F1X3
This description applies to Westermo TD-32AC / TD-32DC modems.
5.30.1DIP switch settings.
SW1:1234
ILS Modem0101Leased line, switch off echo and result code. All AT
commands ignored, including +++
PC-Modem1000Dial mode, configured from PC-program
SW2:12345678
ILS Modem10010000Enable SW2:2 to 7. Asynchro-
nous. DTR/DSR disconnected.
PC-Modem10000000Enable SW2:2 to 7. Asynchro-
nous. DTR/DSR not disconnected.
SW3:12345678
ILS and PC100000002-wire connection.
SW4:12345678
ILS and PC011011009600 bps. 8 data bits. No stop
bits.
SW5:12345678
ILS-Modem11110000V.32bis, 9600 bps
PC-Modem11110000Auto detect line parameters.
5.30.2Connecting the modem to the cabinet.
The modem must be connected to Remote Port 1
1
or Remote Port 2 on the connection interface board in the ILS cabinet. The baud rate for this port mu st be 9600 bps (this is the default
setting). There is no need to configure any modem init strings in the NM7000 RMS.
5.30.3Connecting the modem to the PC
The modem must be connected to COM1, 2, 3 or 4. The PC RMM modem connection type
should be set to Leased line. The modem init string should be:
AT\N5&D0&C1&L1&WD
(The hangup string is ignored in the DOS version)
1. It is recommended to use Remote Port 2 on the ILS, because this port has a 16 byte FIFO
that reduces the CPU load caused by the serial communication handlin g. If the modem must
be connected to Remote Port 1, a cable with R TS (pin 4) con nected to DSR (pin 6) both i n the
modem-end, must be used. Also, RTS must not be connected in the ILS-end.
This description applies to Westermo TD-32AC / TD-32DC modems.
5.31.1DIP switch settings.
SW1:1234
ILS Modem0000Dial-up connection. Enable AT commands. Auto
answer.
PC-Modem1000Dial-up connection. Enable AT commands.
SW2:12345678
ILS Modem10010000Enable SW2:2 to 7. Asynchro-
nous. DTR/DSR disconnected.
PC-Modem10000000Enable SW2:2 to 7. Asynchro-
nous. DTR/DSR not disconnected.
SW3:12345678
ILS and PC100000002-wire connection.
SW4:12345678
ILS and PC011011009600 bps. 8 data bits. No stop
bits.
SW5:12345678
ILS and PC11110000
Auto detect line parameters.
5.31.2Connecting the modem to the cabinet.
The modem must be connected to Remote Port 1
1
or Remote Port 2 on the connection interface board in the ILS cabinet. The baud rate for this port must be 9600 bps. The init string for
the port should be:
ATE0&C1&D2&W
5.31.3Connecting the modem to the PC
The modem must be connected to COM1, 2, 3 or 4 . The connect ion type i n the NM70 00 RMM
software must be set to Normal dial-up, the baud rate to 9600 bp s and the init string should
be:
ATX3&C1&D2\N5&W
In the DOS version, the Use DTR to hangup check box should also be checked (the hang-up
string will then be ignored).
1.It is recommended to use Remote Port 2 on the ILS, because this port has a 16 byte FIFO
that reduces the CPU load caused by the serial communication handlin g. If the modem must
be connected to Remote Port 1, a cable with R TS (pin 4) con nected to DSR (pin 6) both i n the
modem-end, must be used. Also, RTS must not be connected in the ILS-end.
No adjustments / alignments shall be done during a Periodic Maintenance. The
parameters should be recorded and compared with their nominal values only.
Over some time, the parameters may change due to natural variations. Variations of less
than 50% of their alarm limits, are considered normal. If a parameter is changing more
than 50% of it's alarm limit, it is likely due to a discrepancy / degeneration of a module,
cable etc. In this case, the reason for the discrepancy should be searched for.
If the result calls for changing a module, rephasing of cables etc., a flight check is oft en
required to verify that para meters are inside operating tolerances.
6.1Weekly inspections
Can be relaxed to 2-Weekly or Monthly intervals if allowed by the Authority.
Test equipment
•NM3710 Field Test Set with portable antenna.
•PC with RMM and printer
6.1.1Field modulation measurements
On the established reference marks for Course Line (THR CL), Sector Width (THR 90)) and
(THR 150) carry out modulation measurements, record the values in Table 6-18.
6.1.2Monitor Parameters recordings
(Can be changed to Monthly intervals if allowed by the Authority.)
•Tx1 to antenna. Open the RMM program in level 1.
•Open Monitor Window 1.
•Open Front Panel.
•Make a paper copy by “Print Screen” function. (Table 6-1)
•Tx2 to antenna. (Executed by tower personnel if ILS is not taken out of service.)
•Make a second paper copy by “Print Screen” function. (Table 6-2)
•Compare Monitor parameter values with previous records.
6.2Monthly inspections
Test equipment
•Field Test Set
6.2.1Field DDM measurements
Using the Field Test Set carry out DDM measurements versus azimuth angles according to
established marks in the near field area of the antenna array.
•Turn off Tx1 COU Mod 90Hz in the Modulation Switches wi ndow. Observe that a transfer
to Tx2 takes place.
a) Check that the text: Aut o ChgOver to TX2 appears in the event list.
Turn off Tx2 COU Mod 90Hz in the Modulation Switches window. Observe that the system
turns off.
b) Check that the text: Auto ShtDown frm TX2 appears in the event list.
c) Check that measurement for PA 27V level < 1V in Maintenance Monitor Windows for COURSE TX1 LPA/GPA and COURSE TX2 LPA/GPA.
Turn on Tx1 COU Mod 90Hz and Tx2 COU Mod 90Hz in the Modulation Switches window
again.
6.4Annual inspections
Test Equipment
•PC including NM7000 RMM program
•NM3710 ILS Field Test Set, including attenuator 20 dB or 30 dB
•Oscilloscope
•Frequency counter 10 Hz to 350 MHz
•RF Signal generator including calibrated ILS modulation
•Digital voltmeter
•Wattmeter with probe covering 108 - 112 MHz, 20 watts.
•Stop watch
6.4.1RF Frequency check.
Set the
AUTOMATIC/MANUAL switch to MANUAL and REMOTE/LOCAL to LOCAL.
6.4.1.1 Course Transmitter Frequency
Switch on Course Tx1.
•Connect the Frequency Counter to CSB Cou test connector in the Change Over Section
via a 50 ohm test cable. (Use 50 ohm input port of the counter)
•If necessary switch off the 90 Hz and 150 Hz modulation through the Local Display and
(Each modulator develops it's own waveform shape due to spreads in insertion phases).
The dynamic maximum point should be approximately -4 volt.
6.4.3.2 SBO waveform checks.
Switch on Course Tx1.
•Connect the Oscilloscope to SBO test connector on Course Tx1 Section via 50 ohm test
cable.
•Set Oscilloscope input mode to DC.
•Check that the waveform resembles the graph below, left.
For best signal resolution on the scope, i t i s advis abl e to set the osci llosc ope i n doubl e sweep
rate and FREE RUN trigger mode ("kissing p atter n" mode). The two hal ves of the 30 Hz waveform will then overlap and look like one waveform when perfect power balance is achieved.
DEMODULATED SBO: 1dB POWER
DEMODULATED SBO: NORMALIMBALANCE (150Hz<90Hz)
Figure 6-1 SBO Waveforms
6.4.3.3 RF Phase
•Connect the Oscilloscope to SBO test connector on Course Tx1 Section via 50 ohm test
cable.
NOTE:
Set the scope's input mode to DC.
Set the oscilloscope in normal trigger mode such that the waveform below can be observed.
The observed waveform should resemble the NORMAL graph on the left.
•Connect the Oscilloscope to CSB test connector on Course Tx1 Section via 50 ohm test
cable.
•Set the Oscilloscope input mode to DC.
Look for the intermediate peak s of the CSB demodul ated wavefo rm. Check t hat the wavefo rm
resembles the graph below (left). The intermediate waveform peaks shal l have the same maximum and minimum level, indicating 90/150 Hz zero phase start.
•Connect the Field Test Set to the CSB Cou test connector on the Change Over Section
through a 20 or 30 dB attenuator.
•From the local panel/display or from the RMM program set Cou Ident modulation to CON-
TINUOUS.
•Use a double shielded 50 ohm coaxial test cable. On the Field Test Set read IDENT MOD.
The nominal reading should be 10% MOD, alternatively the same value as previous
recording. (If voice modulation is applied the Ident modulation shall be 5% modulation
depth.)
6.4.3.7 Voice modulation check. (If voice ident is used)
Switch on Tx1.
•Connect the Oscilloscope to CSB test connecto r on Course Tx1 Section via 50 ohm test
cable.
•From the local panel/display connect the audible transducer by command “Set ident
speaker”.
•Listen to the voice message and check that the waveform has a peak amplitude approxi-
mately six times the Ident waveform peak amplitude seen on the oscilloscope.
6.4.4CSB output power.
a) Course Tx CSB. Switch off the transmitters. Select Tx1 as Main Tx.
•Insert the power meter into the Course CSB output path of the Cabinet. Switch on Tx1.
Read and record the Tx1 power in Table 6-14.
•Change over to Tx2. Read and record the Tx2 power in Table 6-14.b) Clearance Tx CSB. Switch off the transmitters. Select Tx1 as Main Tx.
•Insert the power meter into the Clearance CSB output path of the Cabinet. Switch on Tx1.
Read and record the Tx1 power in Table 6-14.
•Change over to Tx2. Read and record the Tx2 power in Table 6-14.
6.4.5Monitor Alarm Limit checks.
After completion and recording of the alarm parameters into Table 6-15, compare and check
that the values are the same as in the previous records.
External RF Signal Generator applied as signal source.
Set the
SELECT
AUTOMATIC/MANUAL switch to MANUAL, the REMOTE/LOCAL to LOCAL and MAIN
to Tx1.
Switch on Course Tx1 and Clearance Tx1.
Check Monitor 1 first.
6.4.5.1 CL DDM ALARM
Apply a test signal from the RF Signal Generator to the CL input connector in the back of the
cabinet.(Disconnect first the sig nal cable from the Monitor network). Adjust the signal to the
same RF level and frequency as the normal signal applied to the monitor.
Modulation shall be: 0.0% DDM, 40.0% SDM.
Verify modulation setting (DDM, SDM) by checking the test signal with the Field Test Set.
Use the Local Display and Keyboard or the RMM Program to display the measured val ues
for CL DDM.
•Change DDM from the generator slowl y toward s 150 Hz do minance unt il a poi nt where th e
letter A (indicates that monitor has an ALARM for the displayed parameter) is shown
beside the measured parameter for Monitor 1.
•Check the new generator DDM value with the Field Test Set.
•Record in Table 6-15 line a).
•Change DDM from the generator slowly towards 90 Hz dominance until a point where the
letter A (indicates that monitor has an ALARM for the displayed parameter) is shown
beside the measured parameter for Monitor 1.
•Check the new generator DDM value with the Field Test Set.
•Record in Table 6-15 line aa).
Repeat the procedure for Monitor 2.
6.4.5.2 CL SDM ALARM
The test setup is the same as in a).
Use the Local Display and Keyboard or the RMM Program to display the measured val ues
for CL SDM
•Increase SDM from the generator slowly until a point where the let ter A (indicates that
monitor has an ALARM for the displayed p arameter) i s shown beside the measur ed par ameter for Monitor 1.
•Check the new generator SDM value with the Field Test Set.
•Record in Table 6-15 line b).
•Decrease SDM from the generator slowly unti l a point where the letter A (indicates that
monitor has an ALARM for the displayed p arameter) i s shown beside the measur ed par ameter for Monitor 1.
•Check the new generator SDM value with the Field Test Set.
•Record in Table 6-15 line bb).
•Repeat the procedure for Monitor 2.
6.4.5.3 CL RF ALARM
The test setup is the same as in a).
Use the Local Display and Keyboard or the RMM Program to display the measured val ues
for CL SDM.
•Decrease slowly RF level on the generator until a point where the letter A (indicates that
monitor has an ALARM for the displayed p arameter) i s shown beside the measur ed par ameter for Monitor 1.
•Record the RF level reading at alarm point in Table 6-15 line c).
Repeat the procedure for Monitor 2.
6.4.5.4 DS DDM ALARM
Apply a test signal from the RF Signal Generator to the DS input connector in the back of the
cabinet.(Disconnect first the sig nal cable from the Monitor network). Adjust the signal to the
same RF level and frequency as the normal signal applied to the monitor.
Modulation shall be: 15.5%DDM, 40.0%SDM.
Verify modulation setting (DDM, SDM) by checking the test signal with the Field Test Set.
Use the Local Display and Keyboard or the RMM Program to display the measured val ues
for DS DDM.
•Increase DDM from the generator slowly above nominal value unt il a poi nt where the le tter
A (indicates that monitor has an ALARM for the displayed parameter) is shown beside the
measured parameter for Monitor 1.
•Check the new generator DDM value with the Field Test Set.
•Record in Table 6-15 line d).
•Decrease DDM from the generator slowly below nominal value until a point where the lett er
A (indicates that monitor has an ALARM for the displayed parameter) is shown beside the
measured parameter for Monitor 1.
•Check the new generator DDM value with the Field Test Set.
•Record in Table 6-15 line dd).
Repeat the procedure for Monitor 2.
6.4.5.5 NF DDM ALARM
Apply a test signal from the RF Signal Generator to the NF input connector in the back of the
cabinet.(Disconnect first the signal cable from the Near Field Antenna). Adjust the signal to
the same RF level and frequency as the normal signal applied to the monitor.
Modulation shall be: 0.0% DDM, 40.0% SDM.
Verify modulation setting (DDM, SDM) by checking the test signal with the Field Test Set.
Use the Local Display and Keyboard or the RMM Program to display the measured val ues
for NF DDM
•Change DDM from the generator slowl y toward s 150 Hz do minance unt il a poi nt where th e
letter A (indicates that monitor has an ALARM for the displayed parameter) is shown
beside the measured parameter for Monitor 1.
•Check the new generator DDM value with the Field Test Set.
•Record in Table 6-15 line e).
•Change DDM from the generator slowly towards 90 Hz dominance until a point where the
letter A (indicates that monitor has an ALARM for the displayed parameter) is shown
beside the measured parameter for Monitor 1.
•Check the new generator DDM value with the Field Test Set.
•Record in Table 6-15 line ee).
Repeat the procedure for Monitor 2.
6.4.5.6 CLR DDM ALARM
Apply a test signal from the RF Sign al Generat or to the CLR input connector i n the back of the
cabinet.(Disconnect first the sig nal cable from the Monitor network). Adjust the signal to the
same RF level and frequency as the normal signal applied to the monitor.
Modulation shall be: Same as commisioned values DDM and SDM from CLR output of Monitor network.
Verify modulation setting (DDM, SDM) by checking the test signal with the Field Test Set.
Use the Local Display and Keyboard or the RMM Program to display the measured val ues
for CLR DDM.
Make an alarm by removing the NF cable to the Cabinet. After the specified delay, the transmitter shall change to Tx2, then after the same time lapse Tx2 shall shut down.
Record in Table 6-16 line a) the measured delay time from cable removal until the standby
transmitter has switched off.
6.4.720 seconds inhibit
After shutdown check that an immediate attempted restart of a trans mitter is inhibited at least
20 seconds from the time the standby Tx is shut off.
Record in Table 6-16 line b) the measured time lapse from the standby transmitter has
switched off until the main transmitter is again radiating to the antenna.
6.4.8Battery maintenance
Check with digital voltmeter that the voltage across the battery terminals is between 27.0 V
and 27.8 V. Record in Table 6-17 line a) the measured voltage.
Switch off the mains 220 V. After approximately 10 minutes battery operation with transmi tter
to air check that the volt age across the bat tery te rminals is not less than 23 V. Record in Table
6-17 line b) the voltage.
Switch back to mains 220 V. After 24 hours check that the battery terminal voltage is again
back between 27.0 to 27.8 V. Record in Table 6-17 line c) the voltage. The 27 V can be monitored from the RMM facility.
6.4.9Final check LLZ
Check that all electrical contacts and connectors are not loose.
Carry out a visual check outside and inside of the equipment.
Check that all switches and keys are in NORMAL position.
Check that the System lamp status is NORMAL after the switches AUTO/MANUAL and