T
Surdial
M
X
Equipped with various monitoring and profile functions to
pursue optimized treatments
Special Features
●
15 inch clear touch display
●
On-line HDF (Double needle and Single needle with double blood pumps)
●
Machine assisted loading pump segments of bloodline
Specifications
General data
Dimension 480(W)×895(D)×1625(H)
Weight Approx. 120kg
Power supply AC230V±10%, 50/60Hz
Power consumption Max. 2000VA
Water supply Temperature: 5 to 30
Extracorporeal circuit
Arterial pressure monitoring
Venous pressure monitoring
Arterial blood pump
Single needle pump(option)
ON-LINE HDF pump(option)
Air bubble detector Ultrasonic sensor
Heparin pump Delivery range : 0.0 to 10.0 mL/hr
Hydraulic circuit
Ultrafiltration Volumetric control
Dialysate flow rate Range : 100 to 800 mL/min
Dialysate temperature
AC110V±10%, 50/60Hz
Pressure: 0.05 to 0.74 MPa
Flow rate: 900mL/min or more
Measurement range: -500 to 500 mmHg
Accuracy
Measurement range: -500 to 500 mmHg
Accuracy
Flow range: 0,10 to 600mL/min
Accuracy: ±10%
Flow range: 0,10 to 500mL/min
Accuracy: ±10%
Two conditions create an air bubble alarm: any
single bubble greater than 10 μL in size and
when bubbles of 0.3 μL or larger, accumulate
over a 1-minute period into a volume of 1, 10,
50 or 300 μL.
Bolus range : 0.5 to 10.0 mL
Accuracy : ±10%
Syringe size : 10,20,30mL
UF rate : 0,0.10 to 4.00L/hr
Accuracy : ±30g/hr
Range : 34.0 to 39.0
℃
: ±10mmHg
: ±10mmHg
℃
Concentration range Bicarbonate dialysis
Water +B : 24 to 70 mmol/L
Water +A+B : 125 to 165 mmol/L
(Na
Stability : 0.3mS/cm or less
Acetate dialysis
Water +A : 125 to 165 mmol/L
(Na+electrolyte concentration)
Stability : 0.3mS/cm or less
Blood leak detector Optical
Rinse Program Water rinse
Chemical disinfection
Chemical decalcification
Hot rinse
Hot water disinfection
Hot citric disinfection
IHR
Function
Acetate and Bicarbonate dialysis
Isolated ultra filtration
Single needle with single blood pump
UF profile
Concentration profile
Bicarbonate powder cartridge holder
Leak Detector (1x extracorporeal, 1x hydraulic)
On-Line Kt/V measurement
Clean Treatment Start (CTS)
Dialysate Infusion Function (DIF)
Option
Blood pressure monitoring (BPM)
Blood volume monitor (BVM)
Acetate bag connector nozzle
Front hook for Acetate solution bag
Acetate bag spike nozzle
Central system with 2 solution ports
BPM remote button
Nurse call button
Central alarm output
* ON-LINE HDF/HF, priming/bolus/reinfusion
* Single Needle with Double Pump
* Patient card
* Network
* Bicarbonate nozzle
+
electrolyte concentration)
*: Factory installed option