Equipped with various monitoring and profile functions to pursue optimized treatments
Special Features
On-line HDF (Double needle and Single needle with double blood pumps)
Machine assisted loading pump segments of bloodline
Specifications
General data
Dimension 480(W)×895(D)×1625(H)
Weight Approx. 120kg
Power supply AC230V±10%, 50/60Hz
Power consumption Max. 2000VA
Water supply Temperature: 5 to 30
Extracorporeal circuit
Arterial pressure monitoring
Venous pressure monitoring
Arterial blood pump Single needle pump(option)
ON-LINE HDF pump(option)
Air bubble detector Ultrasonic sensor
Heparin pump Delivery range : 0.0 to 10.0 mL/hr
Hydraulic circuit
Ultrafiltration Volumetric control
Dialysate flow rate Range : 100 to 800 mL/min
Dialysate temperature
AC110V±10%, 50/60Hz
Pressure: 0.05 to 0.74 MPa Flow rate: 900mL/min or more
Measurement range: -500 to 500 mmHg Accuracy
Measurement range: -500 to 500 mmHg Accuracy
Flow range: 0,10 to 600mL/min Accuracy: ±10%
Flow range: 0,10 to 500mL/min Accuracy: ±10%
Two conditions create an air bubble alarm: any single bubble greater than 10 μL in size and when bubbles of 0.3 μL or larger, accumulate over a 1-minute period into a volume of 1, 10, 50 or 300 μL.
Bolus range : 0.5 to 10.0 mL Accuracy : ±10% Syringe size : 10,20,30mL
UF rate : 0,0.10 to 4.00L/hr Accuracy : ±30g/hr
Range : 34.0 to 39.0
: ±10mmHg
: ±10mmHg
Concentration range Bicarbonate dialysis
Water +B : 24 to 70 mmol/L Water +A+B : 125 to 165 mmol/L
(Na
Stability : 0.3mS/cm or less
Acetate dialysis
Water +A : 125 to 165 mmol/L
(Na+electrolyte concentration)
Stability : 0.3mS/cm or less
Blood leak detector Optical
Rinse Program Water rinse
Chemical disinfection Chemical decalcification Hot rinse Hot water disinfection Hot citric disinfection IHR
Function
Acetate and Bicarbonate dialysis Isolated ultra filtration Single needle with single blood pump UF profile Concentration profile Bicarbonate powder cartridge holder Leak Detector (1x extracorporeal, 1x hydraulic) On-Line Kt/V measurement Clean Treatment Start (CTS) Dialysate Infusion Function (DIF)
Option
Blood pressure monitoring (BPM) Blood volume monitor (BVM) Acetate bag connector nozzle Front hook for Acetate solution bag Acetate bag spike nozzle Central system with 2 solution ports BPM remote button Nurse call button Central alarm output * ON-LINE HDF/HF, priming/bolus/reinfusion * Single Needle with Double Pump * Patient card * Network * Bicarbonate nozzle
+
electrolyte concentration)
*: Factory installed option