NIPRO SURDIAL 55plus Specifications

T
SURDIAL
M
55Plus
Compact and simple designed to allow easy settings and operations
Special Features
180 degrees swivable touch panel display
Single needle system with one clamp
UF profile and Conductivity profile
Detection of micro air bubbles
Specifications
General data
Dimension 280(W)×420(D)×1365(H)
Weight Approx. 72kg
Power Supply AC230V±10%, 50/60Hz
AC110V±10%, 50/60Hz
Power consumption Max. 2000VA
Water Supply Temperature:
15 to 30
5 to 30℃ (with heat exchanger, option) Pressure: 0.05 to 0.74 MPa Flow rate: 900mL or more
Extracorporeal circuit
Arterial pressure monitoring
Venous pressure monitoring
Measurement range: -500 to 500 mmHg Accuracy
: ±10mmHg
Measurement range: -500 to 500 mmHg Accuracy
: ±10mmHg
Air Bubble Detector Ultrasonic Sensor
Two conditions create an air bubble alarm: any single bubble greater than 10 μL in size and when bubbles of 0.3 μL or larger, accumulate over a 1-minute period into a volume of 1, 10, 50 or 300 μL.
Arterial blood pump Flow range: 0,10 to 600mL /min
Accuracy: ±10%
Heparin pump Delivery range: 0.0 to 20.0 mL /hr
Bolus range: 0.1 to 9.9 mL Accuracy
: ±10%
Syringe size: 10,20,30mL
Hydraulic circuit
Ultrafiltration
Volumetric control UF rate:0,0.10 to 5.00L /hr Accuracy: ±30g/hr
Dialysate flow rate 300 to 800 mL /min
Dialysate temperature
34.0 to 39.0
Concentration range Bicarbonate dialysis
Water +B: 5 to 65 mmol/L Water +A+B: 130 to 160 mmol/L
(Na
Stability
: 0.3mS/cm or less
Acetate dialysis
Water +A: 130 to 160 mmol/L
(Na+electrolyte concentration)
Stability: 0.3mS/cm or less
Blood leak detector Optical
Rinse Program Water rinse
Chemical disinfection Chemical decalcification Hot rinse Hot citric disinfection Tepid rinse
Function
Acetate and Bicarbonate dialysis Isolated ultra filtration Single needle with single blood pump UF profile Concentration profile Dose finder (Kt/V calculator)
10.4” color touch display
Option
Blood pressure monitoring (BPM) LAN connector Front hook for solution bag CF flushing unit Dialysate solution tank tray Acid rinsing port unit Potential equalization conductor * Heat exchanger unit(Factory Option) * Bicarbonate powder cartridge system * Central system for A solution (Factory Option)
+
electrolyte concentration)
*: Factory installed option