Nikon SMZ1000, SMZ800 Instruction Manual

Page 1
M248 E 06.2.NF.3
Stereoscopic Zoom Microscope
SMZ1000 / SMZ800
Instructions
Page 2
Page 3
Thank you for purchasing the Nikon product. This instruction manual is written for the users of Nikon Stereoscopic Zoom Microscopes. To ensure correct usage, read this manual carefully before operating the instrument.
• It is prohibited to reproduce or transmit this manual in part or
whole without Nikon’s expressed permission.
• The contents of this manual are subject to change without notice.
• Although every effort has been made to ensure the accuracy of
this manual, if you note any points that are unclear or incorrect, contact your nearest Nikon representative.
• Some of the products described in this manual may not be
included in the set you have purchased.
• If the equipment is used in a manner not specified by the
manufacturer, the protection provided by the equipment may be impaired.

Warning/Caution Symbols Used in This Manual

Though Nikon products are designed to provide you with the utmost safety during use, incorrect usage or disregard of the instructions may cause personal injury or property damage. For your own safety, read the instruction manual carefully and thoroughly before using the product. Do not discard this manual. Always keep it near the product for easy reference. Inside this instruction manual, safety instructions are indicated with the symbols shown below. Be sure to follow the instructions marked with these symbols for your safety.
Symbol Meaning
WARNING
CAUTION
Disregarding instructions marked with this symbol may lead to death or serious injury.
Disregarding instructions marked with this symbol may lead to injury or property damage.
1
Page 4

Be sure to always follow these guidelines.

WARNING

1. Intended use of this product
This product is intended only for microscopy. Do not use it for any other purpose.
2. Do not disassemble
Disassembly may result in damage to the instrument. Never disassemble any part except as described in this operation manual. Contact your Nikon representative if you notice any malfunction of this instrument.
3. Check the input voltage
When using an illuminator, check that the input voltage displayed on the power supply of the illuminator matches the operating voltage. Contact your Nikon representative if the displayed voltage does not match the operating voltage. Use of an improperly matched illuminator may result in damage to equipment.
4. Power cord
Always turn the power switch off and unplug the power cord when replacing the lamp of an illuminator as failure to do so may result in electric shock or equipment damage.
5. Heat from the light source
When using an illuminator, do not place cloth or paper or highly flammable materials, such as gasoline, benzene, thinner or alcohol, near the lamp as there is a danger of fire.
• Always use lamps that are of the specified rating only.
For the lamp rating, please refer to the operation manual.
• Always turn the power switch off and unplug the
power cord when replacing the lamp of a illuminator as failure to do so may result in electric shock or equipment damage.
2
Page 5

CAUTION

1. Caution when replacing lamps
2. Check the light source
The lamp and the surrounding area become very hot during and immediately after using the illuminator. Be careful not to burn yourself. In addition, when replacing a lamp be sure the lamp has cooled sufficiently.
Always use lamps that are of the specified rating only. Use of a lamp having a different specification may result in damage to equipment. For the lamp rating, please refer to the operation manual.
3
Page 6
1. Installation location
Note the following points when installing the stereoscopic microscope.
• Install the microscope in a location with the temperature between
0° and 40°C, and humidity of less than 80%. If installed in a hot and humid location, mold may form on the lenses or condensation may occur inside, resulting in reduced performance or damage to the microscope.
• Do not install the microscope in a location subject to direct
sunlight.
• Install the microscope in a location that is not subject to vibration.
• Install the microscope in a location that is free from dust and dirt.
2. Handle the microscope carefully
The stereoscopic microscope is a precision optical instrument. Handle it carefully and do not subject it to impact. Shock from impact during transportation or operation as well as forcible operations may cause damage to the instrument.
3. Cleaning the lenses
Do not let dust, fingerprints, etc. to get on the lenses. Dirt on the lenses will adversely affect the view of the image. If any lenses become dirty, clean them as described below.
• Use an air blower to blow dust away. If this does not suffice, brush
away the dust with a soft brush or gently wipe it away with a piece of gauze.
• Only if the lenses become dirty with fingerprints or grease stains,
slightly dampen a piece of soft, clean cotton cloth with absolute alcohol (ethyl alcohol or methyl alcohol) and gently wipe away the dirt. Do not use the same part of the cloth more than once.
• Since absolute alcohol is highly flammable, be careful in handling
it so that it does not ignite.
• Observe the manufacturer’s handling instructions when handling
absolute alcohol.
4
Page 7
Be sure to always follow these guidelines.
4. Cleaning painted or plastic parts
Use of silicon cloth is recommended when cleaning painted parts, plastic parts and printed parts. If such a part becomes excessively dirty, wipe it gently with gauze dampened in a mild detergent solution. Do not use organic solvents (such as alcohol, ether, or paint thinner) as this may result in deformation of the part or remove of lettering.
5. Storage
Store the microscope in a location with low humidity where mold is unlikely to form. Do not store the microscope in a location subject to direct sunlight or high temperature and humidity. During storage, place a plastic cover over the equipment to prevent dust accumulation.
6. Regular inspections
Regular inspections are recommended in order to maintain peak performance. Please consult your Nikon representative for details about regular inspections.
5
Page 8

Contents

Warning/Caution Symbols Used in This Manual ...........................1
Be sure to always follow these guidelines. ....................................2
WARNING ..........................................................................................2
CAUTION ........................................................................................... 3
I Nomenclature ...................................................................................8
II Assembly .........................................................................................12
III Usage ................................................................................................ 15
1 Preparations For Observation ......................................................15
1. Adjust the torque of the focus knob.......................................15
2. Adjust the interpupillary distance. ......................................... 15
3. Adjust the diopter. ................................................................. 16
2 Focusing ...................................................................................... 16
1. Check the working distance................................................... 16
2. Focus on the sample. ............................................................. 16
3 Zoom ........................................................................................... 17
1. Change the zooming magnification. ......................................17
4 If You Cannot Focus On The Sample Though The Zooming
Body Is At The Highest Position. ................................................18
IV Using Accessories ........................................................................19
1 Reticles ........................................................................................19
2 Eyepiece Tube ............................................................................. 19
1. P-BTL low eye-level eyepiece tube .......................................19
2. P-BERG tilted eyepiece tube .................................................20
3 P-IER Eyelevel Riser .................................................................. 20
4 Plan 1X ERGO Objective ...........................................................21
5 SM-S4L 4 X 4 Stage L ................................................................22
V Illuminator ....................................................................................... 23
1 Halogen Illuminators ...................................................................23
1. Lamp ......................................................................................24
2. Attaching G-LS to the stand [C-PS & C-PSC] ...................... 25
3. Arm for halogen illuminators ................................................25
4. Power supplies .......................................................................27
6
Page 9
5. Lighting area adjustment [G-LS]........................................... 29
6. Attaching filters [C-DSLS].................................................... 29
2 P-ICI2 Coaxial Episcopic Illuminator .........................................29
VI Photomicrographic and TV-related Accessories ..................32
1. Photomicrographic magnification ......................................... 33
2. Attaching the beam splitter .................................................... 33
3. Photo reticles ......................................................................... 34
4. Photomicrography procedure ................................................ 34
Table 1: Total Magnification and Real Field...................................... 36
Table 2: Observable Sample Heights [mm] .......................................38
7
Page 10
Nomenclature
I
The figures on p.8 & 9 show a SMZ1000 zooming body, with P-BT binocular eyepiece tube, C-W 10X eyepieces, Plan Apo 1X objective, C-FMC focusing mount, and C-PS160 plain stand.
Diopter ring
Refer to “Adjust the diopter” on p.16.
Eyepieces
10X, 15X, 20X and 30X eyepieces are available.
Eyepiece tube clamp screw
Zooming body clamp screw
Fixes the zooming body to the focusing mount.
Zooming body
Outside diameter of the part to be mounted the focusing mount: ø76
Objective
Plan 1X, Achro 0.5X, Plan 1X ERGO, Plan Apo 0.5X, Plan Apo 1X, ED Plan 1.5X and ED Plan 2X objectives are available. HR Plan Apo series objectives for SMZ1500 can be used also, but the magnifi­cation in this case becomes 1.25 times the magnification indicated on the objective itself. Refer to Table 1 on p.36.
Groove for mounting accessories
Holds the Fluorescent ring illuminator, Fiber optic ring illuminator and the like. Outside diameter: ø60
Stage plate clamp screw
Fixes the stage plate to the stand base.
Eyepiece sleeve
P-BT Binocular eyepiece tube
P-BTL low eye-level eyepiece tube and P­BERG tilted eyepiece tube are also available.
Zooming knob
Changes the magnification of the sample image. Refer to “Zoom” on p.17.
Coarse focus knob
Refer to “Focusing” on p.16.
Fine focus knob
Refer to “Focusing” on p.16.
Stand
C-PS160 Stand base
Various stands are available.
Screw for mounting accessories
Holds the 1/4 lambda plate and the like. The screw on the inside diameter: M55 × 0.75
Stage plate
The black or the white side can be chosen according to the sample.
8
Page 11
Zooming knob clicks ON/OFF screw (inside)
Refer to “ON/OFF of the Zooming Knob Clicks” on p.17.
Clamp tool
Used for the stage plate clamp screw, eyepiece tube clamp screw, etc. Same size as the small hexagonal wrench supplied with the zooming body. Stored in the top of the pillar.
Pillar
I
C-FMC Focusing mount
Screw hole for mounting an Epi-arm for halogen illuminator
Located on both sides of the focusing mount. Holds the G-EIA flexible arm or Epi-arm.
Screw hole for mounting accessories
Located on both sides of the stand base. M5
Torque adjustment ring (Innermost ring)
Adjusts the torque of the focus knob. (Refer to “Adjust the focus knob torque” on p.15.)
Focusing mount clamp screw
Fixes the focusing mount to the pillar.
Pillar clamp screw (Inside)
9
Page 12
The illustrations on p.10 & 11 show a SMZ800 zooming body, with P-BT binocular eyepiece tube, C-W 10X eyepieces, Plan 1X objective, and C-PS plain focusing stand.
Diopter ring
Refer to “Adjust the diopter” on p.16.
Eyepiece
10X, 15X, 20X and 30X eyepieces are available.
Eyepiece tube clamp screw
Zooming body clamp screw
Fixes the zooming body to the arm.
Zooming body
Outside diameter of the part to be mounted on the arm: ø76
Objective
Plan 1X, Achro 0.5X, Plan 1X ERGO, Plan Apo 0.5X, Plan Apo 1X, ED Plan
1.5X and ED Plan 2X objectives are available. HR Plan Apo series objectives for SMZ1500 can be used also, but the magnification in this case becomes 1.25 times the magnification indicated on the objective itself. Refer to Table 1 on p.36.
Groove for mounting accessories
Holds the Fluorescent ring illuminator, Fiber optic ring illuminator and the like. Outside diameter: ø60
Screw for mounting accessories
Holds the 1/4 lambda plate and the like. The screw on the inside diameter: M55
×
0.75
Eyepiece sleeve
P-BT Binocular eyepiece tube
P-BTL low eye-level eyepiece tube and P­BERG tilted eyepiece tube are also available.
Zooming knob
Changes the magnification of the sample image. Refer to “Zoom” on p.17.
Focus knob
Refer to “Focusing” on p.16.
Arm
Arm fixing screw
Holds the arm. The position of the arm can be chosen from either high or low.
Vertical slider
Hole for clip
Holds the clip for retaining the sample.
C-PS Focusing stand
Various stands available.
Stage plate
The black or the white side can be chosen according to the sample.
10
Page 13
Screw hole for mounting an Epi-arm for halogen illuminator.
Screw holes are located on both sides of the arm. Holds the G-EIA flexible arm or Epi-arm.
Screw hole for mounting 6V-10W halogen illuminator
Holds the G-LS 6V-10W halogen illuminator.
Nomenclature
Zooming knob clicks ON/OFF screw (inside)
Refer to “ON/OFF of the Zooming Knob Clicks” on p.17.
Hexagonal wrench hanger
Holds the supplied hexagonal wrenches.
Hexagonal wrench (Large)
Supplied with the focusing stand. Used for changing the height of the arm.
Hexagonal wrench (Small)
Supplied with the zooming body. Used for the eyepiece tube clamp screw and zooming knob clicks ON/OFF screw.
I
Pillar
11
Page 14
Assembly
II
• An equipment model name in the bold-faced brackets [ ] indicates
☞
1
2
3
that the following description is exclusively for that model.
[C-PS160] Attach the pillar to the stand.
Insert the pillar straight into the stand with the groove of the pillar located down as far as the pillar reaches to the bottom. Clamp the pillar using the clamp tool.
Place the stand on the level surface.
Mount the stage plate.
[C-PS160]
Fit the stage plate into the stand base. Tighten the stage plate camp screw using the clamp tool to fix the stage plate.
Pillar
Groove
12
[C-PS & C-PSC]
Fit the stage plate into the stand base while pushing it against the rim in the direction shown by the arrow in the illustration.
[C-PS160]
4
Fix the C-FMC focusing mount to the pillar.
Face the focusing mount to the front and tighten the focusing mount clamp screw.
Note) When fixing the focusing mount to the
upper part of the pillar, securely tighten the clamp screw so as not to fall the focusing mount.
Page 15
[C-PS & C-PSC]
5
The stand arm can be lowered. (If you do not need to lower the arm, jump to step 6.)
Use the hexagonal wrench (large - supplied with the focusing stand) to loosen the arm fixing screw. Reattach the arm using the screw hole at the lower side of the vertical slider. The arm will become 55 mm lower than its original position. Before tightening the screw, make sure that the two pins on the arm fit in the grooves on the vertical slider.
Mount the zooming body.
6
Lightly tighten the zooming body clamp screw to hold the zooming body on the focusing mount or arm.
Note) Do not over-tighten the zooming body
clamp screw since it may cause damage to the instrument.
II
Groove
Pin
Screw the objective into the bottom of
7
the zooming body.
Mount the eyepiece tube.
8
Align the groove on the eyepiece tube with the positioning pin on the zooming body, push the eyepiece tube toward the back to attach. Tighten it to the zooming body using the supplied hexagonal wrench (small) or the clamp tool supplied with the C-PS160.
Positioning pin
Groove
13
Page 16
Assembly
Insert the eyepieces into the eyepiece sleeves.
9
Be sure that it is inserted all the way until it touches the end of the sleeve.
Note) When inserting the eyepiece, hold the rubber cover of the eyepiece.
Holding the diopter ring causes a failure. Especially, be careful with the 10X eyepieces because their rubber covers obstruct the view of the sleeve end.
The hexagonal wrenches and clamp tool can be stored in the
10
pillar of the stand.
Refer to the figures on p.9 and p.11.
II
14
Page 17
III
Increasing the torque.
Usage

Preparations For Observation

1
Adjust the torque of the focus knob.
1
Adjust the torque of the focus knob so as not to fall down the zooming body on its own weight.
[C-FMC]
Turn the torque adjustment ring using the clamp tool. Turning the ring in the direction of the arrow indicated on the side surface of the focusing mount will increase the torque.
Note) • Do not twist the left and right focus
knobs.
• Do not turn the knob beyond its turning limit.
• Do not turn the fine focus knob while holding the coarse focus knob or vice versa. Otherwise, the damage to the instrument will result.
Increasing the torque.
TOR
QUE
[C-PS and C-PSC]
Rotating the left and right focus knobs in the opposite direction as shown in the illustration will increase the torque.
Adjust the interpupillary distance.
2
This adjustment should be performed every time the observer is changed since the interpupillary distance differs between individuals. Adjust the interpupillary distance so that the view field for each eye is merged into one. Move while holding each sleeve with both hands.
(To reduce the torque, turn the knob in the direction opposite to the arrow.)
15
Page 18
Adjust the diopter.
3
This adjustment should be performed every time the observer is changed since the eyesight differs between individuals.
1 Turn the diopter rings on both eyepieces to set them at the 0 position
(match the 0 line with the index line).
2 Turn the zooming knob to the highest magnification. Focus on the
sample using the focus knob. (Refer to “Focusing”.)
3 Turn the zooming knob to the lowest magnification. Peering
through the left eyepiece with the left eye, focus on the sample using the diopter ring on the left eyepiece. Then, peer through the right eyepiece with your right eye and focus on the sample using the diopter ring on the right eyepiece.
4 Repeat steps 2 and 3 until the image is kept focused even though the
zooming magnification is changed. This adjustment ensures sharp image throughout the zooming range.

Focusing

2
Check the working distance.
1
The distance between the focus plane and the bottom of the objective is called the working distance. Focusing is easier if beforehand the distance between the specimen and the bottom of the objective is set to the working distance. The working distance differs depending on the objective being used. Please refer to Table 1 on p.36.
16
Focus on the sample.
2
Turning the focus knob moves the zooming body up and down and brings focus on the sample.
[C-FMC]
Using the fine focus knob will permit the delicate focusing. Especially, it is effective for the high magnification observation.
Page 19
Usage

Zoom

3
Change the zooming magnification.
1
Turning the zooming knobs on the left and right side of the zooming body will change the magnification of the sample image.
■ Total Magnification
The zooming knob has the indication of the zooming magnification. The total magnification can be found by multiplying the magnification of the objective times that of the eyepieces.
Note) When the HR Plan Apo series objective for the SMZ1500 is used,
the total magnification will be obtained by multiplying by 1.25.
■ ON/OFF of the Zooming Knob Clicks
The clicks on the zooming knob can be eliminated.
1 Insert the supplied hexagonal wrench
(small) or the clamp tool supplied with the C-PS160 into the screw hole on the rear of the zooming body.
2 Turning the inside screw in the
direction opposite to the arrow eliminates the clicks (OFF). Turning the same screw in the direction of the arrow puts on back the clicks (ON).
III
17
Page 20
Usage
If You Cannot Focus On The Sample Though
4
The Zooming Body Is At The Highest Position.
When you use the 0.5X objective or observe a tall sample, you may not be able to focus on the sample though turning the focus knob to raise the zooming body to its highest position.
[C-PS160]
Exchange the pillar for the C-EP extension pillar available on order. Attach the C-EP extension pillar to the stand in the same manner as for the standard pillar. (Hold the side having the groove downward. Refer to item 1. on p.12.) Put the stopper ring on the pillar and then the focusing mount. Tighten the focusing mount clamp screw.
Note) Contact the bottom surface of the focusing mount with the stopper
ring and securely tighten each clamp screw.
Extension pillar
III
18
Stopper ring
Correct Wrong
[C-PS & C-PSC]
Use the C-ER auxiliary adapter available as an option. (This adapter can’t be used for the SMZ1000.)
• Refer to the Table 2 on p.38 for the sample heights which may be
☞
viewed.
Auxiliary adapter
Page 21
IV
Separately sold reticle
Reticle
Pattern surface
Field ring
Using Accessories

Reticles

1
Your reticles may be attached to the eyepiece. Once remove the field ring (or lens room of the 20X and 30X) from the eyepiece. Attach the reticle with its pattern surface facing down to the eyepiece and reattach the field ring (or lens room).
• Refer to the Table 1 on p.36 for the size
☞
of the reticle which may be attached.
Note) The 20X and 30X eyepieces adopt the
internal focusing system, and the primary image has the magnification. Use the reticle with 1.3X pattern for 20X eyepiece, and with 1.4X pattern for 30X eyepiece. A reticle with cross hairs, scaled cross hairs, or photographic mask is available on order for 10X eyepiece.

Eyepiece Tube

2
P-BTL low eye-level eyepiece tube
1
When using a beam splitter or coaxial episcopic illuminator, it is possible to lower the eye level (eye level during observation) by using this eyepiece tube in place of the binocular eyepiece tube. The installation procedure for this eyepiece tube is the same as for the P-BT binocular eyepiece tube.
P-BTL
19
Page 22
P-BERG tilted eyepiece tube
2
This eyepiece tube allows the angle of the eyepiece sleeve to be continuously adjusted as desired by the observer. It is also possible to change the eye level by rotating the interpupillary distance adjuster (shown in gray in the illustration) 180°. The installation procedure for this eyepiece tube is the same as for the P-BT binocular eyepiece tube.
P-BERG
CAUTION

P-IER Eyelevel Riser

3
When the eye level is too low, it is possible to raise the eye level by 25 mm by attaching a P-IER Eyelevel riser between the zooming body and the eyepiece tube. It is possible to use up to three Eyelevel risers at once. Avoid using the four or more Eyelevel risers at once or using the other intermediate tube together such as a beam splitter or coaxial episcopic illuminator. It may cause the vignetting around the image periphery.
Be careful not to put your fingers between the interpupillary distance adjuster and the lower part of the eyepiece tube when adjusting the angle of the sleeve.
Eyelevel riser
20
Page 23
Using Accessories

Plan 1X ERGO Objective

4
Using the Plan 1X ERGO objective allows you to continuously change the eye level through a 40 mm range while maintaining a constant working distance (50 mm).
1 Rotate the upper knurled ring on the
Plan 1X ERGO objective to attach it to the zooming body.
2 Rotate the focus knobs on the focusing
stand to raise or lower the microscope to set to the height best suited for viewing.
3 Rotate the lower knurled ring to focus
on the specimen. Use the microscope as usual after setting the eye level.
Note) When attaching to the SMZ1000, vignetting will occur in the lateral
direction of the viewfield at the high eye level with 1X or lower zoom magnification.
Plan 1X ERGO
IV
21
Page 24

SM-S4L 4 X 4 Stage L

4 X 4 stage
Stage
clamp screw
Stage adapter fixing screw
Stage adapter
5
SM-S4L 4 X 4 stage L may be attached by using the C-4SA stage adapter. In this case, use the C-ER auxiliary adapter or C-EP extension pillar because the position of the sample will become higher. (Refer to “4. If you cannot focus on the sample though the zooming body is at the highest position.”.)
1 Remove the stage plate from the stand
base, and attach the C-4SA stage adapter in place using the fixing screws supplied with the adapter.
2 Sufficiently loosen the stage clamp
screw of the SM-S4L 4 X 4 stage L. Fit and mount the stage onto the stage adapter with the side equipped with the stage clamp screw facing front. Tighten the stage clamp screw.
Using Accessories
IV
22
Page 25
Illuminator
V

Halogen Illuminators

1
There are two illuminators available: the G-LS of 6V 10W and the C­DSLS of 6V 20W.
Note) These illuminators are not adequate to observe the sample having a
high reflective surface. Use the P-ICI2 Coaxial Episcopic Illuminator on p.29 for this purpose.
G-LS
Lamp Socket A (6V10W)
Lamp house
C-DSLS
Illuminator A (6V20W)
Lamp house
Filter cap
ø33 mm filter attachable
Lamp socket
(may be moved
back and forth)
Mounting screw
Lamp socket
Holder and Epi arm mounting screw hole
Plug
Plug
23
Page 26
1
Lamp
CAUTION
The lamp and the surrounding area become very hot during and immediately after using the illuminator. Be careful not to burn yourself. In addition, be sure the lamp has cooled sufficiently before replacing a lamp.
Pull the lamp socket from the lamp house and insert the lamp straight into the socket as far as it will go. Only use halogen lamps of the specified rating. Next, attach the lamp socket to the lamp house.
C-DSLS
Confirmation window
Lamp
(Philips JCR/M 6V-20W)
G-LS
Socket
Lamp
(GE 787 6V-10W)
24
Light shielding tube
Socket Short protrusion
Holder mounting screw hole
Note) Do not touch the glass part of a lamp with your bare hands as your
skin may burn and stick to the lamp, possibly reducing its level of light output or causing damage to the lamp. When replacing a
Page 27
Mounting screw
Arm
Holder
Protru­sion
G-LS
Illuminator
lamp, handle it through a piece of cloth or through its packaging until replacement is complete. If you get fingerprints on the lamp, wipe gently with a piece of clean cloth dampened with alcohol.
Attaching G-LS to the stand [C-PS & C-PSC]
2
Set so that the protrusion on the lamp house aligns with the inside of the screw hole on the
G-LS
arm of the stand (screw hole for mounting 6V10W halogen illuminator) and secure it in place from the outside of the arm using the mounting screw.
Screw hole for mounting 6V10W halogen illuminator
Arm for halogen illuminators
3
It is possible to freely adjust the direction of illumination by using a special arm.
V
Mounting screw
1 G-EIA flexible arm [G-LS]
Set so that the protrusion on the lamp house aligns with the inside of the hole in the holder of the flexible arm and secure it in place from the outside of the holder using the mounting screw.
Flexible arm
Secure by screwing the mounting screw supplied with the arm into the arm mounting screw hole in the stand (screw hole for mounting an Epi-arm).
Mounting screw
25
Page 28
2 Epi arm
Slightly loosen the clamp knob on the arm. Screw in the mounting screw at the end of the Epi arm into the screw hole of the stand (screw hole for mounting an Epi-arm) while turning the joint. Next, securely tighten the nut using the supplied wrench. Epi arm may be attached also to the screw hole for mounting accessories of C-PS160.
Screw the mounting screw at the other end of the Epi arm into the screw hole on the lamp house in the same manner as described above. Set the lamp house in a position that makes for easy viewing and tighten the clamp knob.
Joint
Mounting screw
Arm
Clamp knob
Simultaneously clamps two arms/joints.
Nut
Spanner
26
Page 29
Illuminator
Power supplies
4
Use the Power Supply XN A for regions using 100/120 V power, and the Power Supply TN-PSE30W A for regions using 230 V power. Insert the plug of the Halogen illuminator into the output connector located on the rear of the power supply.
Input voltage display
Check that the voltage displayed matches the voltage used in your area. If the voltages do not match, do not use the power supply and contact your Nikon representative.
4
5
3
6
V
O
L
T
A
G
E
Power cord
Power switch
Brightness adjuster
Changes the illumination brightness when rotated.
Fuse
Always use the fuse specified.
(Rear view)
Output connector
Connect the lamphouse cable here.
V
Power Supply XN A (only for regions with 100/120 V power)
Input voltage display
Brightness adjuster
Changes the illumination brightness when rotated.
Check that the voltage displayed matches the voltage used in your area. If the voltages do not match, do not use the power supply and contact your Nikon representative.
(Rear view)
Output
Power switch
Power cord
connector
Connect the lamphouse cable here.
AC input connector
Connect the specified power cord here.
Power Supply TN-PSE30W A (only for regions with 230 V power)
27
Page 30
Power Supply XN A (only for regions with 100/120 V power)
Input rating Power supply for 100V regions only:
100 V AC, 50/60 Hz, 30 W
Power supply for 120V regions only:
120 V AC, 50/60 Hz, 30 W
Output rating 3-6 V AC, 3.3 A
Fuse rating 250 V, 1 A
Power Supply TN-PSE30W A (only for regions with 230 V power)
Input rating 230V AC, 50/60Hz, 0.3A
Voltage fluctuation ±10%
Output rating AC 6V 5A
Internal fuse Fast acting type, F1A/250V, 5.2 × 20 mm
Operating Altitude: 2000 m max. environment Temperature: 0° - 40°C
(indoor use only)
Relative humidity: 85% max.
(no condensation) Pollution level: Degree 2 Installation: Category 2
28
Storage Temperature: -20° - 60°C environment Relative humidity: 90% max.
(no condensation)
Protection class Class 1
Power cord Use only the following power supply cord.
Using the wrong power cord could result in danger or fire. The protection Class 1 equipment should be connected to PE (protective earth) terminal.
• For 220-240V AC area
Approved according to EU/EN standards, 3 conductor grounding Type H05VV-F, 3 m long maximum, rated at 250V AC minimum.
Page 31
Illuminator
Lighting area adjustment [G-LS]
5
It is possible to adjust the lighting area by moving the lamp socket back and forth. Adjust the lighting area to allow easy viewing.
Attaching filters [C-DSLS]
6
Rotate and remove the filter cap at the end of the lamp house and insert a filter. Use filters that have a diameter of 33 mm and a width of
1.5 ~ 2 mm.

P-ICI2 Coaxial Episcopic Illuminator

2
Light guide mount
Light guide clamp screw
Main body
Eyepiece tube clamp screw
Luminous section
Light guide
Filter
ø33 mm
Filter cap
Fiber illuminator
Fiber illuminator (C-FI115 for 100/120V region and C-FI230 for 230V region) available on order is necessary for the light source. Read also the Instruction Manual of the Fiber Illuminator C-FI115/C-FI230.
V
Incident section
Magnification of the P-ICI2 Coaxial Episcopic Illuminator is 1.5X. Total magnification is obtained by multiplying 1.5 by the total magnification of the basic set of the microscope.
Note) Usable magnification for the P-ICI2
Coaxial Episcopic Illuminator is restricted. Use the magnification within the range shown in the right
Objective
Acro 0.5X 3X or higher Plan 1X 1.5X or higher Plan 1X ERGO 1.5X or higher Plan Apo 0.5X 4X or higher Plan Apo 1X 2X or higher ED Plan 1.5X 2X or higher ED Plan 2X Not usable HR Plan Apo 0.5X 2.5X or higher HR Plan Apo 1X 2.5X or higher HR Plan Apo 1.6X Not usable
Zooming
magnification
29
Page 32
table. If the magnification out of the
Groove
Positioning pin
specified range is used, uneven illumination will occur or the seeing of the sample image will be degraded.
Mount the P-ICI2 onto the zooming body
1
aligning the groove on the bottom of the P-ICI2 main body with the positioning pin on the zooming body while tilting the P-ICI2 toward the rear. Tighten the eyepiece tube clamp screw.
Note) When the other intermediate tube such
as a beam splitter is used together, attach it on the P-ICI2.
Screw in the 1/4 lambda plate to the end of
2
the objective. When the C-FIR ring fiber light guide is used together, use the other 1/4 lambda plate separately available for ring fiber illuminator.
Insert the incident section of the light guide
3
into the light guide mount of the fiber illuminator and tighten the clamp screw. Insert the luminous section of the light guide into the light guide mount of the P-ICI2 and tighten the clamp screw.
Luminous section
1/4 lambda plate
30
Page 33
The optical axis of the 1/4 lambda plate is at
Bright
Bright
Bright
Bright
Dark
Dark Dark
Dark
4
the 45° position from the red dot. Brightness is the maximum when the red dot is positioned at the front and switches from bright to dark by each 45° position. Adjust the plate to the position where the brightness and contrast are best suited for the observation.
Insert the filter P-ICI33 NCB14 into the fiber
5
illuminator to improve the color reproducibility when taking a photomicrogaph. Refer to the instruction manual of Fiber Illuminator.
Illuminator
Red dot
V
31
Page 34
VI
Photomicrographic and TV-related Accessories
It is possible to photograph images or display them on a TV monitor by attaching a beam splitter between the zooming body and the eyepiece tube. There are two beam splitters available: the S2 P-IBSS2 with one port and the D2 P-IBSD2 with two ports. Read also the instruction manual of the instrument such as a photomicrographic equipment or TV camera to be used.
F-mount camera
Photomicrographic
equipment FX-III series
C-mount camera
for digital cameras
such as DS-5M
F-mount camera adapter
Digital camera
DS-2M
C-mount TV adapter
LV-TV tube
Optical path switching knob Optical path switching knob
Beam splitter D2
0.55x relay lens
Eyepiece
Rear port Port
Side port
CF PLI projection lens (5X, 4X, 2.5X, 2X)
Photomicrographic equipment adapter
Photo reticle
Binocular eyepiece tube
Beam splitter S2
32
Page 35
Photomicrographic and TV-related accessories
Photomicrographic magnification
1
Photomicrographic magnification = (PLI projection lens magnification)
× (Zoom magnification) × (Objective magnification)
• When the P-ICI2 Coaxial Episcopic Illuminator is used:
Photomicrographic magnification = (PLI projection lens magnification)
× (Zoom magnification) × (Objective magnification) × 1.5
• When the HR Plan Apo series objective is used:
Photomicrographic magnification = (PLI projection lens magnification)
× (Zoom magnification) × (Objective magnification) × 1.25
Attaching the beam splitter
2
Mount the beam splitter onto the zooming body aligning the groove on the bottom of the beam splitter with the positioning pin on the zooming body while tilting the beam splitter toward the rear. Tighten the eyepiece tube clamp screw.
[P-IBSS2] Mount it so as to position the
port at either the left or right.
VI
Groove
Pin
[P-IBSD2] 1) Mount it so as to position the
rear port at the rear.
2) Attach the side port to the beam splitter main body with the positioning pin on the side port fitted into the groove on the beam splitter main body. Tighten the side port clamp screw.
P-IBSS2 Port location
Side port
Pin
Groove
P-IBSD2 Attaching side port
33
Page 36
Photo reticles
3
Attaching the C-FP photo reticle to the C-W10X eyepiece permits you to make focusing for phtotomicrography through the eyepiece. Select the correct frame indicating the area being photographed according to the magnification of the PLI projection lens. For more accurate focusing and photographing area, use the finder of the photomicrographic equipment.
2X projection lens, 35-mm full size
2.5X projection lens, 35-mm half size
Photomicrography procedure
4
1 Switching the optical path
[P-IBSS2] Set the optical path switching knob to PHOTO/BINO.
[P-IBSD2] Set the optical path switching knob to the left or right.
Setting it to the left allows you to use only the rear port and setting it to the right, both the rear and side ports.
100%
100%
100%
100%
2.5X projection lens, 35-mm full size
5X projection lens, 35-mm full size
4X projection lens, 35-mm full size
50%
50%
50%
50%
34
Page 37
Photomicrographic and TV-related accessories
2 Focusing
The following is the focusing procedure using the photo reticle. For the focusing procedure using the finder of the photomicrographic equipment, refer to the instruction manual of the phtotomicrographic equipment.
1 Once remove the field ring of the
eyepiece and attach the photo mask reticle. (Refer to p.19 "1. Reticles".)
2 Turn the diopter ring to focus on the
reticle.
3 Turn the focus knob to focus on both
the reticle and the image.
VI
Note) [P-IBSD2]
You can not observe the image through the left eyepiece if the
Adjust so that the double
crosshairs is clearly
divided in two
optical path switching knob is set to the left. In this case, use the finder of the photomicrographic equipment for focusing. Focusing with the right eyepiece or focusing with the optical path switching knob set to the position other than left may result in defocusing.
3 Adjusting the frame position
Turn the eyepiece to align each frame of the photo mask and the photomicrographic equipment finder. Note that the image seen through the finder is reversed from that seen through the eyepiece.
Finder and eyepiece images
Eyepiece
Finder
35
Page 38
Magnification 1.3X Magnification 1.4X
Eyepieces
10 X 15 X 20 X 30 X

Table 1: Total Magnification and Real Field

Field number 22 Field number 16 Field number 12.5 Field number 7
Working
distance
Total Real field Total Real field Total Real field Total Real field
Objectives [mm] Reticle diameter ø25 Reticle diameter ø19 Reticle diameter ø19 Reticle diameter ø12
magnification [mm] magnification [mm] magnification [mm] magnification [mm]
36
SMZ1000
Acro 0.5 X 189 4 ~ 40 X 55.0 ~ 5.5 6 ~ 60 X 40.0 ~ 4.0 8 ~ 80 X 31.3 ~ 3.1 12 ~ 120 X 17.5 ~ 1.8
Plan 1 X 78 8 ~ 80 X 27.5 ~ 2.8 12 ~ 120 X 20.0 ~ 2.0 16 ~ 160 X 15.6 ~ 1.6 24 ~ 240 X 8.8 ~ 0.9
Plan 1 X ERGO 50 8 ~ 80 X 27.5 ~ 2.8 12 ~ 120 X 20.0 ~ 2.0 16 ~ 160 X 15.6 ~ 1.6 24 ~ 240 X 8.8 ~ 0.9
Plan Apo 0.5 X 123 4 ~ 40 X 55.0 ~ 5.5 6 ~ 60 X 40.0 ~ 4.0 8 ~ 80 X 31.3 ~ 3.1 12 ~ 120 X 17.5 ~ 1.8
Plan Apo 1 X 70 8 ~ 80 X 27.5 ~ 2.8 12 ~ 120 X 20.0 ~ 2.0 16 ~ 160 X 15.6 ~ 1.6 24 ~ 240 X 8.8 ~ 0.9
ED Plan 1.5 X 44.5 12 ~ 120 X 18.3 ~ 1.8 18 ~ 180 X 13.3 ~ 1.3 24 ~ 240 X 10.4 ~ 1.0 36 ~ 360 X 5.8 ~ 0.6
ED Plan 2 X 32.5 16 ~ 160 X 13.8 ~ 1.4 24 ~ 240 X 10.0 ~ 1.0 32 ~ 320 X 7.8 ~ 0.8 48 ~ 480 X 4.4 ~ 0.4
HR Plan Apo 0.5 X 136 5 ~ 50 X 44.0 ~ 4.4 7.5 ~ 75 X 32.0 ~ 3.2 10 ~ 100 X 25 ~ 2.5 15 ~ 150 X 14 ~ 1.4
HR Plan Apo 1 X 54 10 ~ 100 X 22.0 ~ 2.2 15 ~ 150 X 16.0 ~ 1.6 20 ~ 200 X 12.5 ~ 1.3 30 ~ 300 X 7 ~ 0.7
HR Plan Apo 1.6 X 24 16 ~ 160 X 13.8 ~ 1.4 24 ~ 240 X 10.0 ~ 1.0 32 ~ 320 X 7.8 ~ 0.8 48 ~ 480 X 4.4 ~ 0.4
Page 39
Eyepieces
Table 1: Total Magnification and Real Field
Magnification 1.3X Magnification 1.4X
SMZ800
10 X 15 X 20 X 30 X
Field number 22 Field number 16 Field number 12.5 Field number 7
Working
distance
Total Real field Total Real field Total Real field Total Real field
Objectives [mm] Reticle diameter ø25 Reticle diameter ø19 Reticle diameter ø19 Reticle diameter ø12
magnification [mm] magnification [mm] magnification [mm] magnification [mm]
Acro 0.5 X 189 5.0 ~ 31.5 X 44.0 ~ 7.0 7.5 ~ 47.3 X 32.0 ~ 5.1 10.0 ~ 63.0 X 25.0 ~ 4.0 15.0 ~ 94.5 X 14.0 ~ 2.2
Plan 1 X 78 10.0 ~ 63.0 X 22.0 ~ 3.5 15.0 ~ 94.5 X 16.0 ~ 2.5 20.0 ~ 126.0 X 12.5 ~ 2.0 30.0 ~ 189.0 X 7.0 ~ 1.1
Plan 1 X ERGO 50 10.0 ~ 63.0 X 22.0 ~ 3.5 15.0 ~ 94.5 X 16.0 ~ 2.5 20.0 ~ 126.0 X 12.5 ~ 2.0 30.0 ~ 189.0 X 7.0 ~ 1.1
Plan Apo 0.5 X 123 5.0 ~ 31.5 X 44.0 ~ 7.0 7.5 ~ 47.3 X 32.0 ~ 5.1 10.0 ~ 63.0 X 25.0 ~ 4.0 15.0 ~ 94.5 X 14.0 ~ 2.2
Plan Apo 1 X 70 10.0 ~ 63.0 X 22.0 ~ 3.5 15.0 ~ 94.5 X 16.0 ~ 2.5 20.0 ~ 126.0 X 12.5 ~ 2.0 30.0 ~ 189.0 X 7.0 ~ 1.1
ED Plan 1.5 X 44.5 15.0 ~ 94.5 X 14.7 ~ 2.3 22.5 ~ 141.8 X 10.7 ~ 1.7 30.0 ~ 189.0 X 8.3 ~ 1.3 45.0 ~ 283.5 X 4.7 ~ 0.7
ED Plan 2 X 32.5 20.0 ~ 126.0 X 11.0 ~ 1.7 30.0 ~ 189.0 X 8.0 ~ 1.3 40.0 ~ 252.0 X 6.3 ~ 1.0 60.0 ~ 378.0 X 3.5 ~ 0.6
HR Plan Apo 0.5 X 136 6.3 ~ 39.4 X 35.2 ~ 5.6 9.4 ~ 59.1 X 25.6 ~ 4.1 12.5 ~ 78.8 X 20.0 ~ 3.2 18.8 ~ 118.1 X 11.2 ~ 1.8
HR Plan Apo 1 X 54 12.5 ~ 78.8 X 17.6 ~ 2.8 18.8 ~ 118.1 X 12.8 ~ 2.0 25.0 ~ 157.5 X 10.0 ~ 1.6 37.5 ~ 236.3 X 5.6 ~ 0.9
HR Plan Apo 1.6 X 24 20.0 ~ 126.0 X 11.0 ~ 1.7 30.0 ~ 189.0 X 8.0 ~ 1.3 40.0 ~ 252.0 X 6.3 ~ 1.0 60.0 ~ 378.0 X 3.5 ~ 0.6
37
Page 40

Table 2: Observable Sample Heights [mm]

SMZ1000 + C-PS160 + C-FMC
Table 2: Observable Sample Heights
Mount Mount up When using
Objectives normal side down C-EP
position position extension pillar
Acro 0.5X 0 ~ 24 — 0 ~ 124 0 ~ 13
Plan 1X 0 ~ 97 0 ~ 42 0 ~ 197 0 ~ 86
Plan 1X ERGO 0 ~ 114 0 ~ 59 0 ~ 214 0 ~ 103
Plan Apo 0.5X 0 ~ 25 — 0 ~ 125 0 ~ 14
Plan Apo 1X 0 ~ 102 0 ~ 47 0 ~ 202 0 ~ 91
ED Plan 1.5X 0 ~ 111 0 ~ 56 0 ~ 211 0 ~ 100
ED Plan 2X 0 ~ 108 0 ~ 53 0 ~ 208 0 ~ 97
HR Plan Apo 0.5X 0 ~ 33 — 0 ~ 133 0 ~ 22
HR Plan Apo 1X 0 ~ 103 0 ~ 48 0 ~ 203 0 ~ 92
HR Plan Apo 1.6X 0 ~ 112 0 ~ 57 0 ~ 212 0 ~ 101
When using
SM-S4L
and C-EP
extension pillar
SMZ800 + C-PS
Objectives
Acro 0.5X 0 ~ 2 — 0 ~ 58 —
Plan 1X 0 ~ 75 0 ~ 20 26 ~ 131 0 ~ 20
Plan 1X ERGO 0 ~ 92 0 ~ 37 3 ~ 148 0 ~ 37
Plan Apo 0.5X 0 ~ 3 — 0 ~ 59 —
Plan Apo 1X 0 ~ 81 0 ~ 26 32 ~ 137 0 ~ 25
ED Plan 1.5X 0 ~ 89 0 ~ 34 40 ~ 145 0 ~ 34
ED Plan 2X 0 ~ 87 0 ~ 32 38 ~ 143 0 ~ 31
HR Plan Apo 0.5X 0 ~ 11 — 0 ~ 67 —
HR Plan Apo 1X 0 ~ 82 0 ~ 27 33 ~ 138 0 ~ 27
HR Plan Apo 1.6X 0 ~ 90 0 ~ 35 41 ~ 146 0 ~ 35
Arm normal Arm lower When using
position position auxiliary adapter
When using
SM-S4L and
auxiliary adapter
38
Loading...