The Safety and Warranty Information section provides details about cautionary symbols used in the manual,
safety markings used on the instrument, and the information about the Warranty including Customer Service
contact information.
Safety Information and the Manual
Throughout this manual, the words Caution and Warning indicate potentially dangerous or hazardous situations
which, if not avoided, could result in death, serious or minor injury, or damage to the product. Specifically:
Indicates a potentially hazardous situation which can result in minor or moderate injury or damage to the
product or equipment.
Indicates a potentially dangerous situation which can result in serious injury or death.
Indicates visible and/or invisible laser radiation. Avoid direct exposure to the beam.
General Safety Considerations
If any of the following conditions exist, or are even suspected, do not use the instrument until safe operation can
be verified by trained service personnel:
Visible damage
Severe transport stress
Prolong storage under adverse conditions
Failure to perform intended measurements or functions
If necessary, return the instrument to ILX Lightwave, or authorized local ILX Lightwave distributor, for service or
repair to ensure that safety features are maintained (see the contact information on page 6).
All instruments returned to ILX Lightwave are required to have a Return Material Authorization (RMA) number
assigned by an official representative of ILX Lightwave Corporation. See “Returning an Instrument” on page 5 for
more information.
November 2017 5 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 8
Safety Symbols
This section describes the safety symbols and classifications.
Technical specifications including electrical ratings and weight are included within the manual. See the Table of Contents to locate the specifications and other product information. The following classifications are standard
across all ILX Lightwave products:
Indoor use only.
Ordinary protection: This product is NOT protected against the harmful ingress of moisture.
Class I Equipment (grounded type).
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
Pollution Degree II
Installation (overvoltage) Category II for transient overvoltages
Maximum Relative Humidity: <80% RH, non-condensing
Operating temperature range of 0°C to 40°C
Storage and transportation temperature of -40°C to 70°C
Maximum altitude of 3000m (9843 ft).
This equipment is suitable for continuous operation.
Safety Marking Symbols
This section provides a description of the safety marking symbols that appear on the instrument. These symbols
provide information about potentially dangerous situations which can result in death, injury, or damage to the
instrument and other components.
Caution:
Refer to
manual
Caution:
Risk of Electric
Shock
Earth Ground
Terminal
Protective
Conductor
Terminal
On: In position of push control.
The (I) denotes that the mains
are on.
Alternating
Current
Caution:
Hot Surface
Of: Out position of push control.
The (O) denotes that the mains
are on.
Visible and/or
invisible laser
radiation
Frame or
Chassis
Terminal
November 2017 6 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 9
Warranty
ILX Lightwave warrants this instrument to be free from defects in material and workmanship for a period of one
year from the date of shipment. During the warranty period, ILX Lightwave will repair or replace the unit, at our
option, without charge.
Limitations
This warranty does not apply to fuses, lamps, defects cause by abuse, modifications, or use of the product for
which it was not intended.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty of
merchantability or fitness for any particular purpose. ILX Lightwave shall not be liable for an incidental, special, or
consequential damages.
If a problem occurs, please contact ILX Lightwave with the instrument’s serial number, and thoroughly describe
the nature of the problem.
Returning an Instrument
If an instrument is to be shipped to ILX Lightwave for repair or service, be sure to:
1. Obtain a Return Material Authorization number (RMA) from ILX Lightwave Customer Service.
2. Attach a tag to the instrument identifying the owner and indicating the required service or repair. Include
the instrument serial number from the rear panel of the instrument.
3. Attach the anti-static protective caps that were shipped with the instrument and place the instrument in a
protective anti-static bag.
4. Place the instrument in the original packing container with at least 3 inches (7.5cm) of compressible
packaging material. Shipping damage is not covered by this warranty.
5. Secure the packing box with fiber reinforced strapping tape or metal bands.
6. Send the instrument, transportation pre-paid, to ILX Lightwave. Clearly write the Return Material
Authorization number on the outside of the box and on the shipping paperwork. ILX Lightwave
recommends the shipment be insured.
If the original shipping container is not available, place the instrument in a container with at least 3 inches
(7.5cm) of compressible packaging material on all sides.
Repairs are made and the instrument returned transportation pre-paid. Repairs are warranted for the
remainder of the original warranty or for 90 days, whichever is greater.
Claims for Shipping Damage
Upon receiving the instrument, inspect it immediately for any damage or shortages on the packing list. If the
instrument is damaged, file a claim with the carrier. The factory will supply a quotation for estimated costs of
repair. The customer is responsible for negotiation and settlement with the carrier for the amount of damage.
November 2017 7 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 10
Comments, Suggestions, and Problems
To ensure getting the most out of ILX Lightwave products, direct any product operation, service related
questions, or comments to ILX Lightwave Customer Support. Contact ILX Lightwave in whatever way is most
convenient:
ILX Lightwave
31950 Frontage Road
Bozeman, Montana, USA 59715
When contacting ILX Lightwave, please have the following information:
Model Number
Serial Number
End-User Name
Company
Phone Number
Email Address
Description of the problem
If ILX Lightwave determines that a return to the factory is necessary, a Return Material Authorization (RMA) will
be issued. Please mark this number on the outside of the shipping box.
The customer or carrier is responsible for any shipping damage when returning an instrument to ILX Lightwave.
ILX Lightwave recommends insuring the shipment. If the original shipping container is not available, place the
instrument in a container with at least 3 inches (7.5cm) of compressible packaging material on all sides.
November 2017 8 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 11
Chapter 1 :
Introduction and Specifications
This chapter is an introduction to the LDX-3232 High Compliance Current Source:
Setup Information
Safety Considerations
Maintenance Information
Instrument Specifications
If any of the following symptoms exist, or are even suspected, remove the LDX-3232 from service. Do not
use the LDX-3232 until trained service personnel can verify safe operation.
-
Visible damage
-
Severe transport stress
-
Prolonged storage under adverse conditions
-
Failure to perform intended measurements or functions
If necessary, return the LDX-3232 to ILX Lightwave for service and repair to ensure that safety features are
maintained.
Product Overview
The LDX-3232 instrument is a high compliance voltage laser diode current source. The LDX-3232 current
source provides a high stability output with multiple laser protection features and a fully redundant current
limit. The LDX-3232 offers the added benefits of dual laser interlocks and a GPIB interface to allow full
automation of testing.
Initial Inspection
Upon receiving the LDX-3232 instrument, verify that the following items were shipped with the instrument:
LDX-3232 High Compliance Current Source
USB Flash Drive containing LDX-3232 User Manual
Power Cord
Shipping Kit
November 2017 9 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 12
Installing the LDX-3232 High Compliance Current Source
Grounding Requirements
The LDX-3232 High Compliance Current Source comes with a three conductor AC power cable. The
power cable must either be plugged into an approved three-contact electrical outlet or used with a threecontact or two-contact adapter with the grounding wire connected to an electrical ground (safety ground).
The LDX-3232's power jack and supplied power cable meet IEC safety standards.
AC Line Power Requirements
The LDX-3232 instrument can be operated from a single phase power source delivering nominal line
voltages of 100, 120, or 220-240 VAC (all values RMS), at 50/60 Hz. The line power voltage can vary
±10% but cannot exceed 250 VAC. Maximum power consumption is 60 VA (Volt-Amps). Setting operating
voltage can only be done at the factory. Before operating the instrument, verify that the voltage indicated
on the back panel of the instrument matches the power-line voltage used in the operating environment.
Before connecting the LDX-3232 High Compliance Current Source to a power source, verify that the AC
power source matches the setting of the LDX-3232's voltage printed on the rear panel of the instrument.
To avoid electrical shock hazard, connect the instrument to properly earth-grounded, 3-prong receptacles
only. Failure to observe this precaution can result in severe injury or death.
GPIB Connector
The IEEE 488 GPIB interface connector is located on the rear panel, directly above the power input
module and fuse (See Figure 1.2). Attach the GPIB cable to the 24-pin connector located on the rear
panel. The connector is tapered to ensure proper orientation. The two screws on the cable connector need
only be finger tightened.
A total of 15 devices can be connected together on the same GPIB interface bus. The cables have single
male/female connectors on each end so that several cables can be stacked. This allows more than one
cable to be attached to any one device. However, the maximum length of the GPIB cables must not
exceed 20 meters (65 feet) total or 2 meters (6.5 feet) per device.
The GPIB Address
The talk and listen addresses on the LDX-3232 High Compliance Current Source are identical. This GPIB
address is read locally by pressing the (GPIB) LOCAL button with the address displayed on the LED
display. The instrument comes from the factory configured with the GPIB address set to 1. The LDX-3232's
GPIB address can be changed locally (via front panel). A procedure for changing the address can be found
in the section Changing the GPIB Address in Chapter 3.
Tilt-Foot Adjustment
The LDX-3232 High Compliance Current Source comes standard with folding front legs and two rear feet
for use as a bench top instrument. Extending the front feet so that the instrument front panel sits up makes
it easier to view the LED displays. To use them, place the unit on a stable base and rotate the front legs
downward until they lock into position.
November 2017 10 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 13
Operating the LDX-3232 Precision Current Source
The following photos show the functional keypad groupings, back panel connectors, and so on. Use these
figures to become familiar with the LDX-3232. After that, use Chapter 2 for fundamentals of instrument
operation.
Figure 1.1 LDX-3232 Front View
Figure 1.2 LDX-3232 Rear View
November 2017 11 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 14
Specifications
Table 1.1 Specifications
2000mA Range 4000mA Range
Output Current Range 0 to 2000 mA 0 to 4000 mA
Setpoint Resolution
Setpoint Accuracy ±0.15% of SP +2 mA
Compliance Voltage 0 - 15 V, adjustable
Temperature Coefficient <100 ppm/°C
Short Term Stability (1 hour)
Long Term Stability (24 hours)
Noise and Ripple (µA rms)
High Bandwidth Mode
Low Bandwidth Mode
Transients
Operational
1 kV EFT/Surge
Range 0 - 15 V
Resolution 60 mV
Accuracy ±2.5%
Range 1 to 2020 mA 1 to 4040 mA
Resolution 10 mA 20 mA
Accuracy ±20 mA ±40 mA
Type Differential
PD Reverse Bias 0 - 5V, adjustable
PD Current Range 5 to 10,000 µA
Output Stability
Accuracy, setpoint (% of FS) ±0.05%
Input
Transfer Function 200 mA / V 400 mA / V
Bandwidth (3 dB)
High Bandwidth Low Bandwidth
Drive Current Output
2
3
4
5
40 µA 80 µA
<20 ppm
<40 ppm
<20 µA
<20 µA
1
<40 µA
<20 µA
6
7
8
9
<4 mA
<8 mA
Compliance Voltage Adjust
Drive Current Limit Settings
Photodiode Feedback
±0.02%
External Analog Modulation
0 - 10 V, 1 k
DC to 250 kHz
DC to 10 kHz
November 2017 12 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 15
Trigger Output
Type TTL
Pulse Width 13 µs
Delay 12 ms
Measurement (Display)
Output Current Range 0 to 2000.0 mA 0 to 4000.0 mA
Output Current Resolution10
0.1 mA
Photodiode Current Range 0 to 10,000 µA
Photodiode Current Resolution 1 µA
Photodiode Current Accuracy ±4 µA
PD Responsivity Range11
0.00 to 1000.00 µA/mW
PD Responsivity Resolution 0.01 µA/mW
Optical Power Range 0.00 to 5000.0 mW
Optical Power Resolution 0.1 mW
Forward Voltage Range 0.000 to 10.000 V
Forward Voltage Resolution 1 mV
Forward Voltage Accuracy12
±2 mV
General
Size (H x W x D) 4” x 8.5” x 13.4”; 102 mm x 216 mm x 340 mm
Weight ~ 20 pounds (9 kg)
Power 100/120/230-240 V~; 2/2/1A 50/60 Hz
Temperature 0°C – 40°C operating; -40°C – 70°C storage
Humidity < 90% relative humidity, non-condensing
Laser Safety Features Key switch, interlock and output delay (meets CDRH US
21 CFR 1040.10)
Display Type 5-digit, green LED
Notes
1.
All values relate to a one-hour warm-up period at room temperature, 25°C.
2.
Based on resolution of digital-to-analog converters used in the design.
3.
Over any 1-hour period, half-scale output.
4.
Over any 24-hour period, half-scale output.
5.
Measured electrically with a 1 ohm load at half-scale output; evaluating spectral noise density over a 150kHz bandwidth from 125kHz to 275 kHz.
6.
Maximum output current transient from normal operational situations (e.g., power on-off, current on-off), as well as accidental situations (e.g., power line
plug removal).
7.
Maximum output current transient resulting from a 1000 V power-line transient spike. Tested to ILX Lightwave Technical Standard #LDX-00196.
8.
Maximum monitor photodiode current drift over any 30 minute period. Assumes zero drift in responsivity of photodiode.
9.
50% modulation at mid-scale output. Higher bandwidth is possible with smaller magnitude modulation signal.
10.
Similar resolution available over GPIB.
11.
Responsivity value is user-defined and is used to calculate the optical power.
12.
Four wire voltage measured at the load. Voltage measurement accuracy while driving calibration load. Accuracy is dependent upon load and cable used.
November 2017 13 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 16
Available Options and Accessories
Options and accessories available for the LDX-3232 High Compliance Current Source include the
following:
Table 1.2 Accessories List
DESCRIPTION MODEL NUMBER
Low Noise CW Filter, 9-pin DSUB LNF-320
Driver to Mount Cable, 5 Amp, DB9 Male to DB9 Male CC-305S
Driver to Unterminated Cable, 5 Amp, DB9 Male to Bare Wire CC-306S
Laser Diode Mount, 14-pin DIL LDM-4982
Laser Diode Mount, 14-pin Butterfly LDM-4894
Laser Diode Mount, Requires Mounting Kit, High Power LDM-4442
Single Rack Mount Kit, LDX-3200 Series & LDT-5900 Series RM-139
Dual Rack Mount Kit, LDX-3200 Series & LDT-5900 Series RM-140
Please contact ILX Lightwave or visit www.newport.com/ilxlightwave for information on additional options for
applications.
November 2017 14 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 17
Chapter 2 :
Operation
This chapter introduces the operation of the LDX-3232 High Compliance Current Source. It
offers instructions for connecting a laser diode to the current source and describes powering up
the instrument. This chapter also includes:
Operation Procedures
Interlock Information
Applying Power to the LDX-3232
To turn on the LDX-3232, rotate the key in the POWER section of the front
panel from O to I. This action will initiate the power on sequence. If the LDX3232 does not appear to turn on, verify that it is connected to line power. If
line power is not the problem, remove the power cord and check the voltage
selection block in addition to the line power fuse.
The Power On Sequence
During the power-up sequence, the following takes place:
For about three seconds all indicators are lit up; all of the 7-segment displays indicate 8.
All indicator lights are turned off for three seconds.
Firmware version will be displayed.
Instrument performs a communications self-test between the hardware and processor.
After this test, the instrument is ready to operate and is configured to the state it was in
when the power was last shut off.
Connecting to the Laser
When connecting a laser to the LDX-3232, it is recommended that the instrument be powered
up with the output off. In this condition, a low impedance shunt is active across the output
terminals. When disconnecting devices, it is only necessary to turn the current source output off.
It is also recommended that the connections to the LDX-3232 output be made using twisted wire
pairs with an earth-grounded shield (see Figures 2.1 - 2.4). The output terminals of the
instrument are left floating relative to earth ground to suppress AC power-on/power-off
transients that may occur through an earth- ground path. If the output circuit is earth-grounded
at some point (such as through the laser package and mount), the user must be careful to avoid
multiple earth grounds in the circuit. Multiple earth grounds may provide circuit paths that induce
spurious currents in the photodiode feedback circuit and output leads.
November 2017 15 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 18
Note: Cable connections to the laser must be secure. Loose connections can cause
momentary open circuits that can damage the laser.
Figures 2.1 – 2.4 show the possible configurations of connecting laser diodes and photodiodes
with the LDX-3232 High Compliance Laser Current Source.
Figure 2.1 Common Laser Cathode & Photodiode Cathode
Figure 2.2 Common Laser Cathode & Photodiode Anode
Figure 2.3 Common Laser Anode & Photodiode Cathode
November 2017 16 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 19
Figure 2.4Common Laser Anode & Photodiode Anode
The 9-pin connector marked CURRENT SOURCE on the back panel is used to connect the
laser diode to the LDX-3232. There are connections provided for laser cathode and anode,
photodiode cathode and anode, chassis ground and interlock. The pinout diagram for this
connector is shown in Figure 2.5.
Figure 2.5Back Panel LD Connector
Interlock Connections
In order for the laser output to be enabled a short circuit must exist between the Interlock Pins
(Pins 1 and 2) of the connector. The short can be a direct short across the pins or a switch to
prevent laser operation until the switch is closed. If a short does not exist between these two
pins, the INTERLOCK LED will illuminate on the front panel and the laser output will be
disabled.
Note: The BNC Interlock Connection from the BNC connector on the rear panel will also
make the INTERLOCK LED illuminate. Please see the notes in chapter 5 on use of the
BNC interlock connection to learn more about its operation.
The interlock terminals on the LASER connector, pins 1 and 2, must be kept isolated
from all other connections including earth ground.
Photodiode Connections
Many laser diode modules contain an internal photodiode that monitors the back- facet emission
of the laser. Usually, this photodiode is internally connected to either the laser anode or
cathode.
November 2017 17 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 20
The photodiode and laser connections to the LDX-3232 are electrically isolated from ground and
each other. If a 4 pin connection is made (no common connections), no additional jumpers are
required. Figures 2.1 - 2.4 show the recommended connections and shielding for 3-pin lasers
(where the common connection is internal to the device). A 4-pin laser should be connected with
the same shielding as shown in Figure 2.1, but the common connection (between the
photodiode and the laser) is optional.
Setting the PD Bias
The LDX-3232 provides an adjustable reverse bias of 0-5 VDC for the photodiode. To set the
photodiode bias to 5 V reverse bias, turn the back panel PD BIAS ADJUST fully clockwise. To
set the photodiode bias to 0 V reverse bias, turn the back panel PD BIAS ADJUST fully counterclockwise.
Grounding Considerations
The LASER outputs of the LDX-3232 High Compliance Current Source are isolated from
chassis ground, allowing either output terminal to be grounded at the user's option. Figures 2.1 -
2.4 show the proper earth-ground shielding for laser diode/photodiode connections.
Front Panel Operation
Figure 2.6LDX-3232 Front Panel
This section describes fundamentals of operation for the LDX-3232 in the two operating modes:
Constant Current (I) and Constant Power (P).
The Display
The digital display is used to show control parameters such as laser drive current (mA), laser
current limit (mA), laser forward voltage (V), voltage limit (V), monitor photodiode current (µA),
and laser optical power (mW) in both set point and measurement modes. It will also display
error codes that relate to LASER operation. Error indicators underneath the display indicate
control errors and are explained in the section titled “ERROR INDICATORS”. All laser control
parameters are displayed on this digital display with the appropriate annunciator illuminated.
November 2017 18 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 21
Setting Up the LDX-3232 Precision Current Source
The following section describes the functionality of the front panel.
Mode Selection
The LDX-3232 has three modes in which it can be driven:
Constant Current – Low Bandwidth
Constant Current – High Bandwidth
Constant Power
Pressing the (MODE) SELECT button cycles between these three modes.
Pressing the (MODE) OUTPUT button will enable or disable the output of
instrument in the mode which is selected.
Range Selection
The LDX-3232 has two ranges in which it can be driven:
2000mA
4000mA
Pressing the button in the RANGE section twice will cycle between the
modes. The button presses must be quick (less than 1 second) to switch
between ranges. The LASER output must be OFF to change ranges.
Display Selection
The LDX-3232 can be configured to display the following properties:
Laser Diode Current (mA)
Laser Diode Power (mW)
Laser Diode Voltage (V)
Photodiode Current (µA)
Pressing the SELECT button in this section will cycle between the
available display properties and the indicator will be lit.
When set point mode is active, the corresponding indicator which is being adjusted will flash to
indicate that the unit is in set point mode rather than measurement display.
To only view the set point, press and hold down the display SELECT button for two or more
sections. When the (DISPLAY) SELECT button is released, the previous measurement mode
will be restored.
Adjust Knob
All set points are adjusted using the adjustment knob when operating
from the front panel of the LDX-3232. When turning the adjustment
knob, the instrument automatically changes to set point mode, as
shown by the flashing indicator.
After an adjustment has been made, the set point mode will time out
after three seconds and will return to the original measurement display mode.
The FINE button is used to make fine control adjustments. When the indicator is not lit, the
instrument will make coarse control adjustments.
November 2017 19 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 22
Parameter Selection
The LDX-3232 has multiple parameters which can be adjusted:
Current Limit (LIM I)
Power Limit (LIM P)
Photodiode Responsivity in µA/mW (CAL PD)
Compliance Voltage Limit (COMP V)
Saving Parameter Setups to a Bin (SAVE)
Recalling Parameter Setups from a Bin (RECALL)
To adjust these values, press the (PARAMETER) SELECT button until the indicator next to the
parameter which is to be adjusted is lit, then press the (PARAMETER) SET button. Adjust the
value using the ADJUST KNOB. After three seconds, the new parameter value is stored in nonvolatile memory.
Using the LDX-3232 in Constant Power (P) Mode
Before using the LDX-3232 in Constant Power (P) Mode, the photodiode responsivity (CAL PD),
which is used to convert between photodiode current and optical power of the laser diode, must
be known.
A method to calculate the responsivity is suggested below.
1. Measure (with a calibrated detector) the output power of the device.
2. Measure the corresponding photodiode current.
3. Calculate the responsivity by dividing the photodiode current by the optical power noting
the units required are µA/mW.
4. Input this value using the CAL PD Parameter Selection as described above.
Note: The LDX-3232 High Compliance Current Source can be put into a special mode
with the CAL PD parameter set to zero. When the CAL PD value is zero, the LASER
output will be controlled to the IPD set point value. Therefore, the instrument is in
Constant IPD mode. The display will indicate -.- for the power measurement.
Error Indicators
The LDX-3232 High Compliance Current Sources indicate general operational error conditions.
Each error condition results in an action as shown in the table below.
Table 2.1 LDX-3232 Error Indicators
Error Condition Action
Interlock Output off; INTERLOCK light flashes at 1 Hz
Open Circuit
Current Limit CURRENT LIMIT light flashes at 1 Hz
Voltage Limit
Optical Output
Power Limit
Output off; OPEN CIRCUIT light flashes at 1 Hz; if this condition was caused by
an excessively high compliance voltage, the VOLTAGE LIMIT light will also be
turned on.
VOLTAGE LIMIT light flashes at 1 Hz; this indicator warns that the output is within
0.25 V of LIM V setting; if the voltage limit (LIM V) is reached, the LASER output
will shut off.
Output off; POWER LIMIT light flashes at 1 Hz
November 2017 20 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 23
Automatic Laser Output Shut Off Conditions
These conditions will automatically cause the instrument to disable the current source output:
LASER High Power Limit
LASER Interlock / Key Lock Turned Off
LASER Open Circuit or Voltage Limit
November 2017 21 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 24
November 2017 22 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 25
Chapter 3 :
Remote Operation
This chapter details the fundamentals of operating the LDX-3232 High Compliance Current
Source remotely through the GPIB interface.
Current Source Command Set
Command Paths
Reading the GPIB Address
Before operating the LDX-3232 instrument from a remote location, its GPIB address must be known.
Simply press the LOCAL button in the section labeled GPIB on the front panel. The instrument will
display the address. The default address from the factory is address 01.
Changing the GPIB Address
Every device on the GPIB bus must have a unique address. If it is necessary to change the
address, press and hold in the (PARAMETER) SET button while displaying the GPIB address.
Turn the ADJUST KNOB until the desired address value is displayed, then release the SET
button. The new GPIB address will then be stored in non-volatile memory. The allowable
address range is 0-30 for primary GPIB addressing. Extended GPIB addressing is not
implemented on the LDX-3232 at this time.
Changing Operation from Local to Remote
Sending a command over the GPIB automatically puts the instrument in
REMOTE mode. The REMOTE indicator will identify when the Current Source
is in remote operation mode. When the instrument is in REMOTE mode,
pressing the LOCAL button returns the instrument to LOCAL control mode.
The TALK/LISTEN indicator is illuminated when the instrument is
communicating over the GPIB bus.
November 2017 23 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 26
LDX-3232 Current Source Command Set
This section contains information regarding GPIB-IEEE488 Command Commands, most used
device specific commands, and command syntax. Error and status registers are also introduced
in this chapter.
Command Syntax
Any command or command query must contain all of the letters that are shown in the LDX-3232
Command Path Structure, Figure 3.1. However, the Current Source's command parser is NOT
case sensitive, so upper or lower case may be used in any combination. The lower case letters
shown with the Device Dependent commands in Chapter 4 are optional letters and may be used
for clarity. For example, the following commands are equal:
LAS:DIS 1
LAS:Dis 1
LAS:DISPLAY 1
In this example, only the first three letters “DIS” are required, while the other letters, “play” are
optional. These optional letters must be used in the proper sequence as shown.
The syntax of the LDX-3232 High Compliance Current Source commands follows the rules laid
out in the GPIB-IEEE488 standard. Colons (:) indicate the start of a new command path, while
semicolons (;) indicate a separation of commands within a command string. A leading colon on
a command may be used to return the LDX-3232 High Compliance Current Source command
parser to the command path root (see Figure 3.1).
Spaces or white space may be used to separate commands and/or data (after the command
header or query question mark). Spaces or white space must be used to separate the command
(header) from the first parameter (or program data). The following examples show valid syntax
for commands with the LDX-3232 High Compliance Current Source:
The following are examples of invalid syntax for the LDX-3232 High Compliance Current
Source. These command strings would produce an erroneous result, as explained:
LAS:MODE MDP Missing colon, MODE? expected
LAS:MODE:ILBW DEC Missing semicolon, DEC command generates an error
LAS:DIS ? Space not allowed before question mark, DIS command expected
Las:LDI33;dis? Space missing between LDI command and the parameter value, 33
November 2017 24 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 27
For further clarity in programming, the (Boolean) values of one (1) and zero (0) may be used or
their names as indicated in the following table:
Table 3.1Substitute Parameter Names
Substitute Name Value
ON
OFF
OLD
NEW
TRUE
FALSE
SET
RESET
1
0
1
0
1
0
1
0
If multiple parameters are expected, they should be separated with commas:
LAS:INC 100,50
A query has no space between the mnemonic and the question mark:
LAS:LDI?
The LDX-3232 High Compliance Current Source uses a terminator of <NL><^END> (new line
with EOI). For users whose GPIB driver defaults expect a carriage return in the terminator,
<CR><NL><^END>, the TERM command may be used for convenience (see TERM command
in Chapter 4). For more information, refer to the GPIB driver’s configuration manual.
The command structure is illustrated in Figure 3.1. Table 4.1 lists all of the LDX-3232 High
Compliance Current Source’s device-dependent commands, with the full path shown for each
command and a brief explanation of its usage.
November 2017 25 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 28
Command Paths
The LDX-3232 High Compliance Current Source device-dependent commands are structured
into a tree format (see Figure 3.1). Each of the legal paths is listed below, followed by its list of
path options, each of that is followed by the commands themselves. It is recommended that the
first-time user begin learning the commands by using the full path notation. Once familiar with
the commands, taking advantage of the shortcuts allowed for command paths may be useful.
Figure 3.1LDX-3232 Command Path Structure
November 2017 26 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 29
IEEE488.2 Command Commands
The IEEE488.2 Common Commands and Queries are distinguished by the * that begins each
mnemonic. The diagrams below show the syntax structure for common commands, common
commands with numeric data required, and command queries.
Figure 3.2Common Command Structure
Numeric data is required with:
*PSC (1=on, 0=off)
*RCL (0 to 10, see front panel RECALL function)
*SAV (1 to 10, see front panel SAVE function)
*ESE (0 to 255, see Figure 3.3)
*SRE (0 to 255, see Figure 3.3)
*PUD (used at factory only)
A list of all of the GPIB-IEEE488 Common Commands supported by the LDX-3232 follows:
For more information on these commands, refer to an ANSI/IEEE 488-1987 standards
reference.
Note: Care should be taken to set the GPIB time-out appropriately for use with the *WAI,
DELAY, or *OPC? commands. If the time for a response after one of these commands
exceed the GPIB time-out period, a bus (timeout) error will occur. Usually, after this
timeout error, the LDX-3232 will generate a query error (E302). This error code is
reported via the ERR? query.
November 2017 27 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 30
Status Reporting
This section contains information that is relevant for understanding instrument error and status
reporting. It also contains information regarding the use of the instrument status for generating
interrupts for interrupt driven programs or subroutines. Understanding the Operation Complete
definition for the instrument is useful for programming synchronization. The Output Off Register
section also contains information on setting some of the conditions that will force the laser
current source output off.
The following sections describe the Event and Condition registers, Operation Complete Flag,
Output Off registers, and Error Messages.
Event and Condition Registers
In addition to the required GPIB-IEEE488 status reporting structures, the LDX-3232 High
Compliance Current Source remote interface provides Event and Condition Registers for
LASER operations. The Event Registers are used to report events that occur during the
operation of the LDX-3232 High Compliance Current Source. Events differ from conditions in
that events signal an occurrence once, and are not reset until the Event Register is queried, the
status cleared by the *CLS command, or the instrument is powered off. Conditions reflect the
current state of the device and therefore may change many times during operation. Querying a
Condition Register does not change its contents.
Figure 3.3 shows the status reporting scheme of the LDX-3232 High Compliance Current
Source. Each of the registers that may be accessed by a command or query has the
appropriate command or query written above or below the register representation. For example,
the LASER Condition Register may be queried via the LASer:COND? query.
The condition or event registers are logically ANDed with their respective enable registers.
These bits are then logically ORed to form a summary message in the status byte for that
particular register.
November 2017 28 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 31
Operation Complete Definition
Note that Bit 0 of the Standard Event Status Register contains the status of the Operation
Complete flag. Enabling this bit via the *ESE command allows the user to update bit 5 of the
status byte. Then, if the SRE mask has bit 5 set, and the user issues an *OPC command, the
SRQ (bus) signal will be generated upon completion of the currently processed commands. This
bus signal (SRQ) may be used to initiate service request routines, which depend on the
completion of all previous commands.
For example, the user may turn the current source on (LAS:OUT 1), enable an SRQ on
Operation Complete (set *ESE 1 and *SRE 5), and have an SRQ handling routine in the user's
software, that reads a new measurement after the output on state has been reached. This
allows the use of the operation complete features of the LDX-3232, without the need for
program looping or polling that can tie up the GPIB.
Operation Complete on the LDX-3232 is defined as:
1. No operations to the LASER current source hardware are pending.
2. No EEPROM (non-volatile) memory write cycles are in progress.
Figure 3.3 LDX-3232 Status Reporting Schematic Diagram
Note: If the GPIB times out while waiting for a response, either set the GPIB time-our longer or use SRQ
generated interrupts in the program. See the GPIB manual for time-out configuration or SRQ programming
setup.
The *OPC, *OPC? and *WAI commands should not be used inside a calibration routine.
November 2017 29 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 32
Output Off Register
The Output Off Enable Register allows determination of conditions and events in the LDX-3232
High Compliance Current Source can cause the current source output to be turned off. This
reigster is configured in a manner similar to the status reporting registers. However, the register
output is not reported in the Status Byte Register. Rather, it sets the hardware that controls the
output switching. The events and conditions that may be set to cause the current source output
to be turned off are shown in Figure 3.4.
The default (factory) settings for this register are shown in the table below. These settings are
not affected by the *PSC (Power-On Status Clear) command.
Figure 3.4 LDX-3232 Output Off Register & Defaults
Command Timing and Completion
This section describes whether a command is performed in an overlapped or sequential
manner. In other words, it states whether the next command may begin while the current
command is being executed, or if the next command must wait until the current command is
completed before its execution begins. The conditions for setting the operation complete flag
are given in the Chapter 3 section titled Operation Complete Definition.
All LDX-3232 device-dependent commands are executed in an overlapped manner, except the
DELAY command that is sequential. The operation complete flag is set after the conditions
outlined in the Operation Complete Definition have been satisfied.
The *WAI command (common command) is an example of a sequential command that forces
the next command to wait until the no-operation-pending flag is true. This is essentially the
same as waiting for the OPC flag to become true, because the no-operations-pending flag is
used to set the OPC flag (bit 0 of the Standard Event Status Register).
Commands that change the status of the instrument limits or change its mode or current range,
step value, or status enable registers will not have their OPC flag set until all present writing to
non-volatile memory has been completed. This is done to ensure that the OPC flag is never set
prematurely.
November 2017 30 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 33
Whenever there is any output (response) data in the Output Queue, bit 4 is set in the Status
Byte Register. Whenever there is any error message in the Error Queue, bit 7 is set in the
Status Byte Register.
Error Messages
In the event of a hardware error condition, error messages will be displayed on the display. In
most cases, the error message will appear for three seconds. In some cases the error code
display will remain until the user changes display modes. In the case of multiple error
messages, the display may sequentially show each message for three seconds. In addition to
the hardware errors, GPIB errors may be read via the ERR? query. Table 3.2 lists the numerical
error ranges by function. Table 3.3 contains all of the error messages that may be generated.
Not all of these messages may appear on the front panel displays. Some refer to GPIB activities
only.
In remote operation, the errors can be read by issuing the ERR? query. When this is done, all of
the error messages that are resident in the error queue are returned (up to 10 may be stored).
Reading the error queue via GPIB clears the error queue.
Table 3.2 Error Code Classifications
Error Code Range Area of Operation
E-001 to E-099 Internal Program Errors
E-100 to E-199 Parser Errors
E-200 to E-299 Execution Control Errors
E-300 to E-399 GPIB Errors
E-500 to E-599 LASER Control Errors
Table 3.3 LDX-3232 Error Codes
Error Code Explanation
E-001 Memory allocation failure
E-103 <DEFINITE LENGTH ARBITRARY BLOCK PROGRAM DATA> length too long
E-104 <NON-DECIMAL NUMERIC PROGRAM DATA> type not defined
E-105 <DECIMAL PROGRAM DATA> exponent not valid
E-106 <DECIMAL PROGRAM DATA> digit expected
E-123 Command not found
E-124 <program mnemonic> Lookup, failed because query/command type match failed
E-126 Too few or too many program data elements
E-201 <PROGRAM DATA> value out of range
E-202 <PROGRAM DATA> will not convert to valid type
E-203 Security violation, command is not available without clearance
E-205 <PROGRAM DATA> is not a Boolean value or word
E-207 <PROGRAM DATA> will not convert to an unsigned 16-bit value
E-209 <PROGRAM DATA> will not convert to an unsigned 32-bit value
November 2017 31 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 34
E-210 <PROGRAM DATA> will not convert to a floating point value
E-211 <PROGRAM DATA> will not convert to a character value
E-213 <PROGRAM DATA> is incorrect block data length
E-214 <PROGRAM DATA> length exceeds maximum
E-302
Query error; device was addressed to talk, but GPIB controller failed to read all of the
<RESPONSE MESSAGE>
Error Code Explanation
E-501 LASER interlock disabled output
E-503 LASER open circuit disabled output
E-504 LASER current limit disabled output
E-505 LASER voltage limit disabled output
E-507 LASER power limit disabled output
E-510 LASER out of tolerance disabled output
E-511 LASER control error disabled output
E-512 Analog section status is all 1’s or all 0’s (power down)
E-513 Serial EEPROM checksum error
E-515 Laser output must be off to change ranges
E-516 Incorrect configuration for calibration sequence to start
E-519
Setting a measurement is only valid during the calibration phase for that measurement;
user has tried to calibrate a measurement without first entering the required calibration
mode.
November 2017 32 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 35
Chapter 4 :
Command Reference
This chapter is a guide to all the device dependent commands for the LDX-3232 High
Compliance Current Source.
Overview of Remote Commands
List of Commands in Alphabetical Order
Figure 4.1 shows the format for the device command descriptions in this chapter. The
commands that emulate local (front panel) operation are denoted by the solid box next to the
Local label in the upper left corner of the command description.
NONE Returns the status of the laser display for monitor photodiode power
measurement.
NONE Sets the status of the laser display to display parameters.
NONE Sets the display to show the setpoint for the presently selected display
mode.
NONE Returns the status of the setpoint mode display.
1
Sets the value of the condition status enable register.
NONE Returns the value of the condition status enable register.
1
Sets the value of the status event enable register.
NONE Returns the value of the status event enable register.
1
Sets the value of the status outoff enable register.
NONE Returns the value of the status outoff enable register.
NONE Returns the value of the status event register.
1
Sets the output current.
NONE Returns the output current.
0, 1, or 2 Increments the selected laser control mode set point by a number of
steps.
1
Inputs a voltage value for voltage calibration.
NONE Returns the value of the measured voltage.
1
Sets the current limit value for the 2000mA range.
NONE Returns the current limit value for the 2000mA range.
1
Sets the current limit value for 4000mA range.
NONE Returns the current limit value for 4000mA range.
1
Sets the monitor photodiode power limit value.
NONE Returns the monitor photodiode power limit value.
1
Sets the compliance voltage limit value.
NONE Returns the compliance voltage limit value.
1
Sets the value of the photodiode current measurement when CALPD
= 0.
NONE Returns the value of the photodiode current measurement when
CALPD = 0.
1
Sets the status of the front panel modulation input.
NONE Returns the status of the front panel modulation input.
1
Sets the value of the photodiode power measurement when CALPD >
0.
NONE Returns the value of the photodiode power measurement when
CALPD > 0.
NONE Returns the instrument output mode.
NONE Sets the instrument to high bandwidth constant current output mode.
NONE Sets the instrument to low bandwidth constant current output mode.
NONE Sets the instrument to constant power output mode.
1
Sets the status of the laser output.
NONE Returns the status of the laser output.
1
Sets the output current range.
NONE Returns the output current range.
November 2017 34 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 37
LAS:SET:LDI?
LAS:SET:MDI?
LAS:SET:MDP?
LAS:STEP
LAS:STEP?
LAS:TOL
LAS:TOL?
MES
MES?
RAD
RAD?
TERM
TERM?
TIME?
TIMER?
NONE Returns the output current setpoint value.
NONE Returns the monitor photodiode current setpoint.
NONE Returns the monitor photodiode power setpoint.
1
Sets the step size value.
NONE Returns the step size value.
2
Sets the current tolerance and time window.
NONE Returns the current tolerance and time window.
1
Sets a stored string.
NONE Returns a stored string.
1
Sets the radix type for status, condition, and event query response
data.
NONE Returns the radix type for status, condition, and event query response
data.
1
Sets the default terminator to include a carriage return.
NONE Returns the default terminator to include a carriage return.
NONE Returns the duration of time since the instrument was powered up.
NONE Returns the duration of time since the last TIMER? query was issued.
November 2017 35 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 38
LDX-3232 Device Dependent Command R eference
The following pages contain a reference for the device-dependent commands of the LDX-3232
High Compliance Current Source. This reference contains useful information for both local and
remote operation of the LDX-3232.
In some reference, parentheses are used to signify the labeled are for a particular button or LED
indicator on the front panel. For example, (PARAMETER) SET refers to the button labeled
“SET” in the PARAMETER area of the front panel.
DELAY <nrf> FRONT PANELREMOTE
The DELAY command causes the execution of commands to be delayed by a user-defined
time. This command is similar to the *WAI common command, except that execution resumes
after the specified number of milliseconds instead of waiting for the Operation-Complete flag to
be set.
Description Sets a single command execution delay.
Parameters An <nrf value> that represents the delay time in milliseconds.
Notes The Operation-Complete flag is held false until the delay period elapses,
but the *OPC? Query will not execute until the delay period has elapsed.
This command is useful for creating delays that don’t require a lot of
program code and don’t tie up the GPIB during execution. Care should be
taken to set the GPIB time-out appropriately for use with this command.
The delay time is approximate, with an error of about +10%.
Examples Write: DELAY 500
Action: The next command or query is not executed until ~0.5
seconds have elapsed from the time this command is
executed.
ERRors?FRONT PANELREMOTE
The ERRors? query returns a list of command and device errors that have occurred since the
last query. These errors are notated by a number that corresponds to the type of error that
occurred. See Chapter 3 for information regarding error handling.
Description Returns a list of errors that have occurred since the last query.
Parameters None.
Examples Query: ERR?
Response: 0
No errors reported.
Query: ERR?
Response: 201
Out of range error was reported.
November 2017 36 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 39
LASer:CALMD <nrf> FRONT PANEL REMOTE
LASer:CALMD?
Description Sets/returns photodiode feedback responsivity.
Parameters An <nrf> value between 0 and 1000 in µA/mW.
Notes If the parameter is set to 0, the LDX-3232 will operate in a constant IPD
mode, when Constant P (MDP) mode is selected.
Examples Write: LAS:CALMD 1
Action: Sets the CAL PD parameter to 1 µA/mW.
Query: LAS:CALMD?
Response: 2.5
Photodiode feedback responsivity is 2.5µA/mW.
LASer:CAL:LDI FRONT PANEL REMOTE
LASer:CAL:LDI?
Description Initiates laser current calibration.
Queries if the instrument is ready for a current calibration value to be
Action: Enters the LASER photodiode current calibration mode.
Query: LAS:CAL:MDI?
Response: 1
Instrument is ready for user to enter calibration value.
LASer:COND? FRONT PANELREMOTE
Description Returns the value of the status condition register.
Notes Response is the sum of the following:
1
LASER Current Limit 256 Output is shorted
2
LASER Voltage Limit 512 Output is outside tolerance limit
4
N/A 1024 Output on/off state
8
Power limit 2048 Ready for calibration data state
16 Interlock disabled 4096 N/A
32 N/A 8192 N/A
64 Compliance Voltage Limit 16384 N/A
128 Open circuit 32768 N/A
The LASER condition status is constantly changing, while the event
status is only cleared when the event status is read or the *CLS
command is issued.
Examples Query: LAS:COND?
Response: 513
LASER limit current and out of tolerance LASER conditions
currently exist.
LASer:DECFRONT PANELREMOTE
Description Decrements the selected laser control mode set point by one step.
Parameters None.
Notes The minimum time to complete one decrement is about 20 ms.
Examples Write: LAS:DEC
Action: The laser source current set point is decremented.
LASer:DISplay <bool> FRONT PANELREMOTE
Description Enables or disables the display and some LEDS.
Parameters A <bool value>; - where 1 = on, 0 = off.
Examples Write: LAS:DIS 1
Action: Turns the LASER display on.
November 2017 38 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 41
LASer:DISplay? FRONT PANELREMOTE
Description Returns the value shown on the display.
Parameters None.
Notes Returns the actual (6-character) string from the output buffer to the
display. If the display is disabled, it returns …
Examples Query: LAS:DIS?
Response: -99.9
LASER display shows -99.9
LASer:DISplay:LDI FRONT PANEL REMOTE
LASer:DISplay:LDI?
Description Sets/returns the status of the laser display for constant current
measurement.
Parameters None.
Notes The LDI measurement display mode is turned off automatically when
another display selection is enabled.
Examples Write: LAS:DIS:LDI
Action: Enables the display for current measurement.
Query: LAS:DIS:LDI?
Response: 0
LDI measurement is not displayed.
LASer:DISplay:LDV FRONT PANEL REMOTE
LASer:DISplay:LDV?
Description Sets/returns the status of the laser display for forward voltage
measurement.
Parameters None.
Notes The LDV measurement mode is turned off automatically when another
display selection is enabled.
Examples Write: LAS:DIS:LDV
Action: Enables the display for LDV measurement.
Query: LAS:DIS:LDV?
Response: 1
LDV measurement is displayed.
November 2017 39 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 42
LASer:DISplay:MDI FRONT PANEL REMOTE
LASer:DISplay:MDI?
LASer:DISplay:MDP
LASer:DISplay:MDP?
Description Sets/returns the status of the laser display for monitor photodiode
current/power measurement.
Parameters None.
Notes The MDI/MDP measurement display is turned off automatically when
another display selection is enabled.
Examples Write: LAS:DIS:MDI
Action: Enables the display for photodiode current measurement.
Query: LAS:DIS:MDP?
Response: 1
Monitor photodiode power measurement is displayed.
LASer:DISplay:PARAM FRONT PANEL REMOTE
Description Sets the status of the laser display to display parameters.
Parameters None.
Notes This command has the same effect as pressing the (PARAMETER)
SELECT button while in LASER mode.
Each time the command is issued, the next LASER parameter will be
selected and displayed.
Examples Write: LAS:DIS:PARAM
Action: Displays the next parameter value.
LASer:DISplay:SET FRONT PANEL REMOTE
Description Sets the display to show the setpoint for the presently selected display
mode.
Parameters None.
Notes Using this command has the same effect as physically holding the
(DISPLAY) button down continuously.
Examples Write: LAS:DIS:SET
Action: Enables the display for the set point of the selected mode.
LASer:DISplay:SET? FRONT PANELREMOTE
Description Returns the status of the setpoint mode display.
Parameters None.
Notes The set point display will not time out when REMOTE operation is used.
(It will be continuously displayed.)
Examples Query: LAS:DISPLAY:SET?
Response: 0
The set point is not shown on the display.
November 2017 40 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 43
LASer:ENABle:COND <nrf> FRONT PANELREMOTE
LASer:ENABle:COND?
Description Sets/returns the value of the condition status enable register.
Parameters An <nrf value> whose sum/response represents the sum of the enabled
bits:
1
LASER Current Limit 256 Output is shorted
2
LASER Voltage Limit 512 Output is outside tolerance limit
4
N/A 1024 Output on/off state
8
Power limit 2048 Ready for calibration data state
16 Interlock disabled 4096 N/A
32 N/A 8192 N/A
64 Compliance Voltage Limit 16384 N/A
128 Open circuit 32768 N/A
Examples Write: LAS:ENAB:COND 129
Action: Enables the open circuit and current limit conditions.
Query: LAS:ENAB:COND?
Response: 17
Current limit and interlock disable conditions are enabled.
LASer:ENABle:EVEnt <nrf> FRONT PANELREMOTE
LASer:ENABle:EVEnt?
Description Sets/returns the value of the status event enable register.
Parameters An <nrf value> whose sum represents the bits that are enabled:
1
LASER Current Limit 256 Output is shorted
2
LASER Voltage Limit 512 Output changed to be in/out of tolerance
4
N/A 1024 Output on/off state changed
8
Power limit 2048 New measurements taken
16 Interlock changed state 4096 N/A
32 N/A 8192 N/A
64 Compliance Voltage Limit 16384 N/A
128 Open circuit 32768 N/A
Examples Write: LAS:ENAB:EVENT 136
Action: Enables the open circuit and power limit events to be
reported.
Query: LAS:ENAB:EVE?
Response: 1040
Output on/off state and interlock state changed events were
reported.
November 2017 41 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 44
LASer:ENABle:OUTOFF <nrf> FRONT PANELREMOTE
LASer:ENABle:OUTOFF?
Description Sets/returns the value of the status outoff enable register.
Parameters An <nrf value> whose sum represents the enabled bits:
1
LASER Current Limit 256 N/A
2
LASER Voltage Limit 512 Output is out of tolerance*
4
N/A 1024
8
Power limit (with output on) 2048
16 N/A 4096 N/A
32 N/A 8192 N/A
64 N/A 16384 N/A
128 N/A 32768 N/A
N/A
N/A
Examples Write: LAS:ENAB:OUTOFF 9
Action: Enables the power limit and current limit conditions to turn off
output if the conditions occur.
Query: LAS:ENAB:OUTOFF?
Response: 1
Current limit condition will turn off the output if it occurs.
LASer:EVENT? FRONT PANELREMOTE
Description Returns the value of the status event register.
Parameters None.
Notes An <nrf value> whose sum represents the following:
Examples Query: LAS:EVE?
Response: 513
1
LASER Current Limit 256 Output is shorted
2
LASER Voltage Limit 512 Output changed to be in / out of tolerance
4
N/A 1024
8
Power limit 2048
16 Interlock disabled 4096 N/A
32 N/A 8192 N/A
64 Compliance Voltage Limit 16384 N/A
128 Open Circuit 32768 N/A
Output on / off state changed
Measurement ready
Output tolerance changed and current limit events have
occurred.
November 2017 42 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 45
LASer:INC <nrf,nrf> FRONT PANELREMOTE
Description Increments the selected laser control mode set point by a number of
steps.
Parameters None, 1, or 2.
First optional parameter is the number of steps.
Second optional parameter is the time in milliseconds between steps.
The minimum time to complete one decrement is about 20 ms.
Examples Write: LAS:INC
Action: The laser source current set point is incremented one step.
Write: LAS:INC 0
Action: Does nothing.
Write: LAS:INC 3,5000
Action: The laser source current set point is incremented by 0.3
mA with 5 seconds between steps.
LASer:LDI <nrf> FRONT PANEL REMOTE
Description Sets/returns the output current.
Parameters An <nrf value> in mA.
Examples Write: LAS:LDI 400
Action: Sets the laser output current to 400mA.
Query: LAS:LDI?
Response: 100
Laser output current is 100mA.
LASer:LDV <nrf> FRONT PANELREMOTE
Description Inputs a voltage value for voltage calibration.
Parameters An <nrf value> in volts.
Examples Write: LAS:LDV 4
Action: Enters a value of 4.000 V.
LASer:LDV? FRONT PANELREMOTE
Description Returns the value of the measured voltage.
Parameters None.
Notes This measurement is updated approximately once every 600ms.
Examples Query: LAS:LDV?
Response: 3.03
Measured output voltage is 3.03 V.
November 2017 43 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 46
LASer:LIMit:I2 <nrf> FRONT PANEL REMOTE
LASer:LIMit:I2?
Description Sets/returns the current limit value for the 2000mA range.
Parameters An <nrf value> in mA.
Notes The current limit is in effect in all modes of operation (ILBW, IHBW, or P).
The current limit in use depends on the LASER output range selection.
Examples Write: LAS:LIM:I2 180
Action: Current limit is set to 180mA.
Query: LAS:LIM:I2?
Response: 50
Current limit is set to 50mA.
LASer:LIMit:I4 <nrf> FRONT PANEL REMOTE
LASer:LIMit:I4?
Description Sets/returns the current limit value for the 4000mA range.
Parameters An <nrf value> in mA.
Notes The current limit is in effect in all modes of operation (ILBW, IHBW, or P).
The current limit in use depends on the LASER output range selection.
Examples Write: LAS:LIM:I4 1800
Action: Current limit is set to 1800mA.
Query: LAS:LIM:I4?
Response: 1500
Current limit is set to 1500mA.
LASer:LIMit:MDP <nrf> FRONT PANEL REMOTE
LASer:LIMit:MDP?
Description Sets/returns the monitor photodiode power limit value.
Parameters An <nrf value> in mW.
Notes When constant MDP mode is used, the output level is limited only by the
LIM Ix value. Exceeding the power limit will either cause the output to
shut off (default) or generate a warning.
Examples Write: LAS:LIM:MDP 10
Action: Monitor photodiode power limit is set to 10mW.
Query: LAS:LIM:MDP?
Response: 5
Monitor photodiode power limit is set to 5mW.
November 2017 44 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 47
LASer:LIMit:V <nrf> FRONT PANEL REMOTE
LASer:LIMit:V?
Description: Sets/returns the compliance voltage limit value.
Parameters An <nrf value> in volts.
Examples Write: LAS:LIM:V 5.0
Action: Compliance voltage limit is set to 5.0V.
Query: LAS:LIM:V?
Response: 4.5
Compliance voltage limit is set to 4.5V.
LASer:MDI <nrf> FRONT PANEL REMOTE
LASer:MDI?
Description Sets/returns the value of the photodiode current measurement when
CALPD = 0.
Parameters An <nrf value> in µA.
Notes If the CALMD (CAL PD) parameter is not set to 0, the LAS:MDI value will
not be used. In this case, the measured MDI would be converted to MDP
(PPD, by the CAL PD factor), and the MDP (PPD) set point would be
used.
Examples Write: LAS:MDI 40
Action: Output is controlled to provide a constant 40µA photodiode
feedback current.
Query: LAS:MDI?
Response: 200
Photodiode feedback current measurement is 200µA.
LASer:MDLN <bool> FRONT PANELREMOTE
LASer:MDLN?
Description Sets/returns the status of the front panel modulation input.
Parameters A <bool value> - where 1 = on, 0 = off.
Notes After turning the modulation off, it will remain off until turned back on via
this command or by resetting the instrument.
Examples Write: LAS:MDLN 1
Action: Turns the modulation input on and enables modulation via
the front panel BNC connector.
Query: LAS:MDLN?
Response 0
Modulation input is disabled.
November 2017 45 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 48
LASer:MDP <nrf> FRONT PANEL REMOTE
LASer:MDP?
Description Sets/returns the value of the photodiode power measurement when
CALPD > 0.
Parameters An <nrf value> in mW.
Notes If the CALMD (CAL PD) parameter is set to 0, the LAS:MDP value will not
be used. In this case, the measured MDP would be invalid, and the MDI
(IPD) set point would be used.
Examples Write: LAS:MDP 40
Action: Output is controlled to provide a constant 40mW
photodiode feedback power.
Query: LAS:MDP?
Response: 200
Photodiode feedback power measurement is 200mW.
ILBW Low bandwidth constant current mode.
IHBW High bandwidth constant current mode.
MDP Constant power mode.
Examples Write: LAS:MODE:IHBW
Action: Sets output mode to high bandwidth constant current
mode.
LASer:OUTput <bool> FRONT PANEL REMOTE
LASer:OUTput?
Description Sets/returns the status of the laser output.
Parameters A <bool value>; 1=on, 0=off.
Examples Write: LAS:OUT ON
Action: Turns on the laser output.
Query: LAS:OUT?
Response: 0
Laser output is off.
November 2017 46 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 49
LASer:RANge <nrf> FRONT PANEL REMOTE
LASer:RANge?
Description Sets/returns the output current range.
Parameters An <nrf value> that represents the output current range.
2 = 2000 mA
4= 4000 mA range
Notes This range setting effects the Laser Drive Current output current range
only. The Laser Drive Current output should be “off” when this command
is issued. If the Laser Diode Current output is “on” when this command is
issued, the LDX-3232 will generate error E515, and the range will not be
changed.
Examples Write: LAS:RAN 4
Action: Sets the current range to 4000mA.
Query: LAS:RAN?
Response: 2
Current range is 2000mA.
LASer:SET:LDI? FRONT PANEL REMOTE
Description Returns the output current setpoint value.
Parameters None.
Examples Query: LAS:SET:LDI?
Response: 50.0
LASer:SET:MDI? FRONT PANEL REMOTE
Description Returns the monitor photodiode current setpoint.
Parameters None.
Examples Query: LAS:SET:MDI?
Response: 30.0,
Monitor photodiode feedback current is set point is 30µA.
LASer:SET:MDP? FRONT PANEL REMOTE
Description Returns the monitor photodiode power setpoint.
Parameters None.
Examples Query: LAS:SET:MDP?
Response: 10.0
Monitor photodiode feedback power set point is 10.0mW
November 2017 47 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 50
LASer:STEP <nrf> FRONT PANELREMOTE
LASer:STEP?
Description Sets/returns the step size value.
Parameters An <nrf value> value between 1 to 9999 in 0.1mA.
Notes The step of one corresponds to the smallest display resolution for the
mode. For example, a step of one means 0.1mA, 0.1mW, or 1µA (if
CALPD=0).
Examples Write: LAS:STEP 100
Action: Sets the step size to 10mA, 10mW, or 100µA. Dependent
on instrument setup.
Query: LAS:STEP?
Response: 3
Step size is 0.3mA, 0.3mW, or 3µA. Dependent on
instrument setup.
LASer:TOLerance <nrf,nrf> FRONT PANELREMOTE
LASer:TOLerance?
Description Sets/returns the current tolerance and time window.
Parameters Two <nrf values>
The first is current tolerance in mA with a range of 0.1 to 100.0mA.
The second is the time window in seconds with a range of 0.001 to
50.000 seconds.
Notes Operation complete flag will bet set when the conditions are met.
The LDX-3232 defaults to a tolerance of 1.0 mA for three seconds, unless
changed by the LASer:TOLerance command.
If the LDX-3232 is operated in P mode, the current tolerance parameter is
not used. Instead a fixed value of 50µA is used for the IPD current. A
fixed value of 1mW is used for PPD power, and only the time window
parameter may be adjusted.
If the tolerance is set too tight it may never be achieved. This is due to the
calibration of the setpoint and measurement values.
Examples Write: LAS:TOL 0.5,10
Action: Current source will be in tolerance when the current is
within 0.5 mA for 10.000 seconds.
Query: LAS:TOL?
Response: 1.0,20.0
Tolerance setting of 1mA with a time window of 20
seconds.
November 2017 48 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 51
MESsage <string> FRONT PANELREMOTE
MESsage?
Description Sets/returns a stored string.
Parameters An ASCII string that is 1 - 16 bytes in length.
Examples Write: MESSAGE “This is a test.”
Action: “This is a test.“ will be stored in non-volatile memory.
Query: MESSAGE?
Response: Test 3
“Test 3” is stored in non-volatile memory.
RADix <string> FRONT PANELREMOTE
RADix?
Description Sets/returns the radix type for status, condition, and event query
response data.
Parameters BIN Binary
OCT Octal
DEC Decimal
HEX Hexadecimal
Notes DECimal is the default type.
Examples Write: RAD DEC
Action: Decimal radix is selected.
Query: RAD?
Response: HEX
Hexadecimal radix is selected.
TERM <bool> FRONT PANELREMOTE
TERM?
Description Sets/returns the default terminator to include a carriage return.
Parameters A <bool value>, 0=FALSE, non-zero=TRUE
Notes An altered terminator will be in the form <CR><NL><^END>. This technically
takes the LDX-3232 out of GPIB/IEEE488 specification, but may be done for
convenience when using nonstandard GPIB controllers. This termination will
be sent with all output until the TERM 0 command is sent or the LDX-3232 is
powered off.
Examples Write: TERM 1
Action: Sets <CR><NL><^END> as the output terminator.
Query: TERM?
Response: 0
<NL><^END> is set as the output terminator.
November 2017 49 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 52
TIME?FRONT PANELREMOTE
Description Returns the duration of time since the instrument was powered up.
Parameters None.
Examples Query: TIME?
Response: 1:02.36
Instrument has been powered up for one minute and 2.36
seconds.
TIMER?FRONT PANELREMOTE
Description: Returns the duration of time since the last TIMER? query was issued.
Parameters None.
Notes Each time the TIMER? query is issued, the timer is reset to 0 and the elapsed
time since the last TIMER? query is returned.
Examples Query: TIMER?
Response: 00:02:00.31
Two minutes and 0.31 seconds have elapsed since the last
TIMER? query was issued.
November 2017 50 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 53
Chapter 5 :
Functions and Features
This chapter introduces LDX-3232 High Compliance Current Source functions and operating
features such as event triggering, analog modulation of laser injection current, and save and
recall.
Saving and Recalling
Modulating the Current Source
Saving and Recalling from the F ront Panel
For applications where alternation between particular instrument configurations is needed, the
LDX-3232 Current Source offers a SAVE and RECALL feature. The SAVE feature allows easy
storage of all front panel settings for any given instrument configuration. These settings, that are
stored in one of ten memory bins, can be retrieved at any time with the RECALL function. This
saves setup time and it reduces the chance of setup error for tests that are periodically
repeated.
To enter the SAVE/RECALL mode, press the SELECT button in the PARAMETER section until
the SAVE indicator LED becomes lit and the unit displays the SAVE bin on the display. If
another bin number is desired, press and hold the SET button and rotate the adjust knob until
the desired bin number is displayed. Releasing the SET button completes the SAVE operation
and all of the instrument parameters are saved to the indicated bin number as they appear on
the front panel.
To recall a previously saved instrument setup, press the SELECT button until the RECALL
indicator led becomes lit. The instrument will display a RECALL bin number on the display. If
another bin number is desired, press and hold the SET button and rotate the adjust knob until
the desired bin number is displayed. Releasing the SET button completes the RECALL
operation. The instrument is restarted and the front panel parameters are reconfigured to the
new parameters saved under that bin number.
Table 5.1LDX-3232 Default Settings (Bin 0)
GPIB mode in LOCAL via front panel
REMOTE via GPIB
PARAMETERS not selected DISPLAY enabled, in I mode
CAL PD = 0 µA / mW Constant I, low bandwidth mode selected
LIM I (high range) = 1000 mA I setpoint = 0 mA
LIM I (low range) = 500 mA I
LIM V = 15 volts P
LIM P = 5000 mW RECALL BIN number = 0
The factory default values for front panel parameters may be restored by recalling bin zero (0).
Output off
setpoint = 0 µA
PD
setpoint = 0 mW
PD
November 2017 51 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 54
Saving and Recalling Under Remote Operation
For saving under remote operation, use the command *SAV x, where x is the desired bin
number (1-10). For a recall under remote operation, use the command *RCL x, where x is the
desired bin number (0 - 10).
Using the Laser Compliance V oltage Adjustment
For applications where critical protection of the laser is required, ILX Lightwave recommends
the following. Using a 1 resistor or other non-critical load, set the LASER voltage limit as
described in Chapter 2.
Then with the LASER output on, slowly increase the LASER current and watch the LASER
voltage measurement. Press the (DISPLAY) SELECT push button to see the voltage. At about
0.25 V below the LASER voltage limit value, the VOLTAGE LIMIT warning indicator will become
lit on the display. Continue to adjust the current upwards until the voltage limit is reached and
the output is shut off. Record the voltage at that the LASER output shuts off. It should be within
the specification of the LASER voltage limit accuracy. However, there is typically some offset
error. For example, if the LASER voltage limit is set to 5.0 V, the actual voltage where it shuts
off may be 4.9 V. This error term of 0.1 V should be considered when setting the LASER
compliance voltage limit.
The closer the LASER voltage limit shut off point is to the operating voltage of the laser, the
faster the circuit will work in the event of an open circuit. Some experimentation may be
necessary for optimum results.
Using the LDX-3232 Current Source’s Trigger Function
For applications where synchronous initiation of measurement tasks is required from a remote
instrument with the LDX-3232 Current Source, the Current Source offers a trigger output signal.
The TTL pulse is initiated with any remote change in set point of the laser current source.
The trigger output is available via a standard BNC connector on the rear panel of the LDX-3232
Current Source. See Figure 1.2. The TTL pulse is approximately 13µs wide and has a delay
time of about 10-12ms after the current set point is changed via GPIB. This delay does not
include the GPIB transmission time. The time for any command to be sent via GPIB depends on
the interface hardware and software control. The current set point may be changed by the
LAS:LDI, LAS:INC, or LAS:DEC commands.
Note: The minimum step change time required for a corresponding trigger pulse is
20ms. Care should be used with respect to the timing of any set point commands in
relation to the actual hardware function. The INC and DEC commands have a minimum
step time of about 20ms that allows for the hardware initiation of a set point change and
the settling of the output. The LAS:LDI command does not have any fixed delay for
settling. With this command, some trigger pulses may be missed if the program step
time is less than the 20ms minimum.
A one shot trigger pulse will occur on power up of the instrument due to the states of the
processor I/O.
November 2017 52 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 55
Using the LDX-3232’s BNC Interlock Connection
The LDX-3232 BNC interlock connection is designed to allow an external emergency input
signal to turn the current output off. The BNC interlock is TTL compatible and configured to be
normally closed (as opposed to the DB-9 interlock connection that is normally open). The BNC
interlock connection, when left unconnected, is pulled to a logic high with a 100kΩ resistor.
High values (+5 V) will keep the current source in the operational state, while low values (0 V)
will force the output off. Once the BNC interlock forces the output off, the output will stay off
until the interlock is closed again, and the output is re-enabled.
BNC Interlock Usage Example
When using the LDX-3232 Current Source to drive a temperature controlled laser diode, the
BNC interlock can be used to accept an error signal from the laser diode temperature controller
unit. Linking the LDX-3232 Current Source to a temperature controller* with the ability to signal
when its temperature control has gone outside of some tolerance window will allow the LDX3232 to safely shutoff its output before thermal damage occurs to the laser diode.
*The ILX Lightwave LDT-5900 series of temperature controllers include an output BNC
connector that is compatible with the BNC interlock feature of the LDX-3232. The output
BNC connector of the LDT-5900 has a TTL compatible output signal that will signal
when the temperature control has gone outside of a programmable temperature window.
Entering Photodiode Responsivity V alues
The photodiode responsivity (CAL PD) is used to convert between photodiode current and optical
power of the laser diode.
To enter the CAL PD parameter, use the following steps:
1. Calculate photodiode responsivity:
a. Measure (with a calibrated detector) the output power of the device.
b. Measure the corresponding photodiode current (I
c. Calculate the responsivity by dividing photodiode current by the optical power.
2. Select CAL PD with the (PARAMETER) SELECT button.
3. Enter the calculated photodiode responsivity from step 1.
a. If the CAL PD value is 0, the instrument will be put into a special mode where the
current output will be controlled by the I
value.
PD
This value will determine the output power of the instrument while in constant power mode.
PD
).
November 2017 53 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 56
Modulating the Laser Current Source
The LDX-3232 Current Source allows a modulated signal to be superimposed on the source
output current. The allowable bandwidth of the modulated analog signal, defined as the 3dB roll
off point, is dependent on the output current range and the mode (constant current low/high
bandwidth). See Chapter 1 for external analog modulation bandwidth specifications.
To set up the current source for modulated laser injection current, use the following steps:
1. Connect the modulation source signal to the instrument via the (MOD)
EXTERNAL BNC connector on the front panel.
2. Set the LDX-3232 to the appropriate bandwidth constant current mode by
pressing the (MODE) SELECT button on the front panel.
a. Low bandwidth mode uses a low pass filter to reduce noise. This
results is slower rise/fall times.
b. High bandwidth mode does not use a low pass filter. Use this
mode to get faster rise/fall times.
3. Select the appropriate current source range. The output must be off in order to change
ranges.
4. Turn on the LDX-3232 output.
5. Turn on the modulation source output.
a. If the modulation source output is turned on before the LDX-3232 output, it can
trigger limit errors
Notes: The limit clamp function for DC output current also applies to the modulated
current, protecting the laser in any mode.
The current limit setting is independent of the voltage drop of the device connected to
the laser output, and therefore, no dummy load is required for precise adjustment of the
current limit. Furthermore, since the current limit circuitry is fully independent of the main
current control, the current limit can be adjusted safely, even while the current source
output is active.
When the LDX-3232 current output is off, an internal short is placed across the output.
This short will prevent the modulation signal from reaching the device under test. Thus, it
is safe to disconnect the device when the LDX-3232's output is off, regardless of the
presence of the modulation signal.
November 2017 54 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 57
Chapter 6 :
Calibration and Troubleshooting
This chapter described the calibration of the LDX-3232 Series Current Source. Descriptions of
the required test instruments, calibration conditions, and the detailed procedures for calibration
are included. A troubleshooting guide is offered for some of the more common failure
symptoms.
Local Calibration
Remote Calibration
Troubleshooting Guide
Potentially lethal voltages exist within the LDX-3232 High Compliance Current Source.
To avoid electric shock, the user should not perform any of the maintenance on the
instrument unless qualified to do so. Qualified service personnel are required to wear
protective eyeglasses and anti-static wristbands while working on the LDX-3232 High
Compliance Current Source circuit boards. High voltages are present on and around the
instrument’s printed circuit boards.
High voltages are present on and around the printed circuit boards of the LDX-3232 High
Compliance Current Source.
Calibration Overview
There are several calibrations to be performed to completely calibrate the LDX-3232 Current
Source. The Laser Diode Current Source calibration consists of calibrating the current source in
both ranges and both modes (low and high bandwidth), calibrating the photodiode monitor
feedback and finally the forward voltage measurement. The current limits are calibrated
internally by the instrument as part of the calibration process.
The LDX-3232 Current Source can be calibrated with the case closed. It should be calibrated
every 12 months or whenever performance verification indicates that calibration is necessary,
such as differences between set point and measurement display values that exceed the
accuracy specification. Calibrate the instrument under laboratory conditions. Recommended
calibration temperature is at 23°C ± 1.0°C. When necessary, however, the LDX-3232 Current
Source may be calibrated at its intended use temperature if this is within the specified operating
temperature range of 0 to 40°C.
Finally, the LDX-3232 Current Source should be allowed to warm up for at least one hour before
beginning calibration.
November 2017 55 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 58
Recommended Equipment
Recommended test equipment for calibrating the LDX-3232 Current Source is listed in Table
6.1. Equipment other than that shown in the table may be used if the specifications meet or
exceed those listed.
Table 6.1 Recommended Test Equipment
DESCRIPTION MFG / MODEL SPECIFICATION
DMM HP 3457A DC Amps (@ 1.0A): +0.02%
Resistance (@ 10): 0.02%
0.1 A or 0.1 mW resolution
It will be necessary to connect various loads and circuits to the outputs of each current source
for the calibration procedure. A schematic is shown in Figure 6.1 for the photodiode calibration
circuit with the required components listed in the table below. Devices required for other
calibration loads are also listed in the table below.
Table 6.2 Calibration Circuit Components
Laser Current Source Calibration
Current Source Calibration, all models
Voltage Measurement Calibration
PD Monitor Calibration (See Schematic in Figure 6.1)
There are three calibration adjustments required for the LASER current source of the LDX-3232
Current Source. They are:
Calibration of the constant current source for both bandwidths and ranges.
Calibration of the laser voltage measurement.
Calibration of the constant light power (I
The LDX-3232 Current Source implements a two-point calibration for the Laser current source.
Two currents (approximately 80% and 20% of FS) are applied to a test load, and the resulting
actual currents are fed back (by the user) to the Current Source. The Laser Current Source
calibration program uses the two sets of data to calculate new calibration constants. These
calibration constants are used to set the actual current output of the current source.
Calibration can be done either locally via the front panel of the instrument or remotely through
GPIB commands.
Note: The operation complete flag (bit 0 of the ESR) may be used to trigger a service request. This type of
interrupt is enabled by setting bit 0 of the SRE register (via the *ESE command) and bit 5 of the SRE
register (via the *SRE command). Service request (SRQ) handling depends on the GPIB hardware in use.
Refer to the GPIB hardware user’s manual for details on how service requests are handled.
After a calibration value is entered, the *OPC? query may be used to determine when the calibration
sequence is done. However, the *OPC, *WAI commands, or *OPC? query should not be issued until after
the expected calibration value is entered, or the system will hang. This happens because the LDX-3232
Current Source will wait indefinitely for an input, yet not allow any input until the calibration is finished.
Current Source Calibration
) feedback circuits.
PD
The following procedure is for calibrating each range of the current source. Calibration must be
performed on both bandwidth modes. A total of four calibrations will be performed on the current
source. Start with the low range, low bandwidth mode. Configure the Laser Current Source in
the following state:
Table 6.3Current Calibration Configuration Setup
Function State GPIB Command
LASER MODE I (constant current low bandwidth)
I (constant current high bandwidth)
RANGE low (lowest per instrument model) LAS:RAN <nrf>
CURRENT LIMIT 90% of FS (full scale of range) LAS:LIM:Ix <nrf>
SETPOINT 80% of FS LAS:LDI <nrf>
LAS:MODE:ILBW
LAS:MODE:IHBW
a. Measure and record the resistance of a 1, 20 W resistor (A 4-wire resistance
measurement is recommended.) Connect the load resistor across the LASER output
terminals (Laser Anode and Laser Cathode) on the 9 pin connector of the rear panel.
b. Connect the calibrated DMM across the load resistor. Configure the multimeter to
measure voltage in volts. The actual Laser current will be calculated in the following
manner using Ohm's Law.
=
November 2017 57 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 60
where V is the accurately measured voltage across the resistor, and R is the
accurately measured load resistance in Ohms. I is the actual Laser current in
Amps.
Note: Check the current measuring range of the multimeter. Laser Output current may be
measured in mA by connecting the meter directly to the outputs of the Laser Current Source.
c. The laser output must be turned on to begin laser current calibration mode:
i. Local: Press the (MODE) ON button.
ii. Remote: Write the LAS:OUTPUT ON command.
d. Enter LASER current calibration mode:
i. Local: Press (GPIB) LOCAL and (DISPLAY) SELECT buttons at the
same time.
ii. Remote: Write the LAS:CAL:LDI command.
The display will blank and the Current Source will drive the current output to the set point
and the Display will indicate output current in mA.
e. Calculate the actual current through the load as described in step b.
i. Local: Press and hold in the (PARAMETER) SET button and turn the
adjust knob until the display indicates the calculated actual
current.
ii. Remote: Write the LAS:LDI <nrf> command, where <nrf> is the actual
measured output current.
Note: Querying the LAS:CAL:LDI? query and receiving a
response of 1 means that the instrument is ready to receive a
calibration value.
f. The current source will drive the current output to approximately 25% of the first
calibration set point when the following actions occur:
i. Local: Releasing the (PARAMETER) SET button to store the first
calibration value.
ii. Remote: Writing the LAS:LDI <nrf> command to store the first calibration
value.
g. Repeat step e.
h. Once the second actual value is entered, the new calibration constants will be calculated
and stored into non-volatile memory. In low bandwidth mode calibration mode, the LDX3232 Current Source will also perform current limit calibration as indicated by the
CURRENT LIMIT LED flashing.
i. If performing a remote calibration, the *OPC? query may be used after the LAS:LDI
<nrf> command is sent to determine when the calibration is completed.
i. Turn the current source output off:
i. Local: Press the (MODE) ON button.
ii. Remote: Write the LAS:OUTPUT OFF command.
j. Change the Mode to IHBW (high bandwidth constant current mode):
i. Local: Press the (MODE) SELECT button until the I
LED is lit.
HBW
ii. Remote: Write the LAS:MODE:IHBW command.
k. Repeat the calibration procedure as described above.
l. Change the output range to high range and repeat the calibration procedure as
described above.
November 2017 58 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 61
IPD Current Calibration
The following procedure is for calibrating the LASER IPD (PD monitor) set point and
measurement. This procedure calibrates the feedback circuits for constant I
modes. When these values are reached and are stable, the user enters the actual value of the
monitor current, measured with a DMM. The LDX-3232 Series Current Source then
automatically calibrates the LASER feedback circuits.
Configure the Current Source in the following manner for the Laser Current Source calibration:
Table 6.4Power Calibration Configuration Setup
Function State GPIB COMMAND
MODE P (constant power) LAS:MODE:MDP
CAL PD set equal to zero LAS:CALMD 0
CURRENT LIMIT 90% of FS (full scale of range) LAS:LIM:Ix
I
SETPOINT
PD
80% of FS LAS:MDI <nrf>
a. With the LASER output off, connect a calibrated ammeter to the PD Anode output of the
Laser Current Source output connector (9 pin on rear panel), and connect the circuit of
Figure 6.1 to the LASER and PD outputs with the correct load resistor for the Current
Source model under calibration (See Table 6.2).
Note: If a precision ammeter is not available, use a zero-Ohm jumper in place of the
ammeter. Then, place a calibrated DMM (with 0.1 mV resolution) to measure the voltage
across the resistor, R1, as shown in Figure 6.1. Calculate the current in the following
steps by using Ohm's Law:
=
where V is the accurately measured voltage across the resistor, and R is the
accurately measured load resistance in Ohms. I is the actual Laser current in
Amps.
b. The laser output must be turned on to begin the power calibration mode:
i. Local: Press the (MODE) ON button.
ii. Remote: Write the LAS:OUTPUT ON command.
After the output is on, verify proper operation. The I
measurement (via the DMM) and
PD
set point should be close, and the unit should not be in current limit.
c. Enter I
Current Calibration Mode:
PD
i. Local: Press the (GPIB) LOCAL and FINE buttons at the same.
ii. Remote: Write the LAS:CAL:MDI command.
After a few seconds the display will show the I
set point value.
PD
d. Wait for the value on the display to stabilize (has not changed by more than one digit for
several seconds). When this has occurred, the LDX-3232 Current Source is ready for
the actual I
value to be entered.
PD
i. Local: Press and hold in the (PARAMETER) SET button and turn the
adjust knob until the display shows the actual monitor current, as
shown on the calibrated ammeter (or the calculated I
Step a).
and constant PPD
PD
value from
PD
November 2017 59 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 62
ii. Remote: Write the LAS:MDI <nrf> command, where <nrf> is the actual
monitor current, as shown on the calibrated ammeter (or the
calculated I
value from Step a.)
PD
e. The current source will drive the current output to approximately 25% of the first
calibration set point when the following actions occur:
i. Local: Releasing the (PARAMETER) SET button to store the first
calibration value.
ii. Remote: Writing the LAS:MDI <nrf> command to store the first calibration
value.
f. Repeat Step d.
g. The current source will calculate the calibration constants and store them to non-volatile
memory when the following actions occur:
i. Local: Releasing the (PARAMETER) SET button to store the second
calibration value.
ii. Remote: Writing the LAS:MDI <nrf> command to store the second
calibration value.
November 2017 60 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 63
Laser Forward Voltage Measurement Calibration
The following procedure is for calibrating the LASER voltage measurement. Configure the
Current Source in the following manner for the Laser Current Source Voltage calibration:
Table 6.5 Voltage Calibration Configuration Setup
Function State GPIB Command
LASER MODE I (constant current low bandwidth) LAS:MODE:ILBW
RANGE 2000 mA LAS:RAN 2
CURRENT LIMIT Set to FS (full scale of range) LAS:LIM:I2 2000
I SETPOINT 800 mA LAS:LDI 800
a. With the current source output off, connect a calibrated voltmeter, in parallel with a
resistor (15, 5W), to the LASER output (Laser Anode and Laser Cathode) on the rear
panel of the Current Source.
b. The laser output must be turned on to beg laser voltage calibration mode:
i. Local: Press the (MODE) ON button.
ii. Remote: Write the LAS:OUTPUT ON command.
c. Enter LASER voltage calibration mode:
i. Local: Press the (GPIB) LOCAL and RANGE buttons at the same time.
ii. Remote: Write the LAS:CAL:LDV command.
The display will blank and the Current Source will drive the LASER current output to the
respective set point and the display will indicate forward voltage in Volts.
d. Enter the measured voltage (in volts):
i. Local: Press and hold in the (PARAMETER) SET button and turn the
adjust knob until the display indicates the measured voltage.
ii. Remote: Write the LAS:LDV <nrf> command, where <nrf> is the actual
voltage, as shown on the calibrated ammeter.
e. The current source will drive the current output to approximately 25% of the first
calibration set point when the following actions occur:
i. Local: Releasing the (PARAMETER) SET button to store the first
calibration value.
ii. Remote: Writing the LAS:LDV <nrf> command to store the first calibration
value.
f. Repeat Step d to enter the second measured voltage.
g. The Current Source will store the calibration constants and store them to non-volatile
memory when the following actions occur:
i. Local: Releasing the (PARAMETER) SET button to store the second
calibration value.
ii. Remote: Writing the LAS:LDV <nrf> command to store the second
calibration value.
This ends the section on calibration of the LDX-3232 Current Source.
November 2017 61 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Page 64
Troubleshooting Guide
This section is a guide to troubleshooting the LDX-3232 Current Sources. More common
symptoms are listed here, and appropriate troubleshooting actions are given. The user is
recommended to start at the beginning of this guide. Read the symptom descriptions, and follow
the steps for the corrective actions that apply. If a problems encountered are beyond the scope
of this guide, contact an ILX Lightwave customer service representative.
Table 6.6 Troubleshooting Guide
Symptom Corrective Actions
GENERAL
LDX-3232 unit will not power up Check AC power line voltage and power cord connection.
Power on, but display is frozen and
buttons don’t operate
Displays E514 on power up on
display
Instrument resets at power up or
any LASER mode change
Power on, but no current output Check interlock pins on LASER input connector on instrument rear panel; these pins
Output current at limit; can’t be
lowered
Output goes off intermittently Check the interlock circuit; an intermittent interlock will turn the output off.
Unable to adjust output Check the I LIMIT parameter for the output range in use; see that it is set above 0 mA.
Power mode operation has high output
current, but little or no power
measured
Output exceeds power limit The “power limit” is not a hardware limit. It only serves as a warning that the power
Open circuit error occurs during
calibration
Calibration is aborted unintentionally Calibration modes will be aborted if an open circuit is detected.
VOLTAGE LIMIT indicator blinks This indicates a voltage limit error; check laser connections. High impedance may
Open circuit error E503 or voltage limit
error E505 prevents output from
reaching desired value
Instrument may have lost power (AC line) briefly. Cycle power to restart.
Hold (GPIB) LOCAL button, RANGE button and FINE button as the instrument powers
up. This clears the laser board EEPROM, LLLLL will be displayed on the display; the laser current source must be recalibrated.
LASER CURRENT SOURCE
must be shorted either directly or through a switch.
If OPEN CIRCUIT indicator is lit, check the load connections and then try again.
Check the OUTPUT ON button, the corresponding LED should be lit.
If POWER mode is used, check the monitor diode (feedback) connections; try
reversing the polarity of the monitor photodiode; check the photodiode bias adjustment
on the rear panel.
If in I or I
below the limit may require several turns of the adjust knob if the setpoint is much
greater than the desired limit setting.
Check that the AC power cord connection is secure; power line drop-outs may reset
the unit and when power is restored, the output will be off.
Check back panel PD BIAS; if set too low, may act as an open feedback loop; if in
doubt, set the PD BIAS to mid-range (2.5 V)
measurement has exceeded the limit setpoint. Normally this limit will cause the output to
shut off. This feature can be disabled via the LAS:ENABLE:OUTOFF command.
Check load connections; check that measuring meter does not auto-range (use nonauto-ranging modes).
cause this condition.
The LDX-3232 instrument has an adjustable laser compliance voltage. Check to see if
the LASER voltage limit setting is too low (Chapter 2); check laser connections.
mode, check the current setpoint and I LIMIT setting; setting the output
HBW
November 2017 62 70037802 LDX-3232 Manual
09-FM030
1/30/13
Rev. 01
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.