Table B.2 10kΩ Thermistor Temperature Range ................................................ 50
Table B.3 10kΩ Thermistor Voltage vs. Resolution ............................................. 50
November 2017 4 70019907 LDT-5500B Manual
Page 7
Safety and Warranty Information
The Safety and Warranty Information section provides details about cautionary symbols used in the
manual, safety markings used on the instrument, and the information about the Warranty including
Customer Service contact information.
Safety Information and the Manual
Throughout this manual, the words Caution and Warning indicate potentially dangerous or hazardous
situations which, if not avoided, could result in death, serious or minor injury, or damage to the
product. Specifically:
Indicates a potentially hazardous situation which can result in minor or moderate injury or
damage to the product or equipment.
Indicates a potentially dangerous situation which can result in serious injury or death.
Indicates visible and/or invisible laser radiation. Avoid direct exposure to the beam.
General Safety Considerations
If any of the following conditions exist, or are even suspected, do not use the instrument until safe
operation can be verified by trained service personnel:
Visible damage
Severe transport stress
Prolong storage under adverse conditions
Failure to perform intended measurements or functions
If necessary, return the instrument to ILX Lightwave, or authorized local ILX Lightwave distributor, for
service or repair to ensure that safety features are maintained (see the contact information on page
6).
All instruments returned to ILX Lightwave are required to have a Return Material Authorization (RMA)
number assigned by an official representative of ILX Lightwave Corporation. See “Returning an
Instrument” on page 5 for more information.
November 2017 5 70019907 LDT-5500B Manual
Page 8
Safety Symbols
This section describes the safety symbols and classifications.
Technical specifications including electrical ratings and weight are included within the manual. See
the Table of Contents to locate the specifications and other product information. The following
classifications are standard across all ILX Lightwave products:
Indoor use only.
Ordinary protection: This product is NOT protected against the harmful ingress of moisture.
Class I Equipment (grounded type).
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
Pollution Degree II
Installation (overvoltage) Category II for transient overvoltages
Maximum Relative Humidity: <80% RH, non-condensing
Operating temperature range of 0°C to 40°C
Storage and transportation temperature of -40°C to 70°C
Maximum altitude of 3000m (9843 ft).
This equipment is suitable for continuous operation.
Safety Marking Symbols
This section provides a description of the safety marking symbols that appear on the instrument.
These symbols provide information about potentially dangerous situations which can result in death,
injury, or damage to the instrument and other components.
Visible
Caution:
Refer to
manual
Caution:
Risk of
Electric
Shock
Earth
Ground
On: In position of push
control. The (I) denotes that
the mains are on.
Terminal
Protective
Conductor
Terminal
Alternating
Current
Caution:
Hot Surface
Of: Out position of push
control. The (O) denotes that
the mains are on.
and/or
invisible
laser
radiation
Frame or
Chassis
Terminal
November 2017 6 70019907 LDT-5500B Manual
Page 9
Warranty
ILX Lightwave warrants this instrument to be free from defects in material and workmanship for a
period of one year from the date of shipment. During the warranty period, ILX Lightwave will repair or
replace the unit, at our option, without charge.
Limitations
This warranty does not apply to fuses, lamps, defects cause by abuse, modifications, or use of the
product for which it was not intended.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty of
merchantability or fitness for any particular purpose. ILX Lightwave shall not be liable for an
incidental, special, or consequential damages.
If a problem occurs, please contact ILX Lightwave with the instrument’s serial number, and
thoroughly describe the nature of the problem.
Returning an Instrument
If an instrument is to be shipped to ILX Lightwave for repair or service, be sure to:
1. Obtain a Return Material Authorization number (RMA) from ILX Lightwave Customer Service.
2. Attach a tag to the instrument identifying the owner and indicating the required service or
repair. Include the instrument serial number from the rear panel of the instrument.
3. Attach the anti-static protective caps that were shipped with the instrument and place the
instrument in a protective anti-static bag.
4. Place the instrument in the original packing container with at least 3 inches (7.5cm) of
compressible packaging material. Shipping damage is not covered by this warranty.
5. Secure the packing box with fiber reinforced strapping tape or metal bands.
6. Send the instrument, transportation pre-paid, to ILX Lightwave. Clearly write the Return
Material Authorization number on the outside of the box and on the shipping paperwork. ILX
Lightwave recommends the shipment be insured.
If the original shipping container is not available, place the instrument in a container with at least 3
inches (7.5cm) of compressible packaging material on all sides.
Repairs are made and the instrument returned transportation pre-paid. Repairs are warranted for
the remainder of the original warranty or for 90 days, whichever is greater.
Claims for Shipping Damage
Upon receiving the instrument, inspect it immediately for any damage or shortages on the
packing list. If the instrument is damaged, file a claim with the carrier. The factory will supply a
quotation for estimated costs of repair. The customer is responsible for negotiation and
settlement with the carrier for the amount of damage.
November 2017 7 70019907 LDT-5500B Manual
Page 10
Comments, Suggestions, and Problems
To ensure getting the most out of ILX Lightwave products, direct any product operation, service
related questions, or comments to ILX Lightwave Customer Support. Contact ILX Lightwave in
whatever way is most convenient:
ILX Lightwave
31950 Frontage Road
Bozeman, Montana, USA 59715
When contacting ILX Lightwave, please have the following information:
Model Number
Serial Number
End-User Name
Company
Phone Number
Email Address
Description of the problem
If ILX Lightwave determines that a return to the factory is necessary, a Return Material Authorization
(RMA) will be issued. Please mark this number on the outside of the shipping box.
The customer or carrier is responsible for any shipping damage when returning an instrument to ILX
Lightwave. ILX Lightwave recommends insuring the shipment. If the original shipping container is not
available, place the instrument in a container with at least 3 inches (7.5cm) of compressible
packaging material on all sides.
November 2017 8 70019907 LDT-5500B Manual
Page 11
Chapter 1 :
Introduction and Specifications
This chapter contains operation and maintenance information for the LDT-5500B Series
Temperature Controller:
Product Overview
Specifications
Product Overview
The LDT-5500B Series Temperature Controller is a microprocessor-based, precision
thermoelectric temperature controller designed for temperature control of laser diodes, detectors
and other temperature sensitive devices. The LDT- 5500B Series can be used for laser diode
testing, laser diode frequency stabilization, IR detector cooling, and the determination of
characteristics of electronic devices. The LDT-5500B Series combines high analog stability with
the versatility of a microprocessor-based instrument. The internal microprocessor controls the
operation of the LDT-5500B Series and performs the non-linear conversion of thermistor
resistance to temperature based on user-defined constants.
The LDT-5500B Series can be configured to operate with a wide variety of thermistor
temperature sensors and TE modules, as well as AD590 series and LM335 series temperature
sensors.
Features of the LDT-5500B Series include:
Intuitive front panel layout.
Large and easy-to-read green LED display.
Display resolution of 0.1 degree Celsius.
Output current limit control to safely operate TE module based devices.
Configurable for a variety of thermal sensors.
LDT-5525B is capable of 4A, 24W; LDT-5545B is capable of 5A, 50W.
Easy to use remote USB operation.
Closed-case calibration.
November 2017 9 70019907 LDT-5500B Manual
Page 12
Available Options and Accessories
Options and accessories available for the LDT-5500B Series Temperature Controller include the
following:
Table 1.1 Accessories List
DESCRIPTION MODEL NUMBER
Single Rack Mount Kit, LDX-3500 Series & LDT-5500 Series RM-134
Dual Rack Mount Kit, LDX-3500 Series & LDT-5500 Series RM-135
Laser Didoe Mount, 3-pin TO-Can, Temperature Controlled LDM-4405
Laser Didoe Mount, 3 & 4-pin TO-Can, Temperature Controlled LDM-4990
Telecommunications Laser Diodes Mounts
(For TO-Can, DIL, and Butterfly packages)
TEC Controller to Unterminated Cable, 5A, DB15 Male to Bare Wire CC-501S
TEC Controller to Mount Cable, DB15 Male to DB9 Female CC-505S
Calibrated 10kΩ Thermistor TS-510
Uncalibrated 10kΩ Thermistor TS-520
RTD Temperature Sensor Converter TSC-599
Unipolar Control Adapter UCA-350
Not a complete list, visit website or contact sales staff for full list.
Please contact ILX Lightwave or visit www.newport.com/ilxlightwave for information on additional
options for applications.
LDM-498X
November 2017 10 70019907 LDT-5500B Manual
Page 13
Specifications
Table 1.2 Specifications
LDT-5525B LDT-5545B
TEMPERATURE CONTROL
OUTPUT
Temperature Control Range
Temperature Setpoint Resolution 0.1oC
Temperature Setpoint Accuracy
Thermistor
AD590
LM335
Short Term Stability (1 hour)
Long Term Stability (24 hours)
Temperature Coefficient
TEC OUTPUT
Type
Control Algorithm
Compliance Voltage
Maximum Output Current
Maximum Output Power
Current Noise and Ripple
CURRENT LIMIT
Current Limit Range
Current Limit Set Accuracy
TEMPERATURE SENSOR
Thermistor 2-wire NTC
IC Sensors AD590/LM335
RTD Sensor
Thermistor Sensing Current
IC Sensor Bias
Usable Thermistor Range
User Calibration
Thermistor
IC Sensor
Analog Output
Transfer Function
10 A thermistor
100 A thermistor
AD590
LM335
TEC MEASUREMENT (DISPLAY)
Display Type
Temperature Range
7
1
2
3
3
-99oC to 199.9oC
o
C
±0.2
o
±0.2
C
o
±0.2
C
<±0.006oC
<±0.01oC
Bipolar current source
Smart Integrator, Hybrid PI
>6V DC (@ 4A)
4.0A
24W4
6
<1 mA rms
0 to 4.04A
>10V DC (@ 5A)
5.0A
50W5
0 to 5.05A
±50 mA
2-wire RTD 100, 500, or 1 k
10/100 A
AD590 = 8V
LM335 = 0.6 mA
250 to 450,000
Steinhart-Hart, 3 constants
Slope and offset, two point
0 to 5V
8
10k/V
10k/V
100A/V
1V/V
4 digit green LED
o
-99
C to 199.9 oC
November 2017 11 70019907 LDT-5500B Manual
Page 14
Temperature Resolution
10 A thermistor
100 A thermistor
AD590
LM335
3
Accuracy
Thermistor Resistance Range
10 A thermistor
100 A thermistor
Thermistor Resistance Resolution
10 A thermistor
100 A thermistor
Thermistor Resistance Accuracy
10 A thermistor
100 A thermistor
TE Current Range
TE Current Resolution
TE Current Accuracy
o
0.1
C
o
C
0.1
o
0.1
C
o
0.1
C
±0.5oC typical
2.5 to 450k
0.25 to 45.0k
0.1 k
0.01 k
±0.05% of FS
±0.05% of FS
-4.00 to 4.00A
0.01A
±0.03A
CONNECTIONS
TEC I/O 15 pin D-sub Rear Panel
Analog Output BNC Front Panel
Communications USB Rear Panel
GENERAL
Power, VAC (50-60 Hz) VAC ±10%
Current Draw
Size
100-120VAC 0.86A
230VAC 0.42A
88mm x 185mm x 304mm
3.5” x 7.3” x 12”
Weight 3.6 kg (8 pounds)
Ambient Temperature Range
Operating: 0 to 40
Storage: -40 to 70 oC
Humidity <80% relative, non-condensing
Warm-Up 1 hour to rated accuracy
1.
Actual temperature control range depends primarily on the thermal load, sensor, and TE module used.
2.
Accuracy figures are quoted for a typical 10k thermistor and 100 A current setting. Accuracy figures are relative to
the calibration standard. Both resolution and accuracy are dependent upon the user-defined configuration of the
instrument.
3.
Stability is a strong function of the thermal environment of the temperature sensor and the TE module. Ambient air
currents in particular can cause fluctuations of 0.1
4.
Output power rated into a 1.5 load.
5.
Output power rated into a 2.0 load. @100VAC, Reduced to 40W.
6.
Measured with the instrument in ITE mode at half scale output over a bandwidth of 10 Hz to 10 MHz.
7.
With use of optional TSC -599 Temperature Sensor Converter.
8.
0 to 5V representing measured temperature.
o
C in an exposed mounting configuration.
-5.00 to 5.00A
100-120VAC 1.16A
230VAC 0.55A
o
C
November 2017 12 70019907 LDT-5500B Manual
Page 15
Chapter 2 :
Operation
This chapter describes how to install, adjust, and operate the LDT-5500B Series Temperature
Controller. It is divided into sections covering installation, familiarization and adjustment, and
normal operating procedures. This chapter also includes:
Front Panel Features
Front Panel Operations
Installation Procedures
AC Po wer Considerations
The LDT-5500B Series Controllers can be configured to operate at nominal line voltages of 100,
120, and 230 VAC (±10%). This can only be done at the factory and need not be changed
before operating the instrument. Before operating the instrument, verify that the voltage
indicated on the back panel of the instrument matches the power-line voltage used in the
operating environment.
To avoid electrical shock hazard, connect the instrument to properly earthgrounded, 3- prong receptacles only. Failure to observe this precaution can result
in severe injury or death.
Rack Mounting
The LDT-5500B Series Temperature Controller may be rack mounted by installing a rack mount
flange on either side of the enclosure. All rack mount accessory kits contain detailed mounting
instructions. Refer to Chapter 1 for applicable rack mount accessory part numbers.
Pow er Up Sequence
With the LDT-5500B Series Temperature Controller connected to an AC power source, pressing the
POWER switch will supply power to the instrument and start the power up sequence.
During the power-up sequence, the following takes place.
For two seconds, all indicators light up, and all of the 7-segment displays indicate “8”. All lights are
then turned off for two seconds. Next, two versions of firmware are displayed on the front panel: first
version number displayed is mainboard; the second is the front panel. After this, the sensor positions
are displayed for two seconds. After the power-up sequence, the unit is configured to the state it was
in when the power was last shut off (except for the display mode which defaults to the control mode).
November 2017 13 70019907 LDT-5500B Manual
Page 16
USB
The LDT-5500B may be controlled by PC using USB. The USB Type B connector is located on the
rear panel above the sensor select switch. See the Figure 2.2. The USB drivers (from either the ILX
Lightwave website or from the USB Flash Drive/CD) must be installed prior to making USB
connection. Please refer to Chapter 3 for more detailed instructions on operating the instrument
through USB.
Introduction to the LDT -5500B Series F ront Panel
The LDT-5500B Series Temperature Controller's front panel contains displays and controls for the
Temperature Controller hardware. Each of the labeled areas on the front panel is described in this
chapter.
Refer to Figure 2.1 for the following discussions of the LDT-5500B Series Temperature Controller
front panel sections. The key words are in CAPITAL LETTERS for quick identification.
Figure 2.1 LDT-5500B Series Front Panel
Adjustments
The adjustment knob is used for entering values and can be disabled using the ENABLE (adjust
enable) button and indicator. In order to make any adjustment, the ENABLE indicator must be lit.
Pressing the ENABLE button toggles the ENABLE indicator on or off.
Display
The four digit display is used to show measurements, temperature and output set point, and
parameter values. Whenever a set point is being adjusted, the corresponding enunciator will blink.
After the set point value is reached, the enunciator will continue to blink for three seconds. After
which, the set point will be recorded.
The display SELECT button is used to select the measured current (ITE), sensor resistance, or
temperature. The set point type is determined by the MODE selection. Repeatedly pressing the
November 2017 14 70019907 LDT-5500B Manual
Page 17
display SELECT switch will cycle the display from ITE to temperature to resistance (with thermistor
sensors only) to set point.
When in ITE mode, the set point will be TE current in Amps.
When in R mode, the set point will be thermistor resistance in k. R mode is not available if the back
panel SENSOR SELECT switch is set to LM335 or AD590. R mode operation may offer improved
set point resolution (over T mode), depending on the desired temperature set point.
Note: If the control method or sensor current is changed, the set point will be stored.
When in T mode the set point will be temperature in °C.
Parameters
The LDT-5500B Series Temperature Controller allows adjustment of the following parameters, LIM I
(TE current limit), LIM T (temperature limit), GAIN (sensor feedback amplifier gain) and CONST
(sensor calibration values).
The LDT-5500B Series will limit the ITE output to the LIM I value, regardless of the set point or
control mode.
The temperature is limited (via the sensor feedback) to the LIM T value. If the sensor reads a
temperature which is greater than LIM T, the output will be disabled.
The GAIN value is used to control the sensor feedback gain, and thus the temperature settling time
and overshoot. GAIN values can be adjusted in discrete steps from 1 to 300. If the GAIN is set low,
the actual temperature will take longer to reach the temperature set point. If the GAIN is set too high,
the actual temperature may oscillate around the set temperature.
The optimum GAIN setting depends on the thermal load, temperature set point, and the type of TE
cooler. Set the GAIN to its lowest value and then try increasing it until the temperature oscillates
around the set temperature. Then, reduce the GAIN one step.
November 2017 15 70019907 LDT-5500B Manual
Page 18
Parameter Setup
The parameter SELECT button is used to view parameters. Repeatedly pressing the parameter
SELECT switch will cycle through the parameters.
When a parameter is selected for viewing, its value will remain on the display for three seconds. The
parameter value can be changed by pressing the SET button and turning the adjust knob. The
parameter LED will not blink when being adjusted. Three seconds after the parameter value has
been reached, the value will be stored and the display will revert to the last measurement mode.
Thermistor Sensor Cal
These are the constants of the Steinhart-Hart equation that the user enters to calibrate the TEC for
thermistor temperature conversions. The Steinhart-Hart equation is used to derive temperature from
the non-linear resistance of an NTC (Negative Temperature Coefficient) thermistor. For information
on setting C
sensor current selection, see Appendix B.
, C2, and C3 for thermistors, see Appendix A. For information on thermistor selection and
1
The range of values for C
To read C
, C2 or C3, press the parameter SELECT button until it sequences to CONST. While on the
1
CONST parameter, the SELECT button cycles through C
, C2, and C3 are -9.99 to +9.99.
1
, C2, and finally C3. The display will indicate
1
which constant is currently selected, and can be adjusted or viewed by pressing the parameter SET
button, and turn the ADJUST knob until the correct value is displayed.
Appendix A contains an explanation of the Steinhart-Hart equation and a computer program to
determine these values for any thermistor.
IC Sensors
These are the constants entered to calibrate the TEC for AD590 and LM335 temperature
conversions.
When a linear sensor device (such as an AD590 or LM335) is used, a linear equation is used. If a
linear sensor's calibration is not known, set C
sensor calibration, see Appendix C.
Appendix C contains information on sensor calibration constants for AD590 and LM335 sensors.
Since these devices are used over their linear range, the constants C
determine a linear approximation of the temperature, rather than the Steinhart-Hart non-linear
approximation which applies for thermistors. The appropriate algorithms are automatically
implemented whenever the sensor type is selected via the back panel SENSOR SELECT switch.
However, C
and C2 must be changed by the user.
1
= 0.00, C2 = 1.00. For more information on linear
1
and C2 are used in this case to
1
November 2017 16 70019907 LDT-5500B Manual
Page 19
Output and Control Mode
The MODE section contains the mode SELECT button, which is used to select the instrument control
mode. Repeatedly pressing the mode SELECT button cycles through the current (I TE), sensor
reference (R), or temperature (T) control modes. The LED indicators show the selected mode. The
MODE section also contains the OUTPUT button and indicator. The ON indicator is lit whenever the
output is on. Pressing the OUTPUT button enables the output of the instrument as indicated by the
LED indicator. With the output enabled, any thermal load connected to the instrument will be
controlled to the mode respective set point.
Conditions Which Will Automatically Disable the OUTPUT
1.
Temperature Limit
2.
Sensor Open (While Output On)
3.
TEC Module Open (While Output On)
4.
SENSOR SELECT Switch Moved (While Output On)
5.
Operating Mode Changed (While Output On)
6.
Short condition or low resistance (LDT-5545B only)
Error Indicators
The ERROR indicators become lit when the corresponding conditions occur. The SENSOR
OPEN light comes on whenever the sensor connections are open. The TE OPEN indicator
becomes lit whenever an open circuit (or a high impedance condition) occurs on the TE module
output when the output is on. When a TE OPEN condition occurs, the output will be shut off and
the indicator will remain on until the problem is resolved and the output is turned on again.
The T LIMIT light remains lit whenever the temperature limit is reached. The I LIMIT light will
remain lit whenever the I TE current limit is reached.
If a shorting condition occurs or a load which has < 1 of resistance is used, the output will be
disabled and both the I LIMIT LED and TE OPEN LED will remain lit. This error can only occur
on the LDT-5545B due to the minimum load resistance specification.
Analog Output
An analog output signal is available at the ANALOG OUTPUT connector (BNC) on the front
panel. This signal is a voltage between 0 - 5.0 volts which is proportional to the measurement.
For example, an analog output signal of 2.5 volts (+0.5 volts) would represent a measurement
of 50% of full scale.
November 2017 17 70019907 LDT-5500B Manual
Page 20
A
Figure 2.2 LDT-5500B Series Back Panel
Back Panel Controls and Connections
Refer to Figure 2.2 for the following discussions of back panel controls and connectors. There
are no user serviceable parts in the instrument, including the external fuses in the AC power
entry module.
Sensor Select Switch
The SENSOR SELECT switch is used to select sensor type and, in the case of thermistor
sensor, the source current level. Table 2.1 shows the SENSOR SELECT positions and
corresponding position code. When the sensor switch is changed during TEC mode operation,
the new sensor position code will be indicated on the TEC display for three seconds.
Table 2.1 SENSOR SELECT Switch Positions
SWITCH POSITION CODE
100 A
10 A
LM335 -03-
D590 -04-
-01-
-02-
The 10A and 100A designations are for the thermistor current source level. When using a
thermistor, the supply current depends on the thermistor operating temperature range and the
required temperature resolution. Guidelines for setting this switch are contained in Appendix B.
The AD590 sensor operates as a current source which is proportional to the sensed
temperature. The LM335 sensor operates as a voltage source which is proportional to the
sensed temperature. Both of these sensors are approximately linear over their operating
ranges. When they are used, the constants C
and C2 are used for a two-point conversion. For
1
more information on setting the constants for use with these sensors, see Appendix C.
November 2017 18 70019907 LDT-5500B Manual
Page 21
TEC Connector
In the lower right hand corner, when facing the back panel, there is a 15-pin D-connector for the
TEC MODULE. This connector is used for the input and output connections, as shown by the
pin-out diagram of Figure 2.3. ILX TEC interconnect cables connect directly to the 15-pin Dconnector.
Figure 2.3 Back Panel TEC Connector
USB Connector
The USB Type B connector is located just to the left of center of the back panel. Attach the USB
cable to the instrument in the proper orientation. A USB cable is provided in the shipping kit.
See Chapter 3 for driver installation information.
TEC Grounding Considerations
The TEC outputs of the LDT-5500B Series are isolated from chassis ground, allowing either
output terminal to be grounded at the user's option.
Note: For the TEC connector, if any one terminal pin is grounded, then no other terminal
pin should be grounded; damage to thermal load or the temperature controller may
occur.
November 2017 19 70019907 LDT-5500B Manual
Page 22
General Operating Procedures
The following sections present some guidelines for operation, as well as some common
operating procedures.
Warm-Up and Environmental Considerations
Operate the LDT-5500B Series Temperature Controller at an ambient temperature in the range
of 0 to +40°C. Storage temperatures should be in the range of -40 to +70°C. To achieve rated
accuracy, let the LDT-5500B Series Temperature Controller warm up for about 1 hour before
use.
Temperature Mode Operation
The LDT-5500B Series Temperature Controller can be operated in several modes, constant
current (ITE), constant thermistor resistance (R), or constant temperature (T). This example is
for constant temperature (T) mode.
a. Plug the LDT-5500B Series Temperature Controller into an AC power source supplying
the correct voltage and frequency for the unit under operation (refer to the back panel for
the correct ratings).
b. Turn on the LDT-5500B Series Temperature Controller. The OUTPUT stage will be
disabled at power up and the unit will automatically configure its parameters to the state
which existed when the power was last shut off.
c. If the ENABLE indicator on the front panel is not lit, press the ENABLE switch so the
indicator is lit (adjustment enabled). Press the mode SELECT switch until T mode is
selected.
d. Check the setting of the SENSOR SELECT switch for the desired operation (10A or
100A). The sensor code will be displayed for two seconds during the power-up
sequence.
e. Press the parameter SELECT switch and the values of I LIMIT, T LIMIT, GAIN, and
CONST to ensure that they are adequate for the thermal load connected to the
controller.
f. Use the ADJUST knob to change the temperature set point until the desired value is
displayed.
Note: In some cases, a temperature set point resolution greater than 0.1
be attained by using R mode with the appropriate resistance value.
g. Enable the TEC output by pressing the output on switch. The unit will automatically
control the temperature to the set point.
h. When the unit is powered off, the state of the unit at power-down is saved in non-volatile
memory.
o
C may
November 2017 20 70019907 LDT-5500B Manual
Page 23
Resistance Mode Operation
a. Plug the LDT-5500B Series Temperature Controller into an AC power source supplying
the correct voltage and frequency for the unit under operation (refer to the back panel for
the correct ratings).
b. Turn on the LDT-5500B Series Temperature Controller. The OUTPUT stage will be
disabled at power up and the unit will automatically configure its parameters to the state
which existed when the power was last shut off.
c. If the ENABLE indicator on the front panel is not lit, press the ENABLE switch so the
indicator is lit (adjustment enabled). Press the mode SELECT switch until R mode is
selected.
d. Check the setting of the SENSOR SELECT switch for the desired operation (10A or
100A). The sensor code will be displayed for two seconds during the power-up
sequence.
e. Press the parameter SELECT switch and check the values of I LIMIT, T LIMIT, GAIN,
and CONST to ensure that they are adequate for the thermal load connected to the
controller.
f. Use the ADJUST knob to change the resistance set point until the desired value is
displayed.
g. Enable the TEC output by pressing the OUTPUT ON switch. The unit will automatically
control the thermal load to the set point resistance.
If the exact resistance is unknown (to control to a desired temperature), press the
DISPLAY switch to view the measured temperature. Readjust the resistance set point
and recheck the temperature until the desired result is attained.
h. When the unit is powered off, the state of the unit at power-down is saved in non-volatile
memory.
November 2017 21 70019907 LDT-5500B Manual
Page 24
November 2017 22 70019907 LDT-5500B Manual
Page 25
Chapter 3 :
Remote Operation
This chapter details the fundamentals of operating the LDT-5500B Temperature Controller
through the USB interface.
USB Driver Installation
Command Syntax
To begin using the USB port on the LDT-5500B Series Temperature Controller, please install
the Virtual Com Port Driver which is available on the companion CD or from our website.
A standard USB A/B cable is necessary to connect the instrument to a PC and is included in the
shipping kit. Once connected, the ILX Virtual COM Port driver provides a simple programmable
interface to the USB port. The LDT-5500B USB Front Panel software is a simple control
software package which allows full control of the instrument.
COM Port Settings
-
Baud Rate: 115,200
-
Data Bits: 8
-
Parity: None
-
Stop Bits: 1
-
Flow Control: None
USB Driver Installation
Insert the companion USB Flash Drive/CD into the PC. This USB Flash Drive/CD is included
with the LDT-5500B Series Temperature controller shipment. The ILX Virtual COM Port Installer
should run automatically but if it does not, the executable can be found at [Drive]:\Driver
Installation.bat. Follow the instructions of the installer to complete the installation.
Once installation is complete, plug the instrument into the computer with the USB cable. If the
Found New Hardware Wizard appears, select “Automatically Install Recommended Software”,
click NEXT, and wait for the installation to complete.
Command Syntax
This section details the syntax of the commands as expected by the LDT-5500B.
Letters
Any remote command or query must contain all of the letters of the command. The LDT-5500B
does not distinguish between capital and lower case letters.
White Space
Spaces or white space may be used to separate data but may not be used as a separation
between command and query question mark. Spaces or white space must be used to separate
the command from the first parameter.
November 2017 23 70019907 LDT-5500B Manual
Page 26
Command Termination
Each USB command or query must always be terminated with the new line character: '\n' or
0x0A.
Boolean Parameter Values
Boolean values (0 and 1) are used to represent On/Off or True/False.
Command Timing and Completion
All commands are executed in a sequential manner.
Error Messages
Error messages are reported using the available commands to query for individual errors such as
TEMPERATURE LIMIT, SENSOR OPEN, TE CURRENT LIMIT, and TE OPEN. An error query is
available to indicate that the output has turned off due to an output short condition or low
resistive load. A response of 0 (false) indicates that there is no error. A response of 1 (true)
indicates that an error is currently present, or the output was previously disabled due to this error
such as a TE OPEN error.
November 2017 24 70019907 LDT-5500B Manual
Page 27
p
Chapter 4 :
Command Reference
This chapter is a guide to all the commands for the LDT-5500B Series Temperature Controller.
Overview of Remote Commands
List of Commands in Alphabetical Order
Table 4.1 shows the format for the device command descriptions in this chapter. The
commands that emulate local (front panel) operation are denoted by “FRONT PANEL” in bold
text in the upper right hand corner of the command description.
NONE Returns the measured temperature sensor value
NONE Returns the position of the SENSOR SELECT switch
NONE Returns the constant ITE (TEC current) set point
NONE
NONE Returns the constant T (temperature) set point
1
Used to set the TEC constant T (temperature) set point
NONE Returns the TEC measured temperature value
NONE
NONE Returns the model of the instrument (5525B or 5545B)
Sets the constant sensor resistance; only available in constant R mode
Returns the constant R (temperature sensor value) set point
Used to return the current status of the temperature limit error
November 2017 26 70019907 LDT-5500B Manual
Page 29
LDT -5500B Series De vice-Dependent Commands
The following pages contain a reference for the commands of the LDT-5500B Temperature
Controller. This reference contains useful information for both local and remote operation of the
LDT-5500B series.
In some references, parentheses are used to signify the labeled area for a particular switch or
LED indicator on the front panel.
For example, (TEC DISPLAY) SET refers to the switch labeled Set in the TEC DISPLAY area of
the front panel.
*IDN? FRONT PANELREMOTE
Description Requests the instrument to identify itself.
Parameters None.
Notes Returns a comma delimited standard format ASCII identification string,
from information stored in the instrument during manufacture.
*RSTFRONT PANELREMOTE
Description Performs a device reset.
Parameters None.
Notes Resets the device to factory default settings (as the unit was received).
CURRENT_LIMIT?FRONT PANELREMOTE
Description Returns the current status of the ITE current limit error.
Parameters None.
Notes This error will not disable the output. The ITE current will be clamped at
the ITE Limit set point value.
Examples Query: CURRENT_LIMIT?
Response: 0
ITE current output is not clamped at set ITE Current Limit.
FET_OVER_POWER?FRONT PANELREMOTE
Description Returns the current status of the Overpower Pass Elements error.
Parameters None.
Notes This error will disable the output. This error can indicate that the output
has been shorted, allowing too much power to be dissipated across the
pass elements. This shorting shutdown feature is built into firmware to
prevent any damage occurring to the instrument and any equipment
connected. This error can also occur if the resistance of the load is less
than 1.
Examples Query: FET_OVER_POWER?
Response: 0
Output not disabled due to a shorting or low resistive load.
November 2017 27 70019907 LDT-5500B Manual
Page 30
SENSOR_OPEN?FRONT PANELREMOTE
Description Returns the status of the sensor open error.
Parameters None.
Notes This error will disable the output and occurs if the connection for the
thermistor or linear temperature sensor is open.
Examples Query: SENSOR_OPEN?
Response: 0
Temperature sensing connections are connected.
TE_OPEN?FRONT PANELREMOTE
Description Returns the current status of the sensor open error.
Parameters None.
Notes This error will disable the output. The status of this error is reset when the
output is enabled.
Examples Query: TE_OPEN?
Response: 0
TE is connected correctly and the output is enabled.
TEC:C1 FRONT PANELREMOTE
TEC:C1?
Description Sets/returns the value of constant C
.
1
Parameters Value between 9.999 and -9.99.
Examples Write: TEC:C1 1.125
Action: Sets constant C
for presently selected sensor to 1.125.
1
Query: TEC:C1?
Response: 2.100
Constant C
is set to 2.100.
1
TEC:C2 FRONT PANELREMOTE
TEC:C2?
Description Sets/returns the value of constant C
.
2
Parameters Value between 9.999 and -9.99.
Examples Write: TEC:C2 1.125
Action: Sets constant C
for presently selected sensor to 1.125.
2
Query: TEC:C2?
Response: 2.100
Constant C
is set to 2.100.
2
November 2017 28 70019907 LDT-5500B Manual
Page 31
TEC:C3 FRONT PANELREMOTE
TEC:C3?
Description Sets/returns the value of constant C
.
3
Parameters Value between 9.999 and -9.99.
Notes When the LM335 or AD590 sensors are selected via the SENSOR
SELECT switch, this constant is not used.
Examples Write: TEC:C3 1.125
Action: Sets constant C
for presently selected sensor to 1.125.
3
Query: TEC:C3?
Response: 2.100
Constant C
is set to 2.100.
3
TEC:CAL:ITEFRONT PANELREMOTE
Description Starts remote TEC current user calibration mode.
Parameters None.
Notes Calibration mode can be disabled by sending the “TEC:OUT 1” command
or pressing the output on button.
Examples Write: TEC:CAL:ITE
Action: Sets the unit into remote TEC current calibration.
TEC:CAL:SENFRONT PANELREMOTE
Description Starts remote SENSOR user calibration mode. Calibration mode is
dependent upon which setting the SENSOR SELECT switch is set.
Parameters None.
Notes Calibration mode can be disabled by sending the “TEC:OUT 1” command
or pressing the output on button.
Examples SENSOR SELECT switch set to AD590.
Write: TEC:CAL:SEN
Action: Sets the unit into remote calibration mode for AD590.
TEC:CAL?FRONT PANELREMOTE
Description Determines if the LDT-5500B Series Temperature Controller is ready for
a value to be entered during the calibration cycle.
Parameters None.
Notes Query can be used to poll the LDT-5500B Series Temperature Controller
after the unit has entered into remote calibration mode. If the response is
1, the LDT-5500B is ready to receive a calibration value.
Examples Query: TEC:CAL?
Response: 1
Unit is ready for the user to enter in a measured value.
November 2017 29 70019907 LDT-5500B Manual
Page 32
TEC:GAIN FRONT PANELREMOTE
TEC:GAIN?
Description Sets/returns the value of the control loop gain.
Parameters 1, 3, 10, 30, 100, or 300.
Notes If the user enters a gain value which does not equal 1, 3, 10, 30, 100, or
300, the instrument will ignore the command.
Examples Write: TEC:GAIN 10
Action: Sets the TEC control loop gain to 10.
Query: TEC:GAIN?
Response: 300
TEC control loop gain is set to 300.
TEC:ITE FRONT PANEL REMOTE
TEC:ITE?
Description Sets/returns the TEC control current set point. TEC:ITE is also used to
enter the TEC current calibration value.
Parameters A value which represents the ITE set point current, in Amps.
Notes The TEC:ITE set point command is only used in constant ITE mode.
The TEC:ITE? query can be used in all modes of operation.
In ITE current calibration mode, the value represents the measured
current value in Amps.
Examples Write: TEC:ITE: 2.5
Action: Sets TEC current set point to 2.5 Amps in ITE mode.
Query: TEC:ITE?
Response: 1.23
TEC output current is 1.230 Amps.
TEC:LIM:ITE FRONT PANEL REMOTE
TEC:LIM:ITE?
Description Sets/returns the TE current limit value.
Parameters A value representing the limit value of the TE current, in Amps.
Notes ITE limit value is in effect for all modes of TEC operation. The ITE limit
will limit the TEC current in both positive and negative current directions.
Examples Write: TEC:LIM:ITE 3.5
Action: TEC current limit is set to 3.500 Amps.
Query: TEC:LIM:ITE?
Response: 4.0
TEC current limit is 4.000 Amps.
November 2017 30 70019907 LDT-5500B Manual
Page 33
TEC:LIM:THI FRONT PANEL REMOTE
TEC:LIM:THI?
Description Sets/returns the temperature limit value.
Parameters A value which represents the upper bound of the load temperature, in °C.
Notes Temperature limit value must be in the range -99.9°C to 199.9°C. This
value also becomes the maximum temperature set point. This value will
force the TEC output to be shut off if the temperature limit is reached.
Examples Write: TEC:LIM:THI 100.0
Action: Sets the temperature limit to 100.0°C.
Query: TEC:LIM:THI?
Response: 35.0
Temperature limit it set to 35°C.
TEC:MODE:ITE FRONT PANEL REMOTE
Description Sets constant TE current control mode.
Parameters None.
Notes This mode keeps the output current constant, regardless of load
temperature variations. Changing modes causes the output to be forced
off, and the new mode’s set point value will be displayed.
Examples Write: TEC:MODE:ITE
Action: Sets TEC controller for constant TEC current operation
reference operating mode.
TEC:MODE:R FRONT PANEL REMOTE
Description Sets constant thermistor resistance reference mode.
Parameters None.
Notes Since sensor resistance is a function of temperature, this mode also
controls the temperature of the thermal load connected to the TEC. This
mode bypasses the use of the constants for temperature calculation. This
mode also allows for finer control of temperature in cases where the
thermistor’s temperature / resistance profile is not known. Changing
modes cause the output to be forced off, and the new mode’s set point
value will be displayed.
Examples Write: TEC:MODE:R
Action: Sets the TEC controller for constant thermistor resistance
reference operating mode.
November 2017 31 70019907 LDT-5500B Manual
Page 34
TEC:MODE:T FRONT PANEL REMOTE
Description Sets constant temperature control mode.
Parameters None.
Notes Since the load temperature is derived from sensor resistance, constant R
and T modes are related. In T mode, the set point is converted to a
resistance for reference using the appropriate constants and conversion
model. Changing modes causes the output to be forced off, and the new
mode’s set point value will be displayed.
Examples Write: TEC:MODE:T
Action: Sets the TEC controller for constant load temperature
operating mode.
TEC:MODE? FRONT PANEL REMOTE
Description Returns the selected TEC control mode.
Parameters None.
Notes The TEC mode is also the parameter which is controlled. The TEC output
is kept at the set point..
Examples Query: TEC:MODE?
Response: R
TEC is in constant (thermistor) resistance mode.
TEC:OUT FRONT PANEL REMOTE
TEC:OUT?
Description Sets/returns the status of the TEC output.
Parameters 1 = on, 0 = off
Notes After the output is enabled, it may be useful to wait until the output is
stable (within tolerance) before performing further operations, but it is not
necessary.
Examples Write: TEC:OUT 1
Action: Enables the TEC output
Query: TEC:OUT?
Response: 1
TEC output is enabled.
TEC:R FRONT PANEL REMOTE
TEC:R?
Command Sets/returns the constant thermistor resistance set point.
Parameters A value which represents the thermistor resistance set point value, in k.
Notes The R set point is used to control the TEC output in R mode only. TEC
load temperature is derived from the thermistor resistance.
Examples Write: TEC:R 25.2
Action: Sets the set point thermistor resistance to 25.2 k.
Query: TEC:R?
Response: 10.534
Measured TEC thermistor resistance is 10.534 k.
November 2017 32 70019907 LDT-5500B Manual
Page 35
TEC:SEN? FRONT PANEL REMOTE
Description Returns the position of the SENSOR SELECT switch.
1 = 100A thermistor; 2 = 10A thermistor; 3 = LM335; 4 = AD590.
Parameters None.
Notes The sensor code is displayed on the TEC display whenever the back
panel SENSOR SELECT switch position is changed. The sensor
selection must be made locally at the back panel SENSOR SELECT
switch. If the response is 0, the sensor type is undetermined and a
hardware error exists. TEC load temperature is derived from the
thermistor resistance.
Examples Query: TEC:SENS?
Response: 1
SENSOR SELECT switch is set to 100A thermistor.
TEC:SET:ITE? FRONT PANEL REMOTE
Description Returns the constant TE current set point value.
Parameters None.
Notes The TEC output is controlled to this set point value only when the TEC is
in constant ITE mode.
Examples Query: TEC:SET:ITE?
Response: 3.0
ITE set point is 3.000 Amps.
TEC:SET:R? FRONT PANEL REMOTE
Description Returns the constant (thermistor) resistance set point value.
Parameters None.
Notes The TEC output is controlled to this set point value only when the TEC is
in constant R mode.
Examples Query: TEC:SET:R?
Response: 3.4
R set point is 3.400k.
TEC:SET:T? FRONT PANEL REMOTE
Description Returns the constant temperature set point value, in °C.
Parameters None.
Notes The TEC output is controlled to this set point value only when the TEC is
in constant T mode.
Examples Query: TEC:SET:T?
Response: 22.0
Temperature set point is 22.0°C.
November 2017 33 70019907 LDT-5500B Manual
Page 36
TEC:T FRONT PANEL REMOTE
TEC:T?
Description Sets/returns the TECs constant temperature set point.
Parameters A value which represents the temperature set point, in °C.
Notes The TEC temperature is controlled to this set point only when the TEC is
operated in T mode. If the maximum temperature limit is set (see
TEC:LIM:T), this limit becomes the maximum set point value as well.
Examples Write: TEC:T 15.0
Action: Sets the TEC temperature set point to 15.0°C.
Query: TEC:T?
Response: 29.2
Measured load temperature is 29.2°C.
TEMP_LIMIT? FRONT PANEL REMOTE
Description Returns the current status of the temperature limit error.
Parameters None.
Notes This error is always updated. This means if the output was disabled
because of a temperature limit, the user may not know if this query is
performed AFTER the actual temperature is below the temperature limit
set point.
Examples Query: TEMP_LIMIT?
Response: 0
Measured temperature is below the set temperature limit.
MODEL? FRONT PANEL REMOTE
Description Returns the instrument model number.
Response Returns an ASCII identification string of the model number.
Examples Query: MODEL?
Response: 5545B
Instrument is an LDT-5545B.
November 2017 34 70019907 LDT-5500B Manual
Page 37
Chapter 5 :
Maintenance and Troubleshooting
This chapter describes how to maintain and troubleshoot the LDT-5500B Series Temperature
Controller.
Calibration
Disassembly
Troubleshooting
The service procedures described in this chapter are for use by qualified personnel.
Potentially lethal voltages exist within the LDT-5500B Series Temperature Controller.
To avoid electric shock, the user should not perform any of the procedures described in
this chapter unless qualified to do so.
High voltages are present on and around the printed circuit boards of the LDT-5500B Series
Temperature Controller.
Calibration Overview
The LDT-5500B Series Temperature Controller should be calibrated every 12 months or
whenever performance verification indicates that calibration is necessary.
All calibrations can be done with the case closed. The instrument is calibrated by changing the
internally stored digital calibration constants.
November 2017 35 70019907 LDT-5500B Manual
Page 38
Recommended Equipment
Recommended test equipment for calibrating the LDT-5500B Series Temperature Controller is
listed in Table 5.1. Equipment other than that shown in the table may be used if the
specifications meet or exceed those listed.
Table 5.1 Recommended Calibration Equipment List
DESCRIPTION MFG / MODEL SPECIFICATION
DMM HP 3457A DC Amps (@ 1.0A): ±0.02%
Resistance (@ 10): 0.02%
0.1 A or 0.1 mW resolution
Resistors Metal Film
High Power
15 k (for ITE calibration)
4 k and 40 k (for 100 A calibration)
4 k and 400 k (for 10 A calibration)
4 k and 10 k (for LM335 calibration)
10 k and 20 k (for AD590 calibration)
1, 20W, low TCR (for 5525B ITE calibration)
1.5, 50W, low TCR (for 5545B ITE calibration)
Environmental Conditions
Calibrate this instrument under laboratory conditions. Recommended calibration temperature is
at 23°C ± 1.0°C. When necessary, however, the LDT-5500B Series Temperature Controller may
be calibrated at its intended use temperature if this is within the specified operating temperature
range of 0 to 40°C.
Warm Up
The LDT-5500B Series Temperature Controller should be allowed to warm up for at least 1 hour
before calibration.
November 2017 36 70019907 LDT-5500B Manual
Page 39
Calibration Adjustments
There are two calibration adjustments that need to be made for the LDT-5500B Series
Temperature Controller. They are calibration of sensor measurement and calibration of the ITE
current measurement and limit circuits.
If a problem arises during calibration which prevents its normal completion, the calibration may
be aborted with no ill effects by simply pressing the OUTPUT switch. This is possible because
the calibration values are not saved to non-volatile memory until the last step of each calibration
procedure.
Thermistor Calibration
The following procedure is for calibrating the 100A and 10A constant current sources so that
the thermistor resistance measurements for these ranges will be accurate. This procedure
calibrates the resistance measurements of the thermistor. This procedure does not calculate
C1, C2, and C3. For information on calibrating the thermistor sensor, see Appendix A.
Calibration may be aborted by pressing the OUTPUT switch.
a. Set the SENSOR SELECT switch (back panel) to the 100 A position. Set C1 to 0.99,
C2 to 2.57, and C3 to 0.855.
b. Measure and record the exact resistance of the 4k, 40k, and 400k metal film
resistors. A 4-point probe resistance measurement is recommended.
c. Connect the 4k (for 100A) or 40k(for 10A) metal film resistor to the sensor input of
the LDT-5500B Series Temperature Controller (pins 7 and 8).
d. Enter the sensor calibration mode by pushing the parameter SET and display SELECT
buttons at the same time. After this, the display will indicate the sensor resistance in k
Allow the measurement to settle for about three seconds.
e. Press and hold in the ENABLE button and turn the ADJUST knob until the display
indicates the same resistance recorded for the 4k metal film resistor.
f. Release the ENABLE switch and wait for the 7-segment display to flash off and then
back on. Replace the 4k resistor with the 40k metal film resistor (for 100A) or 400k
metal film resistor (for 10A). After three seconds, repeat Step e with this resistor.
Once the final sensor measurement is entered, the display will go blank. During this
step, the sensor temperature set point calibration is taking place internally. Allow the
instrument to self-calibrate. Once completed, the calibration constants will be stored to
non-volatile memory, and the display will return to its previous state.
g. Switch the SENSOR SELECT switch to the 10A position and repeat steps c - f.
h. After calibration, the I LIMIT will be automatically set to 2.00 Amps. Reset the I LIMIT to
the desired value.
November 2017 37 70019907 LDT-5500B Manual
Page 40
Remote Thermistor Calibration
a. Set the SENSOR SELECT switch (back panel) to the 100A position. Set C
to 2.57, and C
to 0.855.
3
b. Measure and record the exact resistance of the 4k, 40k, and 400k metal film
resistors. A 4-point probe resistance measurement is recommended.
c. Connect the 4k (for 100A) or 40k (for 10A) metal film resistor to the sensor input of
the LDT-5500B Series Temperature Controller (pins 7 and 8).
d. Enter the sensor calibration mode by sending the command TEC:CAL:SEN. After this,
the display will indicate the sensor resistance in kΩ. Allow the measurement to settle for
about three seconds.
e. Send the command “TEC:R VALUE”, where VALUE indicates the resistance in k
measured in Step b for the 4k resistor.
f. Wait for the 7-segment display to flash off and then back on. Replace the 4k resistor
with the 40k metal film resistor (for 100A) or 400k metal film resistor (for 10A).
After three seconds, repeat Step e with this resistor.
Once the final sensor measurement is entered, the display will go blank. During this
step, the sensor temperature set point calibration is taking place internally. Allow the
instrument to self-calibrate. Once completed, the calibration constants will be stored to
non-volatile memory and the display will return to its previous state.
g. Switch the SENSOR SELECT switch to the 10A position and repeat steps c - f.
h. After calibration, the I LIMIT will be automatically set to 2.00 Amps. Reset the I LIMIT to
the desired value.
to 0.99, C2
1
November 2017 38 70019907 LDT-5500B Manual
Page 41
AD590 Sensor Calibration
The following procedure is for calibrating the AD590 sensor measurement so that the
temperature measurement will be accurate. This procedure calibrates the current measurement
of the AD590. This procedure does not calibrate C
AD590 sensor, see Appendix C.
Calibration may be aborted by pressing the OUTPUT switch.
a. Set the SENSOR SELECT switch (back panel) to the AD590 position. Set C
to 1.00.
b. Connect a precision 20k metal film resistor and a precision ammeter in series at the
sensor input of the LDT-5500B Series Temperature Controller (pins 7 and 8).
c. Enter the sensor calibration mode by pushing the display SELECT and parameter SET
buttons at the same time. After this, the TEC display will indicate sensor reference
current in A. Wait for three seconds for the measurement to settle. The k enunciator
is used to indicate sensor calibration.
d. Press and hold in the ENABLE button and turn the ADJUST knob until the display
indicates the same current as shown on the precision ammeter.
e. Release the ENABLE button and wait for the 7-segment display to flash off and then
back on. Replace the 20k resistor with a 10k metal film resistor. Wait for three
seconds and repeat Step d using the 10k resistor.
Once the final sensor measurement is entered, the display will go blank. During this
step, the sensor temperature set point calibration is taking place internally. Allow the
instrument to self-calibrate. Once completed, the calibration constants will be stored to
non-volatile memory, and the display will return to its previous state.
f. After calibration, I LIMIT will be automatically set to 2.00 Amps. Reset the I LIMIT to the
desired value.
and C2. For information on calibrating the
1
to 0.00, C2
1
November 2017 39 70019907 LDT-5500B Manual
Page 42
Remote AD590 Sensor Calibration
a. Set the SENSOR SELECT switch (back panel) to the AD590 position. Set C
to 0.00, C2 to
1
1.00.
b. Connect a precision 20k metal film resistor and a precision ammeter in series at the sensor
input of the LDT-5500B Series Temperature Controller (pins 7 and 8).
c. Enter the sensor calibration mode by sending the command “TEC:CAL:SEN”. After this, the
TEC display will indicate sensor reference current in A. Wait for three seconds for the
measurement to settle.
d. Send the command “TEC:R VALUE”, where VALUE indicates in A, the current measured
by the precision ammeter.
e. Wait for the 7-segment display to flash off and then back on. Replace the 20k resistor with a
10k metal film resistor. Wait for three seconds, then repeat Step d using the 10kresistor.
Once the final sensor measurement is entered, the display will go blank. During this step, the
sensor temperature set point calibration is taking place internally. Allow the instrument to selfcalibrate. Once completed, the calibration constants will be stored to non-volatile memory,
and the display will return to its previous state.
f. After calibration, I LIMIT will be automatically set to 2.00 Amps. Reset the I LIMIT to the
desired value.
November 2017 40 70019907 LDT-5500B Manual
Page 43
LM335 Sensor Calibration
The following procedure is for calibrating the LM335 sensor measurement so that the
temperature measurement will be accurate. This procedure calibrates the voltage measurement
of the LM335. This procedure does not calibrate C
LM335 sensor, see Appendix C.
Calibration may be aborted by pressing the OUTPUT switch.
a. Set the SENSOR SELECT switch (back panel) to the LM335 position. Set C
to 1.00.
b. Connect a precision 4k metal film resistor and a precision voltmeter in parallel at the
sensor input of the LDT-5500B Series Temperature Controller (pins 7 and 8).
c. Enter the sensor calibration mode by pushing the display SELECT and parameter SET
buttons at the same time. After this, the display will indicate sensor reference voltage in
100’s of mV. Wait for three seconds for the measurement to settle. The k enunciator is
used to indicate sensor calibration.
d. Press and hold in the ENABLE button and turn the ADJUST knob until the display
indicates the same voltage as shown on the precision voltmeter multiplied by 10. For
example, if the voltage across the resistor is 1.9871 Volts, turn the ADJUST knob until the
display reads 19.87.
and C2. For information on calibrating the
1
to 0.00, C2
1
e. Release the ENABLE button and wait for the 7-segment display to flash off and then back
on. Replace the 4k resistor with a 10k metal film resistor. After three seconds, repeat
Step d with the 10k resistor.
Once the final sensor measurement is entered, the display will go blank. During this step,
the sensor temperature set point calibration is taking place internally. Allow the instrument
to self-calibrate. Once completed, the calibration constants will be stored to non-volatile
memory, and the display will return to its previous state.
f. After calibration, I LIMIT will be automatically set to 2.00 Amps. Reset the I LIMIT to the
desired value.
November 2017 41 70019907 LDT-5500B Manual
Page 44
Remote LM335 Sensor Calibration
a. Set the SENSOR SELECT switch (back panel) to the LM335 position. Set C
to 0.00, C2 to
1
1.00.
b. Connect a precision 4k metal film resistor and a precision voltmeter in parallel at the sensor
input of the LDT-5500B Series Temperature Controller (pins 7 and 8).
c. Enter the sensor calibration mode by sending the command “TEC:CAL:SEN”. After this, the
TEC display will indicate sensor reference current in mV. Wait for three seconds for the
measurement to settle.
d. Send the command “TEC:R VALUE”, where VALUE indicates the value measured by the
precision voltmeter multiplied by 10. For example, if the voltage across the resistor is 1.9871
Volts, send the command “TEC:R 19.87”.
e. Wait for the 7-segment display to flash off and then back on. Replace to 20k resistor with a
10 k metal film resistor. Wait for three seconds, then repeat step d using the 10k resistor.
Once the final sensor measurement is entered, the display will go blank. During this step, the
sensor temperature set point calibration is taking place internally. Allow the instrument to selfcalibrate. Once completed, the calibration constants will be stored to non-volatile memory,
and the display will return to its previous state.
f. After calibration, I LIMIT will be automatically set to 2.00 Amps. Reset the I LIMIT to the
desired value.
November 2017 42 70019907 LDT-5500B Manual
Page 45
ITE Current Calibration
The following procedure is for calibrating the ITE constant current source for both polarities of
current. During this procedure the ITE current is driven to a series of pre-determined values.
When each of these values is reached and is stable, the user enters the actual value of the
current, as measured by an external DMM. The LDT-5500B Series Temperature Controller then
automatically calibrates the TEC current source and limits.
Calibration may be aborted by pressing the OUTPUT switch.
a. For the LDT-5525B, connect a 1Ω, 20W resistor across the TEC output terminals (pins 1 and
3). For the LDT-5545B, connect a 1.5Ω, 50W resistor. Use a calibrated DMM to measure the
voltage across the resistor. Calculate the current in the following steps by using Ohm's Law:
=
where V is the accurately measured voltage across the resistor, and R is the accurately
measured load resistance. A 4-point probe resistance measurement is recommended.
b. Enter the calibration mode by pushing the display SELECT and mode SELECT buttons. This
will put the LDT-5500B Series into ITE calibration mode. Wait for three seconds for the output
to settle to about 3 Amps (LDT-5525B); 4 Amps (LDT-5545B).
c. Press and hold in the ENABLE button and turn the ADJUST knob until the display shows the
correct value (absolute value of the ITE measurement, as calculated from Step a.)
d. Release the ENABLE button. Wait three seconds to allow the ITE current to settle at the new
set point.
e. Repeat steps c and d for all six set points, once for each of the (automatically adjusted) set
points. For the LDT-5525B, the set points are: 3A, -3A, 3A, 1A, -1A, and -3A. For the LDT5545B, the set points are: 4A, -4A, 4A, 1A, -1A, and -4A.
After the last set point is entered, the LDT-5500B Series Temperature Controller’s front panel
will go blank while the instrument calibrates the ITE set point and current limits. Allow the
instrument to self-calibrate. Once completed, the calibration constants will be stored to nonvolatile memory, and the display will return to its previous state.
f. After calibration, the I LIMIT will be automatically set to 2.00 Amps. Reset the I LIMIT to the
desired value.
November 2017 43 70019907 LDT-5500B Manual
Page 46
Remote ITE Current Calibration
a. For the LDT-5525B, connect a 1, 20W resistor across the TEC output terminals (pins 1
and 3). For the LDT-5545B, connect a 1.5, 50W resistor. Use a calibrated DMM to
measure the voltage across the resistor. Calculate the current in the following steps by
using Ohm's Law:
where V is the accurately measured voltage across the resistor, and R is the accurately
measured load resistance. A 4-point probe resistance measurement is recommended.
b. Enter into remote ITE current calibration mode by sending the command “TEC:CAL:ITE”.
After this, the TEC display will indicate ITE current in Amps.
c. Remotely send the current value measured and calculated using step a using the
command TEC:ITE <value>
d. Wait three seconds to allow the ITE current to settle at the new set point.
=
e. Repeat Steps c and d for all six set points; once for each of the (automatically set) set
points. For the LDT-5525B, the set points are 3A, -3A, 3A, 1A, -1A and -3A. For the LDT5545B, the set points are 4A, -4A, 4A, 1A, -1A and -4A.
After the last set point is entered, the LDT-5500B Series Temperature Controller’s front
panel will go blank while the instrument calibrates the ITE set point and current limits.
Allow the instrument to self-calibrate. Once completed, the calibration constants will be
stored in non-volatile memory, and the display will return to its previous state.
f. After calibration, the I LIMIT will be automatically set to 2.00 Amps. Reset the I LIMIT to
the desired state.
November 2017 44 70019907 LDT-5500B Manual
Page 47
Troubleshooting
This section is a guide to troubleshooting the LDT-5500B Series Temperature Controller. Some
of the more common symptoms are listed here, and the appropriate troubleshooting actions are
given. It is recommended that the user start at the beginning of this guide. Read the symptom
descriptions, and follow the steps for the corrective actions which apply. If problems are
encountered which are beyond the scope of this guide, contact an ILX Lightwave
representative.
Table 5.2 Troubleshooting Reference List
Symptom Causes and Corrective Actions
LDT-5500B Series unit will not
power up
Power on, but display is frozen
and switches don’t work
Power on, but no TE current
output
Power on, but measured ITE
current is always about 0.0A
Power on, but temperature is
not controlled
Unable to adjust output or
parameter
Unable to switch DISPLAY,
MODE, SENSOR CAL or
PARAMETER modes
Output goes off intermittently
R Mode set point is not saved
Check AC power line voltage and power cord connections.
This may occur if the unit loses power (AC line) briefly. Turn the
power switch off and on again to restart.
If TE OPEN indicator is lit, check the load connections and then
try again. If SENSOR OPEN or TEMP LIMIT indicators are lit,
check the temperature sensor connections and then try again.
Check to ensure that the I LIMIT parameter is not set to 0.0A.
If SENSOR OPEN indicator is lit, check the sensor connections
and then try again.
Check that the back panel SENSOR SWITCH position is set to the
proper sensor type.
Check that the sensor’s temperature calibration constants (C1,
C2, and C3) are correct values for the sensor type.
Check that the GAIN setting is not too low and that the I LIMIT
value is not too low for the thermal load.
Check the adjust ENABLE switch; the indicator must be lit for any
adjustments to be made.
Check the MODE or DISPLAY switch; if they do not respond, the
unit may be in measurement calibration mode (see Chapter 3);
press the OUTPUT switch to abort this mode.
The unit may be in measurement calibration mode (see Chapter
3); press the OUTPUT switch to abort this mode.
Check that the AC power cord connection is secure; power line
drop outs may reset the unit and when the power is restored, the
output will be off.
Check the TE module connections; a high impedance on the TE
load may cause the output to exceed the compliance voltage
momentarily, thus shutting the output off.
The R mode set point value is not independent from the T mode
set point value; if the control mode is changed from R mode to T
mode, the R set value will change to a value which corresponds to
the temperature, based on C1, C2, and C3.
November 2017 45 70019907 LDT-5500B Manual
Page 48
November 2017 46 70019907 LDT-5500B Manual
Page 49
Appendix A:
Steinhart-Hart Equation
Reference the ILX Lightwave Application Note “Thermistor Calibration and the Steinhart-Hart
Equation” for information on this equation.
November 2017 47 70019907 LDT-5500B Manual
Page 50
November 2017 48 70019907 LDT-5500B Manual
Page 51
Appendix B:
Sensing Current and Thermistor
Selection
Choosing the right sensing current depends on the range of temperature that needs to be
measured and the resolution required at the highest measured temperature. To correctly select
the proper sensing current, two aspects must be understood: how the thermistor and the LDT5500B Series Temperature Controller interact, and how temperature range and resolution
values are inherent in the nature of thermistors.
Thermistor Range
Thermistors can span a wide temperature range, but their practical range is limited by their nonlinear resistance properties. At high temperatures, the thermistor resistance changes less for an
equivalent temperature change at lower temperatures (the thermistor becomes less sensitive).
Consider the temperature and sensitivity figures in Table B.1 below for a 10kΩ thermistor.
Table B.1 Thermistor Sensitivity
TEMPERATURE SENSITIVITY
-20°C 5600Ω / °C
25°C 439Ω / °C
50°C 137Ω / °C
In the LDT-5500B Series Temperature Controller, the practical upper temperature limit is the
temperature at which the thermistor becomes insensitive to temperature changes. The lower
end of the temperature range is limited by the maximum input voltage of the LDT-5500B Series
Temperature Controller.
Thermistor resistance and voltage are related through Ohms Law (V = IR). The LDT-5500B
Series Temperature Controller supplies current to the thermistor, either 10A or 100A. As the
thermistor resistance changes, a changing voltage signal is available to the thermistor inputs of
the LDT-5500B Series. The LDT- 5500B Series' measurement system will over-range when the
input voltage exceeds about 4.5 volts. Figure B.1 graphically shows the lower temperature and
upper voltage limits for a typical 10kΩ thermistor. (A 10kΩ thermistor has a resistance of 10k
at 25°C). The practical temperature ranges for a typical 10kΩ thermistor with the LDT-5500B
Series are given in Table B.2, below. These temperature ranges may vary from thermistor to
thermistor, even though both thermistors are nominally 10kΩ. This is due to manufacturing
tolerances in the thermistor, and is compensated for by determining C
the thermistor). The practical temperature ranges for a 10kΩ thermistor are also shown as solid
bars at the bottom of Figure B.1.
, C2, and C3 (calibrating
1
November 2017 49 70019907 LDT-5500B Manual
Page 52
Table B.2 10kΩ Thermistor Temperature Range
SENSING CURRENT TEMPERATURE RANGE
10A
100A
-30 to 30
10 to 70°C
°
C
C THERMISTOR)
THERMISTOR
VOLTAGE
5.00
4.50
4.00
3.00
2.00
1.00
(USING TYPICAL* 10 k @ 25
0.00
-60
* ILX default values for C1 and C2
-40
-20
0
20
DEGREES C
- Denotes practical range with typical 10 K thermistor
- Denotes measurable range with t ypical 10K thermistor
60 80 100
40
10 A
100 A
Figure B.1Thermistor Temperature Range
Temperature Resolution
Measurement resolution must also be considered since the measurement resolution decreases
as the thermistor temperature increases. A temperature controller (such as the LDT-5500B
Series) has a limited measurement resolution. A temperature change of one degree centigrade
will be represented by a greater resistance increase at a lower temperature than at a higher
temperature because of the non-linear resistance of the thermistor. Resolution figures for a
typical 10kΩ thermistor are given in Table B.3, below.
Table B.310kΩ Thermistor Voltage vs. Resolution
TEMPERATURE
-20°C 56mV /
25°C 4.4mV /
50°C 1.4mV /
VOLTAGE AT 10 A
°
C 0.018°C / mV
°
C 0.23°C / mV
°
C 0.70°C / mV
For a typical 10kΩ thermistor, a temperature change from -20°C to -19°C will be represented by
a measurement change of about 56mV (if supplied with 10A). The same thermistor
measurement will only change about 1.4mV from 49 to 50°C. For that case, with the LDT-5500B
Series, the temperature measurement resolution would be reduced to about 0.2°C. If the 100A
setting were used instead, the thermistor measurement would change by 14mV from 49 to
50°C, providing the maximum resolution of 0.1°C (with the LDT-5500B Series).
RESOLUTION
November 2017 50 70019907 LDT-5500B Manual
Page 53
Therefore, the sensor current chosen may impact the temperature measurement resolution as
well as the set point control accuracy.
Selecting the Sensing Current
To select the current setting for a typical 10kΩ thermistor, determine the lowest temperature
needed to sample and set the SENSOR SELECT switch according to the range limits in Table
B.2. If the temperature of interest is below 10°C, the switch will likely need to be set to the 10A
setting.
If the required temperatures are between 10°C and 30°C, either SENSOR SELECT setting
(100A or 10A) will work with a 10kΩ thermistor. However, the 100A setting provides greater
measurement resolution, and therefore better control.
Note: Generally, it is best to use the 100A SENSOR SELECT setting for all
measurements of 10°C or greater with a typical 10kΩ thermistor.
Selecting and Using Thermistors
The type of thermistor chosen will depend primarily on the operating temperature range. These
guidelines for selecting the range and resolution will apply to any thermistor. From Figure B.1 it
can be seen that 10kΩ thermistors are generally a good choice for most laser diode applications
where high stability is required near room temperatures. Similarly, 10kΩ thermistors are often a
good choice for detector cooling applications where operating temperatures are from -30°C to
room temperature.
If a different temperature range is required or the required accuracy can't be achieved with
either switch setting, select another thermistor. Thermistor temperature curves, supplied by the
manufacture, show the resistance verses temperature range for many other thermistors. ILX
Lightwave Corporation will also offer help for specific applications.
November 2017 51 70019907 LDT-5500B Manual
Page 54
November 2017 52 70019907 LDT-5500B Manual
Page 55
Appendix C:
AD590 and LM335 Sensor Calibration
The LDT-5500B Series Temperature Controller uses two constants (C1 and C2) for calibrating
linear thermal sensing devices, such as the AD590, and the LM335. C1 is used as the linear or
zero offset value, and C
a nominal value of 0, and C
switch is in the AD590, or LM335 positions.
In order to calibrate a linear sensor device, the sensor must be operated at an accurately
known, stable temperature. For example, the sensor may be calibrated at 0
placed in ice water until its temperature is stable. A highly accurate temperature probe,
thermometer, environmental chamber, etc., may also be used to determine the known
temperature for calibration. This appendix contains one and two point calibration methods for
linear sensor devices. These methods will work for either type of device.
is used as the slope or gain adjustment. Therefore, C1 should be set to
2
should be set to a nominal value of 1, when the SENSOR SELECT
2
o
C if the sensor is
November 2017 53 70019907 LDT-5500B Manual
Page 56
AD590 Sensor
The AD590 is a linear thermal sensor which acts as a constant current source. It produces a
current, i, which is directly proportional to absolute temperature, over its useful range (-50°C to
+150°C). This nominal value can be expressed as:
1μ
=
where i is the nominal current produced by the AD590, and K is in Kelvin.
The LDT-5500B Series Temperature Controller uses i to determine the nominal temperature, T
by the formula:
=
1μ
where T
is in °C.
n
The temperature, T
which is displayed by the LDT-5500B Series Temperature Controller, is
d,
first calibrated as follows:
=+∗
where C
and C2 are the constants stored by the user in the LDT-5500B Series
1
Temperature Controller for the AD590.
The AD590 measurement is calibrated, at the factory, with C
The AD590 grades of tolerance vary, but typically this means that without adjusting C
temperature accuracy is ±1°C over its rated operating range. If C
the temperature accuracy is ±0.2°C over its rated operating range. However, the AD590 is not
perfectly linear, and even with C
accurately known there is a non-linear absolute temperature
1
error associated with the device. This non-linearity is shown in Figure C.1, reprinted from
Analog Devices specifications, where the error associated with C
− 273.15
= 1 and C1 = 0 (nominal values).
2
1
and C2 are also calibrated,
1
is assumed to be zero.
1
or C2, the
,
n
Figure C.1 AD590 Nonlinearity
If a maximum absolute error of 0.8°C is tolerable (over the entire temperature range), the one
point calibration of C
should be used (see page 54). If C1 is calibrated at 25°C, and the
1
intended operating range is 0 to 50°C, a maximum error of about ±0.2°C may be expected over
November 2017 54 70019907 LDT-5500B Manual
Page 57
that operating range. If a greater accuracy is desired, the two point method of determining C1
and C
therefore the constant C
should used (see page 49). Note however, the absolute error curve is non-linear;
2
will vary over different temperature ranges.
2
LM335 Sensor
The LM335 is a linear thermal sensor which acts as a constant voltage source. It produces a
voltage, V, which is directly proportional to absolute temperature, over its useful range (-40°C to
+100°C). This nominal value can be expressed as:
10
=
where V is the nominal voltage produced by the LM335 and K is in Kelvin.
The LDT-5500B Series Temperature Controller uses V to determine the nominal temperature,
T
, by the formula:
n
=
1
where T
is in °C.
n
The temperature, T
, which is displayed by the LDT-5500B Series Temperature Controller, is
d
first calibrated as follows:
=+∗
where C
and C2 are the constants stored by the user in the LDT-5500B Series
1
Temperature Controller for the LM335.
When the LDT-5500B Series is shipped from the factory, the LM335 measurement system is
calibrated, but the sensor (C
and C2) is not. Nominally, C1 = 0, and C2 = 1. In that case, the
1
temperature accuracy is typically ±1°C over the rated operating range. With C
calibrated also, the temperature accuracy is typically ±0.3°C over the rated operating range. The
temperature accuracy may be improved over a narrow temperature range by a two-point
calibration of C
known (and C
and C2. However, the LM335 is not perfectly linear, and even with C1 accurately
1
uncalibrated) there is a non-linear absolute temperature error associated with
2
the device. This non-linearity caused error is typically ±0.3°C, with the error associated with C
assumed to be zero.
If a maximum absolute error of ±1°C is tolerable, no calibration of C
C
= 0, C2 = 1. If a maximum absolute error of ±0.5°C is tolerable, the one point calibration of C1
1
may be used (see page 48). If a greater accuracy is desired, the two point method of
determining C
with the constant C
and C2 should used (see page 49). Note however, the absolute error associated
1
may vary over different temperature ranges.
2
− 273.15
and C2
1
1
or C2 is required, just set
1
November 2017 55 70019907 LDT-5500B Manual
Page 58
One Point Calibration Method
This procedure will work for any linear temperature sensor. The accuracy of this procedure
depends on the accuracy of the known temperature, externally measured. It is used to
determine the zero offset of the device, and it assumes that the gain offset (slope) is known and
is correct.
1. Allow the LDT-5500B Series Temperature Controller to warm up for at least one hour.
Set the SENSOR SELECT switch for the desired sensor type, and RECALL the
constants for the particular device to be calibrated.
2. Select the C
3. Place the sensor at an accurately known and stable temperature, Ta. Connect the
sensor to pins 7 and 8 of the LDT-5500B Series' 15-pin connector. Set the LDT-5500B
Series for normal constant temperature (T mode) operation. Allow the LDT-5500B Series
Temperature Controller to stabilize at the known temperature, T
temperature, T
4. Determine the new value of C
andreplaceC
parameter. Read and record the value of C1.
1
and read the displayed
a
.
d
, C1n, from the formula:
1
=+−
with C1n by selecting the C1 parameter and entering the new C1n value.
1
November 2017 56 70019907 LDT-5500B Manual
Page 59
Two Point Calibration Method
This procedure will work for any linear temperature sensor. The accuracy of this procedure
depends on the accuracy of the known temperatures, externally measured. It is used to
determine the zero offset of the device and the gain offset (slope).
1. Allow the LDT-5500B Series Temperature Controller to warm up for at least one hour.
Set the SENSOR SELECT switch for the desired sensor type, and RECALL the
constants for the particular device to be calibrated.
2. Select the C
Read and record the value of C
3. Place the sensor at an accurately known and stable temperature, T
sensor to pins 7 and 8 of the LDT-5500B Series's 15-pin connector. Set the LDT-5500B
Series for normal constant temperature (T mode) operation. Allow the LDT-5500B Series
Temperature Controller to stabilize at the known temperature, T
temperature, T
4. Repeat Step 3 for another known temperature, T
temperature, T
The two known temperatures should at the bounds of the intended operating range. The
smaller the intended operating range, the better the calibration over that same range.
5. Determine the new value of C
First determine the intermediate values U and V, where
Then C
parameter. Read and record the value of C1. Select the C2 parameter.
1
.
2
. Connect the
a1
and read the displayed
a1
. Record these values.
d1
, and the corresponding displayed
a2
.
d2
(C1n) and C2 (C2n) from the following calculations.
1
and C2n can be determined by the following:
1n
=
−
−
and
=
−∗
=+∗
and
=∗
6. Replace C
Replace C
with C1n by selecting the C1 parameter and entering the new C1n value.
1
with C2n by selecting the C2 parameter and entering the new C2n value.
2
November 2017 57 70019907 LDT-5500B Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.