Details about cautionary symbols
Safety markings used on the instrument
Information about the warranty
Customer service contact information
Safety Information and the Manual
Throughout this manual, you will see the words Caution and Warning indicating potentially
dangerous or hazardous situations which, if not avoided, could result in death, serious or minor injury,
or damage to the product. Specifically:
Caution indicates a potentially hazardous situation which can result in minor or moderate injury or
damage to the product or equipment.
Warning indicates a potentially dangerous situation which can result in serious injury or death.
Visible and/or invisible laser radiation. Avoid direct exposure to the beam.
General Safety Considerations
If any of the following conditions exist, or are even suspected, do not use the instrument until safe
operation can be verified by trained service personnel:
Visible damage
Severe transport stress
Prolonged storage under adverse conditions
Failure to perform intended measurements or functions
If necessary, return the instrument to ILX Lightwave, or authorized local ILX Lightwave distributor, for
service or repair to ensure that safety features are maintained.
All instruments returned to ILX Lightwave are required to have a Return Authorization Number
assigned by an official representative of ILX Lightwave Corporation. See Returning an Instrument for
more information.
March 2019 v LDT-5910C and LDT-5940C
Safety Symbols
This section describes the safety symbols and classifications.
Technical specifications including electrical ratings and weight are included within the manual. See
the Table of Contents to locate the specifications and other product information. The following
classifications are standard across all ILX Lightwave products:
Indoor use only
Ordinary Protection: This product is NOT protected against the harmful ingress of
moisture.
IEC Class I Equipment (grounded type)
Mains supply voltage fluctuations are not to exceed ±10% of the nominal supply voltage.
Pollution Degree II
Installation (overvoltage) Category II for transient over-voltages
Maximum Relative Humidity: <85% RH, non-condensing
Operating temperature range of 10 °C to 40 °C
Storage and transportation temperature of –40 °C to 70 °C
Maximum altitude: 3000 m (9843 ft.)
This equipment is suitable for continuous operation.
March 2019 vi LDT-5910C and LDT-5940C
Safety Marking Symbols
This section provides a description of the safety marking symbols that appear on the instrument.
These symbols provide information about potentially dangerous situations which can result in death,
injury, or damage to the instrument and other components.
Caution, refer
to manual
Caution, risk of
electric shock
Earth ground
Terminal
Protective
Conductor
Terminal
Alternating current
Caution,
hot surface
Visible and/or
invisible laser
radiation
Frame or chassis
terminal
Warranty
ILX Lightwave Corporation warrants this instrument to be free from defects in material and
workmanship for a period of one year from date of shipment. During the warranty period, ILX will
repair or replace the unit, at our option, without charge.
Limitations
This warranty does not apply to fuses, lamps, defects caused by abuse, modifications, or to use of
the product for which it was not intended.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty of
merchantability or fitness for any particular purpose. ILX Lightwave Corporation shall not be liable for
any incidental, special, or consequential damages.
If a problem occurs, please contact ILX Lightwave Corporation with the instrument's serial number,
and thoroughly describe the nature of the problem.
March 2019 vii LDT-5910C and LDT-5940C
Returning an Instrument
If an instrument is to be shipped to ILX Lightwave for repair or service, be sure to:
Obtain a Return Authorization number (RA) from ILX Customer Service.
Attach a tag to the instrument identifying the owner and indicating the required service or
repair. Include the instrument serial number from the rear panel of the instrument.
Attach the anti-static protective caps that were shipped with the instrument.
Place the instrument in the original packing container with at least 3 inches (7.5 cm) of
compressible packaging material. Shipping damage is not covered by this warranty.
Secure the packing box with fiber reinforced strapping tape or metal bands.
Send the instrument, transportation pre-paid, to ILX Lightwave. Clearly write the return
authorization number on the outside of the box and on the shipping paperwork. ILX
Lightwave recommends you insure the shipment.
If the original shipping container is not available, place your instrument in a container with at
least 3 inches (7.5 cm) of compressible packaging material on all sides.
Repairs are made and the instrument returned transportation pre-paid. Repairs are warranted for the
remainder of the original warranty or for 90 days, whichever is greater.
Claims for Shipping Damage
When you receive the instrument, inspect it immediately for any damage or shortages on the packing
list. If the instrument is damaged, file a claim with the carrier. The factory will supply you with a
quotation for estimated costs of repair. You must negotiate and settle with the carrier for the amount
of damage.
March 2019 viii LDT-5910C and LDT-5940C
Comments, Suggestions, and Problems
To ensure that you get the most out of your ILX Lightwave product, we ask that you direct any
product operation or service related questions or comments to ILX Lightwave Customer Support. You
may contact us in whatever way is most convenient:
Phone (800) 459-9459 or (406) 586-1244
Fax (406) 586-9405
On the web at: ilx.custhelp.com
Or mail to:
ILX Lightwave Corporation
31950 East Frontage Road
Bozeman, Montana, U.S.A 59715-8642
www.ilxlightwave.com
When you contact us, please have the following information:
Model Number
Serial Number
End-user Name
Company
Phone
Fax
Description of what is connected to the ILX Lightwave instrument
Description of the problem
If ILX Lightwave determines that a return to the factory is necessary, you are issued a Return
Authorization (RA) number. Please mark this number on the outside of the shipping box.
You or your shipping service are responsible for any shipping damage when returning the instrument
to ILX Lightwave; ILX recommends you insure the shipment. If the original shipping container is not
available, place your instrument in a container with at least 3 inches (7.5 cm) of compressible
packaging material on all sides.
We look forward to serving you even better in the future!
March 2019 ix LDT-5910C and LDT-5940C
March 2019 x LDT-5910C and LDT-5940C
Chapter 1:
Introduction and Specifications
This chapter is an introduction to theLDT-5910C and LDT-5940C Thermoelectric Temperature
Controllers.
Safety Considerations and unpacking information
Product Overview
Options and accessories
Specifications
Safety Considerations
If any of the following symptoms exist, or are even suspected, remove the LDT-5910C or
LDT-5940C from service. Do not use the LDT-5910C or LDT-5940C until trained service
personnel can verify safe operation.
Visible damage
Severe transport stress
Prolonged storage under adverse conditions
Failure to perform intended measurements or functions
If necessary, return the LDT-5910C or LDT-5940C to ILX Lightwave for service and repair
to ensure that safety features are maintained.
Initial Inspection
When you receive your LDT-5910C or LDT-5940C Thermoelectric Temperature Controller, verity that
the following items were shipped with the instrument
LDT-5900C Series Thermoelectric Temperature Controller Manual
Power Cord
Shipping Kit
Shipping Kit
The shipping kit for the LDT-5910C and LDT-5940C Thermoelectric Temperature Controller includes
a USB A/B cable and a CD containing the ILX Lightwave USB Driver.
March 2019 1 LDT-5910C and LDT-5940C
Product Overview
The LDT-5910C and LDT-5940C are high performance thermoelectric temperature controllers
that offers a flexible, extended temperature control range from -100 °C to +200 °C with a choice
of temperature sensors. The LDT-5910C features a linear, bipolar output providing 32 Watts of
power to drive most thermal loads. This output stage offers the low current noise necessary to
achieve the temperature stability required to meet demanding low noise and frequencystabilized laser applications. The LDT-5940C features a switching output providing 60 Watts of
temperature control power for applications requiring increased cooling or heating capacity. The
instruments support thermistor, platinum RTD, and IC temperature sensors. In addition, a
unique linearized thermistor mode allows for the use of a standard 10 k thermistor over an
extended temperature range from -15 °C to 65 °C. A digital PID control loop delivers fast
settling time while maintaining high stability. For quick setup the LDT-5910C and LDT-5940C
incorporate pre-programmed PID values for common applications. In addition, a fast auto-tune
algorithm is available to automatically adjust the PID constants to support a wide range of
thermal loads. The GPIB and USB interfaces ensure trouble-free remote programming and
readout.
Features of the LDT-5910C and LDT-5940C include:
Precision setpoint resolution of ±0.01 °C with long term temperature stability of ±0.002 °C
(LDT-5910C) and ±0.003 °C (LDT-5940C)
Unique, linearized thermistor mode allows a standard 10 k thermistor to be used over the
range from -15 °C to 65 °C
Pseudo 4-Wire Sensor and TEC Measurement (LDT-5940C only)
Fully programmable PID control loop with twelve preset PID values
Auto-tune algorithm automatically adjusts PID constants for a wide range of loads
External analog temperature control input
Independent heating and cooling current limits
TE voltage measurement
Heat only control mode
Remote commands in LDT-5910C and LDT-5940C are compatible with LDT-5910B remote
commands
GPIB/IEEE 488 and USB Interfaces
Normally open and normally closed output interlock
Variable speed external fan control
If cleaning is required, use a clean dry cloth. Do not use solvents.
March 2019 2 LDT-5910C and LDT-5940C
Figure 1.1 – LDT-5910C and LDT-5940C Front Panel
Figure 1.2 – LDT-5910C Rear Panel
March 2019 3 LDT-5910C and LDT-5940C
Figure 1.4 – LDT-5940C Rear Panel
March 2019 4 LDT-5910C and LDT-5940C
Options and Accessories
Options and accessories available for LDT-5910C and LDT-5940C Thermoelectric Temperature
Controllers include the following:
DESCRIPTION MODEL / PART NUMBER
Single Rack Mount Kit RM-144
Dual Rack Mount Kit RM-145
Temperature Controller Interconnect Cable (terminated with 9-pin DSUB) CC-505S [5910C only]
Temperature Controller Interconnect Cable (unterminated) CC-501S [5910C only]
10 Amp TE Interconnect Cable (unterminated without fan connections) CC-591H [5940C only]
10 Amp TE Interconnect Cable (unterminated with fan connections) CC-594H [5940C only]
5 Amp TE/LDM Interconnect Cable (terminated with 9-pin DSUB) CC-595S [5940C only]
10 Amp TE Interconnect Cable (terminated with 7W2 DSUB) CC-596H [5940C only]
Calibrated 10 k Thermistor TS-510
Uncalibrated 10 k Thermistor TS-520
Uncalibrated 5 k Thermistor TS-521
Uncalibrated 20 k Thermistor TS-523
Uncalibrated 100 k Thermistor TS-525
Uncalibrated AD590 IC Temperature Sensor TS-530
Uncalibrated LM335 IC Temperature Sensor TS-540
Unipolar Control Adapter UCA-350
TO-Can Laser Diode Mount LDM-4405
TO-Can Laser Diode Mount LDM-4407
C-Mount Laser Diode Mount LDM-4409
TO-Can Laser Diode Mount (with optional collimating lens) LDM-4412
TELECOM Laser Diode Mount LDM-4980
Production TO-Can Laser Diode Mount LDM-4990
March 2019 5 LDT-5910C and LDT-5940C
Specifications
March 2019 6 LDT-5910C and LDT-5940C
March 2019 7 LDT-5910C and LDT-5940C
March 2019 8 LDT-5910C and LDT-5940C
Chapter 2:
General Operation
This chapter is an overview of the operation of the LDT-5910C and LDT-5940C Thermoelectric
Temperature Controllers.
Power requirements
Front panel operation
General operating procedures
Grounding Requirements
The LDT-5910C and LDT-5940C Thermoelectric Temperature Controllers come with a three
conductor AC power cable. The power cable must be plugged into an approved three-contact
electrical outlet or used with a three-contact to two-contact adaptor with the grounding wire
connected to an electrical ground (safety ground). ILX Lightwave recommends connecting the
instrument only to properly earth grounded receptacles. The power cord connector and power
cable meet IEC safety standards.
AC Line Power Requirements
The LDT-5910C and LDT-5940C Thermoelectric Temperature Controller can be configured to
operate at nominal line voltages of either 110 - 120 VAC or 220 - 240 VAC and for the LDT-5910CJ
(100V) model 100 – 110 VAC. This is a factory fixed configuration which cannot be changed by the
end user. Check to be sure that the voltage indicated on the back panel of the instrument matches
the power line voltage in your area.
The Power Up Sequence
Prior to power up ensure the fan inlet located on the right hand side of the instrument has
no obstructions that would impede airflow. In addition ensure there is no obstruction in
the rear that may impede airflow.
With the LDT-5910C or LDT-5940C connected to an AC power source, pressing POWER supplies
AC line power to the instrument and starts the following power up sequence where each display last
two to three seconds.
All front panel indicators are ON, all 7-segment displays indicate “8”
All front panel indicators OFF
Display shows the model number, the serial number and the firmware version of the
controller
April 2018 9 LDT-5910C and LDT-5940C
During the front panel indicator test, the LDT-5910C and LDT-5940C performs a self-test to ensure
that the internal hardware and software are communicating. If the LDT-5910C and LDT-5940C
cannot successfully complete the test, an error message is displayed. See Chapter 5 for a complete
list of error messages. After the self test, the LDT-5910C and LDT-5940C configuration is set to the
same state as when the power was last turned off.
To quickly set a different configuration you can use the recall function. See Store and Recall for
more information.
Firmware Upgradeability
The firmware on the LDT-5910C and LDT-5940C can be reinstalled or upgraded via USB. Contact
ILX Lightwave technical support for information on upgrading the software of the LDT-5910C and
LDT-5940C.
GPIB Communication
The IEEE 488 GPIB interface connector is located on the rear panel, directly above the power input
module (See Figure 1.2 on page 3 or Figure 1.4 on page 4). Attach the GPIB cable to the 24-pin
connector located on the rear panel. The connector is tapered to ensure proper orientation. Finger
tighten the two screws on the cable connector.
A total of 15 devices can be connected together on the same GPIB interface bus. The cables have
single male/female connectors on each end so that several cables can be stacked. This allows more
than one cable to be attached to any one device. However, the maximum length of the GPIB cables
must not exceed 20 meters (65 feet) total, or 2 meters (6.5 feet) per device. As good practice the
number of GPIB cables connected to one instrument should be limited to less than 6.
USB Communication
The USB connector is located on the back rear panel, next to the GPIB connector. This USB
connector is the square "B"-style connector. A standard USB A/B cable is required to communicate
with the instrument. To communicate with the instrument using USB, please install the Virtual Com
Port Driver found on the accompanying CD or website prior to connecting the unit to the PC. Please
refer to Chapter 3 for more detailed instructions on operating the instrument through USB.
Tilt Foot Adjustment
The LDT-5910C and LDT-5940C have front legs that extend to make it easier to view the display. To
use them, rotate the legs downward until they lock into position.
Rack Mounting
The LDT-5910C and LDT-5940C Thermoelectric Temperature Controllers may be rack mounted in a
single or dual configuration by installing the appropriate rack mount flange on either side of the
enclosure. All rack mount accessory kits contain detailed mounting instructions. Refer to the Options
and Accessories table in Chapter 1 for applicable rack mount accessory part numbers.
March 2019 10 LDT-5910C and LDT-5940C
Connections
Temperature Control Output: A 15-pin DSUB connector (5910C) or 25-pin DSUB connector
(5940C) is located on the back panel of the instrument. The connections for each case are shown
below.
LDT-5910C LDT-5940C
PIN NUMBER CONNECTION PIN NUMBER CONNECTION
1 TE Module (+) 1 Sensor (+) Sense Terminal
2 TE Module (+) 2 Sensor (-) Sense Terminal
3 TE Module (-) 3 Fan (+)
4 TE Module (-) 4 N/C
5 TE Module Shield 5 N/C
6 Sensor Shield 6 Sensor / TE Module Shield
7 Sensor (+) 7 TE Module (+) Sense Terminal
8 Sensor (-) 8 TE Module (-) Sense Terminal
9 N/C 9 TE Module (+)
10 N/C 10 TE Module (+)
11 N/C 11 N/C
12 N/C 12 TE Module (-)
13 N/C 13 TE Module (-)
14 Fan (+) 14 Sensor (-)
15 Fan (-) 15 Sensor (+)
16 Fan (-)
17 Cable ID 1
18 Cable ID 2
19 N/C
20 N/C
21 TE Module (+)
22 TE Module (+)
23 N/C
24 TE Module (-)
25 TE Module (-)
The output terminals of the LDT-5910C or LDT-5940C should never be shorted
together or loaded with less than 0.1 Ω. Doing so may result in damage to the
instrument.
Analog Control Input: An isolated BNC connector is located on the rear panel of the LDT-5910C
and LDT-5940C provides the capability to adjust the temperature setpoint by applying a voltage
signal. The LDT-5910C and LDT-5940C can accept a signal from -5 V to +5 V, with a gain of 2 °C/V
and bandwidth of 5 Hz. This feature is useful in applications requiring sweeping of the temperature
without using GPIB or USB remote interface.
March 2019 11 LDT-5910C and LDT-5940C
Interlocks: The LDT-5910C and LDT-5940C have normally open and normally closed interlock
contacts. The interlock will change states from the “powered off” condition when the output is
enabled and will change back when the output is disabled; including any error condition that disables
the output.
External Fan Control:
The LDT-5910C and LDT-5940C feature the ability to control an
external fan through either the 15-pin DSUB connector (5910C) or the 25-pin DSUB connector
(5940C). This circuit can provide up to 500 mA and may be controlled from 1 to 12 Vdc. The
circuit will automatically increase the voltage to 12 Vdc start a fan from a stopped condition and
will regulate to the desired voltage once more than 20 mA is drawn.
Pseudo 4-Wire Measurements: The LDT-5940C offers a pseudo 4-wire measurement feature
which helps to minimize errors induced by cable resistance both in thermistor/RTD sensor
modes as well as on the output terminals. This feature is enabled by connecting the appropriate
‘sense’ terminals and is disabled by leaving them disconnected. This feature is not available in
the LDT-5910C.
March 2019 12 LDT-5910C and LDT-5940C
Front Panel Operation
This section describes the fundamentals of operation for the LDT-5910C and LDT-5940C
Thermoelectric Temperature Controllers. The order of descriptions will follow the normal progression
of how the user would typically configure the instrument for use for the first time.
Figure 2.1 – Front Panel Display
Power On / Off
The POWER button applies power to the LDT-5910C or LDT-5940C and starts the power up
sequence described above.
Adjust Knob and Enable Button
The ADJUST knob and ENABLE button is located on
the upper right side of the LDT-5910C and LDT-5940C
front panel. It is used to change the setpoints, enter
parameter values, enter the GPIB address, enter Save
or Recall bin numbers or enter instrument calibration
data. The ENABLE button indicates the two modes of
operation of the knob: Disabled (LED off) and Enabled
(LED on).
Setup
SET Button – Allows the user to adjust settings under PARAM, RECALL and STORE. The SET
button will allow the user to select each menu item and a second SET press will allow the user to
adjust the parameter. When a parameter is being adjusted, the units will flash at a constant
frequency. To return to the main menu, the sub menu will contain one option that reads “Return
<SET>.”
March 2019 13 LDT-5910C and LDT-5940C
PARAM Button – Uses a menu system for temperature controller setup functions including Limits,
Sensor, Sensor Calibration Constants, External Fan Control, Cable Resistance, Analog Input, GPIB
Address, and Display Brightness. The SET button selects each menu item and a second SET press
allows the user to adjust each parameter. The ADJUST knob can be used to cycle through and
select or modify the following parameters:
Limits:
The Limits menu allows the user to select and modify the temperature, current,
and sensor upper and lower limits
Temperature High / Low Limits
Current Positive / Negative Limits
Sensor High / Low Limits
Depending on the sensor selected, the user can configure high and low limits for resistance,
current or voltage.
Sensor
: The Sensor parameter allows the user to select the type of sensor for the
application: Thermistor, RTD, IC-V, or IC-I. For more information on temperature
sensors see the “Sensor Options” later in this chapter.
Thermistor (100 µA, 10 µA, Auto Ranging, Linearized)
RTD (1 mA, 2.5 mA)
IC – Current
IC - Voltage
Sensor Calibration Constants: The calibration constants menu allows the user to enter
sensor constants for thermistor, current and voltage IC, and RTD. For more information see
the “Sensor Options” later in this chapter.
External Fan Control: Allows the user to enable or disable the external fan. The user can
also set the voltage applied to an external fan.
Cable Resistance: To increase the accuracy of the TEC voltage measurement at higher
currents the user can enter a cable resistance. The value entered for cable resistance should
be equal to the round trip resistance between the temperature controller and the load. If the
wiring is identical on both sides of the connection, as it is with ILX cables, this value is double
the resistance value measured between the temperature controller and the load. The 4-wire
capability of the LDT-5940C may be used instead if using a cable that supports this feature.
Cable resistance should be set to 0.000 when using 4-wire capability.
Analog Control Input: Enables and disables the analog input feature, for optimal stability if a
signal is not applied to the input the analog control input should be disabled.
GPIB Address: Adjust the GPIB address from 1 to 30 using the ADJUST knob.
Display Brightness: Adjust the display brightness from 1 to 10 using the ADJUST knob.
STORE Button – Stores instrument parameters for control mode, setpoint, limits, sensor type,
calibration constants, and PID values for bins numbered 1 – 10.
March 2019 14 LDT-5910C and LDT-5940C
RECALL Button – Recalls instrument parameters for control mode, setpoint, limits, sensor type,
calibration constants, and PID values for bins numbered 1 – 10. Recall bin 0 will reset all parameters
to the factory defaults.
Non-volatile memory is used for saving the instrument parameters. When a store operation is
performed, all of the current instrument parameters are stored to a "bin" number (1 - 10). When that
"bin" number is recalled, the instrument configuration is recalled to the stored values.
To enter the STORE/RECALL mode, press either the STORE or RECALL button which will display
the current “bin” number in the bottom of the screen. The current “bin” number will be flashing and
the ADJUST knob can be used to select a new “bin”. The store or recall operation is performed
when the SET button is pressed. If the SET button isn’t pressed after three seconds the LDT-5910C
and LDT-5940C will time out and the new “bin” number will not be stored or recalled.
PID Control
SELECT Button – In PRESET mode, pressing SELECT displays the selected preset gain. In
MANUAL mode, pressing SELECT displays the last PID values used and in AUTO-TUNE mode,
pressing SELECT displays the auto-tuning method implemented. Pressing SELECT repeatedly or
using the ADJUST knob cycles through the presets and PID values.
SET Button – In PRESET mode, pressing SET selects the preset gain range on the display. In
MANUAL mode, pressing SET allows the user to adjust the value on the display. SET in the AUTOTUNE mode starts the auto-tune procedure.
MODE Button –Cycles through PRESET, MANUAL and AUTO-TUNE modes. The selected mode is
indicated by an illuminated LED. The output will not be disabled if the user changes the PID Control
Mode.
Preset: The user can select one of the preprogrammed gain ranges by first selecting the
PRESET LED using the MODE button then pressing the SELECT button or ADJUST knob to
display the correct preset gain for the thermal load and pressing the SET button to select the
gain. The LDT-5910C and LDT-5940C implements the original gain ranges of the LDT5910B (Gain 1, 3, 10, 30, 100, and 300). Presets have also been implemented for specific
ILX Lightwave mounts. When a preset has been selected for a mount, the LDT-5910C and
LDT-5940C will set the PID values, sensor type, and current limits. The following mounts are
compatible with the LDT-5910C and LDT-5940C: LDM-4405, LDM-4407, LDM-4409, LDM4412, LDM-4980, and LDM-4990.
Manual:In MANUAL mode the user can adjust the PID values or select one of the preset
PID values to load into the manual adjust (note: if the user adjusts the preset PID value
loaded into manual PID adjust the change will not permanently affect the preset PID value).
To manually adjust the PID values first select the MANUAL LED by pressing MODE, then
press SELECT to change between P, I, and D. To adjust the PID value, press SET; to move
to the next value press SELECT.
To load a preset PID value use the SELECT button to cycle to the “Adjust Preset”screen,
and then press the SET button, then using the SELECT or ADJUST knob display the preset
PID that you want to load. Once the correct preset is displayed press the SET button to load
the values.
March 2019 15 LDT-5910C and LDT-5940C
Auto-tune – The auto-tune mode will calculate a thermal system's PID coefficients
through an iterative PID temperature control process. To enter the auto-tune mode first
press MODE until the AUTO LED is illuminated, and then SET to begin the auto-tune
procedure. When the LDT-5910C and LDT-5940C successfully completes the auto-tune
the instrument will revert back to MANUAL and the calculated PID values will overwrite
the current PID values in MANUAL mode. If the auto-tune process fails an error will be
displayed on the screen. For more information on the auto-tune process see Appendix B. Additional auto-tune methods may be available in the future via a firmware upgrade,
for more information contact sales at 800-459-9459 / sales@ilxlightwave.com.
There are limitations to the auto-tune feature. Any of the following will cause the
auto-tune algorithm to fail:
Thermal systems requiring the proportional term to be less than 0.5
Noisy temperature measurements
Reaching any output off enable condition (such as temperature or voltage limits) during
the auto-tune operation.
If the auto-tune algorithm fails for a particular thermal system, it may be necessary to
modify the PID coefficients manually.
Mode
SELECT Button – Selects constant temperature, constant sensor and constant current mode. The
selected mode is indicated by an illuminated LED. The output is disabled when the control mode is
changed.
ON Button – Enables and disables the output. An enabled output is indicated by an illumined LED.
A disabled output is indicated by a dark LED.
TEMP – Controls the LDT-5910C or LDT-5940C to a constant temperature
I TE – Sets the LDT-5910C or LDT-5940C to output a constant current
R SNSR – Controls the LDT-5910C or LDT-5940C to a constant sensor value
HEAT – Illuminated when negative current is flowing
COOL– Illuminated when positive current is flowing
I LIMIT – Illuminated if the LDT-5910C or LDT-5940C has reached the user defined
current limit.
Display
The large 7-segment LED displays the measured value in the selected operating mode. When the
setpoint is being adjusted, the 7-segment LED display will show “set” after the dimension and revert
back to reporting the measured value after the three second timeout. The dot matrix display, at the
bottom of the screen, cycles through available measurement and mode setpoint. The available
measurement and setpoint parameters are listed under General Operation Procedures.
The measured values are indicated by an equal sign, as in Figure 2.2 a, and the setpoint values are
indicated by a colon sign, as in Figure 2.2 b.
March 2019 16 LDT-5910C and LDT-5940C
Figure 2.2 a – The measured current Figure 2.2 b – The setpoint current
LEFT Button – Cycles through available measurement and setpoint parameters that can be
displayed in the left hand side of the display.
RIGHT Button – Cycles through available measurement and setpoint parameters that can be
displayed in the right hand side of the display.
ADJUST LED – Indicates the status of the ADJUST knob: an illuminated LED means that the knob is
enabled and a dark LED means that the knob is disabled.
Error Indicators
The LDT-5910C and LDT-5940C indicate operational errors on the measurement display with an
error code. A complete list of error codes can be found in Chapter 5.
General Operating Procedures
The discussion below presents guidelines for operation as well as some common operating
procedures. Remote operations are discussed in the next chapter.
The output terminals of the LDT-5910C or LDT-5940C should never be shorted
together or loaded with less than 0.1 Ω. Doing so may result in damage to the
instrument.
Warm-Up and Environmental Considerations
To achieve the rated accuracy, allow the LDT-5910C or LDT-5940C to warm-up for at least one hour
before use. Operate the controller within the environmental limits specified in Chapter 1. The best
accuracy is achieved near the calibration temperatures.
General Guidelines for Sensor Selection and Safety Limits
This section presents some guidelines to assist in selecting the optimal settings for your
application.
Sensor Options
The LDT-5910C and LDT-5940C Thermoelectric Temperature Controllers can measure
temperature through a variety of sensor options; thermistors, IC sensors (IC-I, IC-V) or RTDs.
THERMISTOR – When a thermistor sensor is selected, the LDT-5910C and LDT-5940C
measures temperature based on a negative temperature coefficient (NTC) thermistor. An NTC
thermistor is a device whose resistance decreases as its temperature increases. The controller
provides a sense current (100 µA or 10 µA) through the thermistor which results in a voltage
across the thermistor. This voltage is used as a feedback signal by the LDT-5910C and LDT5940C digital control loops to maintain a constant temperature. The thermistor should be
connected across the Sensor+ and Sensor- pins (pins 7 and 8).
March 2019 17 LDT-5910C and LDT-5940C
Loading...
+ 65 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.