Newport HT-70 User manual

Newport Medical Instruments, Inc.
Newport HT70 Ventilator
Operating Manual
for Plus Model
OPRHT70-2 Rev. F
06/13
0344
Newport Medical Instruments, Inc. 1620 Sunower Ave. Costa Mesa, CA 92626 Tel: 1.714.427.5811 Tel: 1.800.451.3111 (USA Only) Fax: 1.714.427.0489 Customer Service ext. 282
www.ventilators.com email: Info@ventilators.com
5
OPR360U A0509
5-1
Manual Revision History
HT70 Operating Manual OPRHT70-2
Revisions Date Description
Rev A October 2011 New release
Rev B November 2011 Miscellaneous type corrections,
text change on page 3-3, 7-1, 7-3 regarding Power Pac.
Rev C May 2012 Add Operating Manual Reivison.
Notice regarding the following change: Pulse Oximeter accessory and related functions no longer available.
Rev D July 2012 Software update ver. P05.12.10
and above, Remove references to optional Pulse Oximeter (not available), Change Slope/Rise setting was: 1 is fastest, is: 1 is slowest.
Rev E November 2012 Update to correct image reversal
in F-1 and F-2, callout labels on various drawings converted to numbers and referenced in text or added to table. Remove reference to Aequitron nurse call systems. Replace warranty details with who to contact information.
Rev F June 2013
Add information about the Aequitron Remote Alarm System.
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Contact Information
Thank you for using the Newport HT70 family of ventilators. With the HT70 you not only get a great ventilator, you get the support of Newport Medical. Since 1981 we have maintained a focused commitment to the design, production and sale of ventilators. We have dedicated our efforts to providing ventilators that are easy to use, clinically versatile, and cost effective.
We know that ventilatory support is critical in emergency and critical care situations. But for many of our customers, it is also a part of their daily lifestyle. The HT70 Ventilators offer home care users the expanded mobility that allows them to experience more freedom in their lives than many have ever known before.
We have designed this manual to be comprehensive and still very user friendly. For the best performance from your HT70 Ventilator, please take the time to review this manual completely.
See our contact information on the following page for complete details.
OPRHT70-2 F0613
5
OPR360U A0509
Contact Information
Contact Information
Telephone:
+1.714.427.5811
1.800.451.3111 (US only)
Fax:
+1.714.427.0489
Departments:
Customer Service (Ext. 282) Technical Services (Ext. 500) available 24/7 Clinical Support (Ext. 123) available 24/7
Operational Hours:
Days: Monday through Friday Hours: 8:00 am to 5:00 pm (PST) Emergency After-hours: 24-Hour Clinical and Technical Support
Email:
Customer Service: customers@ventilators.com Clinical Education and Support: clinical@ventilators.com Technical Education and Support: techservice@ventilators.com
Internet:
www.ventilators.com
Shipping Address:
Newport Medical Instruments Attn: Receiving Department 1620 Sunower Avenue Costa Mesa, CA 92626, USA
EC REP
Authorized European Representative Emergo Europe Molenstraat 15 2513 BH, The Hague The Netherlands
OPRHT70-2 F0613
FOLDOUT DRAWINGS
Use the following drawings as reference while reviewing
the manual sections
Unfold to view drawing on reverse side
OPRHT70-2 F0613
Figure F-1 English Version- Front Panel Overview
1
9
10
1. Breath Delivery Indicator LED. Flashes green with every breath delivered by the
ventilator.
2. External Power LED. Lights green whenever external power is connected. This
also indicates that the Internal Dual Battery System is being charged.
3. Manual Inflation button. Press and hold this button to deliver ow to the patient.
The ventilator will deliver ow at the current settings while the button is pressed. Flow delivery is limited to a maximum of 3 seconds or until the High Pressure alarm setting is reached.
11
2
3
4
5
6
12
13
14
15
16
4. Brightness button. Press this button repeatedly to scroll to one of four screen
brightness levels.
5. Device Alert LED. Lights red when there is a device alarm. Take the ventilator out
of service and use an alternate means of ventilation until resolved.
6. Patient Gas Output. Attach patient breathing circuit tubing here.
7. Proximal Pressure Line connector. Attach proximal pressure tubing here.
8. Exhalation Valve Drive Tubing connector. Attach exhalation valve drive tubing here.
9. Alarm Violation LEDs. LEDs in the handle light to indicate alarm conditions.
10. Alarm Silence/Reset button. Press this button to silence the audible alarm for 1
minute. Once an alarm condition has been corrected, press this button to clear/ reset the alarm message and latched indicators.
11. Alarm Silence LED. Remains lit during the one minute alarm silence period.
12. Cancel button. Press this button if you want to cancel changes that have not
already been accepted.
13. Accept button. Press this button to accept/conrm all changes made to control
settings.
F-1
14. s Up / t Down Arrow buttons. Press to change a highlighted parameter up/down
7
8
by one unit. Hold down continuously and the parameter will change at an increasingly quicker pace.
15. Touch Screen User Interface. Touch screen to access alarms and parameter settings.
16. Flow Sensor connector. Attach on-airway ow sensor here.
OPRHT70-2 F0613
Unfold to view drawing on reverse side
OPRHT70-2 F0613
Figure F-2 Symbols Version- Front Panel Overview
1
9
10
1. Breath Delivery Indicator LED. Flashes green with every breath delivered by the
ventilator.
2. External Power LED. Lights green whenever external power is connected. This
also indicates that the Internal Dual Battery System is being charged.
3. Manual Inflation button. Press and hold this button to deliver ow to the patient.
The ventilator will deliver ow at the current settings while the button is pressed. Flow delivery is limited to a maximum of 3 seconds or until the High Pressure alarm setting is reached.
5
4
6
3
2
11
15
12
13
14
4. Brightness button. Press this button repeatedly to scroll to one of four screen
brightness levels.
5. Device Alert LED. Lights red when there is a device alarm. Take the ventilator out
of service and use an alternate means of ventilation until resolved.
6. Patient Gas Output. Attach patient breathing circuit tubing here.
7. Proximal Pressure Line connector. Attach proximal pressure tubing here.
8. Exhalation Valve Drive Tubing connector. Attach exhalation valve drive tubing here.
9. Alarm Violation LEDs. LEDs in the handle light to indicate alarm conditions.
10. Alarm Silence/Reset button. Press this button to silence the audible alarm for 1
minute. Once an alarm condition has been corrected, press this button to clear/ reset the alarm message and latched indicators.
11. Alarm Silence LED. Remains lit during the one minute alarm silence period.
12. Cancel button. Press this button if you want to cancel changes that have not
already been accepted.
13. Accept button. Press this button to accept/conrm all changes made to control
settings.
F-2
14. s Up / t Down Arrow buttons. Press to change a highlighted parameter up/down
7
16
8
by one unit. Hold down continuously and the parameter will change at an increasingly quicker pace.
15. Touch Screen User Interface. Touch screen to access alarms and parameter settings.
16. Flow Sensor connector. Attach on-airway ow sensor here.
OPRHT70-2 F0613
Unfold to view drawing on reverse side
OPRHT70-2 F0613
Figure F-3 HT70 Plus Model Touch Screen (Hospital Domain)
1
1. Screen Selections buttons. Touching any one of these buttons will take you to the new
screen. The More screen includes links to Event, Trends, Wave and Utility screens.
2. Mode selector. Touching this button scrolls through the mode choices. The mode will
not change until you press the Accept button.
3. Breath Type selector. Touching this button toggles the breath type choice. The breath
type will not change until you press the Accept button.
4. Help button. Touching this button enables a tutorial for each feature on the screen.
6
7
8
9
Touch the help button then touch any button for an explanation of that feature.
5. Monitored Data buttons. Touching any one of these four buttons opens a screen with a
view of monitored parameter choices to display in that button.
6. Message display. This area shows all informational and alarm messages and current
NIV selection, mode, and breath type selection. During an alarm violation this area will light red for High Priority, amber for Medium Priority and yellow for Low Priority alarms and display the alarm message.
7. Battery Charge Level indicator. Shows the charge level of the “Power Pac” battery
pack (blue icon) during external power or Power Pac use or the charge level of the Backup Battery (red icon) during Backup Battery use.
8. Pressure Bar. Indicates dynamic pressure in the patient circuit in green, the High and
Low Pressure Alarm settings in red and the peak pressure of the last breath in green.
2
3
4
5
F-3
10
11/12
9. Parameter Setting buttons. Touching any one of these buttons will activate the
parameter to allow adjustments.
10. Patient Effort indicator. Flashes green to show a spontaneous patient effort.
11. Domain button. The HT70 can be set up in one of three Domains: Basic, Transport and
Hospital. Touch to scroll through the Domain choices. Press Accept to conrm choice.
12. AutoLock/Unlock button. This button is only visible if Auto Lock is enabled and the
panel is locked. Touch and hold for 3 seconds to unlock touch screen buttons.
NOTE: While operating on battery power with Power Save enabled and all alarms cleared, the touch screen will go to sleep after 2 minutes. Just touch the screen to bring it back into view.
OPRHT70-2 F0613
OPR360U A0509
5-1
5
Table of Contents
1 Introduction
Brief Device Description ............................................................. 1-1
Intended Use .............................................................................. 1-3
Warnings, Cautions and Notes ................................................... 1-3
2 Overview of Controls, Screens and Connectors
Front Panel Overview ................................................................. 2-1
Touch Screens Overview ............................................................ 2-1
Internal Dual Battery System Overview ...................................... 2-1
Rear Panel Overview .................................................................. 2-2
Right Side Overview ................................................................... 2-3
Left Side Overview ..................................................................... 2-3
Bottom Panel Labeling ............................................................... 2-4
3 Set up and Pre-use Preparations
Unpack the HT70 Ventilator ........................................................ 3-1
Assemble the Ventilator .............................................................. 3-2
Connect to AC Power ................................................................. 3-2
Using the Power Switch ............................................................. 3-3
Make Parameter Changes .......................................................... 3-4
Attach a Patient Circuit ...............................................................3-4
For use with a third party humidier ..................................... 3-5
For use with an HME ............................................................3-8
Using the On-airway Flow Sensor ...................................... 3 -10
Connect Optional Accessories .................................................. 3 -11
Air/Oxygen Entrainment Mixer .............................................3 -11
Low Flow Oxygen Reservoir ............................................... 3 -12
D.C. Auto Lighter Power Adapter ....................................... 3-14
Aequitron Remote Alarm Cable Accessory ........................ 3 -14
4 Navigating the HT70 Screens
Touchscreen (Graphical User Interface) Layout ......................... 4-1
Primary Screen Buttons and Displays........................................ 4 -1
Ventilator Settings Adjustment ...................................................4-4
Start Up Screen Navigation (Standby Condition Only) ..............4-4
Circuit Check Button ............................................................ 4-5
How to Perform a Circuit Check ..................................... 4-5
If the Circuit Check Fails .................................................4-6
Activate Preset Button .......................................................... 4-7
How to Use a Preset ....................................................... 4-7
Start Ventilation Button .............................................................. 4-7
OPRHT70-2 F0613
555
OPR360U A0509
5-1
Table of Contents
Alarms Screen Navigation .......................................................... 4-8
Main Screen Navigation ............................................................. 4-9
More Screen Navigation ............................................................4 -11
More Screen Details ........................................................... 4-13
Events ........................................................................... 4 -13
Trends ........................................................................... 4-14
Waves ........................................................................... 4 -15
O2 Cylinder Data Screen .............................................. 4-16
Calibrate O2 Monitor .....................................................4-17
Utility Screen ............................................................................ 4 -18
Utility Screen Details .......................................................... 4 -19
Time/Altitude Screen .................................................... 4-19
Customize Settings Screen .......................................... 4-20
Custom Presets ............................................................ 4-21
Back Up Ventilation ......................................................4-22
Domain Navigation ................................................................... 4-23
Hospital .............................................................................. 4-23
Transport ............................................................................ 4-24
Basic ................................................................................ 4-25
5 Operating the HT70 Ventilator
Quick Check Procedure ............................................................. 5-1
Introduction .......................................................................... 5-1
Equipment Needed ............................................................... 5-1
Pretest Inspection ................................................................. 5-1
Set Up .................................................................................. 5-1
Standard Test Settings ......................................................... 5-2
Quick Check Procedure ....................................................... 5-2
Pass / Fail Check Off Sheet ................................................. 5-4
Patient Setup Procedure ............................................................5-5
Troubleshooting Guide ............................................................... 5-7
6 Ventilator Alarms
Setting Alarms ............................................................................ 6-1
Alarm Quickset ..................................................................... 6-1
Alarm Indicators ......................................................................... 6-2
Alarm Silence/Reset Button .................................................6-2
Alarm Silence LED ................................................................ 6-2
User Adjustable Alarms .............................................................. 6-2
Backup Ventilation ......................................................................6-6
Automatic Alarms .......................................................................6-6
Battery Alarms ............................................................................ 6-9
OPRHT70-2 F0613
5
5
OPR360U A0509
5-1
Table of Contents
7 Battery Operation
Internal Dual Battery System .......................................................7-1
Power Pac Battery Pack ..............................................................7-1
Backup Battery ........................................................................... 7-2
Conditions that Affect Battery Use Time .................................... 7-2
Check the Battery Charge Level and Battery Time Estimator ... 7-3
Best Use Tips ............................................................................. 7- 3
Battery System Maintenance ..................................................... 7- 4
Power Pac Battery Pack Removal .............................................. 7-4
Battery Alarms Overview ............................................................ 7-5
Power Accessories ..................................................................... 7- 6
8 Cleaning and Maintenance
Cleaning and Disinfecting ........................................................... 8 -1
Ventilator .................................................................................. 8 -1
Accessories ................................................................................ 8-2
Low Flow Oxygen Reservoir ................................................. 8-2
Air/Oxygen Entrainment Mixer .............................................. 8-2
Reusable Breathing Circuits ....................................................... 8-3
Air Intake Filter ............................................................................ 8-3
Proximal Inline Filter ...................................................................8-4
Maintenance Guidelines ............................................................. 8-4
Routine Maintenance ............................................................8-4
6 Month Maintenance ........................................................... 8-5
12 Month Maintenance ......................................................... 8-5
24 Month Maintenance ......................................................... 8-5
15,000 Hour Maintenance ....................................................8-6
General Warnings ....................................................................... 8-6
Factory Maintenance or Repair .................................................. 8-7
Repacking/Return Information ................................................... 8-7
9 Specifications
Front Panel Buttons - Symbols Version ..................................... 9-1
Miscellaneous Reference Symbols ............................................ 9-2
Controls / Monitors .....................................................................9-3
Monitor Data Selections ............................................................. 9-4
Front Panel Membrane Buttons and Indicators .........................9-4
Alarms .................................................................................. 9-5
User Adjustable .................................................................... 9-5
Automatic .............................................................................9-5
Hardware Requirements ............................................................. 9-7
Environment ................................................................................9-8
Size and Weight ..........................................................................9-8
OPRHT70-2 F0613
5
5
OPR360U A0509
5-1
Table of Contents
Factory Default Parameters ........................................................ 9-9
Miscellaneous ............................................................................9-9
(optional) Air / Oxygen Entrainment Mixer .................................. 9-9
(optional) Low Flow Oxygen Reservoir ....................................... 9-9
Regulatory and Agency Standards .......................................... 9 -10
10 Explanations of Modes and Controls
Explanation of Modes and Controls ......................................... 10-1
Foldout Diagrams
English Version - Front Panel Overview ..................................... F-1
Symbols Version - Front Panel Overview ................................... F-2
HT70 Plus Model Touch Screen ................................................. F-3
OPRHT70-2 F0613
Section 1:
Introduction
Introduction
5
5
OPR360U A0509
5-1
Section 1:
Introduction
Brief Device Description .................................... 1-1
Intended Use .......................................................1-3
Warnings, Cautions and Notes ......................... 1-4
5
5
OPR360U A0509
5-1
Introduction
Brief Device Description
The Newport HT70 family of ventilators are state of the art ventilators that combine ruggedness, ease of use and clinical prociency with exceptional mobility to provide ventilatory support for infant, pediatric and adult patients in emergency care, transport, critical care, subacute care and home care applications. They are also ideal for emergency preparedness applications.
The compact, lightweight HT70 ventilator is built for hard work with a durable polymer exterior and robust overall design that stands up to harsh environments.
The HT70 Ventilator denes ease of use with all essential controls at your ngertips using a simple membrane button and touch screen combination. There are no complicated menus or difcult sequences to follow in order to make necessary adjustments for common operations.
A three-tiered management domain system makes it very easy for critical caregivers to manage all controls while providing quick access to the more essential elements in transport situations and signicantly enhanced safety and simplicity in the homecare environment.
1
Sophisticated Clinical Capabilities
In addition to its durability and ease of use, the HT70 ventilator offers the complete array of clinical capabilities needed for managing critical patients.
The twin micro-piston pump’s ability to deliver a variable ow enables the HT70 to provide a full range of operating modes and breath types with servo-controlled, leak-compensated PEEP. Leak compensation helps to improve triggering and avoid auto-triggering when a leak is present. The HT70 may be used with an endotracheal tube, tracheal tube, face mask, nasal mask or prongs, or mouthpiece.
There are 3 models for the HT70 series of ventilators:
HT70S HT70 Basic for use when Pressure Support is not needed. HT70 HT70 Classic, adds Pressure Support and related parameters and Trends screen HT70PM HT70 Plus, adds on-airway ow sensor option with graphics, ow trigger and exhaled volumes
The HT70 Basic and Classic models provide monitoring of inspiratory tidal volume (every breath), inspiratory minute volume, total respiratory rate, peak pressure, mean pressure and baseline (PEEP) pressure. Real-time patient circuit pressure is displayed at all times
OPRHT70-2 F0613
1-1
5
5
OPR360U A0509
5-1
1
Introduction
on the airway pressure gauge on the face panel. A comprehensive alarm system is built-in to alert the user to violations of user-set or ventilator safety limits. An optional built-in oxygen sensor allows monitoring of O2 with high and low O2 alarms.
The HT70 Plus model adds an on-airway ow sensor with onscreen graphics, exhaled tidal and minute volume monitoring/alarms, and ow trigger. This manual describes the HT70 Plus model and will denote features that are not available on the HT70 and HT70S models.
Gas delivery to the patient may be enriched with oxygen (0.21-1.00) using either the optional Air Oxygen Entrainment (50 psi) Mixer or optional Low Flow Oxygen Reservoir.
Exceptional Mobility
The ventilator’s unique design provides maximum mobility and safety for short or long distance transport of critically ill patients and also for patients who are going about their normal activities of daily life. This exceptional mobility is derived from two sources: Newport’s patented, power conserving dual-micro-piston technology which eliminates the need for an external compressed gas source, and the Internal Dual Battery System which allows virtually continuous use from battery power through hot-swappable technology.
The HT70’s micro-pistons use a fraction of the power that is consumed by turbines and blowers. This enables longer battery use time. Our patented system also uses considerably less supplemental oxygen than turbine or blower systems, again improving mobility for transport or homecare use. The superior technology of our micro­piston system over the turbine and blower systems allow the HT70 to ventilate safely over a wide range of environmental conditions and altitudes.
The HT70’s twin micro-piston internal pump is made of mechanically moving components. As with any other gas delivery system made of moving components, it may emit a minor level of noise during operation. This is not a malfunction and does not affect the performance of the ventilator.
The Internal Dual Battery System consists of two independent but coordinated lithium ion batteries, the Power Pac battery, located on the back of the ventilator and the Backup Battery inside the ventilator. The Internal Dual Battery System can provide up to 10 hours of operation at standard settings when new and fully charged. This system assures continued support during transport, daily activities or power outages.
1-2
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Introduction
The detachable Power Pac is ‘hot-swappable’. That is, if more battery time is needed, a depleted Power Pac can easily be removed from the back of the HT70 and replaced with a recharged Power Pac without interrupting ventilation. No tools are needed. The secondary Backup Battery maintains operation without interruption when the Power Pac is swapped out and also provides a minimum of 30 minutes of full operation when all other power sources are depleted. The Power Pac weighs two pounds and is charged anytime the ventilator is connected to an external power source (AC or DC). It can also be charged separately.
The HT70 may be operated from a variety of AC (100-240 VAC @ 50 / 60 Hz) or DC (12-24 VDC) external power sources or from the Internal Dual Battery System. The optional DC Auto Lighter Power Adapter accessory enables connection to an automobile-type DC outlet. Any time the ventilator is connected to external power, both batteries in the Internal Dual Battery System are charging, whether or not the ventilator is in use.
Travel Certified
1
The HT70 has been tested for and meets requirements for use in helicopter and xed wing transport and for use on commercial airlines. Before traveling, be sure to speak with your airline representative about their particular concerns and clear all of your equipment with them well before your departure. The labeling that the FAA requires to be on the ventilator is located on the bottom of the HT70.
Intended Use
Newport HT70 family of ventilators is intended to provide continuous or intermittent positive pressure mechanical ventilatory support for the care of individuals who require mechanical ventilation through invasive or noninvasive interfaces.
Specically, the Newport HT70 family of ventilators is applicable for infant, pediatric and adult patients greater than or equal to 5 kg (11 lbs) in hospital, sub-acute, emergency department, and home care environments as well as for transport and emergency response applications.
NOTE:
Federal law (US) restricts sale by or on the order of a physician.
OPRHT70-2 F0613
1-3
5
5
OPR360U A0509
5-1
1
Introduction
Ventilator Configurations
Newport Medical offers ve congurations for the 3 models in the HT70 family of ventilators. See Table 1. In addition, the front control panel labeling is available in various languages and regional power cords, i.e. North American, European, etc., can be specied. See your Newport Medical Representative for details.
Table 1
Part Number Description Distinguishing Features
HT70PM HT70 Plus
HT70M
HT70S
HT70SM
HT70
HT70, w/o Oxygen Sensor
HT70, with Oxygen Sensor
HT70, Basic w/o Oxygen Sensor
HT70, Basic with Oxygen Sensor
Full featured with Flow sensor, graphics and built-in oxygen monitor with alarms.
Classic features No built-in oxygen monitor
Classic features Includes built-in oxygen monitor with alarms
Classic features except:
• No Pressure Support or Pressure
Support parameters
• No Trends
• No built-in oxygen monitor
Classic features except:
• No Pressure Support or Pressure
Support parameters
• No Trends
Includes built-in oxygen monitor with alarms
Warnings, Cautions, and Notes
Please review all WARNINGS and Cautions outlined in this manual before operating the ventilator.
Use of the product requires full understanding and strict observation of all sections of these instructions. The equipment is only to be used for the purposes specied under Intended Use and in conjunction with appropriate patient observation and monitoring. Observe all WARNINGS and Cautions that appear in this manual and on equipment labels.
WARNING
1-4
A warning describes a condition that can cause injury.
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Introduction
1
Caution:
equipment.
NOTE:
convenient.
General Notes
The Newport HT70 has been designed to accommodate connectivity with nurse call/monitoring systems. Because it is not possible to anticipate every conguration of hardware and software associated with nurse call/monitoring system, it is the user’s responsibility to conrm proper functionality of the system when used in conjunction with the HT70. Verication of alarms, alerts and patient data transmissions is required. If the system performance is not as expected, contact Newport Medical Technical Support for assistance troubleshooting the set-up. Do not use the HT70 ventilator with a nurse call/monitoring system until the functionality of the ventilator/ system combination has been conrmed.
General Cautions
Do not place liquids on or near the ventilator.
A caution describes a condition that can cause damage to
A note emphasizes information that is important or
Damage can occur if the HT70 is exposed to extreme temperatures. Do not store the HT70 in areas where it may be exposed to temperatures below -40° C (-40° F) or above 65° C (149° F).
To avoid the risk of electric shock, the ventilator should not be opened by anyone other than an approved service provider.
General Warnings
The design of the HT70 ventilator, the Operating and Service manuals, and the labeling on the ventilator take into consideration that the purchase and use of the equipment is restricted to trained professionals, and that certain inherent characteristics of the ventilator are known to the operator. Instructions, warnings and caution statements are therefore limited to the specics of the HT70.
This manual excludes references to various hazards which are obvious to medical professionals and operators of this equipment including consequences of product misuse, and potential adverse effects in patients with abnormal conditions.
Transport of patients with the HT70 requires that medical staff have a good working knowledge of the ventilator’s use and problem resolution. Proper emergency backup equipment must be immediately available during transport.
OPRHT70-2 F0613
1-5
5
5
OPR360U A0509
5-1
1
Introduction
Product modication or misuse can be dangerous. Newport Medical Instruments, Inc. disclaims all liability for the consequences of product alterations or modications, as well as for the consequences which might result from the combination of this ventilator with other products, whether supplied by Newport or by other manufacturers, unless such a combination has been specically endorsed by Newport Medical. There is a risk of explosion if used in the presence of ammable anesthetics.
A patient connected to a ventilator requires the constant attention of trained caregivers to the patient’s condition.
Ventilator alarms are a critical element in the safety net of patient care. It is extremely important for patient safety that caregivers immediately identify and correct alarm violations.
Always have an alternate power source and means of ventilation available when the ventilator is in use so that they are easy to access in case of a mechanical or system problem.
If a fault is detected in the ventilator and its life support functions are in doubt, immediately discontinue use; use an alternative method of ventilation until the fault has been corrected. Contact your service provider immediately.
Do not block the Emergency Gas Intake (on the bottom panel) or the Fresh Gas Intake Port (on the right side panel).
Always use appropriate monitors to ensure sufcient oxygenation and ventilation (such as a pulse oximeter and/or a capnograph) when the HT70 Ventilator is in use on a patient.
The optional Air/Oxygen Entrainment Mixer and Low Flow Oxygen Reservoir are designed to operate with medical grade oxygen.
Ensure that the oxygen source is not empty before and during the use of the optional Air/Oxygen Entrainment Mixer or Low Flow Oxygen Reservoir.
To avoid putting stress on the internal pump and compromising gas delivery to the patient, ensure that the Air/Oxygen Entrainment Mixer is not connected to the gas intake port on the ventilator when performing a Circuit Check. Ensure that the oxygen supply is enabled any time the optional Air/Oxygen Entrainment Mixer is secured in place while ventilating.
1-6
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Introduction
Calibrated oxygen monitoring at clinically appropriate levels is required for patient safety when supplemental oxygen is in use. The optional built-in oxygen sensor on the HT70 allows High and Low O2 alarms to be enabled which can be used to assure proper oxygen de livery.
Always plug the HT70 into an external power supply source whenever it is available, even when HT70 is not in use, to keep the Internal Dual Battery System fully charged and to ensure best battery performance. Check battery capacity on the front panel before detaching from external power.
When installing a replacement Power Pac during battery operation, always ensure that the charge level LED on the replacement pack is green, indicating 90% or higher charge level.
Always ensure that the green External Power LED lights when the ventilator is connected to an external AC or DC power source.
To maintain grounding integrity when using AC power, only connect to properly grounded receptacles.
1
Use only the Newport supplied AC Power Supply (p/n PWR3204P) with the HT70 ventilator and HT70 Power Pac (p/n BAT3271A).
Always disconnect the external power supply prior to servicing.
After servicing the HT70, it must pass the Operational Verication Procedure (OVP) before it is returned to patient use. See the HT70 Service Manual.
Do not use electrically conductive breathing circuits. Always use clean and dry breathing circuits.
Always use a clean, dry lter in the following locations: a standard bacteria lter on the gas output, a prox line (bacteria) lter on the proximal pressure tubing port and an intake (bacteria) lter behind the lter cover.
Adding attachments or other components or sub-assemblies to the ventilator breathing circuit system can increase the patient’s work of breathing and/or add resistance to patient exhalation.
Always ensure that the audible alarm loudness level is set at a volume that can be heard by the caregiver. Do not use the ventilator in an environment where audible alarms cannot be heard by the caregivers.
OPRHT70-2 F0613
1-7
5
5
OPR360U A0509
5-1
1
Introduction
The functioning of this machine may be adversely affected by the operation of other medical equipment, such as high frequency surgical(diathermy) equipment, debrillators or short-wave therapy equipment in the vicinity.
This device has undergone EMC testing and found to be in conformance with IEC 60601-1-2:2001 and meets the requirement of CISPR11:2004 (Class B), IEC 61000-3-2:2006, and IEC 61000-3-3:1955 + A1:2001 + A2:2005. These requirements are designed to provide reasonable protection against harmful interference in a typical medical installation, as well as in homecare environments. The equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with these instructions, may cause harmful interference to other devices in the vicinity. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference with other devices, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures;
Reorient or relocate the receiving device.
Increase the separation between the equipment.
Connect the equipment into an outlet on a circuit different from
that to which the devices(s) is connected.
Consult the manufacturer or eld service technician for help.
Copyright Information
© Copyright 2012 Newport Medical Instruments, Inc. All rights reserved. Newport HT70 Ventilator is manufactured in accordance with Newport Medical Instruments, Inc. proprietary information and is protected under U.S. Patent # 7,654,802.
1-8
OPRHT70-2 F0613
Section 2:
Overview
Overview
5
5
OPR360U A0509
5-1
Section 2:
Overview
of Controls, Screens and Connectors
Front Panel Overview ......................................... 2-1
Touch Screens Overview ................................... 2-1
Internal Dual Battery System Overview ........... 2-1
Rear Panel Overview ..........................................2-2
Right Side Overview ...........................................2-3
Left Side Overview .............................................2-3
Bottom Panel Labeling ......................................2-4
5
OPR360U A0509
5-1
Overview of Controls, Screens and Connectors
2
Front Panel Overview
Please refer to Foldout drawing F-1 at the front of the manual. The HT70 front panel consists of easy access membrane buttons, LED indicators and the patient connection manifold. The center touch screen panel provides access to alarm and parameter settings. The HT70 Plus model has the added port for connecting the on-airway ow sensor. This manual describes the HT70 Plus full features.
Touch Screen Overview (Hospital Domain)
Please refer to Foldout drawing F-3 at the front of the manual. The HT70 touch screen includes direct access to essential screens for setting patient parameters and alarms. Simple menu navigation allows access to advanced features and utility screens.
NOTE:
While operating on internal battery power, when the Power Save feature is ON and there are no active alarms, the touch screen will go to sleep after two minutes. Just touch the screen or any membrane button to bring it back into view.
Internal Dual Battery System Overview
The Internal Dual Battery System can provide up to 10 hours of operation when new and fully charged (under standard conditions shown in Section 7) and consists of two independent but coordinated lithium ion batteries; the hot swappable Power Pac battery and the secondary Backup Battery. When external power is lost, the ventilator will run on the Power Pac until the “Switching to Backup Battery” alarm activates. The Backup Battery will then provide a minimum of 30 minutes of emergency back up power. The Backup Battery portion of the system also maintains operation without interruption whenever the Power Pac is swapped out. The Power Pac can be recharged independently from the ventilator. The Power Pac has an LED on the bottom edge to show charge condition. Push the button to see charge condition. Green = approximately 90% or higher charge level, Amber = charge not completed, Red = battery depleted. Always insert the Power Pac onto the HT70 and power it ON to verify
x arn
e ltE o e wrP
a fa lI i
lM aun
nn t o
Bi sen r t s g h
ve ecD i
eA trl
1
1. Backup Battery (inside the case)
2. Power Pac Battery (exchangeable)
3. Release Latch
E tx
oP wr
re na
e
l
nuaM la fa o tl iI nn
B gr hi tn
e s s
D e cv eA
ei r tl
OPRHT70-2 F0613
3
2
2-1
5
OPR360U A0509
5-1
2
Overview of Controls, Screens and Connectors
the actual charge level percentage (shown in the message display). Proper care and maintenance of the Internal Dual Battery System will ensure the longest life and best usage performance. See Section 7 for complete details on the Internal Dual Battery System.
Rear Panel Overview
1. “Power Pac” battery pack
2. External power supply input
3. Remote alarm output
4. Release latch
5. RS-232 output
with central monitoring systems.
6. Serial number label
7. On/Off power switch
ventilator On/Off.
Push to remove battery pack.
External communication port to communicate
Connects to Nurse Call systems.
Momentary switch to power the
2-2
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Overview of Controls, Screens and Connectors
2
Right Side Overview
1. Fresh Gas Intake port / Oxygen Accessories’ connection/ Optional Oxygen Accessories / Bio-filter connection
attachment of the optional Air Oxygen Entrainment Mixer, Low Flow Oxygen Reservoir or Bio-lter.
2. Air Intake Filter cover
lter through the transparent cover.
Allows visual inspection of the air intake
Allows
Left Side Overview
1. Cooling fan cover
2. USB ports (2)
such as:
• aashdrivefordownloadingtheTrends*andEventHistory
les or uploading new software.
Protects the internal fan.
Allows the attachment of optional accessories
*not available on HT70S model
OPRHT70-2 F0613
2-3
5
OPR360U A0509
5-1
2
Overview of Controls, Screens and Connectors
Bottom Panel Labeling
The Bottom Panel of the HT70 includes a label that contains information regarding agency approvals and power ratings. Here you will nd the model number and manufacturing information.
NOTE:
panel near the power switch.
The serial number for the unit is located on the lower rear
Type BF
0344
2-4
OPRHT70-2 F0613
Section 3:
Set-up
Set-up
5
5
OPR360U A0509
5-1
Section 3:
Set up and
Pre-use Preparations
Unpack the HT70 Ventilator .............................. 3-1
Assemble the Ventilator ....................................3-2
Connect to AC Power.........................................3-2
Using the Power Switch ....................................3-3
Make Parameter Changes .................................3-4
Attach a Patient Circuit .....................................3-4
For use with a third party humidifier .......... 3-5
For use with an HME .................................... 3-8
Using the On-airway Flow Sensor .............3-10
Connect Optional Accessories ....................... 3 -11
Air/Oxygen Entrainment Mixer ...................3 -11
Low Flow Oxygen Reservoir ....................... 3-12
DC Auto Lighter Power Adapter ................. 3 -14
Aequitron Remote Alarm Cable .................3-14
5
5
OPR360U A0509
5-1
Set up and Pre-use Preparation
Unpack the HT70 Ventilator
Remove all of the items from the shipping box and inspect each part and component for completeness. Verify that there is no shipping damage. To obtain information about a warranty, if any, contact Technical Services or your local representative.
The Newport HT70 Ventilator, Plus Model assembly includes the following parts:
1 ea. HT70PM-XX-X X Ventilator 1 ea. OPRHT70-2 Operating Manual 1 ea. PWR3204P AC Power Supply with Pinch Release 1 ea. PWRXXXX Country specic Power Cord 1 ea. HT460300 Air Intake Filter, Disposable (pkg. of 5 lters) 1 ea. HT6004701 Prox. Inline Filter, Disposable (pkg. of 5 lters) 1 ea. FLT3302P Bacteria Filter 2 ea. ------------- Adult/Ped Flow Sensor Kit (pkg. of 1)
Optional accessories:
o KIT3420A Adult/Ped Flow Sensor Kit (pkg. of 4) o RSV3215A Low Flow Oxygen Reservoir o MXL70A-XX-XX Air/Oxygen Entrainment Mixer (country specic) o FLT3209P Mixer Filter (for Air/Oxygen Entrainment Mixer) o ADP3203P DC Auto Lighter Power Adapter o BAT3271A Power Pac Battery Pack with LED o PWR3204P AC Power Supply with Pinch Release (use to recharge extra batteries) (requires country-specic power cord) o PWR3207P NA-North American Style Power Cord (use with Power Supply) o PWR3210P UK-British Style Power Cord (use with Power Supply) o PWR3211P EU-Euro Style Power Cord (use with Power Supply) o CBL3223 Remote Alarm Cable (1/4” phone jack connector) o CRT3215A HT70 Cruiser Cart o MNT3208A Single e-Cylinder Mount o MNT3209A Dual e-Cylinder Mount o 10104494 Aequitron Remote Alarm Cable
3
Contact Newport Medical Customer Service for more details on available accessories.
OPRHT70-2 F0613
3-1
5
5
OPR360U A0509
5-1
3
Set up and Pre-use Preparation
Assemble the Ventilator
1. After unpacking the ventilator, check to see that you have all the accessories needed and check for any damage that may have occurred during shipping.
2. Assemble the cart using the instructions provided with the cart.
3. Securely position the ventilator on the pedestal mount of the cart.
Caution:
ventilator and accessories remain secure and are not damaged.
Take care to assemble the cart correctly to assure that the
Connect to A.C. Power
The HT70 comes with an AC Power Supply that includes AC power adapter with pinch-release power plug. The detachable AC power cord can be ordered in country specic congurations. Only use the approved HT70 AC Power Supply to connect the HT70 Ventilator to AC p ower.
1. External power supply input
2. Pinch Release power cable
1
2
4
3
3. AC power cord
4. AC power adapter
Plug the pinch-release power plug from the AC Power adapter into the external power supply input located on the lower left corner of the Power Pac battery pack. Ensure that the cord is to the right of the plug and that it locks in place securely. Plug one end of the power cord into the adapter and the other end into a properly grounded outlet.
To remove the AC Power Supply from the external power supply input, gently pinch the connector to release the locking pin and then pull the plug out.
3-2
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Set up and Pre-use Preparation
3
Pinch release power plug
Caution:
When the HT70 Ventilator is connected to external power, both batteries in the Internal Dual Battery System are charged simultaneously.
NOTE:
ensure that both the Power Pac battery and the secondary Backup Battery are fully charged before disconnecting from external power.
The Power Pac can also be connected to external power independently of the ventilator. Before installation on an HT70 Ventilator, check the Battery Charge LED on the bottom edge of the battery to ensure that the green LED is lit, indicating that the charge level is approximately 90% or above. Insert the Power Pac onto the HT70 and power it ON to verify the actual charge level percentage (shown in the message display area).
Do not twist the power plug or it may be damaged.
Check the Battery Charge gauge on the touch screen to
Using the Power Switch
The momentary-type power switch is located on the left side rear of the ventilator along the bottom edge.
To power the ventilator On: Press the power switch once and wait for the startup screen to appear.
On/Off momentary
power switch
changes and perform the Circuit Check prior to ventilation.
To start ventilation, touch the ‘Start Ventilation’ button at the top of the screen.
To power the ventilator Off:
the power switch once. A message appears that prompts you to ‘Press Accept to Shutdown or Cancel to Ignore’. Pressing the Accept
OPRHT70-2 F0613
The ventilator will be in Standby Condition. You can make settings
Press
3-3
5
5
OPR360U A0509
5-1
3
Set up and Pre-use Preparation
button will power off the ventilator. Pressing Cancel will return the ventilator to its previous state. Press Alarm Silence to silence the Shut Down alarm.
Make Parameter Changes
Most parameters are changed with a simple Touch / Adjust / Accept method:
1. Activate the control by touching it (button will appear highlighted).
2. Use the Up/Down s t buttons to adjust the setting.
3. Press the Accept button to accept the change.
You can make several adjustments before accepting the changes. When you are satised with all of the changes, you can accept them all by pressing the Accept button once. In the case of a mode and or breath type change, select the Main Screen for view, change the mode and/or breath type and then adjust all visible parameters prior to pressing the Accept button.
Before you accept any change, if you want to go back to the previous settings, simply press the Cancel button.
NOTE:
pressed within 20 seconds of the last button touched, the parameter will revert to the original setting.
If a parameter is touched and adjusted but Accept is not
Attach a Patient Circuit
Always use a clean and dry patient circuit.
Always use an inline lter (p/n HT6004701 or equivalent) at the Prox. Line Connector to protect the internal transducers from moisture or other contaminants.
Always use a bacteria lter (p/n FLT3302P or equivalent) on the Gas Output connector.
Always orient the exhalation valve for the correct ow orientation. Valves used in a single limb circuit have arrows that point towards the patient and valves that are used in J style or two limb circuits have arrows that point away from the patient.
When using the on-airway ow sensor, orient it so that the blue tubing is towards the patient.
3-4
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Set up and Pre-use Preparation
The HT70 Ventilator will perform to specication when Newport recommended breathing circuits and exhalation valves are used. Newport cannot guarantee the safe use of breathing circuits or exhalation valves that are not recommended.
For use with a third party humidifier:
When using a humidier with the ventilator, be sure to follow the manufacturer’s instructions for use.
See Figure 1:
1. Attach a bacteria lter to the Gas Output connector on the HT70.
2. Locate the short piece of 22 mm ID circuit tubing. Connect the end that includes the prox line pressure port to the inlet port of the humidier.
3
1. Prox Line Pressure Port
rml aA
ce en l
i
t
S
e s Re
ln a C ce
pcA et
n r t a Ex le r e w o P
a uln a M in ta l n If o
se s nt h gB i
1
2
r
S
A
G
PU U T
O
c
ev c e D i t r e l A
X
O .
PR
E
N
LI
.X
H
E
E V L
A V
T
6
5
2. Inlet port
3. Humidifier
4. Short section breathing circuit
5. Bacteria filter
6. Gas output connector
3
4
Figure 1
OPRHT70-2 F0613
3-5
5
5
OPR360U A0509
5-1
3
Set up and Pre-use Preparation
3. Attach the other end of the breathing circuit to the bacteria lter on the HT70 Gas Output connector.
1. Breathing circuit
2. Outlet port
3. Humidifier
4. Pressure port
5. Prox line
6. Prox inline filter
7. Prox line connector
1
l mra
ee
A
nl c
t
Si
ese
R
ne a lc C
t p e
c
c
n ra e x Et l
r owe P
alM nu a
ioI a l nf nt
sr s
t i Bg neh
GAS
U UT O T
A
ece i v D tA r el
ROX.P
E
I
LN
.
XHE
V L E A V
P
7
6
4
5
2
3
Figure 2
See Figure 2:
4. Locate the 22 mm ID end of the main breathing circuit. Attach this end to the outlet port of the humidier chamber.
5. Attach the Prox Inline Filter with tubing to the Prox. Line connector.
6. Attach one end of the prox tubing to the Prox Inline Filter.
7. Attach the other end of the prox tubing to the prox line pressure port on the circuit tubing that is connected to the inlet port of the humidier chamber.
3-6
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Set up and Pre-use Preparation
1. Exhalation Valve connector
mrAla
c
neSi e l
eeR ts
cn l a e C
tA p
c
lE e axt n r r e ow P
a a luM n l nn iofI at
s egh n ri tB s
eD cv i e t r lAe
XP .RO
LINE
H
E .X
V
L E
VA
SG
A OU PT U T
c e
2. Exhalation Valve tubing
3. Patient wye connector
4. Breathing circuit
5. Exhalation valve
3
1
5
2
4
3
Figure 3
See Figure 3:
8. Attach one end of the exhalation valve tubing (smallest clear tubing) to the Exh. Valve connector.
9. Attach the other end of the exhalation valve tubing to the connector on the exhalation valve at the end of the circuit.
10. Attach Flex Tube (not shown), if used, on the patient wye connector.
11. If using the on-airway ow sensor, plug the connector into the front panel port. Attach the ow sensor with the blue tubing toward the patient onto the patient connection of the circuit. Use the 15/22mm adapter supplied with the ow sensor to attach to the patient interface as needed. Use the circuit clip supplied with the ow sensor to secure the ow sensor lines to the main breathing circuit tubing.
See Figure 4 for completed set up with humidier.
12. If a temperature probe is used, insert probes into ports at either end of the tubing that connects the humidier and the patient wye connector.
OPRHT70-2 F0613
3-7
5
5
OPR360U A0509
5-1
3
Set up and Pre-use Preparation
Figure 4
13. Perform the Circuit Check. If the circuit includes an end cap, use it during the rst step of the Circuit Check. See Section 4 for instructions.
For use with an HME (artificial nose):
See Figure 5:
1. Attach a bacteria lter to the Gas Output connector on the HT70.
2. Locate the 22 mm ID end of the breathing circuit. Attach this end to the bacteria lter.
3. Attach the HME to the patient wye connector.
4. Attach a pressure tee if the on-airway ow sensor is not used.
5. Attach Flex Tube, if used, to the patient side of the pressure tee/ or on-airway ow sensor adapter.
6. Attach the Prox Inline Filter with tubing to the Prox. Line connector.
7. Attach one end of the prox tubing to the Prox Inline Filter.
8. Attach the other end of the prox tubing to the port located on the wye connector (on-airway ow sensor in use) or to the pressure tee adapter on patient side of the HME.
3-8
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Set up and Pre-use Preparation
1. Gas Output connector
ra m l
e
A
c n eSil
teRse
a c lneC
c pec t
Ex rnte al
r w o e P
l an u a M
tI an inol f
n s h est i
Brg
i e cv e D r e l t A
ASG U PT U
O T
A
X.R O
P
E NI
L
XH.E
V L A V E
9
1
2
8
7
2. Bacteria filter
3. Breathing circuit
4. HME
5. Flex Tube (optional)
6. Pressure tee adapter (patient side)
7. Prox line tubing
8. Prox inline filter
9. Prox Line connector
6
3
3
4 5
Figure 5
See Figure 6:
9. Attach one end of the exhalation valve tubing (smallest clear tubing) to the Exh. Valve connector.
10. Attach the other end of the exhalation valve tubing to the connector on the exhalation valve.
1. Breathing circuit
ml r a
A
nc e l e
i S
R tees
cla eC n
ce
lae x r E nt
ew oP r
uaMa ln l in nat I of
g Br tn si esh
G SA O T UU P T
Ac pt
c vDe ei
t r e l A
X R P .O LINE
.
H EX
V V L A E
4
2. Exhalation valve
3. Exhalation tubing
4. Exhalation valve connector
3
2
OPRHT70-2 F0613
1
Figure 6
3-9
5
5
OPR360U A0509
5-1
3
Set up and Pre-use Preparation
11. If using the on-airway ow sensor, plug the connector into the front panel port. Attach the on-airway ow sensor with the blue tubing toward the patient to the wye connector of the circuit. Use the 15/22 mm adapter supplied with the on-airway ow sensor to attach to the patient interface as needed. Use the circuit clip supplied with the on-airway ow sensor to secure the on-airway ow sensor lines to the main breathing circuit tubing.
See Figure 7 for completed setup for HME with ow sensor.
Figure 7
12. Perform the Circuit Check. If the circuit includes an end cap, retain it for use in the Circuit Check. See Section 4 for instructions.
Using the On-airway Flow Sensor
The Newport Flow Sensor is a disposable, single patient use on­airway ow sensor that can be used for pediatric to adult patients. Use the 15/22 mm adapter supplied with the on-airway ow sensor to attach to the patient interface as needed. Use the circuit clip supplied with the on-airway ow sensor to secure the on-airway ow sensor lines to the main breathing circuit tubing.
Set up your patient breathing circuit as described above for the appropriate usage (with or without humidier or HME). Plug the on­airway ow sensor connector into the port on the front panel of the HT70 Plus. Attach the on-airway ow sensor to the patient end of the breathing circuit with the blue tubing toward the patient.
3-10
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Set up and Pre-use Preparation
When it is connected the ventilator will recognize the on-airway ow sensor and enable these added features:
• Flowtrigger  • Exhaledvolumemonitoring  • HighTidalVolumealarm  • Highandlowexpiratoryminutevolumealarms
3
Caution:
circuit, be sure to remove the ow sensor from the circuit. This will protect the ow sensor from medication build up.
When giving nebulizer treatments through the ventilator
Connect Optional Accessories
Connect optional accessories such as the Air/Oxygen Mixer, Low Flow Oxygen Reservoir or Biolter, to the right side of the HT70 at the Fresh Gas Intake Port.
For External DC Power use, connect the DC Auto Lighter Cable into the external power supply input on the rear of the Power Pac battery pack.
WARNING
of HT70. Use only approved accessories.
Air/Oxygen Entrainment Mixer
The Air/Oxygen Entrainment Mixer (p/n MXL70A-XX-XX) which is used to blend atmospheric air with 50 psi medical grade oxygen, attaches to the Fresh Gas Intake Port on the Filter Cover (right side of the ventilator). Attach the mixer’s high pressure hose to an active source of medical grade, 100% oxygen, before attaching the mixer to the ventilator. The mixer should not be attached to the ventilator before the Circuit Check is completed. Make sure that the oxygen source gas is always on while using the mixer during ventilation.
Do not block the Fresh Gas Intake Port on the right side
(optional)
Use the mixer’s control knob to adjust oxygen enrichment of the gas delivered to the patient circuit from 21 to 100 percent. The mixer does not need re-adjustment when the PEEP and Bias Flow settings or patient’s minute volume change.
Use a calibrated oxygen monitor with alarms (such as the one that is built-in to the ventilator) to assure that the O ventilator into the patient circuit matches the prescribed value.
2 that is delivered by the
OPRHT70-2 F0613
3-11
5
5
OPR360U A0509
5-1
3
Set up and Pre-use Preparation
Check the mixer lter (p/n FLT3209P) at the time of setup and weekly during use. Replace when dirty. See section 8 for instructions.
Pneumatic Requirements: Oxygen 35-65 psig (2.4 to 4.5 Bar) Maximum accuracy 40-50 psiq (2.7-3.4 Bar)
Note:
must be medical grade,
Oxygen source gas
Ca elnc
Accept
GAS
TOU PUT
EXH. VALVE
Oxygen Mixer
100% oxygen.
Low Flow Oxygen Reservoir (optional)
The Low Flow Oxygen Reservoir (p/n RSV3215A)
Low Flow Oxygen
Reservoir
which is used to blend
Ca eln
atmospheric air with 1-10
c
Accept
L/min of medical grade oxygen, attaches to the Fresh Gas Intake Port on the
GAS
TO U PU T
EXH . VALVE
Filter Cover (right side of the ventilator). Attach oxygen supply tubing between the oxygen owmeter and the small bore connector on the reservoir. Attach the reservoir to the ventilator.
Use the oxygen owmeter to adjust oxygen enrichment of the gas delivered to the patient circuit. When the Low Flow Reservoir is in use, the percent of oxygen that is delivered from the ventilator into the breathing circuit will vary, depending on the use of Bias Flow and PEEP, the delivered minute volume and the percent O
2 of the source gas.
3-12
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Set up and Pre-use Preparation
3
Oxygen Supply Flow vs Desired % of Oxygen
6
6
1
Min Vol i 25 Liters
2
Min Vol i 20 Liters
3
Min Vol i 15 Liters 7 Desired % of Oxygen Enrichment
4
Min Vol i 10 Liters
Oxygen Supply Flow vs Desired % of Oxygen
For Use With PEEP
2 3 4 51
7
5
Min Vol i 5 Liters
6 Oxygen Supply Flow, L/min
For Use Without PEEP
1 2 3 4 5
7
1
Min Vol i 25 Liters
2
Min Vol i 20 Liters
3
Min Vol i 15 Liters 7 Desired % of Oxygen Enrichment
4
Min Vol i 10 Liters
5
Min Vol i 5 Liters
6 Oxygen Supply Flow, L/min
Use these graphs for estimating the liter ow of supplemental oxygen needed to attain a particular O2 percentage. Note that the rst graph applies when PEEP is on (and NIV off) and the second graph applies when PEEP is off. The graphs are also printed on the Instructions for Use that are packaged with the reservoir.
To keep a constant percentage of O adjustment when PEEP is turned on or off or the Bias Flow setting is
2, the owmeter will need re-
changed (PEEP and NIV on) or the patient’s minute volume changes.
Use a calibrated oxygen monitor with alarms (such as the one that is built-in to the ventilator) to assure that the percentage of O
OPRHT70-2 F0613
2 that
3-13
5
5
OPR360U A0509
5-1
3
Set up and Pre-use Preparation
is delivered by the ventilator into the patient circuit matches the prescribed value.
Pneumatic Requirements: Oxygen 0-10 L/min
WARNING
oxygen will result in lower oxygen percentage levels than what is printed on the graphs. Use a calibrated oxygen monitor to verify the level of oxygen enrichment.
Using an oxygen concentrator in place of medical grade
Caution:
malfunction and damage.
DC Auto Lighter Power Adapter (optional)
The DC Auto Lighter Power Adapter (p/n ADP3203P) is used to plug the HT70 into any vehicle’s DC power outlet (12 VDC to 16 VDC). The external DC source of power will not only power the ventilator but also charge both batteries in the Internal Dual Battery System.
This cable can also be used to connect the HT70 Ventilator to other external batteries (12 VDC to 16 VDC) when combined with an alligator clip adapter that is available from common electronics outlets such as Radio Shack.
To use the DC Auto Lighter Power Adapter, plug the adapter’s pinch­release power plug into the external power supply input located on the lower left corner of the Power Pac battery pack. Ensure that the cord is to the right of the plug and that it locks in place securely. Plug the other end of the adapter into a vehicles autolighter power outlet (12 VDC to 16 VDC).
To remove the DC Auto Lighter Power Adapter from the external power supply input, gently pinch the connector to release the locking pin and then pull the plug out.
Water in the oxygen supply can cause equipment
Aequitron Remote Alarm Cable
The Aequitron Remote Alarm Cable allows the user to connect the HT70 Ventilator to the Aequitron 6217 remote alarm assembly.
This cable must be connected to both the nurse call and USB output ports on the ventilator. The other end of this cable must be connected to the original extension cable (not to the Aequitron unit itself).
The ventilator must be congured to “Norm Close” (see Figure 4 on page 4-18). In addition, see the General Notes regarding nurse call connectivity on page 1-5.
Complete a self-test after the cable has been installed and at regular intervals to ensure the Aequitron system is operating as intended. A self-test consists of inducing an alarm and conrming the Aequitron unit emits an audio alarm, and also conrming the audio alarm ceases once the alarm in the ventilator has been reset.
3-14
OPRHT70-2 F0613
Section 4:
Navigating
Navigating
5
5
OPR360U A0509
5-1
Section 4:
Navigating
the Screens
Touchscreen (Graphical User Interface)
Layout ............................................................ 4-1
Primary Screen Buttons and Displays ............. 4-1
Ventilator Settings Adjustment .........................4-4
Start Up Screen Navigation (Standby
Condition Only) ............................................. 4-4
Circuit Check Button .................................... 4-5
How to Perform a Circuit Check ..............4-5
If the Circuit Check Fails ..........................4-6
Activate Preset Button ................................. 4-7
How to Use a Preset .................................4-7
Start Ventilation Button ............................... 4-7
Alarms Screen Navigation .................................4-8
Main Screen Navigation ....................................4-9
More Screen Navigation .................................. 4 -11
More Screen Details ....................................4 -13
Events .......................................................4-13
Trends .......................................................4-14
Waves .......................................................4-15
O Calibrate O
Utility Screen .................................................... 4-18
Utility Screen Details ...................................4 -19
Time/Altitude Screen ..............................4-19
Customize Settings Screen ....................4-20
Custom Presets ..................................4-21
Back Up Ventilation ............................4-22
2 Cylinder Data Screen .........................4-16
2 Monitor ...............................4-17
5
5
OPR360U A0509
5-1
5
Domain Navigation ...........................................4-23
Hospital ........................................................ 4-23
Transport ..................................................... 4-24
Basic ............................................................ 4-25
5
5
OPR360U A0509
5-1
Navigating the Screens
Touchscreen (Graphical User Interface) Layout
The touchcreen display is color coded so that it is very easy to differentiate between basic ventilation settings which are green, alarm settings which are red and monitored values which are yellow (on blue background). The “More” and “Utility” settings are in blue.
If the power save feature is enabled, the HT70’s touch screen will go dark if not touched for 2 minutes (while running on internal battery). To bring it back into full view just touch the screen or a membrane button.
Primary Screen Navigation
The buttons for accessing Alarms, Main, and More screens as well as the buttons for selecting mode and breath type are consolidated along the left margin. Monitored values are displayed across the bottom margin and the pressure bar graph rises and falls along the right. This leaves plenty of room for the display in the middle of the screen.
The name of the active screen is written in larger letters than the other two. To change to a different screen view, just touch one of the other screen buttons. There is no need to press Accept.
4
Primary Screen Buttons and Displays
Refer to Figure 1
1/10
2
3
4 6/7
5
Figure 1
1.
Start up Screen selection buttons:
Condition there are three additional buttons in the Message and Alerts display window. They disappear when the Start Ventilation button is touched. See Startup Screen Navigation on page 4-4 for a full description.
While in Standby
9
8
OPRHT70-2 F0613
4-1
5
5
OPR360U A0509
5-1
4
Navigating the Screens
2. buttons to open these screens in the center display area. Simply touch the desired button and the screen changes. You do not need to press Accept. See following pages for details on these screens.
additional screens: Trends*, Events, Waves+ and Utility. You can return to the Main screen from any of these screens by touching the screen selection button labeled “Main”.
3. type button to toggle between Volume Control and Pressure Control. Then press Accept. Touch the mode button to scroll through the selections A/CMV, SIMV or SPONT. Press Accept to conrm.
4. or button on the touch screen and the center panel displays an explanation of the features or controls. Touch any button except the help button again to close the tutorial.
5. located at the bottom of the screen. To choose and change the parameters displayed, touch any one of the buttons to select it. The full monitoring screen appears and shows all twelve monitored parameters. Touch the parameter that you would like to display. That parameter will automatically appear in the Monitor Data display button that was selected. You can arrange these parameters in the order you want. The Monitor screen will remain, showing the monitored parameters, until you make an alternate selection, press a different screen button (Alarms, Main or More) or for 2 minutes to allow you to view and check all monitored values. The values on the monitor screen do not update while the screen is displayed.
Screen selection buttons
NOTE:
Breath Type/Mode Selection buttons:
Help button:
Monitor Data display buttons:
From the “More” screen you can choose to view these
Touch the help button, then touch any feature
: Touch the Alarms, Main or More
Monitor Data buttons are
Touch the breath
NOTE:
Accept to complete your change.
6. is determined by the Domain selection. The HT70 user interface can be setup in one of three Domains: Basic, Transport and Hospital. The Basic Domain is a simplied screen for use in the longterm care or homecare setting. The Transport Domain is an specically designed to assist
4-2
For monitor display selections, you do not have to push
Domain button:
The level of accessibility for HT70 controls
*not available in the HT70S models
+only available in the HT70 Plus models
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
during transport applications. The Hospital and Transport Domains provide full access to all ventilation and alarm settings as well as to the special screens and menus. Full access is recommended for use in acute care settings and for the initial setup of patients in long term care before switching to Basic Domain. See Domain Navigation on pg 4-23 for more details.
7. the Utility screen. When Auto Lock is enabled, the touchscreen will automatically lock 20 seconds after the last button was touched. When this occurs, a lock icon will appear in the lower right corner in place of the Domain button. To unlock the screen, simply touch and hold the lock icon for 3 seconds. The screen will automatically relock when no buttons have been touched for 20 seconds.
8. It indicates dynamic pressure in the patient circuit with a green bar that rises and falls. The High and Low Pressure Alarm settings are indicated with red lines and the peak pressure of the last breath is indicated with a green line.
Auto Lock button:
Pressure Bar:
The pressure bar appears on every screen.
The Auto Lock function is enabled from
4
9.
Battery Charge Level display:
the upper right corner of the screen. This icon indicates the percent of charge level of the battery that is currently operating the ventilator - either the Power Pac battery pack (blue icon) or the Backup Battery (red icon).
10. are three startup buttons that appear in this area. They disappear when the ventilator is in ventilating condition. While in Standby or while ventilating, messages, including alarm alerts, are displayed in the message display area of the screen in order of priority. Active alarms are displayed rst followed by latched alarms that have not been cleared by the user.
Reset button (located on the top of the panel). As you clear each alarm message, the alarm message with the next priority will appear in the message display area. Continue pressing to clear all alarm messages. Press and hold for 3 seconds to clear all at once.
Messages and Alerts display:
NOTE:
To clear alarm messages, press the Alarm Silence
A battery icon is displayed in
In Standby Condition there
OPRHT70-2 F0613
4-3
5
5
OPR360U A0509
5-1
4
Navigating the Screens
Ventilator Settings Adjustment
Most ventilator settings are changed by touching a parameter to highlight it, using the up and down arrows to change the set value, and then pressing Accept to conrm the change. Pressing and holding the arrow button makes the change happen more rapidly. Other parameters are changed by toggling, meaning that the same button is touched repeatedly to change the value, then conrmed by pressing Accept. In either case, you can press the Accept button after each setting change or make multiple changes and then press Accept.
If you decide not to make the changes you started, press the Cancel button instead of pressing Accept or just wait and the values will revert back to the original settings.
The mode and mandatory breath type selections determine which Main Screen breath delivery parameters are active and available for adjustment. Follow these steps when setting up ventilation:
1. While the Main screen is in view, start by selecting the mode and mandatory breath type along the left margin of the screen.
2. Adjust all Main screen parameters that are visible.
3. Press the Accept button to implement the mode/breath type and relevant parameters change.
4. Visit the More screen to turn NIV ON or OFF.
5. When NIV is turned ON while PEEP is in use, the Bias Flow adjustment window appears so that you can also adjust Bias Flow.
6. Adjust the other relevant ventilation parameters on the More screen.
7. Visit the Alarm screen to adjust/check alarm parameters.
Startup Screen Navigation (Standby Condition only)
When the HT70 is turned On, it goes through a short self-test before entering the Standby Condition. Make sure that you hear the alarm sound and see the LEDs light during the short self-test. While in Standby, the Startup Screen is available. In addition, ventilator settings can be adjusted. The Startup screen includes three buttons in the top message area: Start Ventilation, Circuit Check and Activate Presets.
4-4
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
4
Startup Screen
Circuit Check Button
Perform a Circuit Check each time the breathing circuit or exhalation valve is replaced. While the ventilator is in Standby Condition, touch the Circuit Check button and follow the instructions on the screen. It is a simple two-step automated process. The Circuit Check is not available during ventilation.
How to perform a Circuit Check
1. Touch the Circuit Check button at the top of the touchscreen and follow the on-screen instructions.
2. For Step 1, occlude the patient connection end of the circuit. (Do not use a test lung.)
3. Press the Accept button to conrm and start the Circuit Check.
OPRHT70-2 F0613
4-5
5
5
OPR360U A0509
5-1
4
Navigating the Screens
4. For Step 2, open the patient connection end of the patient circuit.
5. Press the Accept button to continue the Circuit Check.
6. If the test passes, the message “Circuit Check PASSED Press Accept to Conrm” will be displayed.
7. When the Circuit Check is completed, adjust patient settings as needed and touch the Start Ventilation button when you are ready to begin ventilation.
8. To cancel the Circuit Check and return to the Startup Screen, press the Cancel button.
If the Circuit Check fails:
o The message “Circuit Check FAILED Press Accept to Continue” will be displayed.
o Press the Accept button to return to Startup Screen.
o Check that all breathing circuit connections are properly connected and leak free.
o Verify that the Air Oxygen Entrainment Mixer is not attached to the Fresh Gas Intake port.
o Then touch the Circuit Check button to redo the test.
If the Circuit Check fails repeatedly, try a different circuit.
WARNING
ventilation may result. Use an alternate method of ventilation. Contact Newport Medical Technical Support.
NOTE:
retained after power down.
4-6
Do not use the HT70 if the Circuit Check fails, inadequate
The Circuit Check results are logged into the Event History and
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
Activate Preset Button
4
While in Standby, ventilation and alarm settings can be adjusted manually or they may be adjusted using custom or default pre­programmed parameter sets (presets) for Adult, Pediatric and Infant patients. From the Activate Preset screen you can also touch the New Patient Flag button to enter a “New Patient” ag in the Events History log. You must be in Standby Condition to activate a preset or enter a new patient ag.
How to Use a “Preset”
1. Touch the Activate Preset button.
2. Touch one of the six preset buttons that appear on the screen. Default Preset settings are based on factory set defaults. Custom Preset settings must be established by the user.
3. Press the Accept button to implement the settings.
NOTE:
Alarms buttons and adjust ventilation and alarm parameters.
To establish custom parameters for a preset you must rst set all parameters and alarms as you want them for your patient. Then go to the More/Utility/Custom Settings/Set Custom Preset to select a custom preset button. For instructions on how to customize the Custom Patient Presets see page 4-20.
Start Ventilation button
Touch the Start Ventilation button to exit the Standby Condition and begin ventilation.
NOTE:
the HT70 for the rst time.
WARNING
prior to starting ventilation. Note that during Standby Condition the
OPRHT70-2 F0613
If you do not want to use a Preset, press the Main, More and
Be sure to review all sections of this manual before you use
Ensure that all settings are appropriate for the patient
4-7
5
5
OPR360U A0509
5-1
4
Navigating the Screens
monitored O2 is not representative of the set O2 or the O2 that will be delivered during ventilating condition. After starting ventilation, use a calibrated oxygen monitor (such as the one that is built-in to the ventilator) to verify that the Air Oxygen Entrainment Mixer setting or liter ow attached to the Low Flow Oxygen Reservoir is delivering the prescribed O2.
Alarms Screen Navigation
(May be accessed in all Domains, Standby or Ventilating Condition except where noted) Refer to Figure 2.
See Section 9, Specications, for ranges and more details for each alarm and setting.
To enter Alarms Screen:
To set Alarm limits (not available in Basic Domain):
limit button to activate (highlight) it, use the up and down arrow panel buttons to adjust the limit, then press Accept to conrm changes or press Cancel to return to original settings. You can set multiple limits before pressing Accept.
1.
Touch the Alarms button.
2.
Figure 2
Touch an Alarm
3.
1. Settable Alarms
#P (High Pressure) #Min Vol (High Minute Volume) $P (Low Pressure) $Min Vol (Low Minute Volume) #RR (High Respiratory Rate) Apnea (time adjustment) #O2 (High O2) #VTE (High Exp. Tidal Volume) $O2 (Low O2)
+ only available on the HT70 Plus model when the on-airway ow sensor is in use
4-8
OPRHT70-2 F0613
+
5
OPR360U A0509
5-1
Navigating the Screens
4
2. Alarm Loudness Level 1-10 (10 is loudest)
use the up and down arrow panel buttons to adjust the loudness level, then press Accept. Always set the alarm loudness level high enough to ensure that caregivers will hear any alarm. If necessary attach a remote alarm system (see Utility Screen for remote alarm selections) to ensure that the caregiver can always hear an alarm when it sounds.
3. Alarm Quickset
when there are no active alarms violations, Alarm Quickset will automatically set the alarm limits. Touch this button to enter the Alarm Quickset screen, then press Accept to activate or Cancel to return to the Alarms screen.
When activated, Alarm Quickset monitors settings for 30 seconds and then sets the alarms. If an alarm occurs during the monitoring period, Quickset is canceled. During the 30 second period the touch screen will not respond unless an alarm occurs or the Cancel button is pressed.
Alarm Quickset will only activate when in ventilating condition.
– During Ventilating (not Standby) condition,
– Touch this button,
Main Screen Navigation
(May be accessed in all Domains, Standby or Ventilating Condition)
See Section 9, Specications, for ranges and details for all parameters.
The ventilation parameters that are displayed on the Main Screen are determined by the mode and breath type that are selected with the Mode/Breath Type buttons along the left margin of the touchscreen. Select the Mode and Breath Type rst to see the Main Screen parameters that need adjustment.
OPRHT70-2 F0613
4-9
5
5
OPR360U A0509
5-1
4
Navigating the Screens
Main Screen
To set ventilation parameters (not available in Basic Domain):
Touch a ventilation parameter button to activate (highlight) it, use the up and down arrow panel buttons to adjust the setting, then press Accept to conrm changes or press Cancel to return to original settings. You can set multiple parameters before pressing Accept.
List of all possible ventilation parameter settings on the Main Screen
VT (Tidal Volume) RR (Respiratory Rate) PEEP PS (Pressure Support)* Ptrig PC (Pressure Control) Flow (in volume control) Flow Trig i time
+
NOTE:
Trig and Ptrig settings are active and the rst available detection of patient effort (either ow or pressure ) will trigger a breath.
NOTE:
time are available to be set. The VT (tidal volume) is the controlling setting. Flow and i time setting are inversely related to each other. Changing either Flow or i time causes the inversely related setting to be recalculated to match the new Flow or i time setting and the VT remains steady.
For example, if you change the Flow rate to a higher setting, the i time will decrease to meet the new Flow setting. If you change the
4-10
When an on-airway ow sensor is detected, both the Flow
For Volume Control mandatory breaths, VT, Flow and i
*not available on HT70S models
+ only available on HT70 Plus models
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
i time to a longer time period, the Flow rate will decrease to meet the new i time setting.
Also while in Volume Control, changing the Flow waveform setting on the More Screen will cause the Flow to change on the Main screen. A square wave will use a lower ow and a descending ramp waveform will use a higher ow. Always recheck the settings and monitored values after changing the Flow waveform.
More Screen Navigation
(Hospital and Transport Domain, Standby and Ventilating Condition) Refer to Figure 3.
See Section 9, Specications, for ranges and more details for each parameter.
4
1. 2. 3. 4.
5. 6. 7. 8.
9.
Press the Accept button to conrm changes made to any of the following parameters.
1. Slope Rise touch button, use up/down s t to adjust from 1 - 10 (1 is slowest) (Pressure Control and Pressure Support breaths)
2. PS Exp Threshold * touch button, use up/down s t to adjust from 5 - 85% (5% is longest) (Pressure Support breaths)
10. 11. 12.
Figure 3
OPRHT70-2 F0613
*not available on the HT70S models
4-11
5
5
OPR360U A0509
5-1
4
Navigating the Screens
3. PS Max i Time* touch button, use up/down s t to adjust from 0.1 - 3.0 s (Pressure Support breaths)
4. Flow [Square or Descending] touch button to select square or descending ramp ow (Volume Control breaths)
5. NIV touch button to select NIV (Non Invasive) function ON or OFF for all breaths
6. Bias Flow touch button, use up/down s t to adjust from 3 to 30 L/min. Bias Flow is delivered during exhalation phase when PEEP is ON.
NOTE:
button, you can press Cancel at anytime to revert to previous setting.
7. O set up oxygen cylinder content tracking. (see page 4-16 for details)
8. Calibrate O screen for the internal oxygen sensor (see page 4-17 for details)
For the above settings, if you have not pressed the Accept
2 Cylinder Data Screen touch button to access screen to
2 Monitor touch button to access calibration
9. Events touch button to access Events screen (see page 4-13 for details)
10. Trends* touch button to access Trends screen (see page 4-14 for details)
11. Wave s screen (see page 4-15 for details)
12. Utility Settings touch button to access Utility screen, see description on the following pages
4-12
+
touch button to access Waves
*not available on the HT70S models
+only available on the HT70 Plus models
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
More Screen Details
Events Screen
(Hospital and Transport Domain, Standby and Ventilating Condition)
The Events Screen displays the last 1,000 recordable events. When a new event occurs, the oldest event is cleared. Use the up and down arrow panel buttons s t to scroll through the list of Events. Recordable events include Circuit Check, parameter changes, alarm activate/deactivate, date/time changes, alarm silenced, alarm cleared, calibrations, screen brightness changes, new patient, and power On/Off.
4
To record when a new patient is started, touch the ‘Activate Presets’ button while in the Startup Screen (Standby Condition) and then press the “New Patient Flag” button. The Event History log will record a “New Patient” entry.
As you scroll through Events, the time and parameters in use during that Event are displayed.
NOTE:
Screen.
Date and time format selection is located on the Utility
OPRHT70-2 F0613
4-13
5
5
OPR360U A0509
5-1
4
Navigating the Screens
Trends Screen Navigation*
(Hospital and Transport Domain, Standby and Ventilating Condition)
The Trends Screen displays trended data for monitored parameters. Push the up and down arrow panel buttons to move the cursor to the right or to the left.
As the cursor line moves across the graph, the time display will indicate the time at that point on the graph and the numbers on the line will show the value for each monitored parameter. The yellow vertical bars represent time that the ventilator was off and not collecting data.
There are four sets of parameters to display.
To Select a Trend Set:
To change the displayed parameters, touch the Tren ds Set button on top of the Trends graph to scroll through these choices:
1) Peak pressure, Mean pressure and PEEP
2) Tidal Volume, Resp. Rate Total, Minute Volume
3) Peak Flow, Power Pac, Back up Batt
4) Power Pac Temp, Back up Batt Temp, Internal Temp
*not available on the HT70S models
4-14
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
To Adjust the Time Scale
You can display the trends in time frames of 1, 2, 4, 8, 24 or 72 hours. To scroll through the time scale selections simply touch the hours button on the top of the Trends graph.
Waves Screen
(Hospital and Transport Domain, Standby and Ventilating Condition)
+
4
The Waves Screen displays real time graphics for Pressure, Volume and Flow. If the on-airway ow sensor is not installed, only the pressure graph will be displayed.
Touch the ‘Choice’ button at the top of the central panel to select 1, 2 or 3 graphs for display. When only one or two graphs are chosen, you can select which waveform you want displayed by toggling the ‘Wave’ button(s) at the top of the central panel.
Freeze.
the current waveform. The up/down arrows will move the cursor across the waveforms and display the value for the displayed graph at that point in time. To unfreeze, touch the same button again.
Time Scale (x-axis).
screen anywhere in the central panel. The time axis will turn yellow and you can use the up/down arrows to increase/decrease the time scale. Press Accept to save the new time scale.
Amplitude Scale (y-axis).
simply touch the central panel anywhere in the desired graph twice.
Touch the ‘Freeze’ button to the left of the graphs to freeze
To change the time scale simply touch the
To change the height of any waveform
OPRHT70-2 F0613
+only available on the HT70 Plus models
4-15
5
5
OPR360U A0509
5-1
4
Navigating the Screens
The rst touch will highlight the time scale in yellow and the second touch will highlight the y-axis. The up/down arrows can now be used to increase/decrease the height of the y-axis. Press Accept to save the change.
Oxygen Cylinder Data Screen
(Hospital and Transport Domain, Standby and Ventilating Condition)
Estimated cylinder use time can be displayed on the monitor screen if the relevant cylinder data is entered, the O2 Cylinder Monitor is Enabled, and the O2 Monitor on the Utility Screen is Enabled.
Touch More, touch O pressure units and pressure for the cylinder you are using. Finally, touch O2 Cylinder Monitor to Enable the function, and press Accept. The HT70 will then calculate the estimated duration of your oxygen supply and warn you before your cylinder may run out. Allow the calculated display to stabilize for several minutes before starting a transport and ensure that the estimated time is sufcient for the planned trip or outing.
NOTE:
estimate only. It can be affected by many factors such as leaks in the O2 path. Do not rely solely on this measurement. Check the oxygen cylinder remaining pressure level frequently to conrm actual oxygen consumption rate.
Size: Toggle to the size of oxygen cylinder in use: D, E, H, M, K, 100 L, & 150 L.
The oxygen cylinder use time displayed in the monitor is an
2 Cylinder Data Screen, then enter the size,
4-16
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
Units: Toggle this button to use the desired pressure units for the oxygen cylinder (psi or ATM).
Cylinder Pressure: Touch this button and use the up/down arrows to enter the current pressure in the oxygen cylinder (300-2450 psi or 25-175 AT M).
O
2 Cylinder Monitor: Toggle this button to ‘Enabled’ and press
accept when all data has been entered.
4
NOTE:
function to operate.
Calibrate O
(Hospital and Transport Domain, Ventilating Condition)
This screen allows the user to calibrate the internal oxygen sensor. Either a single point or a two point calibration can be done. This can be done while on a patient if they can tolerate the desired calibration point (room air or 100% oxygen). Touch this button to go to the Calibrate O2 Mon screen.
You must enable the O2 Monitor (Utility Screen) for this
2 Mon Screen
O
2 Cal, 21% O2. Touch this button and follow the onscreen
directions to calibrate at room air. Ensure that no oxygen device is connected to the air intake port on the right side of the ventilator.
O
2 Cal, 100% O2. Touch this button and follow the onscreen
directions to calibrate at 100% oxygen. Ensure that 100% oxygen is being delivered to the air intake port on the right side of the ventilator. Newport suggests using the Low Flow Oxygen Reservoir with 10 L/min. of medical grade 100% oxygen connected to it.
OPRHT70-2 F0613
4-17
5
5
OPR360U A0509
5-1
4
Navigating the Screens
Utility Screen
(Hospital and Transport Domain, Standby and Ventilating Condition except where noted) Refer to Figure 4
Access the Utility Screen through the More Screen as shown above. See Section 9, Specications, for ranges and more details for each parameter.
1. 2. 3. 4.
5. 6. 7. 8.
11.
9.
10.
12.
Figure 4
Press Accept button to conrm changes made to any of the following parameters.
1. Auto Lock touch button to select Autolock function Enabled or Disabled
2. Language touch button to scroll through language selections
3. cmH pressure units
4. Power Save touch button to select Power Save function ON or OFF
5. Export Data only available in Standby Condition. Touch to download data to Flash USB Drive or computer. This is not available while ventilating. Follow the instructions on the screen.
6. Time/Altitude touch button to access Time and Altitude screen (see page 4-19 for details)
2O or mbar touch button to select cmH2O or mbar
4-18
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
7. Remote Nurse Call touch button to select from Norm Open (normally open), Norm Close (normally closed), and Respironics nurse call systems
For use of the Aequitron Remote Alarm System, use Norm Close.
8. Comm touch button to select from Bernoulli enabled, Vuelink enabled or Comm inactive for central monitoring systems
9. Custom Settings touch button to access Backup Ventilation (BUV) and Custom Presets screens (see page 4-20 for details).
10. O Enabled or Disabled
11. Software the software version installed is displayed here
12. Hours the number of hours of ventilator operation is displayed here
2 Monitor touch button to select to O2 Monitor
4
Utility Screen Details
Time/Altitude Screen
(Hospital and Transport Domain, Standby and Ventilating Condition)
Access the Time/Altitude Screen through the More/Utility Screens as shown above.
OPRHT70-2 F0613
4-19
5
5
OPR360U A0509
5-1
4
Navigating the Screens
Touch a button to activate (highlight) it, use the up and down arrow panel buttons to adjust the setting, then press Accept to conrm changes or press Cancel to return to original settings. You can set multiple parameters before pressing Accept. The Hours are displayed in military style; 1 - 12 for AM and 12 - 24 for PM.
Altitude can be displayed in meters or feet. The Altitude adjustment is only for the accuracy of the on-airway ow sensor. If the ow sensor is used, ensure that the altitude is set.
NOTE:
delivery in altitudes up to 15,000 feet. The patented twin micro-piston system is a volume displacement technology that will deliver the set volume regardless of the altitude.
Custom Settings Screen
(Hospital and Transport Domain, Standby and Ventilating Condition)
The HT70 ventilator automatically maintains accurate volume
Access the Custom Settings Screen through the More/Utility Screens as shown above.
This screen allows access to available customization options for BUV (Backup Ventilation) function and the Preset functions. The following options are available on the Custom Settings Screen:
1. Set Custom Presets touch to access menu to dene the available Custom Presets
2. BUV Settings touch to access the BUV Settings Screen
4-20
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
4
Set Custom Presets
To store customized presets for different customized patient protocols:
1. Turn the ventilator on.
2. Make all the changes on the ventilator that you require for your protocol. Be sure to check the settings in the More screen and the Alarms screen.
(from the Custom Settings Screen)
3. Once you are satised with the settings for your patient protocol, return to the More Screen/Utility Screen/Custom Settings screen and touch the Set Custom Presets button.
4. Touch the desired Custom Preset (infant, pediatric or adult).
5. Touch the Accept button to conrm and save your choice.
You can choose “Activate Presets” button to enable your preset parameters from the Startup Screen (see page 4-17) in the Standby condition.
NOTE:
Custom Presets are retained after power down.
OPRHT70-2 F0613
4-21
5
5
OPR360U A0509
5-1
4
Navigating the Screens
BUV Settings Screen
This screen allows you to customize Backup Ventilation (BUV) to your institution’s policies or to revert to factory default for the Backup Ventilation parameters. You can also choose to link Backup Ventilation to the Low Minute Volume alarm (LMV), the Apnea alarm, or to both alarms. See Section 9, Specications, for ranges and more details for each parameter.
(from the Custom Settings Screen)
Press Accept button to conrm changes made to any of the parameters.
The BUV Settings Screen includes the following adjustable parameters:
1. Minimum RR (respiratory rate)
2. Rate Factor (set rate will be multiplied times this number to determine the BUV breath rate)
3. SPONT (mode) delta P (pressure target above set PEEP for breath delivery)
4. SPONT (mode) i-time (i time for BUV breaths delivered while in SPONT mode)
5. BUV Link (with LMV [low minute volume] alarm, Apnea alarm, or both)
6. Revert to Defaults
4-22
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Navigating the Screens
Domain Navigation
The Newport HT70 is designed with the exibility of useful application in the acute care as well as longterm care environments. To make the product easy and safe to use in the full spectrum of applications, we have divided care into three domains: hospital – which means acute care, Transport – anytime the user is on the move with battery and supplemental oxygen supplies, and Basic – for the long term/ home care environments.
Touch the Domain button in the lower right corner of the touch screen to scroll through the domain choices, Hosp (Hospital), Trans (Transport) and Basic (Basic). Press Accept to change to the new domain. If the panel is locked this button is not visible until the panel is unlocked.
Hospital Domain
4
The Hospital Domain has full access to all features and screen selections available on the HT70. This manual describes all of the features and screen selections found in the Hospital Domain. The Transport and Basic Domain features and screen selections work identically to the Hospital Domain with the limitations as noted below.
OPRHT70-2 F0613
4-23
5
5
OPR360U A0509
5-1
4
Navigating the Screens
Transport Domain
This Domain gives preference to transport related monitoring features like O
2 Cylinder use duration and estimated battery use time.
4-24
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Basic Domain
Navigating the Screens
4
This is a simplied screen for use in a homecare or sub-acute facility type environment.The central parameter platform is replaced with a digital clock unless the Main or Alarm buttons are touched. The Breath Type and Mode settings are displayed and the user has access to the Main and Alarm screens and can view all monitored data by touching one of the monitor buttons at the lower margin of the screen. The More screen is not available. No settings can be changed within this Domain. This helps simplify operation and protect against accidental settings changes. If a setting change is needed or access to the More screen is required, simply use the Domain button in the lower right corner of the screen to toggle to the Hospital Domain. Press Accept to conrm your selection.
OPRHT70-2 F0613
4-25
Section 5:
Operating
Operating
5
5
OPR360U A0509
5-1
Section 5:
Operating
the Ventilator
Quick Check Procedure .................................... 5 -1
Introduction ................................................... 5-1
Equipment Needed ....................................... 5-1
Pretest Inspection ........................................ 5 -1
Set Up ........................................................... 5-1
Standard Test Settings ................................ 5-2
Quick Check Procedure ............................... 5-2
Pass / Fail Check Off Sheet ......................... 5-4
Patient Setup Procedure ...................................5-5
Troubleshooting Guide ......................................5 -7
5
OPR360U A0509
5-1
Operating the HT70 Ventilator
Quick Check Procedure
Introduction
This procedure is intended to assist qualied operators to establish a routine program for verifying proper HT70 operation. Perform this Quick Check procedure each time the ventilator is prepared for new patient use. The Quick Check procedure should be performed every 6 months as part of the routine maintenance procedure to conrm proper ventilator operation. Use a copy of the Quick Check Procedure Check-off Sheet found at the end of this section to record the results of each check.
HOMECARE PROVIDERS: This procedure should be performed prior to delivery of the HT70 to a patient’s home.
5
NOTE:
when not used for two minutes. Just touch the screen anywhere to bring it back into view.
WARNING
Equipment needed:
o 1 liter test lung with resistor (LNG800P) o Patient breathing circuit with exhalation valve
Pre-test Inspection
1. Inspect the Air Intake Filter through the lter cover on the right side of HT70. Replace the lter if it is dirty. See Section 8 for replacement.
2. Examine the test lung and patient circuit to ensure that there are no holes that will cause leaks.
3. Verify that the AC power supply is in good condition.
Set Up
1. Connect the AC power supply to an AC power source.
2. Verify that the External Power LED is lit.
3. Turn the ventilator on and verify that the audible alarm sounds and the LEDs light during the self test.
4. Connect a breathing circuit with exhalation valve and on-airway ow sensor if used.
5. Set the ventilator to the following Standard Test Settings and press Start Ventilation.
If Power Save is On, the screen will go to sleep (go blank)
Do not use the HT70 if it fails the Quick Check Procedure.
Standard Test Settings
Mode A/CMV Breath Type Volume Control NIV Off Tidal Volume 500 mL i time 1.0 sec RR 15 b/min Ptrig 1 cmH
OPRHT70-2 F0613
2O / mbar
5-1
5
OPR360U A0509
5-1
5
Operating the Ventilator
Flow Trig Off
$Low P Alarm 5 cmH2O / mbar #High P Alarm 99 cmH2O / 97 mbar $Low Min Vol alarm 0.01 L (minimum setting) #High Min Vol alarm 50 L (maximum setting with NIV Off) #VTE Alarm 1.00 L
PEEP/CPAP 0 cmH2O / mbar
Quick Check Procedure
1. Perform a Circuit Check
o Touch the Circuit Check button at the top of the touchscreen and follow the on-screen instructions. o For Step 1, occlude the patient connection end of the circuit. (Do not use a test lung.) o Press the Accept button to conrm and start the Circuit Check. o For Step 2, open the patient connection end of the patient circuit. o Press the Accept button to continue the Circuit Check. o If the test passes, the message “Circuit Check PASSED Press Accept to Conrm” will be displayed.
If the test fails, the message “Circuit Check FAILED Press Accept to Continue” will be displayed.
o Press the Accept button to return to Startup Screen. o Check that all breathing circuit components/connections are properly connected and leak free. o Verify that the Air Oxygen Entrainment Mixer is not attached to the Fresh Gas Intake port. o Then touch the Circuit Check button to redo the test.
If the Circuit Check fails repeatedly, try a different circuit.
WARNING
ventilation may result. Use an alternate method of ventilation. Contact Newport Medical Technical Support.
o Connect the test lung to the patient connection of the circuit. Touch Start Ventilation.
2. No External Power Alarm Check
o Disconnect the AC power supply. Verify that there is an audible alarm and the alarm LEDs in the HT70 handle ash. Verify that the External Power LED turns off, and the Message Area turns yellow and displays the No External Power alarm message. Conrm that HT70 continues to ventilate. o Press the Alarm Silence/Reset button and conrm that its LED lights yellow, the audible alarm is muted and the message area returns to black. o Press the Alarm Silence/Reset button again and conrm that the alarm message clears. o Reconnect the AC power supply. Verify that the External Power LED lights green.
Do not use the HT70 if the Circuit Check fails, inadequate
5-2
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Operating the Ventilator
3. Alarms and Indicators Check
High #P Alarm
o Set the High Pressure alarm limit to 20 cmH an audible alarm sounds and the High Pressure message displays and that inspiration ends when pressure reaches the high limit. Set the High Pressure alarm limit back to 99 cmH verify that the audible alarm stops and the alarm message remains. Press the Alarm Silence/Reset button to clear the alarm message.
Low $P Alarm
o Disconnect the test lung from the breathing circuit and verify that after two breaths an audible alarm sounds and the Low Pressure Alarm message displays. o Attach the test lung to the breathing circuit and verify that the audible alarm ceases and the alarm message remains. Press Alarm Silence/Reset button to clear the message.
4. Pressure Gauge / PEEP Check
o Verify that the pressure gauge moves up and down with each breath. o Select PEEP and Peak Paw to display in each of two Monitor Data buttons. o Adjust PEEP to 5 cmH displays a PEEP value of 4 to 6 cmH o Select Pressure Control and set PC at 20 cmH Monitor Data button displays a Peak Paw of 17 to 23 cmH
2O. Verify that the Monitor Data button
2O. Reduce PEEP to zero.
2O/mbar. Verify that
2O /mbar and
2O. Verify that the
2O.
5
5. Volume/Minute Volume/Respiratory Rate Monitor Check
o Change Breath Type back to Volume Control and conrm Tidal Volume is set to 500. o Select VT, Min Vol and RR Tot to display in each of three Monitor Data buttons. Verify that VT= 450-550, Min Vol = 6-9 and RR Tot = 13-17.
6. Power Pac Battery Pack and Backup Battery Check
o Unplug the AC power supply, clear the alarm with the Alarm Silence/Reset button. Verify that HT70 continues to ventilate and the Power Pac battery gauge (blue icon) reads at least 80%. If battery charge level is not sufcient, plug into external power to fully charge the Internal Dual Battery System. o Remove the Power Pac battery pack. Verify that HT70 continues to ventilate, the alarm sounds, the alarm LEDs light and the message in the message area indicates that the Backup Battery is in use. o Verify that the battery gauge is now red (for secondary backup battery) and reads at least 80%. If the secondary backup battery charge level is not sufcient, re-insert the Power Pac battery and plug into external power to fully charge the system. o Replace the Power Pac battery pack and verify that the audible alarm clears but the message remains. o Reconnect the AC power supply into the Power Pac battery pack and conrm that the External Power LED turns green.
OPRHT70-2 F0613
5-3
5
OPR360U A0509
5-1
5
Operating the Ventilator
o Press the Alarm Silence/Reset button repeatedly until all alarm messages are cleared.
7. Brightness Check
o Press the Brightness button and verify that it scrolls through 4 levels of brightness. Set the brightness at desired level.
THIS CONCLUDES THE QUICK CHECK PROCEDURE
HT70 Ventilator Quick Check Procedure
Pass / Fail Check-Off Sheet
Preparation for Use Tests Indicate result for each test
1. Circuit Check Procedure Pass _____ Fail ______
2. No External Power Alarm Check Pass _____ Fail ______
3. Alarms & Indicators Check High Low
#P Alarm Pass_____ Fail _____
$P Alarm Pass _____ Fail ______
4. Pressure Gauge / PEEP Check Pass _____ Fail ______
5. Volume/Minute Volume/ Pass _____ Fail ______ Respiratory Rate Monitor Check
6. Power Pac Battery Pack and Pass _____ Fail ______ Backup Battery Check
7. Brightness Check Pass _____ Fail ______
The ventilator is ready for operation when all tests have been successfully completed.
Note any comments on inspection of unit, corrective action taken, or recommendations for further action.
Completed by: ___________________________ Date: _________
Facility: _______________________________________________
Serial #: _______________________________________________
Unit hours: _____________________________________________
5-4
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Operating the Ventilator
Patient Set Up Procedure
5
WARNING
Section 1 prior to using the ventilator.
All ventilator controls and alarm limits must be appropriate for the patient’s condition, according to the therapy prescribed by a physician.
1. Press the momentary power switch located on the back of the ventilator to turn the ventilator On. The ventilator performs a brief self-test to ensure proper microprocessor function. During the self-test, verify that the Startup Screen appears, the LEDs light and the audible alarm sounds briey.
2. Ensure the ventilator, patient circuit and accessories are assembled correctly. (See Section 3)
3. Make sure HT70 has passed the Quick Check Procedure.
4. Perform the Circuit Check. Resolve any issues. (See Section 4)
5. Set all parameters per physician’s prescription using manual adjustment or a Custom or Default Preset. See Section 9 for specications on all settings. a. Select mode and breath type. Then set all parameters on the Main Screen and relevant parameters on the More Screen. b. Select safe/appropriate alarm limits on the Alarms Screen. c. Ensure that the Alarm Loudness is set loudly enough for the alarm to be heard under all circumstances.
Review all of the General Warnings and Cautions in
6. Place a test lung on the patient end of the breathing circuit and press the Start Ventilation button on the touchscreen.
7. Ensure that the ventilator starts operation appropriately.
NOTE:
and volume delivery for pressure breaths will be different than they will be on the patient. Pressure Support breaths will not perform the same way they do on patients. And PEEP may cause auto-triggering. These differences may cause nuisance alarms during this step.
8. When you are ready, remove the test lung and attach the patient connection of the breathing circuit to the patient interface.
9. Monitor the patient settings and check for appropriate alarm settings.
OPRHT70-2 F0613
While ventilating a test lung, peak pressure for volume breaths
5-5
5
OPR360U A0509
5-1
5
Operating the Ventilator
10. Verify that the patient trigger icon lights each time the patient initiates a spontaneous breath. Re-adjust sensitivity (Ptrig or Flow trig if using on-airway ow sensor on HT70 Plus model) as necessary to ensure comfortable triggering without auto- triggering. When using PEEP while ventilating a patient with an airway leak, set NIV to ON and adjust Bias Flow to stabilize the PEEP and eliminate auto-triggering at reasonable trigger settings.
11. Closely monitor the patient and ventilator to ensure appropriate oxygen delivery, and adequate oxygenation and ventilation.
WARNING
ventilating patients.
WARNING
appropriately, the patient should be taken off the ventilator immediately and connected to an alternate method of ventilation. Contact your health care provider or physician immediately.
NOTE:
into an external power source when it is available, even when the ventilator is not in use.
Always ensure adequate monitoring is in place when
If, at any time, the patient is not responding to ventilation
To ensure best battery performance, always plug the HT70
5-6
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Operating the Ventilator
Troubleshooting Guide for Newport HT70
Ventilation and alarm settings are determined by the physician’s prescription. Consult with your physician on ventilation and alarm settings.
• AlarmmessagescanbeclearedwiththeSilence/Resetbutton.
• ReviewtheentireOperatingManualforfulluserinstructions.
• Notethattheminutevolumealarmsareexpiratoryminute
volume alarms when the on-airway ow sensor is in use and they are inspiratory minute volume alarms when the on-airway ow sensor is not in use.
• BackupVentilationincreasestherespiratoryrateinA/CMVand
SIMV and provides pressure controlled breaths in SPONT mode. It may be caused by the violation of a Low Minute Volume Alarm or Apnea Alarm. Resolve the alarm to resolve the back up ventilation.
Problem/ Area
of Concern
Touchscreen doesn’t respond to touch
Touchscreen is dark/blank
Need to view all monitored values
Alarm indicator(s) lit but not blinking/Alarm message in the message window
Water accumulating in the breathing circuit
Probable
Cause
Screen is locked by Auto-lock function (set on Utilities screen)
Using Basic Domain and the parameters are changed to displays rather than buttons
Power Save is ON (Utility Screen) and the screen has “gone to sleep” to save power
Alarm condition is resolved. (Alarm is “latched”.)
Gas is cooling as it travels through the circuit tubing
Touch the lock icon in the lower right corner of the screen for 3 seconds to unlock the screen
Touch the Basic Domain button in the lower right corner of the screen to toggle the domain to Hospital (Hosp), then press Accept.
Touch the screen or push any button to “wake up” the screen
Touch any monitored value at lower edge of the screen
Push Silence/Reset button to clear the indicator and messages one at a time. Hold 3 seconds to clear all messages at once.
Keep tubing as short and warm as possible between the humidier and the airway so that water remains in the vapor state Keep tubing away from cold surfaces Do not aim a cooling fan at the tubing If appropriate, use a heated wire circuit
Resolution
5
OPRHT70-2 F0613
Water trap needs to be emptied
Empty the water trap frequently
5-7
5
OPR360U A0509
5-1
5
Operating the Ventilator
Problem/ Area
of Concern
Audible alarm is too quiet
Power Pac battery doesn’t last long enough
Probable
Cause
Alarm Loudness is set too low
Not fully recharged
Ventilator settings/ patient condition (a large leak during pressure ventilation, a high level of bias ow setting, or an aggressively breathing patient) demand more gas delivery than the standard settings.
Power Pac needs to be replaced
Resolution
Touch Alarms, touch Alarm Loudness. Adjust setting to a higher number and press Accept.
Connect the Power Pac to external AC or DC power for at least 3 hours between uses
The Power Pac battery is functioning normally. Carry at least one extra fully charged Power Pac battery (Part Number: BAT3271A) for ventilator dependent patients and for patients whose ventilation pattern requires higher battery power consumption.
Contact a Newport Authorized Service Provider or Newport Medical Technical Service Department for assistance. Email: techservice@ ventilators.com Tel: 1.714.4275811 Ext. 500
Check Circuit or Prox Line Alarm
5-8
Circuit disconnect
Humidity in Proximal Line
No proximal lter in place
Inspiratory Flow is too low, circuit pressure does not rise fast enough when the breath starts
Reconnect the circuit
Change where Proximal Line is connected to the circuit. Move it from the connection at the patient wye (wet environment) to an adapter placed directly on the inlet of the humidier chamber(dry environment).
Insert approved proximal lter (part number: HT6004701)
As appropriate, increase Flow or change ow pattern in Volume Control or speed up Slope Rise in Pressure Control/Pressure Support.
OPRHT70-2 F0613
5
OPR360U A0509
5-1
Operating the Ventilator
5
Problem/ Area
of Concern
Circuit Check fails
Green “Ext Power” indicator on panel does not light when HT70 is plugged into external AC (wall) or DC (external battery or auto lighter outlet) power
Probable
Cause
Leak in breathing circuit
Connecting test lung instead of occluding patient connection of the circuit in step 1.
Incorrect assembly of circuit/ exhalation valve
Incompatible circuit/exhalation valve
Oxygen connected directly to the circuit
Ventilator needs service
No external power is reaching HT70. HT70 is running from the Internal Battery System.
Resolution
Tighten all circuit & water trap connections
Trim ends of the proximal and exhalation valve tubings to enable a tighter t
Check the integrity of/replace the exhalation valve diaphragm
Remove test lung and occlude the patient connection of the circuit during step 1.
See Operating Manual for proper assembly
Contact Newport Medical to verify if circuit is compatible. Email: clinical@ ventilators.com
Use the Low Flow Oxygen Reservoir or 50 psi Air Oxygen Mixer. Connect oxygen devices to the Fresh Gas Intake port after, not before, the Circuit Check is complete
Contact a Newport Authorized Service Provider or Newport Medical Technical Service Department for assistance. Email: techservice@ ventilators.com Tel: 1.714.4275811 Ext. 500
Check power cord connections (L-shaped pinch t connector on rear of HT70 should angle toward the midline of the ventilator, not away. See diagram on sticker.)
Check that power outlet is active
External battery is depleted, plug into another external battery, auto lighter or AC power
Check / replace the fuse in the External Battery System or Auto DC Cable
Contact a Newport Authorized Service Provider or Newport Medical Technical Service Department for assistance. Email: techservice@ ventilators.com Tel: 1.714.4275811 Ext. 500
OPRHT70-2 F0613
5-9
5
OPR360U A0509
5-1
5
Operating the Ventilator
Problem/ Area
of Concern
Auto­triggering (at a typical trigger setting)
(green trigger light illuminates when there is no patient effort)
In SIMV or SPONT modes, some or all breaths seem to last extra long
Occlusion Alarm/ Sustained Occlusion Alarm
The patient will not be ventilated. Manually ventilate the patient until this is resolved.
High (Peak) Pressure Alarm
This alarm violation stops breath delivery until pressure drops
Probable
Cause
Circuit and/or airway leak
Leak is causing ow to remain high during pressure support breath delivery, so the PS Exp. Threshold (ow cycling off threshold) is never met
The patient’s expiratory gas pathway is occluded or partially occluded
Coughing / need for airway care or bronchodilator treatment
Secretions too dry due to inadequate humidity
Resolution
Check for and resolve leaks if unintentional.
If airway leak is intentional, turn on NIV and adjust Bias Flow/Trigger settings so that the patient can trigger effectively without auto-triggering.
Consider using the Flow Sensor and Flow Trigger.
Eliminate circuit leaks
Set the PS Max i-Time control to limit the breath delivery to a reasonable inspiratory time
Evaluate everything in the patient’s pathway of exhalation to determine what is causing resistance and resolve the issue
Change HME and/or expiratory lter if used
Change the exhalation valve
Unkink expiratory drive line
Replace ow sensor
Perform suctioning/airway care, or if due and prescribed, give prescribed bronchodilator treatment
Use heated humidier with appropriate temperature setting and keep tubing warm
5-10
Pneumatic nebulizer inline
cont.
Contact Newport Medical for assistance Email: clinical@ventilators.com Tel: 1.714.4275811 Ext. 123
cont.
OPRHT70-2 F0613
Loading...