Newport Breeze E150 Service Manual

-
Newport
Medical
Instruments
Newport
Breeze
150
Ventilator
Service
Instructions
SERI 50
Rev.
B
[serial #901
0081
60
and later]
EU REPRESENTATIVE: Newport Medical Instruments Att: Robert Brink Barton-on-Humber
18 Pasture Rd.
South Humberside Tel: 44.4682.31311
DN18
5HN,
England Fax: 44.1652.633399
NEWPORT MEDICAL INSTRUMENTS
Post Office Box 2600
Newport Beach, California 92658 USA
800.451.3111
949.642.3910
949.548.3091
FAX
Customer Service Ext.
282
Table
of
Contents
Section One
1.1
INTRODUCTION
1.2
NEWPORT BREEZE VENTILATOR
PRESSURE RELIEF VALVE EXHALATION VALVE WATER TRAP ASSEMBLY STAND AND POLE ASSEMBLY
1.3
OPTIONAL ACCESSORIES
1.4
NEWPORT BREEZE SPECIFICATIONS
Section
Two
2.1
GENERAL
2.2
POWER SOURCE
2.3
SOLENOIDS AND VALVES
2.4
AIR I OXYGEN MIXER
2.5
PRINTED CIRCUIT BOARDS
2.6
MANOMETER ASSEMBLY
2.7
THEORY
OF
OPERATION
2.7.1
GENERAL
Table
of
Contents
2.7.2
VOLUME LIMITED MODE
2.7.3
PRESSURE LIMITED MODE
2.7.4
SPONTANEOUS MODE
2.7.5
NEBULIZER
Section
Three
3.1
GENERAL CALIBRATION PROCEDURE
3.2
PREVENTATIVE MAINTENANCE
3.3
OPERATIONAL VERIFICATION AND
CALIBRATION PROCEDURE
Drawinas
FIGURE 1-1
VENTILATOR FRONT
VIEW
FIGURE 1-2 PRESSURE
REILEF VALVE
FIGURE 1-3A EXHALATION VALVE ASSEMBLY
FIGURE 1-3B PB EXHALATION VALVE BLOCK ASSEMBLY
FIGURE 1-4
WATER TRAP ASSEMBLY
FIGURE 1-5 STANDING POLE ASSEMBLY
Table
of
Contents
Drawinqs
FIGURE 1-6 MIXER AUXILIARY FLOW OUTLET ASSEMBLY
FIGURE 2-1
AIR
/
OXYGEN MIXER
FIGURE 2-2 MANOMETER ASSEMBLY
FIGURE 2-7 PNEUMATIC TUBING DIAGRAM
FIGURE 4-1 BASE PLATE (TOP VIEW)
FIGURE 4-2 BEZEL (FRONT VIEW)
FIGURE 4-3 FRONT PANEL ASSEMBLY
FIGURE 4-4BACK PANEL (INSIDE VIEW)
FIGURE 4-5 BACK PANEL (OUTSIDE VIEW)
FIGURE 4-6 FLOW CONTROLLER AND LO/HI POTENTIOMETER ASSEMBLY
FIGURE 4-7 SPONTANEOUS FLOW ASSEMBLY
FIGURE 4-8 I.T.
/
R&R POTENTIOMETER ASSEMBLY
FIGURE 4-9 PEEP
/
PIP ASSEMBLY
FIGURE 4-1
0
PNEUMATIC BRACKET ASSEMBLY
1
.I
INTRODUCTION
The Newport Breeze Ventilator can be used to ventilate infant,
pediatric or adult patients. It's versatility provides for a safe and
effective use throughout the entire patient range. This service manual
(Part No. SER150)
contains the necessary
information to enable a
qualified service person
to maintain and
service the
Newport Breeze ventilator. This manual is a companion to
the operating manual
(Part No. OPR150).
Both the operating and service manuals should be read before any service or repair is attempted on the Breeze.
Together the manuals will provide the service person with enough
information to operate, maintain and service the Breeze to an assembly or subassembly level.
We recommend that you do not attempt to make repairs of a complex
nature such as electronic circuit board repairs unless you are a
.fully
qualified service person and possess the appropriate test equipment.
We also recommend that you contact your authorized service
representative or the service department at
NEWPORT MEDICAL
INSTRUMENTS INC.
for circuit board and other complex repairs, as
the
Newport Medical personnel are properly trained and equipped to
provide this service. If at any time the Breeze cannot be restored to properly operating
condition the unit should be returned to the NMI factory for the service or repairs as needed.
The material in this manual is organized as follows:
Section One provides a brief description of the
Newport Breeze
Ventilator, options and accessories.
Section Two covers mechanical drawings and parts. Section Three covers acceptance tests and calibration procedures.
Parts, Accessories and Optional equipment may be ordered by part number through your local NMI dealer or by contacting NMI customer service.
Customer Service
800 451 .3111
or
949 642 .3910 ext. 282
Fax 949
548 3091
Technical Support
800.451 3111
or
949.642 391 0 ext. 500
Fax 949
548 3091
Newport Medical Instruments Inc.
Corres~ondence:
Post Office Box 2600 Newport Beach, CA 92658
Ship ~roduct to:
760
W,
16th Street, Bldg
B
Costa Mesa, CA 92627
1.2
NEWPORT VENTILATOR
Fig
1.1
Shows the Newport Breeze Ventilator.
The Breeze may be operated in either Volume limited or Pressure
limited modes.
Volume limited operation has four basic modes:
Assist Control with Sigh (AIC Sigh) Assist Control
(NC) Synchronized l ntermittent Mandatory Ventilation (SIMV) Spontaneous
Pressure limited operation has three basic modes:
Assist Control Plateau
(NC) PLATEAU)
Synchronized
l
ntermittent Mandatory Ventilation Plateau (SIMV) PLATEAU) Spontaneous
The manual inflation button will deliver a pressure limited or a volume
limited breath as defined by the ventilator settings.
Pressure Relief Valve
The pressure relief valve (Part No.
POP120A) is shown in Figure 1.2.
The pressure relief valve is located on the back of the unit. It functions
as a mechanical pressure safety device which provides protection for the patient against high pressure. This is accomplished by bleeding off gas and limiting pressure in the circuit at a preset pressure. This setting is achieved by turning the knurled knob on the pressure relief valve either clockwise or counterclockwise.
(The pressure relief valve is set to
0 cmH2O before being shipped from
the FACTORY.)
Exhalation Valve
The exhalation valve for adults only (Part No.
EXH100A) is shown in
figure
1.3a.
Figure 1.3b shows the exhalation valve assembly for use on neonatal
through adult range of patients.
f
3
T
\
<
[APNE
A
LO
PRESS
HI
PRESS
NEWPORT BREEZE
NEWPORT BREEZE VENTILATOR
I
I
I
I
I
I
I
f
\
MODE SET F10z FLOW
l
NSP TI ME SET RATE
vPSU a3NTrnL CONTROL
(1
(-1
(-1
(-1
0
LO
BAlT
LO
PRESS
HI
PRESS
SPONT
SlMV
\
0
SPONT. FLOW
0,
f
3
0
EXP. TIME (sec.)
1.
Vol (I.) I:E TOTAL RATE
cb
TRIGGER 0 EFFORT LEVEL
0
0(3c_>,
PRESSURE
@
(-1
MEAN
(cmH20)
@
\B
6,
L
1
0
MANUAL
NEWPORT MEDICAL INSTRUMENTS, INC.
\
I
1
BAT
PWR
SILENCE
Sel
PIP
SPONT
upm)
(set)
(bpm)
0
0
AJC
AJC
NC
SI
lMV
0
0
0
0,
PRESET
0
\
1
(cmH20) (cmH20)
HSG4Ol A
HOUSING
DIA200P
DIAPHRAGM ASSEMBLY
1
/
/
ADP520M
4
/"-
ADAPTER
BASE
SCREWS
(2)
TSKl
OOP
/
THUMB
SCREW
SCLl OOP
SCREW LOCK
Figure
1.3
(a)
Exhalation Valve Assembly (Adult)
The exhalation valve is mounted beneath the Breeze in the square area marked EXH VALVE. The exhalation valve controls the release of the exhaled gas to the atmosphere. The opening and closing of the valve is controlled by the pressure in the pneumatic drive line, marked
EXP. OUTLET, located beneath the unit and to the left of the
exhalation valve. The drive
line(EXP. OUTLET) is connected to the
exhalation valve .by means of a tubing.
Water Tra~ Assembly
The water trap assembly (Part No. WR300A) is shown in Figure 1.4. It
is mounted to the air inlet connector of the mixer on the rear of the unit. The water trap helps reduce condensates and contaminates in
the air supply to the unit.
A
drain valve assembly, located at the
bottom of the water trap is opened by pushing upwards to release
contaminates collected in the bowl. Contaminated air enters the
inside of the microfiber element
(.I micron filter) where condensates
and particles are removed from the air stream. Liquids (oil or water) collect on the elements outer surface and then fall to the bottom of the bowl.
1.2
STAND AND POLE ASSEMBLY
The stand and pole assembly (Part No. SPA450A) is shown in Figure
1.5. The ventilator is mounted on the stand and pole assembly and is secured onto the pole by a knob screw. The base has two locking casters and three non-locking casters. When the casters are in the lock position they assist in keeping the unit stationed in one location.
I
.3
OPTIONAL ACCESSORIES
Auxiliatv Flow Outlet Assemblv
The auxiliary flow outlet assembly (Part No.AXF150A) is an optional
accessory supplied upon request only. This feature allows the user to obtain gas flow for various needs. Figure 1.6 shows a unit with the auxiliary flow outlet assembly.
NOTE:
(AXF150A also includes FLOWMETER FLW300P with
nipple
and tubing.)
1.4
NEWPORT BREEZE SPECIFICATIONS
CONTROLS
Power Switch: F102: FLOW: lnspiratory Time
:
Rate
:
PEEP 1 CPAP
Peak Insp. Pressure Spontaneous Flow Trigger Level
Pneumatic Pressure Relief
Nebulizer
MODES:
VOLUME LIMITED
Assist CMV + Sigh Assist CMV
Sl MV Spontaneous
Manual Breath Alarm Loudness
Alarm Silence
ON-OFF / charge 21
% - 100%
3
-
120 LPM
0.1
-
3.0
sec.
1 - 150 BPM
0 - 60 cmH20
0
-
60 cmH20
0 - 40+ LPM
-10
-
+60 cmH20
0 - 120 cmH20
ON / OFF
PRESSURE LIMITED
SlMV Plateau
A,C Plateau
Spontaneous
2
sec. Max.
72
-
97 dBa
(at a distance of 1 meter.)
60
sec.
Audible & Visual
High Pressure Low Pressure
Apnea
Low Battery
Visual onlv
Battery Power
ALARMS
Audible only
Gas supply source failure
Power failure
System failure
DISPLAYS
F102 FLOW SPONTANEOUS FLOW
INSPIRATORY TIME RESPIRATORY RATE
5
(LOW CPAP ALRAM IN
SPONT.
MODE
ONLY),
10, 15, 30, 60
sec.
(EPROM
REV.
1.1
ONLY)
When
15
minutes of
operating time remains.
Indicates unit operating on battery power
TOTAL RESP. RATE TIDAL VOLUME
Volume Limited Mode only
EXPIRATORY TIME
Pressure Limited Mode only
I:E RATIO MEAN AIRWAY PRESSURE
BASE AIRWAY PRESSURE PEAK AIRWAY PRESSURE
HIGH PRESSURE
LOW PRESSURE
OTHER FEATURES
PRESSURE SENSING
PRESET
Proximal Airway pressure
Displays machine
settings
RS-232 /ALARM
Remote alarm outlet
HOURMETER
0
-
99,999
hours
POWER REQUIREMENTS
A C POWER
BAlTERY POWER
1
00/120/220/240
VAC
+/-
1
0%
47
-
63 Hz, 45W
Max.
Sealed Lead Acid
12V,
3.4AHr
1
Hour minimum
operating time
Recharge time
-
18
hours maximum
GAS REQUIREMENTS
AIR
OXYGEN
40 - 60 PSI (2.81
-
4.22 kg/cm2)
40 - 60 psi (2.81
-
4.22 kg/cm2)
DIMENSIONS & WEIGHT
HEIGHT
11 " (27.94 cm)
WIDTH
13" (33 cm)
DEPTH
12" (30.5 cm)
WEIGHT
32 Lbs. (14.6 kg)
SHIP WEIGHT
64 Lbs. (29 kg)
AGENCY REQUIREMENTS
Meets requirements
of
UL 544, CSA 22.2, IEC
601
-
1 and City
of
Los Angeles - Electrical
Testing Lab.
SECTION TWO
INSTRUMENT DESCRIPTION
2.1
GENERAL
A
simple functional description and principle of operation of the
Newport Breeze is presented in this section.
The function of various
components, different modes and various settings are also explained.
2.2
POWER SOURCE
The ventilator may be powered by an electrical supply of IOOvac,
120vac, 220vac or 240vac
+/-
10% at 47
-
63
Hz, 45W maximum. The voltage adjustment on the power entry module located at the rear of the unit must be set for the supply voltage.
2.3
SOLENOIDS AND VALVES
The Newport Breeze has four solenoids and four valves. The solenoids are electronically controlled and the valves are pneumatically controlled by the solenoids. The function, switching cycles and control
methods of these components are explained in Table 1 and Table 2.
The reader is advised to refer to pneumatic diagram Figure 2.7.
Loading...
+ 48 hidden pages