Newport 3150 User Manual

Page 1
Model 3150 High Power Temperature Controller
User’s Manual
Page 2
Page 3
Newport Corporation Telephone: 905-567-0390 Telephone: 02-924-5518 Telephone: 030-6592111 Telephone: 2-2769-9796 1791 Deere Avenue Facsimile: 905-567-0392 Facsimile: 02-923-2448 Facsimile: 030-6570242 Facsimile: 2-2769-9638 Irvine, CA 92714
France Japan Switzerland United Kingdom
Telephone: 949-863-3144 Telephone: 1-60 91 68 68 Telephone: 03-5379-0261 Telephone: 01-740-2283 Telephone: 01635-521757 Facsimile: 949-253-1800 Facsimile: 1-60 91 68 69 Facsimile: 03-5379-0155 Facsimile: 01-740-2503 Facsimile: 01635-521348
Belgium
Germany
Telephone: 016-402927 Telephone: 06151-36 21-0 Facsimile: 016-402227 Facsimile: 06151-36 21-52
Newport Corporation, Irvine, California, has been certified compliant with ISO 9002 by the British Standards Institution.
Page 4
Limited Warranty
Newport warrants that this product will be free from defects in materials and workmanship for a period of two years from the date of shipment. If any such product proves defective during the applicable warranty period, Newport, at its option, either will repair the defective product without charge for parts and labor or will provide a replacement in exchange for t he defective product.
In order to obtain service under this warranty, the customer must notify Newport of the defect before the expiration of the warranty period and make suitable arrangements for the performance of service. In all cases the customer will be responsible for packaging and shipping the defective product back to the service center specified by Newport, with shipping charges prepaid. Newport shall pay for the return of the product to the customer if the shipment is within the continental United States, otherwise the customer shall be responsible for all shipping charges, insurance, duties and taxes, if the product is returned to any other location.
This warranty shall not apply to any defect, failure or damage caused by improper use or failure to observe proper operating procedures per the product spec i f i c a tion or ope r a tors manual or im prope r or ina de qua te maintenance a nd care. Newport shall not be obligated to furnish service under this warranty 1) to repair damage resulting from attempts by personnel other than Newport’s representatives to repair or service the product; 2) to repair damage resulting from improper use or connection to incompatible equipment; 3) to repair damage resulting from operation outside of the operating or environmental specifications of the product.
THIS PRODUCT IS NOT DESIGNED FOR USE IN MEDICAL INSTRUMENTS WHERE MALFUNCTION OF SUCH PRODUCTS CAN REASONABLY BE EXPECTED TO RESULT IN PERSONAL INJURY OR DEATH.
NEWPORT’S LIABILITY FOR THE MERCHANTABILITY AND USE OF THE PRODUCT IS EXPRESSLY LIMITED TO ITS WARRANTY SET OUT ABOVE. THIS DISCLAIMER AND LIMITED WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL REPRESENTATIONS AND WARRANTIES EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR OF FITNESS FOR PARTICULAR PURPOSE, WHETHER ARISING FROM STATUTE, COMMON LAW, CUSTOM OR OTHERWISE. THE REMEDY SET FORTH IN THIS DISCLAIMER AND LIMITED WARRANTY SHALL BE THE EXCLUSIVE REMEDIES AVAILABLE TO ANY PERSON. NEWPORT SHALL NOT BE LIABLE FOR ANY SPECIAL, DIRECT, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES RESULTING FROM THE USE OF THIS PRODUCT OR CAUSED BY THE DEFECT, FAILURE OR MALFUNCTION OF THIS PRODUCT, NOR ANY OTHER LOSSES OR INJURIES, WHETHER A CLAIM FOR SUCH DAMAGES, LOSSES OR INJURIES IS BASED UPON WARRANTY, CONTRACT, NEGLIGENCE, OR OTHERWISE. BY ACCEPTING DELIVERY OF THIS PRODUCT, THE PURCHASER EXPRESSLY WAIVES ALL OTHER SUCH POSSIBLE WARRANTIES, LIABILITIES AND REMEDIES. NEWPORT AND PURCHASER EXPRESSLY AGREE THAT THE SALE HEREUNDER IS FOR COMMERCIAL OR INDUSTRIAL USE ONLY AND NOT FOR CONSUMER USES AS DEFINED BY THE MAGNUSOM-MOSS WARRANTY ACT OR SIMILAR STATE CONSUMER WARRANTY STATUTE.
©1996, Newport Corporation Irvine, California, USA Part No. 24594-01 IN-08972 Rev. C Printed 11-Apr-00
Page 5
EC DECLARATION OF CONFORMITY
Model 3150 High Power Temperature Controller
We declare that the accompanying product, identified with the “ ” mark, meets all relevant requirements of Directive 89/336/EEC and Low Voltage Directive 73/23/EEC.
Compliance was demonstrated to the following specifications:
EN50081-1 EMISSIONS:
Radiated and conducted emissions per EN55011, Group 1, Class A
EN50082-1 IMMUNITY:
Electrostatic Discharge per IEC 1000-4-2, severity level 3 Rated Emission Immunity per IEC 1000-4-3, severity level 2 Fast Burst Transients per IEC 1000-4-4, severity level 3 Surge Immunity per IEC 1000-4-5, severity level 3
IEC SAFETY:
Safety requirements for electrical equipment specified in IEC 1010-1
.
VP European Operations General Manager-Precision Systems Zone Industrielle 1791 Deere Avenue 45340 Beaune-la-Rolande, France Irvine, Ca. USA
______________________ ______________________
Alain Danielo Jeff Cannon
Page 6
Page 7
i
Table of Contents
1 General Information______________________________________________ 1
1.1 Introduction_______________________________________________________ 1
1.2 Product Overview __________________________________________________ 1
1.3 Available Options and Accessories ____________________________________ 2
1.4 Safety Terms and Symbols___________________________________________ 2
1.4.1 Terms _______________________________________________________________ 2
1.4.2 Symbols _____________________________________________________________ 3
1.5 General Warnings and Cautions ______________________________________ 3
2 System Operation_________________________________________________ 5
2.1 Introduction_______________________________________________________ 5
2.2 Installation________________________________________________________ 5
2.2.1 AC Power Considerations________________________________________________ 5
2.2.2 Tilt-Foot Adjustment ___________________________________________________ 5
2.2.3 Rack Mounting________________________________________________________ 6
2.2.4 Ventilation Requirements________________________________________________ 6
2.2.5 Power-Up Sequence ____________________________________________________6
2.2.6 Quick Start ___________________________________________________________ 6
2.3 Introduction to the 3150 Front Panel __________________________________ 7
2.3.1 Model 3150___________________________________________________________ 7
2.4 General Operation__________________________________________________ 8
2.4.1 Display Elements ______________________________________________________ 8
2.4.1.1 Static Fields ________________________________________________________8
2.4.1.2 Non-Editable Data Fields ______________________________________________8
2.4.1.3 Editable Data Fields __________________________________________________8
2.4.1.3.1 Changing Data Fields ______________________________________________ 8
2.4.1.4 Soft Keys __________________________________________________________9
2.4.2 Function Keys_________________________________________________________ 9
2.4.3 Menu Structure_______________________________________________________ 10
2.4.4 Master Display _______________________________________________________11
2.4.5 Main Menu __________________________________________________________ 11
2.4.6 Configure Menu ______________________________________________________12
2.4.7 System Configure Screen _______________________________________________ 12
2.4.8 Save/Recall Screen____________________________________________________ 14
2.4.9 Calibration Screen ____________________________________________________14
2.4.10 Configure Communications Screen _______________________________________ 14
2.4.10.1 Error Message Control _____________________________________________15
2.5 Rear Panel Familiarization__________________________________________ 15
2.5.1 GPIB Connector ______________________________________________________ 15
Page 8
ii
2.5.2 RS-232 Connector ____________________________________________________ 15
2.5.3 Input Power Connector ________________________________________________ 15
2.5.4 Ground Post
_____________________________________________________ 16
2.6 Warm Up and Environmental Consideration___________________________16
3 Temperature Controller Operation__________________________________ 17
3.1 General Features __________________________________________________17
3.2 TEC Safety Features _______________________________________________17
3.2.1 Conditions Which Will Automatically Shut Off the TEC Output ________________ 17
3.3 The TEC Connectors_______________________________________________18
3.3.1 TEC Interlock _______________________________________________________ 18
3.3.2 TEC Grounding Consideration __________________________________________ 18
3.4 TEC Operation____________________________________________________18
3.4.1 TEC Main Screen_____________________________________________________ 18
3.4.2 TEC Setup Screen ____________________________________________________ 19
3.4.2.1 The

and
Soft Keys ____________________________________________ 20
3.4.2.2 Sensor (Sens) ______________________________________________________ 20
3.4.2.3 Mode ____________________________________________________________ 21
3.4.2.3.1 Constant Temperature Mode (Const T)_______________________________ 21
3.4.2.3.2 Constant Resistance/Reference Mode (Const R)________________________ 21
3.4.2.3.3 Constant Current Mode (Const Ite) __________________________________ 21
3.4.2.3.4 Effects of Calibration on TEC modes ________________________________ 21
3.4.2.4 Gain _____________________________________________________________ 22
3.4.2.5 Limits____________________________________________________________ 22
3.4.2.5.1 TE Current Limit (Lim Ite) ________________________________________ 22
3.4.2.5.2 Temperature Limits (Lim Th and Lim Tl)_____________________________ 23
3.4.2.5.3 Resistance/Reference Limits (Lim Rh/vhI/ih and Lim Rl/vlO/il) ___________ 23
3.4.2.6 Tolerances (Tol Time and Tol Temp) ___________________________________ 23
3.4.2.7 Vpow, Ths, and Tamb _______________________________________________ 23
3.4.2.8 C1, C2, C3, and Ro _________________________________________________ 23
3.4.3 Thermistor and Thermistor Current Selection _______________________________ 24
3.4.3.1 Introduction _______________________________________________________ 24
3.4.3.2 Thermistor Range __________________________________________________ 24
3.4.3.3 Temperature Resolution______________________________________________ 25
3.4.3.4 Selecting Thermistor Current__________________________________________ 26
3.4.3.5 Selecting Thermistors _______________________________________________ 26
3.4.3.6 The Steinhart-Hart Equation __________________________________________ 27
3.4.3.7 Table of Constants __________________________________________________ 29
3.4.4 AD590 and LM335 ___________________________________________________ 29
3.4.4.1 General___________________________________________________________ 29
3.4.4.2 AD590 Sensor _____________________________________________________ 29
3.4.4.3 LM335 Sensor _____________________________________________________ 30
3.4.4.4 Determining C1 and C2 for the AD590 and LM335 ________________________ 31
3.4.4.4.1 One Point Calibration Method______________________________________ 31
3.4.4.4.2 Two Point Calibration Method _____________________________________ 32
Page 9
iii
3.4.5 RTD Sensors_________________________________________________________ 33
3.4.5.1 RTD Constants _____________________________________________________ 33
4 Principles of Operation ___________________________________________ 35
4.1 Introduction______________________________________________________ 35
4.2 TEC Controller Theory of Operation_________________________________ 36
4.2.1 TEC Interface ________________________________________________________ 36
4.2.2 Limit DAC __________________________________________________________ 36
4.2.3 Set Point DAC _______________________________________________________ 37
4.2.4 A/D Converter _______________________________________________________ 37
4.2.5 Sensor Select_________________________________________________________37
4.2.6 Difference Amplifier __________________________________________________37
4.2.7 Proportional Amplifier _________________________________________________ 37
4.2.8 Integrator ___________________________________________________________ 37
4.2.9 Bipolar Output Stage __________________________________________________ 38
4.2.9.1 Current Limiting____________________________________________________38
4.2.9.2 Current Limit Condition Sensing _______________________________________ 38
4.2.9.3 Voltage Controlled Current Source _____________________________________38
4.2.9.4 Voltage Limit Condition Sensing _______________________________________ 38
4.2.10 TEC Control Modes ___________________________________________________ 38
4.2.10.1 T Mode _________________________________________________________ 38
4.2.10.2 R Mode_________________________________________________________ 39
4.2.10.3 I
TE
Mode ________________________________________________________ 39
4.3 Microprocessor Board _____________________________________________ 40
4.3.1 Microprocessor_______________________________________________________40
4.3.2 Memory ____________________________________________________________41
4.3.3 Serial Interface _______________________________________________________ 41
4.3.4 Front Panel Interface___________________________________________________ 41
4.3.5 GPIB Interface _______________________________________________________ 41
4.4 Power Supplies____________________________________________________ 41
4.4.1 TEC Power Supplies___________________________________________________ 41
4.4.2 Main Supply _________________________________________________________ 42
5 Tips and Techniques _____________________________________________ 43
5.1 Introduction______________________________________________________ 43
5.2 TEC Limits_______________________________________________________ 43
6 Maintenance ___________________________________________________ 45
6.1 Introduction______________________________________________________ 45
6.2 Fuse Replacement _________________________________________________ 45
6.3 Cleaning _________________________________________________________ 45
7 Calibration_____________________________________________________ 47
Page 10
iv
7.1 Calibration Overview ______________________________________________47
7.1.1 Environmental Conditions______________________________________________ 47
7.1.2 Warm-Up___________________________________________________________ 47
7.2 TEC Calibration __________________________________________________47
7.2.1 Recommended Equipment______________________________________________ 47
7.2.2 Local Operation Thermistor Calibration ___________________________________ 48
7.2.3 Remote Operation Thermistor Calibration__________________________________ 48
7.2.4 Local Operation AD590 Sensor Calibration ________________________________ 49
7.2.5 Remote Operation AD590 Sensor Calibration_______________________________ 49
7.2.6 Local Operation LM335 Sensor Calibration ________________________________ 50
7.2.7 Remote Operation LM335 Sensor Calibration ______________________________ 50
7.2.8 Local Operation RTD Calibration ________________________________________ 51
7.2.9 Remote Operation RTD Calibration ______________________________________ 51
7.2.10 RTD Lead Resistance Calibration (Offset Null) _____________________________ 51
7.2.11 Local Operation ITE Current Calibration __________________________________ 52
7.2.12 Remote Operation ITE Current Calibration_________________________________ 52
8 Factory Service _________________________________________________ 55
8.1 Introduction ______________________________________________________55
8.2 Obtaining Service__________________________________________________55
9 Error Messages _________________________________________________ 59
9.1 Introduction ______________________________________________________59
10 Specifications _________________________________________________ 61
10.1 Temperature Controller Specifications________________________________61
10.2 General Specifications______________________________________________62
Tables
Table 1 - TEC Connector Pinout_____________________________________________________ 18 Table 2 - Comparison of Curve Fitting Equations _______________________________________ 28 Table 3 - Thermistor Constants______________________________________________________ 29 Table 4 - RTD Constants___________________________________________________________ 33 Table 5 - Recommended Test Equipment_______________________________________________ 48 Table 6 - Error Codes _____________________________________________________________ 59
Figures
Figure 1 - Model 3150 Front Panel____________________________________________________ 7 Figure 2 - A Sample Screen with Various Data Fields _____________________________________ 9 Figure 3 - Model 3150 Menu Structure________________________________________________ 10 Figure 4 - Master Display __________________________________________________________ 11 Figure 5 - Main Menu _____________________________________________________________ 12 Figure 6 - Configure Menu _________________________________________________________ 12
Page 11
v
Figure 7 - Save/Recall Screen _______________________________________________________ 14 Figure 8 - Communications Screen ___________________________________________________ 14 Figure 9 - Rear Panel______________________________________________________________ 15 Figure 10 - TEC Main Screen _______________________________________________________ 19 Figure 11 - TEC Setup Screens ______________________________________________________ 20 Figure 12 - Thermistor Temperature Range_____________________________________________25 Figure 13 - Thermistor Resistance versus Temperature____________________________________28 Figure 14 - AD590 Nonlinearity _____________________________________________________ 30 Figure 15 - 3150 Block Diagram _____________________________________________________ 35 Figure 16 - TEC Board Diagram_____________________________________________________ 36 Figure 17 - Microprocessor Board Block Diagram_______________________________________ 40 Figure 18 - Power Supply Block Diagram______________________________________________ 41
Page 12
Page 13
1
CHAPTER 1
1 General Information
1.1 Introduction
This chapter describes the features, options, accessories, and specifications of the Model 3150.
1.2 Product Overview
PRODUCT FEATURES
GPIB/IEEE 488.2 and RS-232C Interfaces Temperature Controller
350 Watt (up to 15A, and up to 28V, but not more than 350 Watts)
Ultra stable bipolar output
Thermistor, AD590, LM335, and Pt RTD sensors
The Model 3150 Temperature Controller is a result of Newport’s continuing commitment to provide advanced instrumentation at affordable prices.
High Power Temperature Controller Fulfills All Your Thermo Electric (TE) Cooling Needs
The 350 Watt Temperature Controller is offered to meet your most demanding TE cooling needs. It may be operated in one of three modes:
Constant Tempe rature
Constant Resistance
Constant TE Current
Short term stability is better than 0.0005°C while long term stability is better than
0.001°C. Four sensor types are compatible with this TEC:
Thermistors
AD590 series
LM335 series
100
Platinum RTDs
With the sensor’s calibration constants, the actual temperature is displayed in °C on the front panel.
Page 14
2 Chapter 1 General Information
Intuitive Controls and LCD Display Simplify Control and Test Procedures
Improved data presentation and system control are achieved using a LCD display. A MASTER display shows the entire system configuration as well as TEC status. “Soft Keys” guide you through initial system setup routines and operation. Real­time control of an output is accomplished either by entering the set point via the cursor keys or control knob. MENU and FUNC keys access saved system configurations and repetitive procedures. All controls are clearly marked and instructions easily understood for simple operation.
GPIB/IEEE-488.2 and RS232 Interfaces Gives Power to Remotely Control and Collect Data.
For ultimate control a GPIB/IEEE-488.2 interface is available. All control and measurement functions are accessible via the GPIB interface. In addition, a standard serial RS-232C port allows simpler interfacing to a PC. As your instrumentation needs change the Model 3150 Temperature Controller will adapt to all your new applications giving you the ultimate in flexible laboratory equipment.
1.3 Available Options and Accessories
Accessories 3150-02 Temperature Controller Cable 3150-04 Temperature Controller/Mount Cable
300-16 10.0 k
thermistor (± 0.2°C) 300-22 AD592CN IC Sensor 3150-RACK Rack Mount Kit
Newport Corporation also supplies temperature controlled mounts and other accessories. Please consult with your representative for additional information.
1.4 Safety Terms and Symbols
1.4.1 Terms
The following safety terms are used in this manual:
The
WARNING
heading in this manual explains dangers that could result in
personal injury or death.
The
CAUTION
heading in this manual explains hazards that could damage the
instrument.
Page 15
Chapter 1 General Information 3
In addition, a
NOTE
heading gives information to the user that may be beneficial in
the use of this instrument.
1.4.2 Symbols
The following symbols are used in this manual and on the instrument:
Power Off
Power On
!
Refer to the documentation.
Earth Ground
1.5 General Warnings and Cautions
The following general warning and cautions are applicable to this instrument:
WARNING
This instrument is intended for use by qualified personnel who
recognize shock hazards or laser hazards and are familiar with safety
precautions required to avoid possible injury. Read the instruction
manual thoroughly before using, to become familiar with the
instrument’s operations and capabilities.
WARNING
The American National Standards Institute (ANSI) states that a shock
hazard exists when probes or sensors are exposed to voltage levels
greater then 42 VDC or 42V peak AC. Do not exceed 42V between
any portion of the Model 3150 (or any attached detector or probe) and
earth ground or a shock hazard will result.
CAUTION
There are no serviceable parts inside the Model 3150. Work performed
by persons not authorized by Newport Corporation may void the
warranty. For instructions on obtaining warranty repair or service
please refer to Chapter 8 of this manual.
Page 16
Page 17
5
CHAPTER 2
2 System Operation
2.1 Introduction
This chapter describes how to operate the Model 3150. Unless otherwise noted, “3150” refers to the Model 3150.
2.2 Installation
CAUTION
Although ESD protection is designed into t he 3150, operation in a
static-fee work area is recommended.
2.2.1 AC Power Considerations
The 3150 can be configured to operate at a nominal line voltage of 100, 120, 220, or 240 VAC. Normally, this is done at the factory and need not be changed before operating the instrument. However, be sure that the voltage setting is correct on the power input module and correct fuses are installed per section 6.2 before connecting to an AC source. The 3150 is shipped set for 120 VAC and a caution sticker is placed on the input power connector.
CAUTION
Do not exceed 250VAC on the line input.
Do not operate with a line voltage that is not within ±10% of the
line setting. Too low of an input voltage may cause excessive
ripple on the DC supplies. Too high of an input voltage will cause
excessive heating.
WARNING
To avoid electrical shock hazard, connect the
instrument to properly earth-grounded, 3-prong
receptacles only. Failure to observe this precaution
can result in severe injury or death.
2.2.2 Tilt-Foot Adjustment
The 3150 has front legs that extend to make it easier to view the LCD display. To use them, place the 3150 on a stable base and rotate the legs downward until they lock into position.
Page 18
6 Chapter 2 System Operation
2.2.3 Rack Mounting
The 3150 may be rack mounted by using a 3150 rack mount kit. All rack mount accessory kits contain detailed mounting instructions.
2.2.4 Ventilation Requirements
Rear panel area needs 2 to 4 inches of clearance for air circulation.
2.2.5 Power-Up Sequence
With the 3150 connected to an AC power source, set the power switch to “I” to supply power to the instrument and start the power-up sequence.
During the power-up sequence, the following takes place. For about 5 seconds an initialization screen is displayed. The software version is displayed in the lower left corner of the screen. During this time a self-test is performed to ensure that the 3150 hardware and software are communicating. If the 3150 cannot successfully complete this test, an error message will be displayed.
NOTE
In the even the 3150 is powered off, it must remain off for a
minimum of 60 seconds
before reapplying power. Not doing so
may result in the internal power supply failing to start, and an error
message will be reported. There is no risk of damage to the unit if
such a condition occurs.
After this test, the 3150 is configured to the state it was in when the power was last shut off and displays the master display.
2.2.6 Quick Start
After the power-on sequence is complete, the 3150 goes to the Master screen. To set up the TEC, press the
SETUP
softkey button. The up and down cursor keys will allow the selection of all the TEC parameters, using the cursor keys and the dial set the parameter values. When finished, return the TEC to the Master screen by pressing the
MASTER
button.
Enter the desired set point value using the cursor keys or the dial. Press the
TEC On
key to operate the TEC. The LED illuminates to indicate TEC operation. To turn the TEC off, press the
TEC On
key again.
Page 19
Chapter 2 System Operation 7
2.3 Introduction to the 3150 Front Panel
2.3.1 Model 3150
Described below are the functions of each area of the Model 3150 front panel, as shown in Figure 1.
Figure 1 - Model 3150 Front Panel
1.
Power On/Off Switch
- Switches on/off the AC power to the unit.
2.
TEC On LED
- Indicates TEC output is on.
3.
TEC On Button
- Turns the TEC output on/off.
4.
Display Soft Keys
- These are the two dark keys located to the right of the display. T he function of these keys varies depending on what menu is displayed. See section 2.4.1.4 for a complete description of soft keys.
5.
MASTER Key
- switches to the master display from any screen in the system
(see section 2.4.4).
6.
Cursor Control Keys
- Moves cursor up or down between editable data fields. The left arrow decrements values in numerical entry fields, or as a previous choice in a multi-choice entry field. The right arrow increments values in numerical entry fields, or as a next choice in multi-choice entry fields. See section 2.4.1.3 for a description of data fields.
7.
MENU Key
- Switches to the main menu from any screen in the system (see
section 2.4.5).
8.
FUNCTION Key
- Used to execute user macros and special functions (see
section 2.4.2).
9.
SHIFT Key
- Toggles between the outer and inner set of soft keys (see section
2.4.1.4).
10.
Remote LED
- Indicates 3150 is in remote mode.
11.
Knob
- Used to continuously vary certain parameters. The knob has an acceleration factor that causes the rate of change to increase as the knob is turned faster. Turning slowly allows for a fine adjustment at the smallest displayed decimal place.
Page 20
8 Chapter 2 System Operation
2.4 General Operation
2.4.1 Display Elements
The Model 3150 uses a character display to depict information about the current state of the system. The display can be broken down into four basic elements: static fields, non-editable data fields, editable data fields, and soft key labels (see Figure 2).
2.4.1.1 Static Fields
Static fields are elements on the display which do not change from moment to moment. These can include screen titles and error messages.
2.4.1.2 Non-Editable Data Fields
Non-editable data fields are used mainly to display read back information, such as temperature, TEC current, etc. These fields can have a prefix or suffix label, such as “T=” or “I=”, and are periodically updated by the system.
2.4.1.3 Editable Data Fields
Editable data fields are used for TEC and system settings such as ITE set point, temperature set point, display contrast, etc. An editable field has three distinct display states: focused, non-focused, and read-only.
The focused state indicates that the field has the input “focus.” When a field has the focus, a right pointing arr ow () is placed to the left of the field. Any keyboard
entry or knob adjustment will be applied to the field, and only one field at a time on the display can have focus. Move between fields using the up and down arrow keys.
The non-focused state indicates that the field is editable, but does not currently have the focus. These fields are indicated with a right pointing triangle () to the left of
the field. Using the up and down arrows, focus can be moved to these fields.
When the editable data field is in the read-only state, it looks and acts exactly like a non-editable data field. Like the non-editable data field, it cannot have focus, and the up or down arrow keys will skip over the field. This state is used primarily to lockout specific data elements from front panel change when the Model 3150 is in remote mode. Any IEEE-488 or RS-232 communication will place the unit in remote mode, and editable fields that are protected during remote operations change to the read only state.
2.4.1.3.1 Changing Data Fields
A data field can only be changed from the front panel when the field is the focus. Some fields are numeric-based, such as current set point or temperature limits. Other fields are multi-choice fields, such as Yes/No fields. Both types are changed with the left and right arrows or the knob.
Page 21
Chapter 2 System Operation 9
2.4.1.4 Soft Keys
Soft key labels are labels for the two gray buttons located to the immediate right of the display. Each label either indicates the action that is performed when the corresponding key is pressed (such as changing screens), or the state of a data element in the system (such as on/off toggle switch). In the first case, pressing the corresponding soft key will cause the action to happen, such as changing to the setup screen when the
Setup
soft key is pressed from a TEC’s main screen. In the second
case, pressing the soft key will change the state of the associated data value.
Like the editable data fields above, certain soft keys are programmed to enter a “display-only” mode when the unit enters remote mode. Display-only soft keys are displayed in lower case, and will not function until the unit returns to local mode.
On some screens, such as the configure menu (see Figure 6), there are more than two soft key selections. In this case, the active soft keys have a left pointing arrow () to
the right of the soft key label. Pressing the
SHIFT
key will toggle between the outer
and inner two soft keys.
Figure 2 - A Sample Screen with Various Data Fields
2.4.2 Function Keys
For macros and special functions, the
FUNC
key is used both to execute and to enter the setup screen on the particular function. For example, if the 3150 supported a special function assigned to the up arrow key, to enter the setup screen of this function, pr ess and hold the
FUNC
key, then press the up arrow key, then release both. This would enter the setup screen for this function. To execute this function, press and release the
FUNC
key, then press and release the up arrow key. If
functions are not setup/supported for a particular key, the 3150 will beep. The 3150
→→→→
Ts = 25.00 °C OutT T = 23.50 °C Ite= 0.00 A V=0.00 T OutT SETUP
Static Field
Non-editable data field
Focused editable data field
Page 22
10 Chapter 2 System Operation
supports assignment of macros to the arrow keys, the
MASTER
key, the
MENU
key, and the
TEC
key.
2.4.3 Menu Structure
Figure 3 - Model 3150 Menu Structure
Master Display
System
Configure Menu
Save/Recall
Communications
Local
Calibration
Setup
Main Menu
Page 23
Chapter 2 System Operation 11
2.4.4 Master Display
The Master Display is shown in Figure 4. This is the highest level display and indicates the status of the temperature controller.
→→→→
Ts = 25.00 °C T = 23.50 °C Ite= 25.00 A V=0.00 SETUP
Figure 4 - Master Display
The Master Display can be accessed from any screen in the system by pressing
MASTER
.
2.4.5 Main Menu
The Main Menu is shown in Figure 5. This is the second highest menu and is used to access four general system functions:
1.
COMM
- Pressing the adjacent soft key gives access to
the GPIB and RS232 parameters.
2.
LOCAL
- When the unit is in remote mode, either
through GPI B or RS-232C communications, the
Local
soft key will be available. Pressing it returns the 3150 to a local state. When in local mode, this key does not appear on the display. The 3150 is placed in remote mode through GPI B or RS232 communication, or during the execution of a macro or special function.
3.
CONFIG
- Pressing the adjacent soft key gives access to the general configuration menu, with soft keys to access system configure, save/recall, and calibration screens.
Page 24
12 Chapter 2 System Operation
Main Menu COMM
←←←←
LOCAL CONFIG
←←←←
Figure 5 - Main Menu
2.4.6 Configure Menu
Config SYSTEM
←←←←
CAL SAV/RCL
←←←←
Figure 6 - Configure Menu
The configure menu provides access to the system configuration, calibration, and save and recall screens.
2.4.7 System Configure Screen
→→→→
Contrast= 11 %

Brightness= 100 %

Lockout dial= No

Lockout pad= No
→→→→
Audible beep= Yes


Key Rate= Fast

Dial Rate= Fast
The system configure screen controls basic operation of the 3150 system.
Brightness
varies the backlighting intensity.
Contrast
is used to optimize the viewing angle.
Page 25
Chapter 2 System Operation 13
Lockout dial
disables the dial to avoid accidental changes in active data fields if the
dial is bumped.
Lockout pad
locks out the left and right arrow keys, the data entry portion on the
keypad. Navigation keys, such as up and down,
MENU, MASTER
, and
FUNC
continue to work.
Note that both the Lockout dial and Lockout pad settings are temporarily suspended while in the Configure System Screen, allowing the dial and keypad lockout settings to be changed while in this screen..
Audible Beep
controls the system’s audible beeper. The beeper indicates errors,
invalid data entry, and other situations where the 3150 needs to alert the user.
Key Rate
- this controls the speed at which, when a key is held down, it repeats.
Settings are
Slow, Medium
, and
Fast
.
Dial Rate
- like the
Key Rate
setting, this controls the acceleration of the dial as it is
turned. Settings are
Slow, Medium
, and
Fast
.
Page 26
14 Chapter 2 System Operation
2.4.8 Save/Recall Screen
Save/Recall SAV
→→→→
Bin=1
RCL
Figure 7 - Save/Recall Screen
The Save and Recall functions are used to store and retrieve 3150 setup configurations for future use. For example, a specific test setup may be saved for later use, and then another setup may be used presently. Then, when the user desires to perform the specific test, its setup is simply recalled.
Non-volatile memory is used for saving the instrument parameters. When a save operation is performed, all of the parameters which are currently in effect on the 3150 are stored. The user selects a “bin” number for saving the parameters, up to the maximum available in the instrument. Then, when that “bin” number is recalled, the 3150 is restarted and the parameters are reconfigured to the previously stored values. A special “bin 0” is reserved for the reset state. Recalling bin 0 will reset the unit to factory defaults.
2.4.9 Calibration Screen
Pressing the
Cal
soft key displays the TEC calibration screen. See section 7.2 for
TEC calibration.
2.4.10 Configure Communications Screen
→→→→
Err While Rmt= No

GPIB Address= 4

Speed= 9600 Baud

Terminal Mode= No
Figure 8 - Communications Screen
The
GPIB Address
is the IEEE-488 device address assigned to the 3150. Valid
addresses are 1 to 31. See the Computer Interfacing Manual for a description of the
Terminal Mode
and
Speed
.
Page 27
Chapter 2 System Operation 15
2.4.10.1 Error Message Control
Error messages may appear on the display when error conditions occur which force the output off or reflect hardware errors in the 3150. Chapter 9 contains an explanation of the error message which may be reported by the 3150. Display of error messages on the 3150 screen may be disabled while in remote mode by setting
Err While Rmt
to No, or by using the GPIB command
REMERR
to set this value remotely. Errors will continue to accumulate in the error queue, but will not be displayed on-screen.
Each press of the
MASTER
button will clear one error.
2.5 Rear Panel Familiarization
Figure 9 - Rear Panel
2.5.1 GPIB Connector
The GPIB connector, located on the back panel, allows full remote control as described in the Computer Interfacing Manual. It accepts a standard IEEE-488 cable for remote control, and uses Metric lock screws.
2.5.2 RS-232 Connector
The 3150 has an RS-232 connector located on the back panel. See the Computer Interfacing Manual for a more complete description of the RS-232 interface.
2.5.3 Input Power Connector
Accepts a standard line cord for AC input. Also selects one of four AC input settings: 100V, 120V, 220V, and 240V. The cord must be removed to change the setting. A small screwdriver will open the top of the module and expose the rotary switch. Select the range that is closest to your expected nominal RMS line voltage. The voltage selection is set for 120 VAC prior to shipping. A caution sticker is then placed over the input power connector to help insure the customer checks for proper voltage.
Page 28
16 Chapter 2 System Operation
CAUTION
Do not exceed 250 VAC on the line input.
Do not operate with a line voltage that is not within ±10% of the
line setting. Too low of an input voltage may cause excessive
ripple on the DC supplies. Too high of an input voltage will cause
excessive heating.
2.5.4 Ground Post
Provides access to chassis ground, which is also an earth ground as long as a standard 3-wire line cord is used. This is a protective conductor terminal to be used to achieve chassis grounding requirements when the main connectors don’t provide an earth ground terminal. Use a minimum of 18 gauge wire to connect to this terminal.
2.6 Warm Up and Environmental Consideration
Operate the 3150 at an ambient temperature in the range of 0 to +40°C. Storage temperatures should be in the range of -20 to +60°C. To achieve rated accuracy, let
the 3150 warm up for 1 hour. For greatest accuracy, recalibrate when ambient temperature changes more than a few degrees.
CAUTION
Operating above +40°C can cause excessive heating and possible
component failures.
Page 29
17
CHAPTER 3
3 Temperature Controller Operation
3.1 General Features
The 3150 Temperature Controller is a precision thermo electric cooler controller. Features include:
Closed-case calibration
Operational with most thermistors, IC and RTD temperature sensors
Flexible setup with Save/Recall front panel functions
High temperature stability
Current Limit
3.2 TEC Safety Features
3.2.1 Conditions Which Will Automatically Shut Off the TEC Output
High Temperature Limit
Low Temperature Limit
R Limit
TEC Open
Sensor Open
Sensor Select changed
Sensor Shorted
Mode Change
Clearing the appropriate bits in the TEC OUTOFF register can disable each of these conditions. See the Computer Interfacing Manual for additional information.
Page 30
18 Chapter 3 Temperature Controller Operation
3.3 The TEC Connectors
A high power 7W 2 female D-connector is used for input and output connections, as shown by the pin out diagram below.
Pin
Description
3150-02 Cable Color Code
A1 TE+ Red A2 TE- Black 1 Sensor+ Green 2 Sensor- White 3 Interlock 4 Interlock 5 Ground
Table 1 - TEC Connector Pinout
3.3.1 TEC Interlock
On the TEC input/output connector pins 3 and 4 form the interlock path. The TEC interlock functions in a normally open condition, where the output will be shutdown only when pins 3 and 4 are shorted together.
3.3.2 TEC Grounding Consideration
The TEC output is isolated from chassis ground, allowing either output terminal to be grounded at the user's option.
3.4 TEC Operation
3.4.1 TEC Main Screen
The TEC main screen is shown Figure 10 and described below.
A1
A2
1
2
5
3
Page 31
Chapter 3 Temperature Controller Operation 19
→→→→
Ts= 25.00 °C T= 25.00 °C Ite= 0.00 A I V T R C InT SETUP
←←←←
Figure 10 - TEC Main Screen
Is=, Ts=, Rs=, is=, vs=
- Indicates the set point value of current, temperature, resistance, AD590 sensor current, or LM335 sensor voltage, respectively. In the screen shown above, the Ts is shown.
Is, Rs, is,
and vs would be seen when
operating in those modes.
I=, T=, R=, i=, v=
- Indicates the measured value of current, temperature, resistance,
AD590 sensor current, or LM335 sensor voltage, respectively. An
err
indicates a sensor error, usually caused by the sensor not hooked up or the wrong se nsor selected. In the screen shown above, the T is shown. I,
R, i
, and
v
would be seen
when operating in those modes.
SETUP
- Pushing the adjacent soft key activates the setup screen.
The bottom line on the display has 6 “LED” elements, each indicating a particular state of the TEC. They are defined as:
I
When illuminated, indicates the TEC module is in current limit.
V
When illuminated, indicates the TEC module has reached it’s voltage limit.
T
When illuminated, indicates the TEC module is outside the temperature limits defined by T
HI
and TLO in the setup screen.
R
When illuminated, indicates the TEC module is outside the reference limits defined by R
HI/vHI/iHI
and RLO/vLO/iLO in the
setup screen.
C/H
When illuminated, indicates that the TEC is cooling or heating.
OutT/InT
When illuminated, indicates that the TEC is out of tolerance or in tolerance as defined by
Tol Time
and
Tol Temp
in the setup
screen.
3.4.2 TEC Setup Screen
The TEC Setup screen is shown in Figure 11. Each section is described below in detail.
Page 32
20 Chapter 3 Temperature Controller Operation
→→→→
Sens= RTD


Mode= Const T

Gain= 30

Lim Ite= 2.0 A
→→→→
Tol Time= 5.000 S


Tol Temp= 0.20 °C

Lim Tl= 10.00 °C

Lim Th= 50.00 °C
Vpow= 10.00 V

Ths= 24.9 °C Tamb= 23.4 °C
→→→→
C1= 3.9080 x 10-3


C2=-0.5802 x 10-6

C3=-4.272 x 10-12

R0= 100.00
ΩΩΩΩ
Figure 11 - TEC Setup Screens
3.4.2.1 The
and Soft Keys
Pushing the

(previous) soft key returns to the previous screen, while pressing the
(next) soft key advances to the next screen.
3.4.2.2 Sensor (Sens)
Selects the temperature sensor type used in your TEC mount. If the
None
is selected,
only the I
TE
mode is allo wed . This type is i ntended for applica t ions running without
a temperature sensor. See the following sections for discussions of the various sensor
Page 33
Chapter 3 Temperature Controller Operation 21
types. The TEC supports the thermistor sensors (10µA and 100µA range), AD590, LM335, and RTD sensors.
3.4.2.3 Mode
3.4.2.3.1 Constant Temperature Mode (Const T)
This mode holds the TEC at a constant temperature based on feedback from the sensor in the TEC mount, using “
Ts=
” and “T=” variables. In this mode, the 3150 uses a control loop comparing the sensor input to the temperature set point, driving the I
TE
current positive or negative to reach and maintain that set point. The sensor’s
input is converted to temperature for display of actual TEC temperature. The I
TE
current is also displayed in this mode.
3.4.2.3.2 Constant Resistance/Reference Mode ( Const R)
This mode operates identically to the Const T mode, but the sensor input is not converted to temperature, and is displayed in unconverted form. Likewise, the set point is used directly, not converted from temperature. Thermistor and RTD sensors use resistance (“
Rs=
” and “R=” variables), LM335 sensors use millivolts (“
vs=
” and
“v=” variables), and AD590 sensors use microamps (“
is=
” and “i=” variables). Const R is primarily intended for users who know a sensor set point in “sensor” units, not in ºC. I
TE
current is also displayed in these modes.
3.4.2.3.3 Constant Current Mode (Const Ite)
Unlike the modes above, the Const ITE mode allows the operator to explicitly set the amount and dir ection of current flow through the TEC, using “
Is=
” and “
Ite=
” variables. If a sensor has been selected, the TEC temperature will be displayed. Although temperature is not a factor in the amount or dir ection of current flow, the high and low temperature limits are observed, and will shutdown the output if exceeded in Const I
TE
mode, if a sensor is selected. For no temperature limits, set the sensor type to “None.” Use caution when limits are not active, as the temperature may exceed your TEC's thermal limits.
3.4.2.3.4 Effects of Calibration on TEC modes
On startup, the TEC performs an auto-calibration to eliminate most of the error in ADC and DAC values. After this auto-calibration, each sensor type supported by the TEC has an offset calibration, while the I
TE
set point and read back has a two point calibration. These calibration constants are then used to calibrate a set point or read back value. This includes “cross-mode” values, such as displaying actual current while in constant temper ature mode. While the current set point calibration has no effect in Const T mode, the read back calibration is used to more accurately display the actual current.
Page 34
22 Chapter 3 Temperature Controller Operation
3.4.2.4 Gain
The Gain function controls two parameters of the hybrid PI control loop; proportional gain and integration time.
When the actual temperature and the set point are different, an error voltage is generated. This error voltage is directly related to the difference in the actual and set point temperatures. The error voltage is then amplified by the proportional gain. This amplified error voltage controls the amount o f c urrent driven through the T EC. The higher the gain, the more current will be driven for any given temperature difference, with the maximum current being determined by the current limit.
The error voltage also drives an integrator. The integrator’s output also controls the amount of curre nt being driven t hrough the TE C. The integr ator is an ampli fi er whose gain is proportional to time. T he longer a given error voltage is present, the more current will be driven through the TEC, with the maximum current being determined by the current limit. The speed at which the integrator’s output increases is the integration time, which can be “Slow” or “Fast”.
The allowed Gain values are: 0.2 Slow, 0.6 Slow, 1 Slow, 1 Fast, 2 Slow, 3 Fast, 5 Fast, 6 Slow, 10 Slow, 10 Fast, 20 Slow, 30 Fast, 50 Fast, 60 Slow, 100 Fast or 300 Fast. The number actually defines the proportional loop gain. The slow/fast suffix indicates the speed at which the integrator’s output increases. The slow setting allows for larger masses or greater distance between the sensor and the thermo­electric cooler by slowing the speed of the integrator.
Both the proportional gain and the integration time must be matched to the thermal characteristics of the TE cooler and sensor. If the settings are incorrect, the temperature set point will take an excessive amount of time to settle, or it will oscillate around the set point and never settle.
The Gain setting depends on the type of T E cooler that you are using, but we can suggest guidelines for selecting the proper ga in. Set the gain to 1 fast and increase it until the actual temperature oscillates around the set temperature. Then reduce the gain to the next lower value, without changing the suffix.
To read the Gain setting, go to the setup. The display will show the value of the Gain setting. In Consta nt I
TE
mode the Gain setting has no effect.
3.4.2.5 Limits
3.4.2.5.1 TE Current Limit (Lim Ite)
This sets the maximum drive current the TEC will allow. This maximum applies to all modes (constant I
TE
/R/T).
Page 35
Chapter 3 Temperature Controller Operation 23
3.4.2.5.2 Temperature Limits (Lim Th and Lim Tl)
The TEC supports both a low and high temperature limit, and can be programmed to turn the TEC output off in the event those limits are exceeded (default state). The temperature limits are monitored regardless of the mode of the TEC. This has the added sa fety feature of shutting down the TEC in Const I
TE
or Const R mode when the temperature limit is exceeded (see TEC:OUTOFF command in Computer Interfacing Manual for additional details).
Caution: these limits do not apply if the sensor type is set to “None.”
3.4.2.5.3 Resistance/Reference Limits (Li m Rh/ vhI/ih and Lim Rl/vlO/il)
Like the temp erature limits, the TEC also supports both a low and high resistance/reference limit, and can be programmed to turn the TEC output off in the event those limits are exceeded, although by default this is disabled. These limits are monitored only while in Const R/v/i mode.
3.4.2.6 Tolerances (Tol Time and Tol Temp)
The
Tol Time
and
Tol Temp
elements are used for determining when the TEC is “in
tolerance,” where the actual temperature has stayed within
Tol Temp
of the set point
for at least
Tol Time
seconds. The
Tol Time
value is expressed in seconds, and can
range from 0.001 seconds to 50 seconds. The
Tol Temp
value is displayed in ºC (the most common usage), and can range from 0.01 to 10.00. If at any time it goes outside the tolerance range, the time restarts at zero.
As an example, if the
Tol Time
is set to 5 seconds, the
Tol Temp
is set to 0.2ºC,
and the temperature set point was 25.0ºC, the TEC would have to stay within
24.8ºC and 25.2ºC to be within tolerance. Out of tolerance is indicated by a
OutT
status field on the bottom of the TEC Main Screen.
3.4.2.7 Vpow, Ths, and Tamb
The three values monitor the operation of the TE drive section.
Vpow
is the supply
voltage, and can range from approximately 10V to 32V, depending on the TE load.
Ths
is the heatsink temperature, and will shutdown the unit if it exceeds 85ºC.
Tamb
is the internal ambient temperature of the unit.
3.4.2.8 C1, C2, C3, and Ro
See the section of each of the sensors for a description of how C1, C2, C3, and Ro are used.
Page 36
24 Chapter 3 Temperature Controller Operation
3.4.3 Thermistor and Thermistor Current Selection
3.4.3.1 Introduction
Choosing the ri ght sensing current depends on the range of tempera ture you want to measure and the resolution you require at the highest measured temperature. To correctly set the thermistor current you must understand how the thermistor and the 3150 interact.
3.4.3.2 Thermistor Range
Thermistors can span a wide temperature range, but their practical range is limited by their non-linear resistance properties. As the sensed temperature increases, the resistance of the thermistor decreases significantly and the thermistor resistance changes less for an equivalent temperature change. Consider the temperature and sensitivity figures below.
Temperature
Sensitivity
-20°C 5600 ohms/°C 25°C 439 ohms/°C 50°C 137 ohms/°C
In the 3150 the practical upper temperature limit is the temperature at which the thermistor becomes insensitive to temperature changes. The maximum ADC input voltage of the 3150 limits the lower end of the temperature range. Thermistor resistance and voltage are related through Ohm's Law (V = I x R). T he 3150 supplies
current to the thermistor, either 10 µA or 100 µA , and as the resistance changes a changing voltage signal is available to the thermistor inputs of the 3150. The 3150 will over-range when the input voltage exceeds about 5 Volts. Figure 12 graphically shows the lower temperature and upper voltage limits for a typical 10 k Ohm thermistor. The practical temperature ranges for a typical 10 K thermistor (a 10 K
thermistor has a resistance of 10 k Ohms at 25°C) are given in the table below.
Sensing Current
Temperature Range 10 µA -51 to 40°C 100 µA -10 to 100°C
Page 37
Chapter 3 Temperature Controller Operation 25
Figure 12 - Thermistor Temperature Range
3.4.3.3 Temperature Resolution
You must also consider measurement resolution since the resolution decreases as the thermistor temperature increases. The 3150 uses an A/D converter that has a
maximum resolution of about 0.763 in 10uA range. The microprocessor converts this digital number to resistance, stores this resistance, then converts it to a temperature using the Steinhart-Hart equation, and stores this temperature. A temperature change of one degree centigrade will b e r epresented by a greater resistance increase (and therefore more ADC counts) at lower temperatures because
of the non-linear resistance of the thermistor. Resolution figures for a typical 10 k thermistor are given below.
Page 38
26 Chapter 3 Temperature Controller Operation
Temperature
Voltage at 10 µA Resolution
-20 °C 56.0 mV/°C 0.018 °C/mV 25 °C 4.4 mV/°C 0.230 °C/mV 50 °C 1.4 mV/°C 0.700 °C/mV
For this thermistor, a temperature change from -20°C to -19°C will be represented by 737 ADC counts (if supplied with 10µA). The same thermistor will only change about 18 ADC counts from 49°C to 50°C.
3.4.3.4 Selecting Thermistor Current
To select the current setting for a typical 10 K thermistor, determine the lowest temperature you will need to sample and select the current according to the range
limits given above. If the temperature you want to sample is below -10°C you should use the 10µA setting.
With the current set to 10µA the best resolution you will see will be a 1.0°C temperature change. If, for example, the lower limit is 0°C you can choose either setting, but there is a tradeoff in terms of resolution. If you need better than 0.1°C measurement resolution you will have to change to 100µA.
If you need high resolution over a narrow range, for a very accurate measurement, you can set the current setting for the maximum resolution. For example, at a high
temperature of 15°C, you require a measurement resolution of at least 0.05°C. This resolution is within the range of either setting, but at the 10µA setting the resolution is only 0.2°C while at the 100 µA setting the resolution is better than .05 °C.
Generally, it is best to use the 100µA setting for all measurements of -10°C or greater with a 10 K thermistor.
3.4.3.5 Selecting Thermistors
The type of thermistor you choose will depend primarily on the operating temperature range. These guidelines for selecting the range and resolution will apply to any thermistor. 10 K thermistors are generally a good choice for most laser diode applications where high stability is required near room temperatures. Similarly, 10 K thermistors are often a good choice for cooling applications where you want to
operate at temperatures from -40°C to room temperature.
If you require a different temperature range or the accuracy you need can't be achieved with either current setting, select another thermistor. Thermistor temperature curves, supplied by the manufacturer, show the resistance versus temperature range for many other thermistors. Contact a Newport application engineer with your specific application.
Page 39
Chapter 3 Temperature Controller Operation 27
3.4.3.6 The Steinhart-Hart Equation
The Steinhart-Hart equation is used to derive temperature from the non-linear resistance of an NTC (Negative Temperature Coefficient) thermistor.
The following section contains an explanation of the Steinhart-Hart equation and the values of these constants for some common thermistors.
Two terminal thermistors have a non-linear relationship between temperature and resistance. The resistance versus temperature characteristics for a family of similar thermistors is shown in Figure 13. It has been found empirically that the resistance versus temperature relationship for most common negative temperature coefficient (NTC) thermistors can be accurately modeled by a polynomial expansion relating the logarithm of resistance to inverse temperature. The Steinhart-Hart equation is one such expression and is given as follows:
1/T = C1 + C2 (Ln R) + C3 (Ln R)
3
(Eq. 1)
Where T is in KELVIN. To convert T to °C, subtract 273.15.
Once the three constants C1, C2, and C3 are accurately determined, only small errors in the calculation of temperature over wide temperature ranges exist. Table 2 shows the results of using the equation to fit the resistance versus temperature characteristic
of a common 10 k thermistor. The equation will produce temperature calculation errors of less than 0.01°C over the range -20 °C to 50 °C.
Page 40
28 Chapter 3 Temperature Controller Operation
Figure 13 - Thermistor Resistance versus Temperature
---------Error T (°C)----------
R
1
T Actual Third Order
Fit. Eq. 1
2
96974 -20.00 -0.00 55298 -10.00 0.00 32651 0.00 -0.00 19904 10.00 -0.00 12494 20.00 -0.00 10000 25.00 0.00
8056 30.00 0.00 5325 40.00 0.00 3601 50.00 -0.00
Table 2 - Comparison of Curve Fitting Equations
The consta nt s C1, C2, and C3 are expressed in the form n.nnnn, simplifying entry into the 3150.
1
Resistance of a BetaTHERM 10K3 thermistor.
2
Constants C1 = 1.1292 * 10-3, C2 = 2.3411 * 10-4, C3 = 0.8775 * 10-7
Page 41
Chapter 3 Temperature Controller Operation 29
3.4.3.7 Table of Constants
We have listed some common thermistors and included the appropriate calibration constants for the temperature range -20 °C to 50 °C in Table 3. The Model 3150, by
default, uses the BetaTHERM 10K3 thermistor values.
Manufacturer
C1 * 10-3 C2 * 10
-4
C2 * 10
-7
BetaTHERM 10K3 1.129241 2.341077 0.877547
BetaTHERM 0.1K1 1.942952 2.989769 3.504383 BetaTHERM 0.3K1 1.627660 2.933316 2.870016 BetaTHERM 1K2 1.373419 2.771785 1.999768 BetaTHERM 1K7 1.446659 2.682454 1.649916 BetaTHERM 2K3 1.498872 2.379047 1.066953 BetaTHERM 2.2K3 1.471388 2.376138 1.051058 BetaTHERM 3K3 1.405027 2.369386 1.012660 BetaTHERM 5K3 1.287450 2.357394 0.950520 BetaTHERM 10K3 1.129241 2.341077 0.877547 BetaTHERM 10K4 1.028444 2.392435 1.562216 BetaTHERM 30K5 0.933175 2.213978 1.263817 BetaTHERM 30K6 1.068981 2.120700 0.901954 BetaTHERM 50K6 0.965715 2.106840 0.858548 BetaTHERM 100K6 0.827111 2.088020 0.805620
Table 3 - Thermistor Constants
3.4.4 AD590 and LM335
3.4.4.1 General
The 3150 uses two constants (C1 and C2) for calibrating the two linear thermal sensing devices, the AD590 and the LM335. C1 is used as the zero offset value, and C2 is used as the slope or gain adjustment. Therefore, C1 has a nominal value of 0, and C2 has a nominal value of 1 when using the AD590 or LM335. In order to calibrate a linear sensor device, the sensor must be operated at an accurately known,
stable temperature. For example, the sensor may be calibrated at 0 °C if the sensor is placed in ice water until its temperature is stable. A highly accurate temperature probe, thermometer, environmental chamber, etc., may also be used to determine the known temperature for calibration.
3.4.4.2 AD590 Sensor
The AD590 is a linear thermal sensor that acts as a current source. It produces a current, i, which is directly proportional to absolute temperature, over its useful range (-50 °C to + 150 °C). This nominal value can be expressed as:
i = 1 µA / K
where i is the nominal current produced by the AD590, and K is in Kelvin.
The 3150 uses i to determine the nominal temperature, T
n
, by the formula:
Page 42
30 Chapter 3 Temperature Controller Operation
T
n
= (i/(1 µA / K) ) - 273.15
where T
n
is in °C.
The displayed temperature, T
d
= C1 + (C2 * Tn), is then computed, where C1 and C2
are the constants stored in the 3150 for the AD590. The AD590 grades of tolerance vary, but typically without adjusting C1 and C2, the temperature accuracy is ±1°C
over its rated operating range. However, the AD590 is not perfectly linear, and even with C1 accurately known there is a non-linear absolute temperature error associated with the device. This non-linearity is shown in Figure 14, reprinted from Analog Devices specifications, where the error associated with C1 is assumed to be zero.
Figure 14 - AD590 Nonlinearity
If a maximum absolute erro r of 0.8°C is tolerable, the one point calibration of C1 should be used. If a greater accuracy is desired, the two point method of determining C1 and C2 should be used. Note however, the absolute error curve is non-linear, therefore the constant C2 will vary for different measurement points.
3.4.4.3 LM335 Sensor
The LM335 is a linear thermal sensor that acts as a voltage source. It produces a voltage, v, which is directly proportional to absolute temperature, over its useful
range (-40°C to + 100°C). This nominal value can be expressed as:
v = 10mV / K
where v is the voltage produced by the LM335 and K is Kelvin.
The 3150 uses v to determine the nominal temperature, T
n
, by the formula:
Page 43
Chapter 3 Temperature Controller Operation 31
T
n
= (v / ( 10mV / K) ) - 273.15
where T
n
is in °C.
The temperature, T
d
, which is displayed by the 3150 is calculated as follows:
T
d
= C1 + (C2 * Tn)
where C1 and C2 are the constants stored in the 3150 for the LM335.
When the LM335 is calibrated to 25°C, C1 = 0 and C2 = 1, and the temperature accuracy is typically ±0.5°C over the rated operating range. However, the LM335 is
not perfectly linear, and even with C1 accurately known there is a non-linear absolute temperature error associated with the device. This non-linearity caused error
is typically ±0.3°C, with the error associated with C1 assumed to be zero.
If a maximu m absolute error of ±0.3°C can be tolerated, the one point calibration of C1 should be used. If a greater accuracy is desired, the two point method of determining C1 and C2 should be used. Note however, the absolute error associated with the constant C2 may vary ove r different temperature ranges.
3.4.4.4 Determining C1 and C2 for the AD590 and LM335
The nominal values of C1 and C2 are 0 and 1, respectively, for both types of devices. These values should be used initially for determining C1 and C2 in the methods described below.
The One Point method is easiest, but it ignores the non-linearity of the device. It is most useful when a high degree of temperature accuracy is not required.
The Two Point method can achieve a high degree of accuracy over a narrower operating temperature range, but requires two accurate temperature measurements.
3.4.4.4.1 One Point Calibration Method
The calibration described in this section is independent of the calibration pro ced ure described in sections 7.2.4 and 7.2.6. Those sections deal with the internal calibration of the TEC, while the following calibration procedure is for calibrating the external AD590 or LM335 sensor. For the most accurate possible results, both calibration procedures should be performed.
The accuracy of this procedure depends on the accuracy of the externally measured temperature. It is used to determine the zero offset of the device, and it assumes that the gain (slope) is known.
1. Allow the 3150 to warm up for at least one hour. Select the desired sensor type in the setup menu.
Page 44
32 Chapter 3 Temperature Controller Operation
2. Set the C1 parameter to zero. Set the C2 parameter to 1.
3. Place the sensor at an accurately known and stable temperature, T
a
. Connect the sensor to the 3150 for normal Constant temperature operation. Allow the 3150 to stabilize at the known temperature, T
a
and
read the displayed temperature, T
d
.
4. Determine the new value of C1 from the formula:
C1 = T
a
- Td
and enter the new C1 value.
3.4.4.4.2 Two Point Calibrati on Method
The calibration described in this section is independent of the calibration procedure described in sections 7.2.4 and 7.2.6. Those sections deal with the internal calibration of the TEC, while the following calibration procedure is for calibrating the external AD590 or LM335 sensor. For the most accurate possible results, both calibration procedures should be performed.
The accuracy of this procedure depends on the accuracy of the externally measured temperature. It is used to determine the zero offset of the device and the gain (slope).
1. Allow the 3150 to warm up for at least one hour. Select the desired sensor type in the setup menu.
2. Set the C1 parameter to zero. Set the C2 parameter to 1.
3. Place the sensor at an accurately known and stable temperature, T
a1
. Connect the sensor to the 3150 for normal Constant temperature operation. Allow the 3150 to stabilize at the known temperature, T
a1
and read the displayed temperature, T
d1
. Record these values.
4. Repeat Step 3 for another known temperature, T
a2
, and the
corresponding displayed temperature, T
d2
. The two known temperatures should at the bounds of the intended operating range. For best results, make the range between T
a1
and Ta2 as narrow as possible.
5. Determine the new value of C1 and C2 from the following calculations.
C2 = (T
a1
- Ta2) / (Td1 - Td2), and
C1 = T
a1
- (Td1* C2)
6. Enter the new C1 and C2 values.
Page 45
Chapter 3 Temperature Controller Operation 33
3.4.5 RTD Sensors
The following equation is used in temperature to resistance conversions:
R
T
= R0 [1 + C1 x T - C2 x T2 - C3 x (T-100) x T3) for T < 0°C
R
T
= R0 [1 + C1 x T - C2 x T2) for T >= 0°C
where: R
T
is the resistance in at temperature T.
T is the temperature in °C.
3.4.5.1 RTD Constants
The constants entered for an RTD depend on the type of curve it has. Table 4 shows three standard types.
Curve TCR
(
ΩΩΩΩ/ΩΩΩΩ/°°°°
C)
C1 C2 C3 R0
Laboratory .003926 3.9848x10-3 -0.58700x10-6 4.0000x10
-12
100.00
US .003910 3.9692x10
-3
-0.58495x10-6 -4.2325x10
-12
100.00
European .003850 3.9080x10
-3
-0.58019x10-6 -4.2735x10
-12
100.00
Table 4 - RTD Constants
The Ro constant also applies for RTD sensors. It is nominally 100.00 , but can be varied from 95.00 to 105.00 Ω.
Page 46
Page 47
35
CHAPTER 6
4
Principles of Operation
4.1 Introduction
A functional block diagram of the 3150 is shown in Figure 15. In each of the following sections there are functional block diagrams for the various circuit boards of the 3150.
GPIB/RS232
Front Panel
Microprocessor
Parallel Bus
Serial Bus
TEC
Optical
Interface
TEC
Control
TEC
Output
TEC
Main
Power
Supply
Analog
Supply
Switching
Supply
Power Supplies
TEC
TEC
Figure 15 - 3150 Block Diagram
Page 48
36 Chapter 4 Principles of Operation
4.2 TEC Controller Theory of Operation
Figure 16 shows the functionality of the TEC. The following sections detail the theory of operation for each of the blocks in Figure 16.
O ptically
Isolated
Serial Bus
Bipolar Output
Stage
Limit DAC
Proportional Amp
Inte g r a l A mp
Differential
Sensor
Select and
TEC
Sensor Lines
A/D
Converter
Limit Se t P o in t
PI Loop
Set
Set Point
DAC
Actual
Heat/Cool Lines
Current
Amps
Amp
To M icroprocessor
Voltage
Figure 16 - TEC Board Diagram
4.2.1 TEC Interface
The TEC interface provides optically isolated serial communications between the TEC board and the microprocessor. Control signals are passed to the TEC board to set the TEC board status, current limit, and temperature set points. Instructions and data are sent over the serial interface to the optical barrier. Status and data are serially passed back to the microprocessor.
4.2.2 Limit DAC
The microprocessor loads the digitally stored current limit value into the current limit 12-bit DAC. The Limit DAC converts the digital limit signal from the microprocessor to a voltage which becomes the limit voltage for the Bipolar Output Stage. The current limit value is updated at power-up, at a "bin" recall, and whenever the LIM I
TE
value is changed.
Page 49
Chapter 4 Principles of Operation 37
4.2.3 Set Point DAC
The microprocessor loads the digitally stored temperature or current set point value into the set point 16-bit DAC. The Set Point DAC converts a digital set point signal from the microprocessor to a voltage which becomes the set point input to the PI control loop. The TEC set point value is updated at power-up, at a "bin" recall, and whenever a TEC set point value is changed.
4.2.4 A/D Converter
The 16-bit A/D converter measures the sensor voltage, TE current, and TE voltage. The sensor measurement is used by the microprocessor in the calculation of temperature or thermistor resistance. The current measurement is used for the I
TE
value and the voltage measurement is needed for V
TE
.
4.2.5 Sensor Select
Sensor selection is accomplished in the Sensor Select block of the TEC board. Precision 100µA and 10µA current sources may be selected for thermistor control.
RTD, LM335 and AD590 IC temperature sensors may also be selected. The AD590 has a +5 VDC bias voltage, the LM335 has a 1 mA bias current, and the RTD has a precision 1 mA current source.
The output of the Sensor Select block of the TEC board is a voltage which is proportional to the actual temperature. This voltage is fed to the A/D converter which provides a digital measurement to the microprocessor, and to the PI control loop to close the feedback loop when temperature is being controlled.
4.2.6 Difference Amplifier
A differential amp provides a difference signal to the PI control. This signal is the difference between set temperature and actual temperature voltage.
4.2.7 Proportional Amplifier
The Proportional amplifier is part of a digitally controlled gain stage consisting of the analog switches and their associated resistors. The analog switches vary the ratio of resistance in the feedback circuit to change the gain. The gain setting determines how fast the TEC reaches the set point temperature and how quickly it settles to this temperature.
4.2.8 Integrator
The signal from the difference amplifier is sent to an integrator which reduces the difference between the set point temperature and the actual temperature to zero, regardless of the gain setting. An analog switch discharges the integrating capacitor whenever integration is not required to prevent unnecessary difference signal integration.
Page 50
38 Chapter 4 Principles of Operation
4.2.9 Bipolar Output Stage
The Bipolar Output Stage consists of circuits which limit the TEC output, sense the TEC output polarity, sense voltage and current limit conditions, as well as supply the bipolar TEC output. The following sections discuss these functions of the Bipolar Output Stage.
4.2.9.1 Current Limiting
The output of the proportional amplifier and integrator together form the control signal. Output current limiting is effected by bounding the control signal so that it is always less than the limit current. The limit current is set with the front panel controls or through the GPIB. The bipolar current limit levels are established by the output of the current Limit DAC.
4.2.9.2 Current Limit Condition Sensing
Comparators sense the output to determine when output current limiting is occurring. When this condition occurs, the I Limit signal is sent to the microprocessor.
4.2.9.3 Voltage Controlled Current Source
The bounded output control signal is applied to an amplifier. This amplifier and the current sensing amplifier form the o ut put voltage controlled current source. The output of this stage directly drives the externally connected TE cooler.
4.2.9.4 Voltage Limit Condition Sensing
Comparators sense the output to determine when the TEC output compliance voltage limiting is occurring. This condition occurs whenever the TEC output is open or connected to a high resistance. If this condition occurs, the V Limit error signal is passed to the microprocessor.
4.2.10 TEC Control Modes
The 3150 provides three control modes for operation, constant T (temperature), constant R (resistance, voltage, or current), and constant I
TE
(current) modes. Each
of these modes is discussed in the following sections.
4.2.10.1 T Mode
In constant T mode the TEC is driven to the set point temperature. This temperature is monitored by the sensor in the TEC. In the case of a thermistor sensor, the thermistor's resistance is used to determine TEC's temperature by using the Steinhart-Hart conversion equation. The resistance is determined by measuring the
voltage across the thermistor (with a known current of 10µA or 100µA). The I
TE
current is also measured and saved. The TEC's output current is sensed across a resistor and the voltage is converted to an I
TE
current value. VTE is also measured.
Page 51
Chapter 4 Principles of Operation 39
When an LM335 sensor is used, a two-point conversion equation is used to determine the temperature. Its voltage is measured as well as the I
TE
current and VTE
voltage.
When an AD590 sensor is used, another two-point conversion equation is used to determine the temperature. Its reference current is sensed across a resister, and this voltage is measured. I
TE
and VTE are also measured.
4.2.10.2 R Mode
In constant R mode, the TEC is driven to the set point resistance, voltage, or current. This resistance, voltage, or current is measured. I
TE
and VTE are also measured.
4.2.10.3 ITE Mode
In constant ITE mode, the TEC is driven with a constant current, at the ITE set point value. The I
TE
current is sensed across a resistor and the voltage is converted to ITE
current.
Page 52
40 Chapter 4 Principles of Operation
4.3 Microprocessor Board
The Microprocessor Board contains the microprocessor, memory, the serial interface to the TEC, front panel interface, and circuitry that saves the state of the 3150 at power down. The block diagram of the Microprocessor Board is shown in Figure
17.
Power Failure
and Watch Dog
Failure Detection
80188 EB
Microprocessor
RAM ROM
EEPROM
Data Bus
Address Bus
TEC
Interface
Front Panel
Interface
RS 232
Interface
To Front Panel RS 232 Ports To TEC
Power Fail/Reset
Watchdog Reset
Figure 17 - Microprocessor Board Block Diagram
4.3.1 Microprocessor
The 3150 uses a CMOS 80188EB microprocessor to control its internal operations. The 3150 provides a fail-safe timer which generates a reset in the event of a malfunction. A 1 Hz watch-dog pulse is normally present. If for any reason this clock pulse fails to appear it will reset the 3150.
Page 53
Chapter 4 Principles of Operation 41
4.3.2 Memory
The 3150 uses three types of memory. RAM memory is retained only while power is applied to the unit. ROM memory contains the firmware. The third type of memory is electrically erasable programmable memo ry: EEPROM.
EEPROM stores calibration constants and other data which must be retained even when power is removed from the unit, and does not require battery backup. Examples of data stored in this memory include the TEC parameters and calibration constants.
4.3.3 Serial Interface
The 80188 communicates with the TEC controller via a serial bus. Parallel data from the microprocessor is converted to bi-directional serial data. Also provided is the RS-232 communication.
4.3.4 Front Panel Interface
Provides parallel communication with the front panel.
4.3.5 GPIB Interface
Provides parallel communication with the GPIB port.
4.4 Power Supplies
AC power is suppl i ed through the rear panel i nput power connector which provides in-line transient protection and RF filtering. The input power connector contains the fuses and the switch to select series or parallel connection of the transformer primaries for operation at 100 VAC, 120 VAC, 220 VAC, or 240 VAC.
4.4.1 TEC Power Supplies
There are two separate power supplies for the TEC. The linear supply provides analog and digital circuit power. The OEM switching supply provides the power to drive the TEC.
Rectifiers
and Filters
Regulators
Switching
Supply
Transformer
TEC Output
Drive
Main
Supply
Power Entry
Module
Rectifiers
and Filters
Regulators
TEC
Supply
Figure 18 - Power Supply B lock Diagram
Page 54
42 Chapter 4 Principles of Operation
4.4.2 Main Supply
This supply provides digital circuit power for all functions except the TEC. It also provides fan power.
Page 55
43
CHAPTER 5
5
Tips and Techniques
5.1 Introduction
This chapter is intended to further explain specific operational details of the Model 3150, as well as provide application examples.
5.2 TEC Limits
The TEC maintains several limits that control the operation of the unit. There are a total of five limits: I
TE
current limit, high temperature limit, low temperature limit, high resistance/LM335 voltage/AD590 current limit (R/v/i), and low resistance/LM335 voltage/AD590 current limit.
Through the rest of this section, the resistanc e/LM335 voltage/AD590 current mode is collectively referred to as R mode, while the resistance/LM335 voltage/AD590 current limits are collectively referred to as the R limits.
I
TE
Limit
The I
TE
limit controls the maximum amount of current the TEC will drive while in any mode. The limit applies to both the positive and negative current drive. In temperature and R modes, it limits the amount of current that can be driven when controlling to the set point. In I
TE
mode, it limits the set point to less than or equal to
the limit. Unlike the temperature and R limits, the I
TE
limit is controlled by hardware
for immediate response.
Temperature Limits
The high and low temperature limits define the operating temperature range of the TEC, and are monitored while in any mode. The only case where temperature is not monitored is when a sensor type of “None” is selected. Temperature checking for high and low limits is done once per second, and when the system detects that a high or low temperature limit has been exceeded, it will shutdown the TEC output. Note, however, that cle aring the appropriate bits in the TEC O UTOFF register through the GPIB or RS-232 interface can disable this automatic shutdown. The factory default is to shutdown the TEC on a temperature limit.
Resistance, LM335 Voltage, and AD590 Current Limits (R Limits)
In addition to temperature and current limits, the TEC also supports limits based on sensor values. R l imits are monitored only when in the R mode, although as mentioned above, the T limits are monitored in any mode. R checking for high and low limits is done once per second. However, unlike a temperature limit, and when the system detects that a high or low R limit has been exceeded, by default it will not shutdown the TEC output. Note, however, that setting the appropriate bit in the TEC
Page 56
44 Chapter 5 Tips and Techniques
OUTOFF regist er through the GP IB or RS-232 interface can enable this automatic shutdown. The factory default is to not shutdown the TEC on a R limit.
How a Sensor Change Affects R Values
Each time the sensor is changed, the old R limits and set point no longer apply to the new sensor. The 3150 calculates the new values for the upper and lower R limits and set point based on the temperature limits and set point. After the R values are initialized, changing the temperature limits or set point will not affect the corresponding R values until the next time the sensor type is changed.
Page 57
45
CHAPTER 6
6
Maintenance
6.1 Introduction
There are no user serviceable parts inside. Do not attempt to remove the cover.
6.2 Fuse Replacement
The fuses are accessible on the back panel of the 3150. Before replacing a fuse, turn power off and disconnect the line cord. Use only the fuses indicated below.
Line Volta ge
Fuse Replacement 90-110 VAC 6 Amp, 3 AG, Slo-Blo, 250V 108-132 VAC 6 Amp, 3 AG, Slo-Blo, 250V 198-242 VAC 6 Amp, 3 AG, Slo-Blo, 250V 216-250 VAC 6 Amp, 3 AG, Slo-Blo, 250V
6.3 Cleaning
Use mild soap solution on a damp but not wet cloth. Disconnect AC power before cleaning.
Page 58
Page 59
47
CHAPTER 7
7 Calibration
7.1 Calibration Overview
The 3150 performs an automatic calibration on power-up. This removes the majority of calibration error. However, if it is desired to completely calibrate the system, the following procedures will do so.
All calibrations are done with the case closed. The instrument is calibrated by changing the internally stored digital calibration constants.
All calibrations may be performed locally or remotely.
7.1.1 Environmental Conditions
Calibrate this instrument under laboratory conditions. We recommend calibration at 25°C ± 1.0°C. When necessary, however, the 3150 may be calibrated at its intended use temperature if this is within the specified operating temperature range of 0°C to 40°C.
7.1.2 Warm-Up
The 3150 should be allowed to warm up for at least 1 hour before calibration.
7.2 TEC Calibration
7.2.1 Recommended Equipment
Recommended test equipment for calibrating the TEC is listed in Table 5. Equipment other than that shown in the table may be used if the specifications meet or exceed those listed.
Page 60
48 Chapter 7 Calibration
Description
Mfg./Model Specification
DMM HP34401A
DC Amps @ 1.0 A): ±1% Resistance (@ 10 ohms): 0.02%
Resistors Metal Film
20 k for 100µA calibration 200 k for 10µA calibration 3 k for LM335 sensor calibration 16 k for AD590 sensor calibration 100 for RTD sensor calibration
Resistor High Power
1 , 50 W, for current calibration
Connector D-sub 15-pin male
Table 5 - Recommended Test Equipment
7.2.2 Local Operation Thermistor Calibration
a. Measure and record the exact resistance of your metal film resistor. Use
nominal values of 20 k for the 100µA setting, and 200 k for the 10µA setting. With the TEC output off, connect the metal film resistor to the sensor input of the TEC.
b. Press the
Setup
soft key and select the appropriate thermistor (10µA or
100µA) as the sensor type.
c. Go to the calibration display by first pressing the
MENU
button, then
the
Config
soft key, then the
Cal
soft key. At the calibration screen,
press the
Sensor
soft key. Follow the on-screen instructions to complete the calibration. The calibration can be terminated without affecting the stored constants if the
Term
soft key is pressed at any
point prior to completing the calibration.
7.2.3 Remote Operation Thermistor Calibration
a. Measure and record the exact resistance of your metal film resistor. Use
nominal values of 20 k for the 100µA setting, and 200 k for the 10µA setting. With the TEC output off, connect the metal film resistor to the sensor input of the TEC Output connector.
b. Send
TEC:SENS 1
100µA thermistor, or
TEC:SENS 2
for the 1 0µA
thermistor, followed by the
TEC:CAL:SEN
to enter sensor calibration
mode.
Page 61
Chapter 7 Calibration 49
The 3150 will be ready to receive the resistance when, after a
TEC:CAL:SEN?
query is sent, a “1” is returned.
c. Input the actual resistance of the metal film resistor, in k, (as an <nrf
value>) via the
TEC:R <nrf value>
command.
If, at any time prior to
TEC:R,
a command other than
TEC:R
or
TEC:R?
is sent to the 3150, the 3150 will cancel the calibration mode
and then process the command(s).
Once the
TEC:R
value is sent, the
OPC?
que ry may be used to determine when the calibration is completed. The operation complete flag (bit 0 of the Standard Event Status Register) may be used to trigger an interrupt. This type of interrupt is enabled by setting bit 0 of the Service Request Enable register and using the
*OPC
command.
7.2.4 Local Operation AD590 Sensor Calibration
a. With the TEC output off, connect a precision 16 kΩ metal film resistor
and a precision ammeter in series at the sensor input of the TEC Output connector.
b. Press the
Setup
soft key and select the AD590 as the Sensor Type.
c. Go to the calibration display by first pressing the
MENU
button, then
the
Config
soft key, then the
Cal
soft key. At the calibration screen,
press the
Sensor
soft key. Follow the on-screen instructions to complete the calibration. The calibration can be terminated without affecting the stored constants if the
Term
soft key is pressed at any
point prior to completing the calibration.
7.2.5 Remote Operation AD590 Sensor Calibration
a. With the TEC output off, connect a precision 16 kΩ metal film resistor
and a precision ammeter in series at the sensor input of the TEC Output connector.
b. Enter the
TEC:SEN 4
and
TEC:CAL:SEN
to select the AD590 sensor
and enter sensor calibration mode.
The 3150 will be ready to receive the current value when, after a
TEC:CAL:SEN?
query is sent, the response from the 3150 is “1”.
c. Input the actual current measured, in µA, by the external ammeter (as
an <nrf value>) via the
TEC:R <nrf value>
command.
Page 62
50 Chapter 7 Calibration
If, at any time prior to
TEC:R,
a command other than
TEC:R
or
TEC:R?
is sent to the 3150, the 3150 will cancel the calibration mode
and then process the command(s).
Once the
TEC:R
value is sent, the
OPC?
que ry may be used to determine when the calibration is completed. The operation complete flag (bit 0 of the Standard Event Status Register) may be used to trigger an interrupt. This type of interrupt is enabled by setting bit 0 of the Service Request Enable register and using the
*OPC
command.
7.2.6 Local Operation LM335 Sensor Calibration
a. Use a 3 k metal film resistor. With the TEC output off, connect the
metal film resistor in parallel with a precision voltmeter to the sensor input of the TEC Output connector.
b. Press the
Setup
soft key and select the LM335 as the Sensor Type.
c. Go to the calibration display by first pressing the
MENU
button, then
the
Config
soft key, then the
Cal
soft key. At the calibration screen,
press the
Sensor
soft key. Follow the on-screen instructions to complete the calibration. The calibration can be terminated without affecting the stored constants if the
Term
soft key is pressed at any
point prior to completing the calibration.
7.2.7 Remote Operation LM335 Sensor Calibration
a. With the TEC output off, connect a 3 k metal film resistor and a
precision voltmeter in parallel at the sensor input of the TEC Output connector.
b. Enter the
TEC:SEN 3
and
TEC:CAL:SEN
to select the LM335 sensor
and enter sensor calibration mode.
The 3150 will be ready to receive the voltage value when, after a
TEC:CAL:SEN?
query is sent, the response from the 3150 is "1".
c. Input the actual voltage, in mV, measured by the external voltmeter (as
an <nrf value>) via the
TEC:R <nrf value>
command.
If, at any time prior to
TEC:R,
a command other than
TEC:R
or
TEC:R?
is sent to the 3150, the 3150 will cancel the calibration mode
and then process the command(s).
Once the
TEC:R
value is sent, the
OPC?
que ry may be used to determine when the calibration is completed. The operation complete flag (bit 0 of the Standard Event Status Register) may be used to trigger
Page 63
Chapter 7 Calibration 51
an interrupt. This type of interrupt is enabled by setting bit 0 of the Service Request Enable register and using the
*OPC
command.
7.2.8 Local Operation RTD Calibration
a. Measure and record the exact resistance of your 100 metal film
resistor. With the TEC output off, connect the metal film resistor to the sensor input of the TEC Output connector.
b. Press the
Setup
soft key and select the RTD as the Sensor Type.
c. Go to the calibration display by first pressing the
MENU
button, then
the
Config
soft key, then the
Cal
soft key. At the calibration screen,
press the
Sensor
soft key. Follow the on-screen instructions to complete the calibration. The calibration can be terminated without affecting the stored constants if the
Term
soft key is pressed at any
point prior to completing the calibration.
7.2.9 Remote Operation RTD Calibration
a. Measure and record the exact resistance of your 100 metal film
resistor. With the TEC output off, connect the metal film resistor to the sensor input of the TEC Output connector.
b. Send
TEC:SENS 5
to select the RTD sensor, followed by the
TEC:CAL:SEN
to enter sensor calibration mode.
The 3150 will be ready to receive the resistance when, after a
TEC:CAL:SEN?
query is sent, a “1” is returned.
c. Input the actual resistance, in ohms, of the metal film resistor (as an
<nrf value>) via the
TEC:R <nrf value>
command.
If, at any time prior to
TEC:R,
a command other than
TEC:R
or
TEC:R?
is sent to the 3150, the 3150 will cancel the calibration mode
and then process the command(s).
Once the
TEC:R
value is sent, the
OPC?
que ry may be used to determine when the calibration is completed. The operation complete flag (bit 0 of the Standard Event Status Register) may be used to trigger an interrupt. This type of interrupt is enabled by setting bit 0 of the Service Request Enable register and using the
*OPC
command.
7.2.10 RTD Lead Resistance Calibration (Offset Null)
Because the RTD sensor reflects changes in temperature with small changes in resistance, even a small lead resista nc e (resistance c aused by the wire running
Page 64
52 Chapter 7 Calibration
between the TEC and the RTD sensor) can cause significant temperature offset. The lead resistance may be taken out of the RTD reading as follows:
a. With the TEC output off, short the sensor wires as close to the RTD
sensor as possible.
b. Press the
Setup
soft key and select the RTD as the Sensor Type.
c. Go to the calibration display by first pressing the
MENU
button, then
the
Config
soft key, then the
Cal
soft key. At the calibration screen,
press the
RTD Null
soft key. Follow the on-screen instructions to complete the calibration. The calibration can be terminated without affecting the stored constants if the
Term
soft key is pressed at any
point prior to completing the calibration.
7.2.11 Local Operation ITE Current Calibration
The following procedure is for calibrating the ITE constant current source for both polarities of current.
a. With the output off, connect a 1, 50W resistor and a calibrated
ammeter in series across the output terminals. If an ammeter with the appropriate current ratings is unavailable, connect a 1, 50W resistor
across the output terminals and use a calibrated DMM to measure the voltage across the resistor. Calculate the current in the following steps by using Ohm's Law:
I = V / R
where V is the measured voltage across the resistor, and R is the
measured load resistance.
b. Go to the calibration display by first pressing the
MENU
button, then
the
Config
soft key, then the
Cal
soft key. At the calibration screen,
press the
I
TE
soft key. Follow the on-screen instructions to complete the calibration. The calibration can be terminated without affecting the stored constants if the
Term
soft key is pressed at any point prior to
completing the calibration.
7.2.12 Remote Operation ITE Current Calibration
a. With the output off, connect a 1, 50W resistor and a calibrated
ammeter in series across the output terminals. If an ammeter with the appropriate current ratings is unavailable, connect a 1, 50W resistor
across the output terminals and use a calibrated DMM to measure the voltage across the resistor. Calculate the current in the following steps by using Ohm's Law:
Page 65
Chapter 7 Calibration 53
I = V / R
where V is the measured voltage across the resistor, and R is the
measured load resistance.
b. Send
TEC:CAL:ITE
to enter I
TE
calibration mode.
The TEC will be placed in I
TE
mode, limit set to 50% of full scale plus
100 mA, and the I
TE
set point set to 50% of full scale.
The 3150 will be ready to receive the first measured current value
when, after a
TEC:CAL:ITE?
query is sent, a “1” is returned.
c. Input the actual current (as an <nrf value>) via the
TEC:ITE <nrf
value>
command. The 3150 will then drive the current to 25% of the
initial set point.
The 3150 will be ready to receive the second measured current value
when, after a
TEC:CAL:ITE?
query is sent, a “1” is returned.
d. Input the second actual current (as an <nrf value>) via the
TEC:ITE
<nrf value>
command. The 3150 will then drive the current to the
negative current value of the initial set point.
The 3150 will be ready to receive the third measured current value
when, after a
TEC:CAL:ITE?
query is sent, a “1” is returned.
e. Input the third actual current (as an <nrf value>) via the
TEC:ITE
<nrf value>
command. The 3150 will then drive the current to 25% of
the negative current value of the initial set point.
The 3150 will be ready to receive the fourth measured current value
when, after a
TEC:CAL:ITE?
query is sent, a “1” is returned.
f. Input the fourth actual current (as an <nrf value>) via the
TEC:ITE
<nrf value>
command.
If, at any time prior to the last
TEC:ITE,
a command other than
TEC:ITE
or
TEC:ITE?
is sent to the 3150, the 3150 will cancel the
calibration mode and then process the command(s).
Once the
TEC:ITE
value is sent, the
OPC?
que ry may be used to determine when the calibration is completed. The operation complete flag (bit 0 of the Standard Event Status Register) may be used to trigger an interrupt. This type of interrupt is enabled by setting bit 0 of the Service Request Enable register and using the
*OPC
command.
Page 66
Page 67
55
CHAPTER 8
8
Factory Service
8.1 Introduction
This section contains info rmation regar ding obtaining factory service for the Model
3150. The user should not attempt any maintenance or service of this instrument and/or accessories beyond the procedures given in chapters 6 and 7. Any problems which cannot be resolved using the guidelines listed in chapters 6 and 7 should be referred to Newport Corporation factory service personnel. Contact Newport Corporation or your Newport representative for assistance.
8.2 Obtaining Service
To obtain information concerning factory service, contact Newport Corporation or your Newport representative. Please have the following information available:
1. Instrument model number (On front panel)
2. Instrument seri al number (On rea r panel)
3. Description of the problem.
If the instrument is to be returned to Newport Corporation, you will be given a Return Materials Authorization (RMA) number, which you should reference in your shipping documents as well as clearly marked on the outside of the shipping container.
Please fill out the service form, located on the following page, and have the information ready when contacting Newport Corporation. Return the completed service form with the instrument.
Page 68
Page 69
Service Form
Newport Corporation
USA Office: 949/863-3144 FAX: 949/253-1800
Name
RETURN AUTHORIZATION #
Company
(Please obtain prior to return of item) Address Country
Date P.O. Number
Phone Number
Item(s) being returned:
Model #
Serial # Description Reason for return of goods (please list any specific problems)
List all control settings and describe problem
(Attach additional sheets as necessary)
Show a block diagram on the backside of this page of your measurement system including all instruments connected (whether power is turned on or not). Describe signal source. If source is laser, describe output mode, peak power, pulse width, and repetition rate.
Where is measurement being performed?
(factory, controlled laboratory, out-of-doors, etc.) What power line voltage is used?
Variation?
Frequency?
Ambient Temperature? Any additional information. (If special modifications have been made by the user, please describe below)
Page 70
Page 71
59
CHAPTER 9
9
Error Messages
9.1 Introduction
Error messages may appear on the display when error conditions occur in the respective functions of the 3150. For example, a current limit error in the TEC will be displayed.
In remote operation, the current error list can be read by issuing the "ERR?" query. When this is done, a string will be returned containing all of the error messages which are currently in the error message queue.
The errors codes are numerically divided into areas of operation as shown below.
Error Code Range
Area of Operation
E-001 to E-099 Internal Program Errors E-100 to E-199 Parser Errors E-200 to E-299 Execution Control Errors E-300 to E-399 GPIB/RS232 Errors E-400 to E-499 TEC Control Errors
Table 6 contains all of the error messages which may be generated by the 3150. Not all of these messages may be displayed. Some refer to GPIB activities only, for example.
Table 6 - Error Codes
Error Code
Explanation
E-001 Memory allocation failure. E-002 Floating point error
E-104 Numeric type not defined. E-106 Digit expected. E-107 Digit not expected. E-115 Identifier not valid. E-116 Parser syntax error, character was not expected. E-126 Too few or too many program data elements.
E-201 Value out of range. E-214 Length exceeds maximum. E-217 Attempted to recall a bin from a unsaved position. E-218 Additional link could not be added to the link table because the
link table is already full.
E-301 A response message was ready, but controller failed to read it. E-302 3150 is talker, but controller didn't read entire message.
Page 72
60 Chapter 9 Error Messages
Error Code
Explanation
E-303 Input buffer overflow E-304 Output buffer overflow E-305 Parser buffer overflow
E-402 Sensor open disabled output. E-403 TEC open disabled output. E-404 TEC Current limit disabled output. E-405 TEC Voltage limit disabled output. E-406 TEC resistance/reference limit disabled output E-407 TEC high temperature limit disabled output. E-409 Sensor change disabled output. E-410 TEC out of tolerance disabled output. E-415 Sensor short disabled output. E-416 Incorrect Configuration for Calibration Sequence to start. E-417 TEC output must be on to begin calibration. E-418 TEC C1, C2, or C3 constants are bad, all set to default values. E-419 Mode change disabled output. E-431 TEC link condition forced output on E-432 TEC link condition forced output off
E-900 Calculation Error shutdown output E-901 System over temperature shutdown all outputs E-903 Loading of a saved bin shutdown TEC output
Page 73
61
CHAPTER 10
10 Specifications
10.1 Temperature Controller Specifications
Specifications 3150 TEC Output
Maximum Current / Voltage 15 Amps at 23 Volts or 12.5 Amps at 28 Volts Maximum Power 350 Watts TE Current Resolution (mA) 0.458 TE Current Accuracy (mA) ± (3 % set point + 50 mA)
Current Limit
Range 0 to 15 Amps Accuracy ± 150 mA
Ripple/Noise (rms) < 10 mA Short Term Stability (1 hour) < 0.0005 °C Long Term Stability (24 hour) < 0.001 °C Temperature Coeffi cient (°C/°C) < 0.05
Display Range
Temperature -100.00°C to +240.00°C Resistance (10 µA) 0.01 kΩ to 495.000 kΩ Resistance (100 µA) 0.001 kΩ to 49.500 kΩ
Resistance (RTD) 20 Ω to 192 Ω LM335 Voltage 2331 mV to 3731 mV AD590 Current 248.15 µA to 378.15 µA TE Current
±
15.00 Amps
TE Voltage 0.0 to 30.0 Volts
Resolution
Temperature 0.01°C Resistance (10 µA) 10 Ω Resistance (100 µA) 1 Ω Resistance (RTD) 0.01 Ω LM335 Voltage 0.1 mV AD590 Current 0.01 µA TE Current 10 mA TE Voltage 0.1 Volts
Accuracy
3
Temperature ± 0.1 °C Resistance (10 µA) ± (0.04% + 16 Ω) Resistance (100 µA) ± (0.05% + 8 Ω) Resistance (RTD) ± (0.03% + 50 mΩ) LM335 Voltage ± (0.09% + 1 m V) AD590 Current ± (0.005% + 0.5 uA) TE Current ± (0.25% + 30 mA) TE Voltage ± (0.005% + 100 mV)
3
± (% of reading + fixed error)
Page 74
62 Chapter 10 Specifications
Temperature Sensors Thermistors AD590 LM335 RTD (100
ΩΩΩΩ
)
Temp Control Resolution 0.01 °C 0.01 °C 0.01 °C 0.01 °C Temp Control Accuracy ± 0.05 °C ± 0.05 °C ± 0.05 °C ± 0.05 °C
4
Sensor Bias Current or Voltage 10 µ A/100 µA +5 Volts 1 mA 1 mA
Temperature Calibration
Thermistor 1/T = (C1 x 10-3) + (C2 x 10-4)(ln R) + (C3 x 10-7)(ln R)3 AD590 T = C1 + C2 x (I
AD590
/1µA/K - 273.15)
LM335 T = C1 + C2 x (V
LM335
/10mV/K - 273.15)
Pt RTD R
T
= Ro [ 1 + C1 * T + C2 * T2] ; T ≥ 0 °C,
R
T
= Ro [ 1 + C1 * T + C2 * T2 + C3 * T3 * (T - 100)] ; T < 0 °C
Ro = resistance at 0 °C where, Ro=100Ω for a 100Ω Pt RTD.
10.2 General Specifications
Display Type LCD character display, 4 lines by 20 characters Back Lighting Green LED Controls Brightness and Contrast (contrast optimizes viewing angle) Channel Active Green LDD LED indicates that temperature controller output is on. Output Connectors Temperature Controller (TEC) High Power 7W2 female D-sub Chassis Ground 4 mm Banana Jack GPIB Connector 24 pin IEEE-488 RS232 Connector 9-pin male D-sub
Power Requirements 90 to 132V, 6A Max.; 198 to 250V, 3A Max. (user selectable), 50 to 60 Hz Size (H x W x D) 86 mm x 356 mm x 356 mm (3.5 " x 14" x 14") Mainframe Weight 5.9 kg (14 lbs.) Operating Temperature 0 to 40°C, < 80% relative humidity non-condensing Storage Temperature -20°C to 60°C, < 90% relative humidity non-condensing Isolation TEC electrically isolated with respect to earth ground.
In accordance with ongoing efforts to continuously improve our products, Newport Corporation reserves the right to modify product specifications without notice and without liability for such changes.
4
Accuracy is with lead wire resistance calibrated out.
Loading...