Newport 305-A User Manual

NEWPORT Electronics, Inc. warrants this unit to be free of defects in materials and workmanship for a
®
period of one (1) year from the date of purchase. In addition to NEWPORT’s standard warranty period, NEWPORT Electronics will extend the warranty period for one (1) additional year if the warranty card enclosed with each instrument is returned to NEWPORT.
If the unit should malfunction, it must be returned to the factory for evaluation. NEWPORT’s Customer Service Department will issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by NEWPORT, if the unit is found to be defective it will be repaired or replaced at no charge. NEWPORT’s WARRANTY does not apply to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing, operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit shows evidence of having been tampered with or shows evidence of being damaged as a result of excessive corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating conditions outside of NEWPORT’s control. Components which wear are not warranted, including but not limited to contact points, fuses, and triacs.
NEWPORT is pleased to offer suggestions on the use of its various products. However, NEWPORT neither assumes responsibility for any omissions or errors nor assumes liability for any damages that result from the use of its products in accordance with information provided by NEWPORT, either verbal or written. NEWPORT warrants only that the parts manufactured by it will be as specified and free of defects. NEWPORT MAKES NO OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESSED OR IMPLIED, EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION OF LIABILITY: The remedies of purchaser set forth herein are exclusive and the total liability of NEWPORT with respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event shall NEWPORT be liable for consequential, incidental or special damages.
CONDITIONS: Equipment sold by NEWPORT is not intended to be used, nor shall it be used: (1) as a “Basic Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application, or used on humans, or misused in any way, NEWPORT assumes no responsibility as set forth in our basic WARRANTY / DISCLAIMER language, and additionally purchaser will indemnify NEWPORT and hold NEWPORT harmless from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.
Return Requests/Inquiries
Direct all warranty and repair requests/inquiries to the NEWPORT Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO NEWPORT, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM NEWPORT’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The assigned AR number should then be marked on the outside of the return package and on any correspondence.
The purchaser is responsible for shipping charges, freight, insurance and proper packaging to
Warranty/Disclaimer
prevent breakage in transit.
FOR WARRANTY RETURNS, please have the following information available BEFORE contacting NEWPORT:
1. P.O. number under which the product was PURCHASED,
2. Model and serial number of the product under warranty, and
3. Repair instructions and/or specific problems relative to the product.
NEWPORT’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. NEWPORT is a registered trademark of NEWPORT Electronics, Inc.
© Copyright 1998 NEWPORT Electronics, Inc. All rights reserved. This document may not be copied, photocopied, reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without prior written consent of NEWPORT Electronics, Inc.
FOR NON-W NEWPORT for current repair charges. Have the following information available BEFORE contacting NEWPORT:
1. P.O. number to cover the COST of the repair,
2. Model and serial number of product, and
3. Repair instructions and/or specific problems
For immediate technical or application assistance please call:
2229 South Yale Street • Santa Ana, CA • 92704-4426 • U.S.A.
Toll Free: 1-800-639-7678 • e-mail: info@newportUS.com
976 Bergar • Laval (Quebec) • H7L 5A1 • Canada
Toll Free: 1-800-639-7678 • e-mail: info@newport.qc.ca
One Omega Drive • River Bend Technology Centre
Toll Free: 0800-488-488 • e-mail: info@newportinc.co.uk
Postbus 8034 • 1180 LA Amstelveen • The Netherlands
Toll Free: 0800 0993344 • e-mail: info@newportinc.nl
ul. Rude armady 1868 • 733 01 Karvina-Hranice • Czech Republic
Toll Free: 0800-1-666432 • e-mail: info@newport.cz
Daimlerstrasse 26 • D-75392 Deckenpfronn • Germany
Toll Free: 0130 11 21 66 • e-mail: info@newport.de
Toll Free: 0800-4-63978 • e-mail: info@newportelect.fr
En Español: (203) 359-7803 • e-mail: espanol@newportinc.com
NEWPORTnet
Newport Electronics, Inc.
TEL: (714) 540-4914 • FAX: (714) 546-3022
Newport Technologies, Inc.
TEL: (514) 335-3183 • FAX: (514) 856-6886
Newport Electronics, Ltd.
Northbank, Irlam, Manchester
Tel: 44 (161) 777-6611 • FAX: 44 (161) 777-6622
TEL: (31) 20 6418405 • FAX: (31) 20 6434643
TEL: 420 (69) 6311899 • FAX: 420 (69) 6311114
TEL: 49 (07056) 3017 • FAX: 49 (07056) 8540
9, rue Denis Papin • 78190 Trappes • France
TEL: (33) 130-621-400 • FAX: (33) 130-699-120
TEL: 95-800-639-7678 • FAX: 95-203-359-7807
M44 5EX, England
Newport Electronics B.V.
Newport Electronics spol s.r.o.
Newport Electronics GmbH
Newport Electronique S.A.R.L.
Mexico and Latin America
SM
On-Line Service
http://www.newportUS.com
It is our policy to comply with all applicable safety and EMC/EMI regulations worldwide. We are constantly pursuing certification of our products to the European New Approach Directives. We will add the CE mark to every appropriate device upon certification.
This device is marked with the international caution symbol. It is important to read the Setup Guide before installing or commissioning this device as it contains important information relating to safety and EMC.
ARRANTY REPAIRS, consult
relative to the product.
Internet e-mail
info@newportUS.com
NEWPORT
Made in the USA
MODEL 305-A
3 1/2 DIGIT MINI-SIZE VOLTMETER
SPECIFICATIONS
Analog Input
Range ±199.9 mV ±1.999 V ±19.99 V ±199.9 V Input resistance 1 G 1.1 M 1.0 M 1.0 M Bias current 50 pA 5 pA 1 pA 1 pA Maximum input 130 Vp 250 Vp 250 Vp 250 Vp
Noise Rejection
NMR, sig hi to sig lo 56 dB, 50/60 Hz CMR, sig lo to pwr gnd 120 dB, DC to 60 Hz CMV, sig lo to pwr gnd 1500 Vp per HV test; 354 Vp per
IEC spacing
CMV, sig lo to dig gnd 0 volts
Accuracy at 25°C / Display
Maximum error ±0.05% of reading ±1 count Display range ±1999 Span tempco ±0.02% of reading / °C Full-scale step response 1 second Warmup 30 minutes Reading rate 2.5 / second Overrange indication Three least-significant digits blank
Power Options
Input power 24, 100, 115 or 230 Vac ±15%,
49-63 Hz +5 or 7-32 Vdc, non-isolated
Environmental
Operating temperature 0 to 60°C Storage temperature -40 to +85°C Relative humidity 95% at 40°C (non-condensing)
Mechanical
Case type 3/64 DIN (DIN 2472-100) Bezel 1.89 x 3.78 in (48 x 96 mm) Panel cutout 1/8 DIN 1.72 x 3.62 in (45 x 92 mm) Depth behind bezel 4.32 in (110 mm), with connector Weight 8 oz (227 g) Case material 94V-1 UL-rated thermoplastic
CHG ECO DATE BY CHK
C 9352 26 OCT 98 ML
11433ML-01
®
TOP VIEWSIDE VIEW
1.89
[48.0]
BEZEL
BACKPLATE
CASE
3.78
[96.0]
3.76
[95.5]
0.76 [19.3]
4.32
[109.7]
MAX
1/8 DIN PANEL CUTOUT
0.25 [6.4] MAX
0.03 [0.8] MIN
PANEL THICKNESS
3.622 +0.032/-.000 [92.00 +0.81/-0.00]
1.772 +0.024/-.000 [45.0 +0.61/-0.00]
0.06 [1.5]
4PLCS
R
S3
P1
DISPLAY
SPAN
ZERO
W1 W2 W3
S3
A
B
C
DE
1 2 3 4 5 6 7
RANGE SELECTION DIMENSIONS
BACKPLATE
SCREWS
METER
BEZEL
LENS
PANEL WITH 1/8 DIN CUTOUT
Voltage Input Range S3
±199.9 mV dc -
±1.999 V dc A ±19.99 V dc B ±199.9 V dc C
Voltage operation was configured per customer order. 115 and 230 Vac operation use the same transformer and may be reversed with simple wiring installation.
TRANSFORMER # AC VOLTAGE INSTALL REMOVE
48131 115 W1, W3 W2 48131 230 W2 W1, W3
INSTALLATION
1. Remove the 2 phillips screws from behind the display (you don't have to go inside the meter) holding the rectangular backplate to the meter. Remove the backplate and set aside.
2. Cut or punch a hole in the panel where you want the meter to go. The panel can be as thick as 0.25" (6.4mm) to as thin as
0.03" (0.8mm).
3. Insert the meter into the panel cutout.
4. From the rear of the panel, slide the backplate over the case (smooth side out).
5. Install the 2 phillips screws to secure the meter in 1/8 DIN mount. Center the meter in hole prior to tightening screws.
ELECTRICAL CONNECTIONS
POWER AND SIGNAL INPUT
P1 ~ AC POWER - DC POWER
1 SIG HI (+S) SIG HI (+S) 2 SIG LO (-S) SIG LO (-S) 3 Hold Hold 4 Digital Return Digital Return 5 Earth GND (ý) -DC (-) 6 Neutral (N) + DC (+) 7 Line (L) No Connection (N/C)
OPTION DESCRIPTION PART # 230 V ac power input Add suffix "-C1" 100 V ac power input Add suffix "-C5" 24 V ac power input Add suffix "-C8" 5 V dc power input Add suffix "-C3"
AVAILABLE OPTIONS
7-32 V dc power input Add suffix "-C2" Green LED display (instead of Red) Add suffix "-G"
ZERO SPAN
A
B
C
S1
Decimal Point S1
1.999 A
19.99 B
199.9 C 1999 Store jumper above A
CALIBRATION
1. Apply a voltage equivalent to 95% of the range selected to the signal input.
2. Adjust the span pot for a reading of 1900 ±1 count.
3. Reinstall the lens.
Loading...