Newport Corporation warrants this product to be free from defects in material and
workmanship for a period of 1 year from the date of shipment. If found to be
defective during the warranty period, the product will either be repaired or
replaced at Newport’s option.
To exercise this warranty, write or call your local Newport representative, or
contact Newport headquarters in Irvine, California. You will be given prompt
assistance and return instructions. Send the instrument, transportation prepaid,
to the indicated service facility. Repairs will be made and the instrument returned, transportation prepaid. Repaired products are warranted for the balance
of the original warranty period, or at least 90 days.
Limitation of Warranty
This warranty does not apply to defects resulting from modification or misuse
of any product or part. This warranty also does not apply to fuses, batteries,
or damage from battery leakage.
This warranty is in lieu of all other warranties, expressed or implied, including
any implied warranty of merchantability or fitness for a particular use.
Newport Corporation shall not be liable for any indirect, special, or consequential damages.
Statement of Calibration
This instrument has been inspected and tested in accordance with specifications published by Newport Corporation.
The accuracy and calibration of this instrument and photodetector (where
applicable) is traceable to the National Institute for Standards and Technology through equipment which is calibrated at planned intervals by comparison to the certified standards maintained at Newport Corporation.
Copyright 1994, Newport Corporation
Part No. 20303-01, Rev. D
IN-11932 (06-00)
tained from the A/D converter. It may represent
amps or volts and may be analog and or digitally
filtered.
S
d
S – S
d
SS
−
()
d
R
λ
101log
()
()
SS R
−
/
d
−
λ
d
mW
/SS R
STO-REF
−
10 log
()
/SS R
d
STO-REF
Represents the value stored as a reference signal
for subsequent use in signal offset, i.e. ZERO
calculations. Sd=0 when ZERO is off.
Represents the most recent net signal value.
This is the value that is displayed when units are
set to Amps or Volts. Note that Sd = 0 when
ZERO is off.
Measurement calculation when the display units
are Watts or Joules. Rλ is the detector
responsivity associated with the current wavelength setting.
λ
Measurement calculation when the display
units are ten times the (base ten) logarithm of the
ratio of the measured power to 1 mW, i.e. dBm.
Measurement calculation when the display units
are the ratio of measured power to the value
stored by the STO-REF function.
λ
Measurement calculation when the display units
are ten times the (base ten) logarithm of the
ratio of measured power to the value stored by
the STO-REF function.
The Model 2835-C Multi-Function Optical Meter is a high performance
instrument with a wealth of measurement and triggering features designed
to provide measurement sensitivity, flexibility and speed. In spite of its
power, Model 2835-C is also designed to provide simple operation with
direct panel access to basic features and a shallow menu for access to
advanced features.
Great flexibility exists within the command structure of the Model 2835-C
so that even complex measurements can be set up quickly and easily. The
2835-C can react to or provide triggering, act as an exposure or noise meter,
or data log up to 1,000 measurements per channel!
The Model 2835-C is compatible with all of Newport’s Low-Power, High-Power
and Energy detector families. A family tree of the 2835-C compatible detectors
and accessories is shown in Figure 1 below.
The Model 2835-C connects to detectors through a calibration module containing information unique to the detector being used. Calibration modules
are ordered with a detector at the time of purchase and are labeled with the
detector’s model number and serial number. Detectors with calibration
modules have a “/CM” appended to their model number.
EXAMPLE: 818-SL (no calib. module) 818-SL/CM (with calibration module)
Please carefully read this instruction manual before using the Model 2835-C
Multi-Function Optical Meter. Be especially careful to observe the warnings and
cautions throughout this manual. If any operating instructions are not clear,
contact Newport Corporation. This instruction manual contains the necessary
information for operation and maintenance of the Newport Model 2835-C MultiFunction Optical Meter as well as information for troubleshooting and obtaining
service if necessary. This information is divided into the following sections:
Section 1General Information and Functional Description
Section 2System Operation
Section 3Principles of Operation
Section 4Computer Interfacing
1.3Unpacking and Inspection
Section 5Remote Command Reference
Section 6Maintenance, Test, and Troubleshooting
Section 7Factory Service
Appendix ASyntax and Definitions
Appendix BError Messages
Appendix CStatus Reporting System
Appendix DSample Programs
All Model 2835-C Multi-Function Optical Meters are carefully assembled,
tested and inspected before shipment. Upon receiving this instrument, check
for any obvious signs of physical damage that might have occurred during
shipment. Report any such damage to the shipping agent immediately. Retain
the original packing materials in case reshipment becomes necessary.
The Model 2835-C Multi-Function Optical Meter should have some operations
performed before measurements are made. These include:
Connecting AC Power (Section 2.5)
Detector Connection and Setup (Section 2.6)
1.5Optional Accessories and Services
The Newport Catalog presents up-to-date information on detectors, detector
accessories and detector calibration services available for use with the Model
2835-C. Refer to Figure 1 for the Model 2835-C family tree of accessories.
The Model 2835-C is designed to provide quick operation and to avoid a steep
learning curve. This section starts by giving a brief listing of display, key pad
and menu command features. Much of the Model 2835-C’s operation will be
obvious after these descriptions. The manual then explains each key operation and menu command in detail. Reference each of these detailed explanations as required when getting started.
Measurement
Activity
Annunciator
Measurement
Mode
Annunciators
The Model 2835-C incorporates a vacuum fluorescent display, VFD, which can
be clearly observed with most laser goggles and at high angles of incidence.
Figure 2 illustrates and identifies the primary regions and annunciators within
the Model 2835-C’s display.
DISP: CHA CHB
PULSECONTSNGLDC INTGMODE: % P-P
MENUEDITSHIFTSTOREUSR CALEXTATTN ZERO
FILTER: ANLG + DIG
AUTO
Figure 2a – Model 2835-C VFD Display
DISP:
= CHA + CHB
Measurement Display Area
Units Display
MODE:FLTR:
= P-P DC INTG SNGL CONT PULSE
Bar Graph, Menu and Message Display Area
SHIFTMENUEDITSTORE USRCALEXTATTNZEROAUTO
= ANLG + DIG
Display Channel
Annunciators
Signal Filtering
Annunciators
Status
Annunciators
Figure 2b – Description of Model 2835-C Display Regions
Blinking indicates that the meter is making
measurements.
DCMeter is set to make DC signal measurements.
INTGMeter is set to make INTEGRATED signal
measurements.
P-PMeter is set to make Peak-to-Peak signal
measurements.
SNGLMeter will make only one measurement per front
panel or external trigger.
CONTMeter will continuously make measurements until
stopped.
PULSEMeter is set to make pulse energy measurements.
ANLGAnalog signal filtering is on.
2.3Top Level Key Functions
DIGDigital signal filtering is on.
SHIFTIndicates that the next key press will execute a blue
key function.
MENUMeter and display are in menu mode. Measurement
is stopped.
STOREMeter is logging measurements into memory, i.e.
data logging.
USRCALUser (versus calibration module) has supplied the
detector responsivity in use.
EXTExternal trigger input is enabled.
AUTOAutomatic signal gain ranging is on.
ZEROBackground signal subtraction (zeroing) is on.
ATTNThe responsivity in use includes the affect of the
detector’s attenuator.
Operating controls for the Model 2835-C are found on both the front and rear
panels of the Model 2835-C. For measurement operation, only the front panel
controls are used. Rear panel controls are used for AC line power setup,
Section 2.5. The front panel key pad of the Model 2835-C, Figure 3, provides
quick access to measurement functions and menu access to advanced features and setup parameters. Most of the keys on the front panel only affect
the channel indicated by the display channel annunciators. Table 2 and
Sections 2.3.1 through 2.3.21 list and describe each top level key function.
Menu level key functions are discussed in Section 2.4.
(LOCAL)
SHIFT
O I
STO REF
DISP
EXT
R/S
REF SEL
FILTER
MENU
MODE
λ
ZERO
ENTER
UNITS
RANGE
AUTO
ESC
STAT S
CH A
R/S A
CH B
R/S B
Figure 3 – Front Panel Key Pad
Table 2 – Top Level Key Functions and Associated Remote Commands
KeypadRemote CommandsDescription
OI
SHIFTNoneEnables the blue key functions.
DISP
FILTERFILTER_n, FILTER_n?Cycles signal filtering:
ZEROZERO_n, ZERO_n?,Zeroes the display via offset subtraction.
AUTOAUTO_n, AUTO_n?Turns automatic gain ranging on and off.
CH ADISPCH, DISPCH?Selects channel A as the display channel.
NoneTurns the Model 2835-C on and off.
DISP, DISP?Cycles display brightness:
OFF,LOW,NORM and HIGH.
OFF, ANLG, DIG and ANLG+DIG.
STOZERO_n, ZEROVAL_n?
CH BDISPCH, DISPCH?Selects channel B as the display channel.
STO REF STOREF_n, STOREF_n?Stores last reading for future dB and
REL measurements.
REF SELREFSEL_n, REFSEL_n?Selects the reference source to be used
for dB and REL measurements.
λLAMBDA_n, LAMBDA_n?Displays and edits the calibration
wavelength in use.
RANGERANGE_n, RANGE_n?Displays signal gain range in use.
, NoneAdjusts gain range or parameter
values or moves through a list.
R/SRUN, STOPToggles the signal acquisition of
both channels.
MODEMODE_n, MODE_n?Cycles meter bet ween allowed
measurement modes.
UNITSUNITS_n, UNITS_n?Cycles display measurement units
R/S BRUN_B, RUN_B?Starts and stops channel B signal
acquisition.
EXTEXTEnable or disable external triggering.
MENUNoneEnter or exit MENU command mode.
ENTERNoneSelect next lower menu level or enter
edit mode.
ESCNoneEscape to next higher menu level or
escape edit mode.
, NoneZooms bargraph in or out. When in
Edit mode, selects the digit to be
edited by the
, keys.
Note: n in a remote command stands for A (channel A) or B (channel B).
2.3.1
OI
, Power
The OI key toggles the Model 2835-C on and off. To turn the meter on, depress
the OI key in until it clicks and stays in its depressed position. To turn the
meter off, press the key again until it clicks and rebounds to its original length.
2.3.2SHIFT
When SHIFT is pressed, the SHIFT display annunciator lights and the blue key
functions (such as STOREF) are enabled. The next key press will cause that
blue function to be executed and the “shift status” to disable.
2.3.3DISP , Display Brightness
This key cycles the display and the backlit key pad through: OFF, LOW, NORM
and HIGH brightness levels. This allows a user to operate in a dark environment without light pollution from the display, (except for one dim scanning
decimal point and the measurement activity annunciator).When the display is
OFF, the R/S, R/S A, or R/S B keys may be used to manually trigger measurements. Any other key press returns the display to the LOW brightness state
while ignoring the key function.
2.3.4FILTER, Signal Filtering
Press this key to cycle input signal filtering between: OFF, ANLG, DIG,
ANLG+DIG. This function provides methods of lowering the noise observed in
the measurement data and the analog output. When the ANLG annunciator is
lit, a 5Hz low pass filter lowers the noise floor by attenuating high frequency
signal components. ANLG filtering is not available to High Power detectors,
Energy detectors or any P-P modes.
When the DIG display annunciator is lit, measurements pass through a moving
10-sample averaging buffer before being further processed, stored or communicated to the display or computer interfaces. With DIG on, all observable values
represent digitally averaged results relative to the original A/D conversions.
This averaging is independent of the subsequent processing available through
the STATS buffer, Section 2.3.15.
This key turns offset subtraction on and off. When turned on, the ZERO
annunciator lights and the last signal reading is saved as Sd and subtracted
from all subsequent signal readings S. This causes subsequent signal calculations (and the display) to use the value S-Sd instead of S.
Offset subtraction allows one to remove the effects of ambient DC signals, by
zeroing the display before making a measurement. Pressing the ZERO key a
second time turns off the ZERO annunciator and stops offset subtraction.
2.3.6AUTO, Automatic Gain Ranging
The AUTO key toggles automatic signal ranging on and off. When on, the
AUTO annunciator lights and the signal range (amplifier gain) is adjusted to
utilize maximum analog-to-digital converter resolution. When AUTO is turned
off, the AUTO annunciator is turned off and the signal range is left in its
current state.
Signal range can be manually controlled by the , arrow keys. Pressing an
, arrow key when AUTO is on (and STATS, MENU and λ are off), turns
AUTO off and executes the manual range change. See RANGE, Section 2.3.11.
Signal range changes will often not coincide with observable changes to the
display value as the display is scaled by the detector responsivity and so must
adjust independently. Signal range gains will occur in 1 decade steps when auto
ranging and 1 or 1/2 decade steps (depending on MODE) when manual ranging.
When auto ranging in CONT PULSE mode, the arrival rate of pulses must be
above 1 Hz. In P-P CONT mode, the arrival rate of peaks must be above 50 Hz.
AUTO gain ranging is not allowed in SNGL measurement modes. When exiting
a SNGL measurement MODE, AUTO will turn back on if it was on when the
SNGL measurement mode was entered. (See MODE, Section 2.3.13)
2.3.7CH A (B), Display Channel A (or B)
CH A causes channel A to be the display channel. CH B causes channel B to be
the display channel. The display channel annunciators will indicate which
channel is currently the display channel. When a display channel is selected,
the front panel will change to reflect its settings and measurement data.
Operations from the keypad or menu that affect a specific channel will only
affect the currently displayed channel and not the other.
2.3.8STO REF, Store Reference Value
STO REF causes the last measurement, D, to be stored as D
use in relative measurement calculations. When units are relative, REL, and
the stored reference is selected, the displayed value is the ratio D/D
units are logarithmic relative, dB, and the stored reference is selected the
displayed value is the function 10 log (D/D
ref
). D
is always a power reading
ref
stored in the units of Watts. Press STO REF to cause a new D
existing D
value.
ref
for subsequent
ref
. When
ref
to overwrite the
ref
NOTE
When not using remote interface operation and when displaying relative dB
or REL measurements with the stored reference selected, the message
display area will show the STO REF value used in the calculation.
The REF SEL key is used to select the reference value, Dref, to be used for
relative, REL, and logarithmic relative, dB, unit readings. Press the REF SEL
key to display the reference source in the message display area. The EDIT
annunciator will be lit and the current selection will blink. Press , to
select a new reference source. The choices will be:
STO REF - The last stored measurement
USR REF - User defined reference (see the USRREF command in Section 5.3
Device Dependent Commands)
CH A(B) - The other channel
Press ENTER to accept the new reference source and exit the select reference
mode. Press REF SEL or ESC to exit without changing the reference source.
NOTE
The CH A(B) source is only applicable if the other channel is available. If the
other channel is not availble then N/A will appear next to CH A(B) and you
will not be able to select it.
2.3.10 λ, Wavelength
A detector calibration module contains responsivity data for its assigned
detector at discrete wavelengths. By telling the meter which wavelength is
being measured, the correct responsivity value is used in calculating the
measured power or energy. When a wavelength falls between two calibration
points, linear interpolation is used to approximate the true responsivity value.
Press the λ key to display the measurement wavelength in the message
display area. Press ENTER to light the EDIT annunciator and cause the last
digit of the wavelength to blink. Press , to adjust the blinking digit up or
down and , to change which digit blinks. Press the ENTER a second time
to accept the new wavelength and exit the wavelength edit/display mode.
Press λ or ESC to exit without changing the wavelength.
2.3.11 RANGE, Signal Range
RANGE key allows the user to view the amplifier signal range. Signal ranges
step in 1 or 1/2 decade gain increments (MODE dependent) as the RANGE is
changed in order to utilize maximum resolution from the meter’s analog-todigital converters. The available signal ranges are detector and mode dependent. Press RANGE to display the signal range number in the message display
area. If AUTO is on, then the signal range number will change if the detector signal
varies more than a decade in magnitude. Signal range changes can occur without
an effect on the displayed measurement value. Press RANGE a second time to exit
the signal range display mode.
Press the , arrow keys to increase or decrease the signal range. If AUTO
is on, pressing the , arrow keys will disable AUTO and cause the signal
range to change. Manual ranging is useful when working with external analog
recording equipment.
NOTE
Pressing the , arrow keys will disable AUTO ranging and change the signal
range even when the signal range is not being displayed via the RANGE key.
The R/S run-stop key provides front panel control over data acquisition for both
channels simultaneously. If the channel is in SNGL measurement mode, each
R/S key press causes one reading to be taken. In CONT measurement mode, each
R/S key press toggles continuous data acquisition on and off.
The activity annunciator (Figure 2b) flashes to indicate that readings are being
taken on the displayed channel as indicated in the upper right hand corner of
the display. A steady glow indicates that the meter is armed and waiting for a
pulse or P-P waveform to arrive. The indicator is off when data acquisition on
the displayed channel has been stopped. In the special case that both channels are started together in DC CONT mode the readings will be synchronized
together with channel A readings leading channel B readings by about 70 µsec.
Otherwise the channels run independent of each other.
2.3.13 MODE, Measurement Mode
The Model 2835-C provides a number of measurement modes for acquiring
data. At power on, the meter checks the detector’s calibration module to
determine which measurement modes the detector supports. Table 3
describes the measurement modes available for each detector family.
Table 3 – Newport Detector Families and Available Measurement Modes
Low-PowerHigh-PowerEnergy
DC CONT䊉䊉
DC SNGL䊊䊊
INTG䊊䊊
P-P CONT䊊
P-P SNGL䊊
CONT PULSE䊉
SNGL PULSE䊊
䊉 marks the default mode for the detector family.
Press the MODE key repeatedly until the desired mode is indicated in the mode
annunciator area of the display, see Table 4. When selecting a new mode, the
display units will change to appropriate default units. See Section 2.3.14 for
more details about units.
DC CONTMeasurement occurs at a programmable sample rate,
Section 2.4.7.
DC SNGLA measurement is taken every time the meter receives a
trigger up to a 500Hz rate when two detector calibration
modules are plugged into the meter and 1000Hz for one.
(AUTO is disabled in SNGL mode.)
INTGMeasurements occur at 400Hz and are trapezoidally
integrated to get an energy result. An R/S key press or
external trigger sets the display to 0.0000 and sampling
starts. A second R/S keypress or external trigger terminates integration.
P-P CONTAcquisition is driven by the arrival of high-low peak
pairs. A measurement is processed for every high-low
peak pair up to a frequency of 500Hz when two detector
calibration modules are plugged into the meter and
1000Hz for one. NOTE: pair captures experience a
minimum 2 or 1 ms separation respectively.
P-P SNGLA trigger1 arms the meter to capture the next high-low
peak pair. Triggers can occur at a rate of up to 500Hz
when two detector calibration modules are plugged into
the meter and 1000Hz for one. (AUTO is disabled in
SNGL mode.)
CONT PULSEThe meter captures every energy pulse up to a frequency
of 500Hz when two detector calibration modules are
plugged into the meter and 1000Hz for one. Above this
rate, pulse acquisitions experience a minimum 2 or 1 ms
separation respectively.
SNGL PULSEA trigger1 arms the meter to capture the next energy
pulse. Triggers can occur at a rate of up to 500Hz when
two detector calibration modules are plugged into the
meter and 1000Hz for one. (AUTO is disabled in SNGL
mode.)
1
“Trigger” refers to a command to start or stop signal acquisition. Trigger
sources are the R/S key, the external trigger input and remote interface
commands.
2.3.14 UNITS, Display Units
Measurements can be displayed in various units. The set of units available at
any given time is determined by the detector type and the measurement
mode. Press the UNITS key repeatedly to cycle the display through the set of
available units. Table 5 describes the sets of available units for each detector
family as a function of the measurement mode.
Table 5 – Valid Display Units Available to Detector Families by MODE.
FAMILYMODEVAWW/cm
2
JJ/cm2ErgErg/cm2dBmdBREL
Low-PowerDC CONT䊊䊉䊊䊊䊊䊊
Low-PowerDC SNGL䊊䊉䊊䊊䊊䊊
Low-PowerINTG䊉䊊䊊䊊
Low-PowerP-P CONT䊊䊉䊊䊊䊊䊊
Low-PowerP-P SNGL䊊䊉䊊䊊䊊䊊
High-PowerDC CONT䊊䊉䊊䊊䊊䊊
High-PowerDC SNGL䊊䊉䊊䊊䊊䊊
High-PowerINTG䊉䊊䊊䊊
EnergyCONT PULSE䊊䊉䊊䊊䊊䊊
EnergySNGL PULSE䊊䊉䊊䊊䊊䊊
䊉 denotes the default units for the detector family in the given measurement
mode. Some display units are abbreviated. Table 6 lists displayed units
versus actual measurement units. Display units are limited to four characters
in order to provide for display engineering prefixes such as: p, n, µ, m and k
(pico, nano, micro, milli and kilo respectively).
Table 6 – Displayed Unit Abbreviations Versus Actual Measurement Units.
Actual UnitsVAWJErgW/cm2J/cm2Erg/cm2dBmdBREL
Displayed UnitsVAWJERGW/cmJ/cmE/cmdBmdBE±dd
2.3.15 STATS, Moving Statistics
The STATS key causes a list of statistical results from the stats buffer of the
display channel to be displayed in the message display area. The stats buffer
is a moving data window containing the most recent measurements to a depth
1 ≤ N ≤ 100. The default value is N = 10. Display occurs without disrupting
data acquisition or storage and results are continuously updated.
Press the STATS key to enter the stats display list. The first statistic displayed
will be the stats buffer depth N. Press the , adjust keys to move through
the list. The following statistics are available: N, MAX, MIN, MAX-MIN, MEAN
and STD DEV.
When the STATS buffer depth, N, is displayed, it can be adjusted by pressing
ENTER and using the , keys to set a new value for N. Press the ENTER
key to adopt the new value for N and escape the edit mode. The STATS buffer
is cleared whenever a new stats buffer depth N, or MODE or UNITS is established. Press the STATS key a second time to exit the stats display mode.
2.3.16 R/S A(B), Run-Stop Channel A (or B)
The R/S A and R/S B run-stop keys provides front panel control over data
acquisition for each individual channel. When in SNGL measurement mode,
each R/S A(B) key press causes one reading to be taken on the respective
channel. In CONT measurement mode, each R/S A(B) key press toggles
continuous data acquisition on and off on the respective channel.
The activity annunciator (Figure 2b) flashes to indicate that readings are being
taken on the display channel as indicated in the upper righthand corner of the
display. A steady glow indicates that the meter is armed and waiting for a
pulse or P-P waveform to arrive. The indicator is off when data acquisition on
the displayed channel has been stopped.
2.3.17 EXT, External Trigger
EXT enables and disables the triggering of data acquisition through the rear
panel trigger input BNC connector. The meter can be configured to be
triggered by either a rising or falling edge TTL signal. See Section 2.4.9.
Press EXT to light the EXT annunciator and enable the external trigger input to
both channels. The Model 2835-C will still respond to triggers from the R/S, R/S
A, or R/S B keys even when the EXT trigger is enabled. Press the EXT key a
second time to turn off the annunciator and disable the external trigger input.
2.3.18 MENU
The MENU key provides access to advanced features and parameters. Press
the MENU key to stop all data acquisition, light the MENU annunciator, and
display the first item of the top level menu list. Press the MENU key a second
time to immediately exit the menu. Menu structure and functions are presented in Section 2.4. The menu consists of a series of lists and parameter
values. Most parameter values can be edited to configure the meter. Lists are
moved through via the , keys. To move to a lower level list, press the
ENTER key. To escape a lower level list, press the ESC key.
In the menu, all key functions are ignored except for the MENU, ENTER, ESC
and the , keys. Each key will act without having to initially press SHIFT.
Pressing SHIFT will toggle the SHIFT annunciator, but it will not effect anything
else when in MENU mode.
To edit a displayed parameter press ENTER. This enables the edit mode and
causes the parameter (or a digit) to blink. Press the , keys to adjust the
parameter through its allowed list or to count the blinking digit up and down.
When adjusting a numerical value, the , keys can be used to change
which decimal digit blinks. Press ENTER to adopt the new value and escape
the editing mode. Press ESC to leave the editing mode without adopting the
new value.
2.3.19 ENTER
ENTER allows one to move lower into the menu when in menu mode, to enter
editing mode when an editable parameter is being displayed and to adopt a
new parameter value after it has been edited. Editable parameters can be
found both in and out of the menu mode. If ENTER is pressed when a noneditable parameter is being displayed, the key press is ignored.
2.3.20 ESC, Escape
ESC allows one to escape to the next higher level when in menu mode and to
escape editing mode when editing a parameter value without adopting any
changes to the parameter. At the top level of the menu, pressing the ESC key
exits the menu mode. The ESC key also allows one to escape various other
display modes such as STATS or wavelength display or amplifier signal gain
range display modes. When there is nothing to escape from, the ESC key is
ignored.
The , , , keys allow one to adjust various parameter states and values.
In normal operation, the , keys turn AUTO off and adjust the amplifier
signal gain. In any other mode, the , keys cause one to move through a list
or to adjust a blinking digit. The , keys zoom the bargraph ( in, out)
and in edit mode, select which digit of a numerical value will blink.
Menu functions provide control over parameter values and methods of
making measurements. The menu consists of a number of lists and parameter values. Table 7 presents a quick summary of the hierarchy of the
menu functions and parameters. Most settings displayed in the menu and
changes to those settings affect the channel indicated by the display
channel annunciator.
2.4.1Menu Access and Movement
To access the menu press MENU. The MENU annunciator will light and data
acquisition will stop. Menu labels and parameter values are displayed in the
message display area. The first, second and third leftmost decimal points
within this area indicate a current position at the top, second, third or fourth
level of the menu hierarchy, Figure 4. Table 7 presents the menu functions
and their hierarchy.
Top Level
Second Level
Third Level
Fourth Level
Figure 4 – Decimal Point Indication of Menu Hierarchy Position
In the menu, valid keys are limited to those needed to perform operations to
move through the menu or edit a parameter value. In the menu, it is not
necessary to hit the SHIFT key to invoke the action of the menu keys: MENU,
ENTER, ESC or , , , . The SHIFT key will toggle the SHIFT annunciator,
but with no effect on subsequent pressings of other keys. Once in the menu,
the rules for moving through the menu are as follows:
i. Press ENTER to move to the next level down or to enter the edit mode if a
editable parameter is being displayed and to accept a parameter value in
its currently displayed state and exit edit mode.
ii. Press ESC to move to the next level up or to escape the menu when at the
top level or to exit edit mode without changing the parameter value being
edited.
iii. Press MENU to immediately escape the menu regardless of current level