Newport Corporation warrants this product to be free from defects in material and workmanship for a period of 1 year from the date of shipment.
found to be defective during the warranty period, the product will either be
repaired or replaced at Newport's option.
To exercise this warranty, write or call your local Newport representative, or
contact Newport headquarters in Irvine, California. You will be given prompt
assistance and return instructions. Send the instrument, transportation
prepaid, to the indicated service facility. Repairs will be made and the instrument returned, transportation prepaid. Repaired products are warranted for
the balance of the original warranty period, or at least 90 days.
If
Limitation of
This warranty does not apply to defects resulting from modification or misuse
of any product or part. This warranty also does not apply to fuses, batteries,
or damage from battery leakage.
This warranty is in lieu of all other warranties, expressed or implied, including
any implied warranty of merchantability or fitness for a particular use.
Newport Corporation shall not be liable for any indirect, special, or consequential damages.
Warranty
Statement of Calibration
This instrument has been inspected and tested in accordance with specifications published by Newport Corporation.
The accuracy and calibration of this instrument and photodetector (where
applicable) is traceable to the National Institute for Standards and Technology through equipment which is calibrated at planned intervals by comparison to the certified standards maintained at Newport Corporation.
Model 2832-C Controller and Accessories
Model 2832-C VFD Display
Description of Model 2832-C Display Regions
Front Panel Key Pad
Decimal Point Indication of Menu Hierarchy Position
Rear Panel Power Supply Voltage Switches in Positions
Connecting a Detector with its Calibration Module
Model 2832-C Detector Calibration Module Input Port
Model 2832-C Analog Signal Flow Diagram
Model 2832-C Digitized Signal Flow Block Diagram
RS-232 Cable Connections
ment with a wealth of measurement and triggering features designed to
provide measurement sensitivity, flexibility and speed. In spite of its power,
is
Model 28324
access to basic features and a shallow menu for access to advanced features.
Great flexibility exists within the command structure of the Model 28324 so
that even complex measurements can be set up quickly and easily. The
28324 can react to or provide triggering or data log up to
per channel!
The Model 28324
family tree of the 28324 compatible detectors and accessories
1
Figure
below.
also designed to provide simple operation with direct panel
is
compatible with Newport's Low-Power detector family.
The Model 28324 connects to detectors through a calibration module con-
taining information unique to the detector being used. Calibration modules
are ordered with a detector at the time of purchase and are labeled with the
detector's model number and serial number. Detectors with calibration
modules have a "/CM appended to their model number.
EXAMPLE: 818SL (no calib. module) 818SL/CM (with calibration module)
Scope of
Thii
Manual
Please carefully read this instruction manual before using the Model 28324
Be
DualChannel Power Meter.
cautions throughout this manual.
contact Newport Corporation. This instruction manual contains the necessary
information for operation and maintenance of the Newport Model 28324
Channel Power Meter
if
service
Section 1
Section 2 System Operation
Section
Section
Section
Section
Section
Appendix A Syntax and Definitions
Appendix
Appendix
necessary. This information
3
4
5
6
7
B
C
as
especially careful to observe the warnings and
If
any operating instructions are not clear.
well
as
information for troubleshooting and obtaining
is
divided into the following sections:
General Information and Functional Description
Principles of Operation
Computer Interfacing
Remote Command Reference
Maintenance, Test, and Troubleshooting
Factory Service
Error Messages
Status Reporting System
Dual-
Unpacking
and
Inspection
Appendix
All Model 28324 Dualchannel Power Meters are carefully assembled, tested
and inspected before shipment. Upon receiving this instrument, check for any
obvious signs of physical damage that might have occurred during shipment.
Report any such damage to the shipping agent immediately. Retain the
original packing materials in case reshipment becomes necessary.
The Model 2832-C
learning curve. This section starts by giving a brief listing of display, key pad
and menu command features. Much of the Model
obvious after these descriptions. The manual then explains each key operation and menu command in detail. Reference each of these detailed explanations as required when getting started.
The Model
be clearly observed with most laser goggles and at high angles of incidence.
Figure 2 illustrates and identifies the primary regions and annunciators within
the Model 2832-C's display.
28324 incorporates
n n
'a.
I
MODE:%
is
designed to provide quick operation and to avoid a steep
Secondary Channel, Bar Graph, Menu and Message Display Area
Figure 2b - Description of Model 2832C Display Regions
4
I
I
VFD
=
I
Units Display
I
Display
CHA
CHB
+
I
Display Channel
Annunciators
Signal Filtering
Armunciators
Page 20
Table
1
-
Model
Z83ZC
Annunciator Comment
Display Annunciaton
CH A
CH B
DC
SNGL
CONT
ANLG
DIG
SHIFT
MENU
The display
primary display area.
The display is showing Channel B information in
primary display area.
Blinking indicates that the meter is making
measurements.
Meter is set to make
Meter will make only one measurement per front
panel or external trigger.
Meter will continuously make measurements until
stopped.
Analog signal filtering is on.
Digital signal filtering
Indicates that the next key press will execute a blue
key function.
Meter and display are in menu mode. Measurement
is
stopped.
is
showing Channel A information in
DC
signal measurements.
is
on.
STORE
USRCAL
EXT
AUTO
ZERO
A'ITN
Operating controls for the Model
panels of the Model
are
controls
Section
quick access to measurement functions and menu access to advanced features and setup parameters. Most of the keys on the front panel only affect
the channel indicated by the display channel annunciators. Table
Sections
Menu
used. Rear panel controls are used for AC line power setup,
2.5.
The front panel key pad
2.3.1
level key
28324.
through
functions are discussed in Section
2.3.21
is
Meter
data logging.
User (versus calibration module) has supplied the
detector responsivity in use.
External trigger input is enabled.
Automatic signal gain ranging is on.
Background signal subtraction (zeroing) is on.
The responsivity in use includes the affect of the
Displays and edits the calibration
wavelength in use.
Displays signal gain range in use.
Adjusts gain range or parameter
values or moves through a list.
Toggles the signal acquisition of
both channels.
Cycles meter between allowed
measurement modes.
Cycles display measurement units
between those allowed.
Displays statistics.: Max, Min,
Max-Min, Mean and Std.
Starts and stops channel A signal
acquisition.
Dev.
Page 22
R/S
B
RUN-B,
RUN-B?
Starts and stops channel
acquisition.
B
signal
EXT EXT
MENU None Enter or exit
ENTER None
ESC None
4,
None
Note: n in a remote command stands forA (channel A) or B (channel B).
Enable or disable external triggering.
MENU
command mode.
Select next lower menu level or enter
edit mode.
Escape to next higher menu level or
escape edit mode.
Zooms
bargraph in or out. When in
Edit mode, selects the digit to be
edited by the
V,
A
keys.
2.3.1 &A, Power
The
AA
key toggles the Model
the
&A
key in until it clicks and stays in its depressed position. To turn the
meter
off,
press the key again until it clicks and rebounds to its original length.
If
the meter
LOCAL key and puts the meter into the local mode. The remote mode is desig-
nated on the front panel by displaying an asterisk
the secondary channel display area.
is
in the remote mode, depressing this key the first time serves
2832€
on and
If
the meter
off.
To turn the meter on, depress
#f
at the right-hand side
is
in the local-lockout mode,
as
of
a
is
displayed on the front panel.
-
If
the meter
When SHIFT
functions (such as STOREF) are enabled. The next key press will cause that
blue function to be executed and the "shift status" to disable.
2.3.3 DISP:+., Display
This key cycles the display and the backlit key pad through: OFF, LOW, NORM
and HIGH brightness levels. This allows a user to operate in a dark environ-
ment without light pollution from the display, (except for one dim scanning
decimal point and the measurement activity annunciator). When the display is
OFF, the R/S, R/S A, or R/S
ments. Any other key press returns the display to the LOW brightness state
while ignoring the key function.
2.3.4
Press this key to cycle input signal filtering between: OFF, ANLG, DIG,
ANLG+DIG. This function provides methods of lowering the noise observed in
the measurement data and the analog output. When the ANLG annunciator is
lit, a 5Hz low pass filter lowers the noise floor by attenuating high frequency
signal components.
is
not in the remote mode, pressing this key serves as a SHIFT.
is
pressed, the SHIFT display annunciator lights and the blue key
10-sample averaging buffer before being further processed, stored or commu-
nicated to the display or computer interfaces. With DIG on, all observable
values represent digitally averaged results relative to the original
sions. This averaging is independent of the subsequent processing available
through the STATS buffer, Section
is
lit, measurements pass through a moving
A/D
conver-
2.3.15.
2.3.5 ZERO, Offset Subtraction
This key turns offset subtraction on and off. When turned on, the ZERO
annunciator lights and the last signal reading
from all subsequent signal readings S. This causes subsequent signal calcula-
tions (and the display) to use the value
Offset subtraction allows one to remove the effects of ambient DC signals, by
zeroing the display before making a measurement. Pressing the ZERO key a
second time turns off the ZERO annunciator and stops offset subtraction.
is
saved
as
SS, instead of S.
S, and subtracted
2.3.6 AUTO, Automatic Gain
The AUTO key toggles automatic signal ranging on and off. When on, the
AUTO annunciator lights and the signal range (amplifier gain) is adjusted to
utilize maximum analog-to-digital converter resolution. When AUTO
off, the AUTO annunciator
current state.
Signal range can be manually controlled by the
V
,
A
arrow key when AUTO
AUTO off and executes the manual range change. See RANGE, Section 2.3.1
Signal range changes will often not coincide with observable changes to the
as
display value
adjust independently. Signal range gains will occur in
or manual ranging.
AUTO gain ranging
a SNGL measurement MODE, AUTO will turn back on if it was on when the
SNGL measurement mode was entered. (See MODE, Section 2.3.13)
2.3.7
The 2832-C has a primary channel and secondary channel display area. The
primary channel
channel displayed underneath it in slightly smaller characters.
Depressing the CH A key causes channel A to be the primary display channel.
The CH
display channel annunciators will indicate which channel
primary display channel. When a display channel is selected, the front panel
will change to reflect its settings and measurement data. Operations from the
keypad or menu that affect a specific channel will only affect the currently
displayed primary channel and not the other.
CH A
B
the display
is
0,
Display Channel A (or
is
displayed in larger, bold characters, with the secondary
key causes channel B to be the primary display channel. The
is
is
not allowed in SNGL measurement modes. When exiting
Ranging
is
turned
turned off and the signal range
V , A
arrow keys. Pressing an
is
on (and STATS, MENU and hare off), turns
scaled by the detector responsivity and so must
1
is
left in its
decade steps when auto
B)
is
currently the
1.
2.3.8
ST0
REF
use in relative measurement calculations. When units are relative,
the stored reference
units are logarithmic relative, dB, and the stored reference is selected the
displayed value
stored in the units of Watts. Press ST0 REF to cause a new D, to overwrite the
existing
D,
is
value.
the function
10
log (DID,). D,
is
always a power reading
NOTE
When not using remote interface operation and when displaying relative dB
or REL measurements with the stored reference selected, the message
display area will show the ST0 REF value used in the calculation.
2.3.9
The REF SEL key
relative, REL, and logarithmic relative, dB, unit readings. Press the REF SEL
key to display the reference source in the message display area. The EDIT
annunciator will be lit and the current selection will blink. Press
select a new reference source. The choices will be:
ST0 REF
USR REF
Device Dependent Commands)
CH A@) -The other channel
Press ENTER to accept the new reference source and exit the select reference
mode. Press REF SEL or ESC to exit without changing the reference source.
REF
SEL,
Select Reference
is
used to select the reference value, Dref, to be used for
-
The last stored measurement
-
User defined reference
(see
V, A to
the USRREF command in Section
5.3
NOTE
The CH A(B) source
other channel
will not be able to select it.
2.3.10
A detector calibration module contains responsivity data for its assigned
detector at discrete wavelengths. By telling the meter which wavelength is
being measured, the correct responsivity value
measured power or energy. When
points, linear interpolation
A,
Wavelength
is
only applicable
is
not available then N/A will appear next to CH A@) and you
is
used to approximate the true responsivity value.
if
the other channel
a
wavelength
is
available.
is
used in calculating the
falls
between two calibration
If
the
Press the
display area. Press ENTER to light the EDIT annunciator and cause the last
digit of the wavelength to blink. Press
down and
to accept the new wavelength and exit the wavelength edit/display mode.
Press
2.3.1 1 RANGE,
RANGE key allows the user to view the amplifier signal range. Signal ranges
step in
the meter's analog-to-digital converters. The available signal ranges are
detector dependent. Press RANGE to display the signal range number in the
message display
the detector signal varies more than a decade in magnitude. Signal range changes
can occur without an effect on the displayed measurement value. Press RANGE a
key to display the measurement wavelength in the message
V, A to adjust the blinking digit up or
4,)
to change which digit blinks. Press the ENTER a second time
h
or ESC to exit without changing the wavelength.
Signal
1
decade gain increments in order to utilize maximum resolution from
area.
time to exit the signal range display mode.
If
AUTO
Range
is
on, then the signal range number will change
if
Page 25
Press the
is on, pressing the
range to change. Manual ranging
recording equipment.
V , A
arrow keys to increase or decrease the signal range.
V,
A
arrow keys will disable AUTO and cause the signal
If
AUTO
is
useful when working with external analog
NOTE
Pressing the
range even when the signal range is not being displayed via the RANGE key.
The
RJS
channels simultaneously.
R/S
key press causes one reading to
R/S
key press toggles continuous data acquisition on and off.
The activity annunciator (Figure 2b) flashes to indicate that readings are being
taken on the displayed channel as indicated in the upper right hand corner of
the display. The indicator is off when data acquisition on the displayed
channel has been stopped. In the special case that both channels are started
together in DC CONT mode the readings will be synchronized together with
channel A readings leading channel B readings by about 70
channels run independent of each other.
V , A
arrow keys will disable AUTO ranging and change the signal
runstop key provides front panel control over data acquisition for
If
the channel
is
in SNGL measurement mode, each
be
taken.
In
CONT measurement mode, each
psec. Otherwise the
both
2.3.13 MODE, Measurement Mode
The Model 28324 acquires data in either continuous or in a single-step mode.
Press the MODE key until the desired mode is indicated in the mode annuncia-
tor area of the display. Table
on the 2832-C.
3
gives a description of the two modes available
Table 3 - Model 2832C Measurement Modes
Mode Description
DC CONT Measurement occurs at a programmable sample rate,
Section 2.4.7.
DC SNGL
2.3.14
Measurements can be displayed in various units. The set of available units is
described in Table 4. Press the UNITS key to cycle the display through the set
of available units.
UNlTS,
Table
4
-
Valid Display Units Available to Low Power Detectors by MODE.
MODE A W W/cm2 dBm dB
DCSNGL
A measurement is taken every time the meter receives a
trigger up to a 500Hz rate when two detector calibration
denotes the default units for the detector in the given measurement mode.
5
Some display units are abbreviated. Table
measurement units. Display units are limited to four characters in order to
provide for display engineering prefixes such as: p, n,
micro, milli and kilo respectively).
Table
5
-
Displayed Unit Abbreviations Versus Actual Measurement Units.
Actual Units A W W/cm2 dBm dB REL
Displayed Units A W W/cm dBm dB L+dd
lists displayed units versus actual
p,
m and k (pico, nano,
2.3.15 STATS, Moving Statistics
The STATS key causes a list of statistical results from the stats buffer of the
display channel to be displayed in the message display area. The stats buffer
is
a moving data window containing the most recent measurements to a depth
1 I N I 100.
data acquisition or storage and results are continuously updated.
Press the STATS key to enter the stats display list. The first statistic displayed
will be the stats buffer depth
the list. The following statistics are available:
and STD DEV.
When the STATS buffer depth, N, is displayed, it can be adjusted by pressing
ENTER and using the
key to adopt the new value for N and escape the edit mode. The STATS buffer
is
cleared whenever a new stats buffer depth N, or MODE or UNITS is estab
lished. Press the STATS key a second time to exit the stats display mode.
The default value
V,
A
keys to set a new value for N. Press the
is
N
N.
Press the
=
10.
Display occurs without disrupting
V
,
A
adjust keys to move through
N,
MAX, MIN, MAX-MIN, MEAN
ENTER
2.3.16 R/S
The R/S A and R/S B run-stop keys provides front panel control over data
acquisition for each individual channel. When in SNGL measurement mode,
each R/S A@) key press causes one reading to be taken on the respective
channel. In CONT measurement mode, each R/S A@) key press toggles
continuous data acquisition on and off on the respective channel.
The activity annunciator (Figure 2b) flashes to indicate that readings are being
taken on the display channel as indicated in the upper righthand corner of the
display. The indicator
has been stopped.
2.3.17
EXT enables and disables the triggering of data acquisition through the rear
panel trigger input BNC connector. The meter can be configured to be
triggered by either a rising or falling edge
Press EXT to light the
both channels. The Model 2832€ will still respond to triggers from the R/S,
A, or R/S B keys even when the EXT trigger
second time to turn off the annunciator and disable the external trigger input.
A@),
RunStop Channel A (or
EXT,
External Trigger
B)
is
off when data acquisition on the displayed channel
TTL signal. See Section 2.4.9.
EXT
annunciator and enable the external trigger input to
The MENU key provides access to advanced features and parameters. Press
the MENU key to stop all data acquisition, light the MENU annunciator, and
display the first item of the top level menu list. Press the MENU key a second
time to immediately exit the menu. Menu structure and functions are pre-
sented in Section
values. Most parameter values can be edited to configure the meter. Lists are
moved through via the
ENTER key. To escape a lower level list, press the ESC key.
In the menu, all key functions are ignored except for the MENU, ENTER, ESC
and the
Pressing SHIm will toggle the SHIm annunciator, but it will not effect anything
else
To edit a displayed parameter press ENTER. This enables the edit mode and
causes the parameter (or a digit) to blink. Press the
parameter through its allowed list or to count the blinking digit up and down.
When adjusting a numerical value, the
which decimal digit blinks. Press ENTER to adopt the new value and escape
the editing mode. Press ESC to leave the editing mode without adopting the
new value.
MENU
2.4.
The menu consists of a series of lists and parameter
V, A keys. To move to a lower level list, press the
V, A keys. Each key will act without having to initially press SHIFT.
when in MENU mode.
V, A keys to adjust the
4
,
b
keys can be used to change
2.3.19
ENTER allows one to move lower into the menu when in menu mode, to enter
editing mode when an editable parameter
new parameter value after it has been edited. Editable parameters can be
found both in and out of the menu mode.
editable parameter is being displayed, the key press
2.3.20
ESC
escape editing mode when editing a parameter value without adopting any
changes to the parameter. At the top level of the menu, pressing the
exits the menu mode. The
display modes such
range display modes. When there
ignored.
2.3.21
The
In normal operation, the
signal gain. In any other mode, the
or to adjust a blinking digit. The4
and in edit mode, select which digit of a numerical value will blink.
ENTER
is
being displayed and to adopt a
If
ENTER
ESC,
Escape
allows one to escape to the next higher level when in menu mode and to
ESC
key also allows one to escape various other
as
STATS or wavelength display or amplifier signal gain
is
nothing to escape from, the
V , A
,4
,
b
Adjust
V
,
A
,4
,
b
keys allow one to adjust various parameter states and values.
Menu functions provide control over parameter values and methods
making measurements. The menu consists
6
eter values. Table
menu functions and parameters. Most settings displayed in the menu and
changes to those settings affect the channel indicated by the display
channel annunciator.
2.4.1
To access the menu press MENU. The MENU annunciator will light and data
acquisition will stop. Menu labels and parameter values are displayed in the
message display area. The first, second and third
within this
level of the menu hierarchy, Figure
and their hierarchy.
Menu
area
presents a quick summary
Access
indicate a current position at the top, second, third or fourth
and
Movement
Top Level
Second Level
4.
Table 6 presents the menu functions
0
of
0
a number
of
leftmost decimal points
of
lists and param-
the hierarchy
0
of
of
the
. .
Third Level
1.1. 1.
Fourth Level
Figure
4
-
Decimal Point Indication
In the menu, valid keys
move through the menu or edit a parameter value. In the menu, it is not
necessary to hit the SHIn key to invoke the action
ESC
or
ENTER,
but with no effect on subsequent pressings of other keys. Once in the menu,
the rules for moving through the menu are
i. Press ENTER to move to the next level down or to enter the edit mode if a
editable parameter is being displayed and to accept a parameter value in
its currently displayed state and exit edit mode.
ii. Press
top level or to exit edit mode without changing the parameter value being
edited.
iii. Press MENU to immediately escape the menu regardless
within the menu.
iv. Use the
level. Also use these keys to adjust a parameter or decimal value when in
edit mode.
V , A, 4 ,
ESC
to move to the next level up or to escape the menu when at the
V,
A
are
limited to those needed to perform operations to
b
.
The SHIn key will toggle the SHIn annunciator,
keys to move up or down through a menu list on a given
channel for a total
an internal buffer for subsequent viewing or transmission over a computer
interface. A separate buffer is maintained and configured for each channel.
Data is maintained on power down, but lost when
loaded, Section 2.4.3, or when the buffer is cleared via the CLR DATA BUFFER
command or when data with new units is being stored.
The data store buffer operates in two ways: SLIDE
tion, the buffer slides along storing the most recent measurements up to the
of
size
pushed out and lost. In
buffer is full. After this, data acquisition stops and no additional data can
stored without first clearing the buffer via the CLR DATA BUFFER command.
DATA BUFFER CLR is displayed when the buffer is empty.
The size
value
before dropping old data or stopping data storage.
Data storing
STORE annunciator.
When in CONT mode, data acquisition and storage is resumed immediately
the buffer. Beyond this,
of
the buffer is set by the D-BUF-SIZE
dddd
to establish the number
is
enabled by the user via the DATA STORE menu function. Edit
ON,
the
upon exiting the menus if acquisition is active
acquisition
data acquisition and storage requires an initiating R/S key press, external
trigger or a remote RUN command upon exiting the menu.
OEFcondition to enable or disable data storing and the associated
is
not active when the menu is entered, or when in SNGL mode,
measurements. These measurements are stored in
as
data enters the buffer, the oldest data is
FIX
configuration, data storing continues until the data
of
1000
measurements for each
a
new configuration is
or
FIX.
In SLIDE configura-
dddd
menu function. Edit the
data points that the buffer will hold
as
the menu is entered. When
be
can
be
of
dddd.
viewed
is
Buffer data
in the measurement
buffer:
dddd
dddd = 0001
value
via
the
VIEW
area
while the message
Use the
the first,
V , A
i.e.
the oldest datum.
DATA menu command. Data are displayed
area
displays the position within
keys to move through the buffer data. The
the
NOTE
When DATA STORE is off, CONT acquisition mode behavior defaults to the
condition where data acquisition is begun without the requirement
The Model
meter and to recall that configuration for later use even
turned off. This
nonvolatile memory. Configuration buffers are numbered
being a DEFAULT buffer which can only
reset state of all the buffers except the default buffer
28324
provides a method to save the configuration
if
is
accomplished through configuration buffers maintained in
be
recalled but not saved to. The
is
empty. Empty buffers
of
the entire
the meter has been
0
to 9 with buffer
0
cannot be recalled.
A recalled configuration becomes the current configuration of the meter. Any
changes to the current configuration must be saved via SAVE CONFIG or they
is
will be lost when a new configuration
The meter will not recall a configuration that
recalled.
is
not compatible with either
detector calibration module currently plugged into the meter. Configurations
using the same model of detector are compatible while configurations using
different models of detectors are incompatible. The list of configuration
as
parameters stored in a configuration buffer as well
Parameter Default Condition
D-BUF-SIZE 100
Data Store Units Same as UNITS
EXT OFF
EXT TRlG IN
TRIGGER OUT
FALLING
TRIG ON CMPLT
TRIG OUT POL TRIG ACTIVE LO
=
=
0.001
30 Hz
W
TRIG OUT FREQ FREQ
TRIG LEVEL
BAR GRAPH
LVL
OFF
GPlB ADDR 05
BAUD RATE 9600
These parameters adopt the following default values at power up and are not
affected by recalling a configuration or by setting the configuration to default.
6232 ECHO OFF
Local Lockout OFF
Display Brightness NORMAL
Table 9. below lists the menu commands effecting the saving and recalling of
meter configurations.
Table 9 -Meter Configuration Operations
Menu Operation Keypad Commands
Associated Remote Commands
SAVE CONFIG
Adjust SAVE TO
d,
*SAV
and ENTER
RECALL CONFIG Adjust RECALL
d,
*RCL,*RST
and ENTER
2.4.4
AUTO
CAL
The AUTO CAL command causes the 2832-C to perform AID conversions of
amplifier offset voltages (zero errors) arising from aging and temperature
effects. These conversions are then used in subtracting the appropriate error
voltage from each reading during normal operation. The 2832€ automatically
is
performs this procedure every time it powers up (or
reset). To achieve
stable reading at the specified accuracy, AUTO CAL should be executed for
each channel after a minimum 60 minute warm-up period from power-up.
Simply press ENTER when AUTO CAL
is
displayed. The display message area
will display "ONE MOMENT", followed by a buzzer sound indicating that
AUTO CAL is complete. Low Power detectors do not have to be connected to
the 28324 or removed from the radiation source to effectively execute AUTO
C AL.
The Attenuator selects the responsivity value,
detector alone or the value for the
ON the AITN annunciator
is
attenuator
alone responsivity
USR CAL
used. When AITN
is
is
on (Section 2.4.6) the AITN setting has no effect.
used.
is
If
detector-with-attenuator.
lit and the responsivity of the detector-with-
is
OFF, the annunciator
the detector does not have an attentuator, or if
&,
to be the value for the
When AITN
is
off and the detector-
is
2.4.6 User Calibration
The Model 28324 allows one to create a detector responsivity which overrides the responsivities obtained from the detector's calibration module. This
allows one to account for the effects of additional optics and filters in the
measurement path. When USR CAL
the meter adopts the responsivity value displayed by the editable USR RESP
value. Use the PRESENT RESP function to display the current calibration
module responsivity.
The USR RESP units are the same as the PRESENT RESP units. Table 10 lists
the possible user calibration operations.
Table
10
Menu Operation Keypad Commands Associated Remote Commands
USR CALlB
ON,
OFF
Edit USR CALlB USRCALn, USRCAL-n?
ON,
off
is
on, the USR CAL annunciator
-
User Calibration Operations
is
lit and
USR RESP Edit
PRESENT RESP ENTER to view
*These units are not editable.
d.dddEkdd
d.dddEMd
A/W* USRRESP-n, USERRESP-n?
RES
P-n?
A/W*
NOTE
n in a remote command stands for A (channel A) or B (channel B).
2.4.7
The Model 28324 incorporates two analog-tdigital,
with 20,000 count resolution and a second with 4096 count resolution. The
user may select which
and DC SNCL acquisition modes. All other modes use the 4096 count
The 20,000 count
When two detector calibration modules are plugged into the meter, the 4,096
count
detector calibration module is plugged into the meter, the 4,096 count
converter can operate at sample rates up to 1000
The SAMPLE PREC menu command, Table 11, selects which analog-todigital
converter
nels can be adjusted within the limits imposed by the SAMPLE PREC state, see
Table 12. When the SAMPLE PREC state changes, the sampling frequency
defaults to 25
SAMPLE PREC state. The SAMPLE FREQ and the SAMPLE PREC settings affect
both channels, not just the display channel.
DC
Sampling
(A/D)
converters, one
AD will
A/D
converter can operate at sample rates up to 25
A/D
converter can operate at sample rates up to 500
be
used for both channels during
Hz.
Hz.
is
used for both channels. The sampling frequency for both chan-
Hz
if the existing SAMPLE FREQ is incompatible with the new
The Model
activities among other instruments. The trigger output can operate in several
ways: conversion complete, periodic output and comparator output. In
addition, the polarity of the trigger output can be specified.
Trigger
2832C's
20,000
4,096
Keypad Commands Associated Remote Commands
4096
ddd.ddd
Output
counts
counts
Table
12
-
DC
SAMPLING Operations.
20,000,
CNT
HZ
rear panel lTL trigger output allows it to coordinate
0.001
Hz to
25.0
0.001
Hz
to
500.0
SPREC, SPREC?
Hz
Hz or
1000.0
Hz
CAUTION
Precautions should
coordinate activities among
may
meter
cause
false
be
taken whenever the trigger output
instruments.
triggering,
causing
Static
discharges
the trigger output to go active.
is
at or
used
near
to
the
In TRIG ON CMPLT mode, a pulse is output after each reading has been
acquired and completely processed from either channel. When acquiring on
one channel, it indicates that the
width of this pulse
In TRIG
Each pulse width is at least
pulses is
In TRIG ON LEVEL mode, each measurement is compared to a programmable
trigger level. The units
detectors.
output
trigger output becomes active. The TRIG ON LEVEL output is a shift in level
rather than a pulse.
If
between subsequent triggers is less than
initiated twice, then the separation between
RUN command initialization, giving a time
tween channel triggers (depending on sample frequency chosen).
The polarity of the trigger output is programmable
low.
active low then the line will idle high.
AT
FREQ mode, a pulse is output at a user defined frequency or rate.
0.001
If
is
inactive.
both channels are sampling when this function is initiated, the time lag
If
the polarity is active high then the output will idle low.
is
at least
Hz to
1000.0
of
the trigger level always equal Watts for power
a measurement is less than the specified value, then the trigger
If
the measurement exceeds the specified value then the
The Model 2832C's rear panel TTL external trigger input can be enabled or
disabled, Section 2.3.17, and have its edge polarity set. The edge polarity is
accessed via the EXT TRIGGER
falling edge triggering.
The external trigger affects both channels and, like the
acquisition trigger when in SNGL modes and acts like a toggled acquisition gate
when in CONT modes.
The external trigger pulse width must be 2200 ns. In DC SNGL mode, the
delay from the trigger input going active to the start of the
<80 ~s.
LEVEL
Trigger
Edit LVL=
Input
77UG
77UGATFREQ
77UG
77UG
ACTNE
Table
14
ON
GWLT
TRIGOUT, TRIGOUT?
ONLEVEL
0UTom
M,
LO
dddddd
ddd&d
IN
menu function and can be set to rising or
-
EXT
TRIGGER
TRIGOUTPOL, TRIGOUTPOL?
Hz
TRIGOLITFREQ, TRIGOUTFREQ?
W TFUGOUT'LVL, TRIGOUT'LVL?
R/S
key, acts like an
AID conversion
IN
Operations.
is
Menu Function
EXT TRIGGER
2.4.10
A
high resolution analog bar graph is provided for visual feedback when
performing optical alignment tasks. This bar graph uses the message display
area and functions differently depending on the measurement units. When the
measurement units are W, W/cm2, or
display area. When using dBm units, the bar graph
(0 dBm) level, moving towards the left or right with decreasing or increasing
display values, respectively.
The
zoom the bar graph into
subsequent
bar graph may go off-scale when changing from one measurement unit to
another. Pressing either
proper scale.
The bar graph
When the meter is put into the remote mode with the bar graph enabled,
the bar graph is automatically replaced by the second channel reading.
However when the meter
be displayed again.
Bar
4
,
b
keys can be used to zoom the bar graph in and out. Press 4 to
4
Keypad Commands Associated Remote Commands
IN
Edit
RISING,
FALWNG
EDGE EXT?
EXTEDCE, EXTEDCE?, EXT,
Graph
A,
the bar graph
a
lox magnification which auto-centers with each
press. Press b to zoom out to normal lx magnification. The
4,
b
key will bring the bar graph back into the
is
not displayed when the meter is in the remote mode.
Turn on the Model 2832C by depressing the front panel key until it clicks in
and remains in its depressed position. The Model 28324 will perform a power
up self test and then configure itself to its last operating state.
operating state is not compatible with one of the current detectors, the 2832C
adopts the default configuration appropriate to this new detector.
does not pass its self test or fails to respond to front panel key commands,
refer to Section
6,
Maintenance, Test and Troubleshooting.
If
the last
If
the meter
Performing
Basic
Measurements
Basic measurement techniques for using the Model 2832C are covered in the
following sections. Refer to Table 2 and Table 6 (pages
for a review of the Model 2832C's functions and capabilities. The following
instructions assume familiarity with the meters functions. They also include
steps to incorporate background correction and assume that the experimental
setup underfills and does not saturate or damage the detector.
2.8.1
The following process describes the procedure for making basic optical power
measurements while properly removing the influence of ambient light and
other drift effects.
i. Plug in a
module and then turn the meter on. Set MODE to DC CONT, set AUTO on
and set the measurement wavelength to the desired value.
ii.
Cover or otherwise block the source that you will be measuring and then
turn ZERO on.
iii.
Uncover the source so it illuminates the detector and note the display value.
This reading is the optical power observed by the detector due to the source.
6
and
14
Making
DC
Power Measurements
Newport Low-Power detector via its associated calibration
respectively)
NOTE
The process
between when you zero the display and when you make your measurement.
Remember,
detector can see you as a changing ambient DC signal!
Detector data
module specific to the detector in use. At power up (and
uploads information about the detector from the calibration module which
describes the set of operating states available to the detector. A user then
selects among the available operating states when using the meter. Front
panel control and the operating states of the Model
Sections
The analog signal flow
ments are made, and
setting. A block diagram of the analog signal flow
is
introduced to the Model
2.3
and
2.4.
is
independent of whether SNGL or
is
primarily determined by the signal range and the filter
Signals that are captured by one of the analog-todigital converters are further
9.
processed as illustrated in the signal flow block diagram of Figure
signals are acquired by the analog-todigital converters and have the units of
current. These digitized signals move through a number of process steps
which may or may not alter the digitized value depending upon the operating
state of the meter. Each of these possible processing steps
their order of occurrence.
is
discussed in
Raw
Calib.
Module
Figure
9
DIG
-
-
Finer Mwb)
-
Gain
+
ZERO
Responsivily
t
4
---
I I
Front Panel
Figure
9 - Model
Digital
If
average of the most recent 10 digitized values. When
been acquired since the last reset of the digital filter, the output
of all the values received. The digital filter
turned on and whenever the UNITS or the range changes or when the MODE
key is pressed.
Filter
the digital filter annunciator DIG
2832C
Digitized Signal Flow
is
on. Section
2.3.4,
is
reset when the Model
Block
Diagram
the filter output
less
than 10 values have
NOTE
When using the digital filter in SNGL acquisition mode, each measurement is
the average of the last 10 acquisitions independent of how old any of the
measurements are.
1
r
is
the average
D
is
2832€
the
is
Gain
Gain processing accounts for the signal gain of the input amplifiers. The
output is the product of the digitized value and the amplifier gain.
Zero
Offset
Zero offset is active whenever the ZERO annunciator is lit, Section
less
zero offset output is equal to the input value
SS,.
This process scales the input value in accordance with current calibration
wavelength and the responsivity map downloaded from the detector calibration
is
module. The output of this process, i.e. the measurement value,
input value divided by the responsivity associated with the current calibration
wavelength or the user defined calibration value. Different responsivities are
used depending upon
if
the attenuator, AITN, annunciator
is
the digitized
lit, Section 2.4.5.
Units Correction
Units correction adjusts a measurement value to account for the display units
selected. When the display units are equal to the detector signal units, i.e. equal
to the numerator of the responsivity units, the measurement value is not
adjusted. Otherwise the digitized value
responsivity and/or additional unit conversions such
is
adjusted to account for detector
as
W to W/cm2.
NOTE
Per area unit conversions such
by the active area of the detector. This calculation assumes that the entire
active area of the detector
where the entire detector active area
user must insure that these conditions are met before utilizing per area units.
as
W to W/cm2 divide the measurement value
is
uniformly illuminated. Per area measurements
is
not uniformly lit are not accurate. The
Analog
Output
is
The flexibility of the Model 28324 analog signal flow
it may properly measure the signals that various types
Basic optical power measurements are related to a measured detector signal,
S,
in the following way:
Where:
Newport Low-Power semiconductor detectors provide a current signal. The
2832C
ity performance with these detectors.
The Model 2832-C provides a 0 to 2.5 volt, into 50R,BNC analog output for
signal monitoring. The analog output is the actual amplified detector signal
and is uncorrected for the effects of the detector's responsivity, AUTO CAL,
and ZERO.
R,=
Detector responsivity at
S
=
Detector signal
is
capable of lOOfA resolution in order to provide the highest sensitiv-
This section describes detector characteristics, optical and electrical considerations, and environmental influences on optical measurements. In general,
measurement accuracy is limited by the accuracy of the detector calibration.
Accurate measurements, however, are also dependent upon proper set-up,
control of temperature and illumination conditions and understanding the
factors that affect optical measurements.
3.6.1
Newport Corporation calibrates its detectors using secondary standards
directly traceable to the United States National Institute of Science and
Technology (NIST) or to Great Britain's National Physical Laboratory (NPL).
The details and accuracy of the calibration procedure vary with each detector
model but a detailed description of the calibration results is supplied with
each individually calibrated detector. In general, detector calibration accu-
racy varies from
detector will also have some variation in response over its surface. Therefore,
for the most reproducible measurements, light should illuminate the detector
as uniformly as possible over as large an area as practical.
Detector Calibration and Accuracy
2%
to
5%
in absolute terms and varies with wavelength. Each
CAUTION
Avoid focusing a light source onto the detector surface. Inaccurate
readings and possible detector damage
manual
NIST traceability requires that detectors be recalibrated on one year intervals.
As
individual detector responses change with time, especially in the ultravie
let, recalibration is necessary to assure confidence in the accuracy of the
measurement. For the most reproducible measurements, the same detector
should always be used for measurements which are to be directly compared.
for saturation or damage thresholds.
may
result. Consult the detector
3.6.2
Newport Low-Power semiconductor detectors, are affected by temperature.
At long wavelengths, quantum detectors typically lose sensitivity with increas-
ing temperature. Additionally, detector dark current increases exponentially
with temperature.
Observed dark current
detector and a meter's amplifier and is typically larger than the theoretical
dark current limit. Silicon detectors are inherently quieter than germanium
detectors due to their higher internal resistance and lower capacitance. The
noise or drift in the dark current sets a lower bound on the measurement
resolution which can be achieved with any given detector. Cooling a detector
significantly lowers its dark current and dark current noise.
The observed dark currents can also be zeroed at any moment via the ZERO
function. Since dark currents drift with temperature, the ZERO should be
adjusted just prior to taking any measurements.
constant, sensitivity changes and dark current drifts are significantly reduced.
Ambient and stray light striking the detector should be considered when
making a measurement. Ambient light can be distinguished from dark current
(or the detector/meter noise floor) by either turning
and covering the detector face with opaque material such
rubber.
Using the human hand to cover the detector is not advised because it emits a
significant amount
cantly different from ambient. With the detector covered, a reading of the
dark current may be made. Next, remove the material which
detector and take another reading. The difference is the ambient light level.
and
Stray
Light
off
or blocking the source
as
a piece
of
infrared radiation and radiates a temperature signifi-
is
covering the
of
black
NOTE
Changes in ambient light levels can occur from such factors
off,
lights on or
your detector element, then your detector can
or by moving people or equipment. Remember, if you can
see
the light bouncing
as
turning room
off
see
you.
The effects
signal input to the detector.
using an attenuator will often improve the signal to ambient signal noise level.
Wavelength-specific filters, such as optical cutoff, bandpass, or spike filters
can be used
reduce stray light include using apertures, placing the detector in a box or
other housing to shield the surface from light (or air currents) and turning
room and other polluting light sources.
3.6.4 Common Measurement
The most common sources of optical measurement error are listed in Table
below. Other common errors are discussed in the preceding subsections
Section
Type of Error
Radiometry
Ambient Light
Wavelength Calibration
of
ambient light are greatly reduced when using a fiberconnectorized
If
free-space beam measurements are desired,
if
the signal wavelength spectrum permits. Other techniques to
Errors
3.10,
Measurement Considerations.
Table
19
-
Common Measurement
What should be done?
Check that all
the detector.
Check that any ambient light was ZEROed
before the measurement was made.
Check that the Calibration Factor for the
measurement wavelength is properly set.
of
Errors
the light
is
actually hitting
off
19
of
Detector Saturation Check that the optical power density remains
or Damage
threshold.
below the detector's saturation or damage
NOTE
The Model
saturation or damage threshold by displaying "SA in the measurement
display area. Be aware that detectors often experience local saturation or
damage without ever exceeding an overall saturation or damage threshold.
This only applies to semiconductor photodiodes used above maximum
intermittent power.
will indicate when a detector is being operated above its
is
"SAW
29
not displayed. Be sure to
Page 45
Section
4
Computer Interfacing
General Guidelines
The
port conforms with the IEEE Standard
488.2
as possible to the IEEE
As
specified by the IEEE
independent commands and device dependent commands. The device
independent commands have the same meaning to all devices and
as part of the lEEE
with an asterisk
asterisk and have meanings unique to the Model
mand that invokes a response from the meter. All queries are terminated by a
question mark
response to that query be read before other commands are issued. When a
query
checked by means of a serial poll to make sure that the data is available
before reading the response. (See Appendix
2832C
has two computer interface ports: GPlB and
488.1
software standard. The protocol for the
488.2
software standard.
488.2
standard, there are two command types; device
488.2
standard. All device independent commands start
(').
Device dependent commands do not start with an
(?).
It
is
recommended that when a query is made, that the
is
made to the GPlB interface, the
MAV
RS232C.
hardware standard and the lEEE
RS232C
bit in the status byte should be
C.4
Status Byte)
port conforms
28324.
A query is a com-
The GPlB
as
much
are
defined
Computer Interface Terminology
Listed below are key abbreviations and concepts used in the command
reference section of this manual.
<...>
Delimiting Punctuation
For the purposes of this manual, any string enclosed by
be a command, a string or a numerical argument. The punctuation
used to symbolize the typographic limits of the command, string or argument
in question.
<EOI>
An
IEEE
clST>
This status is generated by the status byte and parallel poll enable register. It
is used in responding to parallel polls.
End
or
488.1
signal sometimes sent with the endaf-string character.
the 2832-C automatically goes into remote interface mode. When the meter is
in the remote mode, the right-most character of the secondary channel
display area will display an asterisk,
is
received by the GPIB or RS232 interface ports,
>[
.
The local-lockout mode is designated
-
by displaying
While in remote mode, the keypad
<(LOCAL)> key. To get out of remote mode press the <(LOCAL)> key.
2832-C is in local lockout then the LLO remote command must be used to
disable local lockout before you will be able to return to keypad control.
The message area
l>i(l
at the right side
-
of
the display will indicate when'the keypad is in local
of
the secondary channel display area.
is
disabled except for the power and
If
the
-
lockout state by displaying as the right-most character.
Before communicating with the 2832-C through the RS232 port, proper cable
connections must be made. Figure 10 shows the cable connections for
communicating with the RS232C port on the 28324.
Once cable connections are made, the baud rate and echo mode need to be
set. Valid baud rates are
bits, and stop bits are fixed at no parity,
-
19.2K, 9600,4800,2400, and 1200. The parity, data
8
data bits, and 1 stop bit.
RS232C
Baud Rate
Parity none
Data bits
Stop bits 1
When the echo mode
new line and all characters sent to the 2832-C are echoed back over the
interface. Error messages are reported over the interface immediately.
user
(sending an ASCII decimal 08 code) or by using the DEL key (sending
decimal 127 code).
In echo mode the
dumb terminal type
When echo mode
prompt or echo characters back over the interface. Error messages must be
accessed by using the
Parameters
19.2K, 9600,4800,2400, or 1200
8
is
enabled the 283242 generates a
is
entering commands the line may be edited by using the backspace key
RS232C port
of
device
is
disabled (normal mode) the 2832-C does not generate
DATA TERM. READY
SIGNAL GROUND
DATA SET READY
REQUEST TO SEND
CLEAR TO SEND
RlNG IND.
I
PIN
PIN NO.
CABLE TERMINATORS (RS-232)
9
pin
to
9
pin
9
PIN
PIN NO.
1
2
3
4
5
6
7
8
9
Figure
4.4.1
--+
10
-
RS232
Setting
PIN NO.
9
;ODE
DCD
RXD
TXD
DTR
GND
DSR
RTS
CTS
R I
DESCRIPTION
CARRIER DETECT
RECEIVE DATA
TRANSMIT DATA
DATATERM. READY
SIGNAL GROUND
DATA SET READY
REQUEST TO SEND
CLEAR TO SEND
RING IND.
Cable Connections
Baud Rate and Echo Mode From The Keypad
The baud rate and echo mode are set from within the menu structure. To set
the baud rate select the BAUD RATE option from the REMOTE SETUP menu.
To set the echo mode select the ECHO MODE-n option from the REMOTE
SETUP menu. (See Section 2.4.1 1)
4.4.2
Setting
Baud Rate and Echo Mode From A Remote Interface
The baud rate cannot be set from the remote interface. To set the echo mode
use the ECHO command. Use the ECHO? to see
The 2832-C uses the XON/XOFF handshaking protocol. When the 2832-C'~
232 input buffer is nearly full the 2832-C will send an XOFF (ASCII encoded 19
decimal) character to the remote computer. The XOFF character informs the
remote computer that it needs to suspend transmission until it receives the
XON character from the
will send an XON (ASCII encoded 17 decimal) to signal the remote computer
that it can resume transmission.
The 2832-C will also respond to XOFF and XON characters sent to it from the
remote computer. When the 2832-C receives an XOFF character it will sus-
pend all transmission of characters to the remote computer until the XON
character is received.
A variety of third party GPIB communication hardware and software, such
plug-in GPlB computer boards and Lab View software from National Instruments exists. The 2832-C should work with any of these as long as they
adhere to the
with one of these third party hardware/software packages. The table below
summarizes the lEEE 488.1 capabilities supported by the 2832X.
Description Command Level of Capability
Source Handshake
Acceptor Handshake
Talker
Extended Talker
Listener
Extended Listener
Service Request
Remote Local
Parallel Poll
Device Clear
Device Trigger
Controller
lEEE 488.1 standard. This manual assumes the user is familiar
Table
2832-C. When the buffer empties sufficiently the 28324
20
-
Model
2832C
SH
AH
T6
TEO
L4
LEO
SR
RLO
PP 1
DC
DTO
co
1
1
1
1
IEEE
488.1
Capabilities Summary
complete capability
complete capability
basic talker, serial poll, unaddress if MLA
no capability
basic listener, unaddress if MTA
no capability
complete capability
*no capability
remote configuration
complete capability
no capability
no capability
RS
as
NOTE
In order to better handle the RS232C interface the 2832-C automatically
goes into remote mode when a command is received from either of the
is
remote computer interfaces. The
the local lockout state.
4.6.1
Before communicating with the 2832-C GPlB port the 2832-C GPIB address
must be set. The address can be between 0 and 30. The address can only be
set from the keypad and is accessed through the menu structure. To set the
GPIB address select the REMOTE
Attenuator Calibration Data Enable
Attenuator Calibration Data Enable Query
Attenuator Serial Number Query
Auto Ranging Enable
Auto Ranging Enable Query
Bar Graph Enable
Bar Graph Enable Query
Calibration Date Query
Available Channel Query
Detector Mode Query
Detector Serial Number Query
Display Brightness Select
Display Brightness Query
Display Channel Select
Display Channel Query
Data Store Value Query
Data Store Buffer Behavior Select
Data Store Buffer Behavior Query
Data Store Clear
Data Store Value Count Query
Data Store Enable
Data Store Enable Query
Data Store Buffer Size Select
Data Store Buffer Size Query
Data Store Units Query
Manual Range Level Query
Reference Source Select
Reference Source Query
Responsivity Query
Start Both Channel Acquisition
Start Single Channel Acquisition
Read Both Channel Measurements with Status Query
Read Single Measurement with Status Query
DC Sample Frequency Select
DC Sample Frequency Query
DC Sample Type Select
DC Sample Type Query
Statistics Buffer Maximum Value Query
Statistics Buffer Mean Value Query
Statistics Buffer Minimum Value Query
Statistics Buffer Maximum-Minimum Value Query
Statistics Buffer Standard Deviation Query
Statistics Buffer Size Select
Statistics Buffer Size Query
Stop Both Channel Acquisition
Stop Single Channel Acquisition
Store Reference Reading
Store Reference Query Reading
Store Zero Signal
Trigger Out Mode Select
Trigger Out Mode Query
Trigger Out Frequency Select
Trigger Out Frequency Query
Trigger Out Level Select
Trigger Out Level Query
Trigger Out Polarity Select
Trigger Out Polarity Query
Units Select
Units Query
User Calibration Enable
User Calibration Enable Query
User Defined Reference Value Select
User Defined Reference Value Query
User Responsivity Factor Select
User Responsivity Factor Query
Signal Zeroing Enable
Signal Zeroing Enable Query
Zero Signal Query
This routine performs an auto calibration sequence on both channels. The
calibration sequence measures the amplifier
measurements are stored and subtracted from subsequent measurements.
Acquisition is suspended during auto calibration and no other remote commands
be
processed until the auto calibration sequence is complete.
will
Returns: <number>
<number> represents the integer
is
autocalibration
Related Commands: None
complete.
0.
offsets
This return value is generated when
at all gain settings. These
'CIS
Clear Status
Syntax:
*CIS
Parameters: None
Function:
This command clears the Standard Event Status register, device event register, and
the error queue.
the message available bit @it
and output buffers the MAV bit will not
cancels any outstanding *OPC and OPC? commands.
As
a result
of
this command the status byte
4).
Because this command does not clear the input
be
affected. The *CLS command also
is
cleared except for
Related Commands:
*ESR?, EVENT?, *IST?,
*
STB?
'ERR?
Error Query
Syntax:
Parameters:
Function:
Returns (and removes) oldest error message from the error queue. The error
message
the Appendix B). A maximum of
This query returns the contents of the Event Status Enable register. See the
*ESE command for a description of the Event Status Enable register. See
Appendix C for a detailed description of the Event Status Register.
Returns:
<mask>
to
255.
Related Commands:
<mask>
is
of type <number> and represents an unsigned integer in the range
'CLS, 'ESE, 'ESR?, EVENT?, EVENTEN, EVENTEN?,
'IST?, 'PRE, 'PRE?, 'SRE, 'SRE?, 'STB?
*BR?
Event Status Register Query
Syntax:
Parameters:
*ESR?
None
Function:
This query returns the contents of the Event Status Register and clears the
Event Status Register except for bit
Device Event and Device Event Enable registers. The 'CLS command will also
set this register to
0.
3.
Bit
3
is
determined by the state of the
0
The Event Status register
any bit
is
set in the result of this AND'ing operation the ESB bit @it
is
AND'ed with the Event Status Enable register.
5)
If
of the
Status Byte will be set. The Status Byte can be used in conjunction with the
Service Request Enable register to generate a service request. See Appendix C
for a detailed explanation of the Event Status Register.
is
The Event Status Register
an 8 bit, bit mapped register, with each bit signify-
ing a different condition. The bits are listed below, most significant bit first:
Oh
Operatloncomplete
Request Control (Not Used)
Query Enor
Devlce Enor
Execution Error
Command Error
User Request (Not Used)
Power
On
(Not Used)
Standard Event Status Register
Returns:
<value> is of type <number> and represents an unsigned integer in the range
This command sets the Parallel Poll Enable register bits. The Parallel Poll
Enable register
(Individual Status) message. See Appendix
Event Status Register.
The Parallel Poll Enable Register
Enable Register is an Sbit, bit mapped register, with each bit signifying
different condition. The bits and their significance are detailed below:
is
used in conjunction with the Status Byte to generate the IST
C
for a detailed description of the
is
set to 0 at power-on. The Parallel Poll
a
7654321
1
/ /
I
(
Parallel Poll Enable Register
1
O-
New Data Available
:4eyU:;
Data Available
Not Used
Message Available Mask
Event Status Byte Mask
Master Summary Status Mask
Error Queue Mark
Related Commands: *]ST, *PRE?, *STB?*PRE?
*
PRE?
Parallel Poll Enable Query
Syntax:
Parameters:
*PRE?
None
Function:
This query returns the contents of the Parallel Poll Enable register. See the
*PRE command for a description of the Parallel Poll Enable register. See
C
Appendix
Returns: <mask>
<mask> is
to
255.
Related Commands: *]ST, *PRE, *STB?
for a detailed description of the Event Status Register.
of
type <number> and represents an unsigned integer in the range
<config> is of type <number> which rounds to an integer in the range 0 to
This number specifies a configuration storage buffer. The number 0 stands for
the default configuration buffer.
*RCL
<config>
9.
Function:
This command configures the meter to a configuration previously stored in
non-volatile memory (see *SAV), or to the default configuration for the
if
attached detector. This command will fail
incompatible with the detector attached to the meter or empty. The parameters affected by this command are those listed in Table
Returns: None
Related Commands: *RST, *SAV
the requested configuration
8.
is
'ST
Reset
Syntax:
Parametem
Function:
This sets the meter to the default configuration,
attached to it. It also cancels any outstanding *OPC or *OPC? commands.
*RST
None
see
Table
8,
for the detector
Related Commands:
'
SAV
*RCL, *SAV
Save Meter Configuration
*SAV
Syntax:
<config>
Parametels:
<config>
This number specifies a configuration storage buffer.
is
of type <number> which rounds to an integer in the range 1 to
Function:
This command saves the configuration of the meter to the configuration
specified by the parameter. The parameters saved by this command are
those listed in Table
Related Commands: *RCL, *RST
is written into the Service Request Enable register.
Function:
The Service Request Enabled Register
is
used in conjunction with the Status
Byte to generate service requests on the CPlB bus. See Appendix
C
for a
detailed description of the Event Status Register.
If
a bit is set in the Service Request Enable register and it's corresponding bit
is set in the Status Byte Register, then a service request will be generated once
for the for the given event. When service
is
being requested, bit
6
is
set in the
byte returned by a serial poll of the
2832-C.
The Service Request Enable Register
is
an gbit, bit mapped register, with each
bit signifying a different condition. The bits are listed below, most significant
bit first:
'
*
New Data Available CH A
New Valid Data
Available
CH A
New Data Available CH
B
New Valid Data Available CH
B
Message Available Mask
Event Status Byte Mask
Always Zero
Error Queue Mask
Service Request Enable Register
Related Commands:
*SRE?, *STB?
*SRE?
Sewice Request Enable Query
Syntax:
*SRE?
Parameters:
None
Function:
This query returns the contents of the Service Request Enable register.
See
the *SRE command for a description of the Service Request Enabled register.
See Appendix
C
for a detailed description of the Event Status Register.
This query returns the contents of the Status Byte which records current
system conditions. Appendix C describes the Status Byte in detail.
If
a bit is set in the Status Byte Register and it's corresponding bit is set in the
Service Request Enable Register, then a service request will be generated once
for the given event. When service
returned by a serial poll of the
The Status Byte Register
is
an
ing a different condition. When a bit is set, or has a value of one, then the
is
condition
true. The bits are listed below, most significant bit first:
'
is
being requested, bit
6
is
set in the byte
28324.
&bit, bit mapped register, with each bit signify-
'
OL
New Data Available CH A
New Valid Data Available CH A
New Data Available CH B
New Valid Data Available CH B
Message Available Brl
Event Status Byie Bit
Master Status
Error
Sumrnaly Bii
Oueue Bit
Status Byte Register.
Note that the *CLS common command clears most of the bits in the Status
Byte.
Returns:
<value>
Related Commands:
*
TST?
Self
Syntax:
Parameters:
<value>
is
an integer in the range 0 to
*CLS, *SRE, *SRE?
Test Query
*TSV
None
255
inclusive.
Function:
As
defined in the
to return a result of a self-test, indicating whether or not the unit completed
the self-test without any errors. The self test
or power up. The query does not initiate the self test and only returns a
indicating that the system passed its earlier
The *WAI command causes the device to wait until all pending operations are
complete before processing any commands waiting in the input queue. Care
should be taken when using this command since it
input queue by continuing to write commands to the
command
Note that the *WAI command will essentially "lock out" the remote interfaces
until all pending operations are complete. The *OPC command and *OPC?
query differ from *WAI in this area, as they allow other commands to be
processed while they are waiting for pending operations to complete.
Once the
following operations are in progress:
Single acquisition in progress
Data store in progress with fixed data store buffer
The meter
The *WAI command can only be canceled by a GPlB interface device clear or
The parameter <state> is a <number> which is either 0 or
data storing
is
disabled.
If
the number
is
1,
data storing is enabled.
1.
If
the number is
Function:
0,
This command enables or disables data storing. Note that data storing cannot
is
be enabled when the data store buffer
enabled. The user must first either clear the data store buffer or enable SLIDE
BUFFER behavior and then enable data storing.
Related Commands:
DSSIZE-n?, DSUNITS-n?, DSBUF-n, DSBUF-n? DS-n
DS-n?, DSCLR-n, DSCNT-n?, DSE-n?, DSSIZE-n,
full and FIXED BUFFER behavior is
DSE-n?
Data Store Enable Query
Syntax:
where n is either
Parameters:
DSE-n?
None
A
or
B
Function:
This query returns a value showing whether or not data storing is enabled or
an integer in the range 0 to 3 inclusive. 0 corre-
sponds to no filtering,
1
to analog filter only, 2 to digital averaging filter only,
and
3
to analog and digital filters combined.
Function:
This command selects a filter combination to be used to condition readings
taken from the detector. The choices are to have no filter at all, an analog
filter, digital averaging filter, and both the analog and digital filters. The digital
filter processes every signal reading by averaging it with the last
9
readings
taken.
All detectors signals are filtered to some extent even without the analog filter
feature. By enabling the analog filter the signal
is
filtered by a
5
Hz
low pass
instead of the standard filter.
If
an attempt
is
made to enable the analog filter for a detector that does not
support
it,
an execution error will occur.
Related Commands: FILTER-n?
FILTER-n?
Fdter Query
Syntax:
FILTER-n?
where
n
is
either A or
B
Parameters:
None
Function:
This query returns a value that represents the filter combination currently
being used to condition readings.
Returns:
<filter>
<filter>
is
of type <number> in a range of 0 to 3 with 0 corresponding to no
filter,
1
to analog filter only, 2 to digital averaging filter only, and 3 to analog
and digital filters combined.
Related Commands: FILTER-n
Read Both Channels Measurement Query
Syntax:
Parametem:
R?
None
Function:
This query returns the last valid measurements taken from both channels.
The units of the measurements are the units defined at the time the readings
were made. Because it
ments are actually being taken, the values returned by successive queries may
be multiple reports of a single measurements.
To ensure that fresh data
the Status Byte can be checked either by issuing a serial poll or by using the
'STB? command. A serial poll
The NVDA bit will be
NVDA bit for a specific channel will be reset to
is
queried from that channel and will stay 0 until a new valid reading is taken. A
valid reading
Returns:
or <number>
<number>
If
two channels are running simultaneously then, both are returned separated
by a comma.
is
a reading that would return a status of 0 using the RWS? query.
<number>, <number>
is
a floating point number in the exponential format.
is
possible to make this query faster than measure-
is
being read the NVDA(new valid data available) bit
is
recommended
1
if
the last valid measurement has not been read. The
if
the CPlB interface is being used.
0
immediately after a measurement
in
Related Commands:
UNITS-n, UNITS-n?
'STB?, MODE-n, MODE-n?, R-n?, RWS?, RWS-n?,
R-n?
Read Single Channel Measurement Query
Syntax:
where
Parameters:
R-n?
n
is either
None
A
or
B
Function:
This query returns the last valid measurements taken from the specified channel.
The units of the measurements are the units defined at the time the reading was
made. Because it
actually being taken, the values returned by successive queries may
reports of a single measurements. To ensure that fresh data
NVDA(new valid data available) bit in the Status Byte can be checked either by
issuing a serial poll or by using the
if
mended
valid measurement has not been read. The NVDA bit for the specified channel
will be reset to 0 immediately after a measurement
and will stay
that would return a status of
the GPlB interface
is
possible to make this query faster than measurements are
depend on the detector and mode being used to acquire data.
entitled "Electrical Specifications" at the beginning of this manual for the
signal range that corresponds to the current setting of
is
of type <number>. The range of this parameter will
See
the section
<range>.
Function:
This command
will disable the auto ranging feature.
Related Commands:
is
used to select a new manual signal range. This command
AUTO-n, AUTO-n?, RANGE-n?
RANGE-n?
Signal Range Query
Syntax:
where
Parameters:
RANGE-n?
n
is
either A or
None
B
Function:
This query returns a value that represents the current signal range setting
is
being used whether the meter
Section
2.3.1
1.
auto ranging or in manual ranging mode.
See
Returns:
<range>
used for either auto or manual ranging. See the section entitled "Electrical
Specifications" at the beginning of this manual for the signal range that
corresponds to
Related Commands:
<range>
is
of type <number> and represents the gain level currently being
<range>.
AUTO-n, AUTO-n?, RANGE-n
-EL-n
Reference Source Select
Syntax:
REFSEL-n <sources
Parameters:
<source> is of type <string> in the double quote format. The possible values
-
of <source> are: "STOREF'
The other channel, "USRREF' - User defined reference value
Last stored reference measurement, "OTHERCH"
Function:
This command
logarithmic (dB) relative measurements. This command will cause an error
an
attempt
available.
Returns:
is
used to select a reference source to be used in linear and
This query returns the current reference source that will be used in linear
and
logarithmic (dB) relative measurements. Returns:
<source>
is
of type <string> in the double quote format. The possible values
of <source> are:
"STOREF"
-
Last stored reference measurement
"OTHERCH"
-
The other channel
"USRREF"
-
User defined reference value
Related Commands: REFSEL-n, STOREF-n, USRREF-n, USRREFn?
RESP-n?
Responsivity
Query
Syntax: RESP-n?
where n is either
A
or
B
Parameters: None
Function:
This query returns the responsivity value currently being used by the
2832-C
in measurement calculation.
Returns: <val>
<val>
is
of type <number> and represents a floating point number. The units
of the responsivity <val> depends upon the type of detector family of the
detector in use. See Table
22
for a listing of responsivity units by detector
family.
Related Commands:
ATN-n, ATN-ri?, LAMBDA-n, LAMBDA-n?, USRCALn,
USRCAL-n?, USRRESP-n, USRRESP-n?
RUN
Start Both Channel Acquisition
Syntax: RUN
Parameters:
None
Function:
This command initiates data acquisition on both channels using the currently
defined acquisition mode. In single type acquisition modes this command will
cause one reading to be acquired. In continuous type acquisition modes this
command will initiate the continuous acquisition of readings. The STOP
command will terminate acquisition.
This command initiates data acquisition on the specified channel using the
currently defined acquisition mode. In single type acquisition modes this
command will cause one reading to be acquired. In continuous type acquisi-
of
tion modes this command will initiate the continuous acquisition
The STOP command will terminate acquisition.
Related Commands: STOP, STOP-n, RUN
readings.
RWS?
Read
Both
Channel Measurements With Status Query
Syntax:
Parameters:
RWS?
None
Function:
This query returns the last measurements taken from both channels. The
of
units
were made. Because it
ments are actually being taken, the values returned by successive queries may
be multiple reports
these measurements are the units defined at the time the readings
is
possible to make this query faster than measure-
of
the same measurement.
To ensure that fresh data is being read the NDA(new data available) bit in the
Status Byte can be checked either by issuing a serial poll or by using the
is
*STB? command. A serial poll
used. The NDA bit will be
NDA bit for a specific channel will be reset to
ment
is
queried from that channel and will stay 0 until a new reading is taken.
Returns: <status>, <number>
<status> is an integer that represents the status of the returned value.
<status> will be
and
4
for ranging.
If
more than one channel
status
of
CHA and CHB respectively.
<number>
Related Commands: *STB?, MODE-n, MODE-n?, R?, R-n?, RWS-n?
0
for ok, 1 for over range, 2 for saturated and 3 for data error,
is
is
a floating point number in the exponential format.
recommended
1
if
the last measurement has not been read. The
available, then the first two numbers represent the
This query returns the last measurement taken from the specified channel.
The units of this measurement are the units defined at the time the reading
was made. Because it
ments are actually being taken, the values returned by successive queries may
be multiple reports of the same measurement.
To ensure that fresh data
Status Byte can be checked either by issuing a serial poll or by using the
*STB? command. A serial poll
used. The NDA bit will be 1
NDA bit for the specified channel will be reset to 0 immediately after a mea-
is
surement
taken.
Returns:
<status>
<status> will be 0 for ok, 1 for over range, 2 for saturated and
and
4
<number>
queried from that channel and will stay 0 until a new reading is
<status>,<number>
is
an integer that represents the status of the returned value.
for ranging.
is
a floating point number in the exponential format.
Related Commands:
UNITS-n?
is
possible to make this query faster than measure
is
being read the NDA(new data available) bit in the
is
recommended
if
the last measurement has not been read. The
*STB?, MODE-n, MODE-n?, R?, R-n?, RWS? UNITS-n,
if
the CPlB interface is being
3
for data error,
SFREQ
DC
Sample Frequency Select
Syntax:
SFREQ
<freq>
Parameters:
The parameter <freq>
on the current DC sample precision
is
the range
500Hz when two detector calibration modules are plugged into the meter and
1000Hz for one.
Function:
This command determines how often a signal
ous mode. Both channels sample data at the same rate when in DC continuous mode.
0.001 to 25Hz.
Related Commands:
is
of type <number>. The range of this number depends
This query returns the sample rate used for continuous DC acquisition on
both channels
Returns: <freq>
<freq> is of type <number> in exponent notation with units of hertz or samples
per second.
Related Commands: SFREQ, SPREC, SPREC?
SPREC
DC
Sample Precision Select
Syntax:
SPEC
<prec>
Pararne ters:
The parameter <prec>
4096.
is
of type <number> with acceptable values of 20000 or
Function:
The 2832-C has a 20,000 count high precision A/D and a 4,096 count A/D. This
command determines which of these A/Ds will be used from both channels to
acquire samples during DC single and DC continuous acquisition modes. This
command affects the available range of DC sample frequencies.
Note that the 4,096 count A/D
Related Commands: SFREQ, SFREQ?, SPREC?
is
used during all other acquisition modes.
SPREC?
DC Sample Precision Query
Syntax:
Parameters:
SPREC?
None
Function:
This query returns a string that indicates whether the 20,000 count A/D or the
A/D
4,096 count
on both channels.
Returns: <prec>
<prec>
and 4096
Related Commands: SFREQ, SFREQ?, SPREC?
is
if
will be used during DC CONT or DC SNGL acquisition modes
of type <number> and represents an integer in the range 0 to
Related Commands: STMAX-n?, STMEAN-n?, STMIN-n, STMXMN-n?,
,
STSDEV-n?
STSIZE-n
100.
STOP
Stop Both Channel Acquisition
Syntax:
Parameters:
STOP
None
Function:
This command terminates any acquisition of data currently in progress on
both channels.
has no effect.
Related Commands: RUN, RUN-n, STOP-n
If
data
is
not being acquired on a channel then this command
STOP-n
Stop Single Channel Acquisition
Syntax:
where n
Parameters:
STOP-n
is
either
None
A
or
B
Function:
This command terminates any acquisition of data currently in progress on the
If
specified channel.
has no effect.
Related Commands: RUN, RUN-n, STOP
data is not being acquired on that channel the command
STOREF-n
Store Reference Reading
Syntax:
where
Parameters:
STOREF-n
n
is either
None
A
or
B
Function:
This command takes the latest reading and stores it as a reference reading to
be used when making relative linear and dB measurements. The units of the
reference reading defaults to Watts.
The following troubleshooting guide
with the power meter so that, to the greatest extent possible, the return of
the power meterldetector system to Newport will be unnecessary. For the
problems that cannot be resolved with information in this manual, or for other
situations that are not covered in this section, please
on returning your entire system to Newport for service.
Table
23
-
Syrnptorn/Fault Troubleshooting Guide
is
intended to isolate and solve problems
see
Section 7 for details
Symptom
Blank display.
Display shows "OL"
Display shows "SA"
Display shows
Any of the following messages:
"REGISTER ERR
"REGISTER ERR
"INTG AID ERR A"
"INTG AID ERR B"
"SA A/D ERR A"
"SA
A/D ERR
B
"-
A
B"
- - -
-"
Possible Fault/Correction
Power switch OFF. Turn switch ON.
Power cord connection is absent. Check
power cord connection.
Display
lndicates that the signal is too large for
the selected signal range. Select a higher
RANGE or use an attenuator if one is
available.
lndicates that the signal exceeds the
detector saturation or damage threshold.
Select a detector with higher power or
energy handling capability or use
attenuator if one is available.
Indicates that the meter is in the middle
of a range change or that the current units
don't match the units
Call Newport Corporation's Customer
Service Department and arrange to
return the unit for repair.
and "MISSING OR BAD" CAL MODULE, if one is available, to
localize the problem to the first
CAL MODULE.
If
the first CAL MODULE does not work
after proper seating and reset, but the
second CAL MODULE works, call Newport
Customer Service and arrange for a repair
of the first CAL MODULE.
If
neither the first or second CAL MODULE
works, call Newport Customer Service
and arrange for a meter repair.
also return the first CAL MODULE
Be
if
sure to
cannot otherwise determine that it
working.
The display shows:
"CAL MODULE
"CAL MODULE B"
A
Call Newport Customer Service and or
arrange to return the meter for repair.
and "CHECKSUM ERROR
The display shows:
"BATTERY FAILURE"
Call
Newport Customer Service and
arrange to return the meter for battery
replacement.
you
is
RS232
communication does
not seem to work
No listener error when attempting
GPlB communications
Display value does not change.
Reading is different than
expected.
Check the
RS232
cable pinouts, Figure
the device talking to the meter
8
data bits, no parity, and 1 stop bit. Also
cable connection and
10.
Make sure that
is
setup for
check the echo mode state.
Check the GPlB cable connections and
the GPIB address.
This section contains information regarding obtaining factory service for the
Model
maintenance or service of this instrument and/or accessories beyond the
procedures given in Section
problems which cannot be resolved using the guidelines listed in Section
should be referred to Newport Corporation factory service personnel. Contact Newport Corporation or your Newport representative for assistance.
2832-C
Dual-Channel Power Meter. The user should not attempt any
6:
Maintenance, Test and Troubleshooting. Any
6
Obtaining
Service
The Model
is
warranted for a period of 1 year. After 1 year, the unit should be returned to
Newport Corporation for recalibration and NlST traceability recertification.
To obtain information concerning factory service, contact Newport Corporation or your Newport representative. Please have the following information
available:
1.
2.
3.
If
the instrument
Return Number, which you should reference in your shipping documents.
Please
information ready when contacting Newport Corporation. Return the completed service form with the instrument.
2832-C
lnstrument model number (On front panel)
Instrument serial number (On rear panel)
Description of the problem.
fill
out the service form, located on the following page, and have the
contains no user serviceable parts. Its calibration accuracy
is
to be returned to Newport Corporation, you will be given
The IEEE 488.2 standard defines two types of string data. These strings are
either single or double quoted. For convenience, the 2832-C also recognizes
an unquoted string with certain restrictions as detailed below. Any of these
forms may be used where a <string> parameter
1. <string>, using double quotes.
2. <string>, using single quotes.
3. <string>, using no quotes.
A description of each type of <string> follows:
is
required.
"this is a string"
'this is a string'
thisisashing
1. <string> defined using double quotes.
A double quote indicates that a string follows, and the string
nated by another double quote. A double quote may be embedded
within the string by using two double quotes together:
""
Example: "this string contains a
All characters within the two outer double quotes are considered part
is
of the string.
quote. The string cannot contain the <CR>(ascii decimal 13), <LF>(ascii
decimal lo), or End or ldentify characters.
<string> defined using single quotes.
2.
This form
that a string follows, and the string is terminated by another single
quote. A single quote may be embedded within the string by using two
single quotes together:
Example: 'this string contains a
All characters within the two outer single quotes are considered part of
the string.
quote. The string cannot contain the <CR>(ascii decimal 13), <LF>(ascii
decimal lo), or End or ldentify characters.
is
an error
It
similar to double quoted string. A single quote indicates
is
an error if the string does not terminate with a single