MTP MTP-1325 Operating Manual

Page 1
MTP
Operating Manual
Model MTP-1325
Non-Contact Infrared Thermometers
7p
9
12
11p
3p
6p
5p
9p
10p
MTP Instruments
www. .com
information@itm.com1.800.561.8187
Page 2
2
TABLE OF CONTENTS
INTRODUCTION……………………………………….. 3
FEATURES…………………………………………….. 3
WIDERANGEAPPLICATION …………………………... 3
SAFETY………………………………………………... 4
DISTANCE&SPOTSIZE………………………………… 4
SPECIFICATIONS……………………………………… 5
FRONTPANELDESCRIPTION………………………….. 6
INDICATOR…………………………………………….. 6
MEASURMENTOPERATION…………………………… 7
BATTERYREPLACEMENT…………………………….. 8
NOTES………………………………………………… 8
MAINTENANCE ……………………………………….11
www. .com
information@itm.com1.800.561.8187
Page 3
3
INTRODUCTION
Thank you for purchase of the IR Thermometer. This is capable of non-contact (infrared) temperature measurements at the touch of a button. The built-in laser pointer increases target accuracy while the backlight LCD and handy push-buttons combine for convenient, ergonomic operation. The Non-contact Infrared Thermometers can be used to
measure the temperature of objects’ surface that is improper
to be measured by traditional (contact) thermometer (such as moving object, the surface with electricity current or the objects which are uneasy to be touched.) Proper use and care of this meter will provide years of reliable service.
FEATURES:
Rapid detection function  Precise non-contact measurements  Unique flat surface, modern housing design  Automatic Data Hold  °C/°F switch  MAX temperature displays  Backlight LCD display  Automatic selection range and Display Resolution 0.1º
C(0.1ºF)
WIDE RANGE APPLICATION:
Food preparation, Safety and Fire inspectors, Plastic molding, Asphalt, Marine and Screen printing, measure ink and Dryer temperature, HVAC/R, Diesel and Fleet maintenance.
www. .com
information@itm.com1.800.561.8187
Page 4
4
SAFETY
Use extreme caution when the laser beam is turned on.  Do not let the beam enter your eye, another person’s
eye or the eye of an animal.
Be careful no to let the beam on a reflective surface
strike your eye.
Do not allow the laser light beam impinge on any gas
which can explode.
Distance & Spot Size
As the distance (D) from the object increases, the spot size (S) of the area measured by the unit becomes larger. The relationship between distance and spot size for each unit is
listed below. The focal point for each unit is 914mm (36”).
The spot sizes indicate 90% encircled energy.
300
100
75
25
900
1200
D:S= 12:1
www. .com
information@itm.com1.800.561.8187
Page 5
5
1. SPECIFICATIONS Temperature range D: S
-50 to 450 °C (-58°F ~ 842°F) 12:1 Display resolution 0.1 °C (0.1°F) Accuracy for targets: Assumes ambient operating temperature of 23 to 25 °C (73 to 77°F)
-50 ~ 20°C (-58°F ~ 68°F) ±2.5°C (4.5°F) 20°C ~300°C (68°F ~572°F) ±1.0% ±1.0°C (1.8°F) 300°C ~450°C (572°F ~ 842°F) ±1.5%
Repeatability
-50~20°C (-58~68°F) : ±1.3°C (2.3°F) 20~450°C (68~842°F) : ±0.5% or ±0.5°C (0.9°F) Response time 150ms Spectral response 8~14um Emissivity Fixed at 0.95 Over range indication LCD will show “----” Polarity Automatic (no indication for positive polarity);
Minus (-) sign for negative polarity Diode laser output <1mW,Wavelength 630~670nm, Class 2 laser product Operating temp. 0 to 50°C (32 to 122°F) Storage temp. –10 to 60°C (14 to 140°F) Relative humidity 10%~90%RH operating, <80%RH storage Power supply 9V battery, NEDA 1604A or IEC 6LR61, or equivalent Safety “ CE ” Comply with EMC
www. .com
information@itm.com1.800.561.8187
Page 6
6
Note: Field of View: Make sure that the target is larger than
the unit’s spot size. The smaller the target, the closer
you should be to it. When accuracy is critical, make sure the target is at least twice as large as the spot size.
2. FRONT PANEL DESCRIPTION
IR sensor
LCD Display
Laser pointer beam
Measurement Trigger
Battery Cover
Handle Grip
3. INDICATOR
Data hold
Laser ” on” symbols
°C /°F symbol
Low power symbols
Emissivity symbol and value
Temperature values for the MAX
Symbols for MAX
Current temperature value
www. .com
information@itm.com1.800.561.8187
Page 7
7
Functional Design the switches of C/F sat in a cell switching Switching C/F Select the temperature units (°C or °F) using the oC/oF switch ( )
Max indicate the max record that displays between the pressing and releasing the "ON/OFF" button each time
MEASUREMENT OPERATION
Hold the meter by its Handle Grip and point it toward the
surface to be measured.
Pull and hold the Trigger to turn the meter on and begin
testing. The display will light if the battery is good. Replace the battery if the display does not light.
Release the Trigger and the HOLD display icon will
appear on the LCD indicating that the reading is being held.
The meter will automatically power down after
approximately 7 seconds after the trigger is released.(Unless the unit is locked on)
Note: Measurement considerations
Holding the meter by its handle, point the IR Sensor toward the object whose temperature is to be measured. The meter automatically compensates for temperature deviations from ambient temperature. Keep in mind that it will take up to 30 minutes to adjust to wide ambient temperatures are to be measured followed by high
www. .com
information@itm.com1.800.561.8187
Page 8
8
temperature measurements, some time (several minutes) is required after the low (and before the high) temperature measurements are made. This is a result of the cooling process, which must take place for the IR sensor.
4. BATTERY REPLACEMENT
As battery power is
not sufficient, LCD will display “ ” replacement with one new battery type 9V is required.
② Open battery cover, then take
out the battery from instrument and replace with a new 9-Volt battery and place the battery cover back.
6. NOTES:
How it Works
Infrared thermometers measure the surface
temperature of an object. The unit’s optics sense
emitted, reflected, and transmitted energy, which is
collected and focused onto a detector. The unit’s
electronics translate the information into a temperature reading, which is display on the unit. In units with a laser, the laser is used for aiming purposes only.
Open the battery over
9v battery
Low battery indicator
www. .com
information@itm.com1.800.561.8187
Page 9
9
Field of View
Make sure that the target is larger than the unit’s spot
size. The smaller the target, the closer you should be to it. When accuracy is critical, make sure the target is at least twice as large as the spot size.
Distance & Spot Size
As the distance (D) from the object increases, the spot size (S) of the area measured by the unit becomes larger. See: Fig: 1.
Locating a hot Spot
To find a hot spot aim the thermometer outside the area of interest, then scan across with an up and down motion until you locate hot spot.
Reminders
Not recommended for use in measuring shiny or
polished metal surfaces ( stainless steel, aluminum, etc.).See Emissivity
The unit cannot measure through transparent surfaces
such as glass. It will measure the surface temperature of the glass instead.
③ Steam, dust, smoke, etc., Can prevent accurate
measurement by obstructing the unit’s optics.
Emissivity
Emissivity is a term used to describe the energy-emitting characteristics of materials. Most (90% of typical applications) organic materials and painted or oxidized surfaces have an emissivity of
0.95 (pre-set in the unit). Inaccurate readings will result from measuring shiny or polished metal surfaces. To
www. .com
information@itm.com1.800.561.8187
Page 10
10
compensate, cove the surface to be measured with masking tape or flat black paint. Allow time for the tape to reach the same temperature as the material underneath it. Measure the temperature of the tape or painted surface.
Emissivity Values
Substance
Thermal
emissivity
Substance
Thermal
emissivity
Asphalt
0.90 to 0.98
Cloth (black)
0.98
Concrete
0.94
Human skin
0.98
Cement
0.96
Lather
0.75 to 0.80
Sand
0.90
Charcoal (powder)
0.96
Earth
0.92 to 0.96
Lacquer
0.80 to 0.95
Water
0.92 to 0.96
Lacquer (matt)
0.97
Ice
0.96 to 0.98
Rubber (black)
0.94
Snow
0.83
Plastic
0.85 to 0.95
Glass
0.90 to 0.95
Timber
0.90
Ceramic
0.90 to 0.94
Paper
0.70 to 0.94
Marble
0.94
Chromium oxides
0.81
Plaster
0.80 to 0.90
Copper oxides
0.78
Mortar
0.89 to 0.91
Iron oxides
0.78 to 0.82
Brick
0.93 to 0.96
Textiles
0.90
www. .com
information@itm.com1.800.561.8187
Page 11
11
7. MAINTENANCE
Repairs or service are not covered in this manual and
should only be carried out by qualified trained technician.
Periodically, wipe the body with a dry cloth. Do not use
abrasives or solvents on this instrument.
For service, use only manufacturer’s specified parts.
www. .com
information@itm.com1.800.561.8187
Page 12
12
Les Instruments
MTP
Instruments
Siège Social
4409, rue Charleroi
Montréal-Nord, Québec
H1H 1T6
Téléphone : (514) 326-7167 Télécopieur : (514)
326-7835
Ligne directe : 1-888-326-7167
Site Web : www.mtpinc.com Courrier
électronique : info@mtpinc.com
www. .com
information@itm.com1.800.561.8187
Loading...