Motic BA410 Instruction Manual

We are constantly endeavouring to improve our instruments and to adapt them to the requirements of modern research techniques and testing methods. This involves modification to the mechanical structure and optical design of our instruments. Therefore, all descriptions and illustrations in this instruction manual, including all specifications are subject to change without notice.
1
INFINITY CORRECTED OPTICS
In this optical concept the light beams are parallel after leaving the objective in the direction of the eyepieces. A second optical element, the tube lens (normally located in the eyepiece tube) is used to converge the parallel beams, resulting in an intermediate image. The intermediate image is focussed by the eyepieces, to provide the real image for visual observation. The implementation of a tube lens gives the opportunity to minimize chromatic aberrations and other “optical defects”. Further, in “Infinity Optics” the distance between the objective and tube lens is not as strictly fixed as in the (historically older) “Finite Optics” of 160mm tube length. This allows additional optical components to be inserted between the objective and tube head. Fluorescence attachments, discussion bridges, eye level risers and other options can be added without affecting the image quality. In general “Infinity Optics” provides flexibility and the opportunity to add additional optional features.
2
TABLE OF CONTENTS
Section I Nomenclature 4 II Setting up the instrument 7
Operating environment 7
III Assembling the microscope 7
Verifying Input voltage 7 1 Mechanical stage 8 2 Specimen holder 8 3 Objectives 8 4 Condenser 9 5 Eyepiece tube 9 6 Eyepieces 10 7 Illumination 10 8 Filters 11 9 Power cord 12
IV Microscopy 12
1 Coarse and fine focusing 12 2 Coarse focus torque adjustment 12 3 Coarse focus quick stop 13 4 Stage upper limit stop adjustment 13 5 Beam splitter lever 14 6 Interpupillary distance adjustment 15 7 Diopter adjustment 15 8 Centering the condenser 16 9 Use of aperture diaphragm 17 10 Use of field diaphragm 18 11 Brightness and contrast adjustment 18
V Photomicrographic procedure 19 VI Using Oil Immersion Objectives 20 VII Troubleshooting Table 21 VIII Care and maintenance 23 IX Warning label 26
3
I. Nomenclature
MOTIC BA410 (30W Halogen)
4
MOTIC BA410 (Trinocular) 30W Halogen
5
MOTIC BA410 30W Halogen
6
II. Setting up the instrument
Avoid placing the instrument in locations exposed to direct sunlight, dust, vibration, high temperature, high humidity and where it is difficult to unplug the power supply cord.
Operating environment
y Indoor use y Altitude: Max 2000 meters y Ambient temperature: 15°C to 35°C
y
Maximum relative humidity: 75% for temperature up to 31ºC decreasing linearly to
50% relative humidity at 40ºC
y
Supply voltage fluctuations: Not to exceed ±10% of the normal voltage
y
Pollution degree: 2 (in according with IEC60664)
y Installation / Overvoltage category: 2 (in according with IEC60664)
y
Air Pressure of 75kPa to 106kPa
y Avoid frost, dew, percolating water, and rain
III. Assembling the microscope
Verifying Input voltage
y The automatic voltage selection works with a broad range of settings. However, always use a power cord that is rated for the voltage used in your area and that has been approved to meet local safety standards. Using the wrong power cord could cause fire or equipment damage. y If using an extension cord, use only a power supply cord with a protective earth (PE) wire. y In order to prevent electric shock, always turn the power supply off before connecting the power cord.
7
y Electrical Specifications:
30W Halogen Bulb
Input: 90-240V~, 80VA, 50-60Hz Bulb: 6V 30W Halogen Fuse: 250V T2.5A (If the original fuse is blown, please replace with specified fuse)
1. Mechanical stage
y Loosen the stage clamp screw; fit the stage into the circular dovetail mount. An orientation notch at the back of the stage has to fit to an orientation pin at the stage carrier. Tighten with the clamp screw facing forward.
y Torque adjustment
The torque of the y-axis and x-axis travel knobs can be adjusted by torque adjusting knobs. For y-axis to (+) increase or (-) decrease the torque, turn Knob Y while holding A For x-axis to (+) increase or (-) decrease the torque, turn Knob X while holding B
2. Specimen holder
y Attach the specimen holder, using the two mounting holes. If the stage is to be used with right right-hand controls, attach the specimen holder to the left-hand holes and similarly, with left-hand controls to the right-hand holes.
3. Objectives
y Lower the stage completely. Screw the objectives into the revolving nosepiece so that clockwise rotation of the nosepiece brings the next higher magnification objective into position.
8
Loading...
+ 19 hidden pages