This document contains confidential information that belongs to Mortara Instrument, Inc. No part of this document
may be transmitted, reproduced, used, or disclosed outside of the receiving organization without the express written
consent of Mortara Instrument, Inc. Mortara is a registered trademark of Mortara Instrument, Inc. ELI 100 is a
trademark of Mortara Instrument, Inc.
9500-090-01 A1
Technical Support and Service
Following are telephone numbers and addresses for contacting various technical support and service personnel.
Mortara Instrument, Inc.
7865 N. 86th St.
Milwaukee, WI 53224
Telephone Number: 414-354-1600
Toll-free Telephone Number: 800-231-7437
Toll-free Service Number: 888-MORTARA
Fax: 414-354-4760
E-mail address: sales@mortara.com
24 hour technical support
Over 120 trained field service technicians
Same day shipment of replacement parts
Biomedical training classes
Extended warranties/service contracts
Sales Support/Supplies & Accessories
Mortara Instrument, Inc.
7865 N. 86th St.
Milwaukee, WI 53224
Phone: 414-354-1600
Fax: 414-354-4760
Internet: http://www.mortara.com
European Economic Community Representative
Mortara Rangoni
Via Oradour, 7 40016
San Giorgio di Piano
Bologna, Italy
Phone: 39-051-6645-360
Fax: 39-051-6651-012
Mortara Instrument, Inc., B.V.
(The Netherlands).
H. Dunantplein 6
3731 CL De Bilt
Postbus 131
3720 AC Bilthoven
Telephone number: 31-30-2205050
Fax: 31-30-2201531
i
Notices
Manufacturer’s Responsibility
Mortara Instrument, Inc., is responsible for the effects on safety, and performance only if
• Assembly operations, extensions, re-adjustments, modifications or repairs are carried out by persons
authorized by Mortara Instrument,
• The electrical installation of the relevant room complies with the requirements of appropriate regulations, and
• The ELI 100 is used in accordance with the instructions for use.
• Upon request, Mortara will make available a service manual containing technical information to assist an
appropriately qualified individual with potential service related issues, though it is highly recommended that if
there are concerns regarding the devices performance, Mortara Service is contacted at 1-800-877-8942.
Responsibility of the Customer
The user of this product is responsible for ensuring the implementation of a satisfactory maintenance schedule.
Failure to do so may cause undue failure and possible health hazards.
Equipment Identification
Mortara Instrument equipment is identified by serial numbers on the back or bottom of the device. Care should be
taken so that these numbers are not defaced.
Information pertinent to tracking and manufacturing is found on the bottom of the product and may be called upon
if service of the device is required.
Copyright and Trademark Notices
This document contains information that is protected by copyright. All rights are reserved. No part of this
document may be photocopied, reproduced or translated to another language without prior written consent of
Mortara Instrument, Inc. The information contained in this document is subject to change without notice.
Other Important Information
The information in this document is subject to change without notice.
Mortara Instrument, Inc., makes no warranty of any kind with regard to this material, including, but not limited to
implied warranties of merchant ability and fitness for a particular purpose. Mortara Instrument, Inc., assumes no
responsibility for any errors of omissions that may appears in this document. Mortara Instrument makes no
commitment to update nor to keep current the information contained in this document.
ii
Warranty Information
Your Mortara Warranty
MORTARA INSTRUMENT, INC. (hereinafter referred to as “Mortara”) hereby warrants that Mortara products
(hereinafter referred to as “Products”) shall be free from defects in material and workmanship under normal use,
service and maintenance for the warranty period of such Product from Mortara or an authorized distributor or
representative of Mortara. Normal use, service and maintenance means operation and maintenance in accordance
with appropriate instructions and/or information guides. This Warranty does not apply to damage to the Products
caused by any or all of the following circumstances or conditions:
a) Freight damage;
b) Parts and/or accessories of the Products not obtained from or approved by Mortara;
c) Misapplication, misuse, abuse and failure to follow the Product instruction sheets and/or information
guides;
d) Accident, a disaster affecting the Products;
e) Alterations or modifications to the Products not authorized by Mortara;
f) Other events outside of Mortara’s reasonable control or not arising under normal operating
conditions.
THE REMEDY UNDER THIS WARRANTY IS LIMITED TO THE REPAIR OR REPLACEMENT WITHOUT
CHARGE FOR LABOR OR MATERIALS, OR ANY PRODUCTS FOUND UPON EXAMINATION BY
MORTARA TO HAVE BEEN DEFECTIVE. This remedy shall be conditioned upon receipt of notice by Mortara
of any alleged defects promptly after discovery thereof within the warranty period. Mortara’s obligations under the
foregoing warranty will further be conditioned upon the assumption by the purchaser of the Products (i) of all
carrier charges with respect to any Products returned to Mortara’s principal place or any other place as specifically
designated by Mortara or an authorized distributor or representative of Mortara, and (ii) all risk of loss in transit. It
is expressly agreed that the liability of Mortara is limited and that Mortara does not function as an insurer. A
purchaser of a Product, by its acceptance and purchase thereof, acknowledges and agrees that Mortara is not liable
for loss, harm or damage due directly or indirectly to an occurrence or consequence therefrom relating to the
Products. If Mortara should be found liable to anyone under any theory (except the expressed warranty set forth
herein) for loss, harm or damage, the liability of Mortara shall be limited to the lesser of the actual loss, harm or
damage, or the original purchase price of the Product when sold.
EXCLUDED FROM THE LIMITED WARRANTY SET FORTH ABOVE ARE CONSUMABLE ITEMS SUCH
AS PAPER, BATTERIES, ELECTRODES, PATIENT CABLES, LEAD WIRES AND MAGNETIC STORAGE
MEDIUMS.
EXCEPT AS SET FORTH HEREIN WITH RESPECT TO REIMBURSEMENT OF LABOR CHARGES, A
PURCHASER’S SOLE EXCLUSIVE REMEDY AGAINST MORTARA FOR CLAIMS RELATING TO THE
PRODUCTS FOR ANY AND ALL LOSSES AND DAMAGES RESULTING FROM ANY CAUSE SHALL BE
THE REPAIR OR REPLACEMENT OF DEFECTIVE PRODUCTS TO THE EXTENT THAT THE DEFECT IS
NOTICED AND MORTARA IS NOTIFIED WITHIN THE WARRANTY PERIOD. IN NO EVENT,
INCLUDING THE CLAIM FOR NEGLIGENCE, SHALL MORTARA BE LIABLE FOR INCIDENTAL,
SPECIAL OR CONSEQUENTIAL DAMAGES, OR FOR ANY OTHER LOSS, DAMAGE OR EXPENSE OF
ANY KIND, INCLUDING LOSS OF PROFITS, WHETHER UNDER TORT, NEGLIGENCE OR STRICT
LIABILITY THEORIES OF LAW, OR OTHERWISE. THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING, BUT NOT LIMITED TO THE IMPLIED
WARRANTY OF MERCHANT ABILITY AND THE WARRANTY OF FITNESS FOR A PARTICULAR
PURPOSE.
iii
User Safety Information
Means there is the possibility of personal
Warning:
Caution:
Note:
Federal law restricts this device to sale by or on the order of a physician.
Warning(s)
•Device (electrocardiograph, Class 1) captures and presents data reflecting a patient’s physiological condition
that when reviewed by a trained physician or clinician can be useful in determining a diagnosis. However, the
data should not
•To ensure that electrical safety is maintained during operation from AC (~) power, the device must be plugged
into a Hospital Grade outlet.
•To maintain designed operator and patient safety, peripheral equipment and accessories used that can come in
direct patient contact, must be in compliance with UL 2601-1, IEC 601-1 and IEC 601-2-25.
•To maintain designed operator and patient safety, only use parts and accessories supplied with the device and
available through Mortara Instrument, Inc.
•To avoid the possibility of serious injury or death during patient defibrillation, do not come into contact with
device or patient cables. Additionally, proper placement of defibrillator paddles in relation to the electrodes is
required to minimize harm to the patient.
•To ensure the safety of both the patient and the device, 1.5 meters (5 feet) of open area should surround the
patient.
• A possible explosion hazard exists, do not use the device in the presence of flammable anesthetics.
• ECG electrodes could cause skin irritation and should be examined for signs of irritation or inflammation.
• Before attempting to use the device for clinical applications the operator must read and understand the contents
of the manual and any documents accompanying the device.
•Where the integrity of external PROTECTIVE EARTH CONDUCTOR arrangement is in doubt,
EQUIPMENT shall be operated from its internal ELECTRICAL POWER SOURCE.
•To maintain operator and patient safety, when connecting the ELI 100 to peripheral equipment not in
compliance with UL2601-1, IEC 601-1 or IEC 601-2-25, consideration should be given to the requirements of
IEC 601-1-1.
•The ELI 100 has not been designed for use with high-frequency (HF) surgical equipment and does not provide
a protective means against hazards to the patient.
be used as a sole means for determining a patient’s diagnosis.
injury to you or others.
Means there is the possibility of damage to
the equipment.
Provides information to further assist in the
use of the device.
iv
•When connected to an external modem, a protective earth conductor must be attached to the ELI 100. The
connection may be through the mains power supply cord, or by a separate protective earth conductor attached
to the metal chassis of the ELI 100. The external modem and its power supply must comply with the
requirements of IEC 60950, safety of information technology equipment, including electrical business
equipment.
• Leakage currents can increase if additional devices are connected to the patient.
• All signal input and output (I/O) connectors are intended only for devices complying with EN 60601-1,
excluding the network connectors(s), that must only be used for connections to supporting Mortara Instrument
systems when cabling is supplied or approved by our Mortara Instrument Representative.
•Connecting/disconnecting cable at the back of the device must be done with the device OFF (main power).
Caution(s)
•To prevent possible damage to the keypad, do not use sharp or hard objects to depress keys, only use
fingertips.
•Do not attempt to clean the device or patient cables by submersing into a liquid, autoclaving, or steam
cleaning.
•Wipe the exterior surface of the device and patient cables with a sterilizing disinfectant, then dry with a clean
cloth.
•Conductive parts of the patient cable, electrodes and associated Type CF connections, including the neutral
conductor of the patient cable and electrode, should not come into contact with other conductive parts,
including earth ground.
•The rechargeable internal battery is a sealed lead acid type and it is totally maintenance free. If the battery
appears to become defective, refer to Mortara Instrument Service Department.
•Do not pull or stretch patient cables as this could result in mechanical and/or electrical failures. Patient cables
should be stored after forming them into a loose loop.
•The quality of the signal produced by the electrocardiograph may be adversely affected by the use of other
medical equipment, including but not limited to defibrillators and ultrasound machines.
v
Notes
• Excessive patient movement could interfere with the operation of the device.
• Proper patient preparation is important to proper application of ECG electrodes and operation of the device.
• Patient cables should be checked for cracks or breakage in its exterior properties prior to use.
• There is no known safety hazard if other equipment, such as pacemakers or other stimulators, are used
simultaneously with the ELI 100; however, disturbance to the signal may occur.
•If the ECG input becomes inoperable due to excessive saturation or overload, the display will indicate a lead
fail for the lead(s) that this condition is present and if the signal is being printed the respective lead(s) will print
out as a square wave.
•As defined by IEC 601-1 and IEC 601-2-25, the device is classified as follows:
- Class I equipment
- Type CF applied parts
- Ordinary equipment
- Not suitable for use in the presence of flammable anesthetics
- Continuous operation
• The ELI 100 will automatically turn off (blank screen) if the batteries have been severely discharged.
• After operating the ELI 100 using battery power, always reconnect the power cord and depress the I on the I/O
power switch. This ensures that the batteries will be recharged for the next time you use the ELI 100. The
word CHARGING or AC Power will appear on the LCD screen.
System Description ......................................................................................................................................... 1-1
ELI 100, Illustration ......................................................................................................................................... 1-2
ELI 100 Specifications .................................................................................................................................... 1-3
Setup and Installation ..................................................................................................................................... 1-4
ELI 100, Two Views: Left Side and Rear ........................................................................................................ 1-4
Protected Configuration Screen A .................................................................................................................. 1-8
AC Filter .......................................................................................................................................................... 1-8
Storage Format ............................................................................................................................................... 1-8
Protected Configuration Screen B .................................................................................................................. 1-9
Interpretation Format ...................................................................................................................................... 1-9
Protected Configuration Screen D .................................................................................................................. 1-10
Protected Configuration Screen E .................................................................................................................. 1-11
Units ................................................................................................................................................................ 1-11
Date Format .................................................................................................................................................... 1-11
Auto Delete ..................................................................................................................................................... 1-12
Serial Port ....................................................................................................................................................... 1-12
Phone Number ................................................................................................................................................ 1-13
Protected Configuration Screen K .................................................................................................................. 1-14
Auto Save........................................................................................................................................................ 1-14
Site Number .................................................................................................................................................... 1-15
ID Format ........................................................................................................................................................ 1-16
Protected Configuration Screen N .................................................................................................................. 1-16
Rhythm Lead Selection................................................................................................................................... 1-17
Site Name ........................................................................................................................................................1-17
ELI 100 Supply List..........................................................................................................................................1-21
100
2
Operation
Starting the System .........................................................................................................................................2-1
AC Power Operation........................................................................................................................................2-1
ID Menu Formatting.........................................................................................................................................2-2
Selecting the ID Menu .....................................................................................................................................2-3
ID Menu Prompts.............................................................................................................................................2-3
ID Menu Sequence (Long Format)..................................................................................................................2-5
Taking a Second ECG.....................................................................................................................................3-3
Selecting an E-SCRIBE Patient File................................................................................................................3-4
Printing a Directory ..........................................................................................................................................5-2
Selecting a Patient Name ................................................................................................................................5-4
Erasing a Patient's ECGs ................................................................................................................................5-4
Listing a Patient's ECGs..................................................................................................................................5-5
xi
Selecting a Specific ECG................................................................................................................................ 5-5
Patient ID Screen Description......................................................................................................................... 5-5
Erasing a Specific ECG .................................................................................................................................. 5-6
Maintenance on a Specific ECG ..................................................................................................................... 5-6
Direct Connection Hookup .............................................................................................................................. 5-7
Downloading a Request List ........................................................................................................................... 5-9
Printing the Request List................................................................................................................................5-10
Self-Test Printout Sample, Pg. 1 of 2.............................................................................................................. 6-3
Self-Test Printout Sample, Pg. 2 of 2.............................................................................................................. 6-4
Cleaning and Inspection ................................................................................................................................. 6-5
Inspecting the ELI 100 .................................................................................................................................... 6-5
Cleaning the ELI 100 ...................................................................................................................................... 6-5
Cleaning the Patient Cable ............................................................................................................................. 6-5
Cleaning the Writer Printhead......................................................................................................................... 6-5
System Information LogA
Appendix
System Information Log .................................................................................................................................. A-1
Figure 1-1 ELI 100......................................................................................................................................... 1-2
Figure 2-1 ELI 100 Left and Rear.................................................................................................................. 1-4
Figure 3-1 ELI 100 Rear (Fuse Installation) .................................................................................................. 1-5
The ELI 100 Operator's Manual explains how to operate the ELI 100 interpretive and non-interpretive
electrocardiographs. You can use this manual as a learning tool as well as a source of reference information. It
explains how to:
• Set up the electrocardiograph.
• Use and understand the keyboard, the viewing screen, and the menu sequences.
• Enter, modify, and delete information.
• Acquire a 12-lead electrocardiogram
• Print an ECG rhythm strip
Audience
This manual is written for clinical research professionals. They are expected to have working knowledge of medical
procedures and terminology as required for monitoring cardiac patients.
Conventions
Text that the user must type, such as a:install, appears in bold, Times New Roman font.
Keys, such as ENTER, appear in bold, upper-case, Arial font.
Text that appears on the screen, such as C:\Portrait, appears in normal, Arial font.
System Description
The ELI 100 is a diagnostic electrocardiograph capable of acquiring and printing ECG data in a hospital and/or
clinical environment. If the ELI 100 is equipped with the computerized interpretation option installed and is
enabled, which is at the user’s discretion, the acquired electrocardiogram may assist in allowing the physician to
more quickly determine what care options may be available for the patient.
The printout offers a variety of formats: three, four or six channels in automatic mode and three, six, or twelve
channels during rhythm recording. The ELI 100 can operate on batteries or line power.
1-1
ELI 100_____________________________________________________________________________
The ELI 100 consists of:
• Electrocardiograph
• ELI 100 Operator’s Manual and short form instruction card
ELI 100_____________________________________________________________________________
Setup and Installation
Figure 2-1, ELI 100
Caution: When the fuse is installed, battery will gradually discharge over time. To
guarantee battery life, always connect the power cord and depress the I on the I/O power
switch to ensure the batteries will be recharged for your next use. The word CHARGING or AC
Power will appear on the LCD screen. If fuse is removed, stored data and configuration
settings will be lost.
100The battery fuse is installed in the back of the unit, as pictured above. The fuse is installed so the ELI 100 can
store data, operate on the internal battery, and save the configuration settings.
One end of the fuse is covered with a plastic tip; the other end has a silver cap. If a new fuse needs to be installed,
follow the steps below:
n Hold the fuse by the plastic tip and insert the silver-capped end into the round socket, which is located near
the fuse “Warning” instructions on the back of the ELI 100.
o Rotate the fuse in the socket so that approximately 1/4" of the plastic cap sticks out.
p In a clockwise motion, press the plastic cap into the socket using your thumb or a screwdriver. The cap
NOTE: If you remove the battery fuse after the ELI 100 is in operation, any ECGs stored in the unit will be lost
and the initial and protected configuration settings will change to their default settings. The persistent
site and cart number configuration settings will be preserved.
NOTE: If the main switch is ON and the LCD (Liquid Crystal Display) does not display “charging” or “AC
Power,” have a qualified service technician verify that the line fuse(s) are installed.
Configuration Settings
Three configuration programs define all ELI 100 operational conditions:
1. Initial Configuration
a. Download Custom ID
b. Clock and Date settings
2. Protected Configuration
a. AC Filter, Storage Format
b. Storage Sensitivity
c. Plot Format, Interpretation Format
d. Plot Frequency and Plot Channels
e. Units, Date Format
f. Baud Rate, Auto Delete
g. Serial Port, Caps Lock
h. Include Site # in ID
i. Phone Number
j. Auto Save, Cueing
k. Copies, Retrieve Serials
l. ID format
m. Rhythm Lead Selection
n. Site Name
o. Key Click
3. Persistent Configuration
a. Site Number
b. Cart Number
1-5
ELI 100_____________________________________________________________________________
A
A
Initial Configuration
The initial configuration menu allows the user to download a custom ID and change the date and time settings.
To access the initial configuration menu, press the ALT key and the SPF (0) key simultaneously from the main
menu. The following initial configuration menu appears:
DWNLD
ID
0
LT
Custom ID Format Exit to
CLOCK
Date and Time
2
Screen
EX
3
4
567891
Main Menu
NOTE: To access the Protected Configuration and Persistent Site Number Configuration menus, you must begin
from the initial configuration menu as pictured above. Details to access the protected and persistent
configuration menus follow.
The custom ID format, which is designed in the E-SCRIBE system, can be downloaded to the ELI 100. When you
select (0) DWNLD ID, the following screen appears:
NOTE: The site number must be configured in the cardiograph and recognized as an established, valid site
number at the E-Scribe before downloading the custom ID.
LT
Download custom format?
0
2
YES
3
4
NO
567891
Back to Initial Configuration
Menu
Select (4) YES to proceed with the download – the screen will display the following message: “connection in
progress”. When the download completes, the new CUSTOM format becomes the ID menu format for all future
ECGs until you reconfigure the ID format. If you reset the format option to SHORT, STD, or LONG in the
protected configuration menu, and then wish to change back to an original or a new CUSTOM format, you must
repeat the download process from the initial configuration menu.
NOTE: CONNECTION FAILURE will be displayed if there was an error in downloading the custom ID.
Press the CLOCK key (2) to access the Clock/Date Selection screen shown below.
Current date and time: 00/00/00 00:00
New date and time: 00/00/00 00:00
n The top line of the LCD screen displays the preprogrammed current date and time set for the ELI 100.
o To reset the clock, type in the correct date and time (using a 24-hour clock) in the same format as
displayed.
US EXAMPLE: 12/31/89, 17:55 for December 31, 1989, 5:55 PM
NOTE: the date format (US or EUROPE) can be switched in the protected configuration settings.
The protected configuration menu allows the user to define the ELI 100 operational conditions that do not change on
a daily or patient-to-patient basis. Once you set these default conditions, you will rarely need to use the protected
configuration program again. To access the protected configuration menu, press the ALT key and the C key
simultaneously from the initial configuration menu.
The chart below summarizes the protected configuration menu screens and the available options for each parameter.
Detailed descriptions of the parameters for each menu option(s) follow the summary chart..
Protected Configuration Screen Summary
Screen Parameter Option
A AC Filter
Storage Format
B
C
D
E Units
F Baud Rate
G
H
I
J Modem Baud Rate 1001100 or 2400
K Auto Save
Storage Sensitivity
Plot Format
Interpretation Format
Plot Frequency
Plot Channels
Date Format
Auto Delete
Serial Port
Caps Lock
Include Site # ID
Phone Number
Cueing
60Hz or 50Hz
2.5 seconds or 10 seconds
2.5 micro-volts or 10 micro-volts.
Interpretation or No Interpretation
Reason Statements or No Reason Statements100
40Hz or 150Hz
3+3 (2pg), 3, 4
LB/IN or kg/cm
US or European
38400
XMT, XMT/PLT, IMMED, OFF, PLT
Direct or Modem
On or Off
Yes or No
User-defined (up to 40 characters including special
characters)
ON or OFF
ON or OFF
1-7
ELI 100_____________________________________________________________________________
A
A
L Site Number
Cart Number
Copies
Retrieve Serials
M ID Format LONG, SHORT, CUSTOM, STD
N Rhythm Lead
Selection
O Site Name User-defined (up to 32 alphanumeric characters)
P Key Click Soft, Medium, Loud
To advance from one configuration screen to the next, press the ENTER key. Select (9) EX from any protected
configuration screen to return to the initial configuration screen.
Display only
Display only
0 = Original copy
1 = Original plus one copy
x = Original plus up to nine copies
0 = Most recent ECG (value should always be
set to 0).
Channel 1, Channel 2, Channel 3: I, II, III, aVR,
aVL, aVF, V1-V6
Protected Configuration Screen A
C
Filter:
60Hz
Storage
Format:
2.5sEX
2
LT
1
0
3456789
Exit to
Main Menu
AC Filter
The ELI 100 software removes 60Hz or 50Hz interference. Which setting you select depends on the line frequency
in your country. Always use the 60Hz setting in the United States. If the AC interference is present, check to see
that the proper AC filter is selected.
Storage Format
There are two available storage format options in the ELI 100. One format stores 2.5 seconds of each of the twelve
leads. The other storage format option, which is the default setting, stores the full 10 seconds of all twelve leads.
Selecting this storage format option will reduce the amount of storage. For optimum storage capacity, select 2.5
seconds.
Most users select the 2.5 second storage format since it provides the standard 12-lead record. If the default Plot
Channels option is set to 4 channels (see Configuration Screen C), the 2.5 second format is modified so that 10
seconds of Lead II is stored along with 2.5 second segments of the remaining leads.
The default storage sensitivity is 2.5 uV. For optimum storage capacity, select 10 uV.
Protected Configuration Screen C
Plot
Format:INT
LT
2345678910
Plot Format
The ELI 100 automatically analyzes ECGs and prints the interpretation on the ECG printout. The Plot Format
option allows you to select or suppress the “interpretive” text on the ECG report.
NOTE: A qualified physician should review the computer generated ECG interpretation before the treatment of
any patient.
Interpretation Format
The reason statements indicate why a particular interpretive statement was printed. Reason statements print in
[square brackets] within the interpretive text (if the interpretation is turned on). Turning the reason statement
function on or off does not affect the measurements performed or the interpretive statements selected by the analysis
program. See Appendix B for examples of the reason statements on the sample printouts.
Interp
Format:REAEX
Exit to
Main Menu
1-9
ELI 100_____________________________________________________________________________
A
(
)
(
)
grap
(
p
;
p
)
p
Protected Configuration Screen D
Plot
Freq:
150Hz
Plot
Channels:
3
EX
LT
0
2
1
3456789
Exit to
Main Menu
The ECG plot frequency default setting is 150Hz, or select 40Hz. The recommended setting is 150Hz.
Plot Channels
This option defines the default for the available plot channels. The plot options are:
Format OptionECG Data
STD
Standard
4CH
2.5-second graph of 12 leads in
a 3-channel format
Condensed
in a 4-channel format
contain a labeled median com
4th channel is a 10-second rhythm
of Lead II
stri
h of 12 leads
3 channels
lex
2PG
Two Page
Standard format PLUS a 3-channel
10-second stri
of each Leads V1, II, V5
STD Format Option
The STD (standard) format is a 4" by 11" strip containing a 2.5-second graph of 12 leads in a 3-channel format.
The following print is an example of the STD format:
The patient's name, ID number, race, age, height, weight, gender and medications are printed across the top left
section of the printout. Measurements are printed in the middle to include ventricular rate, P-QRS-T axes, PR
interval, QRS duration and QT/QTc. On the top right of the printout, the ECG interpretation and reason statements
are printed.
The date and time of acquisition, location, plot frequency, and software version are printed across the bottom of the
printout.
Two and one-half seconds of leads I, II, III are printed after the calibration pulse, followed by two and one-half
seconds each of the lead sets aVR, aVL, aVF, V1, V2, V3, V4, V5, V6.
1-11
Loading...
+ 61 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.