Monarch Instrument ACT-1B Operating Manual

Page 1
Printed in the U.S.A.
Copyright 2009 Monarch Instrument, all rights reserved
1071-4843-111R
0909
MONARCH INSTRUMENT
Instruction Manual
ACT-1B Series
Panel Tachometer
15 Columbia Drive
Amherst, NH 03031 USA
Phone: (603) 883-3390
Fax: (603) 886-3300
Website: www.monarchinstrument.com
Page 2
Safeguards and Precautions
1. Read and follow all instructions in this manual carefully, and retain this manual for future reference.
2. Do not use this instrument in any manner inconsistent with these operating instructions or under any conditions that exceed the environmental specifications stated.
3. Be sure the power supplied to this instrument matches the specification indicated on the rear panel of the instrument.
4. Be sure all power is removed before making or removing any connections to or from this instrument.
5. This instrument is not user serviceable. For technical assistance, contact the sales organization from which you
purchased the product or Monarch Instrument directly.
DECLARATION OF CONFORMITY
As Manufacturer:
Monarch Instrument
Division of Monarch International Inc.
15 Columbia Drive, Amherst NH 03031 USA
declares under Monarch’s sole responsibility that the product:
Name: ACT - Panel Tachometer Models: ACT-1B
to which this declaration relates is in conformity with the following standards:
In order to comply with EU Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE): This product
may contain material which could be hazardous to human health and the environment. DO NOT DISPOSE of this product as un­sorted municipal waste. This product needs to be RECYCLED in
accordance with local regulations, contact your local authorities for more information. This product may be returnable to your distributor for recycling - contact the distributor for details.
Monarch Instrument’s Limited Warranty applies. See www.monarchinstrument.com for details.
Warranty Registration and Extended Warranty coverage available online at www.monarchinstrument.com.
EMC: EN50082-1:1997
and therefore conforms with the requirements of Council Directive 89/336/EEC relating to electromagnetic compatibility. (File R-351 4N2).
8th October 1999
Manufacturer (Amherst,NH) Alan Woolfson, VP Engineering (Authorized Signature)
EN50082-2:1995 EN55011:1991 Group 1 Class B
Page 3
@CH_A/LOEND = 12 (or 1_SEC, HALF) Sets low end time. This allows a
min reading of 5 RPM, 60 RPM, or 120
RPM. @CH_A/GATE Shows Gate Speed. (Default is 12) @CH_A/GATE = STD (1/100 Second) or FAST (1/1000 second). Sets
Gate Speed (Default is 1/100)
@DECPT Shows the number of decimal places
displayed @DECPT = NONE, 1, 2, or 3 Sets the maximum number of decimal
places
@DAC1/ FSCAL Shows Analog Out Full Scale @DAC1/FSCAL = xxx.xx Sets the Reading value that the Analog
output will output Full Scale (5V or
20mA). Depends on TYPE. @DAC1/0SCAL Shows Analog Out Zero Scale @DAC1/0SCAL = xxx.xx Sets the Reading value that the Analog
output will output Zero Scale (0V or
4mA). Depends on TYPE. (Default is
0.00)
@OUTPT Shows pulse output polarity @OUTPT = POS or NEG Sets pulse output polarity
@DISPR Shows Display Update Rate @DISPR = HALF or 1_SEC or 1.5_S. This sets the maximum display
update rate to one half a second, 1 second
or 1 ½ seconds between updates.
TABLE OF CONTENTS
TABLE OF CONTENTS
OVERVIEW ..................................................................................... 1
INSTALLATION and POWER .......................................................... 1
Installation ............................................................................... 2
Power ...................................................................................... 2
USB Programming Cable and Software ................................ 3
SENSOR CONNECTIONS ............................................................. 3
OUTPUT OPTIONS ......................................................................... 4
Current Output Option (IO) ...................................................... 5
Analog Output Option (AO) ...................................................... 6
Pulse Repeater Option (PO) ................................................... 6
SPECIFICATIONS ........................................................................... 7
OPTIONS ........................................................................................ 8
ACCESSORIES / SENSORS .......................................................... 9
APPENDIX A - Serial Programming Commands ......................... 10
@SERNO Shows unit Serial Number
11
Page 4
OVERVIEW
The ACT-1B is an economical and easy to use digital tachometer that displays rotational speed directly in RPM or RPS on a 5-digit red 0.56” high LED display using a speed sensor providing a single (or multiple) pulse(s) per revolution. The number of pulses per rev is factory set at time of ordering and can be user programmed from 1 to 999 using the optional USB Programming Cable and PM Remote Software. Power may be either universal 100 to 240 Vac (50/60Hz), or optionally, 12 Vdc or 24 Vdc isolated. The ACT-1B accepts input signals from optical, proximity, magnetic, infrared or laser sensors, or direct TTL or external ac inputs. The ACT-1B is suitable for panel mounting or bench top use with convenient screw terminal connections on the rear panel of the instrument.
If specified at time of order placement, the ACT-1B may be equipped with either an optional 4 to 20 mA current output (IO) or 0 to 5 Vdc analog output (AO) proportional to speed, and/or a TTL pulse repeater output (PO).
INSTALLATION and POWER
The ACT-1B enclosure is a standard 1/8 DIN size requiring a 3.58” wide by
1.74” high (91x44 mm) mounting hole.
ACT Front View ACT Side View
Figure 1 Dimensions in Inches (mm)
1
Appendix A - Serial Programming Commands
Programming the unit requires the optional USB Programming Cable with associated PM Remote software and a PC running Windows XP or later with an available USB port.
All serial commands are @ then two or more characters or words separated by a delimiter “/”. One or two numbers follow some commands. All valid commands respond immediately with an “OK” or data, or “ERR” if incorrect. Default baud rate is 9600.
Communication requires the USB Programming Cable.
@PI Product Information, Shows Product name \n Firmware revision \n @C2 Shows all settings @D0 Sends current display value once @D1 Sends display data continuously (at up to display update rate) @D2 Stops sending data @M X Sends Max reading @M N Sends Min reading @RE 32 Resets Max @RE 64 Resets Min @RE 96 Resets Max and Min
@CH_A/TYPE Shows current type @CH_A/TYPE = RPM Sets scale to 60 so displays in RPM. @CH_A/TYPE = FREQ Sets scale to 1 so displays in hertz. @CH_A/TYPE = SCALE Scale mode. Enter Scale factor. @CH_A/TYPE/SCALE = 30.00 This will set the SCALE factor to 30.00
@CH_A/INPUT Shows sense of the trigger input @CH_A/INPUT = POS (or NEG) Sets the sense of the input trigger @CH_A/LOEND Sets how long (in secs) with no pulses
before the tachometer shows 0.
10
Page 5
ACCESSORIES / SENSORS
T- 5 Reflective Tape - 5 foot (1.5 m) roll, 0.5 inch (10 mm) wide
USB Programming Cable with PM Remote Software on CD:
Enables the user to program the ACT-1B using a PC with USB connection. The software also allows remote monitoring of the RPM using a graphic display or an Excel
ROLS-W Remote Optical Laser Sensor with 8 foot cable
ROS-W Remote Optical Sensor with 8 foot cable
ROS-P-25 Remote Optical Sensor with 25 foot cable (must cut plug off)
ROS-HT-W-25 Remote Optical Sensor for high temperature applications
to 257 °F (125 °C) with 25 foot cable
P5-11 Proximity Sensor
M-190W Magnetic Sensor with 8 foot cable
MT-190W Magnetic Trigger Sensor with Amplifier Module, 8 foot cable
GE-200 Gasoline Engine Electromagnetic Inductive Sensor and
amplifier with 23 feet of cable
IRS-W Infrared Sensor with 8 foot cable
TM
spreadsheet.
Installation
Remove the mounting clips, if fitted, and install the unit into the panel from the front. From the rear of the unit, install the mounting clips on each side and tighten the mounting screws against the rear of the panel.
WARNING: Do not over tighten the mounting screws.
Power
Power to the unit is connected to the terminals under the section labeled POWER on the rear panel. Be sure the power supplied matches the specification indicated on the rear panel. Refer to Figure 2 below.
USB Programming Jack
Figure 2 ACT-1B Rear Panel
If the unit is ac powered (100 - 240 Vac), connect the Live (Hot) wire to the terminal marked L+ and the Neutral (Return) wire to the terminal marked N-.
If the unit is dc powered, connect the dc supply Positive to the terminal marked L+ and the dc supply Negative or Common to the terminal marked N-.
9
2
Page 6
USB Programming Cable and Software
The 3.5 mm connection “hole” in the center bottom of the rear panel is for the optional USB Programming Cable (p/n 6180-031), which comes with Windows ™ compatible PM (Panel Meter) Remote Software on CD. The cable and software combination allows the user to configure: operation mode, analog output scaling, decimal places, display update rate and pulses per input. The user can also view real time data in digital format and/or through Microsoft
®
Excel. Software instructions are
included on the software CD.
This product is designed to be safe for indoor use under the following conditions (per IEC61010-1):
Installation Category II per IEC 664 Pollution Degree Level II per IEC 664 Operating Temperature 32-122 °F [0-50 °C] Humidity Maximum relative humidity 80% for temperature up to 88 °F
[31 °C] decreasing linearly to 50% relative humidity at 104 °F [40 °C]
SENSOR CONNECTIONS
A speed sensor (not included) is connected to the terminals under the section labeled INPUT on the rear panel. Refer to Figures 2 and 3.
Connections and their functions are as follows:
+VA Positive +12 Vdc to provide power to optical, laser, infrared
or amplified magnetic sensors. Maximum load is 75 mA dc. Optionally 5 or 24 Vdc supply may be ordered.
PX+ Positive +12 Vdc supply for use with two-wire proximity
sensors. Maximum load for proper operation with two-wire sensors is 25 mA.
SIG Positive input signal from the speed sensor. Accepts TTL
pulses or ac signals, unipolar and bipolar, from 1.5 Vac to 50 Vac. (Contact the factory for increased sensitivity.) Connect the signal wire from three-wire sensors or the positive side of two-wire magnetic sensors to this terminal. Typical input impedance is 10 Kohms.
COM Common or Negative connection for both signal and power
from most sensors.
Refer to Figure 3 for connection of Monarch standard sensors. The connections are typical for these types of sensors.
3
OPTIONS
IO: 4 to 20 mA current output
AO: 0 to 5 Vdc analog output
NOTE: Full scale RPM must be specified for the above
options when ordering unless the USB Programming Cable and PM Remote Software is ordered as a separate item.
PO: 0 to 5 V TTL compatible pulse output
NOTE: Pulses out per revolution equal pulses in per
revolution.
CAL-N.I.S.T. N.I.S.T. Traceable Certificate of Calibration
8
Page 7
SPECIFICATIONS
Range: 5 to 99,999 RPM Accuracy: Resolution: 1 RPM (user programmable to 0.0001*) Display: 5 digit, 0.56” (14 mm) high red LED Display Update: Twice per second above 120 RPM (user programmable
Dimensions: 1/8 DIN by 4.5” (114 mm) deep Power Supply: Standard: 100 - 240 Vac ±10%, 50/60 Hz, 5VA
Inputs: Universal input for optical, proximity, two-wire or three-
Sensor Excitation: 12 Vdc at 20 mA for proximity sensors, 12 Vdc at 75 mA
Recommended Sensors: Optical - Monarch ROS-W
IO Option: 4 to 20 mA out, 16 bit resolution 12Vdc compliance voltage. See
AO Option: 0 to 5 Vdc out, 5 mA 16 bit resolution. Full scale RPM settings
PO Option: 0 to 5 V TTL pulse, non-inverting, one pulse out for each
* Requires optional USB Programming Cable and PM Remote Software
±1 RPM or 0.005% of reading
to 0.5, 1, or 1.5 second*)
Optional: 12 or 24 Vdc ±10%, Isolated, 5 Watts
wire magnetic, infrared or laser sensors TTL input or 2 Vac to 50 Vac peak to peak input Standard input is 1 pulse per revolution. (user programmable from 1 to 999*)
for all other sensors, Optional 5 or 24 Vdc available
Proximity - Monarch P5-11 Magnetic - Monarch M-190W or MT-190W Infrared - Monarch IRS-W Laser - Monarch ROLS-W
page 7 for maximum load calculation. Full scale RPM settings as specified when ordered. (user programmable*)
as specified when ordered. (user programmable*)
pulse in. (Positive or negative out programmable*)
PX++VA COMSIG
WHITE
M-190W
MAGNETIC
SENSOR
INPUT
BLACK
Shield to
COM
BROWN
Link SIG to COM
PROXIMITY
SENSOR
INPUT
PX++VA SIG COM
P5-11
INPUT
PX++VA SIG COM
BROWN
BLUE
BLACK
ROS-W - OPTICAL SENSOR (SHOWN)
R0LS-W - REMOTE LASER SENSOR
IRS-W - INFRARED SENSOR
MT-190W - AMPLIFIED MAGNETIC SENSOR
BLUE
Shield to
COM
INPUT
SIG
+VA PX+ COM
AC or TTL
SOURCE
Figure 3 Sensor Connections
OUTPUT OPTIONS
The ACT-1B may be equipped with either a Current Output (IO) or an Analog Output (AO), and/or a TTL Pulse Output (PO).
NOTE: Full scale RPM settings must have been specified when ordered
or may be user programmed using the optional USB Programming Cable and PM Remote Software.
The Current or Analog Outputs are connected to the terminals in the section labeled ANALOG OUTPUT OPTION on the rear panel. The actual output is marked. The Pulse Output is connected to the terminals under the section labeled OUT on the rear panel.
CAUTION: The IO or AO COM may NOT be isolated from the other
COM connections.
7
4
Page 8
Current Output Option (IO)
The current output is 4 to 20 mA. This output is a current source and has a 12 Vdc internal compliance voltage. (Optional 24 Vdc may be ordered).
Typical connections are as follows: (See Figure 4.)
Connect the Positive side of the load to the terminal marked OUT and the other (Negative) side of the load to the terminal marked
COM.
Do not use an external voltage supply.
Figure 4 Current Output Option Connections
Analog Output Option (AO)
The analog output is 0 to 5 Vdc.
Connect the Positive side of the signal to the terminal marked OUT, and the Return side of the signal to the terminal marked COM.
NOTE: If your ACT-1B is equipped with either a current output or
an analog output, the full-scale output has been factory preset to the speed range specified at the time of purchase. The output range may be programmed using the optional USB Programming Cable and PM Remote Software.
Pulse Repeater Output Option (PO)
The Pulse Repeater output provides a conditioned TTL positive going 5 V pulse out for each pulse in.
Connect the Positive signal wire (+5 V pulse) to terminal marked +P and the Return to the terminal marked COM in the rear panel section labeled OUT.
NOTE: The polarity of the optional pulse output can be set by the
user using the optional USB Programming Cable and PM Remote Software.
NOTE: With the internal 12 Vdc compliance voltage the maximum
load for the current loop is 500 Ohms.
If the optional 24 Vdc compliance option is ordered the maximum load will be 1000 Ohms.
5
6