Mizuho Osi Jackson 5927, Jackson 5943 User manual

JACKSON SPINAL SURGERY TOPS
AND 5927 RADIOLUCENT IMAGING TOP
OWNER’S MANUAL
Models 5943 & 5943AP
languages in the following order:
(en) English (es) Spanish (fr) French (de) German (it) Italian
MIZUHO OSI
30031 AHERN AVENUE UNION CITY, CA 94587
Bus: 510-429-1500 Toll Free: 800-777-4674 Fax:  510-429- 85 0 0
WWW.MIZUHOSI.COM ·
NEWHIPNEWS.COM · NEWSPINENEWS.COM
MIZUHOSI 2009 NW0504 Rev. D 1
IMPORTANT NOTICES
CAUTION: To ensure safe operation of the equipment, please READ THESE
INSTRUCTIONS COMPLETELY and keep this manual readily available to operating room personnel for future reference.
Carefully observe and comply with all warnings, cautions and instructions which are printed on the equipment or described in this manual.
The WARNING
symbol in this manual is intended to alert the user to the presence
of important operation, maintenance, or safety instructions.
Warning:
Proper pre-operative and intra-operative procedures must be followed to prevent venous stasis and pooling, pressure sore development, neuropathy, improper electrosurgical tissue grounding, hypotension and hypothermia.
NOTE: The application techniques outlined in these instructions are the manufacturer’s suggested techniques. The final disposition of each patient’s care as related to the use of this equipment rests with the attending physician.
This device is to be used by trained personnel only. PROTECTION AGAINST ELECTRICAL SHOCK HAZARD:
This symbol indicates this equipment is an applied part, TYPE B in accordance with IEC 60601-1 and is generally suitable for applications involving external or internal contact with the patient, excluding the heart. The patient circuit is connected to a protective earth ground, and this equipment should be connected only to outlets with a protective earth ground.
This symbol indicates an external ground stud that is required for use when the AC power cable is not connected to a protective earth ground outlet in your operating room or facility.
To protect the patient and hospital staff from any possible hazards, an external ground wire connection is required between the external ground stud and protective earth ground.
Warning:
Medical electrical equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this manual.
MIZUHOSI 2009 NW0504 Rev. D
2
Warning:
Portable and mobile RF communications equipment can affect medical electrical equipment.
Warning:
This equipment should not be used adjacent to or stacked with other equipment and that if adjacent or stacked use is necessary, the equipment should be observed to verify normal operation in the configuration in which it will be used.
Warning:
The use of accessories, transducers and cables other than those specified by MIZUHO OSI may result in increased emissions or decreased immunity of the equipment.
Disposal of Electrical Components In accordance with the WEEE directive, all electrical components, batteries and carbon composite components must be returned to Mizuho OSI for proper disposal. Please contact Mizuho OSI Technical Service Department at 1-800-777­4674 for further information regarding this requirement.
MIZUHOSI 2009 NW0504 Rev. D
3
TABLE OF CONTENTS
OVERVIEW.........................................................................................................................................6
1.0
1.1 General Description ......................................................................................................................6
1.2 Storage..........................................................................................................................................6
1.3 Acceptance & Transfer..................................................................................................................6
1.4 Inspection and Transfer ................................................................................................................6
2.0 COMPONENT IDENTIFICATION.......................................................................................................7
2.1 Jackson Spinal Surgery Top.........................................................................................................7
2.2 Radiolucent Imaging Top..............................................................................................................8
2.3 H-Frame........................................................................................................................................8
2.4 T-Pin..............................................................................................................................................9
2.5 Patient Safety Straps ....................................................................................................................9
2.6 Articulating Arm Board Assembly..................................................................................................9
2.7 Cervical Traction Vector..............................................................................................................10
2.8 Pulley Assembly..........................................................................................................................10
2.9 Coupling Device Identification.....................................................................................................11
3.0 TABLE TOP COUPLING PROCEDURES .......................................................................................12
3.1 5890/5891 Modular Table Bases................................................................................................12
3.2 5890/5891 Patient Transfer Safety Lock.....................................................................................12
3.3 5890/5891 Rotational Friction Control ........................................................................................13
3.4 Twenty-Five Degree Rotation Stop.............................................................................................13
3.5 Installing The H- Frames............................................................................................................14
3.6 Selecting The Mounting Holes For The Table Top.....................................................................15
3.7 5892/5803 Advanced Control Base............................................................................................17
3.8 5892/5803 Table Control Identification.......................................................................................17
3.9 Rotation Safety Lock Switch .......................................................................................................18
3.10 180° Rotation Lock Indicator.....................................................................................................18
3.11 Tilt Drive Status Indicator..........................................................................................................19
3.12 Installing The H-Frames............................................................................................................19
3.13 Selecting The Mounting Holes For The Table Top...................................................................20
4.0 IMAGING PROCEDURES................................................................................................................21
4.1 5890/5891 Modular Table Bases ............................................................................................21
4.2 5892/5803 Advanced Control Bases.......................................................................................21
5.0 SUPINE POSITIONING FOR ANTERIOR SPINE SURGERY AND PROCEDURES .....................23
5.1 5890/5891 Modular Table Bases ............................................................................................23
5.2 5892/5803 Advanced Control Bases.......................................................................................23
6.0 LATERAL POSITIONING FOR LUMBAR & THORACIC SPINE SURGERY A ND PR OCEDURES..
...........................................................................................................................................................25
6.1 5890/5891 Modular Table Bases ............................................................................................25
6.2 5892/5803 Advanced Control Bases.......................................................................................25
7.0 PRONE POSITIONING FOR POSTERIOR SPINE SURGERY AND PROCEDURES....................27
7.1 5890/5891 Modular Table Bases ............................................................................................27
7.2 5892/5803 Advanced Control Bases.......................................................................................27
7.3 5943AP Jackson Spinal Surgery Top with Advanced Control Pad System............................28
7.4 Patient Support Pad Operation................................................................................................28
7.5 Selection Of Hip Pads .............................................................................................................29
7.6 Use Of The Leg Sling..............................................................................................................30
7.7 Selection Of Head Support Method.........................................................................................31
7.8 Transferring The Patient..........................................................................................................33
7.9 Positioning Of The Head .........................................................................................................33
7.10 Positioning Of The Arms..........................................................................................................33
7.11 Positioning Of The Chest.........................................................................................................33
7.12 Positioning Of The Hips And Thighs .......................................................................................34
7.13 Positioning Of The Legs..........................................................................................................35
MIZUHOSI 2009 NW0504 Rev. D
4
ROTATION PROCEDURES.............................................................................................................36
8.0
8.1 Intra-Operative Lateral Rotation or Lateral Tilt........................................................................36
8.2 5890/5891 Modular Table Bases ............................................................................................36
8.3 5892/5803 Advanced Control Bases.......................................................................................37
8.4 180-Degree Patient Rotation...................................................................................................37
8.5 Preparation For Rotation.........................................................................................................38
8.6 Compression Sequence..........................................................................................................39
8.7 Rotation Procedure For The 5890/5891 Modular Table Bases...............................................41
8.8 Rotation Procedure For The 5892/5803 Advanced Control Base...........................................42
8.9 Removing The Radiolucent Imaging Top After Rotation.........................................................42
8.10 Rotating The Patient From Prone To Supine..........................................................................45
9.0 CART AND STORAGE IDENTIFICATION.......................................................................................46
10.0 USE OF OPTIONAL ACCESSORIES..............................................................................................47
10.1 Dual Chest Pads......................................................................................................................47
10.2 Wilson Plus™ Radiolucent Wilson Frame...............................................................................48
11.0 5943 JACKSON SPINAL SURGERY TOP STANDARD COMPONENTS......................................49
12.0 5943APJACKSON SPINAL SURGERY TOP WITH ADVANCED CONTROL PAD SYSTEM.......52
13.0 OPTIONAL ACCESSORIES ............................................................................................................53
14.0 CLEANING........................................................................................................................................56
14.1 Maintenance............................................................................................................................56
14.2 Cleaning and Disinfecting........................................................................................................56
14.3 Table Top.................................................................................................................................56
14.4 Casters....................................................................................................................................56
14.5 Table Pads...............................................................................................................................56
15.0 REPAIR AND TECHNICAL SERVICE.............................................................................................57
15.1 Contact for Parts and Service..................................................................................................57
15.2 Instant Support Value Package...............................................................................................57
15.3 To order Replacement Parts (RP)...........................................................................................57
15.4 To return damaged parts (RGA)..............................................................................................57
15.5 To send a part for repair (RA)..................................................................................................58
15.6 EC Authorized Representative................................................................................................58
MIZUHOSI 2009 NW0504 Rev. D
5
1.0 OVERVIEW
1.1 General Description
The Mizuho OSI Jackson Spinal Table System is comprised of a Modular Table Base with two interchangeable procedure-specific table tops and a variety of table accessories. The two tops are the Jackson Spinal Surgery Top and the Radiolucent Imaging Top.
For detailed information on the operation and components of the Modular Table Bases, Cervical Management Systems and Wilson Plus Radiolucent Frame, refer to the owner’s manual for each of those specific products. A thorough understanding of the use of each of these products is required in order to perform certain procedures using the Jackson Spinal Table System.
This manual requires the reader to be thoroughly familiar with the operation of the appropriate Mizuho OSI Modular Table Base owner’s manual (NW0498).
1.2 Storage
When not in use, store table top in a clean, dry environment with temperature between 32°F and 120°F (0° C to 49°C). It is recommended that the 5803 Base be stowed with the power cord plugged into a proper AC receptacle with the on/off power-switch “on” to assure batteries are fully charged and ready for service.
If recharge is required, charge a minimum of 3 hours. A full charge should be available after 18 hours.
1.3 Acceptance & Transfer
Upon receipt of your table top, remove the table top from the shipping crate according to the unpacking instructions. Visually inspect all surfaces for freight damage.
NOTE: Any freight damage must be reported to the freight carrier immediately upon
delivery. It is the responsibility of the recipient to make freight damage claims.
1.4 Inspection and Transfer
Before use, inspect the device for possible damage, excessive wear or non-functioning parts. Carefully inspect all critical, inaccessible areas, joints, electrical cords and all movable parts for possible damage.
Damaged or defective products should not be used or processed. Contact your local Mizuho OSI sales representative for repair or replacement.
MIZUHOSI 2009 NW0504 Rev. D
6
2.0 COMPONENT IDENTIFICATION
2.1 Jackson Spinal Surgery Top
The Jackson Spinal Surgery Top is 17” (43 cm) wide, 84” (213 cm) long, made from a carbon composite material and has a patient weight limit of 500 pounds (226 kg). The table top is labeled to indicate its patient weight limit. The Jackson Spinal Surgery Top includes the following items as standard components (see Figure 1):
Small Hip Pad 2 2 Regular (Medium) Hip Pad 2 2 Large Hip Pad 4 2 Thigh Pad (Right/Left) 2 2 Chest Pad Assembly 1 1 Head Support Section (Face Plate) 1 1 Leg Support Section (Leg Boards) 2 2 Leg Sling 1 1 Buttock Strap 1 1 Safety Strap 60” 1 1 Safety Strap 90” 4 4 Articulatiing Arm Board 2 2
HEAD END
Head support (face plate)
H-frame
(The table top is shown coupled to the 5803 Advanced Control Modular Table System Base.)
5943 and 5943AP JACKSON SPINAL SURGERY TOPS
(Standard Components)
Components 5943 5943AP
Figure 1: 5943/5943AP Jackson spinal surgery tops (standard components)
Chest pad
Articulating arm boards (2)
Figure 2: 5943 Jackson spinal surgery top with components
Thigh pads (2)
Hip pads (2)
Support sections (leg boards)
Spinal surgery
H-frame
FOOT END
top
MIZUHOSI 2009 NW0504 Rev. D
7
2.2 Radiolucent Imaging Top
The Radiolucent Imaging Top is 21.5” (54.6 cm) wide, 84” (213 cm) long, made from a carbon composite material and has a patient weight limit of 500 pounds (226 kg). Each Radiolucent Imaging Top is labeled to indicate its patient weight limit. The imaging top is radiolucent including views through the radiolucent side rails. It includes a 2” thick Mizuho OSI Tempur-Med® table pad and two (2) side rail mounted safety straps.
HEAD END
H-frame Table pad H-frame
FOOT END
Safety strap (2)
Radiolucent imaging top
Figure 3: Radiolucent Imaging Top component identification
(The table top is shown coupled to the 5803 Advanced Control Modular Table System Base)
NOTE: Mizuho OSI P/N 5855-830 side rail adapters must always be used to attach these types of devices to the table to prevent damage. If retractors, clark sockets, arm boards or other accessories are attached directly to the side rail of the radiolucent imaging top, they may cause damage in the form of cracking or splintering of the carbon fiber table top. Please see section 3.0 in this owner’s manual for more information related to the use of the Radiolucent Imaging Top.
2.3 H-Frame
Each Jackson Spinal Surgery Top and Radiolucent Imaging Top is supplied with two (2) H-frames (see Figure 4). The H-frames are used to connect the table tops to the base with T-pins. Four (4) H-frames are required to attach both table tops for a 180 degree patient rotation (see section 8.0: Rotation Procedures).
Figure 4: H-Frame
MIZUHOSI 2009 NW0504 Rev. D
8
2.4 T-Pin
Each Jackson Spinal Surgery Top and Radiolucent Imaging Top is supplied with four (4) T-pins (see Figure 5). The T-pins are used to connect the H-frames to the crossbars on the table base and to connect the table tops to the H-frames. Eight (8) T-pins are required to attach both table tops for a 180 degree patient rotation (see section 8.0: Rotation Procedures).
Drop lock
Figure 5: T-pin
2.5 Patient Safety Straps
The Jackson Spinal Surgery Top is supplied with six (6) patient safety straps; four (4) 90” straps, one (1) 60” strap, and one (1) buttock strap (see Figure 6). When a patient is positioned on the Jackson Spinal Surgery Top or Radiolucent Imaging Top, verify that the appropriate patient safety straps are applied and secured as described in section
7.0 of this owner’s manual.
Buttock strap
60"
90"
Figure 6: Jackson Spinal Surgery Top patient safety straps
2.6 Articulating Arm Board Assembly
The Jackson Spinal Surgery Top is supplied with two (2) articulating arm boards and mounting brackets (see Figure 7). These arm boards may be attached directly to the spinal top at any location along the length of the frame. The articulating arm board is adjustable in six directions as shown by the arrows in Figure 7.
Armboard assembly
Mounting bracket
Figure 7: Articulating arm board for the Jackson Spinal Surgery Top
MIZUHOSI 2009 NW0504 Rev. D
9
2.7 Cervical Traction Vector
The modular table base is equipped with a cervical traction vector (see Figure 8). The cervical traction vector provides the adjustability of the angle of pull for cervical traction as shown in Figure 9.
Figure 8: Cervical traction vector
2.8 Pulley Assembly
The pulley assembly is permanently attached to the head end column to guide a traction rope. The cervical traction vector attaches to the medial side of the head end H-frame in the mounting holes to achieve the desired angle of traction (see Figure 9). Sandbag weights may be added to provide the appropriate traction load during anterior or posterior cervical surgery or any other time cervical traction is desired. The traction weight is connected to a rope, which passes over the pulley assembly, threads through the pivot shaft of the head end crossbar and connects to the patient traction apparatus (see Figure 10).
Cervical traction vector
Pulley
Traction rope connecting to patient traction apparatus
Pivot shaft
Weight
Figure 9: Cervical traction vector installed in H-frame.
Figure 10: Traction pulley assembly
NOTE: The amount of traction load applied to the patient is completely at the discretion of the attending physician.
MIZUHOSI 2009 NW0504 Rev. D
10
p
p
2.9 Coupling Device Identification
Identify the coupling device on the end of the table top (see Figures 11 and 12). Early versions of the Jackson Spinal Surgery Top and Imaging Tops have fixed mounting tubes. Jackson Spinal Surgery Tops built after October 1999 and Radiolucent Imaging Table Tops built after January 1997 will have gimbal mounting assemblies (Figure 12). Gimbals allow for rotation of the table top when positioned in extreme lateral tilt and trendelenburg or reverse trendelenburg. Both designs couple to the H-frame in the same fashion. The 5890/5891 Modular Table Bases can utilize either version of the coupling devices but the 5892/5803 Advanced Control Bases must use only gimbal mounting assemblies (Figure 12).
WARNING
:
Use of table tops on the 5892/5803 Advanced Control Bases without gimbal mounting assemblies (Figure 12) can result in damage to the table top.
Gimbal mounting tube
Fixed mounting tube
Figure 11: Fixed mounting tube on Jackson S
inal Surgery To
Figure 12: Gimbal mounting assembly on Radiolucent Imaging Top
MIZUHOSI 2009 NW0504 Rev. D
11
3.0 TABLE TOP COUPLING PROCEDURES
)
The Jackson Spinal Surgery Top and Radiolucent Imaging Top may be coupled and interchanged to any modular table base, as needed, to provide flexibility and procedural capability.
NOTE: For complete instructions on the modular table base operation and functions, please refer to the appropriate owner’s manual supplied with the Modular Table Base. A thorough understanding of the operation of the 5890 and 5891 Modular Table Base is required before continuing.
3.1 5890/5891 Modular Table Bases
The following instructions are directed toward the use of a table top on the 5890 or 5891 Modular Table Base. These bases do not have the powered lateral tilt function.
3.2 5890/5891 Patient Transfer Safety Lock
Each 5890/5891 Modular Base is equipped with a patient transfer safety lock at both the head and foot end columns. The patient transfer safety lock is actuated by positioning the lever protruding out of the right side of the control box at the head end column (see Figure 13). A second patient transfer safety lock is located on the side of the foot end control box (see Figure 14). When engaged, the patient transfer safety lock prevents rotation of the crossbar by locking the crossbar in the horizontal position.
Rotational friction
control
Crossbar
Crossbar
Patient transfer
safety lock
Disengage
motion
Patient transfer
safety lock
Disengage
motion
Figure 13: Head end column (5890/5891 bases shown
Figure 14: Foot end column (5890/5891 bases shown)
MIZUHOSI 2009 NW0504 Rev. D
12
3.3 5890/5891 Rotational Friction Control
The rotational friction control is located at both the head and foot end columns. At the head end, the rotation friction control is the large right-angled lever on top of the head end control box (see Figure 13). At the foot end, the rotation friction control is the black lever that is located on the lower side of the control box (see Figure 15). The rotation friction control may be tightened to act as a friction brake or it can be adjusted to hold the table top at any degree of lateral tilt. As a friction brake, it can be loosened to maintain the desired amount of resistance while still allowing the table top to be adjusted to a different degree of tilt. When completely loose, the rotation friction control will facilitate 180 degree rotations. Rotate the rotation friction control clockwise to increase friction and counterclockwise to reduce friction. It is recommended that the rotation friction control be left in the maximum friction position (fully clockwise) whenever table top movement is not intended.
Crossbar
Figure 15: Foot end column (5890/5891 base shown)
Rotation friction control
NOTE: Because the foot end rotation friction control lever has limited range of movement, occasionally it may not be possible to fully loosen or tighten the brake with one turn of the handle. In that case, once the end of the movement range has been reached, pull the handle down. This will disengage the mechanism and allow the handle to be returned to its starting position without affecting the amount of friction. Repeat the process until the desired amount of friction is achieved.
3.4 Twenty-Five Degree Rotation Stop
Each modular base is equipped with a 25-degree rotation stop. The rotation stop is actuated by the lever located on the left side of the control box near the top of the head end column. It can be easily activated with the left hand when standing at the head end of the table (see Figure 16). When engaged, the rotation stop prevents lateral rotation greater than 25 degrees from the horizontal position. The rotation stop should be left in the engaged position and disengaged only during a 180-degree rotation.
Crossbar
Disengage motion
MIZUHOSI 2009 NW0504 Rev. D
25 Degree rotation stop
Figure 16: Head end column 25 degree rotation stop (5890/5891 base shown)
13
p
3.5 Installing The H- Frames
The following coupling procedure assumes the head end and foot end columns do not have either the Jackson Spinal Surgery Top or the Radiolucent Imaging Top in place.
1. Lock the table base in position by engaging the (4) caster locks.
2. Install an H-frame to head end crossbar and secure with a T-pin (see Figures 17 and 18).
Drop lock
Crossbar
HEAD END
Crossbar
Figure 17: H-frame aligned with mounting stud and T-pin prepared for insertion
H-frame T-pin
Mounting stud
T-pin
Figure 18: T-pin insertion into H-frame and crossbar head end column (5890/5891 base shown)
H-frame
WARNING:
Failure to install the T-pin will result in the patient and table assembly being dropped.
3. Install an H-frame to foot end crossbar and secure with a T-pin. Fully retract the foot
end crossbar by sliding it toward the foot end column (see Figure 19).
Drop lock
H-frame
T-
Crossbar
in
ON
T-pin
SLIDE
Rotation friction control
Figure 19: H-frame installed – foot end column (5890/5891 base shown)
MIZUHOSI 2009 NW0504 Rev. D
14
4. Engage the patient transfer safety lock on the head end and foot end columns (see
Figures 13 and 14)
Warning:
For safe patient transfer, the patient transfer safety lock must be engaged prior to transfer in order to prevent the possibility of the table top rotating while the transfer occurs. Failure to do so may result in the patient being dropped.
5. Set the rotational friction control on the head and foot end columns to maximum
friction by fully rotating clockwise (see Figures 13 and 15).
6. Set the 25 degree rotation stop on the head end column (see Figure 16).
Warning:
Prior to coupling the table tops, verify that the two (2) H-frames are properly installed to the crossbar. Confirm the T-pins passes completely through the H­frames and the crossbars and that the T-pin drop locks are in sight and pivot freely. Failure to do so may result in the patient being dropped.
3.6 Selecting The Mounting Holes For The Table Top
When selecting the H-frame mounting hole, it is intended that the center of gravity of the patient be at or below the centerline of the crossbars. For most patients, the fourth open mounting hole below the crossbar is the appropriate position (see Figure 20).
Centerline of crossbars
Center of gravity of patient
Figure 20: Center of gravity axis of patient below the centerline of crossbars
NOTE: Two people are required to complete the following sequence of actions.
7. Select the appropriate mounting holes on the H-frame to support the table top.
Insert the T-pin into the H-frame on the head end column first. Align the mounting tube on the table top to the selected mounting hole on the H-frame. Push the T-pin through the H-frame, completely through the mounting tube and through the opposite side of the H-frame. The T-pin drop lock should be in sight and pivot freely (see Figure 21).
MIZUHOSI 2009 NW0504 Rev. D
15
HEAD END
H-frame
Drop lock
Figure 21: Head end column with coupled Jackson Spinal Surgery Top
T-pin
Jackson Spinal Surgery Top
Warning:
Verify that the T-pin is inserted through the H-frame, passes completely through the table mounting tube and through the opposite side of the H-frame and that the T-pin drop lock is in sight and pivots freely.
8. To couple the foot end of the table top, slide the H-frame and crossbar as necessary
to accommodate the length of the table top (see Figure 22). Align the mounting tube on the table top to the selected hole on the foot end H-frame. Push the T-pin through the H-frame, completely through the mounting tube on the table top and through the opposite side of the H-frame. This should be done at the same level as the mounting hole where the head end was coupled. The T-pin drop lock should be visible and pivot freely.
Jackson Spinal Surgery Top
H-frame
T-pin
Figure 22: Foot end column with coupled Jackson Spinal Top
FOOT END
Drop lock
MIZUHOSI 2009 NW0504 Rev. D
16
Warning:
Failure to follow the above prescribed procedures regarding securing the H-frame to the crossbars and the table top to the H-frames may result in the patient being dropped.
Figure 23: Jackson Spinal Surgery Top coupled to 5890/5891 Modular Table Base
(shown without patient support pads)
3.7 5892/5803 Advanced Control Base
The Jackson Spinal Surgery Top and Radiolucent Imaging Top may be coupled and interchanged to the 5892/5803 Advanced Control Base as needed to provide flexibility and procedural capability. Identify the coupling device on the end of the table top as described in section 2.9, Figures 11 and 12 of this manual.
Warning:
The Jackson Spinal Surgery or Radiolucent Imaging Tops used on the 5892/5803 Advanced Control Base must have gimbals (Figure 12). Earlier versions of these table tops do not have gimbals. Use of table tops without gimbals can result in damage to the table top.
NOTE: For complete instructions on the 5892/5803 Advanced Control Base operation and functions, please refer to the appropriate owner’s manual supplied with the advanced control base. A thorough understanding of the operation of the 5892/5803 advanced control base is required before continuing.
3.8 5892/5803 Table Control Identification
The head end of the 5892/5803 Advanced Control Base is equipped with three indicator lights. These lights are: 1) 180° rotation lock indicator light, 2) tilt drive status light and 3) rotation safety indicator light. Verify that the rotational safety lock switch is in the ON position, and the 180 degree rotation lock lever is in the locked position. All three lights must be illuminated before a patient is transferred to the table.
MIZUHOSI 2009 NW0504 Rev. D
17
Rotation safety lock switch / indicator light
180° rotation lock indicator light
Tilt drive status indicator light
Crossbar
Traction pulley 180° rotation
lock lever
Figure 24: Indicator lights, head end column 5892/5803 Advanced Control Base
3.9 Rotation Safety Lock Switch
The rotation safety lock switch (with the lighted indicator) locks out the lateral tilt function. This feature is designed to prevent unintended lateral tilt of the table top. The rotation safety lock switch should be turned off only when a 180-degree patient rotation is to be performed. If the rotation safety lock switch is not illuminated, press the rocker switch to the ON position.
3.10 180° Rotation Lock Indicator
The 180° rotation lock indicator light, when lit, indicates the 180-degree rotation lock is engaged and the table cannot be rotated 180 degrees. If the 180-degree rotation lock indicator is not on, rotate the rotation lock lever clockwise until the light is illuminated.
NOTE: If a 180 degree patient rotation is to be performed, the rotation lock lever should be rotated counterclockwise approximately one half rotation past the point that the rotation lock indicator light goes out. Rotating the lock lever to this point lessens the drag of the friction control, thereby allowing the table tops to rotate more freely during the patient rotation. If even less friction is desired, the lock lever may be rotated further, counterclockwise, as necessary.
After the patient rotation is complete, rotate the rotation lock lever clockwise until the rotation lock indicator light is illuminated.
Warning:
When the rotation lock indicator light is not illuminated, the table top is UNLOCKED regardless of how much the lock lever has been rotated clockwise or counterclockwise. Always rotate the lock lever until the rotation lock Indicator light is illuminated after a 180° rotation. If increased friction lock on the rotation is required, rotate handle in a clockwise direction until desired friction is met.
Warning:
All three lights must be illuminated before a patient is transferred to the table. If all three lights are not illuminated, do not transfer the patient until they are illuminated.
MIZUHOSI 2009 NW0504 Rev. D
18
3.11 Tilt Drive Status Indicator
The tilt drive status indicator light is lit when the power lateral tilt mechanism is in the center position of the +/- 25 degree powered lateral tilt range. If the tilt drive indicator is not illuminated, operate the lateral tilt function using the hand pendant in the appropriate direction until the indicator is lit. The head end crossbar should be horizontal when the tilt drive status indicator is lit. If the head end crossbar is not horizontal, the internal power tilt mechanism is out of synchronization with the crossbar. Refer to the section on synchronizing the lateral tilt function in the 5892 advanced control base user guide or the 5803 owner’s manual, to reset this function.
3.12 Installing The H-Frames
The following coupling procedure assumes the head end and foot end columns do not yet have either the Jackson Spinal Surgery Top or the Radiolucent Imaging Top in place. The coupling procedure for the Jackson Spinal Surgery Top or the Radiolucent Imaging Top is as follows:
1. Lock table base in position by engaging the (4) caster locks.
2. Install an H-frame to head end crossbar and secure with a T-pin (see Figure 25).
Fully retract the head end crossbar against the head end column.
3. Install an H-frame to foot end crossbar and secure with a T-pin (see Figure 26).
HEAD END
Sliding assembly
Crossbar
FOOT END
T-pin
Drop lock
Figure 25: H-frame installed - head end column 5892/5803 Advanced Control Base
ON
H-frame
ON
H-frame
Figure 26: H-frame installed – foot end column 5892/5803 Advanced Control Base
T-pin
MIZUHOSI 2009 NW0504 Rev. D
19
3.13 Selecting The Mounting Holes For The Table Top
When selecting the H-frame mounting hole, it is intended that the center of gravity of the patient be at or below the centerline of the crossbars. For most patients, the fourth mounting hole below the crossbar is the appropriate position (see Figure 20).
NOTE: Two people are required to complete the following sequence.
4. Select the appropriate mounting holes on the H-frame to support the table top.
Insert the T-pin into the H-frame on the foot end column first. Align the mounting tube on the table top to the selected mounting hole on the H-frame. Push the T-pin through the H-frame, completely through the mounting tube and through the opposite side of the H-frame. The T-pin drop lock should be visible and pivot freely (see Figure 22).
Warning:
Verify that the T-pin is inserted through the H-frame, passes completely through the table mounting tube, and through the opposite side of the H-frame and the T­pin drop lock is in sight and pivots freely. Failure to do so will result in the patient and table assembly being dropped.
5. To couple the head end of the table top; slide the H-frame and crossbar away from
the foot end (shown in Figure 25) as necessary to accommodate the length of the table top. Select the appropriate mounting holes on the H-frame to support the table top. Align the mounting tube on the table top to the selected mounting hole on the H-frame. Push the T-pin through the H-frame, completely through the mounting tube and through the opposite side of the H-frame. This should be done at the same level as the mounting hole where the head end was coupled. The T-pin drop lock should be in sight and pivot freely (see Figure 21).
A table top completely coupled to a table base is shown in Figure 23.
Warning:
Failure to follow the above prescribed procedures regarding securing the H-frame to the crossbars and the table top to the H-frames may result in the patient being dropped.
MIZUHOSI 2009 NW0504 Rev. D
20
4.0 IMAGING PROCEDURES
To perform most imaging procedures, attach the Radiolucent Imaging Top according to the following procedure;
1. Lock the table base in position by engaging the (4) caster locks.
2. Verify that the table controls are locked.
4.1 5890/5891 Modular Table Bases
1. Engage the patient transfer safety locks at both the head and foot end columns (see
section 3.0, Figures 13 and 14).
2. Set the rotational friction control to maximum friction at both the head and foot end
columns (see section 3.0, Figures 13 and 15).
3. Engage the 25° rotation stop (see section 3.0, Figure 16).
4.2 5892/5803 Advanced Control Bases
1. Confirm the rotation safety lock switch is illuminated (see section 3. 0, Figure 24).
2. Confirm rotation lock indicator light is illuminated (see section 3.0, Figure 24).
3. Confirm the tilt drive status indicator is illuminated (see section 3.0, Figure 24).
4.2.1 Attach the Radiolucent Imaging Table Top with table pad (see section 3.0: Table
Top Coupling Procedures), depending on which base is used.
4.2.2 Transfer the patient in a supine position to the Radiolucent Imaging Top. Apply
the patient safety strap and secure.
Figure 27: Patient supine on the Radiolucent Imaging Top with pivoting arm boards
MIZUHOSI 2009 NW0504 Rev. D
21
The radiolucent side rails are designed to support surgical arm boards. We recommend Mizuho OSI P/N 5356 pivoting arm boards as shown in Figure 28. These should be used in conjunction with P/N 5855-830 side rail adapters to prevent damage to the imaging top side rails.
Pivoting armboard
Imaging top pad
Figure 28: Pivoting arm board with side rail adapter on the Radiolucent Imaging Top
P/N 5855-830 Side rail adapter
Imaging top radiolucent side rail
MIZUHOSI 2009 NW0504 Rev. D
22
5.0 SUPINE POSITIONING FOR ANTERIOR SPINE SURGERY AND PROCEDURES
The Radiolucent Imaging Top may be used for anterior spine procedures according to the following procedure:
1. Lock the table base in position by engaging the (4) caster locks.
2. Verify that the table controls are locked.
5.1 5890/5891 Modular Table Bases
1. Engage the patient transfer safety locks at both the head and foot end columns (see
section 3.0, Figures 13 and 14).
2. Set the rotational friction controls to maximum friction at both the head and foot end
columns (see section 3.0, Figures 13 and 15).
3. Engage the 25° rotation stop (see section 3.0, Figure 16).
5.2 5892/5803 Advanced Control Bases
1. Confirm the rotation safety lock switch is illuminated (see section 3.0, Figure 24).
2. Confirm rotation lock indicator light is illuminated (see section 3.0, Figure 24).
3. Confirm the tilt drive status indicator is illuminated (see section 3.0, Figure 24).
5.2.1 Attach the Radiolucent Imaging Table Top with table pad or the 5879-2 Anterior Extension Positioner with 5” bladder (see section 3.0: Table Top Coupling Procedures), depending on which base is used.
NOTE: For anterior cervical surgery where cervical spine extension is required, it is recommended that the Mizuho OSI P/N 5879-2 Anterior Extension Positioner with 5” bladder be used instead of the standard imaging top pad assembly (see Figure 29).
5.2.2 Transfer the patient to a supine position on the Radiolucent Imaging Table Top. Apply the patient safety strap and secure.
5.2.3 The arms may be positioned parallel to the patient, tucked and s upported with padding and a draw sheet.
Cervical management system
Figure 29: Patient table top with the optional anterior extension positioner and cervical management system
MIZUHOSI 2009 NW0504 Rev. D
Imaging top
Anterior extension positioner
23
5.2.4 Prep and drape the patient in the standard fashion. This position provides ideal access to the anterior cervical spine or ideal access to the anterior lumbar spine through an abdominal approach which gives exposure to the lumbar-sacral area. If the surgeon should desire to laterally rotate (tilt) the patient, carefully follow the directions in section 8.0: Rotation Procedures, depending on which base is used.
The option to perform surgery with the patient in cervical traction is readily available. The Mizuho OSI P/N 5979-1 cervical management system is recommended for this function (see Figure 29).
For anterior lower lumbar spine surgery, the Mizuho OSI P/N 973 or 974 inflatable bladder may be used under the lower back to provide controlled lumbar spine extension.
Upon completion of the anterior surgery, it is possible to rotate the patient to a prone position while maintaining cervical traction and without having to remove the patient from the table (see section 8.0: 180-degree patient rotation).
Figure 30: Inflatable bladder p/n: 973 (6”x 10”) and p/n: 974 (8”x10”)
MIZUHOSI 2009 NW0504 Rev. D
24
6.0 LATERAL POSITIONING FOR LUMBAR & THORACIC SPINE SURGERY AND PROCEDURES
A variety of surgical procedures in the lateral decubitus position may be performed on the Radiolucent Imaging Top according to the following procedures:
1. Lock the table base in position by engaging the (4) caster locks.
2. Verify that the table controls are locked.
6.1 5890/5891 Modular Table Bases
1. Engage the patient transfer safety locks at both the head and foot end columns (see section 3.0, Figures 13 and 14).
2. Set the rotational friction controls to maximum friction at both the head and foot end columns (see section 3.0, Figures 13 and 15).
3. Engage the 25° rotation stop (see section 3, Figure 16).
6.2 5892/5803 Advanced Control Bases
1. Confirm the rotation safety lock switch is illuminated (see section 3.0, Figure 24).
2. Confirm rotation lock indicator light is illuminated (see section 3.0, Figure 24).
3. Confirm the tilt drive status indicator is illuminated (see section 3.0, Figure 24).
6.2.1 The Radiolucent Imaging Top should be coupled at the fifth or sixth mounting hole below the centerline of the crossbar for lateral procedures (see section 3.0, Figure 20).
Warning:
The Radiolucent Imaging Top must be coupled with the center of gravity of the patient below the centerline of the crossbars when the patient is positioned laterally. Failure to do so may make the patient unstable during lateral rotation (see Figure 20).
6.2.2 Transfer the patient to a supine position on the Radiolucent Imaging Top; apply a patient safety strap and secure.
6.2.3 Roll the patient on his or her side. Stabilize and maintain the patient in the lateral decubitus position by means of at least two (2) patient safety straps or a lateral positioning device such as the Mizuho OSI P/N 5300 Lateral Positioner or the Mizuho OSI P/N 5301 Deluxe Lateral Positioner (see Figure 31). You may wish to use standard positioning aids, e.g., pillows, an axillary roll and foam or gel padding for additional patient stability and comfort.
MIZUHOSI 2009 NW0504 Rev. D
25
Imaging top as viewed from above
Lateral positioners
Figure 31: Mizuho OSI P/N 5300 Lateral Positioner on Radiolucent Imaging Top
6.2.4 The patient may be laterally rotated to a maximum of 25° in each direction. To do so, use the appropriate table base controls. Do not release the 25° rotation stop or attempt to rotate the patient in the lateral position more than 25 degrees.
Warning:
The 25° rotational stop must always be engaged when using the Radiolucent Imaging Top for lateral positioning. Lateral rotation of a patient in the lateral position to more than 25° may result in the patient being dropped.
NOTE: If increased stability in the lateral position is desired or if more than 25° of lateral rotation is required for thoracic surgery; we recommend the use of the Mizuho OSI 5895 Maximum Access™ Lateral Top (see Figure 32). This top is also used for simultaneous anterior/posterior lumbar surgery. Contact your Mizuho OSI representative for more information regarding this product.
MIZUHOSI 2009 NW0504 Rev. D
26
Figure 32: P/N 5895 Maximum Access™ Lateral Top
7.0 PRONE POSITIONING FOR POSTERIOR SPINE SURGERY AND PROCEDURES
The Jackson Spinal Surgery Top can be used for prone procedures according to the following procedure:
1. Lock the table base in position by engaging the (4) caster locks.
2. Verify that the table controls are locked.
7.1 5890/5891 Modular Table Bases
1. Engage the patient transfer safety locks at both the head and foot end columns (see section 3.0, Figures 13 and 14).
2. Set the rotational friction controls to maximum friction at both the head and foot end columns (see section 3.0, Figures 13 and 15).
3. Engage the 25° rotation stop (see section 3, Figure 16).
7.2 5892/5803 Advanced Control Bases
1. Confirm the rotation safety lock switch is illuminated (see section 3.0, Figure 24).
2. Confirm rotation lock indicator light is illuminated (see section 3.0, Figure 24).
3. Confirm the tilt drive status indicator is illuminated (see section 3.0, Figure 24).
7.2.1 Attach the Jackson Spinal Surgery Top (see section 3.0: Table Top Coupling
Procedures).
Warning:
The Jackson Spinal Surgery Top must be coupled with the center of gravity of the patient at or below the centerline of the crossbars when the patient is prone. Failure to do so may make the patient unstable during lateral rotation (see Figure
20).
MIZUHOSI 2009 NW0504 Rev. D
27
7.2.2 Open the patient care kit. The patient care kit is intended for SINGLE PATIENT
USE ONLY. When using a GentleTouch® Pillow, it is recommended to open the pillow at least 20 minutes prior to use, in order to allow for complete expansion of the pillow.
7.3 5943AP Jackson Spinal Surgery Top with Advanced Control Pad System
The 5996 Advanced Control Pad System™ (ACP) with TEMPUR-MED® is designed specifically for use with the Jackson Spinal Top. Utilizing the ACP System, automatic massage action and periodic pressure point stimulation is provided to the patient at the site of the hip and thigh pads. The ACP System consists of three components that work together:
1. 5996-5 Variable cycle controller 100-120 VAC or 5996-6 Variable cycle controller­230 VAC
2. 5996-200 Mizuho OSI TEMPUR-MED® patient support pads
3. 5996-12 ACP Quick connect tubing
The ACP System is an electronically controlled, pneumatically actuated system with a rating of IPX-4 waterproof rating. It operates at either 100-120V-AC, 50/60hz or at 230V-AC, 50hz. Refer to the 5996 Advanced Control Pad System user guide (NW0491) for complete set-up and use instructions.
7.4 Patient Support Pad Operation
The patient support pads attach to the Jackson Spinal Surgery Top using the lock lever on the pad base. To attach, slip the pad base over the frame of the spinal top and rotate the lever to lock as shown in Figures 33 and 34.
Pad support
Jackson spinal surgery top frame
Clamp post
Rotate to lock
Clamp post
Jackson spinal surgery top frame
Lock lever
Lock lever
Figure 33: Unlocked position
(Support pad shown from below)
Figure 34: Locked position
MIZUHOSI 2009 NW0504 Rev. D
28
7.5 Selection Of Hip Pads
(
Small
Regular
medium)
Large
Figure 35: Identification of available hip pads
7.5.1 For patients under 350 pounds; determine which hip pad set (small, regular or large) will be used, according to the size of the patient. Once installed, confirm that the pad locking levers latch securely.
HEAD END
Head support (face plate)
Chest pad
Figure 36: Table pad set-up for a patient up to 350 pounds
For 350 to 500 pound patients it may be necessary to: use four (4) large hip pads in conjunction with two (2) thigh pads and one (1) chest pad (see Figure 37). Once installed, confirm that the pad locking levers latch securely.
Head support (face plate)
Articulating arm boards
Figure 37: Table pad set-up for a patient of 350 to 500 pounds
Chest pad
7.5.2 Attach the appropriate patient support pads onto the Jackson Spinal Surgery Top. Lock the pads in place by rotating the locking lever onto the clamp post. Once installed, confirm that the pad locking levers latch securely (see Figures 33 and 34).
Thigh pads (2)
Hip pads (2)
Articulating arm boards
Thigh pads (2)
Large hip pads (4)
Jackson spinal top
FOOT END
Spinal surgery
top frame
Leg supports
H-frame
Spinal surgery
top frame
Leg supports
MIZUHOSI 2009 NW0504 Rev. D
29
NOTE: In order to achieve the correct and final position of the pads, it may be necessary to reposition the pads along the table frame after a patient has been transferred. To do so, carefully lift the patient, release the pad locking lever; slide and relocate the pad to the designed position and re-engage the locking lever.
Warning:
Verify that the pad locking levers are latched securely in place. Failure to do so may result in the pad bases breaking when the patient is being transferred on or off the Jackson Spinal Surgery Top.
7.5.3 Apply the GentleTouch® face pillow, chest, hip and thigh pad covers as described in the patient care kit instructions.
7.5.4 Attach either the two leg support sections or the leg sling to the table top. The leg support sections are shown in Figures 36 and 37; the sling is shown in Figure 38.
7.6 Use Of The Leg Sling
The leg sling may be used to support the legs on the Jackson Spinal Surgery Top in place of the leg support sections (see Figure 38). The leg sling can allow for increased hip flexion by allowing the patient’s knees to be lowered below the frame of the Jackson Spinal Surgery Top. When the sling is used, it is recommended not to use the thigh pads as the increased hip flexion angle may increase the pressure on the anterior aspect of the thighs. A patient properly positioned in a leg sling is shown in Figure 43.
Figure 38: Leg sling
The sling should be adjusted for each patient. The straps on the sling consist of a double-sided hook closure that allows for infinite adjustment along the length of the strap. To install the sling, place the sling over the frame of the foot end of the spinal top. Feed the four straps through their respective buckles on the opposite side of the sling and secure. The hook end of the strap may be folded over onto the loop portion of the strap to allow for greater adjustability. The strap may be folded over itself again to shorten it, thereby raising the sling and decreasing the hip flexion angle of the patient.
7.6.1 Place two (2) or three (3) pillows in the leg sling approximately where the patient’s lower legs will rest to help achieve the desired amount of knee flexion. This can also be done when using the leg support sections.
MIZUHOSI 2009 NW0504 Rev. D
30
Loading...
+ 282 hidden pages