Mizuho Osi 7200B User manual

SURGICAL T ABLES
10-04
MODEL 7200B
OPERATORS MANUAL
19
TABLE OF CONTENTS
SPECIAL USER ATTENTION ................................................................................................................ 2
SECTION I INTRODUCTION ................................................................................................................. 5
1-1. General......................................................................................................................................... 5
1-2. Power Requirements.................................................................................................................... 5
1-3. Pendant Control Unit .................................................................................................................... 6
1-4. Foot Control Unit .......................................................................................................................... 6
SECTION II OPERATION ....................................................................................................................... 7
2-1. Electrical Power ........................................................................................................................... 7
2-2. Battery Power .............................................................................................................................. 7
2-3. Automatic Shut-Off....................................................................................................................... 7
2-4. Charging the Battery .................................................................................................................... 8
2-5. Positioning Functions ................................................................................................................... 8
a. Speed Selection ..................................................................................................................... 8
b. Floor Lock/Brake system .......................................................................................................9
c. Trendelenburg ...................................................................................................................... 10
d. Lateral Tilt ............................................................................................................................. 10
e. Back Section ........................................................................................................................ 11
f. Elevation............................................................................................................................... 12
g. Return To Level.................................................................................................................... 12
2-6. Foot Control................................................................................................................................ 12
2-7. Head Section.............................................................................................................................. 13
2-8. Intermediate Back Section ......................................................................................................... 13
2-9. Leg Section. ............................................................................................................................... 14
SECTION IV MAINTENANCE .............................................................................................................. 16
3-1. Preventive Maintenance............................................................................................................. 16
3-2. Cleaning Recommendations ...................................................................................................... 16
10-04
1
SPECIAL USER ATTENTION
The extreme positioning capabilities of the 7200 Series Table requires special attention for possible interference points when using multiple function positioning. As with the operation of any surgical table, a certain amount of care should be exercised to position the patient safely. Although the thick pads and sheets substantially protect the patient, pinch points, located at the joints of the top section should always be considered. BE SURE THAT THE ARMS, HANDS AND FINGERS OF THE PATIENT AND THOSE OF THE OPERATING ROOM PERSONNEL ARE CLEAR OF ALL MOV­ING PARTS BEFORE MOVING THE TABLE. Proper restraints should always be used for patient safety.
Certain accessories such as the Armboards and X-Ray top can be damaged when changing the position of the table top sections. Always look first to see if a desired movement is going to interfere with any accessories in use.
The operator has the ultimate responsibility of preventing damage to the table and surrounding equipment or possible injury to the patient or staff.
The operator must ensure proper positioning is maintained to prevent compromizing respiration, nerve pathways or circulation. In general, common sense will dictate when there is a potential hazard.
The following precautions should be reviewed by all personnel prior to operating the table.
A routine Instructional Program should be implemented by the facility for proper usage instructions for all personnel that may operate this table.
WARNING
Indicates a possibility of personal injury.
WARNING
Prior to operating the table, observe all table caution labels and review the SPECIAL USER ATTENTION sec­tion in the front of this manual.
WARNING
Possible explosion hazard exists if table is used in the presence of FLAMMABLE ANESTHETICS.
NOTE
Pressing the Main Power Switch will turn the power to the table off.
NOTE
If the table is stored for a period greater than 6 months, the batteries should be removed and stored in a dry, clean con­dition at a storage temperature of 68° F (20° C). Batteries should be recharged every 6 months of product storage.
NOTE
When the indicator starts to blink (indi­cating low power in battery) the table will operate for approximately 5 con­tinuous minutes, typically long enough to use the table for the rest of the day.
NOTE
The charging system operates ONLY when the table is in AC120V operation mode. The table can be operated on 120VAC power while the battery is being re­charged.
CAUTION
Indicates a possibility of damage to equipment.
NOTE
Indicates important facts or helpful hints.
2
WARNING
DO NOT unlock brakes when a patient is on the table. An uneven patient weight load may cause instability. Po­tential pinch and/or crush hazard exists
SPECIAL USER ATTENTION
WARNING
To maximize patient safety, utilize proper restraint methods during ex­treme Trendelenburg positioning modes.
WARNING
To avoid injury to surgical staff or dam­age to the table, use extreme care when attempting full Trendelenburg positioning while the table top is in a low position. Some table positions will allow the head section of the table to contact the floor.
WARNING
To maximize patient safety, utilize proper restraint methods during ex­treme lateral tilt positioning.
WARNING
Potential pinch and/or crush hazard exists if hands or fingers are placed between the edges of seat and leg sections when leg sections are raised or lowered.
CAUTION
Do not sit on end of leg sections as damage may result.
WARNING
Consult manufacturer's instructions when using high frequency surgical equipment, cardiac defibrillator and car­diac defibrillator monitors.
WARNING
WARNING
Potential pinch and/or crush hazard exists if hands or fingers are placed between edges of back and seat secrion frames or side rails during extreme back section positioning.
WARNING
Potential pinch and/or crush hazard exists if hands or fingers are placed in gear mechanism (under back section) during back section positioning.
WARNING
Potential pinch and/or crush hazard exists if hands or fingers are placed between elevation sections or eleva­tion sectons and base during elevation­down positioning.
When an antistatic pathway is required, the table has to be used on an antistatic floor.
WARNING
The antistatic properties of the table are dependent on the use of the original pad set which was furnished with the table or an alternate approved replacement.
NOTE
Always follow current AORN Journal Guidelines to ensure proper cleaning and disinfection procedure.
3
SPECIAL USER ATTENTION
WARNING
Always follow OSHA blood-borne patho­gens standards for protective clothing, including gloves, masks and eye pro­tection when cleaning the surgical table.
CAUTION
Thoroughly read and follow the manufacturer's directions for all clean­ing fluids. DO NOT use cleaners con­taining phenolics.
CAUTION
When using spray cleaners DO NOT spray fluids directly into electrical re­ceptacles or micro switches.
CAUTION
Before replacing pads on the table, make sure the pads and all mating surfaces are completely dry. Moisture trapped between the pads and mating surfaces may cause distortion of table tops.
4
Loading...
+ 14 hidden pages