Mizuho Osi 6844 User manual

6844 INSITE Surgery Table T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
Preventive Maintenance (PM) Checklist
INSITE Surgery Table
Work Instruction number: 50603-15 Rev. A

Contact Information:

Date: ________________Model/Serial#: _______________________ Account #: ____________________ Hospital Name: _________________________________ City & State: ________________________________ Contact Name: _______________________ Position: ___________________ Phone: _____________________ Equipment Location: ________________________________________________________________________ Service Representative: _____________________ Service Organization: _______________________________ Service Rep. phone: ________________________ Home location: ____________________________________
Customer contact accepting service: ______________________ __________________ ____________ Name please print Signature Date
Service Representative: ________________________ _______________ ___________________ Name please print Signature Date
Using this report: Please proceed through each step and place an “x” in the “( )” as each item representing Pass Fail
( ) ( ) criteria is inspected. If a step has failed, please indicate with an “x” and note the reasons in the comments
section. If you resolve the failure, place an “x” in the “Pass” area and indicate the corrective action again in the comments section. If you are unable to perform any Pass/Fail check, please contact Mizuho OSI Technical Services for assistance and circle the check.
Biohazards when servicing equipment: Please take appropriate precautions for servicing potentially biohazard contaminated equipment or working in biohazard surroundings. Wear protective surgical gloves and long sleeved gowns at all times. If your gloves or protective clothing tears, replace immediately. Have the hospital staff clean the equipment appropriately prior to servicing. Clean your tools that may become contaminated with biohazards prior to placing back in your tool kit for removal from the location. If you suspect biohazards are present, do not handle until they have been cleaned.
ESD Precautions: Electro-Static Discharge can damage sensitive electronic printed circuit boards. If you are servicing components that are located near or in contact with PCBs or other sensitive devices, please ground yourself with a proper ESD strap.
(If serviced by an outside service organization, the record must be reviewed, signed, and dated by Technical Services. Refer to last page of this document).
50603-15 Rev. A ECN# 100249 Eff Date: 07/20/2012 Page 1 of 5
6844 INSITE Surgery Table T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
Tools Required:
Normal handtools
Materials Required:
NW0635 Rev B. 6844 – INSITE Surgery Table Owner’s Manual
Pass

1. General Overall Condition

1.1 ( ) ( ) Table base is clean, without signs of hazardous contamination, and is consistent with proper care.

Fail
Comments: _________________________________________________________________________
1.2 ( ) ( ) All labels are intact and legible.
Verify attachment to base cover of “Warning – Not for Storage” label. If possible, advise customer of potential table damage if items are stored on this surface.
Comments: _________________________________________________________________________

2. Power Input Inspection

2.1 ( ) ( ) Power Cord is free of wear, corrosion, and insulation cracks. Comments: _________________________________________________________________________
2.2 ( ) ( ) Power Cord has OSI tag (indicating “Hospital Grade”) attached. Comments: _________________________________________________________________________
2.3 ( ) ( ) Power Cord Base Receptacle is properly secured and operational. Comments: __________________________________________________________________________________________________________________________________________________
2.4 ( ) ( ) ON/OFF Power Switch/Circuit Breaker is properly secured; plastic cover is not torn or pierced. Comments: _________________________________________________________________________
3. Basic Controls
3.1 ( ) ( ) Power Switch ON
a. Connect the Power Cord to AC power source and also to a power receptacle. b. Toggle Power Switch to “ON” (located on table-base panel). If illuminated, table is connected
to live AC power. If not illuminated, ta ble may still operate on internal battery power (Hand Pendant buttons and lights will be illuminated).
c. Verify that the Three Circuit Breakers (located on table-base panel) are not tripped. d. Verify that the Hand Pendant Battery Status “OK” LED illuminates green.

Comments: _________________________________________________________________________

50603-15 Rev. A ECN# 100249 Eff Date: 07/20/2012 Page 2 of 5