Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
Andrews Spinal Surgery Table, Manual
Models: 5820
Work Instruction Number 50603-13
Contact Information:
Date: ________________Model/Serial#: _______________________ Account #: ____________________ Hospital Name: ________________________ _________ City & State: ________________________________ Contact Name: _______________________ Position: ___________________ Phone: _____________________ Equipment Location: ___________________________ ____________________________________ _________ Service Representative: _____________________ Service Organization: ____________________________ ___ Service Rep. phone: ________________________ Home location: ____________________________________
Customer contact accepting service: ______________________ __________________ ____________ Name please print Signature Date
Service Representative: ________________________ _______________ ________________ ___ Name please print Signature Date
Using this report: Please proceed through each step and place an “x” in the “( )” as each item representing Pass Fail ( ) ( ) criteria is inspected. If a step has failed, please indicate with an “x” and note the reasons in the comments section. If you resolve the failure place an “x” in the “Pass” area and indicate the corrective action again in the comments section. If you are unable to perform any Pass/Fail check, please contact Mizuhosi Technical Services for assistance and circle the check
Biohazards when servicing equipment: Please take appropriate precautions for servicing potentially contaminated biohazard equipment or working in biohazard surroundings. Please wear protective surgical gloves and log sleeved gowns at all times. If your gloves or protective clothing tears, replace immediately. Have the hospital staff clean the equipment appropriately prior to servicing. Clean your tools that may contain biohazards prior to placing back in your tool kit for removal from the location. If you suspect biohazards are present do not service the equipment until they have been cleaned.
(If serviced by an outside service organization, the record must be reviewed, signed, and dated by Technical Services Dept. Refer to last page of this document).
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 1 of 6
Page 2
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
Surgical bases covered:
5820, Andrews Table
Tools Required:
Digital Volt Multimeter (DVM) with probe and hook leads Grease, Lithium, white, multi-purpose Hex keys, (Allen), ball head, minimum handle length 4 inch (in): 5/64 in, 5/32 in, 3/16 in, 5/16 in. Wrench, open end, 3/8 in. Philip Driver Screw Driver (medium size) Mirror, small inspection. Thread-lock, non-permanent, Loctite, blue, or equiv. Surgical gloves.
Materials Required:
Andrews SST-3000 Spinal Surgery Table User Guide Andrews Spinal Surgery Table Maintenance and Repair Manual -
Pass Fail
1. General Overall Condition.
1.1 ( ) ( ) Table is clean, without signs of hazardous contamination, is consistent with proper care. Comments: _________________________________________________________________________
1.2 ( ) ( ) All sheet metal covers are secure without missing fasteners. Comments: _________________________________________________________________________
1.3 ( ) ( ) Sheet metal covers and paint are free of dents or scratches. Comments: _________________________________________________________________________
1.4 ( ) ( ) All labels and control panel are intact and legible. Comments: _________________________________________________________________________
1.5 ( ) ( ) Table Pads retain good shape, are free of punctures, tears or excessive wear.
1. Torso Pads, left and right.
2. Torso Lift Pad.
3. Lateral (Hip) Support Pads (2x).
4. Seat Pad.
5. Thigh (Femoral) Pad.
6. Tibial Pad.
7. “Head Support” pad.
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 2 of 6
Page 3
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
8. Arm Board Pads (2x). Comments: _________________________________________________________________________
1.6 ( ) ( ) Slings and Straps are free of excessive wear or tears.
1. Hip Sling
2. Thigh Strap Comments: _________________________________________________________________________
2. All Casters Operate Properly.
2.1 ( ) ( ) Casters rotate and roll freely and smoothly full 360 degrees. Comments: _________________________________________________________________________
2.2 ( ) ( ) Caster do not shake or have abnormal uniformity. Comments: _________________________________________________________________________
2.3 ( ) ( ) Rubber wheels are not worn or damaged. Comments: _________________________________________________________________________
2.4 ( ) ( ) Foot end casters lock and remain engaged until unlocked. Comments: _________________________________________________________________________
2.5 ( ) ( ) Head end casters lock from rolling and rotating and remain engaged when caster lock pedal is depressed until released.
Comments: _________________________________________________________________________
3. Hydraulic System
Caution
: The following requires removal of protective cover(s). Please refer to Maintenance and
Repair manual.
3.1 ( ) ( ) Hydraulic fluid level is approximately 1and 1/2 inches from reservoir top. Add AW Hydraulic oil,
viscosity 32 or equivalent, as needed. Comments: _________________________________________________________________________
3.2 ( ) ( )With all control valves in neutral (off) position, pump pedal. No motion of the table occurs.
Comments: _________________________________________________________________________
4. Torso Lift function.
4.1 ( ) ( ) Switch control valve to UP/DOWN/ position, pump the pedal. Table moves smoothly and quietly
through its full range. Comments: _________________________________________________________________________
4.2 ( ) ( ) Switch control valve to center (neutral) position, pump pedal. No motion of the table occurs.
Comments: _________________________________________________________________________
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 3 of 6
Page 4
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
4.3 ( ) ( ) Inspect Torso Lift assembly Chest Board for stress cracks or breakage.
Comments: _________________________________________________________________________
5. Tibial Support function.
5.1 ( ) ( ) Switch control valve to UP/DOWN position, pump pedal, table moves smoothly and quietly through its
full range.
Comments: _________________________________________________________________________
5.2 ( ) ( ) Switch control valve to center (neutral) position, pump pedal. No motion of the table occurs.
Comments: ___________________________ ______________________________________________
6. Lateral Tilt function.
6.1 ( ) ( ) Switch control valve to PT’S RIGHT/PT’S LEFT position, pump pedal. Table moves smoothly and
quietly through its full range. Comments: _________________________________________________________________________
6.2 ( ) ( ) Switch control valve to center (neutral) position, pump pedal. No motion of the table occurs.
Comments: _________________________________________________________________________
7. Table Height function.
7.1 ( ) ( ) Switch control valve to UP/DOWN position, pump pedal. Table moves smoothly and quietly through
its full range. Comments: _________________________________________________________________________
7.2 ( ) ( ) Switch control valve to center (neutral) position, pump pedal. No motion of the table occurs.
Comments: _________________________________________________________________________
8. Trendelenberg/Reverse Trendelenburg function.
8.1 ( ) ( ) Switch control valve to TRENDEL/REV. TRENDEL position. Table moves smoothly and quietly
through its full range. Comments: _________________________________________________________________________
8.2 ( ) ( ) Switch control valve to center (neutral) function, pump pedal. No motion of the table occurs.
Comments: ________________________ _________________________________________________
9. Foot End function.
9.1 ( ) ( ) Switch control valve to UP/DOWN position, pump the pedal. Table moves smoothly and quietly
through its full range. Comments: _________________________________________________________________________
9.2 ( ) ( ) Switch control lever to center (neutral) position, pump the pedal. No motion of the table occurs.
Comments: _________________________________________________________________________
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 4 of 6
Page 5
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
9.3 ( ) ( ) Inspect tibial support lock handle for ease of engagement. Lock handle indicator pin should align with
red arrow on casting. Ensure that the support rods are fully engaged into the table. Comments: _________________________________________________________________________
9.4 ( ) ( ) Inspect tibial Lock pin tips (white nylon) for wear and tear.
Comments: _________________________________________________________________________
10. Foot Assembly operates properly.
10.1( ) ( ) Leg restraint foot extensions are easily adjustable and locking detents and heel clamps
engage and lock securely. Comments: _________________________________________________________________________ Note: Tables prior to s/n 922 and manufactured prior to 4’99 use Keepers (Heel Cup Locks).
10.2( ) ( ) Leg Restraints (Heel Cups) are free of cracks or tears.
Comments: _________________________________________________________________________
Seat Assembly operates properly.
1111..
11.1( ) ( ) Seat post and seat upright posts are free of wear or scoring. Lock handles secures seat firmly.
Comments: _________________________________________________________________________
11.2( ) ( ) Lateral (hip) support pads (2x) adjust easily and lock handles secure firmly. Lock pins at end of rods
are in place. Comments: _________________________________________________________________________
12. Autoglide Torso Lift operates properly.
12.1( ) ( ) Autoglide Torso Lift lock handle disengages Autoglide and easily allows full travel without
binding. Comments: _________________________________________________________________________
12.2( ) ( ) Autoglide Torso Lift lock handle engages Autoglide and easily locks movement.
Comments: _________________________________________________________________________
13. Arm Boards inspection only if attached to the table.
13.1( ) ( ) Armboards attach easily to siderails and lock securely.
Comments: _________________________________________________________________________
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 5 of 6
Page 6
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
This section is for Technical Services use only.
Reviewed: _________________________ Signature: ________________________ Date: _________
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 6 of 6