Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t
Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
Andrews Spinal Surgery Table, Manual
Models: 5820
Work Instruction Number 50603-13
Contact Information:
Date: ________________Model/Serial#: _______________________ Account #: ____________________
Hospital Name: ________________________ _________ City & State: ________________________________
Contact Name: _______________________ Position: ___________________ Phone: _____________________
Equipment Location: ___________________________ ____________________________________ _________
Service Representative: _____________________ Service Organization: ____________________________ ___
Service Rep. phone: ________________________ Home location: ____________________________________
Customer contact accepting service: ______________________ __________________ ____________
Name please print Signature Date
Service Representative: ________________________ _______________ ________________ ___
Name please print Signature Date
Using this report: Please proceed through each step and place an “x” in the “( )” as each item representing Pass
Fail ( ) ( ) criteria is inspected. If a step has failed, please indicate with an “x” and note the reasons in the
comments section. If you resolve the failure place an “x” in the “Pass” area and indicate the corrective action again
in the comments section. If you are unable to perform any Pass/Fail check, please contact Mizuhosi Technical
Services for assistance and circle the check
Biohazards when servicing equipment: Please take appropriate precautions for servicing potentially contaminated
biohazard equipment or working in biohazard surroundings. Please wear protective surgical gloves and log sleeved
gowns at all times. If your gloves or protective clothing tears, replace immediately. Have the hospital staff clean the
equipment appropriately prior to servicing. Clean your tools that may contain biohazards prior to placing back in
your tool kit for removal from the location. If you suspect biohazards are present do not service the equipment until
they have been cleaned.
(If serviced by an outside service organization, the record must be reviewed, signed, and dated by Technical
Services Dept. Refer to last page of this document).
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 1 of 6
Page 2
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t
Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
Surgical bases covered:
5820, Andrews Table
Tools Required:
Digital Volt Multimeter (DVM) with probe and hook leads
Grease, Lithium, white, multi-purpose
Hex keys, (Allen), ball head, minimum handle length 4 inch (in): 5/64 in, 5/32 in, 3/16 in, 5/16 in.
Wrench, open end, 3/8 in.
Philip Driver
Screw Driver (medium size)
Mirror, small inspection.
Thread-lock, non-permanent, Loctite, blue, or equiv.
Surgical gloves.
Materials Required:
Andrews SST-3000 Spinal Surgery Table User Guide
Andrews Spinal Surgery Table Maintenance and Repair Manual -
Pass Fail
1. General Overall Condition.
1.1 ( ) ( ) Table is clean, without signs of hazardous contamination, is consistent with proper care.
Comments: _________________________________________________________________________
1.2 ( ) ( ) All sheet metal covers are secure without missing fasteners.
Comments: _________________________________________________________________________
1.3 ( ) ( ) Sheet metal covers and paint are free of dents or scratches.
Comments: _________________________________________________________________________
1.4 ( ) ( ) All labels and control panel are intact and legible.
Comments: _________________________________________________________________________
1.5 ( ) ( ) Table Pads retain good shape, are free of punctures, tears or excessive wear.
1. Torso Pads, left and right.
2. Torso Lift Pad.
3. Lateral (Hip) Support Pads (2x).
4. Seat Pad.
5. Thigh (Femoral) Pad.
6. Tibial Pad.
7. “Head Support” pad.
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 2 of 6
Page 3
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t
Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
8. Arm Board Pads (2x).
Comments: _________________________________________________________________________
1.6 ( ) ( ) Slings and Straps are free of excessive wear or tears.
2.1 ( ) ( ) Casters rotate and roll freely and smoothly full 360 degrees.
Comments: _________________________________________________________________________
2.2 ( ) ( ) Caster do not shake or have abnormal uniformity.
Comments: _________________________________________________________________________
2.3 ( ) ( ) Rubber wheels are not worn or damaged.
Comments: _________________________________________________________________________
2.4 ( ) ( ) Foot end casters lock and remain engaged until unlocked.
Comments: _________________________________________________________________________
2.5 ( ) ( ) Head end casters lock from rolling and rotating and remain engaged when caster lock
pedal is depressed until released.
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 3 of 6
Page 4
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t
Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
4.3 ( ) ( ) Inspect Torso Lift assembly Chest Board for stress cracks or breakage.
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 4 of 6
Page 5
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t
Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
9.3 ( ) ( ) Inspect tibial support lock handle for ease of engagement. Lock handle indicator pin should align with
red arrow on casting. Ensure that the support rods are fully engaged into the table.
Comments: _________________________________________________________________________
9.4 ( ) ( ) Inspect tibial Lock pin tips (white nylon) for wear and tear.
10.1( ) ( ) Leg restraint foot extensions are easily adjustable and locking detents and heel clamps
engage and lock securely.
Comments: _________________________________________________________________________
Note: Tables prior to s/n 922 and manufactured prior to 4’99 use Keepers (Heel Cup Locks).
10.2( ) ( ) Leg Restraints (Heel Cups) are free of cracks or tears.
50603-13 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 5 of 6
Page 6
Andrews Surgical Table T e c h n i c a l S e r v i c e s D e p a r t m e n t
Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324