Modular Surgical Tabletop T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
Imaging Spinal Surgical Tabletop
Models: 5827, 5927
Work Instruction Number 50603-12
Contact Information:
Date: ________________Model/Serial#: _______________________ Account #: ____________________ Hospital Name: ________________________ _________ City & State: ________________________________ Contact Name: _______________________ Position: ___________________ Phone: _____________________ Equipment Location: ___________________________ ____________________________________ _________ Service Representative: _____________________ Service Organization: ____________________________ ___ Service Rep. phone: ________________________ Home location: ____________________________________
Customer contact accepting service: ______________________ __________________ ____________ Name please print Signature Date
Service Representative: ________________________ _______________ ________________ ___ Name please print Signature Date
Using this report: Please proceed through each step and place an “x” in the “( )” as each item representing Pass Fail ( ) ( ) criteria is inspected. If a step has failed, please indicate with an “x” and note the reasons in the comments section. If you resolve the failure place an “x” in the “Pass” area and indicate the corrective action again in the comments section. If you are unable to perform any Pass/Fail check, please contact Mizuhosi Technical Services for assistance and circle the check.
Biohazards when servicing equipment: Please take appropriate precautions for servicing potentially contaminated biohazard equipment or working in biohazard surroundings. Please wear protective surgical gloves and log sleeved gowns at all times. If your gloves or protective clothing tears, replace immediately. Have the hospital staff clean the equipment appropriately prior to servicing. Clean your tools that may contain biohazards prior to placing back in your tool kit for removal from the location. If you suspect biohazards are present do not service the equipment until they have been cleaned.
(If serviced by an outside service organization, the record must be reviewed, signed, and dated by Technical Services Dept. Refer to last page of this document).
50603-12 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 1 of 3
Modular Surgical Tabletop T e c h n i c a l S e r v i c e s D e p a r t m e n t Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
Surgical bases covered:
5827 Radiolucent Imaging Top, 350lb (pounds) patient weight capacity 5927 Radiolucent Imaging Top, 500lb patient weight capacity
Tools Required:
Cotton swab Lubricant, silicon, spray Grease, Lithium bases, white, multi-purpose
Materials Required:
JJaacckkssoonn SSppiinnaall SSuurrggeerryy && IImmaaggiinngg TTaabbllee TTooppss TTeecchhnniiq
Pass Fail
1. Top Inspection and fit.
1.1 ( ) ( ) Table top is clean, without signs of hazardous contamination, and all labels are intact and legible. Comment: __________________________________________________________________________
1.2 ( ) ( ) Carbon fiber top is free of wear, cracks, delamination or splinters. Particular inspection is required of the siderail-to-top joint. Joint is securely bonded without voids, cracks, or separation.
Comments: _________________________________________________________________________ Diagnostic/repair tip: splinters are easily detected by running a cotton swab along surface. Smooth
splinter area with a light sanding with 400 grit sandpaper then applying a small amount of Polycyanuacyralate (“super-glue”) to seal affected area.
1.3 ( ) ( ) Gimbals are secure, rotate to stops and return to center position. Comments: _________________________________________________________________________ Note: earlier models may not be equipped with gimbals.
1.4 ( ) ( ) Fits in all positions on removable H-frames, T-pins slide into locked position freely. Comments: _________________________________________________________________________
quuee MMaannuuaall
Warning
will cause severe damage.
1.5 ( ) ( ) With top in 3 and Reverse-Trendelenburg.
Comments: _________________________________________________________________________
1.6 ( ) ( )With top in lowest position on H-frames, top rotates freely at 5 degree Trendelenburg and Reverse­ Trendelenburg. Comments: _________________________________________________________________________
50603-12 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 2 of 3
: Rotation of tops not equipped with gimbals beyond 5 degree Trendelenburg and Reverse-Trendelenburg
rd
from highest position on H-frames, top rotates freely at 5 degree Trendelenburg