The information in this document is subject to change without notice and should not be construed as a commitment by Millipore Corporation. Millipore Corporation assumes no responsibility for any errors that may appear in this document. This manual is believed to be complete and accurate at the time of publication. In no event shall Millipore Corporation be liable for incidental or consequential damages in connection with or arising from the use of this manual.
Copyright 2002, Millipore Corporation. All rights reserved. This book or parts thereof may not be reproduced in any form without the written permission of the publishers. Printed in U.S.A.
Millipore and M Air T are registered trademarks of Millipore Corporation. Tri-Clover is a registered trademark of Alfa Laval, Inc.
P36483, Rev. –, 10/02
Introduction
The new M Air T® Compressed Gas and Air Sampling Kit is designed especially to monitor the microbiological quality of compressed gas and air used in the pharmaceutical and food & beverage industries.
Connected to the M Air T Isolator Pump ATBPUMP01, the new sampling kit allows accurate sampling and microbiological monitoring of a compressed gas sample, by direct impaction onto the M Air T pre­filled agar cassette.
The pre-sterilized M Air T agar cassettes used with the M Air T Compressed Gas and Air Sampling Kit are easier to handle, especially when wearing gloves, significantly reducing the risk of inadvertent contamination. The M Air T Compressed Gas and Air Sampling Kit is manufactured from stainless steel, with sanitary Tri-Clover connections, making it autoclavable.
The design of the M Air T Compressed Gas and Air Sampling Kit allows one to sample compressed gas or air up to 5 bar (73 psi) of pressure, without any pressure reducer.
Applications
The M Air T Compressed Gas and Air Sampling Kit, connected to the M Air T Isolator Pump, is suited for a variety of applications, as fol­lows.
®
Pharmaceutical
Evaluation of the microbiological quality of compressed gas or air
used for blanketing, purging, sparging or stirring products
Evaluation of the microbiological quality of confined areas such as
Blow Fill Seal (BFS) sterile showers
Food and Beverage
Evaluation of the microbiological quality of compressed gas or air
used for blanketing, purging, cooling and freezing products
2 www.millipore.com
Component Diagram
Item Description Function
A Pressure reducer Connected to the sampling port, to reduce
incoming pressure B Clamp, 1.5 in. TC To connect pressure reducer to cone adapter C Cone adapter To reduce charge loss and to expand impaction
onto the entire agar surface D Clamp, 4 in. TC To connect cone adapter to impaction sieve E Impaction sieve Captures microorganisms using 611 micro-
perforations
A
B
C
D
E
Precautions/Limitations
Autoclave all parts at 121 °C for 20 minutes before any compressed gas or
air sampling.
Use only with the M Air T Isolator Pump.
Use only M Air T cassettes with this unit.
Do not use for microbiological monitoring of compressed O
Do not sample gas or air at pressures above the specified limit of
5 bar (73 psi).
.
2