Mobile unit E 327 for surgical instruments

Items supplied:

Mobile unit E 327 with 2 levels, H 502, W 535, D 515 mm, with a holder for the magnet strip used in automatic mobile unit recognition (G 7835 and G 7836)

Accessories available to order:

Magnet strip ML/2 for automatic mobile unit recognition

E 142 DIN mesh tray

E 476 holders for micro instruments

E 473 insert for very small items

,Please read the Warning and Safety instructions in the Operating Instructions for your washer-disinfector carefully. You should only process instruments/medical products in this mobile unit if they have been declared as suitable for machine processing by the manufacturer. Please also observe the manufacturer's care instructions. The programme selected must be suitable for the materials from which the medical products to be processed are made. Where possible the programme should include thermal disinfection. The final rinse should, if possible, always be carried out with fully de-mineralised water. If using automatic mobile unit recognition please make sure that the coding is correctly set.

^Hinged instruments must be opened up before being placed in the mesh trays to minimise the risk of surfaces not being accessed during processing.

^To prevent the risk of light instruments such as bulldog clips and other small items being moved about during processing, silicone holders e.g. E 473 and closable inserts e.g. E 476 can be used to keep them secure.

Areas of application:

The E 327 mobile unit is suitable for use in the following washer-disinfectors: G 7882, G 7882 CD, G 7835 and G 7836.

It enables surgical instruments to be cleaned and disinfected in 4 DIN mesh trays. 2 DIN mesh trays can be accommodated on each level.

Loading height (levels numbered from the bottom upwards) [H x W x D in mm]

Level 1

112 x 520 x 510

Level 2

105 x 512 x 480

Before loading the machine and before starting a programme please check:

that the spray arms rotate freely,

that the spray arm jets are clean and free of any blockages,

that the mobile unit is correctly fitted into the cabinet and that it has connected correctly with the water intake.

For validation purposes please make sure that you follow the loading instructions given on the template.

^Containers e.g. beakers and kidney dishes should be placed in the mobile unit, preferably on the top level, with the open side facing downwards, ensuring that all surfaces can be accessed easily and that cleaning solution can flow freely off them. They must also be secure so that they cannot be flipped over by the force of water during processing.

If hollow items are placed such that cleaning solution can collect in recesses etc. they will not get properly clean and rinsing results will be impaired.