Safety First: The primary concern of Midmark
Corporation is that this sterilizer is maintained with
the safety of the patient and staff in mind. To assure
that services and repairs are completed safely and
correctly, proceed as follows:
(1) Read this entire manual before performing any
services or repairs on this sterilizer.
(2) Be sure you understand the instructions
contained in this manual before attempting to
service or repair this sterilizer.
Safety Alert Symbols
Throughout this manual are safety alert symbols that
call attention to particular procedures. These items are
used as follows:
DANGER
A DANGER is used for an imminently
hazardous operating procedure,
practice, or condition which, if not correctly
followed, will result in loss of life or serious
personal injury.
NOTE
A NOTE is used to amplify an operating procedure,
practice or condition.
Warranty Instructions
Refer to the Midmark “Limited Warranty” printed in the
Installation and Operation Manual for warranty information. Failure to follow the guidelines listed below will
void the warranty and/or render the M9 sterilizer
unsafe for operation.
• In the event of a malfunction, do not attempt to
operate the sterilizer until necessary repairs have
been made.
• Do not attempt to disassemble sterilizer, replace
malfunctioning or damaged components, or perform
adjustments unless you are one of Midmark’s
authorized service technicians.
• Do not substitute parts of another manufacturer
when replacing inoperative or damaged components.
Use only Midmark replacement parts.
WARNING
A WARNING is used for a potentially
hazardous operating procedure,
practice, or condition which, if not correctly
followed, could result in loss of life or serious
personal injury.
CAUTION
A CAUTION is used for a potentially
hazardous operating procedure, practice,
or condition which, if not correctly followed, could
result in minor or moderate injury. It may also be
used to alert against unsafe practices.
EQUIPMENT ALERT
An EQUIPMENT ALERT is used for an
imminently or potentially hazardous
operating procedure, practice, or condition which, if
not correctly followed, will or could result in serious,
moderate, or minor damage to unit.
This manual contains detailed troubleshooting, scheduled maintenance, maintenance, and service instructions for the M9 Sterilizers. This manual is intended to
be used by Midmark’s authorized service technicians.
1.2How to Use Manual
A. Manual Use When Performing Scheduled Mainte-
nance.
(1) Perform inspections and services listed in
Scheduled Maintenance Chart (Refer to
para 3.1).
(2) If a component is discovered to be faulty or out
of adjustment, replace or adjust component in
accordance with Maintenance/Service Instructions (Refer to para 4.1).
B. Manual Use When Sterilizer Is Malfunctioning And
Cause Is Unknown.
(1) Perform an operational test on sterilizer (Refer
to para 2.1).
(2) Perform troubleshooting procedures listed in
Troubleshooting Guide (Refer to para 2.2).
(3) If a component is discovered to be faulty or out
of adjustment, replace or adjust component in
accordance with Maintenance/Service Instructions (Refer to para 4.1).
C. Manual Use When Damaged Component Is Known.
(1) Replace or adjust component in accordance
with Maintenance/Service Instructions (Refer to
para 4.1).
1.3Description Of Sterilizer
A. General Description (See Figure 1-1).
The M9 Sterilizers are designed to sterilize instruments.
The major components of the sterilizer consist of a
pressure vessel, condensing tank assembly, heating
element, temperature sensor assembly, condensing
tank water level sensor (early M9 units only), pressure
vessel water level sensor, pulse solenoid, thermostat,
door switch, control PC board, fill solenoid, vent solenoid, bellows assembly, display PC board, and a
pressure relief valve.
B. Theory of Operation (See Figures 5-1 and 5-2).
Theory Of Operation:
Power is supplied to the control PC board when the
sterilizer is plugged in. At this time, the control PC
board will flash the error lamp and display “E 001”. This
is normal and occurs when the sterilizer is first plugged
in or if a power outage has occurred.
When the ON/STANDBY switch is pressed, the error
lamp extinguishes and the program switches are
enabled. The program lamps flash and the ON/
STANDBY indicator lamp illuminates.
The operator determines the desired cycle and selects
the cycle by pressing either: UNWRAPPED,
POUCHES, LIQUIDS, OR PACKS button. Upon
selecting the desired cycle, the control PC board sets
the specific time and temperature parameters for that
cycle into memory. These settings are alternately
displayed on the TIME/TEMP display. Note: The
control PC board will continue to use the same parameters on subsequent cycles until a new type of cycle is
selected.
The operator presses the START button which signals
the control PC board to initiate the cycle.
The control PC board continuously monitors for an open
door condition from the time the START button is
pressed until the start of the drying cycle. It does this
by monitoring the door switch for continuity. If no
continuity is detected after a cycle has been started, the
control PC board will stop the cycle and illuminate the
DOOR AJAR indicator. The DOOR AJAR indicator will
continue to flash until continuity is sensed or until the
STOP or ON/STANDBY button is pressed. Or, if the
door is closed and continuity is sensed, the control PC
board continues the cycle.
On early units (units with an old style control PC board
which has EPROM version M or before) which have a
condensing tank water level sensor, the control PC
board monitors the condensing tank water level when
the sterilizer door is closed. The control PC board
monitors the water level in the condensing tank during
this portion of the cycle only, and it performs this
monitoring by checking for continuity between the
condensing tank water level sensor and the chassis.
Water touches the sensor to complete a ground circuit,
signaling the control PC board that the water level is
sufficient and the cycle may be continued. Otherwise,
the control PC board will stop the cycle and flash the
WATER LOW indicator until water is sensed by the
sensor, the STOP button is pressed, or the ON/
STANDBY button is pressed. On later units (units with
old style control PC board which has EPROM version N
or after or new style control PC board) which do not
have the condensing tank water level sensor, the
control PC board monitors the pressure vessel water
level sensor to determine if a low water condition in the
condensing tank exists. If the pressure vessel water
CONDENSING TANK
WATER LEVEL SENSOR
(early M9 only)
TEMPERATURE
SENSOR ASSEMBLY
PRESSURE VESSEL
WATER LEVEL SENSOR
VENT
SOLENOID
MA207604
level sensor does not sense water within 5 minutes of
starting the FILLING portion of the cycle, the control PC
board stops the cycle, illuminates the WATER LOW
indicator and starts sounding a beeper signal.
The control PC board illuminates the FILLING indicator
and energizes the fill solenoid which allows water to
flow from the condensing tank to the pressure vessel.
The control PC board controls the water level in the
pressure vessel by monitoring the pressure vessel
water level sensor for continuity. When water completes the ground circuit between the pressure vessel
water level sensor and the chassis, the control PC
board de-energizes the fill solenoid.
The control PC board extinguishes the FILLING indicator lamp and illuminates the STERILIZING indicator
lamp. During this portion of the cycle, temperature and
pressure in the pressure vessel are continuously
monitored until the drying portion of the cycle begins. A
precision integrated circuit temperature sensor potted
inside a brass thermowell is used to monitor the tem-
perature inside the pressure vessel. A PC board
mounted piezoresistive pressure sensor, connected by
plumbing to the pressure vessel, is used to monitor the
pressure in the pressure vessel.
The control PC board energizes the heating element to
begin heating the water in the pressure vessel. The
temperature in the pressure vessel is displayed on the
display PC board. As the water begins to boil, air is
bled off through the bellows assembly into the condensing tank. The pressure in the pressure vessel is not
displayed until the temperature in the pressure vessel
reaches 208 °F (98 °C). When the bellows assembly
senses pure steam flowing through it (which is around
215 °F), the valve in the bellows assembly closes,
allowing the pressure in the pressure vessel to build.
When the temperature corresponding to the cycle
selected is reached, the control PC board de-energizes
the heating element and begins to count down the
remaining time left in the cycle. The remaining time is
now displayed on the display PC board instead of the
temperature. During this portion of the cycle, the
control PC board regulates the pressure vessel temperature/pressure by energizing and de-energizing the
heating element as necessary. The programmed
temperatures are the minimum values, and the temperature during the cycle is regulated approximately 2
°F (1.1 °C) above those minimum values.
When the cycle’s set time has elapsed, the control PC
board de-energizes the heating element and energizes
the vent solenoid, unless the liquids cycle was selected.
If the liquids cycle was selected, the control PC board
de-energizes the heating element, but does not energize the vent solenoid until the temperature in the
pressure vessel drops to 223 °F (106 °C).
The control PC board monitors the pressure in the
pressure vessel until it drops to 0.7 psi (4.8 kPa). At
that time, the control PC board goes into a 35 second
delay. At the end of the delay, the control PC board
sounds a series of five beeper signals and energizes
the pulse solenoid for 500 ms (except for the liquids
cycle on a sterilizer that contains an old style control PC
board with EPROM version H or before. If old style
control PC board after version H or if it is a new style
control PC board, the door will open at the end of the
liquids cycle also.), which allows the door to open to the
vent position. The control PC board then de-energizes
the vent solenoid. The control PC board (on units with
old style control PC board which has EPROM version J
or after or new style control PC board), the control PC
board checks that the door switch switches back to its
normally open position after the door opens. Checking
the door switch position ensures that the door switch is
not stuck closed, which would mislead the operator to
think the door was closed when it was actually ajar.
During the sterilizing portion of the cycle, the control PC
board continuously monitors the pressure sensor,
temperature sensor, door switch, STOP button, and the
ON/STANDBY button. If the pressure or temperature
exceeds preset limits or the status of the door switch,
STOP button, or ON/STANDBY button changes, the
control PC board de-energizes the heating element and
initiates an error sequence.
When an error sequence is initiated, the control PC
board energizes the vent solenoid, flashes the ERROR
indicator, sounds a beeper signal, and displays the
error code, corresponding to the cause of the error, on
the TEMP/TIME display. When the pressure in the
pressure vessel drops to 0.7 psi (4.8 kPa), the control
PC board delays 35 seconds. The control PC board
then de-energizes the vent solenoid and stops sounding
the beeper signal. The ERROR indicator lamp continues to flash and the error code continues to be displayed until the STOP or ON/STANDBY button is
pressed.
The control PC board illuminates the DRYING indicator
and extinguishes the STERILIZING indicator. The
control PC board begins a countdown of 30 minutes
and displays this time on the TEMP/TIME display. The
control PC board energizes the heating element for 45
seconds, delays two minutes, energizes the heating
element for 30 seconds, delays two minutes, energizes
the heating coil for a predetermined duration (the
duration can be set on the dip switch assembly for
different time intervals), and then delays for two more
minutes. The control PC board continues to energize
the heating element for xx seconds followed by a two
minute delay for the rest of the drying portion of the
cycle. On units with new style control PC boards, there
is a dip switch assembly which contains two switches
for "dry cycle element duration". There are four possible
combinations for these two switches, which results in
setting the heating element duration during the drying
cycle; they are repeat interval setting #1, setting #2,
setting #3, and setting #4. Setting #1 is shortest repeat
interval, setting #2 is longer, setting #3 is longer yet,
and setting #4 is the longest repeat interval. This
function was added to allow adjustment of "dry cycle
element duration on units which are overheating during
drying cycle. The unit comes factory set at the repeat
interval setting #3. If the sterilizer contains an old style
control PC board with EPROM version L or after or a
new style control PC board, the control PC board
monitors the chamber for excessive pressure during the
drying cycle, which indicates that the door did not open
properly. If excessive pressure is detected, the drying
cycle is stopped. After 30 minutes have elapsed, the
control PC board illuminates the COMPLETE indicator,
extinguishes the DRYING indicator, and sounds the
beeper signal seven times. If the liquids cycle was
selected, the DRYING indicator does not illuminate, the
heating element does not energize, and the COMPLETE
indicator illuminates. If the STOP button is pressed
during the drying portion of the cycle, a new cycle can
be initiated.
Several safety checks are present to prevent the
pressure vessel from exceeding safe pressure and
temperature limits. The control PC board limits the
pressure and temperature from exceeding 34.5 psi (238
kPa) and 277 °F (136 °C) respectively. A pressure
relief valve prevents the pressure from exceeding 35
psi (241 kPa) on older units and 40 psi (275.7 kPa) on
newer units. A bi-metallic thermostat shuts off power to
the control PC board if the pressure vessel temperature
exceeds 295 °F (146 °C).
1.4SPECIFICATIONS
Factual data for the sterilizer is provided in Table 1-1.
Table 1-1. Specifications
DescriptionData
Dimensions (overall):
Length ............................................... 19 in (470 mm)
Width (depth) ....................................... 15 in (381 mm)
Height .................................................. 16 in (406 mm)
Electrical Requirements:
100 VAC Unit ...............................100 VAC 50 - 60 HZ,
100 VAC Unit ......................................... 1425 WATTS,
15 amps @ 100 VAC
115 VAC Unit ......................................... 1425 WATTS,
12 amps @ 120 VAC
230 VAC Unit ......................................... 1500 WATTS,
7 amps @ 240 VAC
Recommended Circuit:
A separate (dedicated) circuit is recommended for
this sterilizer. The sterilizer
to an electrical circuit with other appliances or
equipment unless the circuit is rated for the additional load.
If a part replacement is required, order the part directly
from the factory as follows:
(1) Refer to Figure 1-2 to determine the location of
the model number and serial number of the
sterilizer and record this data. There are
different letter prefixes which proceed the serial
number, depending on the configuration of the
unit. These prefixes are very important and are
needed to order the proper parts.
!"#$%&'(%
Return To Table Of Contents
)"'"*+,(%'-&*.+$%&'
MODEL NUMBER
SERIAL NUMBER
MA2071-01
Figure 1-2. Model Number / Serial Number Location
(2) Refer to the Parts List to determine the item
numbers of the parts, part numbers of the
parts, descriptions of the parts, and quantities
of parts needed and record this data (Refer to
para 6.1).
NOTE
Ask the Purchasing Department of the company that
owns the sterilizer for this information. Otherwise,
this information may be obtained from the dealer that
sold the sterilizer.
(4) Call Midmark with the recorded information and
ask for the Medical Services Department. See
back cover of this manual for the phone number
or use the Fax Order Form (See page 7-2 for
Fax Order Form).
1.6Special Tools
Table 1-2 lists all the special tools needed to repair the
sterilizer, describes how to obtain the special tools, and
describes the purpose of each special tool.
(3) Determine the installation date of the sterilizer
and record this data.
Table 1-2. Special Tool List
Description of Special Tool
Digital Multimeter (must be
capable of displaying 3 digits)
Water Level Sensor WrenchMidmark Corp.
3/32 in. Diameter PunchCommercially AvailableAny TypeUsed to remove / install two roll pins which secure
Pressure Gauge Test HarnessMidmark Corp.
Manufacturer's
Name / Address / Phone
Commercially AvailableAny TypeUsed to check probes, switches, and connections
60 Vista Drive
Versailles, Ohio 45380
(513) 526-3662
60 Vista Drive
Versailles, Ohio 45380
(513) 526-3662
Manufacturer's
Part Number
for proper function by performing continuity
checks.
050-2324-00Used to hold fitting in place so the nut that holds
water level sensor can be loosened / tightened.
door switch in place or roll pin which secures latch
lever to pulse solenoid.
002-0372-00Used to check the pressure in the pressure vessel
during a cycle to diagnose malfunctions and / or
adjust the pressure range potentiometer to a
correct setting.
button. When each button is pressed, its
corresponding lamp should illuminate and stay
illuminated. Also, the PRESSURE (PSI)
display should display the pressure that the
control PC board is reading, and the TEMP (°F)
/ TIME (MIN: SECONDS) display should
display the temperature that the control PC
board is reading.
(9) Disconnect the power cord from the sterilizer.
(10) Connect Pressure Gauge Test Harness to
sterilizer (See Figure 2-2) (Refer to Table 1-2
for special tool).
(11) Drain the water from the condensing tank.
WARNING
The following steps require the steril-
izer to be plugged in to power while
the operational test is being performed. Do not
touch any components inside of the sterilizer.
Failure to comply with these instructions could
result in an electrical shock, which could result
in severe personal injury or death.
(12) Connect the power cord to the sterilizer.
(13) If the sterilizer contains an old style control PC
Board with EPROM version M or before,
perform steps 15 thru 18; then go to step 27. If
the sterilizer contains an old style control PC
board with EPROM version N or after or a new
style control PC board, perform steps 19 thru
26; then go to step 27.
2
TUBE
FITTING
PRESSURE GAUGE
TEST HARNESS
To connect Pressure Gauge Test Harness to sterilizer:
1. Disconnect tube (1) from pressure sensor (2).
2. Connect tube of Pressure Guage Test Harness
to pressure sensor (2).
3. Connect tube (1) to fitting of Pressure Guage
Test Harness.
To disconnect Pressure Gauge Test Harness from
sterilizer:
1. Disconnect tube (1) from fitting of Pressure
Guage Test Harness.
2. Disconnect tube of Pressure Guage Test Harness
from pressure sensor (2).
3. Connect tube (1) to pressure sensor (2).
Figure 2-2. Connecting / Disconnecting Pressure
Gauge Test Harness
(16) Fill condensing tank with distilled water.
1
MA2070-01
NOTE
Ensure that the sterilizer door is open during the
following step.
(14) Press the ON/STANDBY button, followed by
the UNWRAPPED button, followed by the
START button (See Figure 2-1).
NOTE
The condensing tank water level sensor may need to
be dried before proceeding to the next step.
extinguish, the DOOR AJAR lamp should begin
to flash, and the beeper signal should continue
to sound. Go to step 27.
(18) Close the sterilizer door.
(19) Press the ON/STANDBY button, followed by
the UNWRAPPED button, followed by the
START button (See Figure 2-1).
!"#$%&'(%%
Return To Table Of Contents
$"!$%')(+'/($*&01,"!2&&$%')
(20) Observe. The FILLING lamp should illuminate.
After approximately 5 minutes, the FILLING
lamp should extinguish, the WATER LOW lamp
should illuminate and a beeper signal should
sound.
(21) Press the STOP button followed by the ON/
STANDBY button.
(22) Fill condensing tank with distilled water.
(23) Open the sterilizer door.
(24) Press the ON/STANDBY button, followed by
the UNWRAPPED button, followed by the
START button (see Figure 2-1).
(25) Observe. The DOOR AJAR lamp should begin
to flash, and the beeper signal should start to
sound. Go to step 27.
(26) Close the sterilizer door.
(27) Observe. The DOOR AJAR lamp should
extinguish and the beeper sound should stop.
The FILLING lamp should illuminate. After 30 90 seconds, the FILLING lamp should extinguish and the STERILIZING lamp should
illuminate. Should begin to hear water boiling in
the pressure vessel and steam and air releasing through the bellows assembly.
(28) Observe. At 205 °F (96°C), should begin to
hear the bellows assembly close. At approximately 215 °F (102 °C), should hear the
bellows assembly close completely (hissing
sound stops). The PRESSURE (PSI) display
should begin to display pressure.
(30) Observe. When the time elapsed on the TEMP
(°F)/TIME (MIN : SECONDS) display reaches
zero, the vent solenoid should energize.
Should hear steam and air venting into the
condensing coil. The TEMP (°F)/TIME (MIN :
SECONDS) display should show the temperature in the pressure vessel decreasing and the
PRESSURE (PSI) display should show the
pressure in the pressure vessel decreasing.
When the pressure drops to approximately 0.7
psi (5 kPa) on the PRESSURE (PSI) display,
there should be a 35 second pause; then a
series of five beeper signals should sound and
the pulse solenoid should energize. The
sterilizer door should open, but only to the first
door stop.
(31) Stop the sterilizer from proceeding into the
drying portion of the cycle by pressing the
STOP button.
(32) Set the multimeter to read at least 5 VDC.
Connect the red lead of the multimeter to Test
Point (A, Figure 4-28) and the black lead of the
multimeter to Test Point (B).
(33) Observe. The multimeter should read 2.550
VDC ±0.001 VDC.
(34) If sterilizer does not operate correctly as
described in steps 1 thru 34, replace the
malfunctioning component(s) or perform the
adjustment(s) that are necessary to correct the
problem. If necessary, refer to Table 2-1,
Troubleshooting Guide, to determine the exact
cause of the malfunction.
(35) Disconnect power cord from sterilizer.
(29) When the “heat up” portion of the cycle is
complete and the elapsed time is being counted
down on the display PC board, compare the
pressure displayed on the PRESSURE (PSI)
display to the pressure displayed on the
Pressure Gauge Test Harness. The two
pressures should agree with each other within
± 0.5 psi (3.5 kPa).
WATER LOW indicator
illuminates when
condensing tank is full or
does not illuminate when
condensing tank is low.
Silicone tube is cut or
not keeping terminal
from touching nut which
causes grounding (only
on units with old style
control PC board with
EPROM version M or
before).
Condensing tank water
level sensor
malfunctioning or dirty
(only on units with old
style control PC board
with EPROM version M
or before).
WATER LOW lamp on
display PC board
malfunctioning.
!"#$%&'(%%
$"!$%')(*'+($,&-./"!0&&$%')
Check for continuity
between terminal and nut.
Perform continuity check
on condensing tank water
level sensor.
Replace display PC board
with known working display
PC board.
Replace silicone tube. Refer
to para 4.8.
Replace/clean condensing
tank water level sensor.
Refer to para 4.8.
Replace display PC board.
Refer to para 4.21.
Door is
obstructed/binding.
WATER LOW indicator
illuminates, but chamber
is overfilled with water.
Door is hard to open or
close.
Control PC board
malfunctioning.
Temperature sensor is
malfunctioning, causing
water sensor circuit to
malfunction (preventing
water level sensor from
detecting water).
Door panel needs
adjusted.
Door latch binding.Operate door latch by hand
Steam block hitting
inside of door cover.
Unplug condensing tank
water level sensor
connector from Control PC
board, start a cycle, and
observe. WATER LOW
lamp should flash after
approximately 5 minutes.
Replace suspect
temperature sensor with
known working temperature
sensor.
periodically on the sterilizer. These inspections and
services should be performed as often as indicated in
Table 3-1 is a Scheduled Maintenance Chart which lists
the chart.
the inspections and services that should be performed
Table 3-1. Scheduled Maintenance Chart
IntervalInspection or ServiceWhat to Do
Semi-annuallyObvious damageVisually check condition of sterilizer for obvious damage such as: cracks in components, missing
Fasteners/hardwareCheck sterilizer for missing or loose fasteners/hardware. Replace any missing hardware and
Moving partsAll moving parts should be lubricated with high temperature grease.
Warning and instructional
decals
Display overlayCheck for missing, damaged, or illegible display overlay. Replace display overlay if necessary.
Wiring connectionsCheck the integrity of all wiring connections. Clean all dirty connections. Tighten any loose
Free movement of door
latch
Free movement of door
switch spring
Latch leverOperate latch lever by hand to check for resistance. Clean latch lever. Lubricate latch lever with
Door and Dam gasketsRemove gaskets and check for dirt on gaskets, deterioration of gaskets, or voids in gaskets.
FilterClean filter using a mild soap and water solution. Replace filter if necessary. Refer to para 4.19.
Pressure vessel water
level sensor
components, dents in components, leaks, or any other visible damage which would cause sterilizer
to be unsafe to operate or would compromise the performance of the sterilizer. Repair sterilizer if
necessary.
tighten any loose hardware as necessary using Loctite 271 if necessary.
Check for missing or illegible decals. Replace decals as necessary.
connections. Replace any damaged connections.
Clean door latch. Lubricate door latch and door pins with high temperature grease.
Clean door switch spring. Lubricate door switch spring with high temperature grease. Replace
door switch spring if necessary.
high temperature grease.
Clean gaskets using a mild soap and water solution. Replace gaskets if necessary. Refer to para
4.20.
Remove any build-up from pressure vessel water level sensors. Replace sensor if necessary.
Refer to para 4.10.
Condensing tank water
level sensor (early units
only)
TubingRemove tubing and inspect for buildup. Clean, drain, and flush tubing. Replace tubing if
Display PC boardPerform the display PC board lamp/display/button check. Refer to para 4.25. Replace display PC
Pressure relief valvePerform a pressure relief valve check. Refer to para 4.26. Replace pressure relief valve if
Printer (optional
equipment)
Operational testPerform an operational test to determine if the sterilizer is operating within its specifications (Refer
from the outlet before removing any of
the sterilizer covers/panels or making any repairs
to prevent the possibility of electrical shock.
Also, drain the sterilizer to prevent spills during
repairs. Failure to comply with these instructions could result in serious personal injury or
death.
The following paragraphs contain replacement, repair,
and adjustment procedures for the sterilizer.
4.2Top Cover Removal / Installation
A. Removal
(1) Disconnect power cord from the outlet.
(2) If the sterilizer contains a printer, remove printer
(Refer to Operation Manual).
(3) Open sterilizer door.
(4) Remove two screws (1, Figure 4-1) and steam
block (2) from top cover (3).
(5) Remove four screws (4) and two screws (5)
from top cover (3).
5
7A
4
6
3
2
(6) On units which apply, carefully break front of
top cover (3) loose from silicone sealant.
EQUIPMENT ALERT
Lift top cover carefully and slowly. Rib-
bon connector is still connected to display
PC board and any excess pressure exerted on it
could result in a damaged ribbon connector or
display PC board.
(7) Lift rear of top cover (3) and disconnect ribbon
connector (6) from display PC board (7A) or
control PC board (7B) (depending on whether
ribbon connector (6) is part of display PC board
or is a separate component).
(8) Remove top cover (3) from sterilizer.
B. Installation
(1) Coat mating surfaces of front panel lip and top
cover (3) with silicone sealant.
(2) Position top cover (3) over sterilizer, with rear of
top cover raised, and connect ribbon connector
(6) to display PC board (7A) or control PC
board (7B).
(3) Install top cover (3) on sterilizer and secure
using two screws (5) and four screws (4).
(4) Install steam block (2) on top cover (3) and
secure using two screws (1).
(5) Close sterilizer door.
(6) If the sterilizer contains a printer, install printer
neously pulling downward and outward on back
panel.
!"#$%&'(%3
.+%'$"'+'#"(4(!"*3%#"
EQUIPMENT ALERT
Ensure that wires are completely discon-
nected before attempting to remove back
panel. Failure to do so could result in damage to
sterilizer.
NOTE
Units prior to serial number CZ1110/OM1000 have a
fuse holder attached to the back panel. On these
sterilizers, tag and disconnect two wires from the
receptacle and one wire from the fuse holder instead
of performing step 6.
(6) On older units, tag and disconnect three wires
(3) from terminals of receptacle (4) and remove
back panel (2) from sterilizer. On newer units,
separate back panel (2) from unit as far as
possible and lay it on its side.
5
B. Installation
1
NOTE
Units prior to serial number CZ1110/OM1000 have a
fuse holder attached to the back panel. On these
sterilizers, remove tags and connect two wires to the
receptacle and one wire to the fuse holder instead of
performing step 1.
FLAP
4
(1) On older units, remove tags and connect three
wires (3) to terminals of receptacle (4).
(2) Position flap of back panel (2) behind top cover
(5) and secure back panel in place using four
screws (1).
(3) Install left hand side panel (Refer to para 4.4).
(4) Install right hand side panel (Refer to para 4.3).
Removal / Installation (Applies only to
Units with old style Control PC Board
with EPROM Version M or Before)
!"#$%&'(%3
.+%'$"'+'#"(4(!"*3%#"
NOTE
Units with old style control PC board with EPROM
version N or after or new style control PC board do
not have a condensing tank water level sensor.
A. Removal
(1) Drain water from condensing tank.
(2) Remove back panel (Refer to para 4.5).
(3) Remove bellows assembly (Refer to para 4.7).
(4) Disconnect wire (1, Figure 4-7) from termi-
nal (2).
(5) Using Water Level Sensor Wrench to hold
fitting (3), loosen nut (4) (Refer to Table 1-2 for
special tool).
(6) Pull assembled level sensor rod (5) from
fitting (3).
(7) Remove nut (4), silicone tube (6), and crimp (7)
from level sensor rod (5). Discard silicone tube
and crimp.
3
6
4
2
8
7
5
1
MA2054-01
Figure 4-7. Condensing Tank Water Level
Sensor Removal / Installation
(2) Install crimp (7), nut (4), and silicone tube (6)
on level sensor rod (5) and secure by crimping
crimp (7).
(3) Install assembled level sensor rod (5) in
fitting (3).
(8) Using vise grips to hold level sensor rod (5),
remove nut (8) and terminal (2) from level
sensor rod.
B. Installation
(1) Using vise grips to hold level sensor rod (5),
install terminal (2) on level sensor rod and
secure using nut (8).
(4) Using Water Level Sensor Wrench to hold
fitting (3), tighten nut (4) (Refer to Table 1-2 for
special tool).
(5) Connect wire (1) to terminal (2).
(6) Install bellows assembly (Refer to para 4.7).
EQUIPMENT ALERT
The end of the silicone tube must extend
past nut (4) after nut is installed. If it is
not, the level sensor rod will not function properly
because the terminal will be in contact with nut (4).
Units with old style control PC board with EPROM
version N or after or new style control PC board will
only have two wire harnesses to disconnect instead
of three.
(2) Disconnect two / three wire harnesses (1,
Figure 4-8) from two / three terminals (2).
(3) Remove flex guard tubing (3) from two / three
wire harnesses (1).
(4) Remove temperature sensor assembly (4) from
pressure vessel (5).
B. Installation
(1) Coat threads of temperature sensor assembly
(4) with teflon tape.
(2) Install temperature sensor assembly (4) in
pressure vessel (5).
(3) Install flex guard tubing (3) on three wire
harnesses (1).
(4) Connect three wire harnesses (1) to three
terminals (2).
(5) Install back panel (Refer to para 4.5).
4.10Pressure Vessel Water Level Sensor
Removal / Installation
A. Removal
(1) Remove back panel (Refer to para 4.5).
(2) Remove wire tray rack and tray plate (Refer to
para (4-17).
2
5
4
1
3
MA2074-01
Figure 4-8. Temperature Sensor Assembly
Removal / Installation
(3) Disconnect wire (1, Figure 4-9) from termi-
nal (2).
(4) Remove nut (3) and terminal (2) from level
sensor rod (4).
(5) Remove nut (5).
EQUIPMENT ALERT
Do not try to pull level sensor rod out of
back side of pressure vessel. Doing so
will damage level sensor rod.
(6) Push level sensor rod (4) thru fitting (6) and into
(6) Install terminal (2) on level sensor rod (4) and
secure using nut (3).
7
8
4
6
9
10
5
1
Figure 4-9. Pressure Vessel Water Level
Sensor Removal / Installation
B. Installation
(1) If removed, coat fitting (6) with teflon tape and
install fitting in pressure vessel (7).
(2) Install silicone tube (9) and spacer (8) on level
sensor rod (4).
NOTE
The assembled level sensor rod must be installed
from the inside of the pressure vessel.
2
3
MA2055-01
(7) Connect wire (1) to terminal (2).
(8) Install wire tray rack and tray plate (Refer to
para 4.17).
(9) Install back panel (Refer to para 4.5).
4.11Vent Solenoid Removal /
Installation
A. Removal
NOTE
Units prior to serial number CZ1110/OM1000 have a
manifold assembly which secures the vent solenoid.
Refer to the parts list to remove the vent solenoid on
these sterilizers.
(1) Drain water from condensing tank.
(2) Remove back panel (Refer to para 4.5).
(3) Tag and disconnect two wires (1, Figure 4-10)
from terminals (3).
(4) Tag and disconnect two wires (2) from termi-
nals (4).
(5) Disconnect three tubes (5 thru 7) from elbows
(8 thru 10).
(6) Remove two screws (11) from bottom of
manifold assembly (12).
(3) Install assembled level sensor rod (4) in
fitting (6).
EQUIPMENT ALERT
The end of the silicone tube must extend
past nut (5) after nut is installed. If it is
not, the level sensor rod will not function properly
because the terminal will be in contact with nut (5).
Units prior to serial number CZ1110/OM1000 have a
manifold assembly which secures the fill solenoid.
Refer to the parts list to remove the fill solenoid on
these sterilizers.
(1) Drain water from condensing tank.
(2) Remove back panel (Refer to para 4.5).
(3) Tag and disconnect two wires (1, Figure 4-12)
from terminals (3).
(4) Tag and disconnect two wires (2) from
terminals (4).
(5) Disconnect three tubes (5 thru 7) from elbows (8
thru 10).
(6) Remove two screws (11) from bottom of
manifold assembly (12).
NOTE
Spacer (12A) is only on newer units.
(7) Remove manifold assembly (12) and spacer
(12A) from base (13).
(8) Hold screw (1, Figure 4-13) and loosen nut (2).
Turn terminals out of the way.
(9) Remove assembled vent solenoid (3) and
elbow (4) from tee (5).
(10) Remove elbow (6) and tee (5) from fill
solenoid (7).
(11) Remove elbow (8) from fill solenoid (7).
B. Installation
(1) Coat threads of tee (5, Figure 4-13) and elbows
(6 and 8) with teflon tape.
(2) Install elbow (8) on fill solenoid (7).
(3) Install elbow (6) and tee (5) on fill solenoid (7).
(4) Loosen nut (9). Turn terminals of fill solenoid
(7) so wires will connect easily to fill solenoid.
Tighten nut (9).
(8) Turn terminals of vent solenoid (3) so wires will
connect easily to vent solenoid.
(9) Hold screw (1) and tighten nut (2).
EQUIPMENT ALERT
When installing the vent solenoid and the
fill solenoid, the side of the solenoids
marked "IN" need to be connected to the pressure
lines. Reversing solenoids will cause system failure.
(10) Install spacer (12A, Figure 4-12) and manifold
assembly (12) on base (13) and secure using
two screws (11).
(11) Connect three tubes (5 thru 7) to elbows (8
thru 10).
(12) Connect two wires (2) to terminals (4).
(13) Connect two wires (1) to terminals (3).
(14) Install back panel (Refer to para 4.5).
(15) Refill condensing tank with distilled water.
4.13Control PC Board Removal /
Installation
(5) Tag and disconnect two wires (7) from
terminals (8).
(6) Disconnect ribbon connector (9) from connec-
tor (10).
(7) Tag and disconnect neutral wire (11) from
terminal (12).
(8) Tag and disconnect hot wire (13) from termi-
nal (14).
(9) Disconnect printer cable (15) from connec-
tor (16).
(10) Disconnect door switch harness (17) from
connector (18).
(11) Tag and disconnect two wires (19) from pulse
solenoid terminals (20).
(12) Disconnect heater element wire (21) from
connector (22).
NOTE
Positions of locknuts (1) and screws (2) are reversed
on newer units.
(13) Remove two locknuts (1, Figure 4-15), screws
(2), and control PC board (3) and control PC
board bracket (4) as an assembly from
base (5).
A. Removal
(1) Remove right hand side panel (Refer to
para 4.3).
(2) Disconnect pressure tube (1, Figure 4-14) from
pressure sensor (2).
NOTE
Units with old style control PC board with EPROM
version N or after or new style control PC board will
only have two wire harnesses (3) to disconnect
instead of three.
(3) Disconnect two / three wire harnesses (3) from
two / three connectors (4).
(4) Tag and disconnect two wires (5) from termi-
nals (6).
NOTE
Units prior to serial number CZ1110/OM1000 have
the ground wire attached to the control PC board
bracket. When nut (6), starwasher (7), and screw (8)
are removed, ground wire will also be removed.
(14) Remove nut (6), starwasher (7), and screw (8)
from control PC board bracket (4).
(15) Turn three standoff screws (9) 1/4 turn in the
counter-clockwise direction and remove control
PC board (3) and insulator (10) from control PC
board bracket (4).
board bracket (4) as an assembly on base (5)
and secure using two screws (2) and locknuts (1).
10
4
2
6
7
3
8
5
1
Figure 4-15. Control PC Board Removal / Installation
B. Installation
(1) Install insulator (10, Figure 4-15) and control
PC board (3) on control PC board bracket (4)
and secure by turning three standoff screws (9)
1/4 turn in the clockwise direction.
MA2073-01
(4) Connect heater element wire (21, Figure 4-14) to
connector (22).
(5) Connect two wires (19) on pulse solenoid
terminals (20).
(6) Connect wire harness (17) to connector (18).
(7) Connect printer cable (15) to connector (16).
(8) Connect hot wire (13) to terminal (14).
(9) Connect neutral wire (11) to terminal (12).
(10) Connect ribbon connector (9) to connec-
tor (10).
(11) Connect two wires (7) to terminals (8).
(12) Connect two wires (5) to terminals (6).
NOTE
Units with old style control PC board with EPROM
version N or after or new style control PC board will
only have two wire harnesses (3) to connect instead
of three.
(13) Connect two / three wire harnesses (3) to two /
three connectors (4).
(14) Connect pressure tube (1) to pressure sen-
sor (2).
NOTE
Units prior to serial number CZ1110/OM1000 have
the ground wire attached to the control PC board
bracket. When nut (6), starwasher (7), and screw (8)
are removed, ground wire will also be removed.
Make sure ground wire is re-installed upon installation.
(2) Install screw (8), starwasher (7), and nut (6) on
control PC board bracket (4).
NOTE
Positions of locknuts (1) and screws (2) are reversed
on newer units.
(15) Install right hand side panel (Refer to para 4.3).
SECTION IV
Return To Table Of Contents
MAINTENANCE / SERVICE
4.14Door Switch Removal / Installation
A. Removal
NOTE
Units prior to serial number CZ1110/OM1000 have a
different switch weldment. Since this procedure
tampers with the pressure vessel integrity, it should
not be attempted until a Midmark service representative has been contacted.
(1) Open sterilizer door.
(2) Remove right hand side panel (Refer to
para 4.3).
(3) Disconnect wire harness (1, Figure 4-16) from
connector (2).
(4) Using 3/32 in. punch, remove two roll pins (3)
from door switch bracket (4) (Refer to Table 1-2
for special tool).
(5) Remove door switch (5) from door switch
bracket (4).
B. Installation
(1) Position door switch (5) in door switch
bracket (4).
(2) Using 3/32 in. punch, install two roll pins (3) in
door switch bracket (4) (Refer to Table 1-2 for
special tool).
(3) Connect wire harness (1) to connector (2).
(4) Install right hand side panel (Refer to
para 4.3).
(5) Close sterilizer door.
4.15Pulse Solenoid Removal / Installation
A. Removal
(1) Remove control PC board (Refer to para 4.13).
(2) Remove two screws (1, Figure 4-17) from pulse
solenoid (2).
(3) Remove nut (3) and shoulder screw (4), and
then remove pulse solenoid (2) and latch lever
(5) as an assembly from pressure vessel
bracket (6).
Figure 4-16. Door Switch Removal / Installation
(4) Using 3/32 in. punch, remove roll pin (7) and
latch lever (5) from plunger of pulse solenoid
(2) (Refer to Table 1-2 for special tool).
B. Installation
(1) Using 3/32 in. punch, install latch lever (5) on
plunger of pulse solenoid (2) and secure using
roll pin (7) (Refer to Table 1-2 for special tool).
(2) Coat shoulder screw (4) with Loctite 271.
(3) Install latch lever (5) and pulse solenoid (2) as
an assembly on pressure vessel bracket (6).
Secure using shoulder screw (4) and nut (3).
while tightening nuts. Otherwise, damage
to heating element or improper positioning of heating
element may result. Also, spacer assembly must
remain above gaskets. If spacer assembly is
installed under gasket, leaking will result.
6
4
3
5
1
Figure 4-20. Heating Element and Gasket
Removal / Installation
B. Installation
(1) Install two gaskets (10) on heating element (8).
EQUIPMENT ALERT
Over bending may result in broken or
cracked terminals.
8
10
MA2061-01
(6) Install two brass washers (5), lock washers (4),
and nuts (3) on heating element (8).
EQUIPMENT ALERT
Over bending may result in broken or
cracked terminals.
(7) Bend terminals (2) slightly, back to their original
90° position.
(8) Connect two wires (1) to terminals (2).
(9) Turn sterilizer upright.
(10) Install wire tray rack and tray plate (Refer to
para 4.17).
(11) Install base inspection cover (Refer to
para 4.6).
(12) Refill condensing tank with distilled water.
4.19Filter Removal / Installation
NOTE
The following step is necessary to install the heating
element in the pressure vessel.
(2) Straighten terminals (2) slightly from the
present 90° position to approximately a 10°
straighter position.
(3) Install spacer assembly (9) and heating element
4.22Condensing Tank Assembly
Removal / Installation
A. Removal
(1) Drain water from condensing tank.
MA2065-01
2
1
MA2066-01
Figure 4-24. Condensing Tank Assembly Fill Tube
Disconnection / Connection
(8) Loosen hose clamp (1, Figure 4-25) and
remove sight/drain tube (2) and hose clamp
from condensing tank assembly (3).
NOTE
The following step applies only to units with old style
control PC board with EPROM version M or before.
Units with old style control PC board with EPROM
version N or after or new style control PC board do
not have wire (4) and terminal (5).
Positions of locknut (9) and screws (8) are reversed
on newer units.
(11) Remove screw (8), locknut (9), and condensing
tank assembly (3) from base (10).
B. Installation
NOTE
Fill tube must be lined up with fitting at this time
because it will be impossible to line up after screw
and locknut are installed.
!"#$%&'(%3
Return To Table Of Contents
.+%'$"'+'#"(4(!"*3%#"
(1) Position condensing tank assembly (3, Figure
4-25) on base (10) and secure using screw (8)
and locknut (9).
(2) Connect vent tube (6) to fitting (7).
NOTE
The following step applies only to units with old style
control PC board with EPROM version M or before.
Units with old style control PC board with EPROM
version N or after or new style control PC board do
not have wire (4) and terminal (5).
(3) Connect wire (4) to terminal (5).
(4) Install hose clamp (1) and sight/drain tube (2)
on condensing tank assembly (3). Tighten
hose clamp.
(5) Install bellows assembly (Refer to para 4.7).
(6) Install back panel (Refer to para 4.5).
(7) Install top cover (Refer to para 4.2).
(8) Turn sterilizer onto its side.
(11) Refill condensing tank with distilled water.
4.23Pressure Relief Valve Removal /
Installation
A. Removal
(1) Remove condensing tank assembly (Refer to
para 4.22).
(2) Disconnect tube (1, Figure 4-26) from
elbow (2).
(3) Remove elbow (2) from pressure relief
valve (3).
(4) Remove pressure relief valve (3) from pressure
vessel (4).
B. Installation
(1) Coat threads of pressure relief valve (3) and
elbow (2) with teflon tape.
(2) Install pressure relief valve (3) in pressure
vessel (4).
(9) Connect fill tube (1, Figure 4-24) to elbow (2).
(3) Install elbow (2) on pressure relief valve (3).
(4) Connect tube (1) to elbow (2).
(5) Install condensing tank assembly (Refer to
para 4.22).
4.24Pressure / Temperature Potentiometers Adjustments
EQUIPMENT ALERT
The pressure zero, pressure range, and
temperature potentiometers cannot be
adjusted separately. If one potentiometer needs
adjusted, then the entire potentiometer adjustment
procedure must be performed, and it must be
performed in the proper sequence as follows:
DANGER
The following steps require the steril-
izer to be plugged in to power while
the adjustment is being performed. Do not touch
any components except for the adjusting screw
of the potentiometer. Failure to comply with
these instructions could result in an electrical
shock, which could result in severe personal
injury or death.
NOTE
After the following step is performed, the display PC
board lamps will individually illuminate and then
extinguish one at a time in a left to right sequence.
When all the lamps have illuminated and extinguished, proceed with step 4.
(1) Remove right hand side panel.
(2) Open sterilizer door.
TEMP / TIME DISPLAY
TEMP (°F) \ TIME (MIN : SECONDS)
FILLING
PRESSURE (PSI)
PRESSURE
DISPLAY
UNWRAPPED
BUTTON
UNWRAPPED
POUCHES
BUTTON
STERILIZING
POUCHES
LIQUIDS
BUTTON
LAMP
DRYING
LIQUIDS
PACKS
BUTTON
(3) Press and hold down the LIQUIDS and PACKS
buttons simultaneously while connecting the
power cord to the sterilizer (See Figure 4-27).
Release the LIQUIDS and PACKS buttons.
When each button is pressed, its corresponding lamp
will illuminate and stay illuminated. Also, after the
following step, the PRESSURE (PSI) display will
display the pressure that the control PC board is
reading and the TEMP (°F) / TIME (MIN: SECONDS)
display will display the temperature that the control
PC board is reading. When each button has been
pushed, proceed with step 5.
!"#$%&'(%3
.+%'$"'+'#"(4(!"*3%#"
counter-clockwise direction to raise the voltage
setting until the multimeter reading is 2.550
VDC ±0.001 VDC.
(9) Close the sterilizer door.
(10) Connect Pressure Gauge Test Harness to
sterilizer (See Figure 4-29) (Refer to the
sterilizer Service and Parts Manual for this
special tool).
(4) Press and release the UNWRAPPED,
POUCHES, LIQUIDS, PACKS, START, STOP,
AND ON/STANDBY buttons one at a time and
in this order.
(5) Turn adjusting screw of pressure zero potenti-
ometer (1, Figure 4-28) in the counter-clockwise direction until the PRESSURE (PSI)
display reads 0.0 psi (0.0 kPa).
(6) Turn adjusting screw of pressure zero potenti-
ometer (1) in the clockwise direction until the
PRESSURE (PSI) display reads 0.1 psi (1
kPa).
CAUTION
The multimeter leads must be connected
exactly as shown in the illustration.
Connecting a lead to the wrong side of the resistor or
wrong pin of the chip will result in an incorrect
adjustment.
(7) Set the multimeter to read VDC at least the 5
Volts range. Connect the red lead of a multimeter to Test Point (A) and the black lead of a
multimeter to Test Point (B).
(11) Press the START button to start a test/calibra-
tion cycle and wait until the temperature on the
TEMP (°F) / TIME (MIN: SECONDS) display
reaches 272 - 273°F (133-134°C) (See Figure
4-27).
NOTE
Turning the adjusting screw of the pressure range
potentiometer in the clockwise direction raises the
sterilizer pressure reading while turning the adjusting
screw in the counter-clockwise direction lowers the
sterilizer pressure reading.
(12) Turn the adjusting screw of the pressure range
potentiometer (2, Figure 4-28) in the clockwise
or counter-clockwise direction until the reading
on the PRESSURE (PSI) display matches the
reading of the Pressure Gauge Test Harness
within a tolerance of ±0.5 psi (3.5 kPa).
(13) Recheck the multimeter reading. It should still
read 2.550 VDC ±0.001 VDC.
(14) Press the STOP button to end the test/calibra-
tion cycle and allow the sterilizer to vent and
open its door (See Figure 4-27).
NOTE
In the following step, if the digital multimeter being
used has only three digits, adjust to a reading of 2.54
VDC and then adjust and set to where the reading
just changes to 2.55 VDC.
(8) The multimeter should read 2.550 VDC ±0.001
VDC. If multimeter reading is not 2.550 VDC
±0.001 VDC, adjust the adjusting screw of the
temperature potentiometer (3) in a clockwise
direction to lower the voltage setting or a
(16) Disconnect the power cord from the sterilizer
and then repeat steps 3 and 4 of this procedure. The reading should read 0.1 psi (1 kPa)
with the sterilizer door open. If not, repeat
steps 5 and 6 of this procedure.
To connect Pressure Gauge Test Harness to sterilizer:
1. Disconnect tube (1) from pressure sensor (2).
2. Connect tube of Pressure Guage Test Harness
to pressure sensor (2).
3. Connect tube (1) to fitting of Pressure Guage
Test Harness.
To disconnect Pressure Gauge Test Harness from
sterilizer:
1. Disconnect tube (1) from fitting of Pressure
Guage Test Harness.
2. Disconnect tube of Pressure Guage Test Harness
from pressure sensor (2).
3. Connect tube (1) to pressure sensor (2).
1
MA2070-01
Figure 4-29. Connecting / Disconnecting Pressure
Gauge Test Harness
(17) Disconnect the power cord from sterilizer.
(1) Disconnect power cord from the sterilizer.
(2) Press and hold down the LIQUIDS and PACKS
buttons simultaneously, while connecting the
power cord to the sterilizer (See Figure 4-30).
Release the LIQUIDS and PACKS buttons.
(3) Observe the display PC board lamps. In a left
to right sequence, each lamp on the display PC
board should individually illuminate and then
extinguish. If all the display PC board lamps do
not illuminate and then extinguish, replace
display PC board (Refer to para 4.21).
(4) Press and release the UNWRAPPED,
POUCHES, LIQUIDS, PACKS, START, STOP,
AND ON/STANDBY buttons one at a time and
in this order.
(5) Observe the lamps that correspond with each
button. If all lamps do not illuminate and stay
illuminated when their corresponding button is
pressed, replace the display PC board (Refer to
para 4.21).
(6) Observe the pressure and temperature/time
displays. If the pressure and temperature/time
displays do not display pressure and temperature readings, replace the display PC board
(Refer to para 4.21).
complete instructions on operating the
sterilizer. Failure to do so could result in severe
personal injury.
(2) Select the UNWRAPPED cycle and start the
cycle.
DANGER
Pressure Relief Valve will be hot.
Use a screwdriver or rod to pull the
wire ring. Do not pull the wire ring with bare
hands or burns will result. Steam will be vented
from under the rear of the sterilizer. To keep
from being burned, place a steam barrier (a
rolled up towel) around the bottom of the sterilizer.
LIQUIDS
PACKS
BUTTON
PACKS
START
STOP
ON/STANDBY
ON / STANDBY
BUTTON
START
BUTTON
STOP
BUTTON
MA207701
(3) When the “heat up” portion of the cycle is
complete and the elapsed time is being counted
down on the display PC board, pull on wire ring
of pressure relief valve (2) with a screwdriver.
(4) If excessive force was required to open pres-
sure relief valve (2), replace the pressure relief
valve (Refer to para 4.23).
NOTE
If the pressure relief valve does not close completely
when the wire ring is released, pull the wire ring
again and release it quickly so the valve snaps back
into position. Do this until the valve seats properly.
(5) Release wire ring of pressure relief valve (2).
The steam should vent out of the pressure
relief valve until the wire ring is released. When
the wire ring is released, the valve should seat,
stopping the release of steam.
(6) If the pressure relief valve (2) will not re-seat
properly, replace the pressure relief valve
(Refer to para 4.23).
There are four repeat interval dip switch settings;
setting #1, setting #2, setting #3, and setting #4. The
unit comes factory set at the repeat interval setting #3. Setting #1 is shortest repeat interval, setting
#2 is longer, setting #3 is longer yet, and setting #4 is
the longest repeat interval.
SETTING #1
REPEAT INTERVAL
F
F
O
1
2
3
4
#3 SWITCH DEPRESSED
TO RIGHT (ON)
#4 SWITCH DEPRESSED
TO RIGHT (ON)
SETTING #2
REPEAT INTERVAL
F
F
O
1
2
3
4
#3 SWITCH DEPRESSED
TO LEFT (OFF)
#4 SWITCH DEPRESSED
TO RIGHT (ON)
SETTING #3
REPEAT INTERVAL
F
F
O
1
2
3
4
#3 SWITCH DEPRESSED
TO LEFT (OFF)
#4 SWITCH DEPRESSED
TO LEFT (OFF)
SETTING #4
REPEAT INTERVAL
F
F
O
1
2
3
4
#3 SWITCH DEPRESSED
TO RIGHT (ON)
#4 SWITCH DEPRESSED
TO LEFT (OFF)
This procedure only applies to units with a new style
control PC board; units with serial numbers: CZ2457, DA-1005, DB-1150, DX-1970, DY-1139, FD1000, and OM-9449 thru Present.
(3) If unit has has been overheating during the
drying cycle and activating the overheat
thermostat, determine the current repeat
interval dip switch settings for the unit per
Figure 4-32 and then change the dip switch
settings to the next shortest repeat interval.
(4) Run a cycle. Repeat step 3 if overheat thermo-
stat activates again.
(5) Install right hand side panel (Refer to para 4.3).
Figure 5-1, sheets 1 thru 3, illustrates the wiring connections for the sterilizer.
Printer
Printer
Connector
12345
Blue
N.O.
Black
Door Switch
Door Pulse
Solenoid
100V Unit 7.90 ohms
120V Unit 14.0 ohms
240V Unit 52.0 ohms
PCB/Heater
(015-0657-00)
Heating Element
100V Unit 6.81 to 7.52 ohms
120V Unit 9.77 to 10.80 ohms
240V Unit 40.46 to 44.71 ohms
PCB/Printer
Harness
(015-0668-00)
Orange (Ground)
Green (Ground)
Red (Power)
Blue (Busy)
White (Data)
Black
Black
Yellow
Green
Harness
Harness (015-0877-00)
4
5
2
PCB/AC/Thermostat
3
1
Display Panel
PC Board
HotNeutral
Control
PC Panel
Overheat
thermostat
White
Ribbon Cable
Connector
(015-0646-00)
Flex Guard
Tubing
(015-0257-02)
Ground
Green/Yellow or
Green
Green
(015-0679-00)
Black
SECTION V
SCHEMATICS AND DIAGRAMS
Temperature
Probe Assembly
Level
Sensor Assembly
Red
Black (+5 V.)
White (Ground)
Red (V. Out)
Black
Black
Fill Solenoid
100V Unit 137 ohms
120V Unit 140 ohms
240V Unit 818 ohms
L
G
U
A.C. Connector
Receptacle
Black
Chamber Level
Sensor Harness
(015-0654-00)
Green
Reservoir
Sensor Assembly
Black
Red
Red
Red
Vent Solenoid
100V Unit 375 ohms
120V Unit 400 ohms
240V Unit 1960 ohms
Reservoir
Sensor
Harness
(015-0655-00)
Cordset
U.S.A. 015-0640-00
Cont. Europe 015-0688-07
Italian 015-0688-08
Swiss 015-0688-09
United Kingdom 015-0688-10
Australian 015-0688-11
Israeli 015-0688-12
Japanese 015-0363-06
MA246706
Figure 5-1 (Sheet 1 of 3). Wiring Diagram (Applies To Serial Numbers CZ1000 thru CZ1684, DB1000 thru DB1010,
100V Unit 7.90 ohms
120V Unit 14.0 ohms
240V Unit 52.0 ohms
Black
Black
Yellow
Green
PCB/Heater
Harness
HotNeutral
Overheat
thermostat
Ground
Black
Black
Fill Solenoid
100V Unit 137 ohms
120V Unit 140 ohms
240V Unit 818 ohms
Red
Red
Vent Solenoid
100V Unit 375 ohms
120V Unit 400 ohms
240V Unit 1960 ohms
Cordset
U.S.A. 015-0640-00
Cont. Europe 015-0688-07
Italian 015-0688-08
Swiss 015-0688-09
United Kingdom 015-0688-10
Australian 015-0688-11
Israeli 015-0688-12
Japanese 015-0363-06
MA246707
Heating Element
100V Unit 6.81 to 7.52 ohms
120V Unit 9.77 to 10.80 ohms
240V Unit 40.46 to 44.71 ohms
PCB/AC/Thermostat
Harness (015-0877-00)
White
Green/Yellow
Black
L
G
N
A.C. Connector
Receptacle
Figure 5-1 (Sheet 2 of 3) Wiring Diagram (Applies To Serial Numbers CZ1685 thru CZ4590, DB1011 thru DB1149,
100V Unit 6.81 to 7.52 ohms
120V Unit 9.77 to 10.80 ohms
240V Unit 40.46 to 44.71 ohms
Blue
N.O.
Black
Door Pulse
Solenoid
100V Unit 7.90 ohms
120V Unit 14.0 ohms
240V Unit 52.0 ohms
Green
Yellow
120 V. CORDSET
(U.S.A.)
015-1139-00
Black
Black
Overheat
thermostat
Green/Yellow
Black
White
015-1138-00
Hot
Neutral
PCB/AC/Thermostat
Ground
Harness
Black
Black
Fill Solenoid
100V Unit 137 ohms
120V Unit 140 ohms
240V Unit 818 ohms
Overheat
thermostat
Black
Green/Yellow
White
Red
Red
Cont. Europe 015-0688-07
Italian 015-0688-08
Swiss 015-0688-09
United Kingdom 015-0688-10
Australian 015-0688-11
Israeli 015-0688-12
Japanese 015-0363-06
L
G
N
A.C. Connector
Receptacle
Vent Solenoid
100V Unit 375 ohms
120V Unit 400 ohms
240V Unit 1960 ohms
230 V.
Cordset
MA376202
Figure 5-1 (Sheet 3 of 3) Wiring Diagram (Applies To Serial Numbers CZ4591 thru Present,
The illustrated parts list provides information for identifying and ordering the parts necessary to maintain the
sterilizer in peak operating condition. Refer to paragraph
1.5 for parts ordering information.
The parts list also illustrates disassembly and assembly
relationships of parts.
6.2Description of Columns
The
Item
column of the parts list gives a component its
own unique number. The same number is given to the
component in the parts illustration. This allows a part
number of a component to be found if the technician can
visually spot the part on the illustration. The technician
simply finds the component in question on the illustration and notes the item number of that component.
Then, he finds that item number in the parts list. The
row corresponding to the item number gives the technician the part number, a description of the component,
and quantity of parts per subassembly. Also, if a part
number is known, the location of that component can be
determined by looking for the item number of the
component on the illustration.
The
Part No.
that component.
column lists the MIDMARK part number for
SECTION VI
PARTS LIST
The
Description
of the component.
The
Qty.
component that is required for the subassembly. The
letters “AR” denote “as required” when quantities of a
particular component cannot be determined, such as:
adhesive.
Bullets [ • ] in the
column show the indenture level of a component. If a
component does not have a bullet, it is a main component of that illustration. If a component has a bullet, it is
a subcomponent of the next component listed higher in
the parts list than itself that does not have a bullet.
Likewise, if a component has two bullets, it is a subcomponent of the next component listed higher in the parts
list than itself that has only one bullet.
column provides a physical description
column lists the number of units of a particular
Part No.
column and the
Description
6.3Torque Specifications and Important
Assembly Notes
When specific assembly torque specifications, measurements, or procedures have been identified, by our
engineering department, as required to assure proper
function of the unit, those torque specifications, measurements, and procedures will be noted on the parts
illustrations. Adherence to these requirements is
essential.
The Technical Publications Department of Midmark
Corporation takes pride in its publications. We are sure
that our manuals will fill all of your needs when you are
performing scheduled maintenance, servicing, or
repairs on a Midmark product.
Page(s) and Paragraph(s) Needing Changed:
Description of Error or Desired Change:
However, if you find any errors or feel that there should
be a change, addition, or deletion to a manual, please
let us know!
Please fax or mail a copy of this completed comment
sheet to:
Midmark Corporation
ATTN: Tec hnical Publica tions Department
60 Vista Drive
Versailles, Ohio 45380
Fax: (937) 526-5542