Safety First: The primary concern of Midmark
Corporation is that this table is maintained with the
safety of the patient and staff in mind. To assure that
services and repairs are completed safely and correctly,
proceed as follows:
(1) Read this entire manual before performing any
services or repairs on this table.
(2) Be sure you understand the instructions
contained in this manual before attempting to
service or repair this table.
Safety Alert Symbols
Throughout this manual are safety alert symbols that
call attention to particular procedures. These items are
used as follows:
DANGER
A DANGER is used for an imminently
hazardous operating procedure,
practice, or condition which, if not correctly
followed, will result in loss of life or serious
personal injury.
NOTE
A NOTE is used to amplify an operating procedure,
practice or condition.
Warranty Instructions
Refer to the Midmark “Limited Warranty” printed in the
Installation and Operation Manual for warranty information. Failure to follow the guidelines listed below will
void the warranty and/or render the 411 Power Examination Table unsafe for operation.
• In the event of a malfunction, do not attempt to
operate the table until necessary repairs have been
made.
• Do not attempt to disassemble table, replace malfunctioning or damaged components, or perform
adjustments unless you are one of Midmark’s
authorized service technicians.
• Do not substitute parts of another manufacturer
when replacing inoperative or damaged components.
Use only Midmark replacement parts.
WARNING
A WARNING is used for a potentially
hazardous operating procedure,
practice, or condition which, if not correctly
followed, could result in loss of life or serious
personal injury.
CAUTION
A CAUTION is used for a potentially
hazardous operating procedure, practice,
or condition which, if not correctly followed, could
result in minor or moderate injury. It may also be
used to alert against unsafe practices.
EQUIPMENT ALERT
An EQUIPMENT ALERT is used for an
imminently or potentially hazardous
operating procedure, practice, or condition which, if
not correctly followed, will or could result in serious,
moderate, or minor damage to unit.
This manual contains detailed troubleshooting, scheduled maintenance, maintenance, and service instructions for 411 Power Examination Table. This manual is
intended to be used by Midmark’s authorized service
technicians.
1.2How to Use Manual
A. Manual Use When Performing Scheduled Mainte-
nance.
(1) Perform inspections and services listed in
Scheduled Maintenance Chart (Refer to
para 3.1).
(2) If a component is discovered to be faulty or out
of adjustment, replace or adjust component in
accordance with maintenance/service instructions (Refer to para 4.1).
B. Manual Use When Table Is Malfunctioning And
Cause Is Unknown.
The major serviceable components of the table are the
headlock, back actuator, back transmitter, back receiver, back capacitor, foot actuator, foot transmitter,
foot receiver, foot capacitor, tilt actuator, tilt transmitter,
tilt receiver, tilt capacitor, base actuator, base transmitter, base receiver, base capacitor, base down limit
switch, chain assembly, base slide assembly, gas
spring(s), program panel, program panel interface
board, PC logic board, interface board, hand control
panel, hand control PC board, foot control foot
switches, and pan safety limit switch.
B. Theory of Operation (See Figures 5-1 and 5-2
for wiring diagrams / electrical schematics)
Electrical Power:
Line voltage (115 VAC for domestic units and 230 VAC
for export units) is supplied directly to the table's
interface board. There is a transformer and associated
follow-on circuitry on the interface board which reduces
the line voltage to 12 - 14 VDC. The 12-14 VDC
provides power to operate the circuitry of the PC logic
board, hand control, and foot control.
(1) Perform an operational test on table (Refer to
para 2.1).
(2) Perform troubleshooting procedures listed in
Troubleshooting Guide (Refer to para 2.2).
(3) If a component is discovered to be faulty or out
of adjustment, replace or adjust component in
accordance with maintenance/service instructions (Refer to para 4.1).
C. Manual Use When Damaged Component Is Known.
(1) Replace or adjust component in accordance
with maintenance/service instructions (Refer to
para 4.1).
1.3Description Of 411 Power Examination Table
A. General Description (See Figure 1-1).
The 411 Power Examination Table is an examination
table designed specifically for performing general
medical examinations and procedures.
On domestic units, 115 VAC is continuously supplied to
the electrical receptacle. Export units do not have an
electrical receptacle.
Manual Operation of Base, Tilt, Back, and Foot Actuator
Assemblies:
When a manual function is selected using the manual
positioning buttons on the hand control or program
panel, a signal is sent thru the PC logic board to the
interface board. The signal causes the appropriate
relay on the interface board to energize (i.e., if the TILT
UP button is depressed on hand control, the Tilt UP
relay on interface board is energized). Line voltage is
continuously supplied to all relays on interface board.
So, when a relay is energized, the line voltage flows
thru the relay and is applied across the windings of the
actuator motor, causing it to run. When the hand
control button is released, the relay de-energizes,
removing the line voltage from the windings of the
actuator motor and causing it to stop running.
NOTE: THE FOOT AND BACK CAPACITORS MAY BE SWITCHED ON
SOME UNITS. CHECK THE WIRE COLORS OF THESE CAPACITORS
TO DETERMINE FOOT OR BACK CAPACITOR LOCATION. RED WIRES
ARE CONNECTED TO THE BACK CAPACITOR AND YELLOW WIRES
ARE CONNECTED TO THE FOOT CAPACITOR.
Actuator Operation:
The Base, Tilt, Back, and Foot actuators are ball screw
driven. The actuator assemblies contain a pivot point
on the end of the ball screw. If an actuator assembly is
run to the end of its stroke, the ball screw shaft spins
inside the nut, which allows the actuator assembly to
run without damaging or advancing the nut.
The Base, Tilt, Back, and Foot actuators have internal
braking mechanisms which use a friction device to keep
actuator stationary when the actuator is not being run.
Programming Operation:
The model 411 can be programmed for up to eight
operator specified positions. To program a position into
memory, the operator uses the manual positioning
buttons on the hand control or program panel to position the table top in the desired position. Now, the PC
logic board determines the position of each function as
follows: each actuator has a transmitter and a receiver
attached to it; the receiver is mounted to the motor end
of the actuator and is stationary while the transmitter is
mounted to the moving actuator shaft of the actuator.
The PC logic board has the transmitter on each actuator shaft continuously transmit a high frequency sound
wave which is received by its receiver. The PC logic
board compares the transmit point on the wave form
against the receive point on the wave form and uses
this data to determine the distance between the transmitter and receiver for each actuator. The PC logic
board uses this frequency measuring technique to
determine the position of the actuators at all times. The
operator now presses the PROGRAM button followed
by one of the eight PROGRAMMED POSITION buttons. The red L.E.D. under the PROGRAM button will
illuminate when the PROGRAM button is pressed and
will remain illuminated for four seconds to inform the
operator that he / she may now select a PROGRAMMED POSITION button (1 thru 8). After four
seconds, the PROGRAM function will automatically
disable itself if a PROGRAMMED POSITION button
has not been selected. When a PROGRAMMED
POSITION button is pressed, the receive point on the
wave form for each actuator is stored into memory.
Now the operator can easily return to the position
stored in memory. To do so, the operator presses and
holds the PROGRAMMED POSITION button (1 thru 8)
which has the desired position stored in memory. The
PC logic board compares the receive point on the wave
form that is actually being measured by each actuator's
transmitter / receiver set, against the value stored in the
PC logic board's memory. The PC logic board then
energizes the appropriate relays and allows each
actuator to run until the receive point on the wave form
being measured by each actuator's transmitter / receiver set is equal to the value stored in the PC logic
board's memory. When the values are equal, the relays
are de-energized, stopping the actuators.
Auto Return Operation:
When the operator presses the AUTO RETURN button,
the PC logic board energizes the base actuator's down
relay, causing the base actuator to run. The base
actuator lowers until the base down limit switch is
tripped. The PC logic board also senses that the base
down limit switch is tripped and de-energizes the base
actuator relays.
If the operator, at any time, wishes to stop the auto
return function, he / she can press the STOP button.
The PC logic board senses that the STOP button has
been pressed and de-energizes the base actuator
relays, causing the base actuator to stop.
Pan Safety Limit Switch:
The table has a N.O. pan safety limit switch. If the pan
assembly is not pushed into its fully stowed position, the
pan safety limit switch will not be tripped and there will
be an open circuit. If the pan safety limit switch circuit
is open, no power can reach the foot actuator windings,
preventing movement of the FOOT UP and FOOT
DOWN function. This safety feature prevents the
operator from accidentally colliding the foot section into
the treatment pan assembly.
General Information:
All actuator motors have a thermal overload switch
which will activate if the actuator assembly is run
continuously. The actuator motor was not designed for
continuous operation. The normal cool off period for
the thermal overload switches is 10 - 20 minutes.
Each actuator motor has a capacitor which provides
start up power and motor run power.
There is a 0.15 amp fuse, located on the interface
board, which provides over-current protection for the
transformer.
There are four 5 amp slow blow fuses, located on the
interface board, which provide over-current protection
for each function's motors (i.e, Tilt fuse protects tilt
actuator motor).
Under the control console, there is an ON / OFF switch.
Turning the switch to OFF disables movement of the
table, preventing accidental table movement or patients
from operating table. A red CONTROL PANEL
POWER L.E.D. illuminates when the ON / OFF switch
is in the ON position.
1.4Specifications
Factual data for the 411 Power Examination Table is
provided in Table 1-1. Also, see Figure 1-2.
Table 1-1. Specifications
DescriptionData
Weight:
Without Shipping Carton................ 550 lb (249.5 kg)
With Shipping Carton..................... 625 lb (283.5 kg)
Weight Capacity (Maximum)............. 300 lb. (136.0 kg)
A separate (dedicated) circuit is recommended for
this table. The table
electrical circuit with other appliances or equipment
unless the circuit is rated for the additional load.
should not
be connected to an
1.5Parts Replacement Ordering
If a part replacement is required, order the part directly
from the factory as follows:
Shipping Carton....... 58 in. "L" x 42 in. "W" x 31 in. "H"
(147.3 cm x 106.7 cm x 78.7 cm)
Dimensions (See Figure 1-2):
Table Top Length ......................... 70.5 in. (179.0 cm)
Table Top Length (headrest extended).............. 81 in.
(205.7 cm)
Table Top Width ................................ 27 in. (68.6 cm)
Overall Width..................................... 27 in. (68.6 cm)
Table Adjustment:
Back Section ...................................................0 to 85°
Foot Section ...................................................0 to -87°
Tilt Function..................................................... 0 to 42°
Table Top Height (Adjustable): ...................25.75 in. to
FOOT UP ..........................................10 +/- 1 seconds
(1) Refer to Figure 1-3 to determine the location of
the model number and serial number of the
table and record this data.
(2) Refer to the Parts List to determine the item
numbers of the parts, part numbers of the
parts, descriptions of the parts, and quantities
of parts needed and record this data (Refer to
para 6.1).
NOTE
Ask the Purchasing Department of the company that
owns the table for this information. Otherwise, this
information may be obtained from the dealer that
sold the table.
(3) Determine the installation date of the table and
record this data.
(4) Call Midmark with the recorded information and
ask for the Medical Products Technical Services Department. See back cover of this
manual for the phone number or use the Fax
Order Form (See page 7-2 for Fax Order
Form).
Table 1-2 lists all of the special tools needed to repair
the table, how to obtain the special tools, and the
purpose of each special tool.
Table 1-2. Special Tool List
Description of Special Tool
MultimeterCommercially AvailableAny TypeUsed to perform continuity and volt age checks.
Torque WrenchCommercially AvailableAny TypeUsed to t i ght en nuts or s cr ews to specified v alues .
Support BarMidmark Corporation
Pry BarMidmark Corporation
Manufacturer's
Name / Address / Phone
60 Vista Drive
Versai ll es, Ohio 45380
(513) 526-3662
60 Vista Drive
Versai ll es, Ohio 45380
(513) 526-3662
Manufacturer's
Part Number
051-0412-00Used to support t he weight of the t able t op so t he
base actuator may be disconnected from table.
Used in base actuator rem ova l and gas spring(s)
removal procedures.
051-0413-00Used to compress t he gas spri ng( s) so they may
In order to effectively diagnose the malfunction of the
table, it may be necessary to perform an operational
test as follows:
WARNING
Refer to the Operator Manual for
complete instructions on operating the
table. Failure to do so could result in personal
injury.
NOTE
The Operational Test, for the most part, only describes what should happen when the table is
operated. If the table does something other than
described, a problem has been discovered. Refer to
the Troubleshooting Guide to determine the cause of
the problem and its correction.
(1) Plug the table into a grounded, non-isolated,
correctly polarized outlet, that has the proper
voltage output for the table. See Figure 2-1.
NOTE
The pan slide assembly should be pushed in fully for
the following step.
FOOT UP ..........................................10 +/- 1 seconds
(4) Repeat steps 2 and 3 on the table's console. If
the table has a foot control, repeats steps 2 and
3 for the foot control also.
(5) Lower FOOT DOWN function almost all the
way down. Pull the pan slide assembly outward until the pan safety limit switch is no
longer tripped. Press FOOT UP and then
FOOT DOWN button on hand control.
(6) Observe. The foot section of table top
not
move when FOOT UP or FOOT DOWN
buttons are pressed.
(7) Push pan slide assembly inward until pan
safety limit switch is tripped. Press FOOT UP
and then FOOT DOWN button on hand control.
(8) Observe. The foot section of table top
move when FOOT UP or FOOT DOWN buttons
are pressed.
should
should
(2) Press TABLE UP, TABLE DOWN, BACK UP,
BACK DOWN, TILT UP, TILT DOWN, FOOT
UP, and FOOT DOWN buttons on hand control.
(3) Observe. The table top should move in the
direction corresponding to the button which is
being depressed. No section of the table top
should drift on its own after hand control button
is released. No actuator assembly should
make excessive squealing noises. Movement
should be steady and should match the positions and speeds listed below:
Table Positions:
Back Section ................................................... 0 to 85°
Foot Section ...................................................0 to -87°
Tilt Function..................................................... 0 to 42°
Table Top Height (Adjustable): ...................25.75 in. to
41.625 in.
(65.4 cm to 105.7 cm)
NOTE
Once the PROGRAM button is pressed, the operator
has four seconds to press one of the eight PROGRAMMED POSITION buttons.
(9) Press the PROGRAM button followed by the
PROGRAMMED POSITION button labeled "1".
(10) Observe. When the PROGRAM button is
pressed, the red L.E.D. directly above it should
illuminate.
(11) Use the manual positioning buttons to move the
table top to a new position.
NOTE
Steps 12 thru 14 check functionality of PROGRAMMED POSITION buttons 1 thru 8 on console.
(12) Press and hold the PROGRAMMED POSITION
button "1" (located on console) until the table
top stops moving.
Table 2-1 is a Troubleshooting Guide which is used to
determine the cause of the malfunction.
SECTION II
Return To Table Of Contents
TESTING AND TROUBLESHOOTING
Table 2-1. Troubleshooting Guide
ProblemSymptomProbable CauseCheckCorrection
Table will not o p e rate
when any of the eight up
or down functions, auto
return, or programmed
positions ar e selected from
the hand control or
console.
When any hand control or
console but ton is pressed,
nothing happens (actuator
motor(s) does not run or
hum).
Nothing hap pens when a
hand contro l or console
button is pressed, but table
runs when a foot control
switc h i s depresse d.
Power cord is not plugged
into facility wall outlet.
CONTR OL PA NE L PO W ER
switch is OFF.
Facility circuit breaker
providing power t o table is
tripped.
Wire conn ect ions loo se.Check all wiring
0.15 amp transformer
primary fuse on interface
board is blown.
Interface board is
malfunctioning.
PC logic board is
malfunctioning.
PC logic board is
malfunctioning.
Check to see if power
cord is plug ged i n.
Check to see if
CONTROL PANEL
POWER switch is ON.
CONTROL PANEL
POWER L.E.D. should
illuminat e to in dicat e
switch is ON.
Check t o see if f acility
circuit br eaker is tripped.
One way of checking this
is to plug a lamp into wall
outlet that table was
plugged into.
connections from power
cord to terminal block
and from terminal block
to interface board.
Perform cont inuit y check
on wires. Use
multimeter to check for
proper v oltage levels.
Refer to Figure 5-1 or
5-2 for this check.
Perform cont inuit y check
on primary fuse.
Replace suspect
interface board with
known wo rking interface
board.
Replace suspect PC logic
board with known
working PC logic board.
Replace suspect PC logic
board with known
working PC logic board.
Plug power cor d into f a c ility
wall outlet.
Turn CONTROL PANEL
POWER switch to ON. Inform
operator on use of the switch.
If the table does not work
even when the switch is in ON
positi on, replace PC logic
board. Refer to para 4.6.
If circuit breaker is tripped,
determine what caused circu it
breaker to trip, correct the
problem, and then reset /
replace circ uit breaker.
Clean any dirt y con ne ct ions.
Tighten any loose
connec tions. Replace any
damaged connections.
Replace blown prim ary fuse.
See Figure 5-1 or 5-2 for
location of fuse.
Replace interface board.
Refer to para 4.4 or 4.5.
One or more funct ions
cannot be initiated from
the hand control.
Table 2-1. Troubleshooting Guide - Continued
Some funct ions may be
initiated from the hand
control, but at least one
may not . All functions can
be initiated from the
console.
TESTING AND TROUBLESHOOTING
Hand control panel of hand
control is malfunctioning
(switch membrane is
malfunctioning).
Hand control board is
malfunc t ioning.
Hand co ntrol coil cor d has
bad connection or there is a
break in coil cord.
Replace suspect h and
control panel with known
working hand c ont rol
panel.
Replace suspect h and
control board with known
working hand c ont rol
board.
Replace suspect c oi l cord
with known working coil
cord.
SECTION II
Replace hand control panel.
Refer to para 4.8.
Replace hand control board.
Refer to para 4.8.
Replace coil cord.
One or more funct ions
cannot be initiated from
the cons ole.
One or more funct ions
cannot be initiated from
the optional foot control.
Some funct ions may be
initiated from the consol e,
but at least one may not.
All functions c an be
initiated from the hand
control.
Some funct ions may be
initiated from the foot
control, but at least one
may not . All functions can
be initiated from t he hand
contr ol an d console.
Program p anel i s
malfunc t ioning ( swi tch
membrane i s
malfunct ioning ) .
Program panel interface
board is malfunctioning.
Wire conn ect ions loo se.Check all wiring
Foot switch for
non-opera ble function is
malfunctioning.
Wire conn ect ions loo se.Check all wiring
Replace suspect
program panel with
known working program
panel.
Replace suspect
program panel interface
board with known
working program panel
interface board.
connections from
program panel to
program panel interface
board to PC logic board.
Perform cont inuit y check
on wires. Use
multimeter to check for
proper v oltage levels.
Perform a continuity
check o n suspect foot
switch.
connections from foot
control to interface
board. Perform
continu ity check on wires.
Use multimeter to check
for proper voltage levels.
Replace program panel.
Refer to para 4.7.
Replace program panel
interface board. Refer to
para 4.7.
Clean any dirt y con ne ct ions.
Tighten any loose
connec tions. Replace any
damaged connections.
Replace foot switch. Refer to
para 4.27.
Clean any dirt y con ne ct ions.
Tighten any loose
connec tions. Replace any
damaged connections.
FOOT UP function works,
but FOOT DOWN function
does not or FOOT DOWN
function works, but FOOT
UP function does not. All
other functions work.
Programming function is
malfunctioning.
Table 2-1. Troubleshooting Guide - Continued
One function operates
properly, but the other
does not.
A table position cannot be
programmed into
memory.
TESTING AND TROUBLESHOOTING
Wire connections loose.Check all wiring
Interface board is
malfunctioning (relay for up
or down function is
malfunctioning).
Foot actuator assembly is
malfunctioning.
The receiver and
transmitter for a function
are not properly aligned or
are malfunctioning,
resulting in a lost signal.
RF cables for receiver and
transmitter are shorted or
open.
connections to foot
actuator assembly.
Replace suspect
interface board with
known working interface
board.
Replace suspect foot
actuator assembly with
known working foot
actuator assembly.
Refer to Figure 5-1or
5-2 for this check. Check
L.E.D. (D1-Base,
D2-Foot, D3-Back, and
D4-Tilt) on PC logic
board. If L.E.D. is
illuminated, transmitter or
receiver for that function
could be out of
alignment. Align the
receiver and transmitter
for a function and then
see if the table goes to
the proper position when
a PROGRAMMED
POSITION button is
pressed.
Perform continuity check
on RF cables.
SECTION II
Clean any dirty connections.
Tighten any loose
connections. Replace any
damaged connections.
Replace interface board.
Refer to para 4.4 or 4.5.
Replace actuator motor or
foot actuator assembly. Refer
to para 4.21 or 4.14.
Align the receiver and
transmitter of a function. If
this does not correct problem,
replace receiver and
transmitter for the suspect
function. Refer to para 4.11
for back function, para 4.13
for foot function, para 4.15 for
tilt function, para 4.17 for base
function.
Replace any damaged RF
cables.
Program panel is
malfunctioning (switch
membrane is
malfunctioning).
Program panel interface
board is malfunctioning.
PC logic board is
malfunctioning.
Replace suspect
program panel with
known working program
panel.
Replace suspect
program panel interface
board with known
working program panel
interface board.
Replace suspect PC logic
board with known
working PC logic board.
Replace program panel.
Refer to para 4.7.
Replace program panel
interface board. Refer to
para 4.7.
Program ming funct ion is
malfunct ioning - Continued.
When a PROGRAMMED
POSITION button is
pressed, the table top
does not move to the
proper position that was
stored in memory.
The receiver and
transmitter for a function
are not properly aligned or
are m alfunctioning,
res ul t ing in a los t si gna l.
Refer to Figure 2-1 for
this check. Check L.E.D.
(D1-Base, D2-Foot,
D3-Back, and D4-Tilt) on
PC logic board. If L.E.D.
is illumin at ed , t ra nsm it te r
or receiver for that
function coul d be out of
alignment. Align the
receiver and transmitter
for a function and then
see if t he t able goes to
the proper position when
a PROG RAM M ED
POSITION button is
pressed.
Align the r eceiver and
transmitter of a function. If
this does not correct problem,
replace receiver and
transmitter for t he suspect
function. Refer to para 4.11
for back function, para 4. 13
for foot function, para 4.15 for
tilt function, para 4.17 f or base
function.
AUTO RETURN function
does not work.
The Base, Tilt, Back, or Foot
function drifts by its elf.
When AUTO RETURN
“START” button is
pressed, nothi ng
happens.
After base actuator has
lowered table top all the
way down, the base
actuator continues to run.
Function operat es
properly otherwise.
The receiver or transmitter
for a funct ion is
malfunctioning, but its
L.E. D. indicator light is not
illuminated on the interface
board.
PC logic boar d is
malfunctioning.
Program panel is
malfunctioning (switch
membrane is
malfunctioning).
Program panel interface
board is malfunctioning.
Base down lim it swi tch is
malfunctioning (stuck
open).
Base down lim it swi tch is
out of adjustment.
Base down lim it swi tch is
malfunctioning (stuck
closed).
Motor actuator br ake is
malfunctioning.
Replace suspect receiver
and transmitter with
known working receiver
and transmitter.
Replace suspect PC logic
board with known
working PC logic board.
Replace suspect
program panel with
known working program
panel.
Replace suspect
program panel interface
board with known
working program panel
interface board.
Perform cont inuit y check
on base down limit
switch.
Check adjustm ent of
base do wn li m i t switch.
Perform cont inuit y check
on base down limit
switch.
Replace suspect actuator
brake com ponents with
new components.
Replace receiver and
transmitter for suspect
function. Refer to para 4.11
for back function, para 4. 13
for foot function, para 4.15 for
tilt function, para 4.17 f or base
function.
Replace PC logic board.
Refer to para 4.6.
Replace program panel.
Refer to para 4.7.
Replace program panel
interface board. Refer to
para 4.7.
Replace bas e down lim it
switch. Refer to para 4.20.
Adjust base down limi t switch.
Refer to para 4.20.
Replace bas e down lim it
switch. Refer to para 4.20.
Table moves fine for light
patient, but will not move or
moves slowly for very heavy
patient.
Whirling or squeaking noise
is heard when an actuator
assembly is being run.
Excessive sideways play of
table top.
Headrest is not operating
properly.
Stirrups are malfunctioning.Stirrups do not lock into
Heavy patients cause
table to malfunction.
Noisy actuator.Foreign matter on ball
Table is not stable and
can be moved from side
to side.
Headrest drifts downward
while supporting patients
head.
Headrest locking handles
pop out of locked position.
one of the three positions
properly.
TESTING AND TROUBLESHOOTING
Low voltage is being
supplied to table.
Gas spring(s) are weak.Replace suspect gas
Table overloaded with too
heavy of a patient.
Capacitor for suspect
function is weak.
screw threads and / or lack
of lubricant.
Chain assemblies are
loose.
Base slide assembly is
worn or deformed.
Table base is not resting
firmly on floor.
Headrest is out of
adjustment.
Headrest handles handle
stops need adjusted.
Stirrup index pin needs
replaced.
Check voltage at wall
receptacle - should be
115 +/- 5 VAC.
spring(s) with known
working gas spring(s).
Maximum weight
capacity of table is 300
lbs (136.0 kg).
Replace suspect
capacitor with known
working capacitor.
Check for foreign matter
on ball screw threads.
Check for lack of
lubricant on ball screw
threads.
Check tension of chain
assemblies.
Check condition of base
slide assembly.
Check to verify all four
leveling screws are
contacting the floor.
Check adjustment of
headrest.
Check adjustment of
headrest handles handle
stops.
Check for damaged
parts.
SECTION II
Correct low voltage situation
at wall receptacle.
Replace gas spring(s). Refer
to para 4.19.
Inform table operator of
weight limitation.
Replace capacitor.
Clean all foreign matter off of
ball screw threads. Coat ball
screw threads with STP
treatment oil or equivalent. If
actuator assembly is still noisy,
replace it.
Adjust tension of chain
assemblies. Refer to para
4.22.
Replace base slide assembly.
Adjust the leveling screws;
there is one located under
each corner of the base.
Adjust headrest. Refer to
para 4.23 or 4.25.
Adjust headrest handles
handle stops. Refer to para
periodically on the table. These inspections and
services should be performed as often as indicated in
Table 3-1 is a Scheduled Maintenance Chart which lists
the chart.
the inspections and services that should be performed
Table 3-1. Scheduled Maintenance Chart
IntervalInspection or ServiceWhat to Do
Semi-ann ua llyObvious damageVisually che ck condit ion of tabl e for obv ious dam ag e suc h as: crack s in compo ne nt s, mi ssing
Fasteners / hardwareCheck table for missing or loose fasteners / hardware. Rep lace any missing hardware and tighten
Warning and instructional
decals
Pivot points / moving
parts / accessories
Hand controlCheck that hand control works correctly. Make sur e all switch membranes work cor r ectly. If
Program panelCheck that program panel works correctly. Make sure all switch membranes work correctly. If
Foot control (optional)Check t hat foot control works correctly. Make sure all footswit ches operate pr operly. Replace any
Ball screws of actuator
assemblies
Drifting of actuator
assemblies
Foot ext ensionCheck tha t foot extension can be easily removed. Make sure foot section is latched sec urely in
Stirr up ass em bliesCheck that stirrup assem blies lock into one of three positions. Check for wear . Replace worn
Pan safety limit switchCheck t hat pan safety lim it switch is tripped when pan assembly is in fully stowed posi t io n. Adjust
Excessive play in column
assembly
Head restCheck that head r est is adjusted properly and will not drift downwar d on it s own under 45 lbs
Elect r ical receptacle
(Domest ic units only)
UpholsteryCheck all upholstery for rips, tears, or excessive wear. Repla ce cushions as necessary.
AccessoriesCheck t hat all accessories have all of their com ponents and that they fun ct ion properly. I f
Opera tional TestPerform an Operational Test to determine if the table is operating within its specifications (Refer to
components, dents in components, o r any other visible damage which would cause table to be
unsafe to operat e or would comprom ise its performance. Repair table as necessary.
any loose har dw ar e as necessary.
Check f or missing or illegible decals. Re place decals as neces sary.
Lubricate all exposed pivot points, moving parts, and accessories with silicone based lubricant.
necessary, replace hand contr ol panel. Refer to par a 4. 8.
necessary, replace program panel. Ref er to para 4.7.
malfunctioning footswitches. Refer t o para 4.27.
Extend each actuator assem bly and wipe ball screw threads down with a rag to remove foreign
matter. Coat as much of the ball screw threads as possible with STP treatment oil or equivalent.
Run each act uator assemb ly to both ends of its travel a co uple of times t o spread the oil evenly
over all of the ball screw th r eads and then rem ove excess oil. If oil does not correct a squealing
actuat or assembly, replac e actuator assembly.
Check each actuator assembly for d r ift. Replace actuator assem bly brake components as
necessary. Refer to para 4.21.
place when installed.
components as necessary. Refer to para 4.28.
or replace pan safety limit switch if necessary. Refer t o para 4.9.
Check tha t column assembly does not have excessi ve si de t o side play. I f necessary, ti ght en
chain. Refer to par a 4. 22. I f t his does not cor re ct pr oblem, r eplace base slide assembly.
(20.4 kg) of static weight. If necessary, adjust head rest. Ref er t o par a 4. 23 or 4. 25
Check tha t the electrica l receptacle is fun ct ioning properly. Replace r eceptacle as ne cessary.
necessary, repair or replace the access or y.
para 2.1). Replace or adjust any malf unct ioning compo nents .
complete instructions on operating the
table. Failure to do so could result in personal
injury.
NOTE
Perform an operational test on the table after the
repair is completed to confirm the repair was properly
made and that
The following paragraphs contain removal, installation, repair, and adjustment procedures for the table.
all
malfunctions were repaired.
4.2Shrouds Removal / Installation (Applies to Units With Serial Numbers
AV-1000 Thru AV2888)
A. Removal
(1) If possible, raise TILT UP function all the way
up.
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in severe personal injury or death.
(2) Unplug table power cord from wall outlet.
3
2
1
5
ACCESS
HOLES
4
9
8
8
10
OPERAT OR'S
CONSOLE
7
NOTE
Remove guide weldment (5) from side of table which
does not have operator's console. Also, early units
do not have access holes with plugs (1).
(3) Remove two plugs (1, Figure 4-1) from access
holes; then remove two screws (2),
lockwashers (3), two screws (4) and guide
weldment (5) from seat weldment (6).
Remove shroud from side of table which does not
have operator's console.
(5) Remove two screws (9) and either L.H. or R.H.
shroud (8) from inner member weldment (10).
SECTION IV
Return To Table Of Contents
MAINTENANCE / SERVICE
B. Installation
(1) Coat threads of two screws (9) with removable
threadlocking adhesive (Loctite 242).
(2) Install shroud (8) on inner member weldment
(10) and secure with two screws (9).
(3) Secure L.H. and R.H. shrouds (8) together with
six screws (7).
(4) Install guide weldment (5) on seat weldment (6)
and secure with two screws (4), lockwashers
(3), and screws (2).
NOTE
Early units do not have access holes with plugs.
(5) Install two plugs (1) in access holes.
(6) Plug table power cord into wall outlet.
10
9
7
6
8
ACCESS
5
HOLES
1
3
11
2
4.3Shrouds Removal / Installation (Applies to Units With Serial Numbers
AV-2889 Thru Present and DJ-1000
Thru Present)
A. Removal
(1) If possible, raise TILT UP function all the way
up.
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in severe personal injury or death.
(2) Unplug table power cord from wall outlet.
(3) Remove six screws (1, Figure 4-2) and cover
(2) from R.H. and L.H. shrouds (3).
(4) Remove three screws (4) securing R.H. and
L.H. shrouds (3) together.
OPERAT OR'S
4
CONSOLE
3
Figure 4-2. Shrouds Removal / Installation
NOTE
Remove stirrup housing from side of table which
does not have operator's console.
(5) Remove one screw (5), two screws (6),
lockwashers (7), and stirrup housing (8) from
seat weldment (9).
This procedure applies to units with Serial Number
AV-1000 Thru AV-2217 and also AV-2229, AV-2231,
AV-2236, AV-2238, AV-2244, AV-2247, and AV-
2248.
A. Removal
(1) If possible, raise TABLE UP function all the way
up.
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in severe personal injury or death.
B. Installation
NOTE
Four new standoffs, starwashers, and screws are
supplied with the new interface board kit.
(1) Install four new standoffs (11) on brace (10) by
installing standoffs into original mounting holes
for interface board.
(2) Install interface board (5) on standoffs (11) and
secure with four starwashers (8) and new
screws (7).
(3) Connect three wires (6) to terminals of interface
board (5).
(4) Connect four wire harnesses (4) to connectors
of interface board (5).
(5) Install back lower shroud (2) on L.H. and R.H.
support channels (3) and secure with two
screws (1), making sure locator pins are
inserted thru locator holes of back lower
shroud.