Safety First: The primary concern of Midmark
Corporation is that this table is maintained with the
safety of the patient and staff in mind. To assure that
services and repairs are completed safely and correctly,
proceed as follows:
(1) Read this entire manual before performing any
services or repairs on this table.
(2) Be sure you understand the instructions
contained in this manual before attempting to
service or repair this table.
Safety Alert Symbols
Throughout this manual are safety alert symbols that
call attention to particular procedures. These items are
used as follows:
DANGER
A DANGER is used for an imminently
hazardous operating procedure,
practice, or condition which, if not correctly
followed, will result in loss of life or serious
personal injury.
NOTE
A NOTE is used to amplify an operating procedure,
practice or condition.
Warranty Instructions
Refer to the Midmark “Limited Warranty” printed in the
Installation and Operation Manual for warranty information. Failure to follow the guidelines listed below will
void the warranty and/or render the 300 Medical
Examination Table unsafe for operation.
• In the event of a malfunction, do not attempt to
operate the table until necessary repairs have been
made.
• Do not attempt to disassemble table, replace malfunctioning or damaged components, or perform
adjustments unless you are one of Midmark’s
authorized service technicians.
• Do not substitute parts of another manufacturer
when replacing inoperative or damaged components.
Use only Midmark replacement parts.
WARNING
A WARNING is used for a potentially
hazardous operating procedure,
practice, or condition which, if not correctly
followed, could result in loss of life or serious
personal injury.
CAUTION
A CAUTION is used for a potentially
hazardous operating procedure, practice,
or condition which, if not correctly followed, could
result in minor or moderate injury. It may also be
used to alert against unsafe practices.
EQUIPMENT ALERT
An EQUIPMENT ALERT is used for an
imminently or potentially hazardous
operating procedure, practice, or condition which, if
not correctly followed, will or could result in serious,
moderate, or minor damage to unit.
This manual contains detailed troubleshooting, scheduled maintenance, and service instructions for Model
300 Medical Examination Table. This manual is
intended to be used by Midmark’s authorized service
technicians.
1.2How to Use Manual
A. Manual Use When Performing Scheduled Mainte-
nance.
(1) Perform inspections and services listed in
Scheduled Maintenance Chart (Refer to
para 3.1).
(2) If a component is discovered to be faulty or out
of adjustment, replace or adjust component in
accordance with maintenance/service instructions (Refer to para 4.1).
B. Manual Use When Table Is Malfunctioning And
Cause Is Unknown.
(1) Perform an operational test on table (Refer to
para 2.1).
(2) Perform troubleshooting procedures listed in
Troubleshooting Guide (Refer to para 2.2).
(3) If a component is discovered to be faulty or out
of adjustment, replace or adjust component in
accordance with maintenance/service instructions (Refer to para 4.1).
C. Manual Use When Damaged Component Is Known.
(1) Replace or adjust component in accordance
with maintenance/service instructions (Refer to
para 4.1).
The Model 300 Series Medical Examination Table is
primarily used in examination rooms for general examinations and minor procedures. The table top is adjustable from a full horizontal position to a chair position.
Also, when supported with an optional pelvic lift bar, the
lithotomy position can be achieved. There are five
storage drawers which accommodate supplies used
during examinations.
The major serviceable components of the table are the
drawer heater plate (optional) (1, Figure 1-1), heater
on/off switch (optional) (2), electrical receptacle (3),
back lock assembly (4), lift assembly (5), step assembly
(6), and two stirrup assemblies (7).
The Model 300 Series Medical Examination Table is
available in six different configurations and are distinguished by the following model numbers:
Model 300-001 Non-pelvic tilt, drawers on patient's
right side.
Model 300-002 Non-pelvic tilt, drawers on patient's left
side.
Model 300-003 Pelvic tilt w/ heater, drawers on
patient's right side.
Model 300-004 Pelvic tilt w/ heater, drawers on
patient's left side.
Model 300-005 Perineal cut-out in seat section,
drawers on patient's right side.
Model 300-006 Perineal cut-out in seat section,
drawers on patient's left side.
B. Theory of Operation (See Figures 5-1 for wiring
diagrams / electrical schematics.
Electrical Power:
115 VAC line voltage is supplied thru the power cord to
electrical receptacles located on the side of the table.
As long as the power cord is plug into a supply outlet
115 VAC power will be present at the receptacles.
1.3Description Of Model 300 Medical
Examination Table.
Optional Heater Plate:
Some units are equipped with the optional heater plate
for drawer warming. When the operator turns the
(N.O.) Heater On/Off Switch to ON, closing the circuit,
power is supplied to the 60 watt heater plate, heating
the unit. This provides heat in the top, foot end drawer
to warm instruments.
The following standard torque specifications in Table
1-1 apply to the various hardware used on the units
unless otherwise listed elsewhere in service procedures
or parts illustrations:
Table 1-1. Torque Specifications
Hardware Size Torque Values
#6 ........................... 11 to 21 inch / lbs. (1.2 to 2.3 N•M)
#8 ........................... 20 to 30 inch / lbs. (2.2 to 3.3 N•M)
#10 ......................... 32 to 42 inch / lbs. (3.6 to 4.8 N•M)
1/4" ......................... 75 to 85 inch / lbs. (8.5 to 9.6 N•M)
5/16" .................... 18 to 22 foot / lbs. (24.4 to 29.8 N•M)
3/8" ...................... 31 to 35 foot / lbs. (42.0 to 47.5 N•M)
1/2" ...................... 50 to 60 foot / lbs. (67.8 to 81.4 N•M)
1.5Specifications
Factual data for the 300 Medical Examination Table is
provided in Table 1-2. Also, see Figure 1-2.
Power Consumption: ................................ 115 VAC Unit
(126 VAC maximum-110 VAC
minimum), 60 HZ, 6 amps (max.
without heater) or 6.5 amps (max.
with heater) 60 watt heater rating.
Recommended Circuit:
A separate (dedicated) circuit is recommended for
this table. The table
electrical circuit with other appliances or equipment
unless the circuit is rated for the additional load.
should not
be connected to an
1.6Parts Replacement Ordering
If a part replacement is required, order the part directly
from the factory as follows:
NOTE
It is
important
Number be presented when ordering parts, scheduling a service call or seeking technical advice.
that the
entire
Model
and
Serial
Weight:
Without Shipping Carton............... 250 lbs (113.4 kg)
With Shipping Carton........................ 300 lbs (136 kg)
Shipping Carton ...... 60.5 in."L" x 30 in."W" x 42 in."H"
(153.7 cm x 76.2 cm x 106.7 cm)
Dimensions (See Figure 1-2):
Table Top Length .......................... 56 in. (142.24 cm)
Table Top Length (footrest extended). 74 in. (188 cm)
Table Top Width ................................ 27 in. (68.6 cm)
Overall Width ..................................... 27 in. (68.6 cm)
Seat Height (to top of upholstered
seat at foot end) .........................32.125 in. (81.6 cm)
Table Adjustment:
Back Section (manually adjusted) ......... 0 to 70o (± 5o)
Weight Capacity (Maximum)............. 325 lb. (147.4 kg)
(1) Refer to Figure 1-3 to determine the location of
the model number and serial number of the
table and record this data.
(2) Refer to the Parts List to determine the item
numbers of the parts, part numbers of the
parts, descriptions of the parts, and quantities
of parts needed and record this data (Refer to
para 6.1).
NOTE
Ask the Purchasing Department of the company that
owns the table for this information. Otherwise, this
information may be obtained from the dealer that
sold the table.
(3) Determine the installation date of the table and
record this data.
(4) Call Midmark with the recorded information and
ask for the Medical Products Technical Services Department. See back cover of this
manual for the phone number or use the Fax
Order Form (See page 7-2 for Fax Order
Form).
In order to effectively diagnose the malfunction of the
table, it may be necessary to perform an operational
test as follows:
WARNING
Refer to the Operator Manual for
complete instructions on operating the
table. Failure to do so could result in personal
injury.
NOTE
The Operational Test, for the most part, only describes what should happen when the table is
operated. If the table does something other than
described, a problem has been discovered. Refer to
the Troubleshooting Guide to determine the cause of
the problem and its correction.
SECTION II
NOTE
The head board should hold firmly with a 200 lb.
(90.7 kg) load applied to the head end of the table.
SECTION II
TESTING AND TROUBLESHOOTING
(4) Pull footrest assembly out all the way.
Observe.
extend easily. The footrest assembly should hit
stops at the end of its travel, preventing accidental removal.
(5) Lift up on the head board assembly stopping at
several positions.
Observe.
raise smoothly. The back support rod and lock
assembly should hold the head board firmly at
all positions without slipping. The head board
should be able to be raised to 80o.
The footrest assembly should
The head board assembly should
(1) Plug the table into a grounded, non-isolated,
correctly polarized outlet, that has the proper
voltage output for the table.
(2) Using a multimeter, check for proper voltage at
the table's duplex receptacle.
Observe.
present at a 115 VAC unit receptacle.
(3) If the unit has the optional heater, turn the
HEATER ON / OFF switch to ON.
Observe.
should illuminate and after a short wait, the
heater plate should warm the warming drawer.
Turn the HEATER ON / OFF switch to OFF.
There should be 110 to 120 VAC
The HEATER ON / OFF switch
(6) Apply pressure to the head board to check for
drifting. Then press down on one of the back
lock handles while pressing down on the head
board assembly and lower the head board all
the way down.
Observe.
drift when pressure is applied to head board
assembly. The back lock rod assembly should
retract smoothly without binding. The head
board assembly should be able to be lowered
to a horizontal position.
(7) On units with the optional pelvic tilt, lift up on
the seat section slightly.
Observe.
cally raise up into its locked position, supporting
the seat section in the pelvic lift position
(+5o above horizontal).
To lower the seat section to standard position,
raise up slightly on seat section; then grasp
pelvic lift handle and rotate the pelvic lift bar
downward while lowering seat section.
stirrups and move them laterally from the left to
the right. Let the stirrups lock into each of the
four positions. While applying downward
pressure, attempt to move a stirrup laterally to
the left or right.
Observe.
The stirrups should be able to be
extended and moved laterally easily when
raised. The stirrups should lock into each of
the four positions and should not be able to be
moved laterally as long as downward pressure
is applied on the stirrup.
(9) Check operation of each drawer.
ProblemSymptomProbable CauseCheckCorrection
Drawer warmer is not
working. Power is present
at the supply outlet.
Duplex receptacle is not
working.
Heater plate assembly
does not warm up when
heater ON / OFF switch is
turned ON.
Table power cord is
plugged into wall outlet
and power is present, but
there is still no power at
duplex receptacle.
Observe.
drawer rollers should move freely in the drawer
tracks.
(10) Check operation of step.
Observe.
easily. The step should drop into locked
position when it is fully extended and should not
pull out of the table.
2.2Troubleshooting Procedures
Table 2-1 is a Troubleshooting Guide which is used to
determine the cause of the malfunction.
Table 2-1. Troubleshooting Guide
Heater ON / OFF switch is
malfunctioning.
Heater plate assembly is
malfunctioning
Wire connections are loose.Check all wiring
Duplex receptacle is
malfunctioning.
Wire connections are loose.Check all wiring
When heater ON / OFF
switch is turned to ON,
heater switch should
illuminate (this indicates
that there is power
present at output
terminals of switch).
Replace suspect heater
plate assembly with
known working heater
plate assembly or check
for continuity thru heater
plate assembly wires.
connections from power
heater ON / OFF switch
to heater plate assembly.
Perform continuity check
on wires. Use
multimeter to check for
proper voltage levels.
Replace suspect duplex
receptacle with known
working duplex
receptacle.
connections from power
cord to duplex
receptacle. Perform
continuity check on
wires. Use multimeter to
check for proper voltage
levels.
The drawers should slide easily. All
The step should pull out freely and
If heater ON / OFF switch
does not illuminate when
turned ON, replace heater
switch. Refer to para 4.7.
Replace heater plate
assembly. Refer to para 4.8.
Clean any dirty connections.
Tighten any loose
connections. Replace any
damaged connections. Refer
to para 5.1.
Replace duplex receptacle.
Refer to para 4.9.
Clean any dirty connections.
Tighten any loose
connections. Replace any
damaged connections. Refer
to para 5.1.
Duplex receptacle or
optional drawer warmer
not working.
Head Board Assembly
does not operate correctly.
No power is available at
duplex receptacle and
heater ON / OFF switch
does not illuminate when
pressed.
Head Board will not stay in
position, drifts down.
Head Board will not
descend when back lock
handle(s) are depressed.
Head Board descends
rapidly when back lock
handles are depressed.
Power cord is not plugged
into facility wall outlet.
Facility circuit breaker
providing power to table is
tripped / blown.
Wire connections are loose.Check all wiring
Springs on back lock
assembly have broken or
come loose.
The hole in the back lock
assembly for the back
support rod has worn or
elongated.
The back support rod has
worn excessively and will
not hold in the back lock
assembly.
Back lock springs are
disconnected or broken.
Back support rod out of
alignment causing the rod to
bind in the back lock
assembly.
Set screws that secure the
back lock assembly to the
back lock rod are loose or
missing.
Set screw(s) are loose in the
back lock handle(s) prevent
handles from rotating back
lock rod and back lock
assembly.
Lift spring(s) are broken or
disconnected from the lift
lever or the eye bolt(s).
Lift spring(s) tension out of
adjustment.
Check to see if power
cord is plugged in.
Check to see if facility
circuit breaker is tripped /
blown. One way of
checking this is to plug a
lamp into wall outlet that
table was plugged into.
connections from power
cord to duplex receptacle
and heater ON / OFF
switch. Perform
continuity check on
wires. Use multimeter to
check for proper voltage
levels.
Check condition of the
back lock springs.
Check the condition of
the back lock assembly
for excessive wear.
Check for excessive
wear on the back
support rod.
Inspect the back lock
springs.
Check the alignment of
the back support rod and
insure the screws are
tight on the back support
plate or bracket.
Check the set screws to
insure they are in place
and tightened to the
back lock rod.
Check the set screws to
insure they are in place
and tightened to the
back lock rod.
Check the condition of
the lift spring(s).
Check the position of the
wing nuts to insure they
are on the eye bolts
sufficiently.
Plug power cord into facility
wall outlet.
If circuit breaker is tripped,
determine what caused circuit
breaker to trip, correct the
problem, and then reset /
replace circuit breaker.
Clean any dirty connections.
Tighten any loose
connections. Replace any
damaged connections. Refer
to para 5.1.
Replace the springs on the
back lock assembly. Refer to
para 4.5.
Replace the back lock
assembly. Refer to para 4.5.
Replace the back support rod.
Refer to para 4.5.
Connect and / or replace the
back lock springs. Refer to
para 4.5.
Re-align the back support rod
plate or bracket and tighten
the mounting screws. Refer
to para 4.5.
Install and / or tighten the set
screws in the back lock
assembly. Refer to para 4.5.
Install and / or tighten the set
screws in the back lock
handles. Refer to para 4.5.
Reconnect or replace the lift
springs. Refer to para 4.6.
Adjust the spring tension on
the lift springs. Refer to
para 4.6.
periodically on the table. These inspections and
services should be performed as often as indicated in
Table 3-1 is a Scheduled Maintenance Chart which lists
the chart.
the inspections and services that should be performed
Table 3-1. Scheduled Maintenance Chart
IntervalInspection or ServiceWhat to Do
Semi-annuallyObvious damageVisually check condition of table for obvious damage such as: cracks in components, missing
Fasteners / hardwareCheck table for missing or loose fasteners / hardware. Replace any missing hardware and
Warning and
instructional decals
Pivot points / moving
parts / accessories
Step AssemblyCheck step assembly to insure that it slides in and out easily. Insure that the step assembly
Footrest extensionCheck that footrest extension can be easily extended and is not damaged. Refer to para 4.3.
Stirrup assembliesCheck that stirrup assemblies lock into the four different positions. Check for wear. Replace
Electrical receptacleCheck that the electrical receptacle is functioning properly. Replace receptacle as necessary.
Drawer Heater (Optional) Where applicable, check to insure the drawer heater is heating sufficiently to warm the top drawer
Pelvic Tilt (Optional)Check that the spring loaded pelvic tilt is operating properly. Replace springs if necessary.
UpholsteryCheck all upholstery for rips, tears, or excessive wear. Replace cushions as necessary.
AccessoriesCheck that all accessories have all of their components and that they function properly. If
Operational TestPerform an Operational Test to determine if the table is operating within its specifications (Refer
components, dents in components, or any other visible damage which would cause table to be
unsafe to operate or would compromise its performance. Repair table as necessary.
tighten any loose hardware as necessary.
Check for missing or illegible decals. Replace decals as necessary.
Lubricate all exposed pivot points, moving parts, and accessories with silicone based lubricant.
locks in place when extended. Refer to para 4.4.
worn components as necessary. Refer to para 4.14.
Refer to para 4.9.
and that the heater ON / OFF switch indicator light illuminates when the switch is in the ON
position. If necessary, replace the heater switch or heater plate. Refer to para 4.7 or 4.8.
necessary, repair or replace the accessory.
to para 2.1). Replace or adjust any malfunctioning components.
complete instructions on operating the
table. Failure to do so could result in personal
injury.
NOTE
Perform an operational test on the table after the
repair is completed to confirm the repair was properly
made and that
all
malfunctions were repaired.
SECTION IV
MAINTENANCE / SERVICE
SECTION IV
NOTE
To prevent the back lock assembly (5) from snapping
back against the upper wrap weldment (6) when the
back support rod assembly (3) is removed, slightly
depress one of the back lock handles (7) while
removing the back support rod (3). When the back
support rod (3) is clear of the back lock assembly (5)
slowly release the handle (7).
(5) Elevate the head board assembly (8) until the
back support rod assembly (3) is clear of the
upper wrap cover assembly (9).
WARNING
Do not use this table in an explosive or
oxygen-rich atmosphere. To do so
could result in an explosion or fire.
The following paragraphs contain removal, installation, repair, and adjustment procedures for the table.
4.2Upholstery Removal / Installation
A. Removal
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
(1) Unplug the table power cord from the wall
outlet.
(2) For tables without Pelvic Tilt, remove the two
screws (1, Figure 4-1) that secure the seat
section (2) to the upper wrap cross
member (A).
(6) Remove the two screws (10) that secure the
back support rod assembly (3) to the head
board (8) and remove the back support rod
assembly (3). Retain the parts for later installation.
NOTE
On some tables the hinge cover (11) may be stapled
along the edges. If this is the case, the staples will
also have to be removed before removing the hinge
cover.
(7) While supporting the head board assembly (8),
remove the five screws (12) and the head
board assembly (8) from the top hinge (13).
(8) Lift up the seat section (2) and remove the two
screws (14) that secure the seat section (2) to
the top hinge assembly (13) and remove the
seat section (2) and the hinge cover (11).
(9) If replacing the entire top upholstery, remove
the four screws (15, Figure 4-1) and top hinge
(13) from the upper wrap weldment (6).
(10) Extend the footrest and remove the upholstered
footrest pad (16).
(3) Remove the two top, side drawers from the
cabinet to gain access to the bottom of the
back support rod assembly (3).
For ease of installation it works best to remove the
hinge assembly (13, Figure 4-1), lay the head (8) and
seat (2) sections in position on a flat surface with the
lower or back sides facing upward, and install the
hinge cover (11) and top hinge (13). Once this has
been completed install the entire upholstered assembly onto the table.
(1) Lay the new head board (8, Figure 4-1) and
seat section (2) in position on a flat surface with
the face (upholstery side) down.
The pre-drilled holes on the back of the head (8) and
seat section (2) for the top hinge (13) may be hidden
by the excess vinyl of the head and seat sections.
Cut small incisions over each hole, in the excess
vinyl, for insertion of the hinge screws.
(2) Place the hinge cover (11) and the top hinge
(13) in position over the pre-drilled holes
on the backs of the head board (8) and
seat section (2) and secure with the
screws (12 and 14).
Page 21
CAUTION
Return To Table Of Contents
Use care when installing the upholstered
top assembly as it weighs approximately
60 lbs. (27.2 kg). Failure to use proper lifting techniques or assistance could result in personal injury.
(3) Place the upholstered top assembly in position
on the upper wrap assembly (6) and secure the
top hinge (13) with four screws (15).
(4) For tables without Pelvic Tilt, install the two
screws (1) that secure the seat section (2) to
the upper wrap cross member (A).
NOTE
Do not tighten the two screws (10) that secure the
back support rod assembly (3) to the head board
assembly (8) at this time. This will allow the back
support rod assembly (3) to self-align itself and
prevent binding when the back support rod (3) is
inserted thru the back lock assembly (5) and the
head board (8) is raised and lowered.
(5) Place the back support rod assembly (3) on the
back of the head board assembly (8) at the
location of the pre-drilled holes for the bracket
screws (10) and loosely install the screws but
do not tighten them yet.
(6) Insert the back support rod (3) into the back
lock assembly (5).
SECTION IV
MAINTENANCE / SERVICE
4.3Footrest Extension Removal / Installation
A. Removal
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
(1) Unplug the table power cord from the wall
outlet.
(2) Pull out on the footrest assembly (1, Figure 4-2)
until it is fully extended.
(3) Remove the upholstered footrest pad (2) and
treatment pan (3).
NOTE
For tables without Pelvic Tilt, remove the two screws
(4) that secure the seat section (5) to the upper wrap
cross member (A).
(4) Raise the seat section (5) and, using a pliers or
equivalent, bend up the two metal tabs (6) on
the back of the footrest weldment (1) until they
are horizontal.
(7) Raise and lower the head board assembly (8)
several times until the back support rod (3) is
aligned in the back lock assembly (5) and then
tighten the two screws (10) on the back support
rod assembly (3).
(5) Pull the footrest weldment (1) out of the upper
wrap weldment (7).
B. Installation
NOTE
(8) Insert the cotter pin (4) into the end of the back
support rod (3) and bend over the ends of the
cotter pin (4).
(9) Lightly lubricate the back support rod assembly
(3) and raise and lower the head board (8) to
insure proper operation and clean the unit.
Before installing the footrest weldment inspect the
footrest runners (B, Figure 4-2) to insure the nylo
tape glides (8) are in place at the side and top of
each runner. Also, inspect the rubber bumpers (9),
located at the back of the footrest runners to insure
they are intact.
(1) Inspect the nylo tape glides (8, Figure 4-2) and
rubber bumpers (9) and replace if necessary.
Place a light coating of lubricant on the glides.
(2) Insert the footrest weldment (1) into the runners
(B) of the upper wrap weldment (7) and bend
the two tabs (6) downward until vertical.
(3) Install the treatment pan (3) and upholstered
footrest pad (2).
(4) Plug the table power cord into wall outlet.
4.4Step Assembly Removal / Installation
A. Removal
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
MA356400
(1) Unplug the table power cord from the wall
outlet.
(2) Pull out on the step assembly (1, Figure 4-3)
until it is fully extended.
(3) While lightly pulling the step assembly outward,
reach beneath it and push the step stop (2)
upward until it clears the base weldment (3) ;
then remove the step assembly from the
cabinet.
B. Drawer Front Replacement
NOTE
Should it be necessary to replace the drawer front on
the step assembly the step covering must also be
replaced as it is adhered to the drawer front.
(5) Install the drawer front (6) to the step weldment
(5) and secure with eight 6-18 x 1/2 inch
screws (8).
1
5
6
3
94
Before removing the protective cover from the
adhesive backing of the step cover, lay the cover in
place and make some light locations marks to insure
the step cover is installed uniformly on the step
weldment.
(6) Before installing the step cover (1, Figure 4-4)
lay it on the step weldment (2) and make
location marks. The step cover (1) should be
located so that it is positioned 5/16 inch (0.794
cm) to 3/8 inch (0.953 cm) from the outer edges
of the drawer front (3) and step weldment (2).
NOTE
NOTE
7
8
2
Using alcohol or equivalent, clean any residue off the
surface of the step before installing the new step
cover.
(7) Remove the protective cover from the step
cover (4, Figure 4-3) and place it in position,
aligned with the location marks.
(8) Inspect the nylon stem bumpers (9) on the step
weldment (5) and replace if necessary.
MA356500
Figure 4-3. Step Assembly Removal / Installation
(2) Reaching up through the opening at the bottom
of the drawer front (6), push out on the insert
panel (7) and remove it.
(3) Remove the eight 6-18 x 1/2 inch screws (8)
that secure the drawer front (6) to the step
weldment (5) and remove the drawer front.
(4) Clean the surface of the step weldment (5), to
which the step covering (4) will be attached,
with alcohol or a similar cleanser to insure the
adhesive back of the step covering will bond to
the surface of the step weldment.
(9) Inspect the nylon stem bumpers (1, Figure 4-5)
and bumper stops (2) on the runners (A) of the
base weldment (3) and replace if worn.
CInstallation
EQUIPMENT ALERT
Before installing the step assembly into
the cabinet check that the lock tabs (A,
Figure 4-3) on both sides of the bottom of the step
weldment (5) are bent downward approximately 1/8
inch (0.318 cm) but
the nylon stem bumpers (9) next to the tabs. These
tabs (A) lock the step assembly in position when it is
extended.
no farther
then the surface of
CAUTION
Extend the step assembly outward to
insure it locks in place. If necessary
readjust the lock tabs (A, Figure 4-3). Failure to
insure the step assembly locks in place could cause
injury to the users.
(4) Plug the table power cord into wall outlet.
4.5Back Lock Assembly Removal / Installation
A. Removal
(1) Inspect the lock tabs (A, Figure 4-3) and adjust
if necessary.
(2) Place some lubricant, such as vaseline, on the
nylon stem bumpers (9).
NOTE
When inserting step assembly into the cabinet the
step will have to be lifted slightly to clear the lock
tabs at the bottom of the step assembly. It will also
be necessary to push up on the step stop (2, Figure
4-3) in order for the stop to clear the base weldment.
(3) Install the step assembly (1) into the cabinet.
1
A
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
(1) Unplug the table power cord from the wall
outlet.
(2) Remove the two top drawers from the side of
the cabinet to gain access to the bottom of the
back support rod (1, Figure 4-6).
(3) Remove the cotter pin (2) from the back
support rod (1).
NOTE
To prevent the back lock assembly (3) from snapping
back against the upper wrap weldment (4) when the
back support rod (1) is removed, slightly depress one
of the back lock handles (5) while removing the back
support rod (1). When the back support rod (1) is
clear of the back lock assembly (3) slowly release
the handle (5).
(1) Insert the back lock rod (13, Figure 4-6) thru
the bracket (A) of the upper wrap assembly,
two washers (14) and back lock assembly (3).
EQUIPMENT ALERT
Do not over-tighten the set screws in the
back lock handles to prevent stripping the
threads. Torque to 10 to 20 inch / lbs. (1.13 to 2.26
N•M).
(2) Install the back lock handle (5) onto the back
lock rod (13) aligning it with the other back lock
handle (5) and tighten the set screw (11) to 10
to 20 inch / lbs. (1.13 to 2.26 N•M).
(3) Connect the two back lock springs (10) to the
upper wrap (4) and compress the spring hooks
to secure the springs to the upper wrap assembly.
CAUTION
Use care to prevent the head board from
falling when inserting the back support rod
into the back lock assembly. Failure to do so may
cause personal injury.
(4) Lifting the back lock assembly (3), align the
back support rod (1) with the hole in the back
lock assembly and insert the rod thru the
assembly.
(5) With the head board (6) in the elevated or chair
position, rotate the back lock handles (5) so the
pointed end of the handles are facing toward
the foot-end of the table and the handles are
pointed downward approximately 15o to 20o.
EQUIPMENT ALERT
Use care not to overtighten the set screws
in the back lock assembly. Tighten the
set screws 75 to 80 inch / lbs. (0.533 to 0.604 N•M).
(8) If necessary, lift the head board, loosen the set
screw on the back lock assembly and rotate the
handles up or down to adjust their position.
Lower the head board and check the positions.
(9) Raise the head board (6) and tighten both set
screws (12) on the back lock assembly to 75 to
80 inch / lbs. (0.533 to 0.604 N•M).
(10) Raise and lower the head board (6) several
times to insure the back support rod (1) moves
up and down smoothly without binding.
NOTE
If the back support rod (1) binds, loosen the two
screws (15) on the back support rod (1). Raise and
lower the head board (6) several times until it operates smoothly and retighten the screws on the
bracket.
(11) Raise the head board (6) until the back support
rod (1) is free from the back lock assembly (3).
(12) Supporting the head board (6), install the cover
assembly (7) and secure with the four
screws (9).
(13) Rotate the back lock handles (5) to align the
hole in the back lock assembly (3) and insert
the back support rod (1).
(14) Insert the cotter pin (2) in the end of the back
support rod (1) and bend over the ends of the
cotter pin.
(15) Lightly lubricate the back support rod (1), check
operation and install the drawers.
(16) Plug the table power cord into wall outlet.
4.6Lift Lever Removal / Installation
A. Removal
(6) Tighten one of the set screws (12) on the back
lock assembly (3) to secure the back lock
assembly to the back lock rod (13).
(7) Lower the head board (6) to the horizontal
position and observe the position of the back
lock handles (5). The handles should be
parallel with the bottom edge of the head board
or pointed slightly downward.
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
Page 27
(1) Unplug the table power cord from the wall
2
1
12
13
A
12
14
11
10
Return To Table Of Contents
outlet.
(2) Remove the two top drawers from the side of
the cabinet to gain access to the bottom of the
back support rod (1, Figure 4-7).
(3) Remove the cotter pin (2) from the back
support rod (1).
7
9
6
1
SECTION IV
MAINTENANCE / SERVICE
NOTE
To prevent the back lock assembly (3) from snapping
back against the upper wrap weldment (4) when the
back support rod (1) is removed, slightly depress one
of the back lock handles (5) while removing the back
support rod (1). When the back support rod (1) is
clear of the back lock assembly (3) slowly release
the handle (5).
back support rod (1) is clear of the upper wrap
cover assembly (7). Allow the head board (6)
to rest against the seat section (8).
CAUTION
Insure the head board is positioned and
secured to prevent it from falling while
working on the table. Failure to do so could result in
personal injury.
(5) Remove the four screws (9) that secure the
upper wrap cover (7) to the upper wrap assembly (4) and remove the upper wrap cover.
(6) Disconnect the lift springs (10) from the lift
lever (11).
(7) Remove the push-on retaining ring (12) from
one side of the lift support rod (13) and remove
the rod (13) and washers (14) from the lift lever
(11) and upper wrap assembly (4).
1"
(2.5 cm)
1
2
MA357000
Figure 4-8. Lift Lever Installation
NOTE
With the head board in the chair position and no
weight on the table, the head board should descend
1/2 to 3/4 of the way to the horizontal position when
the back lock handles are released.
B. Installation
(1) Place the lift lever (11, Figure 4-7) and two flat
washer (14) in position on the brackets (A) of
the upper wrap assembly (4).
(2) Insert the lift support rod (13) thru the brackets
(A) on the upper wrap assembly and lift lever
(11).
(3) Insure the lift support rod (13) is centered in the
bracket (A) and install the push-on retaining
rings (12).
(4) Connect the lift springs (10) to the lift lever (11).
NOTE
The wing nuts (1, Figure 4-8) should be threaded
onto the eye bolts (2) approximately 1 inch (2.54
cm).
(5) Check the position of the wing nuts (1, Figure
4-8) on the eye bolts (2) of the lift assembly.
(6) Rotate the back lock handles (5, Figure 4-7) to
align the hole in the back lock assembly (3) and
insert the back support rod (1).
(7) Check to insure the head board (6) raises and
lowers smoothly. If necessary, adjust the wing
nuts (1, Figure 4-8) to adjust the tension on the
eye bolts (2).
(8) Raise the head board (6, Figure 4-7) until the
back support rod (1) is free from the back lock
assembly (3).
(9) Supporting the head board (6), install the cover
assembly (7) and secure with the four
screws (9).
(10) Rotate the back lock handles (5) to align the
hole in the back lock assembly (3) and insert
the back support rod (1).
(11) Insert the cotter pin (2) in the end of the back
support rod (1) and bend over the ends of the
cotter pin.
(12) Lightly lubricate the back support rod (1), check
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
(1) Unplug table power cord from wall outlet.
(2) Using a standard screwdriver, carefully pry
heater switch (1, Figure 4-9) out of upper wrap
weldment (2), making sure not to scratch table.
(3) Tag the wires for later assembly and discon-
nect the four wires (3) from heater switch (1).
B. Installation
(1) Connect four wires (3, Figure 4-9) to heater
switch (1) (Refer to the para 5.1).
4.8Heater Plate Removal / Installation
A. Removal
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
(1) Unplug table power cord from wall outlet.
(2) Remove the two drawers from foot end of table.
(3) Remove the heater ON / OFF switch (Refer to
para 4.7).
(4) Pull the footrest section (1, Figure 4-10) out as
far as it will go.
NOTE
For tables without Pelvic Tilt, remove the two screws
(2) that secure the seat section (3) to the upper wrap
cross member (A).
NOTE
Insure that the switch ON ( I ) and OFF ( O ) symbols
are matched with the proper ON or OFF designation
on the decal when installing the switch.
(2) Push heater switch (1) into upper wrap weld-
ment (2) until heater switch "pops" into place.
(3) Plug table power cord into wall outlet.
2
ON
OFF
CAUTION
1
3
HOT
MA357100
(5) Raise up the seat section (3).
(6) Remove the four screws (4) and stirrup guide
(5) from the upper wrap weldment (6).
(7) Remove any wrap-n-tap wire clamps (7)
securing the heater plate wiring (8) to the upper
wrap weldment (6).
(8) Remove the two screws (9), heater plate (10),
and heat shield (11) from bottom side of the
upper wrap weldment (6).
B. Installation
(1) Route the heater plate wires (8, Figure 4-10)
thru the wire hole (B) in the upper wrap
weldment (6).
CAUTION
Insure the heater plate wires are not
pinched by the heater plate during installation. Failure to do so could cause personal injury.
(2) Install the heat shield (11), and the heater plate
(10) on the upper wrap weldment (6) and
secure with the two screws (9).
(3) Install the heater ON / OFF switch (Refer to
para 4.7).
(4) Secure the heater plate wires (8) to the upper
wrap weldment (6) with wrap-n-tap clamps (7).
NOTE
Plug the table in temporarily, turn ON the heater
switch and check to insure the heater plate begins to
warm. Unplug the table and continue on with
installation procedures.
2
6
MA358000
MA358000
(6) Lower the seat section (3). If necessary, for
tables without pelvic tilt, install the two screws
(2) to secure the seat section (3) to the upper
wrap cross member (A).
(7) Push the foot rest (1) into its stowed position.
(8) Install the two drawers in the table.
(9) Plug the table power cord into wall outlet.
(5) Install the stirrup guide (5) on the upper wrap
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
(1) Unplug table power cord from wall outlet.
(2) Extend the stirrups (1, Figure 4-11) out to full
extension.
(3) Extend the footrest (2) out to full extension.
For tables without Pelvic Tilt, remove the two screws
(5) that secure the seat section (6) to the upper wrap
cross member (A).
(5) Raise up the seat section (6).
(6) Remove the screw (7), from inside the upper
wrap assembly (8), that secures the receptacle
cover (4) to the upper wrap assembly (8).
(7) Remove the two screws (9) that secure the
duplex receptacle (10) to the upper wrap
assembly (8).
(8) Pull the duplex receptacle (10) out of the cavity
in the upper wrap assembly (8).
Page 32
SECTION IV
Return To Table Of Contents
MAINTENANCE / SERVICE
NOTE
Before removing any of the electrical leads from the
duplex receptacle place identification tags on them
for installation onto the replacement receptacle. If
the table has a drawer heater there will be five leads
and without the drawer heater there are three leads.
(6) Plug the table power cord into wall outlet and
check that there is power to the receptacle.
4.10Stirrup Components Removal /
Installation
A. Removal
(9) Tag and disconnect the electrical leads (11)
from the duplex receptacle (10) and remove the
receptacle.
B. Installation
CAUTION
When installing the electrical leads (11,
Figure 4-11) on the duplex receptacle (10)
insure the insulation on the leads covers sufficiently
to prevent electrical shorts to the cabinet. The black
leads connect to the side of the receptacle with the
black screws and the white leads connect to the side
with the silver screws. Insure the ground (green)
lead is connected to the ground terminal on the
receptacle and to the table. Failure to do so could
result in personal injury
(1) Connect the electrical leads (11, Figure 4-11) to
the appropriate terminals on the duplex receptacle (10) (Refer to para 5.1).
CAUTION
Insure the receptacle is positioned in the
cavity of the upper wrap assembly so that
none of the electrical terminals or exposed wires are
contacting the upper wrap assembly.
WARNING
Always disconnect the power cord
from the wall outlet before removing
any of the table's covers / shrouds or making any
repairs to prevent the possibility of electrical
shock. Failure to comply with these instructions
could result in personal injury or death.
NOTE
For tables without Pelvic Tilt, remove the two screws
(1, Figure 4-12) that secure the seat section (2) to
the upper wrap cross member (A).
(1) Raise the seat section (2, Figure 4-12) up and
support it.
(2) Pull the foot rest section (3) out to its full
extension.
(3) Pull the stirrup (4) out far enough to access the
stop screw (5) and remove the stop screw (5)
from the stirrup (4).
(4) Pull the stirrup (4) out of the pivot boss (6).
(5) Remove the pivot boss (6) and the stirrup guide
bracket (7) from the upper wrap weldment (8).
(2) Insert the receptacle (10) into its cavity in the
upper wrap assembly (8) and secure with the
two screws (9).
(3) Place the cover (4) on the receptacle (10) and
install the screw (3) on the front cover and the
screw (7) inside the upper wrap
assembly (8).
(4) Push the footrest (2) and stirrups (1) into their
stowed positions.
(5) Lower the seat section (6). If necessary, for
tables without pelvic tilt, install the two screws
(5) to secure the seat section (6) to the upper
wrap cross member (A).
The illustrated parts list provides information for
identifying and ordering the parts necessary to maintain
the unit in peak operating condition. Refer to paragraph 1.5 for parts ordering information.
The parts list also illustrates disassembly and assembly
relationships of parts.
6.2 Description of Columns
The
Item
column of the parts list gives a component its
own unique number. The same number is given to the
component in the parts illustration. This allows a part
number of a component to be found if the technician
can visually spot the part on the illustration. The
technician simply finds the component in question on
the illustration and notes the item number of that
component. Then, he finds that item number in the
parts list. The row corresponding to the item number
gives the technician the part number, a description of
the component, and quantity of parts per subassembly.
Also, if a part number is known, the location of that
component can be determined by looking for the item
number of the component on the illustration.
The
Part No.
for that component.
column lists the MIDMARK part number
SECTION VI
PARTS LIST
The
Description
of the component.
The
Qty.
component that is required for the subassembly. The
letters “AR” denote “as required” when quantities of a
particular component cannot be determined, such as:
adhesive.
Bullets { • } in the
column show the indenture level of a component. If a
component does not have a bullet, it is a main component of that illustration. If a component has a bullet, it
is a subcomponent of the next component listed higher
in the parts list than itself that does not have a bullet.
Likewise, if a component has two bullets, it is a subcomponent of the next component listed higher in the
parts list than itself that has only one bullet.
column provides a physical description
column lists the number of units of a particular
Part No.
column and the
Description
6.3 Torque Specifications and Important
Assembly Notes
When specific assembly torque specifications, measurements, or procedures have been identified, by our
engineering department, as required to assure proper
function of the unit, those torque specifications measurements, and procedures will be noted on the parts
illustrations. Adherence to these requirements is
essential.
The Technical Publications Department of Midmark
Corporation takes pride in its publications. We are sure
that our manuals will fill all of your needs when you are
performing scheduled maintenance, servicing, or
repairs on a Midmark product.
Page(s) and Paragraph(s) Needing Changed:
Description of Error or Desired Change:
However, if you find any errors or feel that there should
be a change, addition, or deletion to a manual, please
let us know!
Please fax or mail a copy of this completed comment
sheet to:
Midmark Corporation
ATTN: Technical Publications Department
60 Vista Drive
Versailles, Ohio 45380
Fax: (937) 526-5542