Micro Motion Brooks Service Tool for QUANTIM Precision Mass Measurement and Control Manuals & Guides

Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
Brooks Service ToolTM for QUANTIM
Precision Mass Measurement and Control
www.BrooksInstrument.com
Brooks Service ToolTM for QUANTIM
Essential Instructions
Read this page before proceeding!
Brooks Instrument designs, manufactures and tests its products to meet many national and international standards. Because these instruments are sophisticated technical products, you must properly install, use and maintain them to ensure they continue to operate within their normal specifications. The following instructions must be adhered to and integrated into your safety program when installing, using and maintaining Brooks Products.
Read all instructions prior to installing, operating and servicing the product. If this instruction manual is not the correct manual, please see back cover for local sales office contact information. Save this instruction manual for future reference.
If you do not understand any of the instructions, contact your Brooks Instrument representative for clarification.
Follow all warnings, cautions and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation and maintenance of the product.
Install your equipment as specified in the installation instructions of the appropriate instruction manual and per applicable local and national codes. Connect all products to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts specified by Brooks Instrument. Unauthorized parts and procedures can affect the product's performance and place the safe operation of your process at risk. Look-alike substitutions may result in fire, electrical hazards or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except when maintenance is being performed by qualified persons, to prevent electrical shock and personal injury.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
Handling Procedure:
1. Power to unit must be removed.
2. Personnel must be grounded, via a wrist strap or other safe, suitable means before any printed circuit card or other internal device is installed, removed or adjusted.
3. Printed circuit cards must be transported in a conductive container. Boards must not be removed from protective enclosure until immediately before installation. Removed boards must immediately be placed in protective container for transport, storage or return to factory.
Comments
This instrument is not unique in its content of ESD (electrostatic discharge) sensitive components. Most modern electronic designs contain components that utilize metal oxide technology (NMOS, SMOS, etc.). Experience has proven that even small amounts of static electricity can damage or destroy these devices. Damaged components, even though they appear to function properly, exhibit early failure.
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
Dear Customer,
Brooks Service ToolTM for QUANTIM
®
The Brooks Service Tool
TM
is a Windows® - based application that provides expanded control of Brooks QUANTIM® QMA and QMB Mass Flow Control (MFC) devices for servicing tasks that include setup and configuration, troubleshooting, and tuning.
We recommend that you read this manual in its entirety as this will enable efficient and proper use of the Brooks Service Tool. Should you require any additional information concerning the Brooks Service Tool, please feel free to contact your local Brooks Sales and Service Office, see back cover for contact information, or visit us on the web at www.BrooksInstrument.com. We appreciate this opportunity to service your fluid measurement and control requirements,and trust that we will be able to provide you with further assistance in future.
Yours sincerely, Brooks Instrument
Contents
Brooks Service ToolTM for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
THIS PAGE WAS INTENTIONALLY
LEFT BLANK
iv
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
Paragraph Page Number Number
Introduction Section 1
1.1 Description .....................................................................................................................................1-1
1.2 How to Use This Manual ................................................................................................................1-2
Theory of Operation Section 2
2 Theory of Operation........................................................................................................................2-1
Installation Section 3
3 Installation ......................................................................................................................................3-1
3.1 Installing the Service Tool Application .............................................................................................3-1
3.2 Connecting the PC to the Device....................................................................................................3-1
3.2.1 Hardware Requirements.............................................................................................................3-2
3.2.2 Connection Procedure................................................................................................................3-3
3.2.3 Disconnecting the PC from the Device .......................................................................................3-6
3.3 Device D-Connector Pin-Outs ........................................................................................................3-7
Basic Operations Section 4
4 Basic Operations ............................................................................................................................4-1
4.1 St arting the Service Tool and Opening a New Session ...................................................................4-2
4.2 Main Window Elements.................................................................................................................. 4-4
4.3 Setting the Main Window Options...................................................................................................4-5
4.4 Putting the Service Tool in Indicator or Control Mode .....................................................................4-8
4.5 Trend Graph Functions...................................................................................................................4-9
4.5.1 Trend Graph Elements ...............................................................................................................4-9
4.5.2 Stopping and S tarting T rending.................................................................................................. 4-11
4.5.3 Scrolling a Graph Axis ............................................................................................................... 4-11
4.5.4 Zooming on a Single Graph Axis ...............................................................................................4-12
4.5.5 Zooming on Both Graph Axes....................................................................................................4-13
4.5.6 Zooming on a Region of the Graph............................................................................................4-14
4.5.7 Displaying X and Y Values at a Particular Point .........................................................................4-15
4.5.8 Copying the Graph to the Clipboard...........................................................................................4-17
4.5.9 Saving the Graph to a File .........................................................................................................4-18
4.5.10 Printing the Graph....................................................................................................................4-18
4.6 Closing the Session and Exiting the Service Tool ..........................................................................4-18
Brooks Service ToolTM for QUANTIM
Contents
®
Viewing and Setting the Device Configuration Section 5
5 Viewing and Setting the Device Configuration ................................................................................5-1
5.1 Setting Input (Setpoint) and Output Upper and Lower Range V alues ............................................. 5-1
5.2 Viewing and Setting Alarms ............................................................................................................ 5-2
5.2.1 Diagnostic Alarm.........................................................................................................................5-3
5.2.2 Configuring Optional Warnings and Alarms ................................................................................5-4
5.2.3 Modifying Standard Alarm Behavior ............................................................................................5-6
5.3 Viewing and Setting Configuration Parameters ............................................................................... 5-7
5.3.1 Viewing the Control Response Parameters ................................................................................5-8
5.3.2 Viewing and Setting the Control Configuration Parameters ........................................................5-9
5.4 Viewing Device Information ...........................................................................................................5-10
5.5 Viewing and Setting Primary and Secondary Output Parameters ..................................................5-10
5.6 Viewing Sensor Parameters ..........................................................................................................5-12
5.7 Viewing Setpoint Parameters ........................................................................................................ 5-13
5.8 Viewing V alve Parameters.............................................................................................................5-14
5.9 Viewing Device V ariables............................................................................................................... 5-15
5.10 Saving the Current Device Configuration.......................................................................................5-17
5.1 1 Restoring a Saved Device Configuration ....................................................................................... 5-17
i
Contents
Brooks Service ToolTM for QUANTIM
Diagnostics Section 6
6 Diagnostics.....................................................................................................................................6-1
6.1 Checking the Warning, Alarm, and Diagnostic Status.....................................................................6-1
6.1.1 Clearing Latched Warning and Alarm Conditions........................................................................6-2
6.2 Testing the Primary and Secondary Outputs...................................................................................6-2
6.3 Testing the Input (Setpoint) ............................................................................................................. 6-5
6.4 Capturing Device Log Data to a File ...............................................................................................6-7
6.4.1 Importing a Data Log File into Excel ...........................................................................................6-9
Troubleshooting, Tuning and Calibration Section 7
7 Troubleshooting, Tuning and Calibration ........................................................................................ 7-1
7.1 Troubleshooting Strategies.............................................................................................................7-2
7.2 Device Tuning and Calibration Procedures..................................................................................... 7-7
7.2.1 Zeroing the Device’s Sensor.......................................................................................................7-7
7.2.1.1 Accessing the Service Tool’s Zero Sensor Function.............................................................7-7
7.2.1.2 Procedure for Zeroing the Sensor ........................................................................................7-8
7.2.2 Optimizing the V alve Behavior for Process Conditions............................................................... 7-10
7.2.2.1 Tuning the V alve Of fset and Span .......................................................................................7-10
7.2.2.2 Cleaning the Orifice for a QUANTIM QMA ..........................................................................7-13
7.2.2.2.1 Removing and Cleaning the QUANTIM QMA’s Orifice..................................................7-14
7.2.2.2.2 Reinstalling the Orifice and Checking the V alve Offset and S p an Values ......................7-15
7.2.2.2.3 Tuning the Valve Response (Proportional and Integral Gain)........................................7-18
7.2.3 Using V alve Override to Control the Valve Opening ...................................................................7-20
7.2.3.1 Using Pin 12 to Control the Valve Override .........................................................................7-21
7.2.3.2 Using the Service Tool’s Valve Override Function ............................................................... 7-22
7.2.4 Cycling the Setpoint ...................................................................................................................7-22
7.2.5 Calibrating the Device Input (Setpoint)....................................................................................... 7-24
7.2.6 Calibrating the Primary and Secondary Outputs ........................................................................ 7-28
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004
Index ................................................................................................................................................... Index-1
Warranty, Local Sales/Service Contact Information...................................................................... Back Cover
ii
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
FIGURES
Figure Page Number Number
4-1 Main Window Elements..................................................................................................................4-4
4-2 Process Variables Options ............................................................................................................. 4-6
4-3 Units of Measure Options ...............................................................................................................4-6
4-4 Trend Graph Options......................................................................................................................4-7
4-5 Setpoint Switch in Control Mode.....................................................................................................4-8
4-6 Trend Graph Main Elements........................................................................................................... 4-9
6-1 Setup for Testing Primary Output Volt age.......................................................................................6-2
6-2 Setup for Testing Primary Output Current.......................................................................................6-3
6-3 Setup for Testing Secondary Output Current ..................................................................................6-3
6-4 Setup for Testing Voltage Setpoint..................................................................................................6-5
6-5 Setup for Testing Current Setpoint..................................................................................................6-6
7-1 Setup for Calibrating to a Voltage Setpoint ....................................................................................7-24
7-2 Setup for Calibrating to a Current Setpoint ....................................................................................7-25
7-3 Setup for Calibrating to a Primary Output Voltage .........................................................................7-28
7-4 Setup for Calibrating to a Primary Output Current .........................................................................7-28
7-5 Setup for Calibrating to a Secondary Output Current.....................................................................7-29
Brooks Service ToolTM for QUANTIM
Contents
®
TABLES
Table Page Number Number
4-1 Options Window Functions.............................................................................................................4-5
4-2 Summary of Trend Graph Toolbar Functions .................................................................................4-10
4-3 Trend Graph Cursor Options .........................................................................................................4-16
5-1 Alarm/W arning Conditions Fields and Options ...............................................................................5-5
5-2 Alarms A ffected by Alarm Disable Options ..................................................................................... 5-6
6-1 Data Log Configuration Window Functions..................................................................................... 6-8
7-1 LED Indications .............................................................................................................................7-2
7-2 Common Start-Up Problems..........................................................................................................7-3
7-3 Common Operating Problems.......................................................................................................7-4
iii
Contents
Brooks Service ToolTM for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
THIS PAGE WAS INTENTIONALLY
LEFT BLANK
iv
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
1 Introduction
1.1 Description
The Brooks Service Tool is a Windows-based application that provides expanded control of Brooks QUANTIM QMA and QMB Mass Flow Control (MFC) devices for servicing tasks that include setup and configuration, troubleshooting, and tuning.
Section 1 - Introduction
Brooks Service ToolTM for QUANTIM
®
Brooks Service Tool Main Window
Using an RS-232 connection between a personal computer (PC) and the device’s service port, you can use the Service Tool to view, control, and modify data such as:
•Flow
• Setpoint
• Analog input/output (I/O) configuration
• I/O ranging
• Valve tuning
• Device I/O calibrations
• Diagnostics
1-1
Section 1 - Introduction
Brooks Service ToolTM for QUANTIM
1.2 How to Use This Manual
It is recommended that you read this manual before installing or using the Service Tool.
The manual is organized into the following sections: Section 1 Introduction
Section 2 Theory of Operation Section 3 Installation Section 4 Basic Operations Section 5 Viewing and Setting the Device Configuration Section 6 Diagnostics Section 7 Troubleshooting, T uning, and Calibration Index Back Cover Warranty, Local Sales/Service Cont act Information
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
1-2
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
2 Theory of Operation
A typical mass flow measurement and control system consists of two components:
Mass flow meter. A mass flow meter incorporates a sensor that measures the amount of mass flow and generates an electrical output signal directly proportional to the mass flow rate.
Control valve. Mating a control valve with a mass flow meter provides a mass flow
controller (MFC). Implementing a mass flow controller requires using the mass flow output
signal as one input and a user-supplied setpoint signal as another input into a comparator circuit. These two signals are then compared and a control signal is generated to operate the opening or closing of the valve. The setpoint signal is, in effect, a request that the mass flow controller maintain a fixed mass flow rate.
Section 2 - Theory of Operation
Brooks Service ToolTM for QUANTIM
®
A setpoint signal can come in many forms. The Brooks QUANTIM QMA and QMB devices are analog—the input/output (I/O) signals between the mass flow device and the user’s control side are represented as a voltage (0–5 Vdc) or current (4–20 mA).
The Brooks Service Tool, which communicates with the device through an RS-232 connection between the PC and the device’s service port connector , uses a digital signal based on a proprietary protocol. The Service Tool has two operating modes:
Indicator, in which the Service Tool is reporting on the data sent to it
from the device, which is being controlled by the normal setpoint source.
Control, in which the Service Tool user specifies the setpoint,
overriding the normal setpoint source.
In either arrangement, the device will attempt to manage the opening and closing of the control valve to maintain the amount of mass flow as requested by the setpoint source.
Note that the Brooks Service Tool is meant to be used as a startup and troubleshooting tool. It is not intended to be permanently connected to the device in your application.
2-1
Section 2 - Theory of Operation
Brooks Service ToolTM for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
THIS PAGE WAS INTENTIONALLY
LEFT BLANK
2-2
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
3 Installation
This section describes how to install the Service Tool application onto a personal computer (PC), and how to connect the PC to the device.
3.1. Installing the Service T ool Application
The Service Tool application can be installed on a Windows PC under the following requirements:
• Windows 95, 98, Me, NT 4, 2000, or XP
• PC running at a minimum of 450 MHz, with at least 64 MB of RAM and a CD-ROM drive
Section 3 - Installation
Brooks Service ToolTM for QUANTIM
®
3.2. Connecting the PC to the Device
To install the Service Tool application, load the Service Tool CD-ROM
into the drive. On the CD-ROM disk, find the installation file, Setup.exe, and double-click it to launch the installation wizard. Follow the installation instructions in the wizard.
The following topics list the hardware requirements and describe the procedure for connecting the PC running the Service Tool to the device’s service port.
3-1
Section 3 - Installation
Brooks Service ToolTM for QUANTIM
3.2.1 Hardware Requirements
The following tools and cables are required to connect the PC serial port to the device’s service port:
• RS-232 to TTL communications interface adapter.
• Standard RS-232 serial cable (between your PC’s serial connector and
the communications interface adapter).
NOTE: If the distance between the PC and the device is short enough, the RS-232
cable is not necessary. Plug the communications interface adapter directly into the PC serial port.
• 3-wire communications cable (between the interface connector and the
device’s service port connector)
• Straight (flat blade) screwdriver
• 3/16-in. nut driver
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
3-2
Communications interface adapter
3-wire communications cable
Installation and Operation Manual
r
r
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
3.2.2 Connection Procedure
To connect the PC to the device’s service port:
1. Turn off power to the device.
2. Remove the standard cable from the device’s 15-pin D-connector.
3. Remove the screw at the center of the top cover and the two 3/
4. Remove the cover.
Brooks Service ToolTM for QUANTIM
16-in. standoffs on the 15-pin D-connector.
Section 3 - Installation
®
Black service port connector located behind D-connector
15-pin D-connecto
D-connecto standoffs
Center cover screw
3-3
Section 3 - Installation
Brooks Service ToolTM for QUANTIM
5. Assemble the RS-232 serial cable, communications interface adapter, and Service Tool 3-wire communications cable.
6. Connect the RS-232 serial cable to a serial port on your computer.
If the distance is short enough, you can connect the communications interface adapter directly to the PC’s serial port, as shown below, and do not need the RS-232 serial cable.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
3-4
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
7. Connect the 3-wire communications cable to the device’s service port, the black connector located behind the D-connector. Align the communications cable’s 5-pin connector with the black wire towards the center of the device (towards the LEDs).
The QUANTIM QMA model’s service port has a 6-pin connector. To attach the communications cable’s 5-pin connector to the 6-pin service port, make the connection so that the unused pin is the innermost pin (the pin closest to the center or LEDs), as shown in the following illustration.
Section 3 - Installation
Brooks Service ToolTM for QUANTIM
®
Service port connector attached (viewed from rear of device)
Unused 6th pin on QUANTIM QMA
8. Reconnect the standard cable to the device’ s D-connector.
This connection is used to apply power and the analog setpoint to the device while using the Service Tool.
9. Turn on power to the device.
10.Verify that the device initializes properly, as indicated by the Status LED being solid green.
You are now ready to use the Service Tool.
3-5
Section 3 - Installation
Brooks Service ToolTM for QUANTIM
3.2.3 Disconnecting the PC from the Device
When you are done using the Service Tool, you should disconnect the PC from the device and reinstall the device’s cover.
To disconnect the PC from the device’s service port:
1. Exit from the Service Tool.
2. Turn off power to the device.
3. Disconnect the standard cable and the 3-wire communications cable from the device.
4. Disconnect the RS-232 serial cable (or interface adapter) from your PC.
5. Place the cover onto the device, and reattach it using the screw at the center of the top cover and the two 3/16-in. standoffs on the 15-pin D-connector.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
6. Reconnect the standard cable to the device’ s 15-pin D-connector. You can now reapply power to the device and begin using it in its process
application.
3-6
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
3.3 Device D-Connector Pin-Outs
The QUANTIM device pin-outs are listed below .
Section 3 - Installation
Brooks Service ToolTM for QUANTIM
®
1
9
Pin No. Function
1 Setpoint Return 2 0 to 5 Vdc Flow Signal Output 3 Not used 4 4 to 20 mA Flow Signal Output 5 +13- to +30-Vdc Power Supply 6 Not used 7 4 to 20 mA Setpoint Input 8 0 to 5 Vdc Setpoint Input
9 Power Supply Common 10 Signal Output Return 11 +5 V Reference Output 12 Valve Override Input
8
15
13 4 to 20 mA Density or Temperature 14 Not used 15 Not used
3-7
Section 3 - Installation
Brooks Service ToolTM for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
3-8
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
4 Basic Operations
This section describes the Service Tool’s basic operations, including:
Starting the tool and opening a new session
Setting the options to determine what variables will be displayed in the main window and how they will be displayed
Putting the tool into Indicator or Control mode, which determines whether the tool can initiate changes in setpoint values
Controlling different aspects of how the trending graphs are displayed in the main window
Capturing the current st ate of the trend graph to the clipboard or a file, and/or by printing it to a hard copy
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
®
Closing the session and exiting the tool
4-1
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
4.1 Starting the Service Tool and Opening a New Session
To start the Service Tool and open a new session:
1. Select the Brooks Service Tool from your Start menu, or double-click its application file. The main window displays, with no session open.
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004
4-2
2. To open a new session, open the File menu and select New Session. The Open Session dialog box displays.
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
3. From the Select Communications Port drop-down list, select the
Comm port to which the Service Tool cable is connected, then click OK.
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
®
The Trend Graph and Status indicators begin showing the data that is being read from the device.
4-3
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
4.2 Main Window Elements
Figure 4-1 indicates the major elements of the main window.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
Maximum flow for which the device was calibrated
Trend Graph controls
Setpoint slider
(used in Control
mode)
Current Setpoint
value
Setpoint
Indicator/Control
switch
Status bar
Figure 4-1 Main Window Elements
Trend Graph
Trend lines of selected variable data, indicated by color
Variable data
4-4
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
4.3 Setting the Main Window Options
To set the main window options, open the File menu and select Options. The Options window displays. From this window, you can
perform the functions described in Table 4-1.
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
®
Table 4-1 Options Window Functions
Function Description
Modify the current settings You can modify the options settings on any of the
three tabs. The options on the three tabs, and which elements on the main window they control, are illustrated in Figures 4-2 through 4-4.
Save current settings as defaults Changes that you make to the option settings are
in effect only for the current session. However, you can save the current settings so that they are used as the default values the next time a session is opened. To do this, select the Save Selections as Defaults option at the bottom of the Options
window. Cancel your setting changes Click the Cancel button. Save your setting changes Click OK.
4-5
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
NOTE: Range values here affect only the trend graph; they do not change the representation of the variable on an analog output.
Figure 4-2 Process Variables Options
Range values determine
variable data values
at 0% and 100% scale
Graph options determine which variable data is trend graphed
Assignment options determine Setpoint type and what variable data is displayed
Select unit of measure
for each variable data type
Select this option to save current settings as default values next time a session is opened
Figure 4-3 Units of Measure Options
Enter precision of measurement data
4-6
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
X-Axis Scroll Method determines how trend graph will advance over time:
Smooth (continuously), or
Page (in time window increments; for example, every 30 seconds)
®
Y-Axis Scale values
determine upper and lower
percentage values
along the Y-axis
Figure 4-4 Trend Graph Options
X-Axis Window determines the time window displayed along the X-axis
4-7
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
4.4 Putting the Service T ool in Indicator or Control Mode
By default, a session opens with the Service Tool in Indicator mode. In this mode, the setpoint is being controlled by the standard setpoint input source. The setpoint data in the Service Tool’s main window “indicates” what the device is currently reporting as its setpoint.
However, you can control the setpoint directly from the Service Tool by moving the Setpoint Switch to Control, then using the slider to set the setpoint or entering a value in the field below the slider. Figure 4-5 shows the Setpoint controls and trend graph in Control mode.
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004
Indicates when Setpoint slider was moved to current position
In Control mode,
use slider to choose setpoint
or
Enter setpoint value here,
then press the Enter key
Setpoint switch in Control mode, setpoint at 51.82 gr/hr
Figure 4-5 Setpoint Switch in Control Mode
4-8
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
4.5 Trend Graph Functions
The following topics cover the functions that you can perform on the trend graph.
4.5.1 Trend Graph Element s
Figure 4-6 shows the Trend Graph’s main element s. Table 4-2 summarizes the Trend Graph’ s toolbar functions.
The data graphs are coordinated by color to the variable data to which they correspond. Note that the selection of what variable data to graph is determined by settings in the main window’s options; refer to “4.3 Setting the Main Window Options” on p. 4-5.
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
®
Trend Graph toolbar
Y-axis, indicating percent
of variable data relative
to upper/lower range settings,
as defined in the main window options
Figure 4-6 Trend Graph Main Elements
Mass Setpoint graph
Valve Drive graph
Mass Flow graph
Current variable data readings
X-axis, indicating elapsed time window since start of session
4-9
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
Table 4-2 Summary of Trend Graph Toolbar Functions
Function Description
Installation and Operation Manual
X-BST-Qm-eng
®
Starts variable data trending, if it has been stopped.
Stops variable data trending.
Scroll up and down the Y-axis, or back and forth across the X-axis.
Zoom in and out of a selected axis.
Zoom in and out of both axes simultaneously.
Part Number: 541B026AAA
July , 2004
Zoom in on a region of the trend graph.
View the X- and Y-axes values at a given point in time the trend graph.
Copy the trend graph to the clipboard.
Save the trend graph as a file.
Print the trend graph.
The following topics provide more detail about each of the Trend Graph toolbar functions.
4-10
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
4.5.2 Stopping and Starting Trending
You can stop the real-time updating of trending data at any time. For example, you may want to analyze the current window of data for troubleshooting. Stopping the updating freezes the data as it is currently displayed.
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
®
4.5.3 Scrolling a Graph Axis
To stop trending, click the
will continue to be collected from the device, but the trend graph is not updated. With trending stopped, you can continue to use the other toolbar functions.
To restart trending, click the button. When you restart trending, the trend graph will jump to display the most recently collected data. To see data collected while the trend graph was stopped, use the X-axis Scroll Tool, covered in the next topic.
The Scrolling tool allows you to scroll up and down the Y-axis, or back and forth across the X-axis. This can be helpful in seeing other areas of the trend graph (for example, if you have zoomed in on it, or if you want to see previous portions of the trend graph).
To scroll on an axis, click the button in the Trend Graph toolbar and move the pointer over the axis on which you want to scroll. The pointer turns into a pointing hand. Click and drag in the direction you want to scroll.
button in the Trend Graph toolbar. Data
Before Y-axis scrolling After Y-axis scrolling
4-11
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
4.5.4 Zooming on a Single Graph Axis
The Zooming on a Single Axis tool allows you to zoom on the X- or Y-axis.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
To zoom on an axis, click the move the pointer over the axis on which you want to scroll. The pointer turns into a double-headed arrow, pointing in the directions relative to the axis it is over. Click and drag in the direction you want to zoom.
Before X-axis zooming After X-axis zooming
button in the Trend Graph toolbar and
4-12
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
4.5.5 Zooming on Both Graph Axes
The Zooming on Both Axes tool allows you to zoom on the X- and Y-axes at the same time.
To zoom on both axes simultaneously, click one of the buttons in the T rend Graph toolbar .
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
®
Before zooming in on both axes After zooming in on both axes
4-13
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
4.5.6 Zooming on a Region of the Graph
The Zooming on a Region tool allows you to zoom on a region on the trending graph, adjusting the X- and Y-axes scales accordingly.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
To zoom on a region of the trend graph, click the Trend Graph toolbar and move the pointer over the trend graph. The pointer turns into an arrow with a box. Click and drag to draw a box around the region on which you want to zoom, then release the mouse button.
With region to zoom selected in box After region was zoomed on
NOTE: When you release the mouse button, the trend graph freezes on the bounded
button in the
4-14
region. Y ou have to click the button to resume trending.
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
4.5.7 Displaying X and Y Values at a Particular Point
The V alues tool allows you to select a point along the Mass Flow trending graph and display the corresponding X- and Y-axis values at that point.
To display the exact X and Y values of the trend graph, click the button in the T rend Graph toolbar and move the pointer over the trend graph. The pointer turns into pointing finger over a colored vertical line— the
cursor
. The X,Y values for the current cursor position are displayed. As you click and drag the cursor across the trend graph, the X,Y values dynamically update.
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
®
X,Y values at current cursor position
If you right-click on the cursor, a pop-up menu displays, allowing you to select from several options that control the cursor’s attributes. These options are described in Table 4-3.
4-15
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
Table 4-3 Trend Graph Cursor Options
Option Description
Channel Allows you choose the channel to which the cursor is
Style Allows you to choose the measurement being made:
Installation and Operation Manual
X-BST-Qm-eng
®
attached. The channels correspond to the variables (Mass Flow, Density, and so on).
Value X-Y. Cursor is a single line.
Value X. Cursor is a single line.
Value Y. Cursor is a single line.
Period. Cursor is two vertical lines connected with a dotted line, showing the time/frequency between the lines.
Peak-Peak. Cursor is two horizontal lines connected by a dotted line showing the percentage difference between the lines (because Y-axis is in percent).
Frequency. Cursor is two vertical lines connected with a dotted line, showing the time/frequency between the lines.
For Styles whose cursor is two lines, you can drag each line separately to make measurements.
Part Number: 541B026AAA
July , 2004
Options>Use Channel Color If selected (the default), the cursor is the color
related to the variable. If not selected, the cursor is yellow.
Options>Hide Hint on Release If selected, the values associated with the cursor are
shown only while the left mouse button is held down.
Flip Alignment Enabled only with single-line cursor styles. If
selected, moves the values displayed to the opposite side of the vertical line.
4-16
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
4.5.8 Copying the Graph to the Clipboard
You can copy the current state of the trend graph to the clipboard, for pasting into another Windows application.
To copy the current state of the trend graph to the clipboard, click the
clipboard. A sample of the trend graph pasted into a document is shown below.
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
button in the Trend Graph toolbar. The trend graph is copied to the
®
4-17
Section 4 - Basic Operations
Brooks Service ToolTM for QUANTIM
4.5.9 Saving the Graph to a File
You can save the current state of the trend graph to a graphic file. To save the current state of the trend graph to a graphic file, click the
button in the Trend Graph toolbar. The Save As dialog box displays.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004
Navigate to the desired directory and enter a name for the file. You can also specify the graphic format—Bitmap (.bmp), Enhanced Metafile (.emf), or JPEG (.jpg)—from the Save as type drop-down list. Click the
Save button to save the file.
4.5.10 Printing the Graph
You can print the current state of the trend graph to a printer.
To print the current state of the trend graph, click the button in the Trend Graph toolbar. The standard Windows Print dialog box displays.
4.6 Closing the Session and Exiting the Service T ool
To close the current session and exit the Service Tool, open the File
menu and choose Exit, or click the window’s close button (at the upper right corner).
4-18
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
5 Viewing and Setting the Device Configuration
This section describes how to view and set device configuration parameters.
NOTE: Any changes you make to the device configuration parameters are saved for
the current and subsequent sessions. Prior to making changes to device configuration settings, it is recommended that you save the current settings. This will allow you to restore the device configuration to a known, working state. For more information about how to save the current settings, refer to “5.10. Saving the Current Device Configuration” on p. 5-17.
5.1 Setting Input (Setpoint) and Output Upper and Lower Range Values
Brooks Service ToolTM for QUANTIM
Section 5 - Viewing and Setting
the Device Configuration
®
The Service Tool’s I/O Ranging function allows you to change the upper and lower range values for the setpoint and output flow values. For example, if you wanted to change the device’s factory-calibrated full-scale flow rate to a lower value, you would use this function. After changing the upper range value, the full value signal (5 V or 20 mA) would correspond to the new value.
Upper range values should never exceed the factory full-scale calibration value.
To change the input (setpoint) and output upper or lower range values, open the Configure menu and choose I/O Ranging. The Input/
Output Ranging Wizard window opens.
5-1
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
Change any of the range values, then click the Close button. Note that, to use the I/O Ranging function, the setpoint and primary output
must be set to the same variable. For example, Mass Setpoint and Mass Flow for output are of the same type; however, Mass Setpoint and Volumetric Flow for output are not.
If you attempt to use the I/O Ranging function and the setpoint and primary output are not set to the same variable, a message window will display indicating that the setpoint and primary output do not match and, therefore, the I/O Ranging function cannot be used for the device.
To correct this condition, go to the Variables and Setpoint tabs in the Configure Parameters window to change the setpoint and primary output to the same variable. For more information, refer to “5.5 Viewing and Setting Primary and Secondary Output Parameters” on p. 5-10 and “5.7 Viewing Setpoint Parameters” on p. 5-13.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
5.2 Viewing and Setting Alarms
Y ou can configure the device to report predefined alarm and warning conditions. The alarm and warning parameters are settable using the Service T ool. All alarm and warning conditions are reported to the Service T ool and indicated by green or red flashing codes on the device’s S tatus and Alarm LEDs. Y ou can also configure alarms and warnings to enable pin 3 of the device’s D-connector.
To access the alarm/warning configuration parameters, open the Configure menu and choose Configure Parameters. The Configure
Parameters window opens.
5-2
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
The following topics provide detailed information about the Diagnostic alarm, which is configured by default, and the optional alarm/warning conditions that be configured.
Section 5 - Viewing and Setting
the Device Configuration
Brooks Service ToolTM for QUANTIM
®
5.2.1 Diagnostic Alarm
For information on viewing the warning and alarm conditions from the Service Tool, refer to “6.1. Checking the Warning, Alarm, and Diagnostic St atus” on p. 6-1.
By default, the Service Tool monitors the device’s Diagnostic alarm status. This is the alarm signal that drives the Status and Alarm LEDs at the top of the device, and that is available to external devices at pin 3 of the device’s D-connector.
The device continuously runs diagnostics. If the diagnostics detect a fatal error, the device goes to a safe state of operation and will not respond to the setpoint. Also, the Status LED turns a solid red and the Alarm LED flashes a red code 7 (7 flashes, followed by brief delay).
NOTE: The fields in the Configure Parameters window’s Alarms:Diagnostics tab are
read-only .
5-3
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
5.2.2 Configuring Optional Warnings and Alarms
In addition to the Diagnostic alarm, you can use the Service Tool to configure the device to report warning and alarm conditions related to the following variables:
• Mass flow
• Volumetric flow
• Density
• Temperature
• Slug flow
• Valve drive
• Setpoint deviation Each of these variables has a tab in the Alarms t ab in the Configuration
Parameters window (the Density tab is shown below). The tab includes the field and options described in Table 5-1.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
5-4
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
Table 5-1 Alarm/Warning Conditions Fields and Options
Field/Option Description
Alarm Condition reported:
Flash Code The code flashed by the Alarm LED to indicate a warning/alarm
Latching If enabled, the warning/alarm condition will persist, even if the
Section 5 - Viewing and Setting
the Device Configuration
Brooks Service ToolTM for QUANTIM
Disabled. No warning/alarm will be set if low or high limits are exceeded.
Warning. A warning condition is set if low or high limits are exceeded. Alarm LED flashes green, but the device will continue normal operation.
Alarm. An alarm condition is set if low or high limits are exceeded. Alarm LED flashes red. Device will continue normal operation, unless the alarm was set by the device diagnostics. If a diagnostic alarm is set, the device enters a safe state, ignoring setpoint.
condition for this variable.
condition goes away.
®
Alarm Contact If enabled, a warning/alarm condition for this variable will cause
the open-collector output at pin 3 of the device’s D-connector to be active.
Alarm Low Limit Lower limit for this variable. If the Alarm option is set to
Warning or Alarm, and the variable’s data value drops below the low limit value, the variable’s warning/alarm condition will be active.
Alarm High Limit Upper limit for this variable. If the Alarm option is set to
Warning or Alarm, and the variable’s data value exceeds the high limit value, the variable’s warning/alarm condition will be active.
Alarm Delay Number of seconds that the warning/alarm condition must be
active before the warning/alarm is indicated on the Alarm LED and Alarm contact (and reported to the Service Tool, if a session is open).
5-5
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
5.2.3 Modifying Standard Alarm Behavior
Y ou can globally modify the behavior of alarms to automatically disable when one or both of the following conditions exist: setpoint is zero and valve override is active.
Setpoint Zero Alarm Disable. When this option is enabled, the alarms defined in T able 5-2 below will not occur if the setpoint is zero. For example, if the Mass Flow Alarm is enabled and you do not want the alarm to occur while the setpoint is zero (flow is commanded to be zero), causing flow to be less than the lower limit, set the Setpoint Zero Alarm Disable option to On (enabled).
Valve Override Alarm Disable. When this option is enabled, the alarms defined in T able 5-2 will not occur if the valve override is in the open or closed state. For example, if the Mass Flow Alarm is enabled and you do not want the alarm to occur while the valve override is set to open, causing flow to exceed the upper alarm limit, then set the V alve Override Alarm Disable option to On (enabled).
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
Table 5-2 Alarms Affected by Alarm Disable Options
Setpoint Zero Valve Override
Alarm Alarm Disable Alarm Disable
Mass Flow Yes Yes Volumetric Flow Yes Yes Density Yes No Temperature No No Slug Flow Yes No V alve Drive Y es Y es Setpoint Deviation Yes Yes
5-6
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
To set the alarm behavior when setpoint is zero or valve override is active, open the Configure menu and choose Configure Parameters.
The Configure Parameters window opens. Click the Alarms:Configuration tab. Choose whether to enable (On) or disable (Off) the option.
Section 5 - Viewing and Setting
the Device Configuration
Brooks Service ToolTM for QUANTIM
®
5.3 Viewing and Setting Configuration Parameters
Y ou can view the current settings of the control parameters from the Configuration Parameters window . Some control parameters can be modified, typically indicated by their field having a white background.
T o access the control parameters, open the Configure menu and choose Configure Parameters. The Configure Parameters window opens. Click the Control tab. The next two topics describe the Control:Response and Control:Configuration tabs.
5-7
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
5.3.1 Viewing the Control Response Parameters
The response parameters on the Control:Response tab are read-only , as shown below. However , you can adjust the proportional and integral gain values; refer to “7.2.2.3 Tuning the V alve Response (Proportional and Integral Gain)” on p. 7-18.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
5-8
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
5.3.2 Viewing and Setting the Control Configuration Parameters
The Control:Configuration tab is shown below .
Brooks Service ToolTM for QUANTIM
Section 5 - Viewing and Setting
the Device Configuration
®
The Setpoint T ype parameter on this t ab is read-only; it must be set at the factory.
The two Softstart Ramp parameters allow you to configure the ramp behavior during a soft start of the device.
T o configure the soft st art ramp behavior, select Ramp Rate from the Softs tart Ramp T ype drop-down list, then enter the ramp rate (in percent
per second) in the Softstart Ramp Rate field.
5-9
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
5.4 Viewing Device Information
To view device information, open the Configure menu and choose Configure Parameters. The Configure Parameters window opens. Click
the Device Info tab.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
The device information parameters are read-only except for the Tag Number field. For a QUANTIM QMA, you can enter the device’s tag number .
5.5 Viewing and Setting Primary and Secondary Output Parameters
Y ou can view the current settings of the primary and secondary output parameters from the Configuration Parameters window . Y ou can also modify the output data assignment, the output configuration (voltage or current), and, for a QUANTIM QMA, what happens to the output in the event of an alarm condition.
To access the primary and secondary output parameters, open the Configure menu and choose Configure Parameters. The Configure
Parameters window opens. Click the Outputs tab. This tab includes Primary and Secondary Output tabs.
5-10
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
Y ou can modify the following settings:
Out Source. Specifies the output data assignment.
Section 5 - Viewing and Setting
the Device Configuration
Brooks Service ToolTM for QUANTIM
®
NOTE: For the primary output, its data assignment must match the setpoint
assignment.
Analog Output Config. S pecifies the type of analog output—voltage or current.
Out Alarm Behavior (QUANTIM QMA only). S pecifies the behavior of the output during an alarm condition.
The Out Alarm Behavior choices are:
Downscale. Output will go to 0 V or 3.8 mA.
Hold. Output will be remain at the value in effect when the alarm condition was detected.
Track. Output will continue to report the value of the measured process variable.
Upscale. Output will go to 5.5 V or 22 mA.
5-11
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
5.6 Viewing Sensor Parameters
To view the sensor parameters, open the Configure menu and choose Configure Parameters. The Configure Parameters window opens. Click
the Sensor tab. All of the fields in the three Sensor tabs are read-only.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
5-12
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
5.7 Viewing Setpoint Parameters
Section 5 - Viewing and Setting
the Device Configuration
Brooks Service ToolTM for QUANTIM
®
T o view the setpoint parameters, open the Configure menu and choose Configure Parameters. The Configure Parameters window opens. Click
the Setpoint tab. All of the fields in this tab are read-only.
5-13
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
5.8 Viewing Valve Parameters
To view the valve parameters, open the Configure menu and choose Configure Parameters. The Configure Parameters window opens. Click
the V alve tab. All of the fields in this tab are read-only. T o adjust the valve parameters, refer to “7.2.2 Optimizing the V alve Behavior for Process Conditions” on p. 7-10.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
5-14
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
5.9 Viewing Device Variables
To view the device variable parameters, open the Configure menu
and choose Configure Parameters. The Configure Parameters window opens. Click the V ariables tab. All of the fields in the four Variables tabs are read-only.
Section 5 - Viewing and Setting
the Device Configuration
Brooks Service ToolTM for QUANTIM
®
5-15
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
5-16
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
5.10 Saving the Current Device Configuration
When you close the Configure Parameters window, any setting changes you made are saved. Prior to making changes to device configuration settings, it is recommended that you save the current settings. This will allow you to restore the device configuration to a known, working state.
T o save the current device configuration settings, open the File menu and choose Save Device Config. The Save Device Configuration File window displays.
Section 5 - Viewing and Setting
the Device Configuration
Brooks Service ToolTM for QUANTIM
®
To accept the default file name and directory, click the Save button. You can also change the file path and name by editing the file name field entry , or by clicking the Browse button to display a Save As dialog box.
5.11 Restoring a Saved Device Configuration
Y ou can restore any previously saved device configuration. To restore a saved current device configuration, open the File menu
and choose Restore Device Config. The Restore Device Configuration File window displays.
Y ou can enter the file p ath and name in the file name field or, to select the desired file, click the Browse button to display an Open dialog box. Once you have specified the file path and name, click the Restore button to restore the configuration from that file.
5-17
Section 5 - Viewing and Setting the Device Configuration
Brooks Service ToolTM for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004
THIS PAGE WAS INTENTIONALLY
LEFT BLANK
5-18
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
6 Diagnostics
This section describes how to:
• View the current state of the device’s warning/alarm conditions and diagnostic results
• Test that the device is correctly reading its input (setpoint), and correctly setting its primary and secondary outputs
• Capture device logging to a file
6.1. Checking the Warning, Alarm, and Diagnostic Status
Section 6 - Diagnostics
Brooks Service ToolTM for QUANTIM
®
The Service T ool monitors the warning or alarm st atus for all of the data variables to which an alarm can be assigned. The Service Tool also reports on the status of the device diagnostics, which run continuously .
To view the current state of warning/alarm conditions and diagnostic results, open the Diagnose menu and choose Alarm Status. The Alarms
and Diagnostics window opens.
6-1
Section 6 - Diagnostics
TM
for QUANTIM
6.1.1 Clearing Latched Warning and Alarm Conditions
As described in “5.2.2 Configuring Optional Warnings and Alarms” on p. 5­4, a Latching option can be set for a variable’s warning/alarm configuration. If latching is enabled for the variable and a warning/alarm condition occurs, the warning/alarm condition will persist even if the condition goes away . However, if you know that it is safe to operate the device, you may want to clear the latched warning/alarm condition.
To clear all latched warning and alarm conditions, click the Clear button in the Alarms and Diagnostics window .
6.2. T esting the Primary and Secondary Output s
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July, 2004Brooks Service Tool
Y ou can use the diagnostic Testing Outputs function to verify that the device is correctly setting its primary and secondary outputs.
Prior to using the Service Tool to test the output, you must connect a digital multimeter (DMM) to the output signal to measure the true output reading. Figures 6-1 through 6-3 show how to set up the DMM for the available output type configurations. Note that the secondary output type can be only current.
Note that all of the D-connector pin-outs are listed in “3.3 Device D-Connector Pin-Outs” on p. 3-7.
15
Pin 2
8
Pin 10
+
_
DMM (Precision Voltmeter)
Pin 5
+
Power Supply
13 - 30 Vdc
_
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. Tie cable shields to ground at one end only.
1
9
DB15 Connector
Pin 9
6-2
Figure 6-1 Setup for Testing Primary Output Voltage
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
Pin 5
Section 6 - Diagnostics
Brooks Service ToolTM for QUANTIM
Pin 4
®
+
Power Supply
13 - 30 Vdc
_
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. R ≤ 500 Ω.
3. Tie cable shields to ground at one end only.
1
9
DB15 Connector
Pin 9
8
15
Pin 10
Figure 6-2 Setup for Testing Primary Output Current
Pin 5
+
Power Supply
13 - 30 Vdc
_
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. R ≤ 500 Ω.
3. Tie cable shields to ground at one end only.
1
9
Pin 9
Pin 10
DB15 Connector
8
15
Pin 13
R
R
+
(Precision
Current Meter)
_
+
(Precision
Current Meter)
_
DMM
DMM
Figure 6-3 Setup for Testing Secondary Output Current
6-3
Section 6 - Diagnostics
TM
for QUANTIM
To test an output:
1. Open the Diagnose menu and choose T est Outputs>Primary Outputs or T est Outputs>Secondary Output s.
The Test Output window opens. The following screen sample shows the T est Output window for the primary output, using current.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004Brooks Service Tool
2. Change the device’s output value by:
Clicking one of the preset value buttons, or
Clicking the User button, which enables the slider , and choosing a value with the slider.
Regardless of which method you choose, the selected value is shown in the field at the bottom of the window. This is the value at which the Service Tool is telling the device to set its output.
3. Verify whether the DMM shows the same value as set using the Service T ool.
4. Try a few additional values to check that the output is correct. If the DMM reading does not match the value being set in the Test Output
window, the device’s output may need to be calibrated. Refer to “7.2.6 Calibrating the Primary and Secondary Outputs” on p. 7-28.
6-4
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
6.3. T esting the Input (Setpoint)
Y ou can use the diagnostic Testing Inputs function to verify that the device is correctly reading its input (setpoint) value.
Prior to using the Service Tool to test the setpoint, you must connect a DMM to the device’s setpoint signal to measure the true reading. Figures 6-4 and 6-5 show how to set up the DMM for the available input types.
Note that all of the D-connector pin-outs are listed in “3.3 Device D-Connector Pin-Outs” on p. 3-7.
Section 6 - Diagnostics
Brooks Service ToolTM for QUANTIM
®
Pin 8
Pin 1
Pin 5
+
Power Supply
13 - 30 Vdc
_
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. Tie cable shields to ground at one end only.
1
9
DB15 Connector
Pin 9
8
15
Figure 6-4 Setup for Testing Voltage Setpoint
+
Precision
Voltage Source
_
+
DMM (Precision Voltmeter)
_
6-5
Section 6 - Diagnostics
TM
for QUANTIM
Power Supply
13 - 30 Vdc
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. Tie cable shields to ground at one end only.
Figure 6-5 Setup for Testing Current Setpoint
Installation and Operation Manual
X-BST-Qm-eng
®
Pin 7
Pin 1
Pin 5
+
1
_
Pin 9
9
DB15 Connector
8
15
Part Number: 541B026AAA
July, 2004Brooks Service Tool
+
Precision
Current
Source
_
+
DMM
(Precision
Current Meter)
_
To test the device’s setpoint reading:
1. Open the Diagnose menu and choose T est Input s>Setpoint. The T est Inputs window opens.
2.
For the QUANTIM QMA
For the QUANTIM QMB
, select the appropriate input type.
, the Service Tool reads the input type and
automatically selects it. This selection is read-only . The device’s reading of the setpoint is displayed in the Input Signal field.
6-6
3. Verify whether the DMM shows the same value as that shown for the
device value in the Input Signal field.
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
NOTE: Input signals below the level defined by the Setpoint Low Signal Cutoff
parameter will be shown as 0 V or 4 mA.
4. Try a few additional setpoint values by changing the setpoint from the
input source, and verify that the value in the Input Signal field reflects those changes.
If the DMM reading does not match the value in the Input Signal field, the device’s setpoint may need to be calibrated. Refer to “7.2.5 Calibrating the Device Input (Setpoint)” on p. 7-24.
6.4 Capturing Device Log Data to a File
Y ou can capture the device’ s log data to a file. The log includes entries for all process variables currently displayed in the trend graph, captured at a rate specified by the user.
Section 6 - Diagnostics
Brooks Service ToolTM for QUANTIM
®
Note that you can use the main window’s Options function to determine which variables are being graphed and, hence, logged. Also, the unit s and precision of the data placed into the log are controlled by the Units of Measure tab in the Options window . For more information about the Options function, refer to “4.3 Setting the Main Window Options” on p. 4-5.
To save the log data to a file, open the Diagnose menu and choose Data Log. The Data Log Configuration window displays. From this window ,
you can perform the functions described in Table 6-1.
6-7
Section 6 - Diagnostics
TM
for QUANTIM
Table 6-1 Data Log Configuration Window Functions
Function Description
Specify the file name and path You can change the file name from the default
Append log entries to the file Select this option if you want the Service Tool to
Specify the log rate Enter the rate, in seconds, at which you want the
Start logging Click the Start button to start saving log entries
Stop logging Click the Stop button to stop saving log entries to
Installation and Operation Manual
X-BST-Qm-eng
®
entry by editing the entry in the file name field, or by clicking the Browse button to open a Save As dialog box.
add new log entries to an existing file, specified in the file name field.
log file to be updated with an entry.
to the specified file.
the specified file.
Part Number: 541B026AAA
July, 2004Brooks Service Tool
A sample listing of a log file is shown below .
6-8
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
6.4.1 Importing a Data Log File into Excel
The data log file is stored in a comma-delimited (*.csv) compatible format, so it can be easily imported into Excel.
To import a data log file into Excel:
1. In Excel, open the Find menu and choose Open.
Section 6 - Diagnostics
Brooks Service ToolTM for QUANTIM
®
2. Navigate to and open the data log file. The T ext Input wizard opens.
6-9
Section 6 - Diagnostics
TM
for QUANTIM
3. In the Original data type panel, select the Delimited option, then
click the Next button.
Step 2 of the wizard displays.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004Brooks Service Tool
4. In the Delimiters panel, select the Comma option, then click the Next
button.
In the Step 2 screen, the data log content is previewed based on your delimiter choice.
6-10
Installation and Operation Manual
X-BST -Qm-eng Part Number: 541B026AAA July, 2004
5. Verify that the data log content s have been logically separated into
columns, then click the Next button.
Step 3 of the wizard displays.
Section 6 - Diagnostics
Brooks Service ToolTM for QUANTIM
®
6. Click the Finish button. The data log file content is opened as an Excel spreadsheet.
6-11
Section 6 - Diagnostics
TM
for QUANTIM
You can now use Excel to analyze the data (for example, using the chart function as shown below), and save the data in a variety of formats, including the native Excel (*.xls) format.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July, 2004Brooks Service Tool
6-12
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
7 Troubleshooting, Tuning, and Calibration
This section provides troubleshooting strategies for responding to device alarms and problem conditions. It also includes procedures for using the Service Tool to make adjustments to the device operation to ensure optimum performance and to correct certain conditions.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
7-1
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
7.1 Troubleshooting S trategies
Tables 7-1 through 7-3 list probable causes and suggest corrective actions for responding to the device’s St atus and Alarm LEDs, startup problems, and problems that occur during operation of the device.
Table 7-1 LED Indications
LED Condition Causes Corrective Action
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004Brooks Service Tool
Status LED solid red, Alarm LED flashes red (flash code 7)
Status LED solid green, Alarm LED flashes red
Status LED solid green, Alarm LED flashes green
Device diagnostics has detected a critical fault. Device is in its safe state.
Alarm condition for a customer­configured fault. For example, flow or density is out of the specified range. Flash code indicates the data parameter that is out of range.
Customer-configured alarm conditions do not cause device to enter safe state.
Warning condition for a customer-configured fault. For example, flow or density is out of the specified range. Flash code indicates the data parameter that is out of range.
Customer-configured warning conditions do not cause device to enter safe state.
1. Use the Service Tool’s Alarms and Diagnostics function to view which diagnostic test detected the critical fault. Refer to “6.1. Checking the Warning, Alarm, and Diagnostic Status” on p. 6-1.
2. Check for external vibrations that could be interfering with the sensor frequency.
3. Check for excessive bubbles in the process line.
4. When condition is corrected, recycle power to the device to clear the alarm.
1. Read the Alarm flash code, which indicates the parameter that is out of range.
OR
Use the Service Tool’s Service Tool’s Alarms and Diagnostics function to see what parameter is out of range. Refer to “6.1. Checking the Warning, Alarm, and Diagnostic Status” on p. 6-1.
2. Determine what caused the parameter to be out of range, and correct the condition.
3. If alarm is latched (that is, does not clear when condition is corrected), use the Clear Latches button in the Service Tool’s Alarms and Diagnostics window to clear the alarm.
1. Read the Alarm flash code, which indicates the parameter that is out of range.
or
Use the Service Tool’s Service Tool’s Alarms and Diagnostics function to see what parameter is out of range. Refer to “6.1. Checking the Warning, Alarm, and Diagnostic Status” on p. 6-1.
2. Determine what caused the parameter to be out of range, and correct the condition.
3. If warning is latched (that is, does not clear when condition is corrected), use the Clear Latches button in the Service Tool’s Alarms and Diagnostics window to clear the warning.
7-2
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
Table 7-2 Common Start-Up Problems
Condition Possible Causes Corrective Action
Brooks Service ToolTM for QUANTIM
Section 7 - Troubleshooting,
T uning, and Calibration
®
Clogged valve orifice Particulates in flow stream,
causing agglomeration of fluid or chemical deposits at the orifice.
Particulates can exist if the process chemical was not properly purged before storage, or if the process fluid is not adequately filtered.
Oscillating mass flow output
1. High-pressure push gas is causing bubbles to dissolve in the fluid. Bubbles then form at the valve due to pressure drop, upsetting valve operation.
2. The piston action of the fluid pump is causing pressure oscillations, upsetting valve operation.
Use an inlet filter.
Consider using a larger orifice (although valve stability
could suffer).
Institute regular preventive maintenance back flushing.
Flush the device before shutting it down.
Use helium as a push gas, with 25- to 50-ft of Teflon tube. Helium is a light gas and can permeate the Teflon tube to release bubbles.
Use pressure regulators on the device’s inlet and outlet to stabilize pressure variations.
7-3
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Table 7-3 Common Operating Problems
Condition Possible Causes Corrective Action
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004Brooks Service Tool
Valve will not open, no full scale flow
Valve will not close or achieve complete shutoff
1. Pressure differential (Delta P) is too low, causing reduced full-scale flow.
2. The orifice is dirty or clogged, causing flow restriction.
3. Analog setpoint source is incorrectly calibrated.
4. Incompatible fluids are causing valve seat to swell.
1. Device is being used inappropriately as a shutoff valve.
Make sure that the device is being operated within its specified conditions, specifically, that the pressures are correct.
1. Use Valve Override to open the valve; this can clear some clogs (refer to “7.2.3 Using Valve Override to Control the Valve Opening” on p. 7-20). You should achieve 110% flow at the correct operating conditions.
2. With the valve open, try reverse flow with a pressure up to 100 psi.
3. Use a cleaning agent or N2 to clean the orifice.
Use the Service Tool’s manual setpoint control to verify the valve operation (refer to “4.4 Putting the Service Tool in Indicator or Control Mode” on p. 4-8). If the valve works properly, then the analog setpoint source could be incorrectly calibrated. Check the analog source signal; refer to “6.3. Testing the Input (Setpoint)” on p. 6-5.
Make sure that the device is being operated within its specified conditions, specifically, the compatibility of your process fluid against the valve seat material used in your device.
The mass flow device is a control valve, and is not designed as a shutoff valve. Some minor leak-by may be acceptable (see your device’s specification).
Use a shutoff valve for positive shutoff.
2. Device is not mounted properly.
3. Delta P is incorrect. Make sure that the device is being operated within its
4. Flow might be allowable leak-by.
5. Analog setpoint source is incorrectly calibrated.
Make sure the device is mounted in the attitude defined in its specification.
specified conditions, specifically, that the pressures are correct.
Use Valve Override to close the valve (refer to “7.2.3 Using Valve Override to Control the Valve Opening” on p. 7-20). There should be less than 1% leak-by (check the device’s specifications for exact leak-by tolerance).
Use the Service Tool’s manual setpoint control to verify the valve operation (refer to “4.4 Putting the Service Tool in Indicator or Control Mode” on p. 4-8). If the valve works properly, then the analog setpoint source could be incorrectly calibrated.
Check the analog source signal; refer to “6.3. Testing the Input (Setpoint)” on p. 6-5.
The Setpoint Low Signal Cutoff parameter may not be set correctly for the application.
7-4
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
Table 7-3 Common Operating Problems (cont.)
Condition Possible Causes Corrective Action
Brooks Service ToolTM for QUANTIM
Section 7 - Troubleshooting,
T uning, and Calibration
®
Valve will not close or achieve complete shutoff (Cont.)
Unstable control or oscillating flow
6. Flow is being indicated when there is no actual flow.
7. Valve offset and span might need to be tuned.
8. There is contamination buildup at the edge of the orifice.
1. Incorrect Delta P. Make sure that the device is being operated within its
2. The process fluid has bubbles.
Close the device’s downstream shutoff to make sure there is no flow. Open the device’s valve to equalize the pressure. If there is still a flow signal, re-zero the meter; refer to “7.2.1 Zeroing the Device’s Sensor” on p. 7-7.
Use the Service Tool’s manual setpoint control to observe flow against the valve drive position (refer to “4.4 Putting the Service Tool in Indicator or Control Mode” on p. 4-8):
At the valve offset, there should be no flow.
At the valve offset + span, there should be 100%
(fullscale) flow.
If either of these conditions do not occur, try tuning the valve offset and span. Refer to “7.2.2.1 Tuning the Valve Offset and Span” on p. 7-10.
1. Use Valve Override to open the valve; this can clear some clogs (refer to “7.2.3 Using Valve Override to Control the Valve Opening” on p. 7-10).
2. With the valve open, try reverse flow with a pressure up to 100 psi.
3. Use a cleaning agent or N2 to clean the orifice.
specified conditions, specifically, that the pressures are correct.
Correct the condition that is causing the bubbles.
3. Unstable pressures are affecting valve response.
4. If a pump is being used to supply fluid, the pump is interfering with the device’s control valve.
Check to see if the pressures are unstable. Install upstream and downstream regulators if necessary.
Use Valve Override to open the valve (refer to “7.2.3 Using Valve Override to Control the Valve Opening” on p. 7-20):
If the flow still oscillates, the pump could be causing the oscillation. Try installing a damping system to remove the effects of the pump.
If the flow oscillation stops, try tuning the valve offset and span, and the valve response. Refer to “7.2.3 Using Valve Override to Control the Valve Opening” on p. 7-10 and “7.2.2.3 Tuning the Valve Response (Proportional and Integral Gain)” on p. 7-18.
7-5
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Table 7-3 Common Operating Problems (cont.)
Condition Possible Causes Corrective Action
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004Brooks Service Tool
Flow does not match setpoint
1. Incorrect Delta P. Make sure that the device is being operated within its specified conditions, specifically, that the pressures are correct.
2. Input/output signals are not at
expected levels.
3. After new installation, wiring
to device is not correct.
1. Make sure the correct signal type is chosen: 0–5 V vs.
4–20 A.
2. Check the input/output signals against the device’s
Calibration Report.
3. Replace the outside analog control with a Brooks 0152i
Power Supply and Control Module (and Brooks cable). Check to see if setpoint and flow match with the 0152i as the setpoint source.
4. Use the Service Tool’s manual setpoint control and
verify whether flow tracks the Service Tool’s digital setpoint (refer to “4.4 Putting the Service Tool in Indicator or Control Mode” on p. 4-8).
5. Verify that the device is accurately reading its input
signal and setting its output signals using the Service Tool’s Diagnostics function. Refer to “6.2. Testing the Primary and Secondary Outputs” on p. 6-2 and "6.3. Testing the Input (Setpoint)" on p. 6-5.
If device worked before the new installation but not after, verify that the wiring to the device was installed properly.
7-6
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
7.2 Device T uning and Calibration Procedures
The Service Tool includes several functions, described in the following topics, that allow you to tune and calibrate the device:
• Zeroing the sensor, to properly set the device to the no-flow state.
• Adjusting the valve offset, span, and response, to optimize the valve behavior during process conditions.
• Overriding the setpoint control of the valve, to fully open or close the valve when investigating the source of flow problems.
• Calibrating the device’s input and output signals, to ensure that the device is accurately reading setpoint and sending its output signal.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
7.2.1 Zeroing the Device’s Sensor
The Service Tool’s Zero Sensor function allows you to zero the device’s sensor when there is a no-flow condition. The device also includes a Zero pushbutton that performs this same function.
7.2.1.1 Accessing the Service Tool’s Zero Sensor Function
To open the Zero Sensor function in the Service T ool, open the Tune
menu and choose Zero Sensor. The Zero Sensor window displays.
7-7
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004Brooks Service Tool
Zero flow configuration values are shown in the read-only fields in the Configure panel.
Clicking the Start button begins the zero sensor process. The process takes about half a minute. When the zeroing completes successfully , the
Status field will display “Passed.”
7.2.1.2 Procedure for Zeroing the Sensor
To zero the device’s sensor:
1. Allow the device to be powered on for 45 minutes so that it achieves its operating temperature.
2. Use the process gas or fluid.
3. Set the Valve Override to open to start flow through the device. For information on using the Valve Override, refer to “7.2.3 Using Valve
Override to Control the V alve Opening” on p. 7-20.
4. If the process fluid is liquid, allow it to flow long enough to purge any
air or gas bubbles from the lines and device.
7-8
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
NOTE: It may be necessary to cycle the Valve Override from open to closed to clear
entrapped air bubbles. If the Valve Override control is not available, then cycle the setpoint between 0 and 100%. Refer to “7.2.4 Cycling the Setpoint” on p. 7-22.
5. Close the downstream shutoff valve.
6. Wait one or two minutes for the flow signal to drop to zero.
7. Set the Valve Override to closed, make sure the setpoint is zero, and
wait again for the signal stabilization.
The device should be full of process gas or fluid with no pressure differential and no vibration.
8. Using the Service Tool’s Zero Sensor function or the device’s Zero
pushbutton, zero the device:
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
If using the Zero Sensor function,
Sensor window. The Status field will display “Passed” when the zeroing operation completes.
If using the pushbutton,
LED flashes red, then release the pushbutton. A successful zeroing operation is indicated by the St atus LED being a steady green.
The device will accept the zero command only when flow is equal to or less than 1% and there is no vibration. Otherwise, the operation will fail, and the St atus LED will be a steady red. If using the Zero Sensor function, a failed indication will be displayed in the Status field of the Zero Sensor window . You will have to recycle power to the device to clear the alarm condition.
press the pushbutton until the device’s St atus
click the Start button in the Zero
7-9
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
7.2.2 Optimizing the Valve Behavior for Process Conditions
The Service Tool includes two functions, described in the following topics, that allow you to tune the valve to optimize its performance:
• Tuning the valve’ s of fset (flow just starting, 1–2%) and span (at
flow + span, flow just at fullscale, 100%).
• Adjusting the valve’s response to make sure it is operating within its
specification.
7.2.2.1 T uning the Valve Offset and Span
Device performance can be affected by contaminants or pressure changes. Tuning the valve offset and sp an can help to optimize the valve response.
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004Brooks Service Tool
At process conditions with process flow:
• The valve offset value corresponds to the valve position (measured as a
percentage from fully closed to fully open) at which flow just begins. The factory default is typically around 20%. See your device’s settings for the actual offset value.
• The valve span value is used to determine the valve position—
offset + span—at which flow is just at fullscale.
• The valve leaktight offset value corresponds to the valve position that
guarantees no flow or leak-through at a setpoint of 0%.
NOTE: The leaktight offset value is set at the factory for 10%. This is applied at zero
setpoint to add extra preload to achieve better shutoff. Do not change this value.
All valve offset values are calibrated at the factory at the conditions specified on the device’s order.
NOTE: It is highly recommended that you use the Service T ool’s Save Device Config
function to save the device’s current configuration before making any changes. This will allow you to restore a working configuration if the valve tuning results in incorrect settings or faulty operation. For more information, refer to “5.10 Saving the Current Device Configuration” on p. 5-17.
7-10
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
You should first check to see whether the offset and span values require adjustment.
To check the offset and span values:
1. Open the Tune menu and choose Valve. The T une Valve window opens.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
2. Note the offset and span values in the Direct Parameter Update
panel, and check whether they are the correct settings for your application.
3. Set inlet and outlet pressure conditions to match those specified on
the device’s order or on its Calibration Report.
4. In the Tune Valve window, select the Manual option to enable the
valve control slider and increment/decrement field.
5. Starting at a value of zero, use the slider or increment button to slowly
increase the valve value until flow begins. At the 1–2% flow rate, the valve value should be very close to the offset value shown in the Direct Parameter Update panel’s Valve Offset field.
7-11
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
6. Continue to increment up until 100% flow is reached. The valve value
should be equal to offset + span [for example, 20% (offset) + 15% (span) = 35% valve drive value].
Allow time for flow to stabilize. It may drop slightly .
You should adjust the offset and span values if one of the following are true:
• The process conditions are now different from that specified on the
device’s order .
• The valve value that produces a flow rate of 1–2% does not match the
current offset value in the Direct Parameter Update panel’s Valve Offset field.
• The valve value that produces a flow rate of 100% does not match the
current offset + span values in the Direct Parameter Update panel’s Valve Offset and Valve Span fields.
Part Number: 541B026AAA
July , 2004Brooks Service Tool
To adjust the offset and span values:
1. Set inlet and outlet pressure conditions to your process or application
specs.
2. In the Tune Valve window, select the Manual option to enable the
valve control slider and increment/decrement field.
3. Starting at a value of zero, use the slider or increment button to slowly
increase the valve value until flow begins. Make sure you have stopped at 1-2% flow.
NOTE: Do not overshoot and then decrease the valve value to locate the offset
position. Because of valve hysteresis, you must start at zero and find the offset value only by increasing up from zero.
4. Press the Save Offset button to save the new offset value. You can also manually enter a new offset value by selecting the Direct
Parameter Update option, entering the value in the V alve Offset field, and pressing the Enter key.
5. Once the offset value is determined and set, continue to slowly
increment the valve value. As flow rate approaches 100%, go slowly until 100% is achieved.
7-12
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
6. Allow flow rate to stabilize. When the flow rate has stabilized, press
the Save Span button to save the new span value.
You can also manually enter a new span value by selecting the Direct Parameter Update option, entering the value in the V alve Span field, and pressing the Enter key .
7. Manually adjust the valve to determine whether or not the newly
saved offset and span values produce the desired flow rates, that is, 1– 2% with the valve drive at offset, and 100% with the valve drive at offset + span.
7.2.2.2 Cleaning the Orifice for a QUANTIM QMA
The QUANTIM QMA has a removable orifice. If the device has become clogged, or you suspect that contamination is interfering with the valve’s function and performance, you can remove the orifice for cleaning.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
Prior to performing the cleaning procedure, observe the following restrictions and guidelines:
• The QUANTIM QMA is a metal sealed device. Never loosen or remove
the QUANTIM QMB assembly bolts. This will compromise the seal integrity.
• Opening the QUANTIM body will void the warranty .
• The QUANTIM valve is a control valve and is not designed as a shutoff
valve. Installation of a shutoff valve on the outlet of the device is recommended to achieve positive shutoff.
• Due to small orifice sizes, filters are highly recommended on the inlet to
prevent contamination from entering the device.
• Sometimes a swelled valve seat will result in no flow response to a
setpoint. Incompatible fluids can cause a seat to swell enough to affect valve control. Check the compatibility of your process fluid against the valve seat material used in your device.
• Before removing the orifice, try cleaning or removing possible clogs by
using reverse flow with an inert gas such as nitrogen and the valve fully open using the Valve Override control (refer to “7.2.3 Using Valve Override to Control the V alve Opening” on p. 7-20). Do not use reverse flow pressures higher than 100 psi.
7-13
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
• It is highly recommended that you use the Service Tool’s Save Device
Config function to save the device’s current configuration before making any changes. This will allow you to restore a working configuration if the valve tuning results in incorrect settings or faulty operation. For more information, refer to “5.10 Saving the Current Device Configuration” on p. 5-17.
7.2.2.2.1 Removing and Cleaning the QUANTIM QMA’ s Orifice
To remove and clean the QUANTIM QMA’s orifice:
1. Locate and remove the external fitting on the outlet side of the device.
NOTE: Do not remove the body end block.
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004Brooks Service Tool
7-14
Remove this fitting
to access orifice
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
2. Insert a 2-mm Allen key into the outlet port until it seats into the
orifice.
Orifice
3. Unscrew the orifice until it is removed from the device body.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
4. Clean the orifice with a compatible fluid and clear it with pressurized
air or N2.
7.2.2.2.2 Reinstalling the Orifice and Checking the Valve Offset and Span Values
The removal and reinstallation of an orifice in the QUANTIM QMA is very delicate and requires some patience to return the device to factory specifications.
1. Turn off the inlet shutoff for no flow.
2. In the Service Tool, open the Tune menu and choose Valve. The T une Valve window opens.
7-15
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004Brooks Service Tool
3. Note the offset and span values in the Direct Parameter Update
panel.
4. Select the Manual option to enable the valve control slider and
increment/decrement field.
5. Use the slider or increment button to set the valve drive value to the
offset value.
6. Insert the orifice into the device outlet.
7. Turn the orifice using a 2-mm copper beryllium Allen key
(nonmagnetic) from the factory until it bottoms out.
NOTE: You can use a 2-mm steel Allen key, but you will have to remove it after each
adjustment to see accurate valve position values, as a steel key will affect the magnetism within the valve.
8. Turn on the inlet process fluid.
9. Adjust for the specified Delta P as stated on the device’s Calibration
Report.
7-16
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
10. Turn the orifice counterclockwise until flow starts, then turn the
orifice clockwise until flow goes back to zero.
This is a very fine adjustment. Af ter each adjustment, remove the Allen key and allow some time for the signal to stabilize at a given flow rate. The point is to adjust the orifice slowly until the flow just gets to zero
1 1. Repeat the orifice adjustment until you just achieve shutoff at the offset value.
12. Increment the valve drive value until you get 100% fullscale flow. The valve value at 100% fullscale flow should be offset + span [for
example, 20% (offset) + 15% (span) = 35% valve drive value]. If the valve value at 100% flow is 10% greater than the original
offset + span value (that is, based on the offset and span values in the Direct Parameter Update panel), then the orifice is probably in too tight. Readjust the orifice.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
13. When you have completed the orifice adjustments, change the valve
drive value to achieve 100% flow, then back to 0 flow , to verify that there is fullscale flow at close to the offset + span value and shutoff at the offset value.
NOTE: The leaktight offset value is set at the factory for 10%. This is applied at zero
setpoint to add extra preload to achieve better shutoff. Do not change this value.
14. Finally, control the device with a setpoint. Monitor the setpoint, flow ,
and valve drive value. Assuming that the current process conditions match the factory calibration
process conditions for the device, you should get fullscale flow at around the factory-set offset + span value. DO NOT change the span value. The valve will open to where it needs to achieve fullscale flow.
For example, for the following factory-set values:
Offset value = 20% Span value = 8%
At a valve value of 20% you should get shutoff, and at a valve value of 28% you should get 100% flow.
7-17
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
The actual valve value to achieve fullscale flow may be a couple of percentage points higher than the factory-set span value. This is acceptable. However, if the actual sp an value is more than 10% higher than the factory-set value, then you may want to adjust the orifice counterclockwise slightly . Do not change the span value to correct for a misadjusted orifice.
At the setpoint setting of 0%, the valve value will be offset minus the leaktight offset (for example, 20% – 10% = 10%).
7.2.2.3 T uning the Valve Response (Proportional and Integral Gain)
The proportional and integral gain settings control the rate at which the valve responds to changes in setpoint. You can adjust these values to tune the valve’s response.
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004Brooks Service Tool
When tuning the valve response, you will use the Cycle Setpoint function to cycle the valve opening between two selected setpoint values. Y ou will then use the trend graph to check that the valve’s response time is within specification, making adjustments accordingly .
To tune the valve response:
1. Make sure the trending graph can display the range of setpoints
between which you want to cycle. For more information about setting the maximum and minimum setpoints displayed on the trend graph, refer to “4.3 Setting the Main Window Options” on p. 4-5.
2. Open the Tune menu and choose Response. The Tune Response Form window opens.
7-18
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
3. Click the Setpoint Cyclical Control button (or open the Control
menu and choose Cycle Setpoint).
The Cycle Setpoint window opens.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
4. Use the fields in the Cycle Setpoint window to establish the cycling
condition you want to use:
Enter the upper setpoint value (in flow units) and cycle time (at least 6 seconds) in the High Setpoint fields.
Enter the lower setpoint value (in flow unit s) and cycle time (at least 6 seconds) in the Low Setpoint fields.
Select the Continuous option to have the cycling continue until you manually stop it, or enter the number of cycles in the Cycle field.
5. When you have completed entering the cycling conditions, click the
Start button to start the cycling process.
7-19
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
6. Observe the mass flow and valve drive graph to see how well the
valve is responding to the changes in the setpoint.
In general, flow response should be as fast as possible with minimum overshoot (less than 5%).
7. If the valve response needs tuning, go to the Tune Response Form
window and change the Proportional Gain and Integral Gain values until the valve response is satisfactory.
When adjusting the Proportional and Integral Gain values in the Tune Response Form window, observe the following guidelines:
• Increasing the proportional gain value and/or decreasing the integral gain
value speeds up the response time. Conversely , decreasing the proportional gain value and/or increasing the integral gain value slows the response time.
• The Proportional Gain value is nominally 0.6. The Integral Gain value is
nominally 0.2. To achieve the desired flow response, first adjust the Proportional Gain value in 0.1 increments, and then adjust the Integral Gain value in 0.05 increments.
Part Number: 541B026AAA
July , 2004Brooks Service Tool
7.2.3 Using Valve Override to Control the Valve Opening
You can use the Valve Override (VOR) control to override the setpoint and control the opening of the valve. Valve Override is available by:
• Applying an input signal to pin 12 of the device’s 15-pin D-connector or
• Using the Service Tool’s Valve Override
The following two topics explain both of these methods.
7-20
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
7.2.3.1 Using Pin 12 to Control the Valve Override
Pin 12 of the device’s 15-pin D-connector is a valve override input:
• Grounding pin 12 causes the device to go into VOR Closed.
• Applying a voltage greater than 5 V causes the device to go into VOR
Open.
• Applying a voltage from 1 to 3 V causes the device to go into VOR
Normal (device tracks setpoint). Note also that simply applying no connection (or “floating”) pin 12 of the D-connector will produce the same result.
Note that an input signal to pin 12 takes precedence of the Service Tool’s V alve Override function. If the device is in VOR Open or VOR Closed due to the signal at pin 12, then in the Service Tool:
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
• The status message “Analog Override Active” will appear at the top of
the Valve Override Control window
• The V alve Override Control buttons will not change
• The Valve Override in the Status Bar is the actual state of the device
7-21
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
7.2.3.2 Using the Service Tool’s Valve Override Function
To use the Service Tool’s Valve Override control, open the Control
menu and choose Valve Override. The Valve Override Control window opens.
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004Brooks Service Tool
7.2.4 Cycling the Setpoint
Set the valve position by:
• Clicking one of the preset value buttons—Normal (tracks setpoint),
Closed, or Open.
or
• Clicking the Manual button and choosing a value using the slider or the
increment/decrement buttons, or by entering the value in the field next to the increment/decrement buttons and pressing the Enter key .
Regardless of which method you choose, the selected value is shown in the field below the slider, and reflected in the valve drive graph.
The Cycle Setpoint function allows you to cycle the setpoint between low and high values that you define. You may need to do this to purge air or gas bubbles from the lines and device. You also use the Cycle Setpoint function when tuning the valve response, as described in “7.2.2.3 Tuning the V alve Response (Proportional and Integral Gain)” on p. 7-18.
7-22
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
To cycle the setpoint:
1. Open the Control menu and choose Cycle Setpoint. The Cycle Setpoint window opens. (Note that, when tuning the valve, you
can access the Cycle Setpoint window by clicking the Setpoint Cyclical
Control button in the Tune Response Form window.)
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
2. Use the fields in the Cycle Setpoint window to establish the cycling
condition you want to use:
Enter the upper setpoint value (in flow units) and cycle time (at least 6 seconds) in the High Setpoint fields.
Enter the lower setpoint value (in flow unit s) and cycle time (at least 6 seconds) in the Low Setpoint fields.
Select the Continuous option to have the cycling continue until you manually stop it, or enter the number of cycles in the Cycle field.
3. When you have completed entering the cycling conditions, click the
Start button to start the cycling process.
The trend graph should show the flow cycling between the values you specified.
With the setpoint cycling, you can:
• Suspend the cycling by clicking the Hold button.
• Stop the cycling by clicking the Stop button.
• Restart the cycling by clicking the Start button.
7-23
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
7.2.5 Calibrating the Device Input (Setpoint)
You calibrate the device’s input to ensure that the device is accurately reading the setpoint from the analog control.
Prior to using the Service Tool to calibrate the device input, you must connect a digital multimeter (DMM) to the device’s setpoint signal, which you will use to verify the actual signal level. Figures 7-1 and 7-2 show how to set up the DMM for the available input types.
Note that all D-connector pin-outs are listed in “3.3 Device D-Connector Pin-Outs” on p. 3-7.
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004Brooks Service Tool
Pin 8
Pin 1
Pin 5
+
Power Supply
13 - 30 Vdc
_
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. Tie cable shields to ground at one end only.
1
9
DB15 Connector
Pin 9
8
15
Figure 7-1 Setup for Calibrating to a Voltage Setpoint
+
Precision
Voltage
Source
_
+
DMM
(Precision
Voltmeter)
_
7-24
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
Power Supply
13 - 30 Vdc
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. Tie cable shields to ground at one end only.
Figure 7-2 Setup for Calibrating to a Current Setpoint
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
+
Precision
Current
Pin 7
Pin 1
Pin 5
+
1
_
Pin 9
9
DB15 Connector
8
15
Source
_
+
DMM
(Precision
Current Meter)
_
®
To calibrate the device’s input:
1. Open the Tune menu and choose Calibrate Inputs. The Calibrate Input Wizard opens.
2.
For a QUANTIM QMA
, select the appropriate input type.
For a QUANTIM QMB
, the Service Tool reads the input type and
automatically selects it. This selection is read-only.
7-25
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
3. Make sure that the DMM is set up as instructed in Step 1 of the wizard
(and shown in Figure 7-1 or 7-2), then click the Next button to advance to Step 2 of the wizard.
Part Number: 541B026AAA
July , 2004Brooks Service Tool
4. Set the analog source to the value shown in Step 2 of the wizard (0 V
or 4 mA). Check the DMM reading to verify that the setpoint from the analog source is accurate.
5. Click the Next button to advance to Step 3 of the wizard.
7-26
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
6. Set the analog source to the value shown in Step 3 of the wizard (5 V
or 20 mA). Check the DMM reading to verify that the setpoint from the analog source is accurate.
7. Click the Next button to advance to Step 4 of the wizard.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
8. As instructed in Step 4 of the wizard, vary the setpoint and verify that
the value in the Input Signal field, which is the value as read by the device, is the same as the value on the DMM.
If the value in the Input Signal field is the same as that on the DMM for several setpoints, the calibration is complete. Click the Finished button to save the calibration data.
If the value in the Input Signal field is not the same as that on the DMM, you can click the Back button to go back and repeat the steps in the wizard.
At any point in the wizard, you can click the Cancel button to cancel the calibration. You will be prompted to confirm the cancellation. If the calibration is cancelled, the most recently saved calibration data is restored.
7-27
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
7.2.6 Calibrating the Primary and Secondary Outputs
You calibrate the device’s outputs to ensure that the device is accurately setting the primary and secondary output signals and, thus, maintaining the correct flow relative to the setpoint.
Prior to using the Service Tool to calibrate the device outputs, you must connect a DMM to the device’s output signal, which you will use to verify the actual signal level. Figures 7-3 through 7-5 show how to set up the DMM for the available output types.
Note that all D-connector pin-outs are listed in “3.3 Device D-Connector Pin-Outs” on p. 3-7.
®
Installation and Operation Manual
X-BST-Qm-eng
Part Number: 541B026AAA
July , 2004Brooks Service Tool
Pin 5
Pin 2
+
Power Supply
13 - 30 Vdc
_
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. Tie cable shields to ground at one end only.
1
9
DB15 Connector
Pin 9
8
15
Pin 10
Figure 7-3 Setup for Calibrating to a Primary Output Voltage
Pin 4
Pin 5
+
Power Supply
13 - 30 Vdc
_
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. R ≤ 500 Ω.
3. Tie cable shields to ground at one end only.
1
9
DB15 Connector
Pin 9
8
15
Pin 10
R
+
(Precision Voltmeter)
_
+
(Precision
Current Meter)
_
DMM
DMM
7-28
Figure 7-4 Setup for Calibrating to a Primary Output Current
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
Pin 5
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
+
Power Supply
13 - 30 Vdc
_
Notes:
1. Refer to the DMM specifications, as provided by the Service Tool.
2. R ≤ 500 Ω.
3. Tie cable shields to ground at one end only.
1
9
Pin 9
Pin 10
DB15 Connector
8
15
Pin 13
R
+
(Precision
Current Meter)
_
Figure 7-5 Setup for Calibrating to a Secondary Output Current
To calibrate one of the device’s outputs:
1. Open the Tune menu and choose Calibrate Outputs>Primary
Output or Calibrate Outputs>Secondary Output.
The Calibrate Output Wizard opens.
DMM
The Service Tool recognizes the output type and automatically selects it.
2. Make sure that the DMM is set up as instructed in Step 1 of the wizard
(and shown in Figure 7-3, 7-4, or 7-5), then click the Next button to advance to Step 3 of the wizard.
7-29
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
The Service Tool forces the device to output the minimum signal level.
Part Number: 541B026AAA
July , 2004Brooks Service Tool
3. As instructed in Step 3 of the wizard, enter the value displayed on the
DMM in the DMM Reading field, then press the Enter key. Continue entering the DMM value until it is within 0.2% of the target value (that is,
0.00–0.01 V or 3.96–4.04 mA).
4. When the DMM value is within 0.2% of the target value, click the Next
button to advance to Step 4 of the wizard.
7-30
The Service Tool forces the device to output the maximum signal level.
Installation and Operation Manual
X-BST-Qm-eng Part Number: 541B026AAA July , 2004
5. As instructed in Step 4 of the wizard, enter the value displayed on the
DMM in the DMM Reading field, then press the Enter key. Continue entering the DMM value until it is within 0.2% of the target value (that is,
4.99–5.01 V or 19.96–20.04 mA).
6. When the DMM value is within 0.2% of the target value, click the Next
button to advance to Step 5 of the wizard.
Section 7 - Troubleshooting,
T uning, and Calibration
Brooks Service ToolTM for QUANTIM
®
7. As instructed in Step 5 of the wizard, click each of the output value
buttons, checking to make sure that the DMM readings are within 0.2% of the selected output values.
If the DMM readings are within 0.2% for all three test values, the calibration is complete. Click the Finished button to save the calibration data.
If any of the DMM readings is not within 0.2% of the corresponding test button’s value when clicked, you can click the Back button to go back and repeat the steps in the wizard.
At any point in the wizard, you can click the Cancel button to cancel the calibration. You will be prompted to confirm the cancellation. If the calibration is cancelled, the most recently saved calibration data is restored.
7-31
Section 7 - T roubleshooting, Tuning, and Calibration
TM
for QUANTIM
Installation and Operation Manual
X-BST-Qm-eng
®
Part Number: 541B026AAA
July , 2004Brooks Service Tool
THIS PAGE WAS
INTENTIONALLY
LEFT BLANK
7-32
Loading...