Micro Motion™ 5700 Transmitters with
Configurable Inputs and Outputs
Configuration and Use Manual
October 2021
Safety messages
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully
before proceeding to the next step.
Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the
instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The following are
available: the EU declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings
and Instructions. In addition the IECEx Installation Instructions for installations outside of the European Union and the CSA
Installation Instructions for installations in North America are available on the internet at www.emerson.com or through your local
Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at
www.emerson.com. For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration
manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government
transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept
your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion
Customer Service department.
Emerson Flow customer service
Email:
• Worldwide: flow.support@emerson.com
• Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South AmericaEurope and Middle EastAsia Pacific
United States800-522-6277U.K. and Ireland0870 240 1978Australia800 158 727
Appendix IDefault values and ranges...................................................................................... 357
Configuration and Use Manual7
ContentsConfiguration and Use Manual
October 2021MMI-20025166
8Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualBefore you begin
MMI-20025166October 2021
1 Before you begin
1.1
This manual helps you configure, commission, use, maintain, and troubleshoot Micro Motion 5700
transmitters with configurable inputs and outputs.
Important
This manual assumes that:
• The transmitter has been installed correctly and completely according to the instructions in the
• Users understand basic transmitter and sensor installation, configuration, and maintenance concepts and
About this manual
transmitter installation manual
procedures
1.2 Hazard messages
This document uses the following criteria for hazard messages based on ANSI standards Z535.6-2011
(R2017).
DANGER
Serious injury or death
WARNING
Serious injury or death
will occur if a hazardous situation is not avoided.
could occur if a hazardous situation is not avoided.
CAUTION
Minor or moderate injury will or could occur if a hazardous situation is not avoided.
NOTICE
Data loss, property damage, hardware damage, or software damage can occur if a situation is not avoided.
There is no credible risk of physical injury.
Physical access
NOTICE
Unauthorized personnel can potentially cause significant damage and/or misconfiguration of end users'
equipment. Protect against all intentional or unintentional unauthorized use.
Physical security is an important part of any security program and fundamental to protecting your system.
Restrict physical access to protect users' assets. This is true for all systems used within the facility.
Configuration and Use Manual9
Before you beginConfiguration and Use Manual
October 2021MMI-20025166
1.3 Related documents
You can find all product documentation on the product documentation DVD shipped with the product or at
www.emerson.com.
See any of the following documents for more information:
• Micro Motion 5700 Product Data Sheet
• Micro Motion 5700 Transmitters with Configurable Inputs and Outputs: Installation Manual
• Coriolis Flow Meter with Micro Motion 5700 Transmitters Safety Manual for Safety Instrumented Systems (SIS)
• Micro Motion Oil and Gas Production Supplement
• Micro Motion Model 5700 Transmitter Net Oil Calculations Multiwell Supplement
• Micro Motion 5700 Transmitters with a Marine Bunker Transfer Package: Application Manual
• Retrofit a Micro Motion 5700 Configurable Inputs and Outputs Display with a Wi-Fi Display Replacement
Instruction
• Replacing the Junction Box for the 4200 Transmitter and the 5700 Transmitter
• Replacing the Sensor Cable for the 4200 Transmitter and the 5700 Transmitter
• Sensor installation manual
1.4
You can use several different communications tools and protocols to interface with the transmitter, use
different tools in different locations, or use different tools for different tasks.
ToolSupported protocols
DisplayNot applicable
ProLink™ III• HART®/Bell 202
Field communicatorHART/Bell 202
For information about how to use the communication tools, see the appendices in this manual.
Communications tools and protocols
• HART/RS-485
• Modbus®/RS-485
• Service port
• Modbus TCP
• Android tablet, iOS tablet, and PC support for Wi-Fi
Tip
You may be able to use other communications tools, such as AMS™ Suite: Intelligent Device Manager, or the
Smart Wireless
manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at
www.emerson.com.
10Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this
Configuration and Use ManualQuick start
MMI-20025166October 2021
2 Quick start
2.1
The transmitter must be powered up for all configuration and commissioning tasks, or for process
measurement.
Procedure
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to
10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long
enough to allow components to reach ambient temperature, allow the electronics to warm up for
approximately 10 minutes before relying on process measurements. During this warm-up period, you may
observe minor measurement instability or inaccuracy.
Power up the transmitter
1. Follow appropriate procedures to ensure that a new device in the control system does not interfere
with existing measurement and control loops.
2. Verify that the cables are connected to the transmitter as described in the installation manual.
3. Verify that all transmitter and sensor covers and seals are closed.
WARNING
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are
tightly closed. For hazardous area installations, applying power while housing covers are removed or
loose can cause an explosion resulting in injury or death.
Turn on the electrical power at the power supply.
4.
The transmitter will automatically perform diagnostic routines. During this period, the Transmitter
Initializing alert is active. The diagnostic routines should complete in approximately 30 seconds.
2.2 Check meter status
Check the meter for any error conditions that require user action or that affect measurement accuracy.
Procedure
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error
conditions. During the power-up sequence, the Transmitter Initializing alert is active. This
alert should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Table 2-1: Status LED and device status
Status LED conditionDevice status
Solid greenNo alerts are active.
Solid yellowOne or more alerts are active with Alert Severity = Out of Specification,
Maintenance Required, or Function Check.
Configuration and Use Manual11
Quick startConfiguration and Use Manual
October 2021MMI-20025166
Table 2-1: Status LED and device status (continued)
Status LED conditionDevice status
Solid redOne or more alerts are active with Alert Severity = Failure.
Flashing yellow (1 Hz)The Function Check in Progress alert is active.
Table 2-2: Wi-Fi status LED and device status
Wi-Fi status LED conditionDevice status
Solid blue lightWi-Fi is active and not connected to a device.
Blinking blue lightWi-Fi is active and connected to a device.
No lightWi-Fi is not active.
2.3 Commissioning wizards
The transmitter menu includes a Guided Setup to help you move quickly through the most common
configuration parameters. ProLink III also provides a commissioning wizard.
By default, when the transmitter starts up, the Guided Setup menu is offered. You can choose to use it or not.
You can also choose whether or not Guided Setup is displayed automatically.
• To enter Guided Setup upon transmitter startup, choose Yes at the prompt.
• To enter Guided Setup after transmitter startup, choose Menu→Startup Tasks.
• To control the automatic display of Guided Setup, choose Menu→Configuration→Guided Setup.
For information on the ProLink III commissioning wizard, see the Micro Motion ProLink III with ProcessVizSoftware User Manual.
As the commissioning wizards are self guided, they are not documented in detail.
2.4
For all configuration tools except the display, you must have an active connection to the transmitter to
configure the transmitter.
Procedure
Identify the connection type to use, and follow the instructions for that connection type in the appropriate
appendix.
Communications toolConnection type to useInstructions
ProLink IIIHART
Make a startup connection to the transmitter
Using ProLink III with the transmitter
Wi-Fi
Configure Wi-Fi
Field communicatorHARTUsing a field communicator with the
transmitter
12Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
The transmitter clock provides timestamp data for alerts, service logs, history logs, and all other timers and
dates in the system. You can set the clock for your local time or for any standard time you want to use.
Tip
You may find it convenient to set all of your transmitter clocks to the same time, even if the transmitters are
in different time zones.
Procedure
Select the time zone that you want to use.
1.
2. If you need a custom time zone, select Special Time Zone and enter your time zone as a difference
from UTC (Coordinated Universal Time).
3. Set the time appropriately for the selected time zone.
Tip
The transmitter does not adjust for Daylight Savings Time. If you observe Daylight Savings Time, you
must reset the transmitter clock manually.
4. Set the month, day, and year.
The transmitter tracks the year and automatically adds a day for leap years.
Field CommunicatorConfigure→Manual Setup→HART→Communications
The transmitter can have both a HART address and a Modbus address. These addresses are used by service
tools and hosts to communicate with the transmitter. The transmitter can also have a tag. The tag identifies
the transmitter and can also be used for HART communications.
Procedure
• If you plan to make HART connections to the transmitter, set the HART address.
Set the transmitter addresses and tags
— Default: 0
— Range: 0–15
Tip
Leave HART Address at the default (0) unless you are in a multidrop environment.
• If you plan to make Modbus connections to the transmitter, set the Modbus address.
Configuration and Use Manual13
Quick startConfiguration and Use Manual
October 2021MMI-20025166
— Default: 1
— Range: 1–15, 23–47, 64–79, 96–110
Tip
— If you need an address that is out of range, you can disable Modbus ASCII Support. When Modbus
ASCII Support is disabled, the Modbus address can be set to 1–127, excluding 111. 111 is reserved for
the service port address. However, you will not be allowed to use Modbus ASCII (7-bit) for connections
to the transmitter. You must use Modbus RTU (8-bit) instead.
— The other Modbus parameters can be left at the default values unless you have trouble making the
connection.
• Set the transmitter tag and/or the long tag.
The transmitter will respond to connection requests that use either the transmitter tag or the long tag.
The long tag is supported only by HART 7. The transmitter accepts both HART 5 and HART 7 connection
requests.
2.7
DisplayMenu→About→Licenses→Licensed Features
ProLink IIIDevice Tools→Device Information→Licensed Features
Field communicatorOverview→Device Information→Licenses
The transmitter license controls the features that are enabled on the transmitter, including both software
applications and I/O channels. You can view the licensed features to ensure that the transmitter was ordered
with the required features.
Licensed features are purchased and available for permanent use. The options model code represents the
licensed features.
A trial license allows you to explore features before purchasing. The trial license enables the specified features
for a limited number of days. This number is displayed for reference. At the end of this period, the feature will
no longer be available.
To purchase additional features or request a trial license, document the Unique ID Number and current
license key from your transmitter and contact customer service. To enable the additional features or trial
license, you will need to install the new license on the transmitter.
14Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualQuick start
MMI-20025166October 2021
You can set several parameters that identify or describe the transmitter and sensor. These parameters are not
used in processing and are not required.
Procedure
1.
Set informational parameters for the transmitter.
a) Set Transmitter Serial Number to the serial number of your transmitter.
The transmitter serial number is provided on the metal tag that is attached to the transmitter
housing.
b) Set Descriptor to any desired description of this transmitter or measurement point.
c) Set Message to any desired message.
d) Verify that Model Code (Base) is set to the base model code of the transmitter.
The base model code completely describes your transmitter, except for the features that can be
licensed independently. The base model code is set at the factory.
e) Set Model Code (Options) to the options model code of the transmitter.
The options model code describes the independent features that have been licensed for this
transmitter. The original options model code is set at the factory. If you license additional
options for this transmitter, Emerson will supply an updated options model code.
For a field communicator, configuring model code options is not available for this release.
2. Set informational parameters for the sensor.
a) Set Sensor Serial Number to the serial number of the sensor connected to this transmitter.
The sensor serial number is provided on the metal tag that is attached to the sensor case.
b) Set Sensor Material to the material used for the sensor.
c) Set Sensor Liner to the material used for the sensor liner, if any.
d) Set Flange Type to the type of flange that was used to install the sensor.
Do not set Sensor Type. Sensor Type is set or derived during characterization.
2.9 Characterize the meter (if required)
DisplayMenu→Configuration→Sensor Parameters
ProLink IIIDevice Tools→Calibration Data
Field communicatorConfigure→Manual Setup→Characterization
Configuration and Use Manual15
Quick startConfiguration and Use Manual
October 2021MMI-20025166
Characterizing the meter adjusts your transmitter to match the unique traits of the sensor it is paired with.
The characterization parameters (also called calibration parameters) describe the sensor’s sensitivity to flow,
density, and temperature. Depending on your sensor type, different parameters are required.
Values for your sensor are provided on the sensor tag or the calibration certificate.
• If your transmitter was ordered with a sensor, it was characterized at the factory. However, you should still
verify the characterization parameters.
• Perform a characterization whenever you replace a core processor.
Procedure
1.
Optional: Specify Sensor Type.
• Straight Tube (T-Series sensors)
• Curved Tube (all sensors except T-Series)
Note
Unlike earlier transmitters, the 5700 derives Sensor Type from the user-specified values for FCF and K1
in combination with an internal ID.
2. Set the flow calibration factor: FCF (also called Flow Cal or Flow Calibration Factor). Be sure to include
all decimal points.
Set the density characterization parameters: D1, D2, TC, K1, K2, and FD. (TC is sometimes shown as
3.
DT.)
4. Apply the changes as required by the tool you are using.
The transmitter identifies your sensor type, and characterization parameters are adjusted as required:
• If Sensor Type changed from Curved Tube to Straight Tube, five characterization parameters are
added to the list.
• If Sensor Type changed from Straight Tube to Curved Tube, five characterization parameters are
removed from the list.
• If Sensor Type did not change, the list of characterization parameters does not change.
5. T-Series sensors only: Set the additional characterization parameters listed below.
Characterization parameter typeParameters
FlowFTG, FFQ
DensityDTG, DFQ1, DFQ2
16Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualQuick start
MMI-20025166October 2021
2.9.1 Sample sensor tags
Figure 2-1: Tag on newer curved-tube sensors (all sensors except T-Series)
2.9.2 Flow calibration parameters (FCF, FT)
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4-character FT value.
They are provided on the sensor tag.
Both values contain decimal points. During characterization, these are entered as a single 10-character string.
The 10-character string is called either Flowcal or FCF.
If your sensor tag shows the FCF and the FT values separately and you need to enter a single value,
concatenate the two values to form the single parameter value, retaining both decimal points.
Concatenating FCF and FT
FCF = x.xxxx FT = y.yy Flow calibration parameter: x.xxxxy.yy
2.9.3
Density calibration parameters are typically on the sensor tag and the calibration certificate.
If your sensor tag does not show a D1 or D2 value:
• For D1, enter the Dens A or D1 value from the calibration certificate. This value is the line-condition
• For D2, enter the Dens B or D2 value from the calibration certificate. This value is the line-condition density
Density calibration parameters (D1, D2, K1, K2, FD, DT, TC)
density of the low-density calibration fluid. Micro Motion uses air. If you cannot find a Dens A or D1 value,
enter 0.001 g/cm3.
of the high-density calibration fluid. Micro Motion uses water. If you cannot find a Dens B or D2 value,
enter 0.998 g/cm3 .
If your sensor tag does not show a K1 or K2 value:
• For K1, enter the first five digits of the density calibration factor. In this sample tag, this value is shown as
12500.
• For K2, enter the second five digits of the density calibration factor. In this sample tag, this value is shown
as 14286.
Configuration and Use Manual17
Quick startConfiguration and Use Manual
October 2021MMI-20025166
Figure 2-2: K1, K2, and TC values in the density calibration factor
If your sensor does not show an FD value, contact customer service.
If your sensor tag does not show a DT or TC value, enter the last four characters of the density calibration
factor. In the sample tag shown above, the value is shown as 4.44.
Do not confuse the Meter Factor line on the pictured sensor tag with any meter factor settings discussed in
this manual.
2.10
Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available
method.
Procedure
• Read the value for Mass Flow Rate on the transmitter display.
Menu → Operations → Process Variable Values
• Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables
panel.
• Connect to the transmitter with a field communicator and read the value for Mass Flow Rate.
Online → Overview → Mass Flow Rate
Postrequisites
If the reported mass flow rate is not accurate:
• Check the characterization parameters.
• Review the troubleshooting suggestions for flow measurement issues.
ProLink IIIDevice Tools→Calibration→Smart Zero Verification and Calibration→Verify Zero
Field communicatorService Tools→Maintenance→Calibration→Zero Calibration→Perform Zero Verify
18Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualQuick start
MMI-20025166October 2021
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field
zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the
following is true:
• The zero is required by site procedures.
• The stored zero value fails the zero verification procedure.
Do not verify the zero or zero the meter if a high-severity alert is active. Correct the problem, then verify the
zero or zero the meter. You may verify the zero or zero the meter if a low-severity alert is active.
Procedure
1.
Prepare the meter:
a) Allow the meter to warm up for at least 20 minutes after applying power.
b) Run the process fluid through the sensor until the sensor temperature reaches the normal
process operating temperature.
c) Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if
available.
d) Verify that the sensor is blocked in, that flow has stopped, and that the sensor is completely full
of process fluid.
2. Start the zero verification procedure, and wait until it completes.
3. If the zero verification procedure fails:
a) Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
b) Verify that the process fluid is not flashing or condensing, and that it does not contain particles
that can settle out.
c) Repeat the zero verification procedure.
d) If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
Configuration and Use Manual19
Quick startConfiguration and Use Manual
October 2021MMI-20025166
20Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20025166October 2021
3 Introduction to configuration and
commissioning
3.1
The transmitter has several features that can help to protect it against intentional or unintentional access and
configuration changes.
• When locked, the mechanical lock switch on the front of the display prevents any configuration changes
• When enabled, the software setting Write Protection prevents any configuration changes. The setting
• If the Universal Service Port (USP) is disabled, the port cannot be used by any service tool to communicate
• When used, the HART Lock prevents any changes by any other HART master.
• When enabled, Security prevents any configuration changes being made from the display unless the
Security and write protection
to the transmitter from any local or remote configuration tool. A transmitter without a display does not
have a lock switch.
can only be enabled if the transmitter does not have a display.
with or make changes to the transmitter.
appropriate password is entered.
3.1.1 Universal Service Port security
This transmitter is equipped with a Universal Service Port that works with USB type A connections, including
compatible flash drives. There are multiple levels of security built into the transmitter service port that you
can configure according to your needs and security standards.
The service port offers the following features that enhance interface security:
• The service port is inaccessible without physical access to the transmitter and requires removal of the
terminal cover.
• The service port can be disabled from the transmitter through software.
• The transmitter has a non-traditional operating system that is not designed to execute programs or run
scripts.
• The display can be password protected to limit access to the USB file menu.
• Overall transmitter security switches such as the lock switch or write-protection disallows configuration
changes from all interfaces including the Universal Service Port.
This transmitter:
• Was designed to be implemented in an industrial automation control system (Level 1 and Level 2 of the
Purdue Reference Architecture Model), with defense in depth security controls.
• Is not intended to be directly connected to an enterprise or to an internet-facing network without a
compensating control in place.
Configuration and Use Manual21
Introduction to configuration and commissioningConfiguration and Use Manual
October 2021MMI-20025166
3.1.2 Lock or unlock the transmitter
If the transmitter has a display, a mechanical switch on the display can be used to lock or unlock the
transmitter. When locked, no configuration changes can be made using any configuration tool.
Figure 3-1: Lock switch on transmitter display
You can determine whether you need to lock or unlock the transmitter by looking at the switch.
• If the switch is in the right position, the transmitter is locked.
• If the switch is in the left position, the transmitter is unlocked.
Note
The top switch is reserved for future use.
Procedure
1.WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is
powered up. Failure to follow these instructions can cause an explosion resulting in injury or death.
If you are in a hazardous area, power down the transmitter.
Remove the transmitter housing cover.
2.
Figure 3-2: Removing the transmitter housing cover
22Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20025166October 2021
3. Using a fine-pointed tool, move the switch to the desired position.
4.
Replace the transmitter housing cover.
5. If necessary, power up the transmitter.
3.1.3 Enable or disable the service port
DisplayMenu→Configuration→Security→Service Port
ProLink IIINot available
Field communicatorConfigure→Manual Setup→Security→Enable/Disable Service Port
The service port is enabled by default, so you can use it for transferring files or connect to it with ProLink III. If
you want to completely prevent it from being used, you can disable it.
Note
Enabling or disabling the service port will not take effect until power has been cycled to the transmitter.
WARNING
Do not use the service port if the transmitter is in a hazardous area because using the service port means
that you must open the transmitter wiring compartment. Opening the wiring compartment in a hazardous
area while the transmitter is powered up can cause an explosion resulting in injury or death.
3.1.4 Set the HART lock
If you plan to use a HART connection to configure the device, you can lock out all other HART masters. If you
do this, other HART masters will be able to read data from the device but will not be able to write data to the
device.
Restriction
• This feature is available only when you are using a field communicator or AMS.
• This feature is available only with a HART 7 host.
2. If you are locking the meter, set Lock Option as desired.
OptionDescription
Permanent Only the current HART master can make changes to the device. The device will remain
locked until manually unlocked by a HART master. The HART master can also change
Lock Option to Temporary.
Temporary Only the current HART master can make changes to the device. The device will remain
locked until manually unlocked by a HART master, or a power-cycle or device reset is
performed. The HART master can also change Lock Option to Permanent.
Configuration and Use Manual23
Introduction to configuration and commissioningConfiguration and Use Manual
October 2021MMI-20025166
OptionDescription
Lock AllNo HART masters are allowed to make changes to the configuration. Before changing
Lock Option to Permanent or Temporary, the device must be unlocked. Any HART
master can be used to unlock the device.
Postrequisites
To avoid future confusion or difficulties, ensure that the device is unlocked after you have completed your
tasks.
Field communicatorConfigure→Manual Setup→Security→Lock/Unlock Device
When enabled, Write-Protection prevents changes to the transmitter configuration. You can perform all
other functions, and you can view the transmitter configuration parameters.
Note
The write protection setting via software methods (such as ProLink III) is available only on transmitters
without a display.
For transmitters with a display, write protection is available only using the lock switch on the display. See Lock
or unlock the transmitter.
Enable or disable software write-protection
Write-protecting the transmitter primarily prevents accidental changes to configuration, not intentional
changes. Any user who can make changes to the configuration can disable write protection.
Field communicatorConfigure→Manual Setup→Display→Display Menus
24Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configure security for the display
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20025166October 2021
When using the display, you can require users to enter a password to do any of the following tasks:
• Enter the main menu
• Change a parameter
• Access alert data through the display
• Start, stop, or reset totalizers or inventories via the context menu
The display password can be the same or different from the totalizer/inventory context menu control
password. If different, the display password is used to reset, start, and stop totalizers or inventories using
Menu → Operations → Totalizers.
Procedure
1.
Configure Password Required as desired.
OptionDescription
At WriteWhen an user chooses an action that leads to a configuration change, they are
prompted to enter the display password.
Enter MenuWhen the menu is selected from the process variable screen, the display password
will be immediately required if Password Required is set.
Never (default) When a user chooses an action that leads to a configuration change, they are
prompted to activate ⇦⇧⇩⇨. This is designed to protect against accidental changes
to configuration. It is not a security measure.
2. If the At Write or Enter Menu option was selected, enable or disable alert security as desired.
OptionDescription
Enabled If an alert is active, the alert symbol ⓘ is shown in the upper right corner of the display but
the alert banner is not displayed. If the operator attempts to enter the alert menu, they are
prompted to enter the display password.
Disabled If an alert is active, the alert symbol ⓘ is shown in the upper right corner of the display and
the alert banner is displayed automatically. No password or confirmation is required to
enter the alert menu.
Restriction
You cannot set Password Required to Never and enable alert security.
• If you did not enable Password Required, alert security is disabled and cannot be enabled.
• Alert security is disabled automatically if you set Password Required to Never after:
— Password Required is initially set to either At Write or Enter Menu
— Alert security is enabled
Configuration and Use Manual25
Introduction to configuration and commissioningConfiguration and Use Manual
October 2021MMI-20025166
3. If Password Required has been set to At Write or Enter Menu, you will be prompted to enter the
desired password.
• Default: AAAA
• Range: Any four alphanumeric characters
• Password Required must be set to At Write or Enter Menu to enable the totalizer/inventory control
context menu password option.
Important
If you enable Password Required but you do not change the display password, the transmitter will post
a configuration alert.
4. Configure Main Menu Available as desired.
OptionDescription
EnabledThe local display Menu option from the process variable screen will be accessible.
Disabled The local display Menu option from the process variable screen will not be accessible.
Important
Once Main Menu Available has been disabled, you cannot enable it from the local display. Use another
configuration tool, such as ProLink III, to re-enable main menu access from the local display.
3.2 Work with configuration files
You can save the current transmitter configuration in two forms: a backup file and a replication file. You can
save the configuration to the SD card on your transmitter or to a USB drive.
Tip
You can use a saved configuration file to change the nature of the transmitter quickly. This might be
convenient if the transmitter is used for different applications or different process fluids.
You can load a configuration file to the transmitter working memory or to the transmitter SD card. You can
load either a backup file or a replication file.
Backup files
Replication files
3.2.1 Save a configuration file using the display
Contain all parameters. They are used to restore the current device if required.
The .spare extension is used to identify backup files.
Contain all parameters except the device-specific parameters, e.g., calibration factors or
meter factors. They are used to replicate the transmitter configuration to other devices.
The .xfer extension is used to identify replication files.
Prerequisites
If you are planning to use the USB drive, the service port must be enabled. It is enabled by default. However, if
you need to enable it, choose Menu→Configuration→Security and set Service Port to On.
26Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20025166October 2021
Procedure
• To save the current configuration to the transmitter SD card as a backup file:
a)
Choose Menu→Configuration→Save/Restore Config→Save Config to Memory.
b) Enter the name for this configuration file.
The configuration file is saved to the transmitter SD card as yourname.spare.
• To save the current configuration to a USB drive, as either a backup file or a replication file:
a)
b)
c) Choose Backup or Replicate.
d) Enter the name for this configuration file.
The configuration file is saved to the USB drive as yourname.spare or yourname.xfer.
• To copy a configuration file from the transmitter SD card to the USB drive:
a)
b)
c) Choose Backup or Replicate.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is
powered up. Failure to follow these instructions can cause an explosion resulting in injury or
death.
Open the wiring compartment on the transmitter and insert a USB drive into the service port.
Choose Menu→USB Options→Transmitter --> USB Drive→Save Active Config to USB Drive.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is
powered up. Failure to follow these instructions can cause an explosion resulting in injury or
death.
Open the wiring compartment on the transmitter and insert a USB drive into the service port.
Choose Menu→USB Options→Transmitter --> USB Drive→Transfer Config File to USB Drive.
d) Select the file that you want to transfer.
The configuration file is copied to the USB drive, using its existing name.
3.2.2 Save a configuration file using ProLink III
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually
or by groups. Therefore, you can use this format for both backup and replication.
Procedure
• To save the current configuration to the transmitter SD card:
b) Select On my 5700 Device Internal Memory and select Next.
c) Select Save.
Configuration and Use Manual27
Introduction to configuration and commissioningConfiguration and Use Manual
October 2021MMI-20025166
d) Enter the name for this configuration file.
e)
Set the file type.
— To save a backup file, set the file type to Backup.
— To save a replication file, set the file type to Transfer.
f) Select Save.
The configuration file is saved to the transmitter SD card as yourname.spare or yourname.xfer.
• To save the current configuration to your PC, in 5700 format:
a) Choose Device Tools→Configuration Transfer→Save Configuration.
b) Select On my computer in 5700 device file format and select Next.
c) Select Save.
d) Browse to the desired location, then enter the name for this configuration file.
e) Set the file type.
— To save a backup file, set the file type to Backup.
— To save a replication file, set the file type to Transfer.
f) Select Save.
The configuration file is saved to the specified location as yourname.spare or yourname.xfer.
• To save the current configuration to your PC, in ProLink III format:
a) Choose Device Tools→Configuration Transfer→Save Configuration.
b) Select On my computer in ProLink III file format and click Next.
c) Select Save.
d) Select the configuration parameters to be included in this file.
— To save a backup file, select all parameters.
— To save a replication file, select all parameters except device-specific parameters.
e) Select Save.
f) Browse to the desired location, then enter the name for this configuration file.
g) Set the file type to ProLink configuration file.
h) Select Start Save.
The configuration file is saved to the specified location as yourname.pcfg.
3.2.3 Load a configuration file using the display
Prerequisites
You must have a backup file or a replication file available for use.
28Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20025166October 2021
If you are planning to use the USB drive, the service port must be enabled. It is enabled by default. However, if
you need to enable it, choose Menu→Configuration→Security and set Service Port to On.
Procedure
• To load either a backup file or a replication file from the transmitter SD card:
a)
Choose Menu→Configuration→Save/Restore Config→Restore Config from Memory.
b) Select Backup or Replicate.
c) Select the file that you want to load.
The file is loaded to working memory and becomes active immediately.
• To load a either a backup file or a replication file from a USB drive:
a)
b) Choose Menu→USB Options→USB Drive --> Transmitter→Upload Configuration File.
c) Select Backup or Replicate.
d) Select the file that you want to load.
e) Choose Yes or No when prompted to apply the settings.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is
powered up. Failure to follow these instructions can cause an explosion resulting in injury or
death.
Open the wiring compartment on the transmitter and insert the USB drive containing the backup
file or replication file into the service port.
— Yes: The file is loaded to working memory and becomes active immediately.
— No: The file is loaded to the transmitter SD card but not to working memory. You can load it
from the SD card to working memory at a later time.
3.2.4 Load a configuration file using ProLink III
You can load a configuration file to the transmitter working memory. You can load a backup file or a
replication file. Two PC file formats are supported: the 5700 format and the ProLink III format.
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually
or by groups. Therefore, you can use this format for both backup and replication.
Procedure
• To load a backup file or replication file from the transmitter SD card:
Field communicatorService Tools→Maintenance→Reset/Restore→Restore Factory Configuration
A file containing the factory configuration is always saved in the transmitter internal memory, and is available
for use.
This action is typically used for error recovery or for repurposing a transmitter.
If you restore the factory configuration, the real-time clock, the audit trail, the historian, and other logs are
not reset.
30Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualIntroduction to configuration and commissioning
MMI-20025166October 2021
3.2.6 Replicate a transmitter configuration
Replicating a transmitter configuration is a fast method to set up similar or identical measurement points.
Procedure
1.
Configure a transmitter and verify its operation and performance.
2. Use any available method to save a replication file from that transmitter.
3. Use any available method to load the replication file to another transmitter.
4. At the replicated transmitter, set device-specific parameters and perform device-specific procedures:
a) Set the clock.
b) Set the tag, long tag, HART address, Modbus address, and related parameters.
c) Characterize the transmitter.
d) Perform zero validation and take any recommended actions.
e) Perform loop tests and take any recommended actions, including mA Output trim.
f) Use sensor simulation to verify transmitter response.
5. At the replicated transmitter, make any other configuration changes.
6. Follow your standard procedures to ensure that the replicated transmitter is performing as desired.
Configuration and Use Manual31
Introduction to configuration and commissioningConfiguration and Use Manual
October 2021MMI-20025166
32Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
4 Configure process measurement
4.1
DisplayMenu→Configuration→Process Measurement→Flow Variables→Flow Direction
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Sensor Direction
Field communicatorConfigure→Manual Setup→Measurements→Flow→Sensor Direction
Sensor Flow Direction Arrow is used to accommodate installations in which the Flow arrow on the sensor
does not match the majority of the process flow. This typically happens when the sensor is accidentally
installed backwards.
Sensor Flow Direction Arrow interacts with mA Output Direction, Frequency Output Direction, and
Totalizer Direction to control how flow is reported by the outputs and accumulated by the totalizers and
inventories.
The Sensor Flow Direction Arrow also affects how flow is reported on the transmitter display and via digital
communications. This includes ProLink III, a field communicator, and all other user interfaces.
Figure 4-1: Flow arrow on sensor
Configure Sensor Flow Direction Arrow
A. Flow arrow
Actual flow direction
B.
Procedure
Set Sensor Flow Direction Arrow as appropriate.
OptionDescription
With ArrowThe majority of flow through the sensor matches the Flow arrow on the sensor. Actual
forward flow is processed as forward flow.
Against Arrow The majority of flow through the sensor is opposite to the Flow arrow on the sensor. Actual
forward flow is processed as reverse flow.
Configuration and Use Manual33
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the
setting of Sensor Flow Direction Arrow. Sensor Flow Direction Arrow controls only whether actual flow is
processed as forward flow or reverse flow.
Related information
Configure mA Output Direction
Configure Frequency Output Direction
Configure Discrete Output Source
Configure totalizers and inventories
Effect of Sensor Flow Direction Arrow on digital communications
4.2
The mass flow measurement parameters control how mass flow is measured and reported. The mass total
and mass inventory are derived from the mass flow data.
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Mass Flow Rate Unit
Field communicatorConfigure→Manual Setup→Measurements→Flow→Mass Flow Unit
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The
default unit used for mass total and mass inventory is derived from this unit.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
Default: g/sec (grams per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one
user-defined special measurement unit. Different communications tools may use different labels for the
units.
Unit description
DisplayProLink IIIField communicator
Grams per secondgram/sg/secg/s
Grams per minutegram/ming/ming/min
Grams per hourgram/hg/hrg/h
Kilograms per secondkg/skg/seckg/s
34Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Label
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
Unit description
DisplayProLink IIIField communicator
Kilograms per minutekg/minkg/minkg/min
Kilograms per hourkg/hkg/hrkg/h
Kilograms per daykg/dkg/daykg/d
Metric tons per minuteMetTon/minmTon/minMetTon/min
Metric tons per hourMetTon/hmTon/hrMetTon/h
Metric tons per dayMetTon/dmTon/dayMetTon/d
Pounds per secondlb/slbs/seclb/s
Pounds per minutelb/minlbs/minlb/min
Pounds per hourlb/hlbs/hrlb/h
Pounds per daylb/dlbs/daylb/d
Short tons (2000 pounds) per minuteSTon/minsTon/minSTon/min
Short tons (2000 pounds) per hourSTon/hsTon/hrSTon/h
Short tons (2000 pounds) per daySTon/dsTon/daySTon/d
Long tons (2240 pounds) per hourLTon/hlTon/hrLTon/h
Long tons (2240 pounds) per dayLTon/dlTon/dayLTon/d
Field communicatorConfigure→Manual Setup→Measurements→Flow→Flow Damping
Flow Damping controls the amount of damping that will be applied to the measured mass flow rate. It affects
flow rate process variables that are based on the measured mass flow rate. This includes volume flow rate and
gas standard volume flow rate.
Flow Damping also affects specialized flow rate variables such as temperature-corrected volume flow rate
(API Referral) and net mass flow rate (concentration measurement). It is not applied to the flow rate received
via the Frequency Input.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value
specifies the time period, in seconds, over which the transmitter will spread changes in the process variable.
At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the
change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
• Default: 0.64 seconds
• Range: 0 seconds to 60 seconds
Note
If a number greater than 60 is entered, it is automatically changed to 60.
36Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
Tip
• A high damping value makes the process variable appear smoother because the reported value changes
slowly.
• A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
• The combination of a high damping value and rapid, large changes in flow rate can result in increased
measurement error.
• Whenever the damping value is non-zero, the reported measurement will lag the actual measurement
because the reported value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the reported value.
• The transmitter automatically rounds off any entered damping value to the nearest valid value. Therefore,
the recommended damping value for gas applications should be 3.2 seconds. If you enter 2.56, the
transmitter will round it off to 3.2.
• For filling applications, Emerson recommends using the default value of 0.04 seconds.
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also affects volume
measurement for gas standard volume data. The transmitter calculates volume data from the damped mass
flow data.
Interaction between Flow Damping and mA Output Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the reported mass flow
value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping controls the rate
of change reported through mA Output. If mA Output Process Variable is set to Mass Flow Rate, and both
Flow Damping and mA Output Damping are set to non-zero values, flow damping is applied first, and the
added damping calculation is applied to the result of the first calculation.
Field communicatorConfigure→Manual Setup→Measurements→Flow→Mass Flow Cutoff
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates
below this cutoff will be reported as 0.
Configure Mass Flow Cutoff
Cutoff
Procedure
Set Mass Flow Cutoff to the value you want to use.
• Default: A sensor-specific value set at the factory. If your transmitter was ordered without a sensor, the
default may be 0.0.
Configuration and Use Manual37
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
• Recommendation: 0.5% of maximum flow rate of the attached sensor. See the sensor specifications.
Important
Do not use your meter for measurement with Mass Flow Cutoff set to 0.0 g/sec. Ensure that Mass Flow
Cutoff is set to the value that is appropriate for your sensor.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data
rather than the reported value.
Volume flow has a separate Volume Flow Cutoff that is not affected by the Mass Flow Cutoff value.
Interaction between Mass Flow Cutoff and mA Output Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. mA
Output Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process
Variable is set to Mass Flow Rate, the mass flow rate reported through mA Output is controlled by the higher
of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events
defined on mass flow).
mA Output Cutoff affects only mass flow values reported through mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
• mA Output Process Variable: Mass Flow Rate
• Frequency Output Process Variable: Mass Flow Rate
• mA Output Cutoff: 10 g/sec
• Mass Flow Cutoff: 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all
internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
• mA Output Process Variable: Mass Flow Rate
• Frequency Output Process Variable: Mass Flow Rate
• mA Output Cutoff: 15 g/sec
• Mass Flow Cutoff: 10 g/sec
Result:
• If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
— The mA Output will report zero flow.
— The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
38Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
• If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.3
Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured and reported. The
volume total and volume inventory are derived from volume flow data.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one
or the other.
4.3.1 Configure Volume Flow Type for liquid applications
Display
ProLink III
Field communicator
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with the following applications:
• API Referral
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Flow
Type → Liquid
Device Tools → Configuration → Process Measurement → Flow → Volume Flow Type → Liquid Volume
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Volume Flow Rate Unit
Field communicatorConfigure→Manual Setup→Measurements→Flow→Volume Flow Unit
Configuration and Use Manual39
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume
flow rate. The unit used for the volume total and volume inventory is based on this unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
Default: l/sec (liters per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement Unit, plus one
user-defined measurement unit. Different communications tools may use different labels for the units.
Unit description
DisplayProLink IIIField communicator
Cubic feet per secondft3/sft3/secCuft/s
Cubic feet per minuteft3/minft3/minCuft/min
Cubic feet per hourft3/hft3/hrCuft/h
Cubic feet per dayft3/dft3/dayCuft/d
Cubic meters per secondm3/sm3/secCum/s
Cubic meters per minutem3/minm3/minCum/min
Cubic meters per hourm3/hm3/hrCum/h
Cubic meters per daym3/dm3/dayCum/d
U.S. gallons per secondgal/sUS gal/secgal/s
U.S. gallons per minutegal/mUS gal/mingal/min
U.S. gallons per hourgal/hUS gal/hrgal/h
U.S. gallons per daygal/dUS gal/daygal/d
Million U.S. gallons per dayMMgal/dmil US gal/dayMMgal/d
Liters per secondL/sl/secL/s
Liters per minuteL/minl/minL/in
Label
Liters per hourL/hl/hrL/h
Million liters per dayMML/dmil l/dayML/d
Imperial gallons per secondImpgal/sImp gal/secImpgal/s
Imperial gallons per minuteImpgal/mImp gal/minImpgal/min
Imperial gallons per hourImpgal/hImp gal/hrImpgal/h
Imperial gallons per dayImpgal/dImp gal/dayImpgal/d
40Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
Unit description
Barrels per second
Barrels per minute
Barrels per hour
Barrels per day
Beer barrels per second
Beer barrels per minute
Beer barrels per hour
Beer barrels per day
Special unitSPECIALSpecialSpecial
(1) Unit based on oil barrels (42 U.S. gallons).
(2) Unit based on U.S. beer barrels (31 U.S. gallons).
Field communicatorConfigure→Manual Setup→Measurements→Optional Setup→Special Units→Volume Special Units
SPECIAL
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer
data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is
calculated from an existing measurement unit using a conversion factor.
Procedure
Specify Base Volume Unit.
1.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Volume Flow Conversion Factor = x ÷ y
4. Enter Volume Flow Conversion Factor.
The original volume flow rate value is divided by this conversion factor.
5. Set Volume Flow Label to the name you want to use for the volume flow unit.
6. Set Volume Total Label to the name you want to use for the volume total and volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the
special measurement unit at any time.
Configuration and Use Manual41
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
Example: Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
Field communicatorConfigure→Manual Setup→Measurements→Flow→Volume Flow Cutoff
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured. All volume flow
rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
• Default: 0.0 l/sec (liters per second)
• Range: 0 l/sec to x l/sec, where x is the sensor’s flow calibration factor, multiplied by 0.0002
Cutoff
Interaction between Volume Flow Cutoff and mAO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured.
mAO Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process
Variable is set to Volume Flow Rate, the volume flow rate reported through mA Output is controlled by the
higher of the two cutoff values.
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the volume flow values
used in other transmitter behavior (e.g., events defined on the volume flow).
mAO Cutoff affects only flow values reported through mA Output.
Example: Cutoff interaction with mAO Cutoff lower than Volume Flow Cutoff
Configuration:
• mA Output Process Variable: Volume Flow Rate
• Frequency Output Process Variable: Volume Flow Rate
42Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
• AO Cutoff: 10 l/sec
• Volume Flow Cutoff: 15 l/sec
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all
internal processing.
Example: Cutoff interaction with mAO Cutoff higher than Volume Flow Cutoff
Configuration:
• mA Output Process Variable: Volume Flow Rate
• Frequency Output Process Variable: Volume Flow Rate
• AO Cutoff: 15 l/sec
• Volume Flow Cutoff: 10 l/sec
Result:
• If the volume flow rate drops below 15 l/sec but not below 10 l/sec:
— The mA Output will report zero flow.
— The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
• If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.4
Configure Gas Standard Volume (GSV) flow
measurement
The gas standard volume (GSV) flow measurement parameters control how gas standard volume flow is
measured and reported.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one
or the other.
4.4.1 Configure Volume Flow Type for gas applications
Display
ProLink III
Field
communicator
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Flow Type →
Gas
Device Tools → Configuration → Process Measurement → Flow → Volume Flow Type → Gas Standard
Volume
Configure → Manual Setup → Measurements → Optional Setup → GSV → Volume Flow Type → Standard Gas
Volume
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with the following applications:
• API Referral
Configuration and Use Manual43
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
• Concentration measurement
• Advanced Phase Measurement — net oil
• Advanced Phase Measurement — liquid with gas
For these applications, set Volume Flow Type to Liquid.
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Standard Density of Gas
Field communicatorConfigure→Manual Setup→Measurements→Optional Setup→GSV→Gas Ref Density
Standard Gas Density is the density of your gas at reference temperature and reference pressure. This is
often called standard density or base density. It is used to calculate the GSV flow rate from the mass flow rate.
Procedure
Set Standard Gas Density to the density of your gas at reference temperature and reference pressure.
You can use any reference temperature and reference pressure that you choose. It is not necessary to
configure these values in the transmitter.
Tip
ProLink III provides a guided method that you can use to calculate the standard density of your gas if you do
not know it.
Configure Standard Gas Density
Gas Density
4.4.3 Configure Gas Standard Volume Flow Measurement Unit
ProLink IIIDevice Tools→Configuration→Process Measurement→Flow→Gas Standard Volume Flow Unit
Field communicatorConfigure→Manual Setup→Measurements→Flow→GSV Flow Unit
Gas Standard Volume Flow Measurement Unit specifies the unit of measure that will be used for the gas
standard volume (GSV) flow rate. The unit used for gas standard volume total and gas standard volume
inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Measurement Unit, be sure that Volume Flow Type is set
to Gas Standard Volume.
Procedure
Set Gas Standard Volume Flow Measurement Unit to the unit you want to use.
Default: SCFM (Standard Cubic Feet per Minute)
44Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Gas Standard Volume Flow Measurement Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow Measurement
Unit, plus one user-defined special measurement unit. Different communications tools may use different
labels for the units.
Unit description
DisplayProLink IIIField communicator
Normal cubic meters per secondNCMSNm3/secNm3/sec
Normal cubic meters per minuteNCMMNm3/minNm3/min
Normal cubic meters per hourNCMHNm3/hrNm3/hr
Normal cubic meters per dayNCMDNm3/dayNm3/day
Normal liter per secondNLPSNLPSNLPS
Normal liter per minuteNLPMNLPMNLPM
Normal liter per hourNLPHNLPHNLPH
Normal liter per dayNLPDNLPDNLPD
Standard cubic feet per secondSCFSSCFSSCFS
Standard cubic feet per minuteSCFMSCFMSCFM
Standard cubic feet per hourSCFHSCFHSCFH
Standard cubic feet per daySCFDSCFDSCFD
Standard cubic meters per secondSCMSSm3/secSm3/sec
Standard cubic meters per minuteSCMMSm3/minSm3/min
Standard cubic meters per hourSCMHSm3/hrSm3/hr
Label
Standard cubic meters per daySCMDSm3/daySm3/day
Standard liter per secondSLPSSLPSSLPS
Standard liter per minuteSLPMSLPMSLPM
Standard liter per hourSLPHSLPHSLPH
Standard liter per daySLPDSLPDSLPD
Special measurement unitSPECIALSpecialSpecial
Define a special measurement unit for gas standard volume flow
Display
ProLink III
Field communicator
Configuration and Use Manual45
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Units →
SPECIAL
Device Tools → Configuration → Process Measurement → Flow → Gas Standard Volume Flow Unit →
Special
Configure → Manual Setup → Measurements → Optional Setup → Special Units → Special Gas Standard
Volume Units
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer
data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is
calculated from an existing measurement unit using a conversion factor.
Procedure
1.
Specify Base Gas Standard Volume Unit.
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be
based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Gas Standard Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Gas Standard Volume Flow Conversion Factor = x ÷ y
4. Enter the Gas Standard Volume Flow Conversion Factor.
The original gas standard volume flow value is divided by this conversion factor.
5. Set Gas Standard Volume Flow Label to the name you want to use for the gas standard volume flow
unit.
6. Set Gas Standard Volume Total Label to the name you want to use for the gas standard volume total
and gas standard volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the
special measurement unit at any time.
Example: Defining a special measurement unit for gas standard volume flow
You want to measure gas standard volume flow in thousands of standard cubic feet per minute.
1. Set Base Gas Standard Volume Unit to SCFM.
2. Set Base Time Unit to minutes (min).
3. Calculate the conversion factor:
a. One thousands of standard cubic feet per minute = 1000 cubic feet per minute
b. Gas Standard Volume Flow Conversion Factor = 1 ÷ 1000 = 0.001
4. Set Gas Standard Volume Flow Conversion Factor to 0.001.
5. Set Gas Standard Volume Flow Label to KSCFM.
6. Set Gas Standard Volume Total Label to KSCF.
4.4.4 Configure Gas Standard Volume Flow Cutoff
Display
ProLink III
Field communicator
46Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Menu → Configuration → Process Measurement → Flow Variables → Volume Flow Settings → Low Flow
Cutoff
Device Tools → Configuration → Process Measurement → Flow → Gas Standard Volume Flow Cutoff
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
Gas Standard Volume Flow Cutoff specifies the lowest gas standard volume flow rate that will reported as
measured. All gas standard volume flow rates below this cutoff will be reported as 0.
Procedure
Set Gas Standard Volume Flow Cutoff to the value you want to use.
• Default: 0.0
• Range: 0.0 to any positive value
Interaction between Gas Standard Volume Flow Cutoff and mA Output
Cutoff
Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the transmitter
will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported through mA
Output. If mA Output Process Variable is set to Gas Standard Volume Flow Rate, the volume flow rate
reported through mA Output is controlled by the higher of the two cutoff values.
Gas Standard Volume Flow Cutoff affects both the gas standard volume flow values reported through
outputs and the gas standard volume flow values used in other transmitter behavior (for example, events
defined on gas standard volume flow).
mA Output Cutoff affects only flow values reported through mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Gas Standard Volume Flow Cutoff
Configuration:
• mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
• Frequency Output Process Variable: Gas Standard Volume Flow Rate
• mA Output Cutoff for the primary mA Output: 10 SLPM (standard liters per minute)
• Gas Standard Volume Flow Cutoff: 15 SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be reported as 0, and 0
will be used in all internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Gas Standard Volume Flow Cutoff
Configuration:
• mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
• Frequency Output Process Variable: Gas Standard Volume Flow Rate
• mA Output Cutoff for the primary mA Output: 15 SLPM (standard liters per minute)
• Gas Standard Volume Flow Cutoff: 10 SLPM
Result:
• If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:
— The primary mA Output will report zero flow.
— The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
Configuration and Use Manual47
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
• If the gas standard volume flow rate drops below 10 SLPM, both outputs will report zero flow, and 0 will be
used in all internal processing.
4.5
The density measurement parameters control how density is measured and reported. Density measurement
is used with mass flow rate measurement to determine liquid volume flow rate.
ProLink IIIDevice Tools→Configuration→Process Measurement→Density→Density Unit
Field communicatorConfigure→Manual Setup→Measurements→Density→Density Unit
Density Measurement Unit controls the measurement units that will be used in density calculations and
reporting.
Restriction
If the API Referral application is enabled, you cannot change the density measurement unit here. The density
measurement unit is controlled by the API table selection.
Procedure
Set Density Measurement Unit to the option you want to use.
Default: g/cm3 (grams per cubic centimeter)
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit. Different
communications tools may use different labels.
Unit description
Specific gravity
Grams per cubic centimeterg/cm3g/cm3g/Cucm
Grams per literg/Lg/lg/L
Grams per milliliterg/mLg/mlg/mL
Kilograms per literkg/Lkg/lkg/L
Kilograms per cubic meterkg/m3kg/m3kg/Cum
Pounds per U.S. gallonlb/gallbs/USgallb/gal
Pounds per cubic footlb/ft3lbs/ft3lb/Cuft
Pounds per cubic inchlb/in3lbs/in3lb/CuIn
(1)
DisplayProLink IIIField communicator
SGUSGUSGU
Label
48Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
Unit description
Degrees API
Short ton per cubic yardSTon/yd3sT/yd3STon/Cuyd
(1) Non-standard calculation. This value represents line density divided by the density of water at 60 °F (15.6 °C).
(2) Non standard calculation, unless the API referral application is enabled. Calculated from line density instead of specific
Field communicatorConfigure→Manual Setup→Measurements→Density→Density Damping
(2)
Configure Density Damping
DisplayProLink IIIField communicator
APIAPIdegAPI
Label
Density Damping controls the amount of damping that will be applied to density data.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value
specifies the time period, in seconds, over which the transmitter will spread changes in the process variable.
At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the
change in the actual measured value.
Procedure
Set Density Damping to the desired value.
• Default: 1.28 seconds
• Range: 0.0 to 60 seconds
Tip
• A high damping value makes the process variable appear smoother because the reported value changes
slowly.
• A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
• The combination of a high damping value and rapid, large changes in density can result in increased
measurement error.
• Whenever the damping value is non-zero, the damped value will lag the actual measurement because the
damped value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the damped value.
• If a number greater than 60 is entered, it is automatically changed to 60.
Configuration and Use Manual49
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
Effect of Density Damping on volume measurement
Density Damping affects liquid volume measurement. Liquid volume values are calculated from the damped
density value rather than the measured density value. Density Damping does not affect gas standard volume
measurement.
Interaction between Density Damping and mA Output Damping
When mA Output is configured to report density, both Density Damping and mA Output Damping are
applied to the reported density value.
Density Damping controls the rate of change in the value of the process variable in transmitter memory. mA
Output Damping controls the rate of change reported through mA Output.
If mA Output Source is set to Density, and both Density Damping and mA Output Damping are set to nonzero values, density damping is applied first, and the mA Output damping calculation is applied to the result
of the first calculation. This value is reported over mA Output.
Field communicatorConfigure→Manual Setup→Measurements→Density→Density Cutoff
Density Cutoff specifies the lowest density value that will be reported as measured. All density values below
this cutoff will be reported as 0.
Procedure
Set Density Cutoff to the value you want to use.
• Default: 0.2 g/cm
• Range: 0.0 g/cm3 to 0.5 g/cm
Configure Density Cutoff
3
3
Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density Cutoff, the
volume flow rate is reported as 0. Density Cutoff does not affect gas standard volume measurement. Gas
standard volume values are always calculated from the value configured for Standard Gas Density or polled
value if configured for polled base density.
50Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
4.6 Configure temperature measurement
The temperature measurement parameters control how temperature data is processed. Temperature data is
used in several different ways, including temperature compensation, API Referral, and concentration
measurement.
ProLink IIIDevice Tools→Configuration→Process Measurement→Temperature→Temperature Unit
Field communicatorConfigure→Manual Setup→Measurements→Temperature→Unit
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
Default: °C (Celsius)
Configure Temperature Measurement Unit
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different
communications tools may use different labels for the units.
Field communicatorConfigure→Manual Setup→Measurements→Temperature→Damping
Temperature Damping controls the amount of damping that will be applied to temperature data from the
sensor. Temperature Damping is not applied to external temperature data.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value
specifies the time period, in seconds, over which the transmitter will spread changes in the process variable.
At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the
change in the actual measured value.
Procedure
Set Temperature Damping to the desired value.
Configuration and Use Manual51
Configure process measurementConfiguration and Use Manual
October 2021MMI-20025166
• Default: 4.8 seconds
• Range: 0.0 to 80 seconds
Note
If a number greater than 80 is entered, it is automatically changed to 80.
Tip
• A high damping value makes the process variable appear smoother because the reported value changes
slowly.
• A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
• The combination of a high damping value and rapid, large changes in temperature can result in increased
measurement error.
• Whenever the damping value is non-zero, the damped value will lag the actual measurement because the
damped value is being averaged over time.
• In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the damped value.
Effect of Temperature Damping on process measurement
Temperature Damping affects all processes and algorithms that use temperature data from the internal
sensor RTD.
Temperature compensation
Temperature compensation adjusts process measurement to compensate for the effect of temperature on
the sensor tubes.
API Referral
Temperature Damping affects API Referral process variables only if the transmitter is configured to use
temperature data from the sensor. If an external temperature value is used for API Referral, TemperatureDamping does not affect API Referral process variables.
Concentration measurement
Temperature Damping affects concentration measurement process variables only if the transmitter is
configured to use temperature data from the sensor. If an external temperature value is used for
concentration measurement, Temperature Damping does not affect concentration measurement process
variables.
4.7
Display
ProLink III
Field communicator
Configure Pressure Measurement Unit
Menu → Configuration → Process Measurement → Pressure → Units
Device Tools → Configuration → Process Measurement → Pressure Compensation → Pressure Unit
52Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement
MMI-20025166October 2021
Pressure Measurement Unit controls the measurement unit used for pressure. This unit must match the unit
used by the external pressure device.
Pressure data is used for pressure compensation and for API Referral. The device does not measure pressure
directly. You must set up a pressure input.
Procedure
Set Pressure Measurement Unit to the desired unit.
Default: psi
4.7.1
Options for Pressure Measurement Unit
The transmitter provides a standard set of measurement units for Pressure Measurement Unit. Different
communications tools may use different labels for the units. In most applications, set Pressure MeasurementUnit to match the pressure measurement unit used by the remote device.
Unit description
DisplayProLink IIIField communicator
Feet water @ 68 °FftH2O @68°FFt Water @ 68°FftH₂O
Inches water @ 4 °CinH2O @4°CIn Water @ 4°CinH₂O @4DegC
Inches water @ 60 °FinH2O @60°FIn Water @ 60°FinH₂O @60DegF
Inches water @ 68 °FinH2O @68°FIn Water @ 68°FinH₂O
Millimeters water @ 4 °CmmH2O @4°Cmm Water @ 4°CmmH₂O @4DegC
Millimeters water @ 68 °FmmH2O @68°Fmm Water @ 68°FmmH₂O
Field communicatorConfigure→Manual Setup→Measurements→Approximate Velocity→Velocity Unit
Velocity Measurement Unit controls the measurement unit used to report velocity.
Procedure
Set Velocity Measurement Unit to the desired unit.
Default: m/sec
4.8.1
The transmitter provides a standard set of measurement units for Velocity Measurement Unit. Different
communications tools may use different labels.
Unit description
Feet per minuteft/minft/minft/min
Feet per secondft/sft/secft/s
Inches per minutein/minin/minin/min
Inches per secondin/sin/secin/s
Meters per hourm/hm/hrm/h
Meters per secondm/sm/secm/s
Options for Velocity Measurement Unit
Label
DisplayProLink IIIField communicator
54Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
5 Configure process measurement
applications
5.1
The API Referral application corrects line density to reference temperature and reference pressure according
to American Petroleum Institute (API) standards. The resulting process variable is referred density.
Restriction
The API Referral application is not compatible with the following applications:
• Gas Standard Volume Measurement (GSV)
• Piecewise linearization (PWL)
• Advanced Phase Measurement
• Concentration measurement
Set up the API Referral application
5.1.1 Set up the API Referral application using the display
Enable the API Referral application using the display
The API Referral application must be enabled before you can perform any setup. If the API Referral application
was enabled at the factory, you do not need to enable it now.
Prerequisites
The API Referral application must be licensed on your transmitter.
Procedure
1.
Choose Menu→Configuration→Process Measurement.
2. Choose Flow Variables→Volume Flow Settings and ensure that Flow Type is set to Liquid.
3. Return to the Process Measurement menu.
4. If the concentration measurement application is displayed in the list, choose Concentration
Measurement and ensure that Enabled/Disabled is set to Disabled.
The concentration measurement application and the API Referral application cannot be enabled
simultaneously.
5. Enable API Referral.
a) Choose Menu→Configuration→Process Measurement→API Referral.
b) Set Enabled/Disabled to Enabled.
Configuration and Use Manual55
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Configure API Referral using the display
The API Referral parameters specify the API table, measurement units, and reference values to be used in
referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process
fluid.
You must know the reference temperature and reference pressure that you want to use.
2. Set API Table to the API table that you want to use to calculate referred density.
Each API table is associated with a specific set of equations. Choose your API table based on your
process fluid and the measurement unit that you want to use for referred density.
Your choice also determines the API table that will be used to calculate the correction factor for volume
(CTPL or CTL).
3. Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
• 230.0 x 10-6 to 930.0 x 10-6 per °F
• 414.0 x 10-6 to 1674.0 x 10-6 per °C
5. If required, set Reference Temperature to the temperature to which density will be corrected in
referred density calculations.
The default reference temperature is determined by the selected API table.
6. If required, set Reference Pressure to the pressure to which density will be corrected in referred
density calculations.
The default reference pressure is determined by the selected API table.
Set up temperature and pressure data for API Referral using the display
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must
decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
56Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART
communications.
If you plan to use the mA Input for temperature data, Channel D must be available for use, and must be wired
to an external temperature device.
If you plan to use the mA Input for pressure data, Channel D must be available for use, and must be wired to
an external pressure device.
Restriction
You can use an mA Input for either external temperature or external pressure, but not both.
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
If you are using an external temperature device, it must use the temperature unit that is configured in the
transmitter.
Procedure
1.
Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal
temperature
PollingThe meter polls an
Temperature data from
the on-board
temperature sensor
(RTD) will be used for
all measurements and
calculations. No
external temperature
data will be available.
external device for
temperature data. This
data will be available in
addition to the internal
temperature data.
a. Choose Menu→Configuration→Process Measurement→
b.
a. Choose Menu→Configuration→Process Measurement→
b.
c. Choose Poll External Device.
d. Select Polled Variable 1 or Polled Variable 2.
e. Set Variable to External Temperature.
f. Set Polling Control to Poll as Primary or Poll as Secondary.
Temperature.
Set External Temperature to Off.
Temperature.
Set External Temperature to On.
OptionDescription
Poll as PrimaryNo other HART masters will be on the
network. A field communicator is not a HART
master.
Poll as SecondaryOther HART masters will be on the network.
A field communicator is not a HART master.
g. Set External Device Tag to the HART tag of the external temperature
device.
Configuration and Use Manual57
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
MethodDescriptionSetup
mA InputAn external device
sends temperature
data to the meter
through mA Input. This
data will be available in
addition to the internal
temperature data.
Digital
communications
A host writes
temperature data to
the meter at
appropriate intervals.
This data will be
available in addition to
the internal
temperature data.
a. Choose Menu→Configuration→Process Measurement→
Temperature.
Set External Temperature to On.
b.
c. Choose Menu→Configuration→Inputs/Outputs→Channel D.
d. Set I/O Type to MAI.
e. Set Power appropriately.
f. Choose I/O Settings.
g. Set Assignment to External Temperature.
h. Set Lower Range Value and Upper Range Value appropriately.
a. Choose Menu→Configuration→Process Measurement→
Temperature.
b.
Set External Temperature to On.
c. Perform the necessary host programming and communications setup
to write temperature data to the transmitter at appropriate intervals.
2. Choose the method to be used to supply pressure data, and perform the required setup.
MethodDescriptionSetup
PollingThe meter polls an
external device for
pressure data.
a. Choose Menu→Configuration→Process Measurement→Pressure
→External Pressure.
Set External Pressure to On.
b.
c. Choose Poll External Device.
d. Select Polled Variable 1 or Polled Variable 2.
e. Set Variable to External Pressure.
f. Set Polling Control to Poll as Primary or Poll as Secondary.
OptionDescription
Poll as PrimaryNo other HART masters will be on the
network. A field communicator is not a HART
master.
Poll as SecondaryOther HART masters will be on the network.
A field communicator is not a HART master.
g. Set External Device Tag to the HART tag of the external pressure
device.
58Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
MethodDescriptionSetup
mA InputAn external device
sends pressure data to
the meter through mA
Input.
Digital
communications
A host writes pressure
data to the meter at
appropriate intervals.
a. Choose Menu→Configuration→Process Measurement→Pressure
→External Pressure.
Set External Pressure to On.
b.
c. Choose Menu→Configuration→Inputs/Outputs→Channel D.
d. Set I/O Type to MAI.
e. Set Power appropriately.
f. Choose I/O Settings.
g. Set Assignment to External Pressure.
h. Set Lower Range Value and Upper Range Value appropriately.
a. Choose Menu→Configuration→Process Measurement→Pressure
→External Pressure.
b.
Set External Pressure to On.
c. Perform the necessary host programming and communications setup
to write pressure data to the transmitter at appropriate intervals.
Postrequisites
Choose Menu→Service Tools→Service Data→View Process Variables and verify the values for External
Temperature and External Pressure.
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For polling:
— Verify the wiring between the meter and the external device.
— Verify the HART tag of the external device.
• For an mA Input:
— Verify the wiring between the meter and the external device.
— Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
— Verify the Lower Range Value and Upper Range Value settings for an mA Input.
— Perform an mA Input trim.
— Adjust the mA Input damping value.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the host is writing to the correct register in memory, using the correct data type.
Configuration and Use Manual59
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
5.1.2 Set up the API Referral application using ProLink III
Enable the API Referral application using ProLink III
Prerequisites
The API Referral application must be licensed and enabled on your transmitter before you can perform any
setup. If the API Referral application was enabled at the factory, you do not need to enable it now.
Procedure
1.
Choose Device Tools→Configuration→Process Measurement→Flow and ensure that VolumeFlow Type is set to Liquid Volume.
C tablesLiquids with a constant base density or known thermal expansion coefficient
(TEC). You will be required to enter the TEC for your process fluid.
D tablesLubricating oils
E tablesNGL (Natural Gas Liquids) and LPG (Liquid Petroleum Gas)
60Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
b) Set Referred Density Measurement Unit to the measurement units that you want to use for
referred density.
c)
Select Apply.
These parameters uniquely identify the API table to be used to calculate referred density. The selected
API table is displayed, and the meter automatically changes the density unit, temperature unit,
pressure unit, and reference pressure to match the API table.
Your choice also determines the API table that will be used to calculate the correction factor for volume
(CTPL or CTL).
Restriction
Not all combinations are supported by the API Referral application. See the list of API tables in this
manual.
3. Refer to the API documentation and confirm your table selection.
Verify that your process fluid falls within range for line density, line temperature, and line
a)
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
• 230.0 x 10-6 to 930.0 x 10-6 per °F
• 414.0 x 10-6 to 1674.0 x 10-6 per °C
5. Set Reference Temperature to the temperature to which density will be corrected in referred density
calculations. If you choose Other, select the temperature measurement unit and enter the reference
temperature.
6. Set Reference Pressure to the pressure to which density will be corrected in referred density
calculations.
Set up temperature and pressure data for API Referral using ProLink III
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must
decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART
communications.
If you plan to use the mA Input for temperature data, Channel D must be available for use, and must be wired
to an external temperature device.
If you plan to use the mA Input for pressure data, Channel D must be available for use, and must be wired to
an external pressure device.
Configuration and Use Manual61
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Restriction
You can use an mA Input for either external temperature or external pressure, but not both.
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
If you are using an external temperature device, it must use the temperature unit that is configured in the
transmitter.
2. Choose the method to be used to supply temperature data, and perform the required setup.
OptionDescriptionSetup
Internal RTD
temperature
data
PollingThe meter polls an external
mA InputAn external device sends
Temperature data from the onboard temperature sensor
(RTD) is used.
device for temperature data.
This data will be available in
addition to the internal RTD
temperature data.
temperature data to the meter
through an mA Input. This data
will be available in addition to
the internal temperature data.
a. Set Line Temperature Source to Internal RTD.
Select Apply.
b.
a. Set Line Temperature Source to Poll for External Value.
b.
Set Polling Slot to an available slot.
c. Set Polling Control to Poll as Primary or Poll as Secondary.
OptionDescription
Poll as PrimaryNo other HART masters will be on the
network. A field communicator is not
a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not
a HART master.
d. Set External Device Tag to the HART tag of the
temperature device.
e.
Select Apply.
a. Set Channel D to operate as an mA Input.
b.
Set mA Input Assignment to External Temperature.
c. Set Temperature Source to mA Input.
Digital
communications
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
a. Set Line Temperature Source to Fixed Value or Digital
Communications.
b. Select Apply.
c. Perform the necessary host programming and
communications setup to write temperature data to the
meter at appropriate intervals.
3. Choose the method you will use to supply pressure data, and perform the required setup.
62Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
OptionDescriptionSetup
PollingThe meter polls an external
device for pressure data.
mA InputAn external device sends
pressure data to the meter
through an mA Input.
Digital
communications
A host writes pressure data to
the meter at appropriate
intervals.
a. Set Pressure Source to Poll for External Value.
Set Polling Slot to an available slot.
b.
c. Set Polling Control to Poll as Primary or Poll as Secondary.
OptionDescription
Poll as PrimaryNo other HART masters will be on the
network. A field communicator is not
a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not
a HART master.
d. Set External Device Tag to the HART tag of the
temperature device.
a. Set Channel D to operate as an mA Input.
b.
Set mA Input Assignment to External Pressure.
c. Set Pressure Source to mA Input.
a. Set Pressure Source to Fixed Value or Digital
Communications.
b.
Perform the necessary host programming and
communications setup to write pressure data to the meter
at appropriate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group
on the ProLink III main window.
The current pressure value is displayed in the External Pressure field. Verify that the value is correct.
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For polling:
— Verify the wiring between the meter and the external device.
— Verify the HART tag of the external device.
• For an mA Input:
— Verify the wiring between the meter and the external device.
— Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
— Verify the Lower Range Value and Upper Range Value settings for an mA Input.
Configuration and Use Manual63
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
— Perform an mA Input trim.
— Adjust the mA Input damping value.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the host is writing to the correct register in memory, using the correct data type.
5.1.3 Set up the API Referral application using a field communicator
Enable the API Referral application using a field communicator
Prerequisites
The API Referral application must be licensed and enabled on your transmitter. If the API Referral application
was enabled at the factory, you do not need to enable it now.
Volume Flow Type must be set to Liquid.
Procedure
1.
Choose Configure→Manual Setup→Measurements→Optional Setup→GSV and ensure that
Volume Flow Type is set to Liquid.
This parameter is available only if API Referral or concentration measurement is not enabled. If you do
not see this parameter, it is already set correctly.
2. If the concentration measurement application is enabled, disable it.
The concentration measurement application and the API Referral application cannot be enabled
simultaneously.
3. Enable the API Referral application.
4. If Advance Phase Measurement→Output Type is other than Disabled, disable it.
The Advance Phase Measurement application and the API Referral application cannot be enabled
simultaneously.
Configure API Referral using a field communicator
The API Referral parameters specify the API table, measurement units, and reference values to be used in
referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process
fluid.
You must know the reference temperature and reference pressure that you want to use.
64Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
3. Specify the API table that you want to use to calculate referred density.
Each API table is associated with a specific set of equations.
a)
Set API Table Number to the number that matches the API table units that you want to use for
referred density.
Your choice also determines the measurement unit to be used for temperature and pressure,
and the default values for reference temperature and reference pressure.
API table number
Measurement
unit for referred
density
Temperature
measurement
unit
Pressure
measurement
unit
Default reference
temperature
Default reference
pressure
5°API°Fpsi (g)60 °F0 psi (g)
(1)
6
°API°Fpsi (g)60 °F0 psi (g)
23SGU°Fpsi (g)60 °F0 psi (g)
(1)
24
SGU°Fpsi (g)60 °F0 psi (g)
53kg/m³°CkPa (g)15 °C0 kPa (g)
(1)
54
59
60
(2)
(2)
kg/m³°CkPa (g)15 °C0 kPa (g)
kg/m³°CkPa (g)20 °C0 kPa (g)
kg/m³°CkPa (g)20 °C0 kPa (g)
(1) Used only with API Table Letter = C.
(2) Used only with API Table Letter = E.
b) Set API Table Letter to the letter of the API table group that is appropriate for your process fluid.
Liquids with a constant base density or known thermal expansion coefficient
(TEC). You will be required to enter the TEC for your process fluid.
DLubricating oils
(2)
E
NGL (Natural Gas Liquids) and LPG (Liquid Petroleum Gas)
(1) Used only with API Table Number= 6, 24, or 54.
(2) Used only with API Table Number = 23, 24, 53, 54, 59, or 60.
API Table Number and API Table Letter uniquely identify the API table. The selected API table is
displayed, and the meter automatically changes the density unit, temperature unit, pressure unit,
reference temperature, and reference pressure to match the API table.
Your choice also determines the API table that will be used to calculate the correction factor for volume
(CTPL or CTL).
Configuration and Use Manual65
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Restriction
Not all combinations are supported by the API Referral application. See the list of API tables in this
manual.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
• 230.0 x 10-6 to 930.0 x 10-6 per °F
• 414.0 x 10-6 to 1674.0 x 10-6 per °C
5.
Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
6. If required, set Reference Temperature to the temperature to which density will be corrected in
referred density calculations.
The default reference temperature is determined by the selected API table.
7. If required, set Reference Pressure to the pressure to which density will be corrected in referred
density calculations.
The default reference pressure is determined by the selected API table. API Referral requires gauge
pressure.
Set up temperature and pressure data for API Referral using a field
communicator
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must
decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value
may produce inaccurate process data.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART
communications.
If you plan to use the mA Input for temperature data, Channel D must be available for use, and must be wired
to an external temperature device.
If you plan to use the mA Input for pressure data, Channel D must be available for use, and must be wired to
an external pressure device.
Restriction
You can use an mA Input for either external temperature or external pressure, but not both.
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
66Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
If you are using an external temperature device, it must use the temperature unit that is configured in the
transmitter.
Procedure
1.
Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal RTD
temperature
data
PollingThe meter polls an external
Temperature data from the onboard temperature sensor
(RTD) is used.
device for temperature data.
This data will be available in
addition to the internal RTD
temperature data.
Liquids with a
constant density
base or known
thermal expansion
coefficient
(5)
API tables
(calculations)
Referred
density
(2)
5A6AUnit: °API
(1)
CTL or
CTPL
Referred density
(API): unit and
range
(3) (4)
Default
reference
temp
Default
reference
pressure
API standard
60 °F0 psi (g)API MPMS 11.1
Range: 0 to 100 °API
23A24AUnit: SGU
60 °F0 psi (g)
Range: 0.6110 to
1.0760 SGU
53A54AUnit: kg/m
3
15 °C0 kPa (g)
Range: 610 to
1075 kg/m³
5B6BUnit: °API
60 °F0 psi (g)API MPMS 11.1
Range: 0 to 85 °API
23B24BUnit: SGU
60 °F0 psi (g)
Range: 0.6535 to
1.0760 SGU
53B54BUnit: kg/m
3
15 °C0 kPa (g)
Range: 653 to
1075 kg/m³
N/A6CUnit: °API60 °F0 psi (g)API MPMS 11.1
N/A24CUnit: SGU60 °F0 psi (g)
N/A54CUnit: kg/m³15 °C0 kPa (g)
5D6DUnit: °API
60 °F0 psi (g)API MPMS 11.1
Range: −10 to
+40 °API
Lubricating oils
23D24DUnit: SGU
Range: 0.8520 to
60 °F0 psi (g)
1.1640 SGU
53D54DUnit: kg/m³
15 °C0 kPa (g)
Range: 825 to
1164 kg/m³
70Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Table 5-1: API tables, process fluids, measurement units, and default reference values (continued)
API tables
Process fluid
NGL (natural gas
liquids) and LPG
(liquid petroleum
gas)
(1) Each API table represents a specialized equation defined by the American Petroleum Institute for a specific combination
of process fluid, line conditions, and output.
(2) Referred density is calculated from line density. You must specify this table, either directly or by selecting the process
fluid and base density measurement unit.
(3) You do not need to specify this table. It is invoked automatically as a result of the previous table selection.
(4) CTL is a correction factor based on online temperature. CTPL is a correction factor based on both line pressure and line
temperature. Calculation of CTL and CTPL for A, B, C, and D table products is in accordance with API MPMS Chapter
11.1. Calculation of CTL and CTPL for E table products is in accordance with API MPMS Chapters 11.2.2, 11.2.4, and
11.2.5.
(5) The Thermal Expansion Coefficient (TEC) replaces the referred density calculation. Use the CTL/CTPL table instead.
5.1.5
Process variables from the API Referral application
(calculations)
Referred
density
23E24EUnit: SGU60 °F0 psi (g)API MPMS 11.2.4
53E54EUnit: kg/m³15 °C0 psi (g)
59E60EUnit: kg/m³20 °C0 psi (g)
(2)
(1)
CTL or
CTPL
Referred density
(API): unit and
range
(3) (4)
Default
reference
temp
Default
reference
pressure
API standard
The API Referral application calculates several different process variables according to API standards.
CTPL
Correction factor based on line temperature and line pressure.
CTL
Referred density
API volume flow
Batch-weighted
average density
Batch-weighted
average
temperature
API volume total
API volume
inventory
Correction factor based on line temperature at saturation conditions.
The measured density after CTL or CTPL has been applied.
The measured volume flow rate after CTL or CTPL has been applied. Also called
corrected volume flow.
One density value is recorded for each unit of flow (e.g., barrel, liter). The average is
calculated from these values. The average is reset when the API totalizer is reset. Not
available unless a totalizer has been configured with Source set to Corrected Volume
Flow.
One temperature value is recorded for each unit of flow (e.g., barrel, liter). The
average is calculated from these values. The average is reset when the API totalizer is
reset. Not available unless a totalizer has been configured with Source set to
Temperature-Corrected Volume Flow.
The total API volume measured by the transmitter since the last API totalizer reset.
Also called corrected volume total. Not available unless a totalizer has been
configured with Source set to Corrected Volume Flow.
The total API volume measured by the transmitter since the last API inventory reset.
Also called corrected volume inventory. Not available unless an inventory has been
configured with Source set to Corrected Volume Flow.
Configuration and Use Manual71
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
5.2 Set up concentration measurement
The concentration measurement application calculates concentration from line density and line temperature.
5.2.1
The procedure for setting up concentration measurement application depends on how your device was
ordered and how you want to use the application. Review this information before you begin.
Requirements for concentration measurement
To use the concentration measurement application, the following conditions must be met:
• The concentration measurement application must be enabled.
• The API Referral application must be disabled.
• The gas piecewise linearization (PWL) application must be disabled.
• The Advanced Phase Measurement application must be disabled or set for the Liquid with Gas application.
• A concentration matrix must be loaded into one of the six slots on the transmitter.
• Temperature Source must be configured and set up.
• One matrix must be selected as the active matrix (the matrix used for measurement).
Requirements for matrices
A matrix is the set of coefficients used to convert process data to concentration, plus related parameters. The
matrix can be saved as a file.
Preparing to set up concentration measurement
Tip
In most cases, the concentration matrix that you ordered was loaded at the factory. If it was not, you have
several options for loading a matrix. You can also build a matrix.
The transmitter requires all matrices to be in .matrix format. You can use ProLink III to load matrices in other
formats:
• .edf (used by ProLink II)
• .xml (used by ProLink III)
The transmitter can store matrices in two locations:
• One of the six slots in memory
• The transmitter SD card
Any matrix in a slot is available for use. In other words, it can be selected as the active matrix and used for
measurement. Matrices on the SD card are not available for use. Matrices must be loaded into a slot before
they can be used for measurement.
All matrices in slots must use the same derived variable. Matrices on the SD card have no requirement for
their derived variables to match.
See the following table for the different ways that you can load matrices.
72Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Table 5-2: Methods to load a matrix file
ActionDisplayProLink IIIField communicator
Load matrix from USB drive to SD card✓
Load matrix from computer to slot✓
Load matrix from SD card to slot✓✓✓
Requirements for derived variables
A derived variable is the process variable that a concentration matrix measures. All other process variables are
calculated from the derived variable. There are eight possible derived variables. Each matrix is designed for
one specific derived variable.
The transmitter can store up to six matrices in six slots. There are additional matrices on the transmitter SD
card. All matrices in the six slots must use the same derived variable. If you change the setting of DerivedVariable, all matrices are deleted from the six slots. Any matrices on the transmitter SD card are not affected.
Tip
Always ensure that Derived Variable is set correctly before loading matrices into slots.
Derived variables and net flow rate
If you want the transmitter to calculate Net Mass Flow Rate, the derived variable must be set to Mass
Concentration (Density). If your matrix is not designed for Mass Concentration (Density), contact customer
support for assistance.
If you want the transmitter to calculate Net Volume Flow Rate, the derived variable must be set to Volume
Concentration (Density). If your matrix is not designed for Volume Concentration (Density), contact customer
support for assistance.
Derived variables based on specific gravity
The following derived variables are based on specific gravity:
• Specific Gravity
• Concentration (Specific Gravity)
• Mass Concentration (Specific Gravity)
• Volume Concentration (Specific Gravity)
If you are using one of these derived variables, two additional parameters can be configured:
• Reference Temperature of Water (default setting: 4 °C)
• Water Density at Reference Temperature (default setting: 999.99988 kg/m³)
These two parameters are used to calculate specific gravity.
You cannot set these parameters from the display. If the default values are not appropriate, you must use
another method to set them.
Configuration and Use Manual73
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Optional tasks in setting up concentration measurement
The following tasks are optional:
• Modifying names and labels
• Configuring extrapolation alerts
5.2.2
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Restriction
This section does not cover building a concentration matrix. For detailed information on building a matrix, see
the Micro Motion Enhanced Density Application Manual.
Set up concentration measurement using the display
Enable concentration measurement using the display
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration
measurement cannot be enabled at the same time:
• Advanced Phase Measurement — gas with liquid
• Advanced Phase Measurement — net oil
• API Referral
• Piecewise linearization (PWL)
• Gas Standard Volume
Procedure
1.
Choose Menu→Configuration→Process Measurement.
2. Choose Flow Variables→Volume Flow Settings and ensure that Flow Type is set to Liquid.
3. Return to the Process Measurement menu.
4. If the API Referral application is displayed in the menu, choose API Referral and ensure that Enabled/Disabled is set to Disabled.
The concentration measurement application and the API Referral application cannot be enabled
simultaneously.
5. If the Advanced Phase Measurement application is displayed in the menu, choose Advanced PhaseMeasurement→Application Setup and ensure that Enabled/Disabled is set to Disabled.
6. Enable concentration measurement.
a) Choose Menu→Configuration→Process Measurement→Concentration Measurement.
b) Set Enabled/Disabled to Enabled.
74Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Load a concentration matrix from a USB drive using the display
At least one concentration matrix must be loaded into one of the six slots on your transmitter. You can load
up to six matrices into slots. You can also copy matrices to the transmitter SD card, and load them into slots at
a later time.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not
need to load any matrices.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is powered
up. Failure to follow these instructions can cause an explosion resulting in injury or death.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The
transmitter SD card and the ProLink III installation include a set of standard concentration matrices. Other
matrices are available from Emerson.
Each concentration matrix file must be in .matrix format.
Tip
• If you have a custom matrix on another device, you can save it to a file, then load it to the current device.
• If you have a matrix file in a different format, you can load it using ProLink III.
The .matrix files must be copied to the root directory of a USB drive.
You must know the derived variable that the matrix is designed to calculate.
Important
• All concentration matrices on your transmitter must use the same derived variable.
• If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the SD card. Set Derived Variable before loading concentration
matrices.
Procedure
1.
Choose Menu→Configuration→Process Measurement→Concentration Measurement→Configure Application and ensure that the setting of Derived Variable matches the derived variable
used by your matrix. If it does not, change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots, but not from the transmitter SD card. Verify the setting of Derived Variable before
continuing.
2. Load the matrix.
Remove the cover from the transmitter wiring compartment, open the snap flap to access the
a)
service port, and insert the USB drive into the service port.
Configuration and Use Manual75
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
b) Choose Menu→USB Options→USB Drive --> Transmitter→Upload Configuration File.
c)
Set Config File Type to Concentration Measurement Matrix.
d) Select the .matrix file that you want to load, and wait for the transfer to complete.
3. Choose Yes or No when you are asked if you want to apply the settings.
The transmitter has six slots that are used to store concentration matrices. Any one of these can be
used for measurement. The transmitter also has the capability to store multiple concentration matrices
on its SD card. These cannot be used for measurement until they are moved to a slot.
Option Description
YesThe matrix is saved to the SD card, and the loading process continues with loading the
matrix into one of the slots.
NoThe matrix is saved to the SD card, and the loading process ends. You must load a matrix into
a slot before you can use it for measurement.
4. If you chose Yes, select the slot to load this matrix into, and wait until the load is complete.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
Postrequisites
If you loaded the matrix into a slot, choose Menu → Configuration → Process Measurement →
Concentration Measurement → Configure Application → Active Matrix and ensure that the matrix is listed.
If you loaded the matrix onto the SD card only, choose Menu → Configuration → Process Measurement →
Concentration Measurement → Load Matrix and ensure that the matrix is listed.
Load a concentration matrix from the SD card using the display
If you have a concentration matrix on the transmitter SD card, you can load it into one of the six slots on your
transmitter. You cannot use the matrix for measurement until it has been loaded into a slot. You can load up
to six matrices into slots.
Prerequisites
You must have one or more concentration matrices stored on the transmitter SD card. The standard matrices
are loaded to the SD card at the factory.
You must know the derived variable that the matrix is designed to calculate.
Procedure
Choose Menu→Configuration→Process Measurement→Concentration Measurement and ensure
1.
that the setting of Derived Variable matches the derived variable used by your matrix. If it does not,
change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots, but not from the transmitter SD card. Verify the setting of Derived Variable before
continuing.
76Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
Matrix.
The transmitter displays a list of all matrices that are on the SD card.
Select the matrix that you want to load.
3.
4. Select the slot that you want to load it into.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
Postrequisites
Choose Menu → Configuration → Process Measurement → Concentration Measurement → Configure
Application → Active Matrix and ensure that the matrix is listed.
Set up temperature data using the display
The concentration measurement application uses line temperature data in its calculations. You must decide
how to provide this data, then perform the required configuration and setup. Temperature data from the onboard temperature sensor (RTD) is always available. Optionally, you can set up an external temperature
device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on
this meter.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the
internal RTD temperature in some areas and an external temperature in others. The transmitter stores the
internal RTD temperature and the external temperature separately. However, the transmitter stores only one
alternate temperature value, which may be either the external temperature or the configured fixed value.
Accordingly, if you choose a fixed temperature for some uses, and an external temperature for others, the
external temperature will overwrite the fixed value.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART
communications.
If you plan to use the mA Input for temperature data, Channel D must be available for use, and must be wired
to an external temperature device.
Procedure
Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal
temperature
Temperature data from the onboard temperature sensor
(RTD) will be used for all
measurements and
calculations. No external
temperature data will be
available.
a. Choose Menu→Configuration→Process Measurement
→Temperature.
Set External Temperature to Off.
b.
Configuration and Use Manual77
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
MethodDescriptionSetup
PollingThe meter polls an external
device for temperature data.
This data will be available in
addition to the internal
temperature data.
mA InputAn external device sends
temperature data to the meter
via the mA Input. This data will
be available in addition to the
internal temperature data.
a. Choose Menu→Configuration→Process Measurement
→Temperature.
Set External Temperature to On.
b.
c. Choose Poll External Device.
d. Select Polled Variable 1 or Polled Variable 2.
e. Set Variable to External Temperature.
f. Set Polling Control to Poll as Primary or Poll as Secondary.
OptionDescription
Poll as PrimaryNo other HART masters will be on the
network. A field communicator is not
a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not
a HART master.
g. Set External Device Tag to the HART tag of the external
temperature device.
a. Choose Menu→Configuration→Process Measurement
→Temperature.
Set External Temperature to On.
b.
c. Choose Menu→Configuration→Inputs/Outputs→
Channel D.
d. Set I/O Type to MAI.
e. Set Power appropriately.
f. Choose I/O Settings.
g. Set Assignment to External Temperature.
h. Set Lower Range Value and Upper Range Value
appropriately.
Digital
communications
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal temperature data.
a. Choose Menu→Configuration→Process Measurement
→Temperature.
b.
Set External Temperature to On.
c. Perform the necessary host programming and
communications setup to write temperature data to the
transmitter at appropriate intervals.
Postrequisites
Choose Menu→Service Tools→Service Data→View Process Variables and verify the value for External
Temperature.
78Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For polling:
— Verify the wiring between the meter and the external device.
— Verify the HART tag of the external device.
• For an mA Input:
— Verify the wiring between the meter and the external device.
— Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
— Verify the Lower Range Value and Upper Range Value settings for an mA Input.
— Perform an mA Input trim.
— Adjust the mA Input damping value.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the host is writing to the correct register in memory, using the correct data type.
Modify matrix names and labels using the display
For convenience, you can change the name of a concentration matrix and the label used for its measurement
unit. This does not affect measurement.
3. Set Matrix Name to the name that will be used for this matrix.
4. Set Concentration Unit to the label that will be used for the concentration unit.
If you want to use a custom label, you can use the display to select Special. However, you cannot use
the display to configure the custom label. You must use another tool to change the label from Special
to a user-defined string.
Modify extrapolation alerts using the display
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control
the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density
or line temperature goes outside the range, the transmitter will extrapolate concentration values. However,
extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is
occurring.
Each concentration matrix has its own extrapolation alert limits.
Configuration and Use Manual79
Configure process measurement applicationsConfiguration and Use Manual
5. Enable or disable the high and low limit alerts for temperature and density as desired.
Example: Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using the display
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Menu → Configuration → Process Measurement → Concentration Measurement →
Configure Application.
2. Set Active Matrix to the matrix you want to use.
5.2.3 Set up concentration measurement using ProLink III
This section guides you through the tasks required to set up, configure, and implement concentration
measurement.
Enable concentration measurement using ProLink III
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration
measurement cannot be enabled at the same time:
• Advanced Phase Measurement — gas with liquid
• Advanced Phase Measurement — net oil
• API Referral
• Piecewise linearization (PWL)
• Gas Standard Volume
80Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Procedure
1.
Choose Device Tools→Configuration→Process Measurement→Flow and ensure that VolumeFlow Type is set to Liquid Volume.
4. Disable API Referral and set the Advance Phase Measurement application to Disabled or Single Liquid.
5. Disable gas Piecewise Linearization (PWL), and set the Advance Phase Measurement application to
Disabled or Single Liquid.
6. Set Concentration Measurement to Enabled and select Apply.
Load a concentration matrix using ProLink III
At least one concentration matrix must be loaded onto your transmitter. You can load up to six.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The ProLink III
installation includes a set of standard concentration matrices. Other matrices are available from Emerson. The
file can be on your computer or in the transmitter internal memory.
The file must be in one of the formats that ProLink III supports. This includes:
• .xml (ProLink III)
• .matrix (5700)
If you are loading an .xml file, you must know the following information for your matrix:
• The derived variable that the matrix is designed to calculate
• The density unit that the matrix was built with
• The temperature unit that the matrix was built with
If you are loading a .matrix file, you must know the derived variable that the matrix is designed to calculate.
Important
• All concentration matrices on your transmitter must use the same derived variable.
• If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the transmitter SD card. Set Derived Variable before loading
concentration matrices.
• ProLink III loads matrices directly to one of the six transmitter slots.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not
need to load any matrices.
Restriction
You cannot use ProLink III to load a matrix to the transmitter SD card.
Configuration and Use Manual81
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Procedure
1.
If you are loading an .xml file, choose Device Tools→Configuration→Process Measurement→LineDensity and set Density Unit to the density unit used by your matrix.
Important
When you load a matrix in one of these formats, if the density unit is not correct, concentration data
will be incorrect. The density units must match at the time of loading. You can change the density unit
after the matrix is loaded.
2. If you are loading an .xml file, choose Device Tools→Configuration→Process Measurement→Line
Temperatureand set Temperature Unit to the temperature unit used by your matrix.
Important
When you load a matrix in one of these formats, if the temperature unit is not correct, concentration
data will be incorrect. The temperature units must match at the time of loading. You can change the
temperature unit after the matrix is loaded.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
4.
In Step 1, ensure that the setting of Derived Variable matches the derived variable used by your
matrix. If it does not, change it as required and select Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots. Verify the setting of Derived Variable before continuing.
5. Load one or more matrices.
In Step 2, set Matrix Being Configured to the location (slot) to which the matrix will be loaded.
a)
b) To load a .xml file from your computer, select Load Matrix from File, navigate to the file, and
load it.
c) To load a .matrix file from your computer, select Load Matrix from My Computer, navigate to
the file, and load it.
d) To load a .matrix file from the transmitter internal memory, select Load Matrix from 5700 Device
Memory, navigate to the file on the transmitter, and load it.
e) Repeat until all required matrices are loaded.
Set reference temperature values for specific gravity using ProLink III
When Derived Variable is set to any option based on specific gravity, you must set the reference temperature
for water, then verify the density of water at the configured reference temperature. These values affect
specific gravity measurement.
This requirement applies to the following derived variables:
• Specific Gravity
• Concentration (Specific Gravity)
82Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
• Mass Concentration (Specific Gravity)
• Volume Concentration (Specific Gravity)
Procedure
1.
Choose Device Tools→Configuration→Process Measurement→Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and select Change
Matrix.
3. Scroll to Step 3, then perform the following actions:
a) Set Reference Temperature for Referred Density to the temperature to which line density will
be corrected for use in the specific gravity calculation.
b) Set Reference Temperature for Water to the water temperature that will be used in the specific
gravity calculation.
c) Set Water Density at Reference Temperature to the density of water at the specified reference
temperature.
The transmitter automatically calculates the density of water at the specified temperature. The
new value will be displayed the next time that transmitter memory is read. You can enter a
different value if you prefer.
4. Select Apply at the bottom of Step 3.
Set up temperature data using ProLink III
The concentration measurement application uses line temperature data in its calculations. You must decide
how to provide this data, then perform the required configuration and setup. Temperature data from the onboard temperature sensor (RTD) is always available. Optionally, you can set up an external temperature
device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on
this meter.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the
internal RTD temperature in some areas and an external temperature in others. The transmitter stores the
internal RTD temperature and the external temperature separately. However, the transmitter stores only one
alternate temperature value, which may be either the external temperature or the configured fixed value.
Accordingly, if you choose a fixed temperature for some uses, and an external temperature for others, the
external temperature will overwrite the fixed value.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART
communications.
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2.
Scroll to Step 4.
3. Choose the method to be used to supply temperature data, and perform the required setup.
OptionDescriptionSetup
Internal
temperature
PollingThe meter polls an external
Temperature data from the onboard temperature sensor
(RTD) will be used for all
measurements and
calculations. No external
temperature data will be
available.
device for temperature data.
This data will be available in
addition to the internal RTD
temperature data.
a. Set Line Temperature Source to Internal.
Click Apply.
b.
a. Set Line Temperature Source to Poll for External Value.
b. Set Polling Slot to an available slot.
c. Set Polling Control to Poll as Primary or Poll as Secondary.
OptionDescription
Poll as PrimaryNo other HART masters will be on the
network. A field communicator is not
a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not
a HART master.
d. Set External Device Tag to the HART tag of the
temperature device.
e.
Click Apply.
mA InputAn external device sends
temperature data to the meter
via the mA Input. This data will
be available in addition to the
internal temperature data.
Digital
communications
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
a. Set Channel D to operate as an mA Input.
b.
Set mA Input Assignment to External Temperature.
c. Set Pressure Source to mA Input.
a. Set Line Temperature Source to Fixed Value or Digital
Communications.
b. Click Apply.
c. Perform the necessary host programming and
communications setup to write temperature data to the
meter at appropriate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group
on the ProLink III main window.
84Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For polling:
— Verify the wiring between the meter and the external device.
— Verify the HART tag of the external device.
• For an mA Input:
— Verify the wiring between the meter and the external device.
— Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
— Verify the Lower Range Value and Upper Range Value settings for an mA Input.
— Perform an mA Input trim.
— Adjust the mA Input damping value.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the host is writing to the correct register in memory, using the correct data type.
Modify matrix names and labels using ProLink III
For convenience, you can change the name of a concentration matrix and the label used for its measurement
unit. This does not affect measurement.
Procedure
1.
Choose Device Tools→Configuration→Process Measurement→Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change
Matrix.
3. Scroll to Step 3, then perform the following actions:
a) Set Concentration Units Label to the label that will be used for the concentration unit.
b) If you set Concentration Units Label to Special, enter the custom label in User-Defined Label.
c) In Matrix Name, enter the name to be used for the matrix.
4. Select Apply at the bottom of Step 3.
Modify extrapolation alerts using ProLink III
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control
the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density
or line temperature goes outside the range, the transmitter will extrapolate concentration values. However,
extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is
occurring.
Configuration and Use Manual85
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Each concentration matrix has its own extrapolation alert limits.
Procedure
1.
Choose Device Tools→Configuration→Process Measurement→Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to
perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change
Matrix.
3. Scroll to Step 4.
4. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert will be posted.
5. Enable or disable the high and low limit alerts for temperature and density, as desired, and click Apply.
Example: Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
The Concentration Measurement window displays. It is organized into steps that allow you to perform
several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Active Matrix to the matrix you want to use and select Change Matrix.
5.2.4 Set up concentration measurement using a field
communicator
This section guides you through most of the tasks related to setting up and implementing the concentration
measurement application.
Restriction
This section does not cover building a concentration matrix. See the Micro Motion Enhanced Density Application
Manual for detailed information on building a matrix.
Enable concentration measurement using a field communicator
The concentration measurement application must be enabled before you can perform any setup. If the
concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration
measurement cannot be enabled at the same time:
86Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
• Advanced Phase Measurement — gas with liquid
• Advanced Phase Measurement — net oil
• API Referral
• Piecewise linearization (PWL)
• Gas Standard Volume
Procedure
1.
Choose Overview→Device Information→Licenses→Enable/Disable Applications and ensure that
Volume Flow Type is set to Liquid.
3. Enable the concentration measurement application.
Load a concentration matrix from the transmitter SD card using a field
communicator
If you have a concentration matrix on the transmitter SD card, you can move it into one of the six slots on your
transmitter. You cannot use the matrix for measurement until it has been loaded into a slot. You can load up
to six matrices into slots.
Prerequisites
You must have one or more concentration matrices loaded onto the transmitter SD card.
You must know the derived variable that the matrix is designed to calculate.
CM Configuration and ensure that the setting of Derived Variable matches the derived variable usedby your matrix. If it does not, change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from
the six slots, but not from the transmitter SD card. Verify the setting of Derived Variable before
continuing.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
4. Enter the name of the matrix file on the SD card, without the .matrix extension.
Example
If the matrix file name is test.matrix, enter test.
Configuration and Use Manual87
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Postrequisites
Choose Configure → Manual Setup → Measurements → Optional Setup → Conc Measurement → CM
Configuration → Active Matrix and ensure that the selected slot contains the matrix that you loaded.
Set reference temperature values for specific gravity using a field
communicator
Field communicatorConfigure→Manual Setup→Measurements→Optional Setup→Concentration Measurement→
When Derived Variable is set to any option based on specific gravity, you must set the reference temperature
for water, then verify the density of water at the configured reference temperature. These values affect
specific gravity measurement.
To check the setting of Derived Variable, choose Configure → Manual Setup → Measurements → Optional
Setup → Conc Measurement → CM Configuration.
Important
Do not change the setting of Derived Variable. If you change the setting of Derived Variable, all existing
concentration matrices will be deleted from transmitter memory.
Procedure
1.
Set Matrix Being Configured to the matrix you want to modify.
2. Choose Reference Conditions, then perform the following actions:
a) Set Reference Temperature to the temperature to which line density will be corrected for use in
the specific gravity calculation.
b) Set Water Reference Temperature to the water temperature that will be used in the specific
gravity calculation.
c) Set Water Reference Density to the density of water at the specified reference temperature.
Configuration Matrix
The transmitter automatically calculates the density of water at the specified temperature. The
new value will be displayed the next time that transmitter memory is read. Optionally, you can
enter a different value.
Provide temperature data using a field communicator
The concentration measurement application uses line temperature data in its calculations. You must decide
how to provide this data, then perform the required configuration and setup. Temperature data from the onboard temperature sensor (RTD) is always available. Optionally, you can set up an external temperature
device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on
this meter.
Procedure
1. Choose the method to be used to supply temperature data, and perform the required setup.
88Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
MethodDescriptionSetup
Internal RTD
temperature data
Digital
communications
Temperature data from the onboard temperature sensor
(RTD) is used.
A host writes temperature data
to the meter at appropriate
intervals. This data will be
available in addition to the
internal RTD temperature data.
a. Choose Configure→Manual Setup→Measurements→
Optional Setup→External Variables
Set Temperature Compensation to Disable.
b.
a. Choose Configure→Manual Setup→Measurements→
Optional Setup→External Variables.
b.
Set Temperature Compensation to Enable.
c. Perform the necessary host programming and
communications setup to write temperature data to the
meter at appropriate intervals.
2. Choose the method to be used to supply temperature data, and perform the required setup.
MethodDescriptionSetup
Internal RTD
temperature data
PollingThe meter polls an external
Temperature data from the onboard temperature sensor
(RTD) is used.
device for temperature data.
This data will be available in
addition to the internal RTD
temperature data.
communications setup to write temperature data to the
meter at appropriate intervals.
Postrequisites
To verify the value for External Temperature, choose Service Tools → Variables → Process → External
Temperature.
Need help?
If the value is not correct:
• Ensure that the external device and the meter are using the same measurement unit.
• For polling:
— Verify the wiring between the meter and the external device.
— Verify the HART tag of the external device.
• For the mA Input:
— Verify the wiring between the meter and the external device.
— Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
— Verify the Lower Range Value and Upper Range Value settings for the mA Input.
— Perform an mA Input trim.
— Adjust the mA Input damping value.
• For digital communications:
— Verify that the host has access to the required data.
— Verify that the host is writing to the correct register in memory using the correct data type.
90Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Modify matrix names and labels using a field communicator
Field communicatorConfigure→Manual Setup→Measurements→Optional Setup→Conc Measurement→Configure
For convenience, you can change the name of a concentration matrix and the label used for its measurement
unit. This does not affect measurement.
Procedure
Set Matrix Being Configured to the matrix you want to modify.
1.
2. Set Matrix Name to the name to be used for the matrix.
3. Set Concentration Unit to the label that will be used for the concentration unit.
4. If you set Concentration Unit to Special, choose Label and enter the custom label.
Matrix
Modify extrapolation alerts using a field communicator
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control
the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density
or line temperature goes outside the range, the transmitter will extrapolate concentration values. However,
extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is
occurring.
Each concentration matrix has its own extrapolation alert limits.
2. Set Matrix Being Configured to the matrix you want to modify.
3. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert will be posted.
4. Choose Configure→Alert Setup→CM Alerts.
5. Enable or disable the high and low alerts for temperature and density, as desired.
Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a
temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if
line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using a field communicator
Field communicatorConfigure→Manual Setup→Measurements→Optional Setup→Conc Measurement→CM
You must select the concentration matrix to be used for measurement. Although the transmitter can store up
to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
Set Active Matrix to the matrix you want to use.
Configuration
Configuration and Use Manual91
Configure process measurement applicationsConfiguration and Use Manual
Field communicatorConfigure→Manual Setup→Inputs/Outputs→Batcher→Setup Presets
A batch preset is a predefined group of batching parameters. You can define and save up to six batch presets.
When you run a batch, you must choose a preset. You can modify the batch target before you run the batch.
All other settings are fixed.
Procedure
Select the preset number that you want to configure.
1.
2. Set the preset name to the name to be used for this preset.
The name can be up to eight characters in length. Valid characters include A-Z and 0-9.
3. Set the preset target to the size of the batch.
When the transmitter has measured the specified amount and applied the selected form of AOC, it
closes the valve.
4. Optional: Set the preset end warning point to the point in the batch at which a Modbus coil will be set.
The end warning allows the transmitter to alert the operator to the end of batch. The operator can
prepare for the end of the batch, e.g., by adjusting the flow rate to avoid overflow.
5. Enable or disable the preset as desired.
Configuration and Use Manual93
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Note
This field is named Preset Status in the display.
OptionDescription
Enabled (On)The preset is available for use. You can select it to run a batch.
Disabled (Off)The preset is not available for use. You cannot select it to run a batch. You
can always enable it at a later time.
Preset 1 is always enabled and cannot be disabled.
Optional: For two-stage batching, set % Target or Quantity for the following parameters:
6.
Two-stage batching is a batch that is run on a gas and liquid mixture flowing through a pipeline.
% TargetThe values for Open Primary, Open Secondary, Close Primary, and End
Warning are each configured as a percent of target.
QuantityThe values for Open Primary and Open Secondary are each configured as
a quantity at which the valve should open. The values for Close Primary
and End Warning are each configured as a quantity that is subtracted
from the target.
ParameterDescription
End WarningThe % target or quantity level that must be reached to alert the operator
that a batch is ending.
Open PrimaryThe % target or quantity level that must be reached to open the primary
valve.
Open SecondaryThe % target or quantity level that must be reached to open the secondary
valve.
Close PrimaryThe % target or quantity level that must be reached to close the primary
valve. When configuring by quantity, the amount is referenced to the end
of the batch and not to the beginning of the batch.
Field communicatorConfigure→Manual Setup→Inputs/Outputs
The transmitter runs a batch by opening and closing a valve. You can optionally configure a Discrete Output
(DO) to send the open and close commands to the valve.
Prerequisites
• Channel B, Channel C, or Channel D must be available for valve control.
• The selected channel must be wired to the valve.
• The selected channel and the valve must be powered by a power supply, not by the transmitter.
94Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure process measurement applications
MMI-20025166October 2021
Procedure
1.
Verify the wiring between the selected channel and the valve.
2. Configure the selected channel to operate as a Discrete Output.
a) Set Channel B, Channel C, or Channel D to operate as a Discrete Output.
b) Set Power Source to External.
3. Configure the channel for valve control.
a) Do one of the following depending on which tool you are using:
ToolSteps
Display1. Select the Discrete Output to be used for valve control.
Choose I/O Settings.
2.
ProLink III1. Navigate to Device Tools→Configuration→I/O→Outputs→
Discrete Output.
2.
Select the Discrete Output to be used for valve control.
Field communicator1. Select the Discrete Output to be used for valve control.
2.
Select Discrete Output x.
b) Set Source to the batch primary valve.
Set Polarity to match the wiring.
c)
The ON state of the Discrete Output must open the valve, and the OFF state must close the
valve.
d) Set Fault Action so that the valve will be closed if a fault occurs.
The appropriate setting is either Upscale or Downscale. The setting to use for your installation
depends on your valve type, your wiring, and your setting for Polarity.
4. Optional: For two-stage batching, assign one of the Discrete Outputs to the secondary valve.
You can also assign one of the Discrete Outputs to a pump when the batcher is configured for single
stage or two stage.
Field communicatorConfigure→Manual Setup→Inputs/Outputs→Batcher→Batcher
Automatic Overshoot Compensation (AOC) adjusts the timing of the valve close command to minimize
overshoot. The valve close command is sent before the target is reached, to compensate for the time
required for the valve to close completely. AOC is applied to all presets and all batches.
There are three options for AOC:
Configuration and Use Manual95
Configure process measurement applicationsConfiguration and Use Manual
October 2021MMI-20025166
Compensation Off
The transmitter sends the close command when the measured batch total reaches
the configured target.
Fixed
Compensation
Value
AOC Algorithm
The transmitter sends the close command when the measured batch total equals the
target minus the value configured for Fixed Overshoot Compensation. This value is
configured in the current mass or volume unit, and applies to all presets.
The transmitter compares the actual measured amount of each batch to the batch
target, and determines the adjustment by calibrating itself according to an internal
algorithm. You can choose to stop AOC calibration when you are satisfied with the
result, or you can set up rolling (continuous) AOC calibration.
Procedure
• To turn compensation off:
ToolSteps
DisplaySet Compensation Mode to Off.
ProLink IIISet AOC Compensation Mode to Compensation Off.
Field communicatorSet Compensation Mode to No Compensation.
• To use the Fixed Compensation Value option:
a)
Set the AOC compensation mode to fixed.
ToolSteps
DisplaySet Compensation Mode to Fixed.
ProLink IIISet AOC Compensation Mode to Fixed Compensation Value.
Field communicatorSet Compensation Mode to Fixed Value.
b) Set the fixed value or the fixed overshoot compensation to the quantity to be subtracted from the
batch target.
Important
Adjust the fixed value:
— Whenever the process changes (e.g., flow rate, process fluid)
— Whenever you change any configuration parameter that can affect flow measurement (e.g., damping)
• To use the AOC Algorithm option:
ToolSteps
DisplaySet Compensation Mode to AOC.
ProLink IIISet AOC Compensation Mode to AOC Algorithm.
Field communicatorSet Compensation Mode to AOC.
96Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure advanced options for process measurement
MMI-20025166October 2021
6 Configure advanced options for process
measurement
6.1
DisplayMenu→Configuration→Process Measurement→Response Time
ProLink IIIDevice Tools→Configuration→Process Measurement→Response Time
Field communicatorNot available
Response Time controls the speed of various internal processes that are involved in retrieving electronic data
from the sensor and converting it to process data.
Response Time affects all process and diagnostic variables.
Restriction
Response Time is configurable only if you are using the enhanced core processor. If you are using the
standard core processor, Response Time is set to Low Filtering and cannot be changed.
Procedure
Set Response Time as desired.
OptionDescription
NormalAppropriate for typical applications.
High Filtering Slower response. Appropriate for applications with significant amount of entrained gas or
Configure Response Time
process noise.
Low Filtering Fastest response. Appropriate for proving or filling applications.
ServiceDo not select unless directed by Emerson personnel.
6.2 Detect and report two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues.
The transmitter provides two methods to detect and report or respond to two-phase flow.
6.2.1 Detect two-phase flow using density
Display
ProLink III
Field communicator
Configuration and Use Manual97
Menu → Configuration → Process Measurement → Density
Device Tools → Configuration → Process Measurement → Density
Configure advanced options for process measurementConfiguration and Use Manual
October 2021MMI-20025166
The transmitter can use line density data to detect two-phase flow (gas in a liquid process or liquid in a gas
process). The density limits are user-specified. When two-phase flow is detected, an alert is posted.
Procedure
1.
Set Two-Phase Flow Low Limit to the lowest density value that is considered normal in your process.
Values below this will cause the transmitter to post a Process Aberration alert.
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the occurrence of twophase flow alerts that are not significant to your process, set Two-Phase Flow Low Limit slightly below
your expected lowest process density.
You must enter Two-Phase Flow Low Limit in g/cm³, even if you configured another unit for density
measurement.
• Default: 0 g/cm³
• Range: 0 g/cm³ to the sensor limit
Set Two-Phase Flow High Limit to the highest density value that is considered normal in your process.
2.
Values above this will cause the transmitter to post a Process Aberration alert.
Tip
To reduce the occurrence of two-phase flow alerts that are not significant to your process, set TwoPhase Flow High Limit slightly above your expected highest process density.
You must enter Two-Phase Flow High Limit in g/cm³, even if you configured another unit for density
measurement.
• Default: 5 g/cm³
• Range: 5 g/cm³ to the sensor limit
Set Two-Phase Flow Timeout to the number of seconds that the transmitter will wait for a two-phase
3.
flow condition to clear before posting the alert.
• Default: 0 seconds, meaning that the alert will be posted immediately
• Range: 0 to 60 seconds
6.2.2 Detect two-phase flow using sensor diagnostics
The transmitter always monitors sensor diagnostics and applies a two-phase flow algorithm. You can assign
an mA Output to report the results of this calculation: single-phase flow, moderate two-phase flow, or severe
two-phase flow. Severe two-phase flow can cause the meter to stop functioning.
Procedure
Set mA Output Source to Two-Phase Flow Detection.
98Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use ManualConfigure advanced options for process measurement
MMI-20025166October 2021
The signal from the mA Output indicates the current state of the process:
Field communicatorConfigure→Manual Setup→Inputs/Outputs→Channel x→Discrete Output x→Flow Switch
Flow Rate Switch is used to indicate that the flow rate has moved past a user-specified setpoint, in either
direction. The flow rate switch is implemented with a user-configurable hysteresis.
Typically, a Discrete Output is assigned as the flow rate switch indicator. The Discrete Output can be wired to
an external device such as a light or a horn.
Prerequisites
A channel must be configured as a Discrete Output, and the Discrete Output must be available for this use.
Procedure
Configure Flow Rate Switch
Set Discrete Output Source to Flow Switch, if you have not already done so.
1.
2. Set Flow Switch Variable to the flow variable that you want to use to control the flow rate switch.
3. Set Flow Switch Setpoint to the value at which the flow switch will be triggered (after Hysteresis is
applied).
Depending on the polarity of the Discrete Output:
• If the flow rate is below this value, the Discrete Output is ON.
• If the flow rate is above this value, the Discrete Output is OFF.
4. Set Hysteresis to the percentage of variation above and below the setpoint that will operate as a
deadband.
Hysteresis defines a range around the setpoint within which the flow rate switch will not change.
• Default: 5%
• Range: 0.1% to 10%
Example
If Flow Switch Setpoint = 100 g/sec and Hysteresis = 5%, and the first measured flow rate is above
100 g/sec, the discrete output is OFF. It will remain OFF unless the flow rate drops below 95 g/sec. If
this happens, the discrete output will turn ON, and remain ON until the flow rate rises above 105 g/sec.
At this point it turns OFF and will remain OFF until the flow rate drops below 95 g/sec.
Related information
Configure a Discrete Output
Configuration and Use Manual99
Configure advanced options for process measurementConfiguration and Use Manual
October 2021MMI-20025166
6.4 Configure events
An event occurs when the real-time value of a user-specified process variable moves past a user-defined
setpoint. Events are used to provide notification of process changes or to perform specific transmitter actions
if a process change occurs.
Your transmitter supports two event models:
• Basic event model
• Enhanced event model
Tip
Use enhanced events rather than basic events. Enhanced events can perform all the functions of basic events,
and they add the following:
• Defining events by range (In Range or Out of Range), in addition to High and Low
• Triggering transmitter actions if an event occurs
A basic event is used to provide notification of process changes. A basic event occurs (is ON) if the real-time
value of a user-specified process variable moves above (HI) or below (LO) a user-defined setpoint. Event status
can be queried via digital communications, and a Discrete Output can be configured to report event status.
You can define up to two basic events.
Procedure
Select the event that you want to configure.
1.
2. Assign a process variable to the event.
3. Specify Event Type.
OptionDescription
HIx > A
The event occurs when the value of the assigned process variable (x) is greater
than the setpoint (Setpoint A), endpoint not included.
LOx < A
The event occurs when the value of the assigned process variable (x) is less than
the setpoint (Setpoint A), endpoint not included.
4. Set a value for Setpoint A.
5. Optional: Configure a Discrete Output to switch states in response to the event status.
100Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.