Micro Motion™ 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual
October 2021
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.
Safety and approval information
This Micro Motion product complies with all applicable European directives when properly installed in accordance with the instructions in this manual. Refer to the EU declaration of conformity for directives that apply to this product. The following are available: the EU declaration of conformity, with all applicable European directives, and the complete ATEX Installation Drawings and Instructions. In addition the IECEx Installation Instructions for installations outside of the European Union and the CSA Installation Instructions for installations in North America are available on the internet at www.emerson.com or through your local Micro Motion support center.
Information affixed to equipment that complies with the Pressure Equipment Directive, can be found on the internet at
www.emerson.com. For hazardous installations in Europe, refer to standard EN 60079-14 if national standards do not apply.
Other information
Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.
Return policy
Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.
Emerson Flow customer service
Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. and Ireland 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 70 413
6666
Mexico +52 55 5809 5010 France +33 (0) 800 917
901
Argentina +54 11 4809 2700 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy +39 8008 77334 China +86 21 2892 9000
Chile +56 2 2928 4800 Central & Eastern +41 (0) 41 7686
111
Peru +51 15190130 Russia/CIS +7 495 995 9559 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
New Zealand 099 128 804
India 800 440 1468
Japan +81 3 5769 6803
2
Configuration and Use Manual Contents
MMI-20025166 October 2021

Contents

Chapter 1 Before you begin........................................................................................................9
1.1 About this manual....................................................................................................................... 9
1.2 Hazard messages.........................................................................................................................9
1.3 Related documents....................................................................................................................10
1.4 Communications tools and protocols........................................................................................ 10
Chapter 2 Quick start............................................................................................................... 11
2.1 Power up the transmitter...........................................................................................................11
2.2 Check meter status....................................................................................................................11
2.3 Commissioning wizards............................................................................................................. 12
2.4 Make a startup connection to the transmitter............................................................................12
2.5 Set the transmitter clock............................................................................................................13
2.6 Set the transmitter addresses and tags...................................................................................... 13
2.7 View the licensed features......................................................................................................... 14
2.8 Set informational parameters.................................................................................................... 14
2.9 Characterize the meter (if required)...........................................................................................15
2.10 Verify mass flow measurement................................................................................................18
2.11 Verify the zero......................................................................................................................... 18
Chapter 3 Introduction to configuration and commissioning....................................................21
3.1 Security and write protection.....................................................................................................21
3.2 Work with configuration files.....................................................................................................26
Chapter 4 Configure process measurement..............................................................................33
4.1 Configure Sensor Flow Direction Arrow .....................................................................................33
4.2 Configure mass flow measurement........................................................................................... 34
4.3 Configure volume flow measurement for liquid applications..................................................... 39
4.4 Configure Gas Standard Volume (GSV) flow measurement........................................................43
4.5 Configure density measurement................................................................................................48
4.6 Configure temperature measurement....................................................................................... 51
4.7 Configure Pressure Measurement Unit ......................................................................................52
4.8 Configure Velocity Measurement Unit ...................................................................................... 54
Chapter 5 Configure process measurement applications.......................................................... 55
5.1 Set up the API Referral application ............................................................................................ 55
5.2 Set up concentration measurement...........................................................................................72
5.3 Configure the batching application............................................................................................92
Chapter 6 Configure advanced options for process measurement............................................ 97
6.1 Configure Response Time ......................................................................................................... 97
6.2 Detect and report two-phase flow............................................................................................. 97
Configuration and Use Manual 3
Contents Configuration and Use Manual
October 2021 MMI-20025166
6.3 Configure Flow Rate Switch .......................................................................................................99
6.4 Configure events..................................................................................................................... 100
6.5 Configure totalizers and inventories........................................................................................ 102
6.6 Configure logging for totalizers and inventories...................................................................... 106
6.7 Configure Process Variable Fault Action .................................................................................. 106
Chapter 7 Configure device options and preferences.............................................................. 111
7.1 Configure the transmitter display............................................................................................ 111
7.2 Configure the transmitter response to alerts............................................................................117
7.3 Control button sensitivity on the display..................................................................................126
Chapter 8 Integrate the meter with the control system.......................................................... 127
8.1 Configure the transmitter channels......................................................................................... 127
8.2 Configure an mA Output..........................................................................................................129
8.3 Configure an mA Input.............................................................................................................138
8.4 Configure a Frequency Output.................................................................................................141
8.5 Configure a Frequency Input....................................................................................................147
8.6 Configure a Discrete Output.................................................................................................... 150
8.7 Configure a Discrete Input....................................................................................................... 154
Chapter 9 Configure digital communications......................................................................... 157
9.1 Configure HART communications ........................................................................................... 157
9.2 Configure Modbus communications........................................................................................163
Chapter 10 Configure, wire, and use a printer for tickets.......................................................... 167
10.1 Ticket types........................................................................................................................... 167
10.2 Wire the printer..................................................................................................................... 168
10.3 Configure the printer............................................................................................................. 169
10.4 Configure a Discrete Input or discrete event.......................................................................... 171
10.5 Print a standard ticket............................................................................................................172
10.6 Print a batch ticket.................................................................................................................172
10.7 Print a transfer ticket............................................................................................................. 173
Chapter 11 Complete the configuration................................................................................... 177
11.1 Test or tune the system using sensor simulation....................................................................177
11.2 Enable or disable software write-protection...........................................................................178
Chapter 12 Transmitter operation............................................................................................ 181
12.1 View process and diagnostic variables................................................................................... 181
12.2 View and acknowledge status alerts...................................................................................... 182
12.3 Read totalizer and inventory values........................................................................................183
12.4 Start, stop, and reset totalizers and inventories..................................................................... 184
Chapter 13 Operation using the batcher...................................................................................187
13.1 Run a batch............................................................................................................................187
13.2 Perform AOC calibration........................................................................................................ 190
4 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Contents
MMI-20025166 October 2021
Chapter 14 Measurement support............................................................................................193
14.1 Use Smart Meter Verification................................................................................................. 193
14.2 Advanced Phase Measurement software................................................................................203
14.3 Piecewise linearization (PWL) for calibrating gas meters........................................................ 204
14.4 Zero the meter...................................................................................................................... 204
14.5 Set up pressure compensation...............................................................................................207
14.6 Validate the meter.................................................................................................................213
14.7 Perform a (standard) D1 and D2 density calibration...............................................................215
14.8 Adjust concentration measurement with Trim Offset ............................................................218
14.9 Adjust concentration measurement with Trim Slope and Trim Offset ................................... 219
14.10 Energy flow..........................................................................................................................221
Chapter 15 Maintenance.......................................................................................................... 223
15.1 Install a new transmitter license.............................................................................................223
15.2 Upgrade the transmitter firmware......................................................................................... 224
15.3 Reboot the transmitter.......................................................................................................... 225
15.4 Battery replacement..............................................................................................................226
Chapter 16 Log files, history files, and service files....................................................................227
16.1 Generate history files.............................................................................................................227
16.2 Generate service files.............................................................................................................233
Chapter 17 Troubleshooting.................................................................................................... 239
17.1 Status LED and device status..................................................................................................239
17.2 API Referral troubleshooting..................................................................................................239
17.3 Batch troubleshooting...........................................................................................................240
17.4 Concentration measurement troubleshooting...................................................................... 241
17.5 Alert when connecting a core processor to a remote 5700 transmitter..................................242
17.6 Density measurement troubleshooting................................................................................. 243
17.7 Discrete Input troubleshooting..............................................................................................245
17.8 Discrete Output troubleshooting...........................................................................................246
17.9 Flow measurement troubleshooting......................................................................................247
17.10 Frequency Input troubleshooting........................................................................................ 250
17.11 Frequency Output troubleshooting..................................................................................... 251
17.12 mA Input troubleshooting................................................................................................... 252
17.13 mA Output troubleshooting................................................................................................ 253
17.14 Status alerts, causes, and recommendations....................................................................... 256
17.15 Perform a core processor resistance test..............................................................................279
17.16 Check batch total against scale reading............................................................................... 281
17.17 Check the cutoffs.................................................................................................................281
17.18 Check the direction parameters...........................................................................................282
17.19 Check the drive gain............................................................................................................ 282
17.20 Check for internal electrical problems..................................................................................283
Configuration and Use Manual 5
Contents Configuration and Use Manual
October 2021 MMI-20025166
17.21 Check Frequency Output Fault Action .................................................................................283
17.22 Check the scaling of the Frequency Output..........................................................................284
17.23 Check Frequency Output Mode ...........................................................................................284
17.24 Check HART burst mode......................................................................................................284
17.25 Check HART communications..............................................................................................284
17.26 Locate a device using the HART 7 Squawk feature................................................................286
17.27 Check grounding................................................................................................................. 287
17.28 Perform loop tests............................................................................................................... 287
17.29 Check Lower Range Value and Upper Range Value ..............................................................293
17.30 Check mA Output Fault Action ............................................................................................294
17.31 Trim mA Output.................................................................................................................. 294
17.32 Check the pickoff voltage.................................................................................................... 295
17.33 Check power supply wiring.................................................................................................. 296
17.34 Check for radio frequency interference (RFI)........................................................................ 297
17.35 Check sensor-to-transmitter wiring..................................................................................... 297
17.36 Check the sensor coils..........................................................................................................298
17.37 Using sensor simulation for troubleshooting....................................................................... 299
17.38 Check the printing............................................................................................................... 300
17.39 Check for two-phase flow (slug flow)................................................................................... 300
17.40 Temperature measurement troubleshooting...................................................................... 301
17.41 Velocity measurement troubleshooting.............................................................................. 302
17.42 Wi-Fi troubleshooting..........................................................................................................303
Appendix A Using the transmitter display................................................................................. 305
A.1 Components of the transmitter display................................................................................... 305
A.2 Access and use the display menus........................................................................................... 307
A.3 Display menu tree................................................................................................................... 309
Appendix B Using ProLink III with the transmitter..................................................................... 311
B.1 Basic information about ProLink III .......................................................................................... 311
B.2 Connect with ProLink III ...........................................................................................................312
Appendix C Configure Wi-Fi......................................................................................................329
C.1 Enable Wi-Fi.............................................................................................................................329
C.2 Wi-Fi status LED.......................................................................................................................330
C.3 Connect to the 5700 Wi-Fi display........................................................................................... 330
C.4 Change Wi-Fi settings..............................................................................................................331
C.5 Cybersecurity.......................................................................................................................... 333
C.6 Add a hidden network for Android and iOS tablets...................................................................333
C.7 View Spectrum approvals........................................................................................................ 334
Appendix D Using a field communicator with the transmitter................................................... 335
D.1 Basic information about field communicators......................................................................... 335
D.2 Connect with a field communicator ........................................................................................336
6 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Contents
MMI-20025166 October 2021
Appendix E Channel combinations............................................................................................343
E.1 Rules for channel combinations............................................................................................... 343
E.2 Valid combinations for channel configuration..........................................................................343
Appendix F Concentration measurement matrices................................................................... 347
F.1 Standard matrices for the concentration measurement application.........................................347
F.2 Derived variables and calculated process variables...................................................................348
Appendix G Environmental compliance.....................................................................................351
G.1 RoHS and WEEE.......................................................................................................................351
Appendix H Example tickets......................................................................................................353
H.1 Print examples........................................................................................................................ 353
Appendix I Default values and ranges...................................................................................... 357
Configuration and Use Manual 7
Contents Configuration and Use Manual
October 2021 MMI-20025166
8 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Before you begin
MMI-20025166 October 2021

1 Before you begin

1.1
This manual helps you configure, commission, use, maintain, and troubleshoot Micro Motion 5700 transmitters with configurable inputs and outputs.
Important
This manual assumes that:
The transmitter has been installed correctly and completely according to the instructions in the
Users understand basic transmitter and sensor installation, configuration, and maintenance concepts and

About this manual

transmitter installation manual
procedures

1.2 Hazard messages

This document uses the following criteria for hazard messages based on ANSI standards Z535.6-2011 (R2017).
DANGER
Serious injury or death
WARNING
Serious injury or death
will occur if a hazardous situation is not avoided.
could occur if a hazardous situation is not avoided.
CAUTION
Minor or moderate injury will or could occur if a hazardous situation is not avoided.
NOTICE
Data loss, property damage, hardware damage, or software damage can occur if a situation is not avoided. There is no credible risk of physical injury.
Physical access
NOTICE
Unauthorized personnel can potentially cause significant damage and/or misconfiguration of end users' equipment. Protect against all intentional or unintentional unauthorized use.
Physical security is an important part of any security program and fundamental to protecting your system. Restrict physical access to protect users' assets. This is true for all systems used within the facility.
Configuration and Use Manual 9
Before you begin Configuration and Use Manual
October 2021 MMI-20025166

1.3 Related documents

You can find all product documentation on the product documentation DVD shipped with the product or at
www.emerson.com.
See any of the following documents for more information:
Micro Motion 5700 Product Data Sheet
Micro Motion 5700 Transmitters with Configurable Inputs and Outputs: Installation Manual
Coriolis Flow Meter with Micro Motion 5700 Transmitters Safety Manual for Safety Instrumented Systems (SIS)
Micro Motion Oil and Gas Production Supplement
Micro Motion Model 5700 Transmitter Net Oil Calculations Multiwell Supplement
Micro Motion Advanced Phase Measurement Application Manual
Micro Motion 5700 Transmitters with a Marine Bunker Transfer Package: Application Manual
Retrofit a Micro Motion 5700 Configurable Inputs and Outputs Display with a Wi-Fi Display Replacement
Instruction
Replacing the Junction Box for the 4200 Transmitter and the 5700 Transmitter
Replacing the Sensor Cable for the 4200 Transmitter and the 5700 Transmitter
Sensor installation manual
1.4
You can use several different communications tools and protocols to interface with the transmitter, use different tools in different locations, or use different tools for different tasks.
Tool Supported protocols
Display Not applicable
ProLink™ III HART®/Bell 202
Field communicator HART/Bell 202
For information about how to use the communication tools, see the appendices in this manual.

Communications tools and protocols

HART/RS-485
Modbus®/RS-485
Service port
Modbus TCP
Android tablet, iOS tablet, and PC support for Wi-Fi
Tip
You may be able to use other communications tools, such as AMS™ Suite: Intelligent Device Manager, or the Smart Wireless manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at
www.emerson.com.
10 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this
Configuration and Use Manual Quick start
MMI-20025166 October 2021

2 Quick start

2.1
The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
Procedure
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.

Power up the transmitter

1. Follow appropriate procedures to ensure that a new device in the control system does not interfere with existing measurement and control loops.
2. Verify that the cables are connected to the transmitter as described in the installation manual.
3. Verify that all transmitter and sensor covers and seals are closed.
WARNING
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion resulting in injury or death.
Turn on the electrical power at the power supply.
4. The transmitter will automatically perform diagnostic routines. During this period, the Transmitter
Initializing alert is active. The diagnostic routines should complete in approximately 30 seconds.

2.2 Check meter status

Check the meter for any error conditions that require user action or that affect measurement accuracy.
Procedure
1. Wait approximately 10 seconds for the power-up sequence to complete. Immediately after power-up, the transmitter runs through diagnostic routines and checks for error
conditions. During the power-up sequence, the Transmitter Initializing alert is active. This alert should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Table 2-1: Status LED and device status
Status LED condition Device status
Solid green No alerts are active.
Solid yellow One or more alerts are active with Alert Severity = Out of Specification,
Maintenance Required, or Function Check.
Configuration and Use Manual 11
Quick start Configuration and Use Manual
October 2021 MMI-20025166
Table 2-1: Status LED and device status (continued)
Status LED condition Device status
Solid red One or more alerts are active with Alert Severity = Failure.
Flashing yellow (1 Hz) The Function Check in Progress alert is active.
Table 2-2: Wi-Fi status LED and device status
Wi-Fi status LED condition Device status
Solid blue light Wi-Fi is active and not connected to a device.
Blinking blue light Wi-Fi is active and connected to a device.
No light Wi-Fi is not active.

2.3 Commissioning wizards

The transmitter menu includes a Guided Setup to help you move quickly through the most common configuration parameters. ProLink III also provides a commissioning wizard.
By default, when the transmitter starts up, the Guided Setup menu is offered. You can choose to use it or not. You can also choose whether or not Guided Setup is displayed automatically.
To enter Guided Setup upon transmitter startup, choose Yes at the prompt.
To enter Guided Setup after transmitter startup, choose Menu Startup Tasks.
To control the automatic display of Guided Setup, choose Menu Configuration Guided Setup.
For information on the ProLink III commissioning wizard, see the Micro Motion ProLink III with ProcessViz Software User Manual.
As the commissioning wizards are self guided, they are not documented in detail.
2.4
For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter.
Procedure
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix.
Communications tool Connection type to use Instructions
ProLink III HART

Make a startup connection to the transmitter

Using ProLink III with the transmitter
Wi-Fi
Configure Wi-Fi
Field communicator HART Using a field communicator with the
transmitter
12 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Quick start
MMI-20025166 October 2021

2.5 Set the transmitter clock

Display Menu Configuration Time/Date/Tag
ProLink III Device Tools Configuration Transmitter Clock
Field communicator Configure Manual Setup Clock
The transmitter clock provides timestamp data for alerts, service logs, history logs, and all other timers and dates in the system. You can set the clock for your local time or for any standard time you want to use.
Tip
You may find it convenient to set all of your transmitter clocks to the same time, even if the transmitters are in different time zones.
Procedure
Select the time zone that you want to use.
1.
2. If you need a custom time zone, select Special Time Zone and enter your time zone as a difference from UTC (Coordinated Universal Time).
3. Set the time appropriately for the selected time zone.
Tip
The transmitter does not adjust for Daylight Savings Time. If you observe Daylight Savings Time, you must reset the transmitter clock manually.
4. Set the month, day, and year.
The transmitter tracks the year and automatically adds a day for leap years.
2.6
Display Menu Configuration Time/Date/Tag
ProLink III Device Tools Configuration Communications Communications (HART)
Field Communicator Configure Manual Setup HART Communications
The transmitter can have both a HART address and a Modbus address. These addresses are used by service tools and hosts to communicate with the transmitter. The transmitter can also have a tag. The tag identifies the transmitter and can also be used for HART communications.
Procedure
If you plan to make HART connections to the transmitter, set the HART address.

Set the transmitter addresses and tags

Default: 0
Range: 0–15
Tip
Leave HART Address at the default (0) unless you are in a multidrop environment.
If you plan to make Modbus connections to the transmitter, set the Modbus address.
Configuration and Use Manual 13
Quick start Configuration and Use Manual
October 2021 MMI-20025166
Default: 1
Range: 1–15, 23–47, 64–79, 96–110
Tip
If you need an address that is out of range, you can disable Modbus ASCII Support. When Modbus
ASCII Support is disabled, the Modbus address can be set to 1–127, excluding 111. 111 is reserved for
the service port address. However, you will not be allowed to use Modbus ASCII (7-bit) for connections to the transmitter. You must use Modbus RTU (8-bit) instead.
The other Modbus parameters can be left at the default values unless you have trouble making the
connection.
Set the transmitter tag and/or the long tag.
The transmitter will respond to connection requests that use either the transmitter tag or the long tag. The long tag is supported only by HART 7. The transmitter accepts both HART 5 and HART 7 connection requests.
2.7
Display Menu About Licenses Licensed Features
ProLink III Device Tools Device Information Licensed Features
Field communicator Overview Device Information Licenses
The transmitter license controls the features that are enabled on the transmitter, including both software applications and I/O channels. You can view the licensed features to ensure that the transmitter was ordered with the required features.
Licensed features are purchased and available for permanent use. The options model code represents the licensed features.
A trial license allows you to explore features before purchasing. The trial license enables the specified features for a limited number of days. This number is displayed for reference. At the end of this period, the feature will no longer be available.
To purchase additional features or request a trial license, document the Unique ID Number and current license key from your transmitter and contact customer service. To enable the additional features or trial license, you will need to install the new license on the transmitter.
2.8
Display Menu Configuration Device Information
ProLink III Device Tools Configuration Informational Parameters
Field communicator Configure Manual Setup Device

View the licensed features

Set informational parameters

14 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Quick start
MMI-20025166 October 2021
You can set several parameters that identify or describe the transmitter and sensor. These parameters are not used in processing and are not required.
Procedure
1.
Set informational parameters for the transmitter.
a) Set Transmitter Serial Number to the serial number of your transmitter.
The transmitter serial number is provided on the metal tag that is attached to the transmitter housing.
b) Set Descriptor to any desired description of this transmitter or measurement point.
c) Set Message to any desired message.
d) Verify that Model Code (Base) is set to the base model code of the transmitter.
The base model code completely describes your transmitter, except for the features that can be licensed independently. The base model code is set at the factory.
e) Set Model Code (Options) to the options model code of the transmitter.
The options model code describes the independent features that have been licensed for this transmitter. The original options model code is set at the factory. If you license additional options for this transmitter, Emerson will supply an updated options model code.
For a field communicator, configuring model code options is not available for this release.
2. Set informational parameters for the sensor.
a) Set Sensor Serial Number to the serial number of the sensor connected to this transmitter.
The sensor serial number is provided on the metal tag that is attached to the sensor case.
b) Set Sensor Material to the material used for the sensor.
c) Set Sensor Liner to the material used for the sensor liner, if any.
d) Set Flange Type to the type of flange that was used to install the sensor.
Do not set Sensor Type. Sensor Type is set or derived during characterization.

2.9 Characterize the meter (if required)

Display Menu Configuration Sensor Parameters
ProLink III Device Tools Calibration Data
Field communicator Configure Manual Setup Characterization
Configuration and Use Manual 15
Quick start Configuration and Use Manual
October 2021 MMI-20025166
Characterizing the meter adjusts your transmitter to match the unique traits of the sensor it is paired with. The characterization parameters (also called calibration parameters) describe the sensor’s sensitivity to flow, density, and temperature. Depending on your sensor type, different parameters are required.
Values for your sensor are provided on the sensor tag or the calibration certificate.
If your transmitter was ordered with a sensor, it was characterized at the factory. However, you should still
verify the characterization parameters.
Perform a characterization whenever you replace a core processor.
Procedure
1.
Optional: Specify Sensor Type.
Straight Tube (T-Series sensors)
Curved Tube (all sensors except T-Series)
Note
Unlike earlier transmitters, the 5700 derives Sensor Type from the user-specified values for FCF and K1 in combination with an internal ID.
2. Set the flow calibration factor: FCF (also called Flow Cal or Flow Calibration Factor). Be sure to include
all decimal points.
Set the density characterization parameters: D1, D2, TC, K1, K2, and FD. (TC is sometimes shown as
3.
DT.)
4. Apply the changes as required by the tool you are using.
The transmitter identifies your sensor type, and characterization parameters are adjusted as required:
If Sensor Type changed from Curved Tube to Straight Tube, five characterization parameters are
added to the list.
If Sensor Type changed from Straight Tube to Curved Tube, five characterization parameters are
removed from the list.
If Sensor Type did not change, the list of characterization parameters does not change.
5. T-Series sensors only: Set the additional characterization parameters listed below.
Characterization parameter type Parameters
Flow FTG, FFQ
Density DTG, DFQ1, DFQ2
16 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Quick start
MMI-20025166 October 2021

2.9.1 Sample sensor tags

Figure 2-1: Tag on newer curved-tube sensors (all sensors except T-Series)

2.9.2 Flow calibration parameters (FCF, FT)

Two separate values are used to describe flow calibration: a 6-character FCF value and a 4-character FT value. They are provided on the sensor tag.
Both values contain decimal points. During characterization, these are entered as a single 10-character string. The 10-character string is called either Flowcal or FCF.
If your sensor tag shows the FCF and the FT values separately and you need to enter a single value, concatenate the two values to form the single parameter value, retaining both decimal points.
Concatenating FCF and FT
FCF = x.xxxx FT = y.yy Flow calibration parameter: x.xxxxy.yy
2.9.3
Density calibration parameters are typically on the sensor tag and the calibration certificate.
If your sensor tag does not show a D1 or D2 value:
For D1, enter the Dens A or D1 value from the calibration certificate. This value is the line-condition
For D2, enter the Dens B or D2 value from the calibration certificate. This value is the line-condition density

Density calibration parameters (D1, D2, K1, K2, FD, DT, TC)

density of the low-density calibration fluid. Micro Motion uses air. If you cannot find a Dens A or D1 value, enter 0.001 g/cm3.
of the high-density calibration fluid. Micro Motion uses water. If you cannot find a Dens B or D2 value, enter 0.998 g/cm3 .
If your sensor tag does not show a K1 or K2 value:
For K1, enter the first five digits of the density calibration factor. In this sample tag, this value is shown as
12500.
For K2, enter the second five digits of the density calibration factor. In this sample tag, this value is shown
as 14286.
Configuration and Use Manual 17
Quick start Configuration and Use Manual
October 2021 MMI-20025166
Figure 2-2: K1, K2, and TC values in the density calibration factor
If your sensor does not show an FD value, contact customer service.
If your sensor tag does not show a DT or TC value, enter the last four characters of the density calibration factor. In the sample tag shown above, the value is shown as 4.44.
Do not confuse the Meter Factor line on the pictured sensor tag with any meter factor settings discussed in this manual.
2.10
Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method.
Procedure
Read the value for Mass Flow Rate on the transmitter display.
Menu Operations Process Variable Values
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the Process Variables
panel.
Connect to the transmitter with a field communicator and read the value for Mass Flow Rate.
Online Overview Mass Flow Rate
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.

Verify mass flow measurement

2.11 Verify the zero

Display Menu Service Tools Verification & Calibration Meter Zero Zero Verification
ProLink III Device Tools Calibration Smart Zero Verification and Calibration Verify Zero
Field communicator Service Tools Maintenance Calibration Zero Calibration Perform Zero Verify
18 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Quick start
MMI-20025166 October 2021
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Do not verify the zero or zero the meter if a high-severity alert is active. Correct the problem, then verify the zero or zero the meter. You may verify the zero or zero the meter if a low-severity alert is active.
Procedure
1.
Prepare the meter:
a) Allow the meter to warm up for at least 20 minutes after applying power.
b) Run the process fluid through the sensor until the sensor temperature reaches the normal
process operating temperature.
c) Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if
available.
d) Verify that the sensor is blocked in, that flow has stopped, and that the sensor is completely full
of process fluid.
2. Start the zero verification procedure, and wait until it completes.
3. If the zero verification procedure fails:
a) Confirm that the sensor is completely blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
b) Verify that the process fluid is not flashing or condensing, and that it does not contain particles
that can settle out.
c) Repeat the zero verification procedure.
d) If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
Configuration and Use Manual 19
Quick start Configuration and Use Manual
October 2021 MMI-20025166
20 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20025166 October 2021

3 Introduction to configuration and commissioning

3.1
The transmitter has several features that can help to protect it against intentional or unintentional access and configuration changes.
When locked, the mechanical lock switch on the front of the display prevents any configuration changes
When enabled, the software setting Write Protection prevents any configuration changes. The setting
If the Universal Service Port (USP) is disabled, the port cannot be used by any service tool to communicate
When used, the HART Lock prevents any changes by any other HART master.
When enabled, Security prevents any configuration changes being made from the display unless the

Security and write protection

to the transmitter from any local or remote configuration tool. A transmitter without a display does not have a lock switch.
can only be enabled if the transmitter does not have a display.
with or make changes to the transmitter.
appropriate password is entered.

3.1.1 Universal Service Port security

This transmitter is equipped with a Universal Service Port that works with USB type A connections, including compatible flash drives. There are multiple levels of security built into the transmitter service port that you can configure according to your needs and security standards.
The service port offers the following features that enhance interface security:
The service port is inaccessible without physical access to the transmitter and requires removal of the
terminal cover.
The service port can be disabled from the transmitter through software.
The transmitter has a non-traditional operating system that is not designed to execute programs or run
scripts.
The display can be password protected to limit access to the USB file menu.
Overall transmitter security switches such as the lock switch or write-protection disallows configuration
changes from all interfaces including the Universal Service Port.
This transmitter:
Was designed to be implemented in an industrial automation control system (Level 1 and Level 2 of the
Purdue Reference Architecture Model), with defense in depth security controls.
Is not intended to be directly connected to an enterprise or to an internet-facing network without a
compensating control in place.
Configuration and Use Manual 21
Introduction to configuration and commissioning Configuration and Use Manual
October 2021 MMI-20025166

3.1.2 Lock or unlock the transmitter

If the transmitter has a display, a mechanical switch on the display can be used to lock or unlock the transmitter. When locked, no configuration changes can be made using any configuration tool.
Figure 3-1: Lock switch on transmitter display
You can determine whether you need to lock or unlock the transmitter by looking at the switch.
If the switch is in the right position, the transmitter is locked.
If the switch is in the left position, the transmitter is unlocked.
Note
The top switch is reserved for future use.
Procedure
1. WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is powered up. Failure to follow these instructions can cause an explosion resulting in injury or death.
If you are in a hazardous area, power down the transmitter.
Remove the transmitter housing cover.
2.
Figure 3-2: Removing the transmitter housing cover
22 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20025166 October 2021
3. Using a fine-pointed tool, move the switch to the desired position.
4.
Replace the transmitter housing cover.
5. If necessary, power up the transmitter.

3.1.3 Enable or disable the service port

Display Menu Configuration Security Service Port
ProLink III Not available
Field communicator Configure Manual Setup Security Enable/Disable Service Port
The service port is enabled by default, so you can use it for transferring files or connect to it with ProLink III. If you want to completely prevent it from being used, you can disable it.
Note
Enabling or disabling the service port will not take effect until power has been cycled to the transmitter.
WARNING
Do not use the service port if the transmitter is in a hazardous area because using the service port means that you must open the transmitter wiring compartment. Opening the wiring compartment in a hazardous area while the transmitter is powered up can cause an explosion resulting in injury or death.

3.1.4 Set the HART lock

If you plan to use a HART connection to configure the device, you can lock out all other HART masters. If you do this, other HART masters will be able to read data from the device but will not be able to write data to the device.
Restriction
This feature is available only when you are using a field communicator or AMS.
This feature is available only with a HART 7 host.
Procedure
Choose Configure Manual Setup Security Lock/Unlock Device.
1.
2. If you are locking the meter, set Lock Option as desired.
Option Description
Permanent Only the current HART master can make changes to the device. The device will remain
locked until manually unlocked by a HART master. The HART master can also change Lock Option to Temporary.
Temporary Only the current HART master can make changes to the device. The device will remain
locked until manually unlocked by a HART master, or a power-cycle or device reset is performed. The HART master can also change Lock Option to Permanent.
Configuration and Use Manual 23
Introduction to configuration and commissioning Configuration and Use Manual
October 2021 MMI-20025166
Option Description
Lock All No HART masters are allowed to make changes to the configuration. Before changing
Lock Option to Permanent or Temporary, the device must be unlocked. Any HART master can be used to unlock the device.
Postrequisites
To avoid future confusion or difficulties, ensure that the device is unlocked after you have completed your tasks.
3.1.5
Display Use the mechanical switch on the display.
ProLink III Device Tools Configuration Write-Protection
Field communicator Configure Manual Setup Security Lock/Unlock Device
When enabled, Write-Protection prevents changes to the transmitter configuration. You can perform all other functions, and you can view the transmitter configuration parameters.
Note
The write protection setting via software methods (such as ProLink III) is available only on transmitters without a display. For transmitters with a display, write protection is available only using the lock switch on the display. See Lock
or unlock the transmitter.

Enable or disable software write-protection

Write-protecting the transmitter primarily prevents accidental changes to configuration, not intentional changes. Any user who can make changes to the configuration can disable write protection.
3.1.6
Display Menu Configuration Security Display Security
ProLink III Device Tools Configuration Transmitter Display Display Security
Field communicator Configure Manual Setup Display Display Menus
24 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs

Configure security for the display

Configuration and Use Manual Introduction to configuration and commissioning
MMI-20025166 October 2021
When using the display, you can require users to enter a password to do any of the following tasks:
Enter the main menu
Change a parameter
Access alert data through the display
Start, stop, or reset totalizers or inventories via the context menu
The display password can be the same or different from the totalizer/inventory context menu control password. If different, the display password is used to reset, start, and stop totalizers or inventories using Menu Operations Totalizers.
Procedure
1.
Configure Password Required as desired.
Option Description
At Write When an user chooses an action that leads to a configuration change, they are
prompted to enter the display password.
Enter Menu When the menu is selected from the process variable screen, the display password
will be immediately required if Password Required is set.
Never (default) When a user chooses an action that leads to a configuration change, they are
prompted to activate ⇦⇧⇩⇨. This is designed to protect against accidental changes to configuration. It is not a security measure.
2. If the At Write or Enter Menu option was selected, enable or disable alert security as desired.
Option Description
Enabled If an alert is active, the alert symbol is shown in the upper right corner of the display but
the alert banner is not displayed. If the operator attempts to enter the alert menu, they are prompted to enter the display password.
Disabled If an alert is active, the alert symbol is shown in the upper right corner of the display and
the alert banner is displayed automatically. No password or confirmation is required to enter the alert menu.
Restriction
You cannot set Password Required to Never and enable alert security.
If you did not enable Password Required, alert security is disabled and cannot be enabled.
Alert security is disabled automatically if you set Password Required to Never after:Password Required is initially set to either At Write or Enter Menu
Alert security is enabled
Configuration and Use Manual 25
Introduction to configuration and commissioning Configuration and Use Manual
October 2021 MMI-20025166
3. If Password Required has been set to At Write or Enter Menu, you will be prompted to enter the desired password.
Default: AAAA
Range: Any four alphanumeric characters
Password Required must be set to At Write or Enter Menu to enable the totalizer/inventory control
context menu password option.
Important
If you enable Password Required but you do not change the display password, the transmitter will post a configuration alert.
4. Configure Main Menu Available as desired.
Option Description
Enabled The local display Menu option from the process variable screen will be accessible.
Disabled The local display Menu option from the process variable screen will not be accessible.
Important
Once Main Menu Available has been disabled, you cannot enable it from the local display. Use another configuration tool, such as ProLink III, to re-enable main menu access from the local display.

3.2 Work with configuration files

You can save the current transmitter configuration in two forms: a backup file and a replication file. You can save the configuration to the SD card on your transmitter or to a USB drive.
Tip
You can use a saved configuration file to change the nature of the transmitter quickly. This might be convenient if the transmitter is used for different applications or different process fluids.
You can load a configuration file to the transmitter working memory or to the transmitter SD card. You can load either a backup file or a replication file.
Backup files
Replication files

3.2.1 Save a configuration file using the display

Contain all parameters. They are used to restore the current device if required. The .spare extension is used to identify backup files.
Contain all parameters except the device-specific parameters, e.g., calibration factors or meter factors. They are used to replicate the transmitter configuration to other devices. The .xfer extension is used to identify replication files.
Prerequisites
If you are planning to use the USB drive, the service port must be enabled. It is enabled by default. However, if you need to enable it, choose Menu Configuration Security and set Service Port to On.
26 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20025166 October 2021
Procedure
To save the current configuration to the transmitter SD card as a backup file:
a)
Choose Menu Configuration Save/Restore Config Save Config to Memory.
b) Enter the name for this configuration file.
The configuration file is saved to the transmitter SD card as yourname.spare.
To save the current configuration to a USB drive, as either a backup file or a replication file:
a)
b)
c) Choose Backup or Replicate.
d) Enter the name for this configuration file.
The configuration file is saved to the USB drive as yourname.spare or yourname.xfer.
To copy a configuration file from the transmitter SD card to the USB drive:
a)
b)
c) Choose Backup or Replicate.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is powered up. Failure to follow these instructions can cause an explosion resulting in injury or death.
Open the wiring compartment on the transmitter and insert a USB drive into the service port.
Choose Menu USB Options Transmitter --> USB Drive Save Active Config to USB Drive.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is powered up. Failure to follow these instructions can cause an explosion resulting in injury or death.
Open the wiring compartment on the transmitter and insert a USB drive into the service port.
Choose Menu USB Options Transmitter --> USB Drive Transfer Config File to USB Drive.
d) Select the file that you want to transfer.
The configuration file is copied to the USB drive, using its existing name.

3.2.2 Save a configuration file using ProLink III

Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually or by groups. Therefore, you can use this format for both backup and replication.
Procedure
To save the current configuration to the transmitter SD card:
a)
Choose Device Tools Configuration Transfer Save Configuration.
b) Select On my 5700 Device Internal Memory and select Next.
c) Select Save.
Configuration and Use Manual 27
Introduction to configuration and commissioning Configuration and Use Manual
October 2021 MMI-20025166
d) Enter the name for this configuration file.
e)
Set the file type.
To save a backup file, set the file type to Backup.
To save a replication file, set the file type to Transfer.
f) Select Save.
The configuration file is saved to the transmitter SD card as yourname.spare or yourname.xfer.
To save the current configuration to your PC, in 5700 format:
a) Choose Device Tools Configuration Transfer Save Configuration.
b) Select On my computer in 5700 device file format and select Next.
c) Select Save.
d) Browse to the desired location, then enter the name for this configuration file.
e) Set the file type.
To save a backup file, set the file type to Backup.
To save a replication file, set the file type to Transfer.
f) Select Save.
The configuration file is saved to the specified location as yourname.spare or yourname.xfer.
To save the current configuration to your PC, in ProLink III format:
a) Choose Device Tools Configuration Transfer Save Configuration.
b) Select On my computer in ProLink III file format and click Next.
c) Select Save.
d) Select the configuration parameters to be included in this file.
To save a backup file, select all parameters.
To save a replication file, select all parameters except device-specific parameters.
e) Select Save.
f) Browse to the desired location, then enter the name for this configuration file.
g) Set the file type to ProLink configuration file.
h) Select Start Save.
The configuration file is saved to the specified location as yourname.pcfg.

3.2.3 Load a configuration file using the display

Prerequisites
You must have a backup file or a replication file available for use.
28 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20025166 October 2021
If you are planning to use the USB drive, the service port must be enabled. It is enabled by default. However, if you need to enable it, choose Menu Configuration Security and set Service Port to On.
Procedure
To load either a backup file or a replication file from the transmitter SD card:
a)
Choose Menu Configuration Save/Restore Config Restore Config from Memory.
b) Select Backup or Replicate.
c) Select the file that you want to load.
The file is loaded to working memory and becomes active immediately.
To load a either a backup file or a replication file from a USB drive:
a)
b) Choose Menu USB Options USB Drive --> Transmitter Upload Configuration File.
c) Select Backup or Replicate.
d) Select the file that you want to load.
e) Choose Yes or No when prompted to apply the settings.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is powered up. Failure to follow these instructions can cause an explosion resulting in injury or death.
Open the wiring compartment on the transmitter and insert the USB drive containing the backup file or replication file into the service port.
Yes: The file is loaded to working memory and becomes active immediately.
No: The file is loaded to the transmitter SD card but not to working memory. You can load it
from the SD card to working memory at a later time.

3.2.4 Load a configuration file using ProLink III

You can load a configuration file to the transmitter working memory. You can load a backup file or a replication file. Two PC file formats are supported: the 5700 format and the ProLink III format.
Note
When you use ProLink III format for configuration files, you can specify configuration parameters individually or by groups. Therefore, you can use this format for both backup and replication.
Procedure
To load a backup file or replication file from the transmitter SD card:
a)
Choose Device Tools Configuration Transfer Load Configuration.
b) Select On my 5700 Device Internal Memory and select Next.
c) Select Restore.
Configuration and Use Manual 29
Introduction to configuration and commissioning Configuration and Use Manual
October 2021 MMI-20025166
d) Set the file type.
To load a backup file, set the file type to Backup.
To load a replication file, set the file type to Transfer.
e)
Select the file that you want to load and select Load.
The parameters are written to working memory, and the new settings become effectively immediately.
To load a backup file or replication file in 5700 format from the PC:
a) Choose Device Tools Configuration Transfer Load Configuration.
b) Select On my computer in 5700 device file format and select Next.
c) Select Restore.
d) Set the file type.
To load a backup file, set the file type to Backup.
To load a replication file, set the file type to Transfer.
e) Navigate to the file you want to load, and select it.
The parameters are written to working memory, and the new settings become effectively immediately.
To load a file in ProLink III format from the PC:
a) Choose Device Tools Configuration Transfer Load Configuration.
b) Select On my computer in ProLink III file format and select Next.
c) Select the parameters that you want to load.
d) Select Load.
e) Set the file type to Configuration file.
f) Navigate to the file you want to load, and select it.
g) Select Start Load.
The parameters are written to working memory, and the new settings become effectively immediately.

3.2.5 Restore the factory configuration

Display Menu Configuration Save/Restore Configuration Restore Config from Memory
ProLink III Device Tools Configuration Transfer Restore Factory Configuration
Field communicator Service Tools Maintenance Reset/Restore Restore Factory Configuration
A file containing the factory configuration is always saved in the transmitter internal memory, and is available for use.
This action is typically used for error recovery or for repurposing a transmitter.
If you restore the factory configuration, the real-time clock, the audit trail, the historian, and other logs are not reset.
30 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Introduction to configuration and commissioning
MMI-20025166 October 2021

3.2.6 Replicate a transmitter configuration

Replicating a transmitter configuration is a fast method to set up similar or identical measurement points.
Procedure
1.
Configure a transmitter and verify its operation and performance.
2. Use any available method to save a replication file from that transmitter.
3. Use any available method to load the replication file to another transmitter.
4. At the replicated transmitter, set device-specific parameters and perform device-specific procedures: a) Set the clock.
b) Set the tag, long tag, HART address, Modbus address, and related parameters.
c) Characterize the transmitter.
d) Perform zero validation and take any recommended actions.
e) Perform loop tests and take any recommended actions, including mA Output trim.
f) Use sensor simulation to verify transmitter response.
5. At the replicated transmitter, make any other configuration changes.
6. Follow your standard procedures to ensure that the replicated transmitter is performing as desired.
Configuration and Use Manual 31
Introduction to configuration and commissioning Configuration and Use Manual
October 2021 MMI-20025166
32 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021

4 Configure process measurement

4.1
Display Menu Configuration Process Measurement Flow Variables Flow Direction
ProLink III Device Tools Configuration Process Measurement Flow Sensor Direction
Field communicator Configure Manual Setup Measurements Flow Sensor Direction
Sensor Flow Direction Arrow is used to accommodate installations in which the Flow arrow on the sensor does not match the majority of the process flow. This typically happens when the sensor is accidentally installed backwards.
Sensor Flow Direction Arrow interacts with mA Output Direction, Frequency Output Direction, and Totalizer Direction to control how flow is reported by the outputs and accumulated by the totalizers and
inventories.
The Sensor Flow Direction Arrow also affects how flow is reported on the transmitter display and via digital communications. This includes ProLink III, a field communicator, and all other user interfaces.
Figure 4-1: Flow arrow on sensor

Configure Sensor Flow Direction Arrow

A. Flow arrow
Actual flow direction
B.
Procedure
Set Sensor Flow Direction Arrow as appropriate.
Option Description
With Arrow The majority of flow through the sensor matches the Flow arrow on the sensor. Actual
forward flow is processed as forward flow.
Against Arrow The majority of flow through the sensor is opposite to the Flow arrow on the sensor. Actual
forward flow is processed as reverse flow.
Configuration and Use Manual 33
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the setting of Sensor Flow Direction Arrow. Sensor Flow Direction Arrow controls only whether actual flow is processed as forward flow or reverse flow.
Related information
Configure mA Output Direction Configure Frequency Output Direction Configure Discrete Output Source Configure totalizers and inventories Effect of Sensor Flow Direction Arrow on digital communications
4.2
The mass flow measurement parameters control how mass flow is measured and reported. The mass total and mass inventory are derived from the mass flow data.

Configure mass flow measurement

4.2.1 Configure Mass Flow Measurement Unit

Display Menu Configuration Process Measurement Flow Variables Mass Flow Settings Units
ProLink III Device Tools Configuration Process Measurement Flow Mass Flow Rate Unit
Field communicator Configure Manual Setup Measurements Flow Mass Flow Unit
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The default unit used for mass total and mass inventory is derived from this unit.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
Default: g/sec (grams per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Unit description
Display ProLink III Field communicator
Grams per second gram/s g/sec g/s
Grams per minute gram/min g/min g/min
Grams per hour gram/h g/hr g/h
Kilograms per second kg/s kg/sec kg/s
34 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Label
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
Unit description
Display ProLink III Field communicator
Kilograms per minute kg/min kg/min kg/min
Kilograms per hour kg/h kg/hr kg/h
Kilograms per day kg/d kg/day kg/d
Metric tons per minute MetTon/min mTon/min MetTon/min
Metric tons per hour MetTon/h mTon/hr MetTon/h
Metric tons per day MetTon/d mTon/day MetTon/d
Pounds per second lb/s lbs/sec lb/s
Pounds per minute lb/min lbs/min lb/min
Pounds per hour lb/h lbs/hr lb/h
Pounds per day lb/d lbs/day lb/d
Short tons (2000 pounds) per minute STon/min sTon/min STon/min
Short tons (2000 pounds) per hour STon/h sTon/hr STon/h
Short tons (2000 pounds) per day STon/d sTon/day STon/d
Long tons (2240 pounds) per hour LTon/h lTon/hr LTon/h
Long tons (2240 pounds) per day LTon/d lTon/day LTon/d
Label
Special unit SPECIAL Special Special
Define a special measurement unit for mass flow
Display Menu Configuration Process Measurement Flow Variables Mass Flow Settings Units
ProLink III Device Tools Configuration Process Measurement Flow Mass Flow Rate Unit Special
Field communicator Configure Manual Setup Measurements Optional Setup Special Units Mass Special Units
Procedure
Specify Base Mass Unit.
1.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Mass Flow Conversion Factor as follows:
a) x base units = y special units
b) Mass Flow Conversion Factor = x ÷ y
4. Enter Mass Flow Conversion Factor.
The original mass flow rate value is divided by this value.
SPECIAL
Configuration and Use Manual 35
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
5. Set Mass Flow Label to the name you want to use for the mass flow unit.
6.
Set Mass Total Label to the name you want to use for the mass total and mass inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for mass flow
If you want to measure mass flow in ounces per second (oz/sec):
1. Set Base Mass Unit to Pounds (lb).
2. Set Base Time Unit to Seconds (sec).
3. Calculate Mass Flow Conversion Factor: a. 1 lb/sec = 16 oz/sec
b. Mass Flow Conversion Factor = 1 ÷ 16 = 0.0625
4. Set Mass Flow Conversion Factor to 0.0625.
5. Set Mass Flow Label to oz/sec.
6. Set Mass Total Label to oz.

4.2.2 Configure Flow Damping

Display Menu Configuration Process Measurement Flow Variables Flow Damping
ProLink III Device Tools Configuration Process Measurement Flow Flow Rate Damping
Field communicator Configure Manual Setup Measurements Flow Flow Damping
Flow Damping controls the amount of damping that will be applied to the measured mass flow rate. It affects flow rate process variables that are based on the measured mass flow rate. This includes volume flow rate and gas standard volume flow rate.
Flow Damping also affects specialized flow rate variables such as temperature-corrected volume flow rate (API Referral) and net mass flow rate (concentration measurement). It is not applied to the flow rate received via the Frequency Input.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value specifies the time period, in seconds, over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
Default: 0.64 seconds
Range: 0 seconds to 60 seconds
Note
If a number greater than 60 is entered, it is automatically changed to 60.
36 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
Tip
A high damping value makes the process variable appear smoother because the reported value changes
slowly.
A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
The combination of a high damping value and rapid, large changes in flow rate can result in increased
measurement error.
Whenever the damping value is non-zero, the reported measurement will lag the actual measurement
because the reported value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the reported value.
The transmitter automatically rounds off any entered damping value to the nearest valid value. Therefore,
the recommended damping value for gas applications should be 3.2 seconds. If you enter 2.56, the transmitter will round it off to 3.2.
For filling applications, Emerson recommends using the default value of 0.04 seconds.
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also affects volume measurement for gas standard volume data. The transmitter calculates volume data from the damped mass flow data.
Interaction between Flow Damping and mA Output Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the reported mass flow value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping controls the rate of change reported through mA Output. If mA Output Process Variable is set to Mass Flow Rate, and both Flow Damping and mA Output Damping are set to non-zero values, flow damping is applied first, and the added damping calculation is applied to the result of the first calculation.
4.2.3
Display Menu Configuration Process Measurement Flow Variables Mass Flow Settings Low Flow
ProLink III Device Tools Configuration Process Measurement Flow Mass Flow Cutoff
Field communicator Configure Manual Setup Measurements Flow Mass Flow Cutoff
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates below this cutoff will be reported as 0.

Configure Mass Flow Cutoff

Cutoff
Procedure
Set Mass Flow Cutoff to the value you want to use.
Default: A sensor-specific value set at the factory. If your transmitter was ordered without a sensor, the
default may be 0.0.
Configuration and Use Manual 37
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
Recommendation: 0.5% of maximum flow rate of the attached sensor. See the sensor specifications.
Important
Do not use your meter for measurement with Mass Flow Cutoff set to 0.0 g/sec. Ensure that Mass Flow Cutoff is set to the value that is appropriate for your sensor.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data rather than the reported value.
Volume flow has a separate Volume Flow Cutoff that is not affected by the Mass Flow Cutoff value.
Interaction between Mass Flow Cutoff and mA Output Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process Variable is set to Mass Flow Rate, the mass flow rate reported through mA Output is controlled by the higher
of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events defined on mass flow).
mA Output Cutoff affects only mass flow values reported through mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 10 g/sec
Mass Flow Cutoff: 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 15 g/sec
Mass Flow Cutoff: 10 g/sec
Result:
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:The mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
38 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.3
Configure volume flow measurement for liquid
applications
The volume flow measurement parameters control how liquid volume flow is measured and reported. The volume total and volume inventory are derived from volume flow data.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.

4.3.1 Configure Volume Flow Type for liquid applications

Display
ProLink III
Field communicator
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with the following applications:
API Referral
Menu Configuration Process Measurement Flow Variables Volume Flow Settings Flow Type Liquid
Device Tools Configuration Process Measurement Flow Volume Flow Type Liquid Volume
Configure Manual Setup Measurements Optional Setup GSV Volume Flow Type Liquid Volume
Concentration measurement
Advanced Phase Measurement — net oil
Advanced Phase Measurement — liquid with gas
For these applications, set Volume Flow Type to Liquid.
Procedure
Set Volume Flow Type to Liquid.
4.3.2 Configure Volume Flow Measurement Unit for liquid
applications
Display Menu Configuration Process Measurement Flow Variables Volume Flow Settings Units
ProLink III Device Tools Configuration Process Measurement Flow Volume Flow Rate Unit
Field communicator Configure Manual Setup Measurements Flow Volume Flow Unit
Configuration and Use Manual 39
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume flow rate. The unit used for the volume total and volume inventory is based on this unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
Default: l/sec (liters per second)
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement Unit, plus one user-defined measurement unit. Different communications tools may use different labels for the units.
Unit description
Display ProLink III Field communicator
Cubic feet per second ft3/s ft3/sec Cuft/s
Cubic feet per minute ft3/min ft3/min Cuft/min
Cubic feet per hour ft3/h ft3/hr Cuft/h
Cubic feet per day ft3/d ft3/day Cuft/d
Cubic meters per second m3/s m3/sec Cum/s
Cubic meters per minute m3/min m3/min Cum/min
Cubic meters per hour m3/h m3/hr Cum/h
Cubic meters per day m3/d m3/day Cum/d
U.S. gallons per second gal/s US gal/sec gal/s
U.S. gallons per minute gal/m US gal/min gal/min
U.S. gallons per hour gal/h US gal/hr gal/h
U.S. gallons per day gal/d US gal/day gal/d
Million U.S. gallons per day MMgal/d mil US gal/day MMgal/d
Liters per second L/s l/sec L/s
Liters per minute L/min l/min L/in
Label
Liters per hour L/h l/hr L/h
Million liters per day MML/d mil l/day ML/d
Imperial gallons per second Impgal/s Imp gal/sec Impgal/s
Imperial gallons per minute Impgal/m Imp gal/min Impgal/min
Imperial gallons per hour Impgal/h Imp gal/hr Impgal/h
Imperial gallons per day Impgal/d Imp gal/day Impgal/d
40 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
Unit description
Barrels per second
Barrels per minute
Barrels per hour
Barrels per day
Beer barrels per second
Beer barrels per minute
Beer barrels per hour
Beer barrels per day
Special unit SPECIAL Special Special
(1) Unit based on oil barrels (42 U.S. gallons). (2) Unit based on U.S. beer barrels (31 U.S. gallons).
(1)
(1)
(1)
(1)
(2)
(2)
(2)
(2)
Display ProLink III Field communicator
bbl/s barrels/sec bbl/s
bbl/min barrels/min bbl/min
bbl/h barrels/hr bbl/h
bbl/d barrels/day bbl/d
Beer bbl/s Beer barrels/sec Beer bbl/s
Beer bbl/min Beer barrels/min Beer bbl/min
Beer bbl/h Beer barrels/hr Beer bbl/h
Beer bbl/d Beer barrels/day Beer bbl/d
Label
Define a special measurement unit for volume flow
Display Menu Configuration Process Measurement Flow Variables Volume Flow Settings Units
ProLink III Device Tools Configuration Process Measurement Flow Volume Flow Rate Unit Special
Field communicator Configure Manual Setup Measurements Optional Setup Special Units Volume Special Units
SPECIAL
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Procedure
Specify Base Volume Unit.
1.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Volume Flow Conversion Factor = x ÷ y
4. Enter Volume Flow Conversion Factor.
The original volume flow rate value is divided by this conversion factor.
5. Set Volume Flow Label to the name you want to use for the volume flow unit.
6. Set Volume Total Label to the name you want to use for the volume total and volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Configuration and Use Manual 41
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
Example: Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
1.
Set Base Volume Unit to Gallons (gal).
2. Set Base Time Unit to Seconds (sec).
3. Calculate the conversion factor:
a. 1 gal/sec = 8 pints/sec
b. Volume Flow Conversion Factor = 1 ÷ 8 = 0.1250
4. Set Volume Flow Conversion Factor to 0.1250.
5. Set Volume Flow Label to pints/sec.
6. Set Volume Total Label to pints.

4.3.3 Configure Volume Flow Cutoff

Display Menu Configuration Process Measurement Flow Variables Volume Flow Settings Low Flow
ProLink III Device Tools Configuration Process Measurement Flow Volume Flow Cutoff
Field communicator Configure Manual Setup Measurements Flow Volume Flow Cutoff
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured. All volume flow rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
Default: 0.0 l/sec (liters per second)
Range: 0 l/sec to x l/sec, where x is the sensor’s flow calibration factor, multiplied by 0.0002
Cutoff
Interaction between Volume Flow Cutoff and mAO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured. mAO Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process Variable is set to Volume Flow Rate, the volume flow rate reported through mA Output is controlled by the
higher of the two cutoff values.
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the volume flow values used in other transmitter behavior (e.g., events defined on the volume flow).
mAO Cutoff affects only flow values reported through mA Output.
Example: Cutoff interaction with mAO Cutoff lower than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
42 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
AO Cutoff: 10 l/sec
Volume Flow Cutoff: 15 l/sec
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mAO Cutoff higher than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 15 l/sec
Volume Flow Cutoff: 10 l/sec
Result:
If the volume flow rate drops below 15 l/sec but not below 10 l/sec:The mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and 0 will be used in all
internal processing.
4.4
Configure Gas Standard Volume (GSV) flow
measurement
The gas standard volume (GSV) flow measurement parameters control how gas standard volume flow is measured and reported.
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.

4.4.1 Configure Volume Flow Type for gas applications

Display
ProLink III
Field communicator
Menu Configuration Process Measurement Flow Variables Volume Flow Settings Flow Type Gas
Device Tools Configuration Process Measurement Flow Volume Flow Type Gas Standard Volume
Configure Manual Setup Measurements Optional Setup GSV Volume Flow Type Standard Gas Volume
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
Gas standard volume measurement is incompatible with the following applications:
API Referral
Configuration and Use Manual 43
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
Concentration measurement
Advanced Phase Measurement — net oil
Advanced Phase Measurement — liquid with gas
For these applications, set Volume Flow Type to Liquid.
Procedure
Set Volume Flow Type to Gas.
4.4.2
Display Menu Configuration Process Measurement Flow Variables Volume Flow Settings Standard
ProLink III Device Tools Configuration Process Measurement Flow Standard Density of Gas
Field communicator Configure Manual Setup Measurements Optional Setup GSV Gas Ref Density
Standard Gas Density is the density of your gas at reference temperature and reference pressure. This is often called standard density or base density. It is used to calculate the GSV flow rate from the mass flow rate.
Procedure
Set Standard Gas Density to the density of your gas at reference temperature and reference pressure.
You can use any reference temperature and reference pressure that you choose. It is not necessary to configure these values in the transmitter.
Tip
ProLink III provides a guided method that you can use to calculate the standard density of your gas if you do not know it.

Configure Standard Gas Density

Gas Density

4.4.3 Configure Gas Standard Volume Flow Measurement Unit

Display Menu Configuration Process Measurement Flow Variables Volume Flow Settings Units
ProLink III Device Tools Configuration Process Measurement Flow Gas Standard Volume Flow Unit
Field communicator Configure Manual Setup Measurements Flow GSV Flow Unit
Gas Standard Volume Flow Measurement Unit specifies the unit of measure that will be used for the gas standard volume (GSV) flow rate. The unit used for gas standard volume total and gas standard volume inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Gas Standard Volume.
Procedure
Set Gas Standard Volume Flow Measurement Unit to the unit you want to use.
Default: SCFM (Standard Cubic Feet per Minute)
44 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Gas Standard Volume Flow Measurement Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may use different
labels for the units.
Unit description
Display ProLink III Field communicator
Normal cubic meters per second NCMS Nm3/sec Nm3/sec
Normal cubic meters per minute NCMM Nm3/min Nm3/min
Normal cubic meters per hour NCMH Nm3/hr Nm3/hr
Normal cubic meters per day NCMD Nm3/day Nm3/day
Normal liter per second NLPS NLPS NLPS
Normal liter per minute NLPM NLPM NLPM
Normal liter per hour NLPH NLPH NLPH
Normal liter per day NLPD NLPD NLPD
Standard cubic feet per second SCFS SCFS SCFS
Standard cubic feet per minute SCFM SCFM SCFM
Standard cubic feet per hour SCFH SCFH SCFH
Standard cubic feet per day SCFD SCFD SCFD
Standard cubic meters per second SCMS Sm3/sec Sm3/sec
Standard cubic meters per minute SCMM Sm3/min Sm3/min
Standard cubic meters per hour SCMH Sm3/hr Sm3/hr
Label
Standard cubic meters per day SCMD Sm3/day Sm3/day
Standard liter per second SLPS SLPS SLPS
Standard liter per minute SLPM SLPM SLPM
Standard liter per hour SLPH SLPH SLPH
Standard liter per day SLPD SLPD SLPD
Special measurement unit SPECIAL Special Special
Define a special measurement unit for gas standard volume flow
Display
ProLink III
Field communicator
Configuration and Use Manual 45
Menu Configuration Process Measurement Flow Variables Volume Flow Settings Units SPECIAL
Device Tools Configuration Process Measurement Flow Gas Standard Volume Flow Unit Special
Configure Manual Setup Measurements Optional Setup Special Units Special Gas Standard Volume Units
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Procedure
1.
Specify Base Gas Standard Volume Unit.
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Gas Standard Volume Flow Conversion Factor as follows:
a) x base units = y special units
b) Gas Standard Volume Flow Conversion Factor = x ÷ y
4. Enter the Gas Standard Volume Flow Conversion Factor.
The original gas standard volume flow value is divided by this conversion factor.
5. Set Gas Standard Volume Flow Label to the name you want to use for the gas standard volume flow
unit.
6. Set Gas Standard Volume Total Label to the name you want to use for the gas standard volume total
and gas standard volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for gas standard volume flow
You want to measure gas standard volume flow in thousands of standard cubic feet per minute.
1. Set Base Gas Standard Volume Unit to SCFM.
2. Set Base Time Unit to minutes (min).
3. Calculate the conversion factor:
a. One thousands of standard cubic feet per minute = 1000 cubic feet per minute
b. Gas Standard Volume Flow Conversion Factor = 1 ÷ 1000 = 0.001
4. Set Gas Standard Volume Flow Conversion Factor to 0.001.
5. Set Gas Standard Volume Flow Label to KSCFM.
6. Set Gas Standard Volume Total Label to KSCF.

4.4.4 Configure Gas Standard Volume Flow Cutoff

Display
ProLink III
Field communicator
46 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Menu Configuration Process Measurement Flow Variables Volume Flow Settings Low Flow Cutoff
Device Tools Configuration Process Measurement Flow Gas Standard Volume Flow Cutoff
Configure Manual Setup Measurements Optional Setup GSV GSV Cutoff
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
Gas Standard Volume Flow Cutoff specifies the lowest gas standard volume flow rate that will reported as measured. All gas standard volume flow rates below this cutoff will be reported as 0.
Procedure
Set Gas Standard Volume Flow Cutoff to the value you want to use.
Default: 0.0
Range: 0.0 to any positive value
Interaction between Gas Standard Volume Flow Cutoff and mA Output Cutoff
Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported through mA Output. If mA Output Process Variable is set to Gas Standard Volume Flow Rate, the volume flow rate reported through mA Output is controlled by the higher of the two cutoff values.
Gas Standard Volume Flow Cutoff affects both the gas standard volume flow values reported through outputs and the gas standard volume flow values used in other transmitter behavior (for example, events defined on gas standard volume flow).
mA Output Cutoff affects only flow values reported through mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Gas Standard Volume Flow Cutoff
Configuration:
mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
Frequency Output Process Variable: Gas Standard Volume Flow Rate
mA Output Cutoff for the primary mA Output: 10 SLPM (standard liters per minute)
Gas Standard Volume Flow Cutoff: 15 SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Gas Standard Volume Flow Cutoff
Configuration:
mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow Rate
Frequency Output Process Variable: Gas Standard Volume Flow Rate
mA Output Cutoff for the primary mA Output: 15 SLPM (standard liters per minute)
Gas Standard Volume Flow Cutoff: 10 SLPM
Result:
If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:The primary mA Output will report zero flow.
The Frequency Output will report the actual flow rate, and the actual flow rate will be used in all
internal processing.
Configuration and Use Manual 47
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
If the gas standard volume flow rate drops below 10 SLPM, both outputs will report zero flow, and 0 will be
used in all internal processing.
4.5
The density measurement parameters control how density is measured and reported. Density measurement is used with mass flow rate measurement to determine liquid volume flow rate.

Configure density measurement

4.5.1 Configure Density Measurement Unit

Display Menu Configuration Process Measurement Density Units
ProLink III Device Tools Configuration Process Measurement Density Density Unit
Field communicator Configure Manual Setup Measurements Density Density Unit
Density Measurement Unit controls the measurement units that will be used in density calculations and reporting.
Restriction
If the API Referral application is enabled, you cannot change the density measurement unit here. The density measurement unit is controlled by the API table selection.
Procedure
Set Density Measurement Unit to the option you want to use.
Default: g/cm3 (grams per cubic centimeter)
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit. Different communications tools may use different labels.
Unit description
Specific gravity
Grams per cubic centimeter g/cm3 g/cm3 g/Cucm
Grams per liter g/L g/l g/L
Grams per milliliter g/mL g/ml g/mL
Kilograms per liter kg/L kg/l kg/L
Kilograms per cubic meter kg/m3 kg/m3 kg/Cum
Pounds per U.S. gallon lb/gal lbs/USgal lb/gal
Pounds per cubic foot lb/ft3 lbs/ft3 lb/Cuft
Pounds per cubic inch lb/in3 lbs/in3 lb/CuIn
(1)
Display ProLink III Field communicator
SGU SGU SGU
Label
48 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
Unit description
Degrees API
Short ton per cubic yard STon/yd3 sT/yd3 STon/Cuyd
(1) Non-standard calculation. This value represents line density divided by the density of water at 60 °F (15.6 °C). (2) Non standard calculation, unless the API referral application is enabled. Calculated from line density instead of specific
gravity.
4.5.2
Display Menu Configuration Process Measurement Density Damping
ProLink III Device Tools Configuration Process Measurement Density Density Damping
Field communicator Configure Manual Setup Measurements Density Density Damping
(2)

Configure Density Damping

Display ProLink III Field communicator
API API degAPI
Label
Density Damping controls the amount of damping that will be applied to density data.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value specifies the time period, in seconds, over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the change in the actual measured value.
Procedure
Set Density Damping to the desired value.
Default: 1.28 seconds
Range: 0.0 to 60 seconds
Tip
A high damping value makes the process variable appear smoother because the reported value changes
slowly.
A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
The combination of a high damping value and rapid, large changes in density can result in increased
measurement error.
Whenever the damping value is non-zero, the damped value will lag the actual measurement because the
damped value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the damped value.
If a number greater than 60 is entered, it is automatically changed to 60.
Configuration and Use Manual 49
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
Effect of Density Damping on volume measurement
Density Damping affects liquid volume measurement. Liquid volume values are calculated from the damped density value rather than the measured density value. Density Damping does not affect gas standard volume measurement.
Interaction between Density Damping and mA Output Damping
When mA Output is configured to report density, both Density Damping and mA Output Damping are applied to the reported density value.
Density Damping controls the rate of change in the value of the process variable in transmitter memory. mA Output Damping controls the rate of change reported through mA Output.
If mA Output Source is set to Density, and both Density Damping and mA Output Damping are set to non­zero values, density damping is applied first, and the mA Output damping calculation is applied to the result of the first calculation. This value is reported over mA Output.
4.5.3
Display Menu Configuration Process Measurement Density Cutoff
ProLink III Device Tools Configuration Process Measurement Density Density Cutoff
Field communicator Configure Manual Setup Measurements Density Density Cutoff
Density Cutoff specifies the lowest density value that will be reported as measured. All density values below this cutoff will be reported as 0.
Procedure
Set Density Cutoff to the value you want to use.
Default: 0.2 g/cm
Range: 0.0 g/cm3 to 0.5 g/cm

Configure Density Cutoff

3
3
Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density Cutoff, the volume flow rate is reported as 0. Density Cutoff does not affect gas standard volume measurement. Gas standard volume values are always calculated from the value configured for Standard Gas Density or polled value if configured for polled base density.
50 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021

4.6 Configure temperature measurement

The temperature measurement parameters control how temperature data is processed. Temperature data is used in several different ways, including temperature compensation, API Referral, and concentration measurement.
4.6.1
Display Menu Configuration Process Measurement Temperature Units
ProLink III Device Tools Configuration Process Measurement Temperature Temperature Unit
Field communicator Configure Manual Setup Measurements Temperature Unit
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
Default: °C (Celsius)

Configure Temperature Measurement Unit

Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different communications tools may use different labels for the units.
Unit description
Display ProLink III Field communicator
Degrees Celsius °C °C degC
Degrees Fahrenheit °F °F degF
Degrees Rankine °R °R degR
Label
Kelvin °K °K Kelvin

4.6.2 Configure Temperature Damping

Display Menu Configuration Process Measurement Temperature Damping
ProLink III Device Tools Configuration Process Measurement Temperature Temperature Damping
Field communicator Configure Manual Setup Measurements Temperature Damping
Temperature Damping controls the amount of damping that will be applied to temperature data from the sensor. Temperature Damping is not applied to external temperature data.
Damping is used to smooth out small, rapid fluctuations in process measurement. The damping value specifies the time period, in seconds, over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value of the process variable (the damped value) will reflect 63% of the change in the actual measured value.
Procedure
Set Temperature Damping to the desired value.
Configuration and Use Manual 51
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166
Default: 4.8 seconds
Range: 0.0 to 80 seconds
Note
If a number greater than 80 is entered, it is automatically changed to 80.
Tip
A high damping value makes the process variable appear smoother because the reported value changes
slowly.
A low damping value makes the process variable appear more erratic because the reported value changes
more quickly.
The combination of a high damping value and rapid, large changes in temperature can result in increased
measurement error.
Whenever the damping value is non-zero, the damped value will lag the actual measurement because the
damped value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less lag time
between the actual measurement and the damped value.
Effect of Temperature Damping on process measurement
Temperature Damping affects all processes and algorithms that use temperature data from the internal sensor RTD.
Temperature compensation
Temperature compensation adjusts process measurement to compensate for the effect of temperature on the sensor tubes.
API Referral
Temperature Damping affects API Referral process variables only if the transmitter is configured to use temperature data from the sensor. If an external temperature value is used for API Referral, Temperature Damping does not affect API Referral process variables.
Concentration measurement
Temperature Damping affects concentration measurement process variables only if the transmitter is configured to use temperature data from the sensor. If an external temperature value is used for concentration measurement, Temperature Damping does not affect concentration measurement process variables.
4.7
Display
ProLink III
Field communicator

Configure Pressure Measurement Unit

Menu Configuration Process Measurement Pressure Units
Device Tools Configuration Process Measurement Pressure Compensation Pressure Unit
Configure Manual Setup Measurements Optional Setup External Pressure/Temperature Pressure Unit
52 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement
MMI-20025166 October 2021
Pressure Measurement Unit controls the measurement unit used for pressure. This unit must match the unit used by the external pressure device.
Pressure data is used for pressure compensation and for API Referral. The device does not measure pressure directly. You must set up a pressure input.
Procedure
Set Pressure Measurement Unit to the desired unit.
Default: psi
4.7.1

Options for Pressure Measurement Unit

The transmitter provides a standard set of measurement units for Pressure Measurement Unit. Different communications tools may use different labels for the units. In most applications, set Pressure Measurement Unit to match the pressure measurement unit used by the remote device.
Unit description
Display ProLink III Field communicator
Feet water @ 68 °F ftH2O @68°F Ft Water @ 68°F ftHO
Inches water @ 4 °C inH2O @4°C In Water @ 4°C inHO @4DegC
Inches water @ 60 °F inH2O @60°F In Water @ 60°F inHO @60DegF
Inches water @ 68 °F inH2O @68°F In Water @ 68°F inHO
Millimeters water @ 4 °C mmH2O @4°C mm Water @ 4°C mmHO @4DegC
Millimeters water @ 68 °F mmH2O @68°F mm Water @ 68°F mmHO
Millimeters mercury @ 0 °C mmHg @0°C mm Mercury @ 0°C mmHg
Inches mercury @ 0 °C inHg @0°C In Mercury @ 0°C inHg
Pounds per square inch psi PSI psi
Bar bar bar bar
Millibar mbar millibar mbar
Label
Grams per square centimeter g/cm2 g/cm2 g/Sqcm
Kilograms per square centimeter kg/cm2 kg/cm2 kg/Sqcm
Pascals Pa pascals Pa
Kilopascals kPA Kilopascals kPa
Megapascals mPA Megapascals MPa
Torr @ 0 °C torr Torr @ 0°C torr
Atmospheres atm atms atm
Configuration and Use Manual 53
Configure process measurement Configuration and Use Manual
October 2021 MMI-20025166

4.8 Configure Velocity Measurement Unit

Display Menu Configuration Process Measurement Velocity Units
ProLink III Device Tools Configuration Process Measurement Velocity Unit
Field communicator Configure Manual Setup Measurements Approximate Velocity Velocity Unit
Velocity Measurement Unit controls the measurement unit used to report velocity.
Procedure
Set Velocity Measurement Unit to the desired unit.
Default: m/sec
4.8.1
The transmitter provides a standard set of measurement units for Velocity Measurement Unit. Different communications tools may use different labels.
Unit description
Feet per minute ft/min ft/min ft/min
Feet per second ft/s ft/sec ft/s
Inches per minute in/min in/min in/min
Inches per second in/s in/sec in/s
Meters per hour m/h m/hr m/h
Meters per second m/s m/sec m/s

Options for Velocity Measurement Unit

Label
Display ProLink III Field communicator
54 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021

5 Configure process measurement applications

5.1
The API Referral application corrects line density to reference temperature and reference pressure according to American Petroleum Institute (API) standards. The resulting process variable is referred density.
Restriction
The API Referral application is not compatible with the following applications:
Gas Standard Volume Measurement (GSV)
Piecewise linearization (PWL)
Advanced Phase Measurement
Concentration measurement

Set up the API Referral application

5.1.1 Set up the API Referral application using the display

Enable the API Referral application using the display
The API Referral application must be enabled before you can perform any setup. If the API Referral application was enabled at the factory, you do not need to enable it now.
Prerequisites
The API Referral application must be licensed on your transmitter.
Procedure
1.
Choose Menu Configuration Process Measurement.
2. Choose Flow Variables Volume Flow Settings and ensure that Flow Type is set to Liquid.
3. Return to the Process Measurement menu.
4. If the concentration measurement application is displayed in the list, choose Concentration
Measurement and ensure that Enabled/Disabled is set to Disabled.
The concentration measurement application and the API Referral application cannot be enabled simultaneously.
5. Enable API Referral.
a) Choose Menu Configuration Process Measurement API Referral.
b) Set Enabled/Disabled to Enabled.
Configuration and Use Manual 55
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Configure API Referral using the display
The API Referral parameters specify the API table, measurement units, and reference values to be used in referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process fluid.
You must know the reference temperature and reference pressure that you want to use.
Procedure
1.
Choose Menu Configure Process Measurement API Referral.
2. Set API Table to the API table that you want to use to calculate referred density.
Each API table is associated with a specific set of equations. Choose your API table based on your process fluid and the measurement unit that you want to use for referred density.
Your choice also determines the API table that will be used to calculate the correction factor for volume (CTPL or CTL).
3. Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
230.0 x 10-6 to 930.0 x 10-6 per °F
414.0 x 10-6 to 1674.0 x 10-6 per °C
5. If required, set Reference Temperature to the temperature to which density will be corrected in
referred density calculations.
The default reference temperature is determined by the selected API table.
6. If required, set Reference Pressure to the pressure to which density will be corrected in referred
density calculations.
The default reference pressure is determined by the selected API table.
Set up temperature and pressure data for API Referral using the display
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value may produce inaccurate process data.
56 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART communications.
If you plan to use the mA Input for temperature data, Channel D must be available for use, and must be wired to an external temperature device.
If you plan to use the mA Input for pressure data, Channel D must be available for use, and must be wired to an external pressure device.
Restriction
You can use an mA Input for either external temperature or external pressure, but not both.
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
If you are using an external temperature device, it must use the temperature unit that is configured in the transmitter.
Procedure
1.
Choose the method to be used to supply temperature data, and perform the required setup.
Method Description Setup
Internal temperature
Polling The meter polls an
Temperature data from the on-board temperature sensor (RTD) will be used for all measurements and calculations. No external temperature data will be available.
external device for temperature data. This data will be available in addition to the internal temperature data.
a. Choose Menu Configuration Process Measurement
b.
a. Choose Menu Configuration Process Measurement
b.
c. Choose Poll External Device.
d. Select Polled Variable 1 or Polled Variable 2.
e. Set Variable to External Temperature.
f. Set Polling Control to Poll as Primary or Poll as Secondary.
Temperature.
Set External Temperature to Off.
Temperature.
Set External Temperature to On.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the network.
A field communicator is not a HART master.
g. Set External Device Tag to the HART tag of the external temperature
device.
Configuration and Use Manual 57
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Method Description Setup
mA Input An external device
sends temperature data to the meter through mA Input. This data will be available in addition to the internal temperature data.
Digital communications
A host writes temperature data to the meter at appropriate intervals. This data will be available in addition to the internal temperature data.
a. Choose Menu Configuration Process Measurement
Temperature.
Set External Temperature to On.
b.
c. Choose Menu Configuration Inputs/Outputs Channel D.
d. Set I/O Type to MAI.
e. Set Power appropriately.
f. Choose I/O Settings.
g. Set Assignment to External Temperature.
h. Set Lower Range Value and Upper Range Value appropriately.
a. Choose Menu Configuration Process Measurement
Temperature.
b.
Set External Temperature to On.
c. Perform the necessary host programming and communications setup
to write temperature data to the transmitter at appropriate intervals.
2. Choose the method to be used to supply pressure data, and perform the required setup.
Method Description Setup
Polling The meter polls an
external device for pressure data.
a. Choose Menu Configuration Process Measurement Pressure
External Pressure.
Set External Pressure to On.
b.
c. Choose Poll External Device.
d. Select Polled Variable 1 or Polled Variable 2.
e. Set Variable to External Pressure.
f. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the network.
A field communicator is not a HART master.
g. Set External Device Tag to the HART tag of the external pressure
device.
58 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Method Description Setup
mA Input An external device
sends pressure data to the meter through mA Input.
Digital communications
A host writes pressure data to the meter at appropriate intervals.
a. Choose Menu Configuration Process Measurement Pressure
External Pressure.
Set External Pressure to On.
b.
c. Choose Menu Configuration Inputs/Outputs Channel D.
d. Set I/O Type to MAI.
e. Set Power appropriately.
f. Choose I/O Settings.
g. Set Assignment to External Pressure.
h. Set Lower Range Value and Upper Range Value appropriately.
a. Choose Menu Configuration Process Measurement Pressure
External Pressure.
b.
Set External Pressure to On.
c. Perform the necessary host programming and communications setup
to write pressure data to the transmitter at appropriate intervals.
Postrequisites
Choose Menu Service Tools Service Data View Process Variables and verify the values for External Temperature and External Pressure.
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
For an mA Input:Verify the wiring between the meter and the external device.
Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
Verify the Lower Range Value and Upper Range Value settings for an mA Input.
Perform an mA Input trim.
Adjust the mA Input damping value.
For digital communications:Verify that the host has access to the required data.
Verify that the host is writing to the correct register in memory, using the correct data type.
Configuration and Use Manual 59
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166

5.1.2 Set up the API Referral application using ProLink III

Enable the API Referral application using ProLink III
Prerequisites
The API Referral application must be licensed and enabled on your transmitter before you can perform any setup. If the API Referral application was enabled at the factory, you do not need to enable it now.
Procedure
1.
Choose Device Tools Configuration Process Measurement Flow and ensure that Volume Flow Type is set to Liquid Volume.
2. Choose Device Tools Configuration Transmitter Options.
3. If the concentration measurement application is enabled, disable it and select Apply.
The concentration measurement application and the API Referral application cannot be enabled simultaneously.
4. Enable API Referral and select Apply.
Configure API Referral using ProLink III
The API Referral parameters specify the API table, measurement units, and reference values to be used in referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process fluid.
You must know the reference temperature and reference pressure that you want to use.
Procedure
1. Choose Device Tools Configuration Process Measurement API Referral.
2. Specify the API table to use to calculate referred density.
Each API table is associated with a specific set of equations.
a) Set Process Fluid to the API table group that your process fluid belongs to.
API table group Process fluids
A tables Generalized crude and JP4
B tables Generalized products: Gasoline, jet fuel, aviation fuel, kerosene, heating oils,
fuel oils, diesel, gas oil
C tables Liquids with a constant base density or known thermal expansion coefficient
(TEC). You will be required to enter the TEC for your process fluid.
D tables Lubricating oils
E tables NGL (Natural Gas Liquids) and LPG (Liquid Petroleum Gas)
60 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
b) Set Referred Density Measurement Unit to the measurement units that you want to use for
referred density.
c)
Select Apply.
These parameters uniquely identify the API table to be used to calculate referred density. The selected API table is displayed, and the meter automatically changes the density unit, temperature unit, pressure unit, and reference pressure to match the API table.
Your choice also determines the API table that will be used to calculate the correction factor for volume (CTPL or CTL).
Restriction
Not all combinations are supported by the API Referral application. See the list of API tables in this manual.
3. Refer to the API documentation and confirm your table selection.
Verify that your process fluid falls within range for line density, line temperature, and line
a)
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
230.0 x 10-6 to 930.0 x 10-6 per °F
414.0 x 10-6 to 1674.0 x 10-6 per °C
5. Set Reference Temperature to the temperature to which density will be corrected in referred density
calculations. If you choose Other, select the temperature measurement unit and enter the reference temperature.
6. Set Reference Pressure to the pressure to which density will be corrected in referred density
calculations.
Set up temperature and pressure data for API Referral using ProLink III
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value may produce inaccurate process data.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART communications.
If you plan to use the mA Input for temperature data, Channel D must be available for use, and must be wired to an external temperature device.
If you plan to use the mA Input for pressure data, Channel D must be available for use, and must be wired to an external pressure device.
Configuration and Use Manual 61
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Restriction
You can use an mA Input for either external temperature or external pressure, but not both.
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
If you are using an external temperature device, it must use the temperature unit that is configured in the transmitter.
Procedure
1.
Choose Device Tools Configuration Process Measurement API Referral.
2. Choose the method to be used to supply temperature data, and perform the required setup.
Option Description Setup
Internal RTD temperature data
Polling The meter polls an external
mA Input An external device sends
Temperature data from the on­board temperature sensor (RTD) is used.
device for temperature data. This data will be available in addition to the internal RTD temperature data.
temperature data to the meter through an mA Input. This data will be available in addition to the internal temperature data.
a. Set Line Temperature Source to Internal RTD.
Select Apply.
b.
a. Set Line Temperature Source to Poll for External Value.
b.
Set Polling Slot to an available slot.
c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not a HART master.
d. Set External Device Tag to the HART tag of the
temperature device.
e.
Select Apply.
a. Set Channel D to operate as an mA Input.
b.
Set mA Input Assignment to External Temperature.
c. Set Temperature Source to mA Input.
Digital communications
A host writes temperature data to the meter at appropriate intervals. This data will be available in addition to the internal RTD temperature data.
a. Set Line Temperature Source to Fixed Value or Digital
Communications.
b. Select Apply.
c. Perform the necessary host programming and
communications setup to write temperature data to the meter at appropriate intervals.
3. Choose the method you will use to supply pressure data, and perform the required setup.
62 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Option Description Setup
Polling The meter polls an external
device for pressure data.
mA Input An external device sends
pressure data to the meter through an mA Input.
Digital communications
A host writes pressure data to the meter at appropriate intervals.
a. Set Pressure Source to Poll for External Value.
Set Polling Slot to an available slot.
b.
c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not a HART master.
d. Set External Device Tag to the HART tag of the
temperature device.
a. Set Channel D to operate as an mA Input.
b.
Set mA Input Assignment to External Pressure.
c. Set Pressure Source to mA Input.
a. Set Pressure Source to Fixed Value or Digital
Communications.
b.
Perform the necessary host programming and communications setup to write pressure data to the meter at appropriate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group on the ProLink III main window.
The current pressure value is displayed in the External Pressure field. Verify that the value is correct.
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
For an mA Input:Verify the wiring between the meter and the external device.
Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
Verify the Lower Range Value and Upper Range Value settings for an mA Input.
Configuration and Use Manual 63
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Perform an mA Input trim.
Adjust the mA Input damping value.
For digital communications:Verify that the host has access to the required data.
Verify that the host is writing to the correct register in memory, using the correct data type.

5.1.3 Set up the API Referral application using a field communicator

Enable the API Referral application using a field communicator
Prerequisites
The API Referral application must be licensed and enabled on your transmitter. If the API Referral application was enabled at the factory, you do not need to enable it now.
Volume Flow Type must be set to Liquid.
Procedure
1.
Choose Configure Manual Setup Measurements Optional Setup GSV and ensure that Volume Flow Type is set to Liquid.
This parameter is available only if API Referral or concentration measurement is not enabled. If you do not see this parameter, it is already set correctly.
2. If the concentration measurement application is enabled, disable it.
The concentration measurement application and the API Referral application cannot be enabled simultaneously.
3. Enable the API Referral application.
4. If Advance Phase Measurement Output Type is other than Disabled, disable it.
The Advance Phase Measurement application and the API Referral application cannot be enabled simultaneously.
Configure API Referral using a field communicator
The API Referral parameters specify the API table, measurement units, and reference values to be used in referred density calculations.
Prerequisites
You will need API documentation for the API table that you select.
Depending on your API table, you may need to know the thermal expansion coefficient (TEC) for your process fluid.
You must know the reference temperature and reference pressure that you want to use.
Procedure
1. Choose Configure Manual Setup Measurements Optional Setup API Referral.
2. Choose API Referral Setup.
64 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
3. Specify the API table that you want to use to calculate referred density.
Each API table is associated with a specific set of equations.
a)
Set API Table Number to the number that matches the API table units that you want to use for referred density.
Your choice also determines the measurement unit to be used for temperature and pressure, and the default values for reference temperature and reference pressure.
API table number
Measurement unit for referred density
Temperature measurement unit
Pressure measurement unit
Default reference temperature
Default reference pressure
5 °API °F psi (g) 60 °F 0 psi (g)
(1)
6
°API °F psi (g) 60 °F 0 psi (g)
23 SGU °F psi (g) 60 °F 0 psi (g)
(1)
24
SGU °F psi (g) 60 °F 0 psi (g)
53 kg/m³ °C kPa (g) 15 °C 0 kPa (g)
(1)
54
59
60
(2)
(2)
kg/m³ °C kPa (g) 15 °C 0 kPa (g)
kg/m³ °C kPa (g) 20 °C 0 kPa (g)
kg/m³ °C kPa (g) 20 °C 0 kPa (g)
(1) Used only with API Table Letter = C. (2) Used only with API Table Letter = E.
b) Set API Table Letter to the letter of the API table group that is appropriate for your process fluid.
API table letter Process fluids
A Generalized crude and JP4
B Generalized products: Gasoline, jet fuel, aviation fuel, kerosene, heating oils,
fuel oils, diesel, gas oil
(1)
C
Liquids with a constant base density or known thermal expansion coefficient (TEC). You will be required to enter the TEC for your process fluid.
D Lubricating oils
(2)
E
NGL (Natural Gas Liquids) and LPG (Liquid Petroleum Gas)
(1) Used only with API Table Number= 6, 24, or 54. (2) Used only with API Table Number = 23, 24, 53, 54, 59, or 60.
API Table Number and API Table Letter uniquely identify the API table. The selected API table is displayed, and the meter automatically changes the density unit, temperature unit, pressure unit, reference temperature, and reference pressure to match the API table.
Your choice also determines the API table that will be used to calculate the correction factor for volume (CTPL or CTL).
Configuration and Use Manual 65
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Restriction
Not all combinations are supported by the API Referral application. See the list of API tables in this manual.
4. If you chose a C table, enter Thermal Expansion Coefficient (TEC) for your process fluid.
Acceptable limits:
230.0 x 10-6 to 930.0 x 10-6 per °F
414.0 x 10-6 to 1674.0 x 10-6 per °C
5.
Refer to the API documentation and confirm your table selection.
a) Verify that your process fluid falls within range for line density, line temperature, and line
pressure.
b) Verify that the referred density range of the selected table is adequate for your application.
6. If required, set Reference Temperature to the temperature to which density will be corrected in
referred density calculations.
The default reference temperature is determined by the selected API table.
7. If required, set Reference Pressure to the pressure to which density will be corrected in referred
density calculations.
The default reference pressure is determined by the selected API table. API Referral requires gauge pressure.
Set up temperature and pressure data for API Referral using a field communicator
The API Referral application uses temperature and, optionally, pressure data in its calculations. You must decide how to provide this data, then perform the required configuration and setup.
Note
Fixed values for temperature or pressure are not recommended. Using a fixed temperature or pressure value may produce inaccurate process data.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART communications.
If you plan to use the mA Input for temperature data, Channel D must be available for use, and must be wired to an external temperature device.
If you plan to use the mA Input for pressure data, Channel D must be available for use, and must be wired to an external pressure device.
Restriction
You can use an mA Input for either external temperature or external pressure, but not both.
The pressure measurement must be gauge pressure, not atmospheric pressure.
The pressure device must use the pressure unit that is configured in the transmitter.
66 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
If you are using an external temperature device, it must use the temperature unit that is configured in the transmitter.
Procedure
1.
Choose the method to be used to supply temperature data, and perform the required setup.
Method Description Setup
Internal RTD temperature data
Polling The meter polls an external
Temperature data from the on­board temperature sensor (RTD) is used.
device for temperature data. This data will be available in addition to the internal RTD temperature data.
a. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature Temperature.
b.
Set External Temperature to Disable.
a. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature Temperature.
b.
Set External Temperature to Enable.
c. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature External Polling.
d. Set Poll Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not a HART master.
e. Choose an unused polling slot.
Set External Device Tag to the HART tag of the external
f.
temperature device.
g. Set Polled Variable to Temperature.
mA Input An external device sends
temperature data to the meter through an mA Input. This data will be available in addition to the internal temperature data.
Configuration and Use Manual 67
a. Choose Configure Manual Setup Inputs/Outputs
Channel D.
b.
Set Assignment to mA Input.
c. Choose mA Input mA Input Settings.
d. Set Var Assignment to External Temperature.
e. Set Upper Range Value and Lower Range Value
appropriately.
f. Set Damping appropriately.
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Method Description Setup
Digital communications
A host writes temperature data to the meter at appropriate intervals. This data will be available in addition to the internal RTD temperature data.
a. Choose Configure Manual Setup Measurements
Optional Setup External Variables External Temperature.
b.
Set Temperature Compensation to Enable.
c. Perform the necessary host programming and
communications setup to write temperature data to the meter at appropriate intervals.
2. Choose the method to be used to supply pressure data, and perform the required setup.
Method Description Setup
Polling The meter polls an external
device for pressure data.
a. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature Pressure.
b.
Set Pressure Compensation to Enable.
c. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature External Polling.
d. Choose an unused polling slot.
e. Set Poll Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not a HART master.
f. Set External Device Tag to the HART tag of the external
pressure device.
Set Polled Variable to Pressure.
g.
68 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Method Description Setup
mA Input An external device sends
pressure data to the meter through an mA Input.
Digital communications
A host writes pressure data to the meter at appropriate intervals.
a. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature Pressure.
b.
Set Pressure Compensation to Enable.
c. Choose Configure Manual Setup Inputs/Outputs
Channel D.
d. Set Assignment to mA Input.
e. Choose mA Input mA Input Settings.
f. Set Var Assignment to External Pressure.
g. Set Upper Range Value and Lower Range Value
appropriately.
h. Set Damping appropriately.
a. Choose Configure Manual Setup Measurements
Optional Setup External Variables External Pressure.
b.
Set Pressure Compensation to Enable.
c. Perform the necessary host programming and
communications setup to write pressure data to the transmitter at appropriate intervals.
Postrequisites
Choose Service Tools Variables Process and verify the values for External Temperature and External Pressure.
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
For am mA Input:Verify the wiring between the meter and the external device.
Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
Verify the Lower Range Value and Upper Range Value settings for the mA Input.
Perform an mA Input trim.
Adjust the mA Input damping value.
For digital communications:Verify that the host has access to the required data.
Configuration and Use Manual 69
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Verify that the host is writing to the correct register in memory, using the correct data type.

5.1.4 API tables supported by the API Referral application

The API tables listed here are supported by the API Referral application.
Table 5-1: API tables, process fluids, measurement units, and default reference values
Process fluid
Generalized crude and JP4
Generalized products (gasoline, jet fuel, aviation fuel, kerosene, heating oils, fuel oils, diesel, gas oil)
Liquids with a constant density base or known thermal expansion coefficient
(5)
API tables
(calculations)
Referred density
(2)
5A 6A Unit: °API
(1)
CTL or CTPL
Referred density (API): unit and range
(3) (4)
Default reference temp
Default reference pressure
API standard
60 °F 0 psi (g) API MPMS 11.1
Range: 0 to 100 °API
23A 24A Unit: SGU
60 °F 0 psi (g)
Range: 0.6110 to
1.0760 SGU
53A 54A Unit: kg/m
3
15 °C 0 kPa (g)
Range: 610 to 1075 kg/m³
5B 6B Unit: °API
60 °F 0 psi (g) API MPMS 11.1
Range: 0 to 85 °API
23B 24B Unit: SGU
60 °F 0 psi (g)
Range: 0.6535 to
1.0760 SGU
53B 54B Unit: kg/m
3
15 °C 0 kPa (g)
Range: 653 to 1075 kg/m³
N/A 6C Unit: °API 60 °F 0 psi (g) API MPMS 11.1
N/A 24C Unit: SGU 60 °F 0 psi (g)
N/A 54C Unit: kg/m³ 15 °C 0 kPa (g)
5D 6D Unit: °API
60 °F 0 psi (g) API MPMS 11.1
Range: 10 to +40 °API
Lubricating oils
23D 24D Unit: SGU
Range: 0.8520 to
60 °F 0 psi (g)
1.1640 SGU
53D 54D Unit: kg/m³
15 °C 0 kPa (g)
Range: 825 to 1164 kg/m³
70 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Table 5-1: API tables, process fluids, measurement units, and default reference values (continued)
API tables
Process fluid
NGL (natural gas liquids) and LPG (liquid petroleum gas)
(1) Each API table represents a specialized equation defined by the American Petroleum Institute for a specific combination
of process fluid, line conditions, and output.
(2) Referred density is calculated from line density. You must specify this table, either directly or by selecting the process
fluid and base density measurement unit.
(3) You do not need to specify this table. It is invoked automatically as a result of the previous table selection. (4) CTL is a correction factor based on online temperature. CTPL is a correction factor based on both line pressure and line
temperature. Calculation of CTL and CTPL for A, B, C, and D table products is in accordance with API MPMS Chapter
11.1. Calculation of CTL and CTPL for E table products is in accordance with API MPMS Chapters 11.2.2, 11.2.4, and
11.2.5.
(5) The Thermal Expansion Coefficient (TEC) replaces the referred density calculation. Use the CTL/CTPL table instead.
5.1.5

Process variables from the API Referral application

(calculations)
Referred density
23E 24E Unit: SGU 60 °F 0 psi (g) API MPMS 11.2.4
53E 54E Unit: kg/m³ 15 °C 0 psi (g)
59E 60E Unit: kg/m³ 20 °C 0 psi (g)
(2)
(1)
CTL or CTPL
Referred density (API): unit and range
(3) (4)
Default reference temp
Default reference pressure
API standard
The API Referral application calculates several different process variables according to API standards.
CTPL
Correction factor based on line temperature and line pressure.
CTL
Referred density
API volume flow
Batch-weighted average density
Batch-weighted average temperature
API volume total
API volume inventory
Correction factor based on line temperature at saturation conditions.
The measured density after CTL or CTPL has been applied.
The measured volume flow rate after CTL or CTPL has been applied. Also called corrected volume flow.
One density value is recorded for each unit of flow (e.g., barrel, liter). The average is calculated from these values. The average is reset when the API totalizer is reset. Not available unless a totalizer has been configured with Source set to Corrected Volume Flow.
One temperature value is recorded for each unit of flow (e.g., barrel, liter). The average is calculated from these values. The average is reset when the API totalizer is reset. Not available unless a totalizer has been configured with Source set to Temperature-Corrected Volume Flow.
The total API volume measured by the transmitter since the last API totalizer reset. Also called corrected volume total. Not available unless a totalizer has been configured with Source set to Corrected Volume Flow.
The total API volume measured by the transmitter since the last API inventory reset. Also called corrected volume inventory. Not available unless an inventory has been configured with Source set to Corrected Volume Flow.
Configuration and Use Manual 71
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166

5.2 Set up concentration measurement

The concentration measurement application calculates concentration from line density and line temperature.
5.2.1
The procedure for setting up concentration measurement application depends on how your device was ordered and how you want to use the application. Review this information before you begin.
Requirements for concentration measurement
To use the concentration measurement application, the following conditions must be met:
The concentration measurement application must be enabled.
The API Referral application must be disabled.
The gas piecewise linearization (PWL) application must be disabled.
The Advanced Phase Measurement application must be disabled or set for the Liquid with Gas application.
A concentration matrix must be loaded into one of the six slots on the transmitter.
Temperature Source must be configured and set up.
One matrix must be selected as the active matrix (the matrix used for measurement).
Requirements for matrices
A matrix is the set of coefficients used to convert process data to concentration, plus related parameters. The matrix can be saved as a file.

Preparing to set up concentration measurement

Tip
In most cases, the concentration matrix that you ordered was loaded at the factory. If it was not, you have several options for loading a matrix. You can also build a matrix.
The transmitter requires all matrices to be in .matrix format. You can use ProLink III to load matrices in other formats:
.edf (used by ProLink II)
.xml (used by ProLink III)
The transmitter can store matrices in two locations:
One of the six slots in memory
The transmitter SD card
Any matrix in a slot is available for use. In other words, it can be selected as the active matrix and used for measurement. Matrices on the SD card are not available for use. Matrices must be loaded into a slot before they can be used for measurement.
All matrices in slots must use the same derived variable. Matrices on the SD card have no requirement for their derived variables to match.
See the following table for the different ways that you can load matrices.
72 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Table 5-2: Methods to load a matrix file
Action Display ProLink III Field communicator
Load matrix from USB drive to SD card
Load matrix from computer to slot
Load matrix from SD card to slot
Requirements for derived variables
A derived variable is the process variable that a concentration matrix measures. All other process variables are calculated from the derived variable. There are eight possible derived variables. Each matrix is designed for one specific derived variable.
The transmitter can store up to six matrices in six slots. There are additional matrices on the transmitter SD card. All matrices in the six slots must use the same derived variable. If you change the setting of Derived Variable, all matrices are deleted from the six slots. Any matrices on the transmitter SD card are not affected.
Tip
Always ensure that Derived Variable is set correctly before loading matrices into slots.
Derived variables and net flow rate
If you want the transmitter to calculate Net Mass Flow Rate, the derived variable must be set to Mass Concentration (Density). If your matrix is not designed for Mass Concentration (Density), contact customer support for assistance.
If you want the transmitter to calculate Net Volume Flow Rate, the derived variable must be set to Volume Concentration (Density). If your matrix is not designed for Volume Concentration (Density), contact customer support for assistance.
Derived variables based on specific gravity
The following derived variables are based on specific gravity:
Specific Gravity
Concentration (Specific Gravity)
Mass Concentration (Specific Gravity)
Volume Concentration (Specific Gravity)
If you are using one of these derived variables, two additional parameters can be configured:
Reference Temperature of Water (default setting: 4 °C)
Water Density at Reference Temperature (default setting: 999.99988 kg/m³)
These two parameters are used to calculate specific gravity.
You cannot set these parameters from the display. If the default values are not appropriate, you must use another method to set them.
Configuration and Use Manual 73
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Optional tasks in setting up concentration measurement
The following tasks are optional:
Modifying names and labels
Configuring extrapolation alerts
5.2.2
This section guides you through most of the tasks related to setting up and implementing the concentration measurement application.
Restriction
This section does not cover building a concentration matrix. For detailed information on building a matrix, see the Micro Motion Enhanced Density Application Manual.

Set up concentration measurement using the display

Enable concentration measurement using the display
The concentration measurement application must be enabled before you can perform any setup. If the concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration measurement cannot be enabled at the same time:
Advanced Phase Measurement — gas with liquid
Advanced Phase Measurement — net oil
API Referral
Piecewise linearization (PWL)
Gas Standard Volume
Procedure
1.
Choose Menu Configuration Process Measurement.
2. Choose Flow Variables Volume Flow Settings and ensure that Flow Type is set to Liquid.
3. Return to the Process Measurement menu.
4. If the API Referral application is displayed in the menu, choose API Referral and ensure that Enabled/ Disabled is set to Disabled.
The concentration measurement application and the API Referral application cannot be enabled simultaneously.
5. If the Advanced Phase Measurement application is displayed in the menu, choose Advanced Phase Measurement Application Setup and ensure that Enabled/Disabled is set to Disabled.
6. Enable concentration measurement.
a) Choose Menu Configuration Process Measurement Concentration Measurement.
b) Set Enabled/Disabled to Enabled.
74 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Load a concentration matrix from a USB drive using the display
At least one concentration matrix must be loaded into one of the six slots on your transmitter. You can load up to six matrices into slots. You can also copy matrices to the transmitter SD card, and load them into slots at a later time.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not need to load any matrices.
WARNING
If the transmitter is in a hazardous area, do not remove the housing cover while the transmitter is powered up. Failure to follow these instructions can cause an explosion resulting in injury or death.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The transmitter SD card and the ProLink III installation include a set of standard concentration matrices. Other matrices are available from Emerson.
Each concentration matrix file must be in .matrix format.
Tip
If you have a custom matrix on another device, you can save it to a file, then load it to the current device.
If you have a matrix file in a different format, you can load it using ProLink III.
The .matrix files must be copied to the root directory of a USB drive.
You must know the derived variable that the matrix is designed to calculate.
Important
All concentration matrices on your transmitter must use the same derived variable.
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the SD card. Set Derived Variable before loading concentration matrices.
Procedure
1.
Choose Menu Configuration Process Measurement Concentration Measurement Configure Application and ensure that the setting of Derived Variable matches the derived variable used by your matrix. If it does not, change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the six slots, but not from the transmitter SD card. Verify the setting of Derived Variable before continuing.
2. Load the matrix.
Remove the cover from the transmitter wiring compartment, open the snap flap to access the
a)
service port, and insert the USB drive into the service port.
Configuration and Use Manual 75
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
b) Choose Menu USB Options USB Drive --> Transmitter Upload Configuration File.
c)
Set Config File Type to Concentration Measurement Matrix.
d) Select the .matrix file that you want to load, and wait for the transfer to complete.
3. Choose Yes or No when you are asked if you want to apply the settings.
The transmitter has six slots that are used to store concentration matrices. Any one of these can be used for measurement. The transmitter also has the capability to store multiple concentration matrices on its SD card. These cannot be used for measurement until they are moved to a slot.
Option Description
Yes The matrix is saved to the SD card, and the loading process continues with loading the
matrix into one of the slots.
No The matrix is saved to the SD card, and the loading process ends. You must load a matrix into
a slot before you can use it for measurement.
4. If you chose Yes, select the slot to load this matrix into, and wait until the load is complete.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
Postrequisites
If you loaded the matrix into a slot, choose Menu Configuration Process Measurement Concentration Measurement Configure Application Active Matrix and ensure that the matrix is listed.
If you loaded the matrix onto the SD card only, choose Menu Configuration Process Measurement Concentration Measurement Load Matrix and ensure that the matrix is listed.
Load a concentration matrix from the SD card using the display
If you have a concentration matrix on the transmitter SD card, you can load it into one of the six slots on your transmitter. You cannot use the matrix for measurement until it has been loaded into a slot. You can load up to six matrices into slots.
Prerequisites
You must have one or more concentration matrices stored on the transmitter SD card. The standard matrices are loaded to the SD card at the factory.
You must know the derived variable that the matrix is designed to calculate.
Procedure
Choose Menu Configuration Process Measurement Concentration Measurement and ensure
1.
that the setting of Derived Variable matches the derived variable used by your matrix. If it does not, change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the six slots, but not from the transmitter SD card. Verify the setting of Derived Variable before continuing.
76 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
2. Choose Menu Configuration Process Measurement Concentration Measurement Load
Matrix. The transmitter displays a list of all matrices that are on the SD card.
Select the matrix that you want to load.
3.
4. Select the slot that you want to load it into.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
Postrequisites
Choose Menu Configuration Process Measurement Concentration Measurement Configure Application Active Matrix and ensure that the matrix is listed.
Set up temperature data using the display
The concentration measurement application uses line temperature data in its calculations. You must decide how to provide this data, then perform the required configuration and setup. Temperature data from the on­board temperature sensor (RTD) is always available. Optionally, you can set up an external temperature device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on this meter.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the internal RTD temperature in some areas and an external temperature in others. The transmitter stores the internal RTD temperature and the external temperature separately. However, the transmitter stores only one alternate temperature value, which may be either the external temperature or the configured fixed value. Accordingly, if you choose a fixed temperature for some uses, and an external temperature for others, the external temperature will overwrite the fixed value.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART communications.
If you plan to use the mA Input for temperature data, Channel D must be available for use, and must be wired to an external temperature device.
Procedure
Choose the method to be used to supply temperature data, and perform the required setup.
Method Description Setup
Internal temperature
Temperature data from the on­board temperature sensor (RTD) will be used for all measurements and calculations. No external temperature data will be available.
a. Choose Menu Configuration Process Measurement
Temperature.
Set External Temperature to Off.
b.
Configuration and Use Manual 77
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Method Description Setup
Polling The meter polls an external
device for temperature data. This data will be available in addition to the internal temperature data.
mA Input An external device sends
temperature data to the meter via the mA Input. This data will be available in addition to the internal temperature data.
a. Choose Menu Configuration Process Measurement
Temperature.
Set External Temperature to On.
b.
c. Choose Poll External Device.
d. Select Polled Variable 1 or Polled Variable 2.
e. Set Variable to External Temperature.
f. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not a HART master.
g. Set External Device Tag to the HART tag of the external
temperature device.
a. Choose Menu Configuration Process Measurement
Temperature.
Set External Temperature to On.
b.
c. Choose Menu Configuration Inputs/Outputs
Channel D.
d. Set I/O Type to MAI.
e. Set Power appropriately.
f. Choose I/O Settings.
g. Set Assignment to External Temperature.
h. Set Lower Range Value and Upper Range Value
appropriately.
Digital communications
A host writes temperature data to the meter at appropriate intervals. This data will be available in addition to the internal temperature data.
a. Choose Menu Configuration Process Measurement
Temperature.
b.
Set External Temperature to On.
c. Perform the necessary host programming and
communications setup to write temperature data to the transmitter at appropriate intervals.
Postrequisites
Choose Menu Service Tools Service Data View Process Variables and verify the value for External Temperature.
78 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
For an mA Input:Verify the wiring between the meter and the external device.
Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
Verify the Lower Range Value and Upper Range Value settings for an mA Input.
Perform an mA Input trim.
Adjust the mA Input damping value.
For digital communications:Verify that the host has access to the required data.
Verify that the host is writing to the correct register in memory, using the correct data type.
Modify matrix names and labels using the display
For convenience, you can change the name of a concentration matrix and the label used for its measurement unit. This does not affect measurement.
Procedure
1.
Choose Menu Configuration Process Measurement Concentration Measurement Configure Matrix.
2. Select the matrix that you want to modify.
3. Set Matrix Name to the name that will be used for this matrix.
4. Set Concentration Unit to the label that will be used for the concentration unit.
If you want to use a custom label, you can use the display to select Special. However, you cannot use the display to configure the custom label. You must use another tool to change the label from Special to a user-defined string.
Modify extrapolation alerts using the display
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density or line temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is occurring.
Each concentration matrix has its own extrapolation alert limits.
Configuration and Use Manual 79
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Procedure
1.
Choose Menu Configuration Process Measurement Concentration Measurement Configure Matrix.
2. Select the matrix that you want to modify.
3. Set Extrapolation Limit to the point, in percent, at which an extrapolation alert will be posted.
4. Choose Menu Configuration Process Measurement Concentration Measurement
Configure Application Extrapolation Alerts.
5. Enable or disable the high and low limit alerts for temperature and density as desired.
Example: Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using the display
You must select the concentration matrix to be used for measurement. Although the transmitter can store up to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Menu Configuration Process Measurement Concentration Measurement
Configure Application.
2. Set Active Matrix to the matrix you want to use.

5.2.3 Set up concentration measurement using ProLink III

This section guides you through the tasks required to set up, configure, and implement concentration measurement.
Enable concentration measurement using ProLink III
The concentration measurement application must be enabled before you can perform any setup. If the concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration measurement cannot be enabled at the same time:
Advanced Phase Measurement — gas with liquid
Advanced Phase Measurement — net oil
API Referral
Piecewise linearization (PWL)
Gas Standard Volume
80 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Procedure
1.
Choose Device Tools Configuration Process Measurement Flow and ensure that Volume Flow Type is set to Liquid Volume.
2. Choose Device Tools Configuration Process Measurement Advance Phase Measurement
APM Status and ensure that Application Status is set to Disable or Liquid with Gas.
3. Choose Device Tools Configuration Transmitter Options.
4. Disable API Referral and set the Advance Phase Measurement application to Disabled or Single Liquid.
5. Disable gas Piecewise Linearization (PWL), and set the Advance Phase Measurement application to
Disabled or Single Liquid.
6. Set Concentration Measurement to Enabled and select Apply.
Load a concentration matrix using ProLink III
At least one concentration matrix must be loaded onto your transmitter. You can load up to six.
Prerequisites
The concentration measurement application must be enabled on your device.
For each concentration matrix that you want to load, you need a file containing the matrix data. The ProLink III installation includes a set of standard concentration matrices. Other matrices are available from Emerson. The file can be on your computer or in the transmitter internal memory.
The file must be in one of the formats that ProLink III supports. This includes:
.xml (ProLink III)
.matrix (5700)
If you are loading an .xml file, you must know the following information for your matrix:
The derived variable that the matrix is designed to calculate
The density unit that the matrix was built with
The temperature unit that the matrix was built with
If you are loading a .matrix file, you must know the derived variable that the matrix is designed to calculate.
Important
All concentration matrices on your transmitter must use the same derived variable.
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the
six slots on the transmitter, but not from the transmitter SD card. Set Derived Variable before loading concentration matrices.
ProLink III loads matrices directly to one of the six transmitter slots.
Tip
In many cases, concentration matrices were ordered with the device and loaded at the factory. You may not need to load any matrices.
Restriction
You cannot use ProLink III to load a matrix to the transmitter SD card.
Configuration and Use Manual 81
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Procedure
1.
If you are loading an .xml file, choose Device Tools Configuration Process Measurement Line Density and set Density Unit to the density unit used by your matrix.
Important
When you load a matrix in one of these formats, if the density unit is not correct, concentration data will be incorrect. The density units must match at the time of loading. You can change the density unit after the matrix is loaded.
2. If you are loading an .xml file, choose Device Tools Configuration Process Measurement Line
Temperatureand set Temperature Unit to the temperature unit used by your matrix.
Important
When you load a matrix in one of these formats, if the temperature unit is not correct, concentration data will be incorrect. The temperature units must match at the time of loading. You can change the temperature unit after the matrix is loaded.
3. Choose Device Tools Configuration Process Measurement Concentration Measurement.
The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
4.
In Step 1, ensure that the setting of Derived Variable matches the derived variable used by your matrix. If it does not, change it as required and select Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the six slots. Verify the setting of Derived Variable before continuing.
5. Load one or more matrices.
In Step 2, set Matrix Being Configured to the location (slot) to which the matrix will be loaded.
a)
b) To load a .xml file from your computer, select Load Matrix from File, navigate to the file, and
load it.
c) To load a .matrix file from your computer, select Load Matrix from My Computer, navigate to
the file, and load it.
d) To load a .matrix file from the transmitter internal memory, select Load Matrix from 5700 Device
Memory, navigate to the file on the transmitter, and load it.
e) Repeat until all required matrices are loaded.
Set reference temperature values for specific gravity using ProLink III
When Derived Variable is set to any option based on specific gravity, you must set the reference temperature for water, then verify the density of water at the configured reference temperature. These values affect specific gravity measurement.
This requirement applies to the following derived variables:
Specific Gravity
Concentration (Specific Gravity)
82 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Mass Concentration (Specific Gravity)
Volume Concentration (Specific Gravity)
Procedure
1.
Choose Device Tools Configuration Process Measurement Concentration Measurement. The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and select Change
Matrix.
3. Scroll to Step 3, then perform the following actions:
a) Set Reference Temperature for Referred Density to the temperature to which line density will
be corrected for use in the specific gravity calculation.
b) Set Reference Temperature for Water to the water temperature that will be used in the specific
gravity calculation.
c) Set Water Density at Reference Temperature to the density of water at the specified reference
temperature.
The transmitter automatically calculates the density of water at the specified temperature. The new value will be displayed the next time that transmitter memory is read. You can enter a different value if you prefer.
4. Select Apply at the bottom of Step 3.
Set up temperature data using ProLink III
The concentration measurement application uses line temperature data in its calculations. You must decide how to provide this data, then perform the required configuration and setup. Temperature data from the on­board temperature sensor (RTD) is always available. Optionally, you can set up an external temperature device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on this meter.
Important
Line temperature data is used in several different measurements and calculations. It is possible to use the internal RTD temperature in some areas and an external temperature in others. The transmitter stores the internal RTD temperature and the external temperature separately. However, the transmitter stores only one alternate temperature value, which may be either the external temperature or the configured fixed value. Accordingly, if you choose a fixed temperature for some uses, and an external temperature for others, the external temperature will overwrite the fixed value.
Prerequisites
If you plan to poll an external device, the primary mA Output (Channel A) must be wired to support HART communications.
Procedure
1.
Choose Device Tools Configuration Process Measurement Concentration Measurement.
Configuration and Use Manual 83
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2.
Scroll to Step 4.
3. Choose the method to be used to supply temperature data, and perform the required setup.
Option Description Setup
Internal temperature
Polling The meter polls an external
Temperature data from the on­board temperature sensor (RTD) will be used for all measurements and calculations. No external temperature data will be available.
device for temperature data. This data will be available in addition to the internal RTD temperature data.
a. Set Line Temperature Source to Internal.
Click Apply.
b.
a. Set Line Temperature Source to Poll for External Value.
b. Set Polling Slot to an available slot.
c. Set Polling Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not a HART master.
d. Set External Device Tag to the HART tag of the
temperature device.
e.
Click Apply.
mA Input An external device sends
temperature data to the meter via the mA Input. This data will be available in addition to the internal temperature data.
Digital communications
A host writes temperature data to the meter at appropriate intervals. This data will be available in addition to the internal RTD temperature data.
a. Set Channel D to operate as an mA Input.
b.
Set mA Input Assignment to External Temperature.
c. Set Pressure Source to mA Input.
a. Set Line Temperature Source to Fixed Value or Digital
Communications.
b. Click Apply.
c. Perform the necessary host programming and
communications setup to write temperature data to the meter at appropriate intervals.
Postrequisites
If you are using external temperature data, verify the external temperature value displayed in the Inputs group on the ProLink III main window.
84 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
For an mA Input:Verify the wiring between the meter and the external device.
Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
Verify the Lower Range Value and Upper Range Value settings for an mA Input.
Perform an mA Input trim.
Adjust the mA Input damping value.
For digital communications:Verify that the host has access to the required data.
Verify that the host is writing to the correct register in memory, using the correct data type.
Modify matrix names and labels using ProLink III
For convenience, you can change the name of a concentration matrix and the label used for its measurement unit. This does not affect measurement.
Procedure
1.
Choose Device Tools Configuration Process Measurement Concentration Measurement. The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change
Matrix.
3. Scroll to Step 3, then perform the following actions:
a) Set Concentration Units Label to the label that will be used for the concentration unit.
b) If you set Concentration Units Label to Special, enter the custom label in User-Defined Label.
c) In Matrix Name, enter the name to be used for the matrix.
4. Select Apply at the bottom of Step 3.
Modify extrapolation alerts using ProLink III
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density or line temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is occurring.
Configuration and Use Manual 85
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Each concentration matrix has its own extrapolation alert limits.
Procedure
1.
Choose Device Tools Configuration Process Measurement Concentration Measurement. The Concentration Measurement window is displayed. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Matrix Being Configured to the matrix you want to modify, and click Change
Matrix.
3. Scroll to Step 4.
4. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert will be posted.
5. Enable or disable the high and low limit alerts for temperature and density, as desired, and click Apply.
Example: Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using ProLink III
You must select the concentration matrix to be used for measurement. Although the transmitter can store up to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
1. Choose Device Tools Configuration Process Measurement Concentration Measurement.
The Concentration Measurement window displays. It is organized into steps that allow you to perform several different setup and configuration tasks. For this task, you will not use all the steps.
2. Scroll to Step 2, set Active Matrix to the matrix you want to use and select Change Matrix.
5.2.4 Set up concentration measurement using a field
communicator
This section guides you through most of the tasks related to setting up and implementing the concentration measurement application.
Restriction
This section does not cover building a concentration matrix. See the Micro Motion Enhanced Density Application Manual for detailed information on building a matrix.
Enable concentration measurement using a field communicator
The concentration measurement application must be enabled before you can perform any setup. If the concentration measurement application was enabled at the factory, you do not need to enable it now.
Prerequisites
The concentration measurement application must be licensed on your transmitter.
Disable the following applications before enabling concentration measurement as concentration measurement cannot be enabled at the same time:
86 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Advanced Phase Measurement — gas with liquid
Advanced Phase Measurement — net oil
API Referral
Piecewise linearization (PWL)
Gas Standard Volume
Procedure
1.
Choose Overview Device Information Licenses Enable/Disable Applications and ensure that Volume Flow Type is set to Liquid.
2. Choose Overview Device Information Licenses Enable/Disable Applications.
3. Enable the concentration measurement application.
Load a concentration matrix from the transmitter SD card using a field communicator
If you have a concentration matrix on the transmitter SD card, you can move it into one of the six slots on your transmitter. You cannot use the matrix for measurement until it has been loaded into a slot. You can load up to six matrices into slots.
Prerequisites
You must have one or more concentration matrices loaded onto the transmitter SD card.
You must know the derived variable that the matrix is designed to calculate.
Procedure
1. Choose Configure Manual Setup Measurements Optional Setup Conc Measurement
CM Configuration and ensure that the setting of Derived Variable matches the derived variable used by your matrix. If it does not, change it as required and click Apply.
Important
If you change the setting of Derived Variable, all existing concentration matrices will be deleted from the six slots, but not from the transmitter SD card. Verify the setting of Derived Variable before continuing.
2. Choose Configure Manual Setup Measurements Optional Setup Conc Measurement
Load Matrix File from IM.
Select the slot that you want to load to.
3.
You can load the matrix into any empty slot, or you can overwrite an existing matrix.
4. Enter the name of the matrix file on the SD card, without the .matrix extension.
Example
If the matrix file name is test.matrix, enter test.
Configuration and Use Manual 87
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Postrequisites
Choose Configure Manual Setup Measurements Optional Setup Conc Measurement CM Configuration Active Matrix and ensure that the selected slot contains the matrix that you loaded.
Set reference temperature values for specific gravity using a field communicator
Field communicator Configure Manual Setup Measurements Optional Setup Concentration Measurement
When Derived Variable is set to any option based on specific gravity, you must set the reference temperature for water, then verify the density of water at the configured reference temperature. These values affect specific gravity measurement.
To check the setting of Derived Variable, choose Configure Manual Setup Measurements Optional Setup Conc Measurement CM Configuration.
Important
Do not change the setting of Derived Variable. If you change the setting of Derived Variable, all existing concentration matrices will be deleted from transmitter memory.
Procedure
1.
Set Matrix Being Configured to the matrix you want to modify.
2. Choose Reference Conditions, then perform the following actions:
a) Set Reference Temperature to the temperature to which line density will be corrected for use in
the specific gravity calculation.
b) Set Water Reference Temperature to the water temperature that will be used in the specific
gravity calculation.
c) Set Water Reference Density to the density of water at the specified reference temperature.
Configuration Matrix
The transmitter automatically calculates the density of water at the specified temperature. The new value will be displayed the next time that transmitter memory is read. Optionally, you can enter a different value.
Provide temperature data using a field communicator
The concentration measurement application uses line temperature data in its calculations. You must decide how to provide this data, then perform the required configuration and setup. Temperature data from the on­board temperature sensor (RTD) is always available. Optionally, you can set up an external temperature device and use external temperature data.
The temperature setup that you establish here will be used for all concentration measurement matrices on this meter.
Procedure
1. Choose the method to be used to supply temperature data, and perform the required setup.
88 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Method Description Setup
Internal RTD temperature data
Digital communications
Temperature data from the on­board temperature sensor (RTD) is used.
A host writes temperature data to the meter at appropriate intervals. This data will be available in addition to the internal RTD temperature data.
a. Choose Configure Manual Setup Measurements
Optional Setup External Variables
Set Temperature Compensation to Disable.
b.
a. Choose Configure Manual Setup Measurements
Optional Setup External Variables.
b.
Set Temperature Compensation to Enable.
c. Perform the necessary host programming and
communications setup to write temperature data to the meter at appropriate intervals.
2. Choose the method to be used to supply temperature data, and perform the required setup.
Method Description Setup
Internal RTD temperature data
Polling The meter polls an external
Temperature data from the on­board temperature sensor (RTD) is used.
device for temperature data. This data will be available in addition to the internal RTD temperature data.
a. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature Temperature.
b.
Set External Temperature to Disable.
a. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature Temperature.
b.
Set External Temperature to Enable.
c. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature External Polling.
d. Set Poll Control to Poll as Primary or Poll as Secondary.
Option Description
Poll as Primary No other HART masters will be on the
network. A field communicator is not a HART master.
Poll as Secondary Other HART masters will be on the
network. A field communicator is not a HART master.
e. Choose an unused polling slot.
Set External Device Tag to the HART tag of the external
f.
temperature device.
g. Set Polled Variable to Temperature.
Configuration and Use Manual 89
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Method Description Setup
mA Input An external device sends
temperature data to the meter via the mA Input. This data will be available in addition to the internal temperature data.
Digital communications
A host writes temperature data to the meter at appropriate intervals. This data will be available in addition to the internal RTD temperature data.
a. Choose Configure Manual Setup Inputs/Outputs
Channel D.
Set Assignment to mA Input.
b.
c. Choose mA Input mA Input Settings.
d. Set Var Assignment to External Temperature.
e. Set Upper Range Value and Lower Range Value
appropriately.
f. Set Damping appropriately.
a. Choose Configure Manual Setup Measurements
Optional Setup External Pressure/Temperature Temperature.
b.
Set External Temperature to Enable.
c. Perform the necessary host programming and
communications setup to write temperature data to the meter at appropriate intervals.
Postrequisites
To verify the value for External Temperature, choose Service Tools Variables Process External Temperature.
Need help?
If the value is not correct:
Ensure that the external device and the meter are using the same measurement unit.
For polling:Verify the wiring between the meter and the external device.
Verify the HART tag of the external device.
For the mA Input:Verify the wiring between the meter and the external device.
Verify the power configuration for Channel D. If external power is required, ensure that the loop is
powered.
Verify the Lower Range Value and Upper Range Value settings for the mA Input.
Perform an mA Input trim.
Adjust the mA Input damping value.
For digital communications:Verify that the host has access to the required data.
Verify that the host is writing to the correct register in memory using the correct data type.
90 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Modify matrix names and labels using a field communicator
Field communicator Configure Manual Setup Measurements Optional Setup Conc Measurement Configure
For convenience, you can change the name of a concentration matrix and the label used for its measurement unit. This does not affect measurement.
Procedure
Set Matrix Being Configured to the matrix you want to modify.
1.
2. Set Matrix Name to the name to be used for the matrix.
3. Set Concentration Unit to the label that will be used for the concentration unit.
4. If you set Concentration Unit to Special, choose Label and enter the custom label.
Matrix
Modify extrapolation alerts using a field communicator
You can enable and disable extrapolation alerts, and set extrapolation alert limits. These parameters control the behavior of the concentration measurement application but do not affect measurement directly.
Each concentration matrix is built for a specific density range and a specific temperature range. If line density or line temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is occurring.
Each concentration matrix has its own extrapolation alert limits.
Procedure
1. Choose Configure Manual Setup Measurements Optional Setup Conc Measurement
Configure Matrix.
2. Set Matrix Being Configured to the matrix you want to modify.
3. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert will be posted.
4. Choose Configure Alert Setup CM Alerts.
5. Enable or disable the high and low alerts for temperature and density, as desired.
Extrapolation alerts in action
If Extrapolation Limit is set to 5%, High Limit (Temp) is enabled, and the active matrix is built for a temperature range of 40 °F (4.4 °C) to 80 °F (26.7 °C), a high-temperature extrapolation alert will be posted if line temperature goes above 82 °F (27.8 °C).
Select the active concentration matrix using a field communicator
Field communicator Configure Manual Setup Measurements Optional Setup Conc Measurement CM
You must select the concentration matrix to be used for measurement. Although the transmitter can store up to six concentration matrices, only one matrix can be used for measurement at any one time.
Procedure
Set Active Matrix to the matrix you want to use.
Configuration
Configuration and Use Manual 91
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166

5.3 Configure the batching application

5.3.1
Display Menu Configuration Process Measurement Batcher Global
ProLink III Device Tools Configuration Batcher Batch Settings
Field communicator Configure Manual Setup Inputs/Outputs Batcher Batcher

Configure global batching parameters

The global batching parameters apply to all batch presets and all batches.
Procedure
Set the flow source to the process variable that will be used to configure and measure the batch.
1.
Note
This field is named Batch Variable in the display.
Options Description
Mass Flow Rate The batch is measured by mass, using the current mass unit.
Volume Flow Rate The batch is measured by volume, using the current volume unit.
API Corrected Vol Flow The batch is measured using the API calculated correction factor.
GSV Flow (only if configured) The batch is measured using the gas standard volume flow rate.
CM Standard Vol Flow (when licensed and enabled)
The batch is measured using the standard volume flow rate at reference temperature.
CM Net Volume Flow (when licensed and enabled)
CM Net Mass Flow (when licensed and enabled)
FL Flow Rate The batch is measured using the Frequency Input.
The batch is measured using the net volume flow rate at reference temperature.
The batch is measured using the net mass flow rate.
Example
The current measurement unit for mass flow rate is g/sec. The batch will be configured and measured in g (grams).
Set the maximum target to the largest batch that will be allowed.
2.
Note
This field is called Preset Max Target with a field communicator.
Default: 999999999.0 kg or the equivalent value in the configured measurement unit
Range: Unlimited
If you try to set the batch target to a larger amount, the transmitter will reject the setting.
3.
Set the maximum batch time to the maximum number of seconds that a batch will be allowed to run.
If the batch does not reach its target before this period expires, the batch is halted automatically and an alert is posted.
92 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Default: 0 seconds
Range: 0 to 86,400 seconds (1 day)
If Maximum Fill Time is set to 0, the control is disabled and no maximum time is applied to batches.
4.
For two-stage batching, set the following parameters:
Parameter Value
Number of Stages Set to 2.
Setting this parameter to 1 disables two-stage batching.
Configure Presets by
% Target The values for Open Primary, Open
Secondary, Close Primary, and End Warning
are each configured as a percent of target.
Quantity The values for Open Primary and Open
Secondary are each configured as a quantity at
which the valve should open. The values for Close Primary and End Warning are each configured as a quantity that is subtracted from the target.

5.3.2 Configure batch presets

Display Menu Configuration Process Measurement Batcher Presets
ProLink III Device Tools Configuration Batcher Preset Configurations
Field communicator Configure Manual Setup Inputs/Outputs Batcher Setup Presets
A batch preset is a predefined group of batching parameters. You can define and save up to six batch presets.
When you run a batch, you must choose a preset. You can modify the batch target before you run the batch. All other settings are fixed.
Procedure
Select the preset number that you want to configure.
1.
2. Set the preset name to the name to be used for this preset.
The name can be up to eight characters in length. Valid characters include A-Z and 0-9.
3. Set the preset target to the size of the batch.
When the transmitter has measured the specified amount and applied the selected form of AOC, it closes the valve.
4. Optional: Set the preset end warning point to the point in the batch at which a Modbus coil will be set.
The end warning allows the transmitter to alert the operator to the end of batch. The operator can prepare for the end of the batch, e.g., by adjusting the flow rate to avoid overflow.
5. Enable or disable the preset as desired.
Configuration and Use Manual 93
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Note
This field is named Preset Status in the display.
Option Description
Enabled (On) The preset is available for use. You can select it to run a batch.
Disabled (Off) The preset is not available for use. You cannot select it to run a batch. You
can always enable it at a later time.
Preset 1 is always enabled and cannot be disabled.
Optional: For two-stage batching, set % Target or Quantity for the following parameters:
6.
Two-stage batching is a batch that is run on a gas and liquid mixture flowing through a pipeline.
% Target The values for Open Primary, Open Secondary, Close Primary, and End
Warning are each configured as a percent of target.
Quantity The values for Open Primary and Open Secondary are each configured as
a quantity at which the valve should open. The values for Close Primary and End Warning are each configured as a quantity that is subtracted from the target.
Parameter Description
End Warning The % target or quantity level that must be reached to alert the operator
that a batch is ending.
Open Primary The % target or quantity level that must be reached to open the primary
valve.
Open Secondary The % target or quantity level that must be reached to open the secondary
valve.
Close Primary The % target or quantity level that must be reached to close the primary
valve. When configuring by quantity, the amount is referenced to the end of the batch and not to the beginning of the batch.

5.3.3 Configure a DO for valve control

Display Menu Configuration Inputs/Outputs
ProLink III Device Tools Configuration I/O Channels
Field communicator Configure Manual Setup Inputs/Outputs
The transmitter runs a batch by opening and closing a valve. You can optionally configure a Discrete Output (DO) to send the open and close commands to the valve.
Prerequisites
Channel B, Channel C, or Channel D must be available for valve control.
The selected channel must be wired to the valve.
The selected channel and the valve must be powered by a power supply, not by the transmitter.
94 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure process measurement applications
MMI-20025166 October 2021
Procedure
1.
Verify the wiring between the selected channel and the valve.
2. Configure the selected channel to operate as a Discrete Output.
a) Set Channel B, Channel C, or Channel D to operate as a Discrete Output.
b) Set Power Source to External.
3. Configure the channel for valve control.
a) Do one of the following depending on which tool you are using:
Tool Steps
Display 1. Select the Discrete Output to be used for valve control.
Choose I/O Settings.
2.
ProLink III 1. Navigate to Device Tools Configuration I/O Outputs
Discrete Output.
2.
Select the Discrete Output to be used for valve control.
Field communicator 1. Select the Discrete Output to be used for valve control.
2.
Select Discrete Output x.
b) Set Source to the batch primary valve.
Set Polarity to match the wiring.
c)
The ON state of the Discrete Output must open the valve, and the OFF state must close the valve.
d) Set Fault Action so that the valve will be closed if a fault occurs.
The appropriate setting is either Upscale or Downscale. The setting to use for your installation depends on your valve type, your wiring, and your setting for Polarity.
4. Optional: For two-stage batching, assign one of the Discrete Outputs to the secondary valve.
You can also assign one of the Discrete Outputs to a pump when the batcher is configured for single stage or two stage.

5.3.4 Configure AOC

Display Menu Configuration Process Measurement Batcher AOC
ProLink III Device Tools Configuration Batcher Batching Options
Field communicator Configure Manual Setup Inputs/Outputs Batcher Batcher
Automatic Overshoot Compensation (AOC) adjusts the timing of the valve close command to minimize overshoot. The valve close command is sent before the target is reached, to compensate for the time required for the valve to close completely. AOC is applied to all presets and all batches.
There are three options for AOC:
Configuration and Use Manual 95
Configure process measurement applications Configuration and Use Manual
October 2021 MMI-20025166
Compensation Off
The transmitter sends the close command when the measured batch total reaches the configured target.
Fixed Compensation Value
AOC Algorithm
The transmitter sends the close command when the measured batch total equals the target minus the value configured for Fixed Overshoot Compensation. This value is configured in the current mass or volume unit, and applies to all presets.
The transmitter compares the actual measured amount of each batch to the batch target, and determines the adjustment by calibrating itself according to an internal algorithm. You can choose to stop AOC calibration when you are satisfied with the result, or you can set up rolling (continuous) AOC calibration.
Procedure
To turn compensation off:
Tool Steps
Display Set Compensation Mode to Off.
ProLink III Set AOC Compensation Mode to Compensation Off.
Field communicator Set Compensation Mode to No Compensation.
To use the Fixed Compensation Value option:
a)
Set the AOC compensation mode to fixed.
Tool Steps
Display Set Compensation Mode to Fixed.
ProLink III Set AOC Compensation Mode to Fixed Compensation Value.
Field communicator Set Compensation Mode to Fixed Value.
b) Set the fixed value or the fixed overshoot compensation to the quantity to be subtracted from the
batch target.
Important
Adjust the fixed value:
Whenever the process changes (e.g., flow rate, process fluid)
Whenever you change any configuration parameter that can affect flow measurement (e.g., damping)
To use the AOC Algorithm option:
Tool Steps
Display Set Compensation Mode to AOC.
ProLink III Set AOC Compensation Mode to AOC Algorithm.
Field communicator Set Compensation Mode to AOC.
96 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure advanced options for process measurement
MMI-20025166 October 2021

6 Configure advanced options for process measurement

6.1
Display Menu Configuration Process Measurement Response Time
ProLink III Device Tools Configuration Process Measurement Response Time
Field communicator Not available
Response Time controls the speed of various internal processes that are involved in retrieving electronic data from the sensor and converting it to process data.
Response Time affects all process and diagnostic variables.
Restriction
Response Time is configurable only if you are using the enhanced core processor. If you are using the standard core processor, Response Time is set to Low Filtering and cannot be changed.
Procedure
Set Response Time as desired.
Option Description
Normal Appropriate for typical applications.
High Filtering Slower response. Appropriate for applications with significant amount of entrained gas or

Configure Response Time

process noise.
Low Filtering Fastest response. Appropriate for proving or filling applications.
Service Do not select unless directed by Emerson personnel.

6.2 Detect and report two-phase flow

Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues. The transmitter provides two methods to detect and report or respond to two-phase flow.

6.2.1 Detect two-phase flow using density

Display
ProLink III
Field communicator
Configuration and Use Manual 97
Menu Configuration Process Measurement Density
Device Tools Configuration Process Measurement Density
Configure Manual Setup Measurements Density Slug Low Limit
Configure Manual Setup Measurements Density Slug High Limit
Configure Manual Setup Measurements Density Slug Duration
Configure advanced options for process measurement Configuration and Use Manual
October 2021 MMI-20025166
The transmitter can use line density data to detect two-phase flow (gas in a liquid process or liquid in a gas process). The density limits are user-specified. When two-phase flow is detected, an alert is posted.
Procedure
1.
Set Two-Phase Flow Low Limit to the lowest density value that is considered normal in your process.
Values below this will cause the transmitter to post a Process Aberration alert.
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the occurrence of two­phase flow alerts that are not significant to your process, set Two-Phase Flow Low Limit slightly below your expected lowest process density.
You must enter Two-Phase Flow Low Limit in g/cm³, even if you configured another unit for density measurement.
Default: 0 g/cm³
Range: 0 g/cm³ to the sensor limit
Set Two-Phase Flow High Limit to the highest density value that is considered normal in your process.
2.
Values above this will cause the transmitter to post a Process Aberration alert.
Tip
To reduce the occurrence of two-phase flow alerts that are not significant to your process, set Two­Phase Flow High Limit slightly above your expected highest process density.
You must enter Two-Phase Flow High Limit in g/cm³, even if you configured another unit for density measurement.
Default: 5 g/cm³
Range: 5 g/cm³ to the sensor limit
Set Two-Phase Flow Timeout to the number of seconds that the transmitter will wait for a two-phase
3.
flow condition to clear before posting the alert.
Default: 0 seconds, meaning that the alert will be posted immediately
Range: 0 to 60 seconds

6.2.2 Detect two-phase flow using sensor diagnostics

Display
ProLink III
Field communicator
Menu Configuration Inputs/Outputs Channel x I/O Settings Source
Device Tools Configuration I/O Outputs mA Output x
Configure Manual Setup Inputs/Outputs Channel x mA Output x mAOxSource
The transmitter always monitors sensor diagnostics and applies a two-phase flow algorithm. You can assign an mA Output to report the results of this calculation: single-phase flow, moderate two-phase flow, or severe two-phase flow. Severe two-phase flow can cause the meter to stop functioning.
Procedure
Set mA Output Source to Two-Phase Flow Detection.
98 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs
Configuration and Use Manual Configure advanced options for process measurement
MMI-20025166 October 2021
The signal from the mA Output indicates the current state of the process:
12 mA: Single-phase flow
16 mA: Moderate two-phase flow
20 mA: Severe two-phase flow
6.3
Display Menu Configuration Alert Setup Enhanced Events Flow Rate Switch
ProLink III Device Tools Configuration I/O Outputs Discrete Output Source Flow Switch Indication
Field communicator Configure Manual Setup Inputs/Outputs Channel x Discrete Output x Flow Switch
Flow Rate Switch is used to indicate that the flow rate has moved past a user-specified setpoint, in either direction. The flow rate switch is implemented with a user-configurable hysteresis.
Typically, a Discrete Output is assigned as the flow rate switch indicator. The Discrete Output can be wired to an external device such as a light or a horn.
Prerequisites
A channel must be configured as a Discrete Output, and the Discrete Output must be available for this use.
Procedure

Configure Flow Rate Switch

Set Discrete Output Source to Flow Switch, if you have not already done so.
1.
2. Set Flow Switch Variable to the flow variable that you want to use to control the flow rate switch.
3. Set Flow Switch Setpoint to the value at which the flow switch will be triggered (after Hysteresis is
applied).
Depending on the polarity of the Discrete Output:
If the flow rate is below this value, the Discrete Output is ON.
If the flow rate is above this value, the Discrete Output is OFF.
4. Set Hysteresis to the percentage of variation above and below the setpoint that will operate as a
deadband.
Hysteresis defines a range around the setpoint within which the flow rate switch will not change.
Default: 5%
Range: 0.1% to 10%
Example
If Flow Switch Setpoint = 100 g/sec and Hysteresis = 5%, and the first measured flow rate is above 100 g/sec, the discrete output is OFF. It will remain OFF unless the flow rate drops below 95 g/sec. If
this happens, the discrete output will turn ON, and remain ON until the flow rate rises above 105 g/sec. At this point it turns OFF and will remain OFF until the flow rate drops below 95 g/sec.
Related information
Configure a Discrete Output
Configuration and Use Manual 99
Configure advanced options for process measurement Configuration and Use Manual
October 2021 MMI-20025166

6.4 Configure events

An event occurs when the real-time value of a user-specified process variable moves past a user-defined setpoint. Events are used to provide notification of process changes or to perform specific transmitter actions if a process change occurs.
Your transmitter supports two event models:
Basic event model
Enhanced event model
Tip
Use enhanced events rather than basic events. Enhanced events can perform all the functions of basic events, and they add the following:
Defining events by range (In Range or Out of Range), in addition to High and Low
Triggering transmitter actions if an event occurs

6.4.1 Configure a basic event

Display Not available
ProLink III Device Tools Configuration Events Basic Events
Field communicator Not available
A basic event is used to provide notification of process changes. A basic event occurs (is ON) if the real-time value of a user-specified process variable moves above (HI) or below (LO) a user-defined setpoint. Event status can be queried via digital communications, and a Discrete Output can be configured to report event status.
You can define up to two basic events.
Procedure
Select the event that you want to configure.
1.
2. Assign a process variable to the event.
3. Specify Event Type.
Option Description
HI x > A
The event occurs when the value of the assigned process variable (x) is greater than the setpoint (Setpoint A), endpoint not included.
LO x < A
The event occurs when the value of the assigned process variable (x) is less than the setpoint (Setpoint A), endpoint not included.
4. Set a value for Setpoint A.
5. Optional: Configure a Discrete Output to switch states in response to the event status.
100 Micro Motion 5700 Transmitters with Configurable Inputs and Outputs