Configuration and Use Manual
MMI-20021714, Rev AB
April
Micro Motion® Model 2700 Transmitters with Analog Outputs
2013
Page 2

Safety messages

Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.

Micro Motion customer service

Email
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 318 495 555 New Zealand 099 128 804
Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3238 3677 Italy 8008 77334 China +86 21 2892 9000
Venezuela +58 26 1731 3446 Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudia Arabia 800 844 9564
UAE 800 0444 0684
Page 3

Contents

Contents
Part I Getting Started
Chapter 1 Before you begin ............................................................................................................. 3
1.1 About this manual .........................................................................................................................3
1.2 Transmitter model code ................................................................................................................3
1.3 Communications tools and protocols ............................................................................................3
1.4 Additional documentation and resources ......................................................................................4
Chapter 2 Quick start .......................................................................................................................7
2.1 Power up the transmitter ...............................................................................................................7
2.2 Check flowmeter status .................................................................................................................7
2.3 Make a startup connection to the transmitter ................................................................................9
2.4 Characterize the flowmeter (if required) ......................................................................................10
2.4.1 Sample sensor tags ....................................................................................................... 11
2.4.2 Flow calibration parameters (FCF, FT)
2.4.3 Density calibration parameters (D1, D2, K1, K2, FD, DT, TC) .............................................13
2.5 Verify mass flow measurement ....................................................................................................13
2.6 Verify the zero ............................................................................................................................. 14
2.6.1 Verify the zero using ProLink II ...................................................................................... 14
2.6.2 Verify the zero using ProLink III ..................................................................................... 15
2.6.3 Terminology used with zero verification and zero calibration ........................................16
........................................................................... 12
Part II Configuration and commissioning
Chapter 3 Introduction to configuration and commissioning ......................................................... 21
3.1 Configuration flowchart .............................................................................................................. 21
3.2 Default values and ranges ............................................................................................................23
3.3 Enable access to the off-line menu of the display ......................................................................... 23
3.4 Disable write-protection on the transmitter configuration .......................................................... 23
3.5 Restore the factory configuration ................................................................................................ 24
Chapter 4 Configure process measurement ................................................................................... 25
4.1 Configure mass flow measurement ............................................................................................. 25
4.1.1 Configure Mass Flow Measurement Unit ............................................................................. 25
4.1.2 Configure Flow Damping ..................................................................................................28
4.1.3 Configure Mass Flow Cutoff ..............................................................................................29
4.2 Configure volume flow measurement for liquid applications ....................................................... 31
4.2.1 Configure Volume Flow Type for liquid applications ..........................................................31
4.2.2 Configure Volume Flow Measurement Unit for liquid applications ........................................32
4.2.3 Configure Volume Flow Cutoff ...........................................................................................35
4.3 Configure gas standard volume (GSV) flow measurement ........................................................... 36
4.3.1 Configure Volume Flow Type for gas applications ............................................................. 37
4.3.2 Configure Standard Gas Density ........................................................................................37
4.3.3 Configure Gas Standard Volume Flow Measurement Unit ....................................................... 38
4.3.4 Configure Gas Standard Volume Flow Cutoff ....................................................................... 41
Configuration and Use Manual i
Page 4
Contents
4.4 Configure Flow Direction .................................................................................................................
4.4.1 Options for Flow Direction ................................................................................................43
4.5 Configure density measurement ................................................................................................. 48
4.5.1 Configure Density Measurement Unit ..................................................................................49
4.5.2 Configure slug flow parameters .................................................................................... 50
4.5.3 Configure Density Damping .............................................................................................. 52
4.5.4 Configure Density Cutoff .................................................................................................. 53
4.6 Configure temperature measurement .........................................................................................54
4.6.1 Configure Temperature Measurement Unit .......................................................................... 54
4.6.2 Configure Temperature Damping .......................................................................................55
4.7 Configure the petroleum measurement application .................................................................... 56
4.7.1 Configure petroleum measurement using ProLink II ..................................................... 56
4.7.2 Configure petroleum measurement using ProLink III .................................................... 57
4.7.3 Configure petroleum measurement using the Field Communicator ..............................59
4.7.4 API reference tables ...................................................................................................... 60
4.8 Configure the concentration measurement application .............................................................. 61
4.8.1 Configure concentration measurement using ProLink II ................................................62
4.8.2 Configure concentration measurement using ProLink III ............................................... 64
4.8.3 Configure concentration measurement using the Field Communicator ........................ 67
4.8.4 Standard matrices for the concentration measurement application ..............................69
4.8.5 Derived variables and calculated process variables ........................................................70
4.9 Configure pressure compensation ...............................................................................................72
4.9.1 Configure pressure compensation using ProLink II ........................................................ 72
4.9.2 Configure pressure compensation using ProLink III ....................................................... 74
4.9.3 Configure pressure compensation using the Field Communicator ................................ 76
4.9.4 Options for Pressure Measurement Unit ..............................................................................77
43
Chapter 5 Configure device options and preferences ..................................................................... 79
5.1 Configure the transmitter display ................................................................................................ 79
5.1.1 Configure the language used for the display ................................................................. 79
5.1.2 Configure the process variables shown on the display ...................................................80
5.1.3 Configure the precision of variables shown on the display .............................................82
5.1.4 Configure the refresh rate of data shown on the display ................................................82
5.1.5 Enable or disable automatic scrolling through the display variables .............................. 83
5.1.6 Enable or disable the display backlight ..........................................................................84
5.1.7 Enable or disable Status LED Blinking ................................................................................84
5.2 Enable or disable operator actions from the display ..................................................................... 85
5.2.1 Enable or disable Totalizer Start/Stop from the display ..................................................85
5.2.2 Enable or disable Totalizer Reset from the display ......................................................... 86
5.2.3 Enable or disable the Acknowledge All Alarms display command .................................. 87
5.3 Configure security for the display menus .................................................................................... 87
5.4 Configure response time parameters .......................................................................................... 89
5.4.1 Configure Update Rate .................................................................................................... 89
5.4.2 Configure Calculation Speed (Response Time) .................................................................... 91
5.5 Configure alarm handling ............................................................................................................ 92
5.5.1 Configure Fault Timeout ...................................................................................................92
5.5.2 Configure Status Alarm Severity ........................................................................................ 93
5.6 Configure informational parameters ........................................................................................... 97
5.6.1 Configure Descriptor ....................................................................................................... 97
5.6.2 Configure Message .........................................................................................................98
5.6.3 Configure Date ...............................................................................................................98
5.6.4 Configure Sensor Serial Number ....................................................................................... 99
ii Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 5
Contents
5.6.5 Configure Sensor Material ................................................................................................
5.6.6 Configure Sensor Liner Material ...................................................................................... 100
5.6.7 Configure Sensor Flange Type ........................................................................................100
99
Chapter 6 Integrate the meter with the control system ................................................................103
6.1 Configure the transmitter channels ........................................................................................... 103
6.2 Configure the mA output .......................................................................................................... 104
6.2.1 Configure mA Output Process Variable ............................................................................. 104
6.2.2 Configure Lower Range Value (LRV) and Upper Range Value (URV) ....................................106
6.2.3 Configure AO Cutoff ......................................................................................................108
6.2.4 Configure Added Damping ............................................................................................. 110
6.2.5 Configure mA Output Fault Action and mA Output Fault Level ...............................................111
6.3 Configure the frequency output ................................................................................................ 112
6.3.1 Configure Frequency Output Process Variable ................................................................... 113
6.3.2 Configure Frequency Output Polarity ................................................................................ 114
6.3.3 Configure Frequency Output Scaling Method ..................................................................... 115
6.3.4 Configure Frequency Output Maximum Pulse Width ............................................................ 117
6.3.5 Configure Frequency Output Fault Action and Frequency Output Fault Level ........................... 118
6.4 Configure the discrete output ................................................................................................... 119
6.4.1 Configure Discrete Output Source ....................................................................................120
6.4.2 Configure Discrete Output Polarity ................................................................................... 122
6.4.3 Configure Discrete Output Fault Action ............................................................................. 124
6.5 Configure events ....................................................................................................................... 125
6.5.1 Configure a basic event ...............................................................................................125
6.5.2 Configure an enhanced event ..................................................................................... 126
6.6 Configure digital communications ............................................................................................ 128
6.6.1 Configure HART/Bell 202 communications ................................................................. 129
6.6.2 Configure HART/RS-485 communications .................................................................. 134
6.6.3 Configure Modbus/RS-485 communications .............................................................. 135
6.6.4 Configure Digital Communications Fault Action ...................................................................137
Chapter 7 Completing the configuration ......................................................................................139
7.1 Test or tune the system using sensor simulation ........................................................................139
7.1.1 Sensor simulation ....................................................................................................... 140
7.2 Back up transmitter configuration ............................................................................................. 141
7.3 Enable write-protection on the transmitter configuration ......................................................... 142
Chapter 8 Set up the Weights & Measures application ................................................................. 143
8.1 Weights & Measures application ................................................................................................143
8.2 Set up the Weights & Measures application using ProLink II ....................................................... 144
8.3 Set up the Weights & Measures application using ProLink III ...................................................... 147
Part III Operations, maintenance, and troubleshooting
Chapter 9 Transmitter operation ................................................................................................. 153
9.1 Record the process variables ..................................................................................................... 153
9.2 View process variables ...............................................................................................................154
9.2.1 View process variables using the display (standard option) ......................................... 154
9.2.2 View process variables using the Chinese-language display ........................................ 155
9.2.3 View process variables using ProLink III ....................................................................... 156
9.3 View transmitter status using the status LED ............................................................................. 156
9.4 View and acknowledge status alarms ........................................................................................ 157
Configuration and Use Manual iii
Page 6
Contents
9.4.1 View and acknowledge alarms using the display (standard option) .............................157
9.4.2 View and acknowledge alarms using the Chinese-language display ............................ 160
9.4.3 View and acknowledge alarms using ProLink II ............................................................162
9.4.4 View and acknowledge alerts using ProLink III .............................................................162
9.4.5 View alarms using the Field Communicator ................................................................ 163
9.4.6 Alarm data in transmitter memory ..............................................................................163
9.5 Read totalizer and inventory values ........................................................................................... 164
9.6 Start and stop totalizers and inventories ....................................................................................165
9.6.1 Start and stop totalizers and inventories using the display (standard option) .............. 165
9.7 Reset totalizers ..........................................................................................................................166
9.7.1 Reset totalizers using the display (standard option) .................................................... 167
9.8 Reset inventories ....................................................................................................................... 168
Chapter 10 Operate the transmitter with the Weights & Measures application .............................. 169
10.1 Operate the transmitter when the Weights & Measures application is installed ......................... 169
10.1.1 Approved methods to read or obtain process data ......................................................170
10.1.2 Large totalizer values on the display (OIML applications only) ..................................... 170
10.1.3 Effect of the Weights & Measures application on process measurement and outputs
10.1.4 Effect of the Weights & Measures application on operation and maintenance
functions .................................................................................................................... 172
10.2 Switch between secured and unsecured mode .......................................................................... 174
10.2.1 Switch between secured and unsecured mode using ProLink II ................................... 175
10.2.2 Switch between secured and unsecured mode using ProLink III .................................. 175
10.2.3 Switch between secured and unsecured mode using the switching utility ...................176
10.3 Clear Status Alarm A027: Security Breach .................................................................................. 176
10.4 Replacing the core processor in a Weights & Measures installation ............................................ 177
.......................................................................................................................171
Chapter 11 Measurement support ................................................................................................. 179
11.1 Options for measurement support ............................................................................................ 179
11.2 Use Smart Meter Verification .....................................................................................................179
11.2.1 Smart Meter Verification requirements .......................................................................180
11.2.2 Smart Meter Verification test preparation ...................................................................180
11.2.3 Run Smart Meter Verification ......................................................................................181
11.2.4 View test data .............................................................................................................187
11.2.5 Schedule automatic execution of the Smart Meter Verification test ............................193
11.3 Zero the flowmeter ................................................................................................................... 198
11.3.1 Zero the flowmeter using the display (standard option) ..............................................199
11.3.2 Zero the flowmeter using the Chinese-language display ............................................. 200
11.3.3 Zero the flowmeter using ProLink II .............................................................................201
11.3.4 Zero the flowmeter using ProLink III ............................................................................202
11.3.5 Zero the flowmeter using the Field Communicator ..................................................... 203
11.4 Validate the meter .....................................................................................................................205
11.4.1 Alternate method for calculating the meter factor for volume flow .............................206
11.5 Perform a (standard) D1 and D2 density calibration ...................................................................207
11.5.1 Perform a D1 and D2 density calibration using ProLink II ............................................. 207
11.5.2 Perform a D1 and D2 density calibration using ProLink III ............................................ 209
11.5.3 Perform a D1 and D2 density calibration using the Field Communicator ..................... 210
11.6 Perform a D3 and D4 density calibration (T-Series sensors only) ................................................ 212
11.6.1 Perform a D3 or D3 and D4 density calibration using ProLink II ....................................212
11.6.2 Perform a D3 or D3 and D4 density calibration using ProLink III ...................................213
11.6.3 Perform a D3 or D3 and D4 density calibration using the Field Communicator ............214
11.7 Perform temperature calibration ............................................................................................... 216
iv Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 7
Contents
11.7.1 Perform temperature calibration using ProLink II ........................................................ 216
11.7.2 Perform temperature calibration using ProLink III ....................................................... 217
Chapter 12 Troubleshooting .......................................................................................................... 219
12.1 Status LED states ....................................................................................................................... 220
12.2 Status alarms .............................................................................................................................220
12.3 Flow measurement problems .................................................................................................... 232
12.4 Density measurement problems ............................................................................................... 234
12.5 Temperature measurement problems .......................................................................................235
12.6 Milliamp output problems ......................................................................................................... 236
12.7 Frequency output problems ...................................................................................................... 237
12.8 Use sensor simulation for troubleshooting ................................................................................ 238
12.9 Check power supply wiring ........................................................................................................ 238
12.10 Check sensor-to-transmitter wiring ........................................................................................... 239
12.11 Check grounding ....................................................................................................................... 239
12.12 Perform loop tests ..................................................................................................................... 240
12.12.1 Perform loop tests using the display (standard option) ............................................... 240
12.12.2 Perform loop tests using the Chinese-language display ...............................................241
12.12.3 Perform loop tests using ProLink II .............................................................................. 243
12.12.4 Perform loop tests using ProLink III ............................................................................. 244
12.12.5 Perform loop tests using the Field Communicator ...................................................... 246
12.13 Trim mA outputs ....................................................................................................................... 247
12.13.1 Trim mA outputs using ProLink II ................................................................................ 247
12.13.2 Trim mA outputs using ProLink III ................................................................................248
12.13.3 Trim mA outputs using the Field Communicator .........................................................249
12.14 Check the HART communication loop ....................................................................................... 249
12.15 Check HART Address and Loop Current Mode ..................................................................................
12.16 Check HART burst mode ............................................................................................................250
12.17 Check Lower Range Value and Upper Range Value ........................................................................... 251
12.18 Check mA Output Fault Action ........................................................................................................ 251
12.19 Check for radio frequency interference (RFI) .............................................................................. 251
12.20 Check Frequency Output Maximum Pulse Width ................................................................................. 251
12.21 Check Frequency Output Scaling Method .......................................................................................... 252
12.22 Check Frequency Output Fault Action ............................................................................................... 252
12.23 Check Flow Direction .................................................................................................................... 252
12.24 Check the cutoffs ...................................................................................................................... 252
12.25 Check for slug flow (two-phase flow) ......................................................................................... 253
12.26 Check the drive gain .................................................................................................................. 253
12.26.1 Collect drive gain data ................................................................................................ 255
12.27 Check the pickoff voltage .......................................................................................................... 255
12.27.1 Collect pickoff voltage data ........................................................................................ 256
12.28 Check for electrical shorts ..........................................................................................................256
12.28.1 Check the sensor coils .................................................................................................256
12.29 Check the core processor LED ....................................................................................................258
12.29.1 Core processor LED states ........................................................................................... 260
12.30 Perform a core processor resistance test ................................................................................... 262
250
Appendices and reference
Appendix A Using the standard transmitter display ........................................................................ 265
A.1 Components of the transmitter interface .................................................................................. 265
A.2 Use the optical switches ............................................................................................................ 266
Configuration and Use Manual v
Page 8
Contents
A.3 Access and use the display menu system ...................................................................................267
A.3.1 Enter a floating-point value using the display .............................................................. 268
A.4 Display codes for process variables ............................................................................................271
A.5 Codes and abbreviations used in display menus ........................................................................ 272
A.6 Menu maps for the transmitter display ...................................................................................... 276
Appendix B Using the Chinese-language display ............................................................................. 287
B.1 Components of the transmitter interface .................................................................................. 287
B.2 Use the optical switches ............................................................................................................ 288
B.3 Access and use the display menu system ................................................................................... 289
B.3.1 Enter a floating-point value using the display .............................................................. 290
B.4 Menu maps for the transmitter display ...................................................................................... 294
Appendix C Using ProLink II with the transmitter ........................................................................... 307
C.1 Basic information about ProLink II ..............................................................................................307
C.2 Connect with ProLink II ..............................................................................................................308
C.2.1 ProLink II connection types ......................................................................................... 308
C.2.2 Make a service port connection ...................................................................................309
C.2.3 Make a HART/Bell 202 connection .............................................................................. 310
C.2.4 Make a HART/RS-485 connection ................................................................................315
C.2.5 Make a Modbus/RS-485 connection ............................................................................318
C.3 Menu maps for ProLink II ........................................................................................................... 321
Appendix D Using ProLink III with the transmitter ...........................................................................331
D.1 Basic information about ProLink III .............................................................................................331
D.2 Connect with ProLink III ............................................................................................................. 332
D.2.1 ProLink III connection types ........................................................................................ 332
D.2.2 Make a service port connection ...................................................................................333
D.2.3 Make a HART/Bell 202 connection .............................................................................. 334
D.2.4 Make a HART/RS-485 connection ................................................................................339
D.2.5 Make a Modbus/RS-485 connection ............................................................................342
D.3 Menu maps for ProLink III .......................................................................................................... 345
Appendix E Using the Field Communicator with the transmitter .....................................................355
E.1 Basic information about the Field Communicator ......................................................................355
E.2 Connect with the Field Communicator ...................................................................................... 356
E.3 Menu maps for the Field Communicator ....................................................................................359
Appendix F Default values and ranges ............................................................................................ 375
F.1 Default values and ranges ..........................................................................................................375
Appendix G Transmitter components and installation wiring ......................................................... 381
G.1 Installation types ....................................................................................................................... 381
G.2 Power supply terminals and ground ...........................................................................................385
G.3 Input/output (I/O) wiring terminals ........................................................................................... 386
Appendix H NE 53 history ............................................................................................................... 387
H.1 NE 53 history ............................................................................................................................. 387
Index ................................................................................................................................................393
vi Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 9
Part I

Getting Started

Chapters covered in this part:
Before you begin
Quick start
Getting Started
Configuration and Use Manual 1
Page 10
Getting Started
2 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 11
1 Before you begin
Topics covered in this chapter:

About this manual

Transmitter model code

Communications tools and protocols

Additional documentation and resources
1.1 About this manual
This manual provides information to help you configure, commission, use, maintain, and troubleshoot the Micro Motion transmitter.
Important
This manual assumes that the transmitter has been installed correctly and completely, according to the instructions in the transmitter installation manual, and that the installation complies with all applicable safety requirements.

Before you begin

1.2 Transmitter model code
Your transmitter can be identified by the model number on the transmitter tag.
The transmitter has a model number of the following form:
2700(I/R/C/B)**A******
I Integral mount
R 4-wire remote-mount
C 9-wire remote-mount
B Remote core processor with remote transmitter
A Analog outputs option board
1.3 Communications tools and protocols
You can use several different communications tools and protocols to interface with the transmitter. You may use different tools in different locations or for different tasks.
Configuration and Use Manual 3
Page 12
Before you begin
Communications tools, protocols, and related informationTable 1-1:
Communica­tions tool Supported protocols Scope In this manual For more information
Display (stand­ard)
Chinese-lan­guage display
ProLink II HART/RS-485
ProLink III HART/RS-485
Field Commu­nicator
Not applicable Basic configuration and
commissioning
Not applicable Basic configuration and
commissioning
(1)
HART/Bell 202
Modbus/RS-485
Service port
(1)
HART/Bell 202
Modbus/RS-485
Service port
HART/Bell 202 Complete configuration
Complete configuration and commissioning
Complete configuration and commissioning
and commissioning
Complete user informa­tion. See Appendix A.
Complete user informa­tion. See Appendix B.
Basic user information. See Appendix C.
Basic user information. See Appendix D.
Basic user information. See Appendix E.
Not applicable
Not applicable
User manual
Installed with soft-
ware
On Micro Motion
user documentation CD
On Micro Motion
web site (www.mi-
cromotion.com
User manual
Installed with soft-
ware
On Micro Motion
user documentation CD
On Micro Motion
web site (www.mi-
cromotion.com
User manual on Micro Motion web site (www.micromo-
tion.com
Tip
You may be able to use other communications tools from Emerson Process Management, such as AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. The AMS interface is similar to the ProLink II interface. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at www.micromotion.com.

1.4 Additional documentation and resources

Micro Motion provides additional documentation to support the installation and operation of the transmitter.
(1) Devices with the Chinese-language display do not support HART/RS-485.
4 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 13
Before you begin
Additional documentation and resourcesTable 1-2:
Topic Document
Sensor Sensor documentation
Transmitter installation
Hazardous area installation See the approval documentation shipped with the transmitter, or
download the appropriate documentation from the Micro Motion web site at www.micromotion.com.
All documentation resources are available on the Micro Motion web site at
www.micromotion.com or on the Micro Motion user documentation CD.
Configuration and Use Manual 5
Page 14
Before you begin
6 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 15
2 Quick start
Topics covered in this chapter:

Power up the transmitter

Check flowmeter status

Make a startup connection to the transmitter
Characterize the flowmeter (if required)
Verify mass flow measurement
Verify the zero
2.1 Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
Ensure that all transmitter and sensor covers and seals are closed.
1.

Quick start

CAUTION!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion.
2. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. During this period, Alarm 009 is active. The diagnostic routines should complete in approximately 30 seconds. For transmitters with a display, the status LED will turn green and begin to flash when the startup diagnostics are complete. If the status LED exhibits different behavior, an alarm condition is present.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power is been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.
2.2 Check flowmeter status
Check the flowmeter for any error conditions that require user action or that affect measurement accuracy.
Configuration and Use Manual 7
Page 16
Quick start
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alarm A009 is active. This alarm should clear automatically when the power-up sequence is complete.
2.
Check the status LED on the transmitter.
Transmitter status reported by status LEDTable 2-1:
LED state Description Recommendation
Green No alarms are active. Continue with configuration or process meas-
urement.
(3)
(1)
(2)
No alarms are active. One or more previously active alarms have not been acknowledged.
and have been acknowledged.
One or more low-severity alarms are active and have not been acknowledged.
and have been acknowledged.
One or more high-severity alarms are active and have not been acknowledged.
Continue with configuration or process meas­urement. If you choose, you can acknowledge the alarms.
A low-severity alarm condition does not affect measurement accuracy or output behavior. You can continue with configuration or proc­ess measurement. If you choose, you can iden­tify and resolve the alarm condition.
A low-severity alarm condition does not affect measurement accuracy or output behavior. You can continue with configuration or proc­ess measurement. If you choose, you can iden­tify and resolve the alarm condition. You may also acknowledge the alarm.
A high-severity alarm condition affects meas­urement accuracy and output behavior. Re­solve the alarm condition before continuing.
A high-severity alarm condition affects meas­urement accuracy and output behavior. Re­solve the alarm condition before continuing. You may also acknowledge the alarm.
Flashing green
Yellow One or more low-severity alarms are active,
Flashing yellow
Red One or more high-severity alarms are active,
Flashing red
Postrequisites
For information on viewing the list of active alarms, see Section 9.4.
For information on individual alarms and suggested resolutions, see Section 12.2.
(1) If Status LED Blinking is disabled, the LED will show solid green rather than flashing. (2)
If Status LED Blinking is disabled, the LED will show solid yellow rather than flashing.
(3) If Status LED Blinking is disabled, the LED will show solid red rather than flashing.
8 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 17
Quick start

2.3 Make a startup connection to the transmitter

For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter. Follow this procedure to make your first connection to the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix. Use the default communications parameters shown in the appendix.
Communications tool Connection type to use Instructions
ProLink II HART/RS-485
ProLink III HART/RS-485
Field Communicator HART Appendix E
Postrequisites
(Optional) Change the communications parameters to site-specific values.
(4)
(4)
Appendix C
Appendix D
To change the communications parameters using ProLink II:
To change the protocol, baud rate, parity, or stop bits, choose ProLink > Configuration >
RS-485.
To change the address, choose ProLink > Configuration > Device.
To change the communications parameters using ProLink III, choose Device Tools > Configuration > Communications.
To change the communications parameters using the Field Communicator, choose On-Line Menu > Configure > Manual Setup > Inputs/Outputs > Communications.
Important
If you are changing communications parameters for the connection type that you are using, you will lose the connection when you write the parameters to the transmitter. Reconnect using the new parameters.
(4) Devices with the Chinese-language display do not support HART/RS-485. The default connection to use for these devices is Modbus/RS-485.
Configuration and Use Manual 9
Page 18
Quick start

2.4 Characterize the flowmeter (if required)

Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Device > Sensor Type
ProLink III Device Tools > Calibration Data
Field Communicator Configure > Manual Setup > Characterize
Overview
Characterizing the flowmeter adjusts your transmitter to match the unique traits of the sensor it is paired with. The characterization parameters (also called calibration parameters) describe the sensor’s sensitivity to flow, density, and temperature. Depending on your sensor type, different parameters are required. Values for your sensor are provided by Micro Motion on the sensor tag or the calibration certificate.
Offline Maintain > Configuration > Calibrate Sensor
ProLink > Configuration > Flow
ProLink > Configuration > Density
ProLink > Configuration > T Series
Tip
If your flowmeter was ordered as a unit, it has already been characterized at the factory. However, you should still verify the characterization parameters.
Procedure
1.
Specify Sensor Type.
Straight-tube (T-Series)
(5)
Curved-tube (all sensors except T-Series)
2. Set the flow characterization parameters. Be sure to include all decimal points.
For straight-tube sensors, set FCF (Flow Cal or Flow Calibration Factor), FTG, and
(5)
FFQ.
For curved-tube sensors, set Flow Cal (Flow Calibration Factor).
3. Set the density characterization parameters.
For straight-tube sensors, set D1, D2, DT, DTG, K1, K2, FD, DFQ1, and DFQ2.
(5)
For curved-tube sensors, set D1, D2, TC, K1, K2, and FD. (TC is sometimes shown
as DT.)
(5) Devices with the Chinese-language display do not support T-Series sensors.
10 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 19
2.4.1 Sample sensor tags
Tag on older curved-tube sensors (all sensors except T-Series)Figure 2-1:
Quick start
Tag on newer curved-tube sensors (all sensors except T-Series)Figure 2-2:
Configuration and Use Manual 11
Page 20
Quick start
Tag on older straight-tube sensor (T-Series)Figure 2-3:
Tag on newer straight-tube sensor (T-Series)Figure 2-4:
2.4.2 Flow calibration parameters (FCF, FT)
Two separate values are used to describe flow calibration: a 6-character FCF value and a 4­character FT value. They are provided on the sensor tag.
Both values contain decimal points. During characterization, these may be entered as two values or as a single 10-character string. The 10-character string is called either Flowcal or FCF.
If your sensor tag shows the FCF and the FT values separately and you need to enter a single value, concatenate the two values to form the single parameter value.
If your sensor tag shows a concatenated Flowcal or FCF value and you need to enter the FCF and the FT values separately, split the concatenated value:
FCF = The first 6 characters, including the decimal point
FT = The last 4 characters, including the decimal point
12 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 21
Quick start
Example: Concatenating FCF and FT
FCF = x.xxxx FT = y.yy Flow calibration parameter: x.xxxxy.yy
Example: Splitting the concatenated Flowcal or FCF value
Flow calibration parameter: x.xxxxy.yy FCF = x.xxxx FT = y.yy
2.4.3 Density calibration parameters (D1, D2, K1, K2, FD, DT, TC)
Density calibration parameters are typically on the sensor tag and the calibration certificate.
If your sensor tag does not show a D1 or D2 value:
For D1, enter the Dens A or D1 value from the calibration certificate. This value is the
line-condition density of the low-density calibration fluid. Micro Motion uses air. If you cannot find a Dens A or D1 value, enter 0.001 g/cm3.
For D2, enter the Dens B or D2 value from the calibration certificate. This value is the
line-condition density of the high-density calibration fluid. Micro Motion uses water. If you cannot find a Dens B or D2 value, enter 0.998 g/cm3.
If your sensor tag does not show a K1 or K2 value:
For K1, enter the first 5 digits of the density calibration factor. In the sample tag, this
value is shown as 12500.
For K2, enter the second 5 digits of the density calibration factor. In the sample tag,
this value is shown as 14286.
If your sensor does not show an FD value, contact Micro Motion customer service.
If your sensor tag does not show a DT or TC value, enter the last 3 digits of the density calibration factor. In the sample tag, this value is shown as 4.44.

2.5 Verify mass flow measurement

Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method.
Read the value for Mass Flow Rate on the transmitter display.
Connect to the transmitter with ProLink II and read the value for Mass Flow Rate in the
Process Variables window (ProLink > Process Variables).
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in
the Process Variables panel.
Configuration and Use Manual 13
Page 22
Quick start
Connect to the transmitter with the Field Communicator and read the value for Mass
Flow Rate in the Process Variables menu (On-Line Menu > Overview > Primary Purpose Variables).
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues. See
Section 12.3.

2.6 Verify the zero

Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
The zero verification procedure analyzes the Live Zero value under conditions of zero flow, and compares it to the Zero Stability range for the sensor. If the average Live Zero value is within a reasonable range, the zero value stored in the transmitter is valid. Performing a field calibration will not improve measurement accuracy.
2.6.1 Verify the zero using ProLink II
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless one of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Prerequisites
ProLink II v2.94 or later
Important
Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem, then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a low­severity alarm is active.
Procedure
1.
Prepare the flowmeter:
a. Allow the flowmeter to warm up for at least 20 minutes after applying power.
14 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 23
Quick start
b. Run the process fluid through the sensor until the sensor temperature reaches
the normal process operating temperature.
c.
Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if available.
d. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
2. Choose ProLink > Calibration > Zero Verification and Calibration > Verify Zero and wait until
the procedure completes.
3. If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Repeat the zero verification procedure.
d. If it fails again, zero the flowmeter.
For instructions on zeroing the flowmeter, see Zero the flowmeter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
2.6.2 Verify the zero using ProLink III
Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless one of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Prerequisites
ProLink III v1.0 with Patch Build 31, or a later release
Important
Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem, then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a low­severity alarm is active.
Procedure
1.
Prepare the flowmeter:
a. Allow the flowmeter to warm up for at least 20 minutes after applying power.
Configuration and Use Manual 15
Page 24
Quick start
b. Run the process fluid through the sensor until the sensor temperature reaches
the normal process operating temperature.
c.
Stop flow through the sensor by shutting the downstream valve, and then the upstream valve if available.
d. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
2. Choose Device Tools > Device Calibration > Zero Verification and Calibration > Verify Zero and
wait until the procedure completes.
3. If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Repeat the zero verification procedure.
d. If it fails again, zero the flowmeter.
For instructions on zeroing the flowmeter, see Zero the flowmeter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
2.6.3 Terminology used with zero verification and zero calibration
Terminology used with zero verification and zero calibrationTable 2-2:
Term Definition
Zero In general, the offset required to synchronize the left pickoff and the right pickoff under
conditions of zero flow. Unit = microseconds.
Factory Zero The zero value obtained at the factory, under laboratory conditions.
Field Zero The zero value obtained by performing a zero calibration outside the factory.
Prior Zero The zero value stored in the transmitter at the time a field zero calibration is begun. May
be the factory zero or a previous field zero.
Manual Zero The zero value stored in the transmitter, typically obtained from a zero calibration proce-
dure. It may also be configured manually. Also called “mechanical zero” or “stored zero.”
Live Zero The real-time bidirectional mass flow rate with no flow damping or mass flow cutoff ap-
plied. An adaptive damping value is applied only when the mass flow rate changes dra­matically over a very short interval. Unit = configured mass flow measurement unit.
Zero Stability A laboratory-derived value used to calculate the expected accuracy for a sensor. Under
laboratory conditions at zero flow, the average flow rate is expected to fall within the range defined by the Zero Stability value (0 ± Zero Stability). Each sensor size and model has a unique Zero Stability value. Statistically, 95% of all data points should fall within the range defined by the Zero Stability value.
Zero Calibration The procedure used to determine the zero value.
16 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 25
Quick start
Terminology used with zero verification and zero calibration (continued)Table 2-2:
Term Definition
Zero Time The time period over which the Zero Calibration procedure is performed. Unit = seconds.
Field Verification Zero A 3-minute running average of the Live Zero value, calculated by the transmitter. Unit =
configured mass flow measurement unit.
Zero Verification A procedure used to evaluate the stored zero and determine whether or not a field zero
can improve measurement accuracy.
Configuration and Use Manual 17
Page 26
Quick start
18 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 27

Configuration and commissioning

Part II
Configuration and commissioning
Chapters covered in this part:
Introduction to configuration and commissioning
Configure process measurement
Configure device options and preferences
Integrate the meter with the control system
Completing the configuration
Set up the Weights & Measures application
Configuration and Use Manual 19
Page 28
Configuration and commissioning
20 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 29

Introduction to configuration and commissioning

3 Introduction to configuration and
commissioning
Topics covered in this chapter:

Configuration flowchart

Default values and ranges
Enable access to the off-line menu of the display
Disable write-protection on the transmitter configuration
Restore the factory configuration
3.1 Configuration flowchart
Use the following flowchart as a general guide to the configuration and commissioning process.
Some options may not apply to your installation. Detailed information is provided in the remainder of this manual. If you are using the Weights & Measures application, additional configuration and setup are required.
Configuration and Use Manual 21
Page 30
Introduction to configuration and commissioning
Configuration flowchartFigure 3-1:
Configure process measurement
Configure device options and preferences
Test and move to production
Configure mass flow
measurement
Configure volume flow
meaurement
Volume flow type
Liquid
Configure flow direction
Configure density
measurement
Configure temperature
measurement
Gas
Define gas properties
Configure display
parameters
Configure fault handling
parameters
Configure sensor
parameters
Configure device
parameters
Integrate device with control system
Configure the channel(s)
Configure the mA
output(s)
Test or tune transmitter using sensor simulation
Back up transmitter
configuration
Enable write-protection on
transmitter configuration
Done
Configure petroleum
measurement (API)
application (if available)
Configure concentration
measurement application
(if available)
Configure pressure
compensation (optional)
Configure the frequency
output(s)
Configure the discrete
output(s)
Configure events
Configure digital communications
22 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 31
Introduction to configuration and commissioning

3.2 Default values and ranges

See Section F.1 to view the default values and ranges for the most commonly used parameters.

3.3 Enable access to the off-line menu of the display

Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features
Offline Maintain > Configuration > Display
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must enable it if you want to use the display to configure the transmitter.
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from another tool.

3.4 Disable write-protection on the transmitter configuration

Display (standard) OFF-LINE MAINT > CONFG > LOCK
Chinese-language display
ProLink II ProLink > Configuration > Device > Enable Write Protection
ProLink III Device Tools > Configuration > Write-Protection
Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Write Protect
Offline Maintain > Configuration > Lock
Overview
If the transmitter is write-protected, the configuration is locked and you must unlock it before you can change any configuration parameters. By default, the transmitter is not write-protected.
Configuration and Use Manual 23
Page 32
Introduction to configuration and commissioning
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent normal operational use. You can always disable write-protection, perform any required configuration changes, then re-enable write-protection.

3.5 Restore the factory configuration

Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Device > Restore Factory Configuration
ProLink III Device Tools > Configuration Transfer > Restore Factory Configuration
Field Communicator Not available
Overview
Not available
Restoring the factory configuration returns the transmitter to a known operational configuration. This may be useful if you experience problems during configuration.
Tip
Restoring the factory configuration is not a common action. You may want to contact Micro Motion to see if there is a preferred method to resolve any issues.
24 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 33

Configure process measurement

4 Configure process measurement
Topics covered in this chapter:

Configure mass flow measurement

Configure volume flow measurement for liquid applications
Configure gas standard volume (GSV) flow measurement
Configure Flow Direction
Configure density measurement
Configure temperature measurement
Configure the petroleum measurement application
Configure the concentration measurement application
Configure pressure compensation
4.1 Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported.
The mass flow measurement parameters include:
Mass Flow Measurement Unit
Flow Damping
Mass Flow Cutoff
4.1.1 Configure Mass Flow Measurement Unit
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > UNITS > MASS
Chinese-language display
ProLink II ProLink > Configuration > Flow > Mass Flow Units
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Unit
Overview
Offline Maintain > Configuration > Units > Mass Flow Rate
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow
rate. The unit used for mass total and mass inventory is derived from this unit.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
Configuration and Use Manual 25
Page 34
Configure process measurement
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may
use different labels for the units.
Options for Mass Flow Measurement UnitTable 4-1:
Label
Display (stand-
Unit description
Grams per second G/S g/sec g/sec g/sec g/s
Grams per minute G/MIN g/min g/min g/min g/min
Grams per hour G/H g/hr g/hr g/hr g/h
Kilograms per second KG/S kg/sec kg/sec kg/sec kg/s
Kilograms per minute KG/MIN kg/min kg/min kg/min kg/min
Kilograms per hour KG/H kg/hr kg/hr kg/hr kg/h
Kilograms per day KG/D kg/day kg/day kg/day kg/d
Metric tons per minute T/MIN mTon/min mTon/min mTon/min MetTon/min
Metric tons per hour T/H mTon/hr mTon/hr mTon/hr MetTon/h
Metric tons per day T/D mTon/day mTon/day mTon/day MetTon/d
Pounds per second LB/S lbs/sec lbs/sec lbs/sec lb/s
Pounds per minute LB/MIN lbs/min lbs/min lbs/min lb/min
Pounds per hour LB/H lbs/hr lbs/hr lbs/hr lb/h
Pounds per day LB/D lbs/day lbs/day lbs/day lb/d
Short tons (2000 pounds) per minute
Short tons (2000 pounds) per hour
Short tons (2000 pounds) per day
Long tons (2240 pounds) per hour
Long tons (2240 pounds) per day
Special unit SPECL Special special special Spcl
ard)
ST/MIN sTon/min sTon/min sTon/min STon/min
ST/H sTon/hr sTon/hr sTon/hr STon/h
ST/D sTon/day sTon/day sTon/day STon/d
LT/H lTon/hr lTon/hr lTon/hr LTon/h
LT/D lTon/day lTon/day lTon/day LTon/d
Chinese-lan­guage display
ProLink II ProLink III Field Commu-
nicator
26 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 35
Configure process measurement
Define a special measurement unit for mass flow
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Special Units
ProLink III Device Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Mass Special Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Although you cannot define a special measurement unit using the display (standard option), you can use the standard display to select an existing special measurement unit, and to view process data using the special measurement unit.
Offline Maintain > Configuration > Units > Special Mass Flow
Procedure
1.
Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Mass Flow Conversion Factor as follows:
a. x base units = y special units
b. Mass Flow Conversion Factor = x/y
4. Enter Mass Flow Conversion Factor.
5. Set Mass Flow Label to the name you want to use for the mass flow unit.
6. Set Mass Total Label to the name you want to use for the mass total and mass
inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
1. Set Base Mass Unit to Pounds (lb).
2. Set Base Time Unit to Seconds (sec).
Configuration and Use Manual 27
Page 36
Configure process measurement
3. Calculate Mass Flow Conversion Factor:
a.
1 lb/sec = 16 oz/sec
b. Mass Flow Conversion Factor = 1/16 = 0.0625
4. Set Mass Flow Conversion Factor to 0.0625.
5. Set Mass Flow Label to oz/sec.
6. Set Mass Total Label to oz.
4.1.2 Configure Flow Damping
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Flow > Flow Damp
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Flow Damping
Offline Maintain > Configuration > Damping > Mass Flow Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes
in the reported process variable. At the end of the interval, the reported process variable will reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
The default value is 0.8 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table.
Core processor type Update Rate setting Flow Damping range
Standard Normal 0 to 51.2 seconds
Special 0 to 10.24 seconds
Enhanced Not applicable 0 to 51.2 seconds
Tips
A high damping value makes the process variable appear smoother because the reported value
changes slowly.
A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
The combination of a high damping value and rapid, large changes in flow rate can result in
increased measurement error.
28 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 37
Configure process measurement
Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
For gas applications, Micro Motion recommends setting Flow Damping to 2.56 or higher.
The value you enter is automatically rounded down to the nearest valid value. Valid damping values are shown in the following table.
Valid values for Flow DampingTable 4-2:
Core processor type Update Rate setting Valid damping values
Standard Normal 0, 0.2, 0.4, 0.8, ... 51.2
Special 0, 0.04, 0.08, 0.16, ... 10.24
Enhanced Not applicable 0, 0.2, 0.4, 0.8, ... 51.2
Effect of Flow Damping on volume measurement
Flow Damping affects volume measurement for liquid volume data. Flow Damping also affects
volume measurement for gas standard volume data. The transmitter calculates volume data from the damped mass flow data.
Interaction between Flow Damping and Added Damping
In some circumstances, both Flow Damping and Added Damping are applied to the reported mass flow value.
Flow Damping controls the rate of change in flow process variables. Added Damping controls the rate of change reported via the mA output. If mA Output Process Variable is set to Mass Flow Rate, and both Flow Damping and Added Damping are set to non-zero values, flow damping is applied first, and the added damping calculation is applied to the result of the first calculation.
4.1.3 Configure Mass Flow Cutoff
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Flow > Mass Flow Cutoff
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Cutoff
Offline Maintain > Configuration > Low Flow Cutoff > Mass Flow Cutoff
Configuration and Use Manual 29
Page 38
Configure process measurement
Overview
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates below this cutoff will be reported as 0.
Procedure
Set Mass Flow Cutoff to the value you want to use.
The default value for Mass Flow Cutoff is 0.0 g/sec or a sensor-specific value set at the factory. The recommended setting is 0.05% of the sensor's rated maximum flow rate or a value below the highest expected flow rate. Do not set Mass Flow Cutoff to 0.0 g/sec.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the
actual mass data rather than the reported value.
Interaction between Mass Flow Cutoff and AO Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. AO Cutoff defines the lowest flow rate that will be reported via the mA output. If mA Output Process Variable is set to Mass Flow Rate, the mass flow rate reported via the mA
output is controlled by the higher of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events defined on mass flow).
AO Cutoff affects only mass flow values reported via the mA output.
Example: Cutoff interaction with AO Cutoff lower than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
AO Cutoff: 10 g/sec
Mass Flow Cutoff: 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with AO Cutoff higher than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
AO Cutoff: 15 g/sec
Mass Flow Cutoff: 10 g/sec
Result:
30 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 39
Configure process measurement
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
- The mA output will report zero flow.
- The frequency output will report the actual flow rate, and the actual flow rate will
be used in all internal processing.
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0
will be used in all internal processing.

4.2 Configure volume flow measurement for liquid applications

The volume flow measurement parameters control how liquid volume flow is measured and reported.
The volume flow measurement parameters include:
Volume Flow Type
Volume Flow Measurement Unit
Volume Flow Cutoff
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. You must choose one or the other.
4.2.1 Configure Volume Flow Type for liquid applications
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Flow > Vol Flow Type > Liquid Volume
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Liquid
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
Restriction
If you are using the petroleum measurement application, you must set Volume Flow Type to Liquid. Gas standard volume measurement is incompatible with the petroleum measurement application.
Not available
Configuration and Use Manual 31
Page 40
Configure process measurement
Restriction
If you are using the concentration measurement application, you must set Volume Flow Type to Liquid. Gas standard volume measurement is incompatible with the concentration measurement application.
Procedure
Set Volume Flow Type to Liquid.
4.2.2 Configure Volume Flow Measurement Unit for liquid applications
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > UNITS > VOL
Chinese-language display
ProLink II ProLink > Configuration > Flow > Vol Flow Units
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Unit
Offline Maintain > Configuration > Units > Volume Flow Rate
Overview
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume flow rate. The unit used for the volume total and volume inventory is based on this unit.
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement Unit, plus one user-defined measurement unit. Different communications tools may use
different labels for the units.
32 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 41
Configure process measurement
Table 4-3:
Options for Volume Flow Measurement Unit for liquid applications
Label
Unit description
Display (stand­ard)
Chinese-lan­guage display
ProLink II ProLink III Field Commu-
nicator
Cubic feet per second CUFT/S ft3/sec ft3/sec ft3/sec Cuft/s
Cubic feet per minute CUF/MN ft3/min ft3/min ft3/min Cuft/min
Cubic feet per hour CUFT/H ft3/hr ft3/hr ft3/hr Cuft/h
Cubic feet per day CUFT/D ft3/day ft3/day ft3/day Cuft/d
Cubic meters per second M3/S m3/sec m3/sec m3/sec Cum/s
Cubic meters per minute M3/MIN m3/min m3/min m3/min Cum/min
Cubic meters per hour M3/H m3/hr m3/hr m3/hr Cum/h
Cubic meters per day M3/D m3/day m3/day m3/day Cum/d
U.S. gallons per second USGPS US gal/sec US gal/sec US gal/sec gal/s
U.S. gallons per minute USGPM US gal/min US gal/min US gal/min gal/min
U.S. gallons per hour USGPH US gal/hr US gal/hr US gal/hr gal/h
U.S. gallons per day USGPD US gal/day US gal/day US gal/day gal/d
Million U.S. gallons per
MILG/D mil US gal/day mil US gal/day mil US gal/day MMgal/d
day
Liters per second L/S l/sec l/sec l/sec L/s
Liters per minute L/MIN l/min l/min l/min L/min
Liters per hour L/H l/hr l/hr l/hr L/h
Million liters per day MILL/D mil l/day mil l/day mil l/day ML/d
Imperial gallons per sec-
UKGPS Imp gal/sec Imp gal/sec Imp gal/sec Impgal/s
ond
Imperial gallons per mi-
UKGPM Imp gal/min Imp gal/min Imp gal/min Impgal/min
nute
Imperial gallons per hour UKGPH Imp gal/hr Imp gal/hr Imp gal/hr Impgal/h
Imperial gallons per day UKGPD Imp gal/day Imp gal/day Imp gal/day Impgal/d
Barrels per second
(1)
BBL/S barrels/sec barrels/sec barrels/sec bbl/s
Barrels per minute BBL/MN barrels/min barrels/min barrels/min bbl/min
Barrels per hour BBL/H barrels/hr barrels/hr barrels/hr bbl/h
Barrels per day BBL/D barrels/day barrels/day barrels/day bbl/d
Beer barrels per second
(2)
BBBL/S Beer barrels/sec Beer barrels/sec Beer barrels/sec bbbl/s
Beer barrels per minute BBBL/MN Beer barrels/min Beer barrels/min Beer barrels/min bbbl/min
Beer barrels per hour BBBL/H Beer barrels/hr Beer barrels/hr Beer barrels/hr bbbl/h
Beer barrels per day BBBL/D Beer barrels/day Beer barrels/day Beer barrels/day bbbl/d
(1) Unit based on oil barrels (42 U.S. gallons).
Unit based on U.S. beer barrels (31 U.S. gallons).
(2)
Configuration and Use Manual 33
Page 42
Configure process measurement
Options for Volume Flow Measurement Unit for liquid applications (continued)Table 4-3:
Label
Display (stand-
Unit description
Special unit SPECL Special special special Spcl
ard)
Chinese-lan­guage display
ProLink II ProLink III Field Commu-
nicator
Define a special measurement unit for volume flow
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Special Units
ProLink III Device Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Volume Special Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Not available
Note
Although you cannot define a special measurement unit using the display, you can use the display to select an existing special measurement unit, and to view process data using the special measurement unit.
Procedure
1.
Specify Base Volume Unit.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Volume Flow Conversion Factor as follows:
a. x base units = y special units
b. Volume Flow Conversion Factor = x/y
4. Enter Volume Flow Conversion Factor.
5. Set Volume Flow Label to the name you want to use for the volume flow unit.
6. Set Volume Total Label to the name you want to use for the volume total and volume
inventory unit.
34 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 43
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
1.
Set Base Volume Unit to Gallons (gal).
2. Set Base Time Unit to Seconds (sec).
3. Calculate the conversion factor:
a. 1 gal/sec = 8 pints/sec
b. Volume Flow Conversion Factor = 1/8 = 0.1250
4. Set Volume Flow Conversion Factor to 0.1250.
5. Set Volume Flow Label to pints/sec.
6. Set Volume Total Label to pints.
4.2.3 Configure Volume Flow Cutoff
Configure process measurement
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Flow > Vol Flow Cutoff
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Cutoff
Offline Maintain > Configuration > Low Flow Cutoff > Volume Flow Cutoff
Overview
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured. All volume flow rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
The default value for Volume Flow Cutoff is 0.0 l/sec (liters per second). The lower limit is 0. The upper limit is the sensor’s flow calibration factor, in units of l/sec, multiplied by 0.2.
Interaction between Volume Flow Cutoff and AO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will
report as measured. AO Cutoff defines the lowest flow rate that will be reported via the mA output. If mA Output Process Variable is set to Volume Flow Rate, the volume flow rate reported via the mA output is controlled by the higher of the two cutoff values.
Configuration and Use Manual 35
Page 44
Configure process measurement
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the volume flow values used in other transmitter behavior (e.g., events defined on the volume flow).
AO Cutoff affects only flow values reported via the mA output.
Example: Cutoff interaction with AO Cutoff lower than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 10 l/sec
Volume Flow Cutoff: 15 l/sec
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with AO Cutoff higher than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 15 l/sec
Volume Flow Cutoff: 10 l/sec
Result:
If the volume flow rate drops below 15 l/sec but not below 10 l/sec:
- The mA output will report zero flow.
- The frequency output will report the actual flow rate, and the actual flow rate will
be used in all internal processing.
If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and
0 will be used in all internal processing.

4.3 Configure gas standard volume (GSV) flow measurement

The gas standard volume (GSV) flow measurement parameters control how gas standard volume flow is measured and reported.
The GSV flow measurement parameters include:
Volume Flow Type
Standard Gas Density
Gas Standard Volume Flow Measurement Unit
Gas Standard Volume Flow Cutoff
36 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 45
Configure process measurement
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. You must choose one or the other.
4.3.1 Configure Volume Flow Type for gas applications
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Flow > Vol Flow Type
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Standard Gas Volume
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement is used.
Not available
Restriction
If you are using the petroleum measurement application, you must set Volume Flow Type to Liquid. Gas standard volume measurement is incompatible with the petroleum measurement application.
Restriction
If you are using the concentration measurement application, you must set Volume Flow Type to Liquid. Gas standard volume measurement is incompatible with the concentration measurement application.
Procedure
Set Volume Flow Type to Gas Standard Volume.
4.3.2 Configure Standard Gas Density
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Flow > Std Gas Density
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Gas Ref Density
Not available
Configuration and Use Manual 37
Page 46
Configure process measurement
Overview
The Standard Gas Density value is used to convert the measured flow data to the standard reference values.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for Standard Gas Density.
Procedure
Set Standard Gas Density to the standard reference density of the gas you are measuring.
Note
ProLink II and ProLink III provide a guided method that you can use to calculate the standard density of your gas, if you do not know it.
4.3.3 Configure Gas Standard Volume Flow Measurement Unit
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > UNITS > GSV
Chinese-language display
ProLink II ProLink > Configuration > Flow > Std Gas Vol Flow Units
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > GSV Flow Unit
Offline Maintain > Configuration > Units > Gas Std Volume Flow
Overview
Gas Standard Volume Flow Measurement Unit specifies the unit of measure that will be displayed for the gas standard volume flow rate. The measurement unit used for the gas standard volume total and the gas standard volume inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Gas Standard Volume.
Procedure
Set Gas Standard Volume Flow Measurement Unit to the unit you want to use.
The default setting for Gas Standard Volume Flow Measurement Unit is SCFM (Standard Cubic Feet per Minute).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
38 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 47
Configure process measurement
Options for Gas Standard Volume Flow Measurement Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow Measurement Unit, plus one user-defined special measurement unit. Different
communications tools may use different labels for the units.
Options for Gas Standard Volume Measurement UnitTable 4-4:
Label
Display (stand-
Unit description
Normal cubic meters per second
Normal cubic meters per minute
Normal cubic meters per hour
Normal cubic meters per day
Normal liter per second NLPS NLPS NLPS NLPS NLPS
Normal liter per minute NLPM NLPM NLPM NLPM NLPM
Normal liter per hour NLPH NLPH NLPH NLPH NLPH
Normal liter per day NLPD NLPD NLPD NLPD NLPD
Standard cubic feet per second
Standard cubic feet per minute
Standard cubic feet per hour
Standard cubic feet per day
Standard cubic meters per second
Standard cubic meters per minute
Standard cubic meters per hour
Standard cubic meters per day
Standard liter per second SLPS SLPS SLPS SLPS SLPS
Standard liter per minute SLPM SLPM SLPM SLPM SLPM
Standard liter per hour SLPH SLPH SLPH SLPH SLPH
Standard liter per day SLPD SLPD SLPD SLPD SLPD
ard)
NM3/S Nm3/sec Nm3/sec Nm3/sec Nm3/sec
NM3/MN Nm3/min Nm3/min Nm3/sec Nm3/min
NM3/H Nm3/hr Nm3/hr Nm3/hr Nm3/hr
NM3/D Nm3/day Nm3/day Nm3/day Nm3/day
SCFS SCFS SCFS SCFS SCFS
SCFM SCFM SCFM SCFM SCFM
SCFH SCFH SCFH SCFH SCFH
SCFD SCFD SCFD SCFD SCFD
SM3/S Sm3/sec Sm3/S Sm3/sec Sm3/sec
SM3/MN Sm3/min Sm3/min Sm3/min Sm3/min
SM3/H Sm3/hr Sm3/hr Sm3/hr Sm3/hr
SM3/D Sm3/day Sm3/day Sm3/day Sm3/day
Chinese-lan­guage display
ProLink II ProLink III Field Commu-
nicator
Configuration and Use Manual 39
Page 48
Configure process measurement
Options for Gas Standard Volume Measurement Unit (continued)Table 4-4:
Label
Unit description
Special measurement unit
Display (stand­ard)
SPECL Special special special Special
Chinese-lan­guage display
ProLink II ProLink III Field Commu-
nicator
Define a special measurement unit for gas standard volume flow
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Special Units
ProLink III Device Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Special GSV Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Not available
Note
Although you cannot define a special measurement unit using the display, you can use the display to select an existing special measurement unit, and to view process data using the special measurement unit.
Procedure
1.
Specify Base Gas Standard Volume Unit.
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Gas Standard Volume Flow Conversion Factor as follows:
a. x base units = y special units
b. Gas Standard Volume Flow Conversion Factor = x/y
4. Enter the Gas Standard Volume Flow Conversion Factor.
40 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 49
Configure process measurement
5. Set Gas Standard Volume Flow Label to the name you want to use for the gas standard
volume flow unit.
6.
Set Gas Standard Volume Total Label to the name you want to use for the gas standard volume total and gas standard volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for gas standard volume flow
You want to measure gas standard volume flow in thousands of standard cubic feet per minute.
1. Set Base Gas Standard Volume Unit to SCFM.
2. Set Base Time Unit to minutes (min).
3. Calculate the conversion factor:
a. 1 thousands of standard cubic feet per minute = 1000 cubic feet per minute
b. Gas Standard Volume Flow Conversion Factor = 1/1000 = 0.001
4. Set Gas Standard Volume Flow Conversion Factor to 0.001.
5. Set Gas Standard Volume Flow Label to KSCFM.
6. Set Gas Standard Volume Total Label to KSCF.
4.3.4 Configure Gas Standard Volume Flow Cutoff
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Flow > Std Gas Vol Flow Cutoff
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > GSV Cutoff
Overview
Gas Standard Volume Flow Cutoff specifies the lowest gas standard volume flow rate that will reported as measured. All gas standard volume flow rates below this cutoff will be reported as 0.
Procedure
Set Gas Standard Volume Flow Cutoff to the value you want to use.
The default value for Gas Standard Volume Flow Cutoff is 0.0. The lower limit is 0.0. There is no upper limit.
Offline Maintain > Configuration > Low Flow Cutoff > Gas Vol Flow Cutoff
Configuration and Use Manual 41
Page 50
Configure process measurement
Interaction between Gas Standard Volume Flow Cutoff and AO Cutoff
Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the
transmitter will report as measured. AO Cutoff defines the lowest flow rate that will be reported via the mA output. If mA Output Process Variable is set to Gas Standard Volume Flow Rate, the volume flow rate reported via the mA output is controlled by the higher of the two cutoff values.
Gas Standard Volume Flow Cutoff affects both the gas standard volume flow values reported via outputs and the gas standard volume flow values used in other transmitter behavior (e.g., events defined on gas standard volume flow).
AO Cutoff affects only flow values reported via the mA output.
Example: Cutoff interaction with AO Cutoff lower than Gas Standard Volume Flow Cutoff
Configuration:
mA Output Process Variable for the primary mA output: Gas Standard Volume Flow Rate
Frequency Output Process Variable: Gas Standard Volume Flow Rate
AO Cutoff for the primary mA output: 10 SLPM (standard liters per minute)
Gas Standard Volume Flow Cutoff: 15 SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with AO Cutoff higher than Gas Standard Volume Flow Cutoff
Configuration:
mA Output Process Variable for the primary mA output: Gas Standard Volume Flow Rate
Frequency Output Process Variable: Gas Standard Volume Flow Rate
AO Cutoff for the primary mA output: 15 SLPM (standard liters per minute)
Gas Standard Volume Flow Cutoff: 10 SLPM
Result:
If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:
- The primary mA output will report zero flow.
- The frequency output will report the actual flow rate, and the actual flow rate will
be used in all internal processing.
If the gas standard volume flow rate drops below 10 SLPM, both outputs will report
zero flow, and 0 will be used in all internal processing.
42 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 51
4.4 Configure Flow Direction
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Flow > Flow Direction
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Flow Direction
Overview
Flow Direction controls how forward flow and reverse flow affect flow measurement and reporting.
Flow Direction is defined with respect to the flow arrow on the sensor:
Forward flow (positive flow) moves in the direction of the flow arrow on the sensor.
Reverse flow (negative flow) moves in the direction opposite to the flow arrow on
the sensor.
Not available
Configure process measurement
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the setting of the Flow Direction parameter.
Procedure
Set Flow Direction to the value you want to use.
4.4.1 Options for Flow Direction
Options for Flow DirectionTable 4-5:
Flow Direction setting Relationship to Flow Direction ar-
ProLink II ProLink III Field Communicator
Forward Forward Forward Appropriate when the Flow Direction
Reverse Reverse Reverse Appropriate when the Flow Direction
Absolute Value Absolute Value Absolute Value Flow Direction arrow is not relevant.
row on sensor
arrow is in the same direction as the majority of flow.
arrow is in the same direction as the majority of flow.
Configuration and Use Manual 43
Page 52
Configure process measurement
Options for Flow Direction (continued)Table 4-5:
Flow Direction setting Relationship to Flow Direction ar-
ProLink II ProLink III Field Communicator
Bidirectional Bidirectional Bi directional Appropriate when both forward and
Negate Forward Negate Forward Negate/Forward Only Appropriate when the Flow Direction
Negate Bidirectional Negate Bidirectional Negate/Bi-directional Appropriate when both forward and
row on sensor
reverse flow are expected, and for­ward flow will dominate, but the amount of reverse flow will be signifi­cant.
arrow is in the opposite direction from the majority of flow.
reverse flow are expected, and reverse flow will dominate, but the amount of forward flow will be significant.
Effect of Flow Direction on mA outputs
Flow Direction affects how the transmitter reports flow values via the mA outputs. The mA
outputs are affected by Flow Direction only if mA Output Process Variable is set to a flow variable.
Flow Direction and mA outputs
The effect of Flow Direction on the mA outputs depend on Lower Range Value configured for the mA output:
If Lower Range Value is set to 0, see Figure 1.
If Lower Range Value is set to a negative value, see Figure 2.
44 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 53
Configure process measurement
Figure 4-1:
Effect of Flow Direction on the mA output: Lower Range Value = 0
Flow Direction = Forward
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Lower Range Value = 0
Upper Range Value = x
Effect of Flow Direction on the mA output: Lower Range Value < 0Figure 4-2:
Flow Direction = Forward
Flow Direction = Reverse, Negate Forward
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Flow Direction = Reverse, Negate Forward
Flow Direction = Absolute Value, Bidirectional, Negate Bidirectional
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Flow Direction = Absolute Value, Bidirectional, Negate Bidirectional
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Lower Range Value = x
Upper Range Value = x
Example: Flow Direction = Forward and Lower Range Value = 0
Configuration:
Flow Direction = Forward
Lower Range Value = 0 g/sec
Upper Range Value = 100 g/sec
Result:
Under conditions of reverse flow or zero flow, the mA output is 4 mA.
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Configuration and Use Manual 45
Page 54
Configure process measurement
Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA output
Under conditions of forward flow, if the flow rate equals or exceeds 100 g/sec, the
Example: Flow Direction = Forward and Lower Range Value < 0
Configuration:
Flow Direction = Forward
Lower Range Value = 100 g/sec
Upper Range Value = +100 g/sec
Result:
Under conditions of zero flow, the mA output is 12 mA.
Under conditions of forward flow, for flow rates between 0 and +100 g/sec, the mA
Under conditions of forward flow, if (the absolute value of) the flow rate equals or
Under conditions of reverse flow, for flow rates between 0 and 100 g/sec, the mA
Under conditions of reverse flow, if the absolute value of the flow rate equals or
varies between 4 mA and 20 mA in proportion to the flow rate.
mA output will be proportional to the flow rate up to 20.5 mA, and will be level at
20.5 mA at higher flow rates.
output varies between 12 mA and 20 mA in proportion to (the absolute value of) the flow rate.
exceeds 100 g/sec, the mA output is proportional to the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher flow rates.
output varies between 4 mA and 12 mA in inverse proportion to the absolute value of the flow rate.
exceeds 100 g/sec, the mA output is inversely proportional to the flow rate down to
3.8 mA, and will be level at 3.8 mA at higher absolute values.
Example: Flow Direction = Reverse
Configuration:
Flow Direction = Reverse
Lower Range Value = 0 g/sec
Upper Range Value = 100 g/sec
Result:
Under conditions of forward flow or zero flow, the mA output is 4 mA.
Under conditions of reverse flow, for flow rates between 0 and +100 g/sec, the mA
output level varies between 4 mA and 20 mA in proportion to the absolute value of the flow rate.
Under conditions of reverse flow, if the absolute value of the flow rate equals or
exceeds 100 g/sec, the mA output will be proportional to the absolute value of the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher absolute values.
46 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 55
Configure process measurement
Effect of Flow Direction on frequency outputs
Flow Direction affects how the transmitter reports flow values via the frequency outputs. The
frequency outputs are affected by Flow Direction only if Frequency Output Process Variable is set to a flow variable.
Table 4-6:
Effect of the Flow Direction parameter and actual flow direction on frequency outputs
Actual flow direction
Flow Direction setting
Forward Hz > 0 0 Hz 0 Hz
Reverse 0 Hz 0 Hz Hz > 0
Bidirectional Hz > 0 0 Hz Hz > 0
Absolute Value Hz > 0 0 Hz Hz > 0
Negate Forward 0 Hz 0 Hz Hz > 0
Negate Bidirectional Hz > 0 0 Hz Hz > 0
Forward Zero flow Reverse
Effect of Flow Direction on discrete outputs
The Flow Direction parameter affects the discrete output behavior only if Discrete Output Source is set to Flow Direction.
Table 4-7:
Effect of the Flow Direction parameter and actual flow direction on discrete outputs
Actual flow direction
Flow Direction setting
Forward OFF OFF ON
Reverse OFF OFF ON
Bidirectional OFF OFF ON
Absolute Value OFF OFF OFF
Negate Forward ON OFF OFF
Negate Bidirectional ON OFF OFF
Forward Zero flow Reverse
Effect of Flow Direction on digital communications
Flow Direction affects how flow values are reported via digital communications.
Configuration and Use Manual 47
Page 56
Configure process measurement
Table 4-8:
Effect of the Flow Direction parameter and actual flow direction on flow values reported via digital communications
Actual flow direction
Flow Direction setting
Forward Positive 0 Negative
Reverse Positive 0 Negative
Bidirectional Positive 0 Negative
Absolute Value Positive
Negate Forward Negative 0 Positive
Negate Bidirectional Negative 0 Positive
Forward Zero flow Reverse
(3)
0 Positive
Effect of Flow Direction on flow totals
Flow Direction affects how flow totals and inventories are calculated.
Table 4-9:
Effect of the Flow Direction parameter and actual flow direction on flow totals and inventories
Actual flow direction
Flow Direction setting
Forward Totals increase Totals do not change Totals do not change
Reverse Totals do not change Totals do not change Totals increase
Bidirectional Totals increase Totals do not change Totals decrease
Absolute Value Totals increase Totals do not change Totals increase
Negate Forward Totals do not change Totals do not change Totals increase
Negate Bidirectional Totals decrease Totals do not change Totals increase
Forward Zero flow Reverse

4.5 Configure density measurement

The density measurement parameters control how density is measured and reported. Density measurement (along with mass measurement) is used to determine liquid volume flow.
The density measurement parameters include:
Density Measurement Unit
Slug Flow Parameters
Density Damping
Density Cutoff
(3) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
48 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 57
4.5.1 Configure Density Measurement Unit
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > UNITS > DENS
Chinese-language display
ProLink II ProLink > Configuration > Density > Density Units
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit
Overview
Density Measurement Unit specifies the units of measure that will be displayed for density measurement.
Procedure
Set Density Measurement Unit to the option you want to use.
Offline Maintain > Configuration > Units > Density
Configure process measurement
The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter).
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit. Different communications tools may use different labels.
Options for Density Measurement UnitTable 4-10:
Label
Display (stand-
Unit description
Specific gravity unit (not temperature-corrected)
Grams per cubic centime­ter
Grams per liter G/L g/L g/l g/l g/L
Grams per milliliter G/mL g/mL g/ml g/ml g/mL
Kilograms per liter KG/L kg/L kg/l kg/l kg/L
Kilograms per cubic me­ter
Pounds per U.S. gallon LB/GAL lbs/US gal lbs/Usgal lbs/Usgal lb/gal
Pounds per cubic foot LB/CUF lbs/ft3 lbs/ft3 lbs/ft3 lb/Cuft
Pounds per cubic inch LB/CUI lbs/in3 lbs/in3 lbs/in3 lb/CuIn
API gravity D API deg API degAPI degAPI degAPI
ard)
SGU SGU SGU SGU SGU
G/CM3 g/cm3 g/cm3 g/cm3 g/Cucm
KG/M3 kg/m3 kg/m3 kg/m3 kg/Cum
Chinese-lan­guage display
ProLink II ProLink III Field Commu-
nicator
Configuration and Use Manual 49
Page 58
Configure process measurement
Options for Density Measurement Unit (continued)Table 4-10:
Label
Display (stand-
Unit description
Short ton per cubic yard ST/CUY sTon/yd3 sT/yd3 sT/yd3 STon/Cuyd
ard)
Chinese-lan­guage display
ProLink II ProLink III Field Commu-
nicator
4.5.2 Configure slug flow parameters
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Density > Slug High Limit
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Slug Low Limit
Overview
The slug flow parameters control how the transmitter detects and reports two-phase flow (gas in a liquid process or liquid in a gas process).
Not available
ProLink > Configuration > Density > Slug Low Limit
ProLink > Configuration > Density > Slug Duration
Configure > Manual Setup > Measurements > Density > Slug High Limit
Configure > Manual Setup > Measurements > Density > Slug Duration
Procedure
1.
Set Slug Low Limit to the lowest density value that is considered normal in your process.
Values below this will cause the transmitter to perform the configured slug flow action. Typically, this value is the lowest density value in the normal range of your process.
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the occurrence of slug flow alarms that are not significant to your process, set Slug Low Limit slightly below your expected lowest process density.
You must enter Slug Low Limit in g/cm3, even if you configured another unit for density measurement.
The default value for Slug Low Limit is 0.0 g/cm3. The range is 0.0 to 10.0 g/cm3.
2.
Set Slug High Limit to the highest density value that is considered normal in your process.
50 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 59
Configure process measurement
Values above this will cause the transmitter to perform the configured slug flow action. Typically, this value is the highest density value in the normal range of your process.
Tip
To reduce the occurrence of slug flow alarms that are not significant to your process, set Slug High Limit slightly above your expected highest process density.
You must enter Slug High Limit in g/cm3, even if you configured another unit for density measurement.
The default value for Slug High Limit is 5.0 g/cm3. The range is 0.0 to 10.0 g/cm3.
3.
Set Slug Duration to the number of seconds that the transmitter will wait for a slug flow condition to clear before performing the configured slug flow action.
The default value for Slug Duration is 0.0 seconds. The range is 0.0 to 60.0 seconds.
Slug flow detection and reporting
Slug flow is typically used as an indicator of two-phase flow (gas in a liquid process or liquid in a gas process). Two-phase flow can cause a variety of process control issues. By configuring the slug flow parameters appropriately for your application, you can detect process conditions that require correction.
Tip
To decrease the occurrence of slug flow alarms, lower Slug Low Limit or raise Slug High Limit.
A slug flow condition occurs whenever the measured density goes below Slug Low Limit or above Slug High Limit. If this occurs:
A slug flow alarm is posted to the active alarm log.
All outputs that are configured to represent flow rate hold their last “pre-slug flow”
value for the configured Slug Duration.
If the slug flow condition clears before Slug Duration expires:
Outputs that represent flow rate revert to reporting actual flow.
The slug flow alarm is deactivated, but remains in the active alarm log until it is
acknowledged.
If the slug flow condition does not clear before Slug Duration expires, the outputs that represent flow rate report a flow rate of 0.
If Slug Duration is set to 0.0 seconds, the outputs that represent flow rate will report a flow rate of 0 as soon as slug flow is detected.
Configuration and Use Manual 51
Page 60
Configure process measurement
4.5.3 Configure Density Damping
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Density > Density Damping
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes
in the reported process variable. At the end of the interval, the reported process variable will reflect 63% of the change in the actual measured value.
Procedure
Offline Maintain > Configuration > Damping > Density Damping
Set Density Damping to the value you want to use.
The default value is 1.6 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table:
Core processor type Update Rate setting Density Damping range
Standard Normal 0 to 51.2 seconds
Special 0 to 10.24 seconds
Enhanced Not applicable 0 to 40.96 seconds
Tips
A high damping value makes the process variable appear smoother because the reported value
changes slowly.
A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
The value you enter is automaticaly rounded down to the nearest valid value. The valid values for Density Damping depend on the setting of Update Rate.
52 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 61
Configure process measurement
Valid values for Density DampingTable 4-11:
Core processor type Update Rate setting Valid damping values
Standard Normal 0, 0.2, 0.4, 0.8, ... 51.2
Special 0, 0.04, 0.08, 0.16, ... 10.24
Enhanced Not applicable 0, 0.04, 0.08, 0.16, ... 40.96
Effect of Density Damping on volume measurement
Density Damping affects liquid volume measurement. Liquid volume values are calculated
from the damped density value rather than the measured density value. Density Damping does not affect gas standard volume measurement.
Interaction between Density Damping and Added Damping
In some circumstances, both Density Damping and Added Damping are applied to the reported density value.
Density Damping controls the rate of change in the density process variable. Added Damping controls the rate of change reported via the mA output. If mA Output Process Variable is set to Density, and both Density Damping and Added Damping are set to non-zero values, density damping is applied first, and the added damping calculation is applied to the result of the first calculation.
4.5.4 Configure Density Cutoff
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Density > Low Density Cutoff
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Cutoff
Overview
Density Cutoff specifies the lowest density value that will be reported as measured. All density values below this cutoff will be reported as 0.
Procedure
Set Density Cutoff to the value you want to use.
The default value for Density Cutoff is 0.2 g/cm3. The range is 0.0 g/cm3 to 0.5 g/cm3.
Offline Maintain > Configuration > Low Flow Cutoff > Density Cutoff
Configuration and Use Manual 53
Page 62
Configure process measurement
Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density Cutoff, the volume flow rate is reported as 0. Density Cutoff does not affect gas standard
volume measurement. Gas standard volume values are always calculated from the value configured for Standard Gas Density.

4.6 Configure temperature measurement

The temperature measurement parameters control how temperature data from the sensor is reported. Temperature data is used to compensate for the effect of temperature on the sensor tubes during flow measurement .
The temperature measurement parameters include:
Temperature Measurement Unit
Temperature Damping
4.6.1 Configure Temperature Measurement Unit
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > UNITS > TEMP
Chinese-language display
ProLink II ProLink > Configuration > Temperature > Temp Units
ProLink III Device Tools > Configuration > Process Measurement > Temperature
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temperature Unit
Overview
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
The default setting is Degrees Celsius.
Offline Maintain > Configuration > Units > Temperature
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different communications tools may use different labels for the units.
54 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 63
Options for Temperature Measurement UnitTable 4-12:
Chinese-lan-
Display
Unit description
Degrees Celsius °C °C degC °C degC
Degrees Fahrenheit °F °F degF °F degF
Degrees Rankine °R °R degR °R degR
Kelvin °K °K degK °K Kelvin
(standard)
guage dis­play ProLink II ProLink III
4.6.2 Configure Temperature Damping
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Temperature > Temp Damping
ProLink III Device Tools > Configuration > Temperature
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temp Damping
Offline Maintain > Configuration > Damping > Temperature Damping
Configure process measurement
Label
Field Com­municator
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes
in the reported process variable. At the end of the interval, the reported process variable will reflect 63% of the change in the actual measured value.
Procedure
Enter the value you want to use for Temperature Damping.
The default value is 4.8 seconds. The range is 0.0 to 76.8 seconds.
Tips
A high damping value makes the process variable appear smoother because the reported value
changes slowly.
A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
Configuration and Use Manual 55
Page 64
Configure process measurement
The value you enter is automaticaly rounded down to the nearest valid value. Valid values for Temperature Damping are 0, 0.6, 1.2, 2.4, 4.8, … 76.8.
Effect of Temperature Damping on process measurement
Temperature Damping affects the response speed for temperature compensation with
fluctuating temperatures. Temperature compensation adjusts the process measurement to compensate for the effect of temperature on the sensor tube.
Temperature Damping affects petroleum measurement process variables only if the transmitter is configured to use temperature data from the sensor. If an external temperature value is used for petroleum measurement, Temperature Damping does not affect petroleum measurement process variables.
Temperature Damping affects concentration measurement process variables only if the transmitter is configured to use temperature data from the sensor. If an external temperature value is used for concentration measurement, Temperature Damping does not affect concentration measurement process variables.

4.7 Configure the petroleum measurement application

The petroleum measurement application enables Correction for the effect of Temperature on the volume of Liquids (CTL), by calculating and applying a Volume Correction Factor (VCF) to volume measurement. Internal calculations are performed in compliance with American Petroleum Institute (API) standards.
4.7.1 Configure petroleum measurement using ProLink II
1. Choose ProLink > Configuration > API Setup.
2.
Specify the API table to use.
a. In API Chapter 11.1 Table Type, select the API table group.
b. In Units, select the the measurement units you want to use.
These two parameters uniquely specify the API table.
3. If your API table is 53A, 53B, 53D, or 54C, set Reference Temperature to the appropriate
value for your application. Enter the value in °C.
4. If your API table is 6C, 24C, or 54C, set Thermal Expansion Coefficient to the appropriate
value for your application.
5. Determine how the transmitter will obtain temperature data for the petroleum
measurement calculations, and perform the required setup.
56 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 65
Option Setup
Temperature data from the sensor
A user-configured static temperature value
Polling for tempera-
(4)
ture
a. Choose View > Preferences . b.
Disable Use External Temperature.
a. Choose View > Preferences . b.
Enable Use External Temperature. c. Choose ProLink > Configuration > Temperature. d. Set External Temperature to the value to be used.
a. Ensure that the primary mA output has been wired to support
HART polling. b.
Choose View > Preferences . c. Enable Use External Temperature. d. Choose ProLink > Configuration > Polled Variables. e. Choose an unused polling slot. f. Set Polling Control to Poll As Primary or Poll as Secondary, and click
Apply. g. Set External Tag to the HART tag of the external temperature de-
vice. h. Set Variable Type to External Temperature.
Configure process measurement
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
A value written by digital communica­tions
a. Choose View > Preferences . b.
Enable Use External Temperature. c. Perform the necessary host programming and communications
setup to write temperature data to the transmitter at appropri-
ate intervals.
Note
If the Weights & Measures application is implemented and the trans­mitter is secured, digital communications cannot be used to write temperature data to the transmitter.
4.7.2 Configure petroleum measurement using ProLink III
1. Choose Device Tools > Configuration > Process Measurement > Petroleum Measurement.
Specify the API table to use.
2.
a. Select the API table group from API Table Type.
b. Set Petroleum Measurement Units to the measurement units you want to use.
c. Click Apply.
(4) Not available on all transmitters.
Configuration and Use Manual 57
Page 66
Configure process measurement
3.
4. If your API table is 6C, 24C, or 54C, set Thermal Expansion Coefficient to the appropriate
5. Set Temperature Source to the method that the transmitter will use to obtain
These two parameters uniquely specify the API table.
If your API table is 53A, 53B, 53D, or 54C, set Reference Temperature to the appropriate value for your application. Enter the value in °C.
value for your application.
temperature data.
Option Description
Poll for external value
RTD The transmitter will use the temperature data from the sen-
Static or Digital Communications The transmitter will use the temperature value that it reads
(5)
The transmitter will poll an external temperature device, us­ing HART protocol over the primary mA output.
sor.
from memory.
Static: The configured value is used.
Digital Communications: An external host writes trans-
mitter data to transmitter memory.
The same location in memory is used for both options.
External temperature data is used only in petroleum measurement calculations. All other transmitter calculations use temperature data from the sensor.
6.
If you chose RTD, no more configuration is required. Click Apply and exit.
7. If you chose to poll for temperature data:
a. Select the Polling Slot to use.
b. Set Polling Control to Poll as Primary or Poll as Secondary, and click Apply.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network. The Field Communicator
is not a HART master.
c. Set External Device Tag to the HART tag of the external temperature device, and
click Apply.
8.
If you chose to use a static temperature value, set External Temperature to the value to use, and click Apply.
9. If you want to use digital communications, perform the necessary host programming and communications setup to write temperature data to the transmitter at appropriate intervals.
(5) Not available on all transmitters.
58 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 67
Configure process measurement
Note
If the Weights & Measures application is implemented and the transmitter is secured, digital communications cannot be used to write temperature data to the transmitter.
4.7.3 Configure petroleum measurement using the Field Communicator
1. Choose Online > Configure > Manual Setup > Measurements > Set Up Petroleum.
Specify the API table to use.
2.
a. Open the Petroleum Measurement Source menu and select the API table
number.
Depending on your choice, you may be prompted to enter a reference temperature or a thermal expansion coefficient.
b. Enter the API table letter.
These two parameters uniquely specify the API table.
3. Determine how the transmitter will obtain temperature data for the petroleum
measurement calculations, and perform the required setup.
Option Setup
Temperature data from the sensor
A user-configured static temperature value
a. Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > Temperature.
b.
Set External Temperature to Disabled.
a. Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > Temperature.
b.
Set External Temperature to Enabled.
c. Set Correction Temperature to the value to be used.
Configuration and Use Manual 59
Page 68
Configure process measurement
Option Setup
Polling for tempera-
(6)
ture
a. Ensure that the primary mA output has been wired to support
HART polling.
b.
Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > Temperature. c. Set External Temperature to Enabled. d. Choose External Polling. e. Set Poll Control to Poll As Primary or Poll as Secondary. f. Determine whether you will use Polling Slot 1 or Polling Slot 2. g. For the chosen slot, set Ext Dev Tag to the HART tag of the exter-
nal temperature device. h. For the chosen slot, set Polled Variable to Temperature.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
A value written by digital communica­tions
a. Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > Temperature.
b.
Set External Temperature to Enabled. c. Perform the necessary host programming and communications
setup to write temperature data to the transmitter at appropri-
ate intervals.
Note
If the Weights & Measures application is implemented and the trans­mitter is secured, digital communications cannot be used to write temperature data to the transmitter.
4.7.4 API reference tables
API reference tables, associated process fluids, and associated calculation valuesTable 4-13:
Table name Process fluid CTL source data Reference temperature Density unit
5A Generalized crude and
JP4
5B Generalized products Observed density and
5D Lubricating oils Observed density and
Observed density and observed temperature
observed temperature
observed temperature
60 °F (non-configurable) Degrees API
Range: 0 to 100
60 °F (non-configurable) Degrees API
Range: 0 to 85
60 °F (non-configurable) Degrees API
Range: 10 to +40
(6) Not available on all transmitters.
60 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 69
Configure process measurement
API reference tables, associated process fluids, and associated calculation values (continued)Table 4-13:
Table name Process fluid CTL source data Reference temperature Density unit
6C Liquids with a constant
density base or known thermal expansion coef­ficient
23A Generalized crude and
JP4
23B Generalized products Observed density and
23D Lubricating oils Observed density and
24C Liquids with a constant
density base or known thermal expansion coef­ficient
53A Generalized crude and
JP4
53B Generalized products Observed density and
53D Lubricating oils Observed density and
54C Liquids with a constant
density base or known thermal expansion coef­ficient
User-supplied reference density (or thermal ex­pansion coefficient) and observed temperature
Observed density and observed temperature
observed temperature
observed temperature
User-supplied reference density (or thermal ex­pansion coefficient) and observed temperature
Observed density and observed temperature
observed temperature
observed temperature
User-supplied reference density (or thermal ex­pansion coefficient) and observed temperature
60 °F (non-configurable) Degrees API
60 °F (non-configurable) Relative density
Range: 0.6110 to 1.0760
60 °F (non-configurable) Relative density
Range: 0.6535 to 1.0760
60 °F (non-configurable) Relative density
Range: 0.8520 to 1.1640
60 °F (non-configurable) Relative density
15 °C (configurable) Base density
Range: 610 to 1075 kg/m
15 °C (configurable) Base density
Range: 653 to 1075 kg/m
15 °C (configurable) Base density
Range: 825 to 1164 kg/m
15 °C (configurable) Base density in kg/m
3
3
3
3

4.8 Configure the concentration measurement application

The concentration measurement application calculates concentration data from process temperature and density. Micro Motion provides a set of concentration matrices that provide the reference data for several standard industry applications and process fluids. If desired, you can build a custom matrix for your process fluid, or purchase a custom matrix from Micro Motion.
More information about the concentration measurement application is available in the following manual: Micro Motion Enhanced Density Application: Theory, Configuration, and
Use.
Configuration and Use Manual 61
Page 70
Configure process measurement
Note
The concentration measurement application is also known as the enhanced density application.
4.8.1 Configure concentration measurement using ProLink II
This task guides you through loading and setting up a concentration matrix to use for measurement. It does not cover building a concentration matrix.
Note
Concentration matrices can be made available on your transmitter either by loading an existing matrix from a file or by building a new matrix. Up to six matrices can be available on your transmitter, but only one can be used for measurement at any given time. See Micro Motion Enhanced Density Application: Theory, Configuration, and Use for detailed information on building a matrix.
Prerequisites
Before you can configure concentration measurement:
The concentration measurement application must be enabled on your transmitter.
The concentration matrix you want to use must be available on your transmitter, or
it must be available as a file on your computer.
You must know the derived variable that your matrix is designed for.
You must know the density unit used by your matrix.
You must know the temperature unit used by your matrix.
The concentration measurement application must be unlocked.
Procedure
1.
Choose ProLink > Configuration > Density and set Density Units to the density unit used by your matrix.
2. Choose ProLink > Configuration > Temperature and set Temp Units to the temperature unit
used by your matrix.
3. Choose ProLink > Configuration > CM Setup.
4. In Global Config, set Derived Variable to the derived variable that your matrix is
designed for.
Important
All concentration matrices on your transmitter must use the same derived variable. If you
are using one of the standard matrices from Micro Motion, set Derived Variable to Mass Conc (Density). If you are using a custom matrix, see the reference information for your matrix.
If you change the setting of Derived Variable, all existing concentration matrices will be
deleted from transmitter memory. Set Derived Variable before loading concentration matrices.
5. Load one or more matrices.
62 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 71
Configure process measurement
a. In Curve Specific Config, set Curve Configured to the location to which the matrix
will be loaded.
b.
Click Load this curve from a file, navigate to the matrix file on your PC, and load it.
c. Repeat until all required matrices are loaded.
6. Set up extrapolation alarms.
Each concentration matrix is built for a specific density range and a specific temperature range. If process density or process temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may affect accuracy. Extrapolation alarms are used to notify the operator that extrapolation is occurring.
a. In Curve Specific Config, set Curve Configured to the matrix that you want to
configure.
b. Set Alarm Limit to the point, in percent, at which an extrapolation alarm will be
posted.
c. Enable or disable the high and low limit alarms for temperature and density, as
desired.
Restriction
The high and low limit alarms require the enhanced core processor.
Example: If Alarm Limit is set to 5%, Enable Temp High is checked, and the matrix is built for a temperature range of 40 °F to 80 °F, an extrapolation alarm will be posted if process temperature goes above 82 °F
7.
Select the label that will be used for the concentration unit.
a. In Curve Specific Config, set Curve Configured to the matrix that you want to
configure.
b. Select the desired label from the Units list.
c. If you set Units to Special, enter the custom label.
8. Determine how the transmitter will obtain temperature data for the concentration measurement calculations, and perform the required setup.
Option Setup
Temperature data from the sensor
A user-configured static temperature value
a. Choose View > Preferences . b.
Disable Use External Temperature.
a. Choose View > Preferences . b.
Enable Use External Temperature. c. Choose ProLink > Configuration > Temperature. d. Set External Temperature to the value to be used.
Configuration and Use Manual 63
Page 72
Configure process measurement
Option Setup
Polling for tempera-
(7)
ture
a. Ensure that the primary mA output has been wired to support
HART polling. b.
Choose View > Preferences . c. Enable Use External Temperature. d. Choose ProLink > Configuration > Polled Variables. e. Choose an unused polling slot. f. Set Polling Control to Poll As Primary or Poll as Secondary, and click
Apply. g. Set External Tag to the HART tag of the external temperature de-
vice. h. Set Variable Type to External Temperature.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
A value written by digital communica­tions
a. Choose View > Preferences . b.
Enable Use External Temperature. c. Perform the necessary host programming and communications
setup to write temperature data to the transmitter at appropri-
ate intervals.
Note
If the Weights & Measures application is implemented and the trans­mitter is secured, digital communications cannot be used to write temperature data to the transmitter.
9. In Global Config, set Active Curve to the matrix to be used for process measurement.
Concentration process variables are now available on the transmitter. You can view and report them in the same way that you view and report other process variables.
4.8.2 Configure concentration measurement using ProLink III
This task guides you through loading and setting up a concentration matrix to use for measurement. It does not cover building a concentration matrix.
Note
Concentration matrices can be made available on your transmitter either by loading an existing matrix from a file or by building a new matrix. Up to six matrices can be available on your transmitter, but only one can be used for measurement at any given time. See Micro Motion Enhanced Density
Application: Theory, Configuration, and Use for detailed information on building a matrix.
(7) Not available on all transmitters.
64 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 73
Configure process measurement
Prerequisites
Before you can configure concentration measurement:
The concentration measurement application must be enabled on your transmitter.
The concentration matrix you want to use must be available on your transmitter, or
it must be available as a file on your computer.
You must know the derived variable that your matrix is designed for.
You must know the density unit used by your matrix.
You must know the temperature unit used by your matrix.
The concentration measurement application must be unlocked.
Procedure
1.
Choose Device Tools > Configuration > Process Measurement > Density and set Density Unit to the density unit used by your matrix.
2. Choose Device Tools > Configuration > Process Measurement > Temperature and set Temperature Unit to the temperature unit used by your matrix.
3. Choose Device Tools > Configuration > Process Measurement > Concentration Measurement.
4. Set Derived Variable to the derived variable that your matrix is designed for, and click Apply.
Important
All concentration matrices on your transmitter must use the same derived variable. If you
are using one of the standard matrices from Micro Motion, set Derived Variable to Mass Concentration (Density). If you are using a custom matrix, see the reference information for your matrix.
If you change the setting of Derived Variable, all existing concentration matrices will be
deleted from transmitter memory. Set Derived Variable before loading concentration matrices.
5. Load one or more matrices.
a.
Set Matrix Being Configured to the location to which the matrix will be loaded.
b. Click Load Matrix from a File, navigate to the matrix file on your computer, and load
it.
c. Repeat until all required matrices are loaded.
6. Review and configure matrix data.
a. If necessary, set Matrix Being Configured to the matrix you want to view, and click
Change Matrix.
b. Set Concentration Unit to the label that will be used for the concentration unit.
c. If you set Concentration Unit to Special, enter the custom label.
d. If desired, change the matrix name.
e. Review the data points for this matrix.
f. Do not change Reference Temperature or Curve Fit Maximum Order.
Configuration and Use Manual 65
Page 74
Configure process measurement
7.
g. If you changed any matrix data, click Apply.
Set up extrapolation alarms.
Each concentration matrix is built for a specific density range and a specific temperature range. If process density or process temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may affect accuracy. Extrapolation alarms are used to notify the operator that extrapolation is occurring.
a. If necessary, set Matrix Being Configured to the matrix you want to view, and click
Change Matrix.
b. Set Extrapolation Alarm Limit to the point, in percent, at which an extrapolation
alarm will be posted.
c. Enable or disable the high and low limit alarms for temperature and density, as
desired, and click Apply.
Restriction
The high and low limit alarms require the enhanced core processor.
Example: If Extrapolation Alarm Limit is set to 5%, High Extrapolation Limit (Temperature) is enabled, and the matrix is built for a temperature range of 40 °F to 80 °F, an extrapolation alarm will be posted if process temperature goes above 82 °F
8.
Set Temperature Source to the method that the transmitter will use to obtain temperature data.
Option Description
Poll for external value
RTD The transmitter will use the temperature data from the sen-
Static or Digital Communications The transmitter will use the temperature value that it reads
(8)
The transmitter will poll an external temperature device, us­ing HART protocol over the primary mA output.
sor.
from memory.
Static: The configured value is used.
Digital Communications: A host writes transmitter data
to transmitter memory.
Note
If the Weights & Measures application is implemented and the transmitter is secured, digital communications cannot be used to write temperature data to the transmitter.
9. If you chose RTD, no more configuration is required. Click Apply and exit.
10.
If you chose to poll for temperature data:
(8) Not available on all transmitters.
66 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 75
Configure process measurement
a. Select the Polling Slot to use.
b.
Set Polling Control to Poll as Primary or Poll as Secondary, and click Apply.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network. The Field Communicator
is not a HART master.
c. Set External Device Tag to the HART tag of the external temperature device, and
click Apply.
11.
If you chose to use a static temperature value, set External Temperature to the value to use, and click Apply.
12. If you want to use digital communications, click Apply, then perform the necessary host programming and communications setup to write temperature data to the transmitter at appropriate intervals.
13. Set Active Matrix to the matrix to be used for measurement.
Concentration process variables are now available on the transmitter. You can view and report them in the same way that you view and report other process variables.
4.8.3 Configure concentration measurement using the Field Communicator
This task guides you through setting up a concentration matrix to use for measurement. It does not cover loading or building a concentration matrix.
Note
Concentration matrices can be made available on your transmitter either by loading an existing matrix from a file or by building a new matrix. Up to six matrices can be available on your transmitter, but only one can be used for measurement at any given time. See Micro Motion Enhanced Density Application: Theory, Configuration, and Use for detailed information on building a matrix.
Prerequisites
Before you can configure concentration measurement:
The concentration measurement application must be enabled on your transmitter.
You must know the derived variable that your matrix is designed for.
You must know the density unit used by your matrix.
You must know the temperature unit used by your matrix.
The concentration measurement application must be unlocked.
Procedure
1.
Choose Online > Configure > Manual Setup > Measurements > Density and set Density Unit to match the density unit used by your matrix.
Configuration and Use Manual 67
Page 76
Configure process measurement
2. Choose Online > Configure > Manual Setup > Measurements > Temperature and set
3.
4. Set up extrapolation alerts.
Temperature Unit to match the temperature unit used by your matrix.
Choose Online > Configure > Manual Setup > Measurements > Conc Measurement (CM) > CM Configuration.
Each concentration matrix is built for a specific density range and a specific temperature range. If process density or process temperature goes outside the range, the transmitter will extrapolate concentration values. However, extrapolation may affect accuracy. Extrapolation alerts are used to notify the operator that extrapolation is occurring.
a. Choose Online > Configure > Manual Setup > Measurements > Conc Measurement (CM) >
Matrix Configuration.
b. Set Matrix Being Configured to the matrix that you want to configure.
c. Set Extrapolation Alert Limit to the point, in percent, at which an extrapolation alert
will be posted.
d. Choose Online > Configure > Alert Setup > CM Alerts.
e. Enable or disable the high and low limit alarms for temperature and density, as
desired.
Restriction
The high and low limit alarms require the enhanced core processor.
Example: If Alarm Limit is set to 5%, the high-temperature extrapolation alert is enabled, and the matrix is built for a temperature range of 40 °F to 80 °F, an extrapolation alarm will be posted if process temperature goes above 82 °F
5.
Select the label that will be used for the concentration unit.
a. Choose Online > Configure > Manual Setup > Measurements > Conc Measurement (CM) >
Matrix Configuration.
b. Set Matrix Being Configured to the matrix that you want to configure.
c. Set Concentration Units to the desired label.
d. If you set Units to Special, enter the custom label.
6. Determine how the transmitter will obtain temperature data for the concentration
measurement calculations, and perform the required setup.
Option Setup
Temperature data from the sensor
A user-configured static temperature value
a. Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > Temperature.
b.
Disable External Temperature.
a. Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > Temperature.
b.
Enable External Temperature.
c. Set Correction Temperature to the value to be used.
68 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 77
Option Setup
Polling for tempera-
(9)
ture
a. Ensure that the primary mA output has been wired to support
HART polling.
b.
Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > Temperature. c. Enable External Temperature. d. Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > External Polling. e. Set Poll Control to Poll As Primary Host or Poll as Secondary Host. f. Choose an unused polling slot. g. Set External Tag to the HART tag of the external temperature de-
vice. h. Set Polled Variable to Temperature.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
Configure process measurement
A value written by digital communica­tions
a. Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > Temperature.
b.
Enable External Temperature. c. Perform the necessary host programming and communications
setup to write temperature data to the transmitter at appropri-
ate intervals.
Note
If the Weights & Measures application is implemented and the trans­mitter is secured, digital communications cannot be used to write temperature data to the transmitter.
7. Online > Configure > Manual Setup > Measurements > Conc Measurement (CM) > CM Configuration and set Active Matrix to the matrix to be used for measurement.
Concentration process variables are now available on the transmitter. You can view and report them in the same way that you view and report other process variables.
4.8.4 Standard matrices for the concentration measurement application
The standard concentration matrices available from Micro Motion are applicable for a variety of process fluids.
See Table 4-14 for a list of the standard concentration matrices available from Micro Motion, along with the density and temperature measurement units used in calculation, and the unit used to report concentration data.
(9) Not available on all transmitters.
Configuration and Use Manual 69
Page 78
Configure process measurement
Tip
If the standard matrices are not appropriate for your application, you can build a custom matrix or purchase a custom matrix from Micro Motion.
Standard concentration matrices and associated measurement unitsTable 4-14:
Matrix name Description Density unit
Deg Balling Matrix represents percent extract, by
g/cm
3
mass, in solution, based on °Balling. For example, if a wort is 10 °Balling and the extract in solution is 100% su­crose, the extract is 10% of the total mass.
Deg Brix Matrix represents a hydrometer scale
g/cm
3
for sucrose solutions that indicates the percent by mass of sucrose in solution at a given temperature. For example, 40 kg of sucrose mixed with 60 kg of water results in a 40 °Brix solution.
Deg Plato Matrix represents percent extract, by
g/cm
3
mass, in solution, based on °Plato. For example, if a wort is 10 °Plato and the extract in solution is 100% sucrose, the extract is 10% of the total mass.
HFCS 42 Matrix represents a hydrometer scale
g/cm
3
for HFCS 42 (high-fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
HFCS 55 Matrix represents a hydrometer scale
g/cm
3
for HFCS 55 (high-fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
HFCS 90 Matrix represents a hydrometer scale
g/cm
3
for HFCS 90 (high-fructose corn syrup) solutions that indicates the percent by mass of HFCS in solution.
Temperature unit
Concentration unit
°F °Balling
°C °Brix
°F °Plato
°C %
°C %
°C %
4.8.5 Derived variables and calculated process variables
For each derived variable, the concentration measurement application calculates a different set of process variables.
70 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 79
Derived variables and calculated process variables Table 4-15:
Derived Variable Description
Density at reference temperature
Specific gravity The ratio of the density
Mass concentration derived from refer­ence density
Mass concentration derived from specif­ic gravity
Volume concentra­tion derived from reference density
Volume concentra­tion derived from specific gravity
Concentration de­rived from reference density
Mass/unit volume, cor­rected to a given refer­ence temperature
of a process fluid at a given temperature to the density of water at a given temperature. The two given temper­ature conditions do not need to be the same.
The percent mass of solute or of material in suspension in the total solution, derived from reference density
The percent mass of solute or of material in suspension in the total solution, derived from specific gravity
The percent volume of solute or of material in suspension in the total solution, derived from reference density
The percent volume of solute or of material in suspension in the total solution, derived from specific gravity
The mass, volume, weight, or number of moles of solute or of material in suspension in proportion to the to­tal solution, derived from reference density
Configure process measurement
Calculated process variables
Density at reference tempera­ture
Standard volume flow rate
Specific gravity
Concen­tration
Net mass flow rate
Net vol­ume flow rate
Configuration and Use Manual 71
Page 80
Configure process measurement
Derived variables and calculated process variables (continued)Table 4-15:
Calculated process variables
Density at
Derived Variable Description
Concentration de­rived from specific gravity
The mass, volume, weight, or number of moles of solute or of material in suspension in proportion to the to­tal solution, derived from specific gravity
reference tempera­ture
Standard volume flow rate
Specific gravity
Concen­tration

4.9 Configure pressure compensation

Net mass flow rate
Net vol­ume flow rate
Pressure compensation adjusts process measurement to compensate for the pressure effect on the sensor. The pressure effect is the change in the sensor’s sensitivity to flow and density caused by the difference between the calibration pressure and the process pressure.
Tip
Not all sensors or applications require pressure compensation. The pressure effect for a specific sensor model can be found in the product data sheet located at www.micromotion.com. If you are uncertain about implementing pressure compensation, contact Micro Motion customer service.
4.9.1 Configure pressure compensation using ProLink II
Prerequisites
You will need the flow factor, density factor, and calibration pressure values for your sensor.
For the flow factor and density factor, see the product data sheet for your sensor.
For the calibration pressure, see the calibration sheet for your sensor. If the data is
unavailable, use 20 PSI.
Procedure
Choose View > Preferences and ensure that Enable External Pressure Compensation is
1.
checked.
2. Choose ProLink > Configuration > Pressure.
3. Enter Flow Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign.
72 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 81
Configure process measurement
Example:
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
4.
Enter Density Factor for your sensor.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the value, reverse the sign.
Example:
If the density factor is 0.000006 g/cm3/PSI, enter −0.000006 g/cm3/PSI.
5. Enter Cal Pressure for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and defines the pressure at which there is no pressure effect. If the data is unavailable, enter 20 PSI.
6. Determine how the transmitter will obtain pressure data, and perform the required
setup.
Option Setup
A user-configured static pressure val­ue
Polling for pres-
(10)
sure
a. Set Pressure Units to the desired unit. b.
Set External Pressure to the desired value.
a. Ensure that the primary mA output has been wired to support
HART polling.
b.
Choose ProLink > Configuration > Polled Variables. c. Choose an unused polling slot. d. Set Polling Control to Poll As Primary or Poll as Secondary, and click
Apply. e. Set External Tag to the HART tag of the external pressure device. f. Set Variable Type to Pressure.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
(10) Not available on all transmitters.
Configuration and Use Manual 73
Page 82
Configure process measurement
Option Setup
A value written by digital communica­tions
a. Set Pressure Units to the desired unit. b.
Perform the necessary host programming and communications
setup to write pressure data to the transmitter at appropriate in-
tervals.
Note
If the Weights & Measures application is implemented and the trans­mitter is secured, digital communications cannot be used to write temperature data to the transmitter.
Postrequisites
If you are using an external pressure value, verify the setup by choosing ProLink > Process Variables and checking the value displayed in External Pressure.
4.9.2 Configure pressure compensation using ProLink III
Prerequisites
You will need the flow factor, density factor, and calibration pressure values for your sensor.
For the flow factor and density factor, see the product data sheet for your sensor.
For the calibration pressure, see the calibration sheet for your sensor. If the data is
unavailable, use 20 PSI.
Procedure
1.
Choose Device Tools > Configuration > Process Measurement > Pressure Compensation.
2. Set Pressure Compensation Status to Enabled.
3. Enter Flow Calibration Pressure for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and defines the pressure at which there is no pressure effect. If the data is unavailable, enter 20 PSI.
4. Enter Flow Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign.
Example:
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
5. Enter Density Factor for your sensor.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the value, reverse the sign.
74 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 83
Configure process measurement
Example:
If the density factor is 0.000006 g/cm3/PSI, enter −0.000006 g/cm3/PSI.
6.
Set Pressure Source to the method that the transmitter will use to obtain pressure data.
Option Description
Poll for external value
Static or Digital Communications The transmitter will use the pressure value that it reads from
(11)
The transmitter will poll an external pressure device, using HART protocol over the primary mA output.
memory.
Static: The configured value is used.
Digital Communications: A host writes transmitter data
to transmitter memory.
Note
If the Weights & Measures application is implemented and the transmitter is secured, digital communications cannot be used to write temperature data to the transmitter.
7. If you chose to poll for pressure data:
a.
Select the Polling Slot to use.
b. Set Polling Control to Poll as Primary or Poll as Secondary, and click Apply.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network. The Field Communicator
is not a HART master.
c. Set External Device Tag to the HART tag of the external pressure device, and click
Apply.
8.
If you chose to use a static pressure value:
a. Set Pressure Unit to the desired unit.
b. Set Static or Current Pressure to the value to use, and click Apply
9. If you want to use digital communications, click Apply, then perform the necessary host programming and communications setup to write pressure data to the transmitter at appropriate intervals.
Postrequisites
If you are using an external pressure value, verify the setup by checking the External Pressure value displayed in the Inputs area of the main window.
(11) Not available on all transmitters.
Configuration and Use Manual 75
Page 84
Configure process measurement
4.9.3 Configure pressure compensation using the Field Communicator
Prerequisites
You will need the flow factor, density factor, and calibration pressure values for your sensor.
For the flow factor and density factor, see the product data sheet for your sensor.
For the calibration pressure, see the calibration sheet for your sensor. If the data is
unavailable, use 20 PSI.
Procedure
Choose Online > Configure > Manual Setup > Measurements > External Pressure/Temperature >
1.
Pressure.
2. Set Pressure Compensation to Enabled.
3. Enter Flow Cal Pressure for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and defines the pressure at which there is no pressure effect. If the data is unavailable, enter 20 PSI.
4. Enter Flow Press Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign.
Example:
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
5. Enter Dens Press Factor for your sensor.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the value, reverse the sign.
Example:
If the density factor is 0.000006 g/cm3/PSI, enter −0.000006 g/cm3/PSI.
6. Determine how the transmitter will obtain pressure data, and perform the required
setup.
Option Setup
A user-configured static pressure val­ue
a. Set Pressure Unit to the desired unit. b.
Set Compensation Pressure to the desired value.
76 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 85
Option Setup
Polling for pres-
(12)
sure
a. Ensure that the primary mA output has been wired to support
HART polling.
b.
Choose Online > Configure > Manual Setup > Measurements > External
Pressure/Temperature > External Polling. c. Set Poll Control to Poll As Primary Host or Poll as Secondary Host. d. Choose an unused polling slot. e. Set External Tag to the HART tag of the external pressure device. f. Set Polled Variable to Pressure.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
Configure process measurement
A value written by digital communica­tions
a. Set Pressure Unit to the desired unit. b.
Perform the necessary host programming and communications
setup to write pressure data to the transmitter at appropriate in-
tervals.
Note
If the Weights & Measures application is implemented and the trans­mitter is secured, digital communications cannot be used to write temperature data to the transmitter.
Postrequisites
If you are using an external pressure value, verify the setup by choosing Service Tools > Variables > External Variables and checking the value displayed for External Pressure.
4.9.4 Options for Pressure Measurement Unit
The transmitter provides a standard set of measurement units for Pressure Measurement Unit. Different communications tools may use different labels for the units. In most applications, Pressure Measurement Unit should be set to match the pressure measurement unit used by the remote device.
Options for Pressure Measurement UnitTable 4-16:
Label
Display (stand-
Unit description
Feet water @ 68 °F FTH2O Ft Water @ 68°F Ft Water @ 68°F Ft Water @ 68°F ftH2O
Inches water @ 4 °C INW4C In Water @ 4°C In Water @ 4°C In Water @ 4°C inH2O @4DegC
(12) Not available on all transmitters.
ard)
Chinese-lan­guage display
ProLink II ProLink III Field Commu-
nicator
Configuration and Use Manual 77
Page 86
Configure process measurement
Options for Pressure Measurement Unit (continued)Table 4-16:
Label
Display (stand-
Unit description
Inches water @ 60 °F INW60 In Water @ 60°F In Water @ 60°F In Water @ 60°F inH2O @60DegF
Inches water @ 68 °F INH2O In Water @ 68°F In Water @ 68°F In Water @ 68°F inH2O
Millimeters water @ 4 °C mmW4C mm Water @
Millimeters water @ 68 °F mmH2O mm Water @
Millimeters mercury @ 0 °C
Inches mercury @ 0 °C INHG In Mercury @
Pounds per square inch PSI PSI PSI PSI psi
Bar BAR bar bar bar bar
Millibar mBAR millibar millibar millibar mbar
Grams per square centi­meter
Kilograms per square cen­timeter
Pascals PA Pascals pascals pascals Pa
Kilopascals KPA Kilopascals Kilopascals Kilopascals kPa
Megapascals MPA Megapascals megapascals Megapascals MPa
Torr @ 0 °C TORR Torr @ 0°C Torr @ 0°C Torr @ 0°C torr
Atmospheres ATM Atmospheres atms atms atms
ard)
mmHG mm Mercury @
G/SCM g/cm2 g/cm2 g/cm2 g/Sqcm
KG/SCM kg/cm2 kg/cm2 kg/cm2 kg/Sqcm
Chinese-lan­guage display
4°C
68°F
0°C
0°C
ProLink II ProLink III Field Commu-
nicator
mm Water @ 4°C
mm Water @ 68°F
mm Mercury @ 0°C
In Mercury @ 0°C
mm Water @ 4°C mmH2O @4DegC
mm Water @ 68°F mmH2O
mm Mercury @ 0°C
In Mercury @ 0°C inHG
mmHg
78 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 87

Configure device options and preferences

5 Configure device options and
preferences
Topics covered in this chapter:

Configure the transmitter display

Enable or disable operator actions from the display
Configure security for the display menus
Configure response time parameters
Configure alarm handling
Configure informational parameters
5.1 Configure the transmitter display
You can control the process variables shown on the display and a variety of display behaviors.
The transmitter display parameters include:
Display Language
Display Variables
Display Precision
Update Period
Auto Scroll and Auto Scroll Rate
Backlight
Status LED Blinking
5.1.1 Configure the language used for the display
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > LANG
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Language
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Language
Offline Maintain > Configuration > Display > Language
Overview
Display Language controls the language used for process data and menus on the display.
Configuration and Use Manual 79
Page 88
Configure device options and preferences
Procedure
Select the language you want to use.
Tip
For devices with the Chinese-language display, you can use a shortcut key, or an optical switch combination, to change the language without having to access the display menu. The optical switch combination is shown on the front of the display.
The languages available depend on your transmitter model and version.
5.1.2 Configure the process variables shown on the display
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Display
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Display Variables
Not available
Overview
You can control the process variables shown on the display and the order in which they appear. The display can scroll through up to 15 process variables in any order you choose. In addition, you can repeat variables or leave slots unassigned.
Restrictions
You cannot set Display Variable 1 to None. Display Variable 1 must be set to a process variable.
If you have configured Display Variable 1 to track the primary mA output, you cannot change
the setting of Display Variable 1 using this procedure. To change the setting of Display Variable 1, you must change the configuration of mA Output Process Variable for the primary mA output.
Note
If you configure a display variable as a volume process variable and then change Volume Flow Type, the display variable is automatically changed to the equivalent process variable. For example, Volume Flow Rate would be changed to Gas Standard Volume Flow Rate.
Procedure
For each display variable you want to change, assign the process variable you want to use.
80 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 89
Configure device options and preferences
Example: Default display variable configuration
Display variable Process variable assignment
Display Variable 1 Mass flow
Display Variable 2 Mass total
Display Variable 3 Volume flow
Display Variable 4 Volume total
Display Variable 5 Density
Display Variable 6 Temperature
Display Variable 7 External pressure
Display Variable 8 Mass flow
Display Variable 9 None
Display Variable 10 None
Display Variable 11 None
Display Variable 12 None
Display Variable 13 None
Display Variable 14 None
Display Variable 15 None
Configure Display Variable 1 to track the primary mA output
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLY > VAR 1
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options > Display Variable 1 Selection
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Not available
Overview
You can configure Display Variable 1 to track mA Output Process Variable for the primary mA output. When tracking is enabled, you can control Display Variable 1 from the display menu.
Tip
This feature is the only way to configure a display variable from the display menus (standard display option only), and it applies only to Display Variable 1.
Procedure
Configure Display Variable 1 to track the primary mA output.
Offline Maintain > Configuration > Display > Configure 1st Var
Configuration and Use Manual 81
Page 90
Configure device options and preferences
Display Variable 1 will automatically be set to match mA Output Process Variable for the primary mA output. If you change the configuration of mA Output Process Variable, Display Variable 1 will be updated automatically.
5.1.3 Configure the precision of variables shown on the display
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Precision
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Decimal Places
Overview
Setting Display Precision determines the precision (number of decimal places) shown on the display. You can set Display Precision independently for each variable. Setting Display Precision does not affect the actual value of the process variable.
Offline Maintain > Configuration > Display > Num of Decimals
Procedure
1.
Select a process variable.
2. Set Display Precision to the number of decimal places you want shown when the
process variable appears on the display.
For temperature and density process variables, the default value is 2 decimal places. For all other process variables, the default value is 4 decimal places. The range is 0 to 5.
Tip
The lower the selected precision, the greater the process change must be for it to be reflected on the display. Do not set Display Precision value too low or too high to be useful.
5.1.4 Configure the refresh rate of data shown on the display
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > RATE
Chinese-language display
ProLink II ProLink > Configuration > Display > Update Period
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Refresh Rate
Offline Maintain > Configuration > Display > Display Rate
82 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 91
Configure device options and preferences
Overview
You can set Update Period to control how frequently data is refreshed on the display.
Procedure
Set Update Period to the desired value.
The default value is 200 milliseconds. The range is 100 milliseconds to 10,000 milliseconds (10 seconds).
5.1.5 Enable or disable automatic scrolling through the display variables
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > AUTO SCRLL
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options > Display Auto Scroll
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Auto Scroll
Offline Maintain > Configuration > Display > Auto Scroll
Overview
You can configure the display to automatically scroll through the configured display variables or to show a single display variable until the operator activates Scroll. When you set automatic scrolling, you can also configure the length of time each display variable is displayed.
Procedure
1.
Enable or disable Auto Scroll as desired.
Option Description
Enabled The display automatically scrolls through each display variable as specified
by Scroll Rate. The operator can move to the next display variable at any time using Scroll.
Disabled (de­fault)
The display shows Display Variable 1 and does not scroll automatically. The operator can move to the next display variable at any time using Scroll.
2. If you enabled Auto Scroll, set Scroll Rate as desired.
The default value is 10 seconds.
Tip
Scroll Rate may not be available until you apply Auto Scroll.
Configuration and Use Manual 83
Page 92
Configure device options and preferences
5.1.6 Enable or disable the display backlight
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > BKLT
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options > Display Backlight On/Off
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Backlight
Overview
You can enable or disable the display backlight.
Procedure
Enable or disable Backlight.
The default setting is Enabled.
Offline Maintain > Configuration > Display > Backlight
5.1.7 Enable or disable Status LED Blinking
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options > Display Status LED Blinking
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Status LED Blinking
Overview
By default, the status LED blinks (flashes) to indicate unacknowledged alarms. If you disable Status LED Blinking, the status LED does not blink, whether alarms are acknowledged or not. It still changes color to indicate active alarms.
Procedure
Enable or disable Status LED Blinking.
The default setting is Enabled.
Not available
84 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 93
Configure device options and preferences

5.2 Enable or disable operator actions from the display

You can configure the transmitter to let the operator perform specific actions using the display.
You can configure the following:
Totalizer Start/Stop
Totalizer Reset
Acknowledge All Alarms
5.2.1 Enable or disable Totalizer Start/Stop from the display
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > TOTALS STOP
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options > Display Start/Stop Totalizers
ProLink III Device Tools > Configuration > Totalizer Control Methods
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Start/Stop Totalizers
Offline Maintain > Configuration > Display > Start Totals
Offline Maintain > Configuration > Display > Stop Totals
Overview
You can control whether or not the operator is able to start and stop totalizers and inventories from the display.
Restrictions
You cannot start and stop totalizers individually from the display. All totalizers are started or
stopped together.
You cannot start or stop inventories separately from totalizers. When a totalizer is started or
stopped, the associated inventory is also started or stopped.
If the petroleum measurement application is installed on your computer, the operator must
enter the off-line password to perform this function, even if the off-line password is not enabled.
Procedure
1.
Ensure that at least one totalizer is configured as a display variable.
2. Enable or disable Totalizer Reset as desired.
Option Description
Enabled Operators can start and stop totalizers and inventories from the display,
if at least one totalizer is configured as a display variable.
Configuration and Use Manual 85
Page 94
Configure device options and preferences
Option Description
Disabled (default) Operators cannot start and stop totalizers and inventories from the dis-
play.
5.2.2 Enable or disable Totalizer Reset from the display
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > TOTALS RESET
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options > Display Totalizer Reset
ProLink III Device Tools > Configuration > Totalizer Control Methods
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Totalizer Reset
Overview
Offline Maintain > Configuration > Display > Reset Total
You can configure whether or not the operator is able to reset totalizers from the display.
Restrictions
This parameter does not apply to inventories. You cannot reset inventories from the display.
You cannot use the display to reset all totalizers as a group. You must reset totalizers
individually.
If the petroleum measurement application is installed on your computer, the operator must
enter the off-line password to perform this function, even if the off-line password is not enabled.
Procedure
1.
Ensure that the totalizers you want to reset have been configured as display variables.
If the totalizer is not configured as a display variable, the operator will not be able to reset it.
2. Enable or disable resetting the totalizer as desired.
Option Description
Enabled Operators can reset a totalizer from the display, if the totalizer is config-
ured as a display variable.
Disabled (default) Operators cannot reset totalizers from the display.
86 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 95
Configure device options and preferences
5.2.3 Enable or disable the Acknowledge All Alarms display command
Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > ALARM
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options > Display Ack All Alarms
ProLink III Device Tools > Configuration > Transmitter Display > Ack All
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features > Acknowledge All
Overview
You can configure whether or not the operator can use a single command to acknowledge all alarms from the display.
Procedure
1.
Ensure that the alarm menu is accessible from the display.
Offline Maintain > Configuration > Display > Acknowledge All
To acknowledge alarms from the display, operators must have access to the alarm menu.
2. Enable or disable Acknowledge All Alarms as desired.
Option Description
Enabled (default) Operators can use a single display command to acknowledge all alarms
at once.
Disabled Operators cannot acknowledge all alarms at once, they must be ac-
knowledged individually.

5.3 Configure security for the display menus

Display (standard) OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY
Chinese-language display
ProLink II ProLink > Configuration > Display > Display Options
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features
Offline Maintain > Configuration > Display
Configuration and Use Manual 87
Page 96
Configure device options and preferences
Overview
You can control operator access to different sections of the display off-line menu. You can also configure a password to control access.
Procedure
1.
To control operator access to the maintenance section of the off-line menu, enable or disable Off-Line Menu.
Option Description
Enabled (de­fault)
Disabled Operator cannot access the maintenance section of the off-line menu.
2. To control operator access to the alarm menu, enable or disable Alarm Menu.
Option Description
Enabled (default) Operator can access the alarm menu. This access is required to view and
Disabled Operator cannot access the alarm menu.
Operator can access the maintenance section of the off-line menu. This access is required for configuration and calibration, but is not required to view alarms or to access Smart Meter Verification (if applicable).
acknowledge alarms, but is not required for Smart Meter Verification (if applicable), configuration, or calibration.
Note
The transmitter status LED changes color to indicate that there are active alarms, but does not show specific alarms.
3. To require a password for access to the maintenance section of the off-line menu
and the Smart Meter Verification menu, enable or disable Off-Line Password.
Option Description
Enabled Operator is prompted for the off-line password at entry to the Smart Meter
Verification menu (if applicable) or entry to the maintenance section of the off-line menu.
Disabled (de­fault)
No password is required for entry to the Smart Meter Verification menu (if applicable) or entry to the maintenance section of the off-line menu.
4. To require a password to access the alarm menu, enable or disable Alarm Password.
Option Description
Enabled Operator is prompted for the off-line password at entry to the alarm
menu.
Disabled (default) No password is required for entry to the alarm menu.
88 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 97
Configure device options and preferences
If both Off-Line Password and Alarm Password are enabled, the operator is prompted for the off-line password to access the off-line menu, but is not prompted thereafter.
5.
(Optional) Set Off-Line Password to the desired value.
The same value is used for both the off-line password and the alarm password. The default value is 1234. The range is 0000 to 9999.
Tip
Record your password for future reference.

5.4 Configure response time parameters

You can configure the rate at which process data is polled and process variables are calculated.
Response time parameters include:
Update Rate
Calculation Speed (Response Time)
5.4.1 Configure Update Rate
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Device > Update Rate
ProLink III Device Tools > Configuration > Process Measurement > Response > Update Rate
Field Communicator Configure > Manual Setup > Measurements > Update Rate
Overview
Update Rate controls the rate at which process data is polled and process variables are calculated. Update Rate = Special produces faster and “noisier” response to changes in the process. Do not use Special mode unless required by your application.
Tip
For systems with a standard core processor, Special mode can improve performance for applications with entrained air or Empty-Full-Empty conditions. This does not apply to systems with an enhanced core processor.
Prerequisites
Not available
Before setting Update Rate to Special:
Check the effects of Special mode on specific process variables.
Configuration and Use Manual 89
Page 98
Configure device options and preferences
Contact Micro Motion.
Procedure
1.
Set Update Rate as desired.
Option Description
Normal All process data is polled at the rate of 20 times per second (20 Hz).
Special A single, user-specified process variable is polled at the rate of 100 times per sec-
If you change Update Rate, the settings for Flow Damping, Density Damping, and Temperature Damping are automatically adjusted.
All process variables are calculated at 20 Hz.
This option is appropriate for most applications.
ond (100 Hz). Other process data is polled at 6.25 Hz). Some process, diagnostic, and calibration data is not polled.
All available process variables are calculated at 100 Hz.
Use this option only if required by your application.
2.
If you set Update Rate to Special, select the process variable to be polled at 100 Hz.
Effects of Update Rate = Special
Incompatible features and functions
Special mode is not compatible with the following features and functions:
Enhanced events. Use basic events instead.
All calibration procedures.
Zero verification.
Restoring the factory zero or the prior zero.
If required, you can switch to Normal mode, perform the desired procedures, and then return to Special mode.
Process variable updates
Some process variables are not updated when Special mode is enabled.
90 Micro Motion® Model 2700 Transmitters with Analog Outputs
Page 99
Configure device options and preferences
Table 5-1:
Always polled and updated
Mass flow
Volume flow
Gas standard volume flow
Density
Temperature
Drive gain
LPO amplitude
Status [contains Event 1 and Event
2 (basic events)]
Raw tube frequency
Mass total
Volume total
Gas standard volume total
Temperature-corrected volume
total
Temperature-corrected density
Temperature-corrected volume
flow
Batch-weighted average tempera-
ture
Batch-weighted average density
Special mode and process variable updates
Updated only when the petroleum measurement application is disa­bled Never updated
RPO amplitude
Board temperature
Core input voltage
Mass inventory
Volume inventory
Gas standard volume inventory
All other process variables and calibra­tion data. They retain the values held at the time you enabled Special mode.
5.4.2 Configure Calculation Speed (Response Time)
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Device > Response Time
ProLink III Device Tools > Configuration > Process Measurement > Response > Calculation Speed
Field Communicator Not available
Overview
Calculation Speed is used to apply a different algorithm to the calculation of process variables from the raw process data. Calculation Speed = Special produces faster and “noisier” response to changes in the process.
In ProLink II, Calculation Speed is called Response Time.
Restriction
Calculation Speed is available only on systems with the enhanced core processor.
Not available
Configuration and Use Manual 91
Page 100
Configure device options and preferences
Tip
You can use Calculation Speed = Special with either setting of Update Rate. The parameters control different aspects of flowmeter processing.
Procedure
Set Calculation Speed as desired.
Option Description
Normal Transmitter calculates process variables at the standard speed.
Special Transmitter calculates process variables at a faster speed.

5.5 Configure alarm handling

The alarm handling parameters control the transmitter’s response to process and device conditions.
Alarm handling parameters include:
Fault Timeout
Status Alarm Severity
5.5.1 Configure Fault Timeout
Display (standard) Not available
Chinese-language display
ProLink II ProLink > Configuration > Analog Output > Last Measured Value Timeout
ProLink III Device Tools > Configuration > Fault Processing
Field Communicator Configure > Alert Setup > Alert Severity > Fault Timeout
Overview
Fault Timeout controls the delay before fault actions are performed.
Restriction
Fault Timeout is applied only to the following alarms (listed by Status Alarm Code): A003, A004, A005, A008, A016, A017, A033. For all other alarms, fault actions are performed as soon as the alarm is detected.
Not available
ProLink > Configuration > Frequency/Discrete Output > Frequency > Last Measured Value Timeout
92 Micro Motion® Model 2700 Transmitters with Analog Outputs