Micro Motion 1700 Operating Manual

Configuration and Use Manual
MMI-20019028, Rev AB
March 2018
Micro Motion® Model 1700 Transmitters with Analog Outputs
Configuration and Use Manual
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step.

Other information

Full product specifications can be found in the product data sheet. Troubleshooting information can be found in the configuration manual. Product data sheets and manuals are available from the Micro Motion web site at www.emerson.com.

Return policy

Follow Micro Motion procedures when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Micro Motion will not accept your returned equipment if you fail to follow Micro Motion procedures.
Return procedures and forms are available on our web support site at www.emerson.com, or by phoning the Micro Motion Customer Service department.

Emerson Flow customer service

Email:
Worldwide: flow.support@emerson.com
Asia-Pacific: APflow.support@emerson.com
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 704 136 666 New Zealand 099 128 804
Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684

Contents

Contents
Part I Getting started
Chapter 1 Before you begin ............................................................................................................. 3
1.1 About this manual ......................................................................................................................... 3
1.2 Transmitter model code ................................................................................................................ 3
1.3 Communications tools and protocols ............................................................................................ 4
1.4 Additional documentation and resources ...................................................................................... 4
Chapter 2 Quick start .......................................................................................................................5
2.1 Power up the transmitter ...............................................................................................................5
2.2 Check meter status ........................................................................................................................6
2.2.1 Transmitter status reported by LED .................................................................................6
2.3 Make a startup connection to the transmitter ................................................................................7
2.4 (Optional) Adjust digital communications settings ........................................................................7
2.5 Verify mass flow measurement ......................................................................................................7
2.6 Verify the zero ............................................................................................................................... 8
2.6.1 Terminology used with zero verification and zero calibration ..........................................9
Part II Configuration and commissioning
Chapter 3 Introduction to configuration and commissioning ......................................................... 13
3.1 Configuration flowchart .............................................................................................................. 13
3.2 Default values and ranges ............................................................................................................15
3.3 Enable access to the off-line menu of the display ......................................................................... 15
3.4 Disable write-protection on the transmitter configuration .......................................................... 15
3.5 Restore the factory configuration ................................................................................................ 16
Chapter 4 Configure process measurement ................................................................................... 17
4.1 Configure mass flow measurement ............................................................................................. 17
4.1.1 Configure Mass Flow Measurement Unit ...................................................................... 17
4.1.2 Configure Flow Damping ..............................................................................................20
4.1.3 Configure Mass Flow Cutoff ..........................................................................................21
4.2 Configure volume flow measurement for liquid applications ....................................................... 22
4.2.1 Configure Volume Flow Type for liquid applications ......................................................23
4.2.2 Configure Volume Flow Measurement Unit for liquid applications ................................ 23
4.2.3 Configure Volume Flow Cutoff ..................................................................................... 26
4.3 Configure GSV flow measurement ...............................................................................................27
4.3.1 Configure Volume Flow Type for gas applications ......................................................... 28
4.3.2 Configure Standard Density of Gas ...............................................................................28
4.3.3 Configure Gas Standard Volume Flow Unit ...................................................................30
4.3.4 Configure Gas Standard Volume Flow Cutoff ................................................................33
4.4 Configure Flow Direction ............................................................................................................ 34
4.4.1 Options for Flow Direction ........................................................................................... 35
4.5 Configure density measurement .................................................................................................39
4.5.1 Configure Density Measurement Unit .......................................................................... 39
Configuration and Use Manual i
Contents
4.5.2 Configure two-phase flow parameters ..........................................................................40
4.5.3 Configure Density Damping .........................................................................................42
4.5.4 Configure Density Cutoff ..............................................................................................43
4.6 Configure temperature measurement .........................................................................................44
4.6.1 Configure Temperature Measurement Unit ..................................................................44
4.6.2 Configure Temperature Damping ................................................................................ 44
4.6.3 Effect of Temperature Damping on process measurement ........................................... 45
4.6.4 Configure Temperature Input ...................................................................................... 45
4.7 Configure pressure compensation ...............................................................................................46
4.7.1 Configure pressure compensation using ProLink III ...................................................... 46
4.7.2 Configure pressure compensation using the Field Communicator ................................47
4.7.3 Options for Pressure Measurement Unit .......................................................................49
Chapter 5 Configure device options and preferences ..................................................................... 51
5.1 Configure the transmitter display ................................................................................................ 51
5.1.1 Configure the language used for the display ................................................................. 51
5.1.2 Configure the process variables and diagnostic variables shown on the display .............51
5.1.3 Configure the number of decimal places (precision) shown on the display .................... 53
5.1.4 Configure the refresh rate of data shown on the display ................................................ 54
5.1.5 Enable or disable automatic scrolling through the display variables .............................. 54
5.1.6 Enable or disable the display backlight .......................................................................... 55
5.1.7 Enable or disable Status LED Blinking ............................................................................55
5.2 Enable or disable operator actions from the display .....................................................................56
5.2.1 Enable or disable Totalizer Start/Stop from the display ..................................................56
5.2.2 Enable or disable Totalizer Reset from the display ......................................................... 57
5.2.3 Enable or disable the Acknowledge All Alerts display command .................................... 57
5.3 Configure security for the display menus .....................................................................................58
5.4 Configure response time parameters .......................................................................................... 59
5.4.1 Configure Update Rate ................................................................................................. 60
5.4.2 Configure Response Time ............................................................................................. 61
5.5 Configure alert handling ..............................................................................................................62
5.5.1 Configure Fault Timeout .............................................................................................. 62
5.5.2 Configure Status Alert Severity .....................................................................................63
5.6 Configure informational parameters ........................................................................................... 67
5.6.1 Configure Sensor Serial Number ...................................................................................67
5.6.2 Configure Sensor Material ............................................................................................ 67
5.6.3 Configure Sensor Liner Material ....................................................................................68
5.6.4 Configure Sensor Flange Type ...................................................................................... 68
5.6.5 Configure Descriptor ....................................................................................................69
5.6.6 Configure Message ...................................................................................................... 69
5.6.7 Configure Date .............................................................................................................69
Chapter 6 Integrate the meter with the control system ..................................................................71
6.1 Configure the transmitter channels ............................................................................................. 71
6.2 Configure the mA Output ............................................................................................................72
6.2.1 Configure mA Output Process Variable ........................................................................ 72
6.2.2 Configure Lower Range Value (LRV) and Upper Range Value (URV) ............................... 74
6.2.3 Configure AO Cutoff .....................................................................................................75
6.2.4 Configure Added Damping ...........................................................................................77
6.2.5 Configure mA Output Fault Action and mA Output Fault Level ..................................... 78
6.3 Configure the Frequency Output ................................................................................................. 79
6.3.1 Configure Frequency Output Polarity ........................................................................... 80
ii Micro Motion Model 1700 Transmitters with Analog Outputs
Contents
6.3.2 Configure Frequency Output Scaling Method ...............................................................81
6.3.3 Configure Frequency Output Fault Action and Frequency Output Fault Level ............... 82
6.4 Configure the Discrete Output .................................................................................................... 83
6.4.1 Configure Discrete Output Source ............................................................................... 84
6.4.2 Configure Discrete Output Polarity .............................................................................. 86
6.4.3 Configure Discrete Output Fault Action ........................................................................86
6.5 Configure events ......................................................................................................................... 87
6.5.1 Configure a basic event ................................................................................................. 88
6.5.2 Configure an enhanced event ....................................................................................... 88
6.6 Configure digital communications .............................................................................................. 90
6.6.1 Configure HART/Bell 202 communications .................................................................. 90
6.6.2 Configure HART/RS-485 communications .................................................................... 95
6.6.3 Configure Modbus/RS-485 communications ................................................................ 96
6.6.4 Configure Digital Communications Fault Action ........................................................... 98
Chapter 7 Complete the configuration .........................................................................................101
7.1 Test or tune the system using sensor simulation ........................................................................101
7.1.1 Sensor simulation ....................................................................................................... 102
7.2 Back up transmitter configuration ............................................................................................. 103
7.3 Enable write-protection on the transmitter configuration ......................................................... 103
Part III Operations, maintenance, and troubleshooting
Chapter 8 Transmitter operation ................................................................................................. 107
8.1 Record the process variables ..................................................................................................... 107
8.2 View process variables ...............................................................................................................108
8.2.1 View process variables using the display .....................................................................108
8.2.2 View process variables and other data using ProLink III ............................................... 109
8.2.3 View process variables using the Field Communicator ................................................109
8.3 View transmitter status using the status LED ............................................................................. 110
8.4 View and acknowledge status alerts .......................................................................................... 111
8.4.1 View and acknowledge alerts using the display .......................................................... 111
8.4.2 View and acknowledge alerts using ProLink III ............................................................ 113
8.4.3 View alerts using the Field Communicator ................................................................. 114
8.5 Read totalizer and inventory values ........................................................................................... 114
8.6 Start and stop totalizers and inventories ....................................................................................114
8.6.1 Start and stop totalizers and inventories using the display ..........................................115
8.7 Reset totalizers ..........................................................................................................................116
8.7.1 Reset totalizers using the display ................................................................................116
8.8 Reset inventories .......................................................................................................................118
Chapter 9 Measurement support ................................................................................................. 119
9.1 Options for measurement support ............................................................................................ 119
9.2 Use Smart Meter Verification (SMV) .......................................................................................... 120
9.2.1 SMV requirements ...................................................................................................... 120
9.2.2 SMV test preparation .................................................................................................. 120
9.2.3 Run SMV ..................................................................................................................... 121
9.2.4 View test data ............................................................................................................. 125
9.2.5 Schedule automatic execution of the SMV test ........................................................... 129
9.3 Use PVR, TBR, and TMR ..............................................................................................................132
9.3.1 PVR, TBR, and TMR applications .................................................................................. 133
9.4 Piecewise linearization (PWL) for calibrating gas meters ............................................................134
Configuration and Use Manual iii
Contents
9.5 Zero the meter .......................................................................................................................... 134
9.6 Validate the meter .....................................................................................................................135
9.6.1 Alternate method for calculating the meter factor for volume flow .............................136
9.7 Perform a (standard) D1 and D2 density calibration ...................................................................137
9.7.1 Perform a D1 and D2 density calibration using ProLink III ............................................ 138
9.7.2 Perform a D1 and D2 density calibration using the Field Communicator ..................... 139
9.8 Perform a D3 and D4 density calibration (T-Series sensors only) ................................................ 140
9.8.1 Perform a D3 or D3 and D4 density calibration using ProLink III .................................. 140
9.8.2 Perform a D3 or D3 and D4 density calibration using the Field Communicator ........... 141
9.9 Perform temperature calibration ...............................................................................................142
9.9.1 Perform temperature calibration using the display ..................................................... 143
9.9.2 Perform temperature calibration using ProLink III .......................................................143
9.9.3 Perform temperature calibration using the Field Communicator ................................ 145
Chapter 10 Troubleshooting .......................................................................................................... 147
10.1 Status LED states ....................................................................................................................... 148
10.2 Status alerts, causes, and recommendations ............................................................................. 148
10.3 Flow measurement problems ................................................................................................... 159
10.4 Density measurement problems ............................................................................................... 161
10.5 Temperature measurement problems .......................................................................................162
10.6 Milliamp output problems ......................................................................................................... 163
10.7 Frequency Output problems ......................................................................................................164
10.8 Using sensor simulation for troubleshooting ............................................................................. 165
10.9 Check power supply wiring ........................................................................................................165
10.10 Check sensor-to-transmitter wiring ........................................................................................... 166
10.11 Check grounding ....................................................................................................................... 167
10.12 Perform loop tests ..................................................................................................................... 167
10.12.1 Perform loop tests using the display ...........................................................................167
10.12.2 Perform loop tests using ProLink III .............................................................................169
10.12.3 Perform loop tests using the Field Communicator ......................................................170
10.13 Check the HART communication loop ....................................................................................... 171
10.14 Check HART Address and mA Output Action ............................................................................. 172
10.15 Check HART burst mode ............................................................................................................173
10.16 Check the trimming of the mA Output ...................................................................................... 173
10.17 Check Lower Range Value and Upper Range Value ....................................................................173
10.18 Check mA Output Fault Action ..................................................................................................173
10.19 Check for radio frequency interference (RFI) ..............................................................................174
10.20 Check Frequency Output Scaling Method ................................................................................. 174
10.21 Check Frequency Output Fault Action .......................................................................................174
10.22 Check Flow Direction ................................................................................................................ 175
10.23 Check the cutoffs ...................................................................................................................... 175
10.24 Check for two-phase flow (slug flow) ......................................................................................... 175
10.25 Check the drive gain .................................................................................................................. 176
10.25.1 Collect drive gain data ................................................................................................ 177
10.26 Check the pickoff voltage .......................................................................................................... 177
10.26.1 Collect pickoff voltage data ........................................................................................ 178
10.27 Check for internal electrical problems ....................................................................................... 178
10.27.1 Check the sensor coils .................................................................................................179
10.28 Check the core processor LED ....................................................................................................181
10.28.1 Core processor LED states ...........................................................................................184
10.29 Perform a 700 core processor resistance test .............................................................................186
iv Micro Motion Model 1700 Transmitters with Analog Outputs
Contents
Appendices and reference
Appendix A Using the transmitter display .......................................................................................189
A.1 Components of the transmitter interface .................................................................................. 189
A.2 Use the optical switches ............................................................................................................ 190
A.3 Access and use the display menu system ................................................................................... 191
A.3.1 Enter a floating-point value using the display .............................................................. 192
A.4 Display codes for process variables ............................................................................................ 195
A.5 Codes and abbreviations used in display menus ........................................................................ 196
Appendix B Using ProLink III with the transmitter ...........................................................................201
B.1 Basic information about ProLink III ............................................................................................201
B.2 Connect with ProLink III ............................................................................................................ 202
B.2.1 Connection types supported by ProLink III ..................................................................202
B.2.2 Connect with ProLink III to the service port ................................................................. 203
B.2.3 Make a HART/Bell 202 connection .............................................................................. 204
B.2.4 Make a HART/RS-485 connection ................................................................................209
B.2.5 Connect with ProLink III to the RS-485 port ................................................................. 212
Appendix C Using a Field Communicator with the transmitter ........................................................217
C.1 Basic information about the Field Communicator ..................................................................... 217
C.2 Connect with the Field Communicator ..................................................................................... 218
Appendix D Default values and ranges ............................................................................................ 221
D.1 Default values and ranges ..........................................................................................................221
Appendix E Transmitter components and installation wiring ......................................................... 227
E.1 Installation types ....................................................................................................................... 227
E.2 Power supply terminals and ground .......................................................................................... 230
E.3 Input/output (I/O) wiring terminals ........................................................................................... 231
Appendix F NE 53 history ............................................................................................................... 233
F.1 NE 53 history ............................................................................................................................. 233
Configuration and Use Manual v
Contents
vi Micro Motion Model 1700 Transmitters with Analog Outputs
Part I
Getting started
Chapters covered in this part:
Before you begin
Quick start
Getting started
Configuration and Use Manual 1
Getting started
2 Micro Motion Model 1700 Transmitters with Analog Outputs
1 Before you begin
Topics covered in this chapter:

About this manual

Transmitter model code

Communications tools and protocols
Additional documentation and resources
1.1 About this manual
This manual helps you configure, commission, use, maintain, and troubleshoot Micro Motion Model 1700 transmitters with analog outputs.
Important
This manual assumes that the following conditions apply:
The transmitter has been installed correctly and completely according to the instructions in
the transmitter installation manual
The installation complies with all applicable safety requirements
The user is trained in local and corporate safety standards

Before you begin

1.2
Transmitter model code
You can verify that this manual pertains to your transmitter by ensuring the model code on the transmitter tag matches the format.
Example:
The transmitter has a model number of the following form: 1700(R/I/E/B/C/M/P)**A******
R
4-wire remote-mount with aluminum housing
I
Integral mount
E
4-wire remote mount transmitter with 9-wire remote enhanced core processor
B
4-wire remote mount transmitter with 9-wire remote core processor
C
9-wire remote-mount with integral core processor and aluminum housing
M
4-wire remote mount with stainless steel housing
P
9-wire remote mount transmitter with integral core processor and stainless steel housing
A
Analog outputs option board
Configuration and Use Manual 3
Before you begin

1.3 Communications tools and protocols

You can use several different communications tools and protocols to interface with the transmitter, use different tools in different locations, or use different tools for different tasks.
Tool Supported protocols
ProLink III HART/RS-485
HART/Bell 202
Modbus/RS-485
Service port
Field Communicator HART/Bell 202
For information about how to use the communication tools, see the appendices in this manual.
Tip
You may be able to use other communications tools, such as AMS Suite: Intelligent Device Manager, or the Smart Wireless THUM™ Adapter. Use of AMS or the Smart Wireless THUM Adapter is not discussed in this manual. For more information on the Smart Wireless THUM Adapter, refer to the documentation available at www.emerson.com.
1.4

Additional documentation and resources

Topic Document
Hazardous area installa­tion
Product Data Sheet
Sensor Sensor documentation
Transmitter installation
All documentation resources are available at www.emerson.com or on the user documentation DVD.
See the approval documentation shipped with the transmitter, or download the appropriate documentation at www.emerson.com.
Micro Motion Series 1000 and Series 2000 Transmitters with MVD™ Tech‐ nology Product Data Sheet
Micro Motion® Model 1700 and 2700 Installation Manual
4 Micro Motion Model 1700 Transmitters with Analog Outputs
2 Quick start
Topics covered in this chapter:

Power up the transmitter

Check meter status
Make a startup connection to the transmitter
(Optional) Adjust digital communications settings
Verify mass flow measurement
Verify the zero
2.1 Power up the transmitter
The transmitter must be powered up for all configuration and commissioning tasks, or for process measurement.
1. Ensure that all transmitter and sensor covers and seals are closed.

Quick start

DANGER!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers and seals are tightly closed. For hazardous area installations, applying power while housing covers are removed or loose can cause an explosion.
2. Turn on the electrical power at the power supply.
The transmitter will automatically perform diagnostic routines. The transmitter is self-switching and will automatically detect the supply voltage. When using DC power, a minimum of 1.5 amps of startup current is required. During this period, Alert 009 is active. The diagnostic routines should complete in approximately 30 seconds. For transmitters with a display, the status LED will turn green and begin to flash when the startup diagnostics are complete. If the status LED exhibits different behavior, an alert is active.
Postrequisites
Although the sensor is ready to receive process fluid shortly after power-up, the electronics can take up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if power has been off long enough to allow components to reach ambient temperature, allow the electronics to warm up for approximately 10 minutes before relying on process measurements. During this warm-up period, you may observe minor measurement instability or inaccuracy.
Configuration and Use Manual 5
Quick start

2.2 Check meter status

Check the meter for any error conditions that require user action or that affect measurement accuracy.
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alert A009 is active. This alert should clear automatically when the power-up sequence is complete.
2. Check the status LED on the transmitter.
Related information
View and acknowledge status alerts
2.2.1 Transmitter status reported by LED
Transmitter status reported by status LEDTable 2-1:
LED state Description Recommendation
Solid green No alerts are active. Continue with configuration or process meas-
urement.
Flashing green (if ena­bled)
Solid yellow One or more low-severity alerts are active. A low-severity alert condition does not affect
Flashing yellow (if ena­bled)
Solid red One or more high-severity alerts are active. A high-severity alert condition affects meas-
Flashing red (if ena­bled)
Unacknowledged corrected condition (no alert)
Calibration in progress, or Known Density Ver­ification in progress.
One or more low-severity alerts are active and have not been acknowledged.
One or more high-severity alerts are active and have not been acknowledged.
Continue with configuration or process meas­urement. Acknowledge the alert if you choose.
measurement accuracy or output behavior. You can continue with configuration or proc­ess measurement, but Micro Motion still rec­ommends identifying and resolving the alert condition.
A low-severity alert condition does not affect measurement accuracy or output behavior. You can continue with configuration or proc­ess measurement, but Micro Motion still rec­ommends identifying and resolving the alert condition.
urement accuracy and output behavior. Re­solve the alert condition before continuing.
A high-severity alert condition affects meas­urement accuracy and output behavior. Re­solve the alert condition before continuing. Acknowledge the alert if you choose.
If Status LED Blinking is disabled, all LEDs will show a solid color rather than flashing.
6 Micro Motion Model 1700 Transmitters with Analog Outputs
Quick start

2.3 Make a startup connection to the transmitter

For all configuration tools except the display, you must have an active connection to the transmitter to configure the transmitter. Follow this procedure to make your first connection to the transmitter.
Identify the connection type to use, and follow the instructions for that connection type in the appropriate appendix. Use the default communications parameters shown in the appendix.
Communications tool Connection type to use Instructions
ProLink III HART/RS-485
Modbus/RS-485
Service port
Field Communicator HART/Bell 202 Appendix C
Appendix B

2.4 (Optional) Adjust digital communications settings

Change the communications parameters to site-specific values.
Important
If you are changing communications parameters for the connection type that you are using, you will lose the connection when you write the parameters to the transmitter. Reconnect using the new parameters.
Procedure
1. To change the communications parameters using ProLink III, choose Device Tools >
Configuration > Communications.
2. To change the communications parameters using the Field Communicator, choose
On-Line Menu > Configure > Manual Setup > Inputs/Outputs > Communications.
2.5

Verify mass flow measurement

Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method.
Read the value for Mass Flow Rate on the transmitter display.
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in
the Process Variables panel.
Connect to the transmitter with the Field Communicator and read the value for Mass
Flow Rate.
Configuration and Use Manual 7
Quick start
On-Line Menu > Overview > Primary Purpose Variables
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.

2.6 Verify the zero

Verifying the zero helps you determine if the stored zero value is appropriate to your installation, or if a field zero can improve measurement accuracy.
The zero verification procedure analyzes the Live Zero value under conditions of zero flow, and compares it to the Zero Stability range for the sensor. If the average Live Zero value is within a reasonable range, the zero value stored in the transmitter is valid. Performing a field calibration will not improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Procedure
1. Allow the flowmeter to warm up for at least 20 minutes after applying power.
2. Run the process fluid through the sensor until the sensor temperature reaches the
normal process operating temperature.
3. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
4. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
5. From ProLink III, choose Device Tools > Calibration > Zero Verification and
Calibration > Verify Zero and wait until the procedure completes.
6. Observe the drive gain, temperature, and density readings. If they are stable, check
the Live Zero or Field Verification Zero value. If the average value is close to 0, you should not need to zero the meter.
7. If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Remove or reduce sources of electromechanical noise if appropriate.
8 Micro Motion Model 1700 Transmitters with Analog Outputs
d. Repeat the zero verification procedure.
e. If it fails again, zero the meter.
Postrequisites
Restore normal flow through the sensor by opening the valves.
Related information
Zero the meter
2.6.1 Terminology used with zero verification and zero calibration
Term Definition
Zero In general, the offset required to synchronize the left pickoff and the right
pickoff under conditions of zero flow. Unit = microseconds.
Factory Zero The zero value obtained at the factory, under laboratory conditions.
Field Zero The zero value obtained by performing a zero calibration outside the fac-
tory.
Prior Zero The zero value stored in the transmitter at the time a field zero calibration
is begun. May be the factory zero or a previous field zero.
Manual Zero The zero value stored in the transmitter, typically obtained from a zero
calibration procedure. It may also be configured manually. Also called “mechanical zero” or “stored zero”.
Live Zero The real-time bidirectional mass flow rate with no flow damping or mass
flow cutoff applied. An adaptive damping value is applied only when the mass flow rate changes dramatically over a very short interval. Unit = con­figured mass flow measurement unit.
Zero Stability A laboratory-derived value used to calculate the expected accuracy for a
sensor. Under laboratory conditions at zero flow, the average flow rate is expected to fall within the range defined by the Zero Stability value (0 ± Zero Stability). Each sensor size and model has a unique Zero Stability val­ue. Statistically, 95% of all data points should fall within the range defined by the Zero Stability value.
Zero Calibration The procedure used to determine the zero value.
Zero Time The time period over which the Zero Calibration procedure is performed.
Unit = seconds.
Field Verification Zero A 3-minute running average of the Live Zero value, calculated by the
transmitter. Unit = configured mass flow measurement unit.
Zero Verification A procedure used to evaluate the stored zero and determine whether or
not a field zero can improve measurement accuracy.
Quick start
Configuration and Use Manual 9
Quick start
10 Micro Motion Model 1700 Transmitters with Analog Outputs
Configuration and commissioning
Part II
Configuration and commissioning
Chapters covered in this part:
Introduction to configuration and commissioning
Configure process measurement
Configure device options and preferences
Integrate the meter with the control system
Complete the configuration
Configuration and Use Manual 11
Configuration and commissioning
12 Micro Motion Model 1700 Transmitters with Analog Outputs

Introduction to configuration and commissioning

3 Introduction to configuration and
commissioning
Topics covered in this chapter:

Configuration flowchart

Default values and ranges
Enable access to the off‐line menu of the display
Disable write‐protection on the transmitter configuration
Restore the factory configuration
3.1 Configuration flowchart
Use the following flowchart as a general guide to the configuration and commissioning process.
Some options may not apply to your installation. Detailed information is provided in the remainder of this manual. If you are using the Weights & Measures application, additional configuration and setup are required.
Configuration and Use Manual 13
Introduction to configuration and commissioning
Configuration flowchartFigure 3-1:
Configure process measurement
Configure mass flow
measurement
Configure volume flow
meaurement
Volume flow type
Liquid
Configure flow direction
Configure density
measurement
Configure temperature
measurement
Gas
Define gas properties
Configure device options and preferences
Configure display
parameters
Configure fault handling
parameters
Configure sensor
parameters
Configure device
parameters
Integrate device with control system
Configure the channel(s)
Configure the mA
output(s)
Test and move to production
Test or tune transmitter using sensor simulation
Back up transmitter
configuration
Enable write-protection on
transmitter configuration
Done
Configure pressure
compensation (optional)
Configure PVR, TMR,
or TBR (if available)
Configure the frequency
output(s)
Configure the discrete
output(s)
Configure events
Configure digital communications
14 Micro Motion Model 1700 Transmitters with Analog Outputs
Introduction to configuration and commissioning

3.2 Default values and ranges

See Section D.1 to view the default values and ranges for the most commonly used parameters.

3.3 Enable access to the off-line menu of the display

Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features
Overview
By default, access to the off-line menu of the display is enabled. If it is disabled, you must enable it if you want to use the display to configure the transmitter.
3.4
Restriction
You cannot use the display to enable access to the off-line menu. You must make a connection from another tool.

Disable write-protection on the transmitter configuration

Display OFF-LINE MAINT > CONFG > LOCK
ProLink III Device Tools > Configuration > Write-Protection
Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Write Protect
Overview
If the transmitter is write-protected, the configuration is locked and you must unlock it before you can change any configuration parameters. By default, the transmitter is not write-protected.
Tip
Write-protecting the transmitter prevents accidental changes to configuration. It does not prevent normal operational use. You can always disable write-protection, perform any required configuration changes, then re-enable write-protection.
Configuration and Use Manual 15
Introduction to configuration and commissioning

3.5 Restore the factory configuration

Display Not available
ProLink III Device Tools > Configuration Transfer > Restore Factory Configuration
Field Communicator Service Tools > Maintenance > Reset/Restore > Restore Factory Configuration
Overview
Restoring the factory configuration returns the transmitter to a known operational configuration. This may be useful if you experience problems during configuration.
Important
You cannot restore factory configurations with a 700 core.
Tip
Restoring the factory configuration is not a common action. You may want to contact customer support to see if there is a preferred method to resolve any issues.
16 Micro Motion Model 1700 Transmitters with Analog Outputs

Configure process measurement

4 Configure process measurement
Topics covered in this chapter:

Configure mass flow measurement

Configure volume flow measurement for liquid applications
Configure GSV flow measurement
Configure Flow Direction
Configure density measurement
Configure temperature measurement
Configure pressure compensation
4.1 Configure mass flow measurement
The mass flow measurement parameters control how mass flow is measured and reported.
4.1.1 Configure Mass Flow Measurement Unit
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > MASS
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Unit
Overview
Mass Flow Measurement Unit specifies the unit of measure that will be used for the mass flow rate. The unit used for mass total and mass inventory is derived from this unit.
Any selected measurement unit, (mass, volume or gas standard volume), is automatically applied to both the mA and Frequency Outputs.
Procedure
Set Mass Flow Measurement Unit to the unit you want to use.
The default setting for Mass Flow Measurement Unit is g/sec (grams per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Configuration and Use Manual 17
Configure process measurement
Options for Mass Flow Measurement Unit
The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Unit description
Grams per second
Grams per minute
Grams per hour
Kilograms per second
Kilograms per minute
Kilograms per hour
Kilograms per day
Metric tons per minute
Metric tons per hour
Metric tons per day
Pounds per second
Pounds per minute
Pounds per hour
Pounds per day
Short tons (2000 pounds) per minute
Short tons (2000 pounds) per hour
Short tons (2000 pounds) per day
Long tons (2240 pounds) per hour
Long tons (2240 pounds) per day
Special unit
Label
Display ProLink III Field Communica-
tor
G/S g/sec g/s
G/MIN g/min g/min
G/H g/hr g/h
KG/S kg/sec kg/s
KG/MIN kg/min kg/min
KG/H kg/hr kg/h
KG/D kg/day kg/d
T/MIN mTon/min MetTon/min
T/H mTon/hr MetTon/h
T/D mTon/day MetTon/d
LB/S lbs/sec lb/s
LB/MIN lbs/min lb/min
LB/H lbs/hr lb/h
LB/D lbs/day lb/d
ST/MIN sTon/min STon/min
ST/H sTon/hr STon/h
ST/D sTon/day STon/d
LT/H lTon/hr LTon/h
LT/D lTon/day LTon/d
SPECL special Spcl
18 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Define a special measurement unit for mass flow
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Mass Special Units
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Although you cannot define a special measurement unit using the display, you can use the display to select an existing special measurement unit, and to view process data using the special measurement unit.
Procedure
1. Specify Base Mass Unit.
Base Mass Unit is the existing mass unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
3. Calculate Mass Flow Conversion Factor as follows:
a. x base units = y special units
b. Mass Flow Conversion Factor = x ÷ y
The original mass flow rate value is divided by this value.
4. Enter Mass Flow Conversion Factor.
5. Set Mass Flow Label to the name you want to use for the mass flow unit.
6. Set Mass Total Label to the name you want to use for the mass total and mass
inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for mass flow
You want to measure mass flow in ounces per second (oz/sec).
1. Set Base Mass Unit to Pounds (lb).
2. Set Base Time Unit to Seconds (sec).
3. Calculate Mass Flow Conversion Factor:
Configuration and Use Manual 19
Configure process measurement
a. 1 lb/sec = 16 oz/sec
b. Mass Flow Conversion Factor = 1 ÷ 16 = 0.0625
4. Set Mass Flow Conversion Factor to 0.0625.
5. Set Mass Flow Label to oz/sec.
6. Set Mass Total Label to oz.
4.1.2 Configure Flow Damping
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Flow Damping
Overview
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Procedure
Set Flow Damping to the value you want to use.
The default value is 0.8 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table.
Update Rate setting
Normal
Special
Damping range
0 to 51.2 seconds
0 to 40.96 seconds
The value you enter is automatically rounded off to the nearest valid value. For example, if the damping is currently set to 0.8 seconds, any value entered up to 1.2 seconds will be rounded down to 0.8 seconds, and any value entered from 1.21 to 1.59 seconds will be rounded up to 1.6 seconds.
Update Rate setting
Normal
Special
Valid damping values
0.0, 0.2, 0.4, 0.8, 1.6, 3.2, 6.4, 12.8, 25.6, 51.2
0.0, 0.04, 0.08, 0.16, 0.32, 0.64, 1.28, 2.56,
5.12, 10.24, 20.48, 40.96
20 Micro Motion Model 1700 Transmitters with Analog Outputs
Effect of flow damping on volume measurement
Flow damping affects volume measurement for liquid volume data. Flow damping also affects volume measurement for gas standard volume data. The transmitter calculates volume data from the damped mass flow data.
Interaction between Flow Damping and mA Output Damping
In some circumstances, both Flow Damping and mA Output Damping are applied to the reported mass flow value.
Flow Damping controls the rate of change in flow process variables. mA Output Damping controls the rate of change reported via the mA Output. If mA Output Process Variable is set to Mass Flow Rate, and both Flow Damping and mA Output Damping are set to non-zero values, flow damping is applied first, and the added damping calculation is applied to the result of the first calculation.
4.1.3 Configure Mass Flow Cutoff
Configure process measurement
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Mass Flow Cutoff
Overview
Mass Flow Cutoff specifies the lowest mass flow rate that will be reported as measured. All mass flow rates below this cutoff will be reported as 0.
Procedure
Set Mass Flow Cutoff to the value you want to use.
The default value for Mass Flow Cutoff is 0.0 g/sec or a sensor-specific value set at the factory. The recommended value is 0.5% of the nominal flow rate of the attached sensor. See the sensor specifications. Leaving Mass Flow Cutoff at 0.0 g/sec is not recommended.
Effect of Mass Flow Cutoff on volume measurement
Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data rather than the reported value.
Volume flow has a separate Volume Flow Cutoff that is not affected by the Mass Flow Cutoff value.
Configuration and Use Manual 21
Configure process measurement
Interaction between Mass Flow Cutoff and mA Output Cutoff
Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported via the mA output. If mA Output Process Variable is set to Mass Flow Rate, the mass flow rate reported via the mA Output is controlled by the higher of the two cutoff values.
Mass Flow Cutoff affects all reported values and values used in other transmitter behavior (e.g., events defined on mass flow).
mA Output Cutoff affects only mass flow values reported via the mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 10 g/sec
Mass Flow Cutoff: 15 g/sec
4.2
Result: If the mass flow rate drops below 15 g/sec, mass flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mA Output Cutoff higher than Mass Flow Cutoff
Configuration:
mA Output Process Variable: Mass Flow Rate
Frequency Output Process Variable: Mass Flow Rate
mA Output Cutoff: 15 g/sec
Mass Flow Cutoff: 10 g/sec
Result:
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
- The mA Output will report zero flow.
- The Frequency Output will report the actual flow rate, and the actual flow rate
will be used in all internal processing.
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow, and 0
will be used in all internal processing.

Configure volume flow measurement for liquid applications

The volume flow measurement parameters control how liquid volume flow is measured and reported.
22 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
Note
If you need to switch from gas standard volume to liquid volume, polling for base density will automatically be disabled.
4.2.1 Configure Volume Flow Type for liquid applications
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Liquid
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement will be used.
4.2.2
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type to Liquid if you are using any of the following applications:
Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to Liquid.
Configure Volume Flow Measurement Unit for liquid applications
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > VOL
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Unit
Overview
Volume Flow Measurement Unit specifies the unit of measurement that will be displayed for the volume flow rate. The unit used for the volume total and volume inventory is based on this unit.
Configuration and Use Manual 23
Configure process measurement
Prerequisites
Before you configure Volume Flow Measurement Unit, be sure that Volume Flow Type is set to Liquid.
Procedure
Set Volume Flow Measurement Unit to the unit you want to use.
The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
Options for Volume Flow Measurement Unit for liquid applications
The transmitter provides a standard set of measurement units for Volume Flow Measurement Unit, plus one user-defined measurement unit. Different communications tools may use different labels for the units.
Unit description
Cubic feet per second
Cubic feet per minute
Cubic feet per hour
Cubic feet per day
Cubic meters per second
Cubic meters per minute
Cubic meters per hour
Cubic meters per day
U.S. gallons per second
U.S. gallons per minute
U.S. gallons per hour
U.S. gallons per day
Million U.S. gallons per day
Liters per second
Liters per minute
Liters per hour
Million liters per day
Imperial gallons per second
Imperial gallons per minute
Label
Display ProLink III Field Communicator
CUFT/S ft3/sec Cuft/s
CUF/MN ft3/min Cuft/min
CUFT/H ft3/hr Cuft/h
CUFT/D ft3/day Cuft/d
M3/S m3/sec Cum/s
M3/MIN m3/min Cum/min
M3/H m3/hr Cum/h
M3/D m3/day Cum/d
USGPS US gal/sec gal/s
USGPM US gal/min gal/min
USGPH US gal/hr gal/h
USGPD US gal/day gal/d
MILG/D mil US gal/day MMgal/d
L/S l/sec L/s
L/MIN l/min L/min
L/H l/hr L/h
MILL/D mil l/day ML/d
UKGPS Imp gal/sec Impgal/s
UKGPM Imp gal/min Impgal/min
24 Micro Motion Model 1700 Transmitters with Analog Outputs
Label
Configure process measurement
Unit description
Imperial gallons per hour
Imperial gallons per day
Barrels per second
Barrels per minute
Barrels per hour
Barrels per day
Beer barrels per second
Beer barrels per minute
Beer barrels per hour
Beer barrels per day
(1)
(1)
(1)
(1)
(2)
(2)
(2)
(2)
Special unit
(1) Unit based on oil barrels (42 U.S. gallons). (2) Unit based on U.S. beer barrels (31 U.S. gallons).
Display ProLink III Field Communicator
UKGPH Imp gal/hr Impgal/h
UKGPD Imp gal/day Impgal/d
BBL/S barrels/sec bbl/s
BBL/MN barrels/min bbl/min
BBL/H barrels/hr bbl/h
BBL/D barrels/day bbl/d
BBBL/S Beer barrels/sec bbbl/s
BBBL/MN Beer barrels/min bbbl/min
BBBL/H Beer barrels/hr bbbl/h
BBBL/D Beer barrels/day bbbl/d
SPECL special Spcl
Define a special measurement unit for volume flow
Display
ProLink III Device Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Volume Special Units
Not available
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Although you cannot define a special measurement unit using the display, you can use the display to select an existing special measurement unit, and to view process data using the special measurement unit.
Procedure
1. Specify Base Volume Unit.
Base Volume Unit is the existing volume unit that the special unit will be based on.
2. Specify Base Time Unit.
Base Time Unit is the existing time unit that the special unit will be based on.
Configuration and Use Manual 25
Configure process measurement
3. Calculate Volume Flow Conversion Factor as follows:
4. Enter Volume Flow Conversion Factor.
5. Set Volume Flow Label to the name you want to use for the volume flow unit.
6. Set Volume Total Label to the name you want to use for the volume total and
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for volume flow
You want to measure volume flow in pints per second (pints/sec).
1. Set Base Volume Unit to Gallons (gal).
2. Set Base Time Unit to Seconds (sec).
a. x base units = y special units
b. Volume Flow Conversion Factor = x ÷ y
The original volume flow rate value is divided by this conversion factor.
volume inventory unit.
4.2.3
3. Calculate the conversion factor:
a. 1 gal/sec = 8 pints/sec
b. Volume Flow Conversion Factor = 1 ÷ 8 = 0.1250
4. Set Volume Flow Conversion Factor to 0.1250.
5. Set Volume Flow Label to pints/sec.
6. Set Volume Total Label to pints.
Configure Volume Flow Cutoff
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Volume Flow Cutoff
Overview
Volume Flow Cutoff specifies the lowest volume flow rate that will be reported as measured. All volume flow rates below this cutoff are reported as 0.
Procedure
Set Volume Flow Cutoff to the value you want to use.
The default value for Volume Flow Cutoff is 0.0 l/sec (liters per second). The lower limit is
0.
26 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Interaction between Volume Flow Cutoff and mAO Cutoff
Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured. mAO Cutoff defines the lowest flow rate that will be reported via the mA Output. If mA Output Process Variable is set to Volume Flow Rate, the volume flow rate reported via the mA Output is controlled by the higher of the two cutoff values.
Volume Flow Cutoff affects both the volume flow values reported via the outputs and the volume flow values used in other transmitter behavior (e.g., events defined on the volume flow).
mAO Cutoff affects only flow values reported via the mA Output.
Example: Cutoff interaction with mAO Cutoff lower than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 10 l/sec
Volume Flow Cutoff: 15 l/sec
4.3
Result: If the volume flow rate drops below 15 l/sec, volume flow will be reported as 0, and 0 will be used in all internal processing.
Example: Cutoff interaction with mAO Cutoff higher than Volume Flow Cutoff
Configuration:
mA Output Process Variable: Volume Flow Rate
Frequency Output Process Variable: Volume Flow Rate
AO Cutoff: 15 l/sec
Volume Flow Cutoff: 10 l/sec
Result:
If the volume flow rate drops below 15 l/sec but not below 10 l/sec:
- The mA Output will report zero flow.
- The Frequency Output will report the actual flow rate, and the actual flow rate
will be used in all internal processing.
If the volume flow rate drops below 10 l/sec, both outputs will report zero flow, and
0 will be used in all internal processing.

Configure GSV flow measurement

The gas standard volume (GSV) flow measurement parameters control how volume flow is measured and reported in a gas application.
Configuration and Use Manual 27
Configure process measurement
Restriction
You cannot implement both liquid volume flow and gas standard volume flow at the same time. Choose one or the other.
4.3.1 Configure Volume Flow Type for gas applications
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Volume Flow Type > Standard Gas
Volume
Overview
Volume Flow Type controls whether liquid or gas standard volume flow measurement is used.
Restriction
Gas standard volume measurement is incompatible with some applications. Set Volume Flow Type to Liquid if you are using any of the following applications:
4.3.2
Production Volume Reconciliation (PVR)
Procedure
Set Volume Flow Type to Gas Standard Volume.
Configure Standard Density of Gas
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > Gas Ref Density
Overview
The Standard Density of Gas value is the gas density at standard reference conditions. Use it to convert the measured mass flow data to volume flow at reference conditions.
Prerequisites
Ensure that Density Measurement Unit is set to the measurement unit you want to use for Standard Density of Gas.
Procedure
From the Source field, choose the method to supply gas base density data and perform the required setup.
28 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Option Description
Fixed Value or Digital Communications
Poll for external value The meter polls an external HART device for gas base density data in order
A host writes gas base density data to the meter at appropriate intervals.
Continue to Configure fixed value or digital communications.
to then compute gas standard volume from the mass flow and gas base density.
Continue to Poll for external value.
Configure fixed value or digital communications
Prerequisites
Section 4.3.2
Procedure
1. Set Standard Density of Gas to the standard reference density of the gas you are measuring.
Note
ProLink III provides a guided method that you can use to calculate your gas base density, if you do not know it.
2. Continue to Section 4.3.3.
Poll for external value
Prerequisites
Section 4.3.2
Procedure
1. Set Polling Slot to an available slot.
2. Set Polling Control n as one of the following options:
The n is the value you selected in the Polling Slot field.
If there is another master, and if that master is primary, then set this field to secondary. If the other master is secondary, then set this field to primary.
Option
Poll as Primary
Poll as Secondary
Description
No other HART masters will be on the network.
Other HART masters will be on the network.
3. Set External Device Tag n to the HART tag of the device being polled.
The n is the value you selected in the Polling Slot field.
Configuration and Use Manual 29
Configure process measurement
The device being polled (slave) cannot have special units set for density.
Otherwise, the master will reject the base density and report the following alarm:
A115: No External Input or Polled Data Alert
On the slave side, setup the HART Primary Variable for Base Density. The master
will reject anything other than Base Density for the HART Primary Variable and trigger an A115 alarm.
The density units on the transmitter and the polled device can be different as
long as they can be classified as density units; for example, kg/m3 and g/cm3. The transmitter converts the polled units into compatible specified units.
For wiring and setup instructions for a polled device, refer to the Micro Motion Gas
Density Meters (GDM) Installation manual or the Micro Motion Specific Gravity Meters (SGM) Installation manual.
4. Continue to Section 4.3.3.
4.3.3 Configure Gas Standard Volume Flow Unit
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > GSV
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > GSV Flow Unit
Overview
Gas Standard Volume Flow Unit specifies the unit of measure that will be displayed for the gas standard volume flow. The measurement unit used for the gas volume total and the gas volume inventory is derived from this unit.
Prerequisites
Before you configure Gas Standard Volume Flow Unit, be sure that Volume Flow Type is set to Gas Standard Volume.
For polling, the first transmitter (master) requests density from a second transmitter (slave) via HART communications. Special units for GSV are allowed on the master side, but the device being polled (slave) cannot have special units set for density, otherwise the master will reject the base density and report an A115: No External Input or Polled Data Alert.
Procedure
Set Gas Standard Volume Flow Unit to the unit you want to use.
The default setting for Gas Standard Volume Flow Unit is SCFM (Standard Cubic Feet per Minute).
Tip
If the measurement unit you want to use is not available, you can define a special measurement unit.
30 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Options for Gas Standard Volume Flow Unit
The transmitter provides a standard set of measurement units for Gas Standard Volume Flow Unit, plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Label
Unit description
Normal cubic meters per second
Normal cubic meters per minute
Normal cubic meters per hour
Normal cubic meters per day
Normal liters per second
Normal liters per minute
Normal liters per hour
Normal liters per day
Standard cubic feet per second
Standard cubic feet per minute
Standard cubic feet per hour
Standard cubic feet per day
Standard cubic meters per second
Standard cubic meters per minute
Standard cubic meters per hour
Standard cubic meters per day
Standard liters per second
Standard liters per minute
Standard liters per hour
Standard liters per day
Special measurement unit
Display ProLink III Field Communicator
NM3/S Nm3/sec Nm3/sec
NM3/MN Nm3/sec Nm3/min
NM3/H Nm3/hr Nm3/hr
NM3/D Nm3/day Nm3/day
NLPS NLPS NLPS
NLPM NLPM NLPM
NLPH NLPH NLPH
NLPD NLPD NLPD
SCFS SCFS SCFS
SCFM SCFM SCFM
SCFH SCFH SCFH
SCFD SCFD SCFD
SM3/S Sm3/sec Sm3/sec
SM3/MN Sm3/min Sm3/min
SM3/H Sm3/hr Sm3/hr
SM3/D Sm3/day Sm3/day
SLPS SLPS SLPS
SLPM SLPM SLPM
SLPH SLPH SLPH
SLPD SLPD SLPD
SPECL special Special
Define a special measurement unit for gas standard volume flow
Display
ProLink III Device Tools > Configuration > Process Measurement > Flow > Special Units
Field Communicator Configure > Manual Setup > Measurements > Special Units > Special GSV Units
Configuration and Use Manual 31
Not available
Configure process measurement
Overview
A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
Note
Although you cannot define a special measurement unit using the display, you can use the display to select an existing special measurement unit, and to view process data using the special measurement unit.
Procedure
1. Specify Base Gas Standard Volume Unit.
2. Specify Base Time Unit.
3. Calculate Gas Standard Volume Flow Conversion Factor as follows:
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be based on.
Base Time Unit is the existing time unit that the special unit will be based on.
a. x base units = y special units
b. Gas Standard Volume Flow Conversion Factor = x ÷ y
4. Enter the Gas Standard Volume Flow Conversion Factor.
The original gas standard volume flow value is divided by this conversion factor.
5. Set Gas Standard Volume Flow Label to the name you want to use for the gas standard volume flow unit.
6. Set Gas Standard Volume Total Label to the name you want to use for the gas standard volume total and gas standard volume inventory unit.
The special measurement unit is stored in the transmitter. You can configure the transmitter to use the special measurement unit at any time.
Example: Defining a special measurement unit for gas standard volume flow
You want to measure gas standard volume flow in thousands of standard cubic feet per minute.
1. Set Base Gas Standard Volume Unit to SCF.
2. Set Base Time Unit to minutes (min).
3. Calculate the conversion factor:
a. 1 thousands of standard cubic feet per minute = 1000 cubic feet per minute
b. Gas Standard Volume Flow Conversion Factor = 1 ÷ 1000 = 0.001 standard
4. Set Gas Standard Volume Flow Conversion Factor to 0.001.
5. Set Gas Standard Volume Flow Label to MSCFM.
32 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
6. Set Gas Standard Volume Total Label to MSCF.
4.3.4 Configure Gas Standard Volume Flow Cutoff
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > GSV > GSV Cutoff
Overview
Gas Standard Volume Flow Cutoff specifies the lowest gas standard volume flow rate that will reported as measured. All gas standard volume flow rates below this cutoff will be reported as 0.
Procedure
Set Gas Standard Volume Flow Cutoff to the value you want to use.
The default value for Gas Standard Volume Flow Cutoff is 0.0. The lower limit is 0.0. There is no upper limit.
Interaction between Gas Standard Volume Flow Cutoff and mA Output Cutoff
Gas Standard Volume Flow Cutoff defines the lowest Gas Standard Volume flow value that the transmitter will report as measured. mA Output Cutoff defines the lowest flow rate that will be reported via the mA Output. If mA Output Process Variable is set to Gas Standard Volume Flow Rate, the volume flow rate reported via the mA Output is controlled by the higher of the two cutoff values.
Gas Standard Volume Flow Cutoff affects both the gas standard volume flow values reported via outputs and the gas standard volume flow values used in other transmitter behavior (e.g., events defined on gas standard volume flow).
mA Output Cutoff affects only flow values reported via the mA Output.
Example: Cutoff interaction with mA Output Cutoff lower than Gas Standard Volume Flow Cutoff
Configuration:
mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow
Rate
Frequency Output Process Variable: Gas Standard Volume Flow Rate
mA Output Cutoff for the primary mA Output: 10 SLPM (standard liters per minute)
Gas Standard Volume Flow Cutoff: 15 SLPM
Result: If the gas standard volume flow rate drops below 15 SLPM, the volume flow will be reported as 0, and 0 will be used in all internal processing.
Configuration and Use Manual 33
Configure process measurement
Example: Cutoff interaction with mA Output Cutoff higher than Gas Standard Volume Flow Cutoff
Configuration:
mA Output Process Variable for the primary mA Output: Gas Standard Volume Flow
Frequency Output Process Variable: Gas Standard Volume Flow Rate
mA Output Cutoff for the primary mA Output: 15 SLPM (standard liters per minute)
Gas Standard Volume Flow Cutoff: 10 SLPM
Result:
If the gas standard volume flow rate drops below 15 SLPM but not below 10 SLPM:
If the gas standard volume flow rate drops below 10 SLPM, both outputs will report
Rate
- The primary mA Output will report zero flow.
- The Frequency Output will report the actual flow rate, and the actual flow rate
will be used in all internal processing.
zero flow, and 0 will be used in all internal processing.

4.4 Configure Flow Direction

Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Flow
Field Communicator Configure > Manual Setup > Measurements > Flow > Flow Direction
Overview
Flow Direction controls how forward flow and reverse flow affect flow measurement and reporting.
Flow Direction is defined with respect to the flow arrow on the sensor:
Forward flow (positive flow) moves in the direction of the flow arrow on the sensor.
Reverse flow (negative flow) moves in the direction opposite to the flow arrow on
the sensor.
Tip
Micro Motion sensors are bidirectional. Measurement accuracy is not affected by actual flow direction or the setting of the Flow Direction parameter.
Procedure
Set Flow Direction to the value you want to use.
The default setting is Forward.
34 Micro Motion Model 1700 Transmitters with Analog Outputs
4.4.1 Options for Flow Direction
Configure process measurement
Flow Direction setting
Forward Forward
Reverse Reverse
Absolute Value Absolute Value
Bidirectional Bi directional
Negate Forward Negate/Forward Only
Negate Bidirectional Negate/Bi-directional
Relationship to Flow Direction arrow on sensorProLink III Field Communicator
Appropriate when the Flow Direction ar­row is in the same direction as the major­ity of flow.
Appropriate when the Flow Direction ar­row is in the opposite direction from the majority of flow.
Flow Direction arrow is not relevant.
Appropriate when both forward and re­verse flow are expected, and forward flow will dominate, but the amount of re­verse flow will be significant.
Appropriate when the Flow Direction ar­row is in the opposite direction from the majority of flow.
Appropriate when both forward and re­verse flow are expected, and reverse flow will dominate, but the amount of for­ward flow will be significant.
Effect of Flow Direction on mA Outputs
Flow Direction affects how the transmitter reports flow values via the mA Outputs. The mA Outputs are affected by Flow Direction only if mA Output Process Variable is set to a flow variable.
Flow Direction and mA Outputs
The effect of Flow Direction on the mA Outputs depends on Lower Range Value configured for the mA Output:
If Lower Range Value is set to 0, see Figure 4‐1.
If Lower Range Value is set to a negative value, see Figure 4‐2.
Configuration and Use Manual 35
Configure process measurement
Effect of Flow Direction on the mA Output: Lower Range Value = 0Figure 4-1:
Flow Direction = Forward
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Lower Range Value = 0
Upper Range Value = x
Effect of Flow Direction on the mA Output: Lower Range Value < 0Figure 4-2:
Flow Direction = Forward
20
Flow Direction = Reverse, Negate Forward
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Flow Direction = Reverse, Negate Forward
20
Flow Direction = Absolute Value, Bidirectional, Negate Bidirectional
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
Flow Direction = Absolute Value, Bidirectional, Negate Bidirectional
20
12
mA output
4
-x 0 x
Reverse flow Forward flow
12
mA output
4
-x 0 x
Reverse flow Forward flow
12
mA output
4
-x 0 x
Reverse flow Forward flow
Lower Range Value = x
Upper Range Value = x
Example: Flow Direction = Forward and Lower Range Value = 0
Configuration:
Flow Direction = Forward
Lower Range Value = 0 g/sec
Upper Range Value = 100 g/sec
36 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Result:
Under conditions of zero flow, the mA Output is 4 mA.
Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA Output
varies between 4 mA and 20 mA in proportion to the flow rate.
Under conditions of forward flow, if the flow rate equals or exceeds 100 g/sec, the
mA Output will be proportional to the flow rate up to 20.5 mA, and will be level at
20.5 mA at higher flow rates.
Example: Flow Direction = Forward and Lower Range Value < 0
Configuration:
Flow Direction = Forward
Lower Range Value = 100 g/sec
Upper Range Value = +100 g/sec
Result:
Under conditions of zero flow, the mA Output is 12 mA.
Under conditions of forward flow, for flow rates between 0 and +100 g/sec, the mA
Output varies between 12 mA and 20 mA in proportion to (the absolute value of) the flow rate.
Under conditions of forward flow, if (the absolute value of) the flow rate equals or
exceeds 100 g/sec, the mA Output is proportional to the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher flow rates.
Under conditions of reverse flow, for flow rates between 0 and 100 g/sec, the mA
Output varies between 4 mA and 12 mA in inverse proportion to the absolute value of the flow rate.
Under conditions of reverse flow, if the absolute value of the flow rate equals or
exceeds 100 g/sec, the mA Output is inversely proportional to the flow rate down to
3.8 mA, and will be level at 3.8 mA at higher absolute values.
Example: Flow Direction = Reverse
Configuration:
Flow Direction = Reverse
Lower Range Value = 0 g/sec
Upper Range Value = 100 g/sec
Result:
Under conditions of zero flow, the mA Output is 4 mA.
Under conditions of reverse flow, for flow rates between 0 and +100 g/sec, the mA
Output level varies between 4 mA and 20 mA in proportion to the absolute value of the flow rate.
Under conditions of reverse flow, if the absolute value of the flow rate equals or
exceeds 100 g/sec, the mA Output will be proportional to the absolute value of the flow rate up to 20.5 mA, and will be level at 20.5 mA at higher absolute values.
Configuration and Use Manual 37
Configure process measurement
Effect of flow direction on Frequency Outputs
Flow direction affects how the transmitter reports flow values via the Frequency Outputs. The Frequency Outputs are affected by flow direction only if Frequency Output Process Variable is set to a flow variable.
Table 4-1:
Effect of the flow direction parameter and actual flow direction on
Frequency Outputs
Actual flow direction
Flow Direction setting
Forward
Reverse
Bidirectional
Absolute Value
Negate Forward
Negate Bidirectional
Forward Zero flow Reverse
Hz > 0 0 Hz 0 Hz
0 Hz 0 Hz Hz > 0
Hz > 0 0 Hz Hz > 0
Hz > 0 0 Hz Hz > 0
0 Hz 0 Hz Hz > 0
Hz > 0 0 Hz Hz > 0
Effect of flow direction on Discrete Outputs
The flow direction parameter affects the Discrete Output behavior only if Discrete Output Source is set to Flow Direction.
Table 4-2: Discrete Outputs
Effect of the flow direction parameter and actual flow direction on
Actual flow direction
Flow Direction setting
Forward
Reverse
Bidirectional
Absolute Value
Negate Forward
Negate Bidirectional
Forward Zero flow Reverse
OFF OFF ON
OFF OFF ON
OFF OFF ON
OFF OFF ON
ON OFF OFF
ON OFF OFF
Effect of flow direction on digital communications
Flow direction affects how flow values are reported via digital communications. The following table describes the effect of the flow direction parameter and actual flow direction on flow values reported via digital communications.
38 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Effect of the flow direction on flow valuesTable 4-3:
Actual flow direction
Flow Direction setting
Forward
Reverse
Bidirectional
Absolute Value
Negate Forward
Negate Bidirectional
(1) Refer to the digital communications status bits for an indication of whether flow is positive or negative.
Forward Zero flow Reverse
Positive 0 Negative
Positive 0 Negative
Positive 0 Negative
Positive
Negative 0 Positive
Negative 0 Positive
(1)
0 Positive
Effect of flow direction on flow totals
Flow direction affects how flow totals and inventories are calculated.
Actual flow direction
Flow Direction setting
Forward
Reverse
Bidirectional
Absolute Value
Negate Forward
Negate Bidirectional
Forward Zero flow Reverse
Totals increase Totals do not change Totals do not change
Totals do not change Totals do not change Totals increase
Totals increase Totals do not change Totals decrease
Totals increase Totals do not change Totals increase
Totals do not change Totals do not change Totals increase
Totals decrease Totals do not change Totals increase
(1)

4.5 Configure density measurement

The density measurement parameters control how density is measured and reported.
4.5.1
Configuration and Use Manual 39
Configure Density Measurement Unit
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > DENS
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Unit
Overview
Density Measurement Unit controls the measurement units that will be used in density calculations and reporting.
Configure process measurement
Procedure
Set Density Measurement Unit to the option you want to use.
The default setting for Density Measurement Unit is g/cm3 (grams per cubic centimeter).
Options for Density Measurement Unit
The transmitter provides a standard set of measurement units for Density Measurement Unit. Different communications tools may use different labels.
Label
Unit description
Specific gravity
Grams per cubic centimeter
Grams per liter
Grams per milliliter
Kilograms per liter
Kilograms per cubic meter
Pounds per U.S. gallon
Pounds per cubic foot
Pounds per cubic inch
Degrees API
Short ton per cubic yard
(1) Non‐standard calculation. This value represents line density divided by the density of water at 60 °F.
4.5.2
(1)
Configure two-phase flow parameters
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Slug Low Limit
Display ProLink III Field Communicator
SGU SGU SGU
G/CM3 g/cm3 g/Cucm
G/L g/l g/L
G/mL g/ml g/mL
KG/L kg/l kg/L
KG/M3 kg/m3 kg/Cum
LB/GAL lbs/Usgal lb/gal
LB/CUF lbs/ft3 lb/Cuft
LB/CUI lbs/in3 lb/CuIn
D API degAPI degAPI
ST/CUY sT/yd3 STon/Cuyd
Configure > Manual Setup > Measurements > Density > Slug High Limit
Configure > Manual Setup > Measurements > Density > Slug Duration
Overview
The two-phase flow parameters control how the transmitter detects and reports two­phase flow (gas in a liquid process or liquid in a gas process).
Note
Two-phase flow is also referred to as slug flow.
40 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Procedure
1. Set Two-Phase Flow Low Limit to the lowest density value that is considered normal in your process.
Values below this will cause the transmitter to post Alert A105 (Two-Phase Flow).
Tip
Gas entrainment can cause your process density to drop temporarily. To reduce the occurrence of two-phase flow alerts that are not significant to your process, set Two-Phase Flow Low Limit slightly below your expected lowest process density.
You must enter Two-Phase Flow Low Limit in g/cm³, even if you configured another unit for density measurement.
The default value for Two-Phase Flow Low Limit is 0.0 g/cm³. The range is 0.0 to
10.0 g/cm³.
2. Set Two-Phase Flow High Limit to the highest density value that is considered normal in your process.
Micro Motion recommends leaving Two-Phase Flow High Limit at the default value.
Values above this will cause the transmitter to post Alert A105 (Two-Phase Flow).
You must enter Two-Phase Flow High Limit in g/cm³, even if you configured another unit for density measurement.
The default value for Two-Phase Flow High Limit is 5.0 g/cm³. The range is 0.0 to
10.0 g/cm³.
3. Set Two-Phase Flow Timeout to the number of seconds that the transmitter will wait for a two-phase flow condition to clear before posting the alert.
The default value for Two-Phase Flow Timeout is 0.0 seconds, meaning that the alert will be posted immediately. The range is 0.0 to 60.0 seconds.
Detecting and reporting two-phase flow
Two-phase flow (gas in a liquid process or liquid in a gas process) can cause a variety of process control issues. By configuring the two-phase flow parameters appropriately for your application, you can detect process conditions that require correction.
Micro Motion recommends leaving Two-Phase Flow High Limit at the default value.
A two-phase flow condition occurs whenever the measured density goes below Two-Phase Flow Low Limit or above Two-Phase Flow High Limit. If this occurs:
A two-phase flow alert is posted to the active alert log.
All outputs that are configured to represent flow rate hold their last pre‐alert value
for the number of seconds configured in Two-Phase Flow Timeout.
If the two-phase flow condition clears before Two-Phase Flow Timeout expires:
Outputs that represent flow rate revert to reporting actual flow.
Configuration and Use Manual 41
Configure process measurement
The two-phase flow alert is deactivated, but remains in the active alert log until it is
acknowledged.
If the two-phase flow condition does not clear before Two-Phase Flow Timeout expires, the outputs that represent flow rate report a flow rate of 0.
If Two-Phase Flow Timeout is set to 0.0 seconds, the outputs that represent flow rate will report a flow rate of 0 as soon as two-phase flow is detected.
4.5.3 Configure Density Damping
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Damping
Overview
Density Damping controls the amount of damping that will be applied to the line density value.
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Tip
Density damping affects all process variables that are calculated from line density.
Procedure
Set Density Damping to the value you want to use.
The default value is 1.6 seconds. The range depends on the core processor type and the setting of Update Rate, as shown in the following table:
Update Rate setting
Normal
Special
Tips
A high damping value makes the process variable appear smoother because the reported value
changes slowly.
A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
Damping range
0 to 51.2 seconds
0 to 40.96 seconds
42 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
The value you enter is automatically rounded off to the nearest valid value. The valid values for Density Damping depend on the setting of Update Rate.
Update Rate setting Valid damping values
Normal
Special
0.0, 0.2, 0.4, 0.8, 1.6, 3.2, 6.4, 12.8, 25.6, 51.2
0.0, 0.04, 0.08, 0.16, 0.32, 0.64, 1.28, 2.56,
5.12, 10.24, 20.48, 40.96
Effect of Density Damping on volume measurement
Density Damping affects liquid volume measurement. Liquid volume values are calculated from the damped density value rather than the measured density value. Density Damping does not affect gas standard volume measurement.
4.5.4 Configure Density Cutoff
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Density
Field Communicator Configure > Manual Setup > Measurements > Density > Density Cutoff
Overview
Density Cutoff specifies the lowest density value that will be reported as measured. All density values below this cutoff will be reported as 0.
Procedure
Set Density Cutoff to the value you want to use.
For most applications, the default setting (0.2 g/cm³) is sufficient. The range is 0.0 g/cm³ to 0.5 g/cm³.
Effect of Density Cutoff on volume measurement
Density Cutoff affects liquid volume measurement. If the density value goes below Density Cutoff, the volume flow rate is reported as 0. Density Cutoff does not affect gas standard volume measurement. Gas standard volume values are always calculated from the value configured for Standard Gas Density or polled value if configured for polled base density.
Configuration and Use Manual 43
Configure process measurement

4.6 Configure temperature measurement

The temperature measurement parameters control how temperature data from the sensor is reported.
4.6.1 Configure Temperature Measurement Unit
Display OFF-LINE MAINT > OFF-LINE CONFG > UNITS > TEMP
ProLink III Device Tools > Configuration > Process Measurement > Temperature
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temperature Unit
Overview
Temperature Measurement Unit specifies the unit that will be used for temperature measurement.
Procedure
Set Temperature Measurement Unit to the option you want to use.
The default setting is Degrees Celsius.
Options for Temperature Measurement Unit
The transmitter provides a standard set of units for Temperature Measurement Unit. Different communications tools may use different labels for the units.
Unit description
Degrees Celsius
Degrees Fahrenheit
Degrees Rankine
Kelvin
Display ProLink III
°C °C degC
°F °F degF
°R °R degR
°K °K Kelvin
4.6.2 Configure Temperature Damping
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Temperature
Field Communicator Configure > Manual Setup > Measurements > Temperature > Temp Damping
Label
Field Communica­tor
44 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Overview
Temperature Damping controls the amount of damping that will be applied to the line temperature value, when the on-board temperature data is used (RTD).
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Tip
Temperature Damping affects all process variables, compensations, and corrections that use temperature data from the sensor.
Procedure
Enter the value you want to use for Temperature Damping.
The default value is 4.8 seconds. The range is 0.0 to 38.4 seconds.
4.6.3
Tips
A high damping value makes the process variable appear smoother because the reported value
changes slowly.
A low damping value makes the process variable appear more erratic because the reported value
changes more quickly.
Whenever the damping value is non-zero, the reported measurement will lag the actual
measurement because the reported value is being averaged over time.
In general, lower damping values are preferable because there is less chance of data loss, and less
lag time between the actual measurement and the reported value.
The value you enter is automatically rounded off to the nearest valid value. Valid values for Temperature Damping are 0, 0.6, 1.2, 2.4, 4.8, … 38.4.
Effect of Temperature Damping on process measurement
Temperature Damping affects all processes and algorithms that use temperature data from the internal sensor RTD.
Temperature compensation
Temperature compensation adjusts process measurement to compensate for the effect of temperature on the sensor tubes.
4.6.4
Configuration and Use Manual 45
Configure Temperature Input
Temperature data from the on-board temperature sensor (RTD) is always available. Optionally, you can set up an external temperature device and use external temperature data.
Configure process measurement
Choose Device Tools > Configuration > Process Measurement > Temperature > Source.

4.7 Configure pressure compensation

Pressure compensation adjusts process measurement to compensate for the pressure effect on the sensor. The pressure effect is the change in the sensor’s sensitivity to flow and density caused by the difference between the calibration pressure and the process pressure.
Tip
Not all sensors or applications require pressure compensation. The pressure effect for a specific sensor model can be found in the product data sheet located at www.emerson.com. If you are uncertain about implementing pressure compensation, contact customer service.
Prerequisites
You will need the flow factor, density factor, and calibration pressure values for your sensor.
For the flow factor and density factor, see the product data sheet for your sensor.
For the calibration pressure, see the calibration sheet for your sensor. If the data is
unavailable, use 20 PSI.
4.7.1
Configure pressure compensation using ProLink III
1. Choose Device Tools > Configuration > Process Measurement > Pressure Compensation.
2. Set Pressure Compensation Status to Enabled.
3. Set Pressure Unit to the appropriate unit.
If you will use an external pressure value, set Pressure Unit to match the pressure unit used by the external pressure device.
4. Enter Flow Calibration Pressure for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and defines the pressure at which there is no pressure effect. If the data is unavailable, enter 20 PSI.
5. Enter Flow Factor for your sensor.
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign.
Example: If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
6. Enter Density Factor for your sensor.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the value, reverse the sign.
46 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure process measurement
Example:
If the density factor is 0.000006 g/cm3/PSI, enter −0.000006g/cm3/PSI.
7. Set Pressure Source to the method that the transmitter will use to obtain pressure data.
Option Description
Poll for external value
Fixed Value or Digital Communi­cations
8. If you chose to poll for pressure data:
a. Select the Polling Slot to use.
b. Set Polling Control to Poll as Primary or Poll as Secondary, and click Apply.
The transmitter will poll an external pressure device, using HART protocol over the primary mA Output.
The transmitter will use the pressure value that it reads from memory.
Fixed Value: The configured value is used.
Digital Communications: A host writes transmitter data
to transmitter memory.
4.7.2
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network. The Field Communicator
is not a HART master.
c. Set External Device Tag to the HART tag of the external pressure device, and click
Apply.
d. Ensure that the primary mA Output is wired to support HART communications
with the external pressure device.
9. If you chose to use a fixed pressure value:
a. Set Fixed Value to the value to use, and click Apply
10. If you want to use digital communications, click Apply, then perform the necessary host programming and communications setup to write pressure data to the transmitter at appropriate intervals.
Postrequisites
If you are using an external pressure value, verify the setup by checking the External Pressure value displayed in the Inputs area of the main window.
Configure pressure compensation using the Field Communicator
1. Choose Online > Configure > Manual Setup > Measurements > External Pressure/ Temperature > Pressure.
Configuration and Use Manual 47
Configure process measurement
2. Set Pressure Compensation to Enabled.
3. Enter Flow Cal Pressure for your sensor.
4. Enter Flow Press Factor for your sensor.
5. Enter Dens Press Factor for your sensor.
The calibration pressure is the pressure at which your sensor was calibrated, and defines the pressure at which there is no pressure effect. If the data is unavailable, enter 20 PSI.
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign.
Example:
If the flow factor is 0.000004 % per PSI, enter −0.000004 % per PSI.
The density factor is the change in fluid density, in g/cm3/PSI. When entering the value, reverse the sign.
Example:
If the density factor is 0.000006 g/cm3/PSI, enter −0.000006g/cm3/PSI.
6. Determine how the transmitter will obtain pressure data, and perform the required setup.
Option
A user-configured static pressure value
Polling for pressure a. Ensure that the primary mA Output has been wired to support
Setup
a. Set Pressure Unit to the desired unit. b. Set Compensation Pressure to the desired value.
HART polling.
b. Choose Online > Configure > Manual Setup > Measurements >
External Pressure/Temperature > External Polling . c. Set Poll Control to Poll As Primary Host or Poll as Secondary Host. d. Choose an unused polling slot. e. Set External Tag to the HART tag of the external pressure device. f. Set Polled Variable to Pressure.
Tip
Poll as Primary: No other HART masters will be on the network.
Poll as Secondary: Other HART masters will be on the network.
The Field Communicator is not a HART master.
A value written by digital communica­tions
a. Set Pressure Unit to the desired unit. b. Perform the necessary host programming and communications
setup to write pressure data to the transmitter at appropriate in-
tervals.
48 Micro Motion Model 1700 Transmitters with Analog Outputs
Postrequisites
If you are using an external pressure value, verify the setup by choosing Service Tools > Variables > External Variables and checking the value displayed for External Pressure.
4.7.3 Options for Pressure Measurement Unit
The transmitter provides a standard set of measurement units for Pressure Measurement Unit. Different communications tools may use different labels for the units. In most applications, Pressure Measurement Unit should be set to match the pressure measurement unit used by the remote device.
Label
Configure process measurement
Unit description
Feet water @ 68 °F
Inches water @ 4 °C
Inches water @ 60 °F
Inches water @ 68 °F
Millimeters water @ 4 °C
Millimeters water @ 68 °F
Millimeters mercury @ 0 °C
Inches mercury @ 0 °C
Pounds per square inch
Bar
Millibar
Grams per square centimeter
Kilograms per square centimeter
Pascals
Kilopascals
Megapascals
Torr @ 0 °C
Atmospheres
Display ProLink III Field Communicator
FTH2O Ft Water @ 68°F ftH2O
INW4C In Water @ 4°C inH2O @4DegC
INW60 In Water @ 60°F inH2O @60DegF
INH2O In Water @ 68°F inH2O
mmW4C mm Water @ 4°C mmH2O @4DegC
mmH2O mm Water @ 68°F mmH2O
mmHG mm Mercury @ 0°C mmHg
INHG In Mercury @ 0°C inHG
PSI PSI psi
BAR bar bar
mBAR millibar mbar
G/SCM g/cm2 g/Sqcm
KG/SCM kg/cm2 kg/Sqcm
PA pascals Pa
KPA Kilopascals kPa
MPA Megapascals MPa
TORR Torr @ 0°C torr
ATM atms atms
Configuration and Use Manual 49
Configure process measurement
50 Micro Motion Model 1700 Transmitters with Analog Outputs

Configure device options and preferences

5 Configure device options and
preferences
Topics covered in this chapter:

Configure the transmitter display

Enable or disable operator actions from the display
Configure security for the display menus
Configure response time parameters
Configure alert handling
Configure informational parameters
5.1 Configure the transmitter display
You can control the process variables shown on the display and a variety of display behaviors.
5.1.1 Configure the language used for the display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > LANG
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Language
Overview
Display Language controls the language used for process data and menus on the display.
Procedure
Select the language you want to use.
The languages available depend on your transmitter model and version.
5.1.2
Configure the process variables and diagnostic variables shown on the display
Display Not available
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Display Variables
Configuration and Use Manual 51
Configure device options and preferences
Overview
You can control the process variables and diagnostic variables shown on the display, and the order in which they appear. The display can scroll through up to 15 variables in any order you choose. In addition, you can repeat variables or leave slots unassigned.
Restrictions
You cannot set Display Variable 1 to None or to a diagnostic variable. Display Variable 1 must
be set to a process variable.
If you have configured Display Variable 1 to track the primary mA Output, you cannot change
the setting of Display Variable 1 using this procedure. To change the setting of Display Variable 1, you must change the configuration of mA Output Process Variable for the primary mA Output.
Note
If you configure a display variable as a volume process variable and then change Volume Flow Type, the display variable is automatically changed to the equivalent process variable. For example, Volume Flow Rate would be changed to Gas Standard Volume Flow Rate.
Procedure
For each display variable you want to change, assign the process variable you want to use.
Example: Default display variable configuration
Display variable
Display Variable 1 Mass flow
Display Variable 2 Mass total
Display Variable 3 Volume flow
Display Variable 4 Volume total
Display Variable 5 Density
Display Variable 6 Temperature
Display Variable 7 External pressure
Display Variable 8 Mass flow
Display Variable 9 None
Display Variable 10 None
Display Variable 11 None
Display Variable 12 None
Display Variable 13 None
Display Variable 14 None
Display Variable 15 None
Process variable assignment
52 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
Configure Display Variable 1 to track the primary mA Output
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLY > VAR 1
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Display Variables
Overview
You can configure Display Variable 1 to track mA Output Process Variable for the primary mA Output. When tracking is enabled, you can control Display Variable 1 from the display menu.
Tip
This feature is the only way to configure a display variable from the display menus, and it applies only to Display Variable 1.
Procedure
5.1.3
Configure Display Variable 1 to track the primary mA Output.
Display Variable 1 will automatically be set to match mA Output Process Variable for the primary mA Output. If you change the configuration of mA Output Process Variable, Display Variable 1 will be updated automatically.
Configure the number of decimal places (precision) shown on the display
Display Not available
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Decimal Places
Overview
You can specify the number of decimal places (precision) that are shown on the display for each process variable or diagnostic variable. You can set the precision independently for each variable.
The display precision does not affect the actual value of the variable or the value used in calculations.
Procedure
1. Select a variable.
2. Set Number of Decimal Places to the number of decimal places you want shown when the process variable or diagnostic variable appears on the display.
Configuration and Use Manual 53
Configure device options and preferences
For temperature and density process variables, the default value is 2 decimal places. For all other variables, the default value is 4 decimal places. The range is 0 to 5.
Tip
The lower the precision, the greater the change must be for it to be reflected on the display. Do not set the precision too low or too high to be useful.
5.1.4 Configure the refresh rate of data shown on the display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > RATE
ProLink III Device Tools > Configuration > Transmitter Display > Display Variables
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Refresh Rate
Overview
You can set Refresh Rate to control how frequently data is refreshed on the display.
5.1.5
Procedure
Set Refresh Rate to the desired value.
The default value is 200 milliseconds. The range is 100 milliseconds to 10,000 milliseconds (10 seconds).
Enable or disable automatic scrolling through the display variables
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > AUTO SCRLL
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Auto Scroll
Overview
You can configure the display to automatically scroll through the configured display variables or to show a single display variable until the operator activates Scroll. When you set automatic scrolling, you can also configure the length of time each display variable is displayed.
Procedure
1. Enable or disable Auto Scroll as desired.
54 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
Option Description
Enabled
Disabled (de-
fault)
The display automatically scrolls through each display variable as specified by Scroll Rate. The operator can move to the next display variable at any time using Scroll.
The display shows Display Variable 1 and does not scroll automatically. The operator can move to the next display variable at any time using Scroll.
2. If you enabled Auto Scroll, set Scroll Rate as desired.
The default value is 10 seconds.
Tip
Scroll Rate may not be available until you apply Auto Scroll.
5.1.6 Enable or disable the display backlight
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > BKLT
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Backlight
5.1.7
Overview
You can enable or disable the display backlight.
Procedure
Enable or disable Backlight.
The default setting is Enabled.
Enable or disable Status LED Blinking
Display Not available
ProLink III Device Tools > Configuration > Transmitter Display > General
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Status LED Blink-
ing
Overview
By default, the status LED blinks (flashes) to indicate unacknowledged alarms. If you disable Status LED Blinking, the status LED does not blink, whether alarms are acknowledged or not. It still changes color to indicate active alarms.
Procedure
Enable or disable Status LED Blinking.
Configuration and Use Manual 55
Configure device options and preferences
The default setting is Enabled.

5.2 Enable or disable operator actions from the display

You can configure the transmitter to let the operator perform specific actions using the display.
Enable or disable Totalizer Start/Stop from the display (Section 5.2.1)
Enable or disable Totalizer Reset from the display (Section 5.2.2)
Enable or disable the Acknowledge All Alerts display command (Section 5.2.3)
5.2.1 Enable or disable Totalizer Start/Stop from the display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > TOTALS STOP
ProLink III Device Tools > Configuration > Totalizer Control Methods
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Start/Stop Total-
izers
Overview
You can control whether or not the operator is able to start and stop totalizers and inventories from the display.
Restrictions
You cannot start and stop totalizers individually from the display. All totalizers are started or
stopped together.
You cannot start or stop inventories separately from totalizers. When a totalizer is started or
stopped, the associated inventory is also started or stopped.
If the petroleum measurement application is installed, the operator must enter the off-line
password to perform this function, even if the off-line password is not enabled.
Procedure
1. Ensure that at least one totalizer is configured as a display variable.
2. Enable or disable Totalizer Reset as desired.
Option
Enabled
Disabled (default) Operators cannot start and stop totalizers and inventories from the dis-
Description
Operators can start and stop totalizers and inventories from the display, if at least one totalizer is configured as a display variable.
play.
56 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
5.2.2 Enable or disable Totalizer Reset from the display
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > TOTALS RESET
ProLink III Device Tools > Configuration > Totalizer Control Methods
Field Communicator Configure > Manual Setup > Display > Display Variable Menu Features > Totalizer Reset
Overview
You can configure whether or not the operator is able to reset totalizers from the display.
Restrictions
This parameter does not apply to inventories. You cannot reset inventories from the display.
You cannot use the display to reset all totalizers as a group. You must reset totalizers
individually.
If the petroleum measurement application is installed, the operator must enter the off-line
password to perform this function, even if the off-line password is not enabled.
Procedure
1. Ensure that the totalizers you want to reset have been configured as display
variables.
If the totalizer is not configured as a display variable, the operator will not be able to reset it.
2. Enable or disable resetting the totalizer as desired.
Option
Enabled
Disabled (default) Operators cannot reset totalizers from the display.
Description
Operators can reset a totalizer from the display, if the totalizer is config­ured as a display variable.
5.2.3 Enable or disable the Acknowledge All Alerts display command
Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > ALARM
ProLink III Device Tools > Configuration > Transmitter Display > Ack All
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features > Acknowledge All
Overview
You can configure whether or not the operator can use a single command to acknowledge all alerts from the display.
Configuration and Use Manual 57
Configure device options and preferences
Procedure
1. Ensure that the alert menu is accessible from the display.
To acknowledge alerts from the display, operators must have access to the alert menu.
2. Enable or disable Acknowledge All Alerts as desired.
Option Description
Enabled (default) Operators can use a single display command to acknowledge all alerts at
once.
Disabled
Operators cannot acknowledge all alerts at once. Each alert must be ac­knowledged separately.

5.3 Configure security for the display menus

Display OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY
ProLink III Device Tools > Configuration > Transmitter Display > Display Security
Field Communicator Configure > Manual Setup > Display > Offline Variable Menu Features
Overview
You can control operator access to different sections of the display off-line menu. You can also configure a password to control access.
Procedure
1. To control operator access to the maintenance section of the off-line menu, enable
or disable Off-Line Menu.
Option
Enabled (de­fault)
Disabled
Description
Operator can access the maintenance section of the off-line menu. This ac­cess is required for configuration and calibration, but is not required to view alerts or to access Smart Meter Verification (if applicable).
Operator cannot access the maintenance section of the off-line menu.
2. To control operator access to the alert menu, enable or disable Alert Menu.
Option
Enabled (default) Operator can access the alert menu. This access is required to view and
Description
acknowledge alerts, but is not required for Smart Meter Verification (if applicable), configuration, or calibration.
58 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
Option Description
Disabled
Note
The transmitter status LED changes color to indicate that there are active alerts, but does not show specific alerts.
Operator cannot access the alert menu.
3. To require a password for access to the maintenance section of the off-line menu
and the Smart Meter Verification menu, enable or disable Off-Line Password.
Option Description
Enabled
Disabled (de-
fault)
Operator is prompted for the off-line password at entry to the Smart Meter Verification menu (if applicable) or entry to the maintenance section of the off-line menu.
No password is required for entry to the Smart Meter Verification menu (if applicable) or entry to the maintenance section of the off-line menu.
5.4
4. To require a password to access the alert menu, enable or disable Alert Password.
Option
Enabled
Disabled (default) No password is required for entry to the alert menu.
Description
Operator is prompted for the off-line password at entry to the alert menu.
If both Off-Line Password and Alert Password are enabled, the operator is prompted for the off-line password to access the off-line menu, but is not prompted thereafter.
5. Set Off-Line Password to the desired value.
The default value is 1234. The range is 0000 to 9999.
The same value is used for both the off-line password and the alert password.
Tip
Record your password for future reference.

Configure response time parameters

You can configure the rate at which process data is polled and process variables are calculated.
Configuration and Use Manual 59
Configure device options and preferences
5.4.1 Configure Update Rate
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Response > Update Rate
Field Communicator Configure > Manual Setup > Measurements > Update Rate
Overview
Update Rate controls the rate at which process data is polled and process variables are calculated. Update Rate = Special produces faster and “noisier” response to changes in the process. Do not use Special mode unless required by your application.
Tip
For systems with a standard core processor, Special mode can improve performance for applications with entrained air or Empty-Full-Empty conditions. This does not apply to systems with an enhanced core processor.
Prerequisites
Before setting Update Rate to Special:
Check the effects of Special mode on specific process variables.
Contact customer support.
Procedure
1. Set Update Rate as desired.
Option
Normal
Special
Description
All process data is polled at the rate of 20 times per second (20 Hz).
All process variables are calculated at 20 Hz.
This option is appropriate for most applications.
A single, user-specified process variable is polled at the rate of 100 times per sec­ond (100 Hz). Other process data is polled at 6.25 Hz. Some process, diagnostic, and calibration data is not polled.
All available process variables are calculated at 100 Hz.
Use this option only if required by your application.
If you change Update Rate, the settings for Flow Damping, Density Damping, and Temperature Damping are automatically adjusted.
2. If you set Update Rate to Special, select the process variable to be polled at 100 Hz.
60 Micro Motion Model 1700 Transmitters with Analog Outputs
Effects of Update Rate = Special
Incompatible features and functions
Special mode is not compatible with the following features and functions:
Enhanced events. Use basic events instead.
All calibration procedures.
Zero verification.
Restoring the factory zero or the prior zero.
If required, you can switch to Normal mode, perform the desired procedures, and then return to Special mode.
Process variable updates
Some process variables are not updated when Special mode is enabled.
Special mode and process variable updatesTable 5-1:
Always polled and updated
Mass flow
Volume flow
Gas standard volume flow
Density
Temperature
Drive gain
LPO amplitude
Status [contains Event 1 and Event
2 (basic events)]
Mass total
Volume total
Live zero
Gas standard volume total
Configure device options and preferences
Updated only when the petroleum measurement application is disa­bled Never updated
RPO amplitude
Core input voltage
Mass inventory
Volume inventory
Gas standard volume inventory
All other process variables and calibra­tion data. They retain the values held at the time you enabled Special mode.
5.4.2 Configure Response Time
Display Not available
ProLink III Device Tools > Configuration > Process Measurement > Response > Response Time
Field Communicator Not available
Overview
Response Time is used to apply a different algorithm to the calculation of process variables from the raw process data.
Configuration and Use Manual 61
Configure device options and preferences
Restriction
Response Time is available only on systems with the enhanced core processor.
Procedure
Set Response Time as desired.
Option Description
Normal (Legacy)
Special (Legacy)
Normal - Optimal Filtering
Low Filtering - Fastest Re­sponse
High Filtering - Smoothest Output
Service
Transmitter calculates process variables at the standard speed. This op­tion is selected if this parameter was configured on an earlier version of ProLink III software.
Transmitter calculates process variables at a faster speed. This option is selected if this parameter was configured on an earlier version of ProLink III software.
Transmitter calculates process variables at standard filtering and speed.
Transmitter calculates process variables at the fastest speed.
Transmitter calculates process variables at the smoothest (least noisy) response to changes in the process.
For factory use only.

5.5 Configure alert handling

The alert handling parameters control the transmitter’s response to process and device conditions.
5.5.1
Configure Fault Timeout
Display Not available
ProLink III Device Tools > Configuration > Fault Processing
Field Communicator Configure > Alert Setup > Alert Severity > Fault Timeout
Overview
Fault Timeout controls the delay before fault actions are performed.
Restriction
Fault Timeout is applied only to the following alerts (listed by Status Alert Code): A003, A004, A005, A008, A016, A017, A033. For all other alerts, fault actions are performed as soon as the alert is detected.
62 Micro Motion Model 1700 Transmitters with Analog Outputs
Procedure
Set Fault Timeout as desired.
The default value is 0 seconds. The range is 0 to 60 seconds.
If you set Fault Timeout to 0, fault actions are performed as soon as the alert condition is detected.
The fault timeout period begins when the transmitter detects an alert condition. During the fault timeout period, the transmitter continues to report its last valid measurements.
If the fault timeout period expires while the alert is still active, the fault actions are performed. If the alert condition clears before the fault timeout expires, no fault actions are performed.
5.5.2 Configure Status Alert Severity
Display Not available
ProLink III Device Tools > Configuration > Alert Severity
Field Communicator Configure > Alert Setup > Alert Severity > Set Alert Severity
Configure device options and preferences
Overview
Use Status Alert Severity to control the fault actions that the transmitter performs when it detects an alert condition.
Restrictions
For some alerts, Status Alert Severity is not configurable.
For some alerts, Status Alert Severity can be set only to two of the three options.
Tip
Use the default settings for Status Alert Severity unless you have a specific requirement to change them.
Procedure
1. Select a status alert.
2. For the selected status alert, set Status Alert Severity as desired.
Configuration and Use Manual 63
Configure device options and preferences
Option Description
Fault
Informa­tional
Ignore
Actions when fault is detected:
The alert is posted to the Alert List.
Outputs go to the configured fault action (after Fault Timeout has expired, if
applicable).
Digital communications go to the configured fault action (after Fault Time-
out has expired, if applicable).
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
Outputs return to normal behavior.
Digital communications return to normal behavior.
The status LED (if available) returns to green and may or may not flash.
Actions when fault is detected:
The alert is posted to the Alert List.
The status LED (if available) changes to red or yellow (depending on alert se-
verity).
Actions when alert clears:
The status LED (if available) returns to green and may or may not flash.
No action
Status alerts and options for Status Alert Severity
Status alerts and Status Alert Severity Table 5-2:
Alert code Status message Default severity Notes Configurable?
A001 EEPROM Error (Core Pro-
cessor)
A002 RAM Error (Core Processor)
A003 No Sensor Response
A004 Temperature Overrange
A005 Mass Flow Rate Overrange
A006 Characterization Required
A008 Density Overrange
A009 Transmitter Initializing/
Warming Up
A010 Calibration Failure
A011 Zero Calibration Failed:
Low
A012 Zero Calibration Failed:
High
A013 Zero Calibration Failed:
Unstable
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
No
No
Yes
No
Yes
Yes
Yes
Yes
No
Yes
Yes
Yes
64 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
Status alerts and Status Alert Severity (continued)Table 5-2:
Alert code Status message Default severity Notes Configurable?
A014 Transmitter Failure
A016 Sensor RTD Failure
A017 T-Series RTD Failure
A018 EEPROM Error (Transmit-
ter)
A019 RAM Error (Transmitter)
A020 No Flow Cal Value
A021 Incorrect Sensor Type (K1)
A022 Configuration Database
Corrupt (Core Processor)
A023 Internal Totals Corrupt
(Core Processor)
A024 Program Corrupt (Core
Processor)
A025 Boot Sector Fault (Core
Processor)
A026 Sensor/Transmitter Com-
munications Failure
A028 Core Processor Write Fail-
ure
A031 Low Power
A032 Meter Verification in Pro-
gress: Outputs to Fault
A033 Insufficient Right/Left Pick-
off Signal
A034 Meter Verification Failed
A035 Meter Verification Aborted
A100 mA Output 1 Saturated
A101 mA Output 1 Fixed
A102 Drive Overrange
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Fault
Varies Applies only to transmitters with
Fault
Fault
Fault
Informational
Informational
Informational
Applies only to flowmeters with the standard core processor.
Applies only to flowmeters with the standard core processor.
Applies only to flowmeters with the standard core processor.
Applies only to flowmeters with the standard core processor.
Applies only to flowmeters with the enhanced core processor.
Smart Meter Verification.
If outputs are set to Last Measured Value, severity is Info. If outputs are set to Fault, severity is Fault.
Applies only to flowmeters with the enhanced core processor.
Applies only to transmitters with Smart Meter Verification.
Applies only to transmitters with Smart Meter Verification.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Can be set to either Informational or Ignore, but cannot be set to Fault.
No
Yes
Yes
No
No
Yes
No
No
No
No
No
No
No
No
No
Yes
Yes
Yes
Yes
Yes
Yes
Configuration and Use Manual 65
Configure device options and preferences
Status alerts and Status Alert Severity (continued)Table 5-2:
Alert code Status message Default severity Notes Configurable?
A103 Data Loss Possible (Totals
and Inventories)
A104 Calibration in Progress
A105 Slug Flow
A106 Burst Mode Enabled
A107 Power Reset Occurred
A108 Basic Event 1 On
A109 Basic Event 2 On
A110 Frequency Output Satura-
ted
A111 Frequency Output Fixed
A112 Upgrade Transmitter Soft-
ware
A113 mA Output 2 Saturated
A114 mA Output 2 Fixed
A115 No External Input or Polled
Data
A118 Discrete Output 1 Fixed
A119 Discrete Output 2 Fixed
A131 Meter Verification in Pro-
gress: Outputs to Last Measured Value
A132 Sensor Simulation Active
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Informational
Applies only to flowmeters with the enhanced core processor.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Normal transmitter behavior; oc­curs after every power cycle.
Applies only to basic events. Yes
Applies only to basic events. Yes
Can be set to either Informational or Ignore, but cannot be set to Fault.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Applies only to systems with trans­mitter software earlier than v5.0.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Applies only to transmitters with Smart Meter Verification.
Applies only to flowmeters with the enhanced core processor.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
Yes
To Informational or Ignore only
66 Micro Motion Model 1700 Transmitters with Analog Outputs
Configure device options and preferences
Status alerts and Status Alert Severity (continued)Table 5-2:
Alert code Status message Default severity Notes Configurable?
A141 DDC trigger(s) have com-
pleted
Informational
Applies only to flowmeters with the enhanced core processor.
Can be set to either Informational or Ignore, but cannot be set to Fault.
Yes

5.6 Configure informational parameters

The informational parameters can be used to identify or describe your meter. They are not used in process measurement and they are not required.
5.6.1 Configure Sensor Serial Number
Display Not available
ProLink III Device Tools > Configuration > Informational Parameters > Sensor
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Sensor Serial Num-
ber
5.6.2
Overview
Sensor Serial Number lets you store the serial number of the sensor component of your flowmeter in transmitter memory. This parameter is not used in processing and is not required.
Procedure
1. Obtain the sensor serial number from your sensor tag.
2. Enter the serial number in the Sensor Serial Number field.
Configure Sensor Material
Display Not available
ProLink III Device Tools > Configuration > Informational Parameters > Sensor
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Tube Wetted Mate-
rial
Overview
Sensor Material lets you store the type of material used for your sensor’s wetted parts in transmitter memory. This parameter is not used in processing and is not required.
Configuration and Use Manual 67
Configure device options and preferences
Procedure
1. Obtain the material used for your sensor’s wetted parts from the documents shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor.
2. Set Sensor Material to the appropriate option.
5.6.3 Configure Sensor Liner Material
Display Not available
ProLink III Device Tools > Configuration > Informational Parameters > Sensor
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Tube Lining
Overview
Sensor Liner Material lets you store the type of material used for your sensor liner in transmitter memory. This parameter is not used in processing and is not required.
5.6.4
Procedure
1. Obtain your sensor’s liner material from the documents shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor.
2. Set Sensor Liner Material to the appropriate option.
Configure Sensor Flange Type
Display Not available
ProLink III Device Tools > Configuration > Informational Parameters > Sensor
Field Communicator Configure > Manual Setup > Info Parameters > Sensor Information > Sensor Flange
Overview
Sensor Flange Type lets you store your sensor’s flange type in transmitter memory. This parameter is not used in processing and is not required.
Procedure
1. Obtain your sensor’s flange type from the documents shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor.
2. Set Sensor Flange Type to the appropriate option.
68 Micro Motion Model 1700 Transmitters with Analog Outputs
5.6.5 Configure Descriptor
Display Not available
ProLink III Device Tools > Configuration > Informational Parameters > Transmitter
Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Descriptor
Overview
Descriptor lets you store a description in transmitter memory. The description is not used in processing and is not required.
Procedure
Enter a description for the transmitter or device
You can use up to 16 characters for the description.
5.6.6 Configure Message
Configure device options and preferences
5.6.7
Display Not available
ProLink III Device Tools > Configuration > Informational Parameters > Transmitter
Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Message
Overview
Message lets you store a short message in transmitter memory. This parameter is not used in processing and is not required.
Procedure
Enter a short message for the transmitter or device.
Your message can be up to 32 characters long.
Configure Date
Display Not available
ProLink III Device Tools > Configuration > Informational Parameters > Transmitter
Field Communicator Configure > Manual Setup > Info Parameters > Transmitter Info > Date
Overview
Date lets you store a static date (not updated by the transmitter) in transmitter memory. This parameter is not used in processing and is not required.
Configuration and Use Manual 69
Configure device options and preferences
Procedure
Enter the date you want to use, in the form mm/dd/yyyy.
Tip
ProLink III provides a calendar tool to help you select the date.
70 Micro Motion Model 1700 Transmitters with Analog Outputs

Integrate the meter with the control system

6 Integrate the meter with the control
system
Topics covered in this chapter:

Configure the transmitter channels

Configure the mA Output
Configure the Frequency Output
Configure the Discrete Output
Configure events
Configure digital communications
6.1 Configure the transmitter channels
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B
ProLink III Device Tools > Configuration > I/O > Outputs
Field Communicator Configure > Manual Setup > Inputs/Outputs > Channels > Channel B
Overview
You can configure Channel B on your transmitter to operate as a Frequency Output or a Discrete Output. The channel configuration must match the wiring at the transmitter terminals.
Prerequisites
To avoid causing process errors:
Configure the channels before configuring the outputs.
Before changing the channel configuration, ensure that all control loops affected by
the channel are under manual control.
Procedure
Set Channel B as desired.
Option
Frequency Output Channel B will operate as a Frequency Output.
Discrete Output Channel B will operate as a Discrete Output.
Description
Configuration and Use Manual 71
Integrate the meter with the control system
Postrequisites
For each channel that you configured, perform or verify the corresponding input or output configuration. When the configuration of a channel is changed, the channel’s behavior will be controlled by the configuration that is stored for the selected input or output type, and the stored configuration may not be appropriate for your process.
After verifying channel and output configuration, return the control loop to automatic control.

6.2 Configure the mA Output

The mA Output is used to report the configured process variable. The mA Output parameters control how the process variable is reported.
Your transmitter has one mA Output: Channel A.
Restriction
The process variable assigned to the primary mA Output is automatically assigned to the Frequency Output. You cannot assign a different process variable.
6.2.1
Important
Whenever you change an mA Output parameter, verify all other mA Output parameters before returning the meter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
Configure mA Output Process Variable
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH A > AO
ProLink III Device Tools > Configuration > I/O > Outputs > mA Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output
Overview
Use mA Output Process Variable to select the variable that is reported over the mA Output. This variable is applied automatically to the Frequency Output.
Prerequisites
If you plan to configure the output to report volume flow, ensure that you have set
Volume Flow Type as desired: Liquid or Gas Standard Volume.
If you are using the HART variables, be aware that changing the configuration of mA
Output Process Variable will change the configuration of the HART Primary Variable (PV) and the HART Tertiary Variable (TV).
72 Micro Motion Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
If you have configured Display Variable 1 to track mA Output Process Variable, be
aware that changing the configuration of mA Output Process Variable will change the contents of Display Variable 1.
Procedure
Set mA Output Process Variable as desired.
The default setting is Mass Flow Rate.
Postrequisites
If you changed the setting of mA Output Process Variable, verify the settings of Lower Range Value (LRV) and Upper Range Value (URV).
Options for mA Output Process Variable
The transmitter provides a basic set of options for mA Output Process Variable, plus several application-specific options. Different communications tools may use different labels for the options.
Standard mA Output process variablesTable 6-1:
Process variable
Gas standard volume flow rate
Mass flow rate
Volume flow rate
PVR mA Output process variablesTable 6-2:
Process variable
Uncorrected oil flow
Uncorrected water cut
Uncorrected water flow
Corrected oil flow
Corrected water cut
Corrected water flow
Shrinkage factor corrected net oil at line
Shrinkage factor corrected net oil at 60F
Label
Display ProLink III Field Communicator
GSV F Gas Standard Volume Flow Rate Gas vol flo
MFLOW Mass Flow Rate Mass flo
VFLOW Volume Flow Rate Vol flo
Label
Display ProLink III Field Communicator
OIL Oil Flow Rate At Line Oil Flow Rate at Line
WATER% Water Cut At Line Water Cut at Line
WATER Water Flow Rate At Line Water Flow Rate at Line
OIL60 Oil Flow Rate At Reference Oil Flow Rate at Reference
WCT60% Water Cut At Reference Water Cut at Reference
WTR60 Water Flow Rate At Reference Water Flow Rate at Reference
SFOIL SF Oil Flow Rate At Line Shrinkage Factor Oil Flow Rate
at Line
SFO60 SF Oil Flow Rate At Reference Shrinkage Factor Oil Flow Rate
at Reference
Configuration and Use Manual 73
Integrate the meter with the control system
PVR mA Output process variables (continued)Table 6-2:
Label
Process variable
Shrinkage factor corrected volume of mix at 60F
Display ProLink III Field Communicator
SFM60 SF Volume Flow Rate At Refer-
ence
Shrinkage Factor Volume Flow Rate at Reference
6.2.2 Configure Lower Range Value (LRV) and Upper Range Value (URV)
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH A > 4 mA
OFF-LINE MAINT > OFF-LINE CONFG > IO > CH A > 20 mA
ProLink III Device Tools > Configuration > I/O > Outputs > mA Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output > mA Output Settings > PV
LRV
Configure > Manual Setup > Inputs/Outputs > mA Output > mA Output Settings > PV
URV
Overview
The Lower Range Value (LRV) and Upper Range Value (URV) are used to scale the mA Output, that is, to define the relationship between mA Output Process Variable and the mA Output level.
Note
For transmitter software v5.0 and later, if you change LRV and URV from the factory default values, and you later change mA Output Process Variable, LRV and URV will not reset to the default values.
Prerequisites
Ensure that mA Output Process Variable is set to the desired process variable. Each process variable has its own set of LRV and URV values. When you change the values of LRV and URV, you are configuring values for the currently assigned mA Output process variable.
Ensure that the measurement unit for the configured process variable has been set as desired.
Procedure
Set LRV and URV as desired.
LRVis the value of mA Output Process Variable represented by an output of 4 mA. The
default value for LRV depends on the setting of mA Output Process Variable. Enter LRV in the measurement units that are configured for mA Output Process Variable.
74 Micro Motion Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
URV is the value of mA Output Process Variable represented by an output of 20 mA.
The default value for URV depends on the setting of mA Output Process Variable. Enter URV in the measurement units that are configured for mA Output Process Variable.
Tip
For best performance:
Set LRV ≥ LSL (lower sensor limit).
Set URV ≤ USL (upper sensor limit).
Set these values so that the difference between URV and LRV is ≥ Min Span (minimum span).
Defining URV and LRV within the recommended values for Min Span, LSL, and USL ensures that the resolution of the mA Output signal is within the range of the bit precision of the D/A converter.
The mA Output uses a range of 4–20 mA to represent mA Output Process Variable. Between LRV and URV, the mA Output is linear with the process variable. If the process variable drops below LRV or rises above URV, the transmitter posts an output saturation alert.
Default values for Lower Range Value (LRV) and Upper Range Value (URV)
Each option for mA Output Process Variable has its own LRV and URV. If you change the configuration of mA Output Process Variable, the corresponding LRV and URV are loaded and used.
Table 6-3:
Process variable LRV URV
All mass flow variables –200.000 g/sec 200.000 g/sec
All liquid volume flow variables –0.200 l/sec 0.200 l/sec
Gas standard volume flow –423.78 SCFM 423.78 SCFM
Default values for Lower Range Value (LRV) and Upper Range Value (URV)
6.2.3 Configure AO Cutoff
Display Not available
ProLink III Device Tools > Configuration > I/O > Outputs > mA Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output > mA Output Settings > PV
MAO Cutoff
AO Cutoff (Analog Output Cutoff) specifies the lowest mass flow rate, volume flow rate, or gas standard volume flow rate that will be reported through the mA Output. Any flow rates below AO Cutoff will be reported as 0.
Configuration and Use Manual 75
Integrate the meter with the control system
Restriction
AO Cutoff is applied only if mA Output Process Variable is set to Mass Flow Rate, Volume Flow Rate, or Gas Standard Volume Flow Rate. If mA Output Process Variable is set to a different process variable, AO
Cutoff is not configurable, and the transmitter does not implement the AO cutoff function.
Procedure
Set AO Cutoff as desired.
The default value for AO Cutoff is 0.0 g/sec.
Tip
For most applications, the default value of AO Cutoff should be used. Contact customer service before changing AO Cutoff.
Interaction between AO Cutoff and process variable cutoffs
When mA Output Process Variable is set to a flow variable (for example, mass flow rate or volume flow rate), AO Cutoff interacts with Mass Flow Cutoff or Volume Flow Cutoff. The transmitter puts the cutoff into effect at the highest flow rate at which a cutoff is applicable.
Example: Cutoff interaction
Configuration:
mA Output Process Variable = Mass Flow Rate
Frequency Output Process Variable = Mass Flow Rate
AO Cutoff = 10 g/sec
Mass Flow Cutoff = 15 g/sec
Result: If the mass flow rate drops below 15 g/sec, all outputs representing mass flow will report zero flow.
Example: Cutoff interaction
Configuration:
mA Output Process Variable = Mass Flow Rate
Frequency Output Process Variable = Mass Flow Rate
AO Cutoff = 15 g/sec
Mass Flow Cutoff = 10 g/sec
Result:
If the mass flow rate drops below 15 g/sec but not below 10 g/sec:
- The mA Output will report zero flow.
- The Frequency Output will report the actual flow rate.
If the mass flow rate drops below 10 g/sec, both outputs will report zero flow.
76 Micro Motion Model 1700 Transmitters with Analog Outputs
6.2.4 Configure Added Damping
Display Not available
ProLink III Device Tools > Configuration > I/O > Outputs > mA Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output > mA Output Settings > PV
Added Damping
Overview
Added Damping controls the amount of damping that will be applied to the mA Output.
Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the process variable. At the end of the interval, the internal value will reflect 63% of the change in the actual measured value.
Added Damping affects the reporting of mA Output Process Variable through the mA Output only. It does not affect the reporting of that process variable via any other method (e.g., a Frequency Output or digital communications), or the value of the process variable used in calculations.
Integrate the meter with the control system
Note
Added Damping is not applied if the mA Output is fixed (for example, during loop testing) or if the mA Output is reporting a fault. Added Damping is applied while sensor simulation is active.
Procedure
Set Added Damping to the desired value.
The default value is 0.0 seconds. The range is 0.0 to 440 seconds.
When you specify a value for Added Damping, the transmitter automatically rounds the value down to the nearest valid value.
Note
Added Damping values are affected by the setting of Update Rate and 100 Hz Variable.
Valid values for Added Damping Table 6-4:
Setting of Update Rate Process variable
Normal
Special
N/A 20 Hz 0.0, 0.1, 0.3, 0.75, 1.6, 3.3, 6.5, 13.5, 27.5, 55,
100 Hz variable (if assigned to the mA Output)
Update rate in effect Valid values for Added Damping
110, 220, 440
100 Hz 0.0, 0.04, 0.12, 0.30, 0.64, 1.32, 2.6, 5.4, 11, 22,
44, 88, 176, 350
Configuration and Use Manual 77
Integrate the meter with the control system
Valid values for Added Damping (continued)Table 6-4:
Setting of Update Rate Process variable
100 Hz variable (if not as­signed to the mA Output)
All other process variables
Interaction between mA Output Damping and process variable damping
When mA Output Source is set to a flow rate variable, mA Output Damping interacts with Flow Damping. If multiple damping parameters are applicable, the effect of damping the process variable is calculated first, and the mA output damping calculation is applied to the result of that calculation.
Example: Damping interaction
Configuration:
Flow Damping = 1 second
Update rate in effect Valid values for Added Damping
6.25 Hz 0.0, 0.32, 0.96, 2.40, 5.12, 10.56, 20.8, 43.2, 88, 176, 352
6.2.5
mA Output Source = Mass Flow Rate
mA Output Damping = 2 seconds
Result: A change in the mass flow rate will be reflected in the mA Output over a time period that is greater than 3 seconds. The exact time period is calculated by the transmitter according to internal algorithms which are not configurable.
Configure mA Output Fault Action and mA Output Fault Level
Display Not available
ProLink III Device Tools > Configuration > Fault Processing
Field Communicator Configure > Manual Setup > Inputs/Outputs > mA Output > MA0 Fault Settings
Overview
mA Output Fault Action controls the behavior of the mA Output if the transmitter encounters an internal fault condition.
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed.
78 Micro Motion Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Procedure
1. Set mA Output Fault Action to the desired value.
The default setting is Downscale.
Restriction
If Digital Communications Fault Action is set to NAN (not a number), you cannot set mA Output Fault Action or Frequency Output Fault Action to None. If you try to do this, the transmitter will not accept the configuration.
2. If you set mA Output Fault Action to Upscale or Downscale, set mA Output Fault Level as desired.
Postrequisites
CAUTION!
If you set mA Output Fault Action or Frequency Output Fault Action to None, be sure to set Digital Communications Fault Action to None. If you do not, the output will not report actual process data, and this may result in measurement errors or unintended consequences for your process.
Options for mA Output Fault Action and mA Output Fault Level
Option
Upscale
Downscale (default) Goes to the configured fault level Default: 2.0 mA
Internal Zero
None
mA Output behavior mA Output Fault Level
Goes to the configured fault level Default: 22.0 mA
Goes to the mA Output level associated with a process variable value of 0 (zero), as determined by Lower Range Value and Upper Range Value settings
Tracks data for the assigned process vari­able; no fault action

6.3 Configure the Frequency Output

The Frequency Output is used to report a process variable. The Frequency Output parameters control how the process variable is reported. Your transmitter may have zero or one Frequency Output: Channel B can be configured as a Frequency Output or a Discrete Output. The default assignment for Channel B is Frequency Output.
Range: 21.0 to 24.0 mA
Range: 1.0 to 3.6 mA
Not applicable
Not applicable
Configuration and Use Manual 79
Integrate the meter with the control system
Restriction
The process variable assigned to the primary mA Output is automatically assigned to the Frequency Output. You cannot assign a different process variable.
Important
Whenever you change a Frequency Output parameter, verify all other Frequency Output parameters before returning the flowmeter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
6.3.1 Configure Frequency Output Polarity
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET FO > FO POLAR
ProLink III Device Tools > Configuration > I/O > Outputs > Frequency Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Settings > FO Po-
larity
Overview
Polarity option
Active High
Active Low
Frequency Output Polarity controls how the output indicates the ON (active) state. The default value, Active High, is appropriate for most applications. Your receiving device might require an Active Low setting.
Procedure
Set Frequency Output Polarity as desired.
The default setting is Active High.
Options for Frequency Output Polarity
Reference voltage (OFF) Pulse voltage (ON)
0 As determined by power supply,
pull-up resistor, and load. See the installation manual for your transmitter.
As determined by power supply, pull-up resistor, and load. See the installation manual for your transmitter.
0
80 Micro Motion Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
6.3.2 Configure Frequency Output Scaling Method
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET FO > FO SCALE
ProLink III Device Tools > Configuration > I/O > Outputs > Frequency Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Scaling
Overview
Frequency Output Scaling Method defines the relationship between output pulse and flow units. Set Frequency Output Scaling Method as required by your frequency receiving device.
Procedure
1. Set Frequency Output Scaling Method.
Option
Frequency=Flow (default) Frequency calculated from flow rate
Pulses/Unit
Units/Pulse
Description
A user-specified number of pulses represents one flow unit
A pulse represents a user-specified number of flow units
2. Set additional required parameters.
If you set Frequency Output Scaling Method to Frequency=Flow, set Rate Factor
and Frequency Factor.
If you set Frequency Output Scaling Method to Pulses/Unit, define the number of
pulses that will represent one flow unit.
If you set Frequency Output Scaling Method to Units/Pulse, define the number of
units that each pulse will indicate.
For all scaling methods, the transmitter puts out a fixed number of pulses per unit, and at the same time, the Frequency Output signal varies in proportion to flowrate.
Calculate frequency from flow rate
The Frequency=Flow option is used to customize the Frequency Output for your application when you do not know appropriate values for Units/Pulse or Pulses/Unit.
If you specify Frequency=Flow, you must provide values for Rate Factor and Frequency Factor:
Rate Factor
Frequency Factor
The maximum flow rate that you want the Frequency Output to report.
A value calculated as follows:
FrequencyFactor = x N
RateFactor
T
where:
Configuration and Use Manual 81
Integrate the meter with the control system
T
Factor to convert selected time base to seconds
N
Number of pulses per flow unit, as configured in the receiving device
The resulting Frequency Factor must be within the range of the Frequency Output (0 to 10,000 Hz :
If Frequency Factor is less than1 Hz, reconfigure the receiving device for a higher
pulses/unit setting.
If Frequency Factor is greater than 10,000 Hz, reconfigure the receiving device for a
lower pulses/unit setting.
6.3.3 Configure Frequency Output Fault Action and Frequency Output Fault Level
Display Not available
ProLink III Device Tools > Configuration > Fault Processing
Field Communicator Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Fault Parame-
ters > FO Fault Action
Configure > Manual Setup > Inputs/Outputs > Frequency Output > FO Fault Parame-
ters > FO Fault Level
Overview
Frequency Output Fault Action controls the behavior of the Frequency Output if the transmitter encounters an internal fault condition.
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed.
Procedure
1. Set Frequency Output Fault Action as desired.
The default value is Downscale (0 Hz).
2. If you set Frequency Output Fault Action to Upscale, set Frequency Fault Level to the desired value.
The default value is 15000 Hz. The range is 10 to 15000 Hz.
82 Micro Motion Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Options for Frequency Output Fault Action
Options for Frequency Output Fault Action Table 6-5:
Label Frequency Output behavior
Upscale
Downscale
Internal Zero
None (default) Tracks data for the assigned process variable; no fault action
CAUTION!
If you set mA Output Fault Action or Frequency Output Fault Action to None, be sure to set Digital Communications Fault Action to None. If you do not, the output will not report actual process data, and this may result in measurement errors or unintended consequences for your process.
Goes to configured Upscale value:
Range: 10 Hz to 15000 Hz
Default: 15000 Hz
0 Hz
0 Hz
6.4
Restriction
If Digital Communications Fault Action is set to NAN (not a number), you cannot set mA Output Fault Action or Frequency Output Fault Action to None. If you try to do this, the transmitter will not accept the configuration.

Configure the Discrete Output

The Discrete Output is used to report specific meter or process conditions. The Discrete Output parameters control which condition is reported and how it is reported. Your transmitter may have zero or one Discrete Output: Channel B can be configured as a Frequency Output or a Discrete Output.
Restriction
Before you can configure the Discrete Output, you must configure a channel to operate as a Discrete Output.
Important
Whenever you change a Discrete Output parameter, verify all other Discrete Output parameters before returning the meter to service. In some situations, the transmitter automatically loads a set of stored values, and these values may not be appropriate for your application.
Configuration and Use Manual 83
Integrate the meter with the control system
6.4.1 Configure Discrete Output Source
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET DO > DO SRC
ProLink III Device Tools > Configuration > I/O > Outputs > Discrete Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Assignment
Overview
Discrete Output Source controls which device condition or process condition is reported via the Discrete Output.
Procedure
Set Discrete Output Source to the desired option.
The default setting for Discrete Output Source is Flow Direction.
Options for Discrete Output Source
Option
(2)
(1)
D EV x
EVNT1
EVNT2
E1OR2
FL SW
FLDIR
ZERO
FAULT
Discrete Event 1–5
Event 1–2
Flow Switch
Flow Direction
Calibration in Pro­gress
Fault
(1) Events configured using the enhanced event model. (2) Events configured using the basic event model.
Enhanced Event 1
Enhanced Event 2
Enhanced Event 3
Enhanced Event 4
Enhanced Event 5
Event 1
Event 2
Event 1 or Event 2 Status
Flow Switch Indicator
Forward Reverse In­dicator
Calibration in Pro­gress
Fault Indication
Label
Field Communi­cator
Discrete Event x ON Site-specific
Event 1
Event 2
Event 1 or Event 2
Flow Switch ON Site-specific
Forward/Reverse Forward flow 0 V
Calibration in Pro­gress
Fault ON Site-specific
State
OFF 0 V
ON Site-specific
OFF 0 V
OFF 0 V
Reverse flow Site-specific
ON Site-specific
OFF 0 V
OFF 0 V
Discrete Output voltageDisplay ProLink III
84 Micro Motion Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Important
If you assign Flow Switch to the Discrete Output, you should also configure Flow Switch Variable, Flow Switch Setpoint, and Hysteresis.
Related information
Configure an enhanced event Fault indication with the Discrete Output
Configure Flow Switch parameters
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > SET DO > CONFIG FL SW
ProLink III Device Tools > Configuration > I/O > Outputs > Discrete Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > Flow Switch Source
Configure > Manual Setup > Inputs/Outputs > Discrete Output > Flow Switch Set-
point
Configure > Manual Setup > Inputs/Outputs > Discrete Output > Hysteresis
Overview
Flow Switch is used to indicate that the flow rate (measured by the configured flow variable) has moved past the configured setpoint, in either direction. The flow switch is implemented with a user-configurable hysteresis.
Procedure
1. Set Discrete Output Source to Flow Switch, if you have not already done so.
2. Set Flow Switch Variable to the flow variable that you want to use to control the flow switch.
3. Set Flow Switch Setpoint to the value at which the flow switch will be triggered (after Hysteresis is applied).
If the flow rate is below this value, the Discrete Output is ON.
If the flow rate is above this value, the Discrete Output is OFF.
4. Set Hysteresis to the percentage of variation above and below the setpoint that will operate as a deadband.
Hysteresis defines a range around the setpoint within which the flow switch will not change. The default is 5%. The valid range is 0.1% to 10%.
Example: If Flow Switch Setpoint = 100 g/sec and Hysteresis = 5%, and the first measured flow rate is above 100 g/sec, the Discrete Output is OFF. It will remain OFF unless the flow rate drops below 95 g/sec. If this happens, the Discrete Output will turn ON, and remain ON until the flow rate rises above 105 g/sec. At this point it turns OFF and will remain OFF until the flow rate drops below 95 g/sec.
Configuration and Use Manual 85
Integrate the meter with the control system
6.4.2 Configure Discrete Output Polarity
Display OFF-LINE MAINT > OFF-LINE CONFG > IO > CH B > DO > POLAR
ProLink III Device Tools > Configuration > I/O > Outputs > Discrete Output
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Polarity
Overview
Discrete Outputs have two states: ON (active) and OFF (inactive). Two different voltage levels are used to represent these states. Discrete Output Polarity controls which voltage level represents which state.
Procedure
Set Discrete Output Polarity as desired.
The default setting is Active High.
Options for Discrete Output Polarity
Polarity option
Active High
Active Low
Description
When asserted (condition tied to DO is
true), the circuit provides a pull-up to 24 V.
When not asserted (condition tied to DO is
false), the circuit provides 0 V.
When asserted (condition tied to DO is
true), the circuit provides 0 V.
When not asserted (condition tied to DO is
false), the circuit provides a pull-up to 24 V.
6.4.3 Configure Discrete Output Fault Action
Display Not available
ProLink III Device Tools > Configuration > Fault Processing
Field Communicator Configure > Manual Setup > Inputs/Outputs > Discrete Output > DO Fault Action
Overview
Discrete Output Fault Action controls the behavior of the Discrete Output if the transmitter encounters an internal fault condition.
86 Micro Motion Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Note
For some faults only: If Fault Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed.
CAUTION!
Do not use Discrete Output Fault Action as a fault indicator. If you do, you may not be able to distinguish a fault condition from a normal operating condition. If you want to use the Discrete Output as a fault indicator, set Discrete Output Source to Fault and set Discrete Output Fault Action to None.
Procedure
Set Discrete Output Fault Action as desired.
The default setting is None.
Related information
Fault indication with the Discrete Output
6.5
Options for Discrete Output Fault Action
Label
Upscale
Downscale
None (default) Discrete Output is controlled by its assignment
Fault: Discrete Output is ON (site-specific voltage)
No fault: Discrete Output is controlled by its assign-
ment
Fault: Discrete Output is OFF (0 V)
No fault: Discrete Output is controlled by its assign-
ment
Discrete Output behavior
Fault indication with the Discrete Output
To indicate faults via the Discrete Output, set Discrete Output Source to Fault. Then, if a fault occurs, the Discrete Output is always ON and the setting of Discrete Output Fault Action is ignored.

Configure events

An event occurs when the real-time value of a user-specified process variable moves past a user-defined setpoint. Events are used to provide notification of process changes or to perform specific transmitter actions if a process change occurs.
Your transmitter supports two event models:
Basic event model
Configuration and Use Manual 87
Integrate the meter with the control system
Enhanced event model
6.5.1 Configure a basic event
Display Not available
ProLink III Device Tools > Configuration > Events > Basic Events
Field Communicator Not available
Overview
A basic event is used to provide notification of process changes. A basic event occurs (is ON) if the real-time value of a user-specified process variable moves above (HI) or below (LO) a user-defined setpoint. You can define up to two basic events. Event status can be queried via digital communications, and a Discrete Output can be configured to report event status.
Procedure
6.5.2
1. Select the event that you want to configure.
2. Specify Event Type.
Option
HI
LO
Description
x > A
The event occurs when the value of the assigned process variable (x) is greater than the setpoint (Setpoint A), endpoint not included.
x < A
The event occurs when the value of the assigned process variable (x) is less than the setpoint (Setpoint A), endpoint not included.
3. Assign a process variable to the event.
4. Set a value for Setpoint A.
5. (Optional) Configure a Discrete Output to switch states in response to the event status.
Configure an enhanced event
Display Not available
ProLink III Device Tools > Configuration > Events > Enhanced Events
Field Communicator Configure > Alert Setup > Discrete Events
88 Micro Motion Model 1700 Transmitters with Analog Outputs
Integrate the meter with the control system
Overview
An enhanced event is used to provide notification of process changes and, optionally, to perform specific transmitter actions if the event occurs. An enhanced event occurs (is ON) if the real-time value of a user-specified process variable moves above (HI) or below (LO) a user-defined setpoint, or in range (IN) or out of range (OUT) with respect to two user­defined setpoints. You can define up to five enhanced events.
Procedure
1. Select the event that you want to configure.
2. Specify Event Type.
Option Description
HI
LO
IN
OUT
x > A
The event occurs when the value of the assigned process variable (x) is greater than the setpoint (Setpoint A), endpoint not included.
x < A
The event occurs when the value of the assigned process variable (x) is less than the setpoint (Setpoint A), endpoint not included.
A ≤ x ≤ B
The event occurs when the value of the assigned process variable (x) is in
range, that is, between Setpoint A and Setpoint B, endpoints included.
x ≤ A or x ≥ B
The event occurs when the value of the assigned process variable (x) is out of range, that is, less than Setpoint A or greater than Setpoint B, end-
points included.
3. Assign a process variable to the event.
4. Set values for the required setpoints.
For HI and LO events, set Setpoint A.
For IN and OUT events, set Setpoint A and Setpoint B.
5. (Optional) Configure a Discrete Output to switch states in response to the event status.
6. (Optional) Specify the action or actions that the transmitter will perform when the event occurs.
With the display: OFF-LINE MAINT > OFF-LINE CONFG > IO > CH C > SET DI > DI
ACT
With ProLink III: Device Tools > Configuration > I/O > Action Assignment
With the Field Communicator: Configure > Alert Setup > Discrete Events > Assign
Discrete Action
Configuration and Use Manual 89
Integrate the meter with the control system
Options for Enhanced Event Action
Label
Action
Standard
None (default)
Start sensor zero
Start/stop all totalizers
Reset mass total
Reset volume total
Reset gas standard volume total
Reset all totals
Meter verification
Start meter verification test
Note
Before assigning actions to an enhanced event, check the status of the event. If it is ON, all assigned actions will be performed when the new configuration is implemented. If this is not accepatable, wait until an appropriate time to assign actions to the event.
Display ProLink III Field Communicator
NONE None None
START ZERO Start Sensor Zero Perform auto zero
START STOP Start/Stop All Totalization Start/stop totals
RESET MASS Reset Mass Total Reset mass total
RESET VOL Reset Volume Total Reset volume total
RESET GSVT Reset Gas Std Volume Total Reset gas standard volume total
RESET ALL Reset All Totals Reset totals
START VERFY Start Meter Verification
Not available
6.6
6.6.1
90 Micro Motion Model 1700 Transmitters with Analog Outputs

Configure digital communications

The digital communications parameters control how the transmitter will communicate using digital communications.
Your transmitter supports the following types of digital communications:
HART/Bell 202 over the primary mA terminals
HART/RS-485 over the RS-485 terminals
Modbus/RS-485 over the RS-485 terminals
Modbus RTU via the service port
Note
The service port responds automatically to a wide range of connection requests. It is not configurable.
Configure HART/Bell 202 communications
HART/Bell 202 communications parameters support HART communications with the transmitter's primary mA terminals over a HART/Bell 202 network.
Integrate the meter with the control system
Configure basic HART parameters
Display OFF-LINE MAINT > OFF-LINE CONFG > COMM
ProLink III Device Tools > Configuration > Communications > Communications (HART)
Field Communicator Configure > Manual Setup > Inputs/Outputs > Communications > HART Communica-
tions
Overview
Basic HART parameters include the HART address, HART tags, and the operation of the primary mA Output.
HART/Bell 202 communications parameters support HART communication with the transmitter's primary mA terminals over a HART/Bell 202 network. The HART/Bell 202 communications parameters include:
HART Address (Polling Address)
mA Output Action
Burst Parameters (optional)
HART Variables (optional)
Procedure
1. Set HART Address to a value that is unique on your network.
Default: 0
Range: 0 to 15
Tips
The default address is typically used unless you are in a multidrop environment.
Devices using HART protocol to communicate with the transmitter may use either HART
Address or HART Tag (Software Tag) to identify the transmitter. Configure either or both, as required by your other HART devices.
2. Ensure that mA Output Action is configured appropriately.
Option
Enabled (Live)
Disabled (Fixed)
Description
The primary mA Output reports process data as configured.
The primary mA Output is fixed at 4 mA and does not report process data.
Configuration and Use Manual 91
Integrate the meter with the control system
Important
If you use ProLink III to set HART Address to 0, the program automatically enables mA Output Action. If you use ProLink III to set HART Address to any other value, the program automatically disables mA Output Action. This is designed to make it easier to configure the transmitter for legacy behavior. Always verify mA Output Action after setting HART Address.
Configure burst parameters
Display Not available
ProLink III Device Tools > Configuration > Communications > Communications (HART)
Field Communicator Configure > Manual Setup > Inputs/Outputs > Communications > Set Up Burst Mode
Overview
Burst mode is a mode of communication during which the transmitter regularly broadcasts HART digital information over the mA Output. The burst parameters control the information that is broadcast when burst mode is enabled.
Tip
In typical installations, burst mode is disabled. Enable burst mode only if you are using a HART Triloop.
Procedure
1. Enable Burst Mode.
2. Set Burst Mode Output as desired.
Label
Source (Primary Variable) PV
Primary Variable (Percent Range/Current)
Process Variables/Current Process variables/current
Transmitter variables Fld dev var
% range/current
DescriptionProLink III Field Communicator
The transmitter sends the primary variable (PV) in the con­figured measurement units in each burst (e.g., 14.0 g/sec,
13.5 g/sec, 12.0 g/sec.
The transmitter sends the PV’s percent of range and the PV’s actual mA level in each burst (e.g., 25%, 11.0 mA.
The transmitter sends PV, SV, TV, and QV values in meas­urement units and the PV’s actual milliamp reading in each burst (e.g., 50 g/sec, 23 °C, 50 g/sec, 0.0023 g/cm3,
11.8 mA.
The transmitter sends four user-specified process variables in each burst.
3. Ensure that the burst output variables are set appropriately.
92 Micro Motion Model 1700 Transmitters with Analog Outputs
Loading...