Proper Environment
Current
- Protection class: IP 65 (applies only when the sensor cable is plugged in) Optical inputs are excluded from protection class. Contamination leads to impairment or failure of the function.
- Operating temperature: 0 °C ... 50 °C (+32 up to +104 °F)
- Storage temperature: -20 °C ... 70 °C (-4 up to +158 °F)
- Humidity: 5 - 95 % (non-condensing)
- Ambient pressure: Atmospheric pressure
Sensor Mounting, Dimensions
The optoNCDT 1750 sensor is an optical system for measurements with micrometer accuracy. Pay attention to careful handling during mounting and operation.
Mount the sensor only to the existing holes on a flat surface. Clamps of any kind are not
permitted.
Use three M4 screws to mount the sensors. The bearing surfaces surrounding the fastening
holes (through-holes) are slightly raised.
Measuring range, Start of Measuring range
1
MR = Measuring range
SMR = Start of measuring range
MMR = Mid of measuring range
EMR = End of measuring range
1) For displacement values without zero setting or mastering.
SMRMR
Voltage Digital value
3 mA 262077
4 mA (SMR) 98232
12 mA (MMR) 131000
20 mA (EMR) 163768
3 mA
5,2 V / 10,2 V
0 V
2,5 V / 5 V
5 V / 10 V
5,2 V / 10,2 V 262078
Mounting
Mount the sensor only to the existing holes on a
i
sensor off
RS422 on
output off
analog on
optoNCDT
-
LASER KLASSE 1
Class 1 Laser Product
flat surface. Clamps of any kind are not permitted. Do not exceed torques. The laser beam must be directed perpendicularly onto the surface of the target. In case of misalignment it is possible that the measurement results will not always be accu­rate.
Sensor mounting with direct reflection
Drawings
75 (2.95)
15 (.59) (1.18)
30
90 °
MR 2 (.08)
a 1 (.04)
b 26.5 (1.04)
c 25 (.98)
d 20 °
e 16.7 (.66)
f 20.7 (.81) g 82.6 (3.25) h 83.7 (3.30)
i 49.5 (1.95)
j 45.6 (1.80)
Mounting
2 Nm
Washer A4.3; ISO 7089 - A2 M4 x 35; ISO 4762-A2 Bolt connection
min 5
97 (3.81)
i
13.4
(.53)
d
j
a
MR
b
Direct reflecting target
Dimensions in mm (inches) MR = Measuring range
Pin Assignment
Signal Pin Description
+U
B
GND 6
5 Supply voltage (11 ... 30 VDC) red
System ground supply, switch signals (Laser on/off, Zero, Limits)
Current 4 ... 20 mA (RB < (UB - 6 V) / 20 mA)
Analog output 13
Voltage 0 ... 5 VDC Voltage 0 ... 10 VDC (Ri = 50 Ohm, I
= 5 mA)
max
AGND 14 Reference potential for analog output
Laser on/off 9
h
g
Multi-function input 10
Switching input, Laser operates when pin 9 is connected to GND
Switching input, TrigIn, Zero/Master, TeachIn, SlaveIn
Switching output 1 8 Error/Limit 1 gray and pink
Switching output 2 7
f
e
c
Sync + 3
Sync - 4 pink
Tx + 1
Tx - 2 brown
Rx + 12
Rx - 11 yellow
View: Solder-pin side male cable connector, insulator
Limit 2, programmable switching characteristic: (NPN, PNP, Push-Pull)
Symmetrical synchronous output (Master) or input (Slave) RS422 level, terminating resistor 120 Ohm switchable, in­put or output depends on selected synchronization mode
RS422 - Output (symmetric) terminate with 120 Ohm receive-site
RS422 - Input (symmetric) internally terminated with 120 Ohm
The PC1700 sensor cable is
1
10
2
12
3
4
5
9
11
14
8
13
7
6
qualified for drag chain use. One end of the cable has a molded cable connector, the other end has braids with ferrules. Connec­tor: ODU MINI-SNAP, 14 poles, B series, size 2, coding 0, IP 68
Supply Voltage, Nominal value: 24 V DC (11 ... 30 V, P < 3 W)
11 ... 30 VDC
5
ILD1750
6
MICRO-EPSILON MESSTECHNIK
Koenigbacher Str. 15 · 94496 Ortenburg
www.micro-epsilon.com
Sensor
Pin
5 red +U
6 black Ground
GmbH & Co. KG
PC1700-x/Y
Color
Supply
B
Use supply voltage for measurement in­struments only. MICRO-EPSILON recom­mends using an optional available power supply unit PS2020 for the sensor.
X9771376.02-A012030SWE
Cable
PC1700-x
black
Coaxial inner
conductor
Screening,
black
red and blue
white and
green
violet
blue
green
gray
Assembly Instructions
optoNCDT 1750-2DR
Proper Use
The optoNCDT 1750-2DR system is designed for use in industrial and laboratory areas. It is used for measuring displacement, distance and position as well as in in-process quality control and dimen­sional testing. The sensor may only be operated within the limits specified in the technical data, see instruction manual, Chap. 3.3. The sensor must be used in such a way that no persons are endangered or machines are damaged in case of malfunctions or total failure of the sensor. Take additional precau­tions for safety and damage prevention for safety-related applications.
Warnings
Avoid unnecessary laser radiation to be exposed to the human body. Switch off the sensor for clean­ing and maintenance, for system maintenance and repair if the sensor is integrated into a system. Caution - use of controls or adjustments or performance of procedures other than those specified may cause harm. Connect the power supply and the display-/output device in accordance with the safety regulations for electrical equipment. The power supply may not exceed the specified limits.
> Risk of injury. Damage to or destruction of the sensor.
Avoid continuous exposure to splashing water on the sensor and the controller. Avoid exposure to aggressive materials (washing agent, cooling emulsions) on the sensor.
> Damage to or destruction of the sensor.
Avoid shock and vibration to the sensor. Protect the sensor cable against damage.
> Damage to or destruction of the sensor , failure of the measuring device.
Laser Class
The optoNCDT 1750-2DR sensors operate with a semiconductor laser with a wavelength of 670 nm (visible/red). The following warning labels must be attached to the cover (front and rear side) of the sensor housing:
LASER KLASSE 1
-
Class 1 Laser Product
Warning label and laser label, ILD1750-2DR Only for USA
Never deliberately look into the laser beam! Consciously close your eyes or turn away immediately if ever the laser beam should hit your eyes.
COMPLIES WITH 21 CFR 1040.10 AND 1040.11
EXCEPT FOR CONFORMANCE WITH
IEC 60825-1 ED. 3., AS DESCRIBED IN
LASER NOTICE NO. 56, DATED MAY 8, 2019
Page 2
Laser On
11... 30 VDC
11... 30 VDC
Type 1
Type 2 Type 3
1
PC1700-x
red-blue
black
GND
ILD1750
+U
H
9
6
If pin 9 is not connected
i
with pin 6, the laser is off.
Analog Output
The current output may not be continu-
i
ously operated in short-circuit opera-
Current output 4 ... 20 mA or
Voltage output 0 ... 5 V or 0 ... 10 V
5
6
5
6
ILD1750
ILD1750
13
14
13
R
i
14
I
out
U
R
out
R
B
L
C
I
C
U
tion without load resistor. This would lead to thermal overload and thus to the automatic overload cut-off of the output.
Current output
RB < (UB -6 V) / 20 mA;
RB max. = 250 Ohm at UB = 11 V
CI ≤ 33 nF
Voltage output
Ri = 50 Ohm, I
Short circuit protection 7 mA
RL > 20 MOhm
CU ≤ 100 nF
= 5 mA,
max
Multi-Function Input
The multi-function input enables triggering, zero setting/mastering and teaching. The function depends on the programming of the input and on the timing of the input signal.
The inputs are not electrically isolated. The maximum switching frequency is 10 kHz.
24 V logic (HTL):
Low level ≤ 3 V; High level ≥
8 V (max 30 V)
5 V logic (TTL):
Low level ≤ 0.8 V; High level ≥ 2 V
internal pull-up resistor, an open input is detected as High.
1
PC1700-x
white-green
black
Type 1
Type 2 Type 3
Connect the input to GND to trigger the function.
GND
10
ILD1750
+U
H
6
RS422 Connection with USB Converter IF2001/USB
Cross the lines for connections between sensor and PC.
Disconnect or connect the D-sub connection between RS422 and USB converter when the
i
sensor is disconnected from power supply only.
Sensor End device (converter)
14-pin cable connector
Tx + (Pin 1) green Rx + (Pin 3)
Tx -(Pin 2) brown Rx -(Pin 4)
Rx + (Pin 12) gray Tx + (Pin 1)
Rx -(Pin 11) yellow Tx -(Pin 2)
GND (Pin 6) black GND (Pin 9)
Symmetric differential signals acc. to EIA-422, not electrically isolated from supply voltage. Use a shielded cable with twisted cores e.g. PC1700-x.
Sensor cable
Type IF2001/USB from MICRO-EPSILON
Quick Guide
Components
Mount the sensor and connect the components.
Source Cable/Supply Interface
PS 2020
Sensor supply is done by peripheral.
PC1700-x/OE
PC1700-x/OE
PC1700-x/IF2008 (IF2008-Y)
PC1700-x/OE
PC1700-x/IF2008
and IF2008-Y adapter cable
PC1750-x/C-Box/RJ45
IF2001/USB
IF2008/PCIE
IF2030/PNET
IF2004/USB
USB
USB Ethernet
C-Box/2A
SPS
PC
Commissioning
Connect the sensor to a PC/notebook via a RS422 connec-
tor. Connect the supply voltage.
Start the program Sensor Connect Vx.x.x. You will find
this program on the provided CD.
You need a web browser (e. g. Mozilla Firefox or Internet Ex­plorer) on a PC/notebook.
Select a browser in the
Browser type dialog.
Select the desired sensor.
Click on the button Con-
nect to selected Sensor and start Browser.
Access via Web Interface
Interactive web pages for programming the sensor now appear in the web browser. The sensor is active and supplies measurement values. The ongoing measurement can be operated by means of function buttons in the area Measurement chart.
In the top navi­gation bar other auxiliary func­tions (settings, measurement chart etc.) are available.
The appearance of the websites can change dependent of the functions. Each page contains descriptions of parameters and so tips for filling the website.
Select a Measuring Rate
Go to the menu Settings > Data recording > Measuring rate.
Start with a medium measuring rate. Select a measuring rate from the list. Confirm with Apply.
Select an Interface
Go to the menu Settings > Output > Output interface.
Defines which interface is used for output of measured values. A parallel output o measured val­ues via multiple channels is not possible. RS422 and analog output cannot be operated simulta­neously. While using the web interface, the output is switched off via RS422.
Place target
Position the target (measurement object) as much as possible in the midrange.
100 %
Signal
50
0
SMR MMR EMR
Measuring object
Measuring range
The State LED on the sensor indicates the position of the target to the sensor.
LED Color Labeling Meaning
off Laser off Laser beam is switched off
green In range Target within measuring range
State
yellow Midrange Target within the midrange
red Error
Target outside the measuring range, too low reflection
Displacement
Store the Settings
Go to the menu Settings > System settings > Load & Stores or click the Save
settings button.
Read the detailed instruction manual before using the sensor. The manual is available online on
www.micro-epsilon.com/download/manuals/man--optoNCDT-1750--en.pdf or on the supplied
CD.