Mettler Toledo XS304, XS204DR, XS205DU, XS225DU, XS64 Data Sheet

...
Short Operating Instructions
For further information, please refer to the operators Manual
METTLER TOLEDO XP / XS Analytical- and Precision Balances
Basic Functions:
Switching on and off Display
1. Swi tch o n di sp lay:
Press the «O n/O ff» key briefly. The display performs a test and is then ready to weigh.
On
Off
2. Switch off Disp lay ; Press and hold the «On/Off» key until the message “OFF” ap­pears in the display. The display then fades and the balance is switched off.
On
Off
User Profile: *only XP
Choose User Profile
B
This key can be used to call up the desired user profile. Different settings can be stored in a user profile.
Example
Zero Display
Ze roi ng: Pr ess t he «G» to set a new zero point.
G
Always use the «G» befo re you s tart the weighing procedure.
Weighing Setup: For more precision / shorter stability time
Weighing filters: Weighing mode
1. Choose System-Set­up
Tare Display
Taring: If you work with a Tare container, first zero the balance. Then put the Tare container on the balance and press «H», to tare the balance.
H
Weighing filters: Environment
1. Choose System-Set­up
C C C
2. Define Weighing pa­rameters
3. Define the Weighing mode
Advise: Solids: «Universal»
Powder, Liquid: «Dosing»
2. Define Weighing pa­rameters
3. Define Environment
Advise: Fume cupboard – un­stable weighing table: «
Unstable»,
«Very unstable»
Stable Weighing table, no drafts «Standard», «Stable»
Weighing­Applications
XS Balance
XP Balance
Choose the applica ­tio n: If yo u w ant to change the application, then press the applica­tion menu button and choose a nother a p­plication.
Weighing filters: Value Release
1. Choose System-Set­up
2. Define Weighing pa­rameters
3. Define Value Release
Advise: Highest Precision: «Very reliable»
Fast Value Release: «Very fast», but less precision
Test:Additional Functions:
Levelling, Internal Adjustment (Factory setting)
1. Level indicator is built­in in the device.
2. Level indicator is not centered.*
3. Release the clamps.*
4. Levelling. After this pro­cess, lock the clamps again.*
5. Internal Adjustment
Pressing this function key triggers a balance adjustment using the bu ilt -in ca lib rat ion wei ght. This ca n b e imp lemented at any time.
* depend the balance
EXCELLENCE PLUS
METTLER TOLEDO Excellence Plus and Excellence Balances
Overview of the Terminal XP Analytical- and Precision balances:
XP Display Ter­minal
1. Button SmartSens: Each of these tw o
contact-free sensors can be assigned a key or menu function.
2. Button Home: This key can be used
to return to the “Home” user profile from any menu level in any application.
3. Button User profile: This key can be used
to call up the desired user profile. Different settings can be stored in a user profile.
Overview of the Terminal XS Analytical- and Precision balances:
XS Display Ter­minal
1
2
3
4
5
4. Status bar: If the left or right
SmartSe ns is as­signed a key func­tion, then the cor­responding symbol is highlighted.
On touch
- e.g. start the internal adjustment
5. System-Setup: Define the bas ic
settings.
6. Select the applica­tion.
7. Button Applications­Setup:
Each appli cat ion
can be optimized to the current task using a number of settings.
One touch
- change the time
8. Printer: Pressin g this key
1
6
7
4
8
transmits data via the interface.
1. Button Home: This key can be used
to return to the “Home” user profile from any menu level in any application.
2. Weighing application: By pressing this symbol
you can choose the desired application.
3. Button Applications­Setup:
Each application can
be ideally adapted to the current task using a number of settings.
© Mettler-Toledo AG 2007 11781027 Printed in Switzerland 0707/7.12
Mettler-Toledo AG, Laboratory & Weighing Technologies, CH-8606 Greifensee, Switzerland Phone +41-44-944 22 11, Fax +41-44-944 30 60, Internet: http://www.mt.com
1
2
4. Printer: Pressin g this key
transmits data via the interface.
On touch
- e.g. start the internal adjustment
3
4
*11781027*