Mettler Toledo ML4002E Instruction Manual

C
C
C
C
C
NewClassic Balances
ML Models
Operating Instructions

Table of contents

Introduction1
Conventions and Symbols Used in These Operating Instructions1.1
Safety Precautions2
Overview3
Components3.1 Operation Keys3.2 Display Panel3.3
Setting up the Balance4
Unpacking and Delivery Inspection4.1 Installing the Components4.2 Selecting the Location and Leveling the Balance4.3 Selecting the Location4.3.1 Leveling the Balance4.3.2 Power Supply4.4 AC Operation4.4.1 Battery Operation4.4.2 Transporting the Balance4.5 Weighing Below the Balance4.6 Adjustment (Calibration)4.7 Adjustment with Internal Weight4.7.1 Adjustment with External Weight4.7.2
7
7
8
9
9 10 11
13
13 14 16 16 16 17 17 17 18 19 19 19 20
Weighing Made Simple5
Switching the Balance On and Off5.1 Performing a Simple Weighing5.2 Zero Setting / Taring5.3 Switching Weight Units5.4 Recall / Recall Weight Value5.5 Weighing with the Weighing-in Aid5.6 Print / Transmit Data5.7
The Menu6
What is in the Menu ?6.1 Menu Operation6.2 Description of Menu Topics6.3 Main Menu6.3.1 Basic Menu6.3.2 Advanced Menu6.3.3 Interface Menu6.3.4
Application "Piece Counting"7
Application "Percent Weighing"8
Application "Check Weighing"9
Application "Statistics"10
22
22 23 23 23 24 24 24
25
25 26 27 27 27 29 32
38
40
42
45
Table of contents
3
Application "Formulation" (Net Total Formulation)11
47
Application "Totaling"12
Application "Dynamic Weighing"13
Application "Multiplication Factor Weighing"14
Application "Division Factor Weighing"15
Application "Density"16
Density Determination of Solids16.1 Density Determination of Liquids16.2 Formulae Used to Calculate Density16.3
Application "Routine Test"17
Application "Diagnostics"18
Repeatability Test18.1 Display Test18.2 Key Test18.3 Motor Test18.4 Balance History18.5 Calibration History18.6 Balance Information18.7 Service Provider Information18.8
Communication with Peripheral Devices19
Function PC-Direct19.1
49
51
53
55
57
57 59 60
63
66
66 67 68 69 69 70 71 72
73
73
Firmware (Software) Updates20
Operating Principle20.1 Update Procedure20.2
Error and Status Messages21
Error Messages21.1 Status Messages21.2
Cleaning and Service22
Interface Specification23
RS232C Interface23.1 MT-SICS Interface Commands and Functions23.2
Technical Data24
General Data24.1 Model-Specific Data24.2 Balances with Readability of 0.1 mg with Draft Shield24.2.1 Balances with Readability of 1 mg with Draft Shield24.2.2 Balances with Readability of 0.01 g24.2.3 Balances with Readability of 0.1 g24.2.4 Dimensions24.3
75
75 75
76
76 77
78
79
79 79
82
82 82 82 83 84 86 88
4
Table of contents
24.3.4
24.3.5
Balances with Readability of 0.1 mg with Draft Shield (235 mm)24.3.1 Balances with Readability of 1 mg with Draft Shield (235 mm)24.3.2 Balances with Readability of 1 mg with "Flex" Draft Shield (105 mm)24.3.3 Balances with Readability of 0.01 g with Square Weighing Pan and
Draft Shield Element Balances with Readability of 0.01 g / 0.1 g with Round Weighing
Pan Balances with Readability of 0.1 g with Square Weighing Pan24.3.6
88 89 90 91
92
93
Accessories and Spare Parts25
Accessories25.1 Spare Parts25.2
Appendix26
Conversion Table for Weight Units26.1
Index27
94
94 97
100
100
101
Table of contents
5

1 Introduction

Thank you for choosing a METTLER TOLEDO balance.
The precision balances of the NewClassic line combine a large number of weighing possibilities with easy operation.
These operating instructions apply to all balance models ML in the NewClassic line. However, the different models have different charcteristics regarding equipment and performance. Special notes in the text indicate where this makes a difference to operation.

1.1 Conventions and Symbols Used in These Operating Instructions

Key designations are indicated by double angular brackets (e.g. « »).
This symbol indicates press key briefly (less than 1.5 s).
This symbol indicates press and hold key down (longer than 1.5 s).
This symbol indicates a flashing display.
This symbol indicates an automatic sequence.
These symbols indicate safety notes and hazard warnings which, if ignored, can cause per sonal danger to the user, damage to the balance or other equipment, or malfunctioning of the balance.
This symbol indicates additional information and notes. These make working with your bal ance easier, as well as ensuring that you use it correctly and economically.
7Introduction

2 Safety Precautions

Always operate and use your balance only in accordance with the instructions contained in this manual. The instructions for setting up your new balance must be strictly observed.
If the balance is not used according to these Operating Instructions, protection of the balance may be impaired and METTLER TOLEDO assumes no liability.
It is not permitted to use the balance in explosive atmosphere of gases, steam, fog, dust and flammable dust (hazardous environments).
For use only in dry interior rooms.
Do not use sharply pointed objects to operate the keyboard of your balance! Although your balance is very ruggedly constructed, it is nevertheless a precision instrument. Treat it with corresponding care.
Do not open the balance: It does not contain any parts which can be maintained, repaired, or replaced by the user. If you ever have problems with your balance, contact your METTLER TOLEDO dealer.
Use only balance accessories and peripheral devices from METTLER TOLEDO; they are opti mally adapted to your balance.
Use only the original universal AC adapter delivered with your balance.
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point specified for electrical and electronic equipment. If you have any questions, please contact the responsible authority or the distributor from which you purchased this device. Should this device be passed on to other parties (for private or professional use), the content of this regu lation must also be related.
Thank you for your contribution to environmental protection.
Safety Precautions
8

3 Overview

C
0.1 mg
C
1 mg
C
1 mg
C
0.01 g

3.1 Components

9Overview
C
0.01g / 0.1g
C
0.1 g
11
12
13
10
14
Name and Function of Components
C
51 42
9
3 6
7
8
1 Display 8 Glass draftshield 2 Operation keys 9 Handle for operation of the draft-shield door 3 Model sticker (with approved models only) 10 Socket for AC Adapter 4 Level indicator 11 RS232C serial interface 5 Weighing pan 12 Kensington slot for anti-theft purposes 6 Draft shield element 13 Slot for second interface (optional) 7 Leveling foot 14 Product label

3.2 Operation Keys

10 Overview
Key Functions
Application Icons
Status Icons
Weighing-in aid
Weight Value Field
Unit Field
No. Key Press briefly (less than 1.5 s) Press and hold (longer than 1.5 s)
1
To change display resolution (1/10d
no function
function) while application is running
2
To navigate back (scroll up) within menu topics or menu selections
Decrease (numerical) parameters within
To select the weighing application
Decrease (numerical) parameters quickly within menu and in applications
menu and in applications
3
To navigate forward (scroll down) within menu topics or menu selections
To select assigned application and enter ing the parameter settings of application.
4 ON//OFF
5
6
7
8
9
Increase (numerical) parameters within menu and in applications
Switch on
Zero/Tare
Cancel and to leave menu without saving (one step back in the menu).
With entries: scroll down
To navigate through menu topics or menu selections
To toggle between unit 1, recall value (if selected), unit 2 (if different from unit 1) and the application unit (if any)
To enter or leave menu selection (from / to menu topic)
To enter application parameter or switch to next parameter
To store parameter
Printout display value
Default application assignment: Piece counting
Increase (numerical) parameters quickly within menu and in applications
Switch off
Switch off
no function
Execute predefined adjusting (calibra tion) procedure
Enter or leave menu (Parameter settings)

3.3 Display Panel

Printout active user menu settings
Transfer data
11Overview
Application Icons
Menu locked Application “Formulation / Net-Total”
Menu setting activated Application “Totaling”
Application "Weighing" Application “Dynamic weighing”
Application "Piece counting" Application “Multiplication factor”
Application "Percent weighing" Application “Division factor”
Application "Check weighing" Application “Density”
Application "Statistics"
Status Icons
Indicates stored value (Memory) Acoustic feedback for pressed keys activated
Indicates Net weight values Weighing range 1 (Dual Range models only)
Adjustments (calibration) started Weighing range 2 (Dual Range models only)
Applications "Diagnostics" and “Routine Test” Charge of battery: full, 2/3, 1/3, discharged
Service reminder
Weight Value Field and Weighing-in aid
Indicates negative values Brackets to indicate uncertified digits
(Battery operated models only)
(approved models only)
Unit Field
Indicates unstable values Marking of nominal or target weight
Indicates calculated values Marking of tolerance limit T+
Marking of tolerance limit T-
g gram ozt troy ounce tls Singapore taels
kg kilogram GN grain tlt Taiwan taels
mg milligram dwt pennyweight tola tola
ct carat mom momme baht baht lb pound msg mesghal
oz ounce tlh Hong Kong taels
12 Overview

4 Setting up the Balance

The balance must be disconnected from the power supply when carrying out all setup and mounting work.

4.1 Unpacking and Delivery Inspection

a) Open the packaging and carefully remove all components.
b) Check the delivered items.
The standard scope of delivery contains the following items:
Components Balances with round weighing pan Balances with
square weighing pan
0.1 mg 1 mg 0.01 g 0.1 g 0.01 g 0.1 g
Draft shield 235 mm – Flex draft shield 105 mm – Weighing pan Ø 90 mm
Ø 120 mm – Ø 160 mm – 170 x 190
mm Draft shield element – Pan support Bottom plate – Protective cover Universal AC adapter (country specific) Start-up or Operating Instructions
1)
Quick Guide CD-ROM
1)
EC declaration of conformity
1)
depending on selected countries
13Setting up the Balance

4.2 Installing the Components

C
2
3
4
1
5
C
2
3
4
1
Balances with readability of 0.1 mg, with draft shield (235 mm)
Place the following components on the balance in the specified order:
Note: Push the side glass door (1) back as far as will go.
Bottom plate (2)
Draft shield element (3)
Pan support (4)
Weighing pan (5)
Balances with readability of 1 mg, with draft shield (235 mm)
Place the following components on the balance in the specified order:
Note: Push the side glass door (1) back as far as will go.
Bottom plate (2)
Pan support (3)
Weighing pan (4)
Setting up the Balance
14
C
2
3
4
1
6
5
Balances with readability of 1 mg, with flex draft
C
5
4
3
2
1
C
2
1
shield (105 mm)
Place the following components on the balance in the specified order:
a) Place bottom plate (1) (push and turn)
b) Place pan support (2) with weighing pan (3).
c) Unfold the draft shield frame (4) and place it cor
rectly.
d) Insert the glass panels (5) into the draft shield
frame.
e) Place draft shield top (6)
Balances with readability of 10 mg with square weighing pan and draft shield element
a) Remove the two screws (1)
b) Remove the plate (2) and retain it.
c) Place draft shield element (3) and fix it with the
two screws.
d) Place pan support (4) with weighing pan (5).
Balances with readability of 0.01 g / 0.1 g with round weighing pan
Place the following components on the balance in the specified order:
Pan support (1)
Weighing pan (2)
15Setting up the Balance
C
2
1
Balances with readability of 0.1 g with square
C
weighing pan
Place the following components on the balance in the specified order:
Pan support (1)
Weighing pan (2)

4.3 Selecting the Location and Leveling the Balance

Your balance is a precision instrument and will thank you for an optimum location with high accuracy and dependability.

4.3.1 Selecting the Location

Select a stable, vibration-free position that is as horizontal as possible. The surface must be able to safely carry the weight of a fully loaded balance.
Observe ambient conditions (see Technical Data).
Avoid the following:
Direct sunlight

4.3.2 Leveling the Balance

Powerful drafts (e.g. from fans or air conditioners)
Excessive temperature fluctuations
The balances have a level indicator and two or four adjustable leveling feet to compensate for slight irregularities in the surface of the weighing bench. The balance is exactly horizontal when the air bubble is in the middle of the level glass.
Note: The balance should be leveled and adjusted each time it is moved to a new location.
Balances with 2 leveling feet
Adjust the two front leveling feet appropriately until the air bubble comes to rest exactly in the middle of the glass:
Air bubble at "12 o'clock" turn both feet clockwise
Air bubble at "3 o'clock" turn left foot clockwise, right
foot counterclockwise
Air bubble at "6 o'clock" turn both feet counterclockwise
Air bubble at "9 o'clock" turn left foot counterclockwise,
right foot clockwise
16 Setting up the Balance
Balances with 4 leveling feet
a) First turn the two rear leveling feet all the way in.
b) Adjust the two front leveling feet as previously described.
c) Turn the rear leveling feet down onto the surface for extra stabiliz
ing safety, so the balance cannot tilt over under eccentric loads.

4.4 Power Supply

Allow your balance to warm up for 30 minutes (0.1 mg models 60 minutes) to enable it to adapt itself to the ambient conditions.

4.4.1 AC Operation

Your balance is supplied with an country-specific AC adapter or with a country-specific power cable. The power supply is suitable for all line voltages in the range: 100 - 240 VAC, 50/60 Hz (for exact specifications, see section "technical data").
First, check the local line voltage is in the range 100 - 240 VAC, 50/60 Hz and whether the power plug fits your local power supply connection. If this is not the case, on no account connect the balance or the AC adapter to the power supply, but contact the responsible METTLER TOLEDO dealer.
Important:
Before operating, check all cables for damage.
Guide the cables so that they cannot become damaged or interfere with the weighing process!
Take care that the AC adapter cannot come into contact with liquids!
The power plug must be always accessible.
Connect the AC adapter to the connection socket on the back of your balance (see figure) and to the power line.

4.4.2 Battery Operation

The Balance can also operates with batteries. Under normal operation conditions, the balance works indepen dently of the AC power line for about 8 to 15 hours (using alkaline batteries). Immediately after the AC power supply is interrupted e.g. by withdrawing the power plug or if there is a power failure, the balance switches automatically to battery operation. Once the AC power supply is restored, the balance reverts automatically to AC operation.
Note: It is also possible to use rechargeable batteries. Charging batteries inside the balance is not possible.
17Setting up the Balance
battery fully When the balance is operating on its batteries, the battery symbol in
2/3 full
1/3 full
battery empty
Inserting / Replacing Batteries
The balance must be disconnected from the power supply when carrying out all setup and mounting work.
Make sure that the balance is off before removing or inserting batteries.
Do not place the balance on the pan support location bolt.
Battery Warnings: Read and follow all warnings and instructions supplied by the battery manufacturer.
Do not mix different types or brands of batteries. Performance of batteries can vary very greatly depending on the manufacturer.
the display lights up. The number of segments that are lit is an indica tor of battery condition (3 = fully charged, 0 = discharged). When the batteries are almost completely discharged, the battery symbol flashes.
If you don't operate the balance with batteries for an extended period, it is recommended to remove the batteries from the balance.
Batteries must be disposed of in an environmentally responsible manner. No attempt must be made to incinerate or disassemble item.
Your balance uses 8 standard AA (LR6) batteries (alkaline batteries preferred)
a) Remove weighing pan, pan support and draft shield element or
draft shield "100 mm" if present.
b) Turn the balance carefully on its side.
c) Open and remove the battery-chamber cover.
d) Insert / replace the batteries with the correct polarity as shown in
the battery holder.
e) Insert and close the battery-chamber cover.
f) Turn the balance carefully to its normal position.
g) Reinstall all components in the reverse order.

4.5 Transporting the Balance

Switch off the balance and remove the power cable and any interface cable from the balance. Refer to the notes in Section "Selecting the location" regarding the choice of an optimal location.
Setting up the Balance
18
Transporting Over Short Distances
A
For balances with a draft shield: Observe the following instructions to transport your bal ance over a short distance to a new location: Never lift the balance using the glass draft
shield. The draft shield is not sufficiently fastened to the balance.
Transporting Over Long Distances
If you would like to transport or send your balance over long distances, use the complete original packaging.

4.6 Weighing Below the Balance

The balances are equipped with a hanger for carrying out weighings below the work surface (weighing below the balance).
Attention:
Do not place the balance on the pan support location bolt.
a) Switch off the balance and remove the power cable and any
interface cable from the balance.
b) Remove weighing pan, pan support and draft shield element
or "Easy draft shield" if present.
c) Turn the balance carefully on its side.
d) Remove one of the caps (A) depending on the models.
e) Then turn the balance to its normal position and simply rein
stall all components in the reverse order.

4.7 Adjustment (Calibration)

To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its location. Adjusting is necessary:
before the balance is used for the first time.
at regular intervals during weighing service.
after a change of location.

4.7.1 Adjustment with Internal Weight

Note: On models with internal weight only (see technical data).
To obtain accurate results, the balance must be connected to the power supply for approximately,
30 minutes for balances with redability of 1 mg to 5 g
60 minutes for balances with redability of 0.01 mg to 0.1 mg
in order to reach operating temperature before adjusting.
19Setting up the Balance
------
ADJ.DONE
0.00
g
ADJ.INT
Requirement: To carry out this operation, in the menu topic "CAL"
- Internal Adjustment --
21.Jan 2009 12:56
METTLER TOLEDO
Balance Type ML4002 SNR 1234567890
Temperature 22.5 °C Diff 3 ppm
Adjustment done
-----------------------
------
2000.00
g
ADJ.DONE
0.00
g
C
0.00
g
------
C
ADJ.EXT
(Adjustment) of advanced menu "ADJ.INT" must be selected.
a) Unload weighing pan
b) Press and hold « » to execute "Internal Adjustment".
The balance adjusts itself automatically. The adjusting is finished when the message "ADJ.DONE" appears briefly on the display. The balance returns to the last active application and is ready for operation.
Sample adjustment printout using internal weight:

4.7.2 Adjustment with External Weight

Note: Because of certification legislation, the approved models cannot be adjusted with an external weight *
(depend on selected countries' certification legislation). * except OIML accuracy class I approved models.
20
Setting up the Balance
To obtain accurate results, the balance must be connected to the power supply for approximately,
30 minutes for balances with redability of 1 mg to 5 g
60 minutes for balances with redability of 0.01 mg to 0.1 mg
in order to reach operating temperature before adjusting.
Requirement: To carry out this operation, in the menu topic "CAL"
(Adjustment) " of advanced menu ADJ.EXT" must be selected.
a) Have required adjustment weight ready.
b) Unload weighing pan.
c) Press and hold « » to execute "External Adjustment". The required
(predefined) adjustment weight value flashes in the display.
d) Place adjustment weight in center of pan. The balance adjusts
itself automatically.
e) When "0.00 g" flashes, remove adjustment weight.
The adjusting is finished when the message "ADJ.DONE" appears briefly on the display. The balance returns to the last active application and is ready for operation.
Sample adjustment printout using external weight:
- External Adjustment --
21.Jan 2009 12:56
METTLER TOLEDO
Balance Type ML4002 SNR 1234567890
Temperature 22.5 °C Nominal 2000.00 g Actual 1999.99 g Diff 5 ppm
Adjustment done
Signature
.......................
-----------------------
21Setting up the Balance

5 Weighing Made Simple

On
Off
8888888
XXXXXXX
0.00
g
SHUTOFF
On
Off
This section shows you how to perform simple weighings and how you can accelerate the weighing process.

5.1 Switching the Balance On and Off

This section shows you how to perform simple weighings and how you can accelerate the weighing process.
Switching On
a) Remove any load from weighing pan.
b) Press «On».
The balance performs a display test (all segments in the display light up briefly), "WELCOME", Software version, Maximum load and Read ability appears briefly. (Startup "FULL" mode only)
The balance is ready for weighing or for operation with the last active application.
Switching Off
Press and hold the «Off» key until "SHUTOFF" appears on the display. Release the key.
When Quickstart is selected (Advanced menu, topic "STARTUP" > "QUICK"): Once your bal ance has been switched off, it is in standby mode. In this case your balance needs no warm­up time in the standby mode and is immediately ready for weighing. If you wish to perform a weighing, you now only need to place the sample on the weighing pan and the balance
immediately displays the result. There is no need to switch it on with the «On/Off» key.
If your balance has been switched off after a preselected time, the display is dimly lit and shows date, time, maximum load and readability.
If your balance has been switched off manually, the display is off.
Note:
Quickstart is not possible with approved balances (only available in selected countries).
Standby mode is available on line powered balances only.
Weighing Made Simple
22

5.2 Performing a Simple Weighing

1250.38
g
0.00
g
1182.03
g
C
#
0.00
g
121.01
g
#
0.01
g
0.00
g
C
C
121.01
g
0.00
g
0.00
g
95.97
g
121.01
g
-
N
et
Net
Net
#
C
C
0.78
oz
22.00
g
*
22.00
g
M
a) Press « » to zero the balance.
Note: If your balance is not in the weighing mode, first press and hold the « » key until “WEIGH“ appears in the display. Release the key. Your balance is in the weighing mode.
b) Place weighing sample on the weighing pan.

5.3 Zero Setting / Taring

c) Wait until the instability detector "
" disappears and the stability
beep sounds.
d) Read the result.
Zero setting
a) Unload the balance.
b) Press « » to set the balance to zero. All weight values are
measured in relation to this zero point (see menu topic "ZERO.RNG").
Note: Use the «
» zeroing key before you start with a weigh
ing.
Taring
If you are working with a weighing container, first set the balance to zero.
a) Place empty container on the balance. The weight is displayed.
b) Press « » to tare the balance.

5.4 Switching Weight Units

"0.00 g" and "Net" appears in the display. "Net" indicates that all weight values displayed are net values.
Note:
If the container is removed from the balance, the tare weight will be shown as a negative value.
The tare weight remains stored until the « » key is pressed again or the balance is switched off.
The « » key can be used at any time to toggle between weight unit "UNIT 1","RECALL" value (if selected), weight unit "UNIT 2" (if different from weight unit 2) and the application unit (if any).
Weighing Made Simple
23

5.5 Recall / Recall Weight Value

121.01
g
0.00
g
0.00
g
M
121.01
g
0.00
g
C
C
0%
100%
Recall stores stable weights with an absolute display value bigger than 10d. Requirement: The function "RECALL" must be activated in the menu.
a) Load weighing sample. The display shows weight value and
stores stable value.
b) Remove weighing sample. When the weight is removed the Dis
play shows zero.
c) Press «
5 seconds together with asterisk (*) and Memory (M) symbols. After 5 seconds the display goes back to zero. This can be repeat ed unlimited times.
Delete last weight value
As soon a new stable weight value is displayed, the old recall value becomes replaced by the new weight value. When pressing « », the recall value is set to 0.
Note: If the power is switched off, the recall value is lost. The recall value can not be printed.

5.6 Weighing with the Weighing-in Aid

The weighing-in aid is a dynamic graphic indicator which shows the used amount of the total weighing range. You can thus recognize at a glance whether the load on the balance approaches the maximum load.

5.7 Print / Transmit Data

Pressing the « » key transmits the weighing results over the interface e.g. to a printer or a PC.
». The display shows last stored stable weight value for
Weighing Made Simple
24

6 The Menu

6.1 What is in the Menu ?

The Menu allows you to match your balance to your specific weighing needs. In the menu you can change the settings of your balance and activate functions. The main menu has 4 different menus and these contains 35 different topics, each of which allows you various selection possibilities. For Menu "PROTECT" see chapter "Description of menu topics" section "Main menu".
Note: See Quick Guide for the graphical overview of the menu (Menu Map) with all setting possibilities.
Menu "BASIC"
Topic Description
DATE Setting the current date. TIME Setting the current time. UNIT 1 Specification of the 1st weight unit in which the balance should show the result. UNIT 2 Specification of the 2nd weight unit in which the balance should show the result. KEY.BEEP Setting the key beep level. STB.BEEP Setting the stability beep level. RESET Call up of the factory settings.
Menu "ADVANCE."
Topic Description
ENVIRON. Matching the balance to the ambient conditions. CAL Settings for the type of adjustment (calibration). DATE.FRM Setting the date format. TIME.FRM Preselection of the time format. RECALL Switching the application "Recall" for storing stable weights on or off. STARTUP Setting the mode which the balance powers up ("FULL" or "QUICK"). SHUTOFF Setting the time after which the balance should be switched off automatically. B.LIGHT Setting the time after which the display backlight should be switched off automatically. A.ZERO Switching the automatic zero correction (Autozero) on or off. ZERO.RNG Setting the zero limit of the zero/tare key. ASSIGN:F Selection of assigned F key application and entering their parameter settings. DIAGNOS. Starting a diagnostic application. SRV.ICON Switching the service reminder (service icon) on or off. SRV.D.RST Reset service date and hours (service reminder)
Menu "INT.FACE"
Topic Description
RS232 Matching the serial interface RS232C to a peripheral unit. HEADER Setting the header for printout of individual values. SINGLE Setting the information for printout of individual values. SIGN.L Setting the footer for printout of individual values. LN.FEED Setting line feed for printout of individual values. ZERO.PRT Setting the auto print function for printing zero. COM.SET Setting the data communication format of the serial interface RS232C. BAUD Setting the transfer speed of the serial interface RS232C. BIT.PAR. Setting the character format (Bit/Parity) of the serial interface RS232C. STOPBIT Setting the character format (stop bit) of the serial interface RS232C. HD.SHK Setting the transfer protocol (Handshake) of the serial interface RS232C.
25The Menu
Topic Description
406.2
g
ADVANCE
BASIC
TIME
DATE
ENVIRON.
STABLE
ENVIRON.
01. 2500
01. 6500
01.6500
RS E.O.L. Setting the end of line format of the serial interface RS232C. RS.CHAR Setting the char set of the serial interface RS232C. INTERVL. Selection of the time interval for the simulated print key press.

6.2 Menu Operation

In this Section you will learn how to work with the menu.
Select Menu
a) Press and hold « » to activate main menu. The first menu
"BASIC" is displayed (except menu protection is active).
b) Press « » repeatedly to change menu (Scrolling down/up «+» /
«–» keys).
c) Press « » to confirm the selection.
Note: The menu selection "BASIC", "ADVANCE." or "INT.FACE" can not be saved. The selection "PROTECT" must be saved.
Select Menu Topic
Press « ». The next menu topic appears in the display. Each time the « » or the «+» key is pressed, the balance switches to the next menu topic; the «–» key to the previous menu topic.
Change Settings in a Selected Menu Topic
a) Press « ».The display shows the current setting in the selected
menu topic. Each time « » or «+» is pressed, the balance switches to the next selection; press «–» to the previous selection. After the last selection, the first is shown again.
b) Press « », the selected setting is accepted but not yet executed.
The settings are executed only after "SAVE:YES" has been con firmed.
Change Settings in a Submenu Selection
The same procedure as for menu topics.
Input Principle of Numerical Values
a) Press « » for input of numerical values.
b) Press « » to select a digit or a value (depending on the appli
cation). The selected digit or the selected value is blinking.
c) For changing digits or values, press «+» to scroll up or «–» to
scroll down.
d) Press « » to confirm the input.
26
The Menu
SAVE:YES
SAVE:NO
Saving Settings and Closing the Menu
a) Press and hold « » to leave menu topic.
b) Press « » to execute "SAVE:YES". Changes are saved.
c) Press «
gle between "SAVE:YES" and "SAVE:NO" press « ».
Cancel
For leaving menu topic or menu selection without saving press «C» (one step back in the menu).
Note: If no entry is made within 30 seconds, the balance reverts to last active application mode. Changes are not saved. If changes are made, the balance asks "SAVE:NO".
» to execute "SAVE:NO". Changes are not saved. To tog

6.3 Description of Menu Topics

In this Section you will find information regarding the individual menu topics and the available selections.

6.3.1 Main Menu

Selecting the menu.
"BASIC" The small "BASIC" menu for simple weighing is displayed.
"ADVANCE." The extended "ADVANCE." menu for further weighing settings is
displayed.
"INT.FACE" The menu "INT.FACE" for all interface parameter settings for
peripheral devices e.g. printer is displayed.
"PROTECT" Menu protection. Protection of balance configurations against
unmeant manipulation.
"OFF" Menu protection is off. (Factory setting)
"ON" Menu protection is on. The menu BASIC, ADVANCE. and INT.FACE
are not displayed. This is indicated with "
Note:
The menu selection "BASIC", "ADVANCE." or "INT.FACE" can not be saved.
" in the display.
To activate "PROTECT" "ON" or "OFF", this selection must be saved.

6.3.2 Basic Menu

"DATE" – Date
Setting the current date according to date format.
Note: A reset of the balance will not change this setting.
"TIME" – Time
Setting the current time according to time format
"+1H" Set the current time forwards by 1 hour (to adjust summer or
winter time). (Factory setting)
"-1H" Set the current time backwards by 1 hour (to adjust summer or
winter time).
"SET.TIME" Enter the current time.
The Menu
27
Note: A reset of the balance will not change this setting.
"UNIT 1" – Weight Unit 1
Depending on requirements, the balance can operate with the following units (depending on the model)
Only those weight units allowed by the appropriate national legislation are selectable.
With approved balances, this menu topic has a fixed setting and cannot be changed.
Conversion table for weight units see chapter Appendix.
Units:
1)
g
kg
mg
2)
3)
Gram dwt Pennyweight
Kilogram mom Momme
Milligram msg Mesghal
ct Carat tlh Tael Hong Kong
lb Pound tls
4)
Tael Singapore
oz Ounce (avdp) tlt Tael Taiwan
ozt Ounce (troy) tola Tola
GN Grain baht Baht
1)
factory setting
2)
not with 0.1 mg and 1 mg balances
3)
with 0.1 mg and 1 mg balances
4)
the Malaysian tael has the same value
"UNIT 2" – Weight Unit 2
If it is required to show the weighing results in weighing mode in an additional unit, the desired second weight unit can be selected in this menu topic (depending on the model). Units see "UNIT 1".
Note: Only those weight units allowed by the appropriate national legislation are selectable.
"KEY.BEEP" – Key Beep
This menu topic allows you to select the volume of the key beep. The according key beep is emitted during the setting.
"MED" Medium level (Factory setting)
"HIGH" High level
"OFF" Beep switched off
"LOW" Low level
"STB.BEEP" – Stability Beep
If the unstable symbol disappears, the stability beep becomes active. This menu topic allows you to preselect the volume of the stability beep.
"LOW" Low level (Factory setting)
"MED" Medium level
"HIGH" High level
"OFF" Beep switched off
28
The Menu
"RESET" – Reset Balance Settings
This menu topic allows you to cal-up the factory settings.
To toggle between "YES?" and "NO?" press « » (or «+» or «–»).
Note: A reset of the balance will not change the "DATE" and "TIME" settings.

6.3.3 Advanced Menu

"ENVIRON." – Environment Settings
This setting can be used to match your balance to the ambient conditions.
"STD." Setting for an average working environment subject to moderate
variations in the ambient conditions. (Factory setting)
"UNSTAB." Setting for a working environment where the conditions are con
tinuously changing.
"STABLE" Setting for a working environment which is practically free from
drafts and vibrations.
"CAL" – Adjustment (calibration)
In this menu topic you can preselect the function of the « external weights by pressing the « » key. If you have attached a printer to your balance, the data of the adjust ment (calibration) are printed out.
» key. Your balance can be adjusted with internal or
"ADJ.OFF" The adjustment is switched off. The « » key has no function.
"ADJ.INT" Internal adjustment: adjustment is performed at a keystroke with
the built-in weight (depending on the model, see technical data).
"ADJ.EXT" External adjustment: adjustment is performed at a keystroke with
a selectable external weight. Note:This function is not available for approved balances *
(depend on selected countries' certification legislation). * except OIML accuracy class I approved models.
"200.00 g" Defining the external adjustment weight: define the weight of
the external adjustment weight (in grams).
Factory setting: depends on the model.
"DATE.FRM" – Date Format
This menu topic allows you to preselect the date format.
The following date formats are available:
Display examples Printing examples
"DD.MM.Y" 01.02.09 01.02.2009
"MM/DD/Y" 02/01/ 02/01/2009
"Y-MM-DD" -02-01 2009-02-01
"D.MMM Y" 1.FEB.09 1.FEB 2009
"MMM D Y" FEB.1.09 FEB 1 2009
Factory setting: "DD.MM.Y"
29The Menu
"TIME.FRM" – Time Format
This menu topic allows you to preselect the time format.
The following date formats are available:
Display examples
"24:MM" 15:04
"12:MM" 3:04 PM
"24.MM" 15.04
"12.MM" 3.04 PM
Factory setting: "24:MM"
"RECALL" – Recall
This menu topic allows you to switch the "RECALL" function on or off. When it is switched on recall stores the last stable weight if the absolute display value was bigger than 10d.
"OFF"
"ON"
Note: The recall value is displayed with an asterisk and cannot be printed.
"STARTUP" – Startup Mode
You can set your balance such that it either immediately starts from the standby mode when you load a weight or it must be switched on with the «ON/OFF» key after which it then performs a display test.
Note: This topic is not visible with approved balances (only available in selected countries).
"QUICK" "Quickstart": The balance can be started directly from the stand
"FULL" Start with display test: You must switch on the balance with the
"SHUTOFF" – Automatic Shutoff
If the automatic shutoff function is activated, the balance automatically switches itself off after a preselected time of inactivity (i.e. with no key being pressed or changes of weight occurring etc.) and is switched to the standby mode.
"RECALL" switched off (Factory setting)
"RECALL" switched on
by mode and is immediately ready for weighing. You can load the weight in the standby mode and the balance immediately
shows the current weighing result. This is the Factory setting Note: Standby mode is available on line powered balances only.
«ON/OFF» key. After it has been switched on, it performs a dis play test for approx. 2 sec. in which all display elements lights up, it shows "WELCOME", software version, maximum load and readability. The balance is ready for weighing.
30
"A.OFF:10' " Automatic shutoff after 10 minutes of inactivity. (Factory setting)
"A.OFF – " Automatic shutoff not activated.
"A.OFF:2' " Automatic shutoff after 2 minutes of inactivity.
"A.OFF:5' " Automatic shutoff after 5 minutes of inactivity.
"B.LIGHT" – Backlight
Under this menu topic, the display backlight can be switched off automatically. If the automatic switch-off is activated, the backlight will turn off automatically after the selected period of inactivity has elapsed. The back light is reactivated when a key is pressed or the weight is changed.
"B.L. ON" Backlight is always on. (Factory setting)
The Menu
"B.L. OFF" Backlight is always off.
"B.L. 30" " Automatic switch-off after 30 seconds inactivity.
"B.L. 1' " Automatic switch-off after 1 minute inactivity.
"B.L. 2' " Automatic switch-off after 2 minutes inactivity.
"B.L. 5' " Automatic switch-off after 5 minutes inactivity.
"A.ZERO" – Automatic Zero Setting
This menu topic allows you to switch the automatic zero setting on or off.
“ON“ "A.ZERO" switched on (factory setting). The automatic zero set
ting continuously corrects possible variations in the zero point that might be caused through small amounts of contamination on the weighing pan.
"OFF" "A.ZERO" switched off. The zero point is not automatically cor
rected. This setting is advantageous for special applications (e.g. evaporation measurements).
Note: With approved balances, this setting is not available (only available in selected countries).
"ZERO.RNG" – Zero Range
This menu topic allows you to set a zero limit for the « « » key will execute a zero. Above this limit the « » key will execute a tare.
"21g" To set the upper limit of the zeroing range as weight in the defini
tion unit of the balance. (Factory setting: 0.5% of weighing range)
» key. Up to and including this limit the
Note: With approved balances, this setting is not available and fixed to 3e (only available in selected countries).
Note: A reset of the balance will not change this setting.
"ASSIGN:F" – Assign Application Key F
At this menu topic you can assign an application to the «F» key. The following applications are available (depending on the model):
"COUNT" Piece counting (Factory setting)
"PERCENT" Percent weighing
"CHECK" Checkweighing
"STAT" Statistics
"FORMULA" Formulation / Net-Total
"TOTAL" Totaling
"DYNAMIC" Dynamic weighing
"FACTOR.M" Multiplication factor
"FACTOR.D" Division factor
"DENSITY" Density
"R.TEST" Routine test
"DIAGNOS." – Diagnostics Application
At this menu topic you can start a diagnostic application. For more information see chapter application "Diag nostics". The following diagnostics are available:
The Menu
31
"REPEAT.T" Repeatability test (models with internal weights only)
"DISPLAY" Display test
"KEYPAD.T" Key test
"CAL.MOT.T" Motor test (models with internal weights only)
"BAL.HIST" Balance history
"CAL.HIST" Calibration history
"BAL.INFO" Balance information
"SRV.PROV" Service provider information
"SRV.ICON" – Service Reminder
This menu topic allows you to switch the service reminder "
" on or off.
"ON" Service reminder "
" switched on.You will be informed after one
Year or 8000 operating hours to call service for recalibration. This will be indicated by the flashing service icon: " ". (Factory
setting)
"OFF" Service reminder " " switched off.
"SRV.D.RST" – Service Date Reset
This menu topic allows you to reset service date and hours. Note: This menu topic is only available if “SRV.ICON” setting “ON” was selected.
To toggle between “YES?” and “NO?” press « » (or «+» or «-»)

6.3.4 Interface Menu

"RS232" – RS232C Interface
At this menu topic you can select the peripheral device connected to the RS232C interface and specify how the data is transmitted.
"PRINTER" Connection to a printer. (Factory setting)
"PC-DIR." Connection to a PC: the balance can send data (as a Keyboard)
1)
"PRT.STAB" If the «
» key is pressed, the next stable weight value will be
printed. (Factory setting)
"PRT.AUTO" Every stable weight value will be printed, without pressing the
« » key.
"PRT.ALL" If the « » key is pressed, the weight value will be printed
regardless of stability.
to the PC used for PC applications e.g. Excel. Note: The balance sends the weight value without the unit to the PC.
"PRT.STAB" If the «
» key is pressed, the next stable weight value will be
sent followed by an enter. (Factory setting)
"PRT.AUTO" Every stable weight value will be sent followed by an enter, with
out pressing the « » key.
"PRT.ALL" If the « » key is pressed, the weight value will be sent followed
by an enter regardless of stability.
32 The Menu
"HOST" Connection to a PC, Barcode Reader etc.: the balance can send
data to the PC and receive commands or data from the PC. Note: The balance sends the complete MT-SICS answer to the PC (see chapter "MT-SICS Interface Commands and Functions".
"SND.OFF" Send mode switched off.(Factory setting)
"SND.STB" If the « » key is pressed, the next stable weight value will be
sent.
"SND.CONT" All weight value updates will be sent regardless of stability, with
out pressing the «
"SND.AUTO" Every stable weight value will be sent, without pressing the « »
key.
"SND.ALL" If the « » key is pressed, the weight value will be sent regard
less of stability.
"2.DISP" Connection of an optional auxiliary display unit.
Note: The transmission parameters cannot be selected. Settings are automatically set.
Attention: If you select 2nd Display "2.DISP", first make sure that no other device is connected at COM1 as an auxiliary display. Other devices could be damaged because of the voltage on con nector Pin 9. Necessary for powering the display (see Chapter "Interface Specification").
» key.
"HEADER" – Options for the Printout Header of individual values
This menu topic allows you to specify the information that is to be printed at the top of the printout for every individual weighing results (after pressing « »). Note: This menu topic is only available if “PRINTER” setting was selected.
"NO" The header is not be printed (Factory setting)
"DAT/TIM" Date and time are printed
"D/T/BAL" Date, time and balance information (Balance type, SNR, Balance
ID) are printed.
Note: Balance ID only if set.
"SINGLE" – Options for Printing out the Result of individual values
This menu topic allows you to specify the information that is to be printed for every individual weighing result (after pressing « »). Note: This menu topic is only available if “PRINTER” setting was selected.
"NET" The value of the Net weight from the current weighing is printed
(Factory setting)
"G/T/N" The values of the Gross weight, the Tare weight and the Net
weight are printed
"SIGN.L" – Options for the Printout Footer for Signature Line of individual values
This menu topic allows you to set a footer for signature at the bottom of the printout for every individual weigh ing result (after pressing « »). Note: This menu topic is only available if “PRINTER” setting was selected.
"OFF" The signature footer is not be printed. (Factory setting)
"ON" The signature footer is printed
33The Menu
"LN.FEED" – Options for Complete the Printout of individual values
This menu topic allows you to specify the number of blank lines to complete the printout (line feed) for every individual weighing result (after pressing « ») . Note: This menu topic is only available if “PRINTER” setting was selected.
"0" Possible numbers of blank lines: 0 to 99 (Factory setting = 0)
"ZERO.PRT" – Options for “PRT.AUTO”
1)
This menu topic allows you to specify the auto print function “PRT.AUTO” for printing zero "YES" or "NO".
"OFF" Zero is not be printed (Zero +/- 3d) (Factory setting)
"ON" Zero is always printed
Note: This menu topic is only available if "PRT.AUTO" fuction of the "PRINTER" or "PC-DIR." was selected.
"COM.SET" – Options for the Data Communication Format (RS232C)(“HOST”)
This menu topic allows you to set the data format depending on which peripheral device is connected. Note: This menu topic is only available if “HOST” setting was selected.
"MT-SICS" The MT-SICS data transfer formats is used. (Factory setting)
For more information see section “MT-SICS Interface Commands and Functions”.
"MT-PM" The following PM balance commands are supported:
S Send value
SI Send immediate value
SIR Send value and repeat
SR Send next value and repeat
SNR Send immediate value and repeat
T Tare
TI Tare immediately
B Base (Negative values are limited up to the current tare
values)
MI Modify ambient vibration
MZ Modify Auto Zero
M Modified settings reset
ID Identify
CA Calibrate
D Display (only symbol N and G available)
34
"SART" The following Sartorius commands are supported:
K Ambient conditions: very stable
L Ambient conditions: stable
M Ambient conditions: unstable
N Ambient conditions: very unstable
O Block keys
P Print key (print, auto print; activate or block)
Q Acoustic signal
R Unblock keys
S Restart/self-test
T Tare key
The Menu
W Calibration/adjustment (depending on the menu setting)
*)
Z Internal calibration/adjustment
**)
f0_ Function key (F)
f1_ Function key (CAL)
s3_ C key
x0_ Perform internal calibration
**)
x1_ Print balance/scale model
x2_ Print weighing cell serial number
x3_ Print software version
*)
may be inaccessible on verified balances/scales
**)
only on models with built-in motorized calibration weight
Functionality mapping
"HOST" settings: Sartorius printer settings:
"SND.OFF" not applicable
"SND.STB" manually print with stability
"SND.ALL" manually print without stability
"SND.CONT" automatically print without stability
"SND.AUTO" similar applicable to automatically print
when load is changed
"BAUD" – Baud rate RS232C
1)
This menu topic allows you to match the data transmission to different serial RS232C receivers. The baud rate (data transfer rate) determines the speed of transmission via the serial interface. For problem-free data trans mission the sending and receiving devices must be set at the same value.
The following settings are available: 600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd, 19200 and 38400 bd. (default: 9600 bd)
Note:
Not visible for 2nd display.
Each device has separate settings.
"BIT.PAR." – Bit/Parity RS232C
1)
At this menu topic you can set the character format for the attached RS232C serial peripheral device.
"8/NO" 8 data bits/no parity (Factory setting)
"7/NO" 7 data bits/no parity
"7/MARK" 7 data bits/mark parity
"7/SPACE" 7 data bits/space parity
"7/EVEN" 7 data bits/even parity
"7/ODD" 7 data bits/odd parity
Note:
Not visible for 2nd display.
Each device has separate settings.
The Menu
35
"STOPBIT" – Stop Bits RS232C
1)
At this menu topic you can set the stop bits of the transmitted data to different RS232C serial receivers.
"1 BIT" 1 Stop bit (Factory setting)
"2 BITS" 2 Stop bits
Note:
Not visible for 2nd display.
Each device has separate settings.
"HD.SHK" – Handshake RS232C
1)
This menu topic allows you to match the data transmission to different RS232C serial receivers.
"XON.XOFF" Software handshake (XON/XOFF) (Factory setting)
"RTS.CTS" Hardware handshake (RTS/CTS)
"OFF" No handshake
Note:
Not visible for 2nd display.
Each device has separate settings.
"RS E.O.L." – End of Line RS232C
1)
At this menu topic you can set the "End of Line" character of the transmitted data to different RS232C serial receivers.
"CR LF" <CR><LF> Carriage Return followed by Line feed (ASCII-Codes
013+010) (Factory setting)
"CR" <CR> Carriage Return (ASCII-Code 013)
"LF" <LF> Line feed (ASCII-Code 010)
Note:
Not visible for 2nd display.
Each device has separate settings.
"RS.CHAR" – Char Set RS232C
1)
At this menu topic you can set the "Character Set" of the transmitted data to different RS232C serial receivers.
"IBM.DOS" Char Set IBM/DOS (Factory setting)
"ANSI.WIN" Char Set ANSI/WINDOWS
Note:
Not visible for 2nd display.
Each device has separate settings.
"INTERVL." – Print Key Simulation
At this menu topic you can activate a simulation of the «
» key. "INTERVL." simulates a print key press every
x seconds.
Range: 0 to 65535 seconds
0 sec: disables the print key simulation
Factory setting: 0 sec
Note: The executed action is according to the configuration of the print key. (see interface setting)
36 The Menu
———————————————————————————
1) Note for 2nd RS232C Interface
If an optional 2nd interface is installed, the menu topic is dis played for each interface, e.g "BAUD.1" for standard interface "BAUD.2" for optional 2nd interface
Only one printer can be set if two RS232 interfaces are existing.
———————————————————————————
The Menu
37

7 Application "Piece Counting"

COUNT
D
A
B C
A
B
C
D
FIX 10
PCS
C
10PCS
------
FIX 10
PCS
A
A
C
------
VAR.REF
25PCS
B
15
PCS
REF.
25
PCS
REF.
B
The "Piece Counting" application allows you to determine the number of pieces put on the weighing pan. Requirement: The function "COUNT" must be assigned to the «F» key (see advanced menu topic "ASSIGN:F", factory setting).
Activate function "COUNT" by pressing and holding the «F» key.
Piece Counting first requires the setting of a reference weight, there are 4 possibilities:
Setting the reference by multiple pieces with fix reference values. Setting the reference by multiple pieces with variable reference
values.
Setting the reference for 1 piece in weighing mode. Setting the reference for 1 piece in manual mode.
Setting possibility
a) Select a number of reference pieces by scrolling with « ». Pos
Setting the reference by multiple pieces
sible numbers* are 5, 10, 20 and 50. * with approved balances in selected countries: min 10
b) Press « » to tare. If needed: place empty container on the
weighing pan and tare again.
c) Add the selected number of reference pieces to container.
d) Press «
Setting possibility
Setting the reference by multiple pieces with variable refer ence values
a) Select "VAR.REF" by scrolling with « ». Press « » to confirm.
b) Select a number of reference pieces by scrolling up («+» key) or
down («-» key). Speed up by press and hold. Possible numbers* are 1 to 999. * with approved balances in selected countries: min 10
c) Press « » to tare. If using: place empty container on the
weighing pan first or tare again.
d) Add the selected number of reference pieces to container.
e) Press «
» to confirm.
» to confirm.
38 Application "Piece Counting"
0.00
g
1
PCS
2.74
g
1
C
------
PCS.WGT
C
Setting possibility
C
2.68
g
*
0
PCS
2.74
g
*
PCS.WGT
D
D
2.74
g
2.74
g
,
812.7
g
75
PCS
Setting the reference for one piece in weighing mode
a) Select "PCS.WGT" by scrolling with « ».
b) Press « » to tare. If needed: place empty container on the
weighing pan and tare again.
c) Add one reference piece to container. The weight of one piece is
displayed.
d) Press «
Note: With approved balances, this setting is not available in selected countries.
Setting possibility
Setting the reference for one piece in manual mode
a) Select "PCS.WGT" by scrolling with « ».
b) Press « » to confirm.
c) Enter the final reference one piece weight by scrolling up («+» key)
or down («-» key). Speed up by press and hold.
d) Press «
Note: With approved balances, this setting is not available in selected countries.
Switching between manual mode and weighing mode
Press « » to switch between manual and weighing mode.
» to confirm.
» to confirm.
Note: By switching from weighing mode to manual mode the weight value will be transferred and can be changed manually.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press «C» to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for piece counting.
Switching between piece count and weight display.
You can use the « » key at any time to switch the display between piece display, weighing unit "UNIT 1", "RECALL" value (if activated) and weighing unit "UNIT 2" (if different from "UNIT 1").
Note:
The "RECALL" value is displayed with an asterisk (*) and icon "M" and can not be printed.
Take into account minimum values: min. reference weight = 10d (10 digits), min. piece weight* = 1d (1 digit)! * with approved balances in selected countries: min 3e
The current reference weight remains stored until the reference setting is changed.
Application "Piece Counting"
39

8 Application "Percent Weighing"

PERCENT
A
B
A
B
0.0%
175.28
g
*
A
SET 100%
A
I00.0
%
175.28
g
SET 100%
C
0.00
g
B
B
175.28
g
175.28
g
,
The "Percent Weighing" application allows you to check a sample weight as percentage to a reference target weight. Requirement: The function "PERCENT" must be assigned to the «F» key (see advanced menu topic "ASSIGN:F".
Activate function percent weighing "PERCENT" by pressing and holding the «F» key.
Percent Weighing first requires the setting of a reference weight that should corresponds to 100%, there are 2 possibilities:
Setting the reference in manual mode (enter 100%).
Setting the reference in weighing mode (weigh 100%).
Setting possibility
Setting the reference by manual mode (enter 100%)
a) Press « » to activate manual mode.
b) Select the reference target weight (100%) by scrolling up («+»
key) or down («–» key). Speed up by press and hold.
c) Press « » to confirm.
Setting possibility
Setting the reference by weighing mode (weigh 100%)
a) Press « » to tare the balance and to activate the weighing
mode. If needed: place empty container on the weighing pan and tare again.
b) Load the reference weight (100%).
Note: Reference weight must be at least +/- 10d.
c) Press « » to confirm.
Switching between manual mode and weighing mode
Press « » to switch between manual and weighing mode.
Note: By switching from weighing mode to manual mode the weight value will be transferred and can be changed manually.
Note: If without any key press within 60 seconds, the balance returns to the previous active application.
On completion of the weighing-in procedure, your balance is ready for percent weighing.
40 Application "Percent Weighing"
,
128.25
g
73.17
%
128250
mg
128.25
g
M
Switching between percent and weight display
You can use the « » key at any time to switch the display between percent display, weighing unit "UNIT 1", "RECALL" value (if activated) and weighing unit "UNIT 2" (if different from UNIT 1).
Note:
The recall value is displayed with an asterisk (*) as well as icon "M" and can not be printed.
The current set weight remains stored until it is redetermined.
Application "Percent Weighing"
41

9 Application "Check Weighing"

CHECK
1A
3
1B
2A
2B
1A
1B
2A
2B
3
SET NOM .
10.00
g
20.28
g
1A
1A
1B
SET NOM .
0.00
g
20.28
g
C
1B
10.00
g
10.00
g
The "Check weighing" application allows you to check the deviation of a sample weight within a tolerance limit to a reference target weight. Requirement: The function "CHECK" must be assigned to the «F» key (see advanced menu topic "ASSIGN:F".
Activate function "CHECK" by pressing and holding the «F» key.
Step 1: Check Weighing first requires the setting of a reference weight that should corresponds to the nominal weight, there are 2 possibilities:
Setting the reference in manual mode (enter nominal weight).
Setting the reference in weighing mode (weigh nominal weight).
Step 2: Check weighing needs the upper and lower limits, there are 2 possibilities::
Setting the upper and lower limits in percentage.
Setting the upper and lower limits by weight.
Step 3: Setting tolerance beep
Activate or deactivate tolerance beep.
Setting possibility:
Setting the reference in manual mode (enter nominal weight)
a) Press « » to activate manual mode.
b) Select the reference target weight by scrolling up («+» key) or
down («–» key). Speed up by press and hold.
c) Press «
Setting possibility:
Setting the reference in weighing mode (weigh nominal weight)
a) Press « » to tare the balance and to activate the weighing
mode. If needed: place empty container on the weighing pan and tare again.
b) Load the nominal weight.
c) Press « » to confirm the nominal weight.
Switching between manual mode and weighing mode
Press « » to switch between manual mode and weighing mode.
» to confirm the nominal weight.
42 Application "Check Weighing"
Note: By switching from weighing mode to manual mode the weight value will be transferred and can be changed manually.
SET.T+
/- %
2.50
%
20.28
g
2A
Step 2, setting possibility:
2A
SET T+/- %
2B
2.50
%
20.28
g
0.25
g
0.36
g
2B
0.25
g
2.50
%
20.28
g
3
T.BEEP: Y
3
Setting the upper and lower limits (in percentage):
a) Press « » to start setting.
b) Press « » to confirm the default limit of +/- 2.5 % or enter the
limit value by scrolling up («+» key) or down («–» key). Press « » to confirm the limits.
Note: Press «
Step 2, setting possibility:
Setting the upper and lower limits by weight:
a) Press «
b) Press « » to switch to UNIT 1.
c) Press « » to confirm the default limit or enter the limit value by
scrolling up («+» key) or down («–» key). Press « » to confirm the limits.
Switching between percentage and weight unit 1
Press « » to switch between setting in percentage and setting by weight .
Step 3:
Setting tolerance beep:
» to switch between "UNIT 1" and Unit "%".
» to start setting.
Note:
On completion of the setting procedure, your balance is ready for checkweighing.
The tolerance beep indicates whether the weighing sample lies within the tolerance by beeping three times.
Note: The beep level corresponds to the setting in menu topic "STB.BEEP" (Basic menu). If "STB.BEEP" is set to "OFF", the tolerance beep level is medium.
To activate tolerance beep press « ». To deactivate tolerance beep press « » to select "N" and press « ».
If without any key press within 60 seconds, the balance returns to the previous active application. Press «C» to cancel and returns to the previous active application.
The nominal weight must be at least 10 digit.
43Application "Check Weighing"
20.15
g
Weighing-in-Aid
The Weighing-in-Aid helps you quickly determine the position of the sample weight regarding the tolerance.
1 Lower limit 2 Target weight 3 Upper limit
44 Application "Check Weighing"

10 Application "Statistics"

CLR. M:NO
STAT
CLEARED
46.36
g
0.00
g
- 1 -
C
999
C
0.00
g
5
N
X
50.530
g
S.DEV
3.961
g
S.REL
7.84
%
MIN
46.36
g
MAX
55.81
g
DIFF
9.45
g
The "Statistics" application allows you to generate statistics of weighing values. 1 to 999 values are possible. Requirement: The function "STAT." must be assigned to the «F» key (see advanced menu topic "ASSIGN:F"). Connect a printer or a PC if present.
a) Activate function "STAT." by pressing and holding the «F» key.
b) To continue the last statistics press « ». For a new statistical
evaluation press « » (or «+», «–») to select "CLR.M:YES" and press « » to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the memory clear question will be not displayed.
Weighing the first sample weight:
a) Press « » to zero/tare the balance if needed.
b) Load the first sample weight.
c) Press « ». The display shows the sample count "- 1 -" and the
current weight is stored as sample and the weight is printed out. Note: When the sample counter is displayed you may press «C» to undo (drop) this sample.
Displayed results:
a) Press « » to show the next
statistical value.
b) Press «C» to cancel displaying
results and to continue weigh ing next sample.
d) Unload the first sample weight.
Weighing further sample weights:
The same procedure as for the first sample weight.
1...999 samples are possible.
The next value will be accepted if the sample weight is in the range 70% –130% of the current average value. "OUT OF RANGE" will be displayed if the sample is not accepted.
Results:
If the numbers of sample are greater than or equal to 2, press « », the results are displayed and printed.
0.5 seconds
number of samples
average
standard deviation
relative standard deviation
lowest value (minimum)
highest value (maximum)
different between the minimum
and the maximum
45Application "Statistics"
SUM
252.65
g
Printout:
------ Statistics ------
21.Jan 2009 12:56
METTLER TOLEDO
Balance Type ML4002 SNR 1234567890
-----------------------­1 46.36 g 2 55.81 g 3 47.49 g 4 53.28 g 5 49.71 g n 5 x 50.530 g s dev 3.961 g s rel 7.84 g Min. 46.36 g Max. 55.81 g Diff 9.45 g Sum 252.65 g
-----------------------
sum of all values
46 Application "Statistics"

11 Application "Formulation" (Net Total Formulation)

CLR.
M:NO
FORMULA
CLEARED
#
-T1-
0.00
g
Net
10.33
g
C
0.00
g
#
Net
Net
C
The "Formulation" (Net Total) application allows you to
weigh in (add and store) up to 999 individual component weights and displays the total. If a printer is connected, the component weights are printed individually and as a total.
tare/pre-tare and store up to 999 container weights and displays the total. If a printer is connected, the tare weights are printed out individually and as a total.
Requirement: The function "FORMULA." must be assigned to the «F» key (see advanced menu topic "ASSIGN:F"). Connect a printer or a PC if present.
a) Activate function formulation "FORMULA." by pressing and holding
the «F» key.
b) Press « » to continue formulation weighing. For a new formula
tion press « » (or «+» or «-») to select "CLR.M:YES" and press « » to clear the memory. Note: If the memory is already cleared (sample and tare/pre-tare counter is zero) the memory clear question will be not displayed.
Tare container (if used):
a) Press « » to zero or tare the balance if needed.
b) Place the empty container on the weighing pan.
c) Press « » . The container is tared and the tare count "- T1
-" is displayed and the tare weight is printed.
Note:
If you pre-tare via MT-SICS (e.g. bar code reader) "- PT1 -" is dis played.
Zero range setting (menu topic "ZERO.RNG") has no effect. The zero-limit is less than or equal 10d.
Weighing the first component weight:
a) Load the first component weight.
b) Press « ». The display briefly shows the component count "- 1 -
" , the current weight is stored as sample and the component weight is printed. The display is set back to zero.
Weighing further component weights:
The same procedure as for the first component weight with the same or new container).
1...999 sample values are possible.
max 999 tare values are possible.
max 999 pre-tare values are possible.
Results:
If the numbers of sample are greater than or equal to 2, press « », the results are displayed and printed.
Displayed results: 0.5 seconds
47Application "Formulation" (Net Total Formulation)
a) Press « » to show the next
8
N
T.TOTAL
452
.
76
g
G.TOTAL
546.79
g
N.TOTAL
94.03
g
------ Formulation -----
21.Jan 2009 12:56
METTLER TOLEDO
Balance Type ML4002 SNR 1234567890
-----------------------­1 T 10.33 g 1 N 8.85 g 2 N 9.23 g 2 T 10.84 g 3 N 7.43 g . . n 8 T Total 452.76 g G Total 546.79 g
N Total 94.03 g
-----------------------
statistical value.
b) Press «C» to cancel displaying
results and to continue weigh ing next component.
Printout:
number of samples
sum of all tare values (T and
PT)
sum of all component gross
weight values
sum of all component net
weight values
48 Application "Formulation" (Net Total Formulation)

12 Application "Totaling"

CLR.M:NO
TOTAL
CLEARED
0.00
g
46.36
g
- 1 -
999
0.00
g
0.00
g
C
C
879
N
TOTAL
8789
.
79
g
The "TOTALING" application allows you to weigh in different samples to add their weight val ues and to totalize them. 1 to 999 samples are possible. Requirement: The function "TOTAL" must be assigned to the «F» key (see advanced menu topic "ASSIGN:F").
a) Activate function "TOTAL" by pressing and holding the «F» key.
b) For a new totaling evaluation press « » (or «+» or «–») to enter
"CLR.M:YES" and press « » to clear the memory. Note: If the memory is already cleared (sample counter is 0) the memory clear question will be not displayed.
c) Press « » to zero or tare the balance.
Weighing in the sample weight:
a) If using a container: place empty container on the weighing pan
and press «
» to zero or tare the balance.
b) Load the first sample weight.
c) Press « ». The display shows the sample count "- 1 -" and the
current weight is stored. Note: When the sample counter is displayed you may press «C» to undo (drop) this sample.
Displayed results:
a) Press « » briefly to show the
totalized value.
b) Press «C» briefly to cancel.
d) Unload the first sample weight. The display shows zero.
Weighing in further sample weights:
The same procedure as for the first sample weight.
1...999 samples are possible.
Results:
If the numbers of sample are greater than or equal to 2, press « », the results are displayed and printed.
0.5 seconds
number of samples
totalized value
49Application "Totaling"
Printout:
------- Totaling -------
21.Jan 2009 12:56
METTLER TOLEDO
Balance Type ML1602 SNR 1234567890
-----------------------­1 46.36 g 2 55.81 g 3 47.49 g 4 53.28 g 5 49.71 g 6 53.93 g . . . n 879 Total 8789.79 g
-----------------------
50 Application "Totaling"

13 Application "Dynamic Weighing"

DYNAMIC
MOD.AUTO
MOD. MAN
1
TIME: 3
TIME: 5
2
The "Dynamic Weighing" application allows you to determine the weights of unstable sam ples or to determine weights under unstable ambient conditions. The balance calculates the weight as the average of a number of weighing operations over a defined time. Requirement: The function "DYNAMIC" must be assigned to the «F» key (see advanced menu topic "ASSIGN:F"). Note: "Switching Units" and "RECALL" Functions are not available in this Application.
Activate function "DYNAMIC" by pressing and holding the «F» key.
Setting "Auto Start" or "Manual Start":
a) Press « » to select the mode:
– "Auto Start ""MOD.AUTO" (default value). The weighing starts automatically on relative stability. However, the weighing sample must weigh at least 5 grams. For weighing samples below 5 g the weighing must be started manually. – "Manual Start" "MOD. MAN"
b) Press « » to confirm the selection.
Setting the weighing time:
a) Press « » to select one of the available time intervals: 3 (default
value), 5, 10, 20, 60 and 120 seconds.
b) Press « » to confirm the selected time interval.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press «C» to cancel and returns to the previous active application.
Your balance is now ready for dynamic weighing:
51Application "Dynamic Weighing"
13.45
g
M.START
TIME: 5
C
TIME: 1
13.39
g
C
a) Press « » to zero if needed.
b) If using a container: place empty container on weighing pan and
press « » to tare the balance.
c) Load sample weight.
d) – If you have selected function "Manual Start" "M.START", press
» to start the weighing.
« – If you have selected function "Auto Start" "A.START", the weigh ing starts automatically on relative stability. For weighing samples below 5 g the weighing must be started manually by pressing « ».
e) Read off result. The result of the dynamic weighing is displayed
with an asterisk (* = calculated value).
f) Unload sample weight.
g) "Manual Start" only, press « » to zero and go back to
"M.START".
Note:
The remaining weighing time (in seconds) is displayed continu ously. You can cancel the countdown by pressing «C».
The weight value remains in the display until the sample weight is removed from weighing pan ("Auto Start" only) or « » is pressed.
52 Application "Dynamic Weighing"

14 Application "Multiplication Factor Weighing"

FACTOR.M
SET.F.MUL
001.0000
009.8100
1
SET.STEP
001.0
005.0
2
The "Multiplication Factor Weighing" application allows you to multiply the weight value (in grams) by a predefined factor (result = factor * weight) and have it calculated to a predefined number of decimal places. Requirement: The function "FACTOR.M" must be assigned to the «F» key (see advanced menu topic "ASSIGN:F").
Activate function "FACTOR.M" by pressing and holding the «F» key.
Setting the factor value:
a) Press «
default value or the factor that was saved most recently.
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scroll up or «–» to scroll down.
d) Press « » to confirm the selected factor (no automatic accep
tance).
Note: Zero for multiplication factor value is outside the allowed range, the error message "FACTOR OUT OF RANGE" will be displayed.
Setting the step value:
"SET.STEP" appears in the display, and the program changes automat ically to allow the display increments to be entered. The smallest pos sible display increment appears as default value, or the last value that was saved.
a) Press « » to execute "SET.STEP".
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scroll up or «–» to scroll down.
d) Press «
tance).
» to execute "SET.F.MUL". Either the factor 1 appears as
» to confirm the selected step (no automatic accep
Note: The allowed range for the step depends on the factor and the resolution of the balance. If it is outside the allowed range the error message "STEP OUT OF RANGE" will be displayed.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press «C» to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for multiplication factor weighing.
Application "Multiplication Factor Weighing"
53
C
9810
1000.02
g
0
C
Weighing procedure
a) Press « » to zero/tare.
b) Load sample weight on weighing pan.
c) Read the result. The appropriate calculation is then made using
the weight of sample and the selected factor, the result being dis played with the selected display step. Note: No units are displayed.
d) Unload sample weight.
Toggling between displaying the calculated value and the measured weight:
You can use the «
» key to toggle between the calculated Value, weight value "UNIT 1", "RECALL" value (if selected) and weight value "UNIT 2" (if different from "UNIT 1").
Application "Multiplication Factor Weighing"
54

15 Application "Division Factor Weighing"

FACTOR.D
SET.F.DIV
001.0000
009.8
100
1
SET.STEP
0.0001
0.0050
2
The "Division Factor Weighing" divide a predefined factor by the weight value (in grams) (result = factor / weight) and have it rounded to a predefined number of decimal places. Requirement: The function "FACTOR.D" must be assigned to the «F» key (see advanced menu topic "ASSIGN:F").
Activate function "FACTOR.D" by pressing and holding the «F» key.
Setting the Factor Value:
a) Press «
default value or the factor that was saved most recently.
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» key to scroll up or «–» to scroll
down.
d) Press « » briefly to confirm the selected factor (no automatic
acceptance).
Note: Zero for division factor value is outside the allowed range, the error message "FACTOR OUT OF RANGE" will be displayed.
Setting the step value:
"SET.STEP" appears in the display, and the program changes automat ically to allow the display increments to be entered. The smallest pos sible display increment appears as default value, or the last value that was saved.
a) Press « » to execute "SET.STEP".
b) Press « » to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scroll up or «–» to scroll down.
d) Press «
tance).
» to execute "SET.F.DIV". Either the factor 1 appears as
» to confirm the selected step (no automatic accep
Note: The allowed range for the step depends on the factor and the resolution of the balance. If it is outside the allowed range the error message "STEP OUT OF RANGE" will be displayed.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press «C» to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for division factor weighing.
Application "Division Factor Weighing"
55
C
0.010
1000.02
g
0.000
C
Weighing procedure
a) Press « » to zero/tare.
b) Load sample weight on weighing pan.
c) Read the result. The appropriate calculation is then made using
the weight of sample and the selected factor, the result being dis played with the selected display step. Note:No units are displayed. To avoid a division by zero, the fac tor division is not calculated at zero.
d) Unload sample weight.
Toggling between displaying the calculated value and the measured weight:
You can use the «
» key to toggle between the calculated Value, weight value "UNIT 1", "RECALL" value (if selected) and weight value "UNIT 2" (if different from "UNIT 1").
Application "Division Factor Weighing"
56

16 Application "Density"

DENSITY
SOLID
LIQUID
WEIGH IN
60.0020
g
¢¢¢¢
H-2-0
ETHANOL
FREE
The "Density" application allows you to determine the density of solid bodies and liquids. Determination of the density uses Archimedes' principle according to which a body immersed in a fluid undergoes an apparent loss in weight which is equal to the weight of the fluid it displaces.
To determine the density of solid bodies, we recommended you to work with the optional density kit which contains all the attachements and aids needed for convenient and precise density determination. To determine the density of liquids, you additionally need a sinker which you can also obtain from your METTLER TOLEDO dealer.
Note for performing of density determinations:
You can also use the hanger for weighing below the balance which belongs to your bal ance.
We recommended you to consult the operating instructions enclosed with the density kit.
If a METTLER TOLEDO printer is attached to your balance, the settings will be automati cally recorded.
Requirement: The function "DENSITY" must be assigned to the «F» key (see advanced menu topic "ASSIGN:F"). Density kit is installed.
Activate function "DENSITY" by pressing and holding the appropriate assigned «F» key.
Setting the method for density determination
a) Select:
"SOLID", the function for the density determination of solids, or "LIQUID", the function for the density determination of liquids with a sinker.
b) Press « » to confirm the selection
Switching the display between user guidance and weighing
Press « » to toggle the display between user guidance and weigh ing.

16.1 Density Determination of Solids

Requirement: The method "SOLID" is set.
Setting the parameter of the auxiliary liquid
a) Select the auxiliary liquid by scrolling with « » (or «-» up / «+»
down): "H-2-O" for distilled water , "ETHANOL" or "FREE" for a freely defin able auxiliary liquid.
b) Press «
» to confirm the selection.
57Application "Density"
20.0
21.8
If you have selected water or ethanol as the auxiliary liquid:
01.00200
01.00000
PRESS EN
WEIGH IN AIR
WEIGH IN LIQUID
5.988
*
a) Enter the current temperature of the auxiliary liquid (read off on
thermometer). Change the value by scrolling up «+» or down «-». The temperature ranges from 10 °C to 30.9 °C.
b) Press « » to confirm the value.
Note: The densities of distilled water and ethanol in the range 10 °C to
30.9 °C are stored in the balance.
If you have selected a freely definable auxiliary liquid:
Enter the density of the auxiliary liquid at the current temperature (read off on thermometer).
a) Press « » to select a digit. The selected digit is blinking.
b) For changing digits, press «+» to scroll up or «-» to scroll down.
c) Press «
Note: If without any key press within 60 seconds or by pressing «C», the balance returns to the previous active application.
On completion of the settings, your balance is ready for performing the density determination of liquids.
Note: Taring the balance is possible at any time.
The balance prompts you: "PRESS ENTER TO START".
Press « » to start. Tare/Zero is executed.
The balance prompts you to weigh the solid in air "WEIGH IN AIR".
a) Load the solid.
b) Press « » to initiate the measurement.
The balance prompts you to weigh the solid in the auxilliary liquid "WEIGH IN LIQUID".
a) Load the solid.
b) Press «
The balance now shows the determined density of the solid.
Note:
This result has already been corrected for the air buoyancy. The buoyancy caused by the two immersed wires (Ø 0.6 mm) can be neglected.
» to confirm the selected value.
» to initiate the measurement.
58 Application "Density"
By pressing «C», the balance returns to "PRESS ENTER TO START".
Result:
Press « », the result will be printed.
Sample printout:
---- Density Solid -----
18.Mar 2010 20:14 Balance Type ML204 SNR 1234567890
------------------------
ID: ................
Liquid: H-2-O 0.99822 g/cm3 Temp. 20.0 °C Weight in air:
60.0020 g Weight in liquid:
49.9997 g Volume of solid:
1.625 cm3
Density: 5.988 g/cm3 ===========
Signature
........................
------------------------
010.0000
012.0000
PRESS EN
WEIGH IN AIR

16.2 Density Determination of Liquids

Requirement: The method "LIQUID" is set.
Note: If without any key press within 60 seconds or by pressing «C», the balance returns to the previous active
application.
On completion of the settings, your balance is ready for performing the density determination of liquids.
Note: Taring the balance is possible at any time.
Setting the displacement volume of your sinker
Press « » to confirm the default value of 10.0 cm3 or change it if needed:
a) Press « » to select a digit. The selected digit is blinking.
b) For changing digits, press «+» to scroll up or «-» to scroll down
c) Press « » to confirm the selected value.
The balance prompts you: "PRESS ENTER TO START".
Press « » to start.
The balance prompts you to weigh the sinker in air "WEIGH IN AIR".
a) Position the sinker.
b) Press « » to initiate the measurement.
Application "Density"
59
WEIGH IN LIQUID
1.000
*
Sample printout:
---- Density Liquid -----
18.Mar 2010 20:14 Balance Type ML204 SNR 1234567890
------------------------
ID: ................
Temp. of liquid:
............
Displaced liquid:
10.0023 g
Density: 1.000 g/cm3 ===========
Signature
........................
------------------------
The balance prompts you to weigh the sinker in the liquid "WEIGH IN LIQUID".
a) Pour the liquid into the beaker. Make sure that the sinker is
immersed by al least 1 cm in the liquid, and that there are no air bubbles in the container.
b) Press « » to initiate the measurement.
The balance now shows the determined density of the liquid at the cur rent temperature (read off on the thermometer).
Note:
This result has already been corrected for the air buoyancy. The buoyancy caused by the immersed wire (Ø 0.2 mm) of the sinker can be neglected.
By pressing «C», the balance returns to "PRESS ENTER TO START".
Result:
Press « », the result will be printed.
60 Application "Density"

16.3 Formulae Used to Calculate Density

The "DENSITY" Application is based on the formulae listed below.
Formulae for determining the density of solids with compensation for air density
= Density of the sample
A = Weight of the sample in air
B = Weight of the sample in the auxiliary liquid
V = Volume of the sample
= Density of the auxiliary liquid
= Density of Air (0.0012 g/cm3)
= Weight correction factor (0.99985), to take the atmospheric buoyancy of the adjustment weight
into account
Formula for determining the density of liquids with compensation for air density
= Density of the liquid
P = Weight of the displaced liquid
V = Volume of the sinker
= Density of air (0.0012 g/cm3)
= Weight correction factor (0.99985), to take the atmospheric buoyancy of the adjustment weight
into account
Density Table for Distilled Water
61Application "Density"
Density Table for Ethanol
Density of C2H5OH according to the "American Institute of Physics Handbook".
62 Application "Density"

17 Application "Routine Test"

R. TEST
TARE:NO
WGT
TEST.
5000.00
g
LMT
CTRL.
2.50
g
The "Routine Test" application allows you to determine the sensitivity of the balance. More about periodic sensitivity tests (routine tests) see: GWP® (Good Weighing Practice) on www.mt.com/gwp. GWP gives clear recommendation for routine testing:
how should I test my balance?
how often?
where can I reduce efforts?
More about test weights see www.mt.com/weights.
Requirement:
The function "R.TEST" must be assigned to «F» key (see advanced menu topic "ASSIGN:F").
It is recommended to connect a printer or a PC to the balance for showing the results.
a) Activate function "R.TEST" by pressing and holding the assigned
b) Select "TARE:NO" (no tare weight used).
«F» key.
If a tare weight is used during the test select "TARE:YES" (use a tare weight). To toggle between "TARE:YES" and "TARE:NO" use « » (or «+» or «–»).
c) Press « » to confirm the selection.
Note:
It is recommended to test the sensitivity without tare load. (factory setting "TARE:NO").
If using tare: Make sure that tare weight plus test weight is not exceeding max. load.
Setting the reference test weight value
The default value of the test weight: Next smaller OIML weight than the maximum load of your balance according to the GWP® recommenda tion.
a) For changing the value, press «+» to scroll up or «–» to scroll
down. Progressing speed by press and hold.
b) Press « » to confirm the value.
Setting the Control Limit
The default value of the control limit: Test weight x weighing process tolerance / 2 Example: 5000 g x 0.1% / 2 = 2.50 g.
a) For changing the value, press «+» to scroll up or «–» to scroll
down. Progressing speed by press and hold.
b) Press «
» to confirm the value.
63Application "Routine Test"
LMT
WARN.
1.25
g
Setting the Warning Limit
PRESS EN
PRESS EN
5000.00
g
----- Routine Test -----
21.Jan 2009 12:56
METTLER TOLEDO
Balance Type ML6001 SNR 1234567890
Sensitivity: Test weight 5000.00 g Value 5000.11 g Warning L. 1.25 g Control L. 2.50 g Warning L. OK Control L. OK
Signature
........................
------------------------
The default value of the warning limit: Warning limit = control limit / safety factor Example: 2.5 g / 2 = 1.25 g.
a) For changing the value, press «+» to scroll up or «–» to scroll
down. Progressing speed by press and hold.
b) Press « » to confirm the value.
Note: The default values of control limit and the warning limit are evaluated according the GWP recommenda tion. These are based under the assumption that the weighing process tolerance is 0.1% and the safety factor is 2.
On completion of the setting procedure, your balance is ready for the routine test procedure.
Note: The test weight must be acclimatized to the ambient tempera ture of the balance.
a) Press « » to start the test.
b) Follow the instructions on the display. If the test weight value is
flashing: Load the test weight (displayed value).
The printout starts after the weighing pan is unloaded.
Exit the current test procedure:
Press and hold « » for executing a new application.
Printout:
What if Warning Limit or Control Limit are "FAILED"?
The "SOP for Periodic Sensitivity Tests (Routine Tests)" provides information about measures when routine tests fail. Find a download version of these SOPs on www.mt.com/gwp, link "GWP® The Program / Routine Opera tion".
64
Application "Routine Test"
Content of SOP:
Preparation
Test procedure
Evaluation
Deviation
If Warning Limit "FAILED"
If Control Limit "FAILED"
Application "Routine Test"
65

18 Application "Diagnostics"

The "Diagnostics" application allows you to carry out predefined diagnostics tests and to view or print predefined sets of balance information. This diagnostics tool helps you find errors faster and more efficiently. Requirement: A printer or a PC is connected to the balance for showing the results.
a) Activate "ADVANCE." menu. (See section menu operation)
b) Activate function "DIAGNOS." by pressing « ».
c) Use « » to select appropriate tests.

18.1 Repeatability Test

The repeatability test allows you to repeat tests with internal weight for a given number of times. Note: On models with internal weights only.
a) Press « » to activate repeatability test "REPEAT.T". "R.TST. 10" appears on the Display.
b) Enter the number of times (blinking) by pressing «+» or «–». Possible values are 5, 10 (default), 20, 50,
100 times.
c) Press «
ed.
d) Press « » to print the test information..
e) Press « » to scroll forward through the displayed list.
f) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Information Displayed:
Displayed for 0.5 s Display
» to start the test. The message "RUNNING REPEAT TEST" is displayed till the tests are complet
"S DEV" * 0.004 g
"MAX.TEMP" 21.2 °C
"MIN.TEMP" 21.0 °C "AVG.TEMP" 21.1 °C
"TOT.TIME" 00:01:26
66 Application "Diagnostics"
Sample Printout:
-- Repeatability Test --
21.Jan 2009 11:34
METTLER TOLEDO
Balance Type ML4002 SNR 1234567890 SW V1.00 Temperature 21.3 °C No. of tests 10
------------------------
1. Time 00:00:00
1. Temp 21.3 °C
2. Time 00:00:04
2. Temp 21.3 °C . . .
-----------------------­s Dev 0.004 g Max Temp 21.3 °C Min Temp 21.3 °C Mean Temp 21.3 °C Total Time 00:00:44
------------------------
Examples:
Repeatability test is a tool to do functional check with the balance. It may be performed:
To check function of balance
To develop the optimal environment settings (see menu topic "ENVIRON."). Measure the time you need to perform repeatability test with each "STABLE", "STD." and "UNSTAB." setting. The setting with the fastest total time suits best for the existing environmental conditions.

18.2 Display Test

The display test allows you to test the display of the balance.
a) Press «
b) Press «
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
during installation to store print out with installation documents.
after preventative maintenance to store print out with installation maintenance report.
when remarkable decrease of weighing performance occurs, so that you can email/fax print out to ser vice support provider for diagnose purposes.
» to start "DISPLAY".
All possible segments and icons on the display will illuminate.
» to print the test information.
67Application "Diagnostics"
Sample Printout:
----- Display Test -----
21.Jan 2009 11:34
METTLER TOLEDO
Balance Type ML204 SNR 1234567890 SW V1.00 Display Test DONE
------------------------
------- Key Test -------
21.Jan 2009 11:34
METTLER TOLEDO
Balance Type ML204 SNR 1234567890 SW V1.00 1/10 d Key OK Print Key OK Minus Key OK Plus Key OK Toggle Key OK Enter Key OK Zero/Tare Key OK Cancel Key OK
------------------------

18.3 Key Test

The key test allows you to test the keys of the balance.
a) Press «
» to start "KEYPAD.T".
b) The message "KEY TEST - PRESS KEY TO BE TESTED" is displayed scrolling during the duration of the key
test. Press every Key briefly. Each press of a key beeps and echoes with "OK” on the display.
c) Second press «C» key to print the test information. The test procedure will be cancelled and the balance will
return to the topic "DIAGNOS.". If a key has not been tested before printing, then the test results will be indi cated with a "----" line.
Sample Information Displayed:
Key Display
« » 1/10.D.OK
« » PRINT.OK
«–» MINUS.OK «+» PLUS OK
« » TOGGL.OK
« » ENTER.OK
«C» C OK
« » 0/T OK
Sample Printout:
68 Application "Diagnostics"

18.4 Motor Test

------ Motor Test ------
21.Jan 2009 11:34
METTLER TOLEDO
Balance Type ML204 SNR 1234567890 SW V1.00 Motor Test OK
------------------------
The motor test allows you to test the calibration motor of the balance. Note: On models with internal weight only.
a) Press « » to start "CAL.MOT.T".
"RUNNING" is displayed during the Motor Test. A motor test is deemed successful when all the motor posi tions have been successfully tested. At the end of the test, the test information will be printed.
b) Press « » for printout.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Printout:

18.5 Balance History

The balance history function allows you to view and print the history of the balance.
a) Press « » to start "BAL.HIST" .
b) Press « » to scroll forward through the displayed list of balance history information.
c) Press «
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Information Displayed:
Operation Time (year:day:hour) 00:018:04 Total load kg (t) 115.7191 kg Number of weighings 1255 Number of key pressed 4931 Number of motor movements 1012 Backlight time (year:day:hour) 00:018:04 Next service due date 01:01:10
» for printout.
Information Display
69Application "Diagnostics"
Sample Printout:
--- Statistical Info ---
21.Jan 2009 11:34
METTLER TOLEDO
Balance Type ML4002 SNR 1234567890 SW V1.00
-----------------------­Operating time 18d 23h Total weight loaded
115.7191 kg Number of weighings 1255 Number of key presses 4931 Motor movements 1012 Backlight operating time 18d 4h Next service due date
01.01.2010
------------------------

18.6 Calibration History

The "Calibration History" function allows you to view and print information of the last 30 (thirty) balance adjust ment. Adjustments made by a service technician and normal user are counted together.
a) Press «
b) Press « » for printout.
c) Press « » key to scroll forward through the displayed list of Adjustments history information.
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Information Displayed:
S = External adjusted service 05:03:09S 01
70 Application "Diagnostics"
I = Internal adjusted 04:03:09I 02
I = Internal adjusted 03:03:09I 28
E = External adjusted user 02:03:09E 29
I = Internal adjusted 02:03:09I 30
» to start "CAL.HIST".
Note Display
-3 PPM
2 PPM
-1 PPM
4 PPM
1 PPM
·
·
·
·
·
·
Sample Printout:
----- Calibration ------
05.Mar 2010 11:34
METTLER TOLEDO
Balance Type ML204 SNR 1234567890
-----------------------­01 05.Mar 2010 11:34 External ADJ SERVICE
23.5°C Diff -3ppm
-----------------------­02 04.Mar 2010 09:00 Internal ADJ
22.4°C Diff 2ppm
-----------------------­. . . 28 03.Mar 2010 10:59 Internal ADJ
22.6°C Diff -1ppm
-----------------------­29 02.Mar 2010 16:34 External ADJ USER
24.6°C Diff 4ppm
-----------------------­30 02.Mar 2010 18:36 Internal ADJ
22.4°C Diff 1ppm
------------------------

18.7 Balance Information

The balance information function allows you to view and print information about your balance.
a) Press «
b) Press « » for printout.
c) Press « » to scroll forward through the displayed list of Balance information.
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample information displayed:
Balance type TYPE ML4002 Max. load MAX 4200 g Software platform PLATFORM RAINBOW Serial number SNR 1234567890 Type definition number TDNR 9.6.3.411 Software version SOFTWARE V1.00 Cell ID CELL ID 1172400044
» to start "BAL.INFO".
Information Display
71Application "Diagnostics"
Information Display
-- Balance Information -
05.Mar 2009 11:34
METTLER TOLEDO
Balance Type ML4002 SNR 1234567890 SW V1.00 Max 4200 g Platform Rainbow TDNR 9.6.3.411 Cell ID 1172400044 Cell Type MMAI4000G2 Tolerance Rev.no. 2 Language English
------------------------
--- Service Provider ---
21.Jan 2009 11:34
METTLER TOLEDO Im Langacher CH-8606 Greifensee Switzerland (+41) 044 944 22 11
------------------------
Cell type CELL TYPE MMAI4000G2 Tolerance revision number TOLERANCE NO2 Language ENGLISH
Sample Printout:

18.8 Service Provider Information

The service provider Information function allows you to print information about your service provider.
a) Press « » to start "SRV.PROV". The service provider information will be displayed.
b) Press « ». The service provider information will be printed and the balance will return to the topic "DIAG
NOS.".
Sample Printout:
72 Application "Diagnostics"

19 Communication with Peripheral Devices

19.1 Function PC-Direct

The numerical value displayed at the balance can be transferred to the cursor position in Windows Applications (e.g. Excel, Word) as by typing with the keyboard.
Note: The units will not be transferred.
Requirements
PC with Microsoft Windows® operating system and serial interface RS232.
Windows Application (e.g. Excel).
Balance to PC connection with cabel RS232 (e.g. No. 11101051 see chapter accessories).
Balance Interface Setting (see Interface Menu):
Topic "RS232": set "PC-DIR." and select the most appropriate option for the desired weighing result.
Save changes.
Settings at the PC
Note: The following examples are based on Windows XP.
a) Click "start".
b) Click "ControlPanel".
c) Click "Accessibility Options" in the Control Panel.
Accessibility Option
a) Click "General" Tab.
b) Enter a check mark at "Use Serial Keys".
c) Click "Settings".
Settings for SerialKeys
a) Select the serial port to be used for connection with the balance.
b) Set the baud rate to 9600
c) Click "OK".
Communication with Peripheral Devices
73
Complete the settings
a) Click "Apply" when active (wait until active).
b) Click "OK" .
Note: If the "serial key" is enabled, applications that use the same port may not function correctly. Remove the check mark from the check box "Use Serial Keys" to disable serial key function.
Checking Operation
a) Start Excel (or another application) at the PC.
b) Activate a cell in Excel.
According to your selected "PC-DIR." option, the displayed values will appear in the column one after the other one in the different rows.
74 Communication with Peripheral Devices

20 Firmware (Software) Updates

METTLER TOLEDO is continuously improving its balance firmware (software) for the benefit of customers. So that the customer can benefit quickly and easily from further developments, METTLER TOLEDO makes the latest firmware versions available on the Internet. The firmware made available on the Internet has been developed and tested by Mettler-Toledo AG using processes that meet the guidelines of ISO 9001. Mettler-Toledo AG does not, however, accept liability for consequences that might arise from using the firmware.

20.1 Operating Principle

You will find all the relevant information and updates for your balance on the METTLER TOLEDO website at the following address:
www.mettler-toledo-support.com
A program known as the "e-Loader II" is loaded onto your computer together with the firmware update. You can use this program to download the firmware to the balance. The "e-Loader II" can also save the settings in your balance before the new firmware is downloaded to it. You can reload the saved settings into the balance manually or automatically after the software is downloaded.
If the selected update includes an application that is not described in these instructions (or that has been updat ed in the meantime) you can download the corresponding instructions in Adobe Acrobat® PDF format.
Requirements
The minimum requirements for obtaining applications from the Internet and downloading them into your bal ance are as follows:
PC with Microsoft Windows® operating system (Version 98, 98SE, ME, NT4.0, 2000, XP or Vista).
Internet connection and web browser (e.g. MS Internet Explorer).
PC to balance connection cable (e.g. No. 11101051 see chapter accessories)

20.2 Update Procedure

Installing the "e-Loader II" software from the Internet onto the PC.
a) Connect to the Internet.
b) Go to the site "www.mettler-toledo-support.com".
c) Enter the information required for registration on the METTLER TOLEDO Balance Support Site.
d) Click the "Customer Support" link and log in.
e) Click your Balance.
f) Click the firmware version you need and install it.
Loading the new firmware into the balance.
Start the "e-Loader II" and follow the instructions, which will take you step-by-step through the installation.
Firmware (Software) Updates
75

21 Error and Status Messages

21.1 Error Messages

Error messages in the display draw your attention to incorrect operation or that the balance could not execute a procedure properly.
Error Message Cause Rectification
NO STABILITY No stability. Ensure more stable ambient condi
tions. If not possible, check settings for environment.
WRONG ADJUSTMENT WEIGHT Wrong adjustment weight on pan
or none at all.
REFERENCE TOO SMALL Reference for piece counting too
small. EEPROM ERROR - PLEASE CON TACT CUSTOMER SERVICE WRONG CELL DATA - PLEASE CON TACT CUSTOMER SERVICE NO STANDARD ADJUSTMENT ­PLEASE CONTACT CUSTOMER SER VICE PROGRAM MEMORY DEFECT ­PLEASE CONTACT CUSTOMER SER VICE TEMP SENSOR DEFECT - PLEASE CONTACT CUSTOMER SERVICE WRONG LOAD CELL BRAND ­PLEASE CONTACT CUSTOMER SER VICE WRONG TYPE DATA SET - PLEASE CONTACT CUSTOMER SERVICE BATTERY BACKUP LOST - CHECK DATE TIME SETTINGS
INITIAL ZERO RANGE EXCEEDED Wrong weighing pan or pan is not
BELOW INITIAL ZERO RANGE Wrong weighing pan or pan is
MEM.FULL Memory full. Clear the memory and start a new
FACTOR OUT OF RANGE Factor is outside the allow range. Select a new factor. STEP OUT OF RANGE Step is outside the allow range. Select a new step. OUT OF RANGE Sample weight is outside the allow
EEPROM (memory) error. Please contact METTLER TOLEDO
Wrong cell data. Please contact METTLER TOLEDO
No standard calibration. Please contact METTLER TOLEDO
Program memory defect. Please contact METTLER TOLEDO
Temperature sensor defect. Please contact METTLER TOLEDO
Wrong load cell brand. Please contact METTLER TOLEDO
Wrong type data set. Please contact METTLER TOLEDO
Backup battery is empty. This bat
tery ensures that the date and time
are not lost when the balance is
disconnected from power.
Overload - The weight on the pan
exceeds the weighing capacity of
the balance.
Underload Check that the weighing pan is
empty.
missing.
range.
Place required adjustment weight in center of pan. Increase reference weight.
customer service.
customer service.
customer service.
customer service.
customer service.
customer service.
customer service. Connect the balance to the power supply for charging the battery (e.g. during the night) or contact METTLER TOLEDO customer ser vice. Reduce the weight on the weighing pan.
positioned correctly. Mount correct weighing pan or unload weighing pan. Mount correct weighing pan.
evaluation.
Unload the pan and load a new sample weight.
76 Error and Status Messages

21.2 Status Messages

Status messages are displayed by means of small icons. The status icons indicate the following:
Status Icon Signification
Service Reminder Your balance is due for servicing. Contact your dealer's customer
service department as soon as possible to have a technician service your balance. (See menu topic "SRV.ICON")
77Error and Status Messages

22 Cleaning and Service

Every now and then, clean the weighing pan, draft shield element, bottom plate, draft shield (depending on the model) and housing of your balance. Your balance is made from high-quality, durable materials and can therefore be cleaned using a damp cloth or with a standard, mild cleaning agent.
Please observe the following notes:
The balance must be disconnected from the power supply
Ensure that no liquid comes into contact with the balance or the AC adapter.
Never open the balance or AC adapter – they contain no components, which can be cleaned, repaired or replaced by the user.
On no account use cleaning agents which contain solvents or abrasive ingredients, as this can result in damage to the operation panel overlay.
Do not use wet, but only damp cloth for cleaning.
Please contact your METTLER TOLEDO dealer for details of the available service options. Reg ular servicing by an authorized service engineer ensures constant accuracy for years to come and prolongs the service life of your balance.
78 Cleaning and Service

23 Interface Specification

15
9 6
IN
GND
OUT
HAND SHAKE
IN
DATA
OUT
RxD
TxD
CTS
RTS
OUT
POWER SUPPLY
+12V
2nd di
splay mode only

23.1 RS232C Interface

Each balance is equipped with an RS232C Interface as standard for the attachment of a peripheral device (e.g. printer or computer).
Schematic Item Specification
Interface type Voltage interface according to EIA
RS-232C/DIN66020 CCITT V24/V.28) Max. cable length 15 m Signal level Outputs:
+5 V … +15 V (RL = 3–7 kΩ)
–5 V … –15 V (RL = 3–7 kΩ)
Inputs:
+3 V … +25 V
–3 V … -25 V Connector Sub-D, 9-pole, female Operating mode Full duplex Transmission mode Bit-serial, asynchronous Transmission code ASCII Baud rates 600, 1200, 2400, 4800, 9600, 19200,
38400 (software selectable) Bits/parity 7-bit/none, 7-bit/even, 7-bit/odd, 8-bit/none
(software selectable) Stop bits 1 stop bit Handshake None, XON/XOFF, RTS/CTS (software selectable) End-of-line <CR><LF>, <CR>, <LF> (software selectable) Power supply for 2nd display
+ 12 V, max 40 mA (software selectable, 2nd
display mode only)

23.2 MT-SICS Interface Commands and Functions

Many of the balances and scales used have to be capable of integration in a complex computer or data acqui sition system.
To enable you to integrate balances in your system in a simple manner and utilize their capabilities to the full, most balance functions are also available as appropriate commands via the data interface.
All new METTLER TOLEDO balances launched on the market support the standardized command set "METTLER TOLEDO Standard Interface Command Set" (MT-SICS). The commands available depending on the functionality of the balance.
Basic information on data interchange with the balance
The balance receives commands from the system and acknowledges the command with an appropriate response.
Command formats
Commands sent to the balance comprise one or more characters of the ASCII character set. Here, the following must be noted:
Enter commands only in uppercase.
The possible parameters of the command must be separated from one another and from the command name by a space (ASCII 32 dec., in this description represented as ).
79Interface Specification
The possible input for "text" is a sequence of characters of the 8-bit ASCII character set from 32 dec to 255 dec.
Each command must be closed by CRLF (ASCII 13 dec., 10 dec.). The characters CRLF, which can be inputted using the Enter or Return key of most entry keypads, are not listed in this description, but it is essential they be included for communication with the balance.
Example
S – Send stable weight value
Command S Get the current stable net weight value.
Response S
S WeightValue Unit Current stable weight value in unit actually set
under unit 1.
S I Command not executable (balance is currently
executing another command, e.g. taring, or timeout as stability was not reached).
S + Balance in overload range.
S - Balance in underload range.
Example
Command S Query a stable weight value.
Response S S 100.00 g The current stable weight value is 100.00 g.
The available MT-SICS commands are listed in the table (depending on the model). For further information please refer to the Reference Manual "MT-SICS 11780711" downloadable from the Internet under
www.mt.com/sics-newclassic.
Description Description
@ Cancel (Reset) M46 Print interval C0 Query/Set adjustment settings PW Piece counting: Query/Set piece weight C1 Start adjustment according to current set
tings
PWR Power on/off (PWR 0 means switch off
completely, if balance is powered by bat tery)
C2 Start adjustment with external weight S Send stable weight value C3 Start adjustment with internal weight SI Send weight value immediately D Display text sent to balance SIR Send weight value immediately and repeat DAT Date query/set SIRU Send weight value with currently displayed
unit immediately and repeat
DW Display weight SIU Send weight value with currently displayed
unit immediately
I0 Commands implemented SM0 Dynamic weighing: cancel all SMx com
mands
I1 MT-SICS level and MT-SICS versions SM1 Dynamic weighing: Start immediately and
send the result
I2 Balance data SM2 Dynamic weighing: start after a minimum
load is exceeded and send result
I3 Software version, type definition number SM3 Dynamic weighing: start after a minimum
load is exceeded, send result and repeat
I4 Query serial number (SNR) SM4 Dynamic weighing: query/set time interval I5 Query SW-identification number SNR Send stable weight value and repeat on
weight change
I10 Query/set balance ID SNRU Send stable weight valuewith currently dis
played unit and repeat on weight change
80 Interface Specification
Description Description
I11 Query balance type SR Send weight value on weight change I14 Query balance information SRU Send stable weight value with currently dis
played unit on weight change
K Keys: set configuration ST Send stable weight value on pressing (print)
key
M02 Query/set environment SU Send stable weight value with currently dis
played unit
M03 Query/set AutoZero T Tare M08 Display brightness TA Get/Set tare weight value M09 Display contrast TAC Clear tare value M11 Beeper: Query/set volume TI Tare immediately M14 List available language TIM Query/set time M15 Query/set language TST0 Query/set test function settings M17 FACT: query/set single time criteria (no pos
sibility to set "weekday"
M22 Custom unit definition
Remarks: no possibility to set "name" of unit
M25 Get application list TST3 Start test function with internal weight M26 Get/set current application UPD Query/set update rate of the host interface M27 Adjustment history Z Zero M30 +/- settings with nominal and tolerance ZI Zero immediately
TST1 Start test function according to current set
tings
TST2 Start test function with external weight
81Interface Specification

24 Technical Data

24.1 General Data

Power Supply
• AC operation: AC/DC Adapter Primary: 100V–240V, 50/60Hz, 0.3 A Secondary: 12VDC, 0.84A (with electronic overload protection) Power supply to the balance: 8–20VDC, 10W
Use only with a tested AC Adapter with SELV output current. Ensure correct polarity
• Battery operation: 8 standard AA (LR6) batteries (alkaline) for 8–15 hours of use.
Protection and Standards
• Overvoltage category: Class III
• Degree of pollution: 2
• Degree of Protection: Protected against dust and water: IP54 in use with weighing pan.
• Standards for safety and EMC: See Declaration of Conformity
• Range of application: For use only in dry interior rooms
Environmental conditions
• Height above mean sea level: up to 4000 m
• Ambient temperature range: 10 to 30 °C
• Relative air humidity: 10% to 80 % at 31 °C, linearly decreasing to 50 % at 40 °C, non condensing
Materials
• Housing: Top Housing: Plastic (ABS) Bottom housing: Die-cast aluminum, lacquered
• Weighing pan: Stainless steel X2CrNiMo 17-12-2 (1.4404)
• Draft shield element: with 0.1 mg models: Stainless steel X2CrNiMo 17-12-2 (1.4404)
• Draft shield: Plastic (ABS), glass
• In-use-cover: Plastic (ABS)

24.2 Model-Specific Data

24.2.1 Balances with Readability of 0.1 mg with Draft Shield

Technical Data
Model ML54 ML104 ML204
Maximum load 52 g 120 g 220 g Readability 0.1 mg 0.1 mg 0.1 mg Taring range 0...52 g 0...120 g 0...220 g Repeatability (sd) 0.1 mg 0.1 mg 0.1 mg Linearity 0.2 mg 0.2 mg 0.2 mg Sensitivity temperature drift (10...30°C) 1.5 ppm/°C 1.5 ppm/°C 1.5 ppm/°C Internal adjustment yes yes yes Adjustment range with external weights 20...52 g 50...120 g 100...220 g
82 Technical Data
Model ML54 ML104 ML204
Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM Minimum weight (acc. to USP) 0.3 g 0.3 g 0.3 g Minimum weight (U=1%, k=2) 0.02 g 0.02 g 0.02 g Minimum weight (OIML) 0.01 g 0.01 g 0.01 g Settling time, typ. 2 s 2 s 3 s Weighing technology MonoBloc MonoBloc MonoBloc Usable height of draft shield [mm] 235 235 235 Weighing pan dimensions (WxD) [mm] Ø 90 Ø 90 Ø 90 Balance dimensions (WxDxH) [mm] 193x290x331 193x290x331 193x290x331 Net Weight [kg] 4.1 4.1 4.1
50 g / F2/4 2 g / E2/2
100 g / F2/4 5 g / E2/2
200 g / F2/4 10 g / F1/3

24.2.2 Balances with Readability of 1 mg with Draft Shield

Technical Data
Model ML203 ML203E ML303
Maximum load 220 g 220 g 320 g Readability 0.001 g 0.001 g 0.001 g Taring range 0...220 g 0...220 g 0...320 g Repeatability (sd) 0.001 g 0.001 g 0.001 g Linearity 0.002 g 0.002 g 0.002 g Sensitivity temperature drift (10...30°C) 3 ppm/°C 3 ppm/°C 3 ppm/°C Internal adjustment yes no yes Adjustment range with external weights 100...220 g 100...220 g 100...320 g Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM Minimum weight (acc. to USP) 3 g 3 g 3 g Minimum weight (U=1%, k=2) 0.2 g 0.2 g 0.2 g Minimum weight (OIML) 0.02 g 0.02 g 0.02 g Settling time, typ. 1.5 s 1.5 s 1.5 s Weighing technology MonoBloc MonoBloc MonoBloc Usable height of draft shield [mm] 230 105 230 Weighing pan dimensions (WxD) [mm] Ø 120 Ø 120 Ø 120 Balance dimensions (WxDxH) [mm] 193x290x331 184x290x188 193x290x331 Net Weight [kg] 4.2 2.8 4.2
200 g / F2/4 10 g / F1/3
200 g / F2/4 10 g / F1/3
200 g / F2/4 20 g / F1/3
Model ML303E ML503
Maximum load 320 g 520 g Readability 0.001 g 0.001 g Taring range 0...320 g 0...520 g Repeatability (sd) 0.001 g 0.001 g Linearity 0.002 g 0.002 g Sensitivity temperature drift (10...30°C) 3 ppm/°C 3 ppm/°C Internal adjustment no yes Adjustment range with external weights 100...320 g 200...520 g
83Technical Data
Model ML303E ML503
Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM Minimum weight (acc. to USP) 3 g 3 g Minimum weight (U=1%, k=2) 0.2 g 0.2 g Minimum weight (OIML) 0.02 g 0.02 g Settling time, typ. 1.5 s 1.5 s Weighing technology MonoBloc MonoBloc Usable height of draft shield [mm] 105 230 Weighing pan dimensions (WxD) [mm] Ø 120 Ø 120 Balance dimensions (WxDxH) [mm] 184x290x188 193x290x331 Net Weight [kg] 2.8 4.2
200 g / F2/4 10 g / F1/3
500 g / F2/4 20 g / F1/3

24.2.3 Balances with Readability of 0.01 g

Technical Data
Model ML802 ML802E ML1502E
Maximum load 820 g 820 g 1520 g Readability 0.01 g 0.01 g 0.01 g Taring range 0...820 g 0...820 g 0...1520 g Repeatability (sd) 0.01 g 0.01 g 0.01 g Linearity 0.02 g 0.02 g 0.02 g Sensitivity temperature drift (10...30°C) 3 ppm/°C 3 ppm/°C 3 ppm/°C Internal adjustment yes no no Adjustment range with external weights 100...820 g 100...820 g 100...1520 g Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM Minimum weight (acc. to USP) 30 g 30 g 30 g Minimum weight (U=1%, k=2) 2 g 2 g 2 g Minimum weight (OIML) 0.5 g 0.5 g 0.5 g Settling time, typ. 1.5 s 1.5 s 1.5 s Weighing technology MonoBloc Strain Gauge Strain Gauge Weighing pan dimensions (WxD) [mm] 170x190 Ø 160 Ø 160 Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84 184x290x84 Net Weight [kg] 3.6 2.2 2.2
500 g / F2/4 20 g / F2/4
500 g / F2/4 20 g / F2/4
1000 g / F2/4 50 g / F2/4
Model ML1602 ML3002 ML3002E
Maximum load 1620 g 3200 g 3200 g Readability 0.01 g 0.01 g 0.01 g Taring range 0...1620 g 0...3200 g 0...3200 g Repeatability (sd) 0.01 g 0.01 g 0.01 g Linearity 0.02 g 0.02 g 0.02 g Sensitivity temperature drift (10...30°C) 3 ppm/°C 3 ppm/°C 3 ppm/°C Internal adjustment yes yes no Adjustment range with external weights 1000...1620 g 1000...3200 g 1000...3200 g Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM
84 Technical Data
1000 g / F2/4 50 g / F2/4
2000 g / F2/4 100 g / F2/4
2000 g / F2/4 100 g / F2/4
Model ML1602 ML3002 ML3002E
Minimum weight (acc. to USP) 30 g 30 g 30 g Minimum weight (U=1%, k=2) 2 g 2 g 2 g Minimum weight (OIML) 0.5 g 0.5 g 0.5 g Settling time, typ. 1.5 s 1.5 s 1.5 s Weighing technology MonoBloc MonoBloc MonoBloc Weighing pan dimensions (WxD) [mm] 170x190 170x190 170x190 Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84 184x290x84 Net Weight [kg] 3.6 3.6 3.4
Model ML4002 ML4002E
Maximum load 4200 g 4200 g Readability 0.01 g 0.01 g Taring range 0...4200 g 0...4200 g Repeatability (sd) 0.01 g 0.01 g Linearity 0.02 g 0.02 g Sensitivity temperature drift (10...30°C) 3 ppm/°C 3 ppm/°C Internal adjustment yes no Adjustment range with external weights 2000...4200 g 2000...4200 g Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM Minimum weight (acc. to USP) 30 g 30 g Minimum weight (U=1%, k=2) 2 g 2 g Minimum weight (OIML) 0.5 g 0.5 g Settling time, typ. 1.2 s 1.2 s Weighing technology MonoBloc MonoBloc Weighing pan dimensions (WxD) [mm] 170x190 170x190 Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84 Net Weight [kg] 3.6 3.4
2000 g / F2/4 200 g / F2/4
2000 g / F2/4 200 g / F2/4
Model ML6002 ML6002E
Maximum load 6200 g 6200 g Readability 0.01 g 0.01 g Taring range 0...6200 g 0...6200 g Repeatability (sd) 0.01 g 0.01 g Linearity 0.02 g 0.02 g Sensitivity temperature drift (10...30°C) 3 ppm/°C 3 ppm/°C Internal adjustment yes no Adjustment range with external weights 2000...6200 g 2000...6200 g Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM Minimum weight (acc. to USP) 30 g 30 g Minimum weight (U=1%, k=2) 2 g 2 g Minimum weight (OIML) 0.5 g 0.5 g Settling time, typ. 1.2 s 1.2 s Weighing technology MonoBloc MonoBloc Weighing pan dimensions (WxD) [mm] 170x190 170x190 Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84 Net Weight [kg] 3.6 3.4
5000 g / F2/4 200 g / F2/4
5000 g / F2/4 200 g / F2/4
85Technical Data

24.2.4 Balances with Readability of 0.1 g

Technical Data
Model ML601E* ML1501E* ML3001E*
Maximum load 620 g 1520 g 3200 g Readability 0.1 g 0.1 g 0.1 g Taring range 0...620 g 0...1520 g 0...3200 g Repeatability (sd) 0.1 g 0.1 g 0.1 g Linearity 0.1 g 0.1 g 0.2 g Sensitivity temperature drift (10...30°C) 5 ppm/°C 5 ppm/°C 5 ppm/°C Internal adjustment no no no Adjustment range with external weights 100...620 g 100...1520 g 1000...3200 g Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM
500 g / F2/4 20 g / F2/4
1000 g / F2/4 50 g / F2/4
2000 g / F2/4
100 g / F2/4 Minimum weight (acc. to USP) 300 g 300 g 300 g Minimum weight (U=1%, k=2) 20 g 20 g 20 g Minimum weight (OIML) 5 g 5 g 5 g Settling time, typ. 1 s 1 s 1 s Weighing technology Strain Gauge Strain Gauge Strain Gauge Weighing pan dimensions (WxD) [mm] Ø 160 Ø 160 Ø 160 Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84 184x290x84 Net Weight [kg] 2.3 2.3 2.3
* Only available in selected countries.
Approved version not available.
Model ML2001 ML4001 ML6001
Maximum load 2200 g 4200 g 6200 g Readability 0.1 g 0.1 g 0.1 g Taring range 0...2200 g 0...4200 g 0...6200 g Repeatability (sd) 0.1 g 0.1 g 0.1 g Linearity 0.2 g / 0.1 g
1)
0.2 g / 0.1 g
1)
0.2 g Sensitivity temperature drift (10...30°C) 5 ppm/°C 5 ppm/°C 5 ppm/°C Internal adjustment yes yes yes Adjustment range with external weights 500...2200 g 500...4200 g 1000...6200 g Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM
2000 g / F2/4 100 g / F2/4
2000 g / F2/4 200 g / F2/4
5000 g / F2/4
200 g / F2/4 Minimum weight (acc. to USP) 300 g 300 g 300 g Minimum weight (U=1%, k=2) 20 g 20 g 20 g Minimum weight (OIML) 5 g 5 g 5 g Settling time, typ. 1 s 1 s 1 s Weighing technology MonoBloc MonoBloc MonoBloc Weighing pan dimensions (WxD) [mm] 170x190 170x190 170x190 Balance dimensions (WxDxH) [mm] 184x290x84 184x290x84 184x290x84 Net Weight [kg] 3.3 3.3 3.3
1)
approved version (OIML)
Model ML6001E
Maximum load 6200 g
86 Technical Data
Model ML6001E
Readability 0.1 g Taring range 0...6200 g Repeatability (sd) 0.1 g Linearity 0.2 g Sensitivity temperature drift (10...30°C) 5 ppm/°C Internal adjustment no Adjustment range with external weights 1000...6200 g Weights for routine testing Large Weight/Class OIML/ASTM Small Weight/Class OIML/ASTM Minimum weight (acc. to USP) 300 g Minimum weight (U=1%, k=2) 20 g Minimum weight (OIML) 5 g Settling time, typ. 1 s Weighing technology Strain Gauge Weighing pan dimensions (WxD) [mm] Ø 160 Balance dimensions (WxDxH) [mm] 184x290x84 Net Weight [kg] 2.3
5000 g / F2/4 200 g / F2/4
87Technical Data

24.3 Dimensions

(235)
134
54
216
19
156
163
144
290
184
193
331
70
83
185
78
ø 90
169
C
126
97

24.3.1 Balances with Readability of 0.1 mg with Draft Shield (235 mm)

Models:
ML54 ML104 ML204
88 Technical Data

24.3.2 Balances with Readability of 1 mg with Draft Shield (235 mm)

(230)
134
54
216
19
156
163
144
290
184
193
331
70
85
185
78
ø 120
169
C
126
97
Models:
ML203 ML303 ML503
89Technical Data

24.3.3 Balances with Readability of 1 mg with "Flex" Draft Shield (105 mm)

C
101
129
54
216
34
166
166
144
290
184
188
70
85
172 £
83
ø 120
169
Models:
ML203E ML303E
90 Technical Data
24.3.4 Balances with Readability of 0.01 g with Square Weighing Pan and Draft Shield
167
200
54
70
84
184
203
170
144
190
290
200
87
180
C
Element
Models:
ML802 ML1602 ML3002 ML3002E ML4002 ML4002E ML 6002 ML 6002E
91Technical Data

24.3.5 Balances with Readability of 0.01 g / 0.1 g with Round Weighing Pan

129
54
216
144
290
184
70
84
ø 160
169
C
Models:
ML802E ML1502E
ML601E ML1501E ML3001E ML6001E
92 Technical Data

24.3.6 Balances with Readability of 0.1 g with Square Weighing Pan

167
200
54
70
84
184
170
144
190
203
290
87
C
Models:
ML2001 ML4001 ML6001
93Technical Data

25 Accessories and Spare Parts

25.1 Accessories

Description Part No.
Density Determination
Density kit ML-DNY-43 for NewClassic ML Balances (d=0.1mg/1mg)
Glass beaker, height 100 mm, Ø 60 mm 00238167
Sinker for density of liquids in conjunction with Density Kit 00210260
Calibrated (sinker + certificate) 00210672
Recalibrated (new certificate) 00210674
Calibrated thermometer with certificate 11132685
11142144
Draft Shields
Printers
"Flex" draft shield "1 mg" (usable heigh 105mm) 12122420
Draft shield ML-DS-21 for models with readability of 0.1g up to 0.01g.
RS-P25 printer with RS232C connection to balance 11124300
12121015
94 Accessories and Spare Parts
RS-P26 printer with RS232C connection to balance (with date
METTLER TOLEDO
000000
and time)
11124303
RS-P28 printer with RS232C connection to balance (with date, time and applications
Cables for RS232C Interface
RS9 – RS9 (m/f): connection cable for PC, length = 1 m 11101051
RS9 – RS25 (m/f): connection cable for PC, length = 1 m 11101052
RS9 – RS9 (m/m): connection cable for devices with DB9 (f) socket, length = 1 m
11124304
21250066
Auxiliary Displays
Power Supplies
RS232 - USB converter – intelligent expansion module for con
11103691
nection to PC
RS232 auxiliary display AD-RS-M7 12122381
AC/DC universal adapter (EU, USA, AU, UK) 100–240 VAC,
11120270
50/60HZ, 0.3 A, 12 VDC 0.84 A
Accessories and Spare Parts
95
Protective Covers
PowerPac-M-12V, for mains independent operation of bal ances, 12 VDC/1 A
Protective cover for models with draft shield "165/235 mm" 12122030
Protective cover for models with round weighing pan 12122032
Protective cover for models with square weighing pan 12122031
12122363
Anti-theft Devices
Software
Adjustment Weights
Steel cable 11600361
LabX direct balance (simple data transfer) 11120340
OIML / ASTM Weights (with calibration certificate) see www.mt.com/weights
96 Accessories and Spare Parts

25.2 Spare Parts

C
4
3
2
1
9
8
10
5
7
6
C
3
2
1
4
8
7
9
6
5
Balances with readability of 0.1 mg, with draft shield (235 mm)
Drawing Pos Description Part No.
1 Weighing pan Ø 90 mm 12122010 2 Pan support 12122042 3 Draft shield element 12122043 4 Bottom plate 12122044 5 Side glass door 12122036 6 Top glass door 12122033 7 Pair of handles 12122035 8 Leveling foot 12122040 9 Battery-chamber cover 12122041 10 Weighing below balance cap 12122029
Balances with readability of 1 mg, with draft shield (235 mm)
Drawing Pos Description Part No.
1 Weighing pan Ø 120 mm 12122037 2 Pan support 12122045 3 Bottom plate 12122044 4 Side glass door 12122036 5 Top glass door 12122033 6 Pair of handles 12122035 7 Leveling foot 12122040 8 Battery-chamber cover 12122041 9 Weighing below balance cap 12122029
97Accessories and Spare Parts
Balances with readability of 1 mg, with flex draft shield (105 mm)
C
2
1
5
3
4
8
7
9
6
C
1
2
3
4
6
5
7
C
1
2
4
3
5
Drawing Pos Description Part No.
1 Weighing pan Ø 120 mm 12122037 2 Pan support 12122045 3 Bottom plate 12122047 4 Draft shield top cover 12122046 5 Draft shield frame 12122039 6 Draft shield glass panels 12122038 7 Leveling foot 12122040 8 Battery-chamber cover 12122041 9 Weighing below balance cap 12122029
Balances with readability of 10 mg with square weighing pan and draft shield element
Drawing Pos Description Part No.
1 Weighing pan 170 mm x 190 mm 12122048 2 Pan support 12122049 3 Draft shield element 12122050 4 Pan support cap 11131029 5 Leveling foot 12122040 6 Battery-chamber cover 12122041 7 Weighing below balance cap 12122029
Balances with readability of 0.01 g / 0.1 g with round weighing pan
Drawing Pos Description Part No.
98 Accessories and Spare Parts
1 Weighing pan Ø 160 mm 12102941 2 Pan support 12122052 3 Leveling foot 12122040 4 Battery-chamber cover 12122041 5 Weighing below balance cap 12122029
Balances with readability of 0.1 g with square weighing pan
C
1
2
5
4
6
3
Drawing Pos Description Part No.
1 Weighing pan 170 mm x 190 mm 12122048 2 Pan support 12122049 3 Pan support cap 11131029 4 Leveling foot 12122040 5 Battery-chamber cover 12122041 6 Weighing below balance cap 12122029
99Accessories and Spare Parts

26 Appendix

26.1 Conversion Table for Weight Units

Kilogram 1 kg = 1000.0 g 1 g = 0.001 kg Milligram 1 mg = 0.001 g 1 g = 1000.0 mg Microgram 1 µg = 0.000001 g 1 g = 1000000.0 µg Carat 1 ct = 0.2 g 1 g = 5.0 ct Pound 1 lb = 453.59237 g 1 g 0.00220462262184878 lb Ounce (avdp) 1 oz = 28.349523125 g 1 g 0.0352739619495804 oz Ounce (troy) 1 ozt = 31.1034768 g 1 g 0.0321507465686280 ozt Grain 1 GN = 0.06479891 g 1 g 15.4323583529414 GN Pennyweight 1 dwt = 1.55517384 g 1 g 0.643014931372560 dwt Momme 1 mom = 3.75 g 1 g 0.266666666666667 mom Mesghal 1 msg 4.6083 g 1 g 0.217 msg Tael Hong Kong 1 tlh = 37.429 g 1 g 0.0267172513291833 tlh Tael Singapore (Malaysia) Tael Taiwan 1 tlt = 37.5 g 1 g 0.0266666666666667 tlt Tola 1 tola = 11.6638038 g 1 g 0.0857353241830079 tola Baht 1 baht = 15.16 g 1 g 0.0659630606860158 baht
1 tls 37.7993641666667 g 1 g 0.0264554714621853 tls
100 Appendix
Loading...