Conventions and Symbols Used in These Operating Instructions1.1
Safety Precautions2
Overview3
Components3.1
Operation Keys3.2
Display Panel3.3
Setting up the Balance4
Unpacking and Delivery Inspection4.1
Installing the Components4.2
Selecting the Location and Leveling the Balance4.3
Selecting the Location4.3.1
Leveling the Balance4.3.2
Power Supply4.4
AC Operation4.4.1
Battery Operation4.4.2
Transporting the Balance4.5
Weighing Below the Balance4.6
Adjustment (Calibration)4.7
Adjustment with Internal Weight4.7.1
Adjustment with External Weight4.7.2
7
7
8
9
9
10
11
13
13
14
16
16
16
17
17
17
19
19
19
19
20
Weighing Made Simple5
Switching the Balance On and Off5.1
Performing a Simple Weighing5.2
Zeroing / Taring5.3
Switching Weight Units5.4
Recall / Recall Weight Value5.5
Weighing with the Weighing-in Aid5.6
Print / Transmit Data5.7
The Menu6
What is in the Menu ?6.1
Menu Operation6.2
Description of Menu Topics6.3
Main Menu6.3.1
Basic Menu6.3.2
Advanced Menu6.3.3
Interface Menu6.3.4
Application "Piece Counting"7
Application "Percent Weighing"8
Application "Check Weighing"9
Application "Statistics"10
22
22
23
23
24
24
24
24
25
25
26
27
27
27
29
32
35
37
39
41
ML Models3
Table of contents
Application "Formulation" (Net Total Formulation)11
Application "Totaling"12
Application "Dynamic Weighing"13
Application "Multiplication Factor Weighing"14
Application "Division Factor Weighing"15
Application "Routine Test"16
Application "Diagnostics"17
Repeatability Test17.1
Display Test17.2
Key Test17.3
Motor Test17.4
Balance History17.5
Calibration History17.6
Balance Information17.7
Service Provider Information17.8
Communication with Peripheral Devices18
Function PC-Direct18.1
Firmware (Software) Updates19
Operating Principle19.1
Update Procedure19.2
43
45
47
49
51
53
56
56
57
58
59
59
60
61
62
64
64
66
66
66
Error and Status Messages20
Error Messages20.1
Status Messages20.2
Cleaning and Service21
Interface Specification22
RS232C Interface22.1
MT-SICS Interface Commands and Functions22.2
Technical Data23
General Data23.1
Model-Specific Data23.2
Balances with Readability of 0.1 mg with Draft Shield23.2.1
Balances with Readability of 1 mg with Draft Shield23.2.2
Balances with Readability of 0.01 g23.2.3
Balances with Readability of 0.1 g23.2.4
Dimensions23.3
Balances with Readability of 0.1 mg with Draft Shield (235 mm)23.3.1
Balances with Readability of 1 mg with Draft Shield (235 mm)23.3.2
Balances with Readability of 1 mg with "Flex" Draft Shield (105 mm)23.3.3
Balances with Readability of 0.01 g with Square Weighing Pan and23.3.4
67
67
67
69
70
70
70
73
73
73
73
74
75
76
79
79
80
81
82
ML Models4
Table of contents
23.3.5
Accessories and Spare Parts24
Appendix25
Index26
Balances with Readability of 0.01 g / 0.1 g with Round Weighing
Pan
Balances with Readability of 0.1 g with Square Weighing Pan23.3.6
Accessories24.1
Spare Parts24.2
Conversion Table for Weight Units25.1
83
84
85
85
88
91
91
92
ML Models5
Introduction
1 Introduction
Thank you for choosing a METTLER TOLEDO balance.
The precision balances of the NewClassic line combine a large number of weighing possibilities with easy opera
tion.
These operating instructions apply to all balance models ML in the NewClassic line. However, the different models
have different charcteristics regarding equipment and performance. Special notes in the text indicate where this
makes a difference to operation.
1.1 Conventions and Symbols Used in These Operating Instructions
Key designations are indicated by double angular brackets (e.g. «»).
This symbol indicates press key briefly (less than 1.5 s).
This symbol indicates press and hold key down (longer than 1.5 s).
This symbol indicates a flashing display.
This symbol indicates an automatic sequence.
These symbols indicate safety notes and hazard warnings which, if ignored, can cause personal
danger to the user, damage to the balance or other equipment, or malfunctioning of the balance.
This symbol indicates additional information and notes. These make working with your balance
easier, as well as ensuring that you use it correctly and economically.
ML Models7
Safety Precautions
2 Safety Precautions
Always operate and use your balance only in accordance with the instructions contained in this manual. The
instructions for setting up your new balance must be strictly observed.
If the balance is not used according to these Operating Instructions, protection of the balance may be
impaired and METTLER TOLEDO assumes no liability.
It is not permitted to use the balance in explosive atmosphere of gases, steam, fog, dust and
flammable dust (hazardous environments).
For use only in dry interior rooms.
Do not use sharply pointed objects to operate the keyboard of your balance! Although your bal
ance is very ruggedly constructed, it is nevertheless a precision instrument. Treat it with corre
sponding care.
Do not open the balance: It does not contain any parts which can be maintained, repaired, or
replaced by the user. If you ever have problems with your balance, contact your METTLER
TOLEDO dealer.
Use only balance accessories and peripheral devices from METTLER TOLEDO; they are optimally
adapted to your balance.
Use only the original universal AC adapter delivered with your balance.
Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to
countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point speci
fied for electrical and electronic equipment. If you have any questions, please contact the
responsible authority or the distributor from which you purchased this device. Should this device
be passed on to other parties (for private or professional use), the content of this regulation
must also be related.
Thank you for your contribution to environmental protection.
ML Models8
3 Overview
C
0.1 mg
C
1 mg
C
1 mg
C
0.01 g
3.1 Components
Overview
ML Models9
C
0.01g / 0.1g
C
0.1 g
11
12
13
10
14
C
5142
9
36
7
8
Overview
Name and Function of Components
1Display8Glass draftshield
2Operation keys9Handle for operation of the draft-shield door
3Model sticker (with approved models only)10 Socket for AC Adapter
4Level indicator11 RS232C serial interface
5Weighing pan 12 Kensington slot for anti-theft purposes
6Draft shield element13 Slot for second interface (optional)
7Leveling foot14 Product label
3.2 Operation Keys
ML Models10
Key Functions
Application Icons
Status Icons
Weighing-in aid
Weight Value Field
Unit Field
No.KeyPress briefly (less than 1.5 s)Press and hold (longer than 1.5 s)
1
●
To change display resolution (1/10d func
no function
tion) while application is running
2
●
To navigate back (scroll up) within menu
topics or menu selections
●
Decrease (numerical) parameters within
●
To select the weighing application
●
Decrease (numerical) parameters quickly
within menu and in applications
menu and in applications
3
●
To navigate forward (scroll down) within
menu topics or menu selections
●
To select assigned application and enter
ing the parameter settings of application.
Overview
4ON//OFF
5
6
7
8
9
●
Increase (numerical) parameters within
menu and in applications
●
Switch on
●
Zero/Tare
●
Cancel and to leave menu without saving
(one step back in the menu).
●
With entries: scroll down
●
To navigate through menu topics or menu
selections
●
To toggle between unit 1, recall value (if
selected), unit 2 (if different from unit 1)
and the application unit (if any)
●
To enter or leave menu selection (from / to
menu topic)
●
To enter application parameter or switch to
next parameter
●
To store parameter
●
Printout display value
●
Default application assignment: Piece
counting
●
Increase (numerical) parameters quickly
within menu and in applications
Balances with readability of 0.1 mg, with draft shield
(235 mm)
Place the following components on the balance in the
specified order:
Note: Push the side glass door (1) back as far as will
go.
●
Bottom plate (2)
●
Draft shield element (3)
●
Pan support (4)
●
Weighing pan (5)
Balances with readability of 1 mg, with draft shield
(235 mm)
Place the following components on the balance in the
specified order:
Note: Push the side glass door (1) back as far as will
go.
●
Bottom plate (2)
●
Pan support (3)
●
Weighing pan (4)
ML Models14
C
2
3
4
1
5
C
5
4
3
2
1
C
2
1
Setting up the Balance
Balances with readability of 1 mg, with flex draft
shield (105 mm)
Place the following components on the balance in the
specified order:
a) Place bottom plate (1) (push and turn)
b) Place pan support (2) with weighing pan (3).
c) Caerfully unfold draft shield (4) and place it cor
rectly. Pay attention to the glass panels!
d) Place draft shield top (5)
Balances with readability of 10 mg with square
weighing pan and draft shield element
a) Remove the two screws (1)
b) Remove the plate (2) and retain it.
c) Place draft shield element (3) and fix it with the two
screws.
d) Place pan support (4) with weighing pan (5).
Balances with readability of 0.01 g / 0.1 g with round
weighing pan
Place the following components on the balance in the
specified order:
●
Pan support (1)
●
Weighing pan (2)
ML Models15
C
2
1
C
Setting up the Balance
Balances with readability of 0.1 g with square weigh
ing pan
Place the following components on the balance in the
specified order:
●
Pan support (1)
●
Weighing pan (2)
4.3 Selecting the Location and Leveling the Balance
Your balance is a precision instrument and will thank you for an optimum location with high accuracy and
dependability.
4.3.1 Selecting the Location
Select a stable, vibration-free position that is as horizontal as possible. The surface must be able to safely carry the
weight of a fully loaded balance.
4.3.2 Leveling the Balance
Observe ambient conditions (see Technical Data).
Avoid the following:
●
Direct sunlight
●
Powerful drafts (e.g. from fans or air conditioners)
●
Excessive temperature fluctuations
The balances have a level indicator and two or four adjustable leveling
feet to compensate for slight irregularities in the surface of the weighing
bench. The balance is exactly horizontal when the air bubble is in the
middle of the level glass.
Note: The balance should be leveled and adjusted each time it is moved
to a new location.
Balances with 2 leveling feet
Adjust the two front leveling feet appropriately until the air bubble comes
to rest exactly in the middle of the glass:
Air bubble at"12 o'clock"turn both feet clockwise
Air bubble at"3 o'clock"turn left foot clockwise, right
foot counterclockwise
ML Models16
4.4 Power Supply
Allow your balance to warm up for 30 minutes (0.1 mg models 60 minutes) to enable it to
adapt itself to the ambient conditions.
4.4.1 AC Operation
Setting up the Balance
Air bubble at"6 o'clock"turn both feet counterclockwise
Air bubble at"9 o'clock"turn left foot counterclockwise,
right foot clockwise
Balances with 4 leveling feet
a) First turn the two rear leveling feet all the way in.
b) Adjust the two front leveling feet as previously described.
c) Turn the rear leveling feet down onto the surface for extra stabilizing
safety, so the balance cannot tilt over under eccentric loads.
Your balance is supplied with an country-specific AC adapter or with a country-specific power cable. The power
supply is suitable for all line voltages in the range: 100 - 240 VAC, 50/60 Hz (for exact specifications, see section
"technical data").
First, check the local line voltage is in the range 100 - 240 VAC, 50/60 Hz and whether the
power plug fits your local power supply connection. If this is not the case, on no account connect the balance or the AC adapter to the power supply, but contact the responsible METTLER
TOLEDO dealer.
Important:
●
Before operating, check all cables for damage.
●
Guide the cables so that they cannot become damaged or interfere with the weighing
process!
●
Take care that the AC adapter cannot come into contact with liquids!
●
The power plug must be always accessible.
Connect the AC adapter to the connection socket on the back of your bal
ance (see figure) and to the power line.
4.4.2 Battery Operation
The Balance can also operates with batteries. Under normal operation conditions, the balance works independently
of the AC power line for about 8 to 15 hours (using alkaline batteries). Immediately after the AC power supply is
ML Models17
Setting up the Balance
interrupted e.g. by withdrawing the power plug or if there is a power failure, the balance switches automatically to
battery operation. Once the AC power supply is restored, the balance reverts automatically to AC operation.
Note: It is also possible to use rechargeable batteries. Charging batteries inside the balance is not possible.
battery fully When the balance is operating on its batteries, the battery symbol in the
display lights up. The number of segments that are lit is an indicator of
battery condition (3 = fully charged, 0 = discharged). When the batteries
2/3 full
are almost completely discharged, the battery symbol flashes.
1/3 full
battery empty
Inserting / Replacing Batteries
The balance must be disconnected from the power supply when carrying out all setup and
mounting work.
●
Make sure that the balance is off before removing or inserting batteries.
●
Do not place the balance on the pan support location bolt.
●
Battery Warnings: Read and follow all warnings and instructions supplied by the battery
manufacturer.
●
Do not mix different types or brands of batteries. Performance of batteries can vary very
greatly depending on the manufacturer.
●
If you don't operate the balance with batteries for an extended period, it is recommended to
remove the batteries from the balance.
Your balance uses 8 standard AA (LR6) batteries (alkaline batteries preferred)
a) Remove weighing pan, pan support and draft shield element or draft
shield "100 mm" if present.
b) Turn the balance carefully on its side.
c) Open and remove the battery-chamber cover.
d) Insert / replace the batteries with the correct polarity as shown in the
battery holder.
e) Insert and close the battery-chamber cover.
f) Turn the balance carefully to its normal position.
g) Reinstall all components in the reverse order.
ML Models18
A
Setting up the Balance
4.5 Transporting the Balance
Switch off the balance and remove the power cable and any interface cable from the balance. Refer to the notes in
Section "Selecting the location" regarding the choice of an optimal location.
Transporting Over Short Distances
For balances with a draft shield: Observe the following instructions to transport your balanceover a short distance to a new location: Never lift the balance using the glass draft shield.
The draft shield is not sufficiently fastened to the balance.
Transporting Over Long Distances
If you would like to transport or send your balance over long distances, use the complete original packaging.
4.6 Weighing Below the Balance
The balances are equipped with a hanger for carrying out weighings below the work surface (weighing below the
balance).
Attention:
●
Do not place the balance on the pan support location bolt.
a) Switch off the balance and remove the power cable and any
interface cable from the balance.
b) Remove weighing pan, pan support and draft shield element or
"Easy draft shield" if present.
c) Turn the balance carefully on its side.
d) Remove one of the caps (A) depending on the models.
e) Then turn the balance to its normal position and simply rein
stall all components in the reverse order.
4.7 Adjustment (Calibration)
To obtain accurate weighing results, the balance must be adjusted to match the gravitational
acceleration at its location. Adjusting is necessary:
●
before the balance is used for the first time.
●
at regular intervals during weighing service.
●
after a change of location.
4.7.1 Adjustment with Internal Weight
Note: On models with internal weight only (see technical data).
To obtain accurate results, the balance must be connected to the power supply for approxi
mately 30 minutes (0.1 mg balances 60 minutes) in order to reach operating temperature
before adjusting.
ML Models19
------
ADJ.DONE
0.00
g
ADJ.INT
-- Internal adjustment -
21.Jan. 2009 12:56
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
Temperature 22.5 °C
Diff 3 ppm
Adjustment done
-----------------------
-------
2000.00
g
ADJ.DONE
0.00
g
C
0.00
g
------
C
ADJ.EXT
Setting up the Balance
Sample adjustment printout using internal weight:
Requirement: To carry out this operation, in the menu topic "CAL"
(Adjustment) of advanced menu "ADJ.INT" must be selected.
a) Unload weighing pan
b) Press and hold « » to execute "Internal Adjustment".
The balance adjusts itself automatically. The adjusting is finished when
the deviation and the message "ADJ.DONE" appears briefly in the dis
play. The balance returns to the last active application and is ready for
operation.
4.7.2 Adjustment with External Weight
Note: Because of certification legislation, the approved models cannot be adjusted with an external weight.
To obtain accurate results, the balance must be connected to the power supply for approxi
mately 30 minutes (0.1 mg balances 60 minutes) in order to reach operating temperature
before adjusting.
Requirement: To carry out this operation, in the menu topic "CAL"
(Adjustment) " of advanced menu ADJ.EXT" must be selected.
a) Have required adjustment weight ready.
b) Unload weighing pan.
c) Press and hold « » to execute "External Adjustment". The required
(predefined) adjustment weight value flashes in the display.
d) Place adjustment weight in center of pan. The balance adjusts itself
automatically.
e) When "0.00 g" flashes, remove adjustment weight.
The adjusting is finished when the deviation value and the message
"ADJ.DONE" appears briefly in the display. The balance returns to the last
active application and is ready for operation.
ML Models20
Sample adjustment printout using external weight:
-- External adjustment -
21.Jan. 2009 12:56
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
Temperature 22.5 °C
Nominal 2000.00 g
Actual 1999.99 g
Diff 5 ppm
Adjustment done
Signature
.......................
-----------------------
Setting up the Balance
ML Models21
On
Off
8888888
XXXXXXX
0.00
g
SHUTOFF
On
Off
Weighing Made Simple
5 Weighing Made Simple
This section shows you how to perform simple weighings and how you can accelerate the
weighing process.
5.1 Switching the Balance On and Off
This section shows you how to perform simple weighings and how you can accelerate the weighing process.
Switching On
a) Remove any load from weighing pan.
b) Press «On».
The balance performs a display test (all segments in the display light up
briefly), "WELCOME", Software version, Maximum load and Readability
appears briefly. (Startup "FULL" mode only)
The balance is ready for weighing or for operation with the last active
application.
Switching Off
Press and hold the «Off» key until "SHUTOFF" appears on the display.
Release the key.
After it has been switched off, your balance is in the standby mode. The display shows date,
time, maximum load and readability. As your balance needs no warm-up time in the standby
mode and is immediately ready for weighing.
When Quickstart "QUICK" (Advanced menu, topic "STARTUP") is selected: If you wish to perform
a weighing, you need now only place the sample on the weighing pan and the balance immedi
ately displays the result. There is no need to switch it on with the «On/Off» key.
Note:
●
Quickstart is not possible with approved balances.
●
Standby mode is available on line powered balances only.
ML Models22
5.2 Performing a Simple Weighing
1250.38
g
0.00
g
1182.03
g
C
#
0.00
g
121.01
g
#
0.01
g
0.00
g
C
C
121.01
g
0.00
g
0.00
g
95.97
g
121
.
01
g
-
Net
Net
Net
#
C
C
a) Press «» to zero the balance.
Note: If your balance is not in the weighing mode, first press and
hold the «» key until “WEIGH“ appears in the display. Release the
key. Your balance is in the weighing mode.
b) Place weighing sample on the weighing pan.
c) Wait until the instability detector " " disappears and the stability
beep sounds.
d) Read the result.
5.3 Zeroing / Taring
Zeroing
a) Unload the balance.
Weighing Made Simple
b) Press «
» to set the balance to zero. All weight values are
measured in relation to this zero point (see menu topic
"ZERO.RNG").
Note: Use the «» zeroing key before you start with a weighing.
Taring
If you are working with a weighing container, first set the balance to zero.
a) Place empty container on the balance. The weight is displayed.
b) Press «
» to tare the balance.
"0.00 g" and "Net" appears in the display. "Net" indicates that all weight
values displayed are net values.
Note:
●
If the container is removed from the balance, the tare weight will be
shown as a negative value.
●
The tare weight remains stored until the «» key is pressed
again or the balance is switched off.
ML Models23
0.78
oz
22.00
g
*
22.00
g
M
121.01
g
0.00
g
0.00
g
M
121.01
g
0.00
g
C
C
0%
100%
Weighing Made Simple
5.4 Switching Weight Units
The «» key can be used at any time to toggle between weight unit
"UNIT 1","RECALL" value (if selected), weight unit "UNIT 2" (if different
from weight unit 2) and the application unit (if any).
5.5 Recall / Recall Weight Value
Recall stores stable weights with an absolute display value bigger than 10d.
Requirement: The function "RECALL" must be activated in the menu.
a) Load weighing sample. The display shows weight value and stores
stable value.
b) Remove weighing sample. When the weight is removed the Display
shows zero.
c) Press «». The display shows last stored stable weight value for 5
seconds together with asterisk (*) and Memory (M) symbols. After 5
seconds the display goes back to zero. This can be repeated unlimit
ed times.
Delete last weight value
As soon a new stable weight value is displayed, the old recall value
becomes replaced by the new weight value. When pressing «»,
the recall value is set to 0.
Note: If the power is switched off, the recall value is lost. The recall value
can not be printed.
5.6 Weighing with the Weighing-in Aid
The weighing-in aid is a dynamic graphic indicator which shows the
used amount of the total weighing range. You can thus recognize at a
glance whether the load on the balance approaches the maximum load.
5.7 Print / Transmit Data
Pressing the «» key transmits the weighing results over the interface
e.g. to a printer or a PC.
ML Models24
6 The Menu
6.1 What is in the Menu ?
The Menu allows you to match your balance to your specific weighing needs. In the menu you
can change the settings of your balance and activate functions. The main menu has 4 different
menus and these contains 27 different topics, each of which allows you various selection pos
sibilities. For Menu "PROTECT" see chapter "Description of menu topics" section "Main menu".
Note: See Quick Guide for the graphical overview of the menu (Menu Map) with all setting possibilities.
Menu "BASIC"
TopicDescription
DATESetting the current date.
TIMESetting the current time.
UNIT 1Specification of the 1st weight unit in which the balance should show the result.
UNIT 2Specification of the 2nd weight unit in which the balance should show the result.
KEY.BEEPSetting the key beep level.
STB.BEEPSetting the stability beep level.
RESETCall up of the factory settings.
The Menu
Menu "ADVANCE."
TopicDescription
SRV.ICONSwitching the service reminder (service icon) on or off.
ENVIRON.Matching the balance to the ambient conditions.
CALSettings for the type of adjustment (calibration).
DATE.FRMSetting the date format.
TIME.FRMPreselection of the time format.
RECALLSwitching the application "Recall" for storing stable weights on or off.
STARTUPSetting the mode which the balance powers up ("FULL" or "QUICK").
SHUTOFFSetting the time after which the balance should be switched off automatically.
B.LIGHTSetting the time after which the display backlight should be switched off automatically.
A.ZEROSwitching the automatic zero correction (Autozero) on or off.
ZERO.RNGSetting the zero limit of the zero/tare key.
ASSIGN:FSelection of assigned F key application and entering their parameter settings.
DIAGNOS.Starting a diagnostic application.
RESETCall up of the factory settings.
Menu "INT.FACE"
TopicDescription
RS232Matching the serial interface RS232C to a peripheral unit.
BAUDSetting the transfer speed of the serial interface RS232C.
BIT.PAR.Setting the character format (Bit/Parity) of the serial interface RS232C.
HD.SHKSetting the transfer protocol (Handshake) of the serial interface RS232C.
RS E.O.L.Setting the end of line format of the serial interface RS232C.
RS.CHARSetting the char set of the serial interface RS232C.
INTERVL.Selection of the time which the balance started an automatic printout.
RESETCall up of the factory settings.
ML Models25
406.2
g
ADVANCE
BASIC
TIME
DATE
ENVIRON.
STABLE
ENVIRON.
01.
2
500
01.
6
500
01.6500
The Menu
6.2 Menu Operation
In this Section you will learn how to work with the menu.
Select Menu
a) Press and hold « » to activate main menu. The first menu "BASIC"
is displayed (except menu protection is active).
b) Press «» repeatedly to change menu (Scrolling down/up «+» /
«–» keys).
c) Press «» to confirm the selection.
Note: The menu selection "BASIC", "ADVANCE." or "INT.FACE" can not be
saved. The selection "PROTECT" must be saved.
Select Menu Topic
Press «». The next menu topic appears in the display. Each time the
«» or the «+» key is pressed, the balance switches to the next menu
topic; the «–» key to the previous menu topic.
Change Settings in a Selected Menu Topic
a) Press «».The display shows the current setting in the selected
menu topic. Each time «» or «+» is pressed, the balance switch
es to the next selection; press «–» to the previous selection. After the
last selection, the first is shown again.
b) Press «», the selected setting is accepted but not yet executed.
The settings are executed only after "SAVE:YES" has been confirmed.
Change Settings in a Submenu Selection
The same procedure as for menu topics.
Input Principle of Numerical Values
a) Press «» for input of numerical values.
b) Press «» to select a digit or a value (depending on the applica
tion). The selected digit or the selected value is blinking.
c) For changing digits or values, press «+» to scroll up or «–» to scroll
down.
d) Press «
» to confirm the input.
ML Models26
SAVE:YES
SAVE:NO
The Menu
Saving Settings and Closing the Menu
a) Press and hold « » to leave menu topic.
b) Press «» to execute "SAVE:YES". Changes are saved.
c) Press «» to execute "SAVE:NO". Changes are not saved. To toggle
between "SAVE:YES" and "SAVE:NO" press «».
Cancel
For leaving menu topic or menu selection without saving press «C» (one
step back in the menu).
Note: If no entry is made within 20 seconds, the balance reverts to last active application mode. Changes are not
saved. If changes are made, the balance asks "SAVE:NO".
6.3 Description of Menu Topics
In this Section you will find information regarding the individual menu topics and the available selections.
6.3.1 Main Menu
Selecting the menu.
"BASIC"The small "BASIC" menu for simple weighing is displayed.
"ADVANCE."The extended "ADVANCE." menu for further weighing settings is dis
played.
"INT.FACE"The menu "INT.FACE" for all interface parameter settings for periph
eral devices e.g. printer is displayed.
"PROTECT"Menu protection. Protection of balance configurations against
unmeant manipulation.
"OFF"Menu protection is off. (Factory setting)
"ON"Menu protection is on. The menu BASIC, ADVANCE. and INT.FACE
are not displayed.
Note:
●
The menu selection "BASIC", "ADVANCE." or "INT.FACE" can not be saved.
●
To activate "PROTECT" "ON" or "OFF", this selection must be saved.
6.3.2 Basic Menu
"DATE" – Date
Setting the current date according to date format.
Note: A reset of the balance will not change this setting.
"TIME" – Time
Setting the current time according to time format
"+1H"Set the current time forwards by 1 hour (to adjust summer or winter
time). (Factory setting)
ML Models27
The Menu
"-1H"Set the current time backwards by 1 hour (to adjust summer or
winter time).
"SET.TIME"Enter the current time.
Note: A reset of the balance will not change this setting.
"UNIT 1" – Weight Unit 1
Depending on requirements, the balance can operate with the following units (depending on the model)
●
Only those weight units allowed by the appropriate national legislation are selectable.
●
With approved balances, this menu topic has a fixed setting and cannot be changed.
●
Conversion table for weight units see chapter Appendix.
Units:
1)
g
kg
mg
2)
3)
GramdwtPennyweight
KilogrammomMomme
MilligrammsgMesghal
ctCarattlhTael Hong Kong
lbPoundtls
4)
Tael Singapore
ozOunce (avdp)tltTael Taiwan
oztOunce (troy)tolaTola
GNGrainbahtBaht
1)
factory setting
2)
not with 0.1 mg and 1 mg balances
3)
with 0.1 mg and 1 mg balances
4)
the Malaysian tael has the same value
"UNIT 2" – Weight Unit 2
If it is required to show the weighing results in weighing mode in an additional unit, the desired second weight unit
can be selected in this menu topic (depending on the model). Units see "UNIT 1".
Note: Only those weight units allowed by the appropriate national legislation are selectable.
"KEY.BEEP" – Key Beep
This menu topic allows you to select the volume of the key beep. The according key beep is emitted during the set
ting.
"MED"Medium level (Factory setting)
"HIGH"High level
"OFF"Beep switched off
"LOW"Low level
"STB.BEEP" – Stability Beep
If the unstable symbol disappears, the stability beep becomes active. This menu topic allows you to preselect the
volume of the stability beep.
"LOW"Low level (Factory setting)
"MED"Medium level
"HIGH"High level
ML Models28
"OFF"Beep switched off
"RESET" – Reset Balance Settings
This menu topic allows you to cal-up the factory settings.
To toggle between "YES?" and "NO?" press «» (or «+» or «–»).
Note: A reset of the balance will not change the "DATE" and "TIME" settings.
6.3.3 Advanced Menu
"SRV.ICON" – Service Reminder
This menu topic allows you to switch the service reminder "
"ON"Service reminder " " switched on.You will be informed after one
Year or 8000 operating hours to call service for recalibration. This
will be indicated by the flashing service icon: " ". (Factory set
ting)
"OFF"Service reminder " " switched off.
"ENVIRON." – Environment Settings
This setting can be used to match your balance to the ambient conditions.
" on or off.
The Menu
"STD."Setting for an average working environment subject to moderate
variations in the ambient conditions. (Factory setting)
"UNSTAB."Setting for a working environment where the conditions are continu
ously changing.
"STABLE"Setting for a working environment which is practically free from
drafts and vibrations.
"CAL" – Adjustment (calibration)
In this menu topic you can preselect the function of the « » key. Your balance can be adjusted with internal or
external weights by pressing the « » key. If you have attached a printer to your balance, the data of the adjustment
(calibration) are printed out.
"ADJ.OFF"The adjustment is switched off. The « » key has no function.
"ADJ.INT"Internal adjustment: adjustment is performed at a keystroke with
the built-in weight (depending on the model, see technical data).
"ADJ.EXT"External adjustment: adjustment is performed at a keystroke with a
selectable external weight.
Note: This function is not available for approved balances.
"200.00 g"Defining the external adjustment weight: define the weight of the
external adjustment weight (in grams).
Factory setting: depends on the model.
"DATE.FRM" – Date Format
This menu topic allows you to preselect the date format.
The following date formats are available:
Display examplesPrinting examples
ML Models29
The Menu
"DD.MM.Y"01.02.0901.02.2009
"MM/DD/Y"02/01/02/01/2009
"Y-MM-DD"-02-012009-02-01
"D.MMM.Y"1.FEB.091.FEB.2009
"MMM.D.Y"FEB.1.09FEB 1 2009
Factory setting: "DD.MM.Y"
"TIME.FRM" – Time Format
This menu topic allows you to preselect the time format.
The following date formats are available:
"24:MM"15:04
"12:MM"3:04 PM
"24.MM"15.04
"12.MM"3.04 PM
Factory setting: "24:MM"
Display examples
"RECALL" – Recall
This menu topic allows you to switch the "RECALL" function on or off. When it is switched on recall stores the last
stable weight if the absolute display value was bigger than 10d.
"OFF"
"ON"
Note: The recall value is displayed with an asterisk and cannot be printed.
"STARTUP" – Startup Mode
You can set your balance such that it either immediately starts from the standby mode when you load a weight or it
must be switched on with the «ON/OFF» key after which it then performs a display test.
Note: This topic is not visible with approved balances.
"QUICK""Quickstart": The balance can be started directly from the standby
"FULL"Start with display test: You must switch on the balance with the
"RECALL" switched off (Factory setting)
"RECALL" switched on
mode and is immediately ready for weighing. You can load the
weight in the standby mode and the balance immediately shows
the current weighing result. This is the Factory settingNote: Standby mode is available on line powered balances only.
«ON/OFF» key. After it has been switched on, it performs a display
test for approx. 2 sec. in which all display elements lights up, it
shows "WELCOME", software version, maximum load and read
ability. The balance is ready for weighing.
"SHUTOFF" – Automatic Shutoff
If the automatic shutoff function is activated, the balance automatically switches itself off after a preselected time of
inactivity (i.e. with no key being pressed or changes of weight occurring etc.) and is switched to the standby
mode.
"A.OFF:10' " Automatic shutoff after 10 minutes of inactivity. (Factory setting)
ML Models30
The Menu
"A.OFF – "Automatic shutoff not activated.
"A.OFF:2' " Automatic shutoff after 2 minutes of inactivity.
"A.OFF:5' " Automatic shutoff after 5 minutes of inactivity.
"B.LIGHT" – Backlight
Under this menu topic, the display backlight can be switched off automatically. If the automatic switch-off is acti
vated, the backlight will turn off automatically after the selected period of inactivity has elapsed. The backlight is
reactivated when a key is pressed or the weight is changed.
"B.L. ON" Backlight is always on. (Factory setting)
"B.L. OFF" Backlight is always off.
"B.L. 30" " Automatic switch-off after 30 seconds inactivity.
"B.L. 1' " Automatic switch-off after 1 minute inactivity.
"B.L. 2' " Automatic switch-off after 2 minutes inactivity.
"B.L. 5' " Automatic switch-off after 5 minutes inactivity.
"A.ZERO" – Automatic Zero Setting
This menu topic allows you to switch the automatic zero setting on or off.
“ON“"A.ZERO" switched on (factory setting). The automatic zero setting
continuously corrects possible variations in the zero point that
might be caused through small amounts of contamination on the
weighing pan.
"OFF""A.ZERO" switched off. The zero point is not automatically correct
ed. This setting is advantageous for special applications (e.g.
evaporation measurements).
Note: With approved balances, this setting is not available.
"ZERO.RNG" – Zero Range
This menu topic allows you to set a zero limit for the «
«» key will execute a zero. Above this limit the «» key will execute a tare.
"WEIGHT"
"PERCENT"
"ASSIGN:F" – Assign Application Key F
At this menu topic you can assign an application to the «F» key. The following applications are available (depend
ing on the model):
"COUNT"Piece counting (Factory setting)
"PERCENT"Percent weighing
"CHECK.W"Checkweighing
"STAT."Statistics
"FORMULA." Formulation / Net-Total
"TOTAL." Totaling
To set the upper limit of the zeroing range as weight in the defini
tion unit of the balance. (Factory setting)
"21g"
n.a. (To set the upper limit of the zeroing range as a percent of the
total range of the balance).
"0%"
» key. Up to and including this limit the
ML Models31
The Menu
"DYNAMIC" Dynamic weighing
"FACTOR.M" Multiplication factor
"FACTOR.D" Division factor
"R.TEST" Routine test
"DIAGNOS." – Diagnostics Application
At this menu topic you can start a diagnostic application. For more information see chapter application "Diagnos
tics".
The following diagnostics are available:
"REPEAT.T"Repeatability test (models with internal weights only)
"DISPLAY"Display test
"KEYPAD.T"Key test
"CAL.MOT.T"Motor test (models with internal weights only)
"BAL.HIST" Balance history
"CAL.HIST" Calibration history
"BAL.INFO" Balance information
"SRV.PROV" Service provider information
"RESET" – Reset Balance Settings
This menu topic allows you to cal-up the factory settings.
To toggle between "YES?" and "NO?" press «
Note: A reset of the balance will not change the "DATE" and "TIME" settings.
» (or «+» or «–»).
6.3.4 Interface Menu
"RS232" – RS232C Interface
At this menu topic you can select the peripheral device connected to the RS232C interface and specify how the
data is transmitted.
"PRINTER"Connection to a printer. (Factory setting)
"PRT.STAB"If the «
printed. (Factory setting)
"PRT.AUTO"Every stable weight value will be printed, without pressing the «»
key.
"PRT.ALL"If the «» key is pressed, the weight value will be printed regard
less of stability.
"PC-DIR."Connection to a PC: the balance can send data (as a Keyboard) to
the PC used for PC applications e.g. Excel.
Note: The balance sends the weight value without the unit to the
PC.
"PRT.STAB"If the «
followed by an enter. (Factory setting)
"PRT.AUTO"Every stable weight value will be sent followed by an enter, without
pressing the «» key.
» key is pressed, the next stable weight value will be
» key is pressed, the next stable weight value will be sent
ML Models32
The Menu
"PRT.ALL"If the «» key is pressed, the weight value will be sent followed
by an enter regardless of stability.
"HOST"Connection to a PC, Barcode Reader etc.: the balance can send
data to the PC and receive commands or data from the PC.
Note: The balance sends the complete MT-SICS answer to the PC
(see chapter "MT-SICS Interface Commands and Functions".
"SND.OFF"Send mode switched off.(Factory setting)
"SND.STB"If the «
» key is pressed, the next stable weight value will be
sent.
"SND.CONT"All weight value updates will be sent regardless of stability, without
pressing the «» key.
"SND.AUTO"Every stable weight value will be sent, without pressing the «
key.
"SND.ALL"If the «» key is pressed, the weight value will be sent regardless
of stability.
"2.DISP"Connection of an optional auxiliary display unit.
Note: The transmission parameters cannot be selected. Settings are
automatically set.
"BAUD" – Baud rate RS232C
This menu topic allows you to match the data transmission to different serial RS232C receivers. The baud rate
(data transfer rate) determines the speed of transmission via the serial interface. For problem-free data transmis
sion the sending and receiving devices must be set at the same value.
The following settings are available:
600 bd, 1200 bd, 2400 bd, 4800 bd, 9600 bd, 19200 and 38400 bd. (default: 9600 bd)
Note:
●
Not visible for 2nd display.
»
●
Each device has separate settings.
"BIT.PAR." – Bit/Parity RS232C
At this menu topic you can set the character format for the attached RS232C serial peripheral device.
"8/NO"8 data bits/no parity (Factory setting)
"7/NO"7 data bits/no parity
"7/EVEN"7 data bits/even parity
"7/ODD"7 data bits/odd parity
Note:
●
Not visible for 2nd display.
●
Each device has separate settings.
"HD.SHK" – Handshake RS232C
This menu topic allows you to match the data transmission to different RS232C serial receivers.
At this menu topic you can set the "End of Line" character of the transmitted data to different RS232C serial
receivers.
"CR LF"<CR><LF> Carriage Return followed by Line feed (ASCII-Codes
013+010) (Factory setting)
"CR"<CR> Carriage Return (ASCII-Code 013)
"LF"<LF> Line feed (ASCII-Code 010)
Note:
●
Not visible for 2nd display.
●
Each device has separate settings.
"RS.CHAR" – Char Set RS232C
At this menu topic you can set the "Character Set" of the transmitted data to different RS232C serial receivers.
"IBM.DOS"Char Set IBM/DOS (Factory setting)
"ANSI.WIN"Char Set ANSI/WINDOWS
Note:
●
Not visible for 2nd display.
●
Each device has separate settings.
"INTERVL." – Print Key Simulation
At this menu topic you can activate a simulation of the «
» key. "INTERVL." simulates a print key press every x
seconds.
Range:0 to 65535 seconds
0 sec:disables the print key simulation
Factory setting: 0 sec
Note: The executed action is according to the configuration of the print key. (see interface setting)
"RESET" – Reset Balance Settings
This menu topic allows you to cal-up the factory settings.
To toggle between "YES?" and "NO?" press «
» (or «+» or «–»).
Note: A reset of the balance will not change the "DATE" and "TIME" settings.
ML Models34
7 Application "Piece Counting"
COUNT
C
A
B
A
B
C
FIX 10
PCS
C
10
PCS
------
FIX 10
PCS
A
A
0.00
g
1
PCS
2.74
g
1
C
------
PCS.WGT
B
B
The "Piece Counting" application allows you to determine the number of pieces put on the
weighing pan.
Requirement: The function "COUNT" must be assigned to the «F» key (see advanced menu top
ic "ASSIGN:F", factory setting).
Activate function "COUNT" by pressing and holding the «F» key.
Piece Counting first requires the setting of a reference weight, there
are 3 possibilities:
Setting the reference by multiple pieces.
Setting the reference for 1 piece in weighing mode.
Setting the reference for 1 piece in manual mode.
Setting possibility
Setting the reference by multiple pieces
Application "Piece Counting"
a) Select a number of reference pieces by scrolling with «». Possible
numbers are 5, 10, 20 and 50.
b) Press «» to tare. If needed: place empty container on the
weighing pan and tare again.
c) Add the selected number of reference pieces to container.
d) Press «» to confirm.
Setting possibility
Setting the reference for one piece in weighing mode
a) Select "PCS.WGT" by scrolling with «».
b) Press «» to tare. If needed: place empty container on the
weighing pan and tare again.
c) Add one reference piece to container. The weight of one piece is dis
played.
d) Press «
» to confirm.
ML Models35
2.68
g
*
0
PCS
2.74
g
*
PCS.WGT
C
C
2
.
74
g
2
.
74
g
,
812.7
g
75
PCS
Application "Piece Counting"
Setting possibility
Setting the reference for one piece in manual mode
a) Select "PCS.WGT" by scrolling with «».
b) Press «
c) Enter the final reference one piece weight by scrolling up («+» key)
or down («-» key). Speed up by press and hold.
d) Press «» to confirm.
Switching between manual mode and weighing mode
Press «» to switch between manual and weighing mode.
Note: By switching from weighing mode to manual mode the weight val
ue will be transferred and can be changed manually.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press «C»
to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for piece counting.
Switching between piece count and weight display.
You can use the «» key at any time to switch the display between
piece display, weighing unit "UNIT 1", "RECALL" value (if activated) and
weighing unit "UNIT 2" (if different from "UNIT 1").
» to confirm.
Note:
●
The "RECALL" value is displayed with an asterisk (*) and icon "M" and can not be printed.
●
Take into account minimum values: min. reference weight = 10d (10 digits), min. piece weight = 1d (1 digit)!
●
The current reference weight remains stored until the reference setting is changed.
ML Models36
8 Application "Percent Weighing"
PERCENT
A
B
A
B
0.0
%
175.28
g
*
A
SET 100
%
A
I00.0
%
175.28
g
SET 100
%
C
0.00
g
B
B
175.28
g
175.28
g
,
The "Percent Weighing" application allows you to check a sample weight as percentage to a
reference target weight.
Requirement: The function "PERCENT" must be assigned to the «F» key (see advanced menu
topic "ASSIGN:F".
Activate function percent weighing "PERCENT" by pressing and holding
the «F» key.
Percent Weighing first requires the setting of a reference weight that
should corresponds to 100%, there are 2 possibilities:
Setting the reference in manual mode (enter 100%).
Setting the reference in weighing mode (weigh 100%).
Setting possibility
Setting the reference by manual mode (enter 100%)
Application "Percent Weighing"
a) Press «» to activate manual mode.
b) Select the reference target weight (100%) by scrolling up («+» key)
or down («–» key). Speed up by press and hold.
c) Press «» to confirm.
Setting possibility
Setting the reference by weighing mode (weigh 100%)
a) Press «» to tare the balance and to activate the weighing
mode. If needed: place empty container on the weighing pan and
tare again.
b) Load the reference weight (100%).
Note: Reference weight must be at least +/- 10d.
c) Press «» to confirm.
Switching between manual mode and weighing mode
Press «» to switch between manual and weighing mode.
Note: By switching from weighing mode to manual mode the weight val
ue will be transferred and can be changed manually.
Note: If without any key press within 60 seconds, the balance returns to the previous active application.
On completion of the weighing-in procedure, your balance is ready for percent weighing.
ML Models37
,
128.25
g
73.17
%
128250
mg
128.25
g
M
Application "Percent Weighing"
Switching between percent and weight display
You can use the «» key at any time to switch the display between
percent display, weighing unit "UNIT 1", "RECALL" value (if activated)
and weighing unit "UNIT 2" (if different from UNIT 1).
Note:
●
The recall value is displayed with an asterisk (*) as well as icon "M"
and can not be printed.
●
The current set weight remains stored until it is redetermined.
ML Models38
9 Application "Check Weighing"
CHECK
2
1A
1B
1A
1B
2
SET NOM .
10.00
g
20.28
g
1A
1A
1B
SET NOM .
0.00
g
20.28
g
C
1B
10.00
g
10.00
g
The "Check weighing" application allows you to check the deviation of a sample weight within
a tolerance limit to a reference target weight.
Requirement: The function "CHECK.W" must be assigned to the «F» key (see advanced menu
topic "ASSIGN:F".
Activate function "CHECK.W" by pressing and holding the «F» key.
Step 1: Check Weighing first requires the setting of a reference
weight that should corresponds to the nominal weight, there are 2
possibilities:
Setting the reference in manual mode (enter nominal weight).
Setting the reference in weighing mode (weigh nominal weight).
Step 2: Check weighing needs the upper and lower limits:
Setting the upper and lower limits.
Application "Check Weighing"
Setting possibility:
Setting the reference in manual mode (enter nominal weight)
a) Press «
b) Select the reference target weight by scrolling up («+» key) or down
(«–» key). Speed up by press and hold.
c) Press «
Setting possibility:
Setting the reference in weighing mode (weigh nominal weight)
a) Press «
mode. If needed: place empty container on the weighing pan and
tare again.
b) Load the nominal weight.
c) Press «» to confirm the nominal weight.
Switching between manual mode and weighing mode
Press «» to switch between manual mode and weighing mode.
» to activate manual mode.
» to confirm the nominal weight.
» to tare the balance and to activate the weighing
Note: By switching from weighing mode to manual mode the weight val
ue will be transferred and can be changed manually.
ML Models39
SET.T+
/-
%
2.50
%
20.28
g
2
2
20.15
g
Application "Check Weighing"
Step 2:
Setting the upper and lower limits (in percentage):
a) Press «» to start setting.
b) Press «» to confirm the default limit of +/- 2.5 % or enter the limit
value by scrolling up («+» key) or down («–» key). Press «» to
confirm the limits.
Note: Press «» to switch between "UNIT 1" and Unit "%".
Note:
●
If without any key press within 60 seconds, the balance returns to the previous active application. Press «C» to
cancel and returns to the previous active application.
●
The nominal weight must be at least 10 digit.
On completion of the setting procedure, your balance is ready for checkweighing.
Weighing-in-Aid
The Weighing-in-Aid helps you quickly determine the position of the sam
ple weight regarding the tolerance.
1 Lower limit
2 Target weight
3 Upper limit
ML Models40
10 Application "Statistics"
CLR. M:NO
STAT
CLEARED
46.36
g
0.00
g
- 1 -
C
999
C
0.00
g
5
N
X
50.530
g
S.DEV
3.961
g
S.REL
7.84
%
MIN
46.36
g
MAX
55.81
g
The "Statistics" application allows you to generate statistics of weighing values. 1 to 999 val
ues are possible. Requirement: The function "STAT." must be assigned to the «F» key (see
advanced menu topic "ASSIGN:F"). Connect a printer or a PC if present.
a) Activate function "STAT." by pressing and holding the «F» key.
b) To continue the last statistics press «». For a new statistical eval
uation press «» (or «+», «–») to select "CLR.M:YES" and press
«» to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the memory
clear question will be not displayed.
Weighing the first sample weight:
a) Press «» to zero/tare the balance if needed.
b) Load the first sample weight.
c) Press «». The display shows the sample count "- 1 -" and the
current weight is stored as sample and the weight is printed out.
Note: When the sample counter is displayed you may press «C» to
undo (drop) this sample.
Application "Statistics"
d) Unload the first sample weight.
Weighing further sample weights:
The same procedure as for the first sample weight.
●
1...999 samples are possible.
●
The next value will be accepted if the sample weight is in the range
70% –130% of the current average value. "OUT OF RANGE" will be
displayed if the sample is not accepted.
Results:
Press «», the results are displayed and printed.
Displayed results:
a) Press «» to show the next
statistical value.
b) Press «C» to cancel displaying
results and to continue weigh
ing next sample.
number of samples
average
standard deviation
relative standard deviation
lowest value (minimum)
highest value (maximum)
0.5 seconds
ML Models41
DIFF
9.45
g
SUM
252.65
g
------ Statistics ------
21.Jan. 2009 12:56
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
-----------------------1 46.36 g
2 55.81 g
3 47.49 g
4 53.28 g
5 49.71 g
n 5
x 50.530 g
s dev 3.961 g
s rel 7.84 g
Min. 46.36 g
Max. 55.81 g
Diff 9.45 g
Sum 252.65 g
-----------------------
Application "Statistics"
Printout:
different between the minimum
and the maximum
sum of all values
ML Models42
CLR.
M:NO
FORMULA
CLEARED
#
-T1-
0.00
g
Net
10.33
g
C
0.00
g
#
Net
Net
C
Application "Formulation" (Net Total Formulation)
11 Application "Formulation" (Net Total Formulation)
The "Formulation" (Net Total) application allows you to
●
weigh in (add and store) up to 999 individual component weights and displays the total. If
a printer is connected, the component weights are printed individually and as a total.
●
tare/pre-tare and store up to 999 container weights and displays the total. If a printer is con
nected, the tare weights are printed out individually and as a total.
Requirement: The function "FORMULA." must be assigned to the «F» key (see advanced menu
topic "ASSIGN:F"). Connect a printer or a PC if present.
a) Activate function formulation "FORMULA." by pressing and holding
the «F» key.
b) Press «» to continue formulation weighing. For a new formulation
press «» (or «+» or «-») to select "CLR.M:YES" and press «»
to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the
memory clear question will be not displayed.
Tare container (if used):
a) Press «» to zero or tare the balance if needed.
b) Place the empty container on the weighing pan.
c) Press «» . The container is tared and the tare count "- T1 -"
is displayed and the tare weight is printed.
Note: If you pre-tare via MT-SICS (e.g. bar code reader) "- PT1 -" is dis
played.
Weighing the first component weight:
a) Load the first component weight.
b) Press «». The display briefly shows the component count "- 1 -" ,
the current weight is stored as sample and the component weight is
printed. The display is set back to zero.
Weighing further component weights:
The same procedure as for the first component weight with the same or
new container).
●
1...999 sample values are possible.
●
max 999 tare values are possible.
●
max 999 pre-tare values are possible.
ML Models43
Results:
Press «», the results are displayed and printed.
8
N
T.TOTAL
452
.
76
g
G.TOTAL
546
.
79
g
N.TOTAL
94
.
03
g
------ Formulation -----
21.Jan. 2009 12:56
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
-----------------------1 T 10.33 g
1 N 8.85 g
2 N 9.23 g
2 T 10.84 g
3 N 7.43 g
.
.
n 999 g
T Total 452.76 g
G Total 546.79 g
N Total 94.03 g
-----------------------
Application "Formulation" (Net Total Formulation)
Displayed results:
a) Press «» to show the next
statistical value.
b) Press «C» to cancel displaying
results and to continue weigh
ing next component.
Printout:
0.5 seconds
number of samples
sum of all tare values (T and
PT)
sum of all component gross
weight values
sum of all component net weight
values
ML Models44
12 Application "Totaling"
CLR.M:NO
TOTAL
CLEARED
0.00
g
46.36
g
- 1 -
999
0.00
g
0.00
g
C
C
879
N
TOTAL
8789
.
79
g
The "TOTALING" application allows you to weigh in different samples to add their weight values
and to totalize them. 1 to 999 samples are possible.
Requirement: The function "TOTAL." must be assigned to the «F» key (see advanced menu topic
"ASSIGN:F").
a) Activate function "TOTAL." by pressing and holding the «F» key.
b) For a new totaling evaluation press «» (or «+» or «–») to enter
Application "Totaling"
"CLR.M:YES" and press «» to clear the memory.
Note: If the memory is already cleared (sample counter is 0) the
memory clear question will be not displayed.
c) Press «
Weighing in the sample weight:
a) If using a container: place empty container on the weighing pan and
press «
b) Load the first sample weight.
c) Press «». The display shows the sample count "- 1 -" and the
current weight is stored.
Note: When the sample counter is displayed you may press «C» to
undo (drop) this sample.
d) Unload the first sample weight. The display shows zero.
Weighing in further sample weights:
The same procedure as for the first sample weight.
●
1...999 samples are possible.
Results:
Press «», the results are displayed and printed.
» to zero or tare the balance.
» to zero or tare the balance.
Displayed results:
a) Press «» briefly to show the
totalized value.
b) Press «C» briefly to cancel.
ML Models45
number of samples
totalized value
0.5 seconds
------- Totaling -------
21.Jan. 2009 12:56
METTLER TOLEDO
Balance Type ML1602
SNR 1234567890
-----------------------1 46.36 g
2 55.81 g
3 47.49 g
4 53.28 g
5 49.71 g
6 53.93 g
.
.
.
n 879
Total 8789.79 g
-----------------------
Application "Totaling"
Printout:
ML Models46
13 Application "Dynamic Weighing"
DYNAMIC
MOD.AUTO
MOD. MAN
1
TIME: 3
TIME: 5
2
The "Dynamic Weighing" application allows you to determine the weights of unstable samples
or to determine weights under unstable ambient conditions. The balance calculates the weight
as the average of a number of weighing operations over a defined time.
Requirement: The function "DYNAMIC" must be assigned to the «F» key (see advanced menu
topic "ASSIGN:F").
Note: "Switching Units" and "RECALL" Functions are not available in this Application.
Activate function "DYNAMIC" by pressing and holding the «F» key.
Setting "Auto Start" or "Manual Start":
a) Press «» to select the mode:
– "Auto Start ""MOD.AUTO" (default value). The weighing starts
automatically on relative stability. However, the weighing sample
must weigh at least 5 grams. For weighing samples below 5 g the
weighing must be started manually.
– "Manual Start" "MOD. MAN"
Application "Dynamic Weighing"
b) Press «» to confirm the selection.
Setting the weighing time:
a) Press «» to select one of the available time intervals: 3 (default
value), 5, 10, 20, 60 and 120 seconds.
b) Press «» to confirm the selected time interval.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press «C»
to cancel and returns to the previous active application.
Your balance is now ready for dynamic weighing:
ML Models47
13.45
g
M.START
TIME: 5
C
TIME: 1
13.39
g
C
Application "Dynamic Weighing"
a) Press «» to zero if needed.
b) If using a container: place empty container on weighing pan and
press «» to tare the balance.
c) Load sample weight.
d) – If you have selected function "Manual Start" "M.START", press
«» to start the weighing.
– If you have selected function "Auto Start" "A.START", the weighing
starts automatically on relative stability. For weighing samples below
5 g the weighing must be started manually by pressing «».
e) Read off result. The result of the dynamic weighing is displayed with
an asterisk (* = calculated value).
f) Unload sample weight.
g) "Manual Start" only, press «
» to zero and go back to
"M.START".
Note:
●
The remaining weighing time (in seconds) is displayed continuously.
You can cancel the countdown by pressing «C».
●
The weight value remains in the display until the sample weight is
removed from weighing pan ("Auto Start" only) or «
pressed.
» is
ML Models48
FACTOR.M
SET.F.MUL
001.0000
009.8
100
1
SET.STEP
001.0
005.0
2
Application "Multiplication Factor Weighing"
14 Application "Multiplication Factor Weighing"
The "Multiplication Factor Weighing" application allows you to multiply the weight value (in
grams) by a predefined factor (result = factor * weight) and have it calculated to a predefined
number of decimal places.
Requirement: The function "FACTOR.M" must be assigned to the «F» key (see advanced menu
topic "ASSIGN:F").
Activate function "FACTOR.M" by pressing and holding the «F» key.
Setting the factor value:
a) Press «
default value or the factor that was saved most recently.
b) Press «
c) For changing digits, press «+» to scroll up or «–» to scroll down.
d) Press «» to confirm the selected factor (no automatic accep
tance).
Note: Zero for multiplication factor value is outside the allowed range, the
error message "FACTOR OUT OF RANGE" will be displayed.
Setting the step value:
"SET.STEP" appears in the display, and the program changes automati
cally to allow the display increments to be entered. The smallest possible
display increment appears as default value, or the last value that was
saved.
a) Press «» to execute "SET.STEP".
b) Press «» to select a digit. The selected digit is blinking.
c) For changing digits, press «+» to scroll up or «–» to scroll down.
d) Press «
Note: The allowed range for the step depends on the factor and the reso
lution of the balance. If it is outside the allowed range the error message
"STEP OUT OF RANGE" will be displayed.
» to execute "SET.F.MUL". Either the factor 1 appears as
» to select a digit. The selected digit is blinking.
» to confirm the selected step (no automatic acceptance).
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press «C»
to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for multiplication factor weighing.
ML Models49
C
9810
1000.02
g
0
C
Application "Multiplication Factor Weighing"
Weighing procedure
a) Press «» to zero/tare.
b) Load sample weight on weighing pan.
c) Read the result. The appropriate calculation is then made using the
weight of sample and the selected factor, the result being displayed
with the selected display step.
Note: No units are displayed.
d) Unload sample weight.
Toggling between displaying the calculated value and the measured
weight:
You can use the «» key to toggle between the calculated Value,
weight value "UNIT 1", "RECALL" value (if selected) and weight value
"UNIT 2" (if different from "UNIT 1").
ML Models50
FACTOR.D
SET.F.DIV
001.0000
009.8
100
1
SET.STEP
0.0001
0.0050
2
Application "Division Factor Weighing"
15 Application "Division Factor Weighing"
The "Division Factor Weighing" divide a predefined factor by the weight value (in grams)
(result = factor / weight) and have it rounded to a predefined number of decimal places.
Requirement: The function "FACTOR.D" must be assigned to the «F» key (see advanced menu
topic "ASSIGN:F").
Activate function "FACTOR.D" by pressing and holding the «F» key.
Setting the Factor Value:
a) Press «» to execute "SET.F.DIV". Either the factor 1 appears as
default value or the factor that was saved most recently.
b) Press «» to select a digit. The selected digit is blinking.
c) For changing digits, press «+» key to scroll up or «–» to scroll down.
d) Press «» briefly to confirm the selected factor (no automatic
acceptance).
Note: Zero for division factor value is outside the allowed range, the error
message "FACTOR OUT OF RANGE" will be displayed.
Setting the step value:
"SET.STEP" appears in the display, and the program changes automati
cally to allow the display increments to be entered. The smallest possible
display increment appears as default value, or the last value that was
saved.
a) Press «
b) Press «
c) For changing digits, press «+» to scroll up or «–» to scroll down.
d) Press «» to confirm the selected step (no automatic acceptance).
Note: The allowed range for the step depends on the factor and the reso
lution of the balance. If it is outside the allowed range the error message
"STEP OUT OF RANGE" will be displayed.
Note: If without any key press within 60 seconds, the balance returns to the previous active application. Press «C»
to cancel and returns to the previous active application.
On completion of the setting procedure, your balance is ready for division factor weighing.
» to execute "SET.STEP".
» to select a digit. The selected digit is blinking.
ML Models51
C
0.010
1000.02
g
0.000
C
Application "Division Factor Weighing"
Weighing procedure
a) Press «» to zero/tare.
b) Load sample weight on weighing pan.
c) Read the result. The appropriate calculation is then made using the
weight of sample and the selected factor, the result being displayed
with the selected display step.
Note: No units are displayed.
d) Unload sample weight.
Toggling between displaying the calculated value and the measured
weight:
You can use the «» key to toggle between the calculated Value,
weight value "UNIT 1", "RECALL" value (if selected) and weight value
"UNIT 2" (if different from "UNIT 1").
ML Models52
16 Application "Routine Test"
R. TEST
TARE:NO
WGT
TEST.
5000.00
g
LMT
CTRL.
2.50
g
The "Routine Test" application allows you to determine the sensitivity of the balance. More
about periodic sensitivity tests (routine tests) see: GWP® (Good Weighing Practice) on
www.mt.com/gwp.
GWP gives clear recommendation for routine testing:
●
how should I test my balance?
●
how often?
●
where can I reduce efforts?
More about test weights see www.mt.com/weights.
Requirement:
●
The function "R.TEST" must be assigned to «F» key (see advanced menu topic "ASSIGN:F").
●
It is recommended to connect a printer or a PC to the balance for showing the results.
a) Activate function "R.TEST" by pressing and holding the assigned «F»
key.
b) Select "TARE:NO" (no tare weight used).
If a tare weight is used during the test select "TARE:YES" (use a tare
weight). To toggle between "TARE:YES" and "TARE:NO" use «»
(or «+» or «–»).
Application "Routine Test"
c) Press «
Note:
●
It is recommended to test the sensitivity without tare load. (factory
setting "TARE:NO").
●
If using tare: Make sure that tare weight plus test weight is not
exceeding max. load.
Setting the reference test weight value
The default value of the test weight: Next smaller OIML weight than the
maximum load of your balance according to the GWP® recommenda
tion.
a) For changing the value, press «+» to scroll up or «–» to scroll down.
Progressing speed by press and hold.
b) Press «» to confirm the value.
Setting the Control Limit
The default value of the control limit:
Test weight x weighing process tolerance / 2
Example: 5000 g x 0.1% / 2 = 2.50 g.
a) For changing the value, press «+» to scroll up or «–» to scroll down.
Progressing speed by press and hold.
» to confirm the selection.
b) Press «» to confirm the value.
ML Models53
LMT
WARN.
1.25
g
PRESS EN
PRESS EN
5000.00
g
----- Routine Test -----
21.Jan. 2009 12:56
METTLER TOLEDO
Balance Type ML6001
SNR 1234567890
Sensitivity:
Test weight 5000.00 g
Value 5000.11 g
Warning L. 1.25 g
Control L. 2.50 g
Warning L. OK
Control L. OK
Signature
........................
------------------------
Application "Routine Test"
Setting the Warning Limit
The default value of the warning limit:
Warning limit = control limit / safety factor
Example: 2.5 g / 2 = 1.25 g.
a) For changing the value, press «+» to scroll up or «–» to scroll down.
Progressing speed by press and hold.
b) Press «» to confirm the value.
Note: The default values of control limit and the warning limit are evaluated according the GWP recommendation.
These are based under the assumption that the weighing process tolerance is 0.1% and the safety factor is 2.
On completion of the setting procedure, your balance is ready for the
routine test procedure.
Note: The test weight must be acclimatized to the ambient temperature
of the balance.
a) Press «» to start the test.
b) Follow the instructions on the display. If the test weight value is
flashing: Load the test weight (displayed value).
The printout starts after the weighing pan is unloaded.
Exit the current test procedure:
Press and hold «
» for executing a new application.
Printout:
What if Warning Limit or Control Limit are "FAILED"?
The "SOP for Periodic Sensitivity Tests (Routine Tests)" provides information about measures when routine tests
fail. Find a download version of these SOPs on www.mt.com/gwp, link "GWP® The Program / Routine
Operation".
ML Models54
Content of SOP:
●
Preparation
●
Test procedure
●
Evaluation
●
Deviation
•
If Warning Limit "FAILED"
•
If Control Limit "FAILED"
Application "Routine Test"
ML Models55
Application "Diagnostics"
17 Application "Diagnostics"
The "Diagnostics" application allows you to carry out predefined diagnostics tests and to view
or print predefined sets of balance information. This diagnostics tool helps you find errors faster
and more efficiently.
Requirement: A printer or a PC is connected to the balance for showing the results.
a) Activate "ADVANCE." menu. (See section menu operation)
b) Activate function "DIAGNOS." by pressing «
c) Use «
» to select appropriate tests.
».
17.1 Repeatability Test
The repeatability test allows you to repeat tests with internal weight for a given number of times.
Note: On models with internal weights only.
a) Press «» to activate repeatability test "REPEAT.T". "R.TST. 10" appears on the Display.
b) Enter the number of times (blinking) by pressing «+» or «–». Possible values are 5, 10 (default), 20, 50, 100
times.
c) Press «» to start the test. The message "RUNNING REPEAT TEST" is displayed till the tests are completed.
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
ML Models56
Sample Printout:
-- Repeatability Test --
21.Jan. 2009 11:34
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
SW V1.00
Temperature 21.3 °C
No. of tests 10
------------------------
1. Value 2813.00 g
1. Time 00:00:00
1. Temp 21.3 °C
2. Value 2813.01 g
2. Time 00:00:04
2. Temp 21.3 °C
.
.
.
-----------------------Max. 2813.01 g
Min. 2813.00 g
x 2813.005 g
s Dev 0.004 g
Max Temp 21.3 °C
Min Temp 21.3 °C
Mean Temp 21.3 °C
Total Time 00:00:44
------------------------
Application "Diagnostics"
Repeatability test is a tool to do functional check with the balance. It may be performed:
●
To check function of balance
•
during installation to store print out with installation documents.
•
after preventative maintenance to store print out with installation maintenance report.
•
when remarkable decrease of weighing performance occurs, so that you can email/fax print out to service
support provider for diagnose purposes.
●
To develop the optimal environment settings (see menu topic "ENVIRON.").
Measure the time you need to perform repeatability test with each "STABLE", "STD." and "UNSTAB." setting. The
setting with the fastest total time suits best for the existing environmental conditions.
17.2 Display Test
The display test allows you to test the display of the balance.
a) Press «
All possible segments and icons on the display will illuminate.
b) Press «» to print the test information.
ML Models57
Examples:
» to start "DISPLAY".
----- Display Test -----
21.Jan. 2009 11:34
METTLER TOLEDO
Balance Type ML204
SNR 1234567890
SW V1.00
Display Test DONE
------------------------
Application "Diagnostics"
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Printout:
17.3 Key Test
The key test allows you to test the keys of the balance.
a) Press «» to start "KEYPAD.T".
b) The message "KEY TEST - PRESS KEY TO BE TESTED" is displayed scrolling during the duration of the key test.
Press every Key briefly. Each press of a key beeps and echoes with "OK” on the display.
c) Second press «C» key to print the test information. The test procedure will be cancelled and the balance will
return to the topic "DIAGNOS.". If a key has not been tested before printing, then the test results will be indicat
ed with a "----" line.
Sample Information Displayed:
KeyDisplay
« »1/10.D.OK
«»PRINT.OK
«–»MINUS.OK
«+»PLUS OK
«»TOGGL.OK
«»ENTER.OK
«C»C OK
«»0/T OK
ML Models58
Sample Printout:
------- Key Test -------
21.Jan. 2009 11:34
METTLER TOLEDO
Balance Type ML204
SNR 1234567890
SW V1.00
1/10 d Key OK
Print Key OK
Minus Key OK
Plus Key OK
Toggle Key OK
Enter Key OK
Zero/Tare Key OK
Cancel Key OK
------------------------
------ Motor Test ------
21.Jan. 2009 11:34
METTLER TOLEDO
Balance Type ML204
SNR 1234567890
SW V1.00
Motor Test OK
------------------------
17.4 Motor Test
Application "Diagnostics"
The motor test allows you to test the calibration motor of the balance.
Note: On models with internal weight only.
a) Press «» to start "CAL.MOT.T".
"RUNNING" is displayed during the Motor Test. A motor test is deemed successful when all the motor positions
have been successfully tested. At the end of the test, the test information will be printed.
b) Press «» for printout.
c) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Printout:
17.5 Balance History
The balance history function allows you to view and print the history of the balance.
a) Press «» to start "BAL.HIST" .
b) Press «» to scroll forward through the displayed list of balance history information.
c) Press «
ML Models59
» for printout.
--- Statistical Info ---
21.Jan. 2009 11:34
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
SW V1.00
-----------------------Operating time 18d 23h
Total weight loaded
115.7191 kg
Number of weighings
1255
Number of key presses
4931
Motor movements
1012
Backlight operating time
18d 4h
Next service due date
01.01.2010
------------------------
Application "Diagnostics"
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample Information Displayed:
InformationDisplay
Operation Time (year:day:hour)00:018:04
Total load kg (t)115.7191 kg
Number of weighings1255
Number of key pressed4931
Number of motor movements1012
Backlight time (year:day:hour)00:018:04
Next service due date01:01:10
Sample Printout:
17.6 Calibration History
The "Calibration History" function allows you to view and print information of the last 30 (thirty) balance adjust
ment. Adjustments made by a service technician and normal user are counted together.
a) Press «» to start "CAL.HIST".
b) Press «» key to scroll forward through the displayed list of Adjustments history information.
c) Press «
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
» for printout.
ML Models60
Sample Information Displayed:
----- Calibration ------
05.Mar. 2009 11:34
METTLER TOLEDO
Balance Type ML204
SNR 1234567890
-----------------------01 05.Mar. 2009 11:34
External ADJ SERVICE
S = External adjusted service05:03:S:01
I = Internal adjusted04:03:I:02
·
·
·
I = Internal adjusted03:03:I:28
E = External adjusted user02:03:E:29
I = Internal adjusted02:03:I:30
Sample Printout:
Application "Diagnostics"
17.7 Balance Information
The balance information function allows you to view and print information about your balance.
a) Press «» to start "BAL.INFO".
b) Press «» to scroll forward through the displayed list of Balance information.
ML Models61
-- Balance Information -
05.Mar. 2009 11:34
METTLER TOLEDO
Balance Type ML4002
SNR 1234567890
SW V1.00
Max. Load 4200 g
Platform Rainbow
TDNR 9.6.3.411.2-03
Cell ID 1172400044
Cell Type MMAI6000G2
Tolerance Rev.no. 2
Language English
------------------------
Application "Diagnostics"
c) Press «» for printout.
d) Press «C» to cancel the test procedure. The balance will return to the topic "DIAGNOS.".
Sample information displayed:
InformationDisplay
Balance typeTYPE MS6002S
Max. loadMAX 6200 g
Software platformPLATFORM RAINBOW
Serial numberSNR 1234567890
Type definition numberTDNR 9.6.3.411
Software versionSOFTWARE V1.00
Cell IDCELL ID 1172400044
Cell typeCELL TYPE MMAI6000G2
Tolerance revision numberTOLERANCE NO2
LanguageENGLISH
Sample Printout:
17.8 Service Provider Information
The service provider Information function allows you to print information about your service provider.
a) Press «» to start "SRV.PROV". The service provider information will be displayed.
b) Press «
"DIAGNOS.".
». The service provider information will be printed and the balance will return to the topic
The numerical value displayed at the balance can be transferred to the cursor position in Windows Applications
(e.g. Excel, Word) as by typing with the keyboard.
Note: The units will not be transferred.
Requirements
●
PC with Microsoft Windows® operating system and serial interface RS232.
●
Windows Application (e.g. Excel).
●
Balance to PC connection with cabel RS232 (e.g. No. 11101051 see chapter accessories).
●
Balance Interface Setting (see Interface Menu):
•
Topic "RS232": set "PC-DIR." and select the most appropriate option for the desired weighing result.
•
Save changes.
Settings at the PC
Note: The following examples are based on Windows XP.
a) Click "start".
b) Click "ControlPanel".
c) Click "Accessibility Options" in the Control Panel.
Accessibility Option
a) Click "General" Tab.
b) Enter a check mark at "Use Serial Keys".
c) Click "Settings".
Settings for SerialKeys
a) Select the serial port to be used for connection with the balance.
b) Set the baud rate to 9600
c) Click "OK".
ML Models64
Communication with Peripheral Devices
Complete the settings
a) Click "Apply" when active (wait until active).
b) Click "OK" .
Note: If the "serial key" is enabled, applications that use the same port may not function correctly. Remove the
check mark from the check box "Use Serial Keys" to disable serial key function.
Checking Operation
a) Start Excel (or another application) at the PC.
b) Activate a cell in Excel.
According to your selected "PC-DIR." option, the displayed values will appear in the column one after the other one
in the different rows.
ML Models65
Firmware (Software) Updates
19 Firmware (Software) Updates
METTLER TOLEDO is continuously improving its balance firmware (software) for the benefit of customers. So that
the customer can benefit quickly and easily from further developments, METTLER TOLEDO makes the latest
firmware versions available on the Internet. The firmware made available on the Internet has been developed and
tested by Mettler-Toledo AG using processes that meet the guidelines of ISO 9001. Mettler Toledo AG does not,
however, accept liability for consequences that might arise from using the firmware.
19.1 Operating Principle
You will find all the relevant information and updates for your balance on the METTLER TOLEDO websile at the fol
lowing address:
www.mettler-toledo-support.com
A program known as the "e-Loader II" is loaded onto your computer together with the firmware update. You can
use this program to download the firmware to the balance. The "e-Loader II" can also save the settings in your bal
ance before the new firmware is downloaded to it. You can reload the saved settings into the balance manually or
automatically after the software is downloaded.
If the selected update includes an application that is not described in these instructions (or that has been updated
in the meantime) you can dowbload the corresponding instructions in Adobe Acrobat® PDF format.
Requirements
The minimum requirements for optaining applications from the Internet and downloading them into your balance
are as follows:
●
PC with Microsoft Windows® operating system (Version 98, 98ME, ME, NT4.0, 2000, XP or Vista).
●
Internet connection and web browser (e.g. MS Internet Explorer).
●
PC to balance connection cable (e.g. No. 11101051 see chapter accessories)
19.2 Update Procedure
Installing the "e-Loader II" software from the Internet onto the PC.
a) Connect to the Internet.
b) Go to the site "www.mettler-toledo-support.com".
c) Enter the information required for registration on the METTLER TOLEDO Balance Support Site.
d) Click the "Customer Support" link and log in.
e) Click your Balance
f) Click the firmware version you need and install it.
Loading the new firmware into the balance
Start the "e-Loader II" and follow the instructions, which will take you step-by-step through the installation.
ML Models66
Error and Status Messages
20 Error and Status Messages
20.1 Error Messages
Error messages in the display draw your attention to incorrect operation or that the balance could not execute a
procedure properly.
Error MessageCauseRectification
NO STABILITYNo stability.Ensure more stable ambient condi
tions. If not possible, check settings
for environment.
WRONG ADJUSTMENT WEIGHTWrong adjustment weight on pan or
none at all.
REFERENCE TOO SMALLReference for piece counting too
small.
EEPROM ERROR - PLEASE CONTACT
CUSTOMER SERVICE
WRONG CELL DATA - PLEASE CON
TACT CUSTOMER SERVICE
NO STANDARD ADJUSTMENT PLEASE CONTACT CUSTOMER SER
VICE
PROGRAM MEMORY DEFECT PLEASE CONTACT CUSTOMER SER
VICE
TEMP SENSOR DEFECT - PLEASE
CONTACT CUSTOMER SERVICE
WRONG TYPE DATA SET - PLEASE
CONTACT CUSTOMER SERVICE
MEM.FULLMemory full.Clear the memory and start a new
FACTOR OUT OF RANGEFactor is outside the allow range.Select a new factor.
STEP OUT OF RANGEStep is outside the allow range.Select a new step.
OUT OF RANGESample weight is outside the allow
Overload - The weight on the pan
exceeds the weighing capacity of the
balance.
UnderloadCheck that the weighing pan is posi
range.
Place required adjustment weight in
center of pan.
Increase reference weight.
customer service.
customer service.
customer service.
customer service.
customer service.
customer service.
customer service.
Reduce the weight on the weighing
pan.
tioned correctly.
evaluation.
Unload the pan and load a new
sample weight.
20.2 Status Messages
Status messages are displayed by means of small icons. The status icons indicate the following:
ML Models67
Error and Status Messages
Status IconSignification
Service Reminder Your balance is due for servicing. Contact your dealer's customer ser
vice department as soon as possible to have a technician service your balance. (See
menu topic "SRV.ICON")
ML Models68
Cleaning and Service
21 Cleaning and Service
Every now and then, clean the weighing pan, draft shield element, bottom plate, draft shield (depending on the
model) and housing of your balance. Your balance is made from high-quality, durable materials and can therefore
be cleaned using a damp cloth or with a standard, mild cleaning agent.
Please observe the following notes:
●
The balance must be disconnected from the power supply
●
Ensure that no liquid comes into contact with the balance or the AC adapter.
●
Never open the balance or AC adapter – they contain no components, which can be
cleaned, repaired or replaced by the user.
●
On no account use cleaning agents which contain solvents or abrasive ingredients, as this
can result in damage to the operation panel overlay.
●
Do not use wet, but only damp cloth for cleaning.
Please contact your METTLER TOLEDO dealer for details of the available service options. Regular
servicing by an authorized service engineer ensures constant accuracy for years to come and
prolongs the service life of your balance.
ML Models69
15
96
IN
GND
OUT
HAND
SHAK
E
IN
DAT
A
OUT
RxD
TxD
CTS
RTS
Interface Specification
22 Interface Specification
22.1 RS232C Interface
Each balance is equipped with an RS232C Interface as standard for the attachment of a peripheral device (e.g.
printer or computer).
+5 V … +15 V (RL = 3–7 kΩ)
–5 V … –15 V (RL = 3–7 kΩ)
Inputs:
+3 V … +25 V
–3 V … -25 V
Full duplex
Bit-serial, asynchronous
ASCII
(software selectable)
ware selectable)
22.2 MT-SICS Interface Commands and Functions
Many of the balances and scales used have to be capable of integration in a complex computer or data acquisition
system.
To enable you to integrate balances in your system in a simple manner and utilize their capabilities to the full,
most balance functions are also available as appropriate commands via the data interface.
All new METTLER TOLEDO balances launched on the market support the standardized command set "METTLER
TOLEDO Standard Interface Command Set" (MT-SICS). The commands available depending on the functionality of
the balance.
ML Models70
Basic information on data interchange with the balance
The balance receives commands from the system and acknowledges the command with an appropriate response.
Interface Specification
Command formats
Commands sent to the balance comprise one or more characters of the ASCII character set. Here, the following
must be noted:
●
Enter commands only in uppercase.
●
The possible parameters of the command must be separated from one another and from the command name
by a space (ASCII 32 dec., in this description represented as ).
●
The possible input for "text" is a sequence of characters of the 8-bit ASCII character set from 32 dec to 255
dec.
●
Each command must be closed by CRLF (ASCII 13 dec., 10 dec.). The characters CRLF, which can be inputted
using the Enter or Return key of most entry keypads, are not listed in this description, but it is essential they be
included for communication with the balance.
Example
S – Send stable weight value
CommandSGet the current stable net weight value.
ResponseS
S WeightValue UnitCurrent stable weight value in unit actually set
under unit 1.
S ICommand not executable (balance is currently
executing another command, e.g. taring, or time
out as stability was not reached).
S +Balance in overload range.
S -Balance in underload range.
Example
CommandSQuery a stable weight value.
ResponseS
S 100.00 gThe current stable weight value is 100.00 g.
The available MT-SICS commands are listed in the table. For further information please refer to the Reference Manu
al "MT-SICS 11780711" downloadable from the Internet under www.mt.com/sics-newclassic.
DescriptionDescription
@Cancel (Reset)M31Mode after restart
C0Query/Set adjustment settings M46Print interval
C1Start adjustment according to current settings PWPiece counting: Query/Set piece weight
C2Start adjustment with external weightPWRPower on/off (PWR 0 means switch off com
pletely, if balance is powered by battery)
C3Start adjustment with internal weightSSend stable weight value
DDisplay text sent to balanceSISend weight value immediately
DATDate query/setSIRSend weight value immediately and repeat
DWDisplay weightSIRUSend weight value with currently displayed
unit immediately and repeat
I0Commands implementedSIUSend weight value with currently displayed
unit immediately
I1MT-SICS level and MT-SICS versionsSM0Dynamic weighing: cancel all SMx com
mands
I2Balance dataSM1Dynamic weighing: Start immediately and
send the result
ML Models71
Interface Specification
DescriptionDescription
I3Software version, type definition numberSM2Dynamic weighing: start after a minimum
I4Query serial number (SNR)SM3Dynamic weighing: start after a minimum
I5Query SW-identification numberSM4Dynamic weighing: query/set time interval
I10Query/set balance IDSNRSend stable weight value and repeat on
I11Query balance typeSNRUSend stable weight valuewith currently dis
I14Query balance informationSRSend weight value on weight change
KKeys: set configurationSRUSend stable weight value with currently dis
M02Query/set environmentSTSend stable weight value on pressing (print)
M03Query/set AutoZeroSUSend stable weight value with currently dis
M08Display brightnessTTare
M09Display contrastTAGet/Set tare weight value
M11Beeper: Query/set volumeTACClear tare value
M14List available languageTITare immediately
M15Query/set languageTIMQuery/set time
M17FACT: query/set single time criteria (no possi
bility to set "weekday"
M22Custom unit definition
Remarks: no possibility to set "name" of unit
M25Get application listTST2Start test function with external weight
M26Get/set current applicationTST3Start test function with internal weight
M27Adjustment historyUPDQuery/set update rate of the host interface
M30+/- settings with nominal and toleranceZZero
load is exceeded and send result
load is exceeded, send result and repeat
weight change
played unit and repeat on weight change
played unit on weight change
key
played unit
TST0Query/set test function settings
TST1Start test function according to current set
tings
ZIZero immediately
ML Models72
Technical Data
23 Technical Data
23.1 General Data
Power Supply
• AC operation:AC/DC Adapter
Primary: 100V–240V, 50/60Hz, 0.3 A
Secondary: 12VDC, 0.84A (with electronic overload protection)
Power supply to the balance: 8–20VDC, 10W
Use only with a tested AC Adapter with SELV output current.
Ensure correct polarity
• Battery operation: 8 standard AA (LR6) batteries (alkaline) for 8–15 hours of use.
Protection and Standards
• Overvoltage category:Class III
• Degree of pollution:2
• Degree of Protection:Protected against dust and water: IP54 in use with weighing pan.
• Standards for safety and EMC:See Declaration of Conformity
• Range of application:For use only in dry interior rooms
Environmental conditions
• Height above mean sea level:up to 4000 m
• Ambient temperature range:10 to 30 °C
• Relative air humidity:10% to 80 % at 31 °C, linearly decreasing to 50 % at 40 °C, noncon
23.2.1 Balances with Readability of 0.1 mg with Draft Shield
Technical Data
ModelML54ML104ML204
Maximum load52 g120 g220 g
Readability0.1 mg0.1 mg0.1 mg
Taring range0...52 g0...120 g0...220 g
Repeatability (sd)0.1 mg0.1 mg0.1 mg
Linearity0.2 mg0.2 mg0.2 mg
Sensitivity temperature drift (10...30°C)1.5 ppm/°C1.5 ppm/°C1.5 ppm/°C
ML Models73
Technical Data
ModelML54ML104ML204
Internal adjustmentyesyesyes
Adjustment range with external weights20...52 g50...120 g100...220 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)0.3 g0.3 g0.3 g
Minimum weight (U=1%, k=2)0.02 g0.02 g0.02 g
Minimum weight (OIML)0.01 g0.01 g0.01 g
Settling time, typ.2 s2 s3 s
Weighing technologyMonoBlocMonoBlocMonoBloc
Usable height of draft shield [mm]235235235
Weighing pan dimensions (WxD) [mm]Ø 90Ø 90Ø 90
Balance dimensions (WxDxH) [mm]193x290x331193x290x331193x290x331
Net Weight [kg]4.14.14.1
50 g / F2/4
5 g / E2/2
23.2.2 Balances with Readability of 1 mg with Draft Shield
Technical Data
ModelML203ML203EML303
Maximum load220 g220 g320 g
Readability0.001 g0.001 g0.001 g
Taring range0...220 g0...220 g0...320 g
Repeatability (sd)0.001 g0.001 g0.001 g
Linearity0.002 g0.002 g0.002 g
Sensitivity temperature drift (10...30°C)3 ppm/°C3 ppm/°C3 ppm/°C
Internal adjustmentyesnoyes
Adjustment range with external weights100...220 g100...220 g100...320 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)3 g3 g3 g
Minimum weight (U=1%, k=2)0.2 g0.2 g0.2 g
Minimum weight (OIML)0.02 g0.02 g0.02 g
Settling time, typ.1.5 s1.5 s1.5 s
Weighing technologyMonoBlocMonoBlocMonoBloc
Usable height of draft shield [mm]230105230
Weighing pan dimensions (WxD) [mm]Ø 120Ø 120Ø 120
Balance dimensions (WxDxH) [mm]193x290x331184x290x188193x290x331
Net Weight [kg]4.22.84.2
200 g / F2/4
10 g / F1/3
100 g / F2/4
5 g / E2/2
200 g / F2/4
10 g / F1/3
200 g / F2/4
10 g / F1/3
200 g / F2/4
20 g / F1/3
ModelML303EML503
Maximum load320 g520 g
Readability0.001 g0.001 g
Taring range0...320 g0...520 g
Repeatability (sd)0.001 g0.001 g
Linearity0.002 g0.002 g
Sensitivity temperature drift (10...30°C)3 ppm/°C3 ppm/°C
ML Models74
Technical Data
ModelML303EML503
Internal adjustmentnoyes
Adjustment range with external weights100...320 g200...520 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)3 g3 g
Minimum weight (U=1%, k=2)0.2 g0.2 g
Minimum weight (OIML)0.02 g0.02 g
Settling time, typ.1.5 s1.5 s
Weighing technologyMonoBlocMonoBloc
Usable height of draft shield [mm]105230
Weighing pan dimensions (WxD) [mm]Ø 120Ø 120
Balance dimensions (WxDxH) [mm]184x290x188193x290x331
Net Weight [kg]2.84.2
200 g / F2/4
10 g / F1/3
500 g / F2/4
2 g / F1/3
23.2.3 Balances with Readability of 0.01 g
Technical Data
ModelML802ML802EML1502E
Maximum load820 g820 g1520 g
Readability0.01 g0.01 g0.01 g
Taring range0...820 g0...820 g0...1520 g
Repeatability (sd)0.01 g0.01 g0.01 g
Linearity0.02 g0.02 g0.02 g
Sensitivity temperature drift (10...30°C)3 ppm/°C3 ppm/°C3 ppm/°C
Internal adjustmentyesnono
Adjustment range with external weights100...820 g100...820 g100...1520 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)30 g30 g30 g
Minimum weight (U=1%, k=2)2 g2 g2 g
Minimum weight (OIML)0.5 g0.5 g0.5 g
Settling time, typ.1.5 s1.5 s1.5 s
Weighing technologyMonoBlocStrain GaugeStrain Gauge
Weighing pan dimensions (WxD) [mm]170x190Ø 160Ø 160
Balance dimensions (WxDxH) [mm]184x290x84184x290x84184x290x84
Net Weight [kg]3.62.22.2
500 g / F2/4
50 g / F2/4
500 g / F2/4
50 g / F2/4
1000 g / F2/4
100 g / F2/4
ModelML1602ML3002ML3002E
Maximum load1620 g3200 g3200 g
Readability0.01 g0.01 g0.01 g
Taring range0...1620 g0...3200 g0...3200 g
Repeatability (sd)0.01 g0.01 g0.01 g
Linearity0.02 g0.02 g0.02 g
Sensitivity temperature drift (10...30°C)3 ppm/°C3 ppm/°C3 ppm/°C
Internal adjustmentyesyesno
ML Models75
Technical Data
ModelML1602ML3002ML3002E
Adjustment range with external weights1000...1620 g1000...3200 g1000...3200 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)30 g30 g30 g
Minimum weight (U=1%, k=2)2 g2 g2 g
Minimum weight (OIML)0.5 g0.5 g0.5 g
Settling time, typ.1.5 s1.5 s1.5 s
Weighing technologyMonoBlocMonoBlocMonoBloc
Weighing pan dimensions (WxD) [mm]170x190170x190170x190
Balance dimensions (WxDxH) [mm]184x290x84184x290x84184x290x84
Net Weight [kg]3.63.63.4
ModelML4002ML4002E
Maximum load4200 g4200 g
Readability0.01 g0.01 g
Taring range0...4200 g0...4200 g
Repeatability (sd)0.01 g0.01 g
Linearity0.02 g0.02 g
Sensitivity temperature drift (10...30°C)3 ppm/°C3 ppm/°C
Internal adjustmentyesno
Adjustment range with external weights2000...4200 g2000...4200 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)30 g30 g
Minimum weight (U=1%, k=2)2 g2 g
Minimum weight (OIML)0.5 g0.5 g
Settling time, typ.1.2 s1.2 s
Weighing technologyMonoBlocMonoBloc
Weighing pan dimensions (WxD) [mm]170x190170x190
Balance dimensions (WxDxH) [mm]184x290x84184x290x84
Net Weight [kg]3.63.4
1000 g / F2/4
100 g / F2/4
2000 g / F2/4
200 g / F2/4
2000 g / F2/4
200 g / F2/4
2000 g / F2/4
200 g / F2/4
2000 g / F2/4
200 g / F2/4
23.2.4 Balances with Readability of 0.1 g
Technical Data
ModelML601E*ML1501E*ML3001E*
Maximum load620 g1520 g3200 g
Readability0.1 g0.1 g0.1 g
Taring range0...620 g0...1520 g0...3200 g
Repeatability (sd)0.1 g0.1 g0.1 g
Linearity0.1 g0.1 g0.2 g
Sensitivity temperature drift (10...30°C)5 ppm/°C5 ppm/°C5 ppm/°C
Internal adjustmentnonono
Adjustment range with external weights100...620 g100...1520 g1000...3200 g
◊
ML Models76
Technical Data
ModelML601E*ML1501E*ML3001E*
◊
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
500 g / F2/4
20 g / F2/4
1000 g / F2/4
100 g / F2/4
2000 g / F2/4
200 g / F2/4
Minimum weight (acc. to USP)300 g300 g300 g
Minimum weight (U=1%, k=2)20 g20 g20 g
Minimum weight (OIML)5 g5 g5 g
Settling time, typ.1 s1 s1 s
Weighing technologyStrain GaugeStrain GaugeStrain Gauge
Weighing pan dimensions (WxD) [mm]Ø 160Ø 160Ø 160
Balance dimensions (WxDxH) [mm]184x290x84184x290x84184x290x84
Net Weight [kg]2.3 2.32.3
* Only available in selected countries.
◊
Approved version not available.
ModelML2001ML4001ML6001
Maximum load2200 g4200 g6200 g
Readability0.1 g0.1 g0.1 g
Taring range0...2200 g0...4200 g0...6200 g
Repeatability (sd)0.1 g0.1 g0.1 g
Linearity0.2 g / 0.1 g
1)
0.2 g / 0.1 g
1)
0.2 g
Sensitivity temperature drift (10...30°C)5 ppm/°C5 ppm/°C5 ppm/°C
Internal adjustmentyesyesyes
Adjustment range with external weights500...2200 g500...4200 g1000...6200 g
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
2000 g / F2/4
200 g / F2/4
2000 g / F2/4
200 g / F2/4
5000 g / F2/4
500 g / F2/4
Minimum weight (acc. to USP)300 g300 g300 g
Minimum weight (U=1%, k=2)20 g20 g20 g
Minimum weight (OIML)5 g5 g5 g
Settling time, typ.1 s1 s1 s
Weighing technologyMonoBlocMonoBlocMonoBloc
Weighing pan dimensions (WxD) [mm]170x190170x190170x190
Balance dimensions (WxDxH) [mm]184x290x84184x290x84184x290x84
Net Weight [kg]3.3 3.33.3
1)
approved version (OIML)
ModelML6001E
Maximum load6200 g
Readability0.1 g
Taring range0...6200 g
Repeatability (sd)0.1 g
Linearity0.2 g
Sensitivity temperature drift (10...30°C)5 ppm/°C
Internal adjustmentno
Adjustment range with external weights1000...6200 g
ML Models77
Technical Data
ModelML6001E
Weights for routine testing
Large Weight/Class OIML/ASTM
Small Weight/Class OIML/ASTM
Minimum weight (acc. to USP)300 g
Minimum weight (U=1%, k=2)20 g
Minimum weight (OIML)5 g
Settling time, typ.1 s
Weighing technologyStrain Gauge
Weighing pan dimensions (WxD) [mm]Ø 160
Balance dimensions (WxDxH) [mm]184x290x84
Net Weight [kg]2.3
5000 g / F2/4
500 g / F2/4
ML Models78
235
134
54
216
19
156
163
144
290
184
193
331
70
90
185
78
ø 90
169
C
Technical Data
23.3 Dimensions
23.3.1 Balances with Readability of 0.1 mg with Draft Shield (235 mm)
Models:
ML54
ML104
ML204
ML Models79
230
134
54
216
19
156
163
144
290
184
193
331
70
85
185
78
ø 120
169
C
Technical Data
23.3.2 Balances with Readability of 1 mg with Draft Shield (235 mm)
Models:
ML203
ML303
ML503
ML Models80
C
105
129
54
216
34
166
166
144
290
184
188
70
80
172 £
83
ø 120
169
Technical Data
23.3.3 Balances with Readability of 1 mg with "Flex" Draft Shield (105 mm)
Models:
ML203E
ML303E
ML Models81
167
200
54
70
84
184
203
170
144
190
290
200
87
180
C
Technical Data
23.3.4 Balances with Readability of 0.01 g with Square Weighing Pan and
Draft Shield Element
Models:
ML802
ML1602
ML3002
ML3002E
ML4002
ML4002E
ML Models82
129
54
216
144
290
184
70
84
ø 160
169
C
Technical Data
23.3.5 Balances with Readability of 0.01 g / 0.1 g with Round Weighing Pan
Models:
ML802E
ML1502E
ML601E
ML1501E
ML3001E
ML6001E
ML Models83
167
200
54
70
84
184
170
144
190
203
290
87
C
Technical Data
23.3.6 Balances with Readability of 0.1 g with Square Weighing Pan
Models:
ML2001
ML4001
ML6001
ML Models84
24 Accessories and Spare Parts
24.1 Accessories
DescriptionPart No.
Draft Shields
"Flex" draft shield "1 mg" (105 mm)12122420
Printers
RS-P25 printer with RS232C connection to balance11124300
Accessories and Spare Parts
RS-P26 printer with RS232C connection to balance (with date
and time)
RS-P28 printer with RS232C connection to balance (with date,
time and applications
Balances with readability of 0.1 g with square weighing pan
DrawingPos DescriptionPart No.
1Weighing pan 170 mm x 190 mm12122048
2Pan support12122049
3Pan support cap11131029
4Leveling foot12122040
5Battery-chamber cover12122041
6Weighing below balance cap12122029
ML Models90
Appendix
25 Appendix
25.1 Conversion Table for Weight Units
Kilogram1 kg= 1000.0g1 g = 0.001kg
Milligram1 mg= 0.001g1 g = 1000.0mg
Microgram1 µg= 0.000001g1 g = 1000000.0µg
Carat1 ct= 0.2g1 g = 5.0ct
Pound1 lb= 453.59237g1 g ≈ 0.00220462262184878lb
Ounce (avdp)1 oz= 28.349523125g1 g ≈ 0.0352739619495804oz
Ounce (troy)1 ozt= 31.1034768g1 g ≈ 0.0321507465686280ozt
Grain1 GN= 0.06479891g1 g ≈ 15.4323583529414GN
Pennyweight1 dwt= 1.55517384g1 g ≈ 0.643014931372560dwt
Momme1 mom = 3.75g1 g ≈ 0.266666666666667mom
Mesghal1 msg≈ 4.6083g1 g ≈ 0.217msg
Tael Hong Kong1 tlh= 37.429g1 g ≈ 0.0267172513291833tlh
Tael Singapore
(Malaysia)
Tael Taiwan1 tlt= 37.5g1 g ≈ 0.0266666666666667tlt
Tola1 tola= 11.6638038g1 g ≈ 0.0857353241830079tola
Baht1 baht= 15.16g1 g ≈ 0.0659630606860158baht
1 tls≈ 37.7993641666667 g1 g ≈ 0.0264554714621853tls
Manual adjustment with external weight20
Manual adjustment with internal weight19
Menu25, 27
Menu Advanced25, 29
Menu Basic25, 27
Menu Interface25, 32
Menu operation26
Menu protection27
Menu topic26, 26, 27
Motor test59
MT-SICS interface commands and functions70
Multiplication Factor Weighing49
N
Net23
Numerical values26
O
Operation keys10
P
PC-Direct64
Percent Weighing37
Performing a simple weighing23
Piece Counting35
Power supply17
Print24
Protection27
Safety precautions8
Saving settings26
Select menu26
Select menu topic26
Selecting the location16
Service29, 69
Service icon29
Service provider information62
Service reminder29
Setting up the balance13
Shutoff22, 30
Software update66
SOP54
Spare parts88
Stability beep28
Startup22, 30
Statistics41
Status icons11
Status messages67
Submenu26
Switching the balance on and off22
Switching weight units24
Symbols and conventions7
T
Taring23
Technical data dimensions79
Technical data general73
Technical data model-specific73
Time27
Time format30
Topic26, 26, 27
Totaling45
Transmit data24
Transporting the balance19
U
Unit28, 28
Unpacking13
W
Warning Limit54
Weighing below the balance19
Weighing made simple22
Weighing-in aid24
Weight unit24, 28, 28, 91
ML Models95
Index
Z
Zero range31
Zeroing23, 31
ML Models96
*11781287*
To protect your METTLER TOLEDO
The global weighing guideline GWP® reduces risks
associated with your weighing processes and helps to
• choose the appropriate balance
• reduce costs by optimizing testing procedures
• comply with the most common regulator y requirements
www.mt.com/GWP
GWP
®
Good Weighing Practice™
product’s future:
METTLER TOLEDO Service assures the
quality, measuring accuracy and preservation
of value of all METTLER TOLEDO
products for years to come.
Please send for full details about our
attractive terms of service.
Thank you.