Safety notes are marked with signal words and warning symbols. These show safety issues and warnings. Ignoring
the safety notes may lead to personal injury, damage to the instrument, malfunctions and false results.
WARNING
CAUTION
Attention
Note
1.2 Product safety information
Intended use
Your balance is used for weighing. Use the balance exclusively for this purpose. Any other type of use and operation
beyond the limits of technical specifications without written consent from Mettler-Toledo AG, is considered as not
intended.
It is not permitted to use the instrument in explosive atmosphere of gases, steam, fog, dust and
flammable dust (hazardous environments).
for a hazardous situation with medium risk, possibly resulting in severe injuries
or death if not avoided.
for a hazardous situation with low risk, resulting in damage to the device or the
property or in loss of data, or minor or medium injuries if not avoided.
(no symbol)
for important information about the product.
(no symbol)
for useful information about the product.
General hazardElectrical shock
General safety information
This balance complies with current industry standards and the recognized safety regulations; however, it can
constitute a hazard in use. Do not open the balance housing: The balance contains no user-serviceable parts. In the
event of problems, please contact a METTLER TOLEDO representative.
Always operate and use your instrument only in accordance with the instructions contained in this manual. The
instructions for setting up your new instrument must be strictly observed.
If the instrument is not used according to these Operating Instructions, protection of the instrument may be
impaired and METTLER TOLEDO assumes no liability.
Staff safety
These operating instructions must be read and understood before using the balance. These operating instructions
must be retained for future reference.
The balance must not be altered or modified in any way. Only use METTLER TOLEDO original spare parts and
accessories.
Safety Information 3Precision and Analytical Balances
Page 4
Safety notes
CAUTION
Damage to the balance
a) Only use indoors in dry locations.
b) Do not use pointed objects to operate the touch screen!
The balance is of a very sturdy design, but is still a precision instrument. It must be handled
with care.
c) Do not open the balance:
The balance contains no user-serviceable parts. In the event of problems, please contact a
METTLER TOLEDO representative.
d) Only use METTLER TOLEDO original accessories and peripheral devices for the balance.
These are specifically designed for the balance.
WARNING
Risk of electric shock
Use only the original universal AC/DC adapter delivered with your balance, and check that the
voltage printed on it is the same as your local power supply voltage. Only plug the adapter into a
socket which is grounded.
Safety Information4Precision and Analytical Balances
Page 5
2 Design and Function
8
10
1
2
3
3
3
1
2
1
2
7
7
7
9
9
5
6
5
4
4
5
6
11
13
14
12
0.1 mg
1 mg
0.01 g~0.1 g
2.1 Components
Display
1
Leveling foot
3
Weighing pan
5
Level indicator
7
Glass draft shield
9
Socket for AC adapter
11
USB Device
13
Operation keys
2
Handle for operation of the draft shield door
4
Draft shield element
6
Kensington slot for anti-theft purposes
8
RS232C serial interface
10
Legal for Trade (LFT) sealing
12
USB Host
14
Design and Function 5Precision and Analytical Balances
Page 6
2.2 Peripheral devices
USB Device
USB Host
RS232C
5
34
12
Interfaces and possible connectivity of peripheral devices:
USB DeviceRS232C
PCPrinter RS-P2x / P5x
Note
The USB Host has currently no function.
For more information about the peripheral devices see at section Accessories.
2.3 Terminal keys
Barcode Reader
RS 2nd Display
PC
No. KeyNameExplanation
1
2
3
4
5
ON/OFFTo switch the instrument on or off.
Capacitive color TFT touch screenGeneral navigation
TareTo tare the balance.
ZeroTo zero the balance.
HomeTo return from any menu level, or other window to
the application home screen.
Design and Function6Precision and Analytical Balances
Page 7
2.4 User Interface
The screen is a capacitive color TFT touch screen. The screen not only displays information, it also allows the user
to enter commands by tapping on certain areas on its surface. You can choose the information displayed on the
screen, change balance settings or perform certain operations on the instrument.
Only those elements which are available for the current dialog appear on the display.
CAUTION
Do not touch the touch screen with pointed or sharp objects!
This may damage the touch screen.
Design and Function 7Precision and Analytical Balances
Page 8
2.4.1 Application home screen
Prepare weight and start
The application home screen appears after the startup of the instrument. It always shows the last application that
was being used before the balance was switched off. The application home screen is the main screen of the
balance, from where every function can be accessed. You can always return to the application home screen by
pressing on the home button [
Information and work bars
NameExplanation
Weighing information barShows the weighing-in aid and general balance information.
1
Work title barShows information about the current activity.
2
Value barShows information about the current weighing process.
3
Main navigationWork-related functions.
4
Information fields
NameExplanation
Weighing-in aidA dynamic graphic indicator shows the used amount of the total
5
Short balance informationReadability and capacity of the balance.*
6
Weighing value fieldShows the value of the current weighing process.
7
Coach text fieldShows instructions for the current weighing process.
8
* For approved balances: Min (minimum capacity) and e (Verification scale interval) are shown in the left upper
corner.
Action buttons
NameExplanation
Main activity configuration
9
Detailed balance InformationShows detailed technical data about the balance.
10
Weighing unitShows the unit of the current weighing process.
11
ActivitiesTo open the activities selection.
12
PrintTo print out results and/or settings (printer required).
Status information fieldShows information about the system status.
15
] in the lower right corner of the screen.
weighing range.
To configure the current application (e.g. Weighing).
Design and Function8Precision and Analytical Balances
Page 9
2.4.2 Input Dialogs
ID 1 - Label
2.4.2.1 Entering Characters and Numbers
The keyboard dialog allows the user to enter characters including letters, numbers and a variety of special
characters.
NameExplanation
Input fieldShows the characters that have been entered.
1
Delete allTo delete all entered characters.
2
SwitchIf a switch appears on the right side in the content title bar, it must be switched on
3
DiscardTo discard the entered data and to exit the dialog.
4
Page indicatorTo navigate when the dialog consists of several steps.
5
ConfirmTo confirm the data entered.
6
DeleteTo delete the last entered character.
7
ShiftTo switch between lower and upper case letters.
8
Specialized tabsTo switch the keyboard mode for entering letters, numbers or special characters.
9
Explanation fieldExtra information about the value to enter (e.g. the maximum number of characters
10
in order to enter a value.
available).
Design and Function 9Precision and Analytical Balances
Page 10
2.4.2.2 Changing the date and time
Time
The dialog (Picker view) allows the user to set the date and time.
NameExplanation
Change date/time formatVarious date/time formats can be chosen.
1
Pick buttonIncrement
2
Picker fieldShows the defined time/date.
3
Pick buttonDecrement
4
Selection tabsTabs of the selectable sub-categories
5
Design and Function10Precision and Analytical Balances
Page 11
2.4.3 Lists and tables
General configuration
Weighing options
Publishing
Devices/connectivity
System settings
Basic elements of a simple list include a content title and a list of sub-elements. Tapping on an element opens a list
of sub-elements or an input dialog.
NameExplanation
List titleTitle of the current list
1
Contextual helpAdditional information about the current process
2
Back buttonTo go one step back.
3
List element titleTitle of the list element
4
Scroll positionThe list can be scrolled.
5
Selection tabsTabs of the selectable sub-categories.
6
Design and Function 11Precision and Analytical Balances
Page 12
2.4.4 Detailed balance information
Balance ME204T
- Tab on [
Balance information
Tap on [
The display shows balance identification defined by the user (see the section system settings), information about
the software and the hardware.
Balance support information
Tap on [
The display shows Support information, Service information (next service due) and Quick support request
Quick support request
Quick support request contains an unique QR code. If you have a QR (Quick Response) code reader on your smart
phone, you can take a picture of the QR code. The smartphone creates an email with all relevant service
information.
Note
Make sure that the QR code can be identified by the smart phone. A program to read the QR codes must be
installed. Make sure that there are no access restrictions, which could block your email program in some way.
] to open the general balance information menu.
] to display balance information.
] to display balance support information.
Design and Function12Precision and Analytical Balances
Page 13
3 Installation and Putting into Operation
This section describes how to put the new instrument into operation.
3.1 Scope of delivery
ComponentsModel
0.1 mg1 mg0.1 g / 0.01 g
Draft shieldhigh, 235 mm––
low, 170 mm––
Weighing pan with pan supportØ 90 mm––
Ø 120 mm––
180 x 180 mm––
Draft shield element–
Pan support––
Protective cover
Universal AC adapter
Declaration of conformity
Quick Guide
Operating instructions: printed or on CD-ROM depending on the
country.
Installation and Putting into Operation 13Precision and Analytical Balances
Page 14
3.2 Installing the components
1
2
1
Balances with readability of 0.1 mg
Balances with readability of 1 mg
Place the following components on the balance in the
specified order:
1 Push the side glass doors back as far as will go.
2 Place draft shield element (1).
3 Place weighing pan (2).
Note
Cleaning the draft shield see Cleaning and Service.
Place the following components on the balance in the
specified order:
1 Push the side glass doors back as far as will go.
2 Place weighing pan (1).
Note
Cleaning the draft shield see Cleaning and Service.
Installation and Putting into Operation14Precision and Analytical Balances
Page 15
Balances with readability of 0.01 g / 0.1 g
1
2
3
Place the following components on the balance in the
specified order:
1 Place draft shield element (1): carefully pull apart the
draft shield element to fix it under the retaining plate.
2 Insert pan support (2).
3 Place weighing pan (3).
Installation and Putting into Operation 15Precision and Analytical Balances
Page 16
3.3 Installing Protective Cover
Note
Make sure using the correct protective cover, see Accessories and Spare Parts
Balances with readability of 0.1mg / 1mgBalances with readability of 0.01g / 0.1g
Install the protective cover according to the illustrations
below, using a screwdriver Philips No. 2.
Install the protective cover according to the illustrations
below, using a flat-head screwdriver.
Installation and Putting into Operation16Precision and Analytical Balances
Page 17
3.4 Selecting the location
Your balance is a sensitive precision instrument; the location where it is placed will have a profound effect on the
repeatability and accuracy of weighing results. Chose a stable surface that is as horizontal as possible. The surface
must be able to safety carry the weight of a fully loaded balance.
Observe ambient conditions (see Technical Data).
Avoid the following:
• Direct sunlight
• Air drafts (e.g. from fans or air conditioners)
• Temperature fluctuations
• Vibrations
Installation and Putting into Operation 17Precision and Analytical Balances
Page 18
3.5 Connecting the instrument
The balance is supplied with a country-specific AC/DC adapter or with a country-specific power cable. The power
supply is suitable for all line voltages in the range: 100 - 240 VAC, 50/60 Hz. For detailed specifications, see the
section Technical Data [}31].
Note
Allow the balance to warm up for 30 minutes (0.1 mg models need 60 minutes) to adapt itself to the ambient
conditions.
Connect the AC/DC adapter to the connection socket on the back of your balance and
to the power line.
Attention
• Check if the local power supply falls within this range. If this is not the case, under no circumstances connect
the AC/DC adapter to the power supply, but contact a METTLER TOLEDO representative.
• The power plug must be accessible at all times.
• Prior to use, check the power cable for damage.
• Route the cable in such a way that it cannot be damaged or cause a hindrance when working.
• Ensure that no liquid ever comes into contact with the AC/DC adapter.
Installation and Putting into Operation18Precision and Analytical Balances
Page 19
3.6 Leveling the balance
Exact horizontal positioning and stable installation are prerequisites for repeatable and accurate weighing results.
Note
To compensate for small irregularities or inclinations (±2 %), the instrument must be leveled and adjusted each
time it is moved to a new location. There are two adjustable leveling feet to compensate for slight irregularities in the
surface of the weighing bench.
Leveling the balance
- Adjust the two front leveling feet until the air bubble comes to rest
exactly in the middle of the glass:
Air bubble at"12 o'clock"turn both feet clockwise
Air bubble at"3 o'clock"turn left foot clockwise, right
Air bubble at"6 o'clock"turn both feet counterclockwise
Air bubble at"9 o'clock"turn left foot counterclockwise,
foot counterclockwise
right foot clockwise
Installation and Putting into Operation 19Precision and Analytical Balances
Page 20
3.7 Transporting the balance
Switch off the balance and remove the power cable and any interface cable from the balance. Refer to the notes in
section Selecting the location [}17] regarding the choice of an optimal location.
Transporting over short distances
CAUTION
For balances with a draft shield:
Never lift the balance by its glass draft shield. The draft shield is not sufficiently fastened to the
balance.
Transporting over long distances
CAUTION
If you would like to transport or ship your balance over long distances, use the complete original
packaging.
3.8 General requirements
3.8.1 Warming up the balance
Before working with the balance, it must be warmed up in order to obtain accurate weighing results. To reach
operating temperature, the balance must be connected to the power supply for at least:
• 30 minutes for balances with a readability of 1mg to 0.1g.
• 60 minutes for balances with a readability of 0.1mg and higher.
3.8.2 Adjusting the balance
To obtain accurate weighing results, the balance must be adjusted to match the gravitational acceleration at its
location and depending on the ambient conditions. After reaching the operation temperature, an adjustment is
necessary in the following cases:
• Before the balance is used for the first time.
• When the balance (readability of 0.1 mg) was disconnected from the power or in case of a power failure.
• After a change of location and after the warming-up phase.
• At regular intervals during weighing service.
Installation and Putting into Operation20Precision and Analytical Balances
Page 21
4 Weighing Made Simple
This section describes how to perform a simple weighing. In addition, the basic concept of the navigation and the
basic functions of the balance are explained.
4.1 Switching on the balance
Switching on the balance for the first time
1 Remove any load from the weighing pan.
2 Connect the balance using the AC/DC adapter to the mains.
ð After the start screen has disappeared, the balance starts with the application home screen.
After the balance has switched on for the first time, it can be switched on by pressing long on [
Note
When the balance is switched on for the first time, the home screen of the application Weighing opens. If the
balance is switched on again, it always starts with the home screen of the application that was last used before
switching off.
].
4.2 Switching off the balance
1 Press on [
2 Tap on [
ð the balance switches off and goes into standby mode.
Note
• After switching on from standby mode, the balance needs no warm-up time and is immediately ready for
weighing.
• If the balance has been switched off after a preselected time, the display is dimly lit and shows the time and the
date.
• If the balance has been switched off manually, the display is off.
To completely switch off the balance, it must be disconnected from the power supply.
] and hold until the dialog Switch-off appears.
].
Weighing Made Simple 21Precision and Analytical Balances
Page 22
4.3 Main settings and activities at a glance
Activities
Main configuration
Application
Balance information
General settings
The diagram below provides an overview of the main settings of an application (in this example, Weighing).
Depending on the application, the selectable options and their content can differ. Every application is based on this
concept.
Weighing Made Simple22Precision and Analytical Balances
Page 23
4.4 Navigation on the touch screen
Activities
Weighing applications
Counting
Percent weighing
Percent
weighing
Quick settings
Date and Time
Brightness
Sound
ON (3 of 3 beeps)
Language
To interact with the balance, use the screen and the operating keys at the bottom of the screen. The navigation on
the screen is similar to the navigation on a smart phone or tablet PC.
4.4.1 Opening an application
To open settings or applications, tap with your finger on the symbol of the application (e.g. [
4.4.2 Scrolling
When the options are too numerous to be shown on one screen, a blue bar on the right side appears. This means
that the user can scroll up and down. For scrolling, simply place your finger somewhere in the list and drag it up or
down.
] Weighing).
Weighing Made Simple 23Precision and Analytical Balances
Page 24
4.4.3 Using shortcuts
To simplify navigation on the capacitive color TFT touch screen, there are a few shortcuts that allow quick access to
the most important areas of the balance. The weighing value field on the application home screen for example
works as a shortcut (see diagram below), so does the weighing unit next to the weighing value field. Depending on
the application, there may be other shortcuts that can be used.
Note
Every setting that can be changed directly via shortcut, can also be changed in the main configuration settings of
the application.
4.5 Performing a simple weighing
When you switch on the balance for the first time, the home screen of the application Weighing opens
automatically. If the balance was used already, the last used application before the balance was switched off
opens. If another application is running, switch to the application Weighing.
Weighing Made Simple24Precision and Analytical Balances
Page 25
2 Place a sample on the weighing pan.
ð The unstability symbol
becomes light blue.
ð After a short period of time, the weighing pan is stable. The unstability symbol disappears and the value
in the weighing value field becomes dark blue again.
ð The weighing process is finished. The result appears on the screen.
appears at the left side of the screen and the value in the weighing value field
4.6 Switching the weighing unit
There are several weighing units available. The default value is country-specific.
The weighing unit can be chosen via the main configuration of the current application or via shortcut. This example
describes how to change the weighing unit via shortcut.
1 Tap on the weighing process unit (shortcut) gram (g).
ð the screen Main unit opens.
Weighing Made Simple 25Precision and Analytical Balances
Page 26
2 Put your finger somewhere in the list and drag up to scroll down.
Main unit
g (Default)
3 Choose another weighing unit (e.g. ounce (oz)) by tapping on it.
4 Tap on [
] to confirm the chosen weighing unit.
ð The weighing unit gram (g) has been changed to ounce (oz).
Note
With approved balances, this menu topic has a fixed setting and cannot be changed.
Weighing Made Simple26Precision and Analytical Balances
Page 27
4.7 Changing the resolution
Resolution
There are several resolutions available. The default resolution is instrument-specific.
The resolution can be changed as follows:
1 Tap on the weighing value field.
2 Tap on 10d - 0.001 g.
3 Confirm the chosen resolution by tapping on [
].
4.8 Zeroing
ð The resolution has been changed.
1 Unload the balance.
2 Press [
All weight values are measured in relation to the zero point.
] to set the balance to zero.
Weighing Made Simple 27Precision and Analytical Balances
Page 28
Note
Press [
] before starting the weighing process.
4.9 Taring
1 Tap on [
2 Place the empty container on the balance.
ð The weight is displayed.
3 Press [
ð The status information field on the left side of the screen shows Net and the weighing value field shows
0.000 g.
] to set the balance to zero.
] to tare the balance.
Weighing Made Simple28Precision and Analytical Balances
Page 29
5 Maintenance
WARNING
Risk of electric shock
– The instrument must be disconnected from the power supply, before cleaning or other maintenance
work to be performed.
– Use only the power cord from METTLER TOLEDO, if it needs replacing.
Please observe the following notes:
Note
On no account use cleaning agents which contain solvents or abrasive ingredients, as this can result in damage to
the touch screen.
5.1 Cleaning and Service
Every now and then, clean the weighing pan, draft shield element, bottom plate, draft shield (depending on the
model) and housing of your balance. Your balance is made from high-quality, durable materials and can therefore
be cleaned using a damp cloth or with a standard, mild cleaning agent.
Please observe the following notes:
• The balance must be disconnected from the power supply
• Ensure that no liquid comes into contact with the balance or the AC adapter.
• Never open the balance or AC adapter – they contain no components, which can be cleaned,
repaired or replaced by the user.
• On no account use cleaning agents which contain solvents or abrasive ingredients, as this
can result in damage to the operation panel overlay.
• Do not use wet, but only damp cloth for cleaning.
Please contact your METTLER TOLEDO dealer for details of the available service options. Regular
servicing by an authorized service engineer ensures constant accuracy for years to come and
prolongs the service life of your balance.
Maintenance 29Precision and Analytical Balances
Page 30
5.2 Draft Shield
Removing or inserting sliding glass doors
It is possible to remove the sliding glass doors for
cleaning or for replacing. In this case you have to
remove the handle first. Installing the handle after
insertion of the glass door.
Note
Front and rear glass panels cannot be removed.
5.3 Disposal
In conformance with the European Directive 2002/96/EC on Waste Electrical and Electronic
Equipment (WEEE) this device may not be disposed of in domestic waste. This also applies to
countries outside the EU, per their specific requirements.
Please dispose of this product in accordance with local regulations at the collecting point
specified for electrical and electronic equipment. If you have any questions, please contact the
responsible authority or the distributor from which you purchased this device. Should this device
be passed on to other parties (for private or professional use), the content of this regulation must
also be related.
Thank you for your contribution to environmental protection.
Maintenance30Precision and Analytical Balances
Page 31
6 Technical Data
6.1 General Data
Power Supply
• AC operation:AC/DC Adapter
Protection and Standards
• Overvoltage categorie:II
• Degree of pollution:2
• Degree of protection:Protected against dust and water.
• Standards for safety and EMC:See Declaration of Conformity
• Range of application:For use only in dry interior rooms
Environmental conditions
• Height above mean sea level:up to 3500 m
• Ambient temperature range:Operating condition for ordinary lab application: +10 to 30 °C
• Relative air humidity:10% up to 80 % at 31 °C, linearly decreasing to 50 % at 40 °C,